Gasman User Manual
Contents
1. D ig Peak display Tad OK qa VIE Zero To perform Autozero select the Zero function from the menu W hen Zero is completed the instrument will return to normal Run mode 3 Run mode The Gasman is now ready to use Familiarise yourself with the gas being monitored in your unit and make sure you understand site health and safety procedures in the event of alarm conditions 12 b CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 12 28 09 2010 14 41 54 Gasman 2 5 Logging Gasman incorporates both data and event logging which can be accessed using the RS232 communications link available with the Single Way Charger Interface part No C01940 using Crowcon Portables PC software See section VIII Data is recorded every minute this rate can be adjusted using the PC software The log can record 900 hours of data at 1 minute intervals Gasman also records the time and date for a number of operating and diagnostic events including e Switch on and switch off e Level 1 Level 2 and TWA alarm on alarm off and the peak level during the alarm e Zero calibration and gas test with success or failure Flammable sensor saver on and off The battery condition is logged every switch on and switch off while the instrument is operating and certain configuration changes are also logged The
2. Gasman Personal Single Gas Monitor WICROWCON sacs CROWCON baga ti Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 1 28 09 2010 14 41 45 Instructions specific for use in hazardous areas The following instructions apply to equipment covered by certificate number Baseefa04ATEX0383 Flammable Gas Baseefa04ATEX0384 Oxygen or Toxic Gas The following information covers all relevant points listed in clause 1 0 6 of the EHSR s of the ATEX directive The certification marking is as follows CEng Baseefa04ATEX0384 Ex ia IIC T4 Ga IECEx BAS 05 0039 4F 20 Cs Ta lt 149F 65 C IR BR BR BR RG BR BR RE GN RIG BER RB RG KR CORO D RARR 6 KIS SR 2 ROR KRG KRDO SRLS RIISIIN PIR Baseefa04ATEX0383 Ex ia d IC T4 Gb IECEx BAS 05 0038 4F 20 C lt Ta lt 149F 65 C FOR INTRINSIC SAFETY PRECAUTIONS SEE MANUAL WARNING UNDERSTAND MANUAL BEFORE OPERATING ONLY CHARGE BATTERY IN A NON HAZARDOUS AREA Manufactured by Crowcon Oxon UK l KX 1 The equipment may be used in Zones 1 and 2 for flammable versions and Zones 0 1 and 2 for toxic and oxygen versions for Group IIA IIB and IIC gases and vapours for Temperature ClassesT1 T2 T3 and T4 2 The equipment is certified for use in ambient temperatures in the range 20 C to 65 C 4 to 149 F The equipment should not be used outside these ranges 3 Compliance with the Essential Health and Safety Requireme
3. ATEX Essential Health and Safety Requirement clause 15 9 Safety certificate no BASEEFA04ATEX0383 Flammable Gas BASEEFA04ATEX0384 Oxygen or Toxic Gas IECEx IECEXBAS05 0038 Flammable Gas IECExBAS05 0039 Oxygen or Toxic Gas Approval codes Europe ATEX Il 1G Ex ia IIC T4 Ga 20 C Ta 65 C Toxic Oxygen ATEX Il 2G Ex ia d IICT4 Gb 20 C Ta 65 C Flammable USA Class Division 1 GroupsA B C and D Canada Class Division 1 GroupsA B C and D 30 b CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 30 28 09 2010 14 42 05 Specification Gasman Standards Safety EN60079 0 EN60079 1 EN60076 11 94 9 EC Canada CSA22 2 152 Operation EN50270 EN50271 IEC61508 EN61779 b CROWCON 31 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 31 28 09 2010 14 42 05 Gasman Accessories and spare parts XI Accessories and spare parts Accessory list Crowcon part number Description Single way chargers C01941 12 V DC input single way charger C01942 Single way charger with 230 V UK style power supply C01943 Single way charger with 230 V EUR style power supply C01944 Single way charger with 110 V US style power supply C01945 Single way charger with 90 260 V in line power supply C01296 Vehicle lighter socket lead C011012 Single way charger interface 230V in line power supply no plug C011010 Single way charger interface 110V in line power supply no plug C01940 S
4. Gasman requires little setting up Follow these simple steps to get your unit ready for use 1 Ensure the unit is in clean air 2 Switch on Press and hold the operator button for approximately 3 seconds until the red LEDs flash The operator display screen will light up and the unit will begin a warm up sequence The user should now check the detector goes through the warm up sequence detailed below In the case of any deviations the Gasman should be returned to your local Crowcon office or Service agent gt CROWCON 3 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 3 28 09 2010 14 41 49 Quickstart guide Gasman warm up sequence a The Gasman will demonstrate Sounder the alarm LEDs sounder Alarm test vibration alerts and the BM operator display screen The ba sen sounder may be silenced by pressing the button Gasman KUNG LEDs Alarm test The Gasman screen will show the following display during the warm up sequence This will take approximately 20 seconds fa A A 7 af v M M N E l IOM hye rr ijas AZ AN NINA eV L LUC gt mm ig fo Of a a ER CT Note these screens may differ depending on what sensor is being used b Autozero If Autozero is enabled default the unit will display the Autozero confirm screen the displa
5. M07630 Gasman UK Iss 10 CS5 indd 24 Gasman Maintenance and Calibration Gasman IR CO2 Personal Detector The concentration of CO in fresh air is 0 04 by volume A correctly zeroed Gasman IR CO monitor will read 0 04 in uncontaminated ambient air Before zeroing the Gasman must be in fresh uncontaminated air well away from a building or any CO emissions Hold the unit well away from the breathing zone of the operator i e at arm s length When Gasman IR CO is zeroed in fresh air it automatically sets the CO baseline level to 0 04 Gasman IR CO can also be set to zero in nitrogen using the PC interface and Portables PC software This is in effect a zero point calibration rather than a normal zero To do this after uploading the instrument configuration file note the existing calibration gas level then set the calibration value to 0 00 apply nitrogen gas and click the Calibrate button After this zero point calibration is completed remember to re set the calibration gas level to its previous level Nitrogen for zeroing should be provided from a suitable gas bottle at a regulated flow of 0 5 I min 1 SCFH connected to the standard flow cap clipped over the sensor aperture Recommended calibration gas is 2 CO in a background of nitrogen Care should be taken to vent the nitrogen flow away from the operator and not allow nitrogen flow for longer than necessary to complete the zeroing If indoors it is recommended t
6. 10 CS5 indd 25 28 09 2010 14 42 04 PC interface and software Gasman Vill PC interface and software Gasman can be connected to a PC using the single way charger unit with optional PC interface The charger unit is fitted with a D type 9 pin RS232 socket which is located at the rear of the charger see diagram below The PC requires Crowcon Portables PC software A USB RS232 adaptor is also available from Crowcon RS 232 link The software provides the user with access to reconfigure alarm levels operation run calibrations print reports and to access data and event log files Set up 1 Install Portables PC software on PC and attach the RS 232 cable to the charger and PC 2 Switch on the Gasman unit and drop it into the charger unit with the display facing forward 3 For information on using the Crowcon Portables PC software see installed help file 26 b CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 26 28 09 2010 14 42 04 Gasman PC interface and software Data logging Data is logged at a rate set by the log period which is configurable via Portables PC software This is set at 1 minute intervals as a default The Gasman is capable of storing 54 000 logs 54 800 events When the memory is filled new data then overwrites the oldest data A log can be extracted from the Gasman and saved using Portables PC software The instrument will record the following
7. 6 Ensure the gasket is in place on the sensor Then slip the sensor onto the sensor housing Click the quick release fixings around the i module board ensuring the i module is held in place firmly and the sensor is still tightly located on the module board 7 Replace the back of the Gasman case and secure with screws 8 Switch on the Gasman unit The new i module will be automatically recognised 9 Crowcon advise carrying out a calibration check when any new sensor is fitted b CROWCON 29 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 29 28 09 2010 14 42 05 ALL ee e Gasman Specification X Specification Dimensions 90 x 48 x24 mm 31 x 1 9 x 1 inches Weight 138 g flammable 129 g oxygen 118 g toxic Housing degree of protection Ingress protection IP65 NEMA 4 Operating temperature 20 C to 55 C 4 F to 131 F Humidity 20 99 RH non condensing for continuous operation Display Custom LCD with backlight Starburst characters for number and text display plus screen icons for status and mode Warm up time 90 seconds maximum Response time typical T90 approx 20 seconds for most toxic and flammable sensors 10 seconds for oxygen Audible Alarms 95 dBA Multiple alarm sounds allow selection of distinctive tones for different alarms Visible Alarms Dual colour red blue flashing LEDs in gas hazard Vibrating Alarm Internal vibrating alarm Repeatability 2 FSD 6 months
8. Event log can record over 4800 events b CROWCON 13 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 13 28 09 2010 14 41 54 M07630 Gasman UK Iss 10 CS5 indd 14 Gasman Batteries lll Batteries 3 1 Rechargeable batteries Recharge time for the Li ion batteries is less then 6 hours less if they are not fully discharged Rechargeable batteries will typically last 12 hours for flammable units 3 2 Gasman Charging Unit There are 3 models of charging units available for Gasman a single way drop in charger unit a single way charger unit with integral PC Interface and a multiway charger unit The multiway charger unit can accommodate up to 5 Gasman units Note the multiway charger unit does not have a PC interface option The charger units are powered by a nominal 12 V dc input Power Supplies are available with UK EUR or US pins alternately a universal 90 260 V power supply is available for any other configuration The multiway charger includes a universal power supply A vehicle lighter socket lead can also be supplied please see section XI Parts and Accessories Communications and charging are only permitted via Crowcon charger charger interface C01940 Charging the batteries 1 Ensure you are ina Power safe area supply 2 Plug the charger power supply into a mains socket and connect to the charger lead 3 To charge your Gasman simply drop the Gasman into the charger in an upright p
9. IEC 61508 to ensure quality and integrity of operation Gasman has been designed to give a truly reliable personal gas monitoring system The internal circuitry includes an independent watchdog The software monitors for any malfunction within the unit and will display an error warning to the user should they occur 8 CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 8 28 09 2010 14 41 51 ol ee o Gasman Operation Il Operation 2 1 Switch on sequence 1 Ensure the unit is in clean air NB For CO2 detectors see zeroing Section VII 2 Switch on Press and hold the operator button for approximately 3 seconds until the red LEDs flash The instrument begins with testing all the LCD segments on the operator display screen the red and blue alarm LEDs sounder and internal vibrator alert The sounder may be silenced by pressing the button The unit enters a warm up mode and displays a sequence of screens see page 3 for more details At the end of warm up the Autozero menu will be displayed The Autozero function can be disabled Battery check confirmation Autozero will not appear sufficient charge in the battery See section VIII PC Interface and software pack Calibration check During the warm up sequence if the next calibration date is due in less than 31 days CAL nn will be displayed where nn is the number of days to the next calibration If calibration i
10. If the gas level is still within alarm limits the button will have no effect The Gasman alarm is set to latch by default The unit will still continue in alarm mode even when the gas level returns to normal until the button is pressed and the alarm is cleared gt CROWCON 5 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 5 28 09 2010 14 41 51 Quickstart guide Gasman 3 Switch off unit and storage 1 Press and hold the button for 5 seconds until the display shows OFF The display will count down to switch off Storing conditions In order to optimise sensor performance and lifetime your Gasman unit should be stored in a safe non hazardous area at 0 20 C and 20 90 RH 4 Additional information For battery recharging information go to section Ill For fixing accessories go to section V For sampling section go to section VI OD For calibration information go to section VII For troubleshooting guide go to section XIl m gt CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 6 28 09 2010 14 41 51 Gasman Introduction I Introduction Thank you for purchasing the new Gasman Personal M onitor Gasman is a portable Single Gas detector designed to be carried or worn by individuals working in hazardous environments such as confined spaces It is suitable for use in classified hazardous areas Gasman monitors a single gas and displays its reading on a display scree
11. events For more information on using Crowcon Portables PC software see the Switch on off Battery status Sensor channel over range Alarms activated cleared including peak values of response Gas test pass fail User triggered event installed help file b CROWCON 21 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 27 28 09 2010 14 42 04 Gasman i module replacement IX i module replacement Installing or replacing an i module 1 Ensure you are in a non hazardous safe area with suitable ESD protection Switch off the unit 2 Remove the back cover by unscrewing the four M 2 5 12 mm Torx T6 screws as shown the in the drawing point Do not touch the charging elastomer connector with your fingers 3 Place the Gasman unit face down on a surface 4 Unclip the i module from its retaining support clip Ease one side out at a time Ensure the elastomer which is retained within the body of the support clip remains in place and should not be touched by hand 5 Unwrap the replacement i module from any packaging and ensure the sensor is fully seated on the module board If replacing an i module with one of the same type instrument specific configuration will be retained If replacing with a different i module its default configuration will be loaded 28 b CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 28 28 09 2010 14 42 05 i module replacement Gasman
12. has been designed from top to bottom to bring you a lighter compact design with single button operation for ease of use maintenance and extreme reliability gt CROWCON 7 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 7 28 09 2010 14 41 51 Introduction Gasman i module gas sensor Gasman uses unique plug and play i module sensor technology Each sensor unit incorporates its own intelligent processor holding sensor configuration and calibration data Different sensors can be purchased and once inserted are immediately ready to run Flammable units are only available with rechargeable batteries Plug and play will ease maintenance time and cost and the intelligent modular system will remove the need to calibrate the sensor Gasman can be reconfigured by purchasing additional pre calibrated i modules from Crowcon or your local office After fitting a new i module it is recommended that a gas test is carried out on the detector before use Reliable anti shock mechanics and robust housing The Gasman housing is built from resilient material giving it strength and flexibility to withstand the hardest of working conditions water and dust tight to IP65 with a non slip grip If the unit is dropped there will be no disruption of power or function ensuring reliability and service for years to come Software The internal software in Gasman has been designed and written in accordance with the requirement of
13. minimum for sampling before readings should be taken CROWCON 19 Gas Detection You Can Trust 28 09 2010 14 42 02 Gasman Gasman gas test accessory kit Gasman gas test accessory kit The Gas Test Accessory is a gas testing kit designed to enable gas testing and one button calibration of your Gasman gas detector using a specially formulated high stability long life single gas mix It can be used with Gasman units having sensors for Flammable Oxygen Carbon Monoxide and Hydrogen Sulphide gases 6 1 Gas testing Gas testing checks the sensor is responding within set limits to an applied gas of known composition This can be performed as often as desired but would usually be carried out each time the Gasman is issued for use The Gasman itself will determine Pass Fail status for the gas test In order to perform successful gas tests you should ensure The gas used has the correct gas concentration and that it is within the validity date specified by the supplier The gas flow path is leak tight It is important to check that the flow cap is properly fitted to the Gasman unit and the outlet tubing is not restricted in any way nor additional tubing length used The GasTest Accessory Kit comprises a gas cylinder containing the gas a Trigger regulator with interconnect tubing a magnet used to activate Test mode a flow cap to attach to the Gasman and a vent line The kit is supplied in a convenient carry
14. selected the instrument shows the highest value for flammable and toxic gases or the lowest value for oxygen since switch on or last Peak Reset This is useful for vertical entry checks where the Gasman can be lowered down the shaft rather than using a sampling tube and for viewing peak exposure at the end of a shift TWA display Shows the 15 minute and 8 hourTWA for toxic gases monitored since last turn on Peak Reset Before a Peak entry test is performed select this menu option to clear any previously stored Peak value Zero Performs Zero on your Gasman unit Please ensure you are in clean air How to display the menu 1 To view the additional display option menu double click the button The menu icons will appear on the screen as shown below s SMG OK Note Only toxic gas monitoring units will display the TWA menu option 2 Press the operator button with a single click to scroll across the list When the underscore appears under your choice double click the operator button This will select the option If Peak orTWA is selected Gasman will display the icon on the operator screen b CROWCON 11 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 11 28 09 2010 14 41 54 Operation Gasman Peak test When performing a Peak test such as a vertical entry check previous readings can be cleared by selecting the Peak Reset menu option 1 Scroll menus 2 Select
15. Calibration gas contact Crowcon required gases depend on sensor combination Carrying and wearing C01952 Harness Plate C01843 Shoulder Strap C01844 Chest Harness Straps C01953 Hard Hat Clip M 02362 Alligator Clip Communications E07532 PC Interface Lead C01832 PC Software C02097 USB to RS232 adaptor Spares consumables 5011818 1 Rear M oulding M 04973 Gasman front and rear moulding set C03329 M etal Clip E01918 Non rechargeable Battery E07621 i module Elastomer E07620 LCD Elastomer M 04682 Sensor Sealing Ring M03705 Clip retaining nut M3 M 03793 Case screw M 05910 Crowcon domed label E015 35 Power supply for charger 230 V for UK only E01536 Power supply for charger 230 V for EUR only E01537 Power supply for charger 110 V for USA only E01552 Power Supply for charger 230 V No plug E01553 Power Supply for charger 115 V No plug E07693 Power supply for charger Universal 90 260 V b CROWCON 33 Gas Detection You Can Trust 28 09 2010 14 42 06 Gasman Troubleshooting guide XII Troubleshooting guide Symptom Cause Action error message Instrument won t switch on Flat battery Recharge or replace battery No confidence beep Function disabled Reconfigure with PC software Gas reading when no gas Zero drifted Restart instrument in clean present air Unstable inaccurate gas Sensor failure Do not use exit hazardous reading area immediately Return instrument for recalibration or sensor repla
16. Trust 28 09 2010 14 42 02 M07630 Gasman UK Iss 10 CS5 indd 19 Flow sampling Gasman VI Flow sampling Attaching the flow cap To perform manual sampling using Gasman a flow cap must be fitted onto the front of the unit over the sensor A flow cap is included with your Gasman unit OS CY SA 1 To fit the flow cap firstly ensure the cap gasket is intact slip the flow cap over the sensor on the front of your Gasman unit until the cap is tightly fitted into place 2 Attach the sampling tube or flow accessory onto a gas nozzle 3 Attach the aspirator bulb onto the remaining gas nozzle 4 To remove the flow cap gently tease the cap from the sensor and lift the cap away from the instrument When using the manual aspirator kit adopt a consistent rhythm whilst using the hand aspirator Crowcon recommend squeezing once per second to achieve a flow rate of approximately 0 5 1 litre min At least 10 pumps per sample are recommended The sampling tube supplied is normally a 2m 6ft length Longer lengths of sampling tube can be provided but will increase the time taken to get a sample from the point of sampling to the Gasman instrument When using an extended length of tubing a response time test is recommended Gas of known concentration should be sampled along the full length of tubing to be used and the time taken for the sensor reading to reach the known gas levels should be noted This time should be used as the
17. alarms will clear automatically when the gas hazard has passed but not until then Alarm mute The sounder can be set to mute for level 1 alarm only pressing the operator button during an alarm condition i e presence of hazardous gas will silence the sounder and stop the vibration alarm The alarm LEDs will continue to flash Alarm sounder tone Different distinctive tones can be selected for the each alarm condition to achieve the best performance for the monitoring conditions available In the event of a Time Weighted Alarm TWA TI I In the event of the 15 minute or the 8 hour TWA being L W triggered Gasman will go into alarm and display a TWA TWH warning with the toxic gas readings Neither the 15 minute nor the 8 hourTWA alarm can be cleared 16 gt CROWCON Gas Detection You Can Trust ml a 28 09 2010 14 41 56 M07630 Gasman UK Iss 10 CS5 indd 17 Alarm indications Gasman In the event of a Flammable overange alarm If flammable gas concentrations exceed 100 LEL Gasman QC locks into alarm and displays 9999 showing an overange condition Gasman will temporarily cut off power to the sensor to prevent burn out and display a progress bar for 200 seconds When the timeout is complete either press the button to continue or switch your unit off and on to restart This option is programmable with Crowcon Portables PC software and the default is switched on If the Gasman goes
18. arise yourself with the gas being monitored in your unit Ensure you understand site health and safety procedures For information on Peak and TWA readings go to section 2 4 Confidence signals Gasman Screen Icons x Warm up OK Flashing icon Gasman running normally AL AL 2 p Battery life ip ja E Autozero Alarms To reassure users the unit is working correctly the Gasman unit will emit a short confidence beep accompanied by a blue LED flash every 10 seconds and the OK icon will flash continuously 2 3 Display symbol guide Battery Fulll A full battery is represented by a battery icon showing full three bars A low battery charge will show 1 bar When zero i J bars are shown the battery icon flashes and the sounder will emit warning beeps If the battery becomes too low for operation the Gasman will switch off A low battery warning will alarm prior to switch off TWA alarm Gasman will display the TWA alarm for toxic gases when the 15 minute or 8 hour Time Weighted Average alarm threshold is passed Gasman will display LTWA and STWA When aTWA alarm threshold is reached the TWA alarm cannot be cleared 10 M07630 Gasman UK Iss 10 CS5 indd 10 gt CROWCON Gas Detection You Can Trust 28 09 2010 14 41 52 Gasman Operation 2 4 Display options Gasman provides four additional selectable displays Peak display When Peak mode is
19. case The Trigger regulator can be operated in two ways 1 squeeze and hold allows gas flow as long as the lever is held in or 2 by lifting the lever the flow is locked on 6 2 How to perform a gas test 1 Ensure your Gasman unit is switched on and in normal operation 2 Fit the flow cap onto the front of the sensor and attach the hose from the Trigger regulator Attach the outlet hose to vent gas away do not extend this hose and do not restrict or allow kinks 3 Touch the magnet horizontally aligned onto the case to the left hand side of the display as shown left Your Gasman unit will activate the GasTest and show TEST on the display 20 b CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 20 28 09 2010 14 42 03 Gasman gas test accessory kit Gasman 4 Gasman will display a progress bar Apply the gas whilst the progress bar is counting down The display will alternate between the screens shown right Gasman will then display PASS or FAIL In the event Gasman displays FAIL please see the troubleshooting guide in the first instance or contact Crowcon or your local service centre 5 To abort the gas test press the button at any time whilst the test is in progress 6 3 How to perform a one button calibration test To perform a one button calibration test you must first perform a Zero on your Gasman unit 1 Ensure you are in clean air Double cli
20. cement Autozero failed Zeroing in Switch off and restart in contaminated clean air atmosphere Cannot Autozero due to Zeroing in Switch off and restart in alarm contaminated clean air atmosphere Calibration expired The calibration Send for calibration due date has passed Display shows empty Battery depleted Charge or change as battery symbol on appropriate switch on 34 b CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 34 28 09 2010 14 42 06 ALL Tl e Appendix Limitations of sensors Gasman Appendix Limitations of sensors Sensor limitations The sensors used in Gasman have limitations common to all such gas sensors and users should be aware of the points listed below Crowcon can advise on particular situations and suggest alternative sensors if the instrument is likely to experience extreme conditions Gasman Flammable uses a catalytic gas sensor which measures the flammability of the gas For this reason readings displayed on the unit will be unreliable over concentrations of approximately 120 LEL Oxygen is necessary for catalytic sensors to operate A pellistor saver is used to disconnect power to the pellistor sensor in the event of over range to prevent burn out This locks out the unit for 200 seconds after which a button press will reconnect power to the pellistor If the sensor power is reconnected when the unit is exposed to an over range gas concentration there is a ri
21. ck the button and select Zero from the Options menu Gasman will perform a Zero To perform a one button Calibration complete the next steps within 15 minutes of completing the Zero 2 Follow steps 1 to 3 given in 6 2 Gasman will display CAL and alternately Press the button to confirm one button Calibration If the button confirmation for calibration is not made within 10 seconds then the process will revert to test as in 6 2 3 Apply calibration gas following step 4 in 6 2 4 To abort the Calibration test press the button at any time whilst the test is in progress Your Gasman unit will adjust the value for the gas channel to match the stored calibration gas value within the sensor i module This will also change the calibration due date which is set as a default of 182 days If your Gasman unit does not calibrate successfully Gasman will display FAIL Your Gasman unit must be sent to Crowcon or a local service centre for re calibration Any Gas Test pass and fail and calibration pass and fail and their values are stored in the Event log b CROWCON 21 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 21 28 09 2010 14 42 03 Gasman Gasman gas test accessory kit 6 4 Gas test calibration troubleshooting Symptom No response to gas Gasman fails gas test Gasman fails calibration Gasman passes gas test but will not enter calibratio
22. conditions However some small items of routine maintenance are recommended General To keep the display panel and operator button free from dirt build up regularly wipe over your Gasman unit with a damp cloth Filter Inspect the front filter at regular intervals for dirt or damage Clean as necessary Zero and calibration Gasman is supplied with an Autozero function on start up This function can be configured to operate automatically on user confirmation see quick start guide or can be disabled This configuration can be set with the Crowcon Portables PC software see section VIII Gasman also has a Zero function in the menu see section 2 4 Crowcon recommends a monthly gas test to confirm sensor operation However please check with your own Health and Safety regulations A test gas of known composition needs to be applied to verify sensor response and alarm function see section 6 1 Instrument calibration of sensor should be performed at 6 month regular intervals Calibration method Gasman calibration can be performed using the Gas Test Kit as per 6 3 or with Crowcon Portables PC software and calibration gas mixtures Appropriate certified calibration gases should be used Calibration gas is applied using the appropriate flow cap Refer to Crowcon Portables PC software help file for further information b CROWCON 23 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 23 28 09 2010 14 42 03
23. dous area Remove the back cover and remove the battery Insert the replacement battery into the instrument and securely fasten the back cover Note When a non rechargeable detector is placed in the charger charger interface the red LED will not illuminate and the detector will not charge Note later versions of the Gasman will have a tab in place to aid removal of the battery b CROWCON 15 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 15 28 09 2010 14 41 55 M07630 Gasman UK Iss 10 CS5 indd 16 Gasman Alarm indications IV Alarm indications Gasman provides two instantaneous alarm levels designated level 1 and level 2 For toxic gas sensors there are also two TWA alarms one for short term exposure STEL based on a 15 minute time weighted average and the second TWA alarm is for long term exposure LTEL based on a 8 hour time weighted average Alarm configurations are set via the Crowcon Portables PC software The following settings can be made Alarm thresholds for each sensor Level 1 and level 2 HL instantaneous alarms H J Alarm trigger This can be set to rising levels of LL gas concentration or as falling Oxygen units are set to falling for deficiency monitoring Alarm latching Alarms can be set to be latched or unlatched Latched alarms will require the button to be pressed in order to clear the alarm once the hazard has passed This is the default setting Unlatched
24. es vapours or liquids present under normal operating conditions An area classified as Zone 2 is not likely to have ignitable concentra tions of flammable gases vapours or liquids present under normal operating conditions Zone 2 M07630 Gasman UK Iss 10 CS5 indd 2 28 09 2010 14 41 47 Crowcon Detection Instruments Ltd 2 Blacklands Way Abingdon 0X14 1DY UK Tel 44 0 1235 557700 Fax 44 0 1235 557718 Www crowcon com Email sales crowcon com Copyright Crowcon Detection Instruments Ltd 2010 All rights are reserved No part of the docu ment may be photocopied reproduced or translated to another language without the prior written consent of Crowcon Detection Instruments Ltd Publication number M07630 Tenth edition October 2010 Safety information Read and understand all instructions in the operation section of this manual before use Do not substitute components as this may impair intrinsic safety and invalidate war ranty Observe all warnings and instructions marked on the unit and within this manual Observe site health and safety procedures for gases being monitored and evacuation procedures Understand the screen display and alarm warnings prior to use If this product is not working properly read the troubleshooting guide or contact your local Crowcon Office or agent Ensure qualified service personnel change sensors and operating system Ensure maintenance serv
25. ice and calibra tion are carried out in accordance with the procedures in the manual and by trained personnel M07630 Gasman UK Iss 10 CS5 indd 3 28 09 2010 14 41 47 M07630 Gasman UK Iss 10 CS5 indd 4 28 09 2010 14 41 47 Gasman Personal Single Gas Monitor Contents UnpackiNO APA 1 Quickstart QUIdE picsa mna sued iis AKA EA AGE hayes 3 INTOQUCOON 2 mam sateen ae Gees Bee cman AA 7 Be Ciel CAR AAP AA 9 I ju AA AA AA 13 IV Alarm indications cee eee eee 15 V FIXING Accessories 0 eee 17 Vi Fow sampling srca EN GANG AE BAH ore ER 18 Vil Maintenance and Calibration 22 amp VIII PC interface and software 25 IX i module replacement 0000 c eee eeeee 27 X SOCCINCAON 58 6 ipo onto ee ew hh Sh WD ed 29 XI Accessories and spare parts 31 XII Troubleshooting guide 33 Appendix Limitations of Sensors 34 M07630 Gasman UK Iss 10 CS5 indd 5 28 09 2010 14 41 47 M07630 Gasman UK Iss 10 CS5 indd 6 28 09 2010 14 41 47 Gasman Unpacking Gasman Personal Single Gas Monitor Thank you for purchasing the new Gasman Personal Gas M onitor Gasman has redefined portable gas monitoring and will give you years of unparalleled service and reliability Please read the instructions carefully before use Keep the manual for future reference Unpacking Remove the Gasman Personal Gas M on
26. ingle Way Combined charger and PC interface C01947 Single way charger interface with 230 V UK style power supply CA C01948 Single way charger interface with 230 V EUR style power supply C01949 Single way charger interface with 110 V US style power supply C01950 Single way charger interface with 90 260 V in line power supply Multiway charger C01951 5 way Multicharger without power supply i modules 5011424 M 0 100 LEL methane S011436 M 0 100 LEL propane S011437 M 0 100 LEL pentane S011439 M 0 100 LEL butane S011440 M 0 100 LEL ethylene S011460 M 0 100 LEL hydrogen S011423 M 0 25 oxygen add USA for USA alarm config S011422 M 0 500 ppm carbon monoxide S011421 M 0 100 ppm hydrogen sulphide add USA for USA alarm config S011425 M 0 10 ppm sulphur dioxide S011429 M 0 1000 ppm hydrogen S011426 M 0 10 ppm nitrogen dioxide S011428 M 0 20 ppm chlorine S011432 M 0 1 ppm ozone S011430 M 0 25 ppm hydrogen cyanide 32 CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 32 28 09 2010 14 42 05 M07630 Gasman UK Iss 10 CS5 indd 33 Accessories and spare parts Gasman S011435 M 0 100 ppm ammonia S011438 M 0 1000 ppm ammonia S011431 M 0 5 ppm phosphine S011434 M 0 1 ppm fluorine S011433 M 0 10 ppm hydrogen fluoride S012015 0 5 Gasman IR carbon dioxide CO i module for use in safe areas only Sampling accessories M 04852 Flow cap C01937 Aspirator Assembly M 01457 Pocket Clip
27. itor from the packaging The Gasman accessories will be located in the bottom of the box Check the contents are complete you should have e Gasman unit with alligator pocket clip A configuration report detailing the sensor installed alarm settings and a calibration certificate Calibration flow cap and tubing User manual Optional battery chargers and other accessories will be packed in a separate box Battery check The Gasman Personal Gas M onitor has two battery options Li ion rechargeable or non rechargeable batteries The label on the back of the detector will have either NR non rechargeable or R rechargeable to indicate which battery option is within the detector Rechargeable units Gasman uses a Li ion battery pack and should arrive with sufficient charge so that the unit can be used straight out the box However if this is the first time you have used the Gasman unit you may need to charge the battery to attain the full operating time The actual operating time will depend on the types of sensor installed The flammable Gasman will operate for at least 12 hours on a fully charged battery b CROWCON 1 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 1 28 09 2010 14 41 48 Unpacking Gasman Warning rechargeable units Only charge the rechargeable Gasman Personal M onitor with a Crowcon Gasman charger Failure to comply could invalidate safety certification warranty and safety and
28. loem 129 3068 G Rotterdam Netherlands Tel 31 10 421 1232 Fax 31 10 421 0542 Email eu crowcon com Web site www crowcon com M07630 Gasman UK Iss 10 CS5 indd 38 USA Office Crowcon Detection Instruments Ltd 21 Kenton Lands Road Erlanger Kentucky 41018 1845 USA Tel 1 800 527 6926 or 1 800 5 CROWCON 1 859 957 1039 Fax 1 859 957 1044 Email salesusa crowcon com Web site www crowcon us Singapore Office Crowcon Detection Instruments Ltd Block 194 Pandan Loop 06 20 Pantech Industrial Complex Singapore 128383 Tel 65 6745 2936 Fax 65 6745 0467 Email sales crowcon com sg Web site www crowcon com 28 09 2010 14 42 06
29. may result in permanent damage to the unit Non rechargeable units Gasman uses a CR2 battery which can be easily field replaced Ensure only the correct types of battery are fitted to maintain compliance with certification see Battery check on page 1 Gasman IR CO Personal Single Gas Monitor The Gasman IR CO personal single gas monitor an infrared gas sensor for CO detection This version of Gasman is not designed or certified for use in a hazardous area and the safety certification data in Section X is not applicable Operation and maintenance of the Gasman IR CO is essentially the same as for other Gasman units but with the following points to note CO is present in ambient air at a background concentration of just under 400ppm 0 0496 In any enclosed environment the ambient CO concentration will increase as a result of respiration and if the room is poorly ventilated concentrations well in excess of 1000ppm 0 1 can occur CO levels in the outdoor air will also be enhanced by any emitted CO e g from vehicle exhausts or boiler flues venting to atmosphere so it is normal to see fluctuating levels of CO on the display whether indoors or outside 2 CROWCON Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 2 28 09 2010 14 41 48 Gasman Quickstart guide Quickstart guide 1 Getting started Gasman familiarisation Gas Sensor Operator Button Operator LCD Display Screen
30. n Alarm warnings are given through a combination of a loud audible alarm a bright visual alarm of blue red flashing LEDs and an internal vibrator Gasman can be fitted with a wide range of modular or plug and play gas sensors The sensor carries an intelligent processor which contains calibration and sensor information Gasman is battery operated and is available with rechargeable batteries or dry cell battery options The dry cell battery option is only available for toxics and oxygen Gasman units A battery charger for single and multiple Gasman units is available for rechargeable units see section XI for more information At Crowcon we recognised the need for a reliable and robust personal monitoring system which is both small light and easy to use Gasman has a single operator button and an intelligent user friendly display with automatic backlight The gas level is continuously monitored providing normal gas readings peak readings and time weighted averages TWA Gasman is available as a diffusion sampling instrument see section XI for sampling accessories Configuration and data event logging is handled by Crowcon Portables PC software the PC communication link is provided as a link on the charger interface unit Gasman s compact shape and design makes it comfortable to wear with a non slip grip for better handling Extra accessories such as pocket clip hard hat clip shoulder strap and chest harness can be purchased Gasman
31. n mode Possible Cause Gas cylinder empty Hose blocked or kinked Gas cylinder empty Gas cylinder out of date Hose blocked or kinked Calibration drifted Gas flow not started immediately Gas cylinder empty Gas cylinder out of date Hose blocked or kinked Calibration drifted Stabilisation time too short M enu Zero not performed Gasman not configured for field button calibration Action Check gauge replace cylinder as needed Ensure no restriction to flow Check gauge replace cylinder as needed Check date and replace as needed Ensure no restriction to flow Calibrate Gasman Repeat test starting gas immediately Check gauge replace bottle as needed Check date and replace as needed Ensure no restriction to flow Calibrate Gasman Reset using PC software Select and perform Zero from menu Send for re configuration Note Remove regulator from gas cylinder when not in use over a prolonged period This will cause the gas to leak For parts list see section XI One button Calibration will check for small drifts of stored calibration value but Crowcon recommends Gasman is sent for full certified calibration at six monthly intervals 22 M07630 Gasman UK Iss 10 CS5 indd 22 gt CROWCON Gas Detection You Can Trust 28 09 2010 14 42 03 M aintenance and Calibration Gasman Vil Maintenance and Calibration Gasman is designed to operate almost maintenance free under most
32. nts has been assured by compliance with EN60079 0 2004 Amds 1 amp 2 EN60079 1 2004 and EN60079 11 2006 as certified by Baseefa Compliance with gas detection performance standards EN50054 EN50057 EN61779 1 EN50104 and EN50270 has been certified by Lloyd s Register 4 Repair of this equipment and gas sensor replacement shall be carried out by the manufacturer or in accordance with the procedures in the manual and by trained personnel 5 If the equipment is likely to come into contact with aggressive substances then it is the responsibility of the user to take suitable precautions that prevent it from being adversely affected which may compromise the certification level 6 The rechargeable battery must only be charged in non hazardous safe areas by connection to the specified Crowcon charger 7 Only the following cell types may be fitted in the battery compartment of the non rechargeable battery pack CR2 of the following brands only Panasonic GP Energiser and M axell Do not fit a Duracell battery The cells must only be changed in a non hazardous safe area 8 The equipment is not certified for use in atmospheres containing more than 21 oxygen Area Classifications An area classified as Zone 0 is an area in which an explosive mixture of gas is continuously present or present for long periods Zone 0 Zone 1 An area classified as Zone 1 is likely to have ignitable concentrations of flammable gas
33. o vent gas out of a window or into an extracted fume cupboard Ozone sensor Due to the reactive nature of ozone O special procedures should be followed when calibrating gas detectors that incorporate an ozone sensor By following the guidelines for calibration below users will maximise the protection level that their gas detectors give them Always calibrate Crowcon ozone detectors with ozone only 24 b CROWCON Gas Detection You Can Trust 28 09 2010 14 42 04 M aintenance and Calibration Gasman Fittings For ozone calibrations all fittings and pipework should be stainless steel brass aluminium or PTFE Fittings or pipework in other plastic materials such as Tygon must not be used Gas flow rate It is important that the flow rate is set correctly as otherwise pressure effects may distort calibration values and prevent the gas detector from working correctly To calibrate Crowcon portable detectors the flow rate should be set between 0 8 1 0 litres minute 0 03 0 04 cubic feet minute Normal safety precautions for handling ozone should always be observed along with any special instructions that accompany the calibration gas cylinder or generator being used The standard flow plate should be used to supply gas at the flow rate stated above If required your local Crowcon representative or service centre will be able to provide further advice b CROWCON 25 Gas Detection You Can Trust M07630 Gasman UK Iss
34. osition with the screen facing outwards see diagram below 14 b CROWCON Gas Detection You Can Trust 12V dc 28 09 2010 14 41 55 Batteries Gasman The Gasman should normally be switched Bane off for charging During supply charging the red LED on the front of the charger EG unit will be lit When TANYA charging is complete the CD 12 V de red LED is extinguished NA During charging the display will show the battery icon sweeping from empty to full On removing Gasman from the charger unit if the Gasman is on the display icon will update in 20 seconds to show actual charge state Chargers with dual LEDs are no longer suitable for use with the current Gasman We recommend that such chargers be replaced with the newer model shown above see Section XI Accessories and spare parts p32 The charging time will be longer if the unit is switched on during charging When Gasman is fully charged and switched on the battery icon will display three full bars 3 3 Changing rechargeable batteries It is recommended that rechargeable batteries should be changed by an authorised Crowcon service centre only 3 4 Non rechargeable batteries Gasman uses a Lithium Cell battery pack which will give up to two years of operation Always switch the Gasman off before opening the case to change the battery To replace the battery ensure you are in a safe non hazar
35. over range at any time it is recommended to carry out a gas test before using it again b CROWCON 17 Gas Detection You Can Trust 28 09 2010 14 41 56 M07630 Gasman UK Iss 10 CS5 indd 18 Gasman Fixing Accessories V Fixing Accessories Clip Accessories Gasman is supplied with an alligator clip Other clip options are Hard hat clip Allows Gasman to be clipped to a hard hat a method of wearing a portable gas detector preferred by many users Pocket Clip Allows Gasman to be clipped to a pocket near the breathing zone of the user Universal harness plate Crowcon provide a universal harness plate which can be used with either a chest harness or a shoulder strap How to wear your Gasman unit Chest harness Shoulder neck Use the M3 fixing usable once the connectors clip has been removed on the back of I your Gasman unit to attach the chest harness plate Create a chest harness by attaching one strap to the top connectors to go around the neck and the other p Waist to link around the waist using the side i connectors Adjust the strap lengths until the Gasman unit is in a comfortable working position Chest harness plate M3 nut Shoulder strap With the universal harness plate in place on the belt clip attach the shoulder strap accessory onto the top connectors Adjust to a comfortable working position See Accessories section XI for full list 18 b CROWCON Gas Detection You Can
36. s already overdue the Gasman unit will display a warning message CAL The instrument can still function but it is strongly recommended the unit is sent for calibration as soon as possible Press the button to continue to Run mode The Gasman can be set using Portables PC software to lock out in the event of the calibration due date expiration in order to prevent further operation of the detector until a calibration is preformed 3 Autozero Press the operator button with a single click to confirm Autozero Flammable and toxic units will be set to read zero NB If Autozero and oxygen units set to read 20 9 If the operator button fails a warning is not pressed within 10 seconds Gasman will proceed Hise oa i i played directly to normal operation without performing a zero Switch off To switch off the unit press and hold the operator button for 5 seconds The display will count down from 5 until the unit counts down to shut off b CROWCON 9 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 9 28 09 2010 14 41 51 Operation 2 2 Run mode The Gasman unit displays the gas reading on the operator display screen A typical display is shown below I lt a OK IA ai The sensor channel will display the current value of the gas being monitored and the unit of measurement the OK symbol will flash indicating unit is working correctly Famili
37. sk of damage to the pellistor sensor Restart should be carried out in a known fresh air environment Depleted oxygen levels can reduce the flammable gas reading and if oxygen levels are below safe breathing limits it should be assumed that the flammable reading is low Electrochemical gas sensors toxic gases or oxygen contain chemicals Extreme levels of humidity can also cause the sensor to be unstable The sensors are rated for an average ambient of 20 9096 R H However they are used from the tropics to deserts to tundra Water should not be allowed to collect on the sensor as this may impede gas diffusion Persistent exposure to high levels of toxic gas will shorten the life of a toxic sensor If the high level gas is corrosive e g hydrogen sulphide damage may occur over time to metal components Sensors may be cross sensitive to other gases If unsure contact Crowcon or your local agent b CROWCON 35 Gas Detection You Can Trust M07630 Gasman UK Iss 10 CS5 indd 35 28 09 2010 14 42 06 M07630 Gasman UK Iss 10 CS5 indd 36 28 09 2010 14 42 06 M07630 Gasman UK Iss 10 CS5 indd 37 28 09 2010 14 42 06 A HALMA COMPANY UK Office Crowcon Detection Instruments Ltd 2 Blacklands Way Abingdon Business Park Abingdon Oxfordshire OX14 1DY United Kingdom Tel 444 0 1235 557700 Fax 44 0 1235 557718 Email sales crowcon com Web site www crowcon com Rotterdam Office Crowcon Detection Instruments Ltd Vlamb
38. y will alternate between ZERO and Press the operator button with a single click to confirm Zero If the operator button is not pressed within the 10 second time out Gasman will proceed directly to Run mode without performing Zero Autozero confirm M07630 Gasman UK Iss 10 CS5 indd 4 Screen Icons Warm up OK Flashing icon Gasman running normally AL 1 AL 2 p Battery life L Alarms KOO Autozero J gt CROWCON Gas Detection You Can Trust 28 09 2010 14 41 50 Gasman Quickstart guide Confidence signals In normal Run mode Gasman will emit a short beep accompanied by a blue LED flash every 10 seconds and the OK icon flashes to show operational health These confidence signals can be disabled using the PC software 2 In the event of an alarm Alarm signals In the event of gas concentrations exceeding the alarm thresholds for the gas being monitored Gasman will activate the alarm signals Alarm signals The red and blue alarm LEDs will flash the sounder will emit a loud fast series of beeps the internal vibrator alarm will activate The operator screen will display the alarm level and gas reading alternately See the figure to the left AL 1 Alarm level one AL 2 Alarm level two 1 When the gas level returns to normal press the operator button This will reset your Gasman unit to normal Run mode