KDS EZFlow 2030 Portable Syringe Pump User`s Manual
Contents
1. 09 2 Set driving rate by infusion volume First use a ruler and measure the length of medication in the syringe This is to count the medical solution in mm not ml Then divide this number by desired completion time hr to determine the driving rate in mm hr Formula mm h mm hr For example The medication in the syringe is 9 ml Load this syringe onto the pump the scale reading of the medication in the syringe is 40 mm in length The desired infusion time is 5 hours 40 mm 5 hr 8 mm hr Therefore set the driving rate to be O 8 START INFUSION Turn on the ON OFF switch to start infusion The running indicator will flash during normal operation while the rate setting key is locked To change the infusion rate you MUST reboot the pump Turn off the ON OFF switch change the rate and turn on again When batteries begin to run down an audible and visual alarm is activated When the battery voltage can not sustain a normal operation LOW BATT and COMPLETE OCCLUSION audible and visu al alarm will both be activated When infusion is completed or an occlusion happens the pump stops operation with an audible and visual alarm To continue infusion you MUST cor rect the error and reboot the pump BOLUS CABLE The pump has a PCA feature As directed by the DVM if the PCA feature is required plug the bolus cable into the pump socket to enable the PCA feature During a normal infusion while additional injections ma
2. When the device senses an occlusion by not being able to drive with a force of 6 5kg 1 kg the indicator is illuminated in red and the alarm buzzer sounds To silence the alarm press the RUN STOP button The battery indicator light ornage indicates the driver is using built in battery power this light and AC power light green will blink together if battery is low AC power light green indicates the pump is connected to mains AC power and is charging the battery LOAD SYRINGE Fill the syringe with desired medication Load the syringe onto the pump Refer to Figure 2 Place the syringe into the bed Be sure to insert the nozzle into the syringe slot Then press and hold the secure button on the driver to bring the driver block into such a position that the extremity of the syringe plunger can be placed into the driver slot Fasten the rubber strip to secure the syringe KDS EZFlow 2030 Operation Procedures Continued SET INFUSION RATE SIMPLIFIED METHOD WITHOUT RULER To determine the infusion rate of the syringe without 1 2 3 4 5 6 Load the empty syringe on the pump as instruction refer to previous page load syringe Note the reading of the pump s cursor location E g 6mm on the pump scale Move the syringe by holding the syringe driver head by 1 ml according to the reading on the syringe Note the reading of the pump s cursor location E g 10 mm on the pump scale Calculate the dif
3. apply to you If a defect arises within the warranty period contact KD Scientific Inc see address below The customer is responsible for shipping charges and must first obtain a Return Material Authorization number RMA before the unit will be accepted If a replacement unit is issued it is covered only for the remainder of the original warranty period dating from the purchase of the original device This warranty gives you specific legal rights You may also have other rights which vary from state to state KD Scientific Inc 84 October Hill Road Holliston MA 01746 Phone 508 429 6809 Fax 508 893 0160 E mail info kdscientific com Web www kdscientific com kd Scientific Copyright 2009 by KD Scientific Inc All rights reserved Printed in USA 8
4. ference on the pump s scale reading E g 10 mm 6 mm 4 mm The difference is the conversion factor between mm and ml of the current loaded syringe per ml E g 4 mm 1 ml the conversion factor is 4 mm ml Set the desired infusion rate by Formula Conversion factor x desired ml hr rate E g If the desired driving rate is 10 ml hr then the equivalent rate on the pump is 4 mm ml x 10ml hr 40 mm hr Therefore set the driving rate to be 4 O on the pump Syringe Driver Head Pump Scale finbimhitn hih finntinn T RATE 1 29 mrvh KDS EZFlow 2030 Operation Procedures Continued SET INFUSION RATE COMPREHENSIVE METHOD WITH RULER The graduations of syringes vary in terms of brands and sizes To ensure an accurate infusion the pump uses driving speed to calculate the flow rate Driving speed range is in 1 mm hr increment from 1 mm hr 99 mm hr Staff gauge range is from O 60 mm There are two ways to calculate desired flow rate 1 Set driving rate by flow rate First use a ruler and measure the ml scale to determine mm ml of each graduation scale Then multiply by desired infusion flow rate ml hr and get the driving speed in mm hr Formula ml hr x mm ml mm hr For example Use a 10 ml syringe for infusion The length of each 1 ml scale is 4 5 mm desired infusion flow rate is 2 ml hr then use the formula to calculate 4 5 mm ml x 2 ml h 9 mm hr Therefore set the driving rate to be
5. kd Scientific KDS EZFlow 2030 Portable Syringe Pump User s Manual Publication 5616 002 REV A KDS EZFlow 2030 Contents COU OM AEE E EEA EEE A E A 2 Application Fields sisssctssscseshisitisssiiiesicssdeviheltspdes sesleesscscdasssussennshsiseausagsostocnbsondensesyecsbundiplesntes 2 Specifications aad Operation Procedures KEY PO ersari atier NAAN EELSETE AEE IE EEEE A EEEE 4 Keypad Description csescsssssssssseesseesseesseesneeseeeseesseeesesseesneesneesuseaneeaeesneesneeseenees 5 Alarmisvand Indicator Lights siccisscacscccrsssaitecceiebelaysseasjssusaudaetessscvssdanspduccsesustebiseees 5 koad Syringe cess casting cteccosarsasndeestttaadens ae A reii e EEES oa a EEEE RETEN ENEE 5 Set Infusion Rate Simplified Method without ruler ssssssssssssssrsssssssrrrssssserrressese 6 Set Infusion Rate Comprehensive Method without ruler 00 sseeeee 7 Siart MEU SHON eesin inor ia EEE EE O I REEERE 7 Bolus Cables unoha A A OERE 7 Storage limited Woman Yerne a A AEREE ERE 8 KDS EZFlow 2030 Caution IMPORTANT User should read this entire manual before operating this portable Syringe pump Keep this manual for your future refrence WARNINGS 1 Critical Do not use in the presence of flammable anesthetics 2 Do not use pump syringe or accessory that shows any sign of damage 3 Only authorized service personnel should repair this pump Manufacturer assumes no responsibility for incidents that
6. minutes interval for safety protection a 3 x AA batteries b 1 x BOLUS cable PCA c 2 x Syringe protecting cover d 1 x Carrying case DC 4 5V 3 x AA batteries 30 days or minimum 50 syringes W 166 x H 60 x D 30 mm 6 5 W x 2 4 H x1 2 D 180 g 0 55 Ib 240g with 3 AA batteries Specified 50 60ml syringe required fill up to 33ml KDS EZFlow 2030 Operation Procedures 1 Type of Protection Against Electric Shock Type BF 2 Degree of Protection Against Ingress of Fluids Drip Proof IPX1 3 Exclamatory mark A CAUTION Please refer to user s manual KEYPAD 1 Low Battery Alarm 4 Syringe Driver Head 2 COMPLETE OCCLUSION Alarm 5 Set Rate 1 99 mm h 3 ON OFF Switch 6 Bolus Cable PCA KDS EZFlow 2030 Operation Procedures Continued KEYPAD DESCRIPTION SET TIME Activated only in STOP mode When pressed sets the desired time SET TIME indicator should illuminate PURGE Only functions in STOP mode When pressed starts purging tubing set RUN STOP Press to start or stop the pump MODE indicator should illuminate ALARMS AND INDICATOR LIGHTS NEAR END When there is approximately one minute of infusion time remaining this indicator flashes and the alarm buzzer sounds To silence the alarm press the RUN STOP button COMPLETE OCCLUSION When infusion is complete the indicator is illuminated in red and the alarm buzzer sounds COMPLETE OCCLUSION
7. occur if the pump is not repaired according to manufacturer authorized procedures 4 Before connecting the syringe to the patient user must prime the syringe and extension tube fo prevent air to the patient 5 Take the batteries out if this product is left unused for a period of time 6 To clean the pump use a damp not wet cloth or sponge Do not immerse pump or power cable into water or other cleaning solutions 7 To preserve our environment and conserve our natural resources recycle batteries at the end of their useful life CAUTION Protective Ground Refer to Manual Terminal Application Fields This Portable Syringe Pump is used to deliver medications from a disposable syringe size from 2 to 50 60 ml This portable user friendly device offers a new alternative for a micro reliable infusion KDS EZFlow 2030 Specifications Applicable Syringes Linear Flow Rate Flow Rate Converted based on syringe sizes Accuracy Alarms Bolus Function Included Accessories Power Supply Battery life Dimensions Weight 2ml 50 6Oml disposable syringes 1 mm hr 99 mm hr linear rate 50 60 ml syringes 0 7 up to 65 ml hr 20ml syringes 0 35 up to 34 ml hr 10 ml syringes 0 1 up to 16 ml hr 5ml syringes 0 1 up to 11 ml hr less than 2 Audible and Visual Alarms of COMPLETE OCCLUSION and LOW BATT Easily adding 1 4 of preset dosage by bolus cable without interruption of current infusion Minimum30
8. y be needed you can push the button on the bolus cale to inject an additional dosage Each PCA dosage is 1 4 of the preset delivery dosage For safety of the patient and to control the frequency of PCA injections the interval between two PCA injections is programmed to be less than 30 minutes Storage This productshould be stored in an environment with ambient temperature of 14 F 131 F 10 C 55 C relative humidity below 93 and air pressure 500hpa 1060hpa Do not store close to corrosive vapour and or harmful impurities KDS EZFlow 2030 Limited Warranty KD Scientific Inc warrants to the first consumer purchaser for a period of one year from the date of purchase that this unit when shipped in its original container will be free from defective workmanship and materials and agree that it will at its option either repair or replace the defective unit This warranty does not extend to misuse neglect or abuse normal wear and tear accident modification or unauthorised repair KD Scientific will not be liable or in any way responsible for any incidental or consequential economic or property damage Some States do not allow the exclusion of incidental or consequential damages so the above exclusion may not apply to you There are no implied warranties of merchantability or fitness for a particular use or of any other nature Some states do not allow this limitation on implied warranty so the above limitation may not