        Field Safety Notice
         Contents
1.      Field Safety Notice    Technical Bulletin  Q CORE MEDICAL   SAPPHIRE INFUSION PUMP  Sapphire Multi Therapy Pump     Hospira L N xxxxxx  Q Core P N xxxxxx  Sapphire Epidural Pump     Hospira L N xxxxxx  Q Core P N xxxxxx  Action Taken by the Company    To address this issue  Q Core Medical is providing an addendum to the User Manual  attached to this  notification   to assist customers in  i  identifying whether the catheter utilised may cause over delivery   ii  better  understanding the effect of back pressure and catheter size on the accuracy of the delivered volume   iil   translating the catheter   s back pressure to the degree of flow deviation that can be expected and  iv  contacting  Q Core  Support aqcore com  to determine if and to what extent specific catheter oump combinations are  susceptible to over delivery     Future versions of the User Manual will incorporate the information contained in the addendum     Actions to be Taken by the User    1  Inform  Inform the healthcare professionals in your organisation about this notification and provide  them with a copy of the notification     2  Add the Addendum to your User Manual  Photocopy a sufficient number of the attached  Addendum and add a copy to each Sapphire Infusion Pump User Manual in your healthcare facility     3  Adjust bolus and basal flow rates  If a catheter size smaller than 20 gauge is used  and over   delivery is experienced during utilisation of Epidural mode with patient bolus     A  Restric
2.  VTBI  lower infusion rate  and higher backpressure  In Epidural  intermittent mode  bolus rate is set to 125 mL h  In PCEA mode  bolus rate can be configured to 125  mL h or 200mL h before starting infusion  for details refer to Epidural Mode Options Menus on Chapter  7      For details about the impact of different back pressures on pump delivery caused by different  catheter sizes when used in Epidural mode  refer to pump accuracy in Chapter 12      Changes to Chapter 12     Technical Specifications     clause     Pump Accuracy       The following section replaces the entire section titled  Pump Accuracy     The updated text is marked in bold    The following graphs and curves were derived from the pump accuracy testing procedures described in  the IEC60601  2 24 standard  Testing was performed under normal conditions  specified in IEC60601  2   24 standard  at room temperature  25  C  72  F      Normal conditions to ensure optimal accuracy of    2 5      e Fluid level should be 50 cm above the pump   e No back pressure due to catheter size or difference in height of pump and infusion site  e Room temperature  25  C    e Barometric pressure of sea level altitude  101 kPa    e IV medication with water like fluid characteristics       D  QCORE 29 Yad Haruzim St     PO  Box 8639   Netanya 4250529     ISRAEL       medical Tel   972 73 2388888    Fax   972 73 2388800    info qcore com     www qcore com    ADDENDUM     USER MANUAL SAPPHIRE MULTI THERAPY  AND DEDICATED INFUSION P
3. Field Safety Notice    Technical Bulletin  Q CORE MEDICAL   SAPPHIRE INFUSION PUMP  Sapphire Multi Therapy Pump     Hospira L N xxxxxx  Q Core P N xxxxxx  Sapphire Epidural Pump     Hospira L N xxxxxx  Q Core P N xxxxxx    16 July 2015    Dear Healthcare Professional and Valuable Customer     Q Core Medical has received complaints from a few customers that when using Epidural mode with patient  bolus the amount of medication delivered may be greater than programmed  In the Sapphire pump  as in all  infusion systems  external factors may cause fluctuations in flow rate accuracy  Investigation of these  complaints determined that the catheters used in these cases were of a gauge smaller than that used for the  Sapphire pump flow accuracy calibration  The Sapphire Infusion System is distributed by Hospira or their  distributors in your country     Cause    The Sapphire pump is calibrated to operate within its accuracy specifications when used with a  Reference  Catheter  that provides a back pressure of 0 32 bar  at 125 mL h bolus delivery rate and 0 57 bar  at 200 mL h  bolus delivery rate  Using catheters that create a lower back pressure may result in over delivery during Epidural  treatments with patient bolus  The actual deviations that occurred during clinical use with either flow rate  as  reported in the complaints  were within the range of 0 5 8 8   over delivery     This deviation in accuracy may occur only when all of the following conditions are concurrently fulfille
4. UMPS    Epidural Mode  the operating conditions to ensure optimal accuracy of    2 5  while utilizing an  Epidural catheter are     e Use of catheter that creates a back pressure of 0 32 bar  4 35 PSI  at 125 mL h and 0 57  bar  8 26 PSI  at 200 mL h  This back pressure is typical for 20 gauge catheters  that  usually have an internal diameter of 0 45 or 0 46 mm    e Catheters with gauge size smaller than 20 may create over delivery  and catheters with  gauge size larger than 20 may create under delivery     In the Sapphire pump  as in all infusion systems  external factors may cause fluctuations in rate  accuracy  Conditions that can cause flow fluctuations include     e Position of the infusion container height  any deviation from 50 cm above the pump     e Fluid characteristics that deviate from water like characteristics  such as density  viscosity and  homogeneity    e Positive and negative pressure  including back pressure    e Environmental temperature above or below 25  C and barometric pressure above or below  101kPa    For example     e A change of     25 cm in the fluid level above the pump may result in deviation in accuracy of      1     e A change of    0 05 bar in back pressure may result in deviation in accuracy of    3     e Using sets with PAV may result in deviation in accuracy of  6      Note 1  In Epidural mode  the inner diameter of the catheter as well as the tubing length may  create back pressure that impacts the accuracy of the pump  and to a lesser 
5. ance is required in determining whether the catheter used may result in over delivery during  delivery with Epidural mode with a bolus  or in calculating the impact that the catheter may have in over delivery     please contact support qcore com  or tick this box O       DB  QCORE 29 Yad Haruzim St     RO  Box 8639    Netanya 4250529    ISRAEL    medical Tel   972 73 2388888   Fax   972 73 2388800   info qcore com     www qcore com    ADDENDUM     USER MANUAL SAPPHIRE MULTI THERAPY  AND DEDICATED INFUSION PUMPS    Changes to Chapter 5   Using the Infusion Modes     clause     Epidural Mode     The following replaces the section between the title  Epidural Mode    and the warning    Epidural Mode  Safety  Precautions     The updated text is marked in bold      This mode enables the pump to deliver epidural infusions  Epidural administration is limited to short  term infusions  up to 96 hours   using indwelling catheters specifically identified for epidural drug  delivery     In Epidural infusion mode  the pump can operate in either of the following sub modes     e PCEA  Patient Controlled Epidural Analgesia   Delivers epidural boluses  either alone or in  addition to a basal preset rate  Alternatively  only a basal infusion  without boluses  can be  programmed    e Intermittent Epidural  Delivers epidural infusions at intermittent pre set intervals     The features of the Epidural mode are designed to accommodate the special requirements of an  epidural infusion  such as lower
6. d      1  Epidural mode  Running an infusion in the Epidural delivery mode     2  Bolus  The patient is given a patient bolus     3  Catheter selection  Using a catheter with a back pressure that is lower  than the back pressure of the     Reference Catheter        The extent of deviation in volume delivered is highly dependent on the catheter selected  together with the  limitations imposed by the pump and the parameters set for the Epidural treatments     Currently  the Q Core Medical User Manual does not include the parameters of the Reference Catheter to allow  precise correlation between the catheter dimensions  the back pressure and the deviation in flow that the  catheter may cause     Risk to Health    No patient injuries or deaths have been reported as a result of this issue  Although Epidural drugs may lead to  adverse effects such as respiratory depression  Q Core s investigation determined that there are no increased  risks to health as a result of the over deliveries reported       Smiths Medical Portex catheter  20 G  length 915 mm  product code 100 382 1 18     24 35 PSI    3 8 26 PSI    4 Using a catheter with back pressure higher than the Q Core Medical Reference Catheter  while using Epidural mode with  patient bolus  leads to under delivery  There have been no complaints regarding under delivery        B  QCOI E 29 Yad Haruzim St     RO  Box 8639    Netanya 4250529    ISRAEL       medical 73 2388888   Fax   972 73 2388800    info qcore com   www qcore com  
7. extent so may the  fluid viscosity  Under delivery or over delivery that may occur due to a change in catheter size   may be estimated based on the following formulas       deviation at 125        back pressure of catheter  bar  at 125 aa 0 32  bar     0 05 Ibar   bar     deviation at 200        back pressure of catheter  bar  at 200 7     0 57  bar     0 05 Ibar   bar     For example  if the backpressure for the catheter in use is 0 2 bar  2 9 PSI  at 125mL h  the  resulting   deviation   7 2        BD  QCOI tE 29 Yad Haruzim St     P O  Box 8639    Netanya 4250529    ISRAEL    medical Tel   972 73 2388888    Fax   972 73 2388800    info qcore com     www qcore com       ADDENDUM     USER MANUAL SAPPHIRE MULTI THERAPY  AND DEDICATED INFUSION PUMPS    If the back pressure of the catheter used is unknown  please consult the set manufacturer  You    may also contact Q Core Medical Support at support qcore com  and Q Core will either  measure or calculate the back pressure and or deviation according to the provided parameters     Practical recommendations   If using a catheter size smaller than 20 gauge and over delivery is experienced during Epidural    mode with patient bolus   A  Restrict the programmed flow rate for bolus delivery to 125 mL h  not 200 mL h   If the treatment is programmed to bolus only infusion  the highest deviation that can  occur is 12  at 125 mL h   And    B  When clinically applicable  use a basal rate of at least 1 mL h  Any basal rate that is  larger 
8. t the programmed flow rate for bolus delivery to 125 mL h  not 200 mL h    lf the treatment is programmed to    bolus only  infusion  the highest deviation that can occur is  12   at 125 mL h     And    B  When clinically applicable  use a basal rate of at least 1 mL h  Any basal rate that is larger than  zero will reduce the amount of over delivery     Note  Over delivery can be further reduced or eliminated by    1  Restricting the number of boluses administered  and or   2  Restricting the volume infused during boluses  and or   3  Programming a basal flow rate over 1 mL h    4 For optimal flow accuracy use a 20 gauge Epidural catheter where suitable     4  Complete the Customer Response Form     A  Complete the attached response form     B  Return the form to the Q Core service department via the e mail address included on the form   Please do so even if you do not have an affected product        B  QCOI E 29 Yad Haruzim St     RO  Box 8639    Netanya 4250529    ISRAEL  9  LJ    medical Tel   972 73 2388888    Fax   972 73 2388800     infoG qcore com    www qcore com    Field Safety Notice    Technical Bulletin  Q CORE MEDICAL   SAPPHIRE INFUSION PUMP  Sapphire Multi Therapy Pump     Hospira L N xxxxxx  Q Core P N xxxxxx  Sapphire Epidural Pump     Hospira L N xxxxxx  Q Core P N xxxxxx    Enquiries and Support     Catheter Enquiries  If further assistance is required to determine  i  whether the catheter you are using may  result in over delivery during Epidural infusion 
9. than zero will reduce the over delivery     Note  Over delivery can be further reduced or eliminated by   1  Restricting the number of boluses administered  and or   2  Restricting the volume infused during boluses  and or   3  Programming a basal flow rate to more than 1 mL h    4  For optimal flow accuracy use a 20 gauge Epidural catheter where suitable    Note 2  When using accessories where the container height deviates from 50 cm above the pump  there  may be deviations in pump accuracy  e g   PCA Lockboxes and the Homecare Large Backpack          QCORE 29 Yad Haruzim St     PO  Box 8639   Netanya 4250529   ISRAEL    medical       Tel   972 73 2388888    Fax   972 73 2388800    info qcore com     www qcore com    
10. with a patient bolus  or  ii  the estimated effect that the catheter  may have in over delivery  please contact support qcore com  or mark the appropriate box in the attached  customer response form     For other enquiries  please contact Hospira using the information provided below     Hospira Contact Contact information    XXXXXX XXXXXX       Tally Eitan    President    Q Core Medical Ltd        BD  QCORE 29 Yad Haruzim St     P O  Box 8639    Netanya 4250529    ISRAEL    medical Tel   972 73 2388888    Fax   972 73 2388800    info qcore com     www qcore com       Field Safety Notice    Technical Bulletin  Q CORE MEDICAL   SAPPHIRE INFUSION PUMP  Sapphire Multi Therapy Pump     Hospira L N xxxxxx  Q Core P N xxxxxx  Sapphire Epidural Pump     Hospira L N xxxxxx  Q Core P N xxxxxx    RESPONSE IS REQUIRED    Please e mail the completed form to QCore6646 stericycle com     Customer Information    Business Name Hospira Customer Number  if applicable     Address Town City County Region Postcode    Contact Name Phone E mail Address    Completed by  Printed Name Signature Date    Actions to be taken   For Hospitals     gt    have received and read the letter dated 16 July 2015 and have provided a copy to each of the users in the  facility  YESO NOO      If NO  state the reason      gt   have copied and attached the Addendum to each Sapphire Infusion Pump User Manual in my healthcare  facility  YESO NOO      If NO  provide the date when this will be completed     If further assist
    