Ultrasonic tester UK1401. User manual. 2 CONTENT 1

image

Contents

1. Taia 13 12 1 RECORD OF MEASURED RESULTS TIME OR VELOCITY OF ULTRASONIC WAVES PROPAGATION AN THE MEMORY 2 13 12 2 REVIEWING OF THE RESULTS WITH LISTING OF GROUPS 14 12 3 RESULTS REVIEWING WITH LISTING CELL NUMBERS IN THE CURRENT 8 15 12 4 RESULTS CORRECTION coe ee ee eee 15 12 5 MEMORY CLEARING 0ccccecececcccececccceccceacuceneccececeacenecceneauaceneaeaustentaesueeaesneceaceeansnenss 16 13 DATA OC TOPO cases cesses a asses is 16 13 1 ADJUSTMENT OF THE INFRARED PORT ccceccececcecccceccceccacececucuceneaceneaueceuueaseneaueneanes 16 13 2 INSTALLATION OF THE PROGRAM FOR DATA TRANSMISSION FROM UK1401 TO THE PC 17 13 3 DATA TRANSFER FROM UK1401 0 0 0 0 0 ccc ccc ceca cececcecccccccceccacececucaceueaceneauenesueaseneaeeneanes 17 13 4 LICENSE 18 13 5 LIMITED WARRANT ccccccceccececcecccccccaceccaceceaceneaueneeneaceneaueneaueaeeneateeauecesueateneaneatanes 19 14 WARRANTY ON THE PRODUCT 2 cccccccscccccccccsccccscccccccccccccccsccccsccccccccsccccscecs 19 15 INSTRUCTIONS ON TESTING THE CONCRETE S SOLIDITY IN MONOLITHIC STRUCTURES WITH THE HELP OF SURFACE SOUNDING ULTRASONIC METHOD 21 AS INTRODUCTION sazcciciateicv
2. screen at starting Press and release these buttons Receive blinking number of cell of the last recorded result With the help of these buttons it is possible to choose and adjust earlier recorded results Put the device on the object the way that probes are perpendicular to the surfaces Push the device with force of about 5 10 kg The device will pass in the active state The change of indications is accompanied with short beeps Hold the device in such state during 15 20 seconds so the indications will change 2 3 times Receive the new result 15 Ultrasonic tester UK1401 User manual Record it n this cell Number of cell will blink all the same That means you can make measurement once again Exit from the correction mode Memory Clearing Go to the main menu The memory should be switched on There should be figures In the line MEMORY With the help of these buttons select the line CLEAR Line should start blinking Function of Data deleting Press Enter button On the screen will appear inscription PRESS ENTER lf the data is not needed just press Enter Remember you will not be able to restore the meanings so be sure that you want to delete all the data from the memory To cancel clearing just press menu The melodious signal will sound The device memory will be cleared The device will return in the measurement mode 13 DATA OUTPUT
3. OF THE BATTERIES At the bottom of rear panel of the electronic module there is the battery cover It is fastened to the case with one screw For replacement of power supply elements remove the battery cover Please pay attention on the polarity at setting the new batteries The results which have been written in the device s memory are not erased when batteries are empty or replaced All device adjustments are saved also 4 RESTORATION STANDARD ADJUSTMENTS AT STARTING UP lf there is no sound signal and no information on the display at switching the device then one of the reasons can be the failure of the processor if it is not the empty or bad batteries or the absence of the electric contact in the circuit In this case the following actions should be undertaken ON OFF Holding the key menu press key on off If the device still doesn t work repeat it once again in 2 3 seconds After such an emergency turning on of the device all the standard settings of the device are restored After that the device is ready for operation Check the adjustments If it is necessary correct meaning hardware delay time and distances between probes value take from passport With the helo of these keyboard combinations the delay time and distances between probes are carried out lf the contrast of the screen is unsatisfactory it can be adjusted with these keys combinations Ultrasonic tester UK1401
4. Press the device with effort of 5 10 kg and fix it in this position Hold it for 5 10 seconds Write down the indications of the device Remove the device from the sample is necessary to repeat this procedure for 20 times After that calculate the average meaning of the measured ultrasound time and compare it with the time that is written on the control sample written The readings of the device should not differ from the written time on more then 0 5 microseconds with taking into the account the temperature coefficient For all the questions please contact the manufacturer Acoustic Control Systems Ltd 119048 Moscow POBox 148 tel fax 7 095 244 3194 244 2535 245 5896 e mail market acsys ru website www acsys ru 20 Ultrasonic tester UK1401 User manual 15 Instructions on testing the concrete s solidity in monolithic structures with the help of surface sounding ultrasonic method Introduction Accordingly to GOST 17624 87 Concrete Ultrasonic methods of solidity estimation the testing of monolithic structures with ultrasonic methods can only be made with through sounding way of testing Operation experience of laboratory of reinforced concrete structures and quality estimation in State Unitary Organization Scientific Research Institute of Reinforced Concrete showed also the possibility of using surface sounding way of ultrasonic testing for nondestructive testing of concrete solidity in monolithic Struc
5. User manual 5 MAINTENANCE The maintenance service of the tester consists basically in clearing the device and probes from the dust and dirty and periodic charge accumulators at complete brightening symbol battery on the screen of the device Accumulators of the given type it is not recommended frequently recharge if they still completely were not unloaded as it can result in reduction of their resource of work The charge accumulators is carried out with the help universal charging unit delivered in the complete set of accessories to the device during 12 16 hours lf to use dry supply elements it is possible to work up to their complete category after which it is necessary elements at once to replace or even to delete from the device what to avoid pollution by its flowed out electro cast Once in the month expediently these elements to get from the device and to examine as electrolyte can sometimes follow and from not unloaded elements It is necessary immediately to replace spoiled elements Practically singular defect which user can remove independently is the infringement of electrical contact in the circuit supper elements It is expressed in absence of any symbols on the indicator after pressing the key on off or in indication of the lowered voltage when the battery obviously normal If it has taken place and you are sure that the supply elements are charged take them from the device and wipe their contacts by spirit or clean
6. average The testing of calibration dependence is made accordingly to the Attachment 4 to GOST 17624 87 In this case if this dependence is build basing on the results of parallel testing with Ultrasonic Methods and Avulsion with Break method the results of testing on cubes are replaced with the results of testing on structure s parts The testing of the calibration dependence can be replaced with the correction of the dependence adjusted for the additionally received testing results In this case the testing of the dependence or its correction is made not less then one time pro month When the environmental temperature falls under 5 the correlation dependence should be corrected or built anew There should be not less then 5 samples or structure s parts when making testing or correction of calibration dependence 15 made Using of universal calibration dependence for approximate detection of concrete s solidity 1 Because of difficulties or impossibility of building the calibration dependence in some cases it is possible to use universal calibration dependence for approximate estimation of concrete s solidity 2 The universal calibration dependence should be built for specific regions after Summarized Data processing that was used for building the calibration dependences for single building objects On the picture is represented the calibration dependence that was based on the summarizing of data from the testing of seve
7. fine emery paper If it has not helped it is necessary to address in the firm manufacturer is not recommended to assort the device without emergency as can result in casual damage of its internal devices At occurrence of malfunctions tester or any questions on its use it is necessary to contact to the representatives firm on telephones specified in the passport of the device 6 DEVICE KEYBOARD Ultrasonic tester UK1401 User manual a Gig Ga Off of the device The device also switches off automatically after removal it from the object s surface and if the operator doesnt not press any key during 4 minutes after the removal Switches on and switches off the backlight Each pressing switches on or off the backlight When used together these keys improve the contrast When used together these keys darken the image of the contrast In the mode of device setting changes setting parameters and starts selected operations from menu In the measurement mode data enters in the free sell memory Allows choose menu line in the setting mode In the flaw detection mode change the amplification coefficient of the receiving path In the memory viewing and correction mode change the result and group numbers Goes to the main device menu Also switched the memory viewing and correction mode Switche
8. ultrasound dissemination are detected After that you need to choose not less then 12 areas including the areas where the velocity time of ultrasound dissemination is minimal maximal and middle After the ultrasound testing these areas are tested with avulsion with break method or the samples from these areas are chosen for testing under pressure The concrete s age on the single areas should not differ on more then 25 from the middle age of concrete in areas of structure or structure itself or group of structures subjected to testing The exception is made for building the calibration dependence for estimation concrete s solidity at engineering testing when the difference in concrete s age is not regulated On each area with the help of magnet device the location of reinforcement in the concrete is detected and then with the help of ultrasonic device the ultrasonic velocity time in it is measured Ultrasonic testing is made not less then 2 times The testing is made in two directions perpendicular to each other The sounding is made under 45 to the reinforcement direction parallel and perpendicular to At sounding in the parallel to the reinforcement direction the sounding line lies between the reinforcement cores on the picture 1 location of the device 2 location of the reinforcement The difference in meanings of singular ultrasound velocity time measurements at every testing area sh
9. with the device are in the rarefied condition as it is recommended by rules of their storage therefore before the beginning of operation of the device they should be charged 19 Ultrasonic tester UK1401 User manual Accordingly to the Article 23 Law of Russian Federation About the providing unity for measuring devices UK1401 ultrasonic tester is calibrated right after its production This calibration is marked in the technical passport of the device with a special stamp The device s calibration is also made during usage of the device either by the manufacturer by the standardizing departments of the enterprises that use ultrasonic tester UK1401 For checking the device in the process of use a special control testing block from Plexiglas is delivered with the device The sizes of this block are 190x40x25 mm On the sample stays the serial number of the device and the right ultrasonic time that the device when operating correctly should read show at the temperature 20 C 1 C The temperature indication is important because the Plexiglas has temperature factor of ultrasound velocity 0 12 C That factor makes influence on the results increasing measurable time for 0 07 microseconds at the increase of temperature on 1 C at base of 150 mm The check should be made as follows Turn on the tester and put it on the control sample Ultrasonic probes should be put in the deepening specially made on one of the sides of the sample
10. 401 User manual On the screen you will see the results of velocity measurement in the object Display after removal of the device from the object surface indicator of the level of receiving signal is light Press Enter for 10 15 seconds after removal the device from the tested object surface The short signal will sound Go to the next cell lf number of result in group is equal 255 the next measurement of the device will record in next group IT IS POSSIBLE TO BEGIN NEW GROUP WITHOUT WAITING FOR WHEN THE CURRENT GROUP IS FULL FILLED Put the device on the object surface the measurement press and held the button for 2 3 seconds Short signal will sound The result will be recorded in the new group under the Ne 001 The number of the free sell in the new group Press and release the buttons Get blinking number of group of the last recorded result With these buttons it is possible to review earlier recorded results in other groups Exit from the reviewing mode 14 Ultrasonic tester UK1401 User manual Results Reviewing with Listing Cell Numbers in the Current Groups screen at starting Press and release the buttons a a i ee Get blinking number of cell of the last recorded result With the help of these buttons it is possible to review earlier recorded results in other groups Exit from the reviewing mode Results Correction
11. TO PC Adjustment of the Infrared Port lf you use the external adapter of IR link of Tekram corporation such as IR 210 primly enough to connect a cable of the adapter to suitable plug on the back panel of the computer lf infrared link is Supposed to be used not only for link with the instrument it is necessary to install the driver and to adjust infrared link as it is described below 4 i A E Infrared Link To unclose Control panel of the computer and to select an icon Infrared link 16 Ultrasonic tester UK1401 User manual In appeared window remove a tick To resolve infrared link through lf you want to use infrared link for operation with other devices it will be necessary to restore the reference value For some high speed infrared adapters it is required to install the SIR compatible mode For details address to a manual on the adapter used by you Installation of the Program for Data Transmission from UK1401 to the PC The required configuration e the operating system MS Windows 95 C Microsoft corp or newer versions e MS Excel97 Microsoft corp or newer version e the built in or external infrared port compatible IrDA standard lf on your computer the program MS Excel is not installed install it After that it is possible to initiate with installation of the program of data transfer For this purpose it is necessary to start setup exe program which is taking place on a customizin
12. Ultrasonic tester UK1401 User manual CONTENT 1 OPERA TION PRINCIPLE O a a aa 3 2 APPEARANC P sce ceca 3 3 SET AND REPLACEMENT OF THE BATTERIES 11 cc ccssccssscccsscccsssccssscccsssccscces 4 4 RESTORATION STANDARD ADJUSTMENTS AT STARTING 4 5 MAINTENANCE onroro AE AEE ELE 5 6 DEVICE KEYBOARD ununa a 5 7 EA A E 6 8 DEVICE SCREEN IN THE DIFFERENT MEASUREMENT 8 8 1 MODE OF VELOCITY AND TIME OF ULTRASONIC WAVES PROPAGATION MEASUREMENT WITH TURNED OFF MEMORY 8 8 2 MODE OF VELOCITY AND TIME OF ULTRASONIC WAVES PROPAGATION MEASUREMENT WITH TURNED ON MEMORY 2 cccccecececccceccccccccecccceccaeaceuececeneacsteneacsteneaceneausnsneatanenss 8 8 3 MODE OF CRACK DEPTH MEASUREMENT 2 sccccececcccececcacaccceacaceceacaceceacuceceacsteceaceneaeas 9 8 4 AUTOMATIC GAIN CONTROL AGC 9 9 OPERATIONS BEFORE MEASUREMENT ccccccscccsscccsccccsccccscccccsccccsccccecscees 10 10 VELOCITY OR TIME OF ULTRASONIC WAVES PROPAGATION DEFINITION 11 11 DEFINITION OF CRACK DEPTH LAYING ON A 0 12 12 OPERATIONS WITH MEMORY
13. ct allows to increase reliability of results of measurements on concrete with strongly destroyed superficial layer with set interstice and fine cracks some In this case it is necessary to choose the greatest possible gain 65 75 dB limiting allowable disorder of results for example no more 3 from average value received at motionlessly established device Ultrasonic tester UK1401 User manual 9 OPERATIONS BEFORE MEASUREMENT Start up the device with pressing On Off button You will hear the melodic signal For 1 5 2 seconds on the screen of the device the number of the version will be seen on the display The device is ready for operation in the last saved condition For selection of TIME VELOCITY CRACK mode go to the main MENU Press Enter until you have the necessary mode screened on the display see below With pressing key once you will go to the memory On or Off selection Press Enter until you have the necessary adjustment see below Press key twice You will go to the sound On or Off selection Press Enter until you have the necessary adjustment see below With this key you can choose the unit system item of the menu METRICS ENGLISH Press Enter until you have the necessary adjustment see below 10 Ultrasonic tester UK1401 User manual Press Arrow Down button to go to AGC On Off adjustments Press Enter until you have the necessary adjustment With pressing Menu again y
14. eiaunicuasaceiciaucnwideaiesatdnaiaeaid ieee lai 21 15 2 APPLICATION OF SURFACE SOUNDING METHOD ccceccececcececcsccccececceceauccesueceseaueeeaues 21 15 3 PREPARATIONS FOR TESTING 21 15 4 USING OF UNIVERSAL CALIBRATION DEPENDENCE FOR APPROXIMATE DETECTION OF CONCRETE S SONDI 23 15 5 TESTING AND DETECTION OF CONCRETE S SOLIDITY IN STRUCTURES 24 16 TABLE OF VELOCITIES OF LONGITUDINAL ULTRASONIC WAVES IN MATERIALS M S oenen a sucenexdecbececscasesescesssueesesosusuesrcaceoncaeseescetece 26 Ultrasonic tester UK1401 User manual 1 OPERATION PRINCIPLE UK1401 ultrasonic tester is applicable for measurement of ultrasound velocity and time of longitudinal ultrasound waves dispersion in solid materials at surface sounding testing method when the purpose of testing is to detect and estimate the solidity and homogeneity of material and structures This estimation is based on the correlation of ultrasound waves velocity in material to its physic and mechanical characteristics and physical statement 2 APPEARANCE Two ultrasonic probes with dry point contact DPC Electronic module in the rectangular plastic case 6 key film keyboard LCD screen with backlight for work in darkness Infrared port for the connection with the external personal computer On 3 Ultrasonic tester UK1401 User manual 3 SET AND REPLACEMENT
15. g diskette Will see the screen of installation Press the button To begin installation 4 The program will check up presence of necessary files and programs justing If everything is all right press appeared button To pattern IR port files After installation the program will offer you to adjust a port of infrared link Data Transfer From UK1401 Include a computer and start the program Excel97 newer aS EFi In the panel of buttons of this program find the button Press it with the help of the left hand mouse button The window UK1401 link will open Transfer in the mode of the menu With the help of keys select to a sink OUTPUT the line should be blink 17 Ultrasonic tester UK1401 User manual Data transfer Arrange the instrument so that window an infrared data link was opposite window the adapter The distance between the instrument and the adapter should be no more than 50 sm It is better in limits 5 10 sm Press Value Unit 2000 mis 250 mis 760 mis On the screen the table should appear In the table of group of results allocate on columns 0 193 inus oo US If it is message about error necessary to close and iterate transmission has appeared The instruments remote controls blinking fluorescent lamps can influence stability of link working number IR lf you have for some reason or other decided not to transmit the data to leave fr
16. he screen is in waiting mode of measuring crack Put the device on the object s surface parallel to crack on the distance 10 15 mm from it Push it with the force of about 5 10 kg The device will activate itself The change of indications is accompanied with short beeps The time of ultrasound propagation in the object without crack in microseconds The time is measured until the device is removed from the surface Press for obtain the meaning as a base value 12 Ultrasonic tester UK1401 User manual The Base value is added in the device memory Put the device on the surface perpendicular to crack Depth of crack meaning measured in metrics units 12 OPERATIONS WITH MEMORY Record of measured results Time or Velocity of Ultrasonic Waves Propagation in the Memory Go to the main menu Choose Memory On or Off item Press Enter until you have the necessary adjustment see below Figure in the MEMORY line means quantity of free cells Return to the measurement mode 001 means number of group 004 means number of the result in this group Put the device on object the way that probes are perpendicular to the surface Push it with force of about 5 10 kg The device will pass in the active state The change of indications is accompanied by short beeps Fix the device in such condition during 15 20 seconds so the indications will change 2 3 times 13 Ultrasonic tester UK1
17. ing The value of this time is in the device memory and is written down its passport distance between probes The devices are made with base 150 or 120 mm however at assembly the deviation in 1 2 can be admitted Exact distance is underlined in the device passport This value should be set at tuning data transfer to the computer deleting result from the memory all result will be in Metrics unit all result will be in English unit shows the number of free memory sells while memory 18 switched automatic gain control Ultrasonic tester UK1401 User manual 8 DEVICE SCREEN IN THE DIFFERENT MEASUREMENT MODES Mode of Velocity and Time of Ultrasonic Waves Propagation Measurement with Turned OFF Memory wee para et se HHH E PEEL CITIT TES Time of ultrasonic waves propagation in the material Indication level of locked on signal in this case device is removed from the object surface Battery symbol while the battery is exhaustion the silhouette is getting lighter Dimension of measuring result Mode of Velocity and Time of Ultrasonic Waves Propagation Measurement with Turned ON Memory Number of the current group up to 255 Number of the result in the group up to 255 Ultrasonic waves velocity propagation in the material Level indication of the locked on signal when the rectangular is getting more darker the sig
18. nal is getting more stronger Loudspeaker symbol Sound indication is switch on Battery symbol while the battery is exhaustion the silhouette is getting lighter Dimension of measuring result Ultrasonic tester UK1401 User manual Mode of Crack Depth Measurement Time of ultrasonic waves propagation in the material with crack Time of ultrasonic waves propagation in the material without crack Crack depth Level indication of the accepted signal when the rectangular is getting more darker the signal is getting more stronger Loudspeaker symbol Sound indication is switch on Battery silhouette while the battery is exhaustion the silhouette is getting lighter 7 Dimension of measuring result crack depth NS Automatic Gain Control AGC Mode At the turned on AGC the ultrasonic tester UK1401 sets the necessary amplification automatically But in this mode the increased press of the device to the objects surface is needed When the AGC is OFF the signal level should be set manually The mode of manual gain control provides more exact readings of the device at minimal pressing it to the object s surface In this mode instead of the indicator of the level of the locked on signal on the screen value of gain in decibels which can be changed by keys upwards and downwards and Manual gain control of the locked on signal and as consequence independence of effort of force of a clip of the device to obje
19. nteen monolithic structures in Moscow The concrete of about 28 day age was tested on solidity The dependence will be the following R 0 016 V 27 3 23 Ultrasonic tester UK1401 User manual The relative residual standard deviation of this dependence is 16 and the correlation coefficient is 0 78 4000 4500 5000 S D S N N lt 3500 The dependence described in the 2 of this article can be used for approximate estimation of concrete s solidity from 15MPa up to 40MPa 4 It is recommended to specify the dependence for specific circumstances of the testing with the following equation i l y concrete s solidity in a specific part of the concrete that is detected with the help of avulsion with break method or the solidity of concrete s sample Ry the same but detected under the dependence given in the item 2 of this part the number of tested areas or the number of concrete s samples that should be not less then 5 items The particular values of A should be in the range 0 7 1 3 y Testing and detection of concrete s solidity in structures 1 The number and location of tested areas on the structures should be set accordingly to the GOST 18105 86 or should be set accordingly to work program agreed with the developer the author of the tested structure or developers of these recommendations Also the number and location of these areas
20. of ACS Ltd Customer shall not disclose provide or otherwise make available such trade secrets or copyrighted materials in any form to any third party 18 Ultrasonic tester UK1401 User manual Customer cannot make any changes in objective codes of programs except the changes made by elements of Software package and described in the Documentation Limited Warranty Acoustic Control Systems Lid warrants the quality of the data on the electronic medium the substantial operation of the Program when used in accordance with Documentation the accordance of the components of the Software package to the specifications and the quality of documents edition Except as expressed above the software is provided AS IS without warranties of any kind express or implied warranty with respect to the programs documentation including any implied warranty of fitness for a particular purpose ACS Ltd does not warrant that any program will be error free or that any defects that may exist in any program can be corrected Each Recipient is solely responsible for determining the appropriateness of using the Program and assumes all risks associated with its exercise of rights under this Agreement including but not limited to the risks and costs of program errors compliance with applicable laws damage to or loss of data programs or equipment and unavailability or interruption of operations In no event shall ACS Itd be liable fo
21. om the mode of reception it is possible having pressed the button To interrupt in the window of data transfer License Agreement The software enclosed with this agreement and related documents edited materials and users manuals collectively the Documentation are licensed exclusively to the Acoustic Control Systems Ltd ACS Ltd Subject to the terms of this Agreement ACS Ltd grants the Customer a non exclusive license to use the Program solely for his own internal purposes and in accordance with the accompanying Documentation if other is not agreed Customer can not distribute this Software that means sell the software or copy rent lease and also can not alter merge modify decompile disassemble the Program any Subset of the program Documentation or any copy in whole or in part Customer can make one 1 copy of the Program exclusively for back up or archive purposes no other copies be made Customer can make his own additional data bases for the program included in the software package if this is agreed Customer can use the Software installed in a single location only on one computer at the same time if no other is agreed You cannot have more then one operating version full or in part on several computers Customer agrees that aspects of the licensed material including the specific design and structure of individual programs constitute trade secrets and copyrighted materials
22. ou can return to the measurement mode with selected imaging view of results 10 VELOCITY OR TIME OF ULTRASONIC WAVES PROPAGATION DEFINITION Go to the main menu t With Arrow Up button go to the very beginning of the menu Press Enter until you have the necessary adjustment see oo Sm ae H BES fam H e s H Waiting mode of velocity measurements will be screened Put the device on object surface the way that probes are perpendicular to the surface Push it to the surface with the force of about 5 10 kg The device will active itself The change of readings is accompanied with short beeps Hold the device in this position during 15 20 seconds so that indications will change 2 3 times lf there are no signals it means that the ultrasonic velocity in the object is less than 1500 or there is a crack in the material between probes 17 Ultrasonic tester UK1401 User manual Measurement of the velocity in the object Display after removal of the device from the object surface In 10 15 seconds after removal of the device from the object surface the horizontal strokes will appear on the screen 11 DEFINITION OF CRACK DEPTH LAYING ON A SURFACE Go to the main menu With Arrow Up button go to the very beginning of the menu Press Enter until you have the necessary adjustment see below Select the view of results screening Return to the measurement mode T
23. ould not exceed 2 from the single average meaning of results of measurements for the tested area The results that do not meet this requirement are not taken into the account at counting the simple average velocity time of ultrasound dissemination in these areas 22 Ultrasonic tester UK1401 User manual 8 The calibration dependence is built taking as unit values the simple average of ultrasound velocity in the area and concrete s solidity of the area detected with the Avulsion with break method or with testing on the samples method 9 Building of the calibration dependence testing it and estimation of the error are made accordingly to the methods given in the Attachment 4 to the GOST 17624 87 The example of building the calibration dependence and estimation of its error are given in the attachment 5 of GOST 17624 87 is admitted to build the linear calibration dependence in the following view R a t b V orR a b where R is solidity V and T are accordingly velocity or time of ultrasound dissemination without the culling of single results using PC software such as MSExcel The correlation coefficient of the calibration dependence should be not less then 0 7 5 the meaning of the standard deviation In some cases after R average agreement with the developers of these recommendations it is admitted to use the ee S MHM calibration dependence with 0 2 R
24. r any loss of profits use business or for any incidental indirect special consequential or exemplary damages whatsoever including but not limited to damages resulting from loss of anticipated savings or lost data 14 WARRANTY ON THE PRODUCT Acoustic Control Systems Ltd ACS Ltd warrants the device for 1 year from the date of sale excluding the transducer The standard warranty period for the transducer is 3 months from the date of sale If something happens to the device during the warranty period for reasons covered by the warranty ACS Ltd at Its option will e Repair the Product or e Replace the product if is unable to repair it This warranty does not cover damages due to external causes including accident usage not in accordance with product instruction misuse The warranty also looses its validity if the customer tried to repair the device himself or under mechanical effect the device became nonrepairable As the tester UK1401 is the complex product and the repair of its electronic units requires special equipment the firm manufacturer does not recommend to repair the device independently Because of this the firm does not deliver the users the electrical circuit of the device and repair documentation Warranties lose force if the user independently tried to repair electronic units of the tester or if the device has undergone to strong mechanical influences as the result of which became pluck out Accumulator delivered
25. rcelain 5340 Caprolan 2787 Cr 6845 Kapron 2640 Zn 4170 Quartz fused 5930 Zirconium 4900 Constantan 5240 Cast iron 3500 5600 Labradorite 44 5450 Ebonite 2400 Brass 4430 Epoxide resin 2080 Brass LS 59 1 4360 Ice 3980 Manganin 4660 Mg 5790 5561 Marble 6150 26
26. s on memory reviewing and correction of the result from current group Switches on memory reviewing and correction of the results with listing group numbers 7 FULL DEVICE MENU TIME VELOCITY CRACK MEMORY LINK SOUND FREQUENCY DELAY BASE OUTPUT CLEAR METRICS ENGLISH AGC O 4000 ON OFF 1 2 3 4 5 6 11 9 150 ON OFF Ultrasonic tester UK1401 User manual List the menu items Active menu libe begins blinking and 20 C TIME VELOCITY CRACK MEMORY LINK SOUND FREQUENCY DELAY BASE OUT DATA CLEAR a ENGLISH MEMORY AGC The item s parameters of the main menu are changed with pressing Enter button On the device screen not all menu items are displayed therefore arrows are showing that there are some more lines higher or below the temperature inside the device in this mode time of ultrasonic waves propagation is determined in this mode velocity of ultrasonic waves propagation is determined in this mode depth crack is determined shows the number of free memory sells while memory is switched for operation in the mode of flaw detection switch on off accompaniment of the different device condition frequency repetition of sound device impulses 1 6 correspond to frequency from 5 to 25 Hz hardware time delay time of passing compensation ultrasound oscillation in a protector It is individual parameters which connect to design features and set at tun
27. should be set taking into account the following conditions Aims of testing detection of real concrete s solidity detection of solidity after taking off cast framework from the structure Special features of structure s operation bending compression and other 24 Ultrasonic tester UK1401 User manual Conditions of testing e Type of structure s reinforcement e Availability or absence of testing sample cubes 2 The velocity time of ultrasound dissemination on each area subjected to testing is made at least two times The deviation of single measurements from the single average should meet the requirements described in the item 7 of Preparations for testing The concrete s solidity is detected by the average meaning of received ultrasound velocity time When choosing the locations of areas subjected to testing you need to take into consideration item 6 of Preparations for Testing 3 When testing the solidity of concrete structures in which the concrete is up to 56 day old the structure s age should not differ from the average age of samples or testing areas used for building the calibration dependence on more then 25 When testing the solidity of older concrete this difference should not exceed the range of parts or sample s ages that were used for building calibration dependence To detect concrete s solidity in explotable structures it is necessary to use the calibration dependence tha
28. t is built right before the testing 4 The concrete s solidity of structures tested area is detected by the calibration dependence that is built accordingly to the part Preparation for Testing The measured ultrasound velocity time should lie between the minimal and maximal meanings of ultrasound velocity in samples or parts of structures that were used for building calibration dependence The received values of concrete s velocity are taken for average solidity of concrete of Ri part of structure 5 For detection of concrete s type class based on the results of testing you need to use GOST 18105 86 requirements For more information and explanation for these recommendations please contact Acoustic Control Systems Ltd 119048 Moscow POBox 148 tel fax 7 095 244 3194 244 2535 245 5896 e mail market acsys ru website www acsys ru 25 Ultrasonic tester UK1401 User manual 16 TABLE OF VELOCITIES OF LONGITUDINAL ULTRASONIC WAVES IN MATERIALS M S Al 6260 Cu 4700 Al alloy D16T 6320 Mo 6290 Basalt 5930 Ni 5630 Concretes 2000 5400 Sn 3320 Berillium 12800 Os 5478 Bronze 3530 Plexiglas 26 0 Vn 6000 Polystyrene 2350 2180 Rubber 1480 W 5460 Pb 2160 Gabbro 38 6320 Ag 3600 Gypsum 4790 Glassceramics 6740 Gneiss 7870 Mica 7760 Granite 4450 Steel3 5930 Diabase 85 5800 Organic glass 2550 Dolomit 9 4450 Soda lime glass 5500 Fe 5850 Ta 4235 Au 3240 Textolite 2920 Limestone 6130 Teflon 1350 Limestone 86 4640 Po
29. ting 1 For determination of concrete s solidity in structures you need preliminarily to build the calibration dependence 2 Calibration dependence is based on the results of parallel testing with two different ways ultrasonic testing and avulsion with break method of the same parts of concrete structures under standard GOST 22690 88 or it can also base on the results of ultrasonic testing of structure s parts and testing of samples cut from the same parts of structure accordingly to GOST 25870 90 is also possible to built the calibration dependence on the results of ultrasonic testing of cube samples and following their testing under press Cubes must be in the same environmental conditions as the structures ultrasonic testing on these cubes must be taken in the same conditions as the testing of structures themselves 21 Ultrasonic tester UK1401 User manual Building of calibration dependences based on the results of samples testing is made accordingly to the standard GOST 17624 87 When you build the calibration dependence based on the results of parallel testing with two methods Ultrasonic method and Avulsion with break method or testing the samples cut from the prepared for testing structures you need to remember the following things The ultrasonic testing is made on the subjected to general testing structures or in the areas of structures The points of the minimal and maximal velocity or time of
30. tures These recommendations were made as addition to GOST 17624 87 and have the main rules of testing the compressive strength of concrete in monolithic structures with the help of surface sounding method Application of surface sounding method 1 The method of surface sounding be used for testing the concrete s solidity at the moment when the cast formwork is taken off or for testing the concrete s solidity in terms determined by the project or technical documentation and also it can be used for engineering monitoring of exploitable or reconstructed monolithic structures 2 The concrete s solidity is detected by the following experimentally established calibration dependences Velocity of ultrasound dissemination at surface sounding concrete s solidity and time of ultrasound dissemination at surface sounding concrete s solidity 3 Surface sounding can be used for testing solidity both of heavy and light concrete of 7 5 to B50 types the calibration dependences should meet the demands described in the item 9 of Preparations for Testing of this Instruction see below 4 Ultrasonic Testing can be made with the help of devices that meet the requirements of Standard GOST 17624 87 of RF and that can provide time and velocity measurements of ultrasound dissemination on 120mm base or larger Is recommended to use the probes with dry point contact for example UK1401 Preparations for Tes