Home
Model 146i Multi-gas Calibrator - User Manual
Contents
1. Figure 7 14 Removing the Glass Chamber Assembly Thermo Fisher Scientific Model 1467 Instruction Manual 1 47 Servicing Reaction Chamber Mixing Chamber Replacement Reaction Chamber Mixing Chamber Replacement A 7 48 Model 146 Instruction Manual Permeation Oven Top Glass Chamber Assembly N ba See Procedure Permeation Oven Bottom _See Expanded View A 8 Expanded View o Figure 7 15 Replacing the Glass Chamber Assembly Use the following procedure to replace the reaction chamber or mixing chamber Equipment Required Reaction chamber or mixing chamber Allen wrench 7 64 inch Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the plumbing 3 Remove two 2 Allen screws from the top of the holder assembly 4 Remove two 2 Allen screws from the side of the holder assembly Thermo Fisher Scientific Servicing Pressure Regulator Replacement 5 Remove the chamber and replace with new chamber Pressure Regulator Use the following procedure to replace the pressure regulator Figure 7 16 Replacement Equipment Required Pressure regulator Wrenches 7 16 and 9 16 inch amounts of static electricity A prope
2. Motherboard Digital Output Board 1 0 Expansion Board Opt Figure 7 8 Replacing the I O Expansion Board Optional 1 0 Expansion Board Digital Output Board Figure 7 9 Rear Panel Board Connectors Thermo Fisher Scientific Model 146 Instruction Manual 7 25 Servicing Digital Output Board Replacement Digital Output Board Replacement A Motherboard Replacement 7 26 Model 146 Instruction Manual Use the following procedure to replace the digital output board Figure 7 8 Equipment Required Digital output board Nut driver 3 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the I O expansion board optional if used See the I O Expansion Board Replacement procedure in this chapter 3 Disconnect the digital output board ribbon cable from the motherboard 4 Using the nut driver remove the two standoffs securing the board to the rear panel Figure 7 9 5 Pop off the digital output board from the mounting studs and remove the board 6 To install the digital output board follow previous steps in reverse Use the following procedure to replace the motherboard Figure 7 8 Equipment Required Motherboard Philips screwdriver Thermo Fis
3. EXPANSION BD J3 mo N oo A WO N 5V 24V 24V Ground Ground Ground RS485 to Expansion Board RS485 to Expansion Board SPARE DATA J4 mo N oo A WO N 5V 24V 24V Ground Ground Ground RS485 to Spare Board RS485 to Spare Board 1 0 Thermo Fisher Scientific J5 oOo ao N oon A wo N gt o Power Fail Relay N C Contact Ground TTL Input 1 TTL Input 2 Ground TTL Input 5 TTL Input 7 TTL Input 8 TTL Input 10 Ground TTL Input 13 Model 146 Instruction Manual 6 11 Troubleshooting Connector Pin Descriptions Table 6 3 Motherboard Connector Pin Descriptions continued Connector Reference Pin Signal Label Designator Description TTL Input 15 Ground Analog Voltage Output 1 Analog Voltage Output 3 Ground Analog Voltage Output 5 Ground gt gt gt gt gt gt gt gt co co N o a gt ow N Ground Power Fail Relay COM N o 21 Power Fail Relay N O Contact 22 Ground 23 TTL Input 3 24 TTL Input 4 25 TTL Input 6 26 Ground 27 TTL Input 9 28 TTL Input 11 29 TTL Input 12 30 TTL Input 14 31 TTL Input 16 32 Ground 33 Analog Voltage Output 2 34 Analog Voltage Output 4 35 Ground 36 Analog Voltage Output 6 37 Ground SER EN J7 1 Serial Enable Jumper 2 3 3V 24V IN J10 1 24V 2 Ground DIGITAL 1 0 J14 1 5V 2 24V 3 24V 6 12 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scien
4. 5 Attach the external vacuum pump vacuum gauge assembly with the vacuum gauge side to the calibrator EXHAUST port 6 Apply vacuum until the gauge stabilizes then trap the vacuum with the shut off valve Model 146 Instruction Manual 7 13 Servicing Mass Flow Controller Replacement 7 Verify that the pressure drops below 180 mm Hg by reading the vacuum gauge attached to the pump assembly 8 If the instrument does not meet the previously described conditions begin troubleshooting beginning at the top of this procedure Mass Flow Controller Use the following procedure to replace a mass flow controller Figure 7 4 Replacement Equipment required Wrenches 7 16 and 9 16 Philips screwdrivers 1 and 2 Flatblade screwdriver New mass flow controller amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Loosen all four fittings to mass flow controllers 7 14 Model 146 Instruction Manual Thermo Fisher Scientific Servicing Mass Flow Controller Replacement 3 Disconnect electrical connections from the top of the flow controllers noting their positions for reconnection Figure 7 4 Figure 7 4 Replacing a Mass Flow Controller 4 Remove plumbing from flow controllers 5 Disconnect fan electrical
5. on page 7 34 Photometer Board Replacement Photometer option on page 7 35 Detector Replacement Photometer option on page 7 35 Photometer Thermistor Replacement on page 7 36 Photometer Temperature Calibration Photometer option on page 7 37 Solenoid Replacement on page 7 38 Pressure Transducer Replacement Photometer option on page 7 39 Pressure Transducer Calibration Photometer option on page 7 40 Ozonator Lamp Replacement Ozonator option on page 7 42 Ozonator Lamp Heater Replacement Ozonator option on page 7 43 1 2 Model 146 Instruction Manual Thermo Fisher Scientific Safety Precautions Thermo Fisher Scientific A Servicing Safety Precautions Ozonator Replacement Ozonator option on page 7 44 Ozonator Board Replacement Ozonator option on page 7 45 Permeation Oven Replacement on page 7 46 Reaction Chamber Mixing Chamber Replacement on page 7 48 Pressure Regulator Replacement on page 7 49 Back Pressure Regulator Replacement on page 7 50 Pressure Regulator s Adjustment on page 7 51 Ambient Temperature Calibration on page 7 53 Service Locations on page 7 54 Read the safety precautions before beginning any procedures in this chapter WARNING The service procedures in this manual are restricted to qualified service representatives If the equipment is operated in a manner not specified by the ma
6. Eu Sur ZERO AR EAST Figure 2 5 Model 1467 Rear Panel Remote The rear panel I O DB37M connector enables flow modes ozone levels Interconnections and gas settings to be remotely controlled using contact closure In addition the connector mimics any input settings with corresponding relay outputs Inputs The Model 1467 is controllable through the 37 pin male rear panel connector Refer to the following tables when selecting the appropriate gas gas span level and ozonator or permeation oven level for the Model 1467 Note These are the default settings that can be changed by the operator A 1 indicates contact closure to ground and 0 indicates no contact assuming the digital input channel is set to Normally Open a 2 6 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Installation Remote Interconnections Pins 3 4 and 23 select a gas Pins 2 5 10 13 16 18 19 22 26 32 35 and 37 are ground Note Gases D E and F are available only when the instrument is equipped with the 6 gas option Gas Bit 3 Gas Bit 2 Gas Bit 1 Gas Type Pin 23 Pin 4 Pin 3 0 0 0 Gas Off 0 0 1 Gas A 0 1 0 Gas B 0 1 Gas C 1 0 0 Gas D 1 0 Gas E 1 1 0 Gas F Pins 24 6 and 25 select a gas span level gas Span Bit 3 Span Bit 2 Span Bit 1 Span Level Pin 25 Pin6 Pin 24 0 0 0 Span 0 0 0 1 Span 1 0 1 0 Span 2 0 1 Span 3 1 0 0 Span 4 1 0 Span 5 1 1
7. Field Name Hex Function 0x03 Starting Address Hi 0x00 Starting Address Lo 0x09 No of Registers Hi 0x00 No of Registers Lo 0x04 Response Field Name Hex Function 0x03 Byte Count 0x06 Register value Hi 10 0x02 Register value Lo 10 0x2B Register value Hi 11 0x00 Register value Lo 11 0x00 Register value Hi 12 0x00 Register value Lo 12 0x64 Register value Hi 13 0x00 Register value Lo 13 0x64 Thermo Fisher Scientific 0x05 Force Write Thermo Fisher Scientific Single Coil MODBUS Protocol Function Codes The contents of register 10 are shown as the two byte values of 0x02 0x2B Then contents of registers 11 13 are 0x00 0x00 0x00 0x64 and 0x00 0x64 respectively The force write single coil function simulates the activation of the digital inputs in the instrument which triggers the respective action This function code is used to set a single action to either ON or OFF The request specifies the address of the action to be forced Actions are addressed starting at zero Therefore action number 1 is addressed as 0 The requested ON OFF state is specified by a constant in the request data field A value of OxFFOO requests the action to be ON A value of 0x0000 requests it to be OFF All other values are illegal and will not affect the output The normal response is an echo of the request returned after the state has been written Request Function code 1 Byte 0x05 Starting Address 2 Bytes 0x0000
8. Read Coils Inputs read the status of the digital outputs relays in the instrument Issuing either of these function codes will generate the same response These requests specify the starting address i e the address of the first output specified and the number of outputs The outputs are addressed starting at zero Therefore outputs numbered 1 16 are addressed as 0 15 The outputs in the response message are packed as one per bit of the data field Status is indicated as 1 Active on and 0 Inactive off The LSB of the first data byte contains the output addressed in the query The other outputs follow toward the high end of this byte and from low order to Model 146 Instruction Manual C 3 MODBUS Protocol Function Codes C 4 Model 146 Instruction Manual high order in subsequent bytes If the returned output quantity is not a multiple of eight the remaining bits in the final data byte will be padded with zeros toward the high order end of the byte The Byte Count field specifies the quantity of complete bytes of data Note The values reported may not reflect the state of the actual relays in the instrument as the user may program these outputs for either active closed or open A Request Function code 1 Byte 0x01 or 0x02 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Quantity of outputs 2 Bytes 1 to maximum allowed by instrument Unit Identifier 1 Byte 0x00 to OxFF Passed back in response Res
9. Cable List Model 146 Instruction Manual Table 7 1 Model 1467 Replacement Parts continued Part Number 102589 00 102030 00 49P766 49P501 8095 8096 8097 8092 8093 8094 4120 103199 00 103205 00 103334 00 103200 00 103207 00 8130 103125 00 103206 00 4291 Description Reaction Chamber Permeation Oven Mixing Chamber Manifold Mass Flow Controller 50 sccm Mass Flow Controller 100 sccm Mass Flow Controller 200 sccm Mass Flow Controller 5 slpm Mass Flow Controller 10 slpm Mass Flow Controller 20 slpm Capillary photometer Capillary ozonator Capillary pressure gauge Capillary pump Solenoid sample reference Solenoid ozonator Solenoid perm 1 Solenoid perm 2 Solenoid reaction chamber Solenoid zero air Solenoid external internal Solenoid span gas Charcoal scrubber assembly external Table 7 2 describes the Model 146i cables See the Troubleshooting chapter for associated connection diagrams and board connector pin descriptions Table 7 2 Model 1467 Cables Part Number 101036 00 101037 00 101048 00 101038 00 Description DC Power Supply 24V Output 115VAC Supply to Measurement Interface Board RS 485 Data AC to Motherboard Thermo Fisher Scientific Servicing External Device Connection Components Table 7 2 Model 1467 Cables continued Part Number 101364 00 1054 00 1035 00 033 00 377 00 1055 00 1706 00 102446 00 101267 00 49P747 gt
10. Check to see that there is a bright blue light when removing the ozone lamp from ozonator with ozonator level set to 100 If not replace lamp see Servicing chapter Unstable ozone output Failure of analyzer measuring ozone Leak in system Zero air flow unstable Lamp failure Defective ozonator board Repair analyzer Check for leaks and repair see Servicing chapter Check zero air flow meter Replace lamp see Servicing chapter Replace ozonator board see Servicing chapter Model 146 Instruction Manual 6 3 Troubleshooting Troubleshooting Guides 6 4 Model 146 Instruction Manual Table 6 1 Troubleshooting General Guide continued Malfunction Permeation oven fails to warm up Possible Cause Oven not ON Not enough time has elapsed since turning oven ON Loose cable connections Oven heater defective Control circuit malfunction Action Turn oven ON Wait one hour from powering up instrument Check cable connections Replace oven Replace measurement interface board see Servicing chapter Temperature display erratic Flow through permeation tube oven not constant Clean capillary readjust regulator see Preventive Maintenance chapter Cell frequency high Light setting too high Defective detector Defective photometer board Defective measurement interface board Lower lamp setting Service Mode gt
11. Lamp Setting Replace detector module Replace photometer board Replace measurement interface board Cell frequency low or zero Light setting too low Contaminated cell Defective detector Defective photometer board Defective measurement interface board Raise lamp setting Service Mode gt Lamp Setting Clean cell Replace detector module Replace photometer board Replace measurement interface board Cell noise excessive Contaminated cell Defective detector Lamp failure Clean cell Replace detector module Replace lamp Pressure transducer does not hold calibration Plumbing leak or bad electrical connection Defective pressure transducer Defective measurement interface board Check plumbing and electrical connections Replace pressure transducer Replace measurement interface board Output signal noisy Defective recorder Sample is varying Replace or repair recorder Run instrument on stable ozone source if quiet no malfunction Thermo Fisher Scientific Table 6 1 Troubleshooting General Guide continued Malfunction Possible Cause Contaminated cell Sticky solenoid valve Troubleshooting Troubleshooting Guides Action Clean cell Replace with known good solenoid valve Photometer does not calibrate properly Leak Bench thermistor defective Pressure transducer out of calibration Dirty system Defective solenoid La
12. POWER_GOOD REAR PANEL AC AC INPUT 3 Cond Pl 2 Pin FRONT PANEL gt POWER SW pose e i P PIG i l TRANSFCRVER wo 3 Pin AAA ke Q0VAC or 240VAC 1 CPrIONS Figure 6 1 Board Level Connection Diagram Common Electronics 6 8 Model 146 Instruction Manual Thermo Fisher Scientific Troubleshooting Board Level Connection Diagrams DETECTOR 6 Pin MASS FLOW 6 Pin CONTROLLER 1 MASS FLOW 6 Pin CONTROLLER 2 TO COMMON ELECTRONICS OZONATOR _ LAMP POWER SUPPLY 2 Pin 8 Pin LAMP 6 Pin HEATER PHOTOMETER 1_ Lamp POWER SUPPLY _ 2Pin 8 Pin LAMP 6 Pin HEATER _ GAS THERM PERM OVEN HEATER 24V 1W OPTIONAL 146i MEASUREMENT re t OVEN THERM INTERFACE BOARD l i PRESSURE 4 Pin AMBIENT TEMP THERM 2 Pin 2 Pin BENCH TEMP THERM 2 Pin SOLENOID 10 Pin CONNECTOR a _ PERMI SOL BOARD 2 Pin GAS A SOL TF PERM2 SOL IPn _ GASB SOL ia _ oz soL T GAS C SOL a C RCSOL aa _ GAS D SOL lt SAMP REF SOL ia _ GAS E SOL IPn _ SPARE SOL T GAS F SOL 2 Pin PAN 2 Pin E Z A SOL aa FAN PR AC EXT INT SOL ae PUMP AC 2 Pin _ Figure 6 2 Board Level Connection Diagram Measurement System Thermo Fisher Scientific Model 146 Instruction Manual 6 9 Troubleshooting Connector Pin Descriptions Connector Pin The connector pin descriptions in Table 6 3 through Table 6 10 can be Descripti
13. Receive set level 1 ok man gas gas gflow value man gas gas zflow value gas a b c d e value float These commands set the manual level for zero and gas flows for gas gas in sccm The set command is optional Send set man gas a gflow 202 00 Receive set man gas a gflow 202 00 man ozon level value This command sets the ozone drive level for the manual level to value in or ppb The set command is optional This command is only available if the ozonator option is installed Send man ozon level 20 Receive man ozon level 20 ok man ozon zflow value This command sets the ozone zero air flow for the ozone manual level to value where value is a floating point in sccm The set command is optional This command is only available if the ozonator option is installed Model 146 Instruction Manual B 33 C Link Protocol Commands Measurement Configuration B 34 Model 1467 Instruction Manual Send set man ozon zflow 1000 Receive set man ozon zflow 1000 ok sel ozon evel This command selects the ozonator output levels 1 2 3 4 or 5 Send sel ozon 1 Receive sel ozon 1 ok sel ozon level set ozon level This command selects the ozonator output levels 0 1 2 3 4 5 or man 0 shuts off the ozonator and man puts it into manual mode The example below sets the ozonator output level 1 This command is only available if the ozonator option is installed Send sel ozon 1 Receive sel ozon 1 ok sel ozon perm o
14. Streaming Data Config porin pn pus fel E MT A EA CES Pel Pi EE EE Streaming Data Interval The Streaming Data Interval screen is used to adjust the streaming data interval The following interval times are available 1 2 5 10 20 30 60 90 120 180 240 and 300 seconds e Inthe Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Streaming Data Interval 3 50 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Concentrations The Concentrations screen allows the user to select the output signal that is tied to the selected streaming data item The selected item is shown by lt after it e Inthe Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Select Item gt Concentrations Other Measurements The Other Measurements screen allows the user to select the output signal that is tied to the selected streaming data item The selected item is shown by lt after it e Inthe Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Select Item gt Other Measurements Thermo Fisher Scientific Model 1467 Instruction Manual 3 51 Operation Instrument Controls Menu Analog Inputs RS 232 RS 485 Selection The Analog Inputs screen allows the user to select the analog input signal none or analog inputs 1
15. This header consists of the following components Transaction Identifier 2 Bytes 0x0000 to OxFFFF Passed back in response Protocol Identifier 2 Bytes 0x00 MODBUS protocol Length 2 Bytes 0x0000 to OxFFFF Number of following bytes Unit Identifier 1 Byte 0x00 to OxFF Passed back in response A Slave address is not required in MODBUS over TCP IP because the higher level protocols include device addressing The unit identifier is not used by the instrument Thermo Fisher Scientific Function Code Data Error Check Function Codes 0x01 0x02 Read Coils Thermo Fisher Scientific Read Inputs MODBUS Protocol Function Codes The function code is a single byte in length The following function codes are supported by the instrument Read Coils 0x01 Read Inputs 0x02 Read Holding Registers 0x03 Read Input Registers 0x04 Force Write Single Coil 0x05 0x06 Read Exception Status If a function code is received that is not in this list an invalid function exception is returned The data field varies depending on the function For more description of these data fields see Function Codes below In MODBUS over Serial an error check is included in the message This is not necessary in MODBUS over TCP IP because the higher level protocols ensure error free transmission The error check is a two byte 16 bit CRC value This section describes the various function codes that are supported by the Model 146i
16. e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Select SREC or LREC gt View Logged Data asec prepne pu Eus Parmy CHP cio A A EE EE Number of Records The Number of Records screen is used to select the starting point to display the number of records back to view risas BEES Pi Taal DH Cil oe hen E BB Fa Bast Banos Haat ake POE Scale AS EE EE The Record Display screen read only displays the selected records DH Guided SOREL PERG Date and Time The Date and Time screen is used to set a start date and time for which to view logged data For example if 20 JUN 2005 10 00 is entered then the first logged data record that is displayed is the first record after this time If set to one minute logging this would be at 20 JUN 2005 10 01 3 42 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu The Record Display screen read only displays the selected records opi Erase Log The Erase Log is used to erase all saved data for the selected record type not all short records and long records e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Erase Log BEE ES BB Fase Boat Banos Select Content The Select Content submenu displays a list of 32 record fields to use and a submenu list of the analog output signal group choices to choose from Choices are Concentrations Othe
17. 15V o oj A Ww N 15V Ground MFC2 o NM A WO N Mass Flow Controller 2 Input Mass Flow Controller 2 Output Ground 15V 15V Ground CONNECTOR BD Oo cao N oon A wo N gt Gas A Solenoid Control Gas C Solenoid Control 24V Gas E Solenoid Control Z A Solenoid Control Gas B Solenoid Control Gas D Solenoid Control 24V Gas F Solenoid Control EXT INT Solenoid Control PERM 1 6 22 Model 146 Instruction Manual 24V Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Table 6 7 Measurement Interface Board Connector Pin Descriptions continued Connector Reference Pin Signal Label Designator Description 2 Perm 1 Solenoid Control PERM 2 J19 24V 2 Perm 2 Solenoid Control OZ J20 24V 2 Ozonator Solenoid Control RC J21 24V 2 Reaction Chamber Solenoid Control SAMP REF J22 24V 2 Sample Reference Solenoid Control SPARE J23 24V 2 Spare Solenoid Control Table 6 8 Solenoid Connector Board Pin Descriptions Connector Reference Pin Signal Label Designator Description BRN Gas A Solenoid Control ORG Gas C Solenoid Control RED 24V GRN Gas E Solenoid Control VIO Z A Solenoid Control WHT Gas B Solenoid Control YEL Gas D Solenoid Control RED 24V BLU Gas F Solenoid Control GRA EXT INT Solenoid Control GAS A J1 24V 2 Gas A Solenoid Control GAS B J2 24V 2 Gas B Soleno
18. 16 list Irec Lists current selection Irec logging data B 17 list sp Lists current selection in the scratchpad list B 17 list srec Lists current selection srec logging data B 17 list stream Lists current selection streaming data output B 17 list var aout Reports list of analog output index numbers and variables B 46 list var din Reports list of digital input index numbers and variables B 46 list var dout Reports list of digital output index numbers and variables B 46 Ir Outputs long records in the format specified in the command B 17 Irec Outputs long records B 18 Irec format Reports sets output format for long records ASCII or binary B 19 Irec layout Reports layout of Irec data B 19 Irec mem size Reports maximum number of long records that can be stored B 20 Irec per Reports sets long record logging period B 20 malloc Irec Reports sets memory allocation for long records B 21 malloc srec Reports sets memory allocation for short records B 21 man gas gflow Reports sets gas flow controller manual flow rate B 33 man gas zflow Reports sets zero flow controller manual flow rate B 33 man ozon level Reports sets ozonator drive level for manual mode B 33 man ozon zflow Reports sets zero flow controller manual flow rate when B 33 ozonator is active me menu Simulates pressing menu pushbutton B 25 mode Reports sets operating mode in local service or remote B 42 no of Irec Reports sets number of long records stored in memory B 20 no of srec Reports set
19. 54 Ibs with all options Permeation Oven Temperature Control Single point 30 35 45 C Temperature Stability 0 1 C Warm up Time Oven 1 hour permeation device can take 24 48 hours to stabilize Carrier Gas Flow 150 sccm nominal Chamber Size Accepts permeation tubes up to 9 cm in total length 1 cm in diameter Model 146 Instruction Manual 1 7 Introduction Specifications 1 8 Model 1461 Instruction Manual Ozone Generator Ozone Output 6 ppm liters Photometer Preset ranges 0 05 0 1 0 2 0 5 1 0 2 0 5 0 ppm Custom ranges 0 05 to 5 0 ppm Zero noise 0 75 ppb 300 second averaging time Lower detectable limit 1 5 ppb 300 second averaging time Response time in control mode lt 180 seconds Linearity 1 of full scale Flow rate 0 6 LPM Stability and Repeatability Remote Operation Meets or exceeds U S EPA requirements for a Transfer Standard External Solenoid outputs Up to 8 24 VDC external solenoids can be controlled by Model 146i Analog outputs 6 voltage outputs 0 100 mV 1 5 10 V User selectable 5 of full scale over under range 12 bit resolution user selectable for measurement input Optional 6 current outputs 0 20 mA or 4 20 mA Digital outputs 1 power fail relay Form C 10 relays Form A user selectable alarm output relay logic 100 mA 200 VDC Digital inputs
20. 60 ppm The reaction chamber is only in the system during gas phase titration This is because the reaction chamber is only needed during gas phase titration the flush time residence time is approximately less than one minute and three flush times are necessary for a response of 95 percent Gas flow through the oven is low approximately 150 sccm to stabilize the permeation tube oven temperature The major portion of the flow bypasses the oven and is directed to the mixing chamber with the stream from the oven This is achieved by using the same back pressure regulator described above for gas phase titration with a capillary of identical size Thermo Fisher Scientific Photometer Detector System Pressure Transducer Ozonator Ozonator Lamp Power Supply Photometer Lamp Power Supply Photometer Temperature Thermistor Pump Sample Reference Solenoid Valves Thermo Fisher Scientific System Description Flow System During standby operation a continuous flow of gas is necessary through the oven The standby flow and operating flow should be the same to 10 percent This is achieved by setting the pressure regulator to the same pressure setting as the back pressure regulator In practice rather than measuring the pressure it is the actual flows that are measured while adjusting the pressure regulator Two solenoids are used to change the flow of gas going through the oven to the main stream or to vent The photomet
21. Connectors Partition Panel Loosen Captive Screw LOWER PANEL lt gt Screws Figure 7 3 Removing the Measurement Bench and Lowering the Partition Panel gt Remove Hidden Screw Equipment Required Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF and unplug the power cord 2 Disconnect the plumbing connections at the rear of the measurement bench Thermo Fisher Scientific Model 146 Instruction Manual 7 9 Servicing Fuse Replacement Fuse Replacement Model 146 Instruction Manual 10 11 If the instrument is mounted in a rack remove it from the rack Remove the cover Disconnect the connectors that pass through the center of the partition panel Remove two screws from the left side of the case Remove one screw from the bottom rear of the case Remove one screw from the top front of the partition panel While holding the case securely loosen the captive screw at the rear of the measurement bench and pull the measurement bench from the rear of the case Remove the screw at the top rear of the partition panel securing the top of partition panel to the measurement bench and lower the panel being careful not to put excessive tension on the c
22. Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the mass flow controllers Refer to Mass Flow Controller Replacement on page 7 14 3 Disconnect the electrical connection from the fan 4 Remove the fan mounting screws 4 and nuts and remove the fan 5 Install the new fan using the previous steps in reverse 7 18 Model 146 Instruction Manual Thermo Fisher Scientific Analog Output Testing Thermo Fisher Scientific 6 Servicing Analog Output Testing Reinstall the mass flow controllers The analog outputs should be tested if the concentration value on the front panel display disagrees with the analog outputs To check the analog outputs you connect a meter to an analog output channel voltage or current and compare the meter reading with the output value displayed on the Test Analog Outputs screen Equipment Required Multimeter Use the following procedure to test the analog outputs l Connect a meter to the channel to be tested Figure 7 7 shows the analog output pins and Table 7 4 identifies the associated channels From the Main Menu press to scroll to Diagnostics press gt to scroll to Test Analog Outputs and press The Test Analog Outputs screen appears Press 4 to scroll to the channel corresponding to the rear panel terminal pins where the meter is connected and press
23. Digital ground 9 Analog5 21 TTL_Input5 10 Analog ground 22 TTL_Input6 11 Analog6 23 TTL_Input7 12 Analog ground 24 Digital ground Thermo Fisher Scientific Model 146 Instruction Manual 2 11 Installation Connecting External Devices 2 12 D O Terminal Board Figure 2 7 shows the recommended method for attaching the cable to the terminal board using the included tie down and spacer Table 2 2 identifies the connector pins and associated signals See Detail A See Detail B Model 146 Instruction Manual Assembled Connector Detail A Figure 2 7 D O Terminal Board Views Table 2 2 D O Terminal Board Pin Descriptions Detail B Pin Signal Description Pin Signal Description 1 Relay1_ContactA 13 Relay7_ContactA 2 Relay1_ContactB 14 Relay7_ContactB 3 Relay2_ContactA 15 Relay8_ContactA 4 Relay2_ContactB 16 Relay8_ContactB 5 Relay3_ContactA 17 Relay9_ContactA 6 Relay3_ContactB 18 Relay9_ContactB 7 Relay4_ContactA 19 Relay10_ContactA 8 Relay4_ContactB 20 Relay10_ContactB 9 Relay5_ContactA 21 Solenoid_Drive_Output1 10 Relay5_ContactB 22 24V 11 Relay6_ContactA 23 Solenoid_Drive_Output2 12 Relay6_ContactB 24 24V Thermo Fisher Scientific 25 Pin Terminal Board Thermo Fisher Scientific Installation Connecting External Devices The 25 pin terminal board is included with the optional I O Expansion Board Figure 2 8 shows the recommen
24. Pressure Analog Input 7 Bench Temperature Analog Input 8 Lamp Temperature Ds Lamp Temperature Intensity Noise Analog Input The Analog Input Configuration menu displays a list of the 8 analog input Configuration channels available for configuration This screen is only displayed if the I O expansion board option is installed Configuration includes entering descriptor units decimal places choice of 1 10 points in the table and corresponding number of points selected e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config Progr Briain eT cS tr os PRAL E Lidi rii i mpini Thermo Fisher Scientific Model 146 Instruction Manual 3 65 Operation Instrument Controls Menu lt O OEH EE PILPG Descriptor The Descriptor screen allows the user to enter the descriptor for the selected analog input channel The descriptor is used in datalogging and streaming data to report what data is being sent out The descriptor may be from 1 to 3 characters in length and defaults to IN1 to IN8 user input channel number e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Descriptor cine Units The Units screen allows the user to enter the units for the selected analog input channel The units are displayed on the diagnostic screen and in datalogging and streaming data The units may be from 1
25. Scientific MODBUS Protocol MODBUS Commands Supported Coil Number Status 20 PERM OVEN 21 USER 1 22 USER 2 23 USER 3 24 USER 4 25 USER 5 26 USER 6 27 USER 7 28 USER 8 29 GEN ALARM 30 AMBIENT TEMP 31 03 LAMP TEMP 32 PERM GAS TEMP 33 PRESSURE 34 PHOT BENCH TEMP 35 PHOT LAMP TEMP 36 PHOT INTENSITY 37 03 LEVEL Table C 2 Read Registers for 146i Register Number Variable 400018 40002 GAS CONC 40003 amp 40004 TARGET GFLOW 40005840006 ACTUAL GFLOW 400078 40008 TARGET ZFLOW 400098 40010 ACTUAL ZFLOW 400118 40012 TARGET TFLOW 400138 40014 ACTUAL TFLOW 400158 40016 03 CONC 40017840018 PERM CONC 400198 40020 PRESSURE 400218 40022 BENCH TEMP 40023840024 LAMP TEMP Model 146 Instruction Manual C 9 MODBUS Protocol MODBUS Commands Supported C 10 Model 146 Instruction Manual Register Number Variable 400258 40026 03 LAMP TEMP 40027840028 INTENSITY 40029840030 NOISE 40031840032 ANALOG 1 40033 amp 40034 ANALOG 2 40035 amp 40036 ANALOG 3 40037 amp 40038 ANALOG 4 40039 amp 40040 ANALOG 5 40041 amp 40042 ANALOG 6 40043 amp 40044 ANALOG 7 40045 amp 40046 ANALOG 8 Table C 3 Write Coils for 1467 Coil Number Action 101 GAS BIT 1 102 GAS BIT 2 103 GAS BIT 3 104 SPAN BIT 1 105 SPAN BIT 2 106 SPAN BIT 3 107 03 PERM BIT 1 108
26. Select Channel gt Point 1 10 The Volts screen allows the user to set the input voltage for the selected table point in the conversion table from 0 00 to 10 50 The default table is a two point table with point 1 0 00 V 000 0 U and point 2 10 00 V 10 0 U e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Select Point gt Volts The User Value screen allows the user to set the output value for the corresponding input voltage for the selected table point in the conversion table from 9999999 to 99999999 The default table is a two point table with point 1 0 00 V 000 0 U and point 2 10 00 V 10 0 U e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Table Point gt User Value Thermo Fisher Scientific Operation Instrument Controls Menu aa ri CORDED RAE mE AS A fee PEE EE Screen Contrast The Screen Contrast screen is used to change the contrast of the display Intensities between 0 and 100 in increments of 10 are available Changing the screen contrast may be necessary if the instrument is operated at extreme temperatures e Inthe Main Menu choose Instrument Controls gt Screen Contrast BE m A 7 En Hina Service Mode The Service Mode screen is used to turn the service mode on or off The service mode locks out any remote actions and includes parameters and
27. This is a powerful screen that allows the user to select the gas span level the permeation level the ozonator level and the photometer mode Like the run screen if any of these options are not installed then their controls are not displayed on the operation screen In addition flow and level settings for the active mode may be edited from this screen The first line of the operation screen allows selection of the dilution gas The Model 146i can support up to three external gases standard GAS A through C or up to six external gases optional GAS A through F The displayed gas names may be changed by the user via the GAS SETUP menu The number to the right of the gas selection is either the total flow or the zero air flow as indicated based on the currently selected span mode The span mode ZERO SPAN 1 through SPAN 5 and MANUAL is selected on the next line The instrument sets the flow parameters to meet the selected criteria For ZERO the gas flow controller is shut off and a previously determined zero air flow is output For SPAN 1 through SPAN 5 the zero and gas flow controllers are set to provide the previously determined total flow and gas concentration For MANUAL the flow rates for the zero and gas flow controllers may be set directly The next line is used to set the permeation oven output level if installed The allowable modes are OFF PERM 1 through 5 and manual The concentration for these modes is displayed on the righ
28. are available for the instrument and Table 9 2 provides the cable color codes For associated part numbers refer to External Device Connection Components in the Servicing chapter Thermo Fisher Scientific Model 146 Instruction Manual 9 11 Optional Equipment Cables 9 12 Model 146 Instruction Manual Note Table 9 2 provides the color coding for both 25 pin cables and 37 pin cables Color codes for pins 1 25 are for 25 pin cables color codes for pins 1 37 are for 37 pin cables Table 9 1 Cable Options Description Cable Length DB37M to open end Six feet DB37F to open end Six feet DB25M to open end Six feet RS 232 Table 9 2 Color Codes for 25 Pin and 37 Pin Cables Pin Color Pin Color 1 BLACK 20 RED BLACK 2 BROWN 21 ORANGE BLACK 3 RED 22 YELLOW BLACK 4 ORANGE 23 GREEN BLACK 5 YELLOW 24 GRAY BLACK 6 GREEN 25 PINK BLACK 7 BLUE End color codes for 25 pin cables continue for 37 pin cables 8 VIOLET 26 PINK GREEN 9 GRAY 27 PINK RED 10 WHITE 28 PINK VIOLET 11 PINK 29 LIGHT BLUE 12 LIGHT GREEN 30 LIGHT BLUE BROWN 13 BLACK WHITE 31 LIGHT BLUE RED 14 BROWN WHITE 32 LIGHT BLUE VIOLET 15 RED WHITE 89 LIGHT BLUE BLACK 16 ORANGE WHITE 34 GRAY GREEN 17 GREEN WHITE 35 GRAY RED 18 BLUE WHITE 36 GRAY VIOLET 19 VIOLET WHITE CH LIGHT GREEN BLACK Thermo Fisher Scientific Optional Equipment Mounting Options Mounting Options T
29. are only available if the photometer option is installed Send alarm bench lamp temp min Receive alarm bench lamp temp min 52 0 deg C set alarm bench lamp temp min value set alarm bench lamp temp max value These commands set the photometer bench lamp temperature minimum and maximum values to value where value is a floating point number representing bench lamp temperature alarm limits in degrees C The example below sets the bench lamp temperature alarm maximum value to 58 C These commands are only available if the photometer option is installed Send set alarm bench lamp temp max 58 Receive set alarm bench lamp temp max 58 ok Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Alarms alarm bench temp min alarm bench temp max These commands report the photometer bench temperature alarm minimum and maximum value current settings The example below reports the bench temperature alarm minimum value is 15 0 C These commands are only available if the photometer option is installed Send alarm bench temp min Receive alarm bench temp min 15 0 set alarm bench temp min value set alarm bench temp max value These commands set the photometer bench temperature minimum and maximum values to value where value is a floating point number representing bias voltage alarm limits in degrees C The example below sets the bench temperature alarm maximum value to 35 C These commands are only available if
30. cables not connected properly Defective measurement interface board Check that all internal cables are connected properly Recycle AC power to instrument If still alarming replace board Alarm 1 0 Exp Status 1 0 expansion board was installed but is not currently installed Re install board if desired and recycle power Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Troubleshooting Guides Table 6 2 Troubleshooting Alarm Messages continued Alarm Message Possible Cause Malfunctioning external device Internal cables not connected properly Defective optional I O expansion board Action Disconnect external devices from board and recycle power If alarm persists replace board Check that all internal cables are connected properly Replace board Model 146 Instruction Manual 6 7 Troubleshooting Board Level Connection Diagrams Board Level Figure 6 1 and Figure 6 2 are board level connection diagrams for the Connection D lagra Ms common electronics and measurement system These illustrations can be used along with the connector pin descriptions in Table 6 3 through Table 6 10 to troubleshoot board level faults EXPANSION I O 25 Pin TONMEASURAVENT SYSTEM DIGITAL OUTPUTS 37 Pin RS232 RS485 LCD DISPLAY RS232 RS485 10 BASE T EXPANSION RS485 15 Pin SPARE DATA 8 Pin ANALOG OUTPUTS DIGITAL INPUTS 37 pin IPJ8 3 Pin
31. choose Service gt Perm Oven Settings gt Cal Gas Thermistor Water Bath The Calibrate Gas Thermistor Water Bath screen is used to view and set the permeation oven gas temperature to a known temperature value e Inthe Main Menu choose Service gt Perm Oven Settings gt Cal Gas Thermistor gt Water Bath 3 96 Model 146 Instruction Manual Thermo Fisher Scientific Operation Service Menu Resistor The Calibrate Gas Thermistor Resistor screen is used to view and set the permeation oven gas thermistor resistance to a known resistor value e Inthe Main Menu choose Service gt Perm Oven Settings gt Cal Gas Thermistor gt Known Resistor Calibrate Oven Thermistor The Calibrate Oven Thermistor screen is used to view and set the permeation oven thermistor resistance to a known resistor value e Inthe Main Menu choose Service gt Perm Oven Settings gt Cal Oven Thermistor Permeation Oven Selection The Permeation Oven Setpoint screen is used to select the permeation oven state to not present or the desired temperature choice of 30 35 and 45 C e Inthe Main Menu choose Service gt Perm Oven Settings gt Perm Oven Selection Thermo Fisher Scientific Model 146 Instruction Manual 3 97 Operation Service Menu Factory Calibrate Gas Thermistor The Factory Calibrate Gas Thermistor menu is used for calibrating the permeation oven gas thermistor to either low point high point or set defaults e Inth
32. concentration of gas being supplied by the B 8 instrument gas index Reports sets the gas index value B 28 gas man Selects manual preset level for gas B 28 gas name Sets the gas name B 29 gas off Selects gas off B 29 gas params Returns parameters for gas tank conc span flow B 29 Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Commands Command Description Page gas solenoid Reports sets solenoid used when gas selected B 29 gas span Selects gas span level 0 1 2 3 4 or 5 B 30 gas span conc Reports sets target concentration for specified gas and span B 30 level gas span flow Reports sets target total flow for specified gas and span B 31 level gas tank conc Reports sets concentration of cylinder connected as specified B 31 gas gflow Reports flow through gas flow controller B 8 he help Simulates pressing help pushbutton B 25 host name Reports sets host name string for Ethernet port B 41 instr name Reports instrument name B 42 instrument id Reports sets instrument id B 42 intensity Reports photometer detector intensity in Hz B 9 intern temp Reports current internal instrument temperature B 15 isc iscreen Retrieves framebuffer data used for the display B 26 11 Reports sets custom level 1 setting of ozonator lamp drive B 32 11 conc Reports sets custom level 1 concentration for photometer B 32 control of ozonator la
33. expendable items are expressly excluded from the warranty under this warranty Notwithstanding the foregoing Products supplied by Seller that are obtained by Seller from an original manufacturer or third party supplier are not warranted by Seller but Seller agrees to assign to Buyer any warranty rights in such Product that Seller may have from the original manufacturer or third party supplier to the extent such assignment is allowed by such original manufacturer or third party supplier In no event shall Seller have any obligation to make repairs replacements or corrections required in whole or in part as the result of i normal wear and tear ii accident disaster or event of force majeure iii misuse fault or negligence of or by Buyer iv use of the Products in a manner for which they were not designed v causes external to the Products such as but not Model 146 Instruction Manual A 1 Warranty limited to power failure or electrical power surges vi improper storage and handling of the Products or vii use of the Products in combination with equipment or software not supplied by Seller If Seller determines that Products for which Buyer has requested warranty services are not covered by the warranty hereunder Buyer shall pay or reimburse Seller for all costs of investigating and responding to such request at Seller s then prevailing time and materials rates If Seller provides repair services or replacement parts that
34. following procedure to replace the photometer board Figure 7 10 Equipment required Photometer board Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a Use the following procedure to replace the photometer board 1 Turn instrument OFE unplug the power cord and remove the cover 2 Remove the ozonator board Refer to Ozonator Board Replacement Ozonator option on page 7 45 3 Remove the cables from the board 4 Remove the two screws from the top of the board 5 Pop off the board from the bottom mounting studs and remove the board 6 To install the new photometer board follow previous steps in reverse 7 Reinstall the ozonator board Use the following procedure to replace the photometer detector Figure 7 12 Equipment Required Detector assembly Model 146 Instruction Manual 7 35 Servicing Photometer Thermistor Replacement Allen wrench 1 16 inch amounts of static electricity A properly grounded antistatic wrist strap must Equipment Damage Some internal components can be damaged by small be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the plumbing and electrical connections 3 Loosen the captive screws 4 securing the photometer to the floor
35. functions that are useful when making adjustments or diagnosing the Model 1467 For more information about the service mode see Service Menu later in this chapter Note The service mode should be turned off when finished as it prevents remote operation A e Inthe Main Menu choose Instrument Controls gt Service Mode Thermo Fisher Scientific Model 146 Instruction Manual 3 69 Operation Diagnostics Menu Date Time The Date Time screen allows the user to view and change the system date and time 24 hour format The internal clock is powered by its own battery when instrument power is off e Inthe Main Menu choose Instrument Controls gt Date Time Diagnostics The Diagnostics menu provides access to diagnostic information and Menu functions This menu is useful when troubleshooting the instrument The analog input readings and analog input voltages are only displayed if the I O expansion board option is installed e Inthe Main Menu choose Diagnostics a pn Eus fel pipi H DH pikor E E 3 70 Model 146 Instruction Manual Thermo Fisher Scientific Operation Diagnostics Menu Prog ram Version The Program Version screen read only shows the version number of the program installed Prior to contacting the factory with any questions regarding the instrument please note the product model name and program version number e Inthe Main Menu choose Diagnostics gt Program Version DIEM RET PT
36. insulation off the lamp handle onto the lamp cord 4 Loosen both lamp clamp screws 5 Carefully slide lamp out of ozonator housing 6 Slide insulation off of old lamp and slide onto new lamp cord 7 Carefully insert new lamp into ozonator housing until it bottoms Pull lamp out approximately 1 16 inch to allow for expansion when the lamp warms up and tighten the Allen screws on the clamp 8 Slide the lamp insulation against the ozonator 9 Re connect the ozonator LAMP connector to the ozonator board 10 Replace the cover and reconnect power Thermo Fisher Scientific Ozonator Lamp Heater Replacement Ozonator option Thermo Fisher Scientific A Servicing Ozonator Lamp Heater Replacement Ozonator option Use the following procedure to replace the ozonator lamp heater Equipment Required Ozonator lamp heater Allen wrench 3 32 inch and 7 64 inch Philips screwdriver Heat conductive compound Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unplug heater cable from the LAMP HEATER and LAMP connectors on the ozonator board Figure 7 10 3 Disconnect plumbing from the ozonator 4 Remove the ozonator by removing the two screws securing the ozonator to the floor plate and lifting the ozo
37. is determined This includes the sample pump and solenoid valves which have a limited life Other operations such as cleaning the optics and checking the pressure calibration should be performed on a regular basis This chapter includes the following maintenance information and replacement procedures e Safety Precautions on page 5 2 e Replacement Parts on page 5 2 e Accessing the Service Mode on page 5 2 e Outside Case Cleaning on page 5 2 e Fan Filter Inspection and Cleaning on page 5 3 e Photometer Cleaning on page 5 3 e Monitoring Detector Frequency and Noise on page 5 4 e Capillary Service on page 5 5 e Ozonator Capillary Replacement Ozonator option on page 5 6 e Pressure Regulator Capillary Replacement on page 5 7 e Pump Capillary Replacement Photometer option on page 5 8 e Pump Rebuilding on page 5 9 Model 146 Instruction Manual 5 1 Preventive Maintenance Safety Precautions Safety Precautions Read the safety precautions before beginning any procedures in this chapter manufacturer the protection provided by the equipment may be impaired yA WARNING If the equipment is operated in a manner not specified by the amounts of static electricity A properly grounded antistatic wrist strap must Equipment Damage Some internal components can be damaged by small be worn while handling any internal component For more information about ap
38. lamp voltage in volts This command is only available if the photometer option is installed Send lamp voltage bench Receive lamp voltage bench 10 3 V lamp voltage 03 This command reports the ozonator lamp voltage in volts This command is only available if the photometer option is installed Send lamp voltage 03 Receive lamp voltage 03 12 3 V oz lamp temp This command reports the current ozonator lamp temperature The example below shows the current ozonator temperature is 68 7 C This command is only available if the photometer option is installed Send oz lamp temp Receive oz lamp temp 68 7 deg C photometer temp This command reports the current photometer bench temperature The first reading is the temperature being used in instrument calculations The second temperature is the actual temperature being measured If temperature compensation is on then both temperature readings are the same If temperature compensation is off a temperature of 0 0 C is shown as the default temperature even though the actual internal temperature is 31 2 C The example below shows that temperature compensation is on and that the internal temperature is 31 2 C Model 146 Instruction Manual B 15 C Link Protocol Commands Datalogging Datalogging B 16 Model 146 Instruction Manual Send photometer temp Receive photometer temp 31 2 deg C actual 31 2 program This command reports programmed events status Send program Rec
39. minimum value B 13 alarm trig conc 03 Reports sets Oz concentration low alarm trigger value B 13 analog iout range Reports sets analog current output range per channel B 43 analog vin Retrieves analog voltage input data per channel B 43 analog vout range Reports sets analog voltage output range per channel B 44 avg time Reports sets averaging time B 8 baud Reports sets baud rate B 40 cal 03 bkg Sets auto calibrates Oz background B 23 cal 03 coef Sets auto calibrates Dz coefficient B 23 clr Irecs Clears away only long records that have been saved B 16 clr records Clears away all logging records that have been saved B 16 clr srecs Clears away only short records that have been saved B 16 contrast Reports sets current front panel screen contrast B 37 copy Irec to sp Copies current Irec selection into the scratch pad B 21 copy sp to Irec Copies current selections in scratch pad into Irec list B 21 copy sp to srec Copies current selections in scratch pad into srec list B 21 Model 146 Instruction Manual B 3 C Link Protocol Commands Commands B 4 Model 1467 Instruction Manual Command Description Page copy sp to stream Copies current selections in scratch pad into stream list B 21 copy srec to sp Copies current srec selection into the scratch pad B 21 copy stream to sp Copies current streaming data selection into the scratch pad B 21 custom Reports sets defined custom range concentration B 36 da
40. photometer O concentration alarm trigger action for minimum alarm to either floor or ceiling The example below shows the O concentration minimum alarm trigger to ceiling according to Table B 3 These commands are only available if the photometer option is installed Send alarm trig conc 03 Receive alarm trig conc 03 1 set alarm trig conc 03 value These commands set the photometer O concentration alarm trigger action for the minimum alarm to value where value is set to either floor or ceiling according to Table B 3 The example below sets the O3 concentration minimum alarm trigger to floor These commands are only available if the photometer option is installed Send set alarm trig conc 0 Receive set alarm trig conc 0 ok Model 146 Instruction Manual B 13 C Link Protocol Commands Diagnostics Diagnostics B 14 Model 1467 Instruction Manual Table B 3 Alarm Trigger Values Value Alarm Trigger 0 Floor 1 Ceiling diag volt mb This command reports the diagnostic voltage measurements on the motherboard The sequence of voltages is Positive 24 positive 15 positive 5 positive 3 3 and negative 3 3 Each voltage value is separated by a space Send diag volt mb Receive diag volt mb 24 1 14 9 4 9 3 2 3 2 diag volt mib This command reports the diagnostic voltage measurements on the measurement interface board The sequence of voltages is Positive 24 positive 15 negative 15 positive 5 positive 3 3 Each vo
41. pressure transducer 5 Remove the back pressure regulator handle 6 Remove two 2 screws holding the regulator to the bracket 7 Remove the back pressure regulator 8 Install a new back pressure regulator and calibrate 9 After calibration tighten the locking nut securing the position of the back pressure regulator handle 10 Remove flow controller assembly Thermo Fisher Scientific Pressure Regulator s Thermo Fisher Scientific Adjustment A A Servicing Pressure Regulator s Adjustment 11 Remove defective flow controller from the bracket by removing the two retaining screws 12 Install the new flow controller by following the previous steps in reverse 13 Check the system for leaks Refer to Leak Test and Pump Checkout in this chapter 14 Recalibrate Use the following procedure to adjust the pressure regulator and the back pressure regulator Equipment required Digital flow meter capable of measuring 150 sccm WARNING The service procedures in this manual are restricted to qualified service representatives If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Configure a digital flow meter capable of m
42. previously selected gas The screen identifies the minimum and maximum acceptable zero air flow rates based on the installed flow controller and lets the user select a flow rate between these two settings If the selected zero air flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Flow Modes gt Gas Dilution gt Gas A F gt Manual gt Zero Air Gas The Manual Gas Flow screen sets the manual gas flow value for the previously selected gas The screen identifies the minimum and maximum acceptable gas flow rates based on the installed flow controller and lets the user select a flow rate between these two settings If the selected gas flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Flow Modes gt Gas Dilution gt Gas A F gt Manual gt Gas DH Guided ML PRA Permeation Oven The Permeation Oven flow mode menu is used to select the permeation oven output level if the permeation oven option is installed Once selected the screen reports the new flow mode The user must press 3 12 Model 146 Instruction Manual Thermo Fisher Scientific Operation Flow Modes to accept and activate these settings The message SAVING is displayed for 3 seconds then removed e In the Main Menu choose Flow Modes gt Permeation Oven gt Level 1 5 Permeation Oven Ga
43. pushbutton B 26 save Stores parameters in FLASH B 38 save params Stores parameters in FLASH B 38 sc screen C series legacy command that reports a generic response B 27 Use iscreen instead sp conc Reports sets photometer span concentration for calibration B 25 sp field Reports sets item number and name in scratch pad list B 22 span index Reports sets span gas level B 36 sr Reports last short record stored B 17 srec Reports maximum number of short records B 18 srec format Reports sets output format for short records ASCII or binary B 19 srec layout Reports current layout of short record data B 19 srec mem size Reports maximum number of short records B 20 srec per Reports sets short record logging period B 20 status Reports gas and span selections B 9 stream per Reports sets current set time period for streaming data B 22 stream time Reports sets a time stamp to streaming data or not B 22 temp comp Reports sets photometer temperature compensation on or off B 37 Model 146 Instruction Manual B 7 C Link Protocol Commands Measurements Measurements B 8 Model 1467 Instruction Manual Command Description Page time Reports sets current time 24 hour time B 38 up Simulates pressing up pushbutton B 26 zflow Reports flow through zero air flow controller B 8 avg time This command reports the photometer averaging time in seconds The example below shows that the averaging time is 300 seconds according to Table B 2 This comma
44. reset the configuration values to factory defaults e Inthe Main Menu choose Service gt Perm Oven Settings gt Factory Cal Oven Therm gt Set Defaults Bad in Be BE Fas fast Enee Moo MO PT DH iy PL et Lamp Setting The Lamp Setting screen is used to adjust the photometer lamp drive The intensity and current lamp drive settings are displayed Adjust the lamp setting until the intensity is about 10000 Hz The lamp setting is visible only if the photometer option is installed and when the calibrator is in service mode For more information on the service mode see Service Mode earlier in the chapter e Inthe Main Menu choose Service gt Lamp Setting 3 100 Model 146 Instruction Manual Thermo Fisher Scientific Operation Service Menu Intensity Check The Intensity Check menu is used to manually control the flow of reference of sample gas through the photometer This enables the intensity and noise reading of the detector to be determined with either reference or sample gas flow Selecting any of these menu choices will disturb the analog outputs The intensity check is visible only if the photometer option is installed and when the calibrator is in service mode For more information on the service mode see Service Mode earlier in the chapter e Inthe Main Menu choose Service gt Intensity Check Reference Gas The Reference Gas screen read only sets the valves to route the reference gas
45. screen allows the user to select the alarm status for the selected solenoid output The selected item is shown by lt after it e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Ext Solenoid Settings gt Select Solenoid gt Instrument State gt Alarms a a En Non Alarm The Non Alarm status screen allows the user to select the non alarm status for the selected solenoid output The selected item is shown by lt after it e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Ext Solenoid Settings gt Select Solenoid gt Instrument State gt Non Alarm PENEN En oE Banas Be Model 1467 Instruction Manual 3 61 Operation Instrument Controls Menu Analog Output The Analog Output Configuration menu displays a list of the analog Configuration output channels available for configuration Channel choices include all voltage channels all current channels voltage channels 1 6 and current channels 1 6 if the I O expansion board option is installed Configuration choices include selecting range setting minimum maximum values and choosing signal to output e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config rusas pr E a T a E ip priori A ML a EA Select Output Range The Select Output Range screen is used to select the hardware range for the selected analog output channel Possibl
46. selectable via software The analog outputs are defaulted based on the measurement range The defaults are calculated by dividing the data values by the full scale range for each of the three parameters and then multiplying each result by the user selected output range Negative concentrations can be represented as long as they are within 5 of full scale The zero and span values may be set by the user to any desired value All electronics operate from a universal switching supply which is capable of auto sensing the input voltage and working over the entire operating range Internal pumps and heaters all operate on 110VAC An optional transformer is required if operating on the 210 250VAC or 90 1 10VAC ranges An on off switch controls all power to the instrument and is accessible on the front panel Thermo Fisher Scientific Motherboard External Connectors Internal Connectors Measurement Interface Board Measurement Interface Board Thermo Fisher Scientific Connectors System Description Electronics The motherboard contains the main processor power supplies a sub processor and serves as the communication hub for the instrument The motherboard receives operator inputs from the front panel mounted function key panel and or over I O connections on the rear panel and sends commands to the other boards to control the functions of the instrument and to collect measurement and diagnostic information The motherboard out
47. sets the minimum and maximum alarm limits Acceptable alarm limits range from 45 to 65 C If the bench temperature reading goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu e Inthe Main Menu choose Alarms gt Phot Lamp Temp EH priori E ES E The Minimum Photometer Lamp Temperature alarm limit screen is used to change the minimum lamp temperature alarm limit The minimum and maximum lamp temperature screens function the same way e In the Main Menu choose Alarms gt Sample Flow gt Min or Max Thermo Fisher Scientific Operation Alarms Menu Photometer Intensity The Photometer Intensity screen displays the current lamp intensity and sets the minimum and maximum alarm limits Acceptable alarm limits range from 0 to 300 000 Hertz If the intensity reading goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu e Inthe Main Menu choose Alarms gt Phot Intensity Min and Max Photometer The Minimum Photometer Intensity alarm limit screen is used to change Intensity Limits the minimum intensity alarm limit The minimum and maximum photometer intensity screens function the same way e Inthe Main Menu choose Alarms gt Sample Flow gt Min or Max WH mri Thermo Fisher Scientific Model 146 Instruction Manual 3 83 Operation Se
48. shows the use of indirect precision specifiers for floating point displays The background value is taken from the 7th element and the precision specifier is taken from the 8th If the asterisk were not present it would indicate instead that 8 digits after the decimal point should be displayed Thermo Fisher Scientific Serial Communication Thermo Fisher Scientific Parameters Appendix C MODBUS Protocol This appendix provides a description of the MODBUS Protocol Interface and is supported both over RS 232 485 RTU protocol as well as TCP IP over Ethernet The MODBUS Commands that are implemented are explained in detail in this document The MODBUS protocol support for the iSeries enables the user to perform the functions of reading the various concentrations and other analog values or variables read the status of the digital outputs of the analyzer and to trigger or simulate the activation of a digital input to the instrument This is achieved by using the supported MODBUS commands listed below For details of the Model 146i MODBUS Protocol specification see the following topics e Serial Communication Parameters on page C 1 e TCP Communication Parameters on page C 2 e Application Data Unit Definition on page C 2 e Function Codes on page C 3 e MODBUS Commands Supported on page C 8 Additional information on the MODBUS protocol can be obtained at http www modbus org References are from
49. the photometer option is installed Send set alarm bench temp max 35 Receive set alarm bench temp max 35 ok alarm conc 03 min alarm conc 03 max These commands report the photometer O concentration alarm minimum and maximum value current setting The example below reports the O concentration alarm minimum is 20 ppb These commands are only available if the photometer option is installed Send alarm conc 03 min Receive alarm conc 03 min 20 0 ppm set alarm conc 03 min value set alarm conc 03 max value These commands set the photometer O3 concentration minimum and maximum values to value where value is a floating point representation of the concentration alarm limits in ppb The example below sets the O concentration alarm maximum value to 5 These commands are only available if the photometer option is installed Send set alarm conc 03 max 5 Receive set alarm conc 03 max 5 ok alarm intensity min alarm intensity max These commands report the photometer detector intensity alarm minimum and maximum value current settings The example below reports the intensity alarm minimum value is 45000 Hz These commands are only available if the photometer option is installed Model 146 Instruction Manual B 11 C Link Protocol Commands Alarms B 12 Model 146 Instruction Manual Send alarm intensity min Receive alarm intensity min 45000 Hz set alarm intensity min value set alarm intensity max value These commands set t
50. through the photometer and displays the intensity and noise e Inthe Main Menu choose Service gt Intensity Check gt Reference Gas i mypnpri PEE Rr Ee Toy dE ES Brane i i PME hae TEEN EE Sample Gas The Sample Gas screen read only sets the valves to route the sample gas through the photometer and displays the intensity and noise e Inthe Main Menu choose Service gt Intensity Check gt Sample Gas Thermo Fisher Scientific Model 1467 Instruction Manual 3 101 Operation Service Menu 0 Response Coefficient Display Pixel Test 3 102 Model 1467 Instruction Manual The O Response Coefficient screen is used to manually adjust the O response coefficient for the photometer option The value is used to adjust how fast the ozonator drive is changed when under photometer control and should usually be left at 1 000 The O response coefficient is visible only if the photometer option is installed and when the calibrator is in service mode For more information on the service mode see Service Mode earlier in the chapter e Inthe Main Menu choose Service gt O3 Response Coefficient The Display Pixel Test is used to test the LCD display The display pixel test is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter e Inthe Main Menu choose Service gt Display Pixel Test Thermo Fisher Sci
51. to 3 characters in length and defaults to V volts e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Units 3 66 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu gengp MO Tp AS BE Fas fast Bosse Bast ade 5 Decimal Places The Decimal Places screen allows the user to select how many digits are displayed to the right of the decimal from 0 to 6 with a default of 2 e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Decimal Places pE EN En REP E e Tit CH ninii Bati ha Be E Eiet fast Bane Bast i PME bee mi i Number of Table Points The Number of Table Points screen allows the user to select how many points are used in the conversion table The points range from 2 to 10 with a default of 2 e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt Select Channel gt Table Points a En REP ED Pr Tit LF EE til WH mri Pin Piri Thermo Fisher Scientific Model 146 Instruction Manual 3 67 Operation Instrument Controls Menu Table Point Volts User Value 3 68 Model 146 Instruction Manual The Table Point submenu allows the user to set up an individual table point e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Input Config gt
52. to maximum allowed by instrument Output Value 2 Bytes 0x0000 or OxFFOO Response Function code 1 Byte 0x05 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Output Value 2 Bytes 0x0000 or OxFFOO Error Response Function code 1 Byte Function code 0x80 Exception code 1 Byte 01 lllegal Function 02 lllegal Address 03 lllegal Data 04 Slave Device Failure Here is an example of a request to write Coil 5 ON Request Field Name Hex Function 05 Output Address Hi 00 Output Address Lo 05 Model 146 Instruction Manual C 7 MODBUS Protocol MODBUS Commands Supported MODBUS Commands Supported C 8 Model 146 Instruction Manual Output Value Hi Output Value Lo Response Field Name Function Dutput Address Hi Dutput Address Lo Output Value Hi Output Value Lo FF 00 Hex 05 00 05 FF 00 The following Table C 1 through Table C 3 lists the MODBUS commands supported for the Model 146i Table C 1 Read Coils for 1467 Coil Number 1 Status GASBIT 1 GAS BIT 2 GAS BIT 3 SPAN BIT 1 SPAN BIT 2 SPAN BIT 3 03 PERM BIT 1 03 PERM BIT 2 OoO ao lrwTa sl wai oy AJJ pro 03 PERM BIT 3 gt o 03 PERM BIT 4 gt PHOTO BIT 1 PHOTO BIT 2 GASA GAS B GAS C GAS D GAS E GAS F gt gt gt gt gt gt gt gt co loo N o a A w N OZONATOR Thermo Fisher Scientific Thermo Fisher
53. to the operating parameters via events C Link or digital inputs this screen will show the new values and any unsaved changes will be lost A e Inthe Main Menu choose Operation or press the OPER soft key e Use l t and to move the cursor up and down to select a mode or setting or change the number in the numeric edit field e Press to save the mode switch to numeric entry or save the numeric entry e Use Ce and gt to change the mode or move the cursor in the numeric edit field e Press M to return to the Main Menu or _ gt to return to the Run screen rio ei Li bel Flow Modes The Flow Modes menu allows the user to set the operational mode of the calibrator in a more intuitive manner than the operation screen Items displayed are determined by the options installed After selecting the desired mode a summary screen is displayed to show the new mode and the user must press to activate the new mode e Inthe Main Menu choose Flow Modes or press the MODE soft key Thermo Fisher Scientific Model 146 Instruction Manual 3 9 Operation Flow Modes Cee TN E Pike EE EE Standby The Standby screen is used to turn the standby mode on or off When Standby mode is on the zero air and all of the gas valves are closed and the ozonator and photometer are turned off A small amount of Zero air is still routed through the permeation oven and photometer The previous operating parameters are restor
54. tube in the Transfer Standard or install a certified permeation tube Allow the permeation tubes in both the 1467 and the Transfer Standard to stabilize for at least 48 hours Carefully calibrate the 146 using the Transfer Standard The output of the Transfer Standard should be connected to the SAMPLE bulkhead on the rear panel of the 1467 Switch the calibrated 1467 into the span mode Model 146 Instruction Manual 9 9 Optional Equipment Six Span Inputs Six Span Inputs Span Gas Mass Flow Controller Zero Gas Mass Flow Controller I O Expansion Board 9 10 Assembly Model 146 Instruction Manual 6 Measure the flow rate into the ZERO bulkhead on the rear panel of the 146 Be sure that the charcoal scrubber is connected Note the flow and measured gas concentration 7 From the flow and measured concentration compute the permeation tube release rate The six span input option increases the number of span ports from three standard to six The span gas mass flow controller governs low gas flow to control gas being diluted Span gas mass flow controllers include e 25 SCCM e 50 SCCM 100 SCCM standard 200 SCCM 500 SCCM The zero gas mass flow controller governs high gas flow to control zero air Zero gas mass flow controllers include e 5SLPM 10 SLPM standard 15 SLPM e 20SLPM The I O expansion board provides six analog current output channels 0 20 mA or 4 20 mA and eight analog
55. two accurate mass flow controllers One governs high flow typically 10 LPM fullscale to control the diluting zero air the other governs low flow typically 100 sccm for controlling the gas to be diluted The two components are mixed in a Teflon mixing chamber Since the mass flow controllers used do not have a positive shutoff a zero air solenoid shuts off zero air when the system is in standby to reduce zero air use The Model 146i uses as many as six different independent gas inlets which are individually controlled by gas solenoids A through F depending on the mode being commanded Ozone Generation The Model 146 produces ozone by exposing air to light at 185 nm The ozone level can be changed by changing the flow through the ozone generator or by changing the light intensity In the Model 146i the selected light intensity level is held constant by temperature controlling the lamp and by using a highly stable lamp power supply The flow is held constant by the mass flow controller Gas Phase Titration Gas phase titration in the Model 1467 is accomplished by combining the outputs of the precision gas dilution system with that of the ozone generating source The gas dilution system is used to produce an accurately known concentration of NO from a known concentration gas cylinder The amount of NO formed is determined by mixing the NO concentration with ozone and measuring the loss of NO it is equal to the NO loss The only caution t
56. using a 1 4 inch piece of Teflon tubing so as not to damage the tube Use a cotton swab to clean the window surfaces through the holes that the tube fits into 9 Replace the photometer components following the previous steps in reverse Note It is recommended that the Model 146i be recalibrated if the photometer has been completely disassembled Always check for leaks after replacing any component Refer to Leak Test and Pump Checkout in the Servicing chapter 10 Re install the instrument cover The Model 1467 measures intensity ratios and not absolute values Therefore a large range of detector frequencies are acceptable for proper operation of the instrument The nominal values are 65 to 120 kHz These frequencies can be monitored from the Intensity Check menu in the Service menu The Intensity Check menu is used to manually control the flow of reference or sample gas through the photometer This enables the detector intensity and noise reading to be determined with either reference or sample gas Thermo Fisher Scientific Capillary Service Thermo Fisher Scientific Preventive Maintenance Capillary Service Note Selecting these menu choices will affect the analog outputs 1 Press 1 to display the Main Menu 2 Usel to scroll to Service press gt to scroll to Intensity Check and press 4 The Intensity Check menu appears Note If Service is not displayed refer to Access
57. voltage inputs 0 10V The DB25 connector on the rear panel provides the interface for these inputs and outputs Thermo Fisher Scientific Optional Equipment Terminal Block and Cable Kits 25 Pin Terminal Board The 25 pin terminal board assembly is included with the optional I O Assembly expansion board Refer Terminal Board PCB Assemblies in the Installation chapter for information on attaching the cable to the connector board For associated part numbers refer to External Device Connection Components in the Servicing chapter Terminal Block and The optional terminal block and cable kits provide a convenient way to Cable Kits connect devices to the instrument These kits break out the signals on the rear panel connector to individual numbered terminals Two types of terminal block and cable kits are available One kit is for the DB37 connectors and can be used for either the analog output connector or the relay output connector The other kit is for the DB25 connector and can be used for the optional I O expansion board For associated part numbers refer to External Device Connection Components in the Servicing chapter Each kit consists of e one six foot cable e one terminal block e one snap track Note Supporting all of the connections on units with the optional I O expansion board requires e two DB37 kits e one DB25 kit Cables Tabie 9 1 identifies the optional individual cables that
58. with smart external devices that may be mounted hundreds of feet from the analyzer using an RS 485 electrical interface Model 146 Instruction Manual 8 15 System Description 1 0 Components 8 16 Model 146 Instruction Manual Thermo Fisher Scientific Ozonator Photometer Zero Air Source Thermo Fisher Scientific Chapter 9 Optional Equipment The Model 146i is available with the following options e Ozonator on page 9 1 e Photometer on page 9 1 e Zero Air Source on page 9 1 e Permeation Oven on page 9 2 e Six Span Inputs on page 9 10 e Span Gas Mass Flow Controller on page 9 10 e Zero Gas Mass Flow Controller on page 9 10 e I O Expansion Board Assembly on page 9 10 e Terminal Block and Cable Kits on page 9 11 e Cables on page 9 11 e Mounting Options on page 9 13 The internal ozone generator provides easy determination of zero precision and Level 1 span checks The ozone level produced is a function of light intensity at 185 nm and gas flow The light intensity is varied by changing the current into the lamp The gas flow is held constant by a pressure regulator followed by a capillary The photometer measures ozone concentration and is used to control ozone output The zero air source option is a convenient system for generating pollutant free zero gas for Oz monitoring requirements This option includes the pump filter filter holder and Si
59. y P 8 p procedure and then continue with the Setting Temperature with Known Resistance procedure In either case use the following procedure for setting the perm oven temperature 1 Unplug POJ3 from the measurement interface board Place a 4 369 KQ resistor across pins 3 and 4 on the board Model 146 Instruction Manual 9 5 Optional Equipment Permeation Oven 2 From the Permeation Oven menu in the Service menu select Cal Oven Thermistor Figure 9 1 conocia PI Lace ip gies E Figure 9 1 Cal Oven Therm Resistor Screen 3 Enter the exact value of the attached resistor and press to save the offset 4 Press Cm to return to the Permeation Oven menu 5 Remove the resistor and re attach the permeation oven to POJ3 Setting Temperature with Water Use the following procedure for setting the measure temperature with water Bath bath 1 Remove the thermistor from the permeation tube oven but leave the thermistor connected to the measurement interface board Insert the thermistor into the water bath next to an NIST traceable thermometer if necessary use an extension cable to reach 2 Turn on the power to the water bath Using an NIST traceable thermometer with a resolution of 0 01 C adjust the water bath to 45 C 9 6 Model 146 Instruction Manual Thermo Fisher Scientific Optional Equipment Permeation Oven 3 From the Permeation Oven menu in the Service menu select Cal Gas Thermist
60. 0 Manual Pins 7 8 27 and 9 select an ozonator or permeation oven level gas Oz Perm Bit 4 Oz Perm Bit 3 02z Perm 2 Oz Perm Bit 1 Ozonator or Pin 9 Pin 27 Pin 8 Pin 7 Permeation Oven Levell 0 0 0 0 Ozon Perm Off 0 0 0 1 Ozon Manual 0 0 1 0 Ozon 1 0 0 1 1 Ozon 2 0 1 0 0 Ozon 3 0 1 0 1 Ozon 4 Model 146 Instruction Manual 2 7 Installation Remote Interconnections 2 8 Outputs Model 146 Instruction Manual 1 1 0 Ozon 5 1 1 Perm 1 0 0 0 Perm 2 0 0 Perm 3 0 1 0 Perm 4 0 1 1 Perm 5 Pins 28 and 29 select the photometer mode Phot Bit 2 Phot Bit 1 Pin 29 Pin 28 0 0 Photometer Off 0 1 Photometer Internal 1 0 Photometer External 1 1 Photometer Ozonator Control Pin 11 sets all analog outputs to zero This can be used to test the analog output functionality Pin 30 sets all analog outputs to full scale values This can be used to test the analog output functionality The Model 1467 is controllable through the 37 pin female rear panel connector The output relays report the status of the 1467 They default to Normally Open meaning that the relays are closed when the condition exists and can be user configured as NO or NC Refer to the following tables for the indicated gas gas span level and ozonator or permeation oven level for the Model 146i Pins 1 2 and 3 to report the selected gas 1 indicates contact closure Gas Bit 3 Gas Bit 2 Gas Bit 1 Gas Type Pins 3 22 Pins 2 21 P
61. 01 ok instr name This command reports the instrument name Send instr name Receive instr name Dynamic Gas Calibrator Dynamic Gas Calibrator instrument id This command reports the instrument id Send instrument id Receive instrument id 51 set instrument id value This command sets the instrument id to value where value is a decimal number between 0 and 127 inclusive Note sending this command via RS 232 or RS 485 will require the host to use the new id for subsequent commands 5 Send set instrument id 52 Receive set instrument id 52 ok mode This command reports what operating mode the instrument is in local service or remote The example below shows that the instrument is in the remote mode Send mode Receive mode remote set mode local set mode remote These commands set the instrument to local or remote mode The example below sets the instrument to the local mode Send set mode local Receive set mode local ok program no This command reports the analyzer s model information and program version number which will be dependant on the current version Thermo Fisher Scientific 1 0 Configuration Thermo Fisher Scientific C Link Protocol Commands 1 0 Configuration Send program no Receive program no iSeries 146i 01 00 00 100 set layout ack This command disables the stale layout layout change indicator that is attached to each response if the layout has changed Send set la
62. 03 PERM BIT 2 109 03 PERM BIT 3 110 03 PERM BIT 4 111 PHOTO BIT 1 112 PHOTO BIT 2 113 AOUTS TO ZERO 114 AOUTS TO FS Thermo Fisher Scientific
63. 15 LD3 LCD Signal 16 LCD Bias Voltage 17 5V 18 Ground 19 Ground 20 LCD_ONOFF LCD Signal 21 Keypad Row 2 Input 22 Keypad Row 1 Input 23 Keypad Row 4 Input 24 Keypad Row 3 Input 25 Keypad Col 2 Select 26 Keypad Col 1 Select 21 Keypad Col 4 Select 28 Keypad Col 3 Select 29 Ground 30 Ground 31 Ground 32 Ground 33 24V 34 24V LCD DATA J2 1 LFLM_5V LCD Signal 2 LLP_5V LCD Signal 3 LCLK_5V LCD Signal 4 LCD_ONOFF_5V LCD Signal 5 5V 6 16 Model 146 Instruction Manual Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Table 6 4 Front Panel Board Connector Pin Diagram continued Connector Reference Pin Signal Label Designator Description Ground LCD Bias Voltage LDO_5V LCD Signal LD1_5V LCD Signal 10 LD2_5V LCD Signal LD3_5V LCD Signal LD4_5V LCD Signal LD5_5V LCD Signal LD6_5V LCD Signal LD7_5V LCD Signal oOo co N om gt gt gt gt gt gt o a gt Se N Ground KEYBOARD J3 1 Keypad Row 1 Input Keypad Row 2 Input Keypad Row 3 Input Keypad Row 4 Input Keypad Col 1 Select Keypad Col 2 Select Keypad Col 3 Select mo N oo A OQ N Keypad Col 4 Select LCD Backlight Voltage 1 NC NC LCD Backlight Voltage 2 LCD BACKLIGHT J4 3 gt WwW N Table 6 5 0 Expansion Board Optional Connector Pin Descriptions Connector Reference Pin Signal Label Designator Description EXPANSION I O J1 3 A
64. 16 digital inputs user select programmable TTL level pulled high Serial Ports 1 RS 232 or RS 485 with two connectors baud rate 1200 115200 Protocols C Link MODBUS and streaming data all user selectable Ethernet connection In non condensing environments RJ45 connector for 10Mbs Ethernet connection static or dynamic TCP IP addressing Thermo Fisher Scientific Lifting A Unpacking and Thermo Fisher Scientific Inspection Chapter 2 Installation Installing the Model 146i includes the following recommendations and procedures e Lifting on page 2 1 e Unpacking and Inspection on page 2 1 e Setup Procedure on page 2 3 e Remote Interconnections on page 2 6 e Connecting External Devices on page 2 10 e Startup on page 2 14 When lifting the instrument use procedure appropriate to lifting a heavy object such as bending at the knees while keeping your back straight and upright Grasp the instrument at the bottom in the front and at the rear of the unit Although one person can lift the unit it is desirable to have two persons lifting one by grasping the bottom in the front and the other by grasping the bottom in the rear Equipment Damage Do not attempt to lift the instrument by the cover or other external fittings a The Model 146i is shipped complete in one container If there is obvious damage to the shipping container when you receive the instrument notify the
65. 2 13 14 15 16 17 18 19 20 Thermo Fisher Scientific Relay 1 Contact a Relay 2 Contact a Relay 3 Contact a Relay 4 Contact a Relay 5 Contact a Relay 6 Contact a Relay 7 Contact a Relay 8 Contact a Relay 9 Contact a Relay 10 Contact a NC Solenoid Drive Output 1 Solenoid Drive Output 2 Solenoid Drive Output 3 Solenoid Drive Output 4 Solenoid Drive Output 5 Solenoid Drive Output 6 Solenoid Drive Output 7 Solenoid Drive Output 8 Relay 1 Contact b Model 146 Instruction Manual 6 19 Troubleshooting Connector Pin Descriptions Table 6 6 Digital Output Board Connector Pin Descriptions continued Connector Label Reference Designator Pin 21 22 23 24 25 26 27 28 29 30 31 32 33 34 35 36 37 Signal Description Relay 2 Contact b Relay 3 Contact b Relay 4 Contact b Relay 5 Contact b Relay 6 Contact b Relay 7 Contact b Relay 8 Contact b Relay 9 Contact b Relay 10 Contact b 24V 24V 24V 24V 24V 24V 24V 24V Table 6 7 Measurement Interface Board Connector Pin Descriptions Connector Reference Pin Signal Label Designator Description AC IN J1 1 AC HOT 2 AC NEUT 3 AC Ground AC PUMP J2 1 AC HOT 2 AC NEUT Switched 3 AC Ground BENCH TEMP J3 1 Bench Temperature Thermistor 2 Ground PRES J4 1 Pressure Sensor Input 2 Ground 3 15V 6 20 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Troubleshooting Connector Pin Descrip
66. 8 that is tied to the selected streaming data item The selected item is shown by lt after it e Inthe Main Menu choose Instrument Controls gt Communication Settings gt Streaming Data Config gt Select Item gt Analog Inputs The RS 232 RS 485 Selection screen allows the user to choose between the RS 232 or RS 485 specification for serial communication Note Disconnect the serial cable before changing RS 232 and RS 485 selection to prevent damage to the connected equipment A e Inthe Main Menu choose Instrument Controls gt Communication Settings gt RS 232 RS 485 Selection 3 52 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu TCP IP Settings The TCP IP Settings menu is used for defining TCP IP settings Note The instrument power must be cycled after this parameter has been changed for the change to take effect a e Inthe Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings pE EN En REP ED Pi Tit m La motel Use DHCP The Use DHCP screen is used to specify whether to use DHCP or not e Inthe Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Use DCHP IP Address The IP Address screen is used to edit the IP address The IP address can only be changed when DHCP is on For more information on DHCP see Use DHCP above e Inthe Main Menu choose Instrument Control
67. CH Crd Voltages The Voltages menu displays the current diagnostic voltage readings This screen enables the power supply to be quickly read for low or fluctuating voltages without having to use a voltage meter The 1 O board is only displayed if the I O expansion board option is installed e Inthe Main Menu choose Diagnostics gt Voltages WH puabri Pin Paid Moro Mme pre OPER p Motherboard Voltages The Motherboard screen read only is used to display the current voltage readings on the motherboard e Inthe Main Menu choose Diagnostics gt Voltages gt Motherboard Voltages Thermo Fisher Scientific Model 146 Instruction Manual 3 71 Operation Diagnostics Menu rusas E E Mihei i Interface Board Voltages The Interface Board screen read only is used to display the current voltage readings on the interface board e In the Main Menu choose Diagnostics gt Voltages gt Interface Board Voltages nico Pirro Py Tce OH falta AA ii Lb A A 1 0 Board Voltages The I O Board screen read only is used to display the current voltage readings on the I O expansion board This menu is only displayed if the I O expansion board option is installed e Inthe Main Menu choose Diagnostics gt Voltages gt I O Board Voltages Cie Hr Ehe EEE 3 72 Model 146 Instruction Manual Thermo Fisher Scientific Operation Diagnostics Menu Flows The Flows screen read only displays the cur
68. Ci a aa 5 1 Monitoring Detector Frequency and NOiSC oooooonccocnconnonconnonannnnnanancnno 5 1 Caplio ia 5 1 Model 146 Instruction Manual vii Contents viii Chapter 6 Chapter 7 Model 146 Instruction Manual Ozonator Capillary Replacement Ozonator option seese 5 1 Pressure Regulator Capillary Replacement oooconnncnnnnconnnnncncenncnnncnno 5 1 Pump Capillary Replacement Photometer option oooocncononnnccnnnnnnono 5 1 Pump Rebasa iess 5 1 PrOUbleSHO OUN Gs ri 6 1 Safety Precautions aos 6 1 Troubleshooting Guide dildo pala 6 1 Board Level Connection Diagrams 0 ia 6 1 Connector Pin Descriptions ui iii 6 1 Seryice LOCAS o alias 6 1 SOMICI a ad AR ad AS air diodes 7 1 Safety Precautions as oTa E sas an E AO RAR ERRE 7 1 Firmware Update e a N et 7 1 Replacement Paris Listar tii 7 1 Cable Dita A RERA 7 1 External Device Connection Components occccococccncnnoncnnnonannnnnnoncnncnnos 7 1 Removing the Measurement Bench and Lowering the Partition Panel 7 1 us e nO 7 1 Accessing the Service Mode isa 7 1 Leak Test and Pump Checkout with Photometer Option 4 7 1 Leak Test and Pump Checkout without Photometer Option 7 1 Mass Flow Controller Replacement vinoiniiian iria 7 1 Pump Replacement Photometer Opos 7 1 Fan Replacement Rear Panel usas ia cie 7 1 Ran Replacement Mass Flow ControlletS ooononcicninnnononnnonnnccnrnnnons 7 1 Analog Output Testi ai 7 1 Analog Output Calibration ini 7 1 Analog
69. Connect the 491 PS OUTPUT to the 146 EXTERNAL SAMPLE port 2 Set flow of the 492 PS to the desired level 3 Determine the ozone concentration at different ozone level settings on the 1467 4 Determine a calibration curve by plotting the results of Step 3 5 Change to the 1467 Photometer Calibration Span 6 For highest confidence reconnect the 497 PS to the 1467 and recheck the ozone calibrations 4 2 Model 146 Instruction Manual Thermo Fisher Scientific Calibration Ozonator Control Check EXHAUST Figure 4 1 Photometer Calibration Flow Schematic Ozonator Control Use the following procedure to perform an ozonator control check Refer to Check the ozonator control check flow schematic Figure 4 2 1 Connect the 146i OUTPUT port to the 497 PS INPUT port 2 Set ozone concentration at different levels 3 Compare the ozone concentration readings on the 1467 and 491 PS Thermo Fisher Scientific Model 146 Instruction Manual 4 3 Calibration Ozonator Control Check Figure 4 2 Ozonator Control Check Flow Schematic 4 4 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Chapter5 Preventive Maintenance This chapter describes the periodic maintenance procedures that should be pP P P performed on the instrument to ensure proper operation Since usage and environmental conditions vary greatly you should inspect the components frequently until an appropriate maintenance schedule
70. ER NC ZERO AIR e e gt PRESSURE R RE CAPILLARY REGULATOR NC HOLDER NO yc Ne y REAR PANEL NC GAS Ce GAS MASS FLOW CONTROLLER nc GAS Be gt NC GAS A e CAPILLARY HOLDER e ZERO AIR MASS FLOW CONTROLLER Ne ZERO AIR ee gt PRESSURE CAP REGULATOR NC FOLDER No c LS BACK PRESSURE REGULATOR PERMEATION TUBE OVEN d MIXING no CHAMBER REACTION CHAMBER NC PERMEATION TUBE REAR PANEL OUTPUT MANIFOLD pm TO ANALYZER DUMP exnaust Figure 1 4 Standard Gas Dilution with Ozonator and Permeation Tube Oven Options Thermo Fisher Scientific Model 146 Instruction Manual 1 5 Introduction Principle of Operation 1 6 Model 146 Instruction Manual EXTERNAL O SAMPLE IN EXTERNAL 0 SAMPLE IN l l l l l l l l l t l l CAPILLARY HOLDER MPERATURE AND E COMPENSATION PRESSURE RELEASE CAPILLARY hii EXT INT PUMP A A x HOLDER UM NA c PHOTOMETER DUMP TO ANALYZER EXHAUST g OZONE l PAN CAPILLARY Os GENERATOR 0 HOLDER l D lt T __ gt l NC a MASS FLOW o Ne MASS E REACTION MIXING BACK PRESSURE REAR REGULATOR PANEL MASS FLOW CONTROLLER NC CAPILLARY HOLDER PRESSURE RELEASE capas VACUUM EXT INT Ne DUMP PERMEATION CAPILLARY TUBE OVEN HOLDER O ANALYZER PERM 2 E
71. Flow Thermo Fisher Scientific Operation Gas Setup Menu Manual Flow The Manual Flow screen displays the zero air and gas flows and lets you manually set the zero air and gas flows e Inthe Main Menu choose Gas Setup gt Gas A F gt Manual Zero Air The Manual Zero Air Flow screen sets the manual zero air flow value for the previously selected gas The screen identifies the minimum and maximum acceptable zero air flow rates and lets the user select a flow rate between these two settings If the selected zero air flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Gas Setup gt Gas A F gt Manual gt Zero Air WH mnie Thermo Fisher Scientific Model 146 Instruction Manual 3 25 Operation Ozonator Setup Menu Gas Ozonator Setup Menu Zero Air 3 26 Model 146 Instruction Manual The Manual Gas Flow screen displays the minimum and maximum gas flows that are allowed for the previously selected gas Minimum and maximum values are calculated based upon the tank concentration and the mass flow controller The desired concentration can be set between the two displayed limits If the concentration entered is outside of the minimum or maximum values the message TOO LOW or TOO HIGH is displayed e Inthe Main Menu choose Gas Setup gt Gas A F gt Manual gt Gas The Model 146 can be equipped with an optiona
72. Input Calibration alcaide 7 1 Calibrating the Input Channels to Zero Volts nesses 7 2 Calibrating the Input Channels to Full Scale oonooincnnnnnnnonionom 7 2 I O Expansion Board Replacement Optional oseese 7 2 Digital Output Board Replacement ias 7 2 Motherboard Replacement to ll add 7 2 Measurement Interface Board Replacement oooonconnnonicnonnonnnnonnannncnnon 7 2 Front Panel Board Replacement da 7 2 LCD Module Replacements indi ais 7 2 Photometer Lamp Replacement Photometer option 00 0 sees 7 2 Photometer Lamp Voltage Adjustment Photometer option 7 2 Photometer Lamp Heater Replacementiunisnini nina inialiars 7 2 Photometer Replacement Photometer Option essees 7 2 Photometer Board Replacement Photometer option oooocconnccnnnnnnos 7 2 Detector Replacement Photometer option 7 2 Photometer Thermistor Replacement luto nie aci 7 2 Thermo Fisher Scientific Thermo Fisher Scientific Contents Photometer Temperature Calibration Photometer option 7 2 Solenoid Replacement dat 7 2 Pressure Transducer Replacement Photometer option 0 0 7 2 Pressure Transducer Calibration Photometer option 7 2 Ozonator Lamp Replacement Ozonator option coonconococnconinnnccnnnnanono 7 2 Ozonator Lamp Heater Replacement Ozonator option sesse 7 2 Ozonator Replacement Ozonator Option cccoonccicconononnnnncnnnanoncnnnacnnono 7 2 Ozonator Board Replacement Ozonator OPti0M ococoncn
73. Instruction Manual i Preface Safety Review the following safety information carefully before using the analyzer This manual provides specific information on how to operate the analyzer Safety and Equipment Damage Alerts Model 1467 Instruction Manual however if the analyzer is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired This manual contains important information to alert you to potential safety hazards and risks of equipment damage Refer to the following types of alerts you may see in this manual Safety and Equipment Damage Alert Descriptions Alert i DANGER WARNING CAUTION Equipment Damage gt gt Description A hazard is present that will result in death or serious personal injury if the warning is ignored A hazard is present or an unsafe practice can result in serious personal injury if the warning is ignored The hazard or unsafe practice could result in minor to moderate personal injury if the warning is ignored The hazard or unsafe practice could result in property damage if the warning is ignored a Safety and Equipment Damage Alerts in this Manual Alert j WARNING Description This instrument contains 120 volts AC or 220 volts AC and 800 volts DC Normal precautions should be used when working on the inside of the instrument with the power connected If the equipment is operated in a manner not specified
74. MODBUS Application Protocol Specification V1 1a MODBUS IDA June 4 2004 The following are the communication parameters that are used to configure the serial port of the iSeries to support MODBUS RTU protocol Number of Data bits 8 Number of Stop bits 1 Parity None Date rate from 1200 115200 Baud 9600 is default Model 146 Instruction Manual C 1 MODBUS Protocol TCP Communication Parameters TCP Communication Parameters Application Data Unit Definition Slave Address MBAP Header C 2 Model 146 Instruction Manual iSeries Instruments support the MODBUS TCP protocol The register definition is the same as for the serial interface TCP connection port for MODBUS 502 Here are the MODBUS ADU Application Data Unit formats over serial and TCP IP Serial Slave Address Function Code Data Error Check TCP IP MBAP Header Function Code Data The MODBUS save address is a single byte in length This is the same as the instrument ID used for C Link commands and can be between 1 and 127 decimal i e 0x01 hex to 0x7F hex This address is only used for MODBUS RTU over serial connections Note Device ID 0 used for broadcast MODBUS commands is not supported Device IDs 128 through 247 i e 0x80 hex to OxF7 hex are not supported because of limitations imposed by C Link A In MODBUS over TCP IP a MODBUS Application Protocol Header MBAP is used to identify the message
75. Menu e Inthe Main Menu choose Alarms gt Pressure Thermo Fisher Scientific Operation Alarms Menu Pri ii Min and Max Pressure The Minimum Pressure alarm limit screen is used to change the minimum Limits temperature alarm limit The minimum and maximum pressure screens function the same way e In the Main Menu choose Alarms gt Pressure gt Min or Max Photometer Bench The Photometer Bench Temperature screen displays the current bench Temperature temperature and sets the minimum and maximum alarm limits Acceptable alarm limits range from 10 to 50 C If the bench temperature reading goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu e Inthe Main Menu choose Alarms gt Phot Bench Temp Thermo Fisher Scientific Model 146 Instruction Manual 3 81 Operation Alarms Menu Min and Max Photometer Bench Temperature Limits Photometer Lamp Temperature Min and Max Photometer Lamp Temperature Limits 3 82 Model 146 Instruction Manual The Minimum Photometer Bench Temperature alarm limit screen is used to change the minimum bench temperature alarm limit The minimum and maximum bench temperature screens function the same way e Inthe Main Menu choose Alarms gt Phot Bench Temp gt Min or Max Cie Hr nice hi CH gyi PL i The Photometer Lamp Temperature screen displays the current lamp temperature and
76. Model 1467 Instruction Manual Dynamic Gas Calibrator Part Number 102482 00 16Apr2013 CE Thermo O 2007 Thermo Fisher Scientific Inc All rights reserved Specifications terms and pricing are subject to change Not all products are available in all countries Please consult your local sales representative for details Thermo Fisher Scientific Air Quality Instruments 27 Forge Parkway Franklin MA 02038 1 508 520 0430 www thermo com aqi Thermo Fisher Scientific WEEE Compliance This product is required to comply with the European Union s Waste Electrical amp Electronic Equipment WEEE Directive 2002 96 EC It is marked with the following symbol Thermo Fisher Scientific has contracted with one or more recycling disposal companies in each EU Member State and this product should be disposed of or recycled through them Further information on Thermo Fisher Scientific s compliance with these Directives the recyclers in your country and information on Thermo Fisher Scientific products which may assist the detection of substances subject to the RoHS Directive are available at www thermo com WEEERoHS WEEE Compliance Thermo Fisher Scientific About This Manual This manual provides information about operating maintaining and servicing the analyzer It also contains important alerts to ensure safe operation and prevent equipment damage The manual is organized into the following chapters and appendices to provide
77. NTS c cccccccscssessesessesesesesseseseseseseeeseeeeees 8 3 Figure 9 1 Cal Oven Therm Resistor Sereen s 9 2 Figure 9 2 Cal Gas Therm Bath SCreen oooooociciocincocincnncncnnoncnncnnnnrncnncnonocnnonos 9 2 Figure 9 3 Cal Gas Therm Resistor SCTeeN ooooooocicciciocinicccnnocononncnnnnononinocnnnnos 9 2 Figure 9 4 Rack Mount Option Assembly o oooonicooninocccnnncininncnnncinonocnnosoo 9 2 Figure 9 5 Bench MOUNTING ocooooconinoninonnnanocononrnononorornnnonocnrnnnonororannonornranonos 9 2 Figure 9 6 EIA Rack Mounting a aa des rada Do de 9 2 Figure 9 7 Retrofit Rack MOUNTING o ococioconononnnncononnnononncornnnonorornrnonororonnonos 9 2 Figure BLAS oes tecs ice co searstece ck cop cv septate de ctuatateys wn ayticracett eh ti tet ceases B 2 Thermo Fisher Scientific Thermo Fisher Scientific Tables Tables Table 1 1 Model 1467 Specifications oooooicoonicioniccncinnonincnncncnnononnocnrnncnnnnos 1 1 Table 2 1 O Terminal Board Pin DescriptiONS ociconcinionincncincnninininncnnnoso 2 2 Table 2 2 D O Terminal Board Pin DescriptiONS ocooniccnnincnninccnnccnnos 2 2 Table 2 3 25 Pin Terminal Board Pin DescriptiONS ococinococccnincncnnnncncninnnooos 2 2 Table 3 1 Front Panel Pushbuttons 0 0 0 cece eee ecesesseseeseseeseseeseseeseeeseeneees 3 4 Table 3 2 Default Analog OUtpUtS oo ccc cece cccesceceseeceseeteseeteseeteseeteseeeees 3 4 Table 3 3 Ana
78. Send 03 coef Receive 03 coef 1 000 set 03 coef value This command sets the O coefficients to user defined values to value where value is a floating point representation of the coefficient The example below sets the O coefficient to 1 005 This command is only available if the photometer option is installed Send set 03 coef 1 005 Receive set 03 coef 1 005 ok 03 bkg This command reports the current Oz background The example below reports that the Oz background is 1 005 ppm This command is only available if the photometer option is installed Send 03 bkg Receive 03 bkg 1 005 ppb set 03 bkg value This command is used to set O background to user defined values to value where value is a floating point representation of the background in ppm The example below sets the O3 background to 15 ppb This command is only available if the photometer option is installed Send set 03 bkg 15 Receive set 03 bkg 15 ok Thermo Fisher Scientific Keys Display Thermo Fisher Scientific C Link Protocol Commands Keys Display resp coef This command reports the current ozonator response coefficient when under photometer control The example below reports that the ozonator response coefficient is 1 000 This command is only available if the photometer option is installed Send resp coef Receive resp coef 1 000 set resp coef value This command sets the ozonator response coefficient to user defined values to value where va
79. Temperature screen is used only prior to calibrating Pressure Temperature the flow controllers These values are used to convert from volumetric flow to standard flow in the flow controller calibration screens The flow controllers are calibrated prior to shipment However if you need to re calibrate the flow controllers you can use this screen to enter the pressure and temperature that existed during re calibration Note When calibrating the flow controller the pressure and temperature of the area immediately around the external flow sensor must be entered for an accurate calibration The pressure is in millimetres mercury and the temperature is degrees Celsius A This adjustment should only be performed by an instrument service technician A e Inthe Main Menu choose Service gt Calibration Pres Temp 3 86 Model 146 Instruction Manual Thermo Fisher Scientific Operation Service Menu CHORDIE PAE TATU ES AA ie i Zero Flow The Zero Flow Calibration menu is used to calibrate the zero air flow Calibration controller at 5 20 35 50 65 80 and 95 percent of full scale Using a NIST traceable flow meter the accuracy of the controller can be calibrated to about 1 as opposed to the 2 accuracy rating when shipped from the manufacturer The zero flow calibration is visible only when the instrument is in service mode For more information of the service mode see Service Mode earlier in the chapter Note This adjustm
80. The Set Analog Outputs screen appears Press to set the output to zero p The Output Set To line displays Zero Check that the meter is displaying a zero value If the meter reading differs by more than one percent the analog outputs should be adjusted Refer to the Analog Output Calibration procedure that follows Press t to set the output to full scale The Output Set To line displays Full Scale Model 146 Instruction Manual 7 19 Servicing Analog Output Testing 7 20 Model 146 Instruction Manual 7 Check that the meter is displaying the full scale value If the meter reading differs by more than one percent the analog outputs should be adjusted Refer to the Analog Output Calibration procedure that follows 8 Press Ce to reset the analog outputs to normal O oa Current Outputs Analog lt q Voltage Inputs 000000000000 0000000000000 O Analog Voltage Dutputs 000000000000000000 10000000000000000000 of Figure 7 7 Rear Panel Analog Input and Output Pins Table 7 4 Analog Output Channels and Rear Panel Pin Connections Voltage Pin Current Pin Channel Channel 1 14 1 15 2 33 2 17 3 15 3 19 4 34 4 21 5 17 5 23 6 36 6 25 Thermo Fisher Scientific Analog Output Thermo Fisher Scientific Calibration Servicing Analog Output Calibration Table 7 4 Analog Output Channels and Rear Panel Pin Connections continued Voltage Pin Current Pin Ch
81. VING is displayed for 3 seconds then removed e Inthe Main Menu choose Flow Modes gt O3 gt Transfer Standard gt Level 1 5 3 16 Model 146 Instruction Manual Thermo Fisher Scientific Operation Flow Modes CONTO CN CUA ee PE EE E Manual 0 Adjust The Manual screen is used to set the ozonator lamp output manually e In the Main Menu choose Flow Modes gt O3 gt Transfer Standard gt Manual pE EN En Pino GPT Photometer Control The ozonator output level can be set using the Ozonator GPT Select screen Alternately if Photometer Control is selected then all ozonator levels are shown in parts per million ppm instead of percent The MANUAL screen is used to set the ozonator output to the manual level in ppm or e In the Main Menu choose Flow Modes gt O3 gt GPT Photometer Control gt Level 1 5 rusas Hna Pi Taal DH Cil Moca 5 a Y eat OE TET Ret a i ae Pi i m Le bt A raii i tt he PPLE A Pat MU Thermo Fisher Scientific Model 146 Instruction Manual 3 17 Operation Flow Modes gengp MOE Ta op ome Gas Selection The Gas Selection screen is used to select a gas to mix with ozone Gas D E and F are available if the six gas option is installed If NONE is selected the screen reports the new flow mode The user must press to accept and activate these settings The message SAVING is displayed for 3 seconds then removed e Inthe Main Menu choose Flow Modes gt O3
82. XHAUST CAPILLARY HOLDER AR 1 MASS FLOW NC MRD REACTION MIXING l CONTROLLER 1i CHAMBER CHAMBER l Ww REAR 8 Ne D Tr Figure 1 6 Standard Gas Dilution System with Ozonator Permeation Tube Oven and Photometer Options Thermo Fisher Scientific Introduction Specifications Specifications Table 1 1 lists the specifications for the Model 1467 Thermo Fisher Scientific Table 1 1 Model 1467 Specifications Gas Dilution Accuracy of each Mass Flow Measurement 1 of full scale 20 to 100 full scale Linearity of Mass Flow Measurement 0 5 of full scale Repeatability of Mass Flow Measurement 2 of reading or 1 of full scale whichever is less 20 to 100 full scale Available Ranges Zero Air 5 10 15 20 sIm Span Gas 25 50 100 200 500 sccm Dilution Ratio Depends upon choice of Mass Flow Controllers Response time as measured at output of manifold to 99 of final value Less than one minute for low flows Faster for high flow for gas phase titration mode response time increases to 4 minutes Meets or exceeds U S EPA Dynamic Parameters specification requirements for gas phase titration Operating temperature 0 45 C Power Requirements 100 VAC 50 60 Hz 115 VAC 50 60 Hz 220 240 VAC 50 60 Hz 275 watts Physical dimensions 16 75 W X 8 62 H X 23 D Weight Approximately
83. ables Replace the measurement bench by following the previous steps in reverse Use the following procedure to replace the fuse Equipment Required l Replacement fuses 250VAC 3 Amp SlowBlow for 100VAC and 110VAC models 250VAC 1 60 Amp SlowBlow for 220 240VAC models Turn instrument OFF and unplug the power cord Thermo Fisher Scientific Accessing the Service Mode Leak Test and Pump Checkout with Photometer Option Thermo Fisher Scientific A A Servicing Accessing the Service Mode 2 Remove fuse drawer located on the AC power connector 3 If either fuse is blown replace both fuses 4 Insert fuse drawer and reconnect power cord If the Service is not displayed on the Main Menu use the following procedure to display it 1 At the Main Menu press to scroll to Instrument Controls gt press gt 4 to scroll to Service Mode gt and press The Service Mode screen appears 2 Press to toggle the Service Mode to ON 3 Press Cm gt Cm to return to the Main Menu 4 Return to the procedure Use the following procedure to check for leaks in the system and to leak test the pump This procedure is for Model 146 instruments that have the photometer option For Model 1467 instruments that do not have the photometer option refer to Leak Test and Pump Checkout without Photometer Option WARNING The service procedures in this manual are restricted to qualified service rep
84. ags CO00000 conc 0 000 tgflo 30 000 agflo 0 000 tzflo 1000 000 azflo 0 000 ttflo 1030 000 atflo 0 000 pres 0 000 Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging 13 00 12 15 05 flags CO00000 conc 0 000 tgflo 30 000 agflo 0 000 tzflo 1000 000 azflo 0 000 ttflo 1030 000 atflo 0 000 pres 0 000 14 00 12 15 05 flags CO00000 conc 0 000 tgflo 30 000 agflo 0 000 tzflo 1000 000 azflo 0 000 ttflo 1030 000 atflo 0 000 pres 0 000 erec format lrec format srec format These commands report the output format for long and short records and dynamic data in various formats such as ASCII without text ASCII with text or binary The example below shows the output format for long records is ASCII with text according to Table B 4 Send lrec format Receive lrec format 01 set erec format format set lrec format format set srec format format These commands set the output format for long and short records and dynamic data according to Table B 4 The example below sets the long record output format to ASCII with text Send set lrec format 1 Receive set lrec format 1 ok Table B 4 Record Output Formats Format Output Format 0 ASCII no text 1 ASCII with text 2 Binary data erec layout lrec layout srec layout These commands report the layout string indicating the data formats for data that is sent out in response to the erec lrec srec and related commands For det
85. ails on how to interpret the strings see Record Layout Definition later in this appendix Model 146 Instruction Manual B 19 C Link Protocol Commands Datalogging B 20 Model 146 Instruction Manual Send lrec layout Receive lrec layout s s lx f f f f t D L ffff flags conc tgflo agflo tzflo azflo ttflo atflo pres lrec mem size srec mem size These commands report the long and short records that can be stored with the current settings and the number of blocks reserved for long and short records To calculate the number of short records per block add 2 to the number of records and then divide by the number of blocks The example below shows that 10 blocks were reserved for long records and the maximum number of long records that can be stored in memory is 2038 Send lrec mem size Receive lrec mem size 2038 recs 10 blocks lrec per srec per These commands report the long and short records logging period The example below shows that the short record logging period is 5 minutes Send srec per Receive srec per 5 min set srec per value set srec per value value 1 5 15 30 60 These commands set the long and short records logging period to value in minutes The example below sets the long record logging period to 15 minutes Send set lrec per 15 Receive set lrec per 15 ok no of lrec no of srec These commands report the number of long and short records stored in the long and short records me
86. ajey uoneouad S 1 19497 los seg uonniig seg PIOMSSEd 18S 7499 S 101 U0D MO du juaiquy U0ISI9A wesboig uonesuadwog dw aj9 9 weiboldg obuey NA uoneaued Aly 0492 OWEN SBD qpuels dnjas s oa uo piomssed 20198 suuely sonsoubeig disun nedi wesbold Piatt Enter peace 43 0 9 a v s po moja uonesedo sey nua uen useJ9s uny u9aJ9s 3591 HES uaal0S dn 193MOd Figure 3 3 Flowchart of Menu Driven Software 3 5 Model 146 Instruction Manual Thermo Fisher Scientific Operation Software Overview Power Up Screen Run Screen 3 6 Model 146 Instruction Manual The Power Up screen is displayed on power up of the Model 1462 The Self Test is displayed while the internal components are warming up and diagnostic checks are performed Thermo SC A oe Ace iSeries The Run screen is the normal operating screen when the menu is not displayed The status bar displays the instrument s current flow selections selected gas span level if ozonator and permeation oven are both off permeation oven and ozonator levels time and alarm status There are 5 values that may be displayed on the run screen that include the following TFLOW total flow in sccm as measured by the zero and gas flow controllers GAS X dilution concentration of gas from selected tank in ppm name is user definable 03 concentration of ozone measured by photometer in ppm PERM concentration from permeation tube in ppm GPT drive level
87. al Thermo Fisher Scientific Servicing Ozonator Board Replacement Ozonator option 7 Re install the instrument cover Capillary Figure 7 13 Replacing the Ozonator Ozonator Board Us the following procedure to replace the ozonator board Figure 7 10 Replacement Ozonator option Esuipmen Required Ozonator board Philips screwdriver amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the connection cables from the board 3 Remove the two screws from the top of the board 4 Pop off the board from the bottom mounting studs and remove the board 5 To install the new ozonator board follow previous steps in reverse Thermo Fisher Scientific Model 146 Instruction Manual 7 45 Servicing Permeation Oven Replacement 7 46 Permeation Oven Replacement A Model 146 Instruction Manual Use the following procedure to replace the permeation oven Equipment Required Permeation oven Flatblade screwdriver Wrenches 1 4 inch and 7 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and rem
88. alibration events can be programmed Next Event is read only and shows the status of the next event number and time there is no setting screen Each event can be assigned gas span ozonator and permeation levels photometer mode and event set time This function is only available when the instrument is NOT in service mode If the service mode is active when the next cycle time occurs the program will not be triggered Turn service mode OFF to run next program cycle time For more information on the service mode see Service Mode later in this chapter Note When events are running none of the settings may be changed via the front panel interface except for the program cycle enable disable which may be toggled to cancel the current program A e Inthe Main Menu choose Program Thermo Fisher Scientific Operation Program Menu Hp bebe i ba PR PCRR AGID AG E Program Cycle The Program Cycle screen is used turn the program mode on or off When enabled the status bar shows the current status of programmed events either ACTIVE or INACTIVE Inactive means enabled but not currently running and active means running current event When disabled the program will not be run at its assigned interval This screen is also used to cancel a running program e Inthe Main Menu choose gt Program gt Program Cycle Period Hours The Period Hours screen is used select how many hours starting between cycles of the program Vali
89. allbration endoricaln tte 9 1 Determining Permeation Rate by Weight Loss ssseeeseseeeeseeeeeeeses 9 1 Determining Release Rate by Transfer Standard ccoo 9 1 SES PAIPA ai 9 1 Span Gas Mass Flow Controller tao 9 1 Zero Gas Mass Flow Controller each er eee 9 1 I O Expansion Board Asse mi Diy it ti 9 1 25 Pin Terminal Board Assembly inciso nica co das 9 1 Terminal Block and Cable Kits coil pisdinos 9 1 A E 9 1 Mounting Opina ae 9 1 Thermo Fisher Scientific Thermo Fisher Scientific Contents A A A AA A 1 C Link Protocol Commands omocmosmosmsmsismsisisisisisiss B 1 Instrument Identification Numb raniia alicia B 1 Commands A Saale ance a a RER B 1 Commands Uist ox nite iat nao eo tet B 1 MESES A settee Ss Gd Geass stones B 1 AJISTE EREE AT E E A EE haan B 1 A B 1 Dato iii A ween B 1 A hare eee eae pear uli Racecar B 1 Keys ORpla yer ienien ana be E ee oor ten ean die B 1 Measurement Configuration ii ust otal ice is Ned B 1 Hardware Contottatio sacada B 1 Communications Configuration meros B 1 OE AO E A NS B 1 Record Layout DA ii B 1 Format Specifier for ASCH Responses ida B 1 Format Specifier for Binary Responses s ssssssssssssseseeeeeerererersrerersres B 1 Format Specifier for Front Panel Layout ssssssseseeeeieeererersesrerersees B 1 MODBUS Protocol ion C 1 Serial Communication Parameters ai ios C 1 TCP Communication Parametros iii C 1 Application Data Unit Definition s sssssesesssesestsssssssrerrrrreresr
90. amp Thermistor 1 Photometer Lamp Heater Control Ground 24V Heater Current Sense D a A WO N Photometer Lamp Thermistor 2 INTF J3 Photometer Lamp Intensity Control Voltage Input 2 Photometer Lamp Intensity Control Voltage Output Photometer Heater Thermistor 24V Ground Photometer Lamp On off Control Ground mo N oon A Ww Ground Se rvice Locations For additional assistance service is available from exclusive distributors worldwide Contact one of the phone numbers below for product support and technical information or visit us on the web at www thermo com aqi 1 866 282 0430 Toll Free 1 508 520 0430 International Thermo Fisher Scientific Model 146 Instruction Manual 6 25 Troubleshooting Service Locations 6 26 Model 146 Instruction Manual Thermo Fisher Scientific Chapter 7 Servicing This chapter explains how to replace the Model 146i subassemblies It assumes that a subassembly has been identified as defective and needs to be replaced For fault location information refer to the Preventive Maintenance chapter and the Troubleshooting chapter in this manual The service mode in the Operation chapter also includes parameters and functions that are useful when making adjustments or diagnosing problems For additional service assistance see Service Locations on page 7 54 at the end of this chapter This chapter includes the following parts informat
91. and press to calibrate the input voltage on the selected channel to zero volts The screen displays 0 00 V as the voltage setting 5 Press y gt c3 to return to the Analog Input Cal screen and repeat Steps 2 through 4 to calibrate other input channels to zero as necessary 6 Continue with the Calibrating the Input Channels to Full Scale procedure that follows Use the following procedure to calibrate the input channels to full scale by applying a known voltage to the channels Equipment Required DC voltage source greater than 0 volts and less than 10 volts 1 Connect the known DC voltage source to the input channel 1 8 to be calibrated Figure 7 7 shows the analog input pins and Table 7 5 identifies the associated channels 2 From the Main Menu press 4 to scroll to Service press gt 4 to scroll to Analog Input Calibration and press Ce The Analog Input Cal screen displays input channels 1 8 Note If Service is not displayed refer to Accessing the Service Mode on page 7 11 then return to the beginning of this step Model 146 Instruction Manual 7 23 Servicing 1 0 Expansion Board Replacement Optional 3 At the Analog Input Cal screen press 4 to scroll to the channel selected in Step 1 and press 4 Press 4 to scroll to Calibrate Full Scale and press The screen displays the current input voltage for the selected channel 5 Use 4 and to en
92. annel Channel Ground 16 18 19 35 37 Current Output Return 16 18 20 22 24 Table 7 5 Analog Input Channels and Rear Panel Pin Connections Input Channel Pin 1 1 o N on A WW N VO aN 0 MM W NN o Ground 4 8 11 Use the following procedure to calibrate the analog outputs if a meter reading in the Analog Output Testing procedure differed by more than one percent or after replacing the optional I O expansion board Equipment Required Multimeter 1 Connect a meter to the channel to be adjusted and set to voltage or current as appropriate Figure 7 7 shows the analog output pins and Table 7 4 identifies the associated channels 2 From the Main Menu press 4 to scroll to Service press gt to scroll to Analog Out Calibration and press Ce The Analog Output Cal screen appears Note If Service is not displayed refer to Accessing the Service Mode on page 7 11 then return to the beginning of this step Model 146 Instruction Manual 7 21 Servicing Analog Input Calibration Analog Input Calibration Calibrating the Input Channels to Zero Volts Model 146 Instruction Manual 3 At the Analog Output Cal screen press to scroll to the voltage channel or current channel corresponding to the rear panel terminal pins where the meter is connected then press 4 4 With the cursor at Calibrate Zero press The Analog Output Cal line displays Zero Note When calibrati
93. anslation button which creates a Model 146 Instruction Manual B 51 C Link Protocol Commands Record Layout Definition B 52 Model 1467 Instruction Manual selection list of twelve Code nn options The number of the user selection is used to create the output command Mode 6 12 13xflocal remote service service 0 1 Tset mode s n This is a line which has a title of Mode and value taken from the sixth field of the data response There is a bitfield extraction of bits 12 through 13 from the source the value type is not important here because the value is being translated to an output string Once the bits have been extracted they are shifted down to the bit zero position Thus the possible values of this example will be 0 through 3 The translation list shows the words which correspond to each input value the zeroth value appearing first 0 gt local 1 gt remote etc The selection list shows that only the first two values in this case are to be shown to the user when the button is pressed The T button indicates full translation input code to string and user selection number to output string xC This is a line that starts a new column the xC or L Comp 6 11x off on Tset temp comp s n This shows that the bitfield end the second part of a bitfield specification is optional The bitfield will be one bit long starting in this case at the eleventh bit Background 7f 8Bd ddd set 03 bkg s n This
94. apillary Mixing Chamber Manifold Mass Flow Controllers 8 6 Model 146 Instruction Manual Controls gas flow though the ozonator Controls gas flow though the permeation oven Directs gas flow to the reaction chamber or to exhaust Directs gas flow to the reaction chamber or to exhaust Shuts off zero air when the system is in standby to minimize zero air usage Used to select either the external ozone sample or the internal ozone sample for measurement Control flow through independent gas inlets depending on the mode being controlled Reduces gas flow to the photometer Reduces gas flow to the ozonator Used in conjunction with the pressure regulator to control flow through the ozonator Controls flow pressure on the photometer Mixes the gas dilutions provided the mass flow controllers Distributes gas from various sources to an analyzer Governs high gas flow to control diluting zero air and governs low gas flow for controlling gas to be diluted Thermo Fisher Scientific Back Pressure Regulator Reaction Chamber Pressure Regulator and Gauge Software Instrument Control Thermo Fisher Scientific System Description Software Controls back pressure flow through the system The reaction chamber is where the sample reacts with ozone and produces excited NO that gives off a photon of energy when it decays The reaction chamber is heated and controlled to approximately 50 C to ensure optimum
95. ard level connection diagrams connector pin descriptions and testing procedures presented in this chapter should be helpful in isolating and identifying problems For additional fault location information refer to the Preventive Maintenance chapter in this manual The service mode in the Operation chapter includes parameters and functions that are useful when making adjustments or diagnosing problems The service mode includes some of the same information found in the Diagnostic menu however readings are updated every second in the service mode compared with every 10 seconds in the Diagnostics menu The Technical Support Department at Thermo Fisher Scientific can also be consulted in the event of problems See Service Locations at the end of this chapter for contact information In any correspondence with the factory please note both the serial number and program number of the instrument This chapter provides the following troubleshooting and service support information e Safety Precautions on page 6 1 e Troubleshooting Guides on page 6 2 e Board Level Connection Diagrams on page 6 8 e Connector Pin Descriptions on page 6 10 e Service Locations on page 6 25 Safety Precautions Read the safety precautions in the Preface and the Servicing chapter before performing any actions listed in this chapter Thermo Fisher Scientific Model 146 Instruction Manual 6 1 Troubleshootin
96. are not covered by the warranty provided in this warranty Buyer shall pay Seller therefor at Seller s then prevailing time and materials rates ANY INSTALLATION MAINTENANCE REPAIR SERVICE RELOCATION OR ALTERATION TO OR OF OR OTHER TAMPERING WITH THE PRODUCTS PERFORMED BY ANY PERSON OR ENTITY OTHER THAN SELLER WITHOUT SELLER S PRIOR WRITTEN APPROVAL OR ANY USE OF REPLACEMENT PARTS NOT SUPPLIED BY SELLER SHALL IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH RESPECT TO THE AFFECTED PRODUCTS THE OBLIGATIONS CREATED BY THIS WARRANTY STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A DEFECTIVE PRODUCT EXCEPT AS EXPRESSLY PROVIDED IN THIS WARRANTY STATEMENT SELLER DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED ORAL OR WRITTEN WITH RESPECT TO THE PRODUCTS INCLUDING WITHOUT LIMITATION ALL IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE SELLER DOES NOT WARRANT THAT THE PRODUCTS ARE ERROR FREE OR WILL ACCOMPLISH ANY PARTICULAR RESULT A 2 Model 146 Instruction Manual Thermo Fisher Scientific Instrument Identification Thermo Fisher Scientific Number Appendix B C Link Protocol Commands This appendix provides a description of the C Link protocol commands that can be used to remotely control a Model 146i analyzer using a host device such as a PC or a datalogger C Link protocol may be used over RS 232 RS 485 or Ethernet C Link functi
97. are proposed changes as opposed to implemented changes To change the selected record format and erase record log file data see Commit Content below e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Select Field gt Other Measurements Wi mihi Pin Paid Analog Inputs The Analog Inputs screen allows the user to select the output signal none or analog inputs 1 8 that is tied to the selected field item The selected item is shown by lt after it Note that at this point pressing indicates that these are proposed changes as opposed to implemented changes To change the selected record format and erase record log file data see Commit Content below e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Select Field gt Analog Inputs Thermo Fisher Scientific Model 146 Instruction Manual 3 45 Operation Instrument Controls Menu Commit Content The Commit Content screen is used to save the selected output signal that is tied to the selected field item If no changes have been made NO CHANGES TO RECORD LIST will appear For more information about selecting the analog output signal group choices see Select Content above e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Commit Content DH Gi MAMA Reset to Default Content The Reset to Default Content screen is
98. art is displayed while the instrument is warming up and performing self checks After the warm up period the Run screen is automatically displayed From the Run screen the Main Menu can be displayed by pressing 1 The Main Menu contains a list of submenus Each submenu contains related instrument controls and settings This chapter describes each submenu and screen in detail Refer to the appropriate sections for more information Thermo Fisher Scientific Operation Software Overview synejaq 19sf 9401S9y 1591 axid ejdsiq j90y asuodsay to 9949 Ayisuaquy Buimes due7 sBuines U9AO Wied 189 Indu Bojeuy smeis Pd O I oul j9ejuog 6yuoy jusuna suj sindjno Bojeuy 1sa1 s9 e15 plouajos snjeis pg uI s9Je s Aejoy 1e9 indino Bojeuy snes an sinduj euB1g w L aJeq 1e9 due youeg Aysuaqul oud se eyon ndul Bojeuy apo 291195 1e9 due juaiquiy dua duie7 104d sGuipesy Indul Bojeuy Se1JuOy ua a19S tenue yu wunasu youn le nss id dw youeg oyd a nss id Biyuog O I suang s 1 ueds dyounojould Piomssed SAQUIY 189 Mo y sey 91NSSIId selnjeradu sBuines wwog JU9AZ IXON uoneaqieo 019Z 0 piomssed abueyo 189 moj OJ9Z dwa USA Wed smoj4 s6unos BurbBojeyea aJ9 D MEN S10 984 UONEIqHeo S 1 19497 enue QUOD YUBL USAQ UONBaUSd yu wnysul 4907 dua 1 S94d 1e9 due due7 O sofeyoA uonesuedwog S id SINOH poned au Bulbersay
99. at the status when all the relays logic is set to normally open Thermo Fisher Scientific Record Layout Thermo Fisher Scientific Definition C Link Protocol Commands Record Layout Definition Send relay stat Receive relay stat open Note If individual relays have been assigned different logic then the response would be a 4 digit hexadecimal string with the least significant byte LSB being relay no 1 A For example Receive relay stat 0x0001 indicates relay no 1 is set to normally open logic all others are normally closed Receive relay stat 0x0005 indicates relay no 1 and 3 are set to be normally open logic all others are normally closed set relay open set relay open value set relay closed set relay closed value These commands set the relay logic to normally open or closed for relay number value where value is the relay between 1 and 16 The example below sets the relay no 1 logic to normally open Note If the command is sent without an appended relay number then all the relays are assigned the set logic of normally open closed 5 Send set relay open 1 Receive set relay open 1 ok The Erec Lrec Srec layouts contain the following e A format specifier for parsing ASCII responses e A format specifier for parsing binary responses In addition to these the Erec Layout contains e A format specifier for producing the front panel displays Values are read in using either the ASCII or binary forma
100. buffer data used for the display on the iSeries instrument It is 19200 bytes in size 2 bits per pixel 4 pixels per byte arranged as 320 by 240 characters The data is sent in RLE encoded form to save time in transmission It is sent as a type 5 binary c_link response with no checksum The RLE encoding consists of a 0 followed by an 8 bit count of consecutive OxFF bytes The following c code will expand the incoming data Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Keys Display void unpackDisplay void far tdib unsigned char far rlescreen int i j k unsigned char far sc4bpp sc2bpp screen ptr ptr screen unsigned char far malloc 19200 RLE decode the screen for i 0 i lt 19200 8282 ptr screen lt 19200 i ptr rlescreen i if rlescreen i 0 unsigned char rlecount unsigned char rlescreen i while rlecount ptr 0 rlecount else if rlescreen i Oxff unsigned char rlecount unsigned char rlescreen i while rlecount ptr Oxff rlecount To convert this data into a BMP for use with windows it needs to be turned into a 4BPP as that is the smallest windows can display Also note that BMP files are upside down relative to this data i e the top display line is the last line in the BMP sec screen This command is meant for backward compatibility on the C ser
101. by the manufacturer the protection provided by the equipment may be impaired The service procedures in this manual are restricted to qualified service personnel only lt The Model 146 is supplied with a three wire grounding cord Under no circumstances should this grounding system be defeated Thermo Fisher Scientific FCC Compliance Thermo Fisher Scientific Preface Safety and Equipment Damage Alerts in this Manual continued Alert Description CAUTION If the LCD panel breaks do not to let the liquid crystal A contact your skin or clothes If the liquid crystal contacts your skin or clothes wash it off immediately using soap and water a Equipment Damage Do not attempt to lift the analyzer by the cover or other A external fittings Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a This adjustment should only be performed by an instrument service technician Handle all printed circuit boards by the edges only Do not remove the panel or frame from the LCD module The LCD module polarizing plate is very fragile handle it carefully Do not wipe the LCD module polarizing plate with a dry cloth it may easily scratch the plate a Do not use Ketonics solvent or aromatic solvent to clean the LCD module use a soft cloth moistened with a naphtha cleaning solvent Do not
102. c format commands For details on how to decode the flag fields within these records see the flags command Send erec Receive erec 14 08 12 15 05 C000000 0 000 4 999 900 999 900 999 900 999 900 50 000 0 000 0 000 0 000 0 000 1 000 90 6 000 1 1 0 000 1 O 0 000 O O 0 000 lrec srec lrec xxxx yy SEEC XXXX yy lrec aa bb 00 pp qq yy srec aa bb 00 pp qq yy xxxx the number of past records yy the number of records to return 1 to 10 aa hours 01 to 24 bb minutes 01 to 59 oo month 01 to 12 pp day 01 to 31 qq year These commands output long or short records The output format is determined by the set lrec format and set srec format commands The logging time is determined by the set lrec per and set srec per commands In the following example there are 740 long records currently stored in memory When the command lrec 100 5 is sent the instrument counts back 100 records from the last record collected record 740 and then returns 5 records 640 641 642 643 and 644 For details on how to decode the flag fields within these records see the flags command Send lrec 5 Receive lrec 100 5 10 00 12 15 05 flags CO00000 conc 0 000 tgflo 30 000 agflo 0 000 tzflo 1000 000 azflo 0 000 ttflo 1030 000 atflo 0 000 pres 0 000 11 00 12 15 05 flags CO00000 conc 0 000 tgflo 30 000 agflo 0 000 tzflo 1000 000 azflo 0 000 ttflo 1030 000 atflo 0 000 pres 0 000 12 00 12 15 05 fl
103. carrier immediately and hold for inspection The carrier is responsible for any damage incurred during shipment Use the following procedure to unpack and inspect the instrument 1 Remove the instrument from the shipping container and set it on a table or bench that allows easy access to both the front and rear Model 146 Instruction Manual 2 1 Installation Unpacking and Inspection 2 Remove the cover to expose the internal components 3 Remove the packing material Figure 2 1 Remove Packing 2 pieces Remove Packing 2 pieces Units without Optional 1 0 Board Units with Optional 1 0 Board Figure 2 1 Remove the Packing Material 4 Ifthe photometer option is installed includes pump remove the three shipping screws Figure 2 2 Shipping Screws 3 Figure 2 2 Remove the Shipping Screws 2 2 Model 146 Instruction Manual Thermo Fisher Scientific Installation Setup Procedure 5 Check for possible damage during shipment 6 Check that all connectors and circuit boards are firmly attached 7 Re install the cover Setup Procedure Use the following procedure to setup the instrument 1 Connect a source of zero air to the inlet port labeled ZERO AIR The zero air source is capable of supplying the full scale flows of the zero air controller at a pressure between 10 and 40 psi 2 Connect the standard gas cylinders to the ports labeled A B and C 3 Install the permeation
104. ce the pump see Pump Replacement Photometer option on page 7 16 in the Servicing chapter Equipment Required Flatblade screwdriver Pump rebuild kit flapper valve and diaphragm amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Loosen the fittings and remove both lines going to the pump Thermo Fisher Scientific Model 146 Instruction Manual 5 9 Preventive Maintenance Pump Rebuilding 3 Remove the four screws from the top plate remove top plate flapper valve and the bottom plate Top Plate E Flapper Plate Bottom Plate A Diaphragm Figure 5 5 Rebuilding the Pump 4 Remove the screw the securing the diaphragm to piston and remove diaphragm 5 Assemble the pump by following the previous steps in reverse make sure the Teflon white side of the diaphragm is facing up and that the flapper valves cover the holes of the top and bottom plate 6 Test the pump for leaks Refer to Leak Test and Pump Checkout in the Servicing chapter 5 10 Model 146 Instruction Manual Thermo Fisher Scientific Chapter6 Troubleshooting This instrument has been designed to achieve a high level of reliability In the event of problems or failure the troubleshooting guidelines bo
105. choices include Solenoid A B or C and Solenoid D E or F if installed e In the Main Menu choose Service gt Gas Flow Calibration gt Solenoid Thermo Fisher Scientific Operation Service Menu i mpini PALAS MPA CORDED OPE TA TEN AS Bae Bast ade Gas Drive The Gas Drive screen is used to calibrate the gas air mass flow controller using a NIST traceable flow meter as a reference The zero drive 5 35 50 65 80 and 95 FS screens function the same way e Inthe Main Menu choose Service gt Gas Flow Calibration gt 5 20 35 50 65 80 or 95 FS Pressure Calibration The Pressure Calibration menu is used to calibrate the pressure sensor to zero span or restore factory default values The pressure calibration is visible only if the photometer option is installed and when the calibrator is in service mode For more information on the service mode see Service Mode earlier in this chapter The pressure sensor s zero counts and span slope are displayed on the menu Note This adjustment should only be performed by an instrument service technician A e Inthe Main Menu choose Service gt Pressure Calibration Thermo Fisher Scientific Model 146 Instruction Manual 3 89 Operation Service Menu Calibrate Pressure Zero The Calibrate Pressure Zero screen calibrates the pressure sensor at zero pressure Note A vacuum pump must be connected to the pressure sensor before pump P performing
106. cientific Servicing Fan Replacement Rear Panel 5 Install the new pump by following the previous steps in reverse Fan Replacement Use the following procedure to replace the rear panel fan Figure 7 6 Rear Panel Equipment Required Fan Philips screwdriver amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFE unplug the power cord and remove the cover 2 Remove the measurement bench Refer to Removing the Measurement Bench and Lowering the Partition Panel on page 7 9 3 Pull the power connector off the fan 4 Remove the fan guard from the fan and remove the filter 5 Remove the four fan mounting screws and remove the fan 6 Install a new fan following the previous steps in reverse Thermo Fisher Scientific Model 1467 Instruction Manual 7 17 Servicing Fan Replacement Mass Flow Controllers 7 Replace the measurement bench Mounting Screws 4 Filter Fan Guard Figure 7 6 Replacing the Rear Panel Fan Fan Replaceme Nt Use the following procedure to replace the cooling fan for the mass flow Mass Flow controllers Figure 7 6 Controllers 5 iment Required Fan Philips screwdriver amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Equipment
107. cnnnnnnnnncnrornononoronos B 2 Table B 11 Analog Current Output Range ValueS o oocinonnincinincccninicnnos B 2 Table B 12 Analog Voltage Output Range Values oo cece ceeeeeeeeeeees B 2 Table B 13 Default Output Assignment ccc ceccscescstesestesesteseeteseetens B 2 Table 6 1 Read Coils for iii AA C 1 Table C 2 Read Registers for 148 2 cccccsccecescecceceseecseseseccescesesesseseceens C 2 Table C 3 Write Coils for 146i oo cececsessessesteseeseeseesesteseeseeseenseeseeseses C 2 Thermo Fisher Scientific Principle of Operation Dilution Ozone Generation Thermo Fisher Scientific Chapter 1 Introduction The Model 1467 Multi gas Calibrator dilutes calibration gases to precise concentrations The diluted gases are used to perform zero precision and Level 1 span checks audits and multipoint calibration of analyzers The design of the Model 1467 meets or exceeds all published U S Environmental Protection Agency requirements for multipoint calibration audit Level 1 span and precision checks For details of the calibrator s principle of operation and product specifications see the following topics e Principle of Operation on page 1 1 e Specifications on page 1 7 All of the components used such as mass flow controllers ozone generator permeation tube oven power supplies solenoid valves and photometer have been used previously for calibration purposes and are known for their accurac
108. connector 6 Remove four screws holding flow controller bracket to floor plate 7 Remove flow controller assembly 8 Remove defective flow controller from bracket by removing the two screws holding flow controller to bracket 9 Install new flow controller by following the above directions in reverse 10 Leak check system and recalibrate Refer to Leak Test and Pump Checkout in this chapter Thermo Fisher Scientific Model 146 Instruction Manual 7 15 Servicing Pump Replacement Photometer option Pump Replacement For instruments with the photometer option use the following procedure to P P 8P Photometer option replace the pump see Figure 7 5 To rebuild the pump see Pump Rebuilding on page 5 9 in the Preventive Maintenance chapter Equipment Required 110V pump Philips screwdriver amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect pump power line from AC PUMP connector on the measurement interface board Pump Bracket Pump Mounting Screws 2 Bottom Figure 7 5 Replacing the Pump 3 Remove both lines from pump 4 Loosen the four captive screws holding the pump to the mounting plate and remove the pump 7 16 Model 146 Instruction Manual Thermo Fisher S
109. ction cannot be cleared replace the capillary Refer to the appropriate capillary replacement procedure that follows Table 5 1 Capillary Cleaning Wire Sizes Capillary Wire Diameter Ozonator 8 mil Photometer 8 mil Pressure Gauge 8 mil Pump 20 mil Refer to Component Layout on page 7 8 to locate these capillaries within the instrument If you need the capillary part number refer to the Replacement Parts List on page 7 5 Ozonator Capi l lary Use the following procedure to replace the ozonator capillary Figure 5 2 Replacement Ozonator option vient Required Ozonator capillary amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unscrew the knurled fitting separate the components and remove the capillary Figure 5 2 5 6 Model 146 Instruction Manual Thermo Fisher Scientific Preventive Maintenance Pressure Regulator Capillary Replacement 3 Insert the new capillary and reconnect the fittings Capillary Figure 5 2 Replacing the Ozonator Capillary Pressure Regulator Use the following procedure to replace the pressure regulator capillary Capillary Figure 5 3 Replacement Equipment Required Pressure regulator capillary amounts of static electricity A properly grounded ant
110. d 5 6 7 8 list Irec list srec list stream list sp These commands report the list of current selections for long record logging data short record logging data streaming data output or the scratch pad sp list The example below shows the list for streaming data output Send list stream Receive list stream field index variable x x time 1 1 conc 2 2 tgflo 3 3 agflo 4 4 tzflo 5 5 azflo 6 6 ttgflo 7 7 atflo sum 242d er xy Ir xy sr xy x 0 1 Reply termination format see set format format command y 0 1 2 Output format see set erec lrec srec format format command These commands report the last long and short records stored or the dynamic data record In the example below the command requests a long record with no checksum in ASCII format with text For details on how to decode the flag fields within these records see the flags command Send 1111 Receive 1111 11 00 12 12 05 flags C0012C9 conc 533 746 tgflo 25 000 agflo 26 689 tzflo 4975 000 azflo 4970 950 ttflo 5000 000 atflo 4997 640 pres 741 699 sum 2583 Model 146i Instruction Manual B 17 C Link Protocol Commands Datalogging B 18 Model 146 Instruction Manual erec This command returns a brief description of the main operating conditions at the time the command is issued i e dynamic data The example below shows a typical response The format is defined by the current settings of format and ere
111. d Not Used 6 Voltage Output Not Used Not Used Not Used Thermo Fisher Scientific Model 146 Instruction Manual B 45 C Link Protocol Commands 1 0 Configuration B 46 Model 146 Instruction Manual Dto A Function Single Range Dual Range Auto Range 7 Current Outputs 0 Not Used Not Used 8 Current Outputs 0 Not Used Not Used 9 Current Outputs Not Used Not Used Not Used 10 Current Outputs Not Used Not Used Not Used 11 Current Outputs Not Used Not Used Not Used 12 Current Outputs Not Used Not Used Not Used list var aout list var dout list var din These commands report the list of index numbers and the variables associated with that index number available for selection in the current mode determined by single dual auto gas mode for analog output digital output and digital inputs The index number is used to insert the variable in a field location in a list using set sp field index The example below reports the list of analog output index numbers and variables Send Receive relay relay stat list var aout list var aout index variable none conc teflo agflo tzflo azflo ttflo atflo ozone perm O pres 11 bncht 12 lmpt 13 o3lt 14 int 15 noise sum 311a Ru ON DUBRWNPR O This command reports the current relay logic normally open or normally closed if all the relays are set to same state that is all open or all closed The example below shows th
112. d by index 1 35 when the input transitions to the designated state high or low Use list din var command to obtain the list of supported index values and corresponding actions Send set din 5 9 high Receive set din 1 9 high ok dout channel This command reports the index number and output variable and the active state assigned to output channel The example below reports the input 2 to be assigned an index number 2 corresponding to local remote with the active state being open Send dout 2 Receive dout 2 2 LOCAL REMOTE open set dout channel index state This command assigns digital output channel to be assigned to the action associated with index and assigns it an active state of state open or closed Send set dout 2 2 open Receive set dout 2 2 open ok dtoa channel This reports the outputs of the 6 or 12 Digital to Analog converters according to Table B 13 The example below shows that the D A 1 is 97 7 full scale Send dtoa 1 Receive dtoa 1 97 7 Note All channel ranges are user definable Ifany customization has been made to the analog output configuration the default selections may not apply A Table B 13 Default Output Assignment DtoA Function Single Range Dual Range Auto Range 1 Voltage Output 0 Not Used Not Used 2 Voltage Output 0 Not Used Not Used 3 Voltage Output Not Used Not Used Not Used 4 Voltage Output Not Used Not Used Not Used 5 Voltage Output Not Used Not Use
113. d values are 0 to 999 If 0 is selected the program will run only once then it will be disabled e Inthe Main Menu choose gt Program gt Period Hours CH mtb Thermo Fisher Scientific Model 146 Instruction Manual 3 37 Operation Program Menu Next Cycle The Next Cycle screen is used view and change the start date and time of the program in 24 hour time e Inthe Main Menu choose gt Program gt Next Cycle Events The Events menu lets you configure events that are part of the program cycle The number of events and total time of enabled events are displayed The target values on the right side of the display is an abbreviated line that describes the settings for that specific event and should show the desired values based on the displayed operation parameters on the left side Specific parameters include gas A F Zero span level for that gas 1 5 Zero ozonator and permeation levels 1 5 OFF photometer mode I internal E external C control and event duration time in minutes Selecting an event allows detailed setup of that event A maximum of 32 events can be controlled from this menu Note When events are running none of the settings may be changed via the front panel interface except for the program cycle enable disable screen A e Inthe Main Menu choose gt Program gt Events ppp HT GEER hi Pike EE EE 3 38 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrumen
114. ded method for attaching the cable to the terminal board using the included tie down and spacer Table 2 3 identifies the connector pins and associated signals See Detail A See Detail B Assembled Connector Figure 2 8 25 Pin Terminal Board Views Table 2 3 25 Pin Terminal Board Pin Descriptions Detail A Detail B Pin Signal Description Pin Signal Description 1 10ut1 13 Analog_In1 2 GND_ISO 14 Analog_In2 3 10ut2 15 Analog_In3 4 GND_ISO 16 GNDD 5 10ut3 17 Analog_In4 6 GND_ISO 18 Analog_In5 7 lOut4 19 Analog_In6 8 GND_ISO 20 GNDD 9 10ut5 21 Analog_In7 10 GND_ISO 22 Analog_In8 11 10ut6 23 GNDD Model 146 Instruction Manual 2 13 Installation Startup Table 2 3 25 Pin Terminal Board Pin Descriptions continued Pin Signal Description Pin Signal Description 12 GND_ISO 24 GNDD Sta rtup Use the following procedure when starting the instrument 1 Turn the power ON 2 Allow 90 minutes for the instrument to stabilize 3 Set instrument parameters such as gas setup and flow modes to appropriate settings For more information about instrument parameters see the Operation chapter 2 14 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Display Chapter 3 Operation The Model 146 can produce accurate gas concentrations by diluting gas from a gas cylinder or a permeation tube ozone transfer standard ozone calibrat
115. direct access to specific operation and service information e Chapter 1 Introduction provides an overview of product features describes the principles of operation and lists the specifications e Chapter 2 Installation describes how to unpack setup and startup the analyzer e Chapter 3 Operation describes the front panel display the front panel pushbuttons and the menu driven software e Chapter 4 Calibration provides the procedures for calibrating the analyzer and describes the required equipment e Chapter 5 Preventive Maintenance provides maintenance procedures to ensure reliable and consistent instrument operation e Chapter 6 Troubleshooting presents guidelines for diagnosing analyzer failures isolating faults and includes recommended actions for restoring proper operation e Chapter 7 Servicing presents safety alerts for technicians working on the analyzer step by step instructions for repairing and replacing components and a replacement parts list It also includes contact information for product support and technical information e Chapter 8 System Description describes the function and location of the system components provides an overview of the software structure and includes a description of the system electronics and input output connections e Chapter 9 Optional Equipment describes the optional equipment that can be used with this analyzer Model 146
116. e Main Menu choose Service gt Perm Oven Settings gt Factory Cal Gas Therm nico LA ip priori PLP Low and High Points The Calibrate Gas Thermistor Low Point screen is used to view and set the permeation oven thermistor resistance The low and high point screens function the same way e Inthe Main Menu choose Service gt Perm Oven Settings gt Factory Cal Gas Therm gt Low or High Point 3 98 Model 146 Instruction Manual Thermo Fisher Scientific Operation Service Menu Set Defaults The Set Defaults screen is used to reset the configuration values to factory defaults e Inthe Main Menu choose Service gt Perm Oven Settings gt Factory Cal Gas Therm gt Set Defaults Pp E Boat Bosse Factory Calibrate Oven The Factory Calibrate Oven Thermistor menu is used for calibrating the Thermistor permeation oven heater thermistor to either low point high point or set defaults e Inthe Main Menu choose Service gt Perm Oven Settings gt Factory Cal Oven Therm Low and High Points The Calibrate Oven Thermistor Low Point screen is used to view and set the permeation oven thermistor resistance The low and high point screens function the same way e Inthe Main Menu choose Service gt Perm Oven Settings gt Factory Cal Oven Therm gt Low or High Point Thermo Fisher Scientific Model 146 Instruction Manual 3 99 Operation Service Menu Set Defaults The Set Defaults screen is used to
117. e glass clean when handling it 4 Place the permeation tube s in the glass chamber 5 Attach the glass chamber to the top assembly by gently pushing the two together with a slight twisting motion 9 2 Model 146 Instruction Manual Thermo Fisher Scientific Optional Equipment Permeation Oven 6 Replace the glass chamber assembly into the oven until the top of the assembly is flush or slightly below the top of the oven following step Equipment Damage Do not use tools to tighten the knurled screw in the 7 Tighten the knurled screw finger tight Do not use tools to tighten 8 Tighten the white plastic retaining collar 9 Replace the oven cover being careful to place the tubing and wiring into the cover slot Computation of The computation of the gas output level is shown in the following Concentrations information Note that is assumed that all devices are properly calibrated and that all flows are corrected to 25 C and 1 atm Permeation Tube R K oO Output ppm Where R permeation rate in ng min Qo flow rate of gas scc min into the charcoal scrubber during span mode K constant for the specific permeant 24 45 MW MW molecular weight Oven Installation and Use the following procedure to install and configure the permeation oven Configuration 1 Physically install the permeation oven and valves into the instrument and connect the cables and plumbing 2 From the Perm Oven Settings
118. e graphics display backlight to a single ribbon cable extending back to the motherboard This board also includes signal buffers for the graphics display control signals and a high voltage power supply for the graphics display backlight External I O is driven from a generic bus that is capable of controlling the following devices e Analog output voltage and current e Analog input voltage e Digital output TTL levels e Digital input TTL levels Note The instrument has spare solenoid valve drivers and I O support for future expansion a Model 146 Instruction Manual 8 11 System Description 1 0 Components 8 12 Analog Voltage Outputs Analog Current Outputs Optional Analog Voltage Inputs Optional Model 146 Instruction Manual The instrument provides six analog voltage outputs Each may be software configured for any one of the following ranges while maintaining a minimum resolution of 12 bits e 0 100mV e 0 1V e 0 5V 0 10V The user can calibrate each analog output zero and span point through firmware At least 5 of full scale over and under range are also supported The analog outputs may be assigned to any measurement or diagnostic channel with a user defined range in the units of the selected parameter The voltage outputs are independent of the current outputs The optional I O Expansion board includes six isolated current outputs These are software configured for any one of the following ra
119. e ranges for the voltage outputs are 0 100 mV 0 1 0 5 0 10 V e In the Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config gt Select Channel gt Select Range 3 62 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Minimum and Maximum The Minimum Value screen is used to edit the zero 0 to full scale 100 Value value in percentages for the selected analog output channel See Table 3 3 for a list of choices The minimum and maximum output value screens function the same way The example below shows the set minimum value screen e Inthe Main Menu choose Instrument Controls gt IO Configuration gt Analog Output Config gt Select Channel gt Set Minimum or Maximum Value ee PEE Table 3 3 Analog Output Zero to Full Scale Table Output Zero Value Full Scale 100 Value Gas Conc Zero 0 10 000 ppm 03 Conc Zero 0 Range setting Perm Conc Zero 0 10 000 ppm Target Gflow 0 0 Gas flow controller setting Actual Gflow 0 0 Gas flow controller setting Target Zflow 0 0 Zero flow controller setting Actual Zflow 0 0 Zero flow controller setting Target Tflow 0 0 Total gas and zero flow controller settings Thermo Fisher Scientific Model 146 Instruction Manual 3 63 Operation Instrument Controls Menu Output Zero Value Full Scale 100 Value Actual Tflow 0 0 Total gas and zero flow controller
120. easuring 150 sccm with a 0 01 inch capillary restricting the meter input 2 For instruments equipped with the ozonator option attach the digital flow meter to the gas line immediately preceding the ozonator capillary holder attached to the solenoid valve on the ozonator For instruments equipped with only the permeation oven option attach the digital flow meter in place of the tee fitting between the back pressure regulator input and the Perm 1 solenoid valve Model 146 Instruction Manual 7 51 Servicing Pressure Regulator s Adjustment 1 52 Model 146 Instruction Manual Connect a source of zero air to the zero air bulkhead connector and the GAS A bulkhead fitting located on the rear panel Activate the Service mode Refer to Accessing the Service Mode on page 7 11 then return to the next step in this procedure From the Main Menu press to scroll to Diagnostics press scroll to Instrument Configuration press and enable S P all appropriate options Set all parameters that apply as follows GASA Maximum flow see Note ZERO 0 000 PPM PERM OFF 0 000 PPM OZONE OFF 0 0 PHOT OFF Note To acquire maximum flow setting go to Service gt Flow Control Selection gt Zero Scale and use the zero flow controller zero scale value listed as the appropriate input for GAS A a 10 11 If the instrument includes an output pressure regulator reduce the output pressure to zero on the
121. eation Rate screen is used to enter the permeation rate of the permeation capsule used Allowable values are 1 0 to 99999 9 If the values selected are lower or higher than allowed TOO LOW or TOO HIGH appears on the screen Note Changing this value may affect allowable limits for permeation concentration levels It is advisable to check these settings A e Inthe Main Menu choose Permeation Oven Setup gt Permeation Rate WH papi Level 1 5 The Permeation Level 1 5 screens are used to set minimum and maximum permeation levels These levels can be selected from the Operation or Flow Modes screens Note When a gas is selected and mixed with the permeation oven output permeation levels cannot be selected A e Inthe Main Menu choose Permeation Oven Setup gt Level 1 5 Thermo Fisher Scientific Model 146 Instruction Manual 3 29 Operation Photometer Setup Menu Photometer Setup Menu Range 0 Range 3 30 Model 146 Instruction Manual The Model 146 can be equipped with an optional internal photometer which is used to measure concentration of ozone The Photometer Setup menu sets up the photometer e Inthe Main Menu choose Photometer Setup The Range menu allows the operator to view the gas units always in ppm to select the O range and to set the custom ranges e Inthe Main Menu choose Photometer Setup gt Range The O Range screen defines the concentration range of the anal
122. ed Send pres comp Receive pres comp on set pres comp onoff These commands turn the pressure compensation on or off The example below turns pressure compensation off This command is only available if the photometer option is installed Send set pres comp off Receive set pres comp off ok span index This command reports the current span index where 0 is zero air 1 5 are pre set span levels and 6 is manual span level Send span index Receive span index 1 set span index index index 0 6 Thermo Fisher Scientific Hardware Configuration Thermo Fisher Scientific C Link Protocol Commands Hardware Configuration This command sets the span index value to index Send set span index 1 Receive set span index 1 ok temp comp This command reports whether temperature compensation is on or off The example below shows the temperature compensation is off This command is only available if the photometer option is installed Send temp comp Receive temp comp off set temp comp onoff These commands turn the temperature compensation on or off The example below turns temperature compensation off This command is only available if the photometer option is installed Send set temp comp off Receive set temp comp off ok contrast This command reports the screen s level of contrast The example below shows the screen contrast is 50 according to Table B 9 Send contrast Receive contrast 5 50 set cont
123. ed when standby mode is turned back off e Inthe Main Menu choose Standby Gas Dilution The Gas Dilution menu is used to select a gas valve flow and span concentration The Gas A B or C menus set the zero air flow span 0 flow and the total flow output for a maximum of five span settings for each gas Additionally the gas menus let you manually set the zero air and gas flows manually for each gas Gas D E and F are available if the six gas option is installed and function the same way e Inthe Main Menu choose gt Flow Modes gt Gas Dilution 3 10 Model 146 Instruction Manual Thermo Fisher Scientific Operation Flow Modes Gas Span Selection The Gas Span Selection screen is used to select the span level and flow for the previously selected gas Once selected the screen reports the new flow mode The user must press to accept and activate these settings The message SAVING is displayed for 3 seconds then removed e Inthe Main Menu choose Flow Modes gt Gas Dilution gt Gas A F gt Zero or Span 1 5 Gas Manual Flow The Gas Manual Flow screen displays the zero air and gas flows and lets the user manually set the zero air and gas flows e Inthe Main Menu choose Flow Modes gt Gas Dilution gt Gas A F gt Manual Wi papi Thermo Fisher Scientific Model 1467 Instruction Manual 3 11 Operation Flow Modes Zero Air The Manual Zero Air Flow screen sets the manual zero air flow value for the
124. eive program on set program onoff This command sets programmed events on or off If set to off programmed events will not be triggered at their normal interval and if programmed events are currently running they will be aborted All the other set commands will be disabled if the programmed events are running except for this command If set to on programmed events will be triggered at their normal interval Send set program on Receive set program on ok clr records This command will clear all long and short records that have been saved Send clear records Receive clear records ok set clr lrecs set clr srecs These commands will clear only the long records or only the short records that have been saved The example below clears short records Send set clr srecs Receive set clr srecs ok list din list dout These commands report the current selection for the digital outputs in the format Output no Index number variable name active state The active state for digital outputs is open or closed The active state for digital inputs is high or low Send list dout Receive list dout output index variable state 1 1 GAS BIT 1 open 2 2 GAS BIT 2 open 3 3 GAS BIT 3 open 4 4 SPAN BIT 1 open Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging 5 SPAN BIT 2 open 6 SPAN BIT 3 open 7 03 PERM BIT 1 open 8 03 PERM BIT 2 open 9 9 03 PERM BIT 3 open 10 10 03 PERM BIT 4 open sum 462
125. enerate the same response These functions read the contents of one or more contiguous Read Input Registers registers These registers are 16 bits each and are organized as shown below All of the values are reported as 32 bit IEEE standard 754 floating point format This uses 2 sequential registers least significant 16 bits first The request specifies the starting register address and the number of registers Registers are addressed starting at zero Therefore registers numbered 1 16 are addressed as 1 15 The register data in the response message are packed as two bytes per register with the binary contents right justified within each byte For each register the first byte contains the high order bits and the second contains the low order bits Request Function code 1 Byte 0x03 or 0x04 Starting Address 2 Bytes 0x0000 to maximum allowed by instrument Quantity of Registers 2 Bytes 1 to maximum allowed by instrument Thermo Fisher Scientific Model 146 Instruction Manual C 5 MODBUS Protocol Function Codes C 6 Model 146 Instruction Manual Response Function code 1 Byte Byte count 1 Byte Register value N x 2 Bytes N Quantity of Registers Error Response Function code 1 Byte Exception code 1 Byte 0x03 or 0x04 2xN N N or N 1 Function code 0x80 01 IIlegal Function 02 Illegal Address 03 Illegal Data 04 Slave Device Failure Here is an example of a request and response to read registers 10 13 Request
126. enoid designations are Solenoid A for Gas A Solenoid B for Gas B and Solenoid C for Gas C the same default designation system applies to Solenoids D through E if the six gas option is installed Allowable selections are SOLENOID A SOLENOID B SOLENOID C SOLENOID D if installed SOLENOID E if installed and SOLENOID F if installed e Inthe Main Menu choose gt Gas Setup gt Gas A F gt Gas Solenoid The Tank Concentration screen lets you enter the concentration in parts per million ppm of each tank gas note that Gas A has been identified in the Name screen as SO and the Tank Concentration screen retains the identification Note Changing this valve may affect allowable limits for span concentration and total flow It is advisable to check these settings A Thermo Fisher Scientific Operation Gas Setup Menu e Inthe Main Menu choose gt Gas Setup gt Gas A F gt Tank Concentration Zero Air The Zero Air Flow screen sets the total flow for zero concentration output The screen identifies the minimum and maximum acceptable zero air flow rates and is used to select a flow rate between these two settings If the selected zero flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Gas Setup gt Gas A F gt Zero WH papi Span 1 5 The Span menu displays the span concentration and total flow ra
127. ent Configuration screen displays information on the Configuration hardware configuration of the instrument Note If the analyzer is in service mode pressing on the item will toggle it yes or no with the exception of purchased options A e Inthe Main Menu choose Diagnostics gt Instrument Configuration 3 76 Model 146 Instruction Manual Thermo Fisher Scientific Operation Alarms Menu Contact Information The Contact Information screen displays the customer service information e Inthe Main Menu choose Diagnostics gt Contact Information Alarms Menu The Alarms menu displays a list of items that are monitored by the instrument If the item being monitored goes outside the lower or upper limit the status of that item will go from OK to either LOW or HIGH respectively If the alarm is not a level alarm the status will go from OK to FAIL The number of alarms detected is displayed to indicate how many alarms have occurred If no alarms are detected the number zero is displayed To see the actual reading of an item and its minimum and maximum limits move the cursor to the item and press Ce Items displayed are determined by the options installed The motherboard status interface board status and I O Expansion board status if installed indicate that the power supplies are working and connections are successful There are no setting screens for these alarms e Inthe Main Menu choose Alar
128. ent should only be performed by an instrument service technician A e In the Main Menu choose Service gt Zero Flow Calibration Zero Air Drive The Zero Air Drive 20 screen is used to calibrate the zero air mass flow controller using a NIST traceable flow meter as a reference The zero drive 5 35 50 65 80 and 95 FS screens function the same way e In the Main Menu choose Service gt Zero Flow Calibration gt 5 20 35 50 65 80 or 95 FS Thermo Fisher Scientific Model 146 Instruction Manual 3 87 Operation Service Menu Gas Flow Calibration Gas Flow Calibration Solenoid Select 3 88 Model 146 Instruction Manual Moca A The Gas Flow Calibration menu is used to calibrate the gas flow controller at 5 20 35 50 65 80 and 95 percent of full scale Using a NIST traceable flow meter the accuracy of the controller can be calibrated to about 1 as opposed to the 2 accuracy rating when shipped from the manufacturer The gas flow calibration is visible only when the instrument is in service mode For more information of the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A e Inthe Main Menu choose Service gt Gas Flow Calibration nic Pirm Er ao Ep A LL A The Gas Flow Calibration Solenoid Select is used to select the solenoid that is active during the gas flow controller calibration Solenoid
129. entific Thermo Fisher Scientific Restore User Defaults nico prir LA is Password Set Password Operation Password The Restore User Defaults screen is used to reset the user calibration and configuration values to factory defaults The restore default user is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter e Inthe Main Menu choose Service gt Restore User Defaults The Password menu allows the user to configure password protection If the instrument is locked none of the settings may be changed via the front panel user interface The items visible under the password menu are determined by the instrument s password status e Inthe Main Menu choose Password The Set Password screen is used to set the password to unlock the front panel The set password is shown if the instrument is unlocked and the password is set e Inthe Main Menu choose Password gt Set Password Model 146 Instruction Manual 3 103 Operation Lock Instrument Lock Instrument The Lock Instrument screen is used to lock the instruments front panel so users can not change any settings from the front panel The lock instrument is shown if the instrument is unlocked and the password is set e Inthe Main Menu choose Password gt Lock Instrument Change Password The Change Password is used to change the password used to unlock the ins
130. enu is used to calibrate zero and span e Inthe Main Menu choose Photometer Setup gt Calibration Calibrate Zero The Calibrate Zero screen is used to automatically adjust the O background or perform a zero calibration Before performing a zero calibration ensure the analyzer samples zero air for at least 5 minutes It is important to note the averaging time when calibrating The longer the averaging time the more accurate the calibration will be To be most accurate use the 300 second averaging time e Inthe Main Menu choose Photometer Setup gt Calibration gt Cal Zero Model 146 Instruction Manual 3 35 Operation Program Menu 3 36 Model Calibrate 0 Coefficient Program Menu 1461 Instruction Manual The Calibrate O Coefficient screen is used to automatically adjust the O coefficient and enter the span concentration The display shows the current O concentration reading and the current Oz range The next line of the display is where the O calibration gas concentration is entered It is important to note the averaging time when calibrating The longer the averaging time the more accurate the calibration will be To be most accurate use the 300 second averaging time e Inthe Main Menu choose Photometer Setup gt Calibration gt Cal 03 Coef The Program menu sets the Model 1461 s internal computer program to conduct a series of calibration events at preset times and durations Up to 32 c
131. er has an airtight chamber that contains the sample and reference gases and includes a photometer lamp at one end and a detector at the other end The photo tube in the detector transmits light intensity information to the measurement interface board for sample measurement computations The pressure transducer measures the pressure of the sample gas The internal ozonator operates on the photolytic principle and the ozone level produced is a function of light intensity at 185 nm and gas flow The light intensity is varied by changing the current into the lamp The gas flow is held constant by a pressure regulator followed by a capillary The ozonator lamp power supply generates high voltage AC and contains heater control circuits for the ozonator lamp The photometer lamp power supply generates high voltage AC and contains heater control circuits for the photometer lamp The photometer temperature thermistor is used for controlling the P P 8 photometer at a consistent temperature The pump draws the sample gas through the photometer The sample reference solenoid valves allow sample gas or reference zero gas to pass through the cell Model 146 Instruction Manual 8 5 System Description Flow System Ozonator Solenoid Perm 1 Solenoid Perm 2 Solenoid Reaction Chamber Solenoid Zero Air Solenoid External Internal Solenoid Span Gas Solenoids Photometer Capillary Ozonator Capillary Pressure Gauge Capillary Pump C
132. er to traverse the various screens menus Figure 3 2 Front Panel Pushbuttons 3 2 Model 146 Instruction Manual Thermo Fisher Scientific Operation Pushbuttons Table 3 1 lists the front panel pushbuttons and their functions Table 3 1 Front Panel Pushbuttons Key Name E Soft Keys Function The E soft keys are used to provide shortcuts that allow the user to jump to user selectable menu screens For more information on processing soft keys see Soft Keys below Cr Run The Cr is used to display the Run screen The Run screen normally displays the current flows parameters and controls of the instrument Cm Menu The Cm is used to display the Main Menu when in the Run screen or back up one level in the menu system For more information about the Main Menu see Main Menu later in this chapter 2 Help The 2 is context sensitive that is it provides additional information about the screen that is being displayed Press 2 for a brief explanation about the current screen or menu Help messages are displayed using lower case letters to easily distinguish them from the operating screens To exit a help screen press MW J or to return to the previous screen or to return to the Run screen Up Down e gt Je Left Right The four arrow pushbuttons t p y e J and move the cursor up down left and right or change values and states in spec
133. erating source Gas Phase Titration GPT is achieved by combining ozone with a known NO concentration and measuring the loss of NO using the NO channel of a chemiluminescence analyzer The amount of NO formed is equal to the measured decrease in the NO level as shown in the following equation NO 0 gt NO 0 In the Model 1467 GPT is accomplished by routing zero air at approximately 150 cc per minute through the internal ozone generator and then precisely mixing this with NO from a gas cylinder connected to one of the gas inlets From there the NO O goes to the reaction chamber whose volume meets the dynamic parameter specification requirements of the U S EPA The gas is then fed into the mixing chamber and out the manifold at the rear of the instrument The Model 146 can also be used as a permeation gas source when equipped with an optional permeation tube oven as shown in Figure 1 3 Accuracy is achieved since both the release rate of the permeation tube gas and the flow of zero air through the zero air mass flow controller are known and the temperature of the permeation tube is stable Temperature stability of the permeation tube is accomplished by allowing only a small amount of zero air to enter the permeation tube oven to be heated The major portion of zero air bypasses the oven and is routed to the mixing chamber The solenoids on either side of the permeation tube oven under microprocessor control are used to route t
134. ere K molecular absorption coefficient 308 cm at 0 C and 1 atmosphere L length of cell 17 90 cm C ozone concentration in parts per million ppm I UV light intensity of sample with ozone sample gas I UV light intensity of sample without ozone reference gas Zero air is supplied to the Model 146 through the zero air bulkhead and is split into two gas streams as shown in Figure 1 5 One gas stream flows through the zero air solenoid and through the zero air mass flow controller then through the ozonator to the manifold and to the external internal solenoid to become the sample gas I The other stream flows from the zero air bulkhead to the sample reference solenoid in the photometer and is used as the reference gas I Alternately the sample gas may be pulled from an external source through the external internal solenoid The sample reference solenoid valve alternates the reference and sample gas streams between the cell every 5 seconds Ozone measured by the photometer is dumped out the exhaust bulkhead Figure 1 5 and Figure 1 6 show the hardware plumbing configuration with optional ozone generator internal permeation oven and photometer The UV light intensities of the zero and sample gases are measured by the detector When the solenoid valves switch the reference and sample gas streams the light intensity is ignored for several seconds to allow the cell to be flushed The Model 1461 calculates the ozone co
135. et gas a name abcd ok sel gas off This command turns off all gas solenoids Send sel gas off Receive sel gas off ok gas gas params gas a b c d e f This command returns the parameters for the specified gas Send gas a params Receive gas a params tank ppm 10 000 span O zflow 1800 000 span 1 tflow 1000 000 ppm 0 400 span 2 tflow 1000 000 ppm 0 800 span 3 tflow 1000 000 ppm 1 200 span 4 tflow 1000 000 ppm 1 600 span 5 tflow 1000 000 ppm 0 400 manual zflow 1000 000 gflow 30 50 gas gas solenoid gas a b c d e This command reports the solenoid number according to Table B 7 which is used when the gas gas is selected The example below shows that solenoid A is associated with gas A Send gas a solenoid Receive gas a solenoid 1 Model 146 Instruction Manual B 29 C Link Protocol Commands Measurement Configuration B 30 Model 146 Instruction Manual set gas gas solenoid number gas a b c dle f number 1 6 This command selects the solenoid number to be used when gas gas is selected The tank for the specified gas should be connected to the specified bulkhead fitting The example below shows that solenoid C is set for gas A according to Table B 7 Send set gas a solenoid 1 Receive set gas a solenoid 1 ok Table B 7 Solenoid Number Values Value Solenoid Value 1 Solenoid A Solenoid B Solenoid C Solenoid D Solenoid E Solenoid F oO or SS wo N sel gas gas span ga
136. et of the mass flow controller 2 Measure barometric pressure and room temperature 3 Connect a suitable flow meter to the mass flow controller outlet 4 Set Model 1467 to Gas Drive or Zero Air Flow Calibration as described in the Operation chapter 5 Set flow controller to 95 percent of full scale then wait until flow meter reading stabilizes 6 Enter the flow meter reading using the flow input screen 7 Repeat Steps 5 and 6 for the remaining flow settings If difficulty is encountered due to a flow controller malfunction contact Thermo Fisher Scientific Thermo Fisher Scientific Model 146 Instruction Manual 4 1 Calibration Photometer Calibration Photometer The ozone generator used on the Model 1467 qualifies as a transfer standard Calibration under current U S Environmental Protection Agency requirements In order to use it as a transfer standard the user must certify the generator following the procedures published in the technical assistance document for the calibration of ozone transfer standards In order to determine a calibration curve for the ozone generator an ozone primary standard must be used Ifan ozone analyzer is also available the generation of the calibration curve becomes easier A procedure for both is given below Calibration Using a Model Use the following procedure to calibrate the Model 146i using a Model 49 49 PS PS Refer to the Photometer Calibration Flow Schematic Figure 4 1 1
137. f the 10 digital outputs and allows togeling of the state to either on 1 or off 0 The relays are restored to their original states upon exiting this screen e Inthe Main Menu choose Diagnostics gt Relay States m Lal mere Solenoid States The Solenoid State screen displays the state of the 10 digital outputs and allows toggling of the state to either on 1 or off 0 The solenoids are restored to their original states upon exiting this screen e Inthe Main Menu choose Diagnostics gt Solenoid States PP HED Eel Test Analog Outputs The Test Analog Outputs menu contains a number of digital to analog converter DAC calibration items Channel choices include all analog outputs 6 voltage channels and 6 current channels if the I O expansion board option is installed e Inthe Main Menu choose Diagnostics gt Test Analog Outputs Thermo Fisher Scientific Model 146 Instruction Manual 3 75 Operation Diagnostics Menu Set Analog Outputs The Set Analog Outputs screen contains three choices Set to full scale set to zero or reset to normal Full scale sets the analog outputs to the full scale voltage zero sets the analog outputs to 0 volts and normal operation The example below shows the selected output state ALL is set to normal e Inthe Main Menu choose Diagnostics gt Test Analog Outputs gt ALL Voltage Channel 1 6 or Current Channel 1 6 DH vided Piet Instrument The Instrum
138. fer to Removing the Measurement Bench and Lowering the Partition Panel on page 7 9 to lower the partition panel then proceed to the next step below 2 Unplug all connectors Note the locations of the connectors to facilitate reconnection 3 Remove the two retaining screws from the top of the board 4 Pop off the measurement interface board from the two mounting studs at the bottom of the board and remove the board Thermistor Measurement Ke Y Interface de Board__ Sy Ozonator Board Photometer Board Figure 7 10 Replacing the Measurement Interface Board 5 To install the measurement interface board follow previous steps in reverse 6 Re install the measurement bench Refer to Removing the Measurement Bench and Lowering the Partition Panel on page 7 9 Front Panel Board use the following procedure to replace the front panel board Figure 7 11 Replacement Equipment Required 7 28 Model 146 Instruction Manual Thermo Fisher Scientific Servicing LCD Module Replacement Eront panel board amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the three ribbon cables and the two wire connector from the front panel board 3 Pop off the board from the two top moun
139. ff This command turns off the ozonator and permeation oven if these options are present Send set ozon perm off Receive set ozon perm off ok perm This command reports the current permeation oven output level where 0 is off This command is only available if the permeation oven option is installed Send perm Receive perm 1 sel perm index set perm index index 0 5 This command selects preset permeation oven levels to index This command is only available if the permeation oven option is installed Send sel perm 1 Receive set perm 1 ok set 03 conc value Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurement Configuration This command sets the ozonator target concentration under photometer control in ppb This command is only available if the photometer option is installed Send set 03 conc 500 Receive set 03 conc 500 ok 03 setting This command reports the ozonator target concentration under photometer control in ppb The example below reports a setting of 100 ppb This command is only available if the photometer option is installed Send 03 setting Receive 03 setting 1 000E 02 ppb range This commands report photometer O range If the mode is incorrect the instrument responds with can t wrong settings The example below reports that the O full scale range is set to 5 0 ppm according to Table B 8 This command is only available if the photometer optio
140. fic Servicing Replacement Parts List Re laceme nt F arts Table 7 1 lists the replacement parts for the Model 146i major P P J I subassemblies Refer to Figure 7 2 to identify the component location ist 8 p Thermo Fisher Scientific Table 7 1 Model 146i Replacement Parts Part Number 100480 00 101491 00 100533 00 100539 00 102014 00 102340 00 102496 00 101681 00 101399 00 101863 00 100880 00 102590 00 102775 00 100895 00 101023 00 102455 00 103209 00 102351 00 102570 00 8540 102554 00 101426 00 102464 01 102464 02 101055 00 100907 00 8630 4510 14007 103075 00 102604 00 Description Front Panel Pushbutton Board Processor Board Motherboard Digital Output Board 1 0 Expansion Board Optional Front Panel Connector Board Front Panel Display DC Power Supply Transformer 220 240VAC Optional Transformer 100VAC Optional Measurement Interface Board Ozonator Ozonator Lamp Ozonator Board Pressure Transducer Air Regulator Assembly Optional Detector Assembly Photometer Lamp Heater Assembly Photometer Photometer Lamp Photometer Board Pump Pump Zero Air Supply 110V Optional Pump Zero Air Supply 220V Optional AC Receptacle Assembly Fan Fan Filter Fuse 250VAC 3 0 Amp SloBlow for 100VAC and 110VAC models Fuse 250VAC 1 60 Amp SloBlow for 220 240VAC models Back Pressure Regulator Pressure Regulator Model 146 Instruction Manual 7 5 Servicing Cable List 1 6
141. figuration set dhcp onoff These commands enables and disables the DHCP service by either on or off Changes to this parameter will only take effect when the analyzer is powered up The example below sets the DHCP service on Note When DHCP is set to on the user supplied addr gw addr dns addr ip and addr nm parameters are not used A Send set dhcp on Receive set dhcp on ok format This command reports the current reply termination format The example below shows that the reply format is 00 which means reply with no checksum according to Table B 10 Send format Receive format 00 set format format This command sets the reply termination format where format is set according to Table B 10 The example below sets the reply termination format to checksum Send set format 01 Receive set format 01 ok Table B 10 Reply Termination Formats Format Reply Termination 00 lt CR gt 01 lt NL gt sum xxxx lt CR gt where xxxx 4 hexadecimal digits that represent the sum of all the characters bytes in the message host name This command reports the host name string Send host name Receive host name analyzero1 set host name string This command sets the host name string where string is 1 13 alphanumeric characters Model 146 Instruction Manual B 41 C Link Protocol Commands Communications Configuration B 42 Model 1461 Instruction Manual Send set host name analyzero1 Receive set host name analyzer
142. g Troubleshooting Guides Troubleshooting Guides The troubleshooting guides presented in this chapter are designed to help isolate and identify instrument problems Table 6 1 provides general troubleshooting information and indicates the checks that you should perform if you experience an instrument problem Table 6 2 lists all the alarm messages you may see on the graphics display and provides recommendations about how to resolve the alarm condition Table 6 1 Troubleshooting General Guide Malfunction Does not start up Possible Cause No power Power supply Action Check that the instrument is plugged into the proper source 115 or 220 volts Check instrument fuse Check power supply voltages Flow controller unstable Leak Flow controller malfunction Perform a leak check see Servicing chapter Replace Flow Controller Solenoid not switching in local mode Instrument in remote mode Solenoid malfunction Defective cable Defective connector Defective solenoid connector board Defective measurement interface board Put instrument in local mode Check 24 VDC on solenoid connector Check solenoid cable Check solenoid for continuity and replace as necessary Replace cable Replace connector Replace solenoid connector board Replace measurement interface board Solenoid not switching in remote mode 6 2 Model 146 Instruction Manual Instrumen
143. g data Send stream per Receive stream per 10 set stream per number value number value 1 2 5 10 20 30 60 90 120 180 240 300 This command sets the time interval between two consecutive streaming data strings to number value in seconds The example below sets the number value to 10 seconds Send set stream per 10 Receive set stream per 10 ok stream time This command reports if the streaming data string will have a time stamp attached to it or not according to Table B 5 Send stream time Receive stream time O Thermo Fisher Scientific Thermo Fisher Scientific Calibration C Link Protocol Commands Calibration set stream time value This command enables value where value is to attach or disable time stamp to streaming data string according to Table B 5 The example below attaches a time stamp to streaming data Send set stream time O Receive set stream time O ok Table B 5 Stream Time Values Value Stream Time 00 Attaches time stamp to streaming data string 01 Disables time stamp to streaming data string set cal 03 bkg This command will auto calibrate the O3 background The example below shows a successful auto calibration of the O3 background This command is only available if the photometer option is installed Send set cal 03 bkg Receive set cal 03 bkg ok set cal 03 coef This command will auto calibrate Oz coefficients based on Oz span gas concentrations If t
144. g the current barometric pressure from the local weather station or airport and comparing it to the pressure reading However since these pressures are usually corrected to sea level it may be necessary to correct the reading to local pressure by subtracting 0 027 mm Hg per foot of altitude Do not try to calibrate the pressure transducer unless the pressure is known accurately Use the following procedure to calibrate the pressure transducer Equipment Required Vacuum pump WARNING The service procedures in this manual are restricted to qualified service representatives If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired a Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Remove the cover 7 40 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Servicing Pressure Transducer Calibration Photometer option 2 Disconnect the tubing from the pressure transducer and connect a vacuum pump known to produce a vacuum less than 1 mm Hg 3 From the Main Menu press to scroll to Service gt press gt to scroll to Pressure Calibration gt and press e The Pressure Sensor Cal screen appears Note If Service is not displayed refer to Accessing the Se
145. gauge Set the back pressure regulator to indicate a flow of 150 sccm on the digital flow meter Change the GAS A flow rate to 400 sccm and observe the digital flow meter to verify that the flow rate remains greater than 125 sccm Disconnect the digital flow meter and all associated plumbing reconnect the instrument plumbing and tighten securely Set GAS A to OFF and set the output pressure regulator to 10 psig Thermo Fisher Scientific Servicing Ambient Temperature Calibration 12 Tighten the locknut on the back pressure regulator and set the lock on the pressure adjust knob on the output pressure regulator Ambient Tempe rature Use the following procedure to calibrate the ambient internal temperature Calibration for the instrument Equipment Required Calibrated thermometer or 10K 1 resistor WARNING The service procedures in this manual are restricted to qualified service representatives a If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Remove the instrument cover 2 Tape the thermistor Figure 7 3 to a calibrated thermometer Note Since the thermistors are interchangeable to an accuracy of 0 2 C and have a value of 10K ohm
146. gt gt gt gt gt o o o o o o gt gt External Device Connection Components instrument Description DC Power Supply Status Monitor Motherboard to Front Panel Board DC Power Supply AC Input AC from Receptacle AC to Power Switch Main AC Receptacle Assembly Measurement Interface Board to Photometer Board Measurement Interface Board to Ozonator Board Fan Power Cable Mass Flow Controller Cable Assembly Table 7 3 lists the standard and optional cables and components used for connecting external devices such as PCs and data loggers to an Series Table 7 3 External Device Connection Components Part Number 101562 00 101556 00 102645 00 102646 00 102659 00 6219 102888 00 102891 00 103084 00 Thermo Fisher Scientific Description Terminal Block and Cable Kit DB25 optional Terminal Block and Cable Kit DB37 optional Cable DB37M to Open End Cable Six Feet optional Cable DB37F to Open End Six Feet optional Cable DB25M to Open End Six Feet optional Cable RS 232 optional Terminal Board PCB Assembly DB37F standard with all instruments Terminal Board PCB Assembly DB37M standard with all instruments Terminal Board PCB Assembly DB25M included with optional 1 0 Expansion Board in all instruments Model 146 Instruction Manual 7 7 Servicing External Device Connection Components Motherboard Ozonator Board Top Digital Output Board Photometer Board Bo
147. gt GPT Photometer Control gt Level 1 5 gt Gas A F Pi Taal DH Cited Haat ake EE Scala Pi EE EE Gas Span Selection The Gas Span Selection screen is used to select the span level for the previously selected gas Gas D E and F are available if the six gas option is installed Once selected the screen reports the new flow mode The user must press to accept and activate these settings The message SAVING is displayed for 3 seconds then removed e Inthe Main Menu choose Flow Modes gt O3 gt GPT Photometer Control gt Level 1 5 gt Gas A F gt Zero or Span 1 5 3 18 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific DEEE hpp Koi iy ii Gas Manual Flow Zero Air Operation Flow Modes prirni E pIo A el Led i E Eiet Boat Bosse BEE ES E Boat banos The Gas Manual Flow screen shows the zero air and gas flows and is used to manually set them for the previously selected gas If the selected zero air flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e In the Main Menu choose Flow Modes gt O3 gt GPT Photometer Control gt Gas A F gt Level 1 5 gt Manual The Manual Gas Zero Air Flow screen sets the manual zero air flow value for the previously selected gas The screen identifies the minimum and maximum acceptable zero air flow rates and lets the user select a flow rate between these tw
148. hat must be used in performing the gas phase titration is to ensure the NO O3 reaction has been completed Empirically certain dynamic parameter constraints have been derived to ensure completion of the reaction These conditions dictate that the residence time in the reaction chamber the gas flow rate and reaction chamber volumes meet the following conditions Thermo Fisher Scientific Model 146 Instruction Manual 8 3 System Description Flow System 8 4 Permeation Tube Oven Model 146 Instruction Manual Condition 1 Py NO gc x ty gt 2 75 ppm min F Condition 2 No RC 7 NO stp E Ro Vac ss te lt 2min Condition 3 R E Fio where Pr the dynamic specification determined empirically to ensure complete reaction of the available O3 ppm min NOIrc the NO concentration in the reaction chamber ppm tr the residence time in the reaction chamber min NO stp the concentration of the NO gas cylinder ppm Fno the NO flow rate sccm Fo the O3 generator air flow rate sccm Vere the volume of the reaction chamber The flows through the Model 146 satisfy these requirements The reaction chamber volume is 150 cc Flow through the ozonator is approximately 150 sccm using the back pressure regulator set to approximately 3 psi and the capillary All excess zero air flow bypasses the ozone generator The dynamic parameter conditions are met for any reasonable NO flow 12 5 100 sccm and NO STD 40
149. he analyzer can be installed in the configurations described in Table 9 3 and shown in Figure 9 4 through Figure 9 7 Table 9 3 Mounting Options Mounting Type Description Bench Positioned on bench includes mounting feet and front panel side trim ears ElA rack Mounted in an ElA style rack includes mounting slides and front panel ElA rack mounting ears Retrofit rack Mounted in a Thermo non ElA rack includes mounting slides and retrofit front panel rack mounting ears Thermo Fisher Scientific Model 146 Instruction Manual 9 13 E F F P F b F P h F b P p P Figure 9 4 Rack Mount Option Assembly 9 14 Model 146 Instruction Manual Thermo Fisher Scientific 16 75 REF Figure 9 5 Bench Mounting Thermo Fisher Scientific Model 146 Instruction Manual 9 15 Figure 9 6 ElA Rack Mounting 9 16 Model 146 Instruction Manual Thermo Fisher Scientific Figure 9 7 Retrofit Rack Mounting Thermo Fisher Scientific Model 146 Instruction Manual 9 17 Optional Equipment Mounting Options 9 18 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Appendix A Warranty Seller warrants that the Products will operate or perform substantially in conformance with Seller s published specifications and be free from defects in material and workmanship when subjected to normal proper and intended usage by properly trained personnel for the period of time set fo
150. he flow of gas going through the oven to either the main air stream or to exhaust It should be noted that zero air must always flow through the permeation tube oven regardless of the 1467 output flow so the oven may be kept at a stable temperature To accomplish this a flow of zero air set by a capillary at approximately 150 cc per minute is continuously fed through the permeation tube oven When only the permeation oven is activated up to five permeation dilution gas levels ppm can be set When the permeation gas source is used in conjunction with the gas dilution system a permeation dilution concentration cannot be specified because the gas dilution system controls the flow rate of zero air In that case only the permeation concentration and not a permeation level is displayed Thermo Fisher Scientific Thermo Fisher Scientific Photometer Introduction Principle of Operation Figure 1 4 shows the hardware plumbing configuration with both the optional ozone generator and permeation oven installed This figure is for illustration purposes only since there is no calibration gas requirement that would necessitate activating both hardware components simultaneously The Model 1467 photometer operates on the principle that ozone O3 molecules absorb UV light at a wavelength of 254 nm The degree to which the UV light is absorbed is directly related to the ozone concentration as described by the Beer Lambert Law T KLC e Lo wh
151. he instrument ID The ID is used to identify the instrument when using the C Link or MODBUS protocols to control the instrument or collect data It may be necessary to edit the ID number if two or more of the same instruments are connected to one computer Valid Instrument ID numbers are from 0 to 127 The Model 1467 has a default Instrument ID of 51 For more information about the Instrument ID see Appendix B C Link Protocol Commands or Appendix C MODBUS Protocol e Inthe Main Menu choose Instrument Controls gt Communication Settings gt Instrument ID BOE prr PRALE Lidi Communication Protocol The Communication Protocol screen is used to change the instrument communication protocol for serial communications e In the Main Menu choose Instrument Controls gt Communication Settings gt Communication Protocol Thermo Fisher Scientific Model 146 Instruction Manual 3 49 Operation Instrument Controls Menu Streaming Data The Streaming Data Configuration menu is used to allow for configuration Configuration of the 8 streaming data output items streaming interval current data format and current timestamp setting The Choose Item Signal submenu displays a list of the analog output signal group choices to choose from Choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board option is installed e Inthe Main Menu choose Instrument Controls gt Communication Settings gt
152. he line is terminated by a n Valid fields are t parse a time specifier 2 bytes D parse a date specifier 3 bytes i ignore one 8 bit character 1 byte e parse a 24 bit floating point number 3 bytes n x E parse a 24 bit floating point number 3 bytes N x f parse a 32 bit floating point number 4 bytes c parse an 8 bit signed number 1 byte C parse an 8 bit unsigned number 1 byte n parse a 16 bit signed number 2 bytes N parse a 16 bit unsigned number 2 bytes m parse a 24 bit signed number 3 bytes M parse a 24 bit unsigned number 3 bytes l parse a 32 bit signed number 4 bytes L parse a 32 bit unsigned number 4 bytes There is an optional single digit d which may follow any of the numeric fields which indicates that after the field has been parsed out the resulting value is to be divided by 104d Thus the 16 bit field OxFFC6 would be interpreted with the format specifier n3 as the number 0 058 Thermo Fisher Scientific Format Specifier for Front Panel Layout Thermo Fisher Scientific Text Value String Value Source C Link Protocol Commands Record Layout Definition The subsequent lines in the ERec Layout response describe the appearance of the full panel The full instrument panel as it appears on the screen has two columns of lines Each line is composed of three major components 1 a text field 2 a value field and 3 a button None of these three components
153. he mode is incorrect the instrument responds with can t wrong settings The example below shows a successful auto calibration of the O3 coefficient This command is only available if the photometer option is installed Send set cal 03 coef Receive set cal 03 coef ok lamp This command reports the current photometer lamp setting in percent The example below shows that the lamp setting is 10 0 This command is only available if the photometer option is installed Send lamp Receive lamp 10 0 set lamp value This command sets the photometer lamp setting to value where value is a floating point number representing a percentage of the photometer lamp setting The example below sets the photometer lamp to 10 0 This command is only available if the photometer option is installed Model 146 Instruction Manual B 23 C Link Protocol Commands Calibration B 24 Model 1461 Instruction Manual Send set lamp 10 0 Receive set lamp 10 0 ok lamp setting This command reports the current photometer lamp setting The example below shows the lamp setting is 10 0 This command is only available if the photometer option is installed Send lamp Receive lamp 10 0 03 coef This command reports Os coefficients If the mode is incorrect the instrument responds with can t wrong settings The example below reports that the O coefficient is 1 000 This command is only available if the photometer option is installed
154. he photometer detector intensity minimum and maximum values to value where value is a floating point number representing intensity alarm limits in hertz The example below sets the intensity alarm maximum value to 150000 Hz These commands are only available if the photometer option is installed Send set alarm intensity max 150000 Receive set alarm intensity max 150000 ok alarm internal temp min alarm internal temp max These commands report the ambient temperature alarm minimum and maximum values The example below reports the internal temperature alarm minimum value is 15 0 C Send alarm internal temp min Receive alarm internal temp min 15 0 deg C set alarm internal temp min value set alarm internal temp max value These commands set the internal temperature minimum and maximum values to value where value is a floating point number representing ambient temperature alarm limits in degrees C The example below sets the internal temperature alarm maximum value to 45 0 C Send set alarm internal temp max 45 Receive set alarm internal temp max 45 ok alarm 03 lamp temp min alarm 03 lamp temp max These commands report the ozonator lamp temperature alarm minimum and maximum values The example below reports the ozonator lamp temperature alarm minimum value is 52 C These commands are only available if the photometer option is installed Send alarm 03 lamp temp min Receive alarm 03 lamp temp min 52 0 deg C set ala
155. hen temperature compensation is off the display shows 0 0 C e Inthe Main Menu choose Instrument Controls gt Temperature Compensation Pressure The Pressure Compensation screen is used to turn the pressure correction Compensation off When pressure compensation is on the first line of the display represents the current chamber pressure When pressure compensation is off the first line of the display shows the factory standard pressure of 760 mmHg e In the Main Menu choose Instrument Controls gt Pressure Compensation 3 40 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu E E En REP E mT BE in E Eiet fat Boas Bast ad Datalogging Settings The Datalogging Settings menu deals with datalogging e Inthe Main Menu choose Instrument Controls gt Datalogging Settings a En REP ED AS BE Fas fast Basse Bast abe a Pi PEE EE Select SREC LREC The Select SREC LREC is used to select short record or long record format for other operations in this menu e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Select SREC LREC PPE EI En BME ED porr A Lf WH mid rib Piri View Logged Data The View Logged Data screen is used to select the start point to view the logged data by number of records or date and time Thermo Fisher Scientific Model 1467 Instruction Manual 3 41 Operation Instrument Controls Menu
156. her Scientific Measurement Interface Board Replacement A Thermo Fisher Scientific Servicing Measurement Interface Board Replacement Nut driver 3 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the I O expansion board optional if used See the I O Expansion Board Replacement procedure in this chapter 3 Remove the digital output board See the Digital Output Board Replacement procedure in this chapter 4 Unplug all connectors from the motherboard Note connector locations to facilitate reconnection 5 Using the nut driver remove the six standoffs securing the board to the rear panel 6 Pop off the motherboard from motherboard support bracket and remove the motherboard 7 To install the motherboard follow previous steps in reverse Use the following procedure to replace the measurement interface board Figure 7 10 Equipment Required Measurement interface board Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Model 146 Instruction Manual 7 27 Servicing Front Panel Board Replacement 1 Re
157. hermo Fisher Scientific Model 146 Instruction Manual 3 91 Operation Service Menu Bench Temperature Calibration Analog Output Calibration 3 92 Model 146 Instruction Manual The Bench Temperature Calibration screen is used to view and set the bench temperature sensor calibration This temperature is used for compensation of the photometer option s ozonator measurement The temperature calibration is visible only if the photometer option is installed and when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A e Inthe Main Menu choose Service gt Bench Temp Calibration The Analog Output Calibration menu is a selection of 6 voltage channels and 6 current channels if I O expansion board option is installed to calibrate and allows the user to select the calibration action zero or span The analog output calibration is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A Thermo Fisher Scientific Operation Service Menu e Inthe Main Menu choose Service gt Analog Out Calibration E EN En REP ED Pr Tit CH bo bet ee da oe m Lo d TP E a E Pio roi Bia E prr CH piini 1 ii L
158. hotometer Replacement Photometer option 7 34 Photometer Replacement Photometer option A Model 146 Instruction Manual 3 Loosen the Allen screws 4 securing the photometer to the floor plate Figure 7 12 and remove the photometer 4 Remove four 4 mounting screws on the lamp heater and remove the heater 5 Install the new lamp heater and secure with the four 4 mounting screws Use the following procedure to replace the photometer Figure 7 12 Equipment required Photometer Allen wrench 7 64 inch Open end wrench 9 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Use the following procedure to replace the photometer 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the plumbing and electrical connections 3 Loosen the four 4 socket head retaining screws securing the photometer to the floor plate Figure 7 12 and remove the photometer 4 Perform the previous steps in reverse to install the new photometer 5 Calibrate the instrument Refer to the Calibration chapter in this manual Thermo Fisher Scientific Photometer Board Replacement Photometer option A Detector Replacement Photometer option Thermo Fisher Scientific Servicing Photometer Board Replacement Photometer option Use the
159. i heeii niai aE EE h 8 2 A A oe rere eae ter hu nRa Pere aa a eras Neea 8 2 Mass Elo wContro lle rs es e a NS 8 2 Model 1467 Instruction Manual ix Contents Chapter 9 x Model 1467 Instruction Manual B ck Pressure Ree ulate tetas 2st ceacccstet distaccaane stato cantar access tik A 8 2 Reaction Chamber iness aiora ia AREA 8 2 Pressure Regulator and Gauge ici pida 8 2 SOLE 0d 8 2 Instrument Controles ar a 8 2 Mo it ring Signals rin ii a Aaa 8 2 Measurement Call IO cds 8 2 OMtPputC Oncol ic cgabtes dunes guucteuateewecess 8 2 Electronics iii Ai it 8 2 Motherboard ars tia 8 2 Measurement Interface Board nie 8 2 Digital Cat put Borra ienaa ne 8 2 I O Expansion Board Optional lada toi 8 2 Front Panel Connector Board rita 8 2 HO COMPONER 8 2 Analog Voltage Outputs iii ita 8 2 Analog Current Outputs Optional po it 8 2 Analog Voltage Inputs Optional ida aaa 8 2 Digital Reliy Cipsa o 8 2 Digital puts cirios ina 8 2 Serial Porras 8 2 RS 232 E O 8 2 RS485 Connections ia aai parada 8 2 Ethernet Connection en Oca aca a E a a 8 2 External Accessory Connector nido 8 3 Optional Equipment sssssssssssssssssnnnnesssunnnnnnnnnnunnununnnennnnunnnnnnnnnnnnnnnnnnnnnnnnnnne 9 1 A nen a ste a a R E RE 9 1 Photos e o e es 9 1 Zero JADE SOECES pri ls 9 1 A A ieinsadsveuteteitriaateiededeutnadens 9 1 Permeation Tube Installation ia 9 1 Computation of Concentrations aii 9 1 Oven Installation and Contursi da 9 1 Permeation Tube Oven C
160. ibrate Zero screen is used to calibrate the zero state of Zero the selected analog input 3 94 Model 146 Instruction Manual Thermo Fisher Scientific Analog Input Calibrate Full Scale Permeation Oven Settings Thermo Fisher Scientific Operation Service Menu e Inthe Main Menu choose Service gt Analog Input Calibration gt Select Channel gt Calibrate Zero Hook up a voltage source of 0 V to the analog input channel Pirin The Analog Input Calibration Full Scale screen is used to calibrate the full scale state of the selected analog input e Inthe Main Menu choose Service gt Analog Input Calibration gt Select Channel gt Calibrate Full Scale Hook up a voltage source of 10 V to the analog input channel The Permeation Oven Settings menu is used for setting up and calibrating the permeation oven option The permeation oven settings are visible only if the permeation oven option installed and when the calibrator is in service mode For more information on the service mode see Service Mode earlier in this chapter e Inthe Main Menu choose Service gt Perm Oven Settings Note This adjustment should only be performed by an instrument service technician A Model 146 Instruction Manual 3 95 Operation Service Menu Calibrate Gas Thermistor The Calibrate Gas Thermistor menu is used for calibrating the permeation oven gas thermistor using a water bath or known resistance e Inthe Main Menu
161. id Control GAS C J3 24V 2 Gas C Solenoid Control GAS D J4 24V Model 146 Instruction Manual 6 23 Troubleshooting Connector Pin Descriptions 6 24 Model 146 Instruction Manual Table 6 8 Solenoid Connector Board Pin Descriptions continued Connector Reference Pin Signal Label Designator Description 2 Gas D Solenoid Control GAS E J5 24V 2 Gas E Solenoid Control GAS F J6 24V 2 Gas F Solenoid Control Z A J7 24V 2 Z A Solenoid Control EXT INT J8 24V 2 EXT INT Solenoid Control Table 6 9 Ozonator Power Supply Connector Pin Descriptions Connector Label Reference Designator Pin Signal Description LAMP J1 Ozonator Lamp Voltage Out Ozonator Lamp Voltage Return LAMP HEATER J2 o NM A WO N Ground Ozonator Lamp Thermistor 1 Ozonator Lamp Heater Control Ground 24V Heater Current Sense Ozonator Lamp Thermistor 2 INTF e mo N on A Ozonator Lamp Intensity Control Voltage Input Ozonator Lamp Intensity Control Voltage Output Ozonator Heater Thermistor 24V Ground Ozonator Lamp On off Control Ground Ground Thermo Fisher Scientific Troubleshooting Service Locations Table 6 10 Photometer Power Supply Connector Pin Descriptions Connector Reference Pin Signal Label Designator Description LAMP J1 1 Photometer Lamp Voltage Out 2 Photometer Lamp Voltage Return LAMP HEATER J2 Ground Photometer L
162. ies Screen information is reported using the iscreen command above Model 146 Instruction Manual B 27 C Link Protocol Commands Measurement Configuration Send screen Receive screen This is an I series Instrument Screen Information not available Measurement ge index E p This command reports the current gas index This selects the gas that will Configuration be diluted according to Table B 6 The example below reports that Gas A is currently selected Send gas index Receive gas index 1 set gas index index index 0 6 This command sets the current gas index value to index The example below selects Gas D according to Table B 6 Send set gas index 4 Receive set gas index 4 ok Table B 6 Gas Index Values Value Gas Index 0 None 1 Gas A 2 Gas B 3 Gas C 4 Gas D 5 Gas E 6 Gas F sel gas gas man gas a b cldle f This command selects the gas at the level preset for manual Send sel gas a man Receive sel gas a man ok B 28 Model 1467 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurement Configuration gas name gas gas a b c d e f This command reports the gas name for gas gas Send gas a name Receive gas a name abcd set gas name gas name gas a b c dle f name 5 character string This command sets the gas name for gas gas to a name 1 5 characters are allowed Send set gas a name abcd Receive s
163. ific screens Enter The is used to select a menu item accept set save a change and or toggle on off functions Soft Keys The Soft Keys are multi functional keys that use part of the display to identify their function at any moment The function of the soft keys allows immediate access to the menu structure and most often used menus and screens They are located directly underneath the display and as the keys functions change this is indicated by user defined labels in the lower part of the display so that the user knows what the keys are to be used for To change a soft key place the menu cursor gt on the item of the selected menu or screen you wish to set Press gt followed by the selected soft key within 1 second of pressing the right arrow key The edit soft key prompt will be displayed for configuration for the new label Thermo Fisher Scientific Model 146 Instruction Manual 3 3 Operation Software Overview Software Overview 3 4 Model 146 Instruction Manual Note Not all menu items may be assigned to soft keys If a particular menu or screen item cannot be assigned the key assignment screen will not come up upon entering right arrow soft key combinations All items under the Service menu including the menu itself cannot be assigned soft keys A The Model 1461 utilizes the menu driven software as illustrated by the flowchart in Figure 3 3 The Power Up screen shown at the top of the flowch
164. inccnnncononccnrnccnrnccnnonos 2 2 Figure 2 8 25 Pin Terminal Board VIBWS oooooococicicocccinnncnccnnncncncnrnrnonocoronoss 2 2 Figure 3 1 Front Panel Display oooooooiococcnoncinnnonoccnoncnococnrnnonororonnononoso 3 4 Figure 3 2 Front Panel PUSADUTLONS o ooiconococnincncocnnnnnncnncnrnononocnrnrnonnnononoso 3 4 Figure 3 3 Flowchart of Menu Driven Software o ooococcicconinconinccninccnnccnonoso 3 4 Figure 3 4 Pin Out of Rear Connector 3 4 Figure 4 1 Photometer Calibration Flow Schematic o onoccnicccnicncooo 4 1 Figure 4 2 Ozonator Control Check Flow Schemaltic o io o o 4 1 Figure 5 1 Inspecting and Cleaning the Fan Fllter ooocooncinoicicninicnccc 5 2 Figure 5 2 Replacing the Ozonator Capillary ooooocioocicinicinincninninccninncoso 5 2 Figure 5 3 Replacing the Pressure Regulator Capillary 5 2 Figure 5 4 Replacing the Pump Capillary o oonoicooninconincininccnnnccnrnicanooo 5 2 Figure 5 5 Rebuilding the PUMP oocooocicinococononoconononononocornnnonorornnnonornrnnonoos 5 2 Figure 6 1 Board Level Connection Diagram Common Electronics 6 1 Figure 6 2 Board Level Connection Diagram Measurement System 6 1 Figure 7 1 Properly Grounded Antistatic Wrist Strap o oococoonic cnic o 7 3 Figure 7 2 Component LayOUt ooococociccccioconinc
165. ing the Service Mode on page 5 2 then return to the beginning of this step 3 Select either Reference Gas or Sample Gas Either the Photometer Reference Gas screen appears or the Photometer Sample Gas screen The noise value displayed after 20 seconds should be below 4 0 Hz for a fully warmed up lamp see Troubleshooting Guides on page 6 2 in the Troubleshooting chapter if the noise is excessive Degradation of detector frequency to below 65 kHz indicates either a dirty cell or low lamp output In addition to degrading the measured detector frequency dirt in the cell can decompose the ozone and give erroneous readings Therefore the cell should first be cleaned and the frequency re measured The Model 146i includes the following capillaries depending on the instrument options Ozonator capillary Photometer capillary Pressure gauge capillary Pump capillary If there are indications that a capillary may be blocked perform the following procedures as appropriate See the associated capillary replacement procedure for a description of how to remove the capillary for inspection and cleaning Model 146 Instruction Manual 5 5 Preventive Maintenance Ozonator Capillary Replacement Ozonator option e Check the ends of the capillary for an obvious obstruction and remove the obstruction e Clear any internal blockage using an appropriate wire for the particular capillary Refer to Table 5 1 e Ifa restri
166. ins 1 20 0 0 0 Gas Off 0 0 1 Gas A 0 1 0 Gas B 0 1 1 Gas C 1 0 0 Gas D Thermo Fisher Scientific Thermo Fisher Scientific Installation Remote Interconnections 1 Gas E 0 Gas F Pins 4 5 and 6 to report the selected gas span level 1 indicates contact closure Span Bit 3 Span Bit 2 Span Bit 1 Span Level Pins 6 25 Pins 5 24 Pins 4 23 0 0 0 Span 0 0 0 Span 1 0 1 0 Span 2 0 1 Span 3 1 0 0 Span 4 1 0 Span 5 1 1 0 Manual Pins 7 8 9 and 10 to report the selected ozonator and permeation oven level 1 indicates contact closure Oz Perm Bit 7 02z Perm Bit 8 Oz Perm 9 Oz Perm Bit10 Ozonator or Pins10 29 Pins 9 28 Pins 8 27 Pins 7 26 Permeation Oven Levell 0 0 0 0 Ozon Perm Off 0 0 0 Ozon Manual 0 0 1 0 Ozon 1 0 0 1 Ozon 2 0 1 0 0 Ozon 3 0 1 0 Ozon 4 0 1 1 0 Ozon 5 0 1 1 Perm 1 1 0 0 0 Perm 2 1 0 0 Perm 3 1 0 1 0 Perm 4 1 0 1 Perm 5 Model 146 Instruction Manual 2 9 Installation Connecting External Devices Connecting External 2 10 Devices Terminal Board PCB Assemblies Model 146 Instruction Manual Pins 12 to 19 are user controllable solenoid outputs The following table lists their default values When the condition is active 24V is provided on the pin Pin 12 Gas A Pin 13 Gas B Pin 14 Gas C Pin 15 Gas D Pin 16 Gas E Pin 17 Gas F Pin 18 Ozonator Pin 19 Perm Oven Several components are available for connecting external devices to iSeries instrument
167. instrument stability Regulates and measures gas flow through the system The processor software tasks are organized into four areas e Instrument Control e Monitoring Signals e Measurement Calculations e Output Communication Low level embedded processors are used to control the various functions on the boards such as analog and digital I O and heater control These processors are controlled over a serial interface with a single high level processor that also controls the front panel user interface The low level processors all run a common piece of firmware that is bundled with the high level firmware and loaded on power up if a different version is detected Each board has a specific address that is used to identify to the firmware what functions are supported on that board This address is also used for the communications between the low level processors and the high level processor Every tenth of a second the frequency counters analog I O and digital I O are read and written to by the low level processor The counters are accumulated over the past second and the analog inputs are averaged over that second The high level processor polls the low level processors once per second to exchange the measurement and control data Model 146 Instruction Manual 8 7 System Description Electronics 8 8 Monitoring Signals Measurement Calculations Output Communication Electronics Model 146 Instruction Manual The instr
168. intenance chapter Check that capillaries are properly installed and O rings are in good shape Replace if necessary Check flow system for leaks See the Servicing chapter Alarm Intensity Defective detector Defective measurement interface board See Cell Frequency High and Cell Frequency Low or Zero in Table 6 1 Replace detector Replace measurement interface board See Cell Frequency High and Cell Frequency Low or Zero in Table 6 1 Alarm Oz Level 1 Check Opt Does not reach the level 1 setting See Table 6 1 for troubleshooting information Oz Level 2 Check Opt Does not reach the level 2 setting See Table 6 1 for troubleshooting information Oz Level 3 Check Opt Does not reach the level 3 setting See Table 6 1 for troubleshooting information Oz Level 4 Check Opt Does not reach the level 4 setting See Table 6 1 for troubleshooting information Alarm 03 Conc Concentration has exceeded range limit Concentration low Check to insure range corresponds with expected value If not select proper range Check user defined low set point set to zero Alarm Motherboard Status Internal cables not connected properly Defective motherboard Check that all internal cables are connected properly Recycle AC power to instrument If still alarming replace board Alarm Interface Status Internal
169. ion and Gas Phase Titration GPT This chapter provides step by step procedures to produce each type of concentration gas and describes the following front panel display front panel pushbuttons and menu driven software For details see the following topics e Display on page 3 1 e Pushbuttons on page 3 2 e Software Overview on page 3 4 e Operation on page 3 8 e Flow Modes on page 3 9 e Gas Setup Menu on page 3 21 e Ozonator Setup Menu on page 3 26 e Permeation Oven Setup Menu on page 3 28 e Photometer Setup Menu on page 3 30 e Program Menu on page 3 36 e Instrument Controls Menu on page 3 39 e Diagnostics Menu on page 3 70 e Alarms Menu on page 3 77 e Service Menu on page 3 84 e Password on page 3 103 The 320 x 240 graphics liquid crystal display LCD shows the sample concentrations instrument parameters instrument controls help and error messages Some menus contain more items than can be displayed at one time For these menus use _ Jand _ 4 to move the cursor up and down to each item Model 146 Instruction Manual 3 1 Operation Pushbuttons Figure 3 1 Front Panel Display your skin or clothes If the liquid crystal contacts your skin or clothes wash it off immediately using soap and water A f CAUTION If the LCD panel breaks do not to let the liquid crystal contact Pushbuttons The Pushbuttons allow the us
170. ion testing adjustment and calibration procedures and component replacement procedures Safety Precautions on page 7 3 Firmware Updates on page 7 4 Replacement Parts List on page 7 5 Cable List on page 7 6 External Device Connection Components on page 7 7 Removing the Measurement Bench and Lowering the Partition Panel on page 7 9 Fuse Replacement on page 7 10 Leak Test and Pump Checkout with Photometer Option on page 7 11 Mass Flow Controller Replacement on page 7 14 Pump Replacement Photometer option on page 7 16 Fan Replacement Rear Panel on page 7 17 Thermo Fisher Scientific Model 146 Instruction Manual 1 1 Servicing Fan Replacement Mass Flow Controllers on page 7 18 Analog Output Testing on page 7 19 Analog Output Calibration on page 7 21 Analog Input Calibration on page 7 22 I O Expansion Board Replacement Optional on page 7 24 Digital Output Board Replacement on page 7 26 Motherboard Replacement on page 7 26 Measurement Interface Board Replacement on page 7 27 Front Panel Board Replacement on page 7 28 LCD Module Replacement on page 7 29 Photometer Lamp Replacement Photometer option on page 7 31 Photometer Lamp Voltage Adjustment Photometer option on page 7 33 Photometer Lamp Heater Replacement on page 7 33 Photometer Replacement Photometer option
171. ir Analog Output Calibrate The Analog Output Calibrate Zero screen is used to calibrate the zero state Zero ofthe selected analog output The operator must connect a meter to the output and adjust the output until it reads 0 0 V on the meter e Inthe Main Menu choose Service gt Analog Out Calibration gt Select Channel gt Calibrate Zero Analog Output Calibrate The Analog Output Calibrate Full Scale screen is used to calibrate the full Full Scale scale state of the selected analog output The operator must connect a meter to the output and adjust output until it reads the value shown in the set output to field e Inthe Main Menu choose Service gt Analog Out Calibration gt Select Channel gt Calibrate Full Scale Thermo Fisher Scientific Model 146 Instruction Manual 3 93 Operation Service Menu Analog Input The Analog Input Calibration menu is a selection of 8 analog inputs if the Calibration I O expansion board option is installed to calibrate and allows the user to select the calibration action zero or span The analog input calibration is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A e Inthe Main Menu choose Service gt Analog Input Calibration y CH pikari ML AA Analog Input Calibrate The Analog Input Cal
172. is required The text field contains statically displayed text The value field displays values which are parsed out of the response to a DATA ERec command It also displays though background changes alarm status The button when pressed triggers input from either a dialog box or a selection list There are five kinds of buttons B I L T and N Each line in the layout string corresponds to one line on the display The layout string describes each of the three major fields as well as translation mechanisms and corresponding commands The first field in the layout string is the text It is delimited by a The string up to the first will be read and inserted in the text field of the line This is followed by a possible string enclosed in quotes that is used to place a string into the value field The value source which is the item or word number in the DATA ERec response appears next This is followed by an optional bitfield designator The datum identified by the value source can be printed as a string s hexadecimal x decimal d or floating point or binary b number Typically bitfield extractions are only done for decimal or hexadecimal numbers Floating point numbers can be followed with an optional precision specifier which will be used as an argument to printf s f format e g a field of 4 would be translated into the printf command of 3f Alternately the special character causes an indirection
173. istatic wrist strap must be worn while handling any internal component a Equipment Damage Some internal components can be damaged by small 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unscrew the knurled fitting separate the components and remove the capillary Figure 5 3 Thermo Fisher Scientific Model 146 Instruction Manual 5 7 Preventive Maintenance Pump Capillary Replacement Photometer option Pump Capillary Replacement Photometer option A 5 8 Model 146 Instruction Manual 3 Insert the new capillary and reconnect the fittings Vir Ul Figure 5 3 Replacing the Pressure Regulator Capillary For instruments with the photometer option use the following procedure to replace the pump capillary Figure 5 4 Equipment Required Pump capillary Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Loosen the fitting connecting the capillary to the pump and remove the capillary Figure 5 4 Thermo Fisher Scientific Preventive Maintenance Pump Rebuilding 3 Insert the new capillary and tighten the fitting Capillary Figure 5 4 Replacing the Pump Capillary Pump Rebuilding Use the following procedure to rebuild the pump Figure 5 5 To repla
174. l internal ozonator The production of ozone is regulated by either changing the flow of zero air or by changing the intensity of the ozone producing lamp The Ozonator Setup menu sets the amount of zero air flow From this menu the ozone generation can be set manually or up to five ozone levels can be set e Inthe Main Menu choose Ozonator Setup ip gies ML PRA The Zero Air screen sets the amount of zero air flow through the ozonator e Inthe Main Menu choose Ozonator Setup gt Zero Air Thermo Fisher Scientific Operation Ozonator Setup Menu Pri ii Level 1 5 Five ozonator levels can be set using the Ozonator Level screen This screen is allows the user to change the custom ozonator levels e In the Main Menu choose Ozonator Setup gt Level 1 5 Manual 0 Lamp The ozonator can be certified for use as a transfer standard To determine a Setting calibration curve an external ozone primary standard such as the Model 49i Primary Standard must be used Due to changes in light intensity flow and ambient temperature the calibration curve is accurate only for a short time after a transfer standard reading is taken Since the setting is not remembered once the instrument is powered down the calibration curve is not retained and must be regenerated on instrument start up This ensures that the calibration is accurate The Ozonator Manual Level screen is used when certifying the ozone generator as a transfe
175. lacing the Glass Chamber Assembly h Tighten knurled screw with finger Do not use tools to tighten i Replace oven cover being careful to place tubing and wire in slot of cover 4 Connect the EXHAUST bulkhead to a suitable vent The exhaust line should be 1 4 inch OD with a minimum ID of 1 8 inch The length of the exhaust line should be less than 10 feet Verify that there is no restriction in this line 5 Connect a suitable recording device to the rear panel connector For detailed information about connecting to the instrument refer to Connecting External Devices on page 2 10 External Device Connection Components on page 7 7 Terminal Block and Cable Kits on page 9 11 Instrument Controls Menu in the Operation chapter For detailed information about troubleshooting a connection refer to Analog Output Testing on page 7 19 Thermo Fisher Scientific Model 146 Instruction Manual 2 5 Installation Remote Interconnections 6 Plug the instrument into an outlet of the appropriate voltage and frequency Under no circumstances should this grounding system be defeated WARNING The Model 1467 is supplied with a three wire grounding cord ra ula E U a 006666 z 2 100000 000000000000 Y 0000000000000 POWER FAII RELAY DIGITAL DIGITAL INPUTS OUTPUTS 000000000000000000 0000000000000000000 ANALOG VOLTAGE OUTPUTS e
176. lastic tubing Model 146 Instruction Manual 9 1 Optional Equipment Permeation Oven Permeation Oven Since the temperature of the permeation tube oven should be stable the gas flow through the oven is small approximately 150 sccm The major portion of the flow bypasses the oven and is directed to the mixing chamber with the stream from the oven This is achieved by using the same backpressure regulator described above for gas phase titration with a capillary of identical size During standby operation a continuous flow of gas is necessary through the oven The standby flow and operating flow should be the same to 10 percent This is achieved by setting the pressure regulator to the same pressure setting as the backpressure regulator In practice rather than measuring the pressure it is the actual flows that are measured while adjusting the pressure regulator Two solenoids under control of the microcomputer are used to change the flow of gas going through the oven to the main stream or to vent Permeation Tube Use the following procedure to install the permeation tube Installation Remove the oven cover 2 Remove the glass chamber assembly by loosening the white plastic retaining collar loosening not removing the knurled screw and gently pulling the assembly upward Completely remove the oven 3 Separate the glass chamber from the top assembly by twisting and gently pulling the glass away from the top Note Keep th
177. lations The second pressure is the actual pressure reading being measured If pressure compensation is on then both pressure readings are the same If pressure compensation is off a pressure of 760 mmHg is used as default pressure even though the actual pressure is 753 4 mmHg The example below shows that the actual reaction chamber pressure is 753 4 mmHg This command is only available if the photometer option is installed Send pres Receive pres 753 4 mmHg actual 753 4 status This command returns the gas and span status of the Model 1467 Send status Receive status remote gas A span 4 flags This reports 8 hexadecimal digits or flags that represent the status of the pressure and temperature compensation status flow mode and alarms To decode the flags each hexadecimal digit is converted to binary as shown in the Figure B 1 It is the binary digits that define the status of each parameter In the example below the instrument is reporting that it is in span 1 with no gases currently selected Send flags Receive flags 00008000 Model 146 Instruction Manual B 9 C Link Protocol Commands Alarms Alarms B 10 Model 1467 Instruction Manual Figure B 1 Flags alarm bench lamp temp min alarm bench lamp temp max These commands report the photometer bench lamp temperature alarm minimum and maximum value current settings The example below reports the bench lamp temperature alarm minimum value is 52 C These commands
178. ll as remote control to other devices such as for controlling valves during calibration The user may select what information is sent out each relay and whether the active state is opened or closed Sixteen digital inputs are available which may be programmed to signal instrument modes and special conditions including e Zero Gas Mode e Span Gas Mode The actual use of these inputs will vary based on instrument configuration The digital inputs are TTL level compatible and are pulled up within the analyzer The active state can be user defined in firmware Two serial ports allow daisy chaining so that multiple analyzers may be linked using one PC serial port The standard bi directional serial interface can be configured for either RS 232 or RS 485 The serial baud rate is user selectable in firmware for standard speeds from 1200 to 19 200 baud The user can also set the data bits parity and stop bits The following protocols are supported e C Link e Streaming Data e Modbus Slave The Streaming Data protocol transmits user selected measurement data via the serial port in real time for capture by a serial printer data logger or PC Model 146 Instruction Manual 8 13 System Description 1 0 Components 8 14 RS 232 Connection RS 485 Connection Model 146 Instruction Manual A null modem crossed cable is required when connecting the analyzer to an IBM Compatible PC However a straight cable one to one may be required
179. log Output Zero to Full Scale Table oo o o 3 4 Table 3 4 Signal Type Group ChOICBS ocoooooccinincinnoninnnnonccnrncnncncnornrnnonrnnnoos 3 4 Table 5 1 Capillary Cleaning Wire Sizes oooocicoocicioninconincnnnncononccnrnccnrnocnnoos 5 2 Table 6 1 Troubleshooting General Guide ooioocionicionincinnncnnnccnrnccnnnos 6 1 Table 6 2 Troubleshooting Alarm Messages oococonooociicococnnnnncnncnrnnonoooos 6 1 Table 6 3 Motherboard Connector Pin Descripti0NS oo cnc icio 6 1 Table 6 4 Front Panel Board Connector Pin Diagram o oniccinicio oo 6 1 Table 6 5 O Expansion Board Optional Connector Pin Descriptions 6 1 Table 6 6 Digital Output Board Connector Pin Descriptions 6 1 Table 6 7 Measurement Interface Board Connector Pin Descriptions 6 1 Table 6 8 Solenoid Connector Board Pin Descriptions 6 1 Table 6 9 Ozonator Power Supply Connector Pin Descriptions 6 1 Table 6 10 Photometer Power Supply Connector Pin Descriptions 6 1 Table 7 1 Model 1467 Replacement Parts ccc cecesesescetetesesteteeseeeeee 7 3 Table 7 2 Model 1467 Cables ccccccccesseseesseseesecesteseeseeseseeseeeeesneseenee 7 3 Table 7 3 External Device Connection Component cocoocicocncoconinccninccninccnnoss 7 3 Table 7 4 Analog Output Cha
180. ltage value is separated by a space Send diag volt mib Receive diag volt mib 24 1 14 9 14 9 4 9 3 2 diag volt iob This command reports the diagnostic voltage measurements on the I O expansion board The sequence of voltages is Positive 24 positive 5 positive 3 3 and negative 3 3 Each voltage value is separated by a space Send diag volt iob Receive diag volt iob 24 1 4 9 3 2 3 2 event event params event 1 32 This command reports the parameters programmed for the specified event event Send event 1 params Receive event 1 params gas span 1 ozon 3 perm 1 photo internal event status This command reports the current event that is being executed 1 32 A 0 response indicates no event is currently running Send event status Receive event status 1 Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Diagnostics intern temp This command reports the current internal instrument temperature in degrees C The example below shows that temperature compensation is on and that the internal temperature is 27 2 C Send intern temp Receive intern temp 27 2 deg C lamp temp This command reports the photometer bench lamp temperature The example below reports the current bench lamp temperature is 55 2 C This command is only available if the photometer option is installed Send lamp temp Receive lamp temp 55 2 deg C lamp voltage bench This command reports the photometer bench
181. lue is a floating point representation of the coefficient The example below sets the ozonator response coefficient to 1 500 This command is only available if the photometer option is installed Send set resp coef 1 500 Receive set resp coef 1 500 ok sp conc This command reports photometer span concentration If the mode is incorrect the instrument responds with can t wrong settings The example below reports the span gas concentration of 50 ppb This command is only available if the photometer option is installed Send sp conc Receive sp conc 5 000E 01 ppb set sp conc value This command sets the span concentrations to user defined values to value where value is a floating point representation of the span concentration in ppb The example below sets the span concentration to 50 ppb This command is only available if the photometer option is installed Send set sp conc 50 Receive set sp conc 50 ok push button do down en enter he help le left me Model 146 Instruction Manual B 25 C Link Protocol Commands Keys Display B 26 Model 146 Instruction Manual menu ri right button do down en enter he help le left me menu ri right ru run up 1 2 3 4 These commands simulate pressing the front panel pushbuttons The numbers represent the front panel soft keys from left to right Send push enter Receive push enter ok isc iscreen This command retrieves the frame
182. m can be specified as a range e Inthe Main Menu choose Photometer Setup gt Range gt Set Custom Ranges Custom Ranges The Custom Ranges screen is used to define the custom ranges The display shows the current custom range The next line of the display is used to set the range To use the custom full scale range be sure to select it Custom range 1 2 or 3 in the O3 Ranges screen For more information about selecting ranges see O3 Range above e Inthe Main Menu choose Photometer Setup gt Range gt Set Custom Ranges gt Custom range 1 2 or 3 Averaging Time The Averaging Time defines a time period 60 to 300 seconds during which Oz measurement is taken The average concentration of the reading is calculated for that time period The front panel display and analog 3 32 Model 146 Instruction Manual Thermo Fisher Scientific Calibration Factors Thermo Fisher Scientific 0 Background Operation Photometer Setup Menu outputs are updated every 5 seconds regardless of the averaging time An averaging time of 300 seconds means that the moving average concentration of the last 300 seconds will be output every 5 seconds The lower the averaging time the faster the front panel display and analog outputs respond to concentration changes Longer averaging times are typically used to smooth output data The following averaging times are available 60 90 120 180 240 and 300 seconds e Inthe Main Menu choo
183. menu in the Service menu select Perm Oven Selection then select 45 C Thermo Fisher Scientific Model 146 Instruction Manual 9 3 Optional Equipment Permeation Oven 9 4 Model 146 Instruction Manual 10 11 12 13 14 Remove the thermistor from POJ1 on the measurement interface board Connect a resistance of approximately 3 8 KQ across pins 1 and 2 of POJ1 Go to the Factory Cal Gas Therm menu from the Perm Oven Settings menu in the Service menu Select Low Point enter the exact value of the resistor and press to calibrate the low resistance point Press to return to the Factory Cal Gas Therm menu Connect a resistance of approximately 5 KQ across pins 1 and 2 of POJ1 Go to the High Point screen enter the exact value of the resistor and press to calibrate the high resistance point Press to return to the Factory Cal Gas Therm menu Remove resistor from POJ1 and re attach the gas thermistor Connect a resistance of approximately 3 8 KQ across pins 3 and 4 of POJ3 Go to the Factory Cal Oven Therm menu from the Perm Oven Settings menu in the Service menu Select Low Point enter the exact value of the resistor and press to calibrate the low resistance point Press j to return to the Factory Cal Oven Therm menu Connect a resistance of approximately 5 KQ across pins 3 and 4 of POJ3 Thermo Fisher Scientific Permeation Tube Oven Calibration Setting Perm Oven Temperat
184. meter Lamp Voltage Adjustment Photometer option Photometer Lamp Heater Replacement Thermo Fisher Scientific A Servicing Photometer Lamp Voltage Adjustment Photometer option Use the following procedure to adjust the photometer lamp intensity to approximately 100 000 Hz Note After turning power ON wait for the lamp to stabilize approximately 15 minutes before beginning the following lamp voltage adjustment procedure 1 From the Main Menu press to scroll to Service gt press gt P P to scroll to Lamp Setting gt and press Ge The Photometer Lamp Setting screen appears Note If Service is not displayed refer to Accessing the Service Mode on page 7 11 then return to the beginning of this step 2 At the Photometer Lamp Setting screen use EIES to increment decrement the Lamp Setting until Intensity indicates approximately 100 000 Hz 3 Press to save the setting Use the following procedure to replace the photometer lamp heater Figure 7 12 Equipment required Photometer lamp heater Allen wrench 7 64 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Turn instrument OFE unplug the power cord and remove the cover 2 Disconnect the plumbing and electrical connections Model 146 Instruction Manual 7 33 Servicing P
185. mory The example below shows that 50 long records have been stored in the memory Send no of lrec Receive no of lrec 50 recs Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Datalogging malloc Irec malloc srec These commands report the currently set memory allocation for long and short records in percent of total memory Send malloc lrec Receive malloc lrec 10 set malloc lrec value set malloc srec value value 0 to 100 These commands set the percent of memory space allocated for long and short records to value where value is a floating point number representing percent The example below sets the memory allocation for long records to 10 Note Issuing these commands will clear all the logging data memory All the existing records should be retrieved using appropriate commands if required A Send set malloc lrec 10 Receive set malloc lrec 10 ok set copy sp to lrec set copy sp to srec set copy sp to stream These commands copy the current selections in scratch pad sp into the long record short record or streaming data list The example below copies the current list in scratch pad into the long records list Send set copy sp to lrec Receive set copy sp to lrec ok set copy lrec to sp set copy srec to sp set copy stream to sp These commands copy the current contents of the long record short record or streaming data list into the scratch pad sp These commands are u
186. mp Temperatute ata aa 3 3 Permeation Oven Temperature idos 3 3 Presa 3 3 Photometer Bench Temperature at 3 3 Photometer Lamp Temperature ii 3 3 Photonieter IAtensityoisi Si 3 3 Service Menusi e o a a Ea a a T AE G 3 3 Flow Control Selection visionar 3 3 Calibration Pressure Temperature uaiacino cesta dios 3 3 Zero Flow Cabral ks Beaetard ieia 3 3 Gas Flow Calibre dede 3 3 Pressure Calibrador ari cae 3 3 Ambient Temperature Calibrationas s canina tds de 3 3 Bench Temperature Calibracion arti 3 3 Analog Output Calibration scasuscesssaecdsineeceediancenonviansueditancuderatecedass 3 3 Analog Input Calibration ar irc 3 3 Permeation yen Sere 3 3 Lamp Seti O E OE 3 3 Intensiy Cheto dile 3 3 O Response Cocfticintiuiinanina id 3 3 Display Pizel Destinia clasicos 3 3 Restore User Deli qe O sad ici 3 3 LS A A A on Ge cad at 2 3 3 EU PassWord aiee sel ae tea ad ds ed 3 4 Lock Instruir taa 3 4 Change Password sisi da uta ade 3 4 Remove Password via 3 4 Unlock A a ae ead ih cata ca e 3 4 oD ee ee re 4 1 Mass Flow Controllers aiiis sisii e teats cancel 4 1 Photometer Ca a cu ah eee 4 1 Calibration Using a Model 49i Pipi ay cae Sates 4 1 Ozenator Control seat aaa edo ee eE 4 1 Preventive Maintenance iaiionacniccainia static 5 1 Safety Precautions id 5 1 Replacement Paraiso a a a S E 5 1 Accessing the Service Mode untada idad 5 1 Outside Case Cleaning ssenarist isani aanika isi aisin 5 1 Fan Filter Inspection and Cleaning ra oa 5 1 Photometer
187. mp drive 12 Reports sets custom level 2 setting of ozonator lamp drive B 32 12 conc Reports sets custom level 2 concentration for photometer B 32 control of ozonator lamp drive 13 Reports sets custom level 3 setting of ozonator lamp drive B 32 13 conc Reports sets custom level 3 concentration for photometer B 32 control of ozonator lamp drive 14 Reports sets custom level 4 setting of ozonator lamp drive B 32 14 conc Reports sets custom level 4 concentration for photometer B 32 control of ozonator lamp drive 15 Reports sets custom level 5 setting of ozonator lamp drive B 32 15 conc Reports sets custom level 5 concentration for photometer B 32 control of ozonator lamp drive lamp Reports sets photometer lamp drive setting B 23 lamp setting Reports photometer lamp drive setting B 24 lamp temp Reports photometer lamp temperature B 15 lamp voltage bench Reports photometer bench lamp voltage in volts B 15 lamp voltage 03 Reports ozonator lamp voltage in volts B 15 Model 146 Instruction Manual B 5 C Link Protocol Commands Commands Command Description Page layout ack Disables CLINK stale layout layout changed indicator B 43 le left Simulates pressing left pushbutton B 25 level Sets the ozonator to level 1 2 3 4 or 5 B 33 list din Lists current selection for digital input B 16 list dout Lists current selection for digital output B
188. mp intensity unstable Perform leak test see Servicing chapter Replace bench thermistor Recalibrate pressure transducer Clean cells and flow components Replace solenoid Replace detector or photometer board Slow response Averaging time Leak Contaminated optical bench Table 6 2 Troubleshooting Alarm Messages Alarm Message Alarm 03 Lamp Temp Possible Cause Defective ozonator board Defective ozonator lamp heater Verify averaging time is set properly Perform leak test see Servicing chapter Clean bench and then condition system overnight Action Replace ozonator board Replace ozonator lamp heater Alarm Lamp Temp Defective photometer board Defective photometer lamp heater Replace photometer board Replace photometer lamp heater Alarm Bench Temp Defective fan Check fan filter Check for 24 VDC on fan connector and replace fan if 24 VDC is applied to fan Clean or replace foam filter see Preventive Maintenance chapter Thermo Fisher Scientific Model 146 Instruction Manual 6 5 Troubleshooting Troubleshooting Guides 6 6 Model 146 Instruction Manual Table 6 2 Troubleshooting Alarm Messages continued Alarm Message Alarm Pressure Possible Cause High pressure indication Action Check the pump for a tear in the diaphragm replace with pump repair kit if necessary Refer to Preventive Ma
189. ms Thermo Fisher Scientific Model 146 Instruction Manual 3 77 Operation Alarms Menu Ambient Temperature Min and Max Ambient Temperature Limits 3 78 Model 146 Instruction Manual The Ambient Temperature screen displays the current ambient temperature and sets the minimum and maximum alarm limits Acceptable alarm limits range from 8 to 47 C If the ambient temperature reading goes beyond either the minimum or maximum alarm limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu e Inthe Main Menu choose Alarms gt Ambient Temp ri Pik The Minimum Ambient Temperature alarm limit screen is used to change the minimum ambient temperature alarm limit The minimum and maximum ambient temperature screens function the same way e Inthe Main Menu choose Alarms gt Ambient Temp gt Min or Max Thermo Fisher Scientific Operation Alarms Menu O3 Lamp The O Lamp Temperature screen displays the current ozonator lamp Temperature temperature and sets the minimum and maximum alarm limits Acceptable alarm limits range from 55 to 85 C If the O lamp temperature reading goes beyond either the minimum or maximum alarm limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu e Inthe Main Menu choose Alarms gt 03 Lamp Temp Min and Max 0 Lamp The Minimum O Lamp Temperature alarm limit screen is used to change Temperat
190. n Manual Send analog vin 1 Receive analog vin 1 75 325 2 796 analog vout range channel This command reports the analog voltage output channel range where channel is between 1 and 6 inclusive according to Table B 12 Send analog vout range 2 Receive analog vout range 2 3 set analog vout range channel range This command sets analog voltage output channel to the range where channel is between 1 and 6 inclusive and range is set according to Table B 12 The example below sets channel 2 to the 0 10 V range Send set analog vout range 2 3 Receive set analog vout range 2 3 ok Table B 12 Analog Voltage Output Range Values Range Output Range 1 0 1V 2 0 100 mV 3 0 10 V 4 0 5 V 0 cannot be set to this but may report Undefined dig in This command reports the status of the digital inputs as a 4 digit hexadecimal string with the most significant bit MSB being input 16 Send dig in Receive dig in oxff7f din channel This command reports the action assigned to input channel and the corresponding active state The example below reports the input 1 to be assigned an index number 3 corresponding to action of set background mode with the active state being high Send din 1 Receive din 1 3 SET BACKGROUND high set din channel index state Thermo Fisher Scientific C Link Protocol Commands 1 0 Configuration This command assigns digital input channel 1 16 to activate the action indicate
191. n is installed Send range Receive range 6 500E 03 ppb set range selection This command selects the O full scale ranges according to Table B 8 The example below sets the O full scale range to 2 0 ppm Send set range 5 Receive set range 5 ok Table B 8 Standard Ranges Selection ppm 0 0 05 0 1 0 2 0 5 1 0 2 0 5 0 C1 C2 C3 gt OoO ao r was asi A I wt rp Model 146 Instruction Manual B 35 C Link Protocol Commands Measurement Configuration B 36 Model 1467 Instruction Manual custom range range 1 2 3 This command reports the user defined value of custom range 1 2 or 3 The example below reports that custom range 1 is defined to 4 50 ppm This command is only available if the photometer option is installed Send custom 1 Receive custom 1 4 500E 03 ppb set custom range range value set custom 1 range value set custom 2 range value set custom 3 range value These commands are used to set the maximum concentration for any of the three custom ranges 1 2 or 3 to range value where value is a floating point number representing concentration in ppm The example below sets the custom 1 range to 4 5 ppm Send set custom 1 range 4 5 Receive set custom 1 range 4 5 ok pres comp This command reports whether pressure compensation is on or off The example below shows that pressure compensation is on This command is only available if the photometer option is install
192. n the photometer is cleaned prior to recalibration The cleanliness of the photometer should also be checked any time the detector frequency drops below 65 kHz since one source of low output is light attenuation due to dirt in the cell Dirt particulates are usually effective ozone removers Use the following procedure to clean the photometer Figure 7 12 static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component For more information about CAUTION Some internal components can be damaged by small amounts of appropriate safety precautions see the Servicing chapter 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the plumbing and electrical connections Thermo Fisher Scientific Model 146 Instruction Manual 5 3 Preventive Maintenance Monitoring Detector Frequency and Noise Monitoring Detector Frequency and Noise 5 4 Model 146 Instruction Manual 3 Loosen the Allen screws 4 securing the photometer to the floor plate Figure 7 12 and remove the photometer 4 Remove four 4 mounting screws on the lamp heater and remove the heater 5 Remove the lamp housing Figure 7 12 and the associated components 6 Remove the detector Refer to Detector Replacement Photometer option on page 7 35 7 Remove the detector housing Figure 7 12 and the associated components 8 Push a piece of lens paper down the tube
193. nalog Voltage Input 1 Analog Voltage Input 2 Analog Voltage Input 3 Ground on Ae N Analog Voltage Input 4 Thermo Fisher Scientific Model 146 Instruction Manual 6 17 Troubleshooting Connector Pin Descriptions Table 6 5 I O Expansion Board Optional Connector Pin Descriptions continued Connector Label Reference Designator Pin 0 0 nN 0 gt as gt gt gt gt gt gt gt co oe N o a gt Se N N N N N N N on F 0 N gt O Signal Description Analog Voltage Input 5 Analog Voltage Input 6 Ground Analog Voltage Input 7 Analog Voltage Input 8 Ground NC NC Ground Current Output 1 Current Output Return Current Output 2 Current Output Return Current Output 3 Current Output Return Current Output 4 Current Output Return Current Output 5 Current Output Return Current Output 6 MOTHER BD 6 18 Model 146 Instruction Manual J2 mo N OQO o A WO N 5V 24V 24V Ground Ground Ground RS485 to Motherboard RS485 to Motherboard Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Table 6 6 Digital Output Board Connector Pin Descriptions Connector Reference Pin Label Designator MOTHER BD J1 Signal Description 5V 24V 24V Ground Ground Ground SPI Reset SPI Input SPI Output SPI Board Select SPI Clock 2 3 4 5 6 7 8 9 10 11 DIGITAL J2 1 OUTPUTS Z 3 4 5 6 7 8 9 10 11 1
194. nalyzer to put it in the remote mode Report commands commands that don t begin with set can be issued either in the remote or local mode For information on changing modes see Chapter 3 Operation The commands can be sent in either uppercase or lowercase characters Each command must begin with the proper instrument identification number ASCII character The command in the example below begins with the ASCII character code 179 decimal which directs the command to the Model 146i and is terminated by a carriage return CR ASCII character code 13 decimal If an incorrect command is sent a bad command message will be received The example below sends the incorrect command set unit ppm instead of the correct command set gas unit ppm Send set unit ppm Receive set unit ppm bad cmd The save and set save params commands stores parameters in FLASH It is important that each time instrument parameters are changed that this command be sent If changes are not saved they will be lost in the event of a power failure Table B 1 lists the 1467 C Link protocol commands The interface will respond to the command strings outlined below Table B 1 C Link Protocol Commands Command Description Page addr dns Reports sets domain name server address for Ethernet port B 39 addr gw Reports sets default gateway address for Ethernet port B 39 addr ip Reports sets IP address for Ethernet
195. nator straight up 5 Remove the ozonator flange by removing three screws securing the bottom flange and pulling the flange off 6 Remove the ozonator lamp heater block from the ozonator by removing four Allen screws 7 Coat new lamp heater block with thin film of heat conductive compound 8 Install new ozonator lamp heater block by following the above procedure in reverse Model 146 Instruction Manual 7 43 Servicing Ozonator Replacement Ozonator option 9 Re install the instrument cover Ozonator Use the following procedure to replace the ozonator Figure 7 13 Replacement Ozonator option Esuipmen Required Wrenches 7 16 and 9 16 inch Philips screwdriver amounts of static electricity A properly grounded antistatic wrist strap must Equipment Damage Some internal components can be damaged by small be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unplug heater cable from the LAMP HEATER and LAMP connectors on the ozonator board Figure 7 10 3 Disconnect plumbing from the ozonator 4 Remove the ozonator by removing the two screws securing the ozonator to the floor plate and lifting the ozonator straight up 5 Install new ozonator by following the above procedure in reverse Ensure all fittings are tight 6 Perform a leak test Refer to Leak Test and Pump Checkout in this chapter 7 44 Model 146 Instruction Manu
196. ncentration from the ratio of zero and sample gas intensities and outputs the concentration to the front panel display the analog outputs and also makes the data available over the serial or Ethernet connection Model 146 Instruction Manual 1 3 Introduction Principle of Operation REAR PANEL NC GAS F e NC GAS E e NC GAS D e NC REAR PANEL GAS Ce GAS MASS FLOW CONTROLLER NC GAS B e OUTPUT MANIFOLD NC MIXING E CAS A e CHAMBER wm TO ANALYZER ZERO AIR DUMP MASS FLOW CONTROLLER NC ZERO AIR Figure 1 1 Standard Gas Dilution System REAR PANEL NC CAS Ce GAS MASS FLOW CONTROLLER NC GAS B e lt e NE REAR PANEL GAS A e gt lt ozone E GENERATOR A CAPILLARY NO MING HOLDER E c NC CHAMBER TO ANALYZER lt gt REACTION DUMP CHAMBER ZERO AIR MASS FLOW CONTROLLER NC ZERO AIR gt 1 BACK PRESSURE REGULATOR Figure 1 2 Standard Gas Dilution System with Ozonator Option 1 4 Model 1461 Instruction Manual Thermo Fisher Scientific BACK PRESSURE REGULATOR MIXING CHAMBER PERMEATION TUBE OVEN NC PERMEATION TUBE Introduction Principle of Operation REAR OUTPUT MANIFOLD PANEL m TO ANALYZER DUMP e EXHAUST Figure 1 3 Standard Gas Dilution System with Permeation Tube Oven Option REAR PANEL NC GAS C lt GAS MASS FLOW CONTROLLER NC GAS B 6 gt NC CAS A e gt ZERO AIR MASS FLOW CONTROLL
197. nd is only available if the photometer option is installed Send avg time Receive avg time 5 300 sec set avg time selection This command sets the averaging time according to Table B 2 The example below sets the averaging time to 120 seconds This command is only available if the photometer option is installed Send set avg time 2 Receive set avg time 2 ok Table B 2 Averaging Times Selection Averaging Time seconds 0 60 seconds 1 90 2 120 3 180 4 240 5 300 gas conc This command reports the current concentration of gas being supplied by the Model146i Send gas conc Receive gas conc actual 100 000 target 1 600 ppm gflow zflow These commands report the current zero air or gas flow Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurements Send gflow Receive gflow actual 15 68 target 16 00 sccm intensity This command reports the photometer detector intensity This command is only available if the photometer option is installed Send intensity Receive intensity 101286 Hz 03 This command reports the O3 concentration measured by the photometer The example below shows that the O3 concentration is 5 ppm This command is only available if the photometer option is installed Send 03 Receive 03 5 000E 00 ppb pres This command reports the photometer bench pressure The first pressure reading is the pressure reading being used in instrument calcu
198. ng the analog output always calibrate zero first and then calibrate full scale a 5 Use as until the meter reads the value shown in the Set Output To line then press to save the value 6 Press Cm to return to the previous screen 7 Press to select Calibrate Full Scale 8 Use e until the meter reads the value shown in the Set Output To line then press to save the value Use the following procedures to calibrate the analog inputs after replacing the optional I O expansion board These procedures include selecting analog input channels calibrating them to zero volts and then calibrating them to full scale using a known voltage source Use the following procedure to calibrate the input channels to zero volts 1 From the Main Menu press to scroll to Service press gt 4 to scroll to Analog Input Calibration and press e The Analog Input Cal screen appears Note If Service is not displayed refer to Accessing the Service Mode on page 7 11 then return to the beginning of this step Thermo Fisher Scientific Calibrating the Input Channels to Full Scale Thermo Fisher Scientific Servicing Analog Input Calibration 2 At the Analog Input Cal screen press to scroll to a channel and press 3 With the cursor at Calibrate Zero press The screen displays the input voltage for the selected channel 4 Make sure that nothing is connected to the channel input pins
199. nges while maintaining a minimum resolution of 11 bits e 0 20 mA e 4 20 mA The user can calibrate each analog output zero and span point through firmware At least 5 of full scale over and under range are also supported The analog outputs may be assigned to any measurement or diagnostic channel with a user defined range in the units of the selected parameter The current outputs are independent of the voltage outputs The current outputs are isolated from the instrument power and ground but they share a common return line Isolated GND The optional I O expansion board includes eight analog voltage inputs These inputs are used to gather measurement data from third party devices such as meteorological equipment The user may assign a label unit and a voltage to user defined unit conversion table up to 16 points All voltage inputs have a resolution of 12 bits over the range of 0 to 10 volts Thermo Fisher Scientific Digital Relay Outputs Thermo Fisher Scientific Digital Inputs Serial Ports System Description 1 0 Components The instrument includes one power fail relay on motherboard and ten digital output relays on the digital output board These are reed relays rated for at least 500 mA 200VDC The power fail relay is Form C both normally opened and normally closed contacts All other relays are Form A normally opened contacts and are used to provide alarm status and mode information from the analyzer as we
200. nnels and Rear Panel Pin Connections 7 4 Table 7 5 Analog Input Channels and Rear Panel Pin Connections 7 4 Table 8 1 RS 232 DB Connector Pin Configurati0N o oocinooninccniccnnncono 8 3 Table 8 2 RS 485 DB Connector Pin Configuration cece 8 3 BA elf ae Eo a O16 2510101 6 9 Sane emanate ater te nee te Et Peet eee eet eee 9 2 Table 9 2 Color Codes for 25 Pin and 37 Pin Cables 9 2 Table 9 3 Mounting Options sf scoters eco ho eel a Su ere 9 2 Table B 1 C Link Protocol Commands ccc eeseesectestestesesseeteeteeteseesee B 2 Table B 2 Averaging TIMES 0 cccceccccccecescssecetescsescesesesesceesesesesesteesesesteeeees B 2 Table B 3 Alarm Trigger Values occ cccceccccccececscecetesesesceteseseecetesseseseeteneness B 2 Table B 4 Record Output Formats o cocoocicoonincocinconincnonncnconononocnnnnrnrnncnconrnnnoos B 2 Table B 5 Stream Time ValueS ooooooicicicocononiocinconnooornornocnocnnnornonornornooos B 2 Table B 6 Gas Index Values ooooocococnicicinonocococoncnnonoonocorooroornonnrnornoroornooos B 2 Table B 7 Solenoid Number Values oocococicicicicononincnocinoonnoonornornocnoooos B 2 Model 1467 Instruction Manual xv Tables xvi Model 146 Instruction Manual Table B 8 Standard Ranges costat tia B 2 Table BL Contrast bevels nate cite ae ea ceases B 2 Table B 10 Reply Termination Formats oooociciciccncncnncncncn
201. nu displays a list of the 10 analog output relays available and allows the user to select the logic state or instrument parameter for the relay selected e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings Model 146 Instruction Manual 3 55 Operation Instrument Controls Menu Logic State The Logic State screen is used to change the I O relay to either normally open or normally closed e Press to toggle and set the logic state open or closed Instrument State The Instrument State menu allows the user to select the instrument state that is tied to the selected relay output A submenu lists signal types of either alarm or non alarm to choose from e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt Select Relay gt Instrument State Fusco EH Guided ML PEA 3 56 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Operation Instrument Controls Menu Alarms The Alarms status screen allows the user to select the alarm status for the selected relay output The selected item is shown by lt after it e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt Select Relay gt Instrument State gt Alarms Bi E Pin Non Alarm The Non Alarm status screen allows the user to select the non alarm status for the selected relay out
202. nufacturer the protection provided by the equipment may be impaired Model 146 Instruction Manual 7 3 Servicing Firmware Updates 7 4 Firmware Updates Model 146 Instruction Manual Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Figure 7 1 If an antistatic wrist strap is not available be sure to touch a grounded metal object before touching any internal components When the instrument is unplugged the chassis is not at earth ground Handle all printed circuit boards by the edges Do not remove the LCD panel or frame from the LCD module The polarizing plate is very fragile handle it carefully a Do not wipe the polarizing plate with a dry cloth it may easily scratch the plate Do not use alcohol acetone MEK or other Ketone based or aromatic solvents to clean the module use a soft cloth moistened with a naphtha cleaning solvent a Do not place the module near organic solvents or corrosive gases Do not shake or jolt the module Figure 7 1 Properly Grounded Antistatic Wrist Strap The firmware can be updated by the user in the field via the serial port or over the Ethernet This includes both the main processor firmware and the firmware in all low level processors Refer to the i Port manual for the firmware update procedure Thermo Fisher Scienti
203. o 3 1 Manual Plow ori 3 1 Ozonator Setup Men ciana e ia i i 3 1 LEC I EEE E EEE E EE ET 3 1 Eevee A O OS 3 1 Manual O Lamp Setting li risa 3 1 Permeation Oven Setup Menu sii iio 3 2 P rimeation A no o a e a iiai Aa 3 2 Permeation Rates oe 3 2 O 3 2 Photo meter Setup Went ia ado siii 3 2 RA SA Na 3 2 Averaging PIE eanne t sstaunndansenidesates ds a 3 2 Calibration Factors citada td 3 2 Calibration meae a Ta E a E gee Ate 3 2 Program Menta its 3 2 Programi Gycl a o o ia 3 2 Period Hours ici 3 2 Net Creen nao 3 2 A E E E EE AE apoantacaut Gade coutvseenueay duos 3 2 Instrument Controls Meri es Ge eeu adc nest As 3 2 Temperature Compensation sins ii dd 3 2 Pressure ComipensaniOn pci iia 3 2 Datalogging Settings iii ratico 3 2 Communication E ptas 3 2 VO Configuration cana ais 3 2 Screen Contrast alii io 3 2 Seryice Mode it ean ety A ORF eter tO N eer 3 2 Dar Tinie ir A A AS 3 2 Diagnostics Men 3 2 Programi VEO A a 3 2 Volare la 3 2 El N leo dais ero tasado aca 3 2 TE PERATURES pst A tol R 3 2 PESTE A A A AA 3 2 Analog Input Readings viii is 3 2 Analog Input Volta ces sai ia 3 2 Digital In putssieiciiciesesvecsereisecsini ace o 3 2 A 3 2 Solenoid States oil 3 2 Test Aina Om Qutp ts Si e ae e 3 3 Instrument Contracts 3 3 Contact Information cid 3 3 A nicest AE cab E EA Goats tee ncaa i 3 3 vi Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Contents Ambient Tepic canina RE 3 3 On La
204. o settings If the selected zero air flow is lower or higher Model 146 Instruction Manual 3 19 Operation Flow Modes 3 20 Gas Photometer Model 146 Instruction Manual than the acceptable range TOO LOW or TOO HIGH appears on the screen e In the Main Menu choose Flow Modes gt O3 gt GPT Photometer Control gt Gas A F gt Level 1 5 gt Manual gt Zero Air Coco P ay A cu mii Fane Haeee FM EE fat Seal cee EE Eee EE EE The Manual Gas Flow screen sets the manual gas flow value for the previously selected gas The screen identifies the minimum and maximum acceptable gas flow rates and lets the user select a flow rate between these two settings If the selected gas flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Flow Modes gt O3 gt GPT Photometer Control gt Gas A F gt Level 1 5 gt Manual gt Gas The Photometer flow mode menu is used to select the photometer measurement mode Available choices include PHOT OFF PHOT INT and PHOT EXT Ozonator control mode may not be selected from this screen and must be selected via the ozonator screen If the instrument is in control mode a warning will display IN OZONATOR CONTROL MODE Once a choice is selected the screen reports that the new mode has been accepted and activated The message SAVING is displayed for 3 seconds then removed e Inthe Main Men
205. ocncnnnnonccnnnnnnono 7 2 Permeation Oven Replacement ici 7 2 Reaction Chamber Mixing Chamber Replacement ooccccionnnnninnconnnnoso 7 3 Pressure Regulator Replacement sist 7 3 Back Pressure Regulator Replacements lus 7 3 Pressure Regulator s AdJUSEMEN Earle diia ia aiii 7 3 Ambient Temperature Calibration minitas 7 3 Service LOCAS ae 7 3 System Description sccsccsscsssssssssseessssssssssssessssesssessesassasestssneesessesatenteatees 8 1 A tated S tied te ag Sac en tes pacer ecko le ck as rel ised S 8 1 Plot 8 1 Precisi n Gas Dilution ini las 8 1 Ozone Genera sti 8 1 Gas Phase o e Nos teu 8 1 Permeation Pube Oyen sae 8 1 Photomieter ini A A aTa 8 1 Detector System eanont nere ce ee aed Eiai E a Mut iE tioa 8 1 Pressure Transductor iia 8 1 A O ede ai i a a a A 8 1 Ozonator Lamp Power Supply seesseeseseessesssisisesisrsisesesesesesesesesesees 8 1 Photometer Lamp Power SUP 8 1 Photometer Temperature Thermistor 8 1 Pump ea eE E a ae tate E O Y 8 1 Sample Reference Solenoid Valves cat 8 1 Ozonator Solenoid irri a 8 1 Permi Solenoid A a e a a e a CR ee 8 1 Pern 2 Soleo id acres ge tice ara ira aain aaa aa a AREAS 8 1 Reaction Chamber Solenoid viril 8 1 Zero Air Solenoid iiae sioi iore e ae Eae eE ESSE 8 1 External Internal Solenoid iiiiiad aii 8 1 Span Gas dolencias dl 8 1 Phiot meter e A a A R N 8 1 Ozonator E A aies eia E e RE ee E le 8 1 Pressure Gauge Caplio io 8 2 Pu mp Gapillaiy cota 8 2 Mixing Chamber sesini nino
206. of ozonator in GPT mode in percent If a particular option is not installed or not being used it is removed from the display The example below shows the Model 1467 Run screen with gas dilution photometer internal measurement permeation oven installed but off and GPT Thermo Fisher Scientific Operation Software Overview pir pTi E Eiane Bust Fanes Boone ppp E it Benos ane a a En REP ED Hp roppu iri CoL m 5s E Eiet Bast Base Bast ae POE Sek et Pih E E Main Menu The Main Menu contains a number of submenus Instrument parameters and settings can be read and modified within the submenus according to their function The measurement values appear above the main menu and submenus in every screen The OZONATOR SETUP PERMEATION OVEN SETUP AND PHOTOMETER SETUP menu items are only displayed if these options are installed The SERVICE menu is visible only when the instrument is in service mode For more information on the service mode see Service Mode later in this chapter e Use and to move the cursor up and down e Press to select a choice e Press M to return to the top of the Main Menu or gt to return to the Run screen Mo MoT CH CER L E Eiet fast Bone ie i Thermo Fisher Scientific Model 146 Instruction Manual 3 7 Operation Operation Operation 3 8 Model 146 Instruction Manual The Operation screen is used to modify the calibrator s operational settings
207. og outputs For example an O range of 0 5 ppm sets the analog output to concentrations between 0 and 5 ppm Available operating ranges include 0 05 0 1 0 2 0 5 1 0 2 0 5 0 C1 C2 and C3 Cl C2 and C3 are custom ranges For more information about custom ranges see Set Custom Ranges below The display shows the current O range The next line of the display is used to change the range e Inthe Main Menu choose Range gt O3 Range Thermo Fisher Scientific Thermo Fisher Scientific Operation Photometer Setup Menu WH mpini Pin Piri The analog outputs are arranged on the rear panel connector as shown in Figure 3 4 See Table 3 3 for channels and pin connections 9 o000000000000000000 oO000000000000000000 Figure 3 4 Pin Out of Rear Connector Table 3 2 Default Analog Outputs Channel Pin Description 1 4 Gas Conc Analog Output 2 33 Gas Conc Analog Output 3 5 None 4 34 None 5 7 None 6 36 None Ground 16 18 19 35 37 Signal Ground Model 146 Instruction Manual 3 31 Operation Photometer Setup Menu Note All channels are user definable If any customization has been made to the analog output configuration the default selections my not apply A Set Custom Ranges The Custom Ranges Menu lists three custom ranges C1 C2 and C3 Custom ranges are user defined ranges In the standard range mode any value between 0 05 ppm and 5 pp
208. ogging Settings gt Configure Datalogging gt Memory Allocation m La nobel Thermo Fisher Scientific Model 146 Instruction Manual 3 47 Operation Instrument Controls Menu Data Treatment The Data Treatment screen is used to select the data type for the selected record whether the data should be averaged over the interval the minimum or maximum used or the current value logged Data treatment doesn t apply to all data just to the concentration measurement All other data points log the current value at the end of the interval e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Configure Datalogging gt Data Treatment ER risas BEES Pi Taal DH Cie Fae Hanne BM EB Fane Bast Bases Haat ade POE Sa e Pi EE EE Communication The Communication Settings menu is used with communications control 8 Settings and configuration e Inthe Main Menu choose Instrument Controls gt Communication Settings poren pn Eus BEES PT Cp kobe E Mihei i Baud Rate The Baud Rate screen is used to set the RS 232 RS 485 interface baud rate Baud rates of 1200 2400 4800 and 9600 19200 38400 57600 and 115200 are available e In the Main Menu choose Instrument Controls gt Communication Settings gt Baud Rate 3 48 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Instrument ID The Instrument ID screen allows the operator to edit t
209. oids Span G All E olenoids Span Gas C Ozonator Solenoid q External Internal Mixing Chamber A Manifold A Solenoid Zero Air El m L Solenoid Ozonator Capillary Pressure Gauge Solenoid Perm 1 Solenoid Perm 2 Mass Flow Controllers Pressure Regulator and Gauge Back Pressure Pressure Fan Reaction Reaction Permeation Oven Regulator Transducer lle Chamber olenoi Flow System 8 2 Model 146 Instruction Manual Figure 8 1 Hardware Components The Model 146i Dynamic Gas Calibrator can produce up to four different flow streams using three hardware modules The basic unit contains only the dilution module There are two optional modules either or both of which can be added to the basic instrument These optional modules are e the ozonator generating module which produces an ozone flow or gas phase titration GPT with the addition of an NO cylinder e the permeation tube oven module The flow for each module is governed by mass flow controllers and solenoids Each system is described independently While many of the components are shared they may only be necessary for one or two functions These components are only described when the system that requires them is being used Thermo Fisher Scientific System Description Flow System Precision Gas Dilution Gas dilution is achieved by using
210. oller is selected Note Changing the flow controller selection may affect allowable limits for span concentration and total flow If is advisable to check these settings A The mass flow controllers are accurate to only 2 upon delivery from the manufacturer The instrument includes factory set calibration curves to improve this accuracy To achieve an equivalent accuracy when installing new controllers see Zero Flow Calibration later in this chapter A e In the Main Menu choose Service gt Flow Control Selection gt Zero Scale Mri Ee Boas Wi rh iyd i de m ii The Gas Flow Controller screen is used to select the gas controller Allowable selections are 25 50 100 200 and 500 sccm The gas flow controller calibration points are reset to defaults when a new controller is selected Model 146 Instruction Manual 3 85 Operation Service Menu Note Changing the flow controller selection may affect allowable limits for span concentration and total flow If is advisable to check these settings A The mass flow controllers are accurate to only 2 upon delivery from the manufacturer The instrument includes factory set calibration curves to improve this accuracy To achieve an equivalent accuracy when installing new controllers see Gas Flow Calibration later in this chapter A e In the Main Menu choose Service gt Flow Control Selection gt Gas ALAR Calibration The Calibrate Pressure
211. olvents or corrosive gases Do not shake or jolt the module 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove the two screws from right side of LCD module viewed front front 3 Disconnect the ribbon cable and the two wire connector from the front panel board 4 Loosen left side retaining screw viewed from front and slide the LCD module out towards the right and rear of the instrument Thermo Fisher Scientific Photometer Lamp Replacement Photometer option Thermo Fisher Scientific Servicing Photometer Lamp Replacement Photometer option 5 Replace the LCD module by following previous steps in reverse The lamp control system of the Model 1467 Photometer option has been designed to operate the lamp conservatively to increase its life However the lamp should be replaced when any one of the following conditions occur e No light output e Inability to adjust lamp position to obtain an output detection frequency of 65 kHz e Noisy output signal which has been traced to an unstable lamp see the Troubleshooting chapter Note It is not necessary to recalibrate the Model 146i since it is a ratio instrument and replacing the lamp does not affect the calibration However after replacing the photometer lamp you must adjust the photometer lamp voltage to ensure proper operation Refer to the Photometer Lamp Voltage Adjustment procedure that follows to adjust the phot
212. ometer lamp voltage yA Use the following procedure to replace the photometer lamp Figure 7 12 Equipment required Allen wrench 1 16 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a Use the following procedure to replace the photometer lamp 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the photometer lamp cable from the LAMP connector on the photometer board 3 Slide the insulation off the end of the lamp 4 Loosen the Allen screw on the lamp clamp Model 146 Instruction Manual 7 31 Servicing Photometer Lamp Replacement Photometer option 7 32 Model 146 Instruction Manual Lamp Retaining Screw Lamp Housing Thermistor Detector Retaining Screw Detector Figure 7 12 Photometer Components 5 Carefully slide the lamp out of the photometer 6 Carefully slide new lamp into place until it bottoms and tighten the lamp clamp screw 7 Plug the new photometer lamp cable into the LAMP connector on the photometer board 8 Re install the instrument cover 9 Reconnect power and turn instrument ON 10 After lamp has stabilized approximately 15 minutes refer to the Photometer Lamp Voltage Adjustment procedure that follows to adjust the photometer lamp voltage Thermo Fisher Scientific Photo
213. on B 40 Model 1467 Instruction Manual set addr ip address This command sets the analyzer s IP address where address consists of four gt numbers ranging from 0 255 inclusive separated by Send set addr ip 192 168 1 200 Receive set addr ip 192 168 1 200 ok addr nm This command reports the IP netmask Send addr nm Receive addr nm 255 255 255 0 set addr nm address This command sets the nm address where address consists of four numbers 9 ranging from 0 255 inclusive separated by Send set addr nm 255 255 255 0 Receive set addr nm 255 255 255 0 ok baud This command reports the current baud rate for the serial port RS232 RS485 The example below reports that the current baud rate is 9600 baud Send baud Receive baud 9600 set baud rate rate 1200 2400 4800 9600 19200 38400 57600 115200 This command sets the instrument baud rate The example below sets the instrument s baud rate to 9600 Note After the command is sent the baud rate of the sending device must be changed to agree with the instrument A Send set baud 9600 Receive set baud 9600 ok dhcp This command reports the current state of use of DHCP on or off DHCP is used to assign an IP address to the analyzer automatically The example below shows that DHCP is on Send dhcp Receive dhcp on Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Communications Con
214. on table within parentheses This is a string of numbers separated by spaces 0 1 The selection table lists the translation table entries which the user may select from when setting the parameter This is not necessarily the same as the entries which may be displayed Then there appears an optional button designator This will be one of B T as T or N B Indicates a button which pops up an input dialog prompting the user for a new value using the designated input format The input format is specified from the B through the subsequent semicolon I Indicates a button which pops up a selection list with input translation That is the values read are translated before they are compared to the selection list options L Indicates a button which pops up a selection list without any translation The output value is number of the selected option T Indicates a button which pops up a selection list with output translation The number of the option selected is used as an index into the translation table to generate an output string N Indicates a button which only sends the subsequent command to the instrument No user prompting happens Thermo Fisher Scientific Thermo Fisher Scientific Examples C Link Protocol Commands Record Layout Definition The following string through an optional or the end of the line is the command which is to be sent to the instrument upon the completion of the button selec
215. on the precision specifier which now becomes a field number can precede the precision specifier this This is useful when formatting for example numbers which have varying precision depending on the mode of the instrument Binary numbers can also have an optional precision specifier which is used to determine how many bits to print For example the specifier b4 will print the lowest four bits of the parsed number I here are serious restrictions on where an s field may appear currentl y y sources 1 and 2 must be s and no others may be s Model 146 Instruction Manual B 49 C Link Protocol Commands Record Layout Definition Alarm Information Translation Table Selection Table Button Designator B 50 Model 1467 Instruction Manual The value source is followed by optional alarm information indicated by a commercial at sign with a source indicator and a starting bit indicator All alarm information is presumed to be two bits long low and high The bitfield extraction is performed on the integer part of the source Typical alarm information would appear as 6 4 Then there appears an optional translation table within braces This is a string of words separated by spaces An example translation table would be Code_0 Code_1 Code_2 Code_3 The value once extracted is used as a zero based index into the translation table to determine the string to display Then there appears an optional selecti
216. oncnccnonccnrncncornncnrnnnnrnonornnnnrncnonos 7 3 Figure 7 3 Removing the Measurement Bench and Lowering the Partition A AE E tan 7 3 Figure 7 4 Replacing a Mass Flow Controller oooooococincicincinincinincininncnooos 7 3 Figure 7 5 Replacing the PuUMP ooococinononcicincnccccnnnoncncnrnnnonononorornnnonororanonnonoso 7 3 Figure 7 6 Replacing the Rear Panel Fan oo ccecccccececesescetesesesteteseseetenenees 7 3 Model 1467 Instruction Manual xiii Figures xiv Model 146 Instruction Manual Figure 7 7 Rear Panel Analog Input and Output Pins 000 0 cece 7 3 Figure 7 8 Replacing the 1 0 Expansion Board Optional 7 3 Figure 7 9 Rear Panel Board Connectors 0 0 2 cecccccccsceceseesestesestesesteseeteseetaveees 7 3 Figure 7 10 Replacing the Measurement Interface Board 7 3 Figure 7 11 Replacing the Front Panel Board and the LCD Module 7 3 Figure 7 12 Photometer COMPONENTS 0 c cccccceccsceseseesestesestesesteseeteseseeseees 7 3 Figure 7 13 Replacing the OzONator oooonocooniccicincininccnnnnononncnnnncnnornonnnocnnnnos 7 3 Figure 7 14 Removing the Glass Chamber Assembly 7 3 Figure 7 15 Replacing the Glass Chamber Assembly 7 3 Figure 7 16 Replacing the Pressure Regulator ocoococioinionnininonncninononocoo 7 3 Figure 8 1 Hardware COMPONE
217. ons can be accessed over Ethernet using TCP IP port 9880 For details see the following topics e Instrument Identification Number on page B 1 e Commands on page B 2 e Measurements on page B 8 e Alarms on page B 10 e Diagnostics on page B 14 e Datalogging on page B 16 e Calibration on page B 23 e Keys Display on page B 25 e Measurement Configuration on page B 28 e Hardware Configuration on page B 37 e Communications Configuration on page B 39 e I O Configuration on page B 43 e Record Layout Definition on page B 47 Each command sent to the instrument must begin with the American Standard Code for Information Interchange ASCII symbol or byte value equivalent of the instrument s identification number plus 128 For example if the instrument ID is set to 25 then each command must begin with the ACSII character code 153 decimal The analyzer ignores any command that does not begin with its instrument identification number If the instrument ID is set to 0 then this byte is not required For more information on changing Instrument ID see Chapter 3 Operation Model 146 Instruction Manual B 1 C Link Protocol Commands Commands Commands Commands List B 2 Model 146 Instruction Manual The instrument must be in the remote mode in order to change instrument parameters via remote However the command set mode remote can be sent to the a
218. ons used along with the board level connection diagrams to troubleshoot board level faults Motherboard Connector Pin Descriptions on page 6 10 Front Panel Board Connector Pin Diagram on page 6 15 I O Expansion Board Optional Connector Pin Descriptions on page 6 17 Digital Output Board Connector Pin Descriptions on page 6 19 Measurement Interface Board Connector Pin Descriptions on page 6 20 Solenoid Connector Board Pin Descriptions on page 6 23 Ozonator Power Supply Connector Pin Descriptions on page 6 24 Photometer Power Supply Connector Pin Descriptions on page 6 25 Table 6 3 Motherboard Connector Pin Descriptions Connector Reference Pin Signal Label Designator Description INTF DATA J1 Ground RS485 to Interface Board RS485 to Interface Board GS nd 10 BASE T J2 Ethernet Output Ethernet Output Ethernet Input NC NC Ethernet Input NC NC Ground RS485 to Interface Board RS485 to Interface Board oOo N on A WwW N INTF DATA J1 wo nN 10 BASE T J2 Ethernet Output 2 Ethernet Output 6 10 Model 146 Instruction Manual Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Table 6 3 Motherboard Connector Pin Descriptions continued Connector Label Reference Designator Pin oOo N oon A q Signal Description Ethernet Input NC NC Ethernet Input NC NC
219. or gt Water Bath Enter the thermistor temperature from the thermometer and press 4 Figure 9 2 E p SRL a L Tra pai CE eo rel Bt Lo on i T Figure 9 2 Cal Gas Therm Bath Screen 4 Remove the thermistor from the water bath dry and replace into the permeation tube oven 5 Make sure the charcoal scrubber is connected to the ZERO bulkhead on the rear panel 6 Wait for the permeation gas temperature reading to stabilize Setting Temperature with Known Use the following procedure to set the gas temperature with an accurate Resistance known resistance 1 Remove the thermistor from POJ1 on the measurement interface board 1 Connect a 4 369 KQ resistor across pins 1 and 2 of POJ1 use a resistance substitution box and an accurate meter if necessary 2 From the Permeation Oven menu select Cal Gas Thermistor then select known resistor Figure 9 3 Thermo Fisher Scientific Model 146 Instruction Manual 9 7 Optional Equipment Permeation Oven STE El weet EEE BE Beane Benue CH mtr Pi nn ied ejects ip guts Piet Figure 9 3 Cal Gas Therm Resistor Screen 3 Enter the exact value of the attached resistor and press to save the offset 4 Remove the resistor and reconnect the gas thermistor 5 Make sure the charcoal scrubber is connected to the ZERO bulkhead on the rear panel 6 Wait for the permeation gas temperature reading to stabilize Determining Permeation U
220. ove the cover 2 Disconnect all electrical connections 3 Disconnect the gas fittings 4 Remove the two 2 screws holding the oven to the floor plate and remove the oven 5 Remove oven glass chamber from oven and install in new oven a Release the latches on the sides of the oven cover and remove cover b Remove glass chamber assembly by loosening not removing knurled screw located at the top of the chamber and gently pulling assembly upward Completely remove assembly from oven Figure 7 14 c Separate glass chamber from top assembly by twisting and gently pulling glass away from top Figure 7 14 Keep glass clean by using Kimwipes or similar material to handle glass d Place permeation tube in chamber e Attach glass chamber to top assembly by gently pushing together with a slight twisting motion Figure 7 14 f Replace glass chamber assembly into oven until top of assembly is flush or slightly below oven top Figure 7 15 Thermo Fisher Scientific Servicing Permeation Oven Replacement g Tighten knurled screw finger tight Do not use tools to tighten h Replace oven cover being careful to place tubing and wire in slot of cover 6 Install the new oven by following the previous steps in reverse 7 Check the system for leaks Refer to Leak Test and Pump Checkout in this chapter S Knurled Screw LA ha Glass Chamber Assembly ven Top Oven Enclosure
221. place the LCD module near organic solvents or corrosive gases Do not shake or jolt the LCD module WARNING Changes or modifications to this unit not expressly approved by the party responsible for compliance could void the user s authority to operate the equipment a Note This equipment has been tested and found to comply with the limits for a Class A digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual may cause harmful interference to radio communications Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense a Model 146 Instruction Manual iii Preface WEEE Symbol The following symbol and description identify the WEEE marking used on the instrument and in the associated documentation Symbol Description Marking of electrical and electronic equipment which applies to Y electrical and electronic equipment falling under the Directive 2002 96 EC WEEE and the equipment that has been put on the market z after 13 August 2005 Where to Get Help Service is available from exclusive distributors worldwide Contact one of
222. plate Figure 7 12 and remove the photometer 4 Loosen the Allen screw securing the detector in detector block and remove the detector 5 Install the new detector by following the above steps in reverse 6 Adjust the photometer lamp intensity if necessary via Service gt Lamp Setting Photometer Use the following procedure to replace the photometer thermistor Thermistor sure 7 12 Re p lacement Equipment Required Photometer thermistor Wrench 1 4 inch amounts of static electricity A properly grounded antistatic wrist strap must Equipment Damage Some internal components can be damaged by small be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect the thermistor electrical connection 7 36 Model 146 Instruction Manual Thermo Fisher Scientific Servicing Photometer Temperature Calibration Photometer option 3 Remove the thermistor and replace it with the new thermistor 4 Connect the thermistor electrical connection Photometer Use the following procedure to calibrate the photometer temperature Temperature Calibration Photometer option Calibrated thermometer or 10K 1 Resistor Equipment Required WARNING The service procedures in this manual are restricted to qualified service representatives If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment ma
223. ponse Function code 1 Byte 0x01 or 0x02 Byte count 1 Byte N Output Status n Byte N N or N 1 N Quantity of Outputs 8 if the remainder not equal to zero then N N 1 Error Response Function code 1 Byte 0x01 or 0x02 Exception code 1 Byte 01 IIlegal Function 02 Illegal Address 03 Illegal Data 04 Slave Device Failure Here is an example of a request and response to read outputs 2 15 Request Field Name Hex Function 0x01 Starting Address Hi 0x00 Starting Address Lo 0x02 Quantity of Outputs Hi 0x00 Quantity of Outputs Lo Ox0D Thermo Fisher Scientific MODBUS Protocol Function Codes Response Field Name Hex Function 0x01 Byte Count 0x03 Output Status 2 10 OxCD Output Status 11 15 Ox0A The status of outputs 2 10 is shown as the byte value 0xCD or binary 1100 1101 Output 10 is the MSB of this byte and output 2 is the LSB By convention bits within a byte are shown with the MSB to the left and the LSB to the right Thus the outputs in the first byte are 10 through 2 from left to right In the last data byte the status of outputs 15 11 is shown as the byte value 0x0A or binary 0000 1010 Output 15 is in the fifth bit position from the left and output 11 is the LSB of this byte The four remaining high order bits are zero filled 0x03 0x04 Read Read holding input registers reads the measurement data from the 8 1np 8 Holding Registers instrument Issuing either of these function codes will g
224. port B 39 addr nm Reports sets IP netmask address for Ethernet port B 40 alarm bench lamp Reports sets photometer bench lamp temperature alarm B 10 temp max maximum value Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Commands Command Description Page alarm bench lamp Reports sets photometer bench lamp temperature alarm B 10 temp min minimum value alarm bench temp Reports sets photometer bench temperature alarm maximum B 11 max value alarm bench temp Reports sets photometer bench temperature alarm minimum B 11 min value alarm conc 03 max Reports sets photometer Oz concentration alarm maximum B 11 value alarm conc 03 min Reports sets photometer 0 concentration alarm minimum B 11 value alarm intensity max Reports sets photometer intensity alarm maximum value B 11 alarm intensity min Reports sets photometer intensity alarm minimum value B 11 alarm internal temp Reports sets ambient temperature alarm maximum value B 12 max alarm internal temp Reports sets ambient temperature alarm minimum value B 12 min alarm 03 lamp temp Reports sets ozonator lamp temperature alarm maximum B 12 max value alarm 03 lamp temp Reports sets ozonator lamp temperature alarm minimum B 12 min value alarm pressure max Reports sets photometer pressure alarm maximum value B 13 alarm pressure min Reports sets photometer pressure alarm
225. propriate safety precautions see the Servicing chapter Replaceme nt Parts see Replacement Parts List on page 7 5 in the Servicing chapter for a list of replacement parts Accessing the Ifthe Service is not displayed on the Main Menu use the following Service Mode procedure to display it 1 At the Main Menu press to scroll to Instrument Controls gt press gt to scroll to Service Mode gt and press The Service Mode screen appears 2 Press to toggle the Service Mode to ON 3 Press aw gt Cm to return to the Main Menu 4 Return to the procedure Outside Case Clean the outside case using a damp cloth being careful not to damage the Cleaning labels on the case clean the outside case a Equipment Damage Do Not use solvents or other cleaning products to 5 2 Model 146 Instruction Manual Thermo Fisher Scientific Preventive Maintenance Fan Filter Inspection and Cleaning Fan Filter Inspection Use the following procedure to inspect and clean the fan filter Figure 5 1 and Cleaning Remove the two fan guards from the fans and remove the filters 2 Flush the filter with warm water and let dry a clean oil free purge will help the drying process or blow the filter clean with compressed air 3 Re install the filter and fan guards Mounting Screws 4 Filter Fan Guard Figure 5 1 Inspecting and Cleaning the Fan Filter Photometer Cleaning Best results are obtained whe
226. put The selected item is shown by lt after it For more information on the I O connections see Remote Interconnections in the Installation chapter e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Output Relay Settings gt Select Relay gt Instrument State gt Non Alarm Model 146 Instruction Manual 3 57 Operation Instrument Controls Menu Digital Input Settings The Digital Input Settings menu displays a list of the 16 digital inputs available and allows the user to select the logic state and instrument parameter for the relay selected Note The digital inputs must be asserted for at least one second for the action to be activated A e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Digital Input Settings mo AE Cabal AA Logic State The Logic State screen is used to change the I O relay to either normally open or normally closed The default state is open which indicates that a relay connected between the digital input pin and ground is normally open and closes to trigger the digital input action If nothing is connected to the digital input pin the state should be left at open to prevent the action from being triggered e Press to toggle and set the logic state open or closed 3 58 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Instrument Action The Instrument Action screen allows the user
227. puts instrument status and measurement data to the front panel mounted graphics display and to the rear panel I O The motherboard also contains I O circuitry and the associated connector to monitor external digital status lines and to output analog voltages that represent the measurement data Connectors located on the motherboard include External connectors include External Accessory e RS 232 485 Communications two connectors Ethernet Communications e I O connector with Power Fail Relay 16 Digital Inputs and 6 Analog Voltage Outputs Internal connectors include e Function key panel and Display e Measurement Interface Board Data e I O Expansion Board Data e Digital Output Board e AC distribution The measurement interface board serves as a central connection area for all measurement electronics in the instrument It contains power supplies and interface circuitry for sensors and control devices in the measurement system It sends status data to the motherboard and receives control signals from the motherboard Connectors located on the measurement interface board include Model 146 Instruction Manual 8 9 System Description Electronics Pressure Sensor Assembly Ozonator Lamp Power Supply 8 10 Model 146 Instruction Manual e Data communication with the motherboard e 24V and 120VAC power supply inputs e Fan and solenoid outputs e 120VAC outputs for the photometer pump e Two mass flow controllers e Pres
228. r Measurements and Analog Inputs if the I O expansion board is installed This is a temporary list of items for the selected record type that must be committed via the datalogging menu before the changes will apply Note that committing any changes to this list Thermo Fisher Scientific Model 146 Instruction Manual 3 43 Operation Instrument Controls Menu will erase all currently logged data as the format of the stored data is changed e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content Concentrations The Concentrations screen allows the user to select the output signal that is tied to the selected field item The selected item is shown by lt after it Note that at this point pressing indicates that these are proposed changes as opposed to implemented changes To change the selected record format and erase record log file data see Commit Content below e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Select Content gt Select Field gt Concentrations 3 44 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Other Measurements The Other Measurements screen allows the user to select the output signal that is tied to the selected field item The selected item is shown by lt after it Items displayed are determined by the options installed Note that at this point pressing indicates that these
229. r broken Leak Test and Pump Use the following procedure to check for leaks in the system and to leak test Checkout without the pump This procedure is for Model 1467 instruments that do not have 5 the photometer option For Model 1467 instruments that have the Photometer Option photometer option refer to Leak Test and Pump Checkout with Photometer Option 7 12 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Servicing Leak Test and Pump Checkout without Photometer Option Equipment Required Vacuum pump Vacuum gauge Shut off valve WARNING The service procedures in this manual are restricted to qualified service representatives If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Remove the cover 2 With the instrument running cap all external ports on the rear panel except the EXHAUST port 3 Ifthe instrument is equipped with a permeation oven activate the permeation oven from the front panel controls Refer to the Operation chapter for additional permeation oven operation information if necessary 4 Connect a shutoff valve between the external vacuum pump intake and the vacuum gauge
230. r standard e Inthe Main Menu choose Ozonator Setup gt Manual Thermo Fisher Scientific Model 146 Instruction Manual 3 27 Operation Permeation Oven Setup Menu Permeation Oven The Model 146 can be equipped with an optional permeation oven The permeation oven supplies an internal source of calibration gas A small Setup Menu amount of zero air flows through the permeation tube and is mixed with the remainder of the zero air flow in the mixing chamber This menu in only available if the PERMEATION OVEN option is enabled in the INSTRUMENT CONFIGURATION menu under DIAGNOSTICS Note Permeation oven temperature selection and calibration may be done via the Service gt Perm Oven Settings menu A e Inthe Main Menu choose Permeation Oven Setup DH Guided Ce a E SO a EA Permeation K The Permeation K constant screen allow the user to set the multiplier for the permeation tube for a given oven temperature Allowable values are 0 000 to 9 999 If the values selected are lower or higher than allowed TOO LOW or TOO HIGH appears on the screen Note Changing this value may affect allowable limits for permeation concentration levels It is advisable to check these settings A e Inthe Main Menu choose Permeation Oven Setup gt Permeation K Fusco Sana hees E iR in Haat h POE ae e Pi EE EE 3 28 Model 146 Instruction Manual Thermo Fisher Scientific Operation Permeation Oven Setup Menu Permeation Rate The Perm
231. rast level This command sets the screen s level of contrast according to Table B 9 The example below sets the contrast level to 50 Send set contrast 5 Receive set contrast 5 ok Table B 9 Contrast Levels Level Contrast Level 0 0 1 10 2 20 3 30 4 40 5 50 6 60 Model 146 Instruction Manual B 37 C Link Protocol Commands Hardware Configuration B 38 Model 146 Instruction Manual Level Contrast Level 7 70 8 80 9 90 10 100 date This command reports the current date The example below reports the date as December 1 2004 Send date Receive date 12 01 04 set date mm dd yy mm month dd day yy year This command sets the date of the analyzer s internal clock The example below sets the date to December 1 2004 Send set date 12 01 04 Receive set date 12 01 04 ok set default params This command sets all the parameters to their default values This does not affect the factory calibrated parameters Send set default params Receive set default params ok save set save params This command stores all current parameters in FLASH memory It is important that each time instrument parameters are changed that this command be sent If changes are not saved they will be lost in the event of a power failure The example below saves the parameters to FLASH memory Send set save params Receive set save params ok time This command reports the c
232. rent flow controller readings The flow is measured by internal flow sensors e Inthe Main Menu choose Diagnostics gt Flows Temperatures The Temperatures screen read only displays the current ambient temperature bench temperature bench lamp temperature ozone lamp temperature and permeation oven temperature e Inthe Main Menu choose Diagnostics gt Temperatures Pressure The Pressure screen read only displays the current pressure reading The pressure is measured by a pressure transducer e Inthe Main Menu choose Diagnostics gt Pressure Thermo Fisher Scientific Model 146 Instruction Manual 3 73 Operation Diagnostics Menu Analog Input Readings Analog Input Voltages Digital Inputs 3 74 Model 146 Instruction Manual The Analog Input Readings screen read only displays the 8 current user scaled analog readings if the I O expansion board option is installed In the Main Menu choose Diagnostics gt Analog Input Readings The Analog Input Voltages screen read only displays the 8 raw analog voltage readings if the I O expansion board option is installed In the Main Menu choose Diagnostics gt Analog Input Voltages The Digital Inputs screen read only displays the state of the 16 digital inputs In the Main Menu choose Diagnostics gt Digital Inputs Thermo Fisher Scientific Operation Diagnostics Menu Relay States The Relay States screen displays the state o
233. resentatives a If the equipment is operated in a manner not specified by the manufacturer the protection provided by the equipment may be impaired Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a Model 146 Instruction Manual 7 11 Servicing Leak Test and Pump Checkout without Photometer Option 1 Disconnect the external SAMPLE input line and plug the external SAMPLE fitting 2 Disconnect the ZERO AIR input lines and plug 3 Plug the VENT and OUTPUT 4 Press to display the Main Menu 5 At Operation press to display the Operation screen 6 Use to scroll to PHOT EXT CTL OFF or INT use gt to change the choice to PHOTO INT if necessary and press to save the setting 7 Press m to display the Main Menu 8 Press 4 to scroll to Diagnostics and press to display the Diagnostics menu 9 Use 4 to scroll to Pressure and press to display the Pressure screen The pressure reading should drop to less than 250 mm Hg If the pump diaphragm is in good condition and the capillary is not blocked it should take less than 20 seconds from the time the inlet is plugged to the time the reading drops below 250 mm Hg If a leak is suspected carefully tighten each fitting until the leak is found and check to see that none of the input lines are cracked o
234. ressure transducer within the instrument and refer to Replacement Parts List on page 7 5 if you need the pressure transducer part number Equipment Required Pressure transducer assembly Philips screwdriver short Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Turn instrument OFF unplug the power cord and remove the cover 2 Disconnect plumbing from the pressure transducer assembly Note the P g P y plumbing connections to facilitate reconnection 3 Disconnect the pressure transducer PRESS connector from the measurement interface board Model 146 Instruction Manual 7 39 Servicing Pressure Transducer Calibration Photometer option 4 Remove the two pressure transducer retaining screws and remove the pressure transducer 5 Install the new pressure transducer assembly by following the previous steps in reverse 6 Calibrate the pressure transducer Refer to the Pressure Transducer Calibration procedure that follows Pressure Transducer An error in the zero setting of the pressure transducer does not introduce a Calibration measurable error in the output concentration reading Therefore if only a Photometer opti on barometer is available and not a vacuum pump only adjust the span setting A rough check of the pressure accuracy can be made by obtainin
235. rly grounded antistatic wrist strap must Equipment Damage Some internal components can be damaged by small be worn while handling any internal component a 1 Turn instrument OFE unplug the power cord and remove the cover 2 Remove plumbing 3 Remove fitting from bottom of the pressure regulator by holding the assembly in place with a 9 16 inch wrench as shown in Figure 7 16 and loosening the pressure regulator fitting 7 16 inch fitting 9 16 inch fitting A DA ED lt 2 Figure 7 16 Replacing the Pressure Regulator 4 Remove the pressure regulator and replace with a new unit 5 Reconnect plumbing and check for leaks Refer to Leak Test and Pump Checkout in this chapter Thermo Fisher Scientific Model 146 Instruction Manual 7 49 Servicing Back Pressure Regulator Replacement 7 50 Back Pressure Regulator Replacement A Model 146 Instruction Manual Use the following procedure to replace the back pressure regulator Equipment Required Back pressure regulator Wrenches 5 32 and 1 2 inch Flatblade screwdriver Philips screwdriver Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Remove plumbing 3 Loosen four 4 captive screws and remove the assembly 4 Remove the
236. rm 03 lamp temp min value set alarm 03 lamp temp max value These commands set the ozonator lamp temperature alarm minimum and maximum values to value where value is a floating point number representing ozonator lamp temperature alarm limits in degrees C The Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Alarms example below sets the ozonator lamp temperature alarm maximum value to 58 C These commands are only available if the photometer option is installed Send set alarm 03 lamp temp max 58 0 Receive set alarm 03 lamp temp max 58 0 ok alarm pressure min alarm pressure max These commands report the photometer bench pressure alarm minimum and maximum value current settings The example below reports that the pressure alarm minimum value is 200 mmHg These commands are only available if the photometer option is installed Send alarm pressure min Receive alarm pressure min 200 mm_Hg set alarm pressure min value set alarm pressure max value These commands set the photometer bench pressure alarm minimum and maximum values to value where value is a floating point number representing pressure alarm limits in millimeters of mercury The example below sets the pressure alarm maximum value to 1000 mmHg These commands are only available if the photometer option is installed Send set alarm pressure max 1000 Receive set alarm pressure max 1000 ok alarm trig conc 03 This command reports the
237. round 32 Ground 33 24V 34 24V RS232 RS485 A P1 A 1 NC 2 Serial Port 1 RX RS485 IN 3 Serial Port 1 TX RS485 OUT 4 NC 5 Ground 6 NC Thermo Fisher Scientific Troubleshooting Connector Pin Descriptions Table 6 3 Motherboard Connector Pin Descriptions continued Connector Reference Label Designator Pin oOo co mw Signal Description Serial Port 1 RTS RS485 OUT Serial Port 1 CTS RS485 IN NC RS232 RS485 B P1 B Oo 0 N oon A wo N NC Serial Port 2 RX RS485 IN Serial Port 2 TX RS485 OUT NC Ground NC Serial Port 2 RTS RS485 OUT Serial Port 2 CTS RS485 IN NC AC IN PJ1 wo N AC HOT AC NEUT AC Ground AC 24VPWR PJ2 0 N AC HOT AC NEUT AC Ground AC INTF BD PJ3 AC HOT AC NEUT AC Ground Table 6 4 Front Panel Board Connector Pin Diagram Connector Reference Label Designator MOTHER J1 BOARD Thermo Fisher Scientific Pin D om A W N Signal Description Ground Ground LCLK LCD Signal Ground Ground LLP LCD Signal Model 146 Instruction Manual 6 15 Troubleshooting Connector Pin Descriptions Table 6 4 Front Panel Board Connector Pin Diagram continued Connector Reference Pin Signal Label Designator Description 7 LFLM LCD Signal 8 LD4 LCD Signal 9 LDO LCD Signal 10 LD5 LCD Signal 11 LD1 LCD Signal 12 LD6 LCD Signal 13 LD2 LCD Signal 14 LD7 LCD Signal
238. rth in the product documentation published specifications or package inserts Ifa period of time is not specified in Seller s product documentation published specifications or package inserts the warranty period shall be one 1 year from the date of shipment to Buyer for equipment and ninety 90 days for all other products the Warranty Period Seller agrees during the Warranty Period to repair or replace at Seller s option defective Products so as to cause the same to operate in substantial conformance with said published specifications provided that a Buyer shall promptly notify Seller in writing upon the discovery of any defect which notice shall include the product model and serial number if applicable and details of the warranty claim b after Seller s review Seller will provide Buyer with service data and or a Return Material Authorization RMA which may include biohazard decontamination procedures and other product specific handling instructions and c then if applicable Buyer may return the defective Products to Seller with all costs prepaid by Buyer Replacement parts may be new or refurbished at the election of Seller All replaced parts shall become the property of Seller Shipment to Buyer of repaired or replacement Products shall be made in accordance with the Delivery provisions of the Seller s Terms and Conditions of Sale Consumables including but not limited to lamps fuses batteries bulbs and other such
239. rvice Menu Service Menu The Service menu appears only when the instrument is in the service mode To put the instrument into the service mode e Inthe Main Menu choose Instrument Controls gt Service Mode Advanced diagnostic functions are included in the service mode Meaningful data should not be collected when the instrument is in the service mode e Inthe Main Menu choose Service popopo pp Ctra AS E Pri i Flow Control Selection The Flow Control Selection menu is used to select the installed zero air 3 84 Model 146 Instruction Manual controller and gas controller The zero air and gas controller screens are used when installing new mass flow controllers for example when installing controllers other than the standard 10 slm for zero air or 100 sccm for gas The flow control selection is visible only when the instrument is in service mode For more information on the service mode see Service Mode earlier in this chapter e Inthe Main Menu choose Service gt Flow Control Selection Thermo Fisher Scientific Zero Air Flow Controller Gas Flow Controller Thermo Fisher Scientific Operation Service Menu Wi rhe MIA CHORDIE piyan TATU Bati haee E Eiet fast Bone i ade The Zero Air Flow Controller screen is used to select the zero air controller Allowable selections are 5 10 15 and 20 slm The zero air flow controller calibration points are reset to defaults when a new contr
240. rvice Mode on page 7 11 then return to the beginning of this step 4 At the Pressure Sensor Cal screen press to select Zero The Calibrate Pressure Zero screen appears 5 Waitat least 10 seconds for the zero reading to stabilize then press to save the zero pressure value 6 Disconnect the pump from the pressure transducer 7 Press Cum to return to the Pressure Sensor Cal screen 8 At the Pressure Sensor Cal screen press to select Span The Calibrate Pressure Span screen appears 9 Wait at least 10 seconds for the ambient reading to stabilize use gt and t to enter the known barometric pressure and press to save the pressure value 10 Reconnect the tubing to the pressure transducer 11 Install the cover Model 146 Instruction Manual 7 41 Servicing Ozonator Lamp Replacement Ozonator option 1 42 Ozonator Lamp Replacement Ozonator option A Model 146 Instruction Manual Use the following procedure to replace the ozonator lamp Figure 7 13 Equipment Required Ozonator lamp Allen wrench 3 32 inch Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unplug the ozonator lamp cable from the LAMP connector on ozonator board Figure 7 10 3 Slide
241. s These connection options include e Individual terminal board PCB assemblies e Terminal block and cable kits optional e Individual cables optional For detailed information on the optional connection components refer to the Optional Equipment chapter For associated part numbers refer to External Device Connection Components on page 7 7 The following terminal board PCB assemblies are available for Series instruments e I O terminal board PCB assembly 37 pin standard e D O terminal board PCB assembly 37 pin standard e 25 pin terminal board PCB assembly included with optional I O Expansion Board Thermo Fisher Scientific Installation Connecting External Devices 1 0 Terminal Board Figure 2 6 shows the recommended method for attaching the cable to the terminal board using the included tie down and spacer Table 2 1 identifies the connector pins and associated signals See Detail A See Detail B N d Sja ne b U SEI S Detail A Detail B Assembled Connector Figure 2 6 0 Terminal Board Views Table 2 1 O Terminal Board Pin Descriptions Pin Signal Description Pin Signal Description 1 Analog 13 Power_Fail_NC 2 Analog ground 14 Power_Fail_COM 3 Analog2 15 Power_Fail_NO 4 Analog ground 16 TTL_Input1 5 Analog3 17 TTL_Input2 6 Analog ground 18 TTL_Input3 7 Analog4 19 TTL_Input4 8 Analog ground 20
242. s a b cldle f This command sets the gas gas to span levels 0 1 2 3 4 or 5 The example below sets gas A to span level 1 Send sel gas a span 1 Receive set gas a span 1 ok gas gas span conc spanindex gas a b c dle f spanindex span index 0 6 1 5 reports the span concentration for each span level and 6 reports the span concentration for manual mode This command reports the target span concentration for the specified gas gas and the span level spanindex The concentration reported is in ppb Send gas a span conc 1 Receive gas a span conc 1 123 45 Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurement Configuration set gas gas span conc spanindex value gas a b c d e f spanindex span index 0 6 1 5 reports the span concentration for each span level and 6 reports the span concentration for manual mode value float This command sets the target span concentration value for the specified gas gas and the span level spanindex The concentration value is in ppb Note Changing span concentration may affect allowable limits for gas span flow It is advisable to check this setting A Send set gas a span conc 1 123 45 Receive set gas a span conc 1 123 45 ok gas gas span flow gas a b c d e f spanindex span index 0 6 0 is for zero flow This command reports the target total flow for the specified gas gas and span level spanindex Send gas a span flow 1 Receive ga
243. s gt Communication Settings gt TCP IP Settings gt IP Address Thermo Fisher Scientific Model 146 Instruction Manual 3 53 Operation Instrument Controls Menu Netmask Default Gateway Host Name 3 54 Model 146 Instruction Manual The Netmask screen is used to edit the netmask In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Netmask The Default Gateway screen is used to edit the gateway address In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Gateway The host name screen is used to edit the host name When DHCP is enabled this name is reported to the DHCP server In the Main Menu choose Instrument Controls gt Communication Settings gt TCP IP Settings gt Host Name Thermo Fisher Scientific 1 0 Configuration Output Relay Settings Thermo Fisher Scientific Operation Instrument Controls Menu pE EN En REP p perm oi Banas Be BE Fas fast Bane Bast ade 5 The I O Configuration menu deals with configuration of the instrument s I O system The analog input configuration is displayed only if the I O expansion board option is installed e Inthe Main Menu choose Instrument Controls gt I O Configuration Note The digital outputs may take up to one second after the assigned state occurs to show up on the outputs A m La nobel The Output Relay Settings me
244. s The Gas Selection screen is used to select the gas to mix with the Mix Selection permeation gas Gas D E and F are available if the six gas option is installed e In the Main Menu choose Flow Modes gt Permeation Oven gt Gas Mix Permeation Oven Gas The Gas Span Select screen is used to select the span level for the previously Span Select selected gas Gas D E and F are available if the six gas option is installed Thermo Fisher Scientific Model 146 Instruction Manual 3 13 Operation Flow Modes Once selected the screen reports the new flow mode The user must press to accept and activate these settings The message SAVING is displayed for 3 seconds then removed e In the Main Menu choose Flow Modes gt Permeation Oven gt Gas Mix gt Gas A F gt Zero or Span 1 5 Permeation Oven Manual The Permeation Oven Gas Manual Flow screen shows the zero air and gas Flow flows and is used to manually set them for the previously selected gas If the selected zero air flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Flow Modes gt Permeation Oven gt Gax Mix gt Gas A F gt Manual 3 14 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Zero Air Gas 0 Operation Flow Modes The Manual Gas Zero Air Flow screen sets the manual zero air flow value for the previously selected gas The
245. s a span flow 1 123 45 set gas gas span flow value gas a b c dle f spanindex span index 0 6 0 is for zero flow value float This command sets the target total flow value for the specified gas gas and span level spanindex Send set gas a span flow 1 123 45 Receive set gas a span flow 1 123 45 ok gas gas tank conc gas a b c d e f This command reports the tank concentration for gas gas Here concentration is reported in ppb Send gas a tank conc Receive gas a tank conc 123 45 set gas gas tank conc conc gas a b c d e f conc float Model 146 Instruction Manual B 31 C Link Protocol Commands Measurement Configuration B 32 Model 146 Instruction Manual This command sets the tank concentration for gas gas Here conc is stored in ppb Note Changing this value may affect allowable limits for gas span concentration and gas span flow It is advisable to check these settings A Send set gas a tank conc 123 45 Receive set gas a tank conc 123 45 ok 11 12 13 14 15 These commands report the percentage of the ozonator lamp drive for each of the five custom levels The example below shows that custom level 1 is 10 This command is only available if the ozonator option is installed Also if the photometer option is installed it must be configured for internal or external measurement mode Send 11 Receive 11 10 set 11 value set 12 value set 13 value set 14 value set 15 value The
246. s an overview of the software structure and includes a description of the system electronics and input output connections and functions Hardware on page 8 1 describes the analyzer components Software on page 8 7 provides an overview of the software organization and detailed information on the software tasks Electronics on page 8 8 describes the system boards assemblies and connectors I O Components on page 8 11 describes the input and output communication functions and components Hardware The Model 146 components include Figure 8 1 Thermo Fisher Scientific Flow system Photometer with photometer lamp optional e Pressure transducer Photometer lamp power supply Photometer temperature thermistor e Pump Sample Reference Solenoid Valves Ozonator assembly optional e Ozonator lamp power supply Permeation oven optional Model 146 Instruction Manual 8 1 System Description Flow System Motherboard Capillary Photometer DC Power Supply Ozonator Board Top Front Panel Connector Board Digital Output Board Photometer Board Bottom Photometer Th il ere al E nterface Boar 1 0 Expansion ED Board Optional GS E Front Panel Displa Capillary Pump dias y Fan Am __ Solenoid Sample Reference Pump bx ee E Capillary Ozonator Solen
247. s at 25 C an alternate procedure is to connect an accurately known 10K resistor to the thermistor input AMB TEMP on the measurement interface board and enter the temperature reading A 1 C change corresponds to a 5 change in resistance thus this alternative procedure can be quite accurate as a check however it clearly is not NIST traceable 3 From the Main Menu press to scroll to Service press gt to scroll to Temperature Calibration and press Ce The Calibrate Ambient Temperature screen appears Thermo Fisher Scientific Model 146 Instruction Manual 7 53 Servicing Service Locations 1 54 Service Locations Model 146 Instruction Manual Note If Service is not displayed refer to Accessing the Service Mode on page 7 11 then return to the beginning of this step 4 at least 10 seconds for the ambient reading to stabilize use gt and to enter the known temperature and peste a to save the temperature value 5 Install the cover For additional assistance Thermo Fisher Scientific has service available from exclusive distributors worldwide Contact one of the phone numbers below for product support and technical information or visit us on the web at www thermo com agj 1 866 282 0430 Toll Free 1 508 520 0430 International Thermo Fisher Scientific Chapters System Description This chapter describes the function and location of the system components provide
248. s number of short records stored in memory B 20 03 Reports ozone concentration measured by photometer B 9 03 bkg Reports sets photometer Oz background B 24 03 coef Reports sets photometer Os coefficient B 24 03 conc Sets ozonator target concentration under photometer control B 34 03 setting Reports ozonator target concentration under photometer B 35 B 6 Model 1467 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Commands Command Description Page control oz lamp temp Reports ozonator lamp temperature B 15 ozon Sets ozonator output level B 34 ozon perm off Sets both ozonator and permeation oven output to zero B 34 perm Reports sets permeation oven output level B 34 photometer temp Reports photometer bench temperature B 15 pres Reports photometer pressure B 9 pres comp Reports sets photometer pressure compensation on or off B 36 program Reports sets program on or off B 16 program no Reports instrument program number B 42 push Simulates pressing a key on the front panel B 25 range Reports sets photometer 0 range B 35 relay Reports sets relay logic status of designated relay s B 46 relay stat Reports relay logic status for the designated relay s B 46 resp coef Reports sets photometer control response coefficient B 25 ri right Simulates pressing right pushbutton B 26 ru run Simulates pressing run
249. screen identifies the minimum and maximum acceptable zero air flow rates and lets the user select a flow rate between these two settings If the selected zero air flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Flow Modes gt Permeation Oven gt Gas Mix gt Gas A F gt Manual gt Zero Air The Manual Gas Flow screen sets the manual gas flow value for the previously selected gas The screen identifies the minimum and maximum acceptable gas flow rates and lets the user select a flow rate between these two settings If the selected gas flow is lower or higher than the acceptable range TOO LOW or TOO HIGH appears on the screen e In the Main Menu choose Flow Modes gt Permeation Oven gt Gas Mix gt Gas A F gt Manual gt Gas The O flow mode menu is used to set the operational mode of the ozonator if the ozonator option is installed Selecting OFF reports the new flow mode The user must press to accept and activate these settings The message SAVING is displayed for 3 seconds then removed Model 146 Instruction Manual 3 15 Operation Flow Modes e Inthe Main Menu choose Flow Modes gt O3 Transfer Standard Five ozonator levels 1 5 can be set using the Ozonator Level Selection screen Once selected the screen reports the new flow mode The user must press to accept and activate these settings The message SA
250. scription 1 0 Components The photometer lamp power supply is similar to the ozonator power supply except for the operating temperature set point and the value of the resistor in series with the output The photometer is always on when the operating temperature is reached Software on off control is not required The digital output board connects to the motherboard and provides solenoid driver outputs and relay contact outputs to a connector located on the rear panel of the instrument Ten relay contacts normally open with power off are provided which are electrically isolated from each other Eight solenoid driver outputs open collector are provided along with a corresponding 24VDC supply pin on the connector The I O expansion board connects to the motherboard and adds the capability to input external analog voltage inputs and to output analog currents via a connector located on the rear panel of the instrument It contains local power supplies a DC DC isolator supply a sub processor and analog circuits Eight analog voltage inputs are provided with an input voltage range of OV to 10VDC Six current outputs are provided with a normal operating range of 0 to 20 mA The front panel connector board interfaces between the motherboard and the front panel mounted function key panel and Graphics display It serves as central location to tie the three connectors required for the function key panel the graphics display control lines and th
251. se Photometer Setup gt Averaging Time The Calibration Factors menu displays the calibration factors and is used to make minor compensations to the O concentration e Inthe Main Menu choose Photometer Setup gt Calibration Factors The O Background screen is used to perform a manual adjustment of the instruments zero background Before performing a background adjustment allow the analyzer to sample zero air until stable readings are obtained e Inthe Main Menu choose Photometer Setup gt Calibration Factors gt O3 Bkg Model 146 Instruction Manual 3 33 Operation Photometer Setup Menu 0 Coefficient The O Coefficient screen allows the O span coefficient to be manually changed while sampling span gas of known concentration Note The concentration value will show ERROR if the measured concentration is not a valid span value either higher than the selected range or 0 or lower A e Inthe Main Menu choose Photometer Setup gt Calibration Factors gt O3 Coef a a ip prer PLP Reset User Calibration The Reset User Calibration Default screen allows the user to reset the Default calibration configuration values to factory defaults e Inthe Main Menu choose Photometer Setup gt Calibration Factors gt Reset User Cal Defaults 3 34 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Operation Photometer Setup Menu Calibration The Calibration m
252. se commands set the percentage of the ozonator lamp drive for each of the five custom levels The example below sets the custom level 1 to 10 This command is only available if the ozonator option is installed Also if the photometer option is installed it must be configured for internal or external measurement mode Send set 11 10 Receive set 11 10 ok 11 conc 12 conc 13 conc l4 conc 15 conc These commands report the concentration setting of each of the five custom levels The example below shows that custom level 1 is 500 ppb This command is only available if the ozonator and photometer options are installed and the photometer is configured for ozonator control mode Thermo Fisher Scientific Thermo Fisher Scientific C Link Protocol Commands Measurement Configuration Send 11 conc Receive 11 conc 500 ppb set 11 value set 12 value set 13 value set 14 value set 15 value These commands set the concentration for each of the five custom levels The example below sets the custom level 1 to 500 ppb The example below shows that custom level 1 is 500 ppb This command is only available if the ozonator and photometer options are installed and the photometer is configured for ozonator control mode Send set level 1 500 Receive set level 1 500 ok set level level This command sets the ozonator output to level where level represents custom levels 1 2 3 4 or 5 The example below sets level 1 Send set level 1
253. se the following procedure to determine the permeation rate by weight Rate by Weight Loss loss 1 Make sure the oven has been calibrated as described in the Permeation Tube Oven Calibration procedure described previously 2 Insert the permeation tube carefully Do not touch with fingers 3 Turn on the instrument 4 Wait 48 72 hours for the permeation tube to stabilize 9 8 Model 146 Instruction Manual Thermo Fisher Scientific Determining Release Rate by Transfer Standard Thermo Fisher Scientific 10 Optional Equipment Permeation Oven Carefully remove the permeation tube from the oven and weigh to an accuracy of 0 1mg Perform this measurement as quickly as possible Replace the permeation tube into the oven of the instrument Repeat Steps 5 and 6 after two weeks Compute the weight loss of the permeation tube from the values determined in Steps 5 through 7 Repeat Steps 5 through 8 until the weight loss has been determined to a precision of 1 2 For the most accurate work use the permeation tube in the same oven that was used to determine the weight loss of permeation tube Use the following procedure to determine the release rate by transfer standard Make sure the oven has been calibrated as described in the Permeation Tube Oven Calibration procedure described previously Also make sure that the Transfer Standard has been properly calibrated Determine the permeation rate for the permeation
254. seful in easy modification of current long record short record or streaming data lists The example below copies the current list of long records into the scratch pad Send set copy lrec to sp Receive set copy lrec to sp ok Model 146 Instruction Manual B 21 C Link Protocol Commands Datalogging B 22 Model 1461 Instruction Manual sp field number This command reports the variable number and name stored at index in the scratch pad list The example below shows that the field 5 in the scratch pad is set to index number 13 which is for the variable pressure Send sp field 5 Receive sp field 5 13 pres set sp field number value number 1 32 is the maximum number of fields in long and short record lists number 1 18 is for streaming data lists This command sets the scratch pad field number item number in scratch pad list to value where value is the index number of a variable in the analog out variable list Available variables and their corresponding index numbers may be obtained using the command list var aout The set sp field command is used to create a list of variables which can then be transferred into the long record short record or streaming data lists using the set copy sp to lrec set copy sp to srec or set copy sp to stream commands Send set sp field 1 34 Receive set sp field 1 34 ok stream per This command reports the currently set time interval in seconds for streamin
255. settings Pressure User set alarm min value User set alarm max value Bench Temperature User set alarm min value User set alarm max value Lamp Temperature User set alarm min value User set alarm max value 03 Lamp Temperature User set alarm min value User set alarm max value Intensity User set alarm min value User set alarm max value Noise 0 0 100 0 Choose Signal to Output The Choose Signal Type to Output screen displays a submenu list of the analog output signal group choices Group choices are Concentrations Other Measurements and Analog Inputs if the I O expansion board option is installed This allows the user to select the output signal to the selected output channel The Concentrations screen is shown below See Table 3 4 below for a list of items for each signal group choice e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Analog Output Config gt Select Channel gt Choose Signal To Output 3 64 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu Table 3 4 Signal Type Group Choices Concentrations Other Measurements Analog Inputs None None None Gas Conc Target Gflow Analog Input 1 Da Conc Actual Gflow Analog Input 2 Perm Conc Target Zflow Analog Input 3 Actual Zflow Analog Input 4 Target Tflow Analog Input 5 Actual Tflow Analog Input 6
256. solenoids Refer to Component Layout on page 7 8 to locate these solenoids within the instrument If you need the solenoid part number refer to the Replacement Parts List on page 7 5 External Internal solenoid e Ozonator solenoid e Perm and Perm 2 solenoids e Reaction chamber solenoid e Sample Reference solenoid e Span gas solenoids e Zero air solenoid Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Use the following general replacement procedure to replace a solenoid Equipment required Solenoid Open end wrenches 5 16 1 2 9 16 and 5 8 inch Flatblade screwdriver Thermo Fisher Scientific Pressure Transducer Replacement Photometer option Thermo Fisher Scientific Servicing Pressure Transducer Replacement Photometer option Philips screwdriver 1 Turn instrument OFE unplug the power cord and remove the cover 2 Unplug the solenoid electrical cable 3 Disconnect all lines from defective solenoid 4 Remove the solenoid and replace with the new solenoid 5 Reconnect the electrical connection and the lines 6 Check for leaks after replacing any solenoid Refer to Leak Test and Pump Checkout in this chapter Use the following procedure to replace the pressure transducer Refer to Component Layout on page 7 8 to locate the p
257. sresesesreseste C 1 Funcion Codesoso eei a RAEE AEEA Ea E antes C 1 MODBUS Commands Supported sa C 1 Model 146 Instruction Manual xi Contents xii Model 1467 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Figures Figures Figure 1 1 Standard Gas Dilution SySteM oooooocicconiconinconincononocccnrncnornncnoos 1 1 Figure 1 2 Standard Gas Dilution System with Ozonator Option 1 1 Figure 1 3 Standard Gas Dilution System with Permeation Tube Oven Option1 1 Figure 1 4 Standard Gas Dilution with Ozonator and Permeation Tube Oven A E E ties hacia N 1 1 Figure 1 5 Standard Dilution System with Photometer Option 1 1 Figure 1 6 Standard Gas Dilution System with Ozonator Permeation Tube Oven and Photometer Options dia 1 1 Figure 2 1 Remove the Packing Material ooooooionocoiciciciocincnninonnnncnrnnnnnonoso 2 2 Figure 2 2 Remove the Shipping Screws o ooconcicincininccnnocinicnnnnononinncnnoncnonoso 2 2 Figure 2 3 Removing the Glass Chamber Assembly o o o 2 2 Figure 2 4 Replacing the Glass Chamber Assembly o o 2 2 Figure 2 5 Model 1467 Rear Panel oooococoocicicoccccincncncnnnnononncnrnononocnrocoronnononoso 2 2 Figure 2 6 O Terminal Board Views oocooocciconiccccnoconinccnnnccnrnccnrnccncnrnnnnoos 2 2 Figure 2 7 D O Terminal Board ViewS ooocoocicocinconincon
258. sure sensor inputs e Detector board inputs e Internal temperature thermistor e Photometer temperature thermistor e Ozonator lamp power supply e Photometer lamp power supply e Permeation oven temperature thermistor and heater control The pressure sensor assembly consists of a board containing an instrumentation amplifier and a pressure transducer with a gas input fitting The pressure transducer output is produced by measuring the pressure difference between the sample gas pressure and ambient air pressure The ozonator lamp power supply board includes the circuits that generate high voltage AC to control the lamp heater A transformer outputting high voltage at approximately 15 kHz drives the lamp A resistor in series with the output limits the lamp current The voltage output is adjustable under software control for lamp intensity control The ozonator lamp can be switched on and off under software control The lamp is heated to a specific temperature to ensure reliable operation The oscillator circuit that drives the high voltage output is inhibited until this operating temperature is reached A power transistor is used to heat the lamp housing Thermistors sense the housing temperature for heater control and for status information Thermo Fisher Scientific Photometer Lamp Power Supply Digital Output Board 1 0 Expansion Board Optional Front Panel Connector Board 1 0 Components Thermo Fisher Scientific System De
259. t If a gas span value is also selected then the output from the permeation oven will be determined by the flow rates required for the gas dilution and will not be the values shown here If installed the ozonator mode is displayed on the next line The allowable modes are OFF OZONE 1 through 5 and manual The value to the right of this shows the output level This level is displayed in units or PPM units depending on the photometer mode The last line shows the photometer mode if installed The allowable modes are OFF INT EXT and CTL In INT mode the photometer is configured to take sample from the manifold and report the ozone concentration In EXT mode the photometer takes sample from a rear panel bulkhead fitting and reports the ozone concentration In CTL mode the photometer is configured the same as INT mode the reaction chamber is bypassed and the ozonator drive signal is controlled to provide a pre set ozone concentration This concentration is maintained by constantly updating the drive signal to the ozonator lamp based on the measured concentration Thermo Fisher Scientific Operation Flow Modes Note Changing the span concentration may affect allowable limits for total flow TELOW It is advisable to check this setting A Detailed information about each of the above options can be found in the appropriate menu screens For more information on options see Chapter 1 Introduction Note If changes are made
260. t Controls Menu Events 1 32 The Events 1 32 screens allow the user to turn events on or off and modify the calibrator s operational settings These screens are similar to the operation screen but the numeric values cannot be changed except duration minutes Duration minutes lets you set the duration of an event and can be set by pressing Y Jandusing f enter a value in minutes Press 4 J to save the value or e to leave the screen User 1 8 allows the instrument to control the user configured digital outputs as part of the event When set to NONE it will leave that event alone it stays the same as it was before the event started If set ON it will turn user outputs on If set OFF it will turn outputs to off e Inthe Main Menu choose gt Program gt Events gt Event 1 32 e Press gt Jor e to toggle and to set following items Instrument Co ntrols The Instrument Controls menu contains a number of items The software M enu controls listed in this menu enable control of the listed instrument functions e Inthe Main Menu choose Instrument Controls Thermo Fisher Scientific Model 146 Instruction Manual 3 39 Operation Instrument Controls Menu Temperature The Temperature Compensation screen is used to turn temperature Compensation compensation on or off When temperature compensation is on the display shows the current ambient instrument temperature measured by a thermistor on the Interface board W
261. t in local mode Defective solenoid connector board Defective measurement interface board Defective cable Put instrument in remote mode Replace solenoid connector board Replace measurement interface board Replace cable Thermo Fisher Scientific Thermo Fisher Scientific Table 6 1 Troubleshooting General Guide continued Malfunction Possible Cause Defective connector Troubleshooting Troubleshooting Guides Action Replace connector Calibration as measured at output of mass flow controller does not agree with calibration as measured at output of instrument Leak Mass flow controller not calibrated Perform a leak check see Servicing chapter Calibrate mass flow controller No ozone output Lamp failure Ozonator heater failure Ozonator board failure Check for blue light when removing ozone lamp from ozonator and replace lamp if light not visible Check to see that ozonator is warm gt 50 C If not warm replace heater or heat power supply see Servicing chapter Replace ozonator board see Servicing chapter Low ozonator output Leak in ozonator or distribution manifold Flow excessively high Ozonator board failure Weak lamp Check for leaks in ozonator and manifold system see Servicing chapter and repair leak Check zero air flow valve and reduce flow Replace ozonator board see Servicing chapter
262. t specifiers and converted to uniform internal representations 32 bit floats or 32 bit integers These values are converted into text for display on the screen using the format specifier for the front panel display Normally the specifier used to parse a particular datum from the input stream will be strongly related to the specifier used to display it such as all of the floating Model 146 Instruction Manual B 47 C Link Protocol Commands Record Layout Definition Format Specifier for ASCII Responses Format Specifier for Binary Responses B 48 Model 146 Instruction Manual point inputs will be displayed with an f output specifier and all of the integer inputs will be displayed with a d specifier The first line of the Layout response is the scanf like parameter list for parsing the fields from an ASCII ERec response Parameters are separated by spaces and the line is terminated by a n the normal line separator character Valid fields are s parse a string d parse a decimal number ld parse a long 32 bit decimal number f parse a floating point number x parse a hexadecimal number lx parse a long 32 bit hex number ignore the field Note Signed versus unsigned for the integer values does not matter it is handled automatically A The second line of the Layout response is the binary parameter list for parsing the fields from a binary response Parameters MUST be separated by spaces and t
263. te settings with a maximum of five span levels The span concentration and or total flow rate can be selected using this menu to change these settings e Inthe Main Menu choose Gas Setup gt Gas A F gt Span 1 5 Thermo Fisher Scientific Model 146 Instruction Manual 3 23 Operation Gas Setup Menu Span Concentration Span Flow 3 24 Model 146 Instruction Manual The Span Concentration screen displays the minimum and maximum concentration for spans 1 through 5 of each gas Minimum and maximum values are calculated based upon the tank concentration and the mass flow controller The desired concentrations can be set between these two limits If the concentration is entered outside of the minimum or maximum values the message TOO LOW or TOO HIGH appears on the screen Note Changing this valve may affect allowable limits for span flow It is advisable to check this setting A e Inthe Main Menu choose Gas Setup gt Gas A F gt Span 1 5 gt Conc The Span Flow screen displays the minimum and maximum flows for spans 1 through 5 of each gas Minimum and maximum values are calculated based upon the tank concentration and the mass flow controller The desired concentrations can be set between these two limits If the concentration is entered outside of the minimum or maximum values the message TOO LOW or TOO HIGH appears on the screen e Inthe Main Menu choose Gas Setup gt Gas A F gt Span 1 5 gt
264. te Reports sets current date B 38 default params Sets parameters to default values B 38 dhcp Reports sets state of use of DHCP on Ethernet port B 40 diag volt ob Reports diagnostic voltage levels for I O expansion board B 14 diag volt mb Reports diagnostic voltage levels for motherboard B 14 diag volt mib Reports diagnostic voltage levels for measurement interface B 14 board dig in Reports status of the digital inputs B 44 din Reports sets active logic state for the digital input channel B 44 do down Simulates pressing down pushbutton B 25 dout Reports sets active logic state for the digital output channel B 45 dtoa Reports outputs of the digital to analog converters per B 45 channel en enter Simulates pressing enter pushbutton B 25 er Returns a brief description of the main operating conditions B 17 in the format specified in the commands erec Returns a brief description of the main operating conditions B 18 in the format specified in the command erec format Reports sets erec format ASCII or binary erec layout Reports current layout of erec data event params Reports current event that is being executed event status Reports the parameters programmed for the specified event o 20 ae gt J AJ oOo wo flags Reports 8 hexadecimal digits or flags that represent the 9 status of the instrument format Reports sets current reply termination format B 41 gas conc Returns current
265. ter the source voltage and press to calibrate the input voltage for the selected channel to the source vo tage 6 Press EE gt to return to the input channels display and repeat Steps 3 5 to calibrate other input channels to the source voltage as necessary 1 0 Expansion Board Use the following procedure to replace the optional I O expansion board Replacement Cisvre 7 3 Optional Note After replacing the optional I O expansion board calibrate the current outputs and the analog voltage inputs See the Analog Output Calibration on page 7 21 and the Analog Input Calibration on page 7 22 in this chapter Equipment Required I O expansion board Nut driver 3 16 inch amounts of static electricity A properly grounded antistatic wrist strap must Equipment Damage Some internal components can be damaged by small be worn while handling any internal component 1 Turn instrument OFF unplug the power cord and remove the cover 2 Unplug the I O expansion board cable from the EXPANSION BD connector on the motherboard 7 24 Model 146 Instruction Manual Thermo Fisher Scientific Servicing 1 0 Expansion Board Replacement Optional 3 Remove the two standoffs holding the I O expansion board connector to the rear panel Figure 7 9 4 Pop off the board from the mounting studs and remove the board 5 To install the 1 O expansion board follow previous steps in reverse Motherboard Support Bracket
266. the phone numbers below for product support and technical information or visit us on the web at www thermo com aqi 1 866 282 0430 1 508 520 0430 iv Model 146i Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Chapter 1 Chapter 2 Chapter 3 Contents INTO UCUON ria 1 1 Pincipleof Operation cagas acopio ia ette bt naa dd elated 1 1 Dilo l 1 1 ON A A AEE 1 1 Gas Phase TO SEA a 1 1 Permeation we At LAR 1 1 Photometrta dada 1 1 e O ncaa acta ta oer ed 1 1 Installation eres ee ee eee 2 1 Lionel dd Shar E E obli iii 2 1 Unpacking and Inspeccion lied 2 1 Setup Procedure ia o E is ae 2 1 Remote Interconnections occocccccononononononononenenonononenenanananananananananananos 2 1 PS AE NO 2 1 CUE a o 2 1 Connecting External Devices inicio raid 2 1 Terminal Board PCB Assemblies cooccononcconnnccconnnonanaconancnonanaconanacos 2 1 Startu pn n a o id 2 1 MA A A AA 3 1 Dip cta add ad ac 3 1 POS DUROS AS o o 3 1 O Tc 3 1 Software OEI SA 3 1 PowerUp SEED A Sa 3 1 Run creta ada las 3 1 M n Ment A dante aA AAR ee bee tinted Soaks AEAEE AENA as SEREA 3 1 CP ERATION aa 3 1 Flow Modesa a ata a oia 3 1 Stnd b e de e O 3 1 Cas DION ia 3 1 Permeacon VE EET E E E ia ATI IAE 3 1 O A O 3 1 A A A Re 3 1 Gas Setup Men iii 3 1 Gas Name AAN AN es 3 1 A A O aasia 3 1 Model 146 Instruction Manual vw Contents Tank Concentration epi o E 3 1 Zer AIT porran a dtauay a Ne 3 1 Span e iron a a ea ar A eaae
267. the zero calibration A e In the Main Menu choose Service gt Pressure Calibration gt Zero Calibrate Pressure Span The Calibrate Pressure Span screen is used to view and set the pressure sensor calibration span point Note The plumbing going to the pressure sensor should be disconnected so the sensor is reading ambient pressure before performing the span calibration The operator should use an independent barometer to measure the ambient pressure and enter the value on this screen before calibrating A e Inthe Main Menu choose Service gt Pressure Calibration gt Span 3 90 Model 146 Instruction Manual Thermo Fisher Scientific Operation Service Menu ty CORDED PAE TA TEN Brave ie i Bati ha 5 E Et fast Restore Default The Restore Default Calibration screen is used to reset the pressure Calibration calibration configuration values to factory defaults e Inthe Main Menu choose Service gt Pressure Calibration gt Set Defaults Ambient The Ambient Temperature Calibration screen is used to view and set the Temperature ambient temperature sensor calibration The ambient temperature calibration is visible only when the instrument is in service mode For more Calibration information on the service mode see Service Mode earlier in the chapter Note This adjustment should only be performed by an instrument service technician A e Inthe Main Menu choose Service gt Ambient Temp Calibration T
268. tific Troubleshooting Connector Pin Descriptions Table 6 3 Motherboard Connector Pin Descriptions continued Connector Label Reference Designator Pin oOo ao N 0 on A Signal Description Ground Ground Ground SPI Reset SPI Input SPI Output SPI Board Select SPI Clock EXT RS485 J15 oOo 0 N oon A wo N gt gt gt gt gt a ow N RS485 to Rear Panel RS485 to Rear Panel 5V 5V 5V Ground Ground Ground NC NC 24V 24V 24V 24V 24V 24V MONITOR J17 24V Power Monitor Ground FRONT PANEL BD J18 YN Oo oO FP W N Ground Ground LCLK LCD Signal Ground Ground LLP LCD Signal LFLM LCD Signal Model 146 Instruction Manual 6 13 Troubleshooting Connector Pin Descriptions 6 14 Model 146 Instruction Manual Table 6 3 Motherboard Connector Pin Descriptions continued Connector Reference Pin Signal Label Designator Description 8 LD4 LCD Signal 9 LDO LCD Signal 10 LD5 LCD Signal 11 LD1 LCD Signal 12 LD6 LCD Signal 13 LD2 LCD Signal 14 LD7 LCD Signal 15 LD3 LCD Signal 16 LCD Bias Voltage 17 5V 18 Ground 19 Ground 20 LCD_ONOFF LCD Signal 21 Keypad Row 2 Input 22 Keypad Row 1 Input 23 Keypad Row 4 Input 24 Keypad Row 3 Input 25 Keypad Col 2 Select 26 Keypad Col 1 Select 27 Keypad Col 4 Select 28 Keypad Col 3 Select 29 Ground 30 Ground 31 G
269. ting studs and remove the board by lifting it up and off the slotted bottom support 4 Replace the front panel board by following previous steps in reverse LCD Module j Retaining Screw LCD Module Screws 2 Front Panel Board Mounting Studs 2 Ribbon Cables 3 ae Bottom Support Figure 7 11 Replacing the Front Panel Board and the LCD Module LCD Module Use the following procedure to replace the LCD module Figure 7 11 Replacement Equipment Required Thermo Fisher Scientific Model 146 Instruction Manual 7 29 Servicing LCD Module Replacement 7 30 Model 146 Instruction Manual LCD module Philips screwdriver CAUTION If the LCD panel breaks do not to let the liquid crystal contact your skin or clothes If the liquid crystal contacts your skin or clothes wash it off immediately using soap and water Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component Do not remove the panel or frame from the module The polarizing plate is very fragile handle it carefully Do not wipe the polarizing plate with a dry cloth it may easily scratch the plate Do not use alcohol acetone MEK or other Ketone based or aromatic solvents to clean the module use a soft cloth moistened with a naphtha cleaning solvent a Do not place the module near organic s
270. tion The command string should normally contain print style formatting to include the user input Ifa is present it indicates a command which is sent to the instrument upon successful completion of the button command to update the value field This is not currently used Some examples n is the C syntax for an end of line character Concentrations n This is a single text only line An This is a single blank line NO 3s n This is a line which appears slightly indented The text field is NO the value is taken from the third element of the data response and interpreted as a string NO 18sBd ddd set no coef s n This is a line which also appears slightly indented The next field is also NO but the value is taken from the eighteenth element of the data response again interpreted as a string A button appears on this line which when pressed pops up an input dialog which will state Please enter a new value for NO using a d ddd format The string entered by the user is used to construct the output command If the user enters for example 1 234 the constructed command will be set no coef 1 234 NO 21f Code_O Code 1 Code 2 Code 3 Code 4 Code 5 Code 6 Code_7 Code _8 Code_9 Code 10 Code 11 Lset range no d n This is a line which appears slightly indented the title is again NO and the value the twenty first element of the data response interpreted as a floating point number There is a no tr
271. tions Table 6 7 Measurement Interface Board Connector Pin Descriptions continued Connector Label Reference Designator Pin Signal Description 15V DET J5 o om A WO N Frequency Input 1 15V Ground Frequency Input 2 15V Ground OZONATOR J6 Ozonator Lamp Intensity Control Voltage Output Ozonator Lamp Intensity Control Voltage Input Ozonator Heater Thermistor 24V Ground Ozonator Lamp On off Control 0 N o o A WwW Ground Ground PHOTOMETER J7 0 N oon A WwW Photometer Lamp Intensity Control Voltage Output Photometer Lamp Intensity Control Voltage Input Photometer Heater Thermistor 24V Ground Photometer Lamp On Off Control Ground Ground FAN 24V Fused Ground FAN 24V Fused Ground 24V IN 24V Ground DATA Ground Model 146 Instruction Manual 6 21 Troubleshooting Connector Pin Descriptions Table 6 7 Measurement Interface Board Connector Pin Descriptions continued Connector Label Reference Designator Pin Signal Description RS485 from Motherboard RS485 from Motherboard AMB TEMP Ambient Temperature Thermistor Ground PERM TEMP Perm Oven Gas Thermistor Ground PERM OVEN Ee N Perm Oven Heater On Off 24V Perm Oven Thermistor Ground MFC1 Mass Flow Controller 1 Input Mass Flow Controller 1 Output Ground
272. to choose the instrument action that is tied to the selected digital input e Inthe Main Menu choose Instrument Controls gt I O Configuration gt Digital Input Settings gt Select Relay gt Instrument Action External Solenoid The Model 146 can drive up to eight external solenoids You can select the Settings external solenoids with the External Solenoids menu Each solenoid can be connected to the following Gas A B or C or D E or F if the six gas option is installed the ozonator or the permeation oven e Inthe Main Menu choose Instrument Controls gt I O Configuration gt External Solenoids Thermo Fisher Scientific Model 146 Instruction Manual 3 59 Operation Instrument Controls Menu Logic State The Logic State screen is used to change the I O solenoid to either normally open or normally closed e Press to toggle and set the logic state open or closed 88 8 P Instrument State The Instrument State menu allows the user to select the instrument state that is tied to the selected solenoid output A submenu lists signal types of either alarm or non alarm to choose from e In the Main Menu choose Instrument Controls gt I O Configuration gt Ext Solenoid Settings gt Select Solenoid gt Instrument State porie p pur ip priori E E 3 60 Model 146 Instruction Manual Thermo Fisher Scientific Thermo Fisher Scientific Operation Instrument Controls Menu Alarms The Alarms status
273. trument s front panel The change password is shown if the instrument is unlocked e Inthe Main Menu choose Password gt Change Password 3 104 Model 146 Instruction Manual Thermo Fisher Scientific Operation Remove Password Remove Password The Remove Password screen is used to erase the current password and disable password protection The remove password is shown if the instrument is unlocked and the password set e Inthe Main Menu choose Password gt Remove Password WH mid Unlock Instrument The Unlock Instrument screen is used to enter the password to unlock the front panel The unlock instrument is shown if the instrument is locked e Inthe Main Menu choose Password gt Unlock Instrument Wi mbit Thermo Fisher Scientific Model 146 Instruction Manual 3 105 Chapter 4 Calibration Mass Flow in order to calibrate the mass flow meter section of the zero or gas mass flow Controllers controller a NIST traceable flow meter is required The term calibration means determining the actual flow versus the flow setting for seven equally spaced flows along the range of the device The Model 146 then corrects the output according to an internal algorithm Calibration may be done with a properly calibrated flow meter For the most accurate calibration procedure use a volumetric NIST traceable calibrator with the following step by step calibration procedure 1 Connect a source of clean dry air to the inl
274. ttom Capillary Photometer Photometer D C Power Supply Front Panel Connector Board Measurement 1 0 Expansion ay Board Optional SSL Interface Board Front Panel Display Capillary Pump Fan Pump EA Solenoids Span Gas Solenoid External Internal Mixing Chamber A Manifold Solenoid Zero mE m Solenoid Sample Reference Capillary Ozonator Ozonator Solenoid Ozonator Capillary Pressure Gauge lr Solenoid Perm 1 r Solenoid Perm 2 Mass Flow Controllers Back Pressure Pressure Regulator Transducer Fan Reaction Chamber Solenoid Reaction Chamber Figure 7 2 Component Layout 7 8 Model 146 Instruction Manual Pressure Regulator Permeation Oven Thermo Fisher Scientific Servicing Removing the Measurement Bench and Lowering the Partition Panel Removing the The measurement bench can be removed and the partition panel can be Measurement Bench lowered to improve access to connectors and components Refer to the and Low ering the following steps when a procedure requires lowering the partition panel see hel Figure 7 3 Partition Panel Remove Screw Z SLIDE BENCH OUT REMOVE SCREWS Pass Through
275. tube using the following procedure applicable only if the permeation option is installed a Remove the instrument cover and locate the permeation oven Component Layout on page 7 8 b Release the latches on the sides of the oven cover and remove cover c Remove glass chamber assembly Figure 2 3 by loosening not removing knurled screw located at the top of the chamber and gently pulling assembly upward Completely remove assembly from oven Thermo Fisher Scientific Model 146 Instruction Manual 2 3 Installation Setup Procedure ay Knurled Screw Glass Chamber Assembly Oven Top Oven Enclosure Figure 2 3 Removing the Glass Chamber Assembly d Separate glass chamber from top assembly by twisting and gently pulling glass away from top Figure 2 3 Keep glass clean by using Kimwipes or similar material to handle glass e Place permeation tube in chamber f Attach glass chamber to top assembly by gently pushing together with a slight twisting motion Figure 2 3 g Replace glass chamber assembly into oven until top of assembly is flush or slightly below oven top Figure 2 4 2 4 Model 1467 Instruction Manual Thermo Fisher Scientific Installation Setup Procedure Permeation Oven Top Glass Chamber Assembly N 1 See Procedure Permeation Oven Bottom _See Expanded View m Expanded View y Figure 2 4 Rep
276. u choose Flow Modes gt Photometer Thermo Fisher Scientific Operation Gas Setup Menu Gas Setup Menu The Gas Setup menu is used to configure flow for each gas The Gas A B or C menus assign a name solenoid and tank concentration to each gas The gas menus set the zero air flow span 0 flow and the total flow output for a maximum of five span settings for each gas Additionally the gas menus manually set the zero air and gas flows for each gas Gas D E and F are available if the six gas option is installed e Inthe Main Menu choose Flow Modes gt Gas Setup Wi pyirpri Gas Name The Gas Name screen is used to identify each gas As initially delivered the instrument lists the gas standards using the letters A B C and D E Fif the six gas option is installed These characters can be changed to represent Thermo Fisher Scientific Model 1467 Instruction Manual 3 21 Operation Gas Setup Menu Gas Solenoid Tank Concentration 3 22 Model 146 Instruction Manual actual chemical symbols of the gases used by each for example SO2 NO CO2 or any other representation up to five characters Once a name is selected the screen reports that the new name has been accepted and activated The message SAVING is displayed for 3 seconds then removed e Inthe Main Menu choose gt Gas Setup gt Gas A F gt Gas Name The Gas Solenoid screen lets you select a solenoid to control each of the gases Default sol
277. ument includes a sample solenoid valve and a reference solenoid valve The instrument can be configured to measure Oz every ten seconds standard cycle or every four seconds fast cycle Every cycle the two solenoid valves switch the sample gas and reference gas streams between the two cells One cell contains sample gas the other cell contains reference gas and vice versa During a standard cycle the cells are purged for seven seconds and measurements are taken for three seconds During a fast cycle the cells are purged for three seconds and measurements are taken for one second During every cycle the instrument calculates the natural logarithm of the sample gas frequency to reference gas frequency ratio in each cell and averages these results to optimize instrument accuracy These logarithms provide the basis for calculating the ozone concentrations The logarithms are stored in a rolling stack of 30 logarithms which are averaged by the instruments averaging time function The background values for Oz are corrected for temperature pressure and span and are subtracted from the ozone calculated to yield a corrected value The front panel display serial and Ethernet data ports and analog outputs are used to communicate the results of the measurement calculations The front panel displays the Oz concentration The display is updated every ten seconds standard cycle or every four seconds fast cycle The analog output ranges are user
278. ure Thermo Fisher Scientific Optional Equipment Permeation Oven 15 Go to the High Point screen enter the exact value of the resistor and press to calibrate the high resistance point 16 Press to return to the Factory Cal Oven Therm menu 17 Remove resistor from POJ3 and re attach the permeation oven 18 Perform a thermistor calibration using the Permeation Tube Oven Calibration procedure that follows There are two general approaches that can be used to calibrate the permeation tube oven The first is to calibrate the temperature indicator very accurately to better than 0 02 C and to use a permeation tube whose weight loss has been previously determined at that temperature Note An error of approximately 0 1 C corresponds to an error of 1 in release rate a The second approach is to note that the thermistors used to measure temperature are interchangeable to better than 0 2 C Thus a 1 resistor of the proper value 4 369 KQ for 45 C can be used to set the span on the measurement interface board The release rate for the permeation tube is then determined by weight loss in the actual oven being used You can use either of the two calibration methods presented here One method involves performing the Setting Perm Oven Temperature procedure and then continuing with the Setting Temperature with Water Bath procedure Alternatively you can perform the Setting Perm Oven Temperature y
279. ure Limits the minimum O lamp temperature alarm limit The minimum and maximum O lamp temperature screens function the same way e Inthe Main Menu choose Alarms gt O3 Lamp Temp gt Min or Max WH moi Thermo Fisher Scientific Model 146 Instruction Manual 3 79 Operation Alarms Menu Permeation Oven Temperature Min and Max Permeation Oven Temperature Limits Pressure 3 80 Model 146 Instruction Manual The Permeation Oven Temperature displays the current permeation gas temperature and sets the minimum and maximum alarm limits Acceptable alarm limits range from25 to 50 C If the permeation oven temperature reading goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main Menu e Inthe Main Menu choose Alarms gt Perm Oven Temp The Minimum Permeation Oven Temperature alarm limit screen is used to change the minimum bench temperature alarm limit The minimum and maximum permeation oven temperature screens function the same way e Inthe Main Menu choose Alarms gt Perm Oven Temp gt Min or Max The Pressure screen displays the current reaction chamber pressure reading and sets the minimum and maximum alarm limits Acceptable alarm limits range from 200 to 1 000 mmHg If the pressure reading goes beyond either the minimum or maximum limit an alarm is activated The word ALARM appears in the Run screen and in the Main
280. urrent time 24 hour time The example below reports that the internal time is 2 15 30 pm Thermo Fisher Scientific Communications Configuration Thermo Fisher Scientific C Link Protocol Commands Communications Configuration Send time Receive time 14 15 30 set time hh mm ss hh hours mm minutes ss seconds This command sets the internal clock 24 hour time The example below sets the internal time to 2 15 pm Note If seconds are omitted the seconds default to 00 A Send set time 14 15 Receive set time 14 15 ok addr dns This command reports the TCP IP address for the domain name server Send addr dns Receive addr dns 192 168 1 1 set addr dns address This command sets the dns address where address consists of four numbers gt ranging from 0 255 inclusive separated by Send set addr dns 192 168 1 1 Receive set addr dns 192 168 1 1 ok addr gw This command reports the default TCP IP gateway address Send addr gw Receive addr gw 192 168 1 1 set addr gw address This command sets the default gateway address where address consists of gt four numbers ranging from 0 255 inclusive separated by Send set addr gw 192 168 1 1 Receive set addr gw 192 168 1 1 ok addr ip This command reports the IP address of the analyzer Send addr ip Receive addr ip 192 168 1 200 Model 146 Instruction Manual B 39 C Link Protocol Commands Communications Configurati
281. used to reset all of the datalogging field items to default values For more information about selecting the analog output signal group choices see Select Content above e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Reset to Default Content DH ind PLE Configure Datalogging The Configure Datalogging menu deals with datalogging configuration for the currently selected record type e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Configure Datalogging 3 46 Model 146 Instruction Manual Thermo Fisher Scientific Operation Instrument Controls Menu E EN En REP ED Pm Tit AS BE Fas fast Basse Bast ade 5 m Lal mob Logging Period Min The Logging Period Min screen is used to select the logging period in minutes for the record format srec or Irec List of choices include off 1 5 15 30 and 60 minutes default e Inthe Main Menu choose Instrument Controls gt Datalogging Settings gt Configure Datalogging gt Logging Period Min y Moro MO RT a El Cue MA Memory Allocation The Memory Allocation Percent screen is used to select the percentage of Percent cach record type for both short records and long records Percentages between 0 and 100 are available in increments of 10 Changing this value results in log erasure for both short records and long records e Inthe Main Menu choose Instrument Controls gt Datal
282. when connecting the analyzer to other remote devices As a general rule when the connector of the host remote device is female a straight cable is required and when the connector is male a null modem cable is required Data Format 1200 2400 4800 9600 19200 38400 57600 or 115200 BAUD 8 data bits 1 stop bit no parity All responses are terminated with a carriage return hex 0D Refer to Table 8 1 for the DB9 connector pin configuration Table 8 1 RS 232 DB Connector Pin Configurations DB9 Pin Function 2 ono co JJ wo RX TX RTS CTS Ground The instrument uses a four wire RS 485 configuration with automatic flow control SD Refer to Table 8 2 for the DB9 connector pin configuration Table 8 2 RS 485 DB Connector Pin Configuration DB9 Pin Function 2 on wo JJ receive receive transmit transmit ground Thermo Fisher Scientific Ethernet Connection External Accessory Thermo Fisher Scientific Connector System Description 1 0 Components An RJ45 connector is used for the 10Mbs Ethernet connection supporting TCP IP communications via standard IPV4 addressing The IP address may be configured for static addressing or dynamic addressing set using a DHCP server Any serial port protocols may be accessed over Ethernet in addition to the serial port The external accessory connector is not used in the Model 146i calibrator This port is used in other models to communicate
283. y and reliability In the Model 1467 these components are integrated into a single microprocessor controlled unit The instrument can be controlled locally via the front panel interface or remotely using a datalogger or computer The standard Model 1467 hardware plumbing configuration comprises gas and zero air flow controllers as shown in Figure 1 1 The basic unit can handle three gas standards controlled by individual solenoids a six gas option is also available The zero air flow and the gas flow are regulated by mass flow controllers The zero air controller is high flow typically 10 slm full scale The gas flow controller is low flow typically 100 sccm A Teflon mixing chamber is used to achieve complete mixing of the two components at the desired concentration level This hardware plumbing configuration allows precision gas dilution The Model 146 can be equipped with an optional internal ozone generator The hardware plumbing configuration of the standard instrument with the ozone generator is shown in Figure 1 2 The generator Model 146 Instruction Manual 1 1 Introduction Principle of Operation Gas Phase Titration Permeation Oven 1 2 Model 1461 Instruction Manual produces ozone by exposing air to light at 185nm The ozone level is changed by varying the intensity of the lamp and varying the zero air flow With all of the gas standard solenoids deactivated the Model 146i can be used as a transfer standard ozone gen
284. y be impaired Equipment Damage Some internal components can be damaged by small amounts of static electricity A properly grounded antistatic wrist strap must be worn while handling any internal component a 1 Remove the instrument cover 2 Tape the thermistor Figure 7 12 to a calibrated thermometer Note Since the thermistors are interchangeable to an accuracy of 0 2 C and have a value of 10K ohms at 25 C an alternate procedure is to connect an accurately known 10K resistor to the thermistor input TEMP on the measurement interface board and enter the temperature reading A 1 C change corresponds to a 5 change in resistance thus this alternative procedure can be quite accurate as a check however it clearly is not NIST traceable 3 From the Main Menu press _ to scroll to Service gt press gt to scroll to Temperature Calibration gt and press The Calibrate Bench Temp screen appears Thermo Fisher Scientific Model 146 Instruction Manual 7 37 Servicing Solenoid Replacement Solenoid Replacement 7 38 Model 146 Instruction Manual Note If Service is not displayed refer to Accessing the Service Mode on page 7 11 then return to the beginning of this step 4 Wait at least 10 seconds for the reading to stabilize use and to enter the known temperature and press to save the temperature value 5 Install the cover The Model 1467 includes the following
285. yout ack Receive set layout ack ok analog iout range channel This command reports the analog current output range setting for channels where channel must be between 1 and 6 inclusive The example below reports current output channel 4 to the 4 20 mA range according to Table B 11 This command responds with feature not enabled if the I O expansion board is not detected Send analog iout range 4 Receive analog iout range 4 2 set analog iout range channel range This command sets analog current output channel to the channel range where channel is between 1 and 6 inclusive and range is set according to Table B 11 The example below sets current output channel 4 to the 0 20 mA range This command responds with feature not enabled if the I O expansion board is not detected Send set analog iout range 4 1 Receive set analog iout range 4 1 ok Table B 11 Analog Current Output Range Values Range Output Range 1 0 20 mA 2 4 20 mA 0 cannot be set to this but may report Undefined analog vin channel This command retrieves the analog voltage input channel data both the calculated value and the actual voltage In the example below the calculated value of channel 1 is 75 325 degrees E volts are 2 796 This command responds with feature not enabled if the I O expansion board is not detected Model 146 Instruction Manual B 43 C Link Protocol Commands 1 0 Configuration B 44 Model 1461 Instructio
Download Pdf Manuals
Related Search
Related Contents
Whirlpool WFW87HEDC Use and Care Manual r USER MANUAL OM, DXR250, DXR270, DXR300, DXR310, 2014-09 GETTING ROUND THE DIARY User's Manual IP SERIAL SERVER Copyright © All rights reserved.
Failed to retrieve file