1.    The patient is wearing a shoe to which the orthosis is not adjusted      gt  Failure of the joint function in Auto mode     gt  Fall    Point out to the patient to only wear a shoe to which the orthosis is adjusted     WARNING    No gait re education for the patient      gt  Fall    Make sure that the patient can use the orthosis  recommend him a physiothera   peutic gait re education and explain him the particularities of the joint     WARNING    The pressure sensor sensitivity is not adjusted correctly      gt  Failure of the joint function in Auto mode     gt  Fall    Adjust the sensor sensitivity to the patient and the shoe     WARNING    A not certified technician uses the multi purpose device      gt  Incorrect adjustments on the orthosis       gt  Fall    Only a certified technician is allowed to use the multi purpose device     WARNING    Moisture and water      gt  Failure of function     gt  Fall    Do not expose the multi purpose device to high humidity and do not immerse it  in water      gt       B    39    40    lI  lt 4 1  lt 14  lt 1  lt     ATTENTION    The activation of a remote control with the second controller has failed      Using the orthosis is no longer possible    Activate the remote control again as described in the manual  In case of  persistent failure  contact our technical support     ATTENTION    Useful life of the orthosis has shortened      gt  Failure of the joint function    Insert fully charged batteries and carry out the battery qualit
2.  e Motion  16 2 Remote Control  16 3 Multi purpose Device  16 4 Controller  16 5 System Knee Joint  17  Software Versions    55    55    56    56    56    58    58    59    59    59    59    59    61    62  62  62  63  63  63  63  64  64  64  68    37    38    1  Declaration of Conformity    We declare that our medical devices as well as our accessories for medical devices are in conformity with the  requirements of the Medical Devices Directive 93 42 EEC  Regarding to these tests  all FIOR  amp  GENTZ products  have the CE marking     2  Warranty    The warranty applies only if the product has been used under the described conditions and for the intended  applications    Because we are not able to check the compliance of handling and dealing with rechargeable batteries  we cannot  assume any liability for any damages or costs caused by batteries     3  Safety Information    This manual is addressed to orthopaedic technicians  That is why the content basically confines to features of  the product  It does not contain any notes about dangers which are obvious to orthopaedic technicians  For  information reasons and for safety of your patients  please note all information provided in this manual including  notes  tables and illustrations  Please note especially the safety instructions indicated by ATTENTION  CAUTION   WARNING  and DANGER  which are listed and explained in the following paragraph  Ignoring this information may  lead to patient injuries and property damage  In ad
3.  piece  and or shoe modifications or  adjustments  Produce the foot piece  stiffly enough and shift the rolling  off line further forward  if necessary     Insert a carbon fibre sole into the  shoe to strengthen the forefoot lever  and shift the rolling off line further  forward  if necessary     Check the placement of the pressure  sensor     Decrease activity level     Problem    Joint does not unlock  with turned on e Motion    Joint unlocks too late    Joint unlocks too late  with turned on p Motion    Joint unlocks too early  with turned on p Motion    Joint switches  unintentionally into Lock  mode    Joint does a ratching  noise during swing phase    Cause    The patient is treated bilaterally i e  he  is wearing an orthosis on his left and his  right leg with a NEURO TRONIC system  knee joint     The patient has a non harmonious gait  pattern i e  he makes one step and then  stops so that no repeating pattern can   develop     The patient s step length is too short   The motion sensors integrated in the  controller do not detect any movement     The patient does not reach the dorsal  stop due to short steps during heel off   Therefore  he cannot apply a knee  extension moment to unlock the joint     Sensor sensitivity is set too low     Pressure sensor placed in the foot piece   The pressure sensor is not loaded after  heel strike because the heel rests at   the edge of the foot piece  Only when  loading the leg  the heel is pressed into  the foot piece and applies pr
4.  switch on the device  You can now see  the three modes AUTO  FREE and LOCK on the display  fig  3      3  You can select the available modes only when the controller is con            nected  Since the multi purpose device and the controller are not    AUTO FREE LOCK             connected via cables  you have to establish a wireless connection  EED  ENA  Please read paragraph 12 2 for this  Re oe  4  Ifyou do not use the multi purpose device for more than 10 minutes  Fig  3       it switches off automatically           Note   When the batteries of the multi purpose device are almost empty  you  will see the message BATTERIES NEARLY EMPTY on the display  fig  4      FIO REG ENTZ     ORTHOP  DIETECHNIK MIT SYSTEM             Fig  4    44       9  Checking the Connection between Controller and Multi purpose Device       The second LED at the controller indicates that the controller and the multi purpose device communicate with  each other  When you select one of the three modes  the following signal appears     Keypress Light signal Meaning  Colour  blue  Signal duration  m    as Iti      The controller gets a  command of the multi   purpose device        One of the three modes is selected    10  Select a Mode    You can select the available modes AUTO  FREE and LOCK with the multi purpose device  Proceed as follows     1  For Auto mode  press the    Left arrow button  When you can see the word AUTO displayed larger in the  multi purpose device  the mode is selected and the 0 L
5.  you have successfully activated the first remote control as described in paragraph 12 4 1  search the    menu item Select other leg in the main menu by browsing with the       Left and right    arrow buttons  When you find the menu item  confirm your selection by pressing the  amp  Set button        2  Search the menu item g  Activate again by browsing with the OO Left and right arrow buttons   Now  pick up the second remote control and repeat point 3 of paragraph 12 4 1     Activating One Remote Control    ATTENTION   V The activation of a remote control with the second controller has failed          Using the orthosis is no longer possible   Activate the remote control again as described in the manual  In case of persis   tent failure  contact our technical support     1  To connect the remote control with a second controller  you must activate it a second time within 60 seconds  after having connected the first controller  see paragraph 12 4 1     2  Search the menu item Select other leg in the main menu by browsing with the       Left and  right arrow buttons  When you find the menu item  confirm your selection by pressing the Set button    3  Search the menu item Activate again by browsing with the    Left and right arrow buttons    4  Repeat point 3 from paragraph 12 4 1  Be sure to press the Auto and Lock button at the same time  If the  LED of the remote control does not blink yellow for a longer period  but instead not at all  red or one time  yellow  the buttons we
6. CEMENT on the display  fig  31   The higher you have to set the  sensitivity  the more likely the sensor is placed incorrectly e g  it is uk  placed in a space and  therefore  does not sense any pressure  However   if you want to set the sensitivity high on purpose  select OK with the    Set button and increase it by pressing the Right arrow button              57    58    12 8 2 Activity Level    You have to adapt the electronics of the orthosis to the patient s activity level  To adjust the activity level   proceed as follows              Go to the menu item Settings as described in paragraph 12 7   2  Search the setting menu    l Activity level by browsing with the       Left and right arrow buttons   3  When you have found the setting menu  confirm your selection by             pressing the Set button  2  4  Increase activity level by pressing the    Right arrow button  There   is a plus sign displayed at this point  fig  32     SL    5  Decrease activity level by pressing the Left arrow button  There   is a minus sign displayed at this point  fig  32          6  Confirm your selection by pressing the Set button  GEO GANE          These are the activity levels available        Fig  32  1  Indoor walker     d    e      The patient has the ability to make transfers and to walk with an orthosis on even surfaces with low  walking speed  Ambulation is possible for a very short distance and duration due to the physical condition of  the patient       2  Restricted outdoor walke
7. ED mode indicator at the controller blinks green     Keypress Display after pressing key Light signal    Colour  green    Signal duration     every 5 sec     ri            OPADIETECHNIK MIT SYSTEM ORTHOP  DIETECHNIK MIT SYSTEM    2  For Free mode  press the GD Set button  When you can see the word FREE displayed larger in the multi   purpose device  the mode is selected and the 0 LED mode indicator at the controller blinks yellow     Note   If the signal function is deactivated  the LED mode indicator at the controller blinks green  For information  about the signal function  please read paragraph 11     Keypress Display after pressing key Light signal       Colour  yellow       Signal duration     every 5 sec     ORTHOPADIETECHNIK MIT SYSTEM                     IT SYSTEM       3  For Lock mode  press the    Right arrow button  When you can see the word LOCK displayed larger in  the multi purpose device  the mode is selected and there is no signal at the o LED mode indicator at  the controller     Keypress Display after pressing key Light signal       No signal       ORTHOPADIETECHNIK MIT 8                    ORTHOP  DIETECHNIK MIT SYSTEM       11  Signal Function for Training Purposes    ATTENTION   The volume of the signal tone for the signal function is still switched on after the patient  has finished gait re education    V    Failure of the joint function   Set the volume of the signal tone for the signal function to O after finishing gait  re education        In order f
8. HOPADIETECHNIK MIT SYSTEM                   Meaning    Test has been interrupted e g   because the batteries have  become loose from the holder or  the battery quality is too bad to  continue the test     Further action    Please insert fully charged batteries  in the controller and select TRY  AGAIN with the Set button     Please disconnect one solenoid     Wait for 2 hours and look again at  the display     Further action    Exchange the rechargeable  batteries for new ones und make  sure that the controller orthosis is  stored so that the batteries cannot  become loose from the holder   Repeat the test     Message   Display        TEST FINISHED  THE BATTERIES  ARE DEFECT  PLEASE CHANGE    Ele  lee ee Ne  BACK    GIER CENNA    ORTHOPADIETECHNIK MIT SYSTEM                TESTING BATTERIES     BATTERY QUALITY IS SUFFICIENT    EEHEHE O JJL   CANCEL                                  GIER EEE    ORTHOPADIETECHNIK MIT SYSTEM          TEST FINISHED   BATTERY QUALITY IS SUFFICIENT    BACK    GEN EEE    ORTHOP  DIETECHNIK MIT SYSTEM             Meaning    Test is finished after 2 hours  maximum  The quality of the  rechargeable batteries is bad     Test lasts more than 2 hours     The battery quality is sufficient     Test is finished after 5 hours   The quality of the rechargeable  batteries is good     Further action    Exchange the rechargeable  batteries for new ones     You can cancel the test and use the  rechargeable batteries or wait until  the test is finished     The test h
9. R  THE SECOND CONTROLLER on the display  Select the required leg by   Left arrow button for the left leg or the Right   arrow button for the right leg  fig  12         pressing the    11  Now  repeat the points 5 to 8 to connect the second controller     Note    If you have connected two controllers  you can switch between the  left and right leg  To do so  search the menu item Select other  leg in the main menu and press the Set button of the multi   purpose device  Now  you can see the currently selected leg on the  display  fig  13      12 3 Controller Update    When updating the multi purpose device  see paragraph 14  the controller  update is downloaded simultaneously  if there is an update  At the menu  item Controller update  you can then see if there is an update        II  Q        bo    If there is an update  you can now update the chosen controller by selecting the menu item Controller    update  If there is no update  this menu item cannot be selected  Always keep all controllers in use up to date           12 4 Activate  Putting into Operation Remote Control     In the menu item Activate you put the remote control for the patient into operation i e  you connect the  controller with the remote control using the multi purpose device     12 4 1 For One Orthosis    To activate the remote control  proceed as follows     1  To go to the main menu  ress the Ga Set button for three seconds     2  Search the menu item g   Activate by browsing with the oo Left and right ar
10. Systemen mbH  Dorette von Stern Stra  e 5   D 21337 L  neburg       49 4131 24445 0   amp   49 4131 24445 57  f  www fior gentz de    amp  info fior gentz de    PB3800 SK T 02 2015    
11. as not pressed    C11  Read user manual    C12  Contact tech  support    C21  Read user manual    C91  Contact tech  support       Meaning    The cable connection from  controller to pressure sensor is  correct     Either you have not pushed the  sensor when you were asked to  or the cable connection from  controller to pressure sensor is  incorrect     Short circuit in the cable from  controller to pressure sensor     You have not pressed the sensor  when you were asked to     Controller BILATERAL is only  connected to one solenoid     Controller UNILATERAL is  connected with two solenoids     Sensor is under pressure     Device internal error    12 6 Le  Sensor Test  only with p Motion     Further action    If you do not have pressed the sensor   carry out the test again by pressing the  sensor when you are asked to    If you have pushed the sensor  check  the cable connection from controller  to pressure sensor  see manual   NEURO TRONIC system knee joint   paragraph 11      Check the cable joints at the controller   If you cannot find a solution  contact  technical support     Carry out the test again and press the  sensor when you are asked to     Check  if the cable connections to the  solenoids  see manual NEURO TRONIC  system knee joint  paragraph 11  are  correct     Contact technical support     During the test  the sensor should not  be under pressure  Make sure that   the pressure sensor is not assembled  deformed and that no outsole material  is adhered under ten
12. as shortened the useful  life of rechargeable batteries   Charge the batteries completely  before using them again     9  If you want to cancel the test  press the Ga  Set button  There is the word CANCEL displayed at this point     12 10 Step Counter    The controller counts the patient s steps with the orthosis  You can check the number of steps of the leg with  orthosis with the multi purpose device  For the total number of steps double the indicated value  Proceed as follows     1  To go to the main menu  press the  2  Search the menu item Step counter by browsing with the    the  amp  Set button     Set button for three seconds     4  You can see the message STEPS WALKED on the display  fig  34      Exit the menu by pressing the  BACK displayed at this point     Set button  There is the word          Left and right arrow buttons   3  When you find the menu item  confirm your selection by pressing       STEPS WALKED  X    BACK    Fig  34                61    2ni Device Info    The menu item il Device info informs you about the version of the multi purpose device and of the controller   To display this information  proceed as follows        1  Togo to the main menu  press the Ga  Set button for three seconds     2  Search the menu item Device info by browsing with the   MULTI PURPOSE DEVICE  V3 14    CONTROLLER  V2 18 XXXX     Left and right arrow buttons     3  When you have found the menu item  confirm your selection by BACK    pressing the  amp  Set button   4  There ar
13. ctivate the signal function by downloading  the latest update and updating the multi purpose device as well as the controller  for updates see paragraph  14 and 12 3   By transferring the current software version to the controller of your NEURO TRONIC knee joint  system  the signal function will also be activated automatically     11 2 Switching off the Signal Function    Switch off the signal tone by setting the volume to O  see paragraph 12 7 4      12  Main Menu    In the main menu of the multi purpose device you can adjust the orthosis and get information about potential  problems  To go to the main menu  press the Set button for three seconds  You see the following menu  items on the display       m Pairing    g  Activate    of Battery quality      Se Cable connection test      le  Sensor test  only with p Motion           A   settings            Step counter    l FIORBEGENTZ    CA Language ORTHOP  DIETECHNIK MIT SYSTEM      1 Device info   MULTI PURPOSE DEVICE  VXX XX    Q Controller update CONTROLLER  VXX XX  BACK                   Fig  5          You can exit a menu item in two ways  If a menu item has more options   exit via Back arrow  fig  5   If a menu item has no other option    and you want to cancel your selection  exit by selecting BACK with the    a        amp  Set button  fig  6               Fig  6    Note    R  If you have connected two controllers  the menu item 25 Select other leg i 5  additionally appears in the main menu  fig  7   Furthermore  under th
14. device refer to a possible malpositioning of the sensor  that is explained more detailed in the production technique of the NEURO TRONIC system knee joint     Result message   Display  Meaning Further action  The pressure sensor is placed    Sensor placement OK    correctly     When walking  no pressure is applied  to the sensor e g  because the shoe is    The pressure sensor is placed narrower than the orthosis shell and     C22  Read user manual    ee therefore  a hollow space results between  these two   This problem can e g  develop because  of a too narrow formed heel part at the  C23  Read user manual Sensor is under pressure  ale Oar MOEN TEN EN OT    the sensor test with unlaced open shoes or  other shoes  If you cannot find a solution   contact technical support     12 7 5 Settings of e Motion    In the menu item a   Settings you can adjust the orthosis  The following setting menus are available with    turned on e Motion              Calibrate             I     LO Turn on p Motion    mi   Volume     aM Signal selection      q Select other leg  only with two connected controllers  see paragraph 12 7 6     Tone selection    sE       To enter a setting menu  proceed as follows        1  To go to the main menu  press the D Set button for three seconds   2  Search the menu item q Settings by browsing with the       Left and right arrow buttons   3  When you have found the menu item  confirm your selection by pressing the Set button  Now you can             53    54    se
15. dition  please read the manual of the NEURO TRONIC system  knee joint as well as the patient information     4  Safety Instructions    4 1 Classification of the Safety Instructions    DANGER       Important information about a possible dangerous situation which  if not avoided   leads to death or irreversible injuries     WARNING    gt  Important information about a possible dangerous situation which  if not avoided   leads to reversible injuries that need medical treatment     CAUTION     Important information about a possible dangerous situation which  if not avoided   leads to light injuries that do not need medical treatment     ATTENTION       Important information about a possible dangerous situation which  if not avoided   leads to damage of the product     TORO    4 2 Structure of the Safety Instructions    The following example demonstrates the structure of the safety instructions     Attention symbol Signal word ATTENTION       Type of danger Mechanical load is too high   Possible consequences s   Wear of the joint parts  for components l    Possible consequences   gt  Maintenance intervals shorten   for the patient  Explain the patient correct use of  the joint and possible dangers     4 3 Overview of all Safety Instructions for the NEURO TRONIC Multi purpose Device       There is detailed information to some of the safety instructions in the course of this manual  Respect the following  safety instructions to avoid patient injuries and property damage     WARNING 
16. e Signal Function  12  Main Menu       12 1  SAD  Language          12 2 m Pairing  Putting into Operation Controller     12 3 Q Controller Update       12 4 g  Activate  Putting into Operation Remote Control   12 4 1 For One Orthosis  12 4 2 For Two Orthoses    12 5 Cable Connection Test    12 5 1 Cable Connection Test Result Messages and Further Action    12 6 Sensor Test  only with p Motion   12 6 1 Le  Sensor Test Result Messages and Further Action    12 7 Settings of e Motion    12 7 1  gt  Calibrate                12 7 2 lO Turn on p Motion             12 7 3    N Tone Selection       Page    38  38  38  38  38  39  39  41  41  42  42  43  43  43  43  44  44  45  45  46  46    47  47  48    49  49  50    51  51  52  53  53  54  54    55    12 7 4    12 7 5    12 7 6    d   J    PR  Le                   Volume  Signal Selection    Select other Leg    12 8 ay Settings of p Motion    12 8 1    12 8 2    12 8 3    12 8 4    12 8 5    12 8 6    12 8 7       A  pa  5  Ai    a     En    PR  Le                               Sensor Sensitivity  Activity Level  Turn on e Motion  Tone Selection  Volume   Signal Selection    Select other Leg       12 9 QQ Battery Quality    12 10   Step Counter    12 11 il Device Info  13  Indication of Battery Status in Multi purpose Device  14  Replacing the Batteries of the Multi purpose Device  15  Updating the NEURO TRONIC Multi purpose Device  16  Advices for an Optimal Functioning of the Orthosis  16 1 Walking with Orthosis in Auto Mode with
17. e e g  the words MULTI PURPOSE DEVICE  V3 14 and CONTROLLER              V2 18 XXXX on the display  fig  35    5  Exit the menu item by pressing the ED Set button  There is the GER ENNEA    ORTHOPADIETECHNIK MIT SYSTEM             word BACK displayed at this point  Fig  35    13  Indication of Battery Status in Multi purpose Device    There is a battery status indicator in the upper right corner of the multi purpose device  It shows the status of  the rechargeable batteries of the controller                 View Meaning  o Rechargeable batteries full  AUTO FREE LOCK  Q       Rechargeable batteries half full  UER  GENE Rechargeable batteries almost  HNIK MIT SYSTEM empty       14  Replacing the Batteries of the Multi purpose Device    ATTENTION   Batteries of the multi purpose device were not replaced correctly   V    Failure of function   Replace the batteries of the multi purpose device properly and as described in  this manual     ATTENTION   V When replacing the batteries of the multi purpose device  an incorrect battery type is  used     gt  Failure of function   Use the battery type specified in this manual for the multi purpose device     The multi purpose device has two batteries of the type CR2450  You can replace the batteries by yourself when  they are empty  To do so  proceed as follows     62    1  Carefully remove the four grey cover caps on the back of the multi purpose device  There are four screws    underneath   2  Unscrew the four screws with a PH1 screwdriv
18. e single gs SELECT OTHER LEG en  menu items  the name of the menu item itself and the chosen leg are blinking    alternately  exception  Back    R  Select other leg    Pairing and  LY    Language    EED   E  1 2 1 Lang uage ORTHOP  DIETECHNIK MIT SYSTEM                      Fig  7    In the menu item Language you can choose between English and German  The German language is preset  in the multi purpose device  To change the language  proceed as follows     5     1  To go to the main menu  press the  amp  Set button for three seconds    2  Search the menu item Language by browsing with the oo Left and right arrow buttons   3   4      You are now in the language menu where you are offered the languages German and English EN     When you find the menu item  confirm your selection by pressing the Set button     Select a language by browsing with the    Left and right arrow buttons and confirm your selec   tion by pressing the Set button   By confirming your selection you will automatically go back to the main menu     12 2 Pairing  Putting into Operation Controller     To connect the controller with the multi purpose device  proceed as follows     ee R ze    To go to the main menu  press the E Set button for three seconds    Search the menu item Pairing by browsing with the       Left and right arrow buttons    When you find the menu item  confirm your selection by pressing the Gap Set button    Now you can see the options SEARCH and BACK on the display  Furthermore  the multi pu
19. e the above mentioned setting menus on the display  This will be described more detailed in the following     12 7 1 Calibrate    In order for the motion sensors in the controller to detect the position of the lower leg  you have to calibrate  the orthosis before fitting     1  Go to the menu item Settings as described in paragraph 12 7   2  Search the setting menu Calibrate by browsing with the       Left and right arrow buttons   3  When you have found the setting menu  confirm your       PLACE THE ORTHOSIS IN VERTI     selection by pressing the Set button  CAL POSITION AND THEN PRESS  4  You can see the message PLACE THE ORTHOSIS IN  lt CONTINUE gt   VERTICAL POSITION AND THEN PRESS  lt CONTINUE gt  FONNNLIERSSERBEIT    on the display  fig  19   This corresponds to the basic  alignment  see paragraph 7 1   Follow the instruction  and make sure that the orthosis is placed on an even  surface and hold it in this position  Select CONTINUE  by pressing the Set button  Fig  19 Fig  20   5  You can see the message PLACE THE ORTHOSIS IN  HORIZONTAL POSITION  THEN PRESS  lt CONTINUE gt  on the display  fig  21   Follow the instruction and make  sure that the orthosis is lying on an even surface with the foot piece pointing upwards and hold it in this  position  Select CONTINUE by pressing the Ga  Set button    6  You can see the message CALIBRATION PROCEDURE COMPLETED on the display  fig  23   The orthosis has  been calibrated successfully and can now be used    7  Exit the settin
20. er   3  Take off the cover of the multi purpose device   4  Remove the two batteries   5  Insert two new batteries of the type CR2450 with the positive pole facing up   6  Put on the cover again  turn in the screws and place the grey cover caps above them   Note     If you want to save battery power  you can connect the multi purpose device via USB cable to a PC or to a  mobile phone charger  The multi purpose device uses no battery power while it is connected  We recommend  to connect the multi purpose device while doing an update in order to increase the batteries    lifespan     15  Updating the NEURO TRONIC Multi purpose Device    The multi purpose device can be updated with the included USB cable at www fior gentz com updates  for  controller updates see paragraph 12 3   Proceed as follows     1  Download the update and open the file    2  Connect the multi purpose device to the computer with the USB cable    3  Turn on the multi purpose device    4  You must not turn off the multi purpose device during the updating procedure    5  If an update is available  you can see the message    Connected with version XXX  The device can be updated  to version XXX    Click on    Start Update    When the updating process was successful  you can see the mes   sage    Ihe device was successfully updated to version XXX      Enter each update with the version number on the back page of this manual     16  Advices for an Optimal Functioning of the Orthosis    16 1 Walking with Orthosis i
21. eset signal tone  Choose   amp  Left or Right arrow button and    confirm by pressing the ED Set button  fig  27    The following settings are available           COMPLETED      lt          TURN ON P MOTION      FIORBIGENTZ    ORTHOP  DIETECHNIK MIT SYSTEM          Fig  24             2      CONFIRM                    CONFIRM                    CONFIRM         55    1  A beep is emitted as long as the joint is free moving    2  A beep is emitted as long as the joint is locked    3  A short beep is emitted when the joint switches from free moving to locked and when it switches from  locked to free moving    4  Two short beeps are emitted when the joint switches from locked to free moving and one short beep is emit   ted when the joint switches from free moving to locked              If you do not have the menu item oth Signal selection in your multi purpose device  proceed as described in       paragraph 11 1        AR  12 7 6 Select other Leg       This additional setting menu makes it quick to switch between the left and   SELECT OTHER LEG  gt   right leg when you have connected two controllers and want to calibrate  the orthosis or turn on the pressure sensor    Switch the leg by pressing the Set button of the multi purpose  device  Now  you can see the currently selected leg on the display  fig           28   Furthermore  under the single setting menus  the name of the setting       menu itself and the choosen leg are blinking alternately     12 8 Settings of p Motion    I
22. essure to  the sensor     Sensor sensitivity is set too high   In stance phase at low body weight relief   activity level is adjusted too low     Rechargeable batteries are empty     Rechargeable batteries are defect     Short circuit in cable     Strong shocks when orthosis is in Free or  Auto mode     Washer for the plunger of the solenoid is  not assembled     Solution    e Motion is not suitable for patients  who are treated bilaterally  Switch  the orthosis to p Motion     Switch the orthosis at first to  p Motion  Gait re education  The  patient needs to feel more secure   After a few days  you can switch to  the e Motion     Switch the orthosis at first to  p Motion  Gait re education  Teach  the patient a harmonious and natural  swing phase  After a few days  you  can switch to the e Motion     KAFO with ankle joint    Adjust the dorsal stop at the system  ankle joint in such a way that the  forefoot lever can apply a knee  extension moment     KAFO without ankle joint    The necessary forefoot lever can be  applied through the foot piece and or  shoe modifications or adjustments   Produce the forefoot piece stiffly  enough and shift the rolling off line  further forward  if necessary     Increase sensor sensitivity     Change the foot piece in such a way  that the heel fits well and loads the  sensor already at low pressure     Decrease sensor sensitivity   Increase activity level     Insert a new battery set and charge  empty batteries     Insert a new battery set a
23. g  ltig ab Version     et  Die Produktbeilage 15   Kae  i ual is valid from V  un Iler V2 64    inheit Contro  Steuereinheit  device V4 01 03    Multifunktionsgerat Multi purpost  Multifunktionsgerat    und Einstellen der Orthese  Multi Purpose Device and Adjusting the Orthosis    NEURO  TRONIC        ORTHOr  DIETECHNIK MIT SYSTEM    Druckdatum  2015 02  Der Inhalt dieser Produktbeilage wird regelm    ig aktualisiert und ist beim Einsatz des Produktes unbedingt zu beachten  Seite 2  Weitere Informationen zu diesem Produkt finden Sie unter  www fior gentz com neuro tronic de    Date printed  2015 02  This manual is regularly updated and should be strictly followed  Page 35  You will find further information at  www fior gentz com neuro tronic gb       36    Table of Contents    1  Declaration of Conformity  2  Warranty  3  Safety Information  4  Safety Instructions  4 1 Classification of the Safety Instructions  4 2 Structure of the Safety Instructions  4 3 Overview of all Safety Instructions for the NEURO TRONIC Multi purpose Device  5  Application  6  Scope of Delivery  7  Adjusting the Orthosis  7 1 Checking the Basic Alignment of the Orthosis  7 1 1 On the Workbench  7 1 2 Statically on the Patient  7 1 3 Dynamically on the Patient  8  Putting into Operation the Multi purpose Device  9  Checking the Connection between Controller and Multi purpose Device  10  Select a Mode  11  Signal Function for Training Purposes  11 1 Activating the Signal Function  11 2 Switching off th
24. g menu Vs Calibrate by selecting EXIT with the  amp  Set button     pe                                  PLACE THE ORTHOSIS IN HORI  CALIBRATION PROCEDURE  ZONTAL POSITION  THEN PRESS COMPLETED   lt CONTINUE gt   CONTINUE CANCEL  lt  EXIT                      Fig  21 Fig  22    12 7 2 Turn on p Motion    WARNING     No gait re education for the patient   AN   gt  Fall     Make sure that the patient can use the orthosis  recommend him a physiothera   peutic gait re education and explain him the particularities of the joint           The p Motion is recommended for patients who are treated bilaterally and for those who have an atactic gait  pattern  Even in patients who do not yet walk that harmonious or do not feel secure in using the orthosis  you  should turn on the p Motion  With this sensor technology  the system knee joint is controlled only by pressure  and not by motion  To activate the pressure sensor  proceed as follows     1  Search the setting menu LO Turn on p Motion by browsing with the  Left and right arrow buttons        2  When you have found the setting menu  confirm your selection by  pressing the Set button    3  You can see the message COMPLETED on the display  fig  24   The  pressure sensor has been successfully turned on and the orthosis is  switched to the p Motion     12 7 3 Tone Selection    Here you can choose the tone for the signal that sounds when you want  to use the signal function for training purposes for the patient  for  information about the 
25. ically on the patient  see paragraph 7 1 2   Checking the basic alignment dynamically on the patient  see paragraph 7 1 3     e Motion p Motion  Calibrate  see paragraph 12 7 1  Sensor sensitivity  see paragraph 12 8 1     Activity level  see paragraph 12 8 2     7 1 Checking the Basic Alignment of the Orthosis    WARNING   The patient is wearing a shoe to which the orthosis is not adjusted   AN   gt  Failure of the joint function in Auto mode     gt  Fall   Point out to the patient to only wear a shoe to which the orthosis is adjusted   42    7 1 1 On the Workbench    Regardless of the plantar flexion  in automatic system knee joints it is obligatory to adjust the dorsal stop  systematically for an optimal function of the orthosis  The dorsal stop influences the unlocking moment of the  NEURO TRONIC system knee joint during mid stance  Furthermore  it causes an extension moment that applies  to the orthosis and the system knee joint  respectively  Such an extension moment is necessary to unlock the  system knee joint    Fix the foot piece of the orthosis firmly in the patient s shoe and put the orthosis on the workbench  The ankle  joint s dorsal stop must be adjusted in such a way that the line of gravity passes through the middle of the  femoral shell and ventrally ahead of the ankle joint  see also leaflet Basic alignment NEURO MATIC or NEURO  TRONIC KAFO  article no   PR2500 DE GB      7 1 2 Statically on the Patient    For checking the correct static alignment of the or
26. l stance  fig  29   Please read paragraph 6 1 2 of the manual NEURO TRONIC system knee joint for this     To change the pressure sensor s sensitivity for unlocking  proceed as follows     Go to the menu item Settings as described in paragraph 12 7    Search the setting menu Afsensor sensitivity by browsing with the oo Left and right arrow buttons   When you have found the setting menu  confirm your selection by pressing the Set button    The preset unlocking value is 60  since this corresponds to the unlocking moment in natural gait     a a i I ae    If it is necessary to make the pressure sensor more sensitive e g  because       the patient does not step firmly  increase the sensitivity by pressing  the    Right arrow button  There is a plus sign displayed at this 60  point  fig  30   The solenoid releases the locking mechanism earlier    CONFIRM  6  Ifitis necessary to make the pressure sensor less sensitive e g  because  the patient does step very firmly  decrease the sensitivity by pressing       o  the    Left arrow button  There is a minus sign displayed at this  point  fig  30   The solenoid releases the locking mechanism later  GrE ENA  7  Confirm your selection by pressing the E Set button        ORTHOPADIETECHNIK MIT SYSTEM             Fig  30  Note    If you want to set the pressure sensor sensitivity above 81  you can                   _   _   _   _   _   _   _   ________  see the message WARNING  SENSITIVITY VERY HIGH   CHECK SENSOR   cuec SENGOR PLACEMENT G  PLA
27. n Auto Mode with e Motion    ATTENTION   Orthosis remains locked when starting to walk   V     Danger of falling   Point out to the patient to make the first step with the leg with orthosis   especially in Auto mode when e Motion is turned on     For safety reasons  the NEURO TRONIC system knee joint remains locked at the first step  When the patient  starts to walk from a standing position  he should make the first step with the leg with orthosis  The electronics  need a stride to switch from standing mode to walking mode in order to avoid disturbed gait or stumbling when  starting to walk     16 2 Remote Control    Problem Further action  LEDs do not blink after pressing a button    Device does not always react when pressinga Contact technical support   button    63    64    16 3 Multi purpose Device    Problem    Display is blank    Device does not always react when pressing a    button    16 4 Controller    Problem    LEDs do not blink after inserting the    rechargeable batteries          No devices found during m    Pairing the    controller    16 5 System Knee Joint    Problem    Joint does not lock    Problem    Joint does not lock    batteries     Further action    Contact technical support     Further action    Check the polarity and quality of the rechargeable    If the problem remains  contact the technical support     Connect the multi purpose device with the controller  within 30 seconds  Check the polarity and quality of the    rechargeable batteries   If the p
28. n the menu item Settings you can adjust the orthosis  The following setting menus are available with  turned on p Motion       Ji Sensor sensitivity      Activity level    LO turn on e Motion      aft   Tone selection      cf  Volume      UN Signal selection                           ns Select other leg  only with two connected controllers  see paragraph 12 7 6     12 8 1 Sensor Sensitivity    WARNING    The pressure sensor sensitivity is not adjusted correctly   AN   gt  Failure of the joint function in Auto mode      gt  Fall    Adjust the sensor sensitivity to the patient and the shoe     56    The pressure sensor s sensitivity is adjusted for unlocking the joint  You can change the sensitivity from 3  to 99  in steps  of three with the multi purpose device  You determine the exact point when the solenoid releases the locking mechanism  so that the joint in fact unlocks right in time before pre swing  directly after terminal stance  fig  29   It is important that  the solenoid releases the locking mechanism before the extension moment of the forefoot lever causes the actual unlocking     E aim    a E E E    free moving locked locked   locked free moving free moving free moving    Initial contact Loading response ET EE Swing phase  Fig  29    Note    If the solenoid does not release the locking mechanism right in time  the extension moment of the forefoot  lever cannot cause the actual unlocking  Therefore  the joint would remain locked in pre swing  directly after  termina
29. n upper and lower part of the  system knee joint directly at heel off     Proximal  dorsal upper thigh band  transmits flexion load when putting the  leg backwards     Extension stop damper is too long   Therefore  the extension stop damper is  compressed so high that the necessary  extension moment to unlock the joint is  greater than the patient can apply     The patient does not reach the dorsal  stop due to short steps during heel off   Therefore  he cannot apply a knee  extension moment to unlock the joint     Forefoot lever of the foot piece does not  achieve its effect on the knee extension     Pressure sensor Is not freed of load   the  sensor is probably placed too far in front     Walking speed lower as expected  activity  level adjusted too high     Solution    Cut the extension stop damper     Gait re education  Teach the patient  a harmonious and natural swing  phase  As in natural gait  the heel  should touch the floor directly in  terminal swing phase     To increase the flexion  mount the  5   joint s upper part to the NEURO  TRONIC system knee joint as well  as the 5   extension stop to the  medial support and check the basic  alignment     Remove material from the upper  thigh band     Cut the extension stop damper     KAFO with ankle joint    Adjust the dorsal stop at the system  ankle joint in such a way that the  forefoot lever can apply a knee  extension moment     KAFO without ankle joint    The necessary forefoot lever can   be applied through the foot
30. nd dispose  defect batteries according to  regulations        Carry out the So Cable connection  test and change cables if necessary     Due to shocks  the magnetic field is  interrupted and the orthosis locks  automatically  Switch to other mode  and then back to the desired     Assemble the washer  If the problem  remains  contact the technical  support     67    17  Software Versionen Software Versions    Von der Firma FIOR  amp  GENTZ auszuf  llen To be filled out by the company FIOR  amp  GENTZ    Ausgelieferte Version  Datum Unterschrift   Delivered version  Date Signature     Multifunktionsgerat Steuereinheit    Multi purpose device Controller       Vom Orthop  dietechniker auszuf  llen To be filled out by the orthopaedic technician    Aktualisiert auf Version  Durch Mitarbeiter  Datum Unterschrift   Updated to version  Orthopaedic technician  Date Signature     Multifunktionsgerat Steuereinheit  Multi purpose device Controller    Der Qualit  tsstandard der Firma FIOR  amp  GENTZ ist durch eine unabh  ngige Zertifizierungsgesellschaft nach  den internationalen Normen ISO 9001 und ISO 13485 gepruft und bescheinigt worden   The quality standard of the FIOR  amp  GENTZ company has been controlled and certified by an independent certification       organisation according to the international standards ISO 9001 and ISO 13485           FIOR S amp  GENTZ    ORTHOP  DIETECHNIK MIT SYSTEM    FIOR  amp  GENTZ Gesellschaft f  r Entwicklung und Vertrieb von orthop  dietechnischen 
31. nsor under the foot piece several times  To get a result  of the SE Cable connection test  select CONTINUE by pressing the Ga  Set button     Note     When you start the Cable connection test  the orthosis switches automatically into Lock mode and  remains in this mode after the test  To change the mode  use the remote control or multi purpose device     12 5 1 Cable Connection Test Result Messages and Further Action    The following test results displayed on the multi purpose device refer to components that are explained more  detailed in the manual of the NEURO TRONIC system knee joint     Result message     Display  Meaning Further action  Connection of one solenoid The cable connection from  OK controller to solenoid is correct     The cable connection from both  Connection of two solenoids er  OK controllers to both solenoids is    correct     The cable connection from Check the cable connection from    Solenoid not connected ee  controller to solenoid is incorrect  controller to solenoid     Short circuit in the cable from Check the cable joints at the controller    Solenoid short circuited controller to solenoid  and at the solenoid     Please connect solenoid and  pressure sensor again  see manual  NEURO TRONIC system knee joint   paragraph 11      The solenoid is connected to the  Cables mixed up  sensor cable joints and the other  way round     51    52    Result message   Display     Sensor connection OK    Sensor not connected    Sensor short circuited    Sensor w
32. off and the orthosis is       COMPLETED    switched to the e Motion  Z TURN ON E MOTION  gt     12 8 4 J Tone Selection             Please read paragraph 12 7 3 for the tone selection                       FIORBIGENTZ  12 8 5 moy Volume ORTHOPADIETECHNIK MIT SYSTEM                      Fig  33  Please read paragraph 12 7 4 for the volume selection              12 8 6 ath Signal Selection       Please read paragraph 12 7 5 for the signal selection   PR  12 8 7 Select other Leg    This additional setting menu makes it quick to switch between the left and right leg when you have connected  two controllers and want to adjust the pressure sensor s sensitivity or the patient s activity   Switch the leg as described in paragraph 12 7 6     12 9 Battery Quality    ATTENTION   Useful life of the orthosis has shortened   V   gt  Failure of the joint function   Insert fully charged batteries and carry out the battery quality test  In case of  persistent failure  contact our technical support     You can check the quality of the rechargeable batteries in the controller with the multi purpose device  The  battery quality test cannot be carried out with two solenoids at the same time  If you have connected two  NEURO TRONIC system knee joints to the controller  you have to disconnect one solenoid to carry out the test     Note   With new rechargeable batteries  the test can last up to five hours  Select an appropriate time for it  Rechar   geable batteries have to be fully charged for the 
33. or the patient to get used to walking with orthosis  you have the possibility to switch on the signal function   It is meant for training purposes and gives you and your patient better control on when the joint is locked or free  moving  The controller sends out a signal and thus informs about the state of the joint  locked or free moving    You can set tone  volume and signal with the multi purpose device  For the different setting possibilities read  paragraphs 12 7 3  12 7 4 and 12 7 5     11 1 Activating the Signal Function  At all controllers that have been delivered with version V2 64 or higher  the signal function is already activated     At controllers with versions lower than V2 64  this function can be activated subsequently   To do so  proceed as follows     1  Switch on the controller and connect it with the multi purpose device  see paragraph 12 2      2  Go to the menu item Device info   45    3  Now  press both arrow buttons at the same time for 6 seconds until you see the message  COMPLETED  The signal function is activated     At e Motion the signal function is now activated with signal tone 1 being preset at maximum volume and with  all other signal tone options available  At p Motion there is only signal tone 2 available up to version V2 60   You have the possibility to make all signal tones available for p Motion by doing a software update at multi   purpose devices with USB port     If you have a multi purpose device with USB port  you can alternatively a
34. plied to the sensor     Pressure sensor placed in the foot piece   The pressure sensor is not loaded directly  after heel strike because the heel does  not entirely rest in the foot piece due to  a too tight fitted distal  dorsal leg band     Pressure sensor placed under the foot  piece  The pressure sensor is not loaded  directly after heel strike because the  patient has changed his shoes and is now  wearing one whose insole form does not  fit with the underside of the foot piece   The pressure sensor is thus placed in a  space     Pressure sensor placed under the foot  piece    The pressure sensor is not loaded directly  after heel strike because the heel rests  at the edge of the insole or the upper    So the heel loads the sensor only after  compression of these areas     Solution    Correct the integrated leg length  compensation     Gait re education  The patient needs  to get greater safety until he dares  to put his own body weight on the  affected leg     Place the pressure sensor in such   a way that it is entirely loaded  A  partial sensor load causes failure  A  pressure sensor that is placed in a  space between foot and foot piece  or between foot piece and shoe can  only lead to partial loads     Change the foot piece in such a way  that the heel fits well and loads the  sensor already at low pressure     Change the leg band in such a way  that the heel touches the entire foot  piece and loads the sensor already at  low pressure     Let the patient wear a shoe 
35. r    BS    The patient has the ability to ambulate with low walking speed and is able to overcome small obstacles  such as curbs  single steps or uneven surfaces     K 3  Unrestricted outdoor walker     gt     The patient has the ability to move with low  high or variable walking speed and overcome environmental  limitations  Additionally  the patient can walk on open terrain and perform other activities which do not apply  an above average load on the orthosis e g  therapy  work     4  Unrestricted outdoor walker with high demands    fe    The patient has the ability or the potential to walk like the unrestricted outdoor walker  Additionally   the increased functional ability can generate high impact  tension and or deformation loads on the orthosis   These patients are mainly athletes and children     12 8 3 Turn on e Motion    If the patient already has a harmonious gait pattern or has improved it through a gait re education  you can turn  off the integrated pressure sensor and turn on the e Motion  With this sensor technology  the orthosis reacts to  movements and changes in position and no longer to pressure  To turn off the pressure sensor  proceed as follows     1  Search the setting menu tum on e Motion by browsing with the OQO  and right arrow buttons     2  When you have found the setting menu  confirm your selection by pressing the Gap Set button        3  You can see the message COMPLETED on the display  fig  33   The  pressure sensor has been successfully turned 
36. re not pressed at the same time and the remote control cannot be activated  In this  case  repeat the paragraphs 12 4 1 and 12 4 2     Note    If you do not connect the remote control with the first and second controller within 60 seconds  the con   nection with the first controller will be deleted and only the second controller is activated i e  the remote  control is still connected with only one controller  In this case  repeat the paragraphs 12 4 1 and 12 4 2   Mind the given time     12 5 se Cable Connection Test       Depending on the sensor technology selected  you can check the cable connections to the solenoid and pressure  sensor on the orthosis with this test  Carry out the test with the orthosis on the workbench and not when the  patient is wearing it  Proceed as follows        1  Togoto the main menu  press the  amp  Set button for three seconds  PRESS SENSOR SEVERAL TIMES AND  2  To start the test  first search the menu item So Cable connection THEN PRESS  lt CONTINUE gt     test by browsing with the    Left and right arrow buttons  CONTINUE  3  When you have found the menu item  confirm your selection by pres   sing the ED Set button   4  If you have selected the e Motion  you will get the result of the Se  Cable connection test already at this point                 5  If you have selected the p Motion  you can see the message PRESS  SENSOR SEVERAL TIMES AND THEN PRESS  lt CONTINUE gt  on the dis  Fig  16  play  fig  16   Follow the instruction and press the se
37. roblem remains  contact the technical support     Cause    The patient touches the floor first with  his toes and not with his heel because  the foot lifting is not strong enough to  keep the foot in 90   position     Cause    The patient touches the floor first with  his toes and not with his heel because  of a broad bending forward of the upper  part of the body which compensates the  flexed position of the knee  Typical gait  characteristics in patients who depend  on orthopaedic appliances such as    crutches  walking frames or parallel bars     Solution    Enhance foot lifting     Solution    Gait re education  Ask the patient to  walk as upright as possible     Problem    Joint does not lock with  turned on p Motion    Joint does not lock with  turned on p Motion after  heel strike    Joint does not lock with  turned on p Motion after  heel strike    Cause    The patient touches the floor first with  his toes and not with his heel because  of an insufficient integrated leg length  compensation  Therefore  no load can be  applied to the pressure sensor     The patient applies not enough or no  pressure to the sensor because he depends  on orthopaedic appliances such as  crutches  walking frames or parallel bars     Pressure sensor Is not entirely or not at  all loaded after heel strike     Pressure sensor placed in the foot piece   The pressure sensor is not loaded after  heel strike because the heel rests at the  edge of the foot piece and  therefore  no  pressure is ap
38. row buttons   3  Press the Auto and Lock button of the remote control at the same time for about four seconds  fig  14    3 1 When the LED of the remote control blinks red  the buttons were not pressed at the same time  The  remote control cannot be activated  Press the Auto and Lock button again at the same time  fig  14            press 4 sec  blinks red                     Fig  14    3 2 When the LED blinks yellow  the remote control can be activated  Connect the controller with the remote    control by pressing the ED Set button of the multi purpose device within ten seconds  fig  15         press 4 sec  10 seconds yellow                Fig  15    49    50    4  The remote control is activated successfully as soon as you can see the message COMPLETED on the multi   purpose device display and the LED of the remote control blinks green once     Display Light signal remote control Meaning       Colour  green       Signal duration  mm  The remote control is activated             successfully   EEH EEN Ma    ORTHOP  DIETECHNIK MIT SYSTEM    12 4 2 For Two Orthoses          When treating both legs  you can decide to connect the controllers with one or two remote controls  If you  activate two remote controls  you can change the mode separately on each controller and system knee joint   respectively  If you activate only one remote control  you change the mode at the same time on both controllers  and system knee joints  respectively     Activating Two Remote Controls    1  When
39. rpose device shows  you that there are not found any devices yet  fig  8     Insert the rechargeable batteries into the battery compartment of the controller  If you have already inser   ted the rechargeable batteries into the controller  briefly remove one battery and reinsert it  The LEDs at the  controller are blinking yellow  blue and orange from left to right  The controller is ready to connect     Light signal  Signal  Insert rechargeable batteries pew Colour duration Meaning   Yellow E    Controller is now ready for   Blue o   ee  connection within 30 sec    Orange u            Select SEARCH at the multi purpose device by pressing the ED Set button  fig  8   When the device has    found the controller  in the upper left corner you can see FOUND VER  XXXX and left of SEARCH you can  confirm by selecting CONFIRM  fig  9        Connect the multi purpose device with the controller within 30 seconds  Select CONFIRM by pressing the    Left arrow button for this  fig  9        The connection has been successfully established as soon as you can see the message COMPLETED on the    47    48       multi purpose device display  fig  10    9  Now  you can see the message WOULD YOU LIKE TO CONNECT A  SECOND CONTROLLER  on the display as well as the options YES and  NO     To connect a second controller  select YES by pressing  the        Left arrow button  Otherwise  select NO by pressing the   Right arrow button    10  If you have pressed YES  you can see the message SELECT SIDE FO
40. signal function see paragraph 11   You have the  choice between two frequencies which help to distinguish between the  two controllers at an orthotic fitting of both legs  1 stands for the lower  frequency and 2 for the higher one  The preset tone is tone 1  Choose the  Left or  confirm by pressing the  ser  Set button  fig  25      desired frequency with the Right arrow button and       If you do not have the menu item N Tone selection in your multi purpose  device  proceed as described in paragraph 11 1     12 7 4 Volume    You can also set the volume in which the signal tone should sound when  using the signal function  for information about the signal function  see paragraph 11   You can select a volume from 0 5  with 5 being the  maximum volume  To switch off the tone  you have to select 0  If you  have activated the signal function yourself  the volume is preset on 5   If the controller was delivered to you with the signal function already  activated  the volume is preset on 0  Choose the desired volume with  the Left or Right arrow button and confirm by pressing the   Set button  fig  26    If you do not have the menu item cf  Volume in your multi purpose                device  proceed as described in paragraph 11 1     12 7 5 Signal Selection    When using the signal function  you can decide between four different  settings regarding the sound of the signal tone  for information about    the desired signal with the    the signal function see paragraph 11   1 is the pr
41. sion onto the  sensor     Contact technical support     With this test  you can check the assembly of the pressure sensor into the orthosis  Carry out the test only after       Proceed as follows     the Cable connection test was successful  The patient is wearing the orthosis during the Le   Sensor test   vA A    1  Togo to the main menu  press the ED Set button for three seconds     2  To start the test  first search the menu item    sing with the    3  When you have found the menu item  confirm your selection by  Set button   4  You can see the message ASK PATIENT TO WALK TWO STEPS  THEN  PRESS  lt CONTINUE gt  on the display  Follow the instruction and let the  patient walk two steps  To get a result of the le  Sensor test  select  CONTINUE by pressing the  Ga     pressing the       Left and right arrow buttons        Set button  fig  17      Le  Sensor test by brow        ASK PATIENT TO WALK TWO STEPS   THEN PRESS  lt CONTINUE gt     CONTINUE                5  After getting the result of the sensor test  exit the menu item    le  Sensor test by selecting EXIT with the    Set button  fig  18   SENSOR PLACEMENT OK       EXIT    Note   When you start the Le Sensor test  the orthosis switches automatically  into Lock mode and remains in this mode after the test  To change the          mode  use the remote control or multi purpose device        Fig  18       12 6 1 le  Sensor Test Result Messages and Further Action    The following test results displayed on the multi purpose 
42. test     To carry out the test  proceed as follows     The patient is not allowed to wear the orthosis during the test   Insert fully charged batteries into the battery compartment of the controller   To go to the main menu  press the Set button for three seconds        Search the menu item oa Battery quality by browsing with the Left and right arrow buttons   When you have found the menu item  confirm your selection by pressing the ED Set button    Now you can see the message THIS WILL TEST CAPACITY OF YOUR BATTERIES  THE TEST CAN TAKE UP TO  5 HOURS on the display     7  To start the test  select CONTINUE by pressing the ser  Set button  The following messages can be displayed     oe eS Se e    59    60    Message   Display        ERROR  PLEASE PUT IN FULLY  CHARGED BATTERIES    TRY AGAIN CANCEL    FIO RIG ENTZ     ORTHOP  DIETECHNIK MIT SYSTEM          BATTERY TEST IMPOSSIBLE WITH  TWO SOLENOIDS  PLEASE  DISCONNECT ONE SOLENOID    TRY AGAIN CANCEL    GIER Een    ORTHOP  DIETECHNIK MIT SYSTEM             TESTING BATTERIES                                                CANCEL    FIO RG ENTZ          ORTHOPADIETECHNIK MIT SYSTEM       Meaning    Batteries in the controller are  not fully charged     You have mounted two NEURO  TRONIC system knee joints to  the orthosis     Battery quality is being tested     8  The following messages can be displayed during or after the test     Message   Display        BATTERY TEST INTERRUPTED    iS Se EAS  Se  Se Se   BACK    GEN EEE    ORT
43. thosis  the patient must wear the orthosis and stand upright  with parallel feet  When viewed from the side  the line of gravity must run from the centre of gravity through  one half of the distance between the mechanical pivot point at ankle height and the metatarsal joints  Usually   this is already given by the correct basic alignment  since the orthosis itself and soft tissue may be deformed  when wearing the orthosis and the line of gravity  therefore  moves automatically into the desired position  If this  is not the case  the dorsal stop must be readjusted  If the dorsal stop is adjusted correctly  a lever arm between  forefoot and lower leg is formed which brings the patient into a stable balance and applies the necessary knee  extension moment     7 1 3 Dynamically on the Patient    For checking the correct dynamic alignment of the orthosis  the patient must wear the orthosis and walk a few  steps with it  The dorsal stop must be adjusted in such a way that a heel off can be clearly seen in mid stance   Due to this  a lever arm between forefoot and lower leg is formed which brings the patient into a stable balance  and applies the necessary knee extension moment  If the heel does not lift  you must limit the ankle joint s  range of motion in dorsiflexion     8  Putting into Operation the Multi purpose Device       1  The multi purpose device is delivered with integrated rechargeable  batteries and can immediately be put into operation    2  Press the ED Set button to
44. to which  the orthosis is adjusted or adapt the  shoe of his choice in such a way  that the pressure sensor is placed  correctly and loaded directly after  heel strike     Change the foot piece in such a way  that the underside fits in the shoe in  the best way  Ideally  the underside  of the foot piece should directly  have contact with the shoe s insole   Spaces between foot piece and  insole have to be compensated in  such a way that the foot piece loads  the sensor already at low pressure  If  this is not possible  take a shoe with  more inside volume     65    66    Problem    Joint locks unintentionally  in slightly flexed position    Joint does not unlock    Joint does not unlock    Joint does not unlock  with turned on p Motion    Cause    Extension stop damper is too long  You  can see ventrally viewed a clearly visible  opening between upper and lower part  of the system knee joint directly before  heel strike     The patient gets into extension with a  strong momentum  If the leg comes into  full extension before the heel touches  the floor  the lower leg bounces off the  extension stop into a slightly flexed  position  You can see ventrally viewed   a Clearly visible opening between upper  and lower part of the system knee joint  directly before heel strike     Leg was put into cast too extended   Therefore  the patient cannot apply the  necessary extension moment to the  orthosis to unlock the joint  You can see  ventrally viewed a clearly visible opening  betwee
45. wing articles  fig  1       NEURO TRONIC system knee joint     NEURO TRONIC component set including controller      NEURO TRONIC remote control for the patient including microprocessor controlled quick charger     NEURO TRONIC multi purpose device for the technician including batteries   The system knee joint and the controller are mounted to the patient s orthosis  You need the multi purpose    device to put the orthosis into operation and to adjust all settings  The patient needs the remote control to use  the orthosis        PATIENT TECHNICIAN       Remote control    Multi purpose device      a      System knee joint       Controller       Microprocessor   controlled quick    charger               Fig  1    41    The NEURO TRONIC multi purpose device is delivered in a case  The case includes the following  fig  2                     Fig  2   Item Article number Description Unit Quantity  1 ET3830 T Multi purpose device for the technician with batteries Set 1  2 ET0710 USB cable Piece 1  ae ET0830 2450 Coin cells  CR2450 Piece 2    7  Adjusting the Orthosis    Respect a proper alignment before putting the orthosis into operation  Depending on the sensor technique you  can make further adjustments of the orthosis with the multi purpose device afterwards   The following parameters influence the functionality of the orthosis and must be adjusted     Alignment of the orthosis  Checking the basic alignment on the workbench  see paragraph 7 1 1   Checking the basic alignment stat
46. y test  In case of  persistent failure  contact our technical support     ATTENTION    Orthosis remains locked when starting to walk        Danger of falling    Point out to the patient to make the first step with the leg with orthosis   especially in Auto mode when e Motion is turned on     ATTENTION    The multi purpose device is soiled      gt  Incorrect reading of the display and probable failure of the joint function   Clean the display with an unsoiled  soft and dry cloth     ATTENTION    The volume of the signal tone for the signal function is still switched on after the patient  has finished gait re education      gt  Failure of the joint function    Set the volume of the signal tone for the signal function to O after finishing gait  re education     ATTENTION    Batteries of the multi purpose device were not replaced correctly      gt  Failure of function    Replace the batteries of the multi purpose device properly and as described in  this manual     ATTENTION    When replacing the batteries of the multi purpose device  an incorrect battery type is  used      gt  Failure of function    Use the battery type specified in this manual for the multi purpose device     5  Application    The multi purpose device is used to adjust the knee joint system NEURO TRONIC  With the multi purpose device   you can adjust orthoses in which a NEURO TRONIC system knee joint is mounted     6  Scope of Delivery   You purchased the knee joint system NEURO TRONIC  It consists of the follo
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Operating Manual  CF Sound Manual  Manuel Tecnique (complète) (PDF - 3,4 Mb)  User Guide for AirPrime Q26 Series Development Kit  ここ ノ ・施工するには、 電気工事士の資格が必要です~ ー施工後~必ず  plaquette filtre presse    Epson PowerLite 7800pNL Product Support Bulletin  6219435 GBA Galileos GAX5 PT-BR.book      Copyright © All rights reserved. 