Contents

1. Increase the separation between the equipment Connect the equipment into an outlet on a circuit different from that to which the other device s are connected Consult the manufacturer or field service technician for help Place the Weighing Platform on a level and stable surface If the equipment is used when the Weighing Platform is unstable because not all feet are on the surface there may be a risk of stumbling or inaccurate measurement Note that portable and mobile RF communications equipment can affect MEDICAL ELECTRICAL EQUIPMENT Consult a physician or a trained health professional for interpretation of measurement results In case of patients who have certain diseases the estimates might be different Incorrect operation or failure of user to maintain the unit spares the manufacturer or his agent of the responsibility for system s non compliance with specifications or responsibility for any damage or injury This manual is made for informational purposes and this manual and product are not meant to be a substitute for the advice provided by your own physician or other medical expert You should not use the information contained in the product for diagnosis or treatment of health problems or prescription of medication by yourself If you have or suspect that you have a medical problem consult with your physician promptly 11 Defective units or accessories must be packed in the replacement cartons to be sh
2. Input weight Input weight by using LCD touch pad The possible input range 10 270 kg Press HEIGHT button Input height 44 Input height by using LCD touch pad The possible input range 50 220 cm Press AGE button Input age Input age by using LCD touch pad The possible input range 1 99yrs Press GENDER button Select gender Select gender by using LCD touch pad Select state of dialysis Select state of dialysis by using LCD touch pad before dialysis during dialysis after dialysis BCA Body composition analysis could be used as a complementary method for assessing nutritional status of HD hemodialysis and CAPD continuous ambulatory peritoneal dialysis patients as body weight or BMI measurements do not reflect fat muscle masses and visceral fat ratios in these patients Use this equipment as directed by an attending physician 4 Measurement Measurement starts after input Check whether clip electrodes are accurately equipped Do not move or talk during measurement 1 When the measurement is incorrect The message Impedance is out of the range appears on the screen If the measurement fails three times in a row the message appears as below The measurement can not complete due to the continuous errors Refer to Solution Result screen After measurement is completed the result for the selected items is displayed on the screen
3. 45 6 Move to the result screen Press Control Dial in the middle of the device or MENU Folder button in the upper right corner of the touch screen to move to another screen Articles are displayed on the right side of the screen as shown in the picture View Graph body water body composition View All and View Impedance are available Select an article and move to the corresponding screen Scanning QR code Press the QR code on the device screen to enlarge the icon and scan the QR code with a mobile device to save the data in the server After scanning the QR code you can check the measurement data with your mobile device and manage it at will from the server in which It is saved 1 Scan the OR code with a OR code reader in a smart mobile device 2 The data saved in the server could be damaged or lost due to a problem with the server Restarting Once the result is displayed on the display it can be printed out in pre printed result sheet After confirming the analyzed result press Initial view to restart The device returns to the initial screen after one minute 46 1 Storage of data 1 Insert USB memory stick into USB A jack placed on the back side of the device 2 The following message appears Do you want to write data to USB memory stick on initial screen Press YES to send data to USB memory 3 Transmitting message will be displayed on the screen When
4. Measuring method Tetra polar electrode method using 8 touch electrodes Frequency Range Measuring site Main items Current Power supply Display Input device Transmitting device Printing device Dimension Weight Measuring range Measuring time Input height Input weight Applicable age 1 5 50 250 550 1000 kHz Whole body and Segmental measurement arms legs and trunk Body Composition Result Weight Standard weight Lean body mass Mass of body fat Skeletal muscle mass Soft lean mass Protein mass Mineral mass Total body water Intra cellular water Extra cellular water Body mass index Percent of body fat Waist to hip ratio Visceral fat level Visceral fat mass Subcutaneous fat mass Visceral fat area Target to Control Control of Body fat Control of Soft lean mass Control of Weight Body composition change 8 times of Percent of body fat Soft lean mass Weight Segmental soft lean mass and percent Segmental mass of body fat Body type Body cell mass Basal metabolic rate Total energy expenditure Age matched of body Study item Segmental impedance and reactance classified by frequency Segmental Phase angle QR code Body water Result Intra cellular water Extra cellular water Total body water E C W T B W T B W L B M Body water change 8 times of Ratio of E C W T B W T B W I C W E C W Segmental total body water Segmental intra cellular water Segmental extra cellular water Segm
5. Select a to enter the menu Enter a domain address or IP you want on the touch pad A Wzalmals lz 15117 maximum 20 digit number can be entered for the domain address T Press DONE button to save the data and return to the ME ma mr initial screen EA EA Ee 01 tole Once you scan QR code after a measurement the data will be stored in the server of the saved domain address If the domain address is not set the data will be stored in the server of Jawon Medical m jawon com You can search the measurement data by creating an ID and PW lt Communication gt Select the connecting method CABLE is for USB cable and WIRELESS is for wireless communication Select MM COMMUNICATION on SYSTEM SETUP screen and press SET button on touch pad Preset CABLE WIRELESS Choose CABLE or WIRELESS on touch pad Press SET button on touch pad to save it Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad CABLE WIRELESS SELECT lt Heightmeter gt Select the connecting heightmeter Select fl HEIGHTMETER on SYSTEM SETUP screen and press SET button on touch pad Preset OFF Choose ON or OFF on touch pad Press SET button on touch pad to save it Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad Heightmeter 39 Once turning on the device you can see the screen same as right picture Body Compositio
6. 930 should assure that it is used in such an environment Electromagnetic environment Immunity test IEC 60601 test level Compliance level guidance Electrostatic Floors should be wood concrete or discharge ES 6kV Contact 6kV Contact ceramic tile If floors are covered with D IEC 8kV Air 8kV Air synthetic material the relative 61000 4 2 humidity should be at least 30 Electrical fast 2kV Power supply 2kV Power W Mains power quality should be that of transition bur lines supply lines st 1kV 1kV environment IEC 61000 4 4 Input output lines Input output lines 1 kV differential 1 kV differential Surge Mains power quality should be that of mode mode IEC 61000 4 a typical commercial or hospital 2 kV common 2 kV common 9 environment mode mode a typical commercial or hospital lt 5 UT lt 5 UT Voltage gt 95 dip in UT gt 95 dip in UT Mains power quality should be that of drops dips for 0 5 cycle for 0 5 cycle a typical commercial or hospital 40 UT 40 UT 60 dip in UT 60 dip in UT for 5 cycles for 5 cycles of input 70 UT 70 UI during power mains interruptions it is power supply 30 dip in UT 30 dip in UT recommended that the COZY 930 be for 25 cycles for 25 cycles l line IEC lt 5 UT lt 5 UT powered from an uninterruptible 61000 4 11 gt 95 dip in UT gt 95 dip in UT power supply or a battery for 5
7. A Port Terminal for connecting the printer option to the device RS 232C Port Terminal for connecting to the external device 19 4 Cart Clip Electrode hanger N ETT N j I Clip Electrode A N BIN A fis iif JAY Handle Electrode Cable Cabinet Cart Wheel Clip Electrode hanger Clip Electrodes are hanging on the hanger Clip Electrode It measures Impedance by being fixed to wrists and ankles as a clip form Electrode Cable It connects clip electrodes to main body Cabinet It keeps charts cables and etc Handle It makes the movement easy Cart Wheel It makes the movement easy by turning 360 20 1 Basic Installation 1 Cart assembly and connection of main body 1 Cart assembly package Driver is not included in the package so prepare it personally 1 U D Cart main body 1ea M4 bolt m size 4ea M5 bolt 10ea 4 Cart wheel 5ea 2l 2 Cart wheel assembly Connect the 5 cart wheels to the cart main body as shown in the picture ni m in N g 1f IS SA VA Tags pou 5 lt 4 es Se AN fa N VON YS N v Fix the connected cart main body and wheels each other with 10 M5 bolts by using driver like the picture above 2 3 Checking the horizontality of clip electrode hanger The clip electrode hanger should be at a level like the picture above If not adjust the angle of the clip electrode
8. caffeine or beverages functioning as diuretics 4 hours before measurement Drink 2 cups of water 2 hours before the measurement 2 Before measurement the subject should be in a stable condition Measure 3 4 hours after a bath a sauna exercise or activity that sweats a lot Or measure before these actions 3 Avoid drinking alcohol 24 hours before the measurement 4 Wear clothes as light as possible 5 Once the subject is on the scale avoid sudden movement from sitting to standing position Body fluid goes down to the lower body and affects the results Thus subjects should be measured after maintaining standing position for 5 minutes 6 Clean both the electrodes and measuring body parts 7 Changes in room temperature may affect the results Measurement should be done in a temperature around 20 C Body composition and weight varies even during a day Therefore the measurement should be performed at the same time every day For a person who stands for a long period of time during the day it is advised to measure in the morning 9 Go to bathroom before measurement Maintain correct position and posture during the measurement In order to keep one s health and the balance of body composition check the changes of body composition through continuous analysis and compare the results Make sure that the body composition should are measured under the same physical and environmental conditions If the condition before the me
9. ee eee eee eee ee ee eee eee ee eee ee eee ee eee 6 5 SAFETY SYMBOLS AND INFORMATION a 8 6 Guidance for Electromagnetic compatibility EMC OCC P PCCP eee ee eee ee eee eee eee ee eee ee eee ee eee ee eee 11 TERMS AND FUNCTION OF EACH PART ss 15 1 Basic Package NRK rer rrr rere reer rr reer er eee ee eee ee eee eee ee 15 2 Options TOCT Cece errr re rerrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrrr rrr rr rrrrr rrr rr rrr rrr reer eee ee eee ee eee ee 16 3 Appearance of the device 17 INSTALLATION EKKE Pec r reece reer rrrerr reece errr r rr rer rr rrr rrr errr reer rr reer rere re reer eee ee ee eee ee eee eee 20 1 Basic Installation tis 20 1 Cart assembly and connection of main body EEK eee eee eee eee eee eee eee eee eee ee ee DD 2 Connecting clip electrode eeessssasaaae eee e eee aan aan naa a nanna aan ennen 24 2 Power Supply KKK KRR eee eee rere eee eee eee ee ee reer ree eee eee eee eee ee ee eee eee ee eee ee ee 25 3 Adjusting the height of the Cart ee 26 4 Peripheral Device Installation tis 27 1 Connecting Bluetooth Option POCCPPCCCOPC CCCP SCC Cee e eee eee ee eee ee eee eee rece ee eee ee eee ee eee ee eee ee ee 27 2 Connecting Computer unnnnnununnnNNENNNNNNNNNNENNNNNMNENNNNNEESNNNNHHESNNENHHENNHNEMENNNNNNHEENENENE 30 3 Connecting Printer Option SS 31 4 Replacing of thermal paper KTKT rere ere eee eee eee ee eee ee ee ee 32 SYSTEM SETUP NRK NRK NRK 33 1 Entering SYSTEM SETUP u nunununuuunuNN
10. image appears press 1 2 3 4 in order Then it will be changed to SYSTEM SETUP mode 2 Menu in SYSTEM SETUP Menu items are displayed The function of each icon is as follows 1 DATE TIME 2 VOLUME 3 PRINT 4 PRINT POSITION 9 DATE TYPE 6 THERMAL PRINT 7 LANGUAGE 8 QR CODE DOMAIN 9 COMMUNICATION 10 HEIGHTMETER 34 3 Selecting a Menu in SYSTEM SETUP SYSTEM SETUP Select menu according to one s desire and confirm changes using SET button 4 Exiting SYSTEM SETUP Press CLOSE button on SYSTEM SETUP phase then back to initial view of the analyzer DATE TIME 5 Moving to SYSTEM SETUP e 2008 Press CLOSE button on the menu view then program EN 03 is returned to SYSTEM SETUP phase __ 20 16 19 oo 6 Setup lt DATE TIME gt It sets date and time year month day hour and minute DATE TIME Select E DATE TIME on SYSTEM SETUP screen and press SET button on touch pad Pre set The date and time of the device released from the manufacturer s factory Set YEAR by pressing V A on touch pad Press MONTH and set Month in the same way to set YEAR Choose DAY HOUR MIN and Set correct date and time in the same way Press SET on touch pad to save the change Return to SYSTEM SETUP screen by pressing CLOSE on touch pad 1 If SET button is pressed before finishing setup of date and time the date and time inputted at
11. survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol 0 14 1 At 80 MHz and 900 MHz the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the COZY 930 is used exceeds the applicable RF compliance level above the COZY 930 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the COZY 930 Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m 4 Recommended separation distances between portable and mobile RF communications eguipment and the COZY 930 The COZY 930 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The customer or the user of the COZY 930 can help prevent electromagnetic in
12. 6 Basic safety and essential performance of Medical Electrical Eguipment Therefore patients must not touch or handle inner side of the system at any time Do not to touch signal input signal output or other connectors and the patient simultaneously The unit has previously been adjusted in the factory for optimum performance Do not attempt to adjust switches or any other things except those specified in this manual for operation Never pour any liquid directly on the scale platform as it may leak and cause internal damage Never jump on the Weighing Platform there may be a risk of stumbling and malfunction of the equipment This equipment has been tested and found to comply with the limits for medical devices according to IEC 60601 1 2 2007 These limits are designed to provide reasonable protection against harmful interference in a typical medical installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be 10 determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures Reorient or relocate the receiving device
13. BODY COMPOSITION ANAYZER USER MANUAL COZY 930 Leading Health Trend BODY WATER ANALYSIS Body Water Volume Intra Cellular Water Extra Cellular Water Extra Cellular Water Total Body Water Total Body Water Lean Body Mass Body Composition Analysis Body Water Protein Mineral Mass of body fat Soft Lean Mass Percent of Body Fat Body Mass Index Abdominal Analysis Waist to Hip Ratio Metabolic Basal Metabolic Rate Total Energy Expenditure Nutrition Protein Mineral Body cell Mass N Segmental Analysis Body Fat Soft Lean Mass Extra Cellular Water Total Body Water Impedance Reactance Phase Angle CE The device bears the CE label in accordance with the provisions of Medical Device Directive 93 42 EEC THE PERSONS RESPONSIBLE FOR PLACING DEVICES ON THE EC MARKET UNDER MDD 93 42 EEC Jawon Medical Co Ltd 29 Gongdan 4 ro Jillyang eup Gyeongsan si Gyeongsangbuk do Korea TEL 82 53 856 0993 FAX 82 53 856 0995 VITAKO Sp z 0 0 REP me UL MALEJ SYRENKI 2 71 790 SZCZECIN POLAND TEL 91 8522900 901 FAX 91 8522910 INTRODUCTION NRK RN RN KRR eee eee ee eee eee eee ee eee ee eee eee 5 1 INTENDED USE NRK errr rrr rere reer rrr errr reer rere ere ree eee eee ee eee ee eee eee 5 2 WORD DEFINITIONS NENNIR 5 3 CLASSIFICATION AND COMPLIANCE OCP PP CCCP CCCP PoC eee eee eee eee ee eee eee eee eee eee eee ee eee eee eee 6 4 SAFETY PRECAUTIONS EEK POSTS PCC CC SCC e Tee ee eee eee eee eee
14. NENNNENNNENNNENNNENNNENNNEMNNEHNNEHHNEHENEHHNEHNHNEHENEHNHNEHENENNN 33 2 Menu in SYSTEM SETUP SS 33 3 Selecting a Menu in SYSTEM SETUP a 34 4 Exiting SYSTEM SETUP EKKS RN 34 5 Moving to SYSTEM SETUP NENNIR 34 6 Setup KKK KRR RN NN KK KIKSI NR N KKK KRR rrrrrrrr rrr rr rrr re terre eee RN NN KRR 35 INITIAL SCREEN SETTING A COTTE ETE 39 MEASUREMENT AND ANALYSIS pununnanonnunnanonnunnupununnanpununnansunununsununnunnununnunuunununpunnununD 40 1 Precautions for Measurement EEE KLIKKAA OSE Cee eee re ee 40 D Correct Posture sunnnnasnunununouunnnanonnunnununuunnunpununnannunununsununnapuunununpununnannunnununpuununanpununnN 41 1 Posture pusunnansunununounnununsununnapuununnupunnununpununnansunununpununnannununnanunuununpununuunsununnanunn 41 2 How to Touch Electrodes TECUUCCCOOCOCCCCCCOOTOCCCCCCOOTOCCCOCCOOTOOTCOCEOETOOOCOCEOETOOOCOCEEE TES eee ee 42 3 Measuring Procedure nenunnanunnunnanunnunannununnanunnunnanuunununnunnnnanpnnannanunnanunpununnunsununnannnn 43 1 Basic Analysis A KKT TKKK KKT TKKTKKK TTKK TT 43 STORAGE OF DATA USING USB MEMORY susunnansunununpununnansununuapununnansnuunnupnnuununpununnanunn 46 RESULT INTERPRETATION nn 47 STORAGE AND MAINTENANCE A OT eee ee ee ee 52 ERROR amp REPAIR TECUUOCCCOOCOCCCOCCOOLOCCOCCCCOCCOOCCCCTOCCOCCCCTCOTOCCCCCCTOPCCC COTTE OTOO TC CCo CET E Te ee ee re 53 1 Kinds of Error amp Repair A TOE EEE 53 2 Error amp Repair A CEE 54 AFTER SERVICE pununnunounununnununnannununuupununnannnnunuspunuun
15. Overweight because of over fat Type Overweight because of much muscle mass Type Sarcopenic obesity Target Decreasing body fat Target Keeping current status Target Increasing muscle mass 1 Percent of Body Fat for Adults Male lessthan15 15 lt 20 20 lt 25 25 lt 30 0 5 2 Body Mass Index kg m for Adults EAST ASIA EU and etc Abdominal Analysis Abdominal fatness is divided into subcutaneous type and visceral type When it comes to body fat experts say that not only the amount of fat is important but also its distribution If visceral fat area is over 100 em it is classified as visceral obesity regardless of P B F W H R or Body weight Waist to hip ratio W H R shows the distribution of fat stored in one s abdomen and hip It is simple but useful in assessing fat distribution Body fat is stored in two distinct ways They are often categorized into and called apple and pear type Apple type shows bigger girth of waist than hip and pear type has bigger girth of hip than waist If body fat in abdomen increases the risk to cardiovascular diseases diabetes etc becomes higher 1 W H R Waist to Hip Ratio W H R is calculated by dividing waist girth by hip girth When W H R is below 0 9 male 0 85 female the risk of visceral obesity is low 2 Visceral Fat Level The degree of visceral obesity is displayed in a level Level 1 4 corresponds to subcutaneous fat type Level 5 8 corresponds
16. ack side of the cart so that the electrode cable cannot touch the floor Turn the adjusting screw to the left to make the screw loose and adjust the height of the cart and turn the screw to the right to make it tight En N e el i a E U fe Lf 4 Peripheral Device Installation 1 Connecting Bluetooth Option This device can be connected with peripheral devices of Jawon Medical Connect optional devices that Jawon provides Bluetooth and etc to the RS 232C port on the back side of the device by using the RS 232C cable that Jawon also provides When choosing a Bluetooth connect the Bluetooth that Jawon provides to the Bluetooth port on the backside of the device In this case users can check and manage the result data on the exclusive mobile device app via Bluetooth connection D Components for Bluetooth Bracket Bluetooth i l Antenna 2 Unscrew the bolts 2EA on the back side of the body with a driver Keep the separate bolts 2EA to assemble a bracket and cable later as show in the picture below 20 3 Connect the antenna to the Bluetooth 1 i Connect the antenna to the Bluetooth Turn the antenna at a right angle to the Bluetooth as shown in the picture Connect the cable for the connection with the main body to the opposite side of the antenna Screw the fixing bolt of the Bluetooth with a driver t
17. al Check Smooth operation Pass Fail a Mainframe Smooth operation with no 2 Recorder Pass Fail abnormal sound Smooth operation and 1 Power cord Pass Fail removal Electrical Check Screen display upon Power supply Pass Fail power on No abnormality and 2 Display Pass Fall flickering Model COZY 930 Copy this sheet for use Accessories Performance If repair is reguired write down so in the Remarks column 61 Service center TEL 82 2 587 4056 82 53 856 0993 lf the problems continue call the service center When you ask for service the manufacturer s label serial number date of Original purchase and explanation of malfunction will be required x For purpose of improvement specifications and subject to change without notice FACTORY 29 Gongdan 4 ro Jillyang eup Gyeongsan si Gyeongsangbuk do Korea TEL 82 53 856 0993 FAX 82 53 856 0995 EUROPEAN REPRESENTATIVE VITAKO Sp z o o UL MALEJ SYRENKI 2 71 790 SZCZECIN POLAND TEL 91 8522900 901 FAX 91 8522910 6Z JW CZE 20151104 Rev 4 0
18. asurement such as volume of a meal meal time and activities exercise sauna drinking lots of beverage urination etc are kept same the reproducibility of a device is obtained Therefore the data can be used to evaluate the change of body composition 41 2 Correct Posture 1 Posture Keep correct posture for accurate results Keep the stable posture for 10 15 minutes before measurement to measure exactly D A subject should be measured after maintaining supine position for 5 minutes 2 Spread both arms 30 aside from the trunk 3 Set both legs apart as much as shoulders and make sure that they are not touched 2 How to Touch Electrodes Maintaining the right posture during the measurement is essential to achieve reliable results It s important to 42 maintain examinee s posture for 10 15 minutes so that impedance can be measured accurately e Clean wrists ankles and electrodes with wet tissue e Wear the clamp electrodes on wrists and ankles Spread arms and legs naturally as shown in the picture below 1 Make sure that the subject s body is not touching other people or conductible materials 2 Make sure arms do not touch the trunk of the body Spread arms to around 15 degree angle away from the trunk and make sure legs do not touch each other 2 Keep stable contact with every electrode If electrode is not properly contacted it may cause the cease of the measurement or decrease reliability of data dur
19. e is an indicator of cellular health and integrity A low phase angle is consistent with an inability of cells to store energy and an indication of breakdown in the selective permeability of cellular membranes A high phase angle is consistent with large quantities of intact healthy cell membranes and body cell mass Phase angle for adults range from 2 20 degrees with normal values of 6 to 8 degrees Energy Expenditure 1 B M R Basal Metabolic Rate B M R refers to the calories required to maintain human body s basic function such as movement of heart brain neural transmission regulating body temperature and so on B M R is in proportion to S L M because body fat stores energy while muscle consumes energy Therefore even if the weight is same between persons the person with more muscle has greater B M R 2 T E E Total Energy Expenditure lt is the sum of basal metabolic rate and calories needed for daily activity Generally it is calculated by multiplying B M R by PAL Physical Activity Level A M B Age Matched of Body It is the estimated physical age of the subject considering body composition analysis result gender and biological age This is calculated by comparing the optimal body composition based on the gender and biological age of the subject with the actual analyzed body composition It can be used to evaluate the subject s health and body development Study 1 Impedance amp Reactance It indicates the segmenta
20. ent or a potential difference 6 Do not put or drop anything on the device and avoid strong impact 7 Do not disassemble or remodel the device 8 If this unit has not been used for a long time use this after confirming with an expert if all function and appearance are in good condition 9 Do not splash any fluid on this device or insert any foreign substances 10 In case of inserting foreign substances or exposing to particular environment this device must be examined by an expert before use 11 Use the power cable plug and fuse that are offered by our company At this time confirm the covering of cable the state of plug connection and other check points e RS 232C cable USB port Adapter 12 When pulling out the power cable turn off the power switch first and then pull the plug out 13 Storage ambient Temperature 25 70 C Humidity lower than 93 non condensing 14 Operation ambient Temperature 5 40 C Humidity 15 93 non condensing 15 Do not store or use this device under 70 kPa 700 mbar or over 106 kPa 1060 mbar of atmospheric pressure 16 Cleaning amp Disinfection 1 Cleaning Use a soft gauze cloth with volatile liquid like alcohol Ethyl or Isopropyl alcohol 70 90 to clean it Clean it every 2 3days Do not use a wet cloth 2 Disinfection After measurement use a soft gauze cloth with volatile liquid like benzene and alcohol Then wipe the enclosure with soft lint Please wipe after every
21. ental ratio of E C W T B W Segmental E C F T B F Nutrition assessment Water balance Phase Angle Study item Segmental impedance and reactance classified by frequency Segmental Phase angle QR code Within 180uA 15 Input AC 100 240 V 50 60 Hz Output DC 12 V 5A adapter 7 Inch wide color LCD Touch pad PC remote control USB port RS 232 cable Thermal printer standard USB port OPTION Assigned by manufacturer 200 x 280 x 72 mm W x D x H 10 mm About 1 8kg 100 950 O Less than 1 minute 40seconds 50 220cm 10 270kg 1 99 years old Ol I Ambient temperature range 5 to 40 C Relative humidity range 15 to Operation ambient 93 non condensing Ambient temperature range 25 to 70 C Relative humidity range lower Storage ambient than 93 non condensing x For purpose of improvement specifications and design are subject to change without notice 08 Warranty osaan SSS Date of purchase Add Name Customer Tel Add Name Dealer market Tel 1 The device passed through the strict quality control and test 2 If the problem occurs within the warranty period contact the service center It will be repaired at a free of charge 3 When the problem falls into Service with Cost case it will be repaired at a cost even under the warranty period 4 Spare parts for repair are held for 5 years after discontinuance of the device Service with Cost A
22. fter the warranty period Malfunction or damage caused by user s negligence Malfunction due to careless handling Any modification or damage caused by unauthorized staff 1 2 3 4 Malfunction due to a natural disaster 5 6 In case entries in this warranty are not stated 7 Consumable parts are related to the frequency of use Therefore the consumable parts are charged at a cost JI Periodic Check List Management No item Inspection Subject Requirements Judgment Remarks Visual Cheek No scratch crack deformation and rust fioo Noama Passa 4 Elecrode No scratch and damage Passa 2 User manual Keptin properpiace Passa K Mechanical Check boe l ol Mainframe 1 Keys Smooth operation Pass Fail 2 Recorder Smooth operation with no Pass Fail i nomena 3 Touch Screen Smooth operation Passa Accessories 1 Power cord Smooth operation and Pass Fail removal Electrical Check 1 Performance 1 Power supply Screen display upon Pass Fail La nn 2 Display No abnormality and Pass Fail a n Model COZY 930 Copy this sheet for use If repair is reguired write down so in the Remarks column Daily Check List Management No No scratch crack 1 Enclosure Pass Fall deformation and rust alico Nodamage Passa 4 Elecode No scratch and damage Passa 2 User manual Keptinpioperpiaos Passa knn Mechanic
23. g V A on touch pad Choose L R by pressing gt lt on touch pad Press SET button on touch pad to save it Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad L R left right is moving to the left and is moving to the right U D up down is moving up and is moving down DT lt DATE TYPE gt This is to set the format of the date Pre set YY MM DD DATE TYPE Select So DATE TYPE to enter the menu Choose the format on touch pad Press SET button on touch pad to save it Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad lt THERMAL PRINT gt It selects the printing mode of thermal printer Pre set ON OFF ON THERMAL PRINT Select E THERMAL PRINT to enter the menu ON OFF Choose ON OFF on touch pad Choose AUTO or MANUAL or OFF on touch pad Press SET button on touch pad to save it Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad lt LANGUAGE gt You can select one of the languages of Korean English IANGUAGE and Japanese Pre set ENGLISH ENGLISH Choose A LANGUAGE to enter the menu JAPAN Select one of the languages you want to set Press SET button on touch pad to save it Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad 38 lt QR CODE DOMAIN gt This is for setting QR code domain
24. hanger by turning the control knob 5 4 Fixing the main head to the cart Ox A JN ARA O CC A NN S W W Fix the main head with M4 bolts by using Si driver like the picture 10 j W b ET A K J 24 2 Connecting clip electrode 1 Put the plug of the electrode cable to the socket of clip like below tt a A i Lo 1 The cables are separated into 4 parts Left Arm Right Arm Left Leg Right Leg so connect them to the correct ports 2 Hand and foot clip electrodes are not necessary to be used separately 25 N Power Supply Connect the power cable to the ADAPTER INPUT placed on the lower back panel of the device Connect the cable Turn on the power switch placed next to Power input then after moment initial screen animation is displayed automatically USB B USB A RS 232C J poy f E Sy TT Tf i l A iS 1 Before connecting a peripheral device to this device the power should be turned off Otherwise this device can malfunction or be damaged due to electric shock 2 Power shouldn t be supplied to the peripheral devices to ensure the safe operation and durable performance 3 This device should be powered with the adapter and cable supplied only from our company 26 3 Adjusting the height of the cart Adjust the height of the cart by using the adjusting screw on the b
25. here the ambient temperature falls below 25 C or exceeds 70 C Where the atmospheric pressure falls below 70kPa 700mbar or exceed 106kPa 1060mbar Where the humidity is over 93 non condensing Where the unit is exposed to spray or splashing water Where the unit is exposed to dust Where the unit is exposed to water vapor Where the unit is exposed to salty atmosphere Where the unit is exposed to explosive gas Where the unit is exposed to excessive shocks or vibrations Where the angle of inclination of mounting surface exceeds 10 degrees Where the unit is exposed to direct sunlight This device needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the ACCOMPANYING DOCUMENTS Cross contamination is possible because this equipment is used with bare hands and feet Refer to the cleaning and disinfecting methods in this manual Measurements may be impaired if this device is used near televisions microwave ovens X ray equipment or other devices with strong electrical fields To prevent such interference use the meter at a sufficient distance from such devices or turn them off Do not disassemble or alter the device under any circumstances as this could result in electric shock or injury as well as adversely affect the precision of measurements This device is specified as Class 1 type BF unit under the standard EN 60601 1 200
26. ing measurement 45 3 Measuring Procedure 1 Basic Analysis 1 Choose a mode by using jog shuttle or touch screen User can measure one among body water body composition impedance and ALL 2 Press START button and the personal information window pops up 3 Input personal information Input personal information in order Input ID weight height age and gender and choose the state of dialysis Confirm the inputted date then press Next button If the inputted information is wrong correct it or press Reset button to revise all Choosing a mode Choose the mode and press OK Press START button and the personal information ad a A window pops up Seok idea ho be tobe nan Input ID Input ID ID can be made up to 20 characters including English number and a space Once you complete the input press WEIGHT button 1 Input ID and press ID SEARCH to call or search the ID saved in database Search Data Display the result with the searched ID Search Resolving The previous measured data of weight height age and gender is automatically inputted with the searched ID The saved data can be changed 2 When transmit the member information from Body Pass Plus to the device ID USAGES screen will appear In this case you can not modify personal information such as height age and gender at the device Modify the personal Information on the program and transmit the member information to the device again
27. ing of visceral muscle such as skeletal muscle heart and stomach 5 Mineral It composes of bone and electrolyte 6 Protein this is a major element that composes soft lean mass together with body water Assessment of Under Optimal and Over in the table is assessed by the optimal range based on standard weight of the subject Body Water Analysis 7 T B W Total Body Water It consists of intra cellular and extra cellular water For healthy adults body water is 45 65 of body weight even though it varies between persons Intra cellular water Quantity of water existing inside cells 9 Extra cellular water Quantity of water existing outside cells Edema Balance of body water is assessed with ratio between extra cellular water and total body water a1 T B W L B M Ratio of total body water to L B M Assessment of E C W T B W Edema is unbalance state of intra and extra cellular water Edema can be increased by salty food malnutrition postpartum exercise temporary fatigue etc It is divided into 3 steps Optimal Borderline Over 48 Body Status Body Status is displayed in a bar graph as shown in the picture It is very convenient to recognize what to do Increase muscle Decrease fat Keep current status etc Under Optimal Over Under Optimal Over Under Optimal Over Weight Weight Weight B M I B M I B M I P B F P B F P B F M B F M B F M B F S L M S L M S L M Type
28. ipped off from you to our company Shipping and insurance costs for return of defective unit must be prepaid by the users 6 Guidance for Electromagnetic compatibility EMC Details about the electromagnetic compatibility EMC of the COZY 930 are given below Before using the COZY 930 be sure to read and understand the following information 1 Guidance and manufacturer s declaration electromagnetic emissions The COZY 930 is intended for use in the electromagnetic environment specified below The customer or the user of the COZY 930 should assure that it is used in such an environment Emissions test _ Compliance Electromagnetic environment guidance The COZY 930 uses RF energy only for its RF emissions G Internal function Therefore its RF emissions are rou CISPR 11 j very low and are not likely to cause any interference in nearby electronic equipment RF emissions Class B CISPR 11 Harmonic The COZY 930 is suitable for use in all emissions Class A establishments including domestic IEC 61000 3 2 establishments and those directly connected to Voltage the public low voltage power supply network that fluctuations supplies buildings used for domestic purposes NU Compliance flicker emissions IEC 61000 3 3 12 2 Guidance and manufacturer s declaration electromagnetic immunity The COZY 930 is intended for use in the electromagnetic environment specified below The customer or the user of the COZY
29. ith information shown below Try to solve the problem over the phone with the personnel of local service department If the problem cannot be solved over the phone return the unit directly to service department x Our company or local distributor will make available on request circuit diagrams component part list descriptions calibration or other information which will assist your appropriately qualified technical personnel to repair those parts of the unit which are designated by our company as repairable How to contact our company Write us at Jawon Medical Co Ltd 29 Gongdan 4 ro Jillyang eup Gyeongsan si Gyeongsangbuk do Korea TEL 82 53 856 0993 FAX 82 53 856 0995 You can also contact the following representative or your local distributor 2 PACKING AND TRANSPORT Our company wraps this device up with the most suitable method to protect it from any impact or damage during shipping and transporting This device can be damaged during delivery if it is packed in other ways except the one our company uses Please handle this device carefully without any impact in packing and delivery If this device needs to be transported wrap this device up again and transport it as follows D Turn off the power 2 Turn off the power of the peripheral devices and disconnect all cables 3 Disassemble the device in reverse order of assembly 4 Pack the device with the original packing materials 5 Transport it carefully 56
30. l impedance and reactance of five body parts left and right arms left and right legs and trunk corresponding to each frequency 1 1000 kHz 2 Phase angle Phase angle is an indicator of cellular health and integrity A low phase angle is consistent with an inability of cells to store energy and an indication of breakdown in the selective permeability of cellular membranes A high phase angle is consistent with large quantities of intact healthy cell membranes and body cell mass Phase angle for adults range from 2 20 degrees with normal values of 6 to 8 degrees ol Blood Pressure When the blood pressure monitor supplied from Jawon Medical Co LTD is connected to the device blood pressure can be measured and the result can be printed out Systolic blood pressure diastolic blood pressure and pulse are printed on result sheet It helps to recognize hypertension assessment related to obesity QR code Scan the QR code with a smart phone to see the result on the website OZ 1 Pay attention to the allowable value of electric current 2 Avoid direct sunlight humidity dust thick oil and salty environment or extreme changes in temperature 3 Do not install or store the device in a place where any chemicals or gas are stored 4 Do not use the device in any unstable vibrating or impact giving area 5 Connect the earth placed on the backside of this device to terminal plate to prevent any electric shock from leaking curr
31. measurement for electrode disinfection 17 Refer to SAFETY PRECAUTIONS Users must wipe while wearing safety equipment such as gloves when disinfecting electrodes Our company does not take responsibility for safety accidents caused by user s carelessness Oo 1 Kinds of Error amp Repair Out of range of impedance Out of range of body fat Out of range of measurement Cant input the height e When the subject s body impedance is deviated from the limit Insufficient contact to electrodes Impedance is out of range Range 100 950 Q e When the subject s P B F is deviated from the limit Incorrect input of personal data P B F is out of range e When the subject s fatness is deviated from the limit Range 100 100 e When the subjects height is deviated from the limit Incorrect input of height 54 e Clean the measuring parts the electrodes palms and soles and try again e Measure again with correct posture e Do not move during measurement e If the same error is repeated please contact our company or its local distributor where this device is purchased e Clean the clip electrodes and try again e If the same error is repeated please contact our company or its local distributor where this device is purchased e Input height correctly e Check the weight is correct e If subject s body fat is out of the range measurement is impossible e If the same erro
32. must be paid to all warnings cautions and notes incorporated herein The following conventions are used throughout the manual to denote information of special emphasis Warning indicates important information about the presence of a hazard which may cause severe personal injury loss of substantial property damage if the warning is ignored Caution indicates important information about the presence of a hazard which may Cause minor personal injury or property damage if the caution is ignored Prohibition indicates import information to signify the prohibited action Notice indicates important information in order to notify installation operation or maintenance of this device Notice is important but not hazard related Hazard warnings are not included here 3 CLASSIFICATION AND COMPLIANCE 1 This device is classified as Class 1 type BF against electric shock Ordinary equipment without protection against ingress of water Equipment not suitable for use in presence of a flammable anesthetic mixture by standard of EN 60601 1 2006 Basic safety and essential performance of Medical Electrical Equipment 2 This device is complied with Class A for Noise Emission Level B for Noise immunity by standard of IEC 60601 1 2 2007 Electromagnetic Compatibility Requirements 4 SAFETY SYMBOLS AND INFORMATION The International Electrotechnical Commission IEC has established a set of symbols for medical electrical equi
33. n Analyzer amp Fluid Monitoring Systems 1 ID search Once you press START button on the initial screen the feleleelelelalelala eleleelelelelelela screen will be changed as follows 888 8 888 8 888 yrs button Then Information Search and Result During After DIALYSIS Search buttons will appear on the screen If you press ID SEARCH i INITIAL VIEW Result Search you can see the results for the ID and print Input an ID you would search on the ID blank and press 1 the results Information Search button has a function to import the input information of the saved ID 2 INDEX search In case of not using ID INDEX search can be used on the left upper side of the screen Press ID part on the screen and enter an index number and press button Then you can search or see the results and print them 40 1 Precaution for measurement The reliability of the results can be assessed by its accuracy The Accuracy of the device is determined by comparing the actual body composition and the results from Body Composition Analyzer The Reproducibility is determined when the device gives the identical results under the same conditions In order to maintain the accuracy of the results the following guidelines should be kept 1 Water volume increases after a meal Therefore measure on an empty stomach Measure 3 4 hours after a meal Avoid beverages containing
34. nt results in particular and shows the core items needed to control body composition The results are indicated systematically and anyone can easily understand the results 3 USB memory It is possible to check the result on PC through USB memory USB memory Bluetooth 4 Bluetooth It allows to do wireless communicate with a PC or mobile device x Model or specification of accessories can be changed according to market supply and demand 17 3 Appearance of the device 1 Front Part Wide Color LCD Operation Button Thermal Printer Start Stop Button Wide Color LCD The touch screen displays the inserted data procedure and results Operation Button It is for choosing or handling items by turning it right or left Start Stop Button It is for starting and stopping the machine Thermal Printer It prints the results on the thermal printer paper 18 2 Side Part Hand Electrode Foot Electrode Connection Port Connection Port Hand Electrode Connection Port Port for connecting electrode cables of right and left hands Foot Electrode Connection Port Port for connecting electrode cables of right and left feet 3 Rear Part ADAPTER 12V 5A USB B USB A RS 232C Power Switch Power Entrance USB A Port RS 232C Port USB B Port Power Switch Switching on or off Power Entrance Connecting adapter USB B Port Terminal for connecting the user s computer to the device USB
35. o fix the cable 4 Switch on the Bluetooth as show in the picture below 29 5 Attach the provided bracket to the Bluetooth and fix them on the back side of the device with a plus driver ea cme S OA a Bracket Combination et Bl 6 Connect the unconnected opposite part of the cable to the RS 232C port on the back side of the device and turn on the device ADAPTER 12V 5A USB B USB A RS 232C Gn i a JOSS y H E A M m 30 2 Connecting Computer Connect the USB B port placed on the rear panel of this device to the USB port in computer with USB cable ADAPTER 12V 5A USB B USB A RS 232C 00 080 oC 1 In order to save search and retrieve subject s data the device should be connected to PC in which Body Pass Plus is installed The result sheet can be printed from the printer connected to the PC 2 Professional consulting software Easy Body Plus option easily assesses and explains body composition analysis This program helps managing body composition easily and systematically It can help the measurer to analyze his her body composition at a glance 3 For the installation of Body Pass Plus and Easy Body Plus refer to its User s manual ol 3 Connecting Printer Option D Connecting the device computer and the printer Connect a computer to the USB B jack placed on the rear panel of the de
36. ot use this equipment because the microcurrent can cause malfunction in the implanted device To prevent fire hazard use only a correctly wired 100 240VAC outlet and do not use a MSO Multiple Socket Outlet that is not in compliance with IEC 60601 1 To reduce the risk of electric shock or product damage never plug in or plug out with wet hands Physically disabled persons should not attempt to take measurements alone but instead should have their caretakers assist them in using the device The unit must be operated only by or under supervision of a qualified person with our company or our distributors If you have experienced any trouble with the unit switch it off immediately and contact our company or its authorized dealer for assistance If you plan to connect any device from other manufacturers electrically or mechanically to the unit contact our company or its authorized dealer for instructions before doing so When you connect computer or other system to the unit RS 232C the attached systems should be those certified by IEC 950 or equivalent standards for data processing equipment Configurations shall comply with the system standard EN 60601 1 2006 Everybody who connects additional equipment to the signal input part or signal output part configures a medical system by standard EN 60601 1 2006 If in doubt consult the A S department of local distributor Avoid the following environments for storage W
37. pment which classifies a connection or warning of any potential hazard The classifications and symbols are shown below Save these instructions for your safety Degree of protection against electric shock TYPE BF e Please observe operating instructions Sa ee T Aw 2 OD e Waste Electrical and Electronic Eguipment WEEE The device could be sent back to the manufacturer for recycling or proper disposal after their useful lives Alternatively the device shall be disposed in accordance with national laws after their useful lives ON OFF key Turn the power ON OFF Class II equipment eR This symbol is used inside system Identifies the point where the safety LDO S ground of the system is fastened to the chassis Do not open This is for factory only Alternating current Date of manufacture Manufacturer TN Non ionizing radiation N p aa a O J CE Ec peP Authorized representative in the European community Keep dry a For indoor use only fo 5 SAFETY PRECAUTIONS This device is designed and manufactured with consideration of the safety of the operator and subject and also the reliability of the unit The following warnings precautions and notes must be observed for safety During measurement of the body composition a microcurrent of 180A flows through the body Individuals who have any kind of implanted active medical devices such as pacemakers should n
38. r is repeated please contact our company or its local distributor where this device is purchased e Input height correctly If the subjects height is out of the range height can t be entered 2 Error amp Repair P B F is measured too low or too high It does not work even when start buttons are correctly pressed e Measure in unstable condition such as right MON E e Measure again in a stable condition with the after the exercise bath LI correct posture sweat or drinking lots of l e Do not move or speak during the measurement water e Clean clip electrodes with soft gauze and try again e Moving or speaking e Clean hands and soles and try again during the measurement e Make sure there are no foreign substances Clip electrodes or between electrodes and measuring body parts measuring parts are dirty e Defective cable between the head and o e Contact our company or its local distributor where the scale MUN this device is purchased e Start buttons are defective e Bad connection e Contact our company or its local distributor where between the head and NUNN this device is purchased the scale D0 1 AFTER SERVICE Or there is any problem with the unit please follow the steps below x Contact our company s Overseas Service Department immediately After gathering the model name Serial Number date of purchase and description of the problem contact our company w
39. sec for 5 sec Magnetic field and environment If the user of the COZY fluctuations 930 requires continued operation l Power frequency magnetic fields should of commercial NN be at levels characteristic of a typical freguency 50 60Hz IEC 61000 4 8 location in a typical commercial or hospital environment UT is the a c mains voltage prior to application of the test level 13 3 Guidance and manufacturer s declaration electromagnetic immunity 2 The COZY 930 is intended for use in the electromagnetic environment specified below The customer or the user of the COZY 930 should assure that it is used in such an environment IEC 60601 test Compliance Electromagnetic environment Immunity test level guidance Portable an mobile communications equipment should be used no closer to any part of the COZY 930 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance d 1 2 P Conducted RF 3 Vrms 3 Vrms IEC 61000 4 6 150 kHz to 80 MHz d 1 2 P 80 MHz to 900 MHz d 2 3 P 900 MHz to 2 5 GHz Radiated RF 3 V m 3 V m a Oa S where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site
40. terference by maintaining a minimum distance between portable and mobile RF communications eguipment transmitters and the COZY 930 as recommended below according to the maximum output power of the communications eguipment Rated maximum Separation distance according to freguency of transmitter output power m of transmitter 150 kHz to 80 MHz 80 MHz to 900 MHz 900 MHz to 2 5 GHz d 1 NP d 1 N P d 1 2 P For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the eguation applicable to the freguency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer 1 At 80 MHz and 900 MHz the separation distance for the higher frequency range applies 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people 15 1 Basic Package weet O 3 User manual 4 Data management USB cable program 6 Power cable Adapter N Thermal printer paper H 8 Roller Seas 0 Electrode gel W G Exclusive cart The model or specification are subject to change according to the market demand 16 2 Option 1 Professional health counseling software CD This program helps managing body composition easily and systematically Also it interprets the measureme
41. that time is saved and SYSTEM SETUP screen appears To cancel any changes attempted press CLOSE then this device saves the previous date and time and SYSTEM SETUP screen appears 2 When software is used in data management measured date is automatically saved as the date set in this device Therefore the date and time set in this device should be checked before use lt VOLUME gt It adjusts the volume of voice guidance Select EN VOLUME on SYSTEM SETUP screen with V A button and press SET on touch pad Pre set 3 Range 0 9 Adjust volume with WV A button on touch pad Press SET on touch pad to save selected value Return to SYSTEM SETUP screen by pressing CLOSE on touch pad 36 lt PRINT Option gt Choose a printing mode among AUTO MANUAL or OFF Pre set OFF Select al PRINT to enter the menu Choose one button on touch pad Press SET button on touch pad to save it s Marat Return to SYSTEM SETUP screen by pressing CLOSE button on touch pad lt PRINT POSITION gt PRINT POSITION It adjusts the printing position in the direction of U D up down and L R left right to fit to the pre formatted result sheet Range 99 for U D and 99 for L R Select E PRINT POSITION to enter the menu Press the arrows which are surrounding the guadrangle as much as adjust Every single press moves print position by about 0 2 mm Choose U D by pressin
42. to balanced type that subcutaneous and visceral fat is balanced Level 9 10 corresponds to borderline type If subjects maintain current lifestyle they will proceed to visceral fat type Level 11 15 corresponds to visceral fat type Level 16 20 corresponds to visceral fat type 3 VFA Visceral Fat Area The optimal range is 50 100 cm male 40 80 cm female 4 V F M Visceral Fat Mass V F M is indicated in kg unit Target to Control It shows the mass of body fat soft lean mass and weight to be controlled This control is divided into 2 steps by comparing current state with the values based on standard weight At the first step of control target is to reach the minimum When assessed Under or maximum When assessed Over value of optimal range At the second step target is to come to the mean value of optimal range Body Composition Change Check to change of body fat soft lean mass and weight for 8 times Segmental Analysis Soft lean mass and mass of body fat of five body parts left and right arms left and right legs and trunk are analyzed Body Type Body type is determined by B M I and P B F Body type is classified into Low fat Low fat Muscular Over Weight Muscular Athletic Thin Standard Over Weight Borderline Low Muscular Over fat Thin Fat Obese Obesell Obese Ill B C M Body Cell Mass It consists of intra cellular water and protein o0 Phase angle Phase angl
43. transmission is complete the message The data writing was completed will be shown Press CLOSE to complete the saving process 2 Data Deletion Press the upper left corner for 5 seconds on initial screen Password window will appear Press 1111 on the password window The following message appears Do you want to delete all data If you want to delete the data press YES to delete otherwise press NO When deleting the data the message All data were deleted will appear Press Close to return to initial screen Deleted data can not be restored 47 This product shows not only result of analysis but also detailed data of body water impedance and body parts Here s the explanation and the criteria of the printed results Personal Data The subject s name ID date height weight age and gender are indicated on the result sheet Body Composition Analysis The body composition analysis is indicated in the ratio based on the subject s weight 1 Weight It is the sum of total body water mineral protein and body fat in the table 2 M B F Mass of Body Fat It is calculated by subtracting lean body mass from weight 3 L B M Lean Body Mass It is calculated by subtracting mass of body fat from body weight Lean body mass consists of fat free mass of body such as muscle organs blood and water 4 S L M Soft Lean Mass It consists of body water and protein It maintains body functions compos
44. uponnunnupunnnnanpununnansnnununnununnununnununD 55 1 AFTTER SERVICE A Ee eee ee ee ee 55 D PACKING AND TRANSPORT punonnannunununpunnnnannununnannnuunnnsunannansunununpunnunupunnunnanunnununn D5 SPECIFICATION Busuunansununnununnsnnunpununnannunununpununnassnnunnnsunuununpununnansunununnunnunupnunununpunnunann 56 WARRANTY A TCE EEE 59 We highly appreciate that you chose our company s product You are kindly requested to be familiar with these directions before using this product and always keep it together with the product In case you are not sure about any directions or problems arising while using the product please contact our service center We will provide you with detailed instructions 1 INTENDED USE This device measures impedance by bioelectrical impedance analysis method and provides lots of information using measured impedance and inputted personal data height age gender weight lt shows body composition of MBF LBM SLM SMM TBW protein mass mineral mass etc and information regarding BMI PBF BMR abdominal analysis Target to control segmental analysis Body composition change etc This device can also be used for the examinee who Is receiving kidney dialysis 2 WORD DEFINITIONS To ensure safe operation and long term performance stability it is essential that you fully understand the functions operating and maintenance instructions by reading this manual before operating your unit Particular attention
45. vice with USB cables Connect the printer to the computer with printer cable The result sheet can be printed out from the printer ADAPTER 12V 5A USB A RS 232C 2 Connecting the device and the printer directly In order to see the measurement result with A4 printer connect the A4 printer offered with this device to the USB A port placed on the rear panel of this device with USB cable and then print the result on A4 sheet ADAPTE RS 232C O 0000 O C I L 5 a 4 Replacing of thermal paper Replace thermal paper while the power is on 1 Push the side button while pulling down the lower button to open the cover 2 Insert the roller in the middle of the thermal paper 3 Insert the thermal paper with the roller in the grooves as shown in the picture 4 Slightly take out the edge of the thermal paper 5 Cover the printer 6 Thermal paper will be printed out and it automatically cuts out 4 1 Solution for thermal printer jam 1 Open the cover placed lower part as show in the picture 2 Push the button for jam inside the printer 3 Remove the jammed paper from the printer 33 SYSTEM SETUP allows users to change the setting of operational parameters Setup contents of this device are subject to change for efficient improvement 1 Entering SYSTEM SETUP Press the left upper edge of the initial screen for 3 4 seconds After it when a keypad