AFM USER MANUAL

image

Contents

1. AFM USER MANUAL UALR Nanotechnology Center John Meredith Jacob A Bice and Fumiya Watanabe July 2009 Turn on screens and open Nanoscope SPM program from desktop Click on warning to OK Select Real Time Mode by clicking this icon on the top left side of the menu bar You should see laser turn on a Set parameters in program ScanContais EL OtherConrols Scan size 1 00 pm Microscope mode Tapping Aspect ratio B k Controls Z limit 4 712 um x offset Main FM igain Y offset 0 00 nm 2 Modulation Enabled FM pgain Scan ange 0 002 j Integral gain 0 5000 Pf iumination Scan rate 1938h Proportional gain Units Metric Tip velocity 3 96 um s LookAhead gain Color table S amples line sas Amplitude setpoint 1 287 y Engage Setpoint Lines sas Drive frequency Serial number Slow scan axis Enabled Drive amplitude 289 9 m Mi a oan sio Cert St caer Data type Height Data type Amplitude Data scale 200 0rm Data scale osoov Line direction Trace Line direction Retrace Scan line Scan line Realtime planefit Line Realtime planefit Line Offline planefit Offset Offline planefit Ofset Highpass filter sot Highpass filter sort Lowpass filter or Lowpass filter or b Make sure table is floating by checking that it is not contacting the legs Page 2 of 9 July 2009 2 MOUNT CANTALEVER on the Holder a Place cantilever holder on stand
2. IMAGE w Once you have a good image that you would like to save you must capture the image a Go to the Capture menu above the toolbar and select Capture Filename b Change the capture filename to the name of your sample using the prompt that comes up and select ok A new number will be appended for each captured image on that sample so there will be no need to change the filename until you change samples c Begin a scan from the bottom or top after adjusting your scan parameters to the desired settings Ej Up El Down The capture does not usually begin in the middle of the scan but you can force the capture by clicking the Capture yellow camera icon twice d Click the yellow capture button shown on the toolbar 12 ADJUSTING SCAN a There are options at the top of the image viewing window typically displayed on the right screen These include Zoom Offset and Execute b When you click Zoom your cursor will then become a moveable box which you can move over an area of interest on the scan A left click will allow you to adjust the size of the box by moving your mouse A right click will then leave the box on the image while allowing you to use your cursor to hit the Execute button This will change the scan size and position so that only the area in the box is now being scanned Remember to reduce the scan rate if you are going to a larger scan area c When you click Offset yo
3. and focus under microscope so that you can see where the probe cantilever will be placed b Press down clamp and move the metal clip back so you can place the probe cantilever on the probe cantilever holder See Fig 7 3b from 3100D manual below c Place cantilever into probe cantilever holder and center it in the slot Do not push it too far back This will result in an undesired mounting angle for the probe cantilever d Hold down clip and move over probe cantilever making sure the body of the probe cantilever does not move Come down on the location as shown on figure below slightly left of the vertical line on the probe cantilever e Verify that the probe cantilever is centered Figure 7 3b Cantilever Holder Stand top view Press here on clamp Figure 7 3c Standard AFM Cantilever Holder SIDE VIEW tip is installed z Cantilever Probe Tip Note You may install the AFM cantilever tip faces down i l i Spring Loaded Probe Clip holder in only one orientation because the pins giectrical Mounting Sockets are asymmetrical 4 pies Probe Cantilever BOTTOM VIEW TIP SIDE Metal Clamp 0 tp Clip Page 3 of 9 July 2009 X Direction Detector Alignment Direction Detector Alignment Scanning Head Screw 3 MOUNT CANTALEVER HOLDER Do not drop the scanner Piezo is very fragile a Tighten scanning head screw on right side of scanning head b Remove cantilever holder from stand and hold i
4. data CROSS SECTIONAL ANALYSIS This will allow you to measure the dimensions of features on the surface You start by making a line across the image then use the two cursers to measure the topography along that line ROUGHNESS ANALYSIS Tells you how rough the surface is This gives a measurement of the deflection in the Z direction Page 9 of 9
5. e button 2 then move the mouse over the abort button in case something goes wrong If something seems to not be right hit the abort button and try again c Once the image begins to appear press the scope trace button Atd Page 7 of 9 July 2009 d Select the amplitude setpoint and use the right arrow key to move the value up until the scope trace lines are no longer together This means the tip is no longer contacting the surface See the Feedback Control Panel above e Then using the left arrow key move the amplitude value down until the tip contacts the surface again Once the lines are tracking each other well move the amplitude setpoint value down two more clicks by pressing the left arrow key twice f Adjust integral gain Increase this gain by right arrow key until you see the unstable noise sets in g Adjust proportional gain Usually this gain is 30 100 of the integral gain value h Press the eyeball button to go back to the image mode i If you want to scan a larger area lower the scan rate first then increase the range If you want to scan smaller area reduce the size first and increase the rate if needed See the Training Notebook for scan rate recommended for scan sizes Scan Control Panel TIP If the scope will not align properly try adjusting the integral and proportional gain As you move one move the other the same amount If you go one way and the scope gets worse try to go the other way 11 CAPTURING
6. he tuning fork shaped button and the press AUTOTUNE b The sweep should look close to the picture below and the curves should have that shape Figure 11 4a Cantilever Tune Control Panel Auto Tune Controls Start frequency 0 000000 kHz Target amplitude End frequency 500 000 kHz Peak offset Minimum Q 0 00 Smash width factor 5 00 Auto Tune Back to Image Mode TIP If you receive an error while tuning the tip is most likely not aligned correctly and you should go back to step 4 and try again 10 TAKING AN IMAGE Before approaching the surface make sure the following 1 Scan Size lum 2 Scan Rate 2 Hz 3 Integral Gain 0 5 4 Proportional Gain 0 7 5 In addition set X amp Y Offsets 0 Scan Angle 0 Scen Controls Scan size 1 00 pm Aspect ratio Feedback Controls X offset 0 00 nm jf Main Y offset 0 00 nm Z Modulation Enabled Scan angle 0 00 2 Integral gain 0 5000 Scan rate 1 98 Hz Proportional gain Tip velocity 3 96 pms Look head gain Samples line 256 Amplitude setpoint 1 287 V Lines 256 Drive frequency 274 954 kHz Slow scan axis Enabled Drive amplitude 289 9 mY Scan Control amp Feedback Control Panel a Press the focus surface button again and move the crosshair to a point of interest and hit ok in the focus surface window if you have not done so already When you are ready click OK to exit b Press the green approach surfac
7. n your right hand keeping in mind how you will have to place it on the scanning head c Carefully lift the scanning head only touching the sides for the head and align the four pins with the four holes in the cantilever holder d Carefully press the cantilever holder onto the four pins It only fits one way with the tip facing towards the camera when the scanning head is mounted Excessive force can destroy the piezo also 4 FIND TIP WITH LASER Do not drop the scanner Piezo is very fragile Laser aiming screws Laser reflection window fo N A Figure 7 3e Dimension Head Laser Control Knobs Rear Right Laser Control Knob X Direction J _ Piezo tube scanner hese TrakScan tracking lens RR _ Removable cantilever holder Knob Adjustment Diagram Front Left Laser Control Knob Y Direction a While holding the scanning head adjust the knobs laser aiming screws on top of the head to find the tip b Sweep the top left knob back and forth to adjust the x direction and find the edge where the laser is blocked Page 4 of 9 July 2009 c When you find the edge where the laser is blocked move the bottom right knob the y direction 1 20 of a turn clockwise sweep in the x direction again d Repeat above two steps until you find a small shadow in the middle of the x direction sweep This shadow is the cantilever tip e If you have found the tip any slight change in the x direction should cause the shadow
8. to disappear and the sum on the bottom of the right screen should disappear f Maximize the sum with the x direction knob Turn the y direction knob counterclockwise until the sum goes away and you are off the tip Then turn back clockwise and maximize the sum Lastly reduce the sum to 95 by turning the y direction knob back counterclockwise Shadow of Tip on Laser 5 REPLACE SCANNER HEAD a Carefully place the scanner head back into its slot on the microscope making sure to not let go until the head is below the line on the mounting bracket Again a slight impact can shutter the piezo b Loosen scanning head screw until textured part of screw and plastic washer do not touch 6 ADJUST PHOTODETECTER a Using knobs on side of scanner head mirror adjusting screws align the red dot to the center of the detector box The bottom knob moves in the x direction and the top knob moves in the y direction b The vertical deflection should be within 02 and 02 when the laser is properly aligned Sum intensity from the photodetector Page 5 of 9 July 2009 Toolbar Buttons Restart Scan from the Move the tip up 2 mm be back te bottom ofthe image E focus and Move on Surface Tuning View image viewer focus on Tip Approach Surface Scope Trace Capture Restart Scan from the top of Abort Capture the image 7 LOCATE AND FOCUS TIP K a Press yellow focus tip button and wait for the tip to stop moving in the vision s
9. ur cursor will become crosshairs that you can move over the image and finalize the position of with a right click When you hit Execute the image parameters will remain the same but the image will be centered at the point designated by the crosshairs 13 When you are done correcting images at this location click Disengage icon To move to a new location you must open the Locate Surface icon again Remember you must re approach with 1 micron 2Hz if you replace the sample Page 8 of 9 July 2009 14 ANALYSIS a Once you have captured the image you go to the image file handler to look at the image This button is on the top right of the window on the left screen Can return to Real Time mode by this con a First select the image that you would like to analyze c Then go to Image gt Select Left Image This will choose the image that is the actual height data Roughness Analysis Cross Sectional Analysis Flatten FLATTEN Flatten will give you the image that you first saw when you scanned This is necessary to get an accurate representation of the surface Do this before any other analysis You will get a warning that you are about to modify the data Click ok and you will NOT SAVE at the end Tip You can draw rectangles to exclude some regions from this routine This is often improves the flattened images 3D IMAGE This gives you a three dimensional representation of the surface This will help you interpret some
10. ystem window Using the ball hold down the focus button and focus the tip c Using the knobs of the side of the camera place the crosshair on the tip slightly behind the very end of the tip 8 LOCATE AND FOCUS SURFACE Fai a Place sample over vacuum chuck and turn on vacuum b Make sure there is enough clearance between the sample and the cantilever holder by moving the scanner head up with the red up arrow button c Rotate the stage so the sample is under the scanning head d Press the red focus and move surface button WARNING As you focus down on the surface the scanning head is moving down Do not allow the tip to touch the surface This could cause serious damage to the microscope which you will be responsible for e While holding the focus button on the ball and move the ball toward you As things come into focus let go of the focus button and move the ball so the stage moves if the focused material does not move with the track ball movement it is dust on the lens and you should continue to focus moving the ball towards you until you find something that moves when you move the stage TIP if your sample is reflective or clear it may be a good idea to move to the edge of your slide and focus on the top of the edge of the slide then move to where the sample is f Place crosshair on point of interest by moving this point by track ball while not holding focus button 9 TUNE THE TIP Page 6 of 9 July 2009 a Press t