Home
        Technical Manual
         Contents
1.       MosBILITYCARE  amp     mobil itg and rehab since 199 2    www mobilitycare net au   1 12 Marriott Street   Oakleigh  MELBOURNE 3166   Phone  03 9568 8383 Fax  03 9568 5114    Invacare   1600 Bed    User s Manual  Version 0310          1600 User s Manual    Table of Contents    OR dey see ee ee ee ee ee eee ee ee ee ee ee    General Information       c ccccccccccccccccccccscsccsceed  Pre Operation Check         ccc ccccccccecsveccvessesee       Operation of Invacare   1600 Series bed                  4  a  Operating the Bed   b  Operating the Castors   c  Operating the Side Rails    Technical Information       cccccccccccccccccccccccc ccc ee o      12     12     General LayOul 6 66 66060ss20ssesscewraweciesentscecte sO  Model Identification          cc ccccvccccccvecvcccscecsce I  Accessories MOUNTING    2  ccc ccccccccccccccccccccccol I  Accessory part numbers        ccccccccccccccccccccvels  Service and Maintenance       cccccccccccccccccccsceeld  Maintenance Chart      cccccccccccccccccccccccccccce l4  Troubleshooting with the electrical system              15  Technical Specifications      ccccccccccccccccccccccccelD  Electrical data    cc cccsccccccc cc ccccccsensscccsseccesle    Warranty  ciccecic cscs cer eserivatessesieseeseseeeeld    2    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invac
2.  to remove lint  grease and dirt  DO NOT wash as this  removes the fine layer of oil     There is a risk of finger entrapment  as there is chance of physical injury         2    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       9  Maintenance Chart     Service and Maintenance of the Invacare   1600 bed must be performed by  personnel who have received the necessary instruction or training     Bed Serial no     Date     Correc operation oib   T  Castor fittings tightened   o   Castor brakes checked     Actuators checked for visible damage        Wiring correct and undamaged  Plugs  undamaged and no interference with bed  movement    Damaged paintwork repaired a  12 Monthly      Check all pivot points es eee       Electrical System   a  The electrical system in this bed is a sealed unit and must be replaced     b  A battery change can be performed by certified maintenance staff and requires the  removal of four screws securing the box to the bed frame and the withdrawal of a plug  from the control box     Toe Operated Castors   Toe operated castors are the non adjustable type  however they can quickly and easily be  replaced by certified maintenance staff     Central Lock Castors    Central Lock Castors are adjustable  Certified ma
3. DES  HORIZONTAL BSH 3093  SAFETY SIDES   VERTICAL BSV VSS425  BUMPER BAR BB 16 1205RG    Head board and Foot board ordering numbers    BZP XXXX XX I XX    Edge Colour    Edge Type    T Mould  2 Polished  Panel Colour    Panel Shape  S Drop in  Std   D Discovery  B Bed Stripper   A Arch  C Colonial  Q Queenanne  H HHE Shape            Footboard Frame  S Drop in  Std   D Discovery   B Bed Stripper  Blank Hard board only    Headboard Frame  S Drop in  Std   D Discovery   B Bed Stripper  Blank Hard board only    P Pair  F Foot  H Head        All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       8  Service and Maintenance     Service and maintenance of the Invacare 1600 Series bed must be performed by personnel who  have received the necessary instruction or training     With normal daily use  service must be carried out according to the service schedule     Ensure that the mains cord is disconnected from the mains supply     The top frame must be supported during service inspections to prevent  accidental lowering        Cleaning     Cleaning can be done using household detergents  brush  cloth and sponge  Dry the bed after  cleaning     DO NOT use acids  alkaline  or any other chemical solvents     For cleaning Actuators use a dry cloth
4. an be locked individually by lowering the brake on the castors     when using any positioning function  The bed must be locked when the  patient is moving in and out of the bed     IN The bed s castors must be locked whilst nursing the patient in bed or       Central locking Treadle bar    c  Operating the Safety Sides     There is a risk of personal injury on operation of safety sides  Before  lowering the bed  the safety side must be down to avoid potential personal  injury due to clearances with the floor  See Figure I        DO NOT place hands between moving parts   DO NOT sit on rails     To Raise   Grasp the top rail of the safety side and lift in an arc  There will be an audible click when it locks     To Lower   Standing beside the bed near the middle of the safety side  hold the black release knob with the  hand nearest to it and the top rail with the other hand        To lower To raise    6    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       Technical Information    4  General Layout    Invacare Australia accepts no liability for use  change or assembly of the  product other than as stated in this User   s Manual        Invacare Australia is certified to ISO 9001 2000   The 1600 bed has been independently tested and co
5. are   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       User Part    I  General Information    Congratulations on your choice of the Invacare   1600 bed   e This bed has been developed for users requiring care in an Aged Care Facility or home care   e The bed combines excellent ergonomic design     e A specially designed 2 or 4 sectioned mattress support provides the user with excellent  comfort and provides working conditions for the carer     Invacare   Australia accepts no liability for any use  change or assembly of the    product other than as stated in this User   s Manual        2  Pre Operation Check    Read This Before Operation     Ignoring This Warning May Result In Damage Or Injury        Disconnect the mains cord from the supply before moving the bed     Ensure the mains cord is kept clear of the wheels when moving the bed  and is never caught in the  bed operating mechanism or in a safety side  It can be dangerous to run over the mains cord     Ensure the mains cord has enough slack when connected that it is not strained when the bed is  raised and lowered     3    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       Positioning Of Bed     Bed fitted without Bumper bar     With the bed at it   s lowes
6. asurements  and angles are stated  without tolerances     soeerrounsr   Invacare reserves the  right to change the  stated measurements  and angles without  warning     n    a        z A  A a    Maximum weight  carrying capacity is  200kg  tested as per  ASNZ standards   includes patient  weight of  65kg         EES 7 a   E           EN SE  Width outside  92cm Length outside  2   5cm Height adjustability  Mattress support Backrest angle   Width inside  80cm Length inside    90cm 20 67cm dimensions   90x80cm 0 70    Thigh angle  Calf rest lift  Tilt option  Max user weight  Safe working load   0 38   0 13   N A 165kg 200kg    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       12  Electrical data    Voltage supply  230V AC  50Hz     20V  Sound level  lt  45db   Protection class  IP54   Insulation class  Il type B    13  Warranty      0 years on welded frame  3 years electrical system  2 years hydraulic   2 years mechanical   2 years castors     year battery     5    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacifi
7. c    NOTES    Local Distributor     Yes  you can     Invacare Australia Pty Ltd Invacare New Zealand Invacare Asia Ltd    Lenton Place  PO Box 5002  North Rocks NSW 2151  Australia 4 Westfield Place  PO Box 62 124  Mt Wellington Auckland  New Zealand 1 Lenton Place  PO Box 5002  North Rocks NSW 2151  Australia  Freephone  800 460 460 Fax  02 8839 5353 Freephone  0508 468 222 Freefax  0508 807 788 Phone   61 2 8839 5330 Fax   61 2 8839 5331  Email  sales invacare com au  Web  www invacare com au Email  sales invacare co nz Web  www  invacare co nz Email  asiasales invacare com Web  www invacare com    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific       
8. ess fit the standard style footboard frame into the  socket holes  This allows rapid removal in an emergency     A bed stripper style footboard is an option which quickly  and easily folds down to allow access to the foot end of  the bed for the purpose of stripping the bed and replacing  covers     Mount bedstripper on the clips at the bed end        c  Self Help Pole    Pole connector  The self help pole is an option to provide extra patient  support  The handle length can be adjusted according to  the need     Press fit the pole into the RHS section located in the  centre at the head end of the bed        Self Help Pole  d  Foot Extension     A foot extension panel can be purchased which extends  the length of the bed by 250mm  A mattress bolster is  then available to sit on top of the foot extension effectively  lengthening the mattress accordingly        Clips on to the foot end of the bed and clamped by knobs     Foot Extension    9    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       e  Safety Rail     There is a risk of entrapment on assembly and operation of safety sides           Raise the bed about one third    2  With the release knob at the head end  position safety  side brackets against the bed frame such that in t
9. he  lowered position the safety side does not go past the  head of the bed    3  Press fit onto bed frame  attach flat bar and tighten the  star knob for secure holding        Safety side brackets    f  Bumper Bar     The bumper bar fits both Standard and Knee Break Beds     Installation      Apply wheel brakes on bed  Raise bed about one third   2  Plug bed into power    3  Using handset controls  lift mattress base to its  highest position    4  Holding the bumper inverted  line up both labels  with holes in the end of base frame    5  Slide the bumper into the frame  at Head End of the  Bed Base  until the locating pins click into the holes   approximately  00mm           Bumper bar    Removal       To remove the bumper  apply wheel brakes on bed    2  Squeeze both sides of locating pins to disengage bar  from holes    3  Squarely pull the bumper out of the bed   s base frame        Bumper bar    10    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       7  Accessory Part Numbers    DENSON EON   FUNCTION PART NUMBER    HEAD BOARD STANDARD SUPPORTS HEAD BZP XXXX XX   XX  FOOT BOARD STANDARD SUPPORTS FOOT BZP XXXX XX   XX    FOOT BOARD BASE STRIPPER   SUPPORTS FOOT AND FOLDS DOWN FOR BZP XXXX XX   XX  MAKING UP THE BED    SAFETY SI
10. ing position  lowering or increasing height of mattress platform  or lifting the back rest      the carer user must ensure that all limbs   appendages must be clear of all bed moving parts  Furthermore the  carer user must ensure that the underside of the bed is clear of  obstructions prior to operation     Linak Handset       Back rest    Button with the symbol shown to  the left  Push the left hand side of the  button to raise  right to lower     2  Hi Low    Button with the symbol shown to  the left  Push the left hand side of the  button to raise  right to lower     3  Thigh  amp  knee section    Button with the symbol shown to  the left  Push the left hand side of the  button to raise  right to lower        Back rest  Buttons with the symbol shown to    the left  Push the top button to raise   bottom to lower     2  Thigh  amp  knee section    Buttons with the symbol shown to  the left  Push the top button to raise   bottom to lower     3  Hi Low    Buttons with the symbol shown to  the left  Push the top button to raise   bottom to lower        All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       b  Operating the Castors     A single treadle bar at the foot end of the bed locks all Central locking Castors   Toe Break Castors c
11. intenance staff can tighten brakes simply by  tightening the grub screw protruding from the brake pad at the rear of the castor just above  the wheel     Upper Posture  Fowler  Removal and Replacement     Removal     Disconnect bed from mains supply     2  Remove bolt fastening fowler Rocker Arm  25mm solid square MS  mild bend one  end  4 lugs attached  to Fowler Frame   Leave rocker arm attached to actuator      3    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       3  Insert 5 0 mm punch into Fowler retaining lug  Effectively hinge for whole fowler frame  located at foot end of fowler frame either end of 25 mm Dia  tube   Push back retaining pin  until pin is free of nylon bush and raise that side of fowler slightly to clear the mount     Fowler frame and sheet should then move sideways and remove from top frame     Replacement of fowler frame is the reverse of the above procedure   Use new pivot bushes and other components if they are worn     Pivot Points   Replace pivot pins  bolts or removable bushes  where fitted  if there is significant play or the  there is wear greater than Imm in diameter     I I  Technical specifications     All measurements are  stated in millimetres   mm  All angles are  stated in degrees    All me
12. mplies with all requirements of AS NZS  200 2 38 1997 for 200kg SWL     The bed serial number is located         Parts of a standard bed   Upper posture   Base frame   Control box  actuators  handset  Castors   Thigh section  4 section only   Leg section  4 section only   Bed ends   Bumper bar    CON DU AWN  gt        7    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       5  Model Identification     Ordering numbers    BAS XXXX KXY VY ZZ          RG   River Gum    100   100mm Castors  125   125mm Castors        C   Central Locking  T   Toebreak    1600   2 Section  1601   4 Section    S Standard    A Aged Care       8    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       6  Accessories Mounting     For an ergonomic working position  raise the bed mattress  support to approx  60cm above the floor level     a  Assembling the Headboard     Press fit the standard style head board frame into the  socket holes  This allows rapid removal in an emergency     b  Assembling the Footboard     Pr
13. t height  position the bed so that the head end buffers are at least  30mm from the wall and some slack is present in the electrical supply cable  Lock the castors  If  this positioning is incorrect or not performed the bed may cause damage to the wall as it moves  through an arc upon raising     Safety Sides     Due to the low height nature of this bed  safety sides are not recommended to be used as the  risk of injury due to falling out of bed is very small  However  if they are fitted then the following  considerations should be made   e Before lowering the bed it is imperative that safety sides when fitted be folded into their     down    position  If this is not performed then a    crushing risk    may be present between  the floor and the safety side catch or clamp  Figure   below demonstrates clearance with  safety side raised   e When safety sides are lowered  the latch  being the lowest point swings up and the risk  is removed           d  T  v  5  R       Figure    Low Height Clearance    4    All information is subject to change without prior notice  Photographs are for illustration purposes only and may vary from product description   Invacare   is a registered trademark and Yes  you can     is a trademark of Invacare   Corporation  2010 Invacare Asia Pacific    1600 User s Manual       3a  Operation of Invacare 1600 Series bed    WARNING  There is a potential risk of personal injury when  operating the bed  Before operating the bed  eg moving the bed into  a sitt
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Hot Shot HG-96168-1 Instructions / Assembly  イージーワイヤレススティックET-UW100 ファームウェアアップデート手順  ファームウェア変更に関するご注意 以下に説明する操作は、カメラから  Instruction Manual CombiPro - Amalgamated Pumping Supplies  コッブ500種鶏飼育管理マニュアル(全ページ)  DC100V系・小形直流電源盤(バッテリ内蔵)  FreeStyle™ - Williamson  spirit tc 5306 - Schulthess Maschinen AG  TruSight Tumor Sample Preparation Guide - Support  Tuff-Cam    Copyright © All rights reserved. 