        User Guide for the Multichannel Confocal 050213
         Contents
1.   to set Mark Points parameters   6  Click Prepare   7  Click Acquire to mark points     Note  If the PA does not occur at the marked points let me know     it may need a new calibration file     
2.  be captured at all XY locations set in the XY   Stage tab    A Z series can be saved so that the same series can be used again    Save Path is used to save the data to a particular location on the computer     Note     Check the power set up in the z series box  If the series is defined by entering values in the z   series calculator the default is that the power will decrease through the series     The T Series tab controls the settings for creating a series of images and data using various parameters such as  XY stage locations  regions of interest  and Z series     A cycle can be set up by changing values in the data table    A new cycle can be added or inserted in the Add Insert Cycle panel by clicking    Single Image      A particular x y stage location can be selected by clicking    Stage Position      Waits can be inserted into the series by clicking    Wait        Iterations indicates the number of times the cycle will repeat    Period is the time from the start of one iteration to the start of the next iteration   Z series can be added using the drop down menu in the data table    To start running the T series you have created  click    Start T Series       Save Path is used to save the data     T Series Preferences  under preferences menu     T Series Execution Order  Determines whether to do all iterations at a given XY stage location before  proceeding to the next XY stage location      1  T Series Iterations    2  XY Stage Locations    3   Cycles     or to perfo
3.  locations     The Misc tab contains the main Save Path for saving your data  The photoactivation mask can also be selected  in the misc tab     The Image Window also contains important controls and settings     To the left of the pictures  are the channel buttons  Ch1 Ch2 Ch3 Ch4 indicate which channel is being  shown in that image window  The small rectangle directly to the right of the channel buttons allow you  to choose the color that that particular channel is shown in    Freeze will freeze the current image in the window  Data will continue to be collected  but it will not be  shown in the frozen window    BOT  brightness over time  can be used to watch the change in pixel intensity over time    LS  line scan  can be used to define and acquire data along a line    PS  point scan  can be used to keep the laser beam focused at a particular spot  Itis possible to record  intensity at this spot over time  Photobleaching is a side effect of point scanning     ROI  region of interest  can be used to choose an area of the image to zoom into  Unlike optical  zooming  region of interest selection does not create a power issue  Image size is chosen automatically  when using ROI    LUT  look up table  can be used to observe the intensity of acquired data  The intensity of an image can  be changed without affecting the raw data  This is a useful way to test whether you can use less power   The button with a graph image can be used to display a plot of intensity values in each chan
4.  the Stage Controls  it is useful to hit both Os in the panel to bring all settings  to    home       The set of arrows on the left can be used to move the stage in the x or y axes  This can also be done  using the joystick    The right set of arrows can be used to set the top and bottom positions of a z stack  the boxes with the  red lines     Step sizes  um  can be changed by entering a number in the appropriate box  Other step sizes can be  chosen from the drop down menu     The tabs halfway down the main control window contain additional controls and experimental setups     The Laser  PMT  DAQ tab controls the power given to the laser and PMTs     The laser setting goes from 1 1000  arbitrary units     The PMTs setting denotes the amount of voltage applied to the PMT    The green bar to the left of the PMT settings will change the menu to set the gains and offsets for each  channel     The Z Series tab controls the settings for creating z stacks     The top and bottom positions as well as step size can be set here or in the Stage Control panel above   The number of slices wanted can be set in this tab    Four values are needed to set z series parameters     start position  stop position  step size  and number of  slices  When any three values are set  the fourth will be automatically calculated    If    Adjust PMT  amp  Laser    is checked  different power and gain settings can be entered in the data table   If    At all XY stage locations    is checked  the z series will
5. For example  if a system has two imaging lasers  one at 488 nm and one at 561 nm this feature allows the  operator to scan a 512x512 image alternating wavelength with each line  The data for track one will be saved as  the Channel One data and the track two data will be saved as channel two     Whenever the Interlaced scan pattern check box is checked  the defined interlaced settings will be used  This  applies to Live Scan  Single Scan  T Series  Z Series   etc     Prairie View Controls  Note  Some sections have extra controls accessible through a green bar located to the left of the section   Image Size is the first panel on the top left corner of the main control window     e Image size is the pixels per image   e 512x 512 is recommended as a starting size for a fast frame rate and quick imaging   e Higher resolution can be used once the object is in view     Image Window Size is located under Image Size     Image Window Size controls the way in which the image is displayed in the window   Fit will change the image to fit within the window    1 1 will show the image at its actual size as defined by Image Size    Smaller will decrease image by 10   click repeatedly to continue decreasing size    Larger will increase image by 10   click repeatedly to contine increasing size      Dwell Time Per Pixel  us  is located under Image Window Size     e Dwell Time designates the amount of time the laser is on each pixel     The value on the left is the minimum dwell time available an
6. Multichannel Confocal User Guide    Note  for more details see the Prairie user manual at http   www  prairie  technologies com resources software Prairie View html    Please report any problems to Julie Last  jalast wisc edu     Turning on the instrument    1     ON Ot ae    10     11    13     2  3  Turn on laser launches  on the right side of the computer cart   Let warm up for 30 minutes   Note  The bottom laser launch  b  is for uncaging and doesn   t need to be turned on for imaging      Turn on main power switches on the front     Use key to turn lasers on     Open shutters  On cart turn on main power  behind the monitor   Turn on the computer  open panel on from of computer tower to find the power button    Turn on the power behind the microscope  on the right side   Open Prairie View software  Select objective either in the software in the Nikon controls window or on the LHS of the microscope   The Prairie View software will automatically update the objective selection   Loading the sample  To load the sample  tilt back the top of the microscope  Caution  The top is heavy  and will not stay up by itself   To look through eyepiece select eye on the front of the microscope  The light can be turned on and off  and intensity adjusted on the left side of the microscope       For confocal imaging select L100 on the front of the microscope and turn off the transmitted light   12     Joystick control   a  If the joystick does not work check to see if the boxes on top of the 
7. can will scan the object once and display that image   New window will open a new imaging window   Running frame average is used with live scanning   o If the box is checked for running frame average  the software will average the number of frames  indicated in the box   o Averaging frames can increase signal to noise ratio and resolve an image   o Averaging frames is only useful if the object is not being moved   Average every N frames is used with single scan   o This function will average the number of frames indicated in the box   o A higher number will result in a higher signal to noise ratio   Hard Shutter is the shutter located in front of the laser box   o When activated  denoted by a red icon   the laser is live   o The Hard Shutter will be activated automatically when acquisition begins   Soft Shutter controls the voltage being applied to the pockels cell   o When activated  denoted by a red icon   laser is being applied to sample     o The Soft Shutter will be activated automatically when acquisition begins   This confocal can only be used in Galvo mode  Scan mode will be inactive    Objective Lens is important to set   o Select the objective lens being used  The objective can be selected in the Nikon controls menu  or on the microscope and the Objective Lens box in Prairie view should update automatically   o All lenses have been calibrated and should be listed in the drop down menu     Stage Control is located under the Scanning panel     Before beginning to move
8. computer cart are on    b  To switch between fine  medium and coarse x y movements rotate the top of the joystick    c  While imaging  moving the joystick up and down will move the sample left and right  but it will  be correct if looking through the eyepiece or if using the stage controls in the software to move  the sample     Available for confocal imaging   Imaging dichroics  488 568 647  488XR  405RDC  405 488 561 640   488  491 633  647   405 5 14 633  Detectors  Three channel PMT  Channel 1  690 60  605 70  Channel 2  605 70  535 30  525 50  Channel 3  525 50  460 50  Channel 1 2 3 dichroics  650 Ip  565 Ip  mirror  Channel 2 3 dichroics  565 Ip  495 Ip  Nikon Objectives  Lenses    10X  N A   0 30  Plan Fluor  W D  16  20X  N A   0 75  PlanApo VC  W D  1 0  40X  N A   1 0  PlanApo  W D  0 16 Oil    60X  N A   1 40  PlanApo VC  W D  0 13  Oil  100X  N A   1 40  PlanApo VC  W D  0 13  Oil    14  Prior to imaging your sample  please check the laser power and PMT gain with the pollen grain slide   Procedure   1  Use pollen grain slide  typically located on the air table  with the 20x objective  2  Choose an area with several pollen grains to image  3  Settings   a  561 power   100  b  Use Ch1 window with the 605 70 filter  set channel color to range check  c  4us dwell time  d  512 x 512 resolution  4  Live scan and increase PMT power until there are a few red pixels  has been around 350 400   5  Note this number on the log sheet    Setting image gains and offset    Gain i
9. d the number on the right is the selected  dwell time    Lower image size   higher dwell time   A starting dwell time of 4 us is suggested     Optical Zoom  mag  is located under Dwell Time     Optical Zoom allows the user to focus in on a smaller field of view    With higher optical zoom  laser is concentrated on a smaller field providing more power and greater  potential for photobleaching   Check laser power and PMTs to avoid this     A limit will be reached at which no more resolution is possible    As the Optical Zoom is raised  the red box in the Pan Control panel will get smaller     Pan Control is located under Optical Zoom     The red box in the Pan Control panel denotes the area of the field of view that is being imaged    With the Pan Control  users can move the laser to specific areas in the total field of view without moving  the stage    To return to the center of field  push 0    The button below the red box indicates pan speed  fine  medium or coarse      Scan Rotation is located under Pan Control     Scan Rotation controls the angle of laser scanning    The top of the scanning box is labeled as the yellow side of the red box in the Pan Control panel   Rotation is possible from 180 to  180 degrees    Clicking Zero will return the Scan Rotation back to 0 degrees     Scanning panel is located on the top right hand corner of the main control window     Scanning controls image acquisition   Live scan will continually scan the object and update the image   Single s
10. ences menu     The Mark Uncage dialog is used to mark locations of points for photo activation or uncaging experiments     These points can be single or in a user defined line or grid     Note  Although the Mark Uncage function can be performed in a PrairieView window  TriggerSync  must be open and running  from the Applications menu  in order to use this function     When Mark is selected  a box appears on the image in the active Image Window and a second  Mark  Uncage  window appears  as shown below   To select points  click on the appropriate option in the Mark  window and follow the steps described below  Because this window is an interface between PrairieView  and TriggerSync  it is necessary to set the Mark parameters using a Mark Points Wizard  which is  accessed by clicking Configure     To Mark a Point     1  Click Point    2  Move point to desired location on image    3  Click Configure    to set Mark Points parameters   4  Click Prepare    5  Click Acquire to mark points     To Mark a Line of Points   1  Click Line     2  Enter number of points in the X Point Density field   3  Move line to desired location on image    4  Click Configure    to set Mark Points parameters    5  Click Prepare     6  Click Acquire to mark points     To Mark a Grid of Points   1  Click Grid   2  Enter number of points in X and Y Point Density fields   3  Move grid to desired location on image by dragging center point   4  Resize and rotate grid by dragging corners   5  Click Configure  
11. nel along a  line that you create on the image     The Photoactivation  PA  tool enhances the functionality of a Z series acquisition or single scans by    allowing the user to apply masks to scan areas  controlling where laser power is applied  These masks  can be on specific z slices  applied to an entire stack or a single scan     To use the PA tool for a stack of images  a Z series must first be acquired  see Z Series tab  While still  in the playback window after acquiring a Z series  select PA    This will open a toolbar on the right side of the Image Window  By using the scroll buttons at the  bottom of the window or cursor keys to change frames and the buttons on the right to edit the masks  the  mask region s  are defined  These are the regions that are scanned by the laser beam  Regions that are  not masked are not scanned  The selected regions are shown as translucent green areas on the image as  shown below  When all masks are defined  disable the PA mask editing mode  Click PA      To use the PA tool for a single image click PA at any time except when in playback mode for a Z series   Define a single slice PA mask as you would a set of masks for a Z series  To use the mask it must be  selected on the Mise tab  To acquire a Z series using these masks  it will be necessary to set up a T   series  being sure to select a set of masks in the PA column  It is important to make sure Save images  generated when Photo Activation  PA  is used  is also checked in the Prefer
12. p  The  focus can be offset by using the perfect focus adjustment knob  on the air table to the right of the microscope      Set up for Live Cell Imaging    The live cell imaging chamber is stored in a box on the bottom shelf of the wall cabinet  The microscope plate  for the chamber is also in the box under the chamber     Use caution when placing the chamber on the microscope stage  The condenser must be raised up and the top of  the microscope tilted back while putting in the chamber  Note  The top of the microscope is heavy due to the  additional PMT so the top will not stay up by itself  Tilt back and place a fitting  located on the microscope  table  in the junction to be sure that the top will not fall     The controller  tubing and bottle for humidity control are located on the table  behind the microscope  Fill the  humidity control bottle above the minimum fill line with DI or filtered water  The top of the bottle can then be  placed in the bottle and the connection made with the chamber     CO2     The tank should be set for a pressure between 15     20 psi  Please do not change this setting  To open  the tank  only open the main valve     Interlaced Scanning    Interlaced scan patterns allow acquisition of frames using up to four different independent laser settings  alternating each line for higher temporal resolution than is possible with using labels to alternate settings each  frame  One line of each track is scanned to acquire one line of the final image     
13. rm each iteration at all XY stage locations before proceeding to the next iteration       1  XY Stage Location    2  T Series Iterations    3  Cycles    TriggerSync Execution  TriggerSync experiments can be executed two different ways  either the  software will wait for the entire experiment to finish before the next cycle starts  or the next cycle can be  started directly after the TriggerSync experiment is started     Start next cycle immediately    The XY Stage tab allows you to set positions in your sample at which to collect data     To add a new position to the XY positions  move the stage to the desired position and click    Add  Current Location      The Plate Definition panel can be used to set up the x  y locations in well plates by entering the distance  between wells and the number of x and y position in the well  This will automatically define all x     y  positions  Currently  x and y need to be switched from the conventional directions or the positions will  be marked incorrectly  90 degrees from expected   Prairie is aware of this problem and is working on a  fix   The Grid Definition panel can be used to automatically set XY positions and aquire a full image  between two points   o The start and stop locations can be set in this panel   o Choose the amount of overlap you wish to see between pictures by entering a value in the     Overlap      box   o The calculated positions can then be added to the grid locations already chosen or used to replace  those grid
14. s adjusted by turning on the PMT voltage  moving the slider bar to the right   Start with the desired PMTs  set to 500 and adjust laser power and PMT voltage to obtain optimal image     Clicking the green tab on the left of the PMT window will allow access to the gain multiplier and the offset   Gain multiplier should only be used if more gain is needed than can be achieved with PMT voltage setting   keep this set to zero   Note     this will amplify both the signal and the noise     Offset can be adjusted in range check mode  right click on the color bar on the right of the channel button to go  to range check   Turn off Laser and PMT powers and adjust the offset to get a minimal number of blue pixels   You should only need to do this once in an imaging session and it should not vary much from the starting  settings     To obtain transmitted light image    To set up for transmitted light     open shutter on 4   PMT and move mirror to direct light to PMT  The lever on  the 4  PMT will be down when the shutter is open  When the mirror is in position red will be observed through  the eyepiece  The transmitted light image will be on channel 4  Note  If using interlace the 4  PMT also needs  to be added in the interlace setup  using any laser      Perfect focus    This microscope has perfect focus  which can be used to keep focus when moving in x     y across the sample  and to adjust for z drift  When the perfect focus is turned on the focus is set to the surface of the coversli
    