Home
ALPHASENSE USER MANUAL 4-20mA Transmitter for Oxygen
Contents
1. ALPHASENSE USER MANUAL Page 1 4 20mA Transmitter for Oxygen Sensors 072 0126 Issue 2 1 INTRODUCTION The Transmitter PCB includes circuitry for an electrochemical oxygen sensor to convert the pA output signal from the sensor to a 2 wire 4 20mA signal The transmitter board is shipped with four mounting pillars and hardware that may be used if required Alphasense 4 20 mA transmitters offer convenience and easy maintenance for gas sensors e Transmitters are shipped pre calibrated for immediate use e Circuitry includes onboard voltage regulator and uses low power two wire transmitter systems allowing the simplest wiring format e Small circuit board size allows smaller sensor housing e Conformally coated circuit board gives environmental protection of circuit e Low power circuitry with excellent performance means no degradation of sensor performance due to electronics e Close PCB tracking and buffer circuit reduces noise pick up and RFI EMI susceptibility Besides periodic sensor re calibration the transmitter electronics require no maintenance throughout the lifetime of the sensor The sensor can be replaced at the end of the two year sensor working life Re calibration is required when the sensor is changed See section 7 Please read these instructions to ensure correct installation use and calibration of your gas sensor transmitter 2 1 Transmitter Specification Diode protection for voltage regulator Alphasens
2. Full calibration requires a nitrogen bottle and accurate blending equipment to generate 15 oxygen to set zero but because the zero and span adjustments are interactive we strongly recommend that the output is checked in ambient air 13 44mA and if the output is less than 12 7mA or greater than 14 1mA then re calibration may be frustrating and we recommend that the transmitter board is returned to Alphasense and a new sensor will be fitted and re calibrated For small re calibrations follow the instructions below The transmitter and sensor were calibrated at 25 C If the sensor will be constantly exposed to a temperature significantly different than 25 C then allow the sensor to stabilise at that temperature then recalibrate as explained below To Recalibrate 1 First ensure that the power supply is connected correctly Ensure that the sensor has been powered for at least one hour before checking calibration 2 Ambient air calibration Either pump or allow clean air to diffuse to the sensor for ten minutes Adjust the zero potentiometer RP2 see figure 3 with a small screwdriver until the reading is within 0 05 mA of the Air Calibration Point shown in Table 3 Full scale oxygen 20 9 oxygen Air Calibration Point mA 02 A2 13 44 Table 3 4 20 mA Transmitter Air Calibration Mass flow oxygen sensors are used in these transmitters With low temperature and pressure dependence these are well suited for fixed system application
3. ar configuration for Figure 4 Mounting pillar configuration for attaching to the lid of an enclosure attaching to the base of an enclosure Long pillars required Pt No 034 0005 00 Figure 5 Side view of mounting to lid of an enclosure Figure 6 Side view of mounting to the base of an enclosure 16 5 a M3 a a 5 a gt y q Pillars Washers MUST be fitted both sides 072 0126 Iss 2 Page 4 3 2 Connection and Wiring Alphasense Limited 4 20mA 02 Iss2 Power to the transmitter board is via a Molex 2 pin mini plug type 22 27 2021 supplied with the transmitter Connect using a screened two core cable to the wires black is ground red is positive by either soldering or using a screw terminal block These leads can be shortened or extended as needed 3 3 Power Supply Your power supply must be between 7 5 and 35 VDC with less than 0 2V ripple Do not supply mains AC power to this unit this will destroy the transmitter and void the warranty The transmitter is protected against incorrect polarity but will not function if you have reversed the power supply wires by connecting the Molex plug incorrectly to the transmitter board socket When selecting the power supply voltage you must not exceed the maximum total loop resistance which includes your measuring resist
4. ating holes Note Long pillars will be required for this configuration Alphasense Pt No 034 0005 00 Figure 1 below shows mounting hole locations dimensions are in mm Figure 2 diagrams the locating holes normally the outer holes are used for mounting while the inner holes are in the same location as the earlier issue of this PCB allowing backward compatibility with the earlier PCB design 14 7 23 20 T p K i 75 SN 3 8 o i i E O E 32 Figure 1 Mounting hole locations Figure 2 Inner and outer mounting holes 072 0126 Iss 2 Page 3 Alphasense Limited 4 20mA 02 Iss2 The pillars are tapped to accept an M3 screw We recommend a screw length that is at least 8mm to ensure rigid fixing It is good practice to hold the pillar when screwing into the pillars to stop the pillar from rotating on the PCB It may be easier to remove the sensor whilst screwing the circuit board pillars to your housing If you move the pillars ensure that if mounting to the lid of your housing that you include the washer between the pillar and PCB to ensure correct height of the pillar assembly See figure 3 Allow 60 minutes after plugging the sensor back into the board for the output to stabilise before calibrating or checking performance Pillar Connector _ Washers Figure 3 Mounting pill
5. e Limited 4 20mA 02 Iss2 072 0126 Iss 2 Page 2 2 2 Range Options Sensor and transmitter boards are shipped from Alphasense pre calibrated You may wish to confirm calibration Standard available range is listed below Sensor Oxygen Concentration Range 02 A2 15 to 25 Table 1 Oxygen Transmitter ranges Although the sensor and transmitter are pre calibrated and the ranges are preset it is possible to modify the range by adjusting the zero and gain potentiometers which changes the circuit gain and offset Contact Alphasense before proceeding 3 Set Up Oxygen sensors are packed separate from the transmitter board to prevent loss of lifetime during shipping and storage The sensor and transmitter are matched so to reconnect the sensor and transmitter board 1 Remove the oxygen sensor from the clear bag 2 Plug the sensor into the board 3 Wait one hour for the sensor to stabilise before checking that the output in ambient air is 13 3 to 13 6mA 3 1 Mechanical Mounting Transmitters should be mounted to your housing using the pillars provided These pillars should first be attached to the PCB using the supplied screws and washers Important note Ensure that a washer is fitted on BOTH sides of the PCB see figure 3 Two sets of mounting holes are provided so that the assembly can be fixed to either the housing top using the locating holes in the corner of the PCB or to the base of the housing using either set of loc
6. or used to change the 4 20 mA current into a measured voltage The transmitter requires a minimum of 7 5 volts to operate therefore the maximum potential drop allowed across your sensing resistor and cable is power supply voltage 7 5V Assuming full scale deflection at 20 mA use Ohm s law to calculate the maximum loop cable plus sensing resistor allowed resistance 4 Correct Usage and Maintenance Ensure that there is a good gas seal between the sensor and the housing also if the sample is pumped then ensure that the flow rate is sufficient Alternatively the sample gas can be allowed to diffuse to the front face of the sensor Alphasense recommends 250 to 500 standard cubic centimetres per minute sccm flow rate if pumped Higher flow rates may be used but beware that pulsing flow and higher pressure drops may lead to erratic readings No maintenance is required but if the top dust oil filter has become badly contaminated then contact Alphasense for replacement dust filter See table 2 below Part Number Description 024 0023 00 Self adhesive dust oil filter 010 0178 00 Cable assembly 034 0005 00 20mm long Spacer Table 2 Replacement part numbers Alphasense Limited 4 20mA 02 Iss2 072 0126 Iss 2 Page 5 5 Calibration The 4 20mA transmitter is shipped pre calibrated to the range shown in 2 1 Periodic re calibration is required for all gas sensors especially in safety critical applications
7. s This type of oxygen sensor has a small non linearity which can normally be ignored Table 4 below lists the output error measured output actual concentration due to non linearity O2 02 A2 15 25 02 range 15 0 20 0 25 0 5 Table 4 Transmitter non linearity Alphasense Limited 4 20mA 02 Iss2 072 0126 Iss 2 Page 6 6 Warranty Transmitters are warranted for two years Alphasense O2 A2 oxygen sensors are warranted for two years If you have any difficulties or problems then contact Customer Support Alphasense Limited Sensor Technology House 300 Avenue West Skyline 120 Great Notley Essex United Kingdom CM77 7AA Tel 44 0 1376 556 700 Fax 44 0 1376 335 899 email sensors alphasense com 7 Attachment Figure 7 Circuit diagram Alphasense Limited 4 20mA 02 Iss2 Qphasense Ltd Sensor Technology House 300 Avenue West Skyline 120 Great Notley Essex United Kingdom CM77 7AA Tel 44 0 1376 556 7008 Fax 44 0 1376 335 899 e mail sensors alphasense com web www alphasense com
Download Pdf Manuals
Related Search
Related Contents
Hardware AZ Analog Drives - Advanced Motion Controls ProLab 4000 pH/ISE/conductivité Brodit Passive Holder,Tilt Swivel HÔTEL DES VENTES DE COULOMMIERS Manuel d`utilisation model/modelo 1696 model/modelo 1696 Samsung AQV18AWAX Manual de Usuario Venom Twin Rechargeable Battery Packs Copyright © All rights reserved.
Failed to retrieve file