1. 8 TYPICAL REPORT TEMPLATES Carat content Procedure Gold Alr 101 8 1 9 Liquid 99 7 3 g Temperature 22 Density 19 36 g cm3 Volume 2 08 cm3 Percent 100 00 Carat content 24 00 K Carat content Procedure Carat K 2 Au Cu Air 217 1 7 9 Liquid 199 9 7 g Temperature 22 Density 12 60 g cm3 Volume 17 24 cm3 Percent 54 14 96 Carat content 12 99 K Carat content Procedure Platinum Pt Ni Air 210 0 0 g Liquid 199 9 9 g Temperature 22 Density 20 93 g cm3 Volume 10 03 cm3 Percent 98 25 PT class 982 46 PT User manual no LMI 64 02 01 13 ENG 21 MANUFACTURER OF ELECTRONIC WEIGHING INSTRUMENTS RADWAG WAGI ELEKTRONICZNE 26 600 Radom Bracka 28 POLAND Phone 48 48 384 88 00 fax 48 48 385 00 10 e mail export radwag com www radwag com Wm 01 9001 2008
2. 7 0 96828 0 96250 0 96190 0 96120 0 96050 fo 35875 0 95906 If a balance is connected with an external thermometer TB 2 series then its software automatically acquires temperature data from the thermometer The temperature value is updated in real time The temperature value is accompanied by a pictogram informing the user on automatic acquiring the temperature value from the plugged thermometer TB 2 series Metal content setup M8 01 Procedure b Golg 02 Liquid WATER 03 Temperature 22 C 04 Density 0 998 g cm 05 Run 12 4 TESTING CARAT CONTENT OF PURE GOLD Testing mode GOLD enables checking the purity of gold for the carat content from 9 0K 37 5 of pure gold to 24 0K 100 of pure gold Testing procedure e Select mode GOLD which opens the main window of the working mode Ek Metal content qnm Gold Weighing in the air Place an object Enter 0 000 4 10056 Esc 0 T e lf the indicated mass value is other than zero press button zeroing function e Place the sample to be tested on the pan assembly weighing in the air and on stabilization of measuring result press CY button The texts on the display change and indicate the following command 151212 Metal content 13 57 20 Gold Weighing in a liquid Place an object Enter 23 446 o 10055 d Unload the tested sample from the pan assembly As the mass indication returns to ZERO place the test
3. Pt Ni platinum nickel Pt Pd platinum palladium Alloyed platinum density table The table represents the density value of alloy minor constituents dependent on platinum class in the tested sample Platinum Content Alloyed metal density g cm Sample density class ei range 100 100 21 45 21 45 21 24 21 66 95 0 100 20 04 20 64 19 84 20 85 Testing procedure e Select procedure PLATINUM which opens the main window of the working mode Es 0 T e lf the indicated mass value is other than zero press l i button zeroing function e Place the sample to be tested on the pan assembly weighing in the air and on stabilization of measuring result press button i7 Unload the tested sample from the pan assembly As the mass indication returns to ZERO place the tested sample on the grill pan located in the container and immersed in the liquid te weighing in liquid On stabilization of measuring result press button A window on testing process completion is displayed After completing the testing procedure the balance automatically sends a report with process summary and measurement result to be printed on a connected printer Multiple pressing of button enables repeated printing of the report 7 DETERMINING PERCENT CONTENT OF MAJOR ALLOY COMPONENT IN A TESTED SAMPLE Testing mode lt ALLOY gt enables determining percent content of major metal in a 2 component or 3 co
4. User Manual User manual no LMI 64 02 01 13 ENG METAL PURITY DETERMINING KIT For a balance PS series BALANCES AND SCALES RADWAG 26 600 Radom Bracka 28 POLAND Phone 48 0 48 38 48 800 fax 48 0 48 385 00 10 export radwag com www radwag com JANUARY 2013 WR N TABLE OF CONTENTS DIMENSION LE 5 M AEN tobe d EE 5 METAL PURITY TESTING MODE e ree nnne nnne 8 cM legen TING i i M P 8 3 2 CARAT CONTENT OF PURE GOLD MENU eene 9 3 3 CARAT CONTENT OF ALLOYED GOLD MENU 10 3 4 PLATINUM CLASS MENU lie i ka aa 11 3 5 CONTENT OF MAJOR ALLOY COMPONENT MENU 11 3 6 DECLARING WATER TEMPERATURE sese 12 TESTING CARAT CONTENT OF PURE GOLD 13 TESTING CARAT CONTENT OF ALLOYED GOLD 15 Modedocal SCTINOS EE 15 Alloyed gold density le 16 TOSUAG eelere Er 16 TESTING PLATINUM CLASS 11er erre nnn nnnm 17 Mode Ee EE 17 Alloyed platinum density table nn 17 TOSUAG geleed 17 DETERMINING PERCENT CONTENT OF MAJOR ALLOY COMPONENT IN A TESTED SAMPLE eee 19 Modedo6al SCHINOS EE 19 Testing procedure NM 19 TYPICAL REPORT TEMPLATES eee ree nnne 21 1 DIMENSIONS 233 Dimensions of a bal
5. an alloy d2 Pe Density of second metal in an alloy d3 Density of third metal in an alloy Refers to ALL 3 alloy only ratio of secondary alloy constituents 20 5 Refers to ALL 3 mode only Available values 4 6 5 5 6 4 7 3 8 2 Other mode settings are coherent with the lt Gold gt procedure 3 6 DECLARING WATER TEMPERATURE When the software calculates density of a tested sample it takes into account data on density of liquid in which the tested sample is weighed Liquid density is dependent on its temperature In case of distilled water the density value is acquired from density tables implemented in the balance software see below table After specifying water temperature its density is automatically updated and used for the calculation 0 0 99986 0 99990 0 99904 0 09996 0 99997 0 99996 0 99904 0 99900 0 99085 0 09978 20 0 99820 0 99799 0 89777 0 99764 0 09780 0 99704 0 99678 0 99651 0 09623 0 99504 30 099565 0 98584 0 80503 0 99470 0 99487 0 99408 0 90368 0 99508 0 99207 0 90250 40 0 99222 0 99168 0 99144 0 99708 0 90068 10 90021 0 98070 0 98036 0 98898 0 98840 a 0 96808 098758 10 98712 0 99668 0 98618 0 90570 0 98821 0 9647 Josee 0 96071 60 028820 ossese sete nsetea 0 96110 0 9085 0 98001 0 97946 0 97890 0 97604 80 osriso 0 97117 0 97054 0 96991 0 96027 0 98862 0 96707 0 96791 o 36665 0 96600 a 0 96532 0 96465 0 9690
6. ance PS X AU series 2 KIT ASSEMBLV A Remove the standard weighing pan from the balance Next remove the rubber mandrels positioning the pan and replace them with mandrels without rubber absorbers that are supplied with the kit for determining carat content B Assembly the kit components in the following sequence basis of the container for liquid item 1 next place weighing pan basis item 2 onto the previously assembled mandrels and finally place the container for liquid item 3 Fill the container with liquid for carrying out measuring processes C On the weighing pan basis place an assembly with attached grill pan D Assemble liquid temperature sensor TB 2 Pay attention that the sensor probe does not touch the assembly C or the grill pan Caution TB 2 thermometer is optional equipment of the kit for determining carat content 3 METAL PURITY TESTING MODE A balance PS X AU series implements the following testing procedures in the metal purity testing mode Gold Testing carat content of pure gold Testing carat content of gold alloy gold with additive Au Ag Cu gold copper silver Au Cu gold copper Au Ag gold silver Testing platinum class platinum with additive Pt Ni platinum nickel Pt Pd platinum palladium Determining percent content of major metal in the tested alloy ALL 2 2 component alloy ALL 3 3 component alloy The parameter settings of each t
7. ed sample on the grill pan located in the container and immersed in the liquid i i weighing in liquid On stabilization of measuring result press button e The texts on the display change and inform on finishing the testing process with simultaneous displaying the carat content of tested gold Td AA Metal content 13 57 20 Gold Wi 2344B VW7 22 72H tacantinue F 13 09 K 10056 After completing the testing procedure the balance automatically sends a report with process summary and measurement result to be printed on a connected printer 44 Multiple pressing of button enables repeated printing of the report 5 TESTING CARAT CONTENT OF ALLOYED GOLD Testing mode lt KARAT gt enables checking one of three gold alloys e Gold alloyed with silver and copper mode Au Ag Cu e Gold alloyed with copper mode Au Cu e Gold alloyed with silver mode Au Ag Mode local settings Before starting the measuring process set testing mode parameters Parameter Initial name settings Select alloyed gold 4 3 1 Alloy Au Ag Cu Au Ag Cu gold copper silver Au Cu gold copper Au Ag gold silver Description ratio of copper and silver 4 3 2 Ratio Cu Ag 7 3 refers to Au Ag Cu alloy only Available values 4 6 5 5 6 4 7 3 Alloyed gold density table The table represents the density value of alloy minor constituents dependent on gold carat content i
8. esting procedure is carried out in the submenu of the selected testing mode 3 1 MODE ACTIVATING 121212 Weighing 13 57 20 0 0000 M2 Checkwelghing M3 Filling 5 M4 Percent W d M5 Animal weighing Ho Mb Density 135428 M7 Formulation ME amp Metal contenet 3 2 CARAT CONTENT OF PURE GOLD MENU Metal content setup Procedure Liquid Temperature Density ILI l IS gt e Platinum choice Alloy Procedure Select a testing procedure from balance menu Liquid select liquid to be used in the measuring process distilled water other liquid with determined density Temperature The temperature value is determined using an external thermometer immersed in the liquid The temperature value will be considered while calculating the carat content of pure gold In case the selected liquid is OTHER enter its density value to balance memory If the balance is connected with an external thermometer TB 2 series then its software automatically acquires temperature data from the thermometer The temperature value is updated in real time The temperature value is accompanied by a pictogram informing the user on automatic acquiring the temperature value from the plugged thermometer TB 2 series Density Liquid density is dependent on temperature In case of distilled water the density value is acquired from density tables saved in balance s memory After specifying distilled water temperature its density is automatically upda
9. mponent alloy Mode local settings Before starting the measuring process set testing mode parameters Parameter Initial Bie Select alloyed metal type ALL 2 2 component alloy BOV Type ieh ALL 3 3 component allov 452 di Ja Densitv of major metal in an allov 1453 a l I Density of second metal in an alloy asal Ir Density of third metal in an alloy Refers to ALL 3 alloy only ratio of secondary alloy constituents 4 5 5 Ratio 515 Refers to ALL 3 mode only Available values 4 6 5 5 6 4 7 3 8 2 Testing procedure e Select mode ALLOY which opens the main window of the working mode m Ese 0 T e lf the indicated mass value is other than zero press i button zeroing function e Place the sample to be tested on the pan assembly weighing in the air and on stabilization of measuring result press button e Unload the tested sample from the pan assembly e As the mass indication returns to ZERO place the tested sample on the grill pan located in the container and immersed in the liquid to weighing in liquid On stabilization of measuring result press button e A window on testing process completion is displayed sou e After completing the testing procedure the balance automatically sends a report with process summary and measurement result to be printed on a connected printer e Multiple pressing of button enables repeated printing of the report 20
10. n the tested sample 3 Gold Content Alloyed metal density g cm Sample caratage 9 density range a Oe Tee om 100 100 19 32 19 32 19 32 19 13 19 51 91 6 100 17 73 17 63 18 06 17 45 18 24 Testing procedure e Select procedure cCARAT which opens the main window of the working mode Esc 0 T e lf the indicated mass value is other than zero press i button zeroing function e Place the sample to be tested on the pan assembly weighing in the 44 air and on stabilization of measuring result press button e Unload the tested sample from the pan assembly e As the mass indication returns to ZERO place the tested sample on the grill pan located in the container and immersed in the liquid __ weighing in liquid On stabilization of measuring result press button e A window on testing process completion is displayed e After completing the testing procedure the balance automatically sends a report with process summary and measurement result to be printed on a connected printer e Multiple pressing of button enables repeated printing of the report 16 6 TESTING PLATINUM CLASS Testing mode cPLATINUM enables checking one of two platinum alloys e Platinum alloyed with nickel mode Pt Ni e Platinum alloyed with palladium mode Pt Pd Mode local settings Before starting the measuring process set testing mode parameters Testing alloyed platinum alloy selection
11. ted In case of OTHER liquid the density value has to be entered manually otart Initiate carat content testing process of pure gold 3 3 CARAT CONTENT OF ALLOYED GOLD MENU Metal content setup Mee 0l Procedure Liquid Temperature Density Alloy Up Ratio Cu Ag Of Run K tol gt e Platinum choice Alloy Out Alloy type e Gold alloyed with silver and copper mode Au Ag Cu e Gold alloyed with copper mode Au Cu e Gold alloyed with silver mode Au Ag rtk ratio of copper and silver refers to Au Ag Cu mode only Available settings 4 6 5 5 6 4 7 3 Other mode settings are coherent with the Gold procedure 3 4 PLATINUM CLASS MENU Metal content setup Procedure Platinum Liquid Temperature Density Alloy Alloy FE Allov tvpe e Platinum alloyed with nickel mode Pt Ni e Platinum alloyed with palladium mode Pt Pd Other mode settings are coherent with the lt Gold gt procedure 3 5 CONTENT OF MAJOR ALLOY COMPONENT MENU Metal content setup MS 01 Procedure Alloy 02 Liquid WODA 03 Temperature 22 C 04 Density 0 998 g cm 05 Alloy type b AI 2 Metal content setup b Alloy type MISBO5SBOL in AL 2 02 dl 8000000 g cm E LEE 03 d 2700000 g cm 04 o 1000000 g cm 05 Ratio 5 6 11 n alloy type Metal alloy selection STEE Hee ALL 2 2 component alloy ALL 3 3 component alloy d ie Density of major metal in