Home

NOVA

image

Contents

1. FILTER OUT WAVE PD300 PD300 PD300 3A IS WAVE PD300 LENGTH 3W UV LENGTH IR 250 N A N A 1mW N A 800nm 20mW 350nm 400nm 30mW 30mW 3mW N A 1 1 3ym 30mW 633nm 20mW 20mW 3mW IW 1 4um 25mW 670nm 13mW 13mW 3mW 2W 1 5um 15mW 800nm 10mW 10mW 2 5mW 3W 1 Bum 10mW 900mm 10mW 10mW 2 5mW 3W 1 8um 25mW 10600m 25mW 25mW 3mW 3W FILTER IN WAVE PD300 PD300 PD300 3A IS WAVE PD300 LENGTH 3W UV LENGTH IR 250 N A NA 100mW N A 800nm 200mWw 350nm 400nm 300mW 3W 300mW NAA 1 1 3um 300mW 633nm 300mW 3W 300mW N A 1 4um 150mW 670nm 200mW 3W 300mW N A 1 5um 80mW 800nm 100mW IW 250mW NA 1 Bum 50mW 900mm 150mW 1 5W 250mW NA 1 Bum 100mW 1060nm 250mW 2 5W 300mW N A Table 2 Maximum Measurable Laser Power as a Function of Wavelength 5 2 Setting up the PD300 and 3A IS to display the user s chosen wavelengths The PD300 and 3A IS series have built in wavelength correction curves for measurements either with the removable filter installed filter in or removed filter out These curves are stored in the head EEROM the correction curves with a resolution of 1nm ensure that the power reading is correct at all laser wavelengths In order to simplify changing from one laser wavelength to another the user can program up to 6 different wavelengths to be available from the screen menu Please use the following procedure to
2. 550g Display 122 x 32 pixel Super twist LCD Display digit height Backlight EL Operates from charger power only Bargraph segments Battery 2X RR Sub C 1 8Ah NiCd Built in Operation between 18 Hrs 10Hrs with charges pyroelectric heads Battery charge time10 14 Hrs 15 30 Hrs if operating Charger DC 11 to 22V or AC 9 to 15Vrms 3 Watt 89 9 2 Head Specifications MAX MAX AVG ABSORB POWER POWER ER WATTS DENSITY TYPE PD300 UV I R PD300 3W 50W cm 3A IS 200W cm Int Sph PD F100A IS 100W 200W cm Int Sph Ther A E BB w w 30 150 A 30 150 W 1000W cm HE HE1 T Ww w B bio u auer PE 2W sowim Peso 20W Tue Table 6 Max Power Specifications of Heads PD Photodiode P P type volume absorber for short pulse lasers HE HE1 volume absorber for high energy pulses EX Excimer type volume absorber PE pyroelectric metallic or black absorber BB broadband surface absorber high power density LP broadband surface absorber for highest power density 90 Absorber Max Energy Density J cm Type Pulse Length 10ns lus 300us P 10 10 10 HE HE1 5 10 100 BB 0 3 0 5 5 EX 0 5 0 6 4 PE Metallic 0 1 0 2 4 PE BB 0 3 0 3 1 PE DIF 1 5 3 40 PE BB DIF 3 3 10 Table 7 Maximum Energy Densities for Various Absorbers 3 A ENERGY DENSITY IN J CM Po 8 eszes a
3. 1 To go from the bargraph power measurement screen to the energy measurement screen press the menu button then press energy Alternatively you can press the leftmost soft key directly from the bargraph screen to go to energy measurement To go from the energy measurement screen to power measurement press the power soft key 2 2 2 Setting and Saving Startup Configuration 1 From the power measurement mode press the menu button located on the right side until more appears Press more Press select until configure is highlighted Press go Press value to choose power or energy for the desired startup screen Press select until the laser type is high lighted Now press value to select the laser type you wish to be the default Press select and value to choose the manual power range you wish to be the default or choose autorange Now press select and value again to choose the default energy range Press exit then all to save all present settings Further details on configuration in section 4 3 2 2 3 Power or Single Shot Energy Measurement Warning Do not exceed maximum head limits for power energy power density and energy density as listed in tables 6 and 7 in section 9 2 Otherwise there is a risk of damaging the absorber To use the Nova to measure Laser power 1 Verify that the display shows power units in W mW etc If the display shows energy units of J mJ et
4. 3 Press value to choose power or energy for the desired startup screen The Nova will now start up in power or energy measurement mode respectively when switched on 4 Press select until the laser type is highlighted Now press value to select the laser type you want to be the default The Nova will now be correctly calibrated for the absorption of the detector at the wavelength of the laser type selected 29 Press select and value to choose the manual power range you wish to be the default or autorange The Nova will now be in the mode selected when switched on with the head in question Now press select and value again to choose the default energy range Now press exit when all selections have been made Now press all if you wish to save all functions presently selected or previous screen to change the defaults of the previous screen only 4 4 Power Measurement The next section describes the procedure for basic power measurement as well as more advanced features with thermal measuring heads 4 4 1 Power Measurement Basic Operation 1 Switch the Nova on by means of the switch on the left side See Figure 4 The Nova will enter the mode of operation last saved with the configure operation For basic operation you need only set up the parameters for the type of measurement you wish Section 4 3 describes the procedure to store your configuration and the Nova will be ready to perform
5. 5 Now press select and value again to choose the default laser wavelength 6 Press exit when all selections have been made and press all to save all current Nova settings If you only want to save the last changes made on the previous screen press previous screen 44 5 4 Selecting Settings from the Nova Screen The Nova can conveniently be set to various chosen parameters from the screen In addition these settings can be saved as startup defaults so the head will be in the desired configuration when turned on 5 4 1 To Set to the Laser Wavelength Being Used Photodiode heads have a different sensitivity at different wavelengths Moreover the filters used in the head have a different transmission at different wavelengths In order to compensate for this each head has a built in calibration curve with 1nm resolution over the measurement range When you choose the correct laser wavelength the correction factor for that wavelength is automatically introduced Note that the laser wavelength presently corrected for is displayed in the upper right corner of the display In order to define the laser wavelengths to choose from See Section 5 2 In order to choose the laser wavelength do the following 1 Fromthe bargraph measurement screen press the menu button twice and press laser until the correct laser type or wavelength is displayed 2 Return to bargraph screen by pressing the menu button again 5 4 2 To
6. 3 Press exit and then press the menu button twice to return to the bargraph measurement screen Measuring loss using the dB offset function Since dBm is a logarithmic measurement the ratio between two measurements will be the difference between the dBm measurements For instance if you want to measure the loss in a fiber optic cable where the measurement before the cable is 1mW 0dBm and the measurement after the cable is 0 1mW 10dBm The ratio is then 1 10 0 1 and the dB loss is 0 10 10dB The dB offset function allows you to easily measure this To do so do as follows 1 When measuring the reference value press dbofst The value changes to 0 dB note that now the units are dB a relative value instead of dBm an absolute value 46 2 Now make your second measurement and the value of the difference in dB ratio in numerical units will be shown Note If there is a zero offset in the reference value you cannot subtract this using the DBOFST function Instead before the start of the measurement go into a numerical scale press offset and subtract the zero offset Then go into the dBm scale and follow steps 1 and 2 above The zero offset activated in the numerical scale will be saved in the dBm scale and you can now use the DBOFST setting to measure true ratio without zero offset problems 5 4 3 Operation with Filter In or Out The PD300 head is equipped with a built in filter so that
7. When the Nova is switched on again it will automatically go into the same mode until a different configuration is saved ica s POWER MORE RANGE GO TO POWER OPTIONS CHANGE MEASUREMENT ENERGY RANGE ee ALSO FLASHES READY FOR NEXT PULSE Figure 8 Energy Mode Screen 4 5 2 Measurement To measure energy of a single pulse set up as above and fire the laser The display will go blank while the energy is being integrated After 2 4 seconds the correct energy will be displayed When the Nova is ready to measure a new pulse ready will appear on the screen and flash on and off See Figure 8 The next pulse will automatically blank the screen and the new value will be displayed If you fire another pulse before ready appears the reading may be inaccurate or may not be displayed 37 4 5 3 Minimum Energy Threshold If the Nova is used in a noisy environment or where there is a high level of background thermal radiation the instrument may trigger spuriously on the noise or background radiation It would then fail to measure the intended pulse Since there is always some degree of noise or background radiation the instrument is designed not to respond to pulses below some preset minimum size This Minimum Energy Threshold is typically set to 0 396 of full scale of the selected range If this level is found to be too sensitive for the user s particular environment it may be altered by the user The threshold should
8. exposed to laser pulses 3 Instrument shows frequency which is too high Possible electromagnetic interference from pulsing laser is causing misreading and or false triggering Check the following 1 Head is mounted to stand using insulated plastic rod provided with instrument and not metal rod 2 Try keeping cable away from bench 3 Move head display further away from EMI 8 6 Maintenance 8 6 1 Maintenance of Thermal Heads 8 6 1 1 Tracing the Signal from the Head to the Nova With the instrument on apply an approximately known amount of power to the head This test can be performed using either a laser or an electrical power supply and the calibration resistor Estimate the approximate signal current that should be developed by the head by multiplying the input power by the head sensitivity shown in the head specification table Unplug the head from the Nova and check that this current appears between pin 1 and 9 of the D type plug 8 6 1 2 Disc Replacement If you have ordered a replacement disc it has been factory calibrated for the wavelength region shown on the calibration sheet sent with the disc 86 It has been calibrated at the wavelength requested or for the 3 wavelengths listed If it is a CAL type disc a CAL Factor is also included to use for electrical calibration checks on your power energy meter Please follow the following steps in disc replacement 1 Remove the screws
9. All wavelengths for diffuser IN have a D suffix e g 106D 1 06um diffuser in 4 f it is necessary to change the pulse length setting from the bargraph screen press the menu button twice press setup and press value to choose the correct pulse length Then press exit 5 If you wish to save these settings as the defaults from the main display press the menu button twice then press more Select config and make sure the settings are those you want Press exit and all to save all present settings including diffuser IN or OUT and pulse length setting 58 6 3 Startup and Configuration of Defaults If you are planning to do repetitive measurements most of the time you can set up the Nova so that it automatically goes into the measurement mode you want upon startup Since these desired settings are saved in the head smart connector you can save different settings for different heads To set the Nova to start up in power or energy measuring mode and to save the chosen Nova configuration 1 From the bargraph measurement screen press the menu button located on the right side until more appears Press more Press select until config is highlighted Press go Press value to choose power or energy for the desired startup screen The Nova will now start up in power or energy measurement mode respectively when switched on Press select until the laser wavelength is highlighted Now
10. Method of Calibration sentent 81 8 3 2 Accuracy of Calibration sse 82 8 3 3 Recalibration from a Known Source of Laser Energy 83 8 4 Error MESSAGES eege ee 84 8 5 Troubleshooting eene 84 8 5 1 Nova UDO asserit etre rre 84 8 5 2 Thermal Heads Energy Measurements ssssssssssssserseeeeee 85 8 5 3 Thermal Heads Power 8 5 4 Pyroelectric Heads eerte 8 6 Maintenance ttt ttt tts eetntec hepatis 8 6 1 Maintenance of Thermal Heads 96 8 6 2 Replacing Battery 88 Chapter 9 Nova Specifications 89 9 1 System Display Specifications sss 89 9 2 Head Specifications EE 90 Chapter 1 Introduction How to Use This Manual The Ophir Nova is a microprocessor based Laser Power Energy Meter providing a broad range of measurements displays and data handling options It operates with thermopile pyroelectric and photodiode heads and uses smart connector technology Just connecting the head configures and calibrates the instrument This manual tells you what you need to know to make full use of the Nova for all your laser measurement needs It includes a Quick Reference Chapter 2 to allow you to perform basic measurements immediately without reading the whole manual The main measurement sections Chapters 4 5 and 6 include a general description and a section detailing operating options Chapter 2 Quick Re
11. Nova will record data until the screen is full or reset or exit is pressed 5 4 7 Average and Measuring Very Low Power Measurement If the laser power is fluctuating the Nova can display the average power readings with averaging periods varying from 1 3s to 1 hour When measuring very low powers such as picowatt measurements using the PD300 IRG or PD300 UV there will be a rather large zero offset coming from the detector as well as a considerable noise fluctuation Nevertheless you can measure these low values by using the average function and pressing offset to eliminate the detector zero offset In order to measure very low powers do as follows 1 Select the lowest manual power range which is larger than the value you wish to measure 2 Go into the average screen by pressing the menu button twice then more then selecting average and pressing go 3 Select average over the appropriate period in seconds and press go 4 Now block the power source you wish to measure wait for a few measurement periods and press offset to subtract the zero offset 5 Now unblock the power source and measure For more about the average function See Sections 4 4 2 5 Note The average function does not work with autorange or dBm scales 5 5 BC20 Heads for Scanned Beams Note Many of the functions of the BC20 head are the same as the PD300 This section only describes the PD300 functions which are differe
12. and Photodiode Head Energy Screen 2 Press the menu button once then press range then the up down soft keys until the proper range is highlighted The correct range is the lowest one that is larger than the pulse energy measured 3 Press exit then press the menu button twice to return to the bargraph measurement screen If you wish to have the frequency displayed press the menu button only once 4 f you wish to display the average of a number of pulses press the menu button until setup is seen Press setup and then select until pulses sample is highlighted Press value to select the number of pulses you wish to average over This function can be permanently saved See also Section 6 2 2 2 and 6 3 61 The Nova is now ready to measure energy pulses The energy reading will operate in the following manner if none selected as the averaging period following every 1 15th of a second period the Nova will display on the screen the next pulse which arrives The Nova will hold the display of the latest pulse until a new one arrives If an averaging period is selected the Nova will collect pulses at up to 15Hz and will then display the average of those pulses When a new pulse triggers the Nova a T will appear in the upper left corner of the screen At over 5 Hg the T stops flashing and appears all the time 6 4 1 2 Minimum Energy Threshold If the pyroelectric heads are used in a electrically or acoustical
13. choose Manual or Automatic Ranging or dBm Autorange The Nova allows you to choose autorange or manual ranging when the reading is in watts In autorange mode you do not have to change scales When the reading of the meter or bar is more than 10096 of full scale the range goes to the next higher one When the reading falls below 9096 of full scale the range changes to one range lower This change occurs after a short delay This provides overlap hysteresis to keep the Nova from flipping back and forth when reading close to the end of the scale 45 Manual Range There are certain disadvantages to autorange since it changes scale even if you don t want it to do so If you want to measure the same range all the time it is better to use manual range The correct range to select is the lowest one which is larger than the largest expected measurement If you wish the Nova to be in manual mode when switched on save this configuration before switching off See Section 5 3 dBm The Nova allows the measurement to be made in units of dBm which is a logarithmic scale dBm units are defined as 10 x log reading in mW At 1mW the reading will be 0 dBm at 100mW it will be 20 dBm etc To choose auto manual or dBm range follow these steps 1 From the bargraph measurement screen press the menu button once 2 Press range and then select the appropriate manual range autorange or dBm by pressing the up or down buttons
14. display will now say in scope mode Select the appropriate measurement range in the usual way You can return to the ordinary measurement mode by pressing exit from the scope mode screen If you wish the instrument to start up in scope mode press the menu button twice and press more Select config and press go Select display and press value It will change to scope Press exit and all to save the new configuration 6 5 Measuring Repeating Pulses of High Energy Because of their construction pyroelectric heads are restricted in the energy density they can withstand particularly for short pulses on the order of nanoseconds If the energy density of your laser exceeds the rating of the pyroelectric absorber there are several options available 1 You can use Pyroelectric heads with a diffuser Ophir has several heads with a diffuser which increases the damage threshold by an order of magnitude See table 7 at the end of this manual You can enlarge your laser beam using a negative lens until the energy density is below damage threshold You should test this using the test slide See section 6 4 You can use the Ophir beam splitter which mounts the PE heads and splits off typically 8 10 of the light If you use this method note that there may be polarization effects You can calibrate the beam splitter by running the laser at an energy below damage threshold and measuring the energy with and without the beam
15. from the front and rear flanges of the absorber head Remove the front flange and raise the rear flange taking care not to tear the wires attached to the connectors Unsolder the wires from the faulty absorber disc attached to the BNC connector in the head remembering which wire is connected to each terminal Pry the absorber disc free from its seat by using a screwdriver in the hole provided at the edge of the seat Smear the absorber disc seat evenly with silicone heat sink compound Feed the wires through the hole and install the new disc It is best to turn the disc back and forth a few times while pressing on it to ensure good contact between the disc grease and head Reassemble the front flange making sure the screws are tight Resolder the two wires to their proper locations on the rear flange and reassemble the rear flange 8 6 1 3 Recalibrating the Head to a new Disc 1 From the bargraph power measurement screen press the menu button twice then press more Press select until calibrate is highlighted Press go then power with CAL Press laser and select the first laser wavelength sent with the disc Now press go and adjust up down until the sensitivity agrees with that sent with the disc Now press exit to save the new value For heads with calibration at several laser wavelengths you now have to adjust the calibration for the other wavelengths 87 From the bargraph power measurement screen p
16. head to display the user s chosen wavelengths The sensitivity of the detector varies somewhat with wavelength The correction curve for the absorber is stored in the head EEROM This correction curve ensures that the power reading is correct at all laser wavelengths In order to simplify changing from one laser wavelength to another the user can program up to 6 different wavelengths to be available from the screen menu Please use the following procedure to set the pyroelectric or PD10 head to your laser wavelengths 1 From the bargraph display press the menu button located on the right side until more appears Press more 2 Press select until wavelength is highlighted Press go 3 Press select change then up and down to change the first wavelength desired Press done Repeat for other wavelengths desired up to 6 Now press exit 4 Now press more then select until config is highlighted Press go exit then all to save the wavelengths you have set If you wish to save other defaults in the config screen See Section 6 3 Note The broadband type heads BB have less variation with wavelength and in those heads fixed wavelength ranges are selected similar to thermal heads This section is not relevant with those heads 55 6 2 2 Setup of Laser Pulse Length and Pulses Sample 6 2 2 1 Laser Pulse Length As was mentioned before the Ophir pyroelectric heads can measure long as well
17. heads have only one time setting for all pulse lengths In that case N A appears 2 4 4 Setting Startup Configuration 1 From the bargraph energy measurement mode press the menu button until more appears Press more Press select until config is highlighted Press go Press value to choose power or energy for the desired startup screen Press select until the laser wavelength is highlighted Now press value to select the laser type you wish to be the default 5 Press select and value to choose the pulse length you wish to be the default 6 Now press select and value again to choose the default energy range 7 Press exit then all to save For further details on configuration See Sections 6 2 and 6 3 2 4 5 Setting up PE DIF diffuser heads to diffuser IN or OUT setting 1 Pressthe menu button twice and press setup 2 Press select until the diffuser setting is highlighted and select IN or OUT as desired Make sure the diffuser is physically installed or not installed on the head Press exit 3 Press the menu button twice and press laser until the correct wavelength is selected Note that only wavelengths compatible with the diffuser setting are visible All wavelengths for diffuser IN have a D suffix e g 106D 1 06um diffuser in 4 f you wish to save these settings as the defaults from the main display press the menu button twice then press more Select config a
18. not however be raised higher than necessary This will cause degradation in the accuracy of energy measurements of pulses below about 4 times the threshold level The factory setting of energy threshold is med for medium If the unit triggers on noise set the threshold to high If you are measuring small energies and the unit does not trigger set the threshold to low To set the energy threshold 1 From the energy measurement screen press more 2 Press select until threshold is highlighted Press go Energy Threshold is Now med will be displayed 3 Press value to select high med or low threshold For the 3A P head optical is also available and is the default choice See below 4 Press exit to return to main measurement screen If you wish to save this value save all with the configure function before switching the Nova off Section 4 3 38 Note to users of 3A P type heads The 3A P type head has special circuitry enabling it to trigger on energy pulses of very low energy This is accomplished by a special photodiode trigger which alerts the Nova that a pulse has been received and to start integration This allows the 3A P to react to smaller pulses than would be measurable just by triggering on the rising thermal signal as in other heads When you select optical this enables the optical trigger However in some cases the optical trigger will not work and you should select one of t
19. on and off fire the laser See Section 4 5 for full details To save configuration 1 To save configuration follow directions in Section 2 2 2 above 2 3 Photodiode Heads 2 3 1 Use of Nova with photodiode type heads To set type of laser being used 1 From bargraph measurement screen press the menu button twice and press laser until the correct laser wavelength is displayed Return to bargraph screen by pressing the menu button again To choose manual or automatic ranging or dBm in power measurement 1 From the bargraph measurement screen press the menu button once Press range and then select the appropriate manual range autorange or dBm logarithmic scale Press exit and then press the menu button twice to return to the bargraph measurement creen Alternately range is accessed directly by pressing the right most soft key from the bargraph screen 2 3 2 Selecting Chosen Wavelength 1 From the power measurement mode press the menu button until more appears Press more Press select until wavelength is highlighted Press go Press change then up and down to select the first wavelength Repeat steps 2 and 3 for other wavelengths desired Up to 6 wavelengths may be selected 2 3 3 Setting Startup Configuration 1 From the power measurement mode press the menu button until more appears Press more Press select until configure is highlighted Press go Now
20. power meter divided by the laser frequency 81 The spectral absorption of the detector coating is measured spectroscopically and the absorption curve is used to correct the calibration for other wavelengths When the user selects his wavelength on the Nova the correction factor for that wavelength is applied The PD10 heads are calibrated in a two step fashion First the photodiode detector filter combination are calibrated against a NIST traceable master in a similar fashion to the PD300 heads over the wavelength range of the head Then the head is calibrated at one wavelength using a 905nm repetitively pulsed laser referenced to a NIST traceable photodiode meter The average energy is set to the average power of the standard power meter divided by the laser frequency 8 3 2 Accuracy of Calibration Since the instruments are calibrated against NIST standards the accuracy is generally 3 at the energy level and wavelength at which the calibration has been performed This accuracy has been verified by checking the scatter of the results when several instruments are calibrated against the same standard The maximum error in measurement will be less than the sum of the specified accuracy linearity and inaccuracy due to errors in the wavelength curve The non linearity is approximately 296 and the error due to wavelength is given in table 5 below In addition to the above errors the reading of pyroelectric heads changes with fre
21. press value to select filter in or out to be the default Press select and value to choose the manual power range you wish to be the default or choose autorange Now press select and value again to choose the startup laser wavelength For further details on configuration See Section 5 3 2 3 4 Power Measurement Warning Do not exceed maximum head limits for power energy power density and energy density as listed in tables 6 and 7 section 9 2 Head Specifications Otherwise there is a risk of damaging the absorber To expand the bargraph scale 10 about the present reading 1 From the bargraph power measurement screen press the menu button Then press zoom Now press the menu button twice again 2 Alternatively press the center button from the bargraph screen without the menu button Press zoom again to return to full scale To offset current reading and set to zero 1 From the bargraph power measurement screen press the menu button press offset then press menu twice to return to the bargraph screen Alternatively press the left upper button from the bargraph screen Press offset again to cancel See Section 5 4 5 for more details To use the Nova to fine tune laser power 1 From the bargraph power measurement screen press the menu button twice then press more Press select until tune is highlighted Press go With the left soft key set the percentage range of the power s
22. press value to select the laser wavelength you want to be the default The Nova will now be correctly calibrated for the absorption of the detector at the wavelength of the laser type selected If you wish to choose another wavelength See Section 6 2 1 Press select and value to choose the default pulse length and energy range Now press exit when all selections have been made Now press all if you want to save all current settings or previous screen if you want to save only the settings previously set In addition to the choices made above the Nova will now save all other functions as they have been chosen at this point such as laser pulse width number of pulses for average power and energy graph settings and so forth Therefore before this point it is best to set up all other functions also 59 6 4 Energy Power or Exposure Measurement Warning Before using the head for power or energy measurement check that your laser energy energy density and average power do not exceed the head ratings as listed in the table with the head specifications at the end of the manual otherwise there is a risk of damaging the absorber With the pyroelectric head you have been supplied a test slide with the same coating as on your pyroelectric detector You can also obtain this slide from your dealer You should use this slide to test the damage threshold with your laser pulses If the slide is damaged then either enlarge your
23. that type of measurement each time it is switched on The normal default mode is the bargraph autoranging power measurement screen If the Nova is in energy mode pressing the power soft key will cause it to enter the power measurement mode Center the laser beam carefully on the absorber surface and read the power 30 4 4 2 Advanced Power Measurement Features The following section describes additional Nova functions in power mode in addition to the basic power measurement mode described above 4 4 2 1 Zoom The Zoom function causes the scale to expand the present scale fivefold Thus if the full scale of the bargraph is 20 watts and your reading is 15 watts pressing zoom will make the bargraph scale range between approximately 13 and 17 watts Variations in power are more easily seen in this mode To access the zoom function press the menu button until zoom appears Then press zoom When it is highlighted it is engaged Now press the menu button twice to return to the bargraph power display Note that the fiducial marks of the bargraph are further apart indicating that zoom is engaged You may also directly engage the zoom function from the bargraph screen by pressing the middle button Note that while the button is pressed the soft key legend appears Pressing zoom again will cancel the function 4 4 2 2 Offset The offset function subtracts background from the signal If the ambient environment has a therm
24. 3 9 Chart Recording 3 10 Beam Splitter Function 4 2 3 43 44 44 1 44 2 4 5 4 5 1 4 5 2 4 5 3 4 5 4 4 5 5 4 5 6 Chapter 5 Operation with Photodiode Type Heads 5 1 To choose Manual or Automatic Ranging in Power Measurement To choose Power or Energy Measurement Startup and Configuration of Defaults E 29 Power Mesasurement EEN 30 Power Measurement Basic Operation Advanced Power Measurement Features Energy Measurement a ET dE Measurement ts ccsccsccsscececsceceacgeacecateiatesnatesdcstoaascotscosasesarebiantoas Minimum Energy Threshold as Measuring Pulses of Very Low Energy 39 Measuring Energy of Rapidly Repeating Pulses 40 Energy E0g reete cta meh Mesas Photodiode Ee 5 6 2 Setting up the PD300 and 3A IS to display the user s chosen Waveletigttis ttt tee erret reete eben Startup Defaults E Selecting Settings from the Nova Screen To Setto the Laser Wavelength Being Used Tune and Powerlog o Average and Measuring Very Low Power Measurement 50 BC20 Heads for Scanned Beams sss 50 General Description s Setting up the startup deiaults cesses 51 Measuring Scanned Beams with the BC20 51 Spectral Response ceo eae eee Special Photodiode Heads E PD 300 BB Radiameter Head PD300 CIE Photometer Head 6 1 Pyroelectric and Photodiode Energy Detector Heads 6 2 Selecting Sett
25. It is calibrated at the fixed wavelengths that are available with that instrument 71 8 2 2 Accuracy of Calibration of PD300 and 3A IS Heads Since the instruments are calibrated against NIST standards the accuracy is generally 2 at the wavelength the calibration has been performed The maximum error in measurement will be less than the sum of the calibration accuracy linearity inaccuracy due to errors in the wavelength curve and variations in gain with temperature The linearity of the photodiode detector is extremely high and errors due to this factor can be ignored The maximum error due to the above factors is given in Table 4 below Wave Error Filter Out Length PD300 PD300 PD300 PD300 3A IS 3W UV IR 200 250nm 5 250 360nm 3 360 400nm 10 10 3 400 950nm 3 3 3 5 950 5 5 5 5 10 1100nm 1100 5 1600nm 1600 7 1800nm Table 4 Maximum Error as a Function of Wavelength and Filter Add 2 to error for filter in 3 for PD300 3W and PD300 UV 8 2 3 Accuracy of calibration of BC20 Heads The sensitivity of various BC20 sensor varies from one head to another as well as with wavelength Therefore each BC20 head is individually calibrated at each specified wavelength against a NIST standard which has been calibrated at the relevant wavelengths 78 Since the instruments are calibrated agains
26. OPHIR Laser Measurement Group LASER POWER ENERGY MONITOR USER MANUAL OPHIR OPTRONICS www ophiropt com Table of Contents Chapter 1 Introduction How to Use This Manual 4 Chapter 2 Quick Beierence EEN 5 2 1 Getting Started 2 2 Thermal Heads 2 2 1 Use of Nova with thermal type heads EE 7 2 2 2 Setting and Saving Startup Configuration 8 2 2 3 Power or Single Shot Energy Measurement H 2 3 Photodiode Heads 10 2 3 1 Use of Nova with photodiode type heads 10 2 3 2 Selecting Chosen VWavelength sss 11 2 3 3 Setting Startup Configuraton sss 11 2 3 4 Power Measurement as 24 Pyroelectric and Photodiode Energy Heads 12 24 1 Selecting Chosen VWavelength sss 24 2 To set type of laser being used 24 3 To set Laser Pulse and Average 2 4 4 Setting Startup Configuration 2 4 5 Setting up PE DIF diffuser heads to diffuser IN or OUT tune 14 2 4 6 Energy or Average Power Measurement 14 Chapter 3 The Nova Display Unit 17 3 1 General Description 17 3 2 Smart Connectors and Multihead Operation 18 3 3 Soft Keys and Menu Button 18 3 3 1 Menu Button au 3 32 Soft E 34 Power Up and Shut Down 3 5 Nova Functions which are independent of Head Type 3 5 1 Line Frequency 3 52 Zero Adjustments e 3 5 3 Baud Rate nee enses terret 3 6 Backliglt niic enini nite REHR ee Cere eet e 3 7 Kickstand 3 8 Charging
27. This means that when the head is plugged in the Nova automatically identifies the head type calibration and configuration The user does not have to adjust anything Note The Nova automatically loads the head information when first turned on so when changing heads the display should be turned off the new head plugged in and then be turned on again When no head is plugged in and the Nova is turned on the display indicates Head Disconnected and gives the user the opportunity to change the power line frequency or rezero the instrument See section 3 4 3 3 Soft Keys and Menu Button 3 3 1 Menu Button The Nova is equipped with soft keys That is the functions of the keys change as indicated by the legend above each key See Figure 3 When the Nova is first switched on the first screen usually has a digital display with a bargraph along the bottom In order to access the soft keys press the menu button located in the second row on the right of the panel Pressing the menu button again will access more functions Pressing it yet again will bring it back in a cyclical manner to the original bargraph screen Note The first set of menu choices can be accessed without pressing menu Simply pressing the appropriate key from the bargraph screen will operate that key even though the label is not visible So if you remember the position of the key you can take a shortcut If the soft key is an alternating toggle type th
28. When a uniform signal such as room light background falls on the detector head the signal from the two detectors cancels On the other hand when a laser beam falls on the head it illuminates only the first detector and therefore is detected Thus the PD300 subtracts most of the background while detecting the desired signal The subtraction is not perfect but usually 9896 of the background signal is eliminated so the detector can usually be used in ordinary laboratory lighting conditions The Nova display unit amplifies this signal and indicates the power level received by the head Due to the superior circuitry of the Nova the noise level is very low and the PD300 series heads with the Nova display have a large dynamic range from nanowatts to hundreds of milliwatts Since many low power lasers have powers on the order of 5 to 30mW and most photodiode detectors saturate at about 2mw most heads of the PD300 series have been constructed with a built in filter so the basic head can measure to 30mW or more without saturation When the additional filter is installed the maximum power is on the order of 300mW or 3W with model PD300 3W 42 The PD300 saturates when the output current exceeds 1 3mA so the exact maximum power depends on the sensitivity of the detector at the wavelength used When saturated the legend SAT will appear on the screen Table 2 gives the actual maximum power as a function of wavelength
29. ailable for the PE50 and PE25 80 High speed broadband type The type with the BBH suffix has a special broadband coating spectrally flat from 190nm to Zum Unlike the BB type it is able to work at high repetition rates to 1000Hz On the other hand its damage threshold is considerably lower on the order of 0 05J cm PD10 type Unlike the other heads the PD10 has a silicon photodiode with a neutral density filter mounted permanently in front of it The filter detector combination are calibrated over the entire wavelength range similarly to the PD300 power heads and therefore the heads have a high accuracy at any wavelength in the range This is an exclusive feature with Ophir energy heads Absorption 96 100 e BB Broadband S Metallic Coating 04 1 10 100 Wavelength nm Graph 2 Absorption of Ophir Pyroelectric Absorbers 8 3 1 Method of Calibration The sensitivity of the various Ophir pyroelectric sensors can vary from one to another as well as with wavelengths Therefore Ophir pyroelectric detectors are individually calibrated against NIST traceable standards In addition the calibration is corrected in the Nova for different wavelengths Ophir pyroelectric detectors are calibrated using a 1 06um repetitively pulsed laser referenced to a NIST traceable thermal power meter The average energy is set to the average power of the standard
30. al background so that the Nova shows a nonzero power reading even when there is no laser you can subtract the background using the zero function For example the Nova display reads 0 1 Watts when the laser is blocked and 20 5 Watts with laser power applied In this case the true power is 20 5 0 1 20 4 Watts To subtract the background press the menu button until offset appears and press offset while the laser is blocked 31 Now press the menu key again to return to the bargraph screen The Nova will now read zero and the 0 1 Watt background will be subtracted from all subsequent readings The laser power reading will thus be 20 4 Watts When offset is engaged the legend is highlighted To cancel press offset again If the offset is engaged and you wish to subtract out a new value of the background press offset twice The first press will cancel the old value and the second will activate a new value f you suspect that the Nova has a permanent zero offset then disconnect the head while itis in the power measurement mode If the Nova still shows a similar offset even when the head is not connected the instrument internal zero should be reset See Section 3 5 2 4 4 2 3 Tune See Figure 5 The Tune function an exclusive Ophir feature makes adjusting your laser to its maximum power easier than ever before Unlike a bargraph or mechanical meter this display shows graphically what came before as well as th
31. al filter which speeds up the effective response time of the head and rejects noise The digital processor is also responsible for controlling the analog processor reading the keypad and driving the display The CPU obtains calibration and capability data from a memory chip in the plug of the detector head and configures itself accordingly Recalibration data and saved settings are also stored in this memory chip 67 Power Supply The power supply board which also holds the display module and the keypad switches provides the internal DC voltages for the analog and digital processors It also contains the battery charging circuit and the AC supply for the backlight All of the power supplies operate in switch mode with an oscillator frequency of 32KHz Electromagnetic Interference The Nova and associated heads have extensive circuitry both to reject outside electromagnetic and electrostatic interference The Nova and associated heads are fully CE qualified and are extremely resistant to EMI If there is still some interference in an unusually high EMI environment it is recommended to use the Nova without the charger plugged in 68 Chapter 8 Calibration Maintenance and Troubleshooting 8 1 Calibration of Thermopile Heads 8 1 1 Absorber types and Method of Calibration of Ophir Power Meters 8 1 1 1 Types of Ophir Laser Absorbers Two types of absorber surface are used in Ophir thermal measuring heads 1 Surface Abso
32. as short pulses In order to do this the user must indicate to the Nova if the laser pulses are going to be longer or shorter than the maximum wavelength in the short pulse mode Some models only have only one maximum pulse length setting which is usually 3ms In that case N A for not applicable appears and the setting cannot be changed Warning If the pulse length is incorrectly set to the short pulse setting for long pulses the reading will be erroneously low If it is set to the long pulse setting for short pulses the reading will be correct but noisy To set up for pulse length please do the following 1 From the bargraph measurement screen press the menu button twice and press setup 2 Now press value to select proper setting for your laser pulse length If your laser pulses are longer than the short pulse setting then the long pulse setting should be selected and if shorter then the short setting should be selected 6 2 2 2 Setup for Number of Pulses per Sample The Nova has the ability to measure a number of pulses and display the average value of the energy of the pulses This function is useful if the laser energy is not stable Then the user can choose to measure a number of pulses and display only the average Note The Pyroelectric and photodiode energy heads are capable of measuring pulses up to very high repetition rates on the order of kilohertz or above 56 However at high repetition rates
33. beam or lower the laser energy until damage is no longer seen Note To measure pyroelectric energies properly it is important that the head is not grounded to the optical bench Make sure that the head is isolated electrically from the ground The PE head has been supplied with an insulating mounting post for this purpose 6 4 1 Energy Average Power and Frequency Measurement Unlike most pyroelectric and photodiode energy meters the Nova measures the pulse frequency to a high degree of accuracy and is therefore able to display pulse frequency and average power as well as pulse energy The following paragraph will describe how to set up and measure these functions 6 4 1 1 Energy Measurement Ordinarily when the instrument is switched on it is in the bargraph energy measurement mode This can be verified by the fiducial marks at the bottom of the screen and the units of mJ uJ etc See figure 9 60 How to set up for pulse energy measurement is described below 1 Make sure the instrument is in the bargraph energy measurement mode indicated by the fiducial marks at the bottom of the screen and the units of mJ uJ etc If itis in power measurement mode with units of mW W etc then press the left soft key or alternatively press the menu button and then the left energy soft key The Nova can be setup to start up in energy or power measurement mode when turned on See section 6 3 3 456 Figure 9 Pyroelectronic
34. blocked and 20 5 mW with laser power applied In this case the true power is 20 5 0 1 20 4 mW To subtract the background press the menu button until offset appears and press offset while the laser is blocked The Nova will now read zero 0 0 and the 0 1 mW background will be subtracted from all subsequent readings The laser power reading will thus be 20 4 mW When offset is engaged the legend is highlighted To cancel press offset again If the offset is engaged and you wish to subtract a new value of the background press offset twice The first press will cancel the old value and the second will activate a new value If you suspect that the Nova has a permanent zero offset then disconnect the head while itis in the power measurement mode If the Nova still shows a nonzero reading even when the head is not connected the instrument internal zero should be reset See section 3 5 2 5 4 6 Tune and Powerlog The Tune function an exclusive Ophir option makes adjusting your laser to its maximum power easier than ever before Unlike a bargraph or mechanical meter this display shows graphically what came before as well as the current reading and the trend This allows you to see at a glance if you have reached maximum power 49 The Nova also has the option of graphing the laser power vs time or successive energy points as long as the Nova has not been turned off While measurements are being taken the
35. c then press power to switch to the power measurement mode To expand the bargraph scale 10 about the present reading 1 From the bargraph power measurement screen press the center button Press the center zoom button again to return to full scale To subtract background and set current reading to zero From the bargraph power measurement screen press the menu button twice then press offset Press the menu button once to return to the bargraph screen Press offset again to cancel See Section 4 4 2 2 for full details To use the Nova to fine tune Laser power 1 From the bargraph power measurement screen press the menu button twice then press more Press select until tune is highlighted Press go Set the percentage range of the power scale to be displayed by repeatedly pressing the left key Setthe horizontal sweep time using the middle soft key See Section 4 4 2 3 for full details To use the Nova to measure Laser energy 1 In measurement mode verify that the units are J mJ etc If not press the menu key until energy appears Press that key to switch to energy measurement mode Alternately press the left soft key directly 2 The energy mode is manual ranging Press range then the up down soft keys until the proper range is highlighted then press exit The correct range is the lowest one that is larger than the pulse energy measured When the Nova screen flashes ready
36. cale to be displayed Setthe horizontal sweep time using the middle soft key See Section 4 4 2 3 for full details 2 4 Pyroelectric and Photodiode Energy Heads 2 4 1 Selecting Chosen Wavelength 1 While the instrument is switched off plug in the head then switch on From the bargraph screen which appears press the menu button until more appears Press more Press select until wavelength is highlighted Press go Press select then change then up and down to change the first wavelength Now press done Repeat steps 2 and 3 for other wavelengths desired up to 6 Save configuration as described in Section 2 4 4 2 4 2 To set type of laser being used 1 From bargraph measurement screen press the menu button twice and press laser until the correct laser type or wavelength is displayed Return to bargraph screen by pressing the menu button again 2 4 3 To set Laser Pulse and Average 1 From the bargraph measurement screen press the menu button twice then press setup 2 Press value to choose the short pulse or long pulse setting depending whether the pulses of your laser are less than or greater than the short setting 3 Press select then value to choose the time period over which you wish to average energy readings Press exit 4 Save configuration as described in section 2 4 4 Warning Incorrect readings will result if pulse length is not set up correctly Note Some
37. calibration factor provided by Ophir Calibration instructions 1 Set the instrument to power If you want to recalibrate energy this can be done from a known energy source See section 8 1 4 3 2 From the bargraph power screen press the right soft key and set to the proper manual power range Press exit 3 Press the menu button twice then more Now press select until calibrate is highlighted Press go 4 Press power with CAL 5 Using the laser button select the proper laser wavelength See note 7 below When laser wavelength is set press go 6 Using a DC power supply apply power to the CAL resistor black and red terminals at the back of the head For highest accuracy use the same power as you intend to measure but do not exceed values in Table 3 Section 8 1 4 1 7 Maximum voltage applied should not exceed resistance of CAL resistor x maximum power 75 8 The voltage should be measured at the input to the head and the current in series with the current loop 9 Press go and adjust up down until the power reading on the screen reads the voltage in volts x current in amps x current calibration factor for your model and laser or as shown on Nova Note 1 For units which have different calibration factors e g C02 YAG or VISIBLE press laser to select correct type of laser before calibration When recalibrating using the CAL resistor the other two settings will change proport
38. d in both cases using an integrating sphere The detector is then calibrated with an argon laser and given a correction factor from this ratio 8 1 2 Linearity and Accuracy of Ophir Thermal Heads 8 1 2 1 Linearity The linearity of most Ophir thermal detectors is specified to be 196 over the specified power range of each particular instrument The linearity is tested by electrical substitution heating Models of up to 30 watts maximum rating are subjected to electrical input up to their maximum rating models of up to 300W maximum rating are heated up to 100 watts and models of 1000W rating and above are heated up to 200 watts On CAL models the built in CAL resistor is used On non CAL models a CAL resistor is temporarily affixed to the sensor For those models for which the linearity is not tested over their entire range randomly chosen sample models are tested periodically over their entire range 72 The test is performed with a high power laser that can cover the entire detector range using a rear leak detector for comparison This rear leak detector is a low power Ophir detector which has previously been tested for linearity Thus in all cases the linearity of the detectors is traceable to electrical standards 8 1 2 2 Total Accuracy of Calibration Since the instruments are calibrated against NIST standards the accuracy is generally 196 at the power level at which the calibration has been performed This accuracy has be
39. e current reading and the trend This allows you to determine if you have reached maximum power z50 7 min EXIT T T Push soft keys to make legends visible Figure 5 Tune Screen The screen is completely autoranging Therefore as soon as the cursor goes over the top or under the bottom edge of the display it rescales to put the cursor back to the middle of the screen This allows you to devote all your attention to tuning the laser without having to worry about the Nova settings 32 To use the Nova to fine tune laser power 1 From the bargraph power measurement screen press the menu button twice then press more 2 Press select until tune is highlighted Press go 3 Press the menu button once to make legends visible Press again to reset and return to measurement screen 4 Set the power expansion scale to 50 120 or 20 as desired using the left soft key The default value is 50 If the laser power will change a lot use a large value if a little use a small value 5 Setthe horizontal sweep time to the rate desired using the middle soft key The default value is 1 minute Both the expansion and time settings are saved for future use when using the configure function See section 4 3 6 Measure the power of the laser and adjust the laser until you determine that the power has reached a maximum 4 4 24 Powerlog See figure 6 The Nova in Powerlog mode has the option of graphing the lase
40. e label will be visible while the key is being pressed 3 3 2 Soft Keys The soft keys accessed by the menu button have functions defined by the legend above the key The legend usually indicates what will happen when pressing the key For example if energy appears above a key pressing that key will change the Nova into energy mode Some functions operate when the key is pressed and are canceled when the key is pressed again The keys show reverse highlighting when operational Pressing the same key again cancels the operation and the highlighting Nova Top View ON NOVA 2 345 SLIDE SOFT wich AAA ON OFF q MENU or e BUTTON Figure 3 Nova Rear Panel View BACK HEAD INPUT LIGHT LCD 12VDC O WHEW GS as 0000000 CONTRAST AC D TYPE RECORDER BACK LIGHT ADJ CHARGER CONNECTOR DC OUT PUSH BUTTON SOCKET FOR INPUT Figure 4 3 4 Power Up and Shut Down To turn the Nova on 1 Push up the slide switch on the left side of the Nova display See Figure 4 2 The unit will switch on and the display will appear If no head is connected a display showing Head Disconnected will appear If a head is connected the appropriate default measurement screen will appear To switch the Nova off 1 Return the Switch to its original position 2 Ifyou wish to save the current Nova configuration use the configure function before switching off For measurements of the same type the Nova does not need to be se
41. ection 2 2 2 or refer to the main body of the manual for details To set line frequency 1 Disconnect the head and switch off then on again Head Disconnected will appear 2 Press select until line freq is highlighted 3 Press go then change to correct frequency 4 Press exit and change will be saved To zero instrument e Disconnect head turn off then on again Head Disconnected will be displayed Press select until zero is highlighted Press go Make sure instrument is not in an electrically noisy environment and is undisturbed Press go and wait for message zeroing completed Press exit For thermal heads zeroing with the head may also be necessary See section 3 5 2 2 2 Thermal Heads 2 2 1 Use of Nova with thermal type heads To set type of laser being used 1 While the Nova is off plug in the head then switch it on again From bargraph measurement screen press the menu button twice and press laser until the correct laser type is displayed Return to bargraph screen by pressing the menu button again To choose manual or automatic ranging in power measurement From the bargraph measurement screen press the menu button once then press range or simply press the right most soft key Select the appropriate manual range or autorange Press exit and then press the menu button twice to return to the bargraph measurement screen To choose power or energy measurement
42. en verified by checking the scatter of the results when several instruments are calibrated against the same standard The maximum error in measurement will be less than the sum of the specified accuracy and linearity Since the linearity is also 1 the maximum error in measurement will generally be less than 296 8 1 3 Calibration Factor For models equipped with the Ophir CAL resistor the calibration factor is determined at the factory by the following procedure After calibration of the sensor as described above in Section 8 1 1 2 electrical power is applied to the CAL resistor The calibration factor is given by the formula CAL factor reading on meter applied electrical power If you wish to check and adjust the power reading apply electrical power to the CAL resistor and measure the voltage and current to obtain the applied power The correct power reading for the meter is the applied electrical power multiplied by the CAL factor given above 73 8 14 Recalibration of Thermopil Heads Using the Nova If your measuring head is equipped with the exclusive Ophir CAL resistor then calibration is performed according to the instructions in Section 8 1 4 2 If you do not have the CAL resistor then you can recalibrate from a known source of power or energy following the instructions in Section 8 1 4 3 On the rear center of the sensor disc is the exclusive Ophir calibration heater The disc can be heated by electrical power
43. eturn to the measurement screen 5 4 4 Zoom The Zoom function causes the scale to expand around the present reading 10 Thus if the full scale of the bargraph is 20mW and your reading is 15mW pressing zoom will make the bargraph scale range between approximately 13 5 and 16 5 mW Variations in power are more easily seen in this mode When zoom is engaged the legend is highlighted To cancel the zoom function press zoom again The zoom function can be useful for laser power tuning and peaking To access the zoom function press the menu button until zoom appears Then press zoom When it is highlighted it is engaged Now press the menu button twice again to return to the bargraph power display Note that the fiducial marks of the bargraph are further apart indicating that zoom is engaged You may also directly engage the zoom function from the bargraph screen by pressing the middle button Note that while the button is pressed the soft key legend appears Pressing zoom again will cancel the function 48 5 4 5 Offset The PD300 PD300 3W and BC20 heads have automatic background subtraction as described in Section 5 1 In addition the offset function can be used to subtract the residual background signal which remains if desired If the Nova shows a nonzero power reading even when there is no laser you can subtract the background using the zero function For example the Nova display reads 0 1mW when the laser is
44. f not press range adjust and press exit 2 From the energy measurement screen press more and then select energy log Press go The histogram screen will appear Now fire a pulse when the ready legend flashes The measurement will immediately begin When the screen is full the measurement will stop and the graph will continue to be displayed Warning Pressing reset will erase all previously stored data and start recording new data 3 The left button is the reset button Press this button to reset the screen and begin a new measurement While pressing the button the legend reset is visible 4 f you wish to see finer variations in the graph press the center zoom button Note Zoom can only be pressed after at least one pulse has been recorded 5 Toreturn to the main energy measurement screen press exit 41 Chapter 5 Operation with Photodiode Type Heads Warning Before using the head for power or energy measurement check that your laser power or energy and energy density does not exceed the head ratings See table 6 5 1 Photodiode Heads When a photon source such as laser is directed at one of the PD300 3A IS or BC20 series photodiode detectors a current is created proportional to the light intensity and dependent on the wavelength The PD300 PD300 3W and BC20 heads have a unique dual detector head patented in which the two detectors are identical and connected back to back
45. ference 2 1 Getting Started The Nova is equipped with soft keys That is the functions of the keys change as indicated by the legend above each key See Figure 2 When the Nova is first switched on the first screen usually has a digital display with a bargraph at the bottom In order to access the soft keys press the menu button located in the second row on the right of the panel Pressing the menu button again will access more functions Pressing it yet again will bring it back in a cyclical manner to the original bargraph screen To connect head to the Nova Display Insert the D type connector of the measuring head cable into the socket marked Head Input on the rear panel of the Nova display See Figure 1 HEAD INPUT EAK uen 0 0 VS A oeoe o VE Can CONTRAST SHALOG BACK LIGHT wi CHARGER CONNECTOR OUTPUT PUSH BUTTON SOCKET FOR INPUT Figure 1 Nova Rear Panel View To switch the Nova on 1 Push up the slide switch on the left side of the Nova display See Figure 2 ON NOVA 2 345 MEEGY 200M OFFRET SLIDE ait AAA i ON OFF q MENU 9m e BUTTON Figure 2 Nova Top View The unit will switch on and the display will appear Note The head must be plugged in before the unit is switched on To switch the Nova off 1 Return the slide switch to its original position 2 If you wish to save the current Nova configuration use the configure function before switching off See below S
46. g on oa a oO o K A i v 2 Lo a a b i Se 3 n Si i T E 2 9 L S 3 X ser Sg sl 4 2 E Y EI m K eg g o x i Lr i og Hi E Se Cw POWER DENSTYW CM BS d Nova User Manual Ophir P N 1J06013 20 March 2005 Rev 3 27 1 91
47. ge or autorange by pressing the up or down buttons 3 Alternately you may directly access range from the bargraph screen by pressing the right most soft key 4 Press exit and then press the menu button again to return to the bargraph measurement screen 28 4 2 3 To choose Power or Energy Measurement The Nova thermopile heads can measure both power and single shot energy See sections 4 4 and 4 5 for details To change from power to energy measurement or vice versa do the following 1 To go from the bargraph power measurement screen power to energy measurement press the menu button then press energy Alternately you can press the left most soft key directly from the bargraph screen to go to energy measurement 2 To go from the energy measurement screen to power measurement press the power soft key 4 3 Startup and Configuration of Defaults If you are planning to do measurements of the same type most of the time you can set up the Nova so that it automatically starts up in the measurement mode desired Since these settings are saved in the head smart connector you can save different settings for different heads To set the Nova to start up in power or energy measuring mode and to save the chosen configuration 1 From the power measurement bargraph screen press the menu button located on the right side until more appears Press more 2 Press select until configure is highlighted Press go
48. he menu button again 21 4 2 2 To choose Manual or Automatic Ranging in Power Measurement Autorange The Nova allows you to choose autorange or manual ranging In autorange mode you do not have to change scales When the reading of the meter or bar is more than 100 of full scale the range goes to the next higher one The ranges are arranged in factors of 1 10 100 etc When the reading falls below 90 of full scale the range changes to one range lower This change only occurs after a few seconds delay This provides overlap hysteresis to keep the Nova from flipping back and forth when reading close to the end of the scale Manual Range There are certain disadvantages to autorange since it changes scale even if you don t want it to do so If you want to measure the same range all the time it is better to use manual range To select manual range press the menu button twice until range appears Press range then press up or down to select the desired manual range The correct range to select is the lowest one which is larger than the largest expected measurement Now press exit to return to the measurement screen If you wish the Nova to be in manual mode when switched on save this configuration before switching off See Section 4 3 To choose auto or manual range follow these steps 1 From the bargraph measurement screen press the menu button once 2 Press range and then select the appropriate manual ran
49. he laser energy until damage is no longer seen 6 1 Pyroelectric and Photodiode Energy Detector Heads When a pulsed heat source such as a laser is directed at the detector head a temperature gradient is created across the pyroelectric crystal mounted in the head An electric charge is produced which is proportional to the energy absorbed The detector head has sophisticated circuitry unique to Ophir patent applied for which determines the baseline before the pulse is received measures the voltage after a pre determined interval amplifies it and holds it for a pre determined time Due to this innovative circuitry Ophir pyroelectric heads can measure very long pulses as well as short ones They can measure low energies as well as high They can also measure at higher repetition rates than was possible before The Nova display amplifies this signal and indicates the energy received by the head as well as the frequency at which the laser is pulsing Using the energy and frequency information the Nova is also able to display average power 54 The photodiode based PD10 head also uses the same circuitry but can measure much lower energies due to the sensitivity of the photodiode 6 2 Selecting Settings from the Nova Screen The Nova can conveniently be set to various chosen parameters from the screen In addition it can be set to startup defaults so it will be in the desired configuration when turned on 6 2 1 Setting up the
50. he proper range is highlighted The correct range is the lowest one that is larger than the pulse energy measured Press exit then press the menu button twice to return to the bargraph measurement screen If you wish to have the frequency displayed press the menu button only once 2 4 6 3 To use the Nova to measure Average Power 1 Verify that the display shows power units in W mW etc If the display shows energy units of J mJ etc then press power to switch to the power measurement mode 2 4 6 4 To expand the bargraph scale fivefold about the present reading From the bargraph energy or power measurement screen press the menu button then press zoom then press the menu button twice or just press the center button from the bargraph screen Press zoom again to return to full scale 2 4 6 5 To measure Total Energy Exposure 1 Press the menu button until more appears Press more Press select until exposure is highlighted Press go Select parameters then press go Press go again The Nova will start summing laser energy exposure and the legend will change to stop When you wish to stop measuring press stop If you wish to reset reading to zero before another reading press reset To return to the bargraph screen press exit Chapter 3 The Nova Display Unit 3 1 General Description The model Nova laser power energy meter represents a new level of sophistication sensitivity compactne
51. he thermal trigger settings low med or high You should use thermal triggering on the 3A P in the following cases 1 The wavelength is outside the range of the of the photodiode i e outside of 0 19 1 0641 m 2 You are measuring the energy of a shuttered pulse of a continuous laser In this case the peak power is too low to trigger the photodiode 3 For some other reason the instrument does not trigger on the optical setting When using the optical trigger the instrument should be able to measure down to 0 1 0 2 mJ depending on wavelength When using the thermal trigger low med or high you should not attempt to measure less than about 1mJ 4 5 4 Measuring Pulses of Very Low Energy When it is necessary to measure pulses of very low energy i e less than 0 5 of the maximum range of the instrument the following two alternative methods allow greater accuracy to be obtained 1 A continuous train of pulses may be fired and the average power measured using power mode The energy per pulse can be calculated by Energy per pulse Average power Pulse Repetition Rate 39 2 A train of a known number of pulses may be fired and the total energy measured in energy mode This train should not exceed 5 seconds duration The energy per pulse can be calculated by Energy per pulse Total Energy Number of Pulses In both of the above methods the pulse repetition rate must exceed 3Hz Higher rates wi
52. ings from the Nova Sereen 6 2 1 Setting up the head to display the user s chosen Wavelengths e enee 622 Setup of Laser Pulse Length and Pulses Sample m 6 3 Startup and Configuration of Defaults 6 4 Energy Power or Exposure Measurement 641 Energy Average Power and Frequency Measurement 60 6 4 2 Pane 6 4 3 Measuring Total Energy Exposure x 6 4 4 Pyro Scope Adapter sette 6 5 Measuring Repeating Pulses of High Energy 65 6 6 Energy og 66 Chapter 7 Circuit Description 11111111111ssrssrsseessssse BI Chapter 8 Calibration Maintenance and Troubleshooting 8 1 Calibration of Thermopile Heads 8 1 1 Absorber types and Method of Calibration of Ophir Power cem 69 812 Linearity and Accuracy of Ophir Thermal Heads 72 8 1 3 Calibration Faetor ap tene 73 814 Recalibration of Thermopil Heads Using the Nova 74 8 2 Calibration of Photodiode type Heads 71 8 2 1 Method of Factory Calibration ae 8 2 2 Accuracy of Calibration of PD300 and 3A IS Heads Latte 78 8 2 3 Accuracy of calibration of BC20 Heads u 78 8 24 Recalibration of PD300 and 3A IS Heads from a Known Source of Laser Power 79 8 2 5 Recalibration of BC20 heads from a Known Source of Laser gu lie 80 8 3 Calibration of Pyroelectric and Photodiode Energy Heads 80 8 3 1
53. instead of by radiation by applying electrical power to the terminals at the rear of the sensor head By accurately measuring the voltage and current applied the calibration can be checked and if necessary readjusted taking into account the calibration factor See Section 8 1 3 8 1 4 1 Guidelines for using the CAL resistor for calibration 1 Do not exceed the maximum electrical power stated in the table Model Power Aprox Max Voltage 3A Series 1 Watt 7 Volts 10A Series 10 Watts 17 Volts 30A Series 30 Watts 35 Volts 150A Series 100 Watts 65 Volts 300W Series 125 Watts 65 Volts 1500W 200 Watts 95 Volts Series and above Table 3 Maximum Power to Apply to Sensor CAL Resistors 74 2 Use the resistor for calibration only Turn off the power as soon as calibration is done 4 Do not operate the calibration resistor when the sensor disc is under radiation the excessive heat may burn out the calibration resistor 5 To eliminate residual non linear effects it is advisable to perform electrical calibration at about the same power level as that to be measured so long as the rating of the heating resistor is not exceeded e 8 1 4 2 Recalibration of Power using the Built in CAL Resistor If your thermal head has the Ophir CAL resistor then you have a traceable absolute calibration source available just by measuring the voltage and current into the CAL resistor and multiplying by the
54. ionately If you wish to change the calibration of only one laser type use the function Power with laser Note 2 The unit retains for your information the current sensitivity of the head in A W in engineering notation The original value is retained at the factory You should record the original setting so you can return to it if you wish 8 1 4 3 Recalibration from a Known Source of Laser Power Energy 1 Setthe instrument to power or energy depending which you wish to recalibrate 2 Settothe proper manual power or energy range 3 From the bargraph power screen press the menu button twice and press more more is directly accessible from the energy measurement screen Now press select till calibrate is highlighted Press go 4 Press power energy with laser 5 Using the laser button select the proper laser wavelength See note 1 below When laser wavelength is set press go 6 Applythe known laser power energy 7 Adjust up down until the power energy reading on the screen equals the known power energy The power energy calibration factor will change accordingly See note 1 8 Press exit The new value will be saved 76 Note 1 For units which have different calibration factors e g C02 YAG or VISIBLE press laser to select correct type of laser before calibration When recalibrating using power with laser function only the calibration of the chosen laser wavelength will change If you w
55. ious screen 5 5 3 Measuring Scanned Beams with the BC20 1 Verify that the head is plugged in 2 Settothe proper power range or autorange 51 3 Set up the startup defaults and choose continuous if you want continuous readings or hold if you want readings to be held for 5s before updating See section 5 5 2 above 4 Hold the wand so the beam strikes the detector vertically While the beam is scanning move the scanner slightly up and down in a direction perpendicular to the scan direction so that the scanned beam is sure to cross the center of the detector The BC20 captures the highest reading so that the most accurate results are obtained if the beam scans across the center of the detector The unit will also measure static beams as would a normal laser power meter 5 5 4 Spectral Response A graph of the approximate relative spectral response of the BC20 is given for purpose of interpolation This is in case the user wishes to use the instrument at a wavelength other than the ones which are factory calibrated 100 p i 70 S 60 50 40 N 20 10 ae 0 340 440 540 640 740 840 940 1040 Wavelength nm 5 6 Special Photodiode Heads 5 6 1 PD 300 BB Radiometer Head In addition to photodiode head for individual wavelength Ophir also supplies heads for measuring the output of broadband light sources The PD300 BB head has s
56. is only changed for the wavelength set 83 8 4 Error Messages The Nova displays various error messages when operated outside its normal range Over range When the power or energy being measured exceeds the range of the measurement scale being used the over range message is displayed but the reading still appears on the display If the power or energy exceeds the maximum by more than 1096 the reading on the display is blanked Low Battery When the battery is almost discharged the flashing message BAT appears When the low battery message appears the Nova should be connected to the charger It will operate normally and charge slowly while connected to the charger When connected to the charger while turned off the Nova will charge faster in about 18 hours Sat When the photodiode current exceeds 1mA and the detector starts to saturate the message sat is displayed 8 5 Troubleshooting 8 5 1 Nova Display Problem Cause Remedy Instrument will not Check that charger operate after being provides 200mA completely discharged and connected to charger Instrument operates with charger but not with battery Battery is low Recharge overnight with the Nova turned off for 14 16 hours If the Nova still doesn t work with battery then the NiCd battery is dead Replace battery 84 8 5 2 Thermal Heads Energy Measurements Problem Cause Remedy Instrument triggers on backgrou
57. ish to change the calibration of all wavelengths proportionately then use the function power with CAL Note 2 When changing power calibration this will also change energy calibration proportionately However changing energy calibration will not change power calibration 8 2 Calibration of Photodiode type Heads Photodiode detectors are inherently very linear but also have a large variation in sensitivity with wavelength In addition the Ophir model PD300 is equipped with both a built in filter and removable filter to allow measurement of higher powers without detector saturation These filters also have a transmission which depends on wavelength Therefore the PD300 Nova has a built in calibration adjustment for wavelength which is described in the next paragraph 8 2 1 Method of Factory Calibration The sensitivity of various Ophir photodiode sensors varies from one head to another as well as with wavelength Therefore Ophir photodiode detectors are individually calibrated against a NIST standard which has been calibrated at several nm intervals over the entire spectral range The spectral sensitivity curve of the detector both for filter out and filter in is fed into the head EEROM and this information is used to set the gain to the proper value at all wavelengths When the user selects his wavelength on the Nova the correction factor for that wavelength is applied The BC20 head does not have a calibration curve
58. larly high damage threshold for pulsed and repetitively pulsed lasers of both the short and long pulse variety and are useful where the highest pulse energies and average powers are used HL high energy long pulse absorber The HL absorber is the absorber of choice for high energy repetitive pulses longer than 2ms For these long pulses it has flatter spectral response than the HE HE1 and withstands higher pulse energy and average power 70 Percent absorption Ke a 0 1 1 10 zc BB therm A EX 14 P 100 i Percent absorption 0 1 1 Wavelength um 40 9 HE m HE1 amp LP x HL Graph 1 Absorption of Ophir Thermal Absorbers vs Wavelength 71 8 1 1 2 Method of Calibration The absorption of the various Ophir thermal absorbers can vary from disc to disc Therefore all Ophir absorbers are individually calibrated against NIST traceable standards Ophir power energy meters with the broadband or P type absorbers are individually calibrated by laser at several wavelengths against a NIST calibrated standard meter The meter can be switched to give the exact calibration at the various wavelengths Argon YAG and where applicable C02 or for some models an average value is given The EX type detector is calibrated by measuring the ratio of absorption in the UV to that at 515nm The total absorption is measure
59. ll generally give improved accuracy but care should be taken not to exceed maximum power ratings 4 5 5 Measuring Energy of Rapidly Repeating Pulses The standard Nova will only measure individual pulses every 5 seconds or so You can also calculate the average energy of rapidly repeating pulses by measuring average power on the power setting and using the formula Average Energy per Pulse Average Power Pulse Repetition Rate For rapidly repeating pulses you can use one of the Ophir pyroelectric heads as long as the pulse energies do not exceed the ratings of the pyroelectric absorbers The pyroelectric heads are compatible with the Nova and just have to be plugged in to be used For higher pulse energies you can use the Ophir model DGX RP to measure both average power and exact individual pulse energy of the rapidly repeating pulses See your Ophir agent or sales representative for details 4 5 6 Energy Log The Nova has the option of displaying successive energy points in a vertical bargraph form While measurements are being taken the Nova will record data until the screen is full or reset is pressed 40 An RS232 interface can be purchased to attach to a PC and then the data points can be transferred to PC either in real time or after storage Contact your Ophir representative for details To use the Nova for graphical display of energy 1 Make sure that the Nova is in the correct energy measurement range I
60. ly noisy environment the instrument may trigger spuriously It will then display a spurious pulse Since there is always some degree of noise or background the instrument is designed not to respond to pulses below some preset minimum size This Minimum Energy Threshold is typically set to 496 of full scale of the selected range Pulses which are very close to background may read sporadically or inaccurately Therefore it is important to always use the lowest energy range which is larger than the energies you are measuring 6 4 1 3 Power Measurement 1 Make sure the instrument is in the bargraph power measurement mode indicated by the fiducial marks at the bottom of the screen and the units of mW W etc If it is in energy measurement mode with units of mJ pJ etc then press the left soft key or alternatively press the menu button and then the left power soft key The Nova can be configured to start up in power measurement mode when turned on See section 6 3 2 Press the menu button once then press range then the up down soft keys until the proper range is highlighted 62 The correct range is the lowest one that is larger than the highest energy being measured Note that the proper range is related to the pulse energy even though we are measuring average power 3 Press exit then press the menu button twice to return to the bargraph measurement screen If you wish to have the frequency displayed press the menu button
61. me up in the normal measurement screen reverse the process in steps 1 and 2 Note 1 The factory default for this screen is 100 0096 and in that case the regular screen comes up on startup and not the attenuator screen Note 2 The attenuator screen only affects the value displayed on that screen and nothing else The power and energy scales remain the same the damage thresholds remain the same etc The attenuator screen is only an extra screen to show a different calculated number introduced by the user 25 Chapter 4 Operation with Thermopile Absorber Heads Warning Before using the head for power or energy measurement check that your laser power energy and energy density do not exceed the head ratings See tables in section 9 2 Head Specifications If the head is a water cooled type ensure that the cooling water is flowing at an adequate rate see table below Also note that the reflectance from the absorber could be as much as 10 and with CO2 lasers the reflected beam can be quite specular so it is advisable to provide a beam stop for the reflected beam with the highest power lasers HEAD LITERS MIN US TYPE PER PRESSURE GALLONS MIN BAR PER MINUTE 8000W 7 1 5 2 5000W 45 0 8 1 2 L1500W 2 5 0 5 0 7 1000W 14 0 4 0 5 300W 1 0 0 3 0 25 Table 1 Minimum Flow Rates For Water cooled Heads 4 1 Thermopile Absorber Heads When a radiant heat source such as a laser is directed a
62. ment is not in an electrically noisy environment and is undisturbed Press go and wait for message zeroing completed Now press exit and saved will be exhibited indicating that the zero configuration has been saved 21 Note for Thermal Heads only For best results with thermal heads it may be necessary to do the procedure once with the head disconnected then afterwards again with the head connected After completing steps 1 5 above Connect the head and make sure it is at room temperature and well shielded from any stray thermal power It may be best advised to lay the head with the absorber face down on the table 5 Switch the instrument off then on again with the head connected 6 Press the menu button twice then more Press select until zero is highlighted 7 Now repeat steps 4 and 5 above 3 5 3 Baud Rate If you are using the Nova for communication with PC the baud rate can be set by selecting baud rate pressing go and selecting the appropriate rate For most purposes the default of 9600 Baud will suffice 3 6 Backlight The backlight illuminates the display from the rear and is operated by a push button on the back panel See Figure 3 Since the backlight consumes considerable power it is operable only when the charger is plugged in 22 The electroluminescent backlight does have a finite lifetime of about 10 000 hours Therefore it is recommended only to use it when necessary To turn
63. n over its entire measurement range of 200 1100nm 57 The PD10 operates in a similar fashion to the PE head except that it has one extra function When measuring very low energies the head will also be sensitive to background light and therefore show a spuriously high reading Therefore we have added a function to measure the background light When measuring low energies block the laser and press bkgrnd You will see a reading of the background light This reading can be subtracted from your laser reading to obtain a more accurate reading Also making the room darker will reduce this background effect 6 2 2 4 PE DIF diffuser heads PE50BB DIF and PE50 DIF ER heads with removable diffusers have to be set up properly to operate with the diffuser IN or OUT Since the wavelength settings with diffuser in are not the same as with diffuser out two things must be changed every time the head is changed from diffuser IN to OUT in addition to changing the pulse length setting if necessary To change from diffuser OUT to IN and back 1 Pressthe menu button twice and press setup 2 Press select until the diffuser setting is highlighted and select IN or OUT as desired Make sure the diffuser is physically installed or not installed on the head Press exit 3 Press the menu button twice and press laser until the correct wavelength is selected Note that only wavelengths compatible with the diffuser setting are visible
64. nd press go Press exit and all to save all present settings including diffuser IN or OUT 2 4 6 Energy or Average Power Measurement Warning Do not exceed maximum head limits for power energy power density and energy density as listed in tables 6 and 7 section 9 2 Head Specifications Otherwise there is a risk of damaging the absorber With the pyroelectric head you have been supplied a test slide with the same coating as on your pyroelectric detector You can also obtain this slide from your dealer You should use this slide to test the damage threshold with your laser pulses If the slide is damaged then either enlarge your beam or lower the laser energy until damage is no longer seen 2 4 6 1 To choose Energy or Average 1 To go from the bargraph energy measurement screen to power measurement press the menu button then press power Alternately you can press the leftmost soft key directly from the bargraph screen to go to power measurement To go from the bargraph power measurement screen to energy measurement press the menu button then press energy Alternately you can press the leftmost soft key directly from the bargraph screen to go to energy measurement 2 4 6 2 To use the Nova to measure Laser Energy and Frequency In measurement mode verify that the units are mJ yJ etc If not press the left most soft key Press the menu button once then press range then the up down soft keys until t
65. nd noise or sometimes fails to catch large pulse Increase threshold level See Section 4 5 3 Instrument does not show ready for a long while after a reading is made Increase threshold level See Section 4 5 3 Non reproducible results when measuring very small energy pulses or no response to pulses at low energy Decrease threshold level See Section 4 5 3 8 5 3 Thermal Heads Power Problem Cause Remedy Instrument Check connections between shows zero the head and the instrument reading in both See signal tracing Section power and 8 6 Check that the sensor disc energy modes is operative Resistance between pins 1 and 9 of the head connector should be about 1 8k If the sensor is defective there will be an open or short circuit Instrument responds while head is cold but suddenly fails as Replace sensor disc See Section 8 6 1 2 it heats up Instrument does If head is very hot allow it to not return cool Disconnect the head completely to zero on power measurement from the instrument If readout unit does not zero adjust zero with head disconnected If the offset persists try zeroing with the head connected as well as described in Section 3 5 2 85 8 5 4 Pyroelectric Heads Problem Cause Remedy 1 Instrument reads incorrectly or erratically Especially on sensitive scale 2 Instrument triggers even without being
66. nt from ordinary PD300 functions For standard PD300 functions please refer to the rest of chapter 5 50 5 5 1 General Description Model BC20 scanned beam laser power meter has become the industry standard for measuring scanned laser beams as well as hard to reach static beams The BC20 smart head has built into its smart plug innovative circuitry to measure scanned as well as static beams of up to 20 milliwatts with a noise level of microwatts The BC20 has the same patented dynamic background subtraction as the PD300 which eliminates over 95 of background light and allows measurement in normal room light 5 5 2 Setting up the startup defaults 1 From the power measurement bargraph screen press the menu button located on the right side until more appears Press more Press select until configure is highlighted Press go Press select then value to choose whether you want the instrument to operate in continuous or hold mode In continuous mode the peak reading of every 1 3 sec period is displayed and in hold mode the peak reading of every 5s period is displayed Note Do not use autorange when in hold mode Now press select and value again to choose the default laser wavelength and the default power range Press exit when all selections have been made and press all to save all current Nova settings If you only want to save the last changes made on the previous screen press prev
67. ode works only with manual range From the bargraph power screen press the menu button and press range Select a manual range and press exit 2 Press the menu button again then press more Press select until average is displayed Press go 3 Now press value to select periodic or continuous mode as described above Press select and value to choose the time period desired Press go 4 The average will be displayed If you wish to stop temporarily press stop After stopping press go to continue If you wish to start over press reset the readings will be set to zero and then press start 5 If there is a zero offset see section 4 4 2 2 then you may subtract the offset by pressing offset 4 4 2 6 Attenuation Screen The Nova allows you to scale your measurement so as to display the original power reading of a laser beam which has been split or attenuated by a beam splitter or attenuator 35 For instance if your original laser power is 100 Watts and you have split off 5 of the power i e 5 Watts and are measuring only the 5 Watts you can display the nonsplit value with the attenuation screen To do so do the following 1 From the main screen press the menu button twice and press more Select atten and press go 2 Select attenuation screen ON and press exit The attenuation screen is now activated 3 Press set and by alternately pressing change and selec
68. on the backlight push the push button while the charger is connected To turn off the backlight push the button again 3 7 Kickstand The Nova is equipped with a kickstand on the underside of the case so the display can be tilted for easy viewing The kickstand folds away in a recessed slot when not in use When the Nova is folded flat it is only 203 x 95 x 35 mm the most compact unit of this type on the market 3 8 Charging The Nova is operated by a rechargeable battery To charge the battery plug the charger into the jack labeled 12VDC on the back panel Figure 3 Note the polarity of the charger The Nova will charge the batteries at about the same rate whether it is on or off and whether the backlight is on or off it fully charges in about 14 hours However it will charge more slowly while operating a pyroelectric type head It is not recommended to leave the charger plugged in for much longer than 14 hours in order to preserve the battery lifetime The unit will operate about 18 hours from one battery charge When the battery is low BAT flashes on and off indicating that the battery needs recharging However even with BAT displayed the unit will operate properly for about one hour before its accuracy will be impaired 23 3 9 Chart Recording The instrument provides an analog voltage output via a 2 5mm pin jack labeled AN OUT on the rear panel See figure 3 This is useful for driving chart reco
69. only once 6 4 2 Zoom The Zoom function causes the scale to expand around the present reading 10 Thus if the full scale of the bargraph is 25mJ and your reading is 15mJ pressing zoom will make the bargraph scale range between approximately 13 5 and 16 5 mJ Variations in energy or power are more easily seen in this mode When zoom is engaged the legend is highlighted To cancel the zoom function press zoom again The zoom function can be useful for laser power tuning and peaking To access the zoom function press the menu button until zoom appears Then press zoom When it is highlighted it is engaged Now press the menu button twice to return to the bargraph power display Note that the fiducial marks of the bargraph are further apart indicating that zoom is engaged You may also directly engage the zoom function from the bargraph screen by pressing the middle button Note While the button is pressed the soft key legend appears Pressing zoom again will cancel the function 6 4 3 Measuring Total Energy Exposure The Nova has the ability to sum the total energy of a number of pulses over a given time period or number of pulses This gives the total energy exposure over that time period For example if the laser is pulsing at 30 times s at 1mJ per pulse and you measure the exposure over 20 seconds then the total exposure is 30 x 1 x 20 600mJ 63 In order to measure exposure do the following 1 Pres
70. pectrally flat response from 400 to 1000 nm and therefore can give the true total power of any broadband light source in that spectral region 52 Note Because the BB heads are spectrally flat wavelength selection is disabled Other than that operation is the same as other PD300 type heads 5 6 2 PD300 CIE Photometer Head The PD300 CIE head has a spectral response similar to that of human eye and can therefore make measurements in eye response units of Lux The PD300 CIE is designed with a small detector where the source overfils the detector It measures the light intensity per unit area in units of Lux or Foot Candles Note Because the CIE heads have a spectral response similar to the human eye wavelength selection is disabled Other than that operation is the same as other PD300 type heads 53 Chapter 6 Operation with Pyroelectric and Photodiode Energy Heads Warning Before using the head for power or energy measurement check that your laser power energy or energy density do not exceed the head ratings as listed in the table with the head specifications Otherwise there is a risk of damaging the absorber With the pyroelectric head you have been supplied a test slide with the same coating as on your pyroelectric detector You can also obtain this slide from your dealer You should use this slide to test the damage threshold with your laser pulses If the slide is damaged then either enlarge your beam or lower t
71. quency The Nova has a built in correction for this error For frequencies above 50 of maximum frequency inaccuracies in this correction can increase the total error by up to 3 The maximum error in measurement will be less than the sum of the above errors and in general will be considerably less 82 Coating Type Wavelength Broadband Metallic PD10 ERROR 190 350nm 1 4 2 400 800nm 1 4 0 1064nm 0 0 2 1 1 5um 1 2 N A 2 3ym 2 8 N A 10 6um 5 15 N A Table 5 Additional Measurement Error due to Wavelength 8 3 3 Recalibration from a Known Source of Laser Energy Setthe instrument to energy 2 Settothe proper energy range 3 From the bargraph power screen press the menu button twice and press more Now press select until cal is highlighted Press go 4 Using the value button select the time period over which you wish to average Now press select and value to choose the correct laser wavelength See Note 1 below Now press go 5 Apply the known laser energy 6 Adjust up down until the energy reading on the screen equals the known energy The energy calibration factor will change accordingly 7 Press exit The new value will be saved Note 1 For metallic and PD10 type heads when the calibration is changed at one laser wavelength the overall calibration of all other wavelengths changes proportionately For BB type the calibration
72. r power against time or successive energy points as long as the Nova has not been turned off While measurements are being taken the Nova will record data until the screen is full or until the center reset or exit button is pressed With the optional Ophir RS232 adapter this data can be read out to PC in real time or in later after storage Contact your Ophir agent for details To use the Nova to graph power vs time 1 From the bargraph power measurement screen press the menu button until range is shown Press range Select the appropriate manual range and exit The Nova must be in manual range to use the powerlog function 2 Press the menu button until more is shown Now press more and select until power log is highlighted Press go 3 Setthe maximum power on the scale by pressing value until the desired value is highlighted 33 Now press select until the minimum setting is highlighted Press select until minimum value on graph is chosen Press select again and then use value to choose the desired horizontal time scale Press go The measurement will begin immediately When the screen is full the measurement will stop and the graph will be held until the center reset button is pressed again The center button is the reset button Press this button to reset the screen and begin a new measurement Press the menu button once to make the legends visible and press the menu button again to make
73. rbers BB broadband absorber On standard high power density broadband Ophir power monitor heads a special refractory coating is used to provide high absorptivity from the UV through the IR This coating can withstand very high power densities up to 20 Kiem without changing calibration The absorption of this coating is above 90 for most of its range as shown in Graph 1 below EX excimer absorber The EX absorber provides high absorption in the UV and it can withstand both the pulse energies and the average power of excimer lasers These discs also have excellent absorption for 10 6 um and other wave lengths They can therefore be used for other types of lasers as well The absorption of the various Ophir absorbers as a function of wavelength is shown in Graph 1 below LP long pulse absorber This absorber has a particularly high damage threshold for long pulse ms or continuous lasers and is therefore offered for use with high power heads It is calibrated for use with YAG laser at 1 064um or CO laser at 10 6um and absorbs about 90 at these wavelengths 69 2 Volume absorbers P pulse type absorber The models with the P suffix for use with pulsed lasers have a special absorbing glass with an absorbance of 95 2 over the operating range Since the surface is a glass its reflectivity does not change even if damaged or melted locally HE HE1 high energy absorber The HE and HE1 types have a particu
74. rders and other analog devices The voltage is proportional to the reading on the display and scaled such that full scale equals 1 00 volts The output can drive up to 2mA into an external device 3 10 Beam Splitter Function If you have a measurement setup with a beam splitter or attenuator and wish to display the laser power or energy before the splitter or attenuator and not the actual value impinging on the measurement head the beam splitter function allows you to do this For example if you are splitting off 4 5 of a laser beam into the meter and you wish to display the full beam do as follows 1 Press the menu button twice and press more then select atten Press go 2 Press select so the screen says attenuation screen on Press exit 3 Press set 90 Now alternately press select and change to set the value of the percent split off to 4 5 Press exit and the value will be saved 4 When you turn on the Nova now with a measuring head the display will come up in the attenuation screen and the value displayed will be 1 attenuation fraction instead of the power or energy actually on the screen For instance in the present example if the laser power on the head is 5 5 Watts 5 5 0 045 122 22 Watts will be displayed If the menu button is pressed again you will be in the normal measurement screen where the actual value measured is displayed 24 5 If you want to have the Nova co
75. ress the menu button twice and press more Press select until calibrate is highlighted Press go Now press power with laser Use the aser button to select one of the laser wavelengths Press go and adjust until the cal factor is the same as that sent with the disc for the wavelength shown Now press exit to save that value Now repeat step 3 for the other laser wavelengths Each time use the laser button to select a different laser wavelength and enter the proper calibration factor sent with the disc for that wavelength 8 6 2 Replacing Battery 1 Remove the bottom of the Nova by unscrewing the 4 Phillips screws Unscrew screw in the center of the PC boards Carefully lift the boards up together Unplug the battery connector from the circuit board Remove the battery and replace it with a new one obtained from your Ophir representative Reassemble the boards and screw in the center screw Now reassemble into the case and close 4 case screws 88 Chapter 9 Nova Specifications 9 1 System Display Specifications pyroelectric OEM 16 ranges Ato D resolution 17 bits plus sign 0 0007 eee Weegee ad Electrical accuracy 0 25 20pA new eee LE ee Electrical input noise 500nV or 1 5pA 0 0015 Dynamicrange range 9 Qdecades 1 10 Qdecades 1 10 107 Analog output 0 1 Volt with 11 bit 0 0596 D e Analog output accuracy 0 2 mV relative to 203H x 95W x 37D mm Mass
76. s the menu button until more appears Press more Press select until exposure is highlighted Press go Setup the desired measurement parameters in the setup screen and press go Press go again The Nova will start summing laser energy exposure and the legend will change to stop When you wish to stop measuring press stop or wait for the timeout selected in the setup screen The Nova will now display the total energy emitted during that period the time elapsed and the number of pulses during the period If you wish to reset reading to zero before another reading press reset To return to the bargraph screen press exit 12 34 uJ 00 00 27 3 227 P GO RESET EXIT Toggle GO STOP Figure 10 Exposure Screen 6 4 4 Pyro Scope Adapter A special adapter can be purchased which is plugged in between the head connector and the Nova head socket This adapter has a cable and BNC connector to attach to an oscilloscope It enables the user to see on the scope pulses proportional to energy up to the maximum pulse rating of the head Unlike the Nova display which samples at the maximum rate the scope adapter shows every single pulse 64 Activate the scope adapter as follows 1 Plug the adapter into the Nova display and then plug the head plug into the adapter Plug the BNC connector into a standard 1MQ scope input Press the menu button twice and then press more Select scope and press go The
77. set the PD300 to your laser wavelengths 43 1 From the power measurement mode with the bargraph display press the menu button located on the right side until more appears Press more 2 Press select until wavelength is highlighted Press go 3 Press change then up and down to select the first wavelength desired Press done Repeat steps 2 and 3 for other wavelengths desired When finished press exit and the new values will be saved 5 3 Startup Defaults If you are planning to do similar measurements most of the time you can set up the Nova so that it automatically goes into the measurement mode you want upon startup Since these desired settings are saved in the head smart connector you can save different settings for different heads The following section describes how to set up the Nova with photodiode heads so that they will start up in the desired configuration Setting up the startup defaults 1 From the power measurement bargraph screen press the menu button located on the right side until more appears Press more 2 Press select until configure is highlighted Press go 3 Press value to select filter in or out as the startup default For details of filter in out See Section 5 43 4 Press select then value to choose the manual power range you wish to be the default autorange or dBm The Nova will now be in the mode selected when switched on with this head connected
78. splitter The Ophir beam splitter is built to facilitate this measurement Ophir has models RP which are specifically designed for pulses with energies up to 100 Joule cm 65 Contact your Ophir dealer for details 6 6 Energy Log This is the same as for thermal heads except you do not have to wait for reset See section 4 5 6 66 Chapter 7 Circuit Description The NOVA has three circuit boards an analog processor a digital processor and a power supply Analog Processor The signal from the detector head enters the analog processor board and passes through EMI protection components to a differential transimpedance preamplifier From there it is further amplified by a programmable gain voltage amplifier and passes to a dual slope integrating analog to digital converter All calibration data for the analog processor is stored ona memory chip on the analog processor board There are no adjustable components trimmers etc in the NOVA except for the display contrast adjustment on the rear panel Digital Processor The digital processor is built around a Motorola MC68332 32 bit central processing unit CPU with a 16 bit bus On the digital processor board there is a socket containing the upgradeable program ROM and a 256Kbit static RAM The processor receives signals from the analog processor and converts them into current in amps When used with thermopile heads the signal is then processed by a sophisticated digit
79. ss and accuracy coupled with ease of operation It can operate with thermal pyroelectric and photodiode heads It has smart connector technology Simply plugging in the head configures and calibrates the Nova to operate with that head The Nova can graph power or energy vs time It displays power measurements in both digital and analog form at the same time It will also autorange so you do not have to set scales or it can manual range if you wish It will remember what mode you were using before you turned it off and will return to that mode when turned on You can zoom in on the present reading or subtract background Although the calibration information is stored in the smart connector you can recalibrate from the screen and store the new calibration in the head You can also zero the Nova atthe touch of a button The main instructions are clearly shown on the screen so you should not have to refer to this manual very often Above all the Nova has advanced circuitry and digital signal processing for excellent sensitivity signal to noise ratio accuracy and response time It also has special circuitry to reject electromagnetic interference 3 2 Smart Connectors and Multihead Operation The Nova display although very compact is versatile and can operate with either thermal pyroelectric or photodiode type laser measuring heads The head configuration and calibration information is stored in an EEROM in the head connector plug
80. t choose the percentage of the beam that you are measuring in our example above you will set this to 005 000 Now if you measure a value the large display will show the value before attenuation and the small display will show the actual value measured In our example the large display will show 100W and the small display 5W Note 1 If the attenuation screen is activated when the Nova is turned on it will come up in the attenuation screen You may switch between the attenuation screen and the regular power or energy screen by pressing the menu button several times Note 2 The attenuation screen can operate with both power and energy measurement measured value attenuation factor power before attenuation Figure 7 Attenuation Screen 36 4 5 Energy Measurement 4 5 1 Setup Switch the Nova on If it is in power mode press the menu button once to make the legends visible and then press the left soft key energy to put it into energy mode Alternately if you remember the position you can press the left soft key without pressing the mode button first See figure 8 The energy mode is manual ranging press range and then press the up and down soft keys to get to the proper range The correct range is the lowest one which is larger than the pulse energy measured If you plan to use the same setting often use the configure function as described in Section 4 3 before switching the Nova off the first time
81. t NIST standards the accuracy is generally 2 at the wavelength the calibration has been performed The maximum error in measurement will be less than the sum of the calibration accuracy and linearity The linearity of the photodiode detector is extremely high and errors due to this factor can be ignored The maximum error due to all factors is the stated accuracy of 3 The calibration is carried out with a fixed laser In addition to the error in calibration the reading drops somewhat with scanning speed The BC20 is specified to deviate from the static reading no more than 3 up to a scanning velocity of 30 000 inch sec 8 2 4 Recalibration of PD300 and 3A IS Heads from a Known Source of Laser Power 1 Set the instrument to the appropriate manual power range 2 From the bargraph power screen press the menu button twice and press more Now press select until calibrate is highlighted Press go 3 Using the laser button select the proper laser wavelength See note 1 below When laser wavelength is set press go 4 Applythe known laser energy 5 Adjust up down until the power on the screen equals the known power The power calibration factor will change accordingly 6 Pressexit The new value will be saved Note 1 The relative readings at different wavelengths are fixed by the wavelength calibration curve stored in the head EEROM When changing the calibration at one wavelength the calibration at all other wa
82. t the absorber head aperture a temperature gradient is created across the thermopile of the enclosed detector disc This generates a voltage proportional to the incident power 26 The display unit amplifies this signal and indicates the power level received by the head At the same time signal processing software causes the display unit to respond faster than the thermal rise time of the detector disc thus reducing the response time of the Nova Energy of a single pulse is measured on the Nova by digitally integrating the pulse power over time 4 2 Selecting Settings from the Nova Screen The Nova can be set to various chosen settings while operating In addition it can be set so that it will be in the desired configuration when turned on the next time 4 2 1 To Set Type of Laser being Used Thermopile heads have somewhat different absorption at different wavelengths In order to compensate for this each head has been calibrated by laser at several wavelengths When you choose the correct laser wavelength the correction factor for that wavelength is automatically introduced Note The laser wavelength correction in use is displayed in the upper right corner of the display In order to choose the laser type do the following 1 From the bargraph power measurement screen press the menu button twice and press laser until the correct laser type or wavelength is displayed 2 Return to the bargraph screen by pressing t
83. t up each time See the appropriate section in the chapter dealing with the head type you intend to use 20 3 5 Nova Functions which are independent of Head Type When no head is connected to the Nova and it is turned on the message Head Disconnected appears In this mode the user can adjust either the power line frequency so the instrument can be adjusted for minimum interference at the power line frequency or the user can zero the instrument to eliminate any offsets 3 5 1 Line Frequency The Nova is factory set for 50 Hz If your country uses 60 Hz you must reset the line frequency To reset the line frequency 1 Turn on the Nova while the head is disconnected 2 Press select until line frequency is highlighted 3 Press go then change to select the correct line frequency 4 Press exit to save the new value and exit 3 5 2 Zero Adjustments In the Nova all adjustments including zeroing internal circuits are done from the software This ensures simple and accurate realignment It is recommended to rezero the Nova every 2 months for best performance The simple zeroing procedure follows 1 If a head is connected disconnect the head turn off the instrument then back on again so the Nova can identify that no head is connected Head Disconnected will be displayed 2 Turn on the Nova and let it run for at least 30 seconds before performing zero adjustment 3 Press go 4 Make sure the instru
84. the display can capture the pulses at rates not exceeding 15Hz The Nova samples individual pulses at a rate of 15Hz from the laser pulse train Thus if the user has a laser pulsing at over 15Hz and wishes to average over a number of pulses the average will be over the number of pulses processed by the Nova and not over those emitted by the laser Example If the user s laser is pulsing at 300Hz and the user selects to average over 1s the Nova will collect 15 pulses of the 300 emitted during that period and will average over those 15 pulses The Nova will then update the display every 1s To set up for averaging over a number of pulses 1 From the bargraph measurement screen press the menu button twice and press setup 2 Now press select until average over is highlighted Now press value until the time period you wish to average over is displayed If you do not want to average but collect individual pulses then select none 3 Press exit If you wish to save the new setting follow the instructions in section 6 3 Note The average function only affects the numerical readings on the screen The bargraph always shows individual pulses 6 2 2 3 PD10 Photodiode Energy Head The PD10 operates in a similar fashion to the pyroelectric PE heads except it has a photodiode detector instead of pyroelectric Because of its great sensitivity it can operate down to about 1nJ of energy It has complete wavelength correctio
85. the legends disappear Warning Pressing reset will erase all previously stored data and start recording new data The left button is the setup button Press this button to go to the setup screen and readjust measurement scales The legends are visible when the buttons are pressed or when pressing the menu button co2 1 234W SETUP RESET EXIT T T Push soft keys to make legends visible Figure 6 Laser Power vs Time Graph 4 4 2 5 Average When a laser output is fluctuating or unstable it is useful to measure the average power over a certain period The Nova gives you this exclusive feature allowing averaging over periods varying from 1 3 s to 1 hour 34 The Nova allows two modes of averaging continuous or periodic mode The periodic mode is meant for lasers which have rapid power fluctuations In this case choose an averaging period from 33 to 30s The Nova averages over each period and displays the result at the end of each period until the next average arrives The continuous mode is meant for lasers having slow variations in output over a relatively longer period In the continuous mode the Nova displays a running average which is continuously updated as time progresses At the end of the period the measurement stops and the last value is displayed until the instrument is reset In continuous mode the measurement period is from 10s to 1hour To use the Nova to display average 1 The average m
86. the photodiode can measure up to 30mW without saturating the detector In addition the PD300 comes with an additional removable filter for measuring up to 300mW Other models of the PD300 series also have built in and removable filters The exact maximum power is reached when the reading reaches full scale or the output current from the head reaches 1 3mA whichever comes first See Table 2 for the exact maximum as a function of wavelength Depending on what powers you wish to measure you should choose to work with the removable filter installed or not For this purpose the Nova has a filter setting and uses the proper correction curve depending on whether the filter is installed or not Warning If the PD300 is used in the filter in setting and the filter is not installed or vice versa the readings will be completely incorrect If the power of your laser exceeds the maximum for filter in you can purchase a thermal head for that wavelength Consult your Ophir agent for details 47 To choose the filter setting 1 From the bargraph measurement screen press the menu button twice 2 Press filter 3 If you wish to work with filter installed press change until display says filter is in Be sure to install removable filter on detector head 4 f you wish to work with the filter out press change until display says filter is out Be sure to remove the filter from the detector head 5 When finished press exit to r
87. velengths will change proportionately 79 8 2 5 Recalibration of BC20 heads from a Known 6 Source of Laser Power Set to the proper manual power or energy range From the bargraph power screen press the menu button twice and press more Now press select till calibrate is highlighted Press go Press power with laser Using the laser button select the proper laser wavelength Press go Apply the known laser power energy Adjust up down until the power energy reading on the screen equals the known power energy The power energy calibration factor will change accordingly Press exit and save 8 3 Calibration of Pyroelectric and Photodiode Energy Heads Several types of absorber are used in Ophir pyroelectric and photodiode energy heads 1 Metallic type The type with no suffix in the name have a partially reflective multilayer metallic coating which absorbs approximately 50 and whose absorption graph is shown in graph 2 below The metallic coating permits very high repetition rates up to 5000Hz as well as high sensitivity Broadband type The type with the BB suffix has a broadband black absorbing coating to provide high absorptivity from the UV through the IR This coating can withstand energy densities up to 0 3J cm for short pulses and 2J m for long pulses without changing calibration The absorption of this coating is above 9096 for most of its range as shown in Graph 2 below This coating is av

Download Pdf Manuals

image

Related Search

Related Contents

ASUS Eee PC 1011PX  X3001 Installation Manual  Truques da Escrita – Howard S. Becker  エレベーター区画用 ユニファイヤーガード スーパーミニ  PIANO OPERATIVO DI SICUREZZA  Samsung DVD-P191 Uživatelská přiručka  

Copyright © All rights reserved.
Failed to retrieve file