Dental Film Digitizer - Vidar Systems Corporation
Contents
1. 1 Lee eee eee Leeecieeeni eene seen n nnn n nhan nana nana nn anna nana 15 Gleaming tie DiS Sr t 15 Fe DIAGN Ne LED Bal rr 24 Appendix Electromagnetic Guidance ccccssscccssseessseecenseecesseeeenseseeneesoeseeeenseesensessones 25 hleler Dnelremednee c c 29 18812 001 Rev D Dental Film Digitizer User s Guide 5 Intended Use The VIDAR dental film digitizer is used for making digital copies of x ray film commonly used in dental practices Images captured with this device are intended for use in primary secondary and over reading applications The target users of the device are medical professional or trained staff 18812 001 Rev D Dental Film Digitizer User s Guide Operating The Dental Film Digitizer Turning the Digitizer On and Off CAUTION Do not apply power by plugging the DC power connector into the digitizer while the power adapter is live Do not remove power by unplugging the DC power connector from the digitizer while the power adapter is live Either of these actions could permanently damage the digitizer the power adapter or both If the film digitizer is already on and ready to scan LED on the front of the digitizer is solid green then skip this section 1 Ensure the PC 1s turned off Note Always apply power to the digitizer before turning on the computer This enables the c
2. Note The digitizer cannot scan films smaller than 6 width x 4 length 15 2cm x 10 2cm Place smaller films in a carrier then feed the carrier into the digitizer Ensure the LED on the front of the digitizer is solid green 2 Place the film or carrier into the digitizer left justified as shown here When the Dental Film Digitizer detects the film carrier it stages the film carrier pulls it in about 1 2 5cm then pushes it back out about 1 2 1 25cm This may take a few seconds CAUTION If the film or carrier is not staged properly it is recommended to use the scanning application s eject command to remove the film Gently pulling the film out of the digitizer will not harm the digitizer but could potentially result in damage to the film Note When a film carrier has been staged it is in the digitizer s light path The ADC Automatic Digitizer Calibration feature requires that the light path be clear of the film carrier for proper background calibration of the digitizer The digitizer will automatically adjust the film or carrier s position to properly proceed with ADC Depending on the location of the film or carrier s leading or trailing edge and the length of time it has been blocking the light path and then it will be a pushed up b pushed down or c ejected to remove it from the light path 3 Using your scanning software execute the Scan command 18812 001 Rev D Dental Film Digitizer User
3. with nitrous oxide Class I Medical Device No Applied Parts This product provides Class I medical device protection against electrical shock WARNING To avoid the risk of electric shock this equipment must only be connected to a supply mains with protective earth Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked HOSPITAL ONLY or HOSPITAL GRADE WARNING Do not modify this equipment without authorization of the manufacture Do not position the product so that it is difficult to disconnect the product from mains power supply Mains power may be removed from the product by disconnecting the power cable at the bottom of the product Mains power may also be disconnected by unplugging the power cord from the external power supply adapter or unplugging the power cord from the wall outlet Power cords used with this device in North America must be rated by Underwriters Laboratories for hospital use Power cords used with this device in Europe must meet the requirements of IEC 227 Designation 53 or IEC 245 Designation 53 This product is not suitable for use in a patient environment Do not use the product or the host computer in the vicinity of a patient Do not touch the product or the host computer while touching the patient The use of portable or mobile communications equipment and or the presence of strong electromagnetic and or x ray fields may interfere with proper op
4. 1 1 Power frequency 50 60 Hz magnetic field IEC 61000 4 8 6 kV contact 8 KV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 UT gt 95 96 dip in Ur for 0 5 cycle 40 Ut 60 96 dip in Ur for 5 cycles 70 96 Ut 30 96 dip in Ur for 25 cycles lt 5 Ur gt 95 96 dip in Ur for 5 sec 6 kV contact 8kV air 2 kV for power supply lines 1 kV for input output lines 1 kV differential mode 2 kV common mode lt 5 UT gt 95 dip in UT for 0 5 cycle 40 Ur 60 96 dip in Ur for 5 cycles 70 Ut 30 96 dip in Ur for 25 cycles lt 5 UT gt 95 96 dip in Ur for 5 sec Floors should be wood concrete or ceramic tile floors are covered with synthetic material the relative humidity should be at least 3096 Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment Mains power quality should be that of a typical commercial or hospital environment If the user of the Dental Film Digitizer requires continued operation during power mains interruptions it is recommended that the Dental Film Digitizer be powered from an uninterruptible power supply or a battery Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital e
5. and 800MHz the higher frequency range applies Note 2 These guidelines may not be applicable in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the Dental Film Digitizer is used exceeds the applicable RF compliance level above the Dental Film Digitizer should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the Digitizer Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m 18812 001 Rev D Dental Film Digitizer User s Guide 27 Recommended Separation Distances Between Portable And Mobile Rf Communications Equipment And The Dental Film Digitizer Table 206 The Dental Film Digitizer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the Dental Film Digitizer can help prevent electromagnetic interference by maintaining a minimum distance bet
6. cover will not seat properly 7 Close digitizer a Place front cover next to main unit b Connect front LED display cable c Rotate front cover down onto main unit 18812 001 Rev D Dental Film Digitizer User s Guide 21 d Using four screws removed in step 3a secure front cover to main unit two screws shown here by arrows other two are on opposite end of front cover e Turn digitizer over 8 Complete assembly a Using 1 5mm Allen key insert four screws into film feed area Tighten screws so that the plastic edge is nearly flush with the blue edge Pushing the plastic flush will help b Replace digitizer on its stand 18812 001 Rev D Dental Film Digitizer User s Guide 22 c Plug USB cable into USB port on bottom of digitizer Back of digitizer USB cable To power adapter d Plug power cable into jack on bottom of digitizer 9 Return digitizer to service a Assure PC is off b Plug digitizer power adapter cord into wall outlet c Turn digitizer power switch on d Observe LED on front of digitizer it should flash light blue for several minutes indicating digitizer is performing internal tests and calibration e When LED stops flashing and remains solid green turn on PC f Digitizer is now ready to scan films 18812 001 Rev D Dental Film Digitizer User s Guide 23 Replacing the LED Bar 18812 001 Rev D Note LED bars can only be replaced on Dental Digitizers with s
7. functions correctly and digitizer does not function Get qualified technical help Ensure that power cable is properly plugged into digitizer power cable connects to bottom of digitizer If digitizer then appears normal wait for solid green LED then reboot PC or if that doesn t solve the problem Ensure that electrical power cord is properly plugged into power adapter If digitizer then appears normal wait for solid green LED then reboot PC or if that doesn t solve the problem Get qualified technical help continued Dental Film Digitizer User s Guide 12 Operational Problems Symptom Film carrier tilts or skews during scanning Corrective action 1 Ensure individual films are loaded properly bottom edge of film carrier parallel to rollers left edge of film carrier up against digitizer body as far left as it will go Review Feeding films and carriers in Operating chapter 2 Ensure that you are feeding the film straight 3 Remove any foreign matter from film entrance slot 4 If problem persists get qualified technical help After film carrier is loaded and staged properly a few minutes later itis pushed out of staging area When film carrier is staged it is over digitizer s light path The digitizer s ADC Automatic Digitizer Calibration feature requires that light path be clear for proper background calibration Thus if film carrier remains in digitizer for m
8. intended for use in the electromagnetic environment specified below The customer or end user of the Dental Film Digitizer should assure that it is used in such an environment Emissions test Compliance Electromagnetic environment guidance HF Emissions The Dental Film Digitizer uses RF energy only for its internal Group 1 function Therefore its RF emissions are very low and are not likely CISPHR 1 1 to cause any interference in nearby electronic equipment RF Emissions CISPR11 Harmonic Emissions Class A The Dental Film Digitizer is suitable for use in all establishments IEC 61000 3 2 other than domestic and those directly connected to the public low k th i ildi f Voltage Fluctuations sie renee ne y network that supplies buildings used for flicker emissions Complies IEC 61000 3 3 18812 001 Rev D Dental Film Digitizer User s Guide 25 Guidance and Manufacturer s Declaration Electromagnetic Immunity Table 202 The Dental Film Digitizer is intended for use in the electromagnetic environment specified below The customer or end user of the Dental Film Digitizer should assure that it is used in such an environment Immunity test o n ies Compliance level Electromagnetic environment guidance ElectroStatic Discharge ESD IEC 61000 4 2 Electrical fast transient burst IEC 61000 4 4 Surge IEC 61000 4 5 Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4
9. User s Guide for VIDAR Dental Film Digitizer Serial numbers 210 000 to 219 999 O Copyright by VIDAR Systems Corporation All rights reserved Terms denoted by the trademark symbol 1M or 9 used in this publication are trademarks or service marks registered in the United States or other countries NOTE The digitizers do not include application specific software Picture Archiving and Communications PAC system Teleradiology Oncology Systems or Computer Aided Detection CAD software The manufacturer of the application software will determine specific indications for use These third party software packages or complete systems are approved separately from a regulatory perspective IDAR r systems corporation A 3D Systems Company 18812 001 Rev D 365 Herndon Parkway Herndon VA 20170 U S A Phone 1 703 471 7070 Fax 1 703 471 7665 Internet www filmdigitizer com Emergo Europe Molenstraat15 2513 BH The Hague The Netherlands Dental Film Digitizer User s Guide Sales 1 800 471 SCAN or 1 800 471 7226 Email sales VIDAR com Technical Support 1 800 471 SCAN or 1 703 471 7070 E mail medtech VIDAR com Phone 31 0 70 345 8570 Fax 31 0 70 346 7299 E mail service emergogroup com 18812 001 Rev D Caution No operator serviceable parts inside Refer servicing to qualified personnel Achtung Geh use nicht ffnen Wartung uno reparatur nur durch eletrofachkr fte Attention Auc
10. eration of this product This product should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary verify normal operation in the configuration in which it will be used Should such interference occur the user is required to provide adequate isolation between the digitizer and the source of the interference Isolation is typically achieved by moving the digitizer away from the source of the interference Correct and safe operation of the digitizer requires familiarity with information that is not marked on the product The following symbols indicate that the operator should consult the manual for additional information 18812 001 Rev D Dental Film Digitizer User s Guide 4 Contents luliziple 1e RS SO eter He B PER 6 Operati Me 7 Turning the Digitizer On and Off ener nnn nnns 7 PADOUIG m REIN E EE EDD aaa 8 Feeding Films and Carriers ccccccccccssseseeeccecceeeeeeeceeeeeesueeeeeceeeeeesaeeaeeeeeeesssaeeeeeeeeesaaas 9 cerle 11 Normal Operaio 9 secco aduenit dete e Senate o nn Deu ndi ioc enn 11 lower Adapter Information s iti pie cto dava onere eet crepat ar brinda D eU Fe du 11 hEerc redBhiedn c 12 zielt 12 Operational PrODIEMS esrar enisi eoo deii ortae Pope ene aa biae eana reo dede obe ae aleeseen 13 Cleaning and Maintaining
11. erial numbers between 210 000 and 219 999 Please contact VIDAR Technical Support to order the Lamp Bar Replacement Kit VIDAR part number 18629 001 m Phone 1 800 471 SCAN 1 800 471 7226 1 703 471 7070 outside the U S m E mail medtech vidar com Instructions for replacing the LED bar are provided with the replacement kit Dental Film Digitizer User s Guide 24 Appendix Electromagnetic Guidance Caution Medical electrical equipment EMC Electro Magnetic Compatibility must be considered before any medical electrical equipment is installed or put into service Follow the information in the accompanying documentation when installing and operating the VIDAR Dental Digitizer Caution Portable or mobile RF communication equipment can effect Medical Electrical equipment Caution Using the Digitizer adjacent to or stacked with other equipment may cause interference between the equipment Before utilizing stacked or adjacent equipment verify proper functionality of all equipment in the actual configuration in which it will operate Caution Connecting the Digitizer to equipment that is not rated CISPR 11 class A or class B may alter the electromagnetic characteristics Caution In a domestic environment this product may cause radio interference in which case the user may be required to take adequate measures Guidance And Manufacturer s Declaration Electromagnetic Equipment Table 201 The Dental Film Digitizer is
12. etween 210 000 and 219 999 IMPORTANT You must follow these instructions when cleaning the diffuser Deviation from these instructions will void the product s warranty and will likely result in costly repairs You will need m 2 medium Phillips screwdriver m 1 small Phillips screwdriver m Small cutters m Soft cloth or foam sheet to protect front of digitizer m Lint free wipes VIDAR part number 15194 m 2cable ties 1 Shut down system a Shut down and turn off PC b Turn digitizer power switch off c Disconnect digitizer power adapter cord from wall outlet d Disconnect USB cable and power cable from bottom of digitizer e Remove digitizer from stand 18812 001 Rev D Dental Film Digitizer User s Guide 15 18812 001 Rev D 2 Prepare digitizer a Using 1 5mm Allen key remove four screws from film feed area b Place digitizer face down on soft cloth or foam sheet Top should be toward you Dental Film Digitizer User s Guide 16 18812 001 Rev D 3 Open digitizer a Remove four screws securing front cover two screws shown here by arrows other two are on opposite end of front cover v Do not remove four screws in feet on rear cover b Holding cover in place flip entire unit over onto its feet bottom toward you c Rotate front cover up d Disconnect front display LED cable e Place cover aside Dental Film Digitizer User s Guide 17 18812 001 R
13. ev D 4 Remove LED bar a Cut the cable tie securing the cables to the LED bar b Remove the two connectors from the LED bar press in on each latch to release the connector from its socket c Remove four screws securing LED bar to main unit frame two screws at each end d Remove the upper screws at each end of the LED bar Dental Film Digitizer User s Guide 18 18812 001 Rev D e Grasp both ends of LED bar then pull gently upward you until it is free of main unit frame 5 Clean diffuser Using lint free wipe clean entire length of diffuser white plastic strip on LED bar Clean diffuser with a single pass for each unused area of the wipe Do not scrub the diffuser CAUTION Clean only with lint free wipes provided by VIDAR part number 15194 Other cloth or paper wipes will leave fibers on the diffuser which can affect performance After cleaning do not touch diffuser Dental Film Digitizer User s Guide 6 Reinstall LED bar a Hold LED bar in both hands with diffuser facing digitizer and with cable tie mount to right b Slide LED bar into main unit frame c Secure LED bar to main unit frame with four screws two screws at each end Tighten screws 18812 001 Rev D Dental Film Digitizer User s Guide 20 e Secure LED bar cables with cable tie Cut off excess cable tie Cable tie must be in orientation shown tail pointing toward digitizer bottom or front
14. ice doesn t work AC power is not available at that wall outlet Get help from building services m If other device does work AC power is available from that outlet Go to step 2 With power applied to digitizer look into digitizer film entrance slot If you see light coming from inside digitizer power is on For this condition LED off internal light on get qualified technical help If you have another Dental Film Digitizer try its power adapter m f digitizer works with other power adapter original power adapter is defective Obtain replacement VIDAR Dental Film Digitizer power adapter from your system integration specialist Do not substitute any other power adapter m If digitizer does not work with a known good power adapter get qualified technical help CAUTION Using a non approved power adapter will void the warranty Film carrier starts and stops during scanning 1 Increase memory allocation for scanning software Especially if scanning at high resolutions memory allocation must be sufficient to accept data stream from digitizer If scanning to disk ensure sufficient space is available on disk drive Ensure PC has enough memory available to support scanning application Close applications not needed for image acquisition from digitizer If problem persists get qualified technical help Digitizer stops scanning and PC locks up 1 Reset entire system Remember that digitizer sh
15. ith the power supply adapter provided by your Service Provider or Vidar Systems Corp 18812 001 Rev D Dental Film Digitizer User s Guide 11 In Case of Difficulty LED indication Flashing dark blue How to use this section 1 Look through the left columns to find a description of the problem you are having 2 Follow the instructions in order in the Corrective action column If the one corrective action doesn t solve the problem then carry out the next corrective action 3 When the instruction is Get qualified technical help then a Contact your system integration specialist the company or person that installed your VIDAR film digitizer b If your system integration specialist isn t available then contact VIDAR Customer Support medtech Q vidar com or 1 703 471 7070 Please record system information before calling and have it available when calling Tip Check www vidar com for current troubleshooting information tools and software updates Error Conditions Condition Normalization error 3 Corrective action 1 Clean lamp diffuser see Cleaning the diffuser in the Cleaning and maintaining chapter later in this manual or if that doesn t solve the problem Replace lamp instructions are provided with the LED Bar Replacement Kit VIDAR part number 18629 or if that doesn t solve the problem Get qualified technical help No LED No LED 18812 001 Rev D and digitizer
16. itself Indicates the need for the user to consult the instructions for use Underwriters Laboratories classification symbol Compliance to EU Directives Authorized Representative in the European Community Indicates the Medical Device Manufacturer This symbol on the product indicates that this product must not be disposed of as unsorted municipal waste Instead it is your responsibility to dispose of your waste equipment by handing it over to a designated collection point for recycling of waste electrical and electronic equipment For more information about where you can dispose of your waste equipment for recycling please contact your local city office your household waste disposal service or the shop where you purchased the product Dental Film Digitizer User s Guide 29 T systems corporation A 3D Systems Company 365 Herndon Parkway Herndon VA 20170 USA Phone 1 703 471 7070 Main fax 1 703 471 1165 Web www vidar com Technical Support Phone 1 703 471 7070 1 800 471 7226 Fax 1 703 471 7665 E mail medtech vidar com
17. lative humidity 20 to 85 non condensing m Atmospheric pressure 696 to 1013hPa 10 000 to O ft Electrical supply m Voltage 100 to 240 VAC m Current 1 0 to 0 42 A m Frequency 50 to 60 Hz 18812 001 Rev D DentalFilmDigiizerUsersGuide 118 Safety and Compliance Information MEDICAL GENERAL MEDICAL EQUIPMENT AS TO ELECTRIC SHOCK FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH CAN CSA C22 2 No 601 1 M90 IEC 60601 1 UL 60601 1 ANSI AAMI ES60601 1 2005 3RD Ed CAN CSA C22 2 No 60601 1 2008 IEC 60601 1 2005 3RD Ed Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC standards i e IEC 60950 for data processing equipment and IEC 60601 1 for medical equipment Furthermore all configurations shall comply with the system standard IEC 60601 1 1 or IEC 60601 1 3 edition Everybody who connects additional equipment to the signal input part or signal output part configures a medical system and is therefore responsible that the system complies with the requirements of IEC 60606 1 1 or IEC 60601 1 3 edition If in doubt consult the technical services department or your local representative This product is rated for continuous use This product is in the Ordinary Equipment Class It provides no protection against the ingress of water This product is not suitable for use in the presence of flammable anesthetic mixtures with air or with oxygen or
18. n the above dimensions m Periapical Film single tooth and surrounding structures e Size0 35 mm x 22 mm e Size 40mm x 24 mm e Size 2 40 5 mm x 30 5 mm m Bite Wing Film interproximal examination e Size0 35 mm x 22 mm e Size 2 40 5 mm x 30 5 mm e Size 3 54 mm x 27 mm m Occlusal Film large areas of the mandible and maxilla e Size 4 57 mm x 76 mm Note VIDAR includes sample acceptable film holders with the film digitizer These holders can be used multiple times however they should not be used if they are torn or bent in any way Additional holders can be purchased from VIDAR from your Rinn Dentsply representative The following part numbers are shipped with the VIDAR Dental Film Digitizer 214567 216081 201163 216200 200138 Paper based or cardboard holders should NEVER be used in the VIDAR Dental Film Digitizer Films in the following sizes can be scanned by placing them directly in the digitizer m Extraoral 5 x 7 8 x 10 10 x 12 m Panoramic Ceph 5 x 12 13 cm x 30 em 6 x 12 15 cm to 30 cm m Other e 24cm x 18 cm to determine sagittal and vertical relationships e 24cm x 22cm hand wrist growth measurement exams Dental Film Digitizer User s Guide IMPORTANT Remove stickers tape staples paper clips etc from films before scanning These may cause serious film feeding problems Failure to remove extraneous items from films will void your warranty Feeding Films and Carriers
19. nvironment NOTE Ur is the A C mains voltage prior to application of the test level 18812 001 Rev D Dental Film Digitizer User s Guide 26 Guidance And Manufacturer s Declaration Electromagnetic Immunity Table 204 The Dental Film Digitizer is intended for use in the electromagnetic environment specified below The customer or end user of the Dental Film Digitizer should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Portable and mobile RF communication equipment should be used no closer to any part of the Dental Film Digitizer including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 V d 1 24 P IEC 61000 4 6 150 kHz to 80 MHz Hadiated RF 3 V m 3 V m d 1 2N P 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz d 22 34 P 800MHzto 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol e Note 1 At 80MHz
20. omputer to recognize the digitizer 2 Turn on the digitizer the switch is located behind the upper right corner of the digitizer body 3 Observe the LED on the front of the digitizer the LED should flash light blue for several minutes This indicates the digitizer 1s performing internal tests and calibration 4 When the LED stops flashing and remains solid green turn on the PC The digitizer is now ready to scan films 18812 001 Rev D Dental Film Digitizer User s Guide 7 About Films 18812 001 Rev D Note For best possible imaging performance please allow the unit to warm up for 30 minutes prior to use You can also choose to leave the digitizer powered on and simply turn the PC on and off as needed CAUTION X ray images displayed on a computer monitor are representative only Dimensional and grayscale inaccuracies may result from the build up of tolerances in the digitizer the display board and the computer For this reason special precautions must be exercised when taking measurements from the digitized image Please refer to the user s manual for your digitizing software for more information if you wish to apply measurements The Dental Digitizer accepts films and carriers within the following dimensions m Minimum size 6 15 24 cm wide x 4 8 89 cm long m Maximum size 13 33 02 cm wide x 13 33 02 cm long Films in the following sizes can be scanned by placing them in a film holder that is withi
21. ore than 5 minutes it will be pushed up 1 4 0 6cm to clear light path Later when you execute your software s Scan command film carrier will be re staged before it is digitized Digitizer has been off for several days First film carrier misfeeds or jams Run three or four films carriers through digitizer load them one at a time then eject each one or perform a scan on each film carrier but this takes longer Can t turn digitizer on or off Can t find power switch The on off switch is located behind the upper left corner of the digitizer body when viewed from the front After applying power LED on front keeps flashing light blue After power is applied digitizer normally requires a little over 5 minutes to normalize and calibrate during which time the LED will flash light blue If LED flashes light blue for more than 15 minutes turn digitizer power off and on again as described later in this chapter then reboot PC Streaks in image Clean LED bar diffuser see Cleaning the diffuser in the Cleaning and maintaining chapter later in this manual 18812 001 Rev D continued Dental Film Digitizer User s Guide 13 Symptom LED is off in spite of using proper power on procedure described later in this chapter Corrective action 1 Ensure wall outlet is providing AC power obtain another electrical device known to be working and plug it into that AC wall outlet m f other dev
22. ould be powered before PC a Turn digitizer off b Shut down PC and turn it off c Turn digitizer on d After digitizer LED turns solid green turn on PC Launch scanning software on PC then try digitizing again If problem persists get qualified technical help Check for correct device driver installation Digitizer was working properly but after installing or reinstalling the scanning application there are Toolkit errors or the digitizer is not detected 1 If you installed a new scanning application or if you reinstalled the existing application the older Toolkit vscsi32 dll may have been installed Run the VIDAR Drivers and Toolkit Installation CD see appropriate section in the Installation Guide for Dental Film Digitizer 2 If problem persists contact medtech vidar com or call Medtech at 1 703 471 7070 Digitizer is not listed in Windows 1 Update to or reinstall STI drivers See the Install device drivers and Control Panel Scanners and toolkit chapter of the Installation Guide for Dental Film Digitizer Cameras 2 Check all USB cables and connectors 18812 001 Rev D Dental Film Digitizer User s Guide 14 Cleaning and Maintaining The Dental Film Digitizer Cleaning the Diffuser Note Use this procedure only when you observe streaks in images or when the digitizer s LED flashes dark blue Note Use this procedure only for Dental Film Digitizers with serial numbers b
23. s Guide The LED will flash green rapidly indicating the scan is in progress When the scan is complete the LED will return to solid green Note After the scan the film carrier may or may not be ejected your scanning software determines what happens If the software s Auto Eject function is activated the film carrier will be pushed out Otherwise the lower rollers in the scanner will hold it until one of the following occurs m Another film carrier is detected m You manually eject the film carrier using the software m he ADC process begins by ejecting the film carrier Note Proper orientation of the film during loading depends on the scanning application program you are using Some scanning programs rotate images 180 for display If images appear upside down you may be able to set the default orientation in the scanning application If this option isn t available then insert films into the digitizer reversed as shown here j o ta WOLLOG 18812 001 Rev D Dental Film Digitizer User s Guide Normal Operation LED indication Flashing light blue Troubleshooting The Dental Film Digitizer Condition Digitizer calibrating please wait Flashing dark blue See Error conditions in the In case of difficulty section later in this chapter Solid green Flashing green Power Adapter Information Ready to scan Scan in progress CAUTION Power the digitizer only w
24. une piece ne peut etre remplacee par l utilisateur Toute operation de maintenance doit etre effectuee par une personne qualiee Atencion Acceso interno solo autorizado a personal tecnico cualificado Attenzione Non appire Rivolgersi a personale qualificado Radio Frequency Emissions This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area can cause harmful interference in which case the user will be required to correct the interference at his or her own expense Product compliance testing was conducted using VIDAR shielded cables Modifications to the digitizer or the VIDAR shielded cables or the use of cables other than those available from VIDAR could void the user s authority to operate the equipment Acceptable shipping conditions m Temperature 15 to 60 C 5 to 140 F m Relative humidity 20 to 85 non condensing Atmospheric pressure 500 to 1060hPa 18 000 to 1 200ft Operating conditions Temperature 15 C to 30 C 60 F to 85 F m Re
25. ween portable and mobile RF communications equipment transmitters and the Dental Film Digitizer as recommended below according to the maximum output power of the Bebe equipment Rated maximum output Separation distance according to frequency of transmitter m power of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 18812 001 Rev D Dental Film Digitizer User s Guide 28 Appendix Symbols 18812 001 Rev D Indicates the OFF condition for a part of equipment not used to identify the OFF position of a switch Indicates the ON condition for a part of equipment not used to identify the ON position of a switch Indicates the need for the user to consult the instructions for use for important cautionary information such as warnings and precautions that cannot for a variety of reasons be presented on the medical device