Radlink GPS Hardware User Guide

image

Contents

1. Touch screen is not responding The user may be touching the inside of the monitor frame blocking the sensors The USB cable for the touch screen may be loose Press edges of the monitor frame to reconnect the touch screen sensors Check the USB cable and make sure it is firmly connected If it is connected unplug the cable and plug it back in Computer response time is slow The computer may have been shut down using the power button rather than via Windows Too many images from previous use are being stored on the computer Turn the computer on and let the computer reload until the Windows home screen appears This process may take several hours Call Radlink Technical Support 11 Troubleshooting Radlink Technical Support For questions call Radlink Technical Support 310 643 6900 ext 2 Problem Reason Solution DR Status light reads Initializing instead of Ready DR Panel will not connect The wireless communication between the panel and the router on the GPS is blocked or slow DR Panel battery may be low Unstable connection between DR panel and GPS Make sure there is no lead in the X ray generator or lead gown between the bucky and the GPS Replace the battery with a fully charged battery 1 Close Radlink Pro Imaging software by clicking Manage 2 Logout 2 Turn off the DR Panel 3 Turn on the DR Panel and wait until the LE
2. charged battery To charge the Vieworks DR panel battery amp With the arrow at the bottom facing front and downward insert the battery into the charger Figure 13 When the battery is fully charged the solid green light at the top of the battery will turn green Figure 14 To insert the battery into the Vieworks DR panel Place the battery in the slot on the back of the DR panel Figure 15 Push the locking mechanisms toward the center of the battery Figure 16 Figure 13 Figure 14 Figure 15 Figure 16 The DR Panel Battery Perkin Elmer The DR Panel Battery Perkin Elmer The Perkin Elmer DR Panel comes with two battery packs and one battery charger When fully charged each battery lasts 2 3 hours For best results start each surgical procedure with a fully charged battery To charge the Perkin Elmer DR Panel battery ED With the sticker at the bottom facing front and downward insert the battery into the charger Figure 17 When the battery is fully charged the four solid green lights will be illuminated If the battery is not fully charged the lights will blink Figure 18 To insert the battery into the Perkin Elmer DR panel Place the battery in the slot on the DR panel The sticker on the battery should be on the opposite side of the sticker on the panel Insert the battery all the way untill it clicks Figure 19 amp To remove the battery push the knobs inward until the batte
3. on the DR panel Press and hold the power button until the panel turns on The green amber and blue lights must all be illuminated Figure 5 Check the battery status of the DR panel by pressing the Windows button Figure 6 amp On the task bar click the FMC icon The battery level will then be displayed at the bottom of the window If battery level is low replace the battery It is recommended to have a fully charged battery at the beginning of a case Figure 7 The battery charge shows in increments of 25 i e a battery with 75 charge will show as 75 Leave the FMC icon open during the case Figure 8 Figure 5 WIRELESS_STATE received 7 value uintValue s value 3 doubleValue Figure 7 Figure 8 Using the Radlink GPS in a Posterior Approach THA Procedure Continued Turning on the Perkin Elmer DR Panel To turn on the Perkin Elmer DR panel press and hold the power button until the panel turns on Figure 9 Once the DR panel has been turned on follow Steps 4 13 to complete the THA procedure Double click the Radlink Pro Imaging icon on the lower monitor E Double click the Pre op Viewer icon on the upper monitor to access the PACS Select the pre operative image to be used as a reference during the case On the lower monitor click on the row for the study if the order is displayed in the modality worklist Otherwise click New Patient and enter the study information Using the Ra
4. Ds stop flashing 4 Launch Radlink Pro Imaging software The GPS will not turn on The battery has been depleted because the GPS was not properly shut down If the GPS is only for anterior approach call Radlink Technical Support for instructions on how to bypass the UPS If the GPS is being used with the DR Panel for posterior cases call Radlink for a new UPS 12 Radlink Inc www radlink com 310 643 6900 815 N Nash St m 3 EI u CA 90245 JARadlink
5. JARadlink Galileo Positioning System Hardware User Guide System Overview Galileo Positioning System with Perkin Elmer Panel Or The Vieworks Panel Upper Monitor shows pre operative image mn E Lower Monitor shows intra operative image DR Panel a ee Keyboard Perkin Elmer Router Or Vieworks Control Box Battery Mn I Power Cord Reel Ethernet Cord Reel BNC Cord Reel Computer Connects to C Arm Monitor Power Button Uninterruptable Power Supply UPS Turning on the Galileo Positioning System GPS Plug the power cord of the unit into an AC power outlet Figure 1 The Uninterruptable Power Supply UPS will turn on The leftmost UPS indicator light should now be green see expanded view below Turn the computer on by pressing the power button Figure 2 If you unplug the GPS the system goes to battery mode the UPS will start beeping and the orange battery light will illuminate which enables the user to move the unit without plugging into the wall for 15 minutes Figure 3 Note The unit can remained unplugged for only 15 minutes before the battery will be depleted When the UPS is on this light will be green Figure 1 Figure 2 Figure 3 Using the Radlink GPS in an Anterior Approach THA Procedure Pull the BNC cable from its cord reel and plug the cable into the Video Output Port on the back of the C Arm monitor by pushing and twisting clockwis
6. dlink GPS in a Posterior Approach THA Procedure Continued amp The SCAN page will appear On the body avatar click the body part to be scanned The DR Status on the upper right corner of screen must be in the READY mode If the DR Status is not in the READY mode refer to Troubleshooting on page 11 Click Start DR Battery Charge READY Mode indicator Start DR Vieworks DR Panel Perkin Elmer DR Panel Figure 10 Using the Radlink GPS in a Posterior Approach THA Procedure Continued Important Do not position the X ray generator between the GPS and the DR Panel Lead from the mobile X ray can interfere with the wireless connection resulting in the inability of the panel to acquire images Take the intraoperative X ray The image will appear on the lower monitor Use Image Orientation buttons to rotate the image Use the Image Appearance buttons to Window Level Zoom in etc Figure 11 Refer to the Surgeon s ChecklistTM user manual to take measurements with the software Figure 12 If it is necessary to shoot another image click SCAN and Start DR and repeat steps 9 12 Image Orientation Image Appearance ZOOM Figure 11 Figure 12 The DR Panel Battery Vieworks The DR Panel Battery Vieworks The Vieworks DR Panel comes with three battery packs and one battery charger When fully charged each battery lasts 2 3 hours For best results start each surgical procedure with a fully
7. e If images need to be pulled from PACS and WiFi access is not available pull the Ethernet cable from its cord reel and plug into an Ethernet port Double click the Radlink Pro Imaging icon on the lower monitor Double click the Pre op Viewer icon on the upper monitor to access the PACS Select the pre operative image to be used as a reference during the case On the lower monitor click on the row for the study if the order is displayed in the modality worklist Otherwise click New Patient and enter the study information Once you are in OC Image page click the Grab Frame tab Figure 4 C Arm images will appear at the bottom of the lower monitor screen Refer to the Surgeon s ChecklistTM user manual for instructions on how to take measurements with the software Figure 4 If images need to be sent to PACS and WiFi access is not available pull the Ethernet cable from its cord reel and plug into an Ethernet port Once cases have been completed for the day unplug the BNC cable from the back of the C Arm monitor by pulling and twisting counter clockwise Gently pull on the BNC cable until it retracts w Using the Radlink GPS in a Posterior Approach THA Procedure Note Your GPS system has been provided with either a Vieworks DR Panel or a Perkin Elmer DR Panel Follow the steps below to turn on the panel Then proceed to steps 4 13 to complete the case Turning on the Vieworks DR Panel Once the GPS is on turn
8. ry ejects Figure 20 Figure 17 Figure 18 Figure 19 Figure 20 Radlink Gps System Shut Down Instructions TO SHUT DOWN THE GPS THE FOLLOWING PROCEDURE MUST BE COMPLETED IN THE FOLLOWING ORDER Click the Windows icon in the lower left hand corner of the lower monitor screen Click Shut down Unplug the power cord from the AC power outlet At this point the Radlink GPS will go to battery mode The solid orange light will illuminate and the UPS will begin to beep Press and hold the power button on the UPS to shut down the unit properly Warning If you do not shut the ups down by pressing and holding the power button step 3 the battery will drain and will need to be replaced For a Posterior case press and hold the power button on the DR panel in order to turn it off Powe pe A Change Batter Mode Indicator Light Figure 21 Troubleshooting Problem Frame Grabber is not pulling images from the C Arm Reason The video output at the back of the C Arm monitor may be damaged or broken The signal mode on the Converter is not set to BNC Frame Grabber is not selected in the software Solution Adjust repair the video output port or try using a different C Arm port Press the Input button on the BNC Converter to switch to BNC mode 1 Open the Radlink Pro Imaging software and click Manage 2 Make sure the Frame Grabber is checked under Hot Buttons 3 Click Save Settings