        User Manual - Alburtylab, Inc.
         Contents
1.           Contact AlburtyLab for a consultation and estimate on custom vertical    elutriator sections     Not equipped to perform specialized aerosol tests  Let us do them for you  AlburtyLab is a  focused aerosol laboratory providing a full range of research  development  and third party  validation services  We also design and build custom wind tunnels  flow systems for testing  aerosol detection and collection systems  aerosol inlets and aerosol generators  AlburtyLab also  offers computational fluid dynamics  CFD  analysis  third party testing validation  and all types    of aerosol testing     Located in Drexel  Missouri  just south of Kansas City  AlburtyLab serves the aerosol research   development and instrumentation communities  Our associates and specialized facilities combine  to help our customers develop and improve their products     AlburtyLab is your source for everything aerosol     Dave Alburty  Aerosol Scientist and President  dalburty  alburtylab com    Zach Packingham  Design Engineer  zpack  alburtylab com    AlburtyLab com  128 E  Main St   Drexel  MO 64742  816 619 3374    AlburtyLab Inc  AG 5025 Operation Manual 11    
2.  nozzles can be adjusted as desired  An individual or a combination of  nozzles can be used by installing a hose clamp on the air lines leading to the unwanted nozzles     Wing Nuts for  Securing Diaphragm    Da Aur Lines for Four Air  is   Injection Nozzles        Umbilica    Figure 5  Fluidized Bed Base Unit with Four Way Air Injection Stage    AlburtyLab Inc  AG 5025 Operation Manual 5    S  AlburtyLab Inc     Serving our customers with innovation and creativity    3 Safety    Flammable or volatile substances should never be aerosolized using the AG   5025  For all other  substances  refer to their Material Safety Data Sheet  MSDS  for proper handling and safety  precautions     There are no inherent safety risks in operating the AG   5025 when used as directed  However  if  operating in close quarters for prolonged periods of time  hearing protection should be used     4 Unpacking and System Setup    This section provides information concerning the accessories shipped with the AG   5025  Fluidized Bed Aerosol Generator and describes basic setup procedures     4 1 Packing List    Table 1 provides a packing list of all items that should have been shipped to you  Please  compare the list to the items you received  If any items are missing  notify AlburtyLab  immediately     Table 1  Accessories Packing List             Qty Description Part No    1 Control Unit AG 2006   1 Fluidized Bed Base Unit AG 2007   1 Four way Elutriation Air Injection Stage AG 2008   1 Elutriator Sect
3.  the single pin heavy duty clamp as  shown in Figure 6 and Figure 7    Place a gasket on top of the Elutriator Section  Figure 5   Place the Elutriator Top Cap on top of the gasket   and tighten the single pin heavy duty clamp        AlburtyLab Inc  AG 5025 Operation Manual 7    SX     AlburtyLab Inc     Serving our customers with innovation and creativity          Figure 6  Figure 7     e Connect the conductive tubing to the 34 inch OD hose barb on the Elutriator Top Cap  with as few bends and as short a length as possible  connect the other end of the  conductive tubing to the aerosol chamber or flow tube  etc     5 Operation    5 1 Signal Settings Guidance   The AlburtyLab AG   5025 has a frequency response of 30 3000 Hz  The particular frequency to  use depends on what is being aerosolized  so some experimentation may be needed to find the     sweet spot    for your application  440 Hz is generally a good starting point     Some dust samples have a tendency to form harmonic patterns on the diaphragm and get stuck   If this occurs  try using a frequency sweep of 100 Hz or more per second  For example  if  harmonic patterns are forming at 330 Hz  try a logarithmic bidirectional sweep between 280 and  380 Hz over a period of 0 7 seconds  A sine waveform is recommended for most applications   Square wave and saw tooth waveforms can also be used  however  they should only be used at  low amplitudes due to the increased strain on the amplifier     5 2 Aerosol Generation  The sy
4. 20VAC connection on the back of the Control  Unit and then into an available power outlet  Switch the On Off toggle switch on the  front of the Control Unit to the On position to verify the unit has power    Connect the    s inch audio jack end of the Monaural Audio Cable to an audio signal  generator or a computer and the 1   4 inch audio jack to one of the Signal In connections   front or rear  as desired for lab layout     Connect the air line portion of the umbilical to one of the Air Out connections on the  Control Unit  front or rear  as desired for lab layout  and connect the other end to the Air  In connection on the Fluidized Bed Base Unit   Connect the audio cable portion of the umbilical to  one of the Signal Out connections on the Control  Unit  front or rear  as desired for lab layout  and  connect the other end to the Signal In connection  on the Fluidized Bed Base Unit    Install the latex diaphragm by loosening the wing  nuts on the Four way Elutriation Air Injection  Stage and placing the diaphragm under the bottom  gasket  While tightening the wing nuts gently pull  the diaphragm on each side  The diaphragm  should be taut but not stretched    Adjust the four way elutriation air injection  nozzles shown in Figure 4 as desired for  dissemination        Pour the sample onto the center of the diaphragm  as shown in Figure 5    Place a gasket on top of the Four way Elutriation  Air Injection Stage    Place the Elutriator Section on top of the gasket  and tighten
5. Control Unit  clockwise to the desired level    e Once the input level is set  it can be turned on and off at the Control Unit by toggling the  Run Standby toggle switch  at the signal generator  or by muting unmuting the computer  output    e Switch the High Low toggle switch on the front of the Control Unit to the desired flow  rate range  The Higher flow rates will allow more particles to escape the elutriator   however  higher flow rates will also allow larger particles to escape    e Set the desired flow rate on the applicable flowmeter  high or low range  on the front of  the Control Unit    e The chart contained in Figure 6 can be used as a starting point for aerosolizing particles  with a known size distribution  The data in this graph was generated for dissemination of  particles of two different densities and using the AG   5025     AlburtyLab Inc  AG 5025 Operation Manual 9    S  AlburtyLab Inc     Serving our customers with innovation and creativity                         70           Density   2 7 g cm3  E 60       P    s    Density   1 2 g cm3             50   amp   A  o 40  2  a 30  2  S  e 20   gt             10   lt    0 T T T T T          0 5 10 15 20 25 30  Flowrate  Lpm     Figure 6  Flowrate Associated with Particle Settling Velocity    6 Cleaning and Maintenance   The Elutriator Section  Four way Elutriation Air Injection Stage  and Elutriator Top Cap are  made of T 304 stainless steel and can be cleaned with most standard cleaners such as bleach   alco
6. O     gt  AlburtyLab Inc     Serving our customers with innovation and creativity       AG 5025 Fluidized Bed Aerosol Generator with Vertical Elutriator    OPERATION MANUAL    1 Product Overview    This section contains a description of the AlburtyLab  AG 5025 fluidized bed aerosol generator and a brief  history of the instrument     1 1 Product Description    AlburtyLab  Inc  designs  builds  and sells the  AG   5025  a custom PITT 3 type fluidized bed  aerosol generator  with a remote amplifier and flow  controller as shown in Figure 0  The drivers for these  units are fully enclosed and the generators are built  of easily decontaminated T304 stainless steel with  latex diaphragms and Teflon or butyl gaskets     Unique air injection methods make these units  exceptionally controllable  Ongoing research  indicates that concentrations in the range of several  mg m  are easily maintained for long periods of time  in closed systems using this generator     Variable waveforms from almost any computer or  audio signal generator can be used to drive the  aerosol generator        Generators in this series range from 0 20 to 25 Lpm  output  with diaphragm diameters ranging from 1 5  in to 12 in  Particle size distributions can be controlled by using well defined input materials   such as Celite R  545  talc  Arizona road dust  fluorescein  grain dust  dry Bacillus globigii  Bg   spores  and dry polystyrene microspheres  You can also aerosolize your own materials or re   aerosolize p
7. articles captured by air sampling     Figure 1  AG 5025 Aerosol Generator    Standard features include    Rugged and dependable industrial grade construction   Extended use periods enabled by fan cooled electronics   Remote flow rate and amplitude controls   Quick  easy  tool free removal of entire aerosol bed  injector section  and elutriator allows   swift  easy material reloading and decontamination   e The aerosol bed can be driven by any audio signal generator  allowing the utilization of  unique wave forms or combinations of wave equations   e Completely sealed sample section can be used in positive or negative pressure systems    AlburtyLab Inc  AG 5025 Operation Manual 1    SX     AlburtyLab Inc     Serving our customers with innovation and creativity    e Versatile umbilical attachments on the front and rear of the Control Unit allow for  convenient placement in your lab     1 2 Instrument History    The first PITT   3 was described by Weyel et al   1984  and was designed to generate a cotton  dust aerosol  In this short communication the author provides the design details of the PITT   3  generator constructed at the University of Pittsburgh  However  none of these PITT   3 type  instruments were commercially available prior to AlburtyLab   s release of the first AG   5025  model in 2007     In 2009  the AlburtyLab AG   5025 was redesigned to include a Control Unit that would allow  the user to control the aerosol generator remotely  This allows for greater flexibil
8. he    AlburtyLab Inc  AG 5025 Operation Manual 4        AlburtyLab Inc     Serving our customers with innovation and creativity       bottom of each of the flowmeters  Selection of the desired flowrate range is made using the  High Low toggle switch located to the left of the Low Range Flowmeter on the front of the  Control Unit     Air flow and audio signal to the Fluidized Bed Base Unit can be turned on and off at the Control  Unit by switching the Run Standby toggle switch located to the right of the On Off toggle switch  on the front of the Control Unit  An indicator lamp above the Run Standby toggle switch is  illuminated whenever it is switched to the Run position     The key components on the Fluidized Bed Base Unit are labeled in Figure 5  Connections for  the Umbilical  composed of an air line and a 1   4 inch monaural audio cable  that connect the  Fluidized Bed Base Unit to the Control Unit are the only connections on the Fluidized Bed Base  Unit     The four wing nuts secure the Four way Elutriation Air Injection Stage to the Fluidized Bed  Base Unit  It also is used to hold the latex diaphragm in place  The latex diaphragm is installed  by loosening these wing nuts and placing the diaphragm under the bottom gasket  While  tightening the wing nuts the diaphragm is gently pulled on each side  The diaphragm should be  taut but not overly stretched     Air is introduced to the Elutriation Section through four air nozzles in the Elutriation Air  Injection Stage  These
9. hol  or hydrogen peroxide  The latex diaphragm can be disinfected using 10  bleach  3   Hydrogen peroxide  or by autoclaving     The only user serviceable maintenance item on the Control Unit is the air filter located on the  back  It should be removed and cleaned with water every 24 hours of operation depending on  conditions  The HEPA filter in the Compressed Air Inlet Line purifies the injection air  This  should not have to be changed unless water is allowed to enter the air line or a drop in the  maximum flow rate is observed  Replacement filters can be obtained from AlburtyLab     7 Specifications    e 120VAC  30W   e Frequency response range  30 3000 Hz   e Flow rate range  0 20 25 Lpm   e Maximum compressed air input  10 PSIG  e 0 02 um internal air filter   e Diaphragm diameter  4 75 inches    8 Bibliography    Weyel  D A   M  Ellakkani  Y  Alarie  and M  Karol  1984  An aerosol generator for the  resuspension of cotton dust  Toxicol  Appl  Pharmacol  76 544 547     AlburtyLab Inc  AG 5025 Operation Manual 10    I  AlburtyLab Inc     Serving our customers with innovation and creativity    AlburtyLab Inc  AG    5025 Part List                         Description AlburtyLab Part No  Price  Complete AG   5025  4 75 in diaphragm  AG 5025  9995 00  Replacement Diaphragm 0 006 Latex  4 75 in  AG LR 4 75  24 95  0 02 um Injection Air Filter AG 2003  129 95  Conductive Silicone Tubing  3 4 in dia   AG 2004  44 95 foot  Signal compressed air umbilical line AG 2016  99 95      
10. ion AG 716 2   1 Elutriator Top Cap AG 716 1   2 Single Pin Heavy Duty Clamp AG HHM OH  3 Gasket AG 40 E5   1 AC Power Cord AG 2013   1 Compressed Air Input with HEPA Filter AG 2003   1 Ys inch to     inch Monaural Audio Cable AG 25 125  1 Umbilical  Air Line and 1   4 inch Monaural Audio Cable  AG 2016   2 Latex Diaphragm AG LR 4 75       4 2 Requirements    The following is required for use of the AG   5025 Fluidized Bed Aerosol Generator   e Compressed air source that can be regulated to 10 PSI and capable of supplying greater  that 30 Lpm  e 120VAC grounded power supply  e   inch ID conductive silicone tubing  e Audio signal generator or a computer with an audio output    4 3 Setup    The system is setup for use as follows     e Switch the On Off toggle switch on the front of the Control Unit to the Off position     AlburtyLab Inc  AG 5025 Operation Manual 6    O    4 4 Sample Loading    The test dust is loaded into the system as follows            AlburtyLab Inc     Serving our customers with innovation and creativity    Switch the Standby Run toggle switch on the front of the Control Unit to the Standby  position    Set the regulator on the compressed air source to 10 PSI    Connect a regulated air supply line to the compressed air input with HEPA filter that was  shipped with the system    Connect the air input with HEPA filter line to one of the Air In connections on the  Control Unit  front or rear  as desired for lab layout     Connect the AC power cord to the 110 1
11. ity when  employing the generator in an enclosed aerosol chamber     2 Description    The primary components of the AG   5025 Fluidized Bed Aerosol Generator with Vertical  Elutriator are labeled in Figure 2                           _ Elutriator Top Cap    Elutriator Section    Four way Elutriation  Ai Injection Stage    Fluidized Bed  Base Unit    Control Unit    Figure 2  Primary Components of the AG   5025    AlburtyLab Inc  AG 5025 Operation Manual           AlburtyLab Inc     Serving our customers with innovation and creativity    The Elutriator Top Cap has a   inch OD hose barb in the center that allows for easy connection  to   inch ID conductive tubing  The connection between the Elutriator Top Cap and the  Elutriator Section is sealed with a gasket and a single pin heavy duty clamp  Regulation of the  flowrate through the system allows separation of lighter particles from heavier ones to be  achieved in the Elutriator Section  Compressed air is introduced to the system through four  nozzles located in the Four way Elutriation Air Injection Stage  Note  The compressed air must  be regulated to a pressure of 10 PSIG or less  The connection between the Four way Elutriation  Air Injection Stage and the Elutriator Section is sealed with a gasket and a single pin heavy duty  clamp  The Fluidized Bed Base Unit houses the speaker which transfers energy into the  Elutriator Section through the latex diaphragm  Four wing nuts on the Four way Elutriation Air  Injection Stage h
12. old the latex diaphragm securely in place against the stainless steel plate on top  of the Fluidized Bed Base Unit  The Umbilical is composed of an air line and a   inch monaural  audio cable and is used to connect the Fluidized Bed Base Unit to the Control Unit  This  Umbilical allows the user to control the AG   5025 remotely     The key components on the front panel and the back panel of the Control Unit are labeled in    Figure 3 and Figure 4  respectively     Amplitude Dial for High Low Flowmeter  Audio Input Toggle Switch    ccm       D  D0    On Off Toggle    Run Standby Toggle Switch    Front Connection       for Air Input    Front Connection Front Connections Low Range High Range  for Audio Input for Umbilical Flowmeter Flowmeter    Figure 3  Front Panel of the Control Unit    AlburtyLab Inc  AG 5025 Operation Manual 3    I  AlburtyLab Inc     Serving our customers with innovation and creativity    Control Unit Air  Filter  m    Back Connections    for Umbilical    Back Connection  for Audio Input       AC Power Cord Back Connection  Connection for Air Input  Figure 4  Back Panel of the Control Unit    An AC Power cord is provided for use with the three prong AC input connector on the lower left  side of the back panel  Power to the Control Unit is controlled by the On Off toggle switch on  the left side of the front panel  An indicator lamp above the On Off toggle switch is illuminated  whenever power to the Control Unit is switched to the On position     Connection
13. s for the compressed air input line are located on the front panel below the On Off  toggle switch and on the lower right side of the back panel  Note  Air should only be introduced  through one of these connections as determined to be the most convenient for lab layout by the  user  The compressed air must be regulated pressure of 10 PSIG or less to avoid bursting the  internal tubing connections     Connections for the audio signal from an audio signal generator or a computer are located to the  right of the air input connections on the front and the back panel  Note  Audio signal should only  be introduced through one of these connections as determined to be the most convenient for lab  layout by the user     Connections for the Umbilical  composed of an air line and an   inch monaural audio cable  that  connect the Control Unit to the Fluidized Bed Base Unit are located in the middle bottom of the  front panel and on the top right of the back panel  The set of connections determined to be the  most convenient for lab layout by the user should be used     The amplitude of the audio signal sent to the Fluidized Bed Base Unit is controlled by using the  dial located in the middle of the front panel     Two flowmeters are located on the right side of the front panel  The Low Range Flowmeter has  a range of 200 to 2500 mL min  0 2 to 2 5 Lpm  and the High Range Flowmeter has a range of 2  to 25 Lpm  The air flowrate through the system is controlled using the knobs located at t
14. stem is prepared for dry powder dissemination as follows     e Turn the Amplitude control dial on the front of the Control Unit all the way down   CCW    e Switch the On Off toggle switch on the front of the Control Unit to the On position     AlburtyLab Inc  AG   5025 Operation Manual 8           AlburtyLab Inc     Serving our customers with innovation and creativity    e Switch the Standby Run toggle switch on the front of the Control Unit to the Run  position  Note  Never switch Control Unit from Standby to Run unless the Fluidized Bed  Base Unit is connected by the umbilical  Damage to the amplifier may result    e Turn on the audio signal generator or start the signal generator software on your  computer  One possible audio signal generator that can be used is SigJenny  SigJenny is  a freeware signal generator program written by NaTCH engineering and is available for  download at their website  http   www natch co uk   SigJenny is simple to use  Set the  desired frequency in the    Frequency 1    box and uncheck the mute box under the left  comer of the graph  For a frequency sweep  check the    Sweep    box and set the     Frequency 1    and    Frequency 2    to the desired values  Some users have experienced  difficulty using SigJenny with Windows Vista  If you experience any problems  contact  AlburtyLab for suggested solutions    e Start the signal generator at approximately 50  volume    e Slowly increase the input level by turning the Amplitude control dial on the 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
ATH-C505iS 取扱説明書  7 - SEW-Eurodrive  Samsung WB WB210  Philips 22PFL3403 22" LCD HD Ready Flat TV  intext:Bedienungsanleitung filetype:pdf  Grundig MONHACO 20 LCD 51-9622 DL User's Manual  Firefriend DF-6513 fireplace  sistema di videosorveglianza ip 8 telecamere    to view the file - Support On Line    Copyright © All rights reserved. 