Contents
1. alarms and alerts type of display type of infusion fluids Nurse call facility Computer interface Mounting method Protection against fluid ingress Electrical safety classification Model Identification Serial number e Software version number MHRA evaluation 03141 April 2004 PRODUCT DATA ARGUS Medical AG Aarestrasse 13 CH 3627 Heimberg Switzerland Yes CE 0120 Switzerland 1450 100mm x 320mm x 180mm 3 2 Kg 2 6 Kg with lower energy battery 230 VAC Mains supply and external DC supply 10 15 V DC internal rechargeable batteries sealed lead acid battery 6V 4 Ah optional 6V 1 2 Ah 8 hours at 5 ml h optional 2 5 hours at 5 ml h 15 hours Lead screw driven by maintenance free stepper motor 10 60 ml syringes from wide range of manufacturers Universal clamp for IV pole included Optional rail clamps and fast click rail clamp Docking station 0 1 to 999 9 ml h 0 1 ml h over complete range Drug library available none available 0 1 999 9 ml 0 1ml increments Yes 0 999 9 ml Stop 2 ml h for rates gt 10 ml h 1 ml h or set rate if lower for rates below 10 ml h 2 infusion delivery time gt 1 hour volume infused gt 2 ml at all flow rates Programmable 200 1200 mbar in 100 mbar increments Standby end of infusion syringe near empty syringe empty syringe not in correct position syringe plunger is not clamped the fitted syringe does not corre
2. The pump can be used on a bench or mounted on a vertical pole or horizontal pole The syringe is held horizontally and fluid is delivered from right to left as is typical thus promoting intuitive handling and syringe loading The lack of integral carry handle can make the pump a little unwieldy to hold An optional pole clamp with carry handle can be purchased to alleviate this problem although this prevents stacking in the docking station The shape of the top surface of the pump casing facilitates holding the pump in one hand The standard pole clamp is straightforward to use MHRA evaluation 03141 April 2004 Ergonomic Assessment Other pole clamp options facilitate fixing the pump to horizontal rails of various sizes These were not assessed during this evaluation The pump is normally powered by a standard external AC source 230 VAC 10 15 50 60 Hz 9 Watt An external DC power source can also be used In the absence or failure of external power an internal sealed lead acid battery automatically takes over the supply No audible signal is given if this occurs The cable connects into the back of the pump but is not captive and in theory can be accidentally dislodged although this did not occur during the period of evaluation BIME Note A captive cable prevents accidental dislodging however it could also have the drawbacks that a the infusion pole may be pulled over and b the cable itself can cause an unnece
3. 600 D MHRA evaluation 03141 April 2004 Technical Assessment Volume to be Infused Actual 0 97 ml in 60m 02s 24 7 ml in 59m 48s Target 1 ml in 60 minutes 25 ml in 60 minutes These results show a high degree of accuracy when delivering pre set volumes of fluid The ability to deliver 1ml accurately over a one hour period is rare among syringe pumps as it is dependent on a fast start up time The rapid start up typically achieved by this pump when using a Codan syringe contributes to this good performance Back pressure Bolus when back Back pressure Accuracy pressure changes 0 mmHg 0 3 Not applicable 100 mmHg 0 9 0 35 ml 100 mmHg 0 3 0 43 ml Resumption Time 2 5 minutes The pump was set to deliver at 5 ml h with the syringe positioned at the level of the cannula outlet After two hours the pump was raised to generate an effective drop in back pressure of 100 mmHg then later lowered to increase backpressure by 100 mmHg from the starting point The pump was not stopped during repositioning Figure 5 shows that a bolus is delivered on raising the pump and the flow pattern is altered When the pump is subsequently lowered flow is interrupted and the flow pattern is further altered No significant effect on long term flow accuracy was caused by the movement of the pump The fluid shift on raising or lowering the pump is slightly greater than has be
4. V Good Good Adequate Inadequate Poor 7a The noise level whilst running 5 4 3 2 1 No comment was Additional comments 7 6 The running indicator clarity 5 4 3 2 1 No comment was Additional comments 7c The display of rate VTBI 5 4 3 2 1 No comment during an infusion was Additional comments 7d The alarm messages were 5 4 3 2 1 No comment Additional comments 7e The alarm tone was 5 4 3 2 1 No comment Additional comments 7f The appropriateness of alarms 5 4 3 2 1 No comment were Additional comments 7g The safeguards against 5 4 3 2 1 No comment tampering were Additional comments 8 Infusion Complete V Good Good Adequate Inadequate Poor 8a The ease of cleaning was 5 4 3 2 1 No comment Additional comments 8b The ease of charging was 5 4 3 2 1 No comment Additional comments ice i General Questions 9a List any features that you would like to see which are not present on this pump 9b List any features that are present but which are unsuitable for your application 9c Are there any additional comments that you would like to make on your experience of using this pump Thank you very much See page one for details on returning completed questionnaires 19 About MHRA evaluation reports MEDICINES AND HEALTHCARE PRODUCTS REGULATORY AGENCY MHRA The M
5. an infusion is running No dose limiting software is offered to restrict delivery of labelled drugs within safe margins but Codan will shortly be launching a fluid management system associated with the docking station that incorporates this facility History logging An event log is provided allowing more than 200 programming events to be stored in the memory of the pump Communications It is possible to configure the pump by RS232 connection to a computer Configurations can also be downloaded to the PC and transferred to subsequent pumps from the PC This makes setting up a library of similarly configured pumps relatively simple The same port can be used as a DC power input socket Monitoring of the pump via this interface is offered Communication with Patient Data Management Systems Datex or Ohmeda etc is also possible USER ASSESSMENT To provide a user assessment of the Codan green stream Argus 600 D syringe pump two Trusts were consulted Each Trust had bought significant numbers of pumps and had been using them for at least 4 months Questionnaires were distrib uted to clinicians in each Trust who used the device regularly in their normal role see Appendix 1 on page 16 for questionnaire details In addition to the questionnaires an interview was held at CODAN green stream SY P ARGUS 600 D User Assessment each Trust with appropriate members of staff to gather overall impressions of the device The results of
6. anomalous behaviour such as a pressure falling below atmospheric is noted Pressure at alarm is also recorded for one of these tests as a rough check of the manufacturer s claims and correct functionality FLUID SIDE OCCLUSION RESPONSE The response of the pump to an upstream occlusion is observed and recorded whether or not an alarm system is provided The response is assessed for patient safety as examples tubing rupture or air entry represent significant hazards The time to alarm is recorded 3 times at each test flow rate Tests are performed at 1 ml h and maximum rate on volumetric pumps only AIR IN LINE DETECTION Significant uncertainties exist in the testing of this feature The air detector sensitivity is set to 50 mcl or minimum if greater Three con secutive test volumes of air are introduced immediately up stream of the infusion pump The infusion pump should alarm for each test vol ume Tests are performed at 125 mi h If the test is failed a larger vol ume of air is injected until a pass is obtained The test is for function ality rather than sensitivity SYRINGE RECOGNITION SYSTEM The syringe pump is tested systematically to find the worst case of incorrect recognition of syringe size and a flow delivery accuracy test is run at 5 ml h to estimate the worst case error that this user error can cause All available syringes recommended for use with the pump are also tested to assess the response to misloading of the
7. barrel plunger or ears LINE TUGGING TEST The delivery line is tugged sharply in a variety of directions and poten tial hazards assessed TESTING OF DEVICE ERGONOMICS Most Adverse Incidents are eventually denoted as user error Good design can contribute to minimising user error Work has been initiat ed to develop a formal ergonomic testing procedure for application to all devices At present User instructions are assessed for clarity and readability conciseness indexing Procedures for using the device are systematically worked through on the bench after other testing is completed Any hazardous potential misuse is noted For all available alarms the reliability readability of text displayed the alarm tone quality the positioning of alarm lights and methods of silencing alarms are tested as part of the ergonomic assessment of the device MHRA evaluation 03141 April 2004 Manufacturer s Data Manufacturer CE Marking on Product Country of origin manufacture Price ex VAT Size H x W x D Weight Power Supply Battery Operation battery capacity battery charging facilities Facilities pumping mechanism syringes accepted accessories flow rate range e flow rate increments programmable infusions drug dosing protocols volume to be infused facility e volume infused indicator e KVO rate claimed accuracy occlusion pressure
8. easily fail to correct this if a different brand is used The set up instructions in the manual supplied with the pump emphasise that both syringe size and brand must be checked by the user before starting an infusion The greatest possible delivery error found for this device was the possibility of installing a 60ml Terumo syringe and registering it as a 60ml Monoject syringe This arrangement yielded a delivery error of 18 8 in a flow test at 5 ml h This is a new test and the result cannot therefore be compared to other devices The results of this test indicate the benefit of having just one available syringe for use in a Trust with this or any syringe pump This largely precludes this type of error from occurring Manufacturer s Comment Line tug testing No parts are provided on the pump to secure the delivery line and thereby reduce vulnerability to tugging on the line However the design of the syringe retaining parts appears to render the pump unusually resistant to dislodging of the syringe by line tugging Strong tugging in a direction that tends to pull the syringe forward does trigger a syringe misload alarm due to movement of the barrel clamp but it was not possible to pull the syringe free of its retaining parts by line tugging There does seem to be confusion over the start up time which Codan perceives as being good and the report confirms this The rapid start up typically achieved by this pump contributes to thi
9. h Figure 6 Time to alarm after occlusion CODAN green stream SY P ARGUS 600 D Technical Assessment 11 syringes tested approximately in proportion to the syringe size as would be expected This can be explained by the faster plunger travel at a given flow rate for a smaller syringe Using a CODAN 50mlI syringe and the recommended narrow bore CODAN extension set the time to alarm and the post occlusion bolus are typical for a current generation syringe pump see Figures 6 and 7 A back off system is provided whereby the syringe plunger is drawn back fractionally when an occlusion is detected to reduce the size of the post occlusion bolus This is a safety feature and pressure monitoring of the pump during occlusion testing verified that it operated successfully When the back off system operates the infused volume ml inf is decremented so that it more closely indicates the actual volume of fluid delivered to the patient The back off appears less pronounced than on some pumps which possess this feature There is a danger however that an over active back off system may generate negative boluses whereby blood is actually drawn back into the delivery line Importantly no negative post occlusion boluses suck back were seen in any tests on this pump Syringe fit An alarm is provided to alert the user to a misloaded syringe though it does not differentiate the specific location of the problem A serie
10. press or automatic user pre programs the pump to deliver a particular volume on demand The option to deliver automatic boluses may be turned off in the pump configuration BIME Note Since the manual technique of delivering a bolus is dependent on user skill whereas the pre set automatic technique depends only on the pump software the automatic method was used to evaluate bolus volume accuracy and should be preferred in clinical practice where accuracy of delivery is required KVO Keep Vein Open delivery options are available The pump supplied for evaluation had conventional KVO settings so that KVO was activated on completion of a VTBI This feature may be configured to deliver at KVO rate any time the STOP key is pressed which would seem CODAN green stream SY P ARGUS 600 D Ergonomic Assessment counter intuitive and is probably not a configuration that would be used much in UK The importance of thorough training is also demonstrated if some programming steps are not performed sufficiently quickly During bolus deliv ery the pump display reverts to a default if no response from the user occurs within a given time The time for this fall back feature is config urable from 5 to 30 seconds The original display then has to be restored and programming contin ued It is possible to disable the pump programming keys with a Data lock facility if configured active The Data lock feature is provided to preven
11. 13 Methodology Where the equipment is intended to administer a discrete bolus vol ume to the patient the accuracy of this feature is measured To test lockout facilities on PCA pumps a 5 minute lockout time is set A bolus is demanded and then subsequent demands are made periodi cally to establish when the next bolus is permitted 0 1ml 1 0ml and 5 0ml target bolus volumes are tested for all pump types BODY TEMPERATURE TEST A 2 hour flow test at 1ml h is performed with the pump and delivery fluid contained within an incubator at a temperature of 37 C PATIENT SIDE OCCLUSION ALARM RESPONSE Time to alarm and bolus on release of occlusion are measured at 1 5 and 25 ml h for syringe pumps 1 5 and 125 ml h for volumetric pumps Pressure at alarm is not measured during these tests Pressure per se is not considered a hazard to the patient The 1 ml h tests using the main recommended disposable are repeated 5 times each for three nominally identical disposables For PCA pumps the number of attempted bolus deliveries before alarm is recorded and the bolus delivered on release of the occlu sion BACKOFF Pumps commonly have a facility to run the pumping mechanism back wards on occurrence of an occlusion alarm Line pressure is recorded at one second intervals for an occlusion response at 1ml h and 5ml h A graphical display of the pressure variation over time is produced to indicate the operation of a backoff system Any
12. HS Description KEY 1 Operation Mode Indicator 2 Infusion Rate Display 3 Infusion Volumes display 4 Pressure display 5 Alarm indicator 6 Alarm cause indicator 7 On off key 8 Rate keys 9 MODE key 10 Start stop key Figure 1 Front view of pump 11 Bolus Prime key 12 Syringe barrel holder 13 Syringe ear guide 14 Drive head 15 Syringe plunger retaining flippers 16 Clutch lever 17 Pole clamp 18 Interface for RS232 or External 12V DC 19 Equipotential plug 20 Staff alert socket 21 AC power socket Figure 2 Rear view of pump MHRA evaluation 03141 April 2004 2 CODAN green stream SY P ARGUS 600 D CONTENTS Description Ergonomic Assessment User Assessment Technical Assessment Manufacturer s Comments Methodology Manufacturer s Data User Assessment Questionnaire About MHRA reports DESCRIPTION Reference is made to the Photographs Figures 1 and 2 on page 2 The operation mode indicator 1 uses icons to display the pump s current power source Mains or battery and also to show if the pump is infusing or is delivering in KVO keep vein open mode One LED display 2 shows set flow rate and another 3 displays volume infused or other parameters dependent on the configuration This icon is labelled ml inf If reconfigured to display volume remaining to be infused confusion can be caused as the label remains as ml inf The default configuration of the pump ensures t
13. ation also invalidate parts of the operating manual and instructions on the top of the pump Manuals A 27 page operations manual is provid ed with the pump giving instructions for its use There is a table of contents but no index Programming steps are described by combining text with diagrams showing typical display appearance Additional brief instructions are print ed on top of the pump CODAN green stream SY P ARGUS 600 D The instructions provided in the operations manual for setting up and using the pump are sufficient for basic infusions but the exact functions of the more advanced programming features such as automatic bolus delivery are difficult to understand Some programming requires many steps and the instructions then become confusing The brief instructions on top of the pump are very abbreviated They do give some useful guidance on the meanings of possible alarm signals A service manual is also provided It provides a list of all the configurable options and the means of accessing them Instructions are also given for viewing either the pump history or configuration by downloading it onto a computer This manual would not be available to ward staff ensuring that access to the configuration can be limited to technical staff only PCA Features The pump is not designed for PCA use Drug library It is possible to label an infusion to allow the user to check the name of the medication being delivered while
14. chasers should satisfy themselves with respect to any modifications that might be made to the product type after MHRA s evaluation The reports are intended to supplement not replace information already available to prospective purchasers CROWN COPYRIGHT 2004 Apart from any fair dealing for the purpose of research or private study or criticism or review as permitted under the Copyright Designs amp Patents Act 1988 this publication may only be reproduced stored or transmitted in any form or by any means with the prior permission in writing of the Controller of Her Majesty s Stationery Office HMSO Enquiries about reproduction should be made to the MHRA at the above address MHRA evaluation 03141 April 2004 20 CODAN green stream SY P ARGUS 600 D ISBN NUMBER 1 84182 8041
15. communications or connection to an external 12V DC power supply 18 an equipotential plug 19 a socket for a nurse call system 20 and a mains power input socket 21 Stacking Argus 600 syringe pumps can be stacked in close proximity to one another on a pole and a docking station is also available which allows power and communication with up to five pumps Interference The pump complies with IEC 60601 1 2 The manufacturer makes no additional claims for this pump regarding electromagnetic compatibility EMC or suitability for use in MRI rooms Costs The list price for the pump is 1450 The user manual states that the pump has been optimised for use with CODAN syringes The list price for Codan 50 60ml syringes is 0 29 minimum order 300 units The list price for the Codan extension sets used in this evaluation part no 71 4349 is 0 60 box size 100 All prices exclude VAT ERGONOMIC ASSESSMENT The Codan green stream SY P ARGUS 600 D is a syringe pump targeted at general ward use Users intending to purchase the SY P 600D for Intensive Care use might also consider purchasing the docking station QUICK 100M whose added functionality includes patient weight dependent dosing and bar code matching of patient and drug Format Size is typical and weight above average for a modern ward based syringe pump 2 9 Kg 30 x 10 x 20 cm The optional short life battery 2 5 hours can reduce the weight by 0 6 Kg
16. configuration is implemented by a numerical system defined within the technical manual Configuration cannot be changed without access to a technician s code which complies with MHRA advice that configuration changes should only be made by technical staff and that all pumps in one clinical area should be configured identically During configuration changes there is no explanatory text on the pump display This has the effect that it is impossible to change configuration without reference to the manual Configuration changes can only therefore be made by appropri ate personnel but the system can also make checking configuration changes difficult The degree of configurability is extraordinarily high in order to permit sale to different European markets The default configuration can be set at purchase to a UK norm Changes from this config uration are possible however and could lead to counter intuitive behaviour of the pump Particular care should therefore be exercised in implement ing configuration changes on this highly flexible pump Examples of behaviour that can be configured which may be potentially confusing are e KVO is activated when the STOP key is pressed e Pump can be configured to enable restarting without requiring an alarm cancellation provided the cause of alarm is rectified ml inf can be configured to read either volume infused or volume remaining to be infused Some changes of configur
17. edical Devices Agency has merged with the Medicines Control Agency to form the Medicines and Healthcare products Regulatory Agency http www mhra gov uk ACKNOWLEDGMENTS This report was prepared by Craig Davey of the Bath Institute of Medical Engineering BIME Wolfson Centre Royal United Hospital Combe Park BA1 3NG Enquiries to Teresa Dunn Evaluation Centre Manager Bath Institute of Medical Engineering BIME Wolfson Centre Royal United Hospital Combe Park BA1 3NG Tel 01225 824103 or e mail t s dunn bath ac uk Martin Glasspool Programme Manager MHRA Hannibal House Elephant and Castle London SE1 6TQ e mail martin glasspool mhra gsi gov uk We would like to thank the pump manufacturer for their co operation in providing a sample of their pump for the period of the evaluation and clinical users who gave their time to evaluate the pump and comment on their findings DISTRIBUTION OF THIS REPORT This report should be distributed to the following departments Intensive Care Anaesthetic and Maternity Units Neonatal Services EBME or MEMS Medical Physics Community Health Centres Hospital Libraries Health Authority Libraries Community Health Visitors Pharmacy Procurement and Supplies HOW TO OBTAIN MHRA evaluation REPORTS MHRA evaluation reports are free of charge to NHS Trusts and Clinics or http nww evaluation nhs uk In England In Scotland In Wales In Northern Ireland MHRA Mr R Stock Ms Lisa Wellsby Mr Brian G
18. en measured for other recently evaluated syringe pumps This difference could be explained by differences between the Codan extension set used when testing this pump and the Vygon extension set normally used for syringe pump evaluations Differences in compliance can mean the same pressure change will cause a greater change in volume for the set and therefore a greater delivery or suck back of fluid at the cannula when pressure change occurs This undesirable fluid shift cannot be eliminated for any pump and the bolus delivered on raising a syringe pump has been shown to have serious MHRA evaluation 03141 April 2004 10 clinical consequences with some drugs For all pumps efforts should be made to retain the pump at the level of the cannula throughout its period of use in order to avoid these flow anomalies The period of flow interruption on raising back pressure to 100 mmHg is referred to as the resumption time This pump shows a resumption time of 2 5 minutes which is typical for a syringe pump Users should further be aware that flow anomalies resulting from changes in back pressure are not significantly dependent on the infusion flow rate A sudden delivery of about 0 4ml could present a significant hazard during a very low flow rate infusion of a potent drug Battery Test The pump contains a sealed lead acid battery 6V 4Ah This battery was fully discharged and recharged before the battery test was performed The pu
19. ents Training was supplied by Please tick as many boxes as apply a manufacturer tuition by a previously trained colleague tuition by a Trust based trainer b c d other Were user instructions available to you during use of the device Yes No If No go to Q 3 V Good Good Adequate Inadequate Poor 2a The overall quality of the user 5 4 3 2 1 No comment instructions for this device was Additional comments 2b How would you assess the 5 4 2 l Necominient readability of the manual s Additional comments 2c The coverage of the 5 4 3 2 1 No comment instructions was Additional comments 2d The user instructions on the 5 4 3 2 1 No comment body of infusion pump were Additional comments 3 Appropriateness of Device V Good Good Adequate Inadequate Poor 3a The device s suitability for your 5 4 3 2 1 No comment application was Additional comments 3b The battery life was 5 4 3 2 1 No comment Additional comments 3c The size of the pump was 5 4 3 2 1 No comment Additional comments 3d The weight of the pump was 5 4 3 2 1 No comment Additional comments 3e The ease of moving pump when 5 4 3 2 1 No comment disconnected from patient and stand was Additional comments 3f The robustness of the pump 5 4 3 2 1 No comment was Additional comments 3g The pole clam
20. et such that the infusion will last one hour Errors for both time and vol ume delivered are measured Tests are performed at 1 ml h and 25 ml h for all pump types CHANGES IN FLUID CONTAINER HEIGHT volumetric pump only The pump is set up to infuse with the drip chamber of the fluid con tainer positioned at a nominal 30cm above the top of the pump After a control period of 2 hours the fluid container is elevated to the maxi mum height obtainable above the pumping mechanism for 1 5 hours and then lowered to the minimum height while still remaining operable or 0 5 m if less for a final 1 5 hours Accuracy is measured and com pared over each of these three periods The pump is not stopped and settings are not altered when moving the fluid container Rate 125 ml h BATTERY PERFORMANCE After appropriate discharge and recharge cycling the delivery accura cy is measured at 1 ml h for syringe pumps and 125 ml h for volumet ric pumps with mains supply disconnected until the infusion pump stops due to battery depletion Long term accuracy over the second and the final hour of the test are calculated to assess any perform ance impairment due to battery depletion The duration of operation is noted The equipment is also tested for accurate delivery upon re application of mains power It is noted whether infusion settings are retained and whether the pump shuts down safely BOLUS VOLUME ACCURACY CODAN green stream SY P ARGUS 600 D
21. evaluation 031 41 MHRA Medicines and Healthcare products Regulatory Agency evaluation April 2004 NUMBER 03141 Syringe Pump Codan green stream SY P ARGUS 600 D Advantages Docking station offers barcoding matches patient with drug and single cable mounting for up to 5 pumps Good long term accuracy Good start up time Good resistance to dislodging of syringe Low profile several pumps can be mounted on one pole Highly configurable but see disadvantages below Disadvantages Display is limited in amount of information displayed Poor short term accuracy Multi function key and configuration options codes are potentially confusing Patient weight dependent dosing requires purchase of docking station Faults during testing Faulty battery SUMMARY The Codan green stream SY P ARGUS 600 D is a simple syringe pump designed for general purpose use and as such has largely appropriate functionality and performance Codan also market the pump for Intensive Care use Evaluation shows that the short term accuracy should elicit caution when selecting the pump for use with short half life drugs at very low flow rates The pump also lacks some of the features that are now routinely present on syringe pumps used in ICU A docking station is available offering additional functionality including a bar coding facility for matching patient and drug This is an excellent safe
22. g the prime function then alarm due to syringe ear mispositioning can be slow to occur Up to two minutes may pass before the fault is registered with smaller syringes 10ml 30ml Syphoning is likely to be prevented in this case by the barrel clamp The above scenario involves a double user error misloading syringe and failing to use the prime function It is not pos sible to start the pump with the barrel clamp dis engaged In testing the pump showed a good ability to resist dislodging of the syringe see page 12 of this report Priming The operations manual advocates prim ing the syringe in the pump prior to starting an infusion but does not state the consequences of not priming The danger of priming fluid into a patient is briefly mentioned Appropriately the primed volume is neither displayed nor added to the measured volume infused ml inf This vol ume should never be delivered to a patient Priming automatically stops after about 2ml have been delivered Starting an infusion Programming the pump is simple provided a basic infusion with standard syringe type is all that is required If more complex outcomes are required for instance confirmation CODAN green stream SY P ARGUS 600 D or changing of syringe type or delivery of a bolus then display limitations force the user to scroll through several options which can become con fusing The confirmation of syringe type might be deemed unnecessary in a Trust
23. hat function and label correspond BIME Note Configurability will be discussed later in this report in general however caution needs to be applied in changing any aspect of the configuration of this pump as discrepancy between actual function and expected function of keys can be generated The remaining LED displays are all alarm indicators These comprise a bar graph showing estimated line pressure changes 4 a large indicator to show that an alarm has been triggered 5 and a further smaller indicator to specify the alarm cause 6 The front panel controls comprise eight keys in a horizontal line There is a single key to turn the pump on or off 7 four keys to set flow rate 8 a MODE key which has several functions 9 a single key to start or stop the pump 10 and a single key to prime the line or deliver a bolus 11 A syringe is loaded horizontally on the front of the pump It is supported by a barrel clamp 12 a syringe ear guide 13 and syringe plunger retaining flippers 15 on the drive head 14 The drive head is positioned and its flippers are opened to accept the syringe by using the clutch lever 16 There is a pole clamp provided on the back of the pump the standard stackable version is CODAN green stream SY P ARGUS 600 D Description illustrated 17 Horizontal rail clamps of various sizes are available as accessories Also provided on the back of the pump are an interface for RS232
24. her pumps BIME Note Multi function keys are commonly regarded as ergonomically undesirable leading to possible user error or confusion They are however fre quently present on current infusion technology due to the limited space and multiple required MHRA evaluation 03141 April 2004 functions As with all current infusion technology the pump should not be used without training being com pleted and competence assessed Audible signals are clear and appropriate Volume can be adjusted in five steps but cannot be turned off Loading the syringe Loading the syringe is easy except when the pump is on the bench a bench support can be obtained as an accessory to make this easier The syringe is loaded behind the barrel retaining clamp the ears located in the slot provided and the plunger gripped by position ing of the flippers It was found during evaluation that the flippers that hold the plunger head per formed adequately except when a 50 ml BD Plastipak was in use frequently failing to secure this type of syringe satisfactorily Codan recom mends the use of Codan syringes Failure to grip the syringe invariably caused alarm as did failure to engage either the barrel or ear detectors The same alarm is activated for both syringe absence and misload most current pumps provide individ ual alarms to indicate which clamping part is at fault If the pump is started with the syringe misloaded and without first operatin
25. id to become established start up time The recorded seven minute time is the worst case seen for this pump using a Codan 60ml syringe some tests showed startup times as short as 1 minute Correct priming procedure was used in all instances BIME Note Users should be aware that the syringe used in a syringe pump can have a significant effect on several performance parameters All syringe pump manufacturers currently facilitate the use of some syringes that do not comply with ISO 7886 2 Syringes for use with power driven syringe pumps with negative consequences for performance A brief assessment of the effect of syringe brand on syringe pump performance will be included in a forthcoming BIME evaluation Codan recommend the use of their own brand 50 ml syringe This does comply with ISO 7886 2 For reference purposes a test using a 60ml BD Plastipak syringe yielded a constancy index of 3 minutes and a start up time of 17 minutes This is the standard syringe used in BIME evaluations when no specific syringe is specified by the pump manufacturer The start up time is poor in this instance though constancy index is improved over that for the Codan 60ml syringe Codan SY P ARGUS 600 D 0 mmHg 100 mm Hg E 2 c _ 100 mm Hg nA arbi ha error ryrie hhh H i Atean kaiaa at 420 480 Time minutes Figure 5 Back pressure test at 5 ml h over 8 hours CODAN green stream SY P ARGUS
26. ing the bolus rate and another felt VTBI programming was not easy Alarms and alerts Most users found the alarm indicators adequate to good one respondent con sidered the alarm indicators to be inadequate Three users all using docking stations said they found it difficult to ascertain which pump was alarming in a stack One other user felt alarm indi cators were too small when viewed from a dis tance Training The training provided by the manufac turer was praised by both Trusts Codan staff were considered flexible supportive and quick to respond to specific requests Questionnaire respondents rated the training from Codan as adequate good or very good Faults No pump faults were reported TECHNICAL ASSESSMENT Long Term Accuracy Flow rates tested 0 1 to 999 9 ml h Greatest overdelivery 0 2 at 5 ml h Greatest underdelivery 3 1 at 1 ml h Accuracy at minimum aie 9 2 at 0 1 ml h The pump was tested using 60ml Codan syringes over the full range of available flow rates It was also tested at 1 and 5 ml h using the following syringes 60ml Terumo 60ml Monoject 60m BD Plastipak 20ml Codan and 30ml Codan Good long term accuracy was seen using all syringes at all tested flow rates as is typical for syringe pumps Figures 3 and 4 show the delivery profile at 1 ml h using a 60ml Codan syringe Current measurement technology cannot adequately test very low flow rates and the
27. l tests excepting the 24 hour test where each of the last 18 hours is assessed Flow accuracy is monitored over the second and last hours of the manufacturer s recommended maximum life of the administration set The test is run at 500 ml h or maximum if lower for volumetric pumps Syringe pumps are run for 48 hours at 1 ml h Any change in the measured hourly delivery rate over the lifetime of the set syringe is noted SHORT TERM ACCURACY Short term accuracy is characterised using a new parameter known as constancy index The constancy index provides a guide to the mini mum half life of drugs suited for administration with the pump under going evaluation Constancy index is measured at 1 ml h for all pumps during a test run of 24 hours duration A discussion of this new concept is included in Evaluation report 02110 and can be provided on request by BIME FLOW ACCURACY UNDER BACK PRESSURE The pump is set up to infuse for a continuous period of 8 hours com prising 2 hours at ambient back pressure followed by 3 hours at 100 mm Hg and then 3 hours at 100 mm Hg At the points when the back pressure is changed the pump is not stopped and settings are not altered The long term accuracy is calculated for each period of the test The parameters resumption time pump height bolus and pump height negative bolus are calculated VOLUME TO BE INFUSED VTBI ACCURACY The infusion pump is set to the required rate with the VTBI facility s
28. mitations The user instructions provide appropriate guidance and some information is provided on the top sur face of the pump As a particular instance where more information would be useful the icons that indicate syringe misloading alarm do not specify which syringe MHRA evaluation 03141 April 2004 Ergonomic Assessment holding part is incorrectly aligned This could increase the amount of time taken to rectify the problem The alarm system is highly configurable This is a potential advantage but should be used carefully due to the possibility of configuring functions that do not correspond with either the labels on the pump or the instruction manual The occlusion detection system is calibrated in pressure units and can be set in increments of 100 mbar from 200 1200 mbar 150 900 mmHg The occlusion alarm pressure level can be reset at any time by the user No confirmation step is required when changing it The graph displaying estimated line pressure is scaled from zero to alarm level As there is no indication of which alarm level is currently set without scrolling through a full list of other pump settings on the display there is potential for misinterpretation of this graph Its function is to indicate how close to occlusion the line pressure is rather than indicate the numerical value of pressure in the line Provided the occlusion pressure has been sensibly set to match infusion conditions it is the rela
29. mp delivered fluid accurately for 7 hours 51 minutes in total 15 minutes before shutdown an audible and visual alarm indicated the need to restore mains power After stopping fluid delivery another audible and visual alarm continued for several minutes before the pump automatically turned off This behaviour was in agreeement with the expected shutdown process detailed in the user instructions and is applicable to both the 4Ah 8 hour battery or the lighter 1 2 Ah option On reconnection to mains power the pump reverted to the set flow rate and specific syringe used in the battery test It was possible to recall all other infusion parameters by pressing the 1 key while turning on the pump as is normal for this pump Accurate fluid delivery was restored Bolus Volume Accuracy Mean elective bolus error 0 1 ml bolus 1 1 1 0 ml bolus 0 3 5 0 ml bolus 0 5 The pump can be configured to deliver elective boluses in automatic and or manual modes In automatic mode a pre set bolus volume is delivered whereas in manual mode the volume depends on the duration of a key press by the user Automatic bolusing was used in this evaluation since manual bolus accuracy is dependent on operator skill Bolus delivery can be performed during an infusion or while the pump is stopped and bolus delivery rate is configurable Testing was carried out at the CODAN green stream SY P ARGUS 600 D defa
30. odfrey Business Services Health Planning amp Quality Division National Assembly for Wales Health Estates Hannibal House Scottish Executive HIMTE 3 Stoney Road Elephant amp Castle Health Department Cathays Park Dundonald London St Andrew s House Cardiff Belfast SE1 6TQ Edinburgh EH1 3DG CF10 3NQ BT16 1US Tel 020 7972 8181 Tel 0131 244 6913 Tel 029 2082 3373 Tel 028 9052 3714 OTHER RECENT MHRA REPORTS ON INFUSION DEVICES 26 web reports on volumetric syringe and ambulatory pumps at http nww mda nhs uk pumpevaluation May 2003 02110 Ambulatory syringe and volumetric pumps summary April 2003 03116 Medex Medfusion 3500 syringe pump November 2003 03117 Micrel Micropump mih ambulatory syringe pump November 2003 03119 Baxter Healthcare 3 Channel Colleague February 2004 03120 Fresenius MVP MS Orchestra volumetric pump December 2003 Forthcoming reports Ergonomic assessment of dose limitation software Alaris Baxter Medex WHAT YOU CAN EXPECT FROM MHRA evaluation REPORTS DISCLAIMER The Device Evaluation Service DES aims to provide independent and objective evaluations of medical devices available on the UK market Specialist centres mainly in the NHS do evaluations under long term contract to and in accordance with protocols approved by the MHRA The evaluations are usually of a unit supplied by a manufacturer We would expect this unit to be representative of the product on the market but cannot guarantee this Prospective pur
31. on If you have any queries about the questionnaire please contact BIME on 01225 821387 or email pumpeval bath ac uk Please return this form to XXXXXXXXXX The results of this survey will be used as the basis of a user assessment forming part of an MHRA evaluation of the XXXXXXXXXX pump Thank you in advance for your assistance Assessor Information User Name Date of birth Gender circle as appropriate Male Female Handedness circle as appropriate Right Handed Left Handed Clinical Specialty Number of months How frequently using this pump do you use this pump weekly The questionnaire is divided up into sections covering different infusion processes All questions require evaluation on a scale of 1 to 5 where Very good could not realistically be any better Good but could be improved Adequate Less than adequate but could be worse N O A Oo Extremely poor There is also space for additional comments These are often the most valuable source of information from the assessment so please take the time to add any comments that occur to you 1 Product Training Was specific training received for this device Yes No If No go to question 2 V Good Good Adequate Inadequate Poor 1 The specific training for this device was 5 4 3 2 No comment Additional comm
32. p was 5 4 3 2 1 No comment Additional comments 17 4 Administration Set V Good Good Adequate Inadequate Poor 4a The suitability of the 5 4 3 2 1 No comment supplied administration set was Additional comments Loading the Set V Good Good Adequate Inadequate Poor 5a The loading procedure for the set syringe was 5 4 3 2 F No comment Additional comments 5b The priming procedure for the set syringe was No comment Additional comments 5c The ease of setting syringe brand and size was syringe pump only No comment Additional comments Setting parameters and Starting Infusion V Good Good Adequate Inadequate Poor 6a The ease of setting the rate 5 4 3 2 1 No comment was Additional comments 6b The ease of setting the VTBI 5 4 3 2 1 No comment was Additional comments 6c The ease of navigating the 5 4 3 2 1 No comment control panels was Additional comments 6d The visual displays were 5 4 3 2 1 No comment Additional comments 6e The clarity of messages 5 4 3 2 1 No comment were Additional comments T Monitoring the Infusion
33. s good performance and some tests showed a start up time as low as 1 minute Constancy index The measurement of this index depends on many different parameters syringe brand temperature etc Codan s own measurements performed in house are better than the ones in the report ARGUS Medical as manufacturer recommends using a smaller size syringe e g 20 30 ml or even 10 ml for critical drugs having a very short half life time MHRA evaluation 03141 April 2004 CODAN green stream SY P ARGUS 600 D METHODOLOGY Tests are undertaken in accordance with the techniques outlined in BS EN 60601 2 24 1998 with variations where indicated A new administration set or syringe is used for each test The parameters constancy index resumption time pump height bolus pump height negative bolus are new concepts and are pro posed variations to the Standard IEC 60601 2 24 Algorithms used for their calculation can be provided on request bime bath ac uk Disclaimer For all tests only one pump and a limited number of disposables is tested The quoted results are a measure of the performance of this one pump It cannot be assumed that other devices of the same type will behave identically LONG TERM ACCURACY Rates tested minimum 1 5 25 125 for volumetric pumps and maxi mum Test durations are generally 2 hours except the 1 ml h test 24 hours and minimum rate test 4 hours Accuracy is assessed over the last hour for al
34. s of trials was undertaken to establish the consequence of incorrect loading of a wide variety of syringes The misload alarm was triggered by Codan green stream SY P ARGUS 600 D e Using Dedicated Codan set E87 P 1 0 0 8 0 6 0 4 0 2 0 0 Bolus ml 10 100 Set Flow Rate ml h Figure 7 Post occlusion bolus volume MHRA evaluation 03141 April 2004 Technical Assessment all deliberately misloaded syringes though the pump took up to two minutes to recognise incorrect placement of the plunger ears for all syringes smaller than 50 ml Free flow is unlikely to occur due to the other two clamps holding barrel and plunger When using 60m BD Plastipak syringes it was found that the plunger was not readily grasped by the drive head flippers It was frequently necessary to reposition the plunger to ensure correct fit A second test establishes the worst case of possible syringe misrecognition This is a new test and is showing that potential for flow error is seen on all syringe pumps if the wrong syringe brand is confirmed through use error Correct syringe recognition is important for the prevention of delivery errors and is dependent on the ability of the barrel clamp sensor to detect small differences in syringe barrel diameter There are no pumps on the market that are able to distinguish all syringe brands successfully This pump defaults to the syringe brand most recently used and the user may
35. spond to the configured one KVO occlusion pre alarm battery low battery depleted battery defective service technical fault Large LED displays for volume delivered and flow rate time etc LEDs behind symbols for most alarms No contra indications known today Yes RS232 serial interface pole rail bench or Docking Station IPX2 Class Type CF 2126326 1 50 14 CODAN green stream SY P ARGUS 600 D Manufacturer s Data PRODUCT SUPPORT Supplier CODAN Limited Eastheath Avenue Wokingham Berkshire RG41 2PR Website address www codan de Provisions for staff training initial in service training Provided FOC free of charge to all staff and users including night staff competency based certification of nurse attendance at training sessions available support material available for use in association with nurse education in IV therapy follow up in service training Provided as requested first line maintenance training Service engineering team available for repair on site if required at less cost to the hospital than workshop repair and minimising downtime full or part time courses available for biomedical staff available by product or according to hospital requirements Contact CODAN for details full service maintenance training Full service maintenance training courses available for biomedical engineering staff available by product or according to hospital requirements contac
36. ssary trip hazard whilst transporting the equipment or using it on battery A docking station is available for the pump allowing single cable powering of up to five pumps which presents a significant ergonomic advantage The docking station was not provided for evaluation Dependent on the size of syringe in use the plunger driver is mostly contained within the pro file of the pump but can extend slightly beyond the pump casing when a syringe is filled with more than 47 ml of fluid This common feature of sever al current syringe pumps leaves the drive mecha nism hypothetically more vulnerable to damage if the pump is dropped User Interface All displays are well lit clear to view and appropriately angled for reading at a range of heights They convey limited information to the user however which sometimes necessi tates multi step programming operations that can become confusing Supplementary information is provided by the fluid management system display that will be supplied in future as part of the dock ing station but this is not yet available in the UK The programming buttons are an appropriate size and neatly laid out in a single row The functions of the majority of these buttons are clear with appropriate text labels provided The MODE key has multiple functions attributed to it depending on the current situation The MODE key does not have an LCD soft key label to help clarify its cur rent purpose as on some ot
37. t CODAN for details Warranty One year based on return to CODAN One year on site warranty service contracts can be purchased at any time while the product is current an additional comprehensive warranty plan is offered Details from CODAN Limited Maintenance Provisions recommended service interval 24 months e contract service maintenance PPM Planned Preventive Maintenance undertaken by supplier s service engineers on site at recommended intervals of 24 months also see extended warranty provisions fixed price repairs avail able on site or at CODAN s UK service centre temporary loan equipment Available against agreement ppan Parts spares availability From stock at CODAN Service Centre cost of parts and materials Prices available on request i manuals e user manual One per pump free of charge e technical service manual One free of charge with first purchase of pumps Additional copies at 20 CODAN green stream SY P ARGUS 600 D 15 MHRA evaluation 03141 April 2004 Appendix 1 MHRA Medicines and Healthcare products Regulatory Agency Please complete the attached questionnaire without conferring with colleagues The results should represent your own opinion All assessors identities are confidential and are not used in the evaluation report We may need to contact you to clarify your responses and have requested some personal details in order to balance the survey populati
38. t accidental or deliberate modification of infusion parameters It can be turned off without the use of a security code but requires several key presses so is unlikely to be achieved by potential tamperers It is also possible to programme a standby time This is a useful feature and allows the pump to be prepared for an infusion and then to stand idle for a fixed time before alerting the user that the standby period has elapsed A separate Timer facility is also provided This can only be set up in STOP mode but is functional in stop or run mode When a pre set time has elapsed an audible alarm is given and if the pump was running it will automatically stop regardless of other programmed parameters Alarms and alerts Comprehensive alarms are provided as listed in the Manufacturer s Data on page 14 of this report Remedies to alarms are listed in the operations manual provided with the pump The alarm indicators are small but bright and their meaning is clarified by adjacent symbols and a label on top of the pump There is also a repeat ing audible beep and the display screens flash Pressing the MODE alarm silence key will stop the flashing display and silence the audible alarm for two minutes The alarm indicator will remain until the situation has been acknowledged and START pressed On occurrence of alarm the user is guided to the cause of alarm but is not offered remedial advice by the pump due to the display li
39. the questionnaires and interviews have been collated and the following sections represent a summary of the findings 13 users gave their views of the pump either by personal interview or by filling in a questionnaire Uses The pump is used in a wide variety of clin ical settings including ITU and general ward use The pump is not currently used for paediatric patients in these Trusts A wide variety of drugs are delivered including inotropes and low flow rate deliveries down to about 0 5 ml h are routine ly used when clinically indicated All questionnaire respondents rated the suitability of the device for their particular applications as good or very good Format The Trusts who assisted with this user assessment had elected to purchase both syringe and volumetric pumps from Codan and comment ed on the user interface advantages when using pumps from the same manufacturer One Trust had purchased docking stations and used them for both Codan syringe pumps and Codan volu metric pumps The docking stations were consid ered useful from a health and safety perspective in that they greatly reduce the number of power cables required All respondents considered the pump s battery life to be adequate or better 2 5 hours on pumps cov ered by this user assessment The decision to purchase pumps with shorter battery life was made to reduce device weight and improve ease of handling About half of respondents rated the pump good or ver
40. tive rise in pressure that is of clinical importance The minimum available alarm level of 150 mm Hg is typical for syringe pumps that do not have in line pressure transducers BIME Note If rapid alarm is required with a syringe pump pumps having in line pressure transducers should be considered The pump provides a back off system to reduce the volume of post occlusion boluses This is a valuable safety feature and was found to function efficiently Configurable options An exhaustive list of con figurable options is not possible here Some options set in the configuration are as follows default setting is given first in each case e OFF button active during an infusion or only if the pump is STOPPED e OFF button operates instantly or requires a prolonged key press e Syringe type is acknowledged at the start of an infusion or not e Infused volume counts up to a preset volume or counts down to zero e KVO Keep Vein Open delivery operates every time the STOP key is pressed or only at the end of a programmed infusion MHRA evaluation 03141 April 2004 6 The alarm system is also configurable For example the audible alarm volume can be adjusted the flashing displays at alarm can be configured not to flash the pump can be set to enable restarting without requiring an alarm to be cancelled if the alarm situation has been rectified and the timing of the syringe near empty alarm can also be adjusted The
41. ty feature Evaluation showed the technical performance of the syringe pump was satisfactory with above mentioned caveat about low flow rates whilst the user interface was found to have some drawbacks Display limitations for instance can necessitate scrolling through crown copyright MAIN FEATURES Power supply 230 VAC 50 60 Hz inter nal rechargeable battery Codan B D Plastipak Monoject Terumo Braun 10 60ml syringes Codan or any standard line 0 1 999 9 ml h Occlusion detection Drive force sensor 150 900 mmHg configurable pressure Graphical pres sure display Comprehensive though lacking clarity and speci ficity in some areas SUMMARY continued parameters to set up or control infusions For simple infusions however the pump is easy to use Syringes accepted list not exhaustive Extension set Flow rate range Alarms and alerts The pump has many configurable options governing both advanced features and basic operation An event log is provided A serial communications port allows download of the event log and down upload of configuration Price ex VAT 1450 List price Manufacturer Argus Medical AG Aarestrasse 13 CH 362 Heimberg Switzerland CE Marking Yes MD Directive Notified Body SGS UNITED KINGDOM LIMITED 0120 Certified to Standard EN 60601 1 1 2 1 4 2 24 EN 61000 4 2 3 4 5 6 7 11 EN 55011 36 Free to the N
42. ult bolus delivery rate with no background infusion Accurate performance was achieved for all elective boluses including the minimum 0 1ml demand volume Since the automatic system has proven accurate in testing it would be recommended that this automatic system be used for all elective bolus deliveries of less than 1 ml BIME Note When delivering manual elective boluses on a syringe pump the user is generally shown the volume delivered with a sensitivity of only 0 1ml This limits a user s ability to deliver small boluses accurately by a manual approach Occlusion Testing Mean occlusion test results at 1 ml h 10m 58s 0 08 ml Time to alarm Post occlusion bolus The pump was tested using a wide range of syringes to establish the speed of response to occlusion and the bolus on release of the occlusion The occlusion alarm pressure level was at the lowest setting 150 mmHg for all tests Sufficient time was given for flow to stabilise at the set flow rate before each occlusion was imposed and the response measured Broadly similar results were recorded for the five 60ml syringes tested Codan BD Plastipak Terumo Monoject and B Braun Omnifix A faster response to occlusion and smaller post occlusion bolus was seen for the 20ml and 30m Codan Codan green stream SY P ARGUS 600 D Using Dedicated Codan set E87 P 20 g 10 2 oe E oOo 10 Set Flow Rate ml
43. where only one type of syringe is available but this is commonly not the case Flow errors result if the wrong syringe brand is used or confirmed When the pump is turned on the last flow rate and syringe brand are recalled as default values These can then be altered if required It is possi ble to recall all previously set parameters from the last infusion by pressing the 1 key while turning on the pump Some programming steps such as clearing an input value are achieved through combination key presses This technique requires familiarity with the pump As with all infusion devices the pump should not be used without training Infusion options The pump can be programmed in ml h or volume over time The QUICK 100M docking station additionally allows units relating to drug concentration or patient weight but these are not possible on the stand alone pump It is possible but not essential to program a vol ume to be infused before starting an infusion Flow rate and infused volume are continuously displayed during an infusion The infused volume can be reset to zero at any time dependent on pump configuration It is also possible to change the flow rate of an infusion without halting an infu sion in progress This requires multiple key press es which effectively safeguards against acciden tal changes and tampering Bolus delivery is possible and can be either manual volume dependent on duration of user button
44. y should be used with caution In general both long and short term accuracy can be improved at very low flow rates by either a using a smaller diameter lower compliance syringe provided it is of equivalent quality b where feasible and clinically acceptable using greater dilutions and higher flow rates This is not always possible particularly in the case of neonatal care where overall volume administered has to be limited Short term accuracy and start up The constancy index indicates the shortest half life of a drug that would be recommended to Codan SY P Argus 600 D 2 ua 240 Set flow rate 1 ml h 720 960 1200 1440 Time minutes Max error over any hour 2 3 Figure 3 Long term accuracy at 1 ml h over 24 hours MHRA evaluation 03141 April 2004 8 CODAN green stream SY P ARGUS 600 D Technical Assessment Codan SY P Argus 600 D E oO 2 u Set flow rate 1 ml h Max error over any hour 2 3 120 Time minutes Figure 4 Long term accuracy at 1 ml h over 2 hours administer with this pump Constancy index values of less than 2 minutes are typical for syringe pumps and the time recorded for this pump is therefore relatively poor 6 5 minutes Constancy Index 6 5 minutes at 1 ml h 7 minutes Start up time A series of tests is used to measure the time taken for a steady delivery of flu
45. y good for size weight and ease of handling and no negative comments were made Pole clamping was rated good or very good by almost all respondents and none consid ered it inadequate All users rated the overall quality of the user instructions as good Codan extension sets were used in one participat ing Trust They were rated as good or very good by all but one respondent who considered it ade quate Staff commented positively on the pump s overall appearance Programming infusion options All but one respondent rated the syringe loading procedure as adequate or better though four users described some loading difficulties all respon dents currently use 50ml BD Plastipak syringes MHRA evaluation 03141 April 2004 Technical Assessment with the pump see syringe loading discussion in the Technical assessment on page 11 of this eval uation All users considered the ease of setting rate as at least adequate with nearly all rating it good or very good The rate and volume displays were considered good or very good by most users and adequate by all Two users rated the ease of navigating the control panel as inade quate two rated it as very good One user found the soft touch keys inadequate at preventing acci dental rate changes All users found the pump running indicator adequate or better most rating it good or very good one requested that it should be made clearer One user highlighted difficulties in chang
Download Pdf Manuals
Related Search
Related Contents
Manuel - funai This document is not a departmental publication. Version 2.1 User`s Installation, Operation & Service Manual Agilent 5975 MSD Serie Benutzerhandbuch HS760・HS970・HS1170 Catalogo General de Equipos Comerciales DEPRO CORROSTOP® SS-LAN-232C4248-2 MANUAL (マニュアル) MR2Qオーナーズマニュアル Copyright © All rights reserved.