Home
Invacare® RPS 440ee Instruction Manual
Contents
1. Inspected By Action Taken Signature 3l All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Inspected By Action Taken Signature Date Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Inspected By Action Taken Signature Date 32 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Notes 33 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trade
2. Invacare RPS 440ee Instruction Manual Version 07 08 DEALER This manual MUST be given to the user of the patient lift USER BEFORE using this patient lift read this manual and save for future reference SERVICE AGENT Only Invacare authorised service agents are permitted to install service or repair Invacare product e e RPS 440ee User s Manual N Safety Precautions Please read and understand the following before using a Invacare patient lifter DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UNDERSTANDING THE WARNINGS CAUTIONS AND INSTRUCTIONS IN THIS MANUAL AND THE OWNER S MANUAL PROVIDED WITH THE SLINGS IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS CAUTIONS AND INSTRUCTIONS CONTACT A HEALTHCARE PROFESSIONAL DEALER OR TECHNICAL PERSONNEL IF APPLICABLE BEFORE ATTEMPTING TO USE THIS EQUIPMENT OTHERWISE INJURY OR DAMAGE MAY RESULT THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE RADIO FREQUENCY INTERFERENCE Most electronic equipment is influenced by Radio Frequency Interference RFI CAUTION should be exercised with regard to the use of portable communication equipment in the area around such equipment If RFI causes erratic behaviour PUSH the Red Emergency Stop Button IMMEDIATELY DO NOT turn the Red Emergency Stop Button OFF while transmission is in progress This product meets EN60601 1 2 Collateral Standard Electromagnetic Compatibility MAINTENANCE Maintena
3. Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Inspected By Action Taken Signature Date Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Inspected By Action Taken Signature Date 28 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Date Inspected By Action Taken Signature Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Date Inspected By Action Taken Signature 29 All information is subject to change without prior notice Photographs are for illustration purposes
4. NOTE Check all parts for shipping damage before using In case of damage DO NOT use the equipment Contact the Carrier Dealer for further instructions NOTE The stand up lift may be operated by one 1 healthcare professional for ALL lifting preparation transferring from and transferring to procedures with a cooperative weight bearing individual able to support the majority of hislher own weight However since medical conditions vary Invacare recommends that the healthcare professional evaluate the need for assistance and determine whether more than one 1 assistant is appropriate in each case to safely perform the transfer DO NOT attempt any transfer without approval of the patient s physician nurse or medical assistant Thoroughly read the instructions in this Owner s Manual observe a trained team of experts perform the lifting procedures and then perform the entire lift procedure several times with proper supervision and a capable individual acting as a patient Individuals that use the Standing Sling MUST be able to support the majority of their own weight otherwise injury can occur DO NOT raise the patient to a full standing position while using the transport sling otherwise injury can occur Invacare standing and transport slings are specifically designed to be used in conjunction with the Invacare Stand Up Lift Slings and accessories designed by other manufacturers or other Invacare slings are not to be utilized as a component
5. OPERATION section of this manual 4 Unhook the standing or transport sling from all attachment points on the stand up lift 5 If equipped unfasten the leg strap from around the patient s legs 6 Instruct the patient to lift their feet off of the footplate NOTE Assist the patient if necessary 7 Remove the standing or transport sling from around the patient 8 Pull the stand up lift away from the bed 18 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Transferring the Patient to a Commode Position the patient over the commode 2 Press the DOWN button and lower the patient onto the commode Invacare recommends locking the rear swivel casters ONLY when positioning or removing the sling standing or transport from around the patient 3 Lock the rear swivel casters Refer to Locking Unlocking the Rear Swivel Casters in the OPERATION section of this manual 4 Perform one 1 of the following A Standing Sling Unhook the standing sling from the attachment points on the stand up lift B Transport Slings Unhook the transport sling from the bottom attachment points on the stand up lift Lift up on the patient s legs and remove the thigh supports from under
6. e RPS 440ee User s Manual Documentation With each lifter you should receive the following TY PRO MED documents WARRAN pA vmi We warrass the product to be free from defects a material and workmanship under normal usc and service and we will vetthue the penods stated below frum the date of purchase repair of replace without cost to the original customer any part assembly or portion thercof whach shall be returned to Pro Med aed from our iaspextion shows lo be defective e this Instruction Manual phe how jer den menle rn urn pi endogena in e Authorised Distributor Pro Med Slings vosd warranty if not laundered as pet instructions on the Sling Label e a Test Certificate and Pa ir cig crai ye aWarranty Registration Card Jm the responsibility of the Purchaser to tram and instruct pew staff on the proper uc of these products Warning Ht us the responsibility of the Purchaser to select the appropriate Pro Med Lifter and Pro Med Sling to address the lifting needs of an individual An incorrect selection may cause iejary I in doube uech input rom a sustably qualified therapest Pro Med disclaims amy responsiblity for accidents mishaps ot injury caused by the combenation of Pro Med products with another manufaa turer s equipment Model Serial No Date of Purchase IMPORTANT This registration card is proof of purchase and must be completed and retumed within 14 days of purchase Vw cortty That Pe machine described
7. patient check again to make sure that the sling is properly connected to the attachment points of the stand up lift If any attachments are NOT properly in place lower the patient back onto the surface and correct this problem During transfer with patient suspended in a sling attached to the lift DO NOT roll stand up lift over objects such as carpet raised carpet bindings door frames or any uneven surfaces or obstacles that would create an imbalance of the stand up lift and could cause the stand up lift to tip over Use push handle on the mast at ALL times to push or pull the stand up lift Casters and axle bolts require inspection every six 6 months to check for tightness and wear After the first 12 months of operation inspect all pivot points and fasteners for wear If the metal is worn the parts MUST be replaced Make this inspection every six 6 months thereafter The electric motor is sealed at the factory and if service is required the motor unit MUST BE returned to the factory for repair DO NOT attempt to open the motor or obtain local service as this will VOID the warranty and may result in damage and a costly repair Consult your Dealer or Invacare for further information Regular maintenance of Stand Up Lift and accessories is necessary to assure proper operation WEIGHT LIMITATION DO NOT exceed maximum weight limitation of the Stand Up Lift The maximum weight limitation is 440 Ibs regardless of any additional weigh
8. Invacare recommends that the healthcare professional evaluate the need for assistance and deter mine whether more than one 1 assistant is appropriate in each case to safely perform the transfer STANDING SLINGS Before lifting the patient make sure the bottom edge of the standing sling is positioned on the lower back of the patient and the patient s arms are outside the standing sling TRANSPORT SLINGS Before lifting the patient make sure the bottom edge of the transport sling is at the base of the spine and the patient s arms are outside the transport sling Invacare does NOT recommend locking the rear casters of the stand up lift when lifting and transferring an individual Doing so could cause the lift to tip and endanger the patient and assistants Invacare DOES recommend that the rear casters be left UNLOCKED during lifting and transferring procedures to allow the stand up lift to stabilize itself when the patient is initially lifted from and transferred to a chair bed or any stationary object Lifting the Patient l Instruct the patient to do the following A Hold onto the hand grips on both sides of the stand up lift B Lean back into the standing or transport sling 2 Make sure of the following A Patient s knees are secure against the knee pad B Patient s feet are positioned on the footplate C The bottom edge of the Standing Sling is positioned on the lower back Transport Sling is at the base of the patient s
9. eee eee eee eens 17 Transferring the Patient to a Bed 2 end osda me ba dp ER ae OEPPERERRE VES E REA 18 Transferring the Patient to a Commode isse sees eR RARE serve eras ERR 19 Litter Maintenance laxe ERE CR ROCR OR RCRCRCORCR CAS a RC e a WER RR ACC E e 20 Troubleshooting 4d E o ROXCCRO ORO ORSON HRN C CR CC o n 22 Care and Maintenance of your Stand Up Lifter 0 cece cece ec enes 23 LUDRICALION si cov aes ecigoiva seer anbvaseene weer ae ay RERO S REALE awn ER EA began 23 WearAnd Damage veo seduce s habeus re XR ESCV P Rd A E ER a 23 S up QT RC e a S e E a a E E E E E E 24 Electric Actuator Replacement esso e ex RR REF RE RR RR 24 Base Adjustimelit 2253s sci ema en ER RR meh baw ew RR Hae RIS Rue OR EUR AR e RR Se 24 Limited Warranty 034 so odes dsenves 44 nA P3 CE RARE WW ER WR A d RARO aes 25 L g Book i449 ae siada ARE ERGO a e CIC dCi DO eRe DOGS Cae ewe ganar 26 3 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e Specifications Model e RPS 440ee User s Manual RPS440 Product Code LPS RPS440 AUS Capacity 200 kg 440 Ib Leg Spread Adjustment Electric Configuration Hoisting Range Reach Hoisting Range Reach 9 10 Hoisting reach at min height
10. of C S P 1 1 Max hoisting reach DLL E 9 pars Hoisting reach at max height of C S P 300 Flexible device C S P Central Suspension Point Spreader bar Lifting machinery Hoisting range erret rts tetti erteneeeesetnee Max height of C S P nentes Min height of C S P Height of C S P at max height hoisting reach NA 12 Base Base Base Radial type Min internal width Min external width Max internal width DUO UTR E T Base Base height clearance Base 2 Min clearance 3 Max height Frame Material Epoxy Coated Steel Castors Single Wheel Size 100mm 4 Brakes Yes Rear Product Weight 52 kg 115 Ib Battery 24VDC Charger Input 100 240 VAC Charger Output Charging Time 29 5VDC Max 6 hours Safety Devices Approx Lifts per Charge 100 200 cycles Motor Anti Entrapment Yes Emergence Stop Yes Emergence Lowering Yes 4 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e Base height clearance Max external length ntt Max internal length treten Max external Width isisisi e RPS 440ee User s Manual Safety Summary
11. of Invacare s stand up lift system Use of these products is prohibited and will void Invacare s stand up lift warranty Use only genuine Invacare standing and transport slings and lift accessories to maintain patient safety and product utility Use the sling that is recommended by the individual s doctor nurse or medical assistant for the comfort and safety of the individual being lifted Before positioning the legs of the stand up lift around the patient make sure the patient s feet are out of the way of the footplate otherwise injury can occur Invacare recommends locking the rear swivel casters ONLY when positioning or removing the sling standing or transport from around the patient Invacare does NOT recommend locking the rear swivel casters of the stand up lift when lifting and transferring an individual Doing so could cause the lift to tip and endanger the patient and assistants Invacare DOES recommend that the rear swivel casters be left UNLOCKED during lifting and transferring procedures to allow the stand up lift to stabilize itself when the patient is initially lifted from and transferred to a chair bed or any stationary object STANDING SLINGS DO NOT use the standing sling in combination with the stand up lift as a transport device It is intended to transfer an individual from one resting surface to another such as a bed to a wheelchair Moving a person using the standing sling in combination with the stand up lift over ANY d
12. only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Inspected By Action Taken Signature Date Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Inspected By Action Taken Signature Date 30 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Date Inspected By Action Taken Signature Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Date
13. you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Contents Special Notes sav nx o m SERRE EEOC ORE HOES CAUCA TRUCO EROR Ro S 2 SPECHICALIONS eoe CX RC b CR CAL MUR ACA CIR UR VCI MAR SOS WEEN One n d 4 Safety SUMMANLY aao ed mh m AERA EAR RR OR REOR Wa RE wm HR A RO RU 5 Key t Symbols iis 9E 06455569054 5 5 6 94 544556569555 ENARRARE NEUE OSES 7 DocumentatlOD soa 3 3 99 ROC OCURRE ROC AUC ACA RS SAAS C9 ORE WO 8 Assembly 222a RR Rx RU RC REOR AUR RS EERE AUR a RUR RE CR DOO Sa RE 9 Operating the Stand Up Lift 1 cc ce wc cee ee ee ee nnn 12 Locking Unlocking the Rear Swivel Casters 0 0 cece eee eee cee eee cease 12 Raising Lowering Stand Up Lift asse o de cw atnenhtatatace nee eae sehen eee dade 12 Closing Opening the Legs of the Base cose eee Rh nn 12 Mounting Battery Charge suom exem es om m dense eens omo et Ro c ce 13 Charging the Batteries s auewa ated EE ERAT Ako en RF erratera snitter tohia 14 Transferring From ss 64 9 664 664 650 664 56 RENO OO AO OUT ORO OR RS 15 Lifting Preparations iaceo don y Rd oder ROGO OR das Reb Rd ox CR Reip eo a Rd 15 Lifting the Patient x Sueco delere Catekny Ra xd PC PIRE SAM ease tas Welt ddr 15 Moving the Patient iis uses e ar A ERRAT RSS EEUE BREE OME Ss BEES l6 Transferring TO 522 4 sanas tane n E E OU ACCRUE RR e n 17 Transferring the Patient to a Wheelchair 0 ee eee
14. E CHECK ALL SLING ATTACHMENTS each time it is used to ensure x x proper connection and patient safety Inspect sling material for wear x x x x Inspect straps for wear x x x x CLEANING Whenever necessary Clean with mild soap and water and remove any adhesive tape marks with eucalyptus oil x x x SHIFTER HANDLE Manual Leg Adjustment Only Operates smoothly x x x Locks adjustable base whenever engaged x x x ON COMPLETION OF CHECK or SERVICE Arrange repair replacement of any faulty parts x x x x Complete and sign log book x x x Repairs Faulty components or workmanship covered by warranty will be repaired where appropriate and replaced when necessary Defects due to mistreatment or neglect will not be covered by warranty but repair can be arranged by Invacare Australia authorised agents at the request of the User Spare Parts Spare parts are available for purchase through one of our Invacare Australia s authorised agents A Major Service undertaken by an authorised service agent every 12 months is a minimum requirement NOTE For individual home use a full inspection is required prior to each new user Regular cleaning will reveal loose or worn parts enhance smooth operation and extend the life expectancy of the lift If applicable an annual load test maximum load to 12576 of the Safe Working Load is required for CE mark maintenance 21 All information is subject to change without prior notice Photographs are for illustr
15. The kneepad is an integral part of the lift and must be attached so that the lifter can function correctly and safely Remove the two hex screws on front of mast behind the actuator Align the kneepad receiver assembly over the screw holes 5 Ensure the assembly is correctly orientated the kneepad release lever should be on top Replace the screws Tighten firmly 9 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Assembly continued Using 6mm and 7mm spanners connect top of actuator to boom using bolt provided 7 Note Do not over tightening this bolt as this will deform the clevis and restrict the pivot The kneepad is fundamental to the safe standing of patients by this lift With the exception of using this lift as a gait trainer the kneepad must be fitted to the lift Attaching the kneepad Attach the kneepad to the kneepad receiver by sliding the two posts into the tube retainers of the kneepad receiver assembly and at the same time turning the kneepad release lever in a clockwise direction Release the lever and slide the kneepad in or out until the mechanism engages and locks the kneepad Attaching the battery Fit battery to control box For more
16. ation purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e Troubleshooting e RPS 440ee User s Manual Symptoms Patient Lift feels loose Casters Brakes noisy or stiff Noisy or dry sound from pivots Electric actuator fails to lift when button is pressed Unusual noise from Actuator Boom will not lower in uppermost position Faults Mast Base joint loose Tie Rods are loose Fluff or debris in bearings Needs lubrication Hand control or actuator connector loose Batteries low RED emergency stop button pressed IN Battery not connected properly in to control box The connecting terminals are damaged Electric actuator in need of service or load is too high Actuator is worn or damaged or spindle is bent Boom require a minimum weight load to lower from the uppermost position Solution Refer to ASSEMBLY section in this manual Refer to the Base Adjustment in the CARE AND MAINTENANCE section in this manual Refer to Casters in the CARE AND MAINTENANCE section in this manual Refer to Lubrication in the CARE AND MAINTENANCE section in this manual Check connections Charge batteries Refer to CHARGING THE BATTERIES in the OPERATION section of this manual Rotate RED emergency stop button CLOCKWISE until it pops out Recon
17. ce Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Transferring To NOTE The stand up lift may be operated by one 1 healthcare professional for ALL lifting preparation transferring from and transferring to procedures with a cooperative weight bearing individual able to support the majority of his her own weight However since medical conditions vary Invacare recommends that the healthcare professional evaluate the need for assistance and determine whether more than one 1 assistant is appropriate in each case to safely perform the transfer Transferring the Patient to a Wheelchair I Move the wheelchair into position Wheelchair wheel locks MUST be in the locked position BEFORE lowering the patient into the wheelchair 2 Lock the rear wheels of the wheelchair 3 Position the patient over the wheelchair 4 Press the DOWN button and lower the patient into the wheelchair Invacare recommends locking the rear swivel casters ONLY when positioning or removing the sling standing or transport from around the patient 5 Lock the rear swivel casters Refer to LOCKING UNLOCKING THE REAR SWIVEL CASTERS in the OPERATION section of this manual 6 Unhook the standing or transport sling from all attachment points on the stand up l
18. e injury or damage may occur Refer to diagram below for correct instructions 3 Place the battery on the battery charger as shown in the following diagram Make sure there is an audible click NOTE The charge LED will illuminate When charging is complete charge LED will stop illuminating NOTE A battery needing to be fully recharged will take approximately four 4 hours 4 Lift UP on the handle on the back of the battery 5 Lift the battery UP and OUT away from the battery charger 6 Reinstall the battery onto the control box as shown Make sure there is an audible click Handle STEPS 1 2 4 and 5 Battery Charger STEP 3 Control Box i STEP 6 NOTE The Battery mounts to the Con trol Box and Battery Charger as shown Audible Click STEPS 3 and 6 i 14 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Transferring From Lifting Preparation NOTE The stand up lift may be operated by one 1 healthcare professional for ALL lifting preparation transferring from and transferring to procedures with a cooperative weight bearing individual able to support the majority of his her own weight However since medical conditions vary
19. e patient into the wheelchair 3 Transferring from a wheelchair Lock the wheel locks of the wheelchair 4 Press the UP button to raise the patient above the surface bed wheelchair or commode being transferred from The patient should be elevated just high enough to clear the surface with their weight fully supported by the lift NOTE The lower center of gravity provides stability making the patient feel more secure and the lift easier to move Moving the Patient STANDING SLINGS DO NOT use the standing sling in combination with the stand up lift as a transport device It is intended to transfer an individual from one resting surface to another such as a bed to a wheelchair Moving a person using the standing sling in combination with the stand up lift over ANY distance is NOT recommended During transfer with patient suspended in a sling attached to the lift DO NOT roll stand up lift over objects such as carpet raised carpet bindings door frames or any uneven surfaces or obstacles that would create an imbalance of the stand up lift and could cause the stand up lift to tip over Use push handle on the mast at ALL times to push or pull the stand up lift I Using the push handle move the stand up lift away from the surface 2 SLOWLY move the patient to the desired surface 3 Lower the patient onto the desired surface Refer to TRANSFERRING TO in this manual l6 All information is subject to change without prior noti
20. er the lift arms and the patient NOTE If the Stand Up Lift is in the full UP position it may be necessary to pull down slightly on the lift arms before the mast will lower EMERGENCY STOP Press the RED button on the control box IN to stop the lift arms and patient from raising or lowering Refer to page 9 NOTE Rotate RED emergency stop button CLOCKWISE until it pops out to disengage EMERGENCY LOWERING In the event of a handset or control box failure the lifter can still be lowered by the following methods By pressing the tip of a ballpoint pen o into the small circle below and to the right of the stop Me 9 button on the control box 2 By pulling up the red tag on the actuator column actuator must be loaded Closing Opening the Leg Base The bottom row of the handset is used to open or close the legs of the base for stability prior to lifting a patient Pressing the left button widens the leg base the right button closes the leg base 12 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Mounting Battery Charger NOTE Refer to your local regulations concerning proper mounting procedures 2 Place the battery charger with mounting bracket on the wal
21. fety of the hoist is jeopardised in any way the Owner should be notified immediately the hoist should be taken out of service and not used again until the problem has been rectified Contact the supplier to arrange servicing A Guide to Maintenance Procedures A regular visual check of the hoist and attachments is essential Invacare Australia Pty Ltd permits servicing only by an Authorised Service Agent in accordance with the following schedule of Service Types Item Pre Delivery Daily Minor Service Major Service After every After every 3 Months of 6 Months of Service Service PRE DELIVERY Confirm all components are included x Inspect parts for possible damage in transit x COMPLETE LIFTER General inspection of Lifter Visually inspect the patient lifter for external damage or wear Scratches and chips Retard rusting by applying spray paint Dulux Special White Code 81880 BASE AND CASTORS x x Inspect for missing hardware Base opens closes with ease Inspect castors and axle bolts for tightness Make sure tightly attached to legs Inspect castors for smooth swivel and roll XX X x X Check castor brakes operate efficiently XXX X X X Overhaul castors and repack with grease Check tightness of bolt in leg bushes XX XXX X X X Check linkages are secure in leg adjuster ball joints MAST Mast MUST be securely assembled to boom x x Inspect for bends or deflections x x x x x x x Check slin
22. g hooks for wear or deflection ELECTRIC ACTUATOR ASSEMBLY Check hand control and cycle to ensure smooth quiet operation Lift adjustment on all Lifters Width also for those with electric leg adjustment Emergency lowering function electric and mechanical Check for wear or deterioration IF DAMAGED RETURN TO FACTORY X XX X Inspect hardware on mast and boom X XXX X X XXX X Remove dust and dirt from actuator with cloth 20 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Item Pre Delivery Daily Minor Service Major Service After every After every 3 Months of 6 Months of Service Service ELECTRIC ACTUATOR ASSEMBLY Repack actuator clevis with grease Top and bottom actuator bolts must not be over tightened the actuator must be able to pivot on the bolt without x resistance Ensure original side packing washers are replaced after regreasing Wipe actuator shaft with oiled rag DO NOT use lubricants such as WWD40 or RP7 x BOOM Check all hardware and swivel bar supports x x x Inspect for bends or deflections x x x Inspect bolted joints of boom for wear x x x Inspect to ensure that the boom is centred between the base legs x x x SLINGS AND HARDWAR
23. ghtness Remove all debris etc from the wheel and swivel bearings If any parts are worn replace these parts immediately If you question the safety of any part of the lift contact your Dealer immediately and advise him her of your problem Lubrication The Invacare lift is designed for minimum maintenance however a six 6 month check and lubrication should ensure continued safety and reliability Keep lift and slings clean and in good working order Any defect should be noted and reported to your dealer as soon as possible The casters MUST swivel and roll smoothly A light grease waterproof auto lubricant may be applied to the ball bearing swivel of the casters once a year Apply more frequently if the casters are exposed to extreme moist conditions Wear and Damage It is important to inspect all stressed parts such as slings spreader bar and any pivot for slings for signs of cracking fraying deformation or deterioration Replace any defective parts immediately and ensure that the lift is not used until repairs are made 23 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Cleaning The sling should be washed regularly in water temperature of 140 F 60 C and a biol
24. he Owner to withdraw the patient lifter from service if inspection reveals that user safety is jeopardised in any way from use of the patient lifter Ensure ongoing servicing after the Warranty Period is maintained by agreement between you and the Owner 26 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual TO BE COMPLETED AFTER EACH INSPECTION Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Date Inspected By Action Taken Signature Service Type Condition Report Pre Delivery Minor Periodic Inspection Major Date Inspected By Action Taken Signature 27 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual
25. hes been received in satafactory conden and That Uwe have been instructed n is proper operation care and manienarce Me cortty Mat vwe recewved a copy of Warranty Registration Card The tear off Registration Card must be completed and returned to Invacare within fourteen days of purchase 8 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Assembly The RPS400ee is packed partly assembled for shipping purposes To assemble your RPS400ee remove all contents from box and place unit on workbench or table Using 16 mm and 17 mm spanners remove the two bolts connecting the mast to the chassis Remove mast and boom assembly from base 1 Position upright and attach mast to base using the previously removed bolts Note Ensure these bolts are firmly tightened Attach the rod ends as shown in 2 When assembling the patient lift it is important to tighten bolts according to instruction Failure to do so may result in damage to the lift or injury to the care giver or client Cut cable tie securing actuator 3 and swing actuator forward so that it is out of the way Swing boom up and over to the front 4 LH Attaching the kneepad receiver Note
26. ift 7 If equipped unfasten the leg strap from around the patient s legs 8 Instruct the patient to lift their feet off of the footplate NOTE Assist the patient if necessary 9 Remove the standing or transport sling from around the patient 10 Pull the stand up lift away from the wheelchair 17 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Transferring the Patient to a Bed NOTE The lower center of gravity provides stability making the patient feel more secure and the lift easier to move NOTE The lift arms will stay in position until the DOWN button is pressed l Position the patient as far over the bed as possible NOTE If patient is being transferred from a surface that is lower than the bed press the UP button to raise the patient above the surface of the bed The patient should be elevated just high enough to clear the bed with their weight fully supported by the lift 2 Press the DOWN button and lower the patient onto the bed Invacare recommends locking the rear swivel casters ONLY when positioning or removing the sling standing or transport from around the patient 3 Lock the rear swivel casters Refer to Locking Unlocking the Rear Swivel Casters in the
27. information about this procedure refer to page 12 of this manual 10 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Ensure control box emergency push button is set to the off position out If not twist the button one quarter of a turn clockwise and the button will pop out e Emergency push button mm o BERRY To mount battery charger on wall refer to page 1 Mounting Battery Charger in this manual All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Operating the Stand Up Lift Locking Unlocking the Rear Swivel Casters Perform one 1 of the following TE P UNLOCK LOCKING Press DOWN on the BOTTOM of the locking Pe i lever i UNLOCKING Press DOWN on the TOP of the locking A lever meme Raising Lowering Stand Up Lift RAISING THE LIFT Press the UP button on the hand control to raise the lift arms and the patient LOWERING THE LIFT Press the DOWN button on the hand control to low
28. istance is NOT recommended STANDING SLINGS Before lifting the patient make sure the bottom edge of the standing sling is positioned on the lower back of the patient and the patient s arms are outside the standing sling 5 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Safety Summary continued STANDING SLINGS The belt MUST be snug but comfortable on the patient otherwise the patient can slide out of the sling during transfer possibly causing injury TRANSPORT SLINGS Before lifting the patient make sure the bottom edge of the transport sling is at the base of the spine and the patient s arms are outside the transport sling DO NOT use any kind of material such as a plastic back incontinence pad or seating cushion between the patient and sling material that may cause the patient to slide out of the sling during transferring ALWAYS use the colour coded strap on the standing sling closest to the patient while still maintaining patient stability and comfort Bleached torn cut frayed or broken slings are unsafe and could result in injury Discard Immediately DO NOT alter slings When elevated a few inches off the surface being transferred from and before moving the
29. itations may not apply to you THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE OR PROVINCIAL LAWS AND REQUIREMENTS 25 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e Log Book Customer Name e RPS 440ee User s Manual Customer Address Product Name Serial No Date of Purchase Warranty Period Distributed By Check List Owner Ensure the patient lifter is serviced regularly refer Lifter Maintenance page 33 Contact an Authorised Invacare Distributor Service Agent immediately if there are any problems with the operation of the patient lifter Ensure the Log Book is completed and signed Record any repairs required Withdraw the patient lifter from service if inspection reveals that user safety is jeopardised in any way from use of the patient lifter Ensure ongoing servicing after the Warranty Period is maintained by agreement between you and an Authorised Invacare Distributor Service Agent Authorised Invacare Distributor Service Agent Ensure the patient lifter is serviced in accordance with the Manufacturer s recommendations Ensure the Log Book is completed and signed Record any repairs required Advise t
30. l at the desired position With a pencil mark the MIDDLE hole position Measure down 6 1 2 inches from the pencil mark and drill one 1 mounting hole Install the BOTTOM mounting screw until there is an approximatel 8inch gap between the screw head and the wall Install the battery charger with mounting bracket onto the bottom mounting screw Drill the remaining two 2 mounting holes Install the two 2 remaining mounting screws through the mounting bracket and into the wall Tighten securely Plug the battery charger into the wall electrical outlet NOTE ON LED should illuminate Mounting Bracket vA Battery Charger with Mounting Bracket Mounting Screws BOTTOM Mounting Screw 13 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Charging the Batteries NOTE Invacare recommends the battery be recharged daily to prolong battery life NOTE An audible alarm will sound beep when battery is low 1 Lift UP on the handle on the back of the battery 2 Lift the battery UP and OUT away from the control box Make sure there is an audible click when mounting battery either on the battery charger or control box to confirm proper mounting Otherwis
31. lease prepay shipping charges LIMITATIONS AND EXCLUSIONS THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE PRODUCTS MODIFIED WITHOUT INVACARE S EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY COMPONENT WITHOUT THE SPECIFIC CONSENT OF INVACARE OR TO A PRODUCT DAMAGED BY CIRCUMSTANCES BEYOND INVACARE S CONTROL AND SUCH EVALUATION WILL BE SOLELY DETERMINED BY INVACARE THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE INSTRUCTIONS IN THIS MANUAL THE FOREGOING WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES IMPLIED WARRANTIES IF ANY INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESSED WARRANTY PROVIDED HEREIN AND THE REMEDY FOR VIOLATIONS OF ANY IMPLIED WARRANTY SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN INVACARE SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER Some states DO NOT allow exclusion or limitation of incidental or consequential damage or limitation on how long an implied warranty lasts so the above exclusions and lim
32. mark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e Local Distributor Yes you can Invacare Australia Pty Ltd Invacare New Zealand Invacare Asia Ltd Lenton Place PO Box 5002 North Rocks NSW 2151 Austraia 4 Westfield Place PO Box 62 124 Mt Wellington Auckland New Zealand Lenton Place PO Box 5002 North Rocks NSW 2151 Australia Freephone 1800 460 460 Fax 02 8839 5353 Freephone 0508 468 222 Freefax 0508 807 788 Phone 6 2 8839 5330 Fax 61 2 8839 5331 Email sales invacare com au Web wwwinvacare com au Email sales invacare co nz Web www invacare co nz Email asiasales invacare com Web wwwinvacare com All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific
33. nce MUST be performed ONLY by qualified personnel WARNING notices as used in this manual apply to hazards or unsafe practices which could result in serious bodily harm CAUTION notices as used in this manual apply to hazards or unsafe practices which could result in minor personal injury or property damage NOTES highlight procedures and contain information which assist the operator in understanding the information contained in this manual SAVETHESE INSTRUCTIONS AND KEEP WITH PATIENT LIFT AT ALL TIMES All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Special Notes ALWAYS be aware of the Lift Arms Injury to the patient and or assistant may occur ALWAYS be aware of the Footrest especially the patient s position on the footrest Injury to the patient and or assistant may occur Device contains Lead Acid Gel Cell batteries DO NOT dispose of batteries in normal household waste They MUST be taken to a proper disposal site Contact your local waste management company for information 2 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes
34. neath the patient If desired unhook the transport sling from the top attachment points on the stand up lift NOTE The patient can remain in the upper portion of the transport sling while using the commode 5 If equipped unfasten the leg strap from around the patient s legs 6 Instruct the patient to lift their feet off of the footplate NOTE Assist the patient if necessary 7 Remove the standing or transport sling from around the patient 8 Pull the stand up lift away from the commode 9 Once the patient is ready to be transferred refer to the LIFTING PREPARATION section of this manual 19 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Lifter Maintenance Safety Inspections A person who is suitably and properly qualified and well acquainted with the design use and care of the hoist should perform periodic inspections Inspections should be carried out both routinely at least as often as specified and as necessary All defects and damages that have lead to corrective action should be noted in a log book see page 26 and signed and dated by the inspector Defects and resulting actions should be reported in writing to your supplier If inspection reveals that the sa
35. nect the battery to the control box Refer to CHARGING BATTERIES in the OPERATION section of this manual Replace the battery pack Refer to CHARGING BATTERIES in the OPERATION section of this manual Refer to Electric Actuator Replacement in the CARE AND MAINTENANCE section of this manual Contact your Dealer Refer to ELECTRIC ACTUATOR REPLACEMENT in the CARE AND MAINTENANCE section of this manual Contact your Dealer Pull down slightly on the lift arms NOTE If problems are not remedied by the suggested means please contact your Dealer or Invacare 22 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Care amp Maintenance of your Patient Lifter NOTE Follow the maintenance procedures described in this manual to keep your patient lift in continuous service The Invacare Patient Lift is designed to provide a maximum of safe efficient and satisfactory service with minimum care and maintenance All parts of the Invacare Lift are made of the best grades of steel but metal to metal contact will wear after considerable use There is no adjustment or maintenance of either the casters or brakes other than cleaning lubrication and checking axle and swivel bolts for ti
36. nty is extended only to the original purchaser user of our products This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Invacare warrants the products manufactured to be free from defects in materials and workmanship for a period of three 3 years on the lift and one 1 year on the slings hydraulic pump electric components from the date of purchase If within such warranty period any such product shall be proven to be defective such product shall be repaired or replaced at Invacare s option This warranty does not include any labour or shipping charges incurred in replacement part installation or repair of any such product Invacare s sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement For warranty service please contact the dealer from whom you purchased your Invacare product In the event you do not receive satisfactory warranty service please write directly to Invacare at the address on the back cover provide dealer s name address date of purchase indicate nature of the defect Invacare Corporation will issue a serialized return authorization The defective unit or parts MUST be returned for warranty inspection using the serial number when applicable as identification within 30 days of return authorization date Do not return products to our factory without our prior consent C O D shipments will be refused p
37. ogical solution A soft cloth dampened with water and a small amount of mild detergent is all that is needed to clean the patient lift The lift can be cleaned with non abrasive cleaners Electric Actuator Replacement Remove the nut washer and shoulder bolt that secure the electric actuator to the mast mounting bracket Rest the boom on your shoulder and remove the nut bolt plastic bushing and pinch guard from the boom mounting bracket Remove the electric actuator assembly Reverse steps for installation Base Adjustment The base adjustment should not require any attention other than Linkage l Check the squareness of the legs when in the Square db CLOSED position 2 Place a square on the inside of the legs and base to lo determine the 90 alignment Eur N 3 Adjust the linkage rods until 90 alignment is o geog achieved OU 1 1 D I 1 DL 1 1 1 t L 1 1 D v 24 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Limited Warranty PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warra
38. spine D The patient s arms are outside of the standing or transport sling The rear casters are unlocked F Make sure the legs are in the maximum open position m I5 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual When elevated slightly off the surface being transferred from and before moving the patient check again to make sure that the sling is properly connected to the attachment points of the stand up lift If any attachments are NOT properly in place lower the patient back onto the surface and correct this problem Adjustments for safety and comfort should be made before moving the patient Invacare standing and transport slings are specifically designed to be used in conjunction with the Invacare Stand Up Lift Slings and accessories designed by other manufacturers or other Invacare slings are not to be utilized as a component of Invacare s stand up lift system Use of these products is prohibited and will void Invacare s stand up lift warranty Use only genuine Invacare standing and transport slings and lift accessories to maintain patient safety and product utility If transferring from a wheelchair Wheelchair wheel locks MUST be in the locked position BEFORE lowering th
39. t limitations on accessories 6 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e e RPS 440ee User s Manual Key to Symbols The following symbols are used on lifter and attachment labels The symbol required to be displayed on regulated products for sale on the European Market It indicates that the product complies with applicable European Directives related to health safety environment and consumer protection SWL Safe Working Load Represents the maximum load the lifter is rated at for safe operation The following symbols are used on sling labels and relate to washing instructions Wash at maximum temperature 60 C 140 F permanent press cycle Do not dry clean Dry cleanable Wash at maximum temperature 85 C 198 F permanent Do not tumble dry press cycle Do not use bleach Do not iron X El E WEO Please refer to individual sling labels for complete instructions regarding washing and laundering 7 All information is subject to change without prior notice Photographs are for illustration purposes only and may vary from product description Invacare is a registered trademark and Yes you can is a trademark of Invacare Corporation 2008 Invacare Asia Pacific e
Download Pdf Manuals