1. InBodys570 User s Manual Please note the important information below before reading this manual A Warning Failure to comply with safety warnings and regulations can cause serious injury or death A Caution Failure to comply with safety cautions and regulations can cause injury or property damage For any problems with the InBody570 or any clinical questions contact Biospace Inc 13850 Cerritos Corporate Dr Unit C Cerritos CA 90703 TEL 1 323 932 6503 FAX 1 323 952 5009 Website http www biospaceamerica com E mail usa biospaceamerica com 1996 Biospace Co Ltd All rights reserved Reproduction adaptation or translation of this manual is prohibited without prior written consent from Biospace Co Ltd under the copyright laws This manual might have typographical errors and its content can be changed without prior notice Biospace Co Ltd shall not be liable for any errors incidental or consequential damages that occurred by not complying with the content of the User s Manual Visit our website http www biospaceamerica com to view and download further information about the functions of the InBody570 the explanation of results and more Biospace Co Ltd reserves the right to modify the appearance specifications and etc of the InBody570 to improve the quality of the product without prior notice for reasons of performance improvement 1996 Biospace Co Ltd All rights reserved BM US
2. A list of printers compatible with the InBody570 can be found at http www inbodyservice com 1 First turn off the InBody570 and then the printer You may experience connection issues in connecting the printer to the InBody570 if the InBody is turned on 2 Plug the USB cable provided with the printer into the USB HOST port on the rear panel of the InBody570 and plug the other end of the USB cable into the printer USB SLAVE PC USB HOST port USB HOST PRINTER STORAGE 3 Turn on the printer 4 Turn on the InBody570 and setup your printer under Settings of the Administrator Menu 09 Printer Setup 5 You can edit your printing settings under Settings of the Administrator Menu from 11 Automatic Printing Options through 15 Printing Alignment 2 Stadiometer If a stadiometer is connected to the InBody570 the height values measured by the stadiometer will be sent directly to the InBody570 x Always connect a stadiometer from BIOSPACE 1 First turn off the InBody570 and then the stadiometer x You may experience connection issues in connecting the stadiometer to the InBody570 if the InBody is turned on 2 Plug the serial cable provided with the stadiometer to the 9 pin stadiometer serial port on the rear panel of the InBody570 Te ic Blood Pre E Stadiometer Monitor 9 pin stadiometer serial port 3 Turn on the stadiometer 4 Tur
3. BIOSPACE complies with the ISO9001 and ISO13485 which are international quality management systems This equipment satisfies the IEC60601 1 EN60601 1 an international safety standard for electronic medical equipment This equipment also satisfies the IEC60601 1 2 EN60601 1 2 an international standard for electromagnetic conformity A Exterior and Functions The following are the names and functions of each part of the InBody Please inspect each component of the InBody570 for damage prior to installation 1 Upper Part Status LED The LED emits a blue light when the InBody570 is turned on Hand electrode Examinee holds the hand electrode so that the 4 fingers wrap the surface of the bottom hand electrode while the thumb is placed on the oval electrode Hand electrode joint and handle Supports the hand electrode and encloses the wiring for the electrode O Body Connects the upper part of the equipment to the lower part LCD screen Shows each stage of the test instructions test results etc You can touch the screen to input the data required for the test configure settings or view test results Number keypad Used for inputting age height and other number based data 0 Text button Shows the virtual keyboard on the screen Used when inputting ID 0 Print button Used for printing the test results 0 Decimal point button Used for inputting the decimal point in ID height age or weight O Gender buttons
4. Used for selecting gender Male or Female 0 Enter button Used to finish inputting data or to save changes in Administrator Menu Delete button Used for deleting inputted data 0 5 Apogu se oo 18 Footplate 0 Front sole electrode The examinee makes contact with this electrode by stepping with the front part of their foot Rear sole electrode The examinee makes contact with this electrode by stepping with the heel of their foot Footplate This is connected to the scale which measures the examinee s weight O Level Indicator Indicates the current horizontal level of the InBody570 Joint Joins the upper part and lower part of the equipment together 0 Leveling screws Used for adjusting the horizontal level of the equipment 19 3 Rear Panel Speaker Provides audible indication for test in progress test complete and successfully saved setting changes O Rear cover Designed to be opened only by BIOSPACE service personnel for checking internal circuits etc 9 pin PC serial port Female RS 232C Used for connecting the InBody570 to Lookin Body installed on the computer The InBody570 can be connected to Lookin Body installed on a computer using one of the ports 0 or O Q 9 pin stadiometer serial port Female RS 232C Used for connecting the InBody570 to a stadiometer Only compatible with a BIOSPACE stadiomet
5. 35 2 L x 43 8 H inch 522 W x 893 L x 1113 H mm Equipment Weight 53 Ibs 24kg Test Time About 50 seconds Operation Environment 50 104 F 10 40 C 30 75 RH 70 106kPa Storage Environment 4 158 F 20 70 C 10 95 RH 50 106kPa No Condensation Testing Weight Range 22 0 551lbs 10 250kg Testing Age Range 3 99 years Testing Height Range 3 ft 1 4in 7ft 2 6in 95 220cm Specifications can be changed without a prior notice 23 www biospaceamerica com Copyright 1996 Biospace Co Ltd All rights reserved BM USA D3 B 130711 Neen
6. is centered Leveling the equipment is necessary for accurate measurement of weight There are a total of 5 leveling screws Level Indicator Leveling screws under the footplate Leveled Not leveled oe XX Em Leveling screws under the footplate A Caution Avoid injuring your hands when rotating the leveling screws under the footplate 4 Connect the adapter to the power input port which is located on the rear panel Connect the adapter to the power cord Plug the power cord into a grounded 3 socket outlet The InBody570 can be used in connection with other test equipment such as a stadiometer or data management software called Lookin Body For more information please refer to A Exterior and Functions 3 Rear Panel in section V Others in this User s Manual Power input port Rear panel Grounded outlet A Warning Do not place the InBody570 in a location making it difficult to disconnect the power cord Do not plug in or pull out the power cord with wet hands There is a risk of an electric shock Always use an outlet connected to the rated power AC 100 240 V Using other power rated outlets may result in fire or malfunction When using a power surge protector make sure that the outlet or the extension cable has adequate power capacity Do not disassemble or modify the equipment includi
7. s Garantie Keine berm ige Kraft auf das Ger t anweden B Safety Information Indicators 9 pin serial port Female RS 232C s LAN port Ethernet 10T Base USB port Safety Symbols Dangerous High Voltage Warning Caution B gt S BF Type Eguipment OC 42V 3 4A Adapter Power On 1 O Power Off Follow instructions for use Suivez les instructions d utilisation WARNING Electric shock hazard do not dismantle Dismantling will void the warranty AVERTISSEMENT Risque de choc lectrique ne pas d monter Le d montage annulera la garantie DANGER Do not use this equipment with electrical medical device such as a pacemaker Ne pas utiliser cet quipement avec des appareils m dicaux lectriques comme un stimulateur cardiaque 6 CAUTION Do not spray any liquid substance directly onto the device Ne pulverisez aucune substances liquids directement sur l appareil 6 CAUTION No excessive force on shoulder joint Ne pas appliquer de force excessive sur les bars articul s 21 C Classification Body Composition Analyzer of Direct Segmental Multi frequency Bioelectrical Impedance Analysis Method Type of protection against electric shock Class I Type of the applied parts BF Type Degree of protection against water infiltration IPXO EMC Immunity Level A EMC Emission Level A Equipment is not suitable for us
8. A D3 B 130711 InBodys70 User s Manual Contents I InBody570 Installation A Product Components B Operating Environment C Installation Instructions D Required Settings E Connecting Printer and Stadiometer F Maintenance II InBody Test A Precautionary Steps B Test Instructions C Test Posture HI Transportation and Storage A Cautions during Transportation B Repacking Instructions C Transportation and Storage Environment IV Frequently Asked Questions FAQ A Regarding the InBody B Regarding the InBody Test V Others A Exterior and Functions B Safety Information C Classification D Specifications O Onn A 11 12 13 14 14 14 15 17 18 21 22 22 I nBody570 Installation A Product Components The InBody570 consists of the following components Please make sure all of the following components are present Please inspect each component of the InBody570 for defects prior to installation 0 InBody570 Keypad Adapter DC 12V 3 4A 1 EA Power cord 1 EA User s Manual 1 EA Upper part Hand electrodes Foot electrodes Footplate B Operating Environment Please make sure that the environment is adequate for the InBody570 installation This equipment is designed for indoor use If installing outdoors the following requirements must be fulfill
9. Body Mass Index BMI and Impedance Values Height Optional Equipment Stadiometers from BIOSPACE Logo Name Address and Contact Information can be shown on the InBody Results Sheet Digital Results LCD Monitor Data management software Lookin Body Results Sheet InBody Results Sheet Voice Guidance Provides audible indication for test in progress test complete and successfully saved settings changes Database Test results can be saved if the member ID is utilized The InBody can save up to 100 000 results Test Mode Self Mode Professional Mode Administrator Menu Settings Configure settings and manage data Troubleshooting Additional information to help use the InBody570 USB Thumb Drive Copy backup or restore the InBody570 data data can be viewed on Excel or Lookin Body data management software 22 Applied Rating Current 400uA 40uA Adapter Manufacturer BridgePower Corp Model JMW 140KA1240F02 Power Input AC 100 240V 50 60Hz 1 2A Power Output DC 12V 3 4A Display Type 800 x 480 7inch Color TFT LCD Internal Interface Touchscreen Keypad External Interface RS 232C 4EA USB HOST 2EA USB SLAVE 1EA LAN Ethernet 10T 1EA Bluetooth 1EA Wi Fi 1EA Compatible Printer Laser Inkjet Printers A list of printers compatible with the InBody570 can be found at http www inbodyservice com Dimension 20 6 W x
10. d 2 this device Do not use this equipment with electrical medical devices must accept any interference received including interference such as a pacemaker that may cause undesired operations i ac USB HOST PRINTER STORAGE Ne pas utiliser cet quipement avec des appareils Follow instructions for use m dicaux lectriques comme un stimulateur cardiaque ON Suivez les instructions d utilisation Verwenden Sie dieses Ger t nicht mit elektrischen I 9 Medinische Ger ten wie z B Herzschrittmachem WARNING i i CAUTION H OFF Electric snock hazard do noraisments Do not spray any liquid substance directly onto the i Dismantling will void the warranty device INE JAA Ne pulverisez aucune substances liquids directement AVERTISSEMENT sur l appareil gt Be sure to use the adapter Risque de choc lectrique ne pas d monter sue keine fl ssigen Substanzen direkt auf offered by Biospace Only Le d montage annulera la garantie i ia CAUTION BEA HHL LAHJA let WARNUNG No excessive force on shoulder joint OAS ASSIA 2 Stromschlaggefahr nicht auseinandernehmen Ne pas appliquer de force excessive sur les bars Falls das Geh use ge ffnet wird erlischt die articul
11. e in the presence of flammable mixtures Specifications Bioelectrical Impedance Analysis BIA Measurement Items 15 Impedance Measurements by Using 3 Different Frequencies 5kHz 50kHz 500kHz at Each of 5 Segments Right Arm Left Arm Trunk Right Leg and Left Leg Bioelectrical Impedance Z Electrode Method Tetrapolar 8 Point Tactile Electrodes Measurement Method Direct Segmental Measurement Bioelectrical Impedance Analysis Method DSM BIA Body Composition Calculation Method No Empirical Estimation Outputs Intercellular Water Extracellular Water Total Body Water Dry Lean Mass Lean Body Mass Body Fat Mass Weight Skeletal Muscle Mass Body Mass Index Percent Body Fat Segmental Lean Analysis Right Arm Left Arm Trunk Right Leg Left Leg Segmental Fat Analysis Right Arm Left Arm Trunk Right Leg Left Leg ECW TBW Body Composition History Weight Skeletal Muscle Mass Percent Body Fat ECW TBW Basal Metabolic Rate Visceral Fat Level Impedance at each segment frequency Indications for use For use only in healthy subjects for Measurement of Estimated Skeletal Muscle Mass Extra Cellular Water ECW Intra Cellular Water ICW Total Body Water TBW ECW TBW Body Fat Mass Percentage of Body Fat PBF Body Lean Dry Lean Metabolic Rates Basal Metabolic Rates Segmental Lean Mass Segmental Fat Mass Segmental Body Fat Visceral Fat Level VFL Actual Weight
12. ed Temperature range 50 104 F 10 40 C Relative humidity 30 75 RH Atmospheric pressure 70 106 kPa C Installation Instructions 1 Open the packing box of the InBody570 and remove the packing pads Then take the InBody570 out of the box Packing pads F Caution If you have any problems installing your InBody570 please contact BIOSPACE for assistance Do not transport the equipment by holding the screen portion or the joints of the hand electrodes Keep the packing materials provided for repacking the equipment in the future Other wastes should be disposed of according to relevant laws and regulations 2 After raising the upper part of the InBody570 all the way remove the polyethylene foam cover A Caution Using the InBody570 on carpet may cause static electricity which could damage the equipment If installing the InBody570 on carpet is unavoidable please use an antistatic mat Install the InBody570 on a leveled non vibrating surface Installing the equipment on an uneven surface may cause the examinee to fall down Test results may also be inaccurate Never clean the hand and foot electrodes with liquid spray or detergent directly The equipment may corrode and or malfunction if the liquid or detergent leaks inside Use the InBody Tissue when cleaning the InBody570 3 Level the InBody570 by rotating the leveling screws under the footplate to the left and right so that the air bubble
13. er LAN port 10T Base Used for connecting the InBody570 to Lookin Body installed on a computer The InBody570 can be connected to Lookin Body installed on a computer using one of the ports O or O USB SLAVE port Used for connecting the InBody570 to Lookin Body installed on a computer The InBody570 can be connected to Lookin Body installed on a computer using one of the ports or O USB HOST port Used for connecting to a printer or a USB Thumb Drive Power input port Used for connecting to the power adapter Always use the specified adapter provided by BIOSPACE as it is a part of the InBody570 Using other adapters may result in malfunction of the InBody570 Power switch Used for turning on off the equipment InBody570 INPUT DC 12V 3 4A The following adapter must be used Model JMW140KA1240F02 Manufacturer BridgePower Corp Input Voltage AC 100 240V 50 60Hz 1 2A Output Voltage DC 12V 3 44 Neil E 113150 i Complies with UL60601 1 R C c us cSAC222No 601 1 0120 BIOSPACE tl Biospace Co Ltd 15 Heugam gil Ipjang myeon Seobuk gu Cheonan si Chungcheongnam do 331 824 KOREA TEL 82 41 581 3003 FAX 82 41 581 3103 Homepage www biospaceamerica com MADE IN KOREA This device complies with part 15 of the FCC Rules i DANGER Operation is subject to the following two conditions 1 this device may not cause harmful interference an
14. itch to turn on the InBody570 o Power switch USB HOST PRINTER STORAGE D Required Settings 1 The InBody570 automatically starts booting when it is turned on While booting it performs a self weight calibration x While booting about 5 minutes make sure there is nothing on top of the footplate Please do not stand on the footplate or place objects on the footplate T 2 Press the Administrator Menu button on the screen which appears when no one is on the footplate InBody Test Stand Barefoot to Begin Administrator Menu 3 Input the password default password 0000 to have access the Administrator Menu Administrator Menu Access If not the Administrator press Exit In the Administrator Menu manage saved test results or change settings Password 4 The Administrator Menu will give you access to many different options such as basic settings touch calibration Results Sheet management printing options and more Administrator Menu 4 ver 1 0 Settings Troubleshooting Exit 101 View Print Delete Test Results 03 Data Backup Restoration 04 Date Time Sound Unit Password 105 Self Mode Professional Mode Professional Mode 08 Bypass Age Gender AgelGender Required just 1234 us E Connecting Printer and Stadiometer 1 Printer In order to print InBody Results Sheets an InBody570 compatible printer is required
15. n in the results Socks InBody Test or stockings must be removed to obtain accurate data Is it okay to wear accessories The ideal condition for the analysis is simply standing with no clothes and jewelry watch rings etc or wearing no accessories However this may not always be possible metal objects while taking Therefore we recommend that the examinee remove as many clothing items the InBody Test and accessories that may affect the weight as possible e Who cannot take the Individuals with medical implant devices such as pacemakers or essential InBody Test or will have support devices such as patient monitoring systems must not use this difficulties taking the equipment The currents will flow through the body during the test which InBody Test may cause malfunctioning of the device or endanger lives Children amputees or the elderly may have trouble testing if they cannot hold the hand electrodes or stand still on the foot electrodes Can a person with metal The ideal test methodology is where the examinee does not wear anything implants in their body metallic Individuals with metallic implants may have skewed test results take the InBody Test due to the conductivity of the metal affecting the results As the weight of clothes and other wear affects the results of the body composition analysis it is strongly recommended to take off any heavy clothing or metallic wear Except for the weight jewelry does
16. n on the InBody570 If the stadiometer is connected to the InBody570 the stadiometer icon EU will appear on the top left corner of the screen when no one is on the footplate 10 F Maintenance A Caution Do not bend the handles of the hand electrodes or rotate them in the reverse direction beyond its limitation Do not extend the handles of the hand electrodes beyond its limitation Do not place any objects on the footplate Do not apply excessive force on the equipment Turn off the equipment if you are not using it for a day or longer Do not allow any liquid substances to contact the equipment directly Keep food and drinks away from the equipment Substances getting inside the equipment can cause critical damage to the electronic components Use a lint free cloth to gently wipe the external surface of the equipment about once every week Be careful not to scratch the LCD screen H InBody Test A Precautionary Steps A Warning Individuals with medical implant devices such as pacemakers or essential support devices such as patient monitoring systems must not use this equipment Safe low level currents will flow through the body during the test which may cause malfunctioning of the device or endanger lives Bioelectrical Impedance Analysis BIA uses safe low level currents which are not harmful to the body However we do not recommend pregnant women test Children and people with limited mobilit
17. ng internal parts without written consent from the manufacturer This may cause electric shock or injury product malfunction inaccurate results and will void the manufacturers warranty Do not directly contact the InBody570 with any other electronic device when the InBody570 is on This may result in an electric shock A Caution If the InBody570 is not plugged into a grounded outlet it may cause damage through electric surges or product malfunction This may affect the test results The test results may be inaccurate if the InBody570 is under electrical interference Do not install the InBody570 near products that generate electrical interference such as fluorescent lights large AC motor equipment treadmill vibration plate refrigerator air conditioner compressor etc high frequency thermal therapy equipments or heating appliances Do not share the power source of the InBody570 with other electrical devices This may affect the test results When connecting the InBody570 with other test equipment turn on the other equipment first When turning off other equipment turn off the InBody570 first This is necessary to minimize electrical surges on the InBody570 Always use the specified adapter provided by BIOSPACE as it is a part of the InBody570 Using other adapters may result in malfunction of the InBody570 Operation of the InBody570 6500 feet above sea level may affect the weight measurement 5 Flip the power sw
18. not effect the body composition analysis as the contact point with the InBody570 are the hands and feet have limited mobility It is impossible to test if an individual cannot maintain contact with the and cannot maintain proper hand or foot electrodes BIOSPACE has a line of products that conduct body posture for the InBody Test composition analysis on bed ridden examinees that allow the patients to How can I still be tested stay in bed For more information please contact BIOSPACE Is the electric current The physiological electric impedance method uses safe low level currents harmful to the body that is not harmful to the body The safety of the InBody has been tested and proven The InBody products have been approved for medical use by the CE and all over the world Many medical institutions around the world are actively using the InBody How often should I take Individuals who are undergoing any programs that may affect their body the InBody Test composition are strongly recommended to have the InBody Test done every two to four weeks Consistent testing will allow individuals to track and monitor their progress over time e What are the precautionary Please refer to A Precautionary Steps in section IL InBody Test in this steps to ensure accuracy of User s Manual the InBody Test 17 V Others The InBody570 is manufactured according to the quality management procedure of BIOSPACE
19. ompletely plugged into the adapter Insert the power cord completely into the adapter The problem may occur if you are using the power adapter that is not provide by BIOSPACE Always connect a power adapter DC 12V 3 4 A provided by BIOSPACE Power adapter Power cord 15 16 Ouestion Answer e My touchscreen is inaccurate or not responsive Calibrate the touchscreen under Settings of the Administrator Menu 23 Touchscreen Alignment Press firmly to optimize touchscreen response If you cannot enter the Administrator Menu due to touchscreen problems please restart the InBody The InBody stores the last touchscreen alignment and will automatically recall the previous touchscreen settings The InBody can also recognize if its touchscreen alignment is off screen and will automatically take the user to the calibration screen after restarting I would like to connect other equipment to the InBody570 Please refer to E Connecting Printer and Stadiometer in section I InBody570 Installation in this User s Manual B Regarding the InBody Test Some of the more common clinical questions are answered below If additional questions or more clarification is desired please contact BIOSPACE Question Answer e Must socks or stockings Bare skin contact is essential in the analysis using the BIA method Socks be removed for the or stockings may cause a varying degree of distortio
20. or if the examinee has too many calluses Avoid contact with the examinee during testing Contact may lead to interference affecting test results 11 B Test Instructions 1 Step on the footplate when the screen below is shown The screens vary according to the Settings of the Administrator Menu 05 Self Mode Professional Mode Professional Mode An examiner is present and guiding the examinee through the InBody Test Self Mode The examinee takes the InBody Test following the instructions that are displayed on screen InBody Test Stand Barefoot to Begin 2 Weight measurement begins Measuring your weight m 178 5 us 3 Input personal information Input height only if using Self Mode Height Age Gender 511 9 lus nD 1 1 0 a aeo alas 2977 0eme 0 Professional Mode Self Mode 4 Maintain proper posture to take the test Refer to C Test Posture for the proper posture 5 The InBody Test begins InBody Information is shown if using Self Mode 0123456789 S119 24 M Yv Viiart 1785 A SMM hae 564 Maintain Yv a eed i a balanced body composition Y a ECWTEM fa 0324 i I to stay healthy 5 Professional Mode Self Mode 12 6 When the test is completed the results will be shown on screen 0123456789 s 119 2 M 5119 Weight a Weight 178 5 Press to view detailed SMM nat Cois 74 r
21. ss SMM Me 74 1 lbs ee X v jai Gag Fs 26 2 retorta 26 2 l SES TRUE ENT mmm 0 373 7 0 f 0 373 d 1 ECW TEW 0 379 Professional Mode Self Mode C Test Posture The examinee must maintain proper posture to have accurate test results The test will proceed when there is good electrical contact Your arms must not touch the sides of your body Keep the arms straight The thighs should not touch Hold the hand electrode so that the 4 fingers wrap the surface of the bottom hand electrode and place the thumb on the oval electrode Place the heels on the rear sole Step on the footplate barefoot electrodes 13 IL Transportation and Storage A Cautions during Transportation When transporting have two people keep the InBody570 parallel to the ground B Repacking Instructions Once the InBody570 is installed avoid transporting the equipment If it must be transported repack it in the following sequence 1 Turn off the InBody570 2 Separate the connected adapter cords and cables from the equipment and cover the InBody570 with the polyethylene foam cover After covering the InBody lower the upper part 3 Put the InBody into the packing box Place the packing pad over the equipment and tape up the packing box Polyethylene foam cover Packing pads A Caution Always use the protective packing materials provided by BIOSPACE
22. when repacking C Transportation and Storage Environment The InBody570 should be transported or stored under the following conditions Temperature range 4 158 F 20 70C Relative humidity 10 95 RH Atmospheric pressure 50 106 kPa No Condensation 14 IV Frequently Asked Questions FAQ Even if no problems arise from the equipment users may still have many questions especially regarding clinical procedures Few common questions and answers are listed below If your questions are not answered here please contact BIOSPACE x Customer contact information can be found under Settings of the Administrator Menu 24 Customer Service Information A Regarding the InBody If a problem arises with the InBody570 you may first attempt to check the Troubleshooting in the Administrator Menu The InBody570 can help you diagnose and solve some problems If your problem cannot be resolved through the Troubleshooting please refer to the possible solutions below Question Answer My InBody570 does not Insert the power plug completely into a grounded 3 socket outlet turn on Grounded 3 socket outlet When using a power surge protector the equipment may not power on if the power switch on the power surge protector is turned off Check the power surge protector which the power plug is connected to Power surge protector power switch The problem may occur if the power cord is not c
23. y should be supervised or assisted when attempting to test on the InBody After an individual with any kind of contagious disease or infection tests on the InBody use an InBody Tissue to clean the equipment A Caution Stand upright for about 5 minutes before testing Taking the test immediately after lying in bed or sitting for a long period of time might result in a slight change in the test results This is because body water tends to move to the lower body as soon as the person stands or gets up Do not eat before testing In cases where the examinee has already eaten the test should be put off for at least two hours after the meal This is because food mass is included in the examinee s weight and thus may result in measurement errors Use the bathroom before testing Waste is not included in the body s compositional elements but the volume of urine and excrement is included in the weight measurement affecting accuracy of the test results Do not exercise before testing Strenuous exercise or sharp movements can cause temporary changes in body composition Even light exercise can change your body composition temporarily Take the test in the morning if possible Body water tends to gravitate towards the lower body throughout the day affecting accuracy of the test results Thoroughly wipe the palms and soles with the InBody Tissue before testing Testing may be difficult if the examinee s palms and soles are too dry