Home

Restore 5000 user manual

image

Contents

1. 27 lite ae tesis o lies E AA ARN 27 About Power Outage and Transportation sessssssssessessesssssssssesssesssssecsucsssscscsussscssnesstesuscssccscssseseccasssucsasesuessscsuceasesuessnecscessessessueeasesuessesanesaeessesaessesseesneesees 28 5 MAINTENANCE AND TROUBLESHOOTING 29 Cleaning he System mir sis ctisecostesiceadichoaistbendeshsuecuiassaassasndssossdcensscensasbovpnszsiucesdsdeasbesunssussnasccuacbtebedaazandsanesedesstactsnagensbuhesecdusetuunbayacsiscaucededasbetavenpdenctenccacbraspiastoosen 29 Setcases item du demo senat cames 32 Restore 5000 3 Restorative Innovations Inc GENERAL enera I 1 Symbols Warnings Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage See the information below for definitions of the signal words DANGER Danger indicates an imminently hazardous situation which if not avoided will result in death or serious injury WARNING Warning indicates a potentially hazardous situation which if not avoided could result in death or serious injury CAUTION Caution indicates a potentially hazardous situation which if not avoided may result in property damage or minor injury or
2. The minimum pressure setting during Alternating Pressure is set to 0 2 mmHg at the factory This means that the mattress will deflate as much as possible during the deflation time when using 10 15 or 20 minute Alternating Pressure times The setting can be changed so the mattress will only deflate to 50 of the selected pressure setting Ensure the Standby LED is lit 2 Perform one of the following Restore 5000 21 Restorative Innovations Inc 4 OPERATION e Press and hold the Mode and Fowler buttons 3 Read the LCD Display The LCD Display will flash 0 if the current setting is 0 2 mmHg 5 if the current setting is 50 of the selected pressure 4 Perform the following to toggle between the two settings e Press the Static Alternating Pressure button 5 Press the Power button to save the setting Firm Soft Buttons Select comfort pressure settings by pressing Firm or Soft 7 button e Soft Button Pressing this button reduces the pressure setting in the mattress e Firm Button Pressing this button increase the pressure setting in the mattress The patient comfort pressure ranges from Soft level 0 6 4 mmHg to Firm level 9 3246 mmHg The Comfort Control LED displays the patient comfort pressure levels from 0 to 9 and provides a guide to the caregiver to set approximate comfort pressure level depending on the patient weight If the patient s weight to height ratio is above average incr
3. both gt IMPORTANT Indicates a hazardous situation that could result in damage to property if it is not avoided Gives useful tips recommendations and information for efficient trouble free use Restorative Innovations Inc 4 Restore 5000 GENERAL 1 2 Limited Warranty PLEASE NOTE THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4 1975 This warranty is extended only to the original purchaser who purchases this product when new and unused This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person This warranty gives you specific legal rights and you may also have other legal rights which vary from state to state Restorative Innovations warrants the mattress and cover when purchased new and unused to be free from defects in materials and workmanship for a period of one year from the date of purchase with a copy of the seller s invoice required for coverage under this warranty Restorative Innovations warrants the control unit when purchased new and unused to be free from defects in materials and workmanship for a period of two years from the date of purchase from Restorative Innovations with a copy of the seller s invoice required for coverage under this warranty If w
4. is fully inflated place the patient on the mattress Ensure that the patient s feet are towards the foot end of the mattress the end with the hose Center the patient on the bed from side to side and head to foot Special positioning may be required with contracted patients to provide comfortable positions After placing the patient make certain no objects will fall under the patient such as feeding tubes IV s etc Wait five minutes for the mattress pressure to stabilize Once the mattress inflates to its normal size set the comfort pressure to the desired comfort level Wait five minutes for the mattress pressure to stabilize Restore 5000 Restorative Innovations Inc 29 4 OPERATION 9 Verify that the patient has not bottomed out by performing the following steps Ensure that the patient is lying flat on his her back in the middle of the mattress Place four fingers between the air cushions directly underneath the sacral region of the patient s body Ensure that there is 3 to 4 finger width clearance between the bottom of the patient and the bed frame Adjust the comfort setting if needed gt VVait five minutes for the mattress pressure to stabilize 7 Repeat STEPS A E until patient has not bottomed out and patient comfort is achieved 10 If the patient feels that the bed is too soft hard press the Soft or Firm button to adjust the comfort settings 11 Use a regular pillow to help support and st
5. launder it using the following method C AUTION Laundry workers should always wash their hands before working with clean bedding DO NOT overload the machine DO NOT use chlorine bleach because it may damage the fabric coating High air temperatures will damage the fabrics and void the warranty i Place the cover in a washing machine ii Wash with warm water below 120 ii Add detergent and disinfectant according to the manufacturer s instructions iv Remove excess water Set the dryer to the lowest setting below 20 F vi Dry the cover until it is completely dry 9 Perform the following steps to clean the control unit and hose fittings Wipe all controls chassis and hose fittings with a quaternary disinfectant solution Using a nylon brush gently clean all crevices as they can harbor microorganisms C Air dry all treated surfaces Restorative Innovations Inc 30 Restore 5000 5 MAINTENANCE AND TROUBLESHOOTING 10 Perform the following steps to clean the mattress components air cushions mattress base etc A Using a clean sponge or paper towel wipe down the mattress components with a dilute detergent solution of quaternary cleaner disinfectant or other germicidal detergent solution B Wipe the mattress components with a clean dry cloth Storing the System Ensure that the control unit is off and disconnect the power cord from the wall outlet 2 Clean the system Refer to Cleani
6. MPONENT WITHOUT THE SPECIFIC CONSENT OF RESTORATIVE INNOVATIONS PRODUCTS DAMAGED BY CIRCUMSTANCES BEYOND RESTORATIVE INNOVATIONS CONTROL PRODUCTS REPAIRED BY ANYONE OTHER THAN RESTORATIVE INNOVATIONS SUCH EVALUATION SHALL BE SOLELY DETERMINED BY RESTORATIVE INNOVATIONS THE FOREGOING EXPRESS WARRANTY IS EXCLUSIVE AND IN LIEU OF ALL OTHER EXPRESS WARRANTIES WHATSOEVER WHETHER EXPRESS OR IMPLIED INCLUDING THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND THE SOLE REMEDY FOR VIOLATIONS OF ANY WARRANTY WHATSOEVER SHALL BE LIMITED TO REPAIR OR REPLACEMENT OF THE DEFECTIVE PRODUCT PURSUANT TO THE TERMS CONTAINED HEREIN THE APPLICATION OF ANY IMPLIED WARRANTY WHATSOEVER SHALL NOT EXTEND BEYOND THE DURATION OF THE EXPRESS WARRANTY PROVIDED HEREIN RESTORATIVE INNOVATIONS SHALL NOT BE LIABLE FOR ANY CONSEQUENTIAL OR INCIDENTAL DAMAGES WHATSOEVER THIS WARRANTY SHALL BE EXTENDED TO COMPLY WITH STATE PROVINCIAL LAWS AND REQUIREMENTS Restore 5000 5 Restorative Innovations Inc 2 OVERVIEVV 2 Overview 2 1 Label Location DANGER EXPLOSION HAZARD DO NOT use in the presence of flammable anesthetics CAUTION Equipment should be connected to a properly grounded receptacle 3 prong Risk of Electrical shock DO NOT remove back Disconnect air hose before administering CPR This label is on the back of the unit and also contains the serial number WARNING Patient entrapment with bed side rails ma
7. When the mattress has reached maximum firmness perform one of the following e to Wheelchair Transfer Slide the patient onto the wheelchair e Wheelchair to Bed Transfer Slide the patient onto the bed 4 10 Preparing for CPR Procedure For this procedure refer to FIGURE 6 on page 28 Press and hold the tabs on hose connector while pulling the hose from the control unit 2 Disconnect the RED CPR connector located on the side of the mattress Restore 5000 Restorative Innovations Inc 27 4 OPERATION Control Unit Connectors LP Mattress Tab RED CPR Connector Connector Hose FIGURE 6 Preparing for CPR Procedure 4 11 About Power Outage and Transportation If the hoses remain connected to the control unit the mattress retains air during a power outage during transportation when the control unit is unplugged or when the control unit is turned Off If the hose becomes disconnected or damaged and the mattress deflates the mattress has a 2 inch foam pad to provide patient support Restorative Innovations Inc 28 Restore 5000 5 MAINTENANCE AND TROUBLESHOOTING aintenance and Troubleshooting 5 1 Cleaning the System WARNING Before cleaning or disassembling the Restore 5000 check the underside of the mattress folds for sharp objects such as scissors needles etc These objects should be removed and discarded before proceeding with further cleaning or disassembly otherwise injury or
8. abilize the patient s head 4 8 Transferring Patient From To a Gurney WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the gurney before transferring the patient between the bed and the gurney Engage the wheel locks of the bed Refer to the owner s manual provided with the bed Engage the wheel locks of the gurney Refer to the owner s manual provided with the gurney Press the Max Inflate or Mode button to achieve maximum mattress pressure Raise or lower the bed to match the gurney height Refer to the owner s manual provided with the bed a When the mattress has reached maximum firmness perform one of the following Bed to Gurney Transfer Slide the patient onto the gurney e Gurney to Bed Transfer Slide the patient onto the bed Restorative Innovations Inc 26 Restore 5000 4 OPERATION 4 9 Transferring Patient From To a Wheelchair WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the wheelchair before transferring the patient between the bed and the wheelchair Engage the wheel locks of the bed Refer to the owner s manual provided with the bed Engage the wheel locks of the wheelchair if applicable Refer to the owner s manual provided with the wheelchair Press the Max Inflate or Mode button to achieve maximum mattress pressure Raise or lower the bed to match the wheelchair height Refer to the owner s manual provided with the bed uk wn
9. attress to the bed frame f X Head End Bed Decle Center of Bed Frame the head and foot ends Secure ALL mattress Ni 25 Foot End Bed straps Secure the straps to the bed deck at the ss x head and foot ends and to the frame at the center 4 of the bed Otherwise damage to the mattress will occur when the head and foot ends are raised ell DO NOT Strap Here The powered mattress comes with ten nylon buckle straps Remove the original foam mattress from the bed 2 necessary lower the side rails to facilitate installation of the mattress 3 Unroll the powered mattress and place it on the bed frame Ensure that the hose is towards the foot end of the bed 4 Use the buckle straps to secure the powered mattress to the bed deck in the following locations e Head End Head End Bed Deck e End Foot End Bed Deck e Center Center of the Bed Frame Restore 5000 Restorative Innovations Inc 15 4 OPERATION Head End Povvered Mattress Buckl Straps Foot End Hose FIGURE 2 Installing the Mattress Mattress Replacement System Restore 5000 Installing the Side Rails A VVARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Op
10. damage to the product may occur Because of potential risk of infectious exposure cleaning with the patient on the bed is not recommended All equipment should be inspected Any item that is visibly soiled with the patient s blood or other body fluids should be properly cleaned or removed Laundry workers should treat all soiled bedding as if it were contaminated with pathogenic microorganisms I Remove the bedding If necessary inflate the mattress Ensure that the control unit is off Unplug the power cord from the wall outlet Ensure that the underside of the mattress is clear of all sharp objects Examine the surface of the control unit and mattress assembly components for visible blood or body fluids SSOP CO a I Perform one of the follovving e If blood is present decontaminate the product i Remove all visible soil with disposable paper towels ii Scrub the area with freshly prepared effective detergent disinfectant solution blood is not present remove any soil from the cover with paper towels Restore 5000 Restorative Innovations Inc 29 5 MAINTENANCE AND TROUBLESHOOTING If soiled the cover should be removed cleaned and decontaminated 8 Perform the following steps to clean the mattress cover A Using a clean sponge or paper towel wipe down the cover surface with dilute detergent solution of quaternary cleaner disinfectant or other germicidal detergent solution B Remove the cover and
11. ding the On button to prevent any tampering with the settings Press the button for approximately 3 5 seconds for the Lock LED to activate The Lock LED illuminates when in this mode Alarm Silence Button Press the Alarm Silence button to silence the alarm that sounds the event of povver failure or vvhen the hose is disconnected from the control unit The Alarm LED illuminates when in this mode Power Fail LED In the event of power outage an alarm sounds and Power Fail LED flashes AMBER The control unit has internal memory and retains the previous settings during the power outage During a power outage the mattress retains the air as long as the mattress is connected to the control unit Low Pressure LED In the event that the mattress hose disconnects an alarm sounds and Low Pressure LED flashes AMBER Once the low pressure problem is fixed the control unit resumes operation in the previously set mode Restorative Innovations Inc 24 Restore 5000 4 6 4 1 On oy 4 OPERATION Powering Up the System For this procedure refer to FIGURE 5 on page 20 Turn on the power to the system by pressing the Power button on the control unit Once the button is released a green led illuminates when the unit is on Placing the Patient on the Mattress Perform the following Press the Mode button to select Max Inflate Mode In this mode the Max Inflate LED lights up After the mattress
12. ease the pressure setting by approximately 20 Fowler Button Press the Fowler button S to activate the patient fowler mode The Fowler LED illuminates when in this mode When this mode is activated the control unit increases the pressure in the mattress to prevent the patient from bottoming out Restorative Innovations Inc 22 Restore 5000 4 OPERATION Mode Button Max Inflate Low Air Loss Press the Mode button gt to select the Max Inflate mode or the Low Air Loss mode e Max Inflate Mode In this mode the mattress inflates rapidly to maximum firmness pressurized to 35 6 mmHg A series of beeps sound every three minutes as a reminder that the Max Inflate mode is active After 45 minutes the Max Inflate mode deactivates and the control unit defaults to the previous setting It takes 20 45 minutes for the mattress to inflate fully inflation time depends on size of mattress It is recommended that Max Inflate setting be used during patient ingress egress patient wound care patient turning or patient cleaning Low Air Loss Mode In this mode the mattress goes into on demand low air loss relief mode 2 Press the Mode button to manually disengage Max Inflate or Low Air Loss mode This will deactivate the Max Inflate and Low Air Loss LEDs Restore 5000 Restorative Innovations Inc 23 4 OPERATION Lock Button Press the Lock button to lock out all control unit functions inclu
13. erator s Manual prior to using this equipment Restorative Innovations Inc 16 Restore 5000 4 OPERATION Refer to the instructions provided with the side Bed Hooks 7 rails for the installation procedure Installing the Power Unit Swing out the bed hooks on the back of the control unit 2 Place the control unit on the footboard If the bed does not have a footboard place the control unit on a flat surface leaving room for the hose to hang down FIGURE 3 Installing the Power Unit 43 Connecting the Hose MN CAUTION Ensure that the hose connecting the control unit to the mattress is routed such that it cannot be stepped on kinked squeezed or otherwise damaged For this procedure refer to FIGURE 4 on page 18 Locate the hose at the foot end of the mattress 1 2 Locate the control unit connectors on the right side of the control unit 3 Squeeze and hold the tabs on the hose connectors 4 Push the hose connectors onto the control connectors until an audible click is heard The audible click indicates that the hose connectors are properly engaged with the control connectors Restore 5000 Restorative Innovations Inc 17 4 OPERATION Control Unit Connectors Connector FIGURE 4 Connecting the Hose Restorative Innovations Inc 18 Restore 5000 4 OPERATION Connecting the Povver Cord VVARNING DO NOT alter plug to fit a non conforming outlet Instead have an electrician in
14. estore 5000 Alternating Pressure Low Air Loss Mattress System User Manual This manual MUST be given to the user of the product BEFORE using this product read this manual and save for future reference Restorative Innovations Inc 2 Restore 5000 CONTENTS GENERAL 4 SYMDOLS cs EN A 4 Limited 5 2 OVERVIEW 6 Label Location moment Typical Product Parameters 3 GENERAL GUIDELINES 9 General Guidelines i Ai e aaao a Aaea aean E aaa EEEa EEEa 9 4 OPERATION 14 Installing the Mattress tit dl i a i a ta E LU 14 Installing th Side actih donee A Manes S 16 Installing the Povver Unit sens Conesa EIs 17 Connecting the Hose Connecting the 19 Using the Front Panell Sete A E E 20 Powering Up the 25 Placing the Patient on the Mattress te ESC 25 Transferring Patient 26 Transferring Patient From To
15. ible and cause a fire Failure to observe this warning can result in severe fire property damage and cause physical injury or death Smoking by visitors in the room will contaminate the system Therefore visitor smoking is NOT permitted Anesthesia Equipment There is an explosion risk if used with flammable anesthetics Oxygen There is a possible fire hazard when used with oxygen administering equipment other than nasal mask or half bed tent type The oxygen tent should NOT extend below mattress support level Restore 5000 Restorative Innovations Inc 11 3 GENERAL GUIDELINES Entrapment May Occur VVARNING Patient entrapment with bed side rails may cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Operator s Manual prior to using this equipment Proper patient assessment and monitoring and proper maintenance and use of equipment is required to reduce the risk of entrapment Variations in bed rail dimensions and mattress thickness size or density could increase the risk of entrapment Visit the FDA website at http www fda gov to learn about the risks of entrapment Refer to the owner s manuals for beds and rails for additional product and safety information After any adjustments repair or service and before use make sure all attaching hardware is tightened secu
16. ithin such warranty period any such product shall be proven to be defective such product shall be repaired or replaced This warranty does not include any labor or shipping charges incurred in replacement part installation or repair of any such product Restorative Innovations sole obligation and your exclusive remedy under this warranty shall be limited to such repair and or replacement For warranty service please contact Restorative Innovations Provide the date of purchase indicate nature of the defect and if the product is serialized indicate the serial number The defective unit or parts must be returned for warranty inspection using the serial number when applicable as identification within thirty days of return authorization date DO NOT return products to our factory without our prior consent C O D shipments will be refused please prepay shipping charges LIMITATIONS AND EXCLUSIONS THE WARRANTY SHALL NOT APPLY TO PROBLEMS ARISING FROM NORMAL WEAR OR FAILURE TO ADHERE TO THE ENCLOSED INSTRUCTIONS IN ADDITION THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BEEN REMOVED OR DEFACED PRODUCTS SUBJECTED TO NEGLIGENCE ACCIDENT IMPROPER OPERATION MAINTENANCE OR STORAGE OR PRODUCTS MODIFIED WITHOUT RESTORATIVE INNOVATIONS EXPRESS WRITTEN CONSENT INCLUDING BUT NOT LIMITED TO MODIFICATION THROUGH THE USE OF UNAUTHORIZED PARTS OR ATTACHMENTS PRODUCTS DAMAGED BY REASON OF REPAIRS MADE TO ANY CO
17. ndicating that the system is in stand by mode Once the Power button is pressed and released a green led illuminates indicating that the control unit is on If the power comes on after a power outage the control unit will go through its system initialization routine for a few seconds and then resume the desired function Static Alternating Pressure Button Press the Alternating Pressure button 62 to select betvveen Static Mode and one of four Alternating Pressure Times The alternating pressure times indicate the frequency of deflation or inflation of half of the air cushions even or odd numbered For example 10 minutes is selected In this example the even numbered air cushions in the mattress will deflate while the odd numbered cushions remain at constant pressure When 10 minutes has elapsed the even numbered air cushions will inflate while the odd numbered cushions remain at constant pressure When 10 minutes has elapsed again the odd numbered air cushions deflate while the even numbered cushions remain at constant pressure The air cushions continue to inflate or deflate this as long as an alternating pressure time has been selected At all times half of the air cushions remain at a constant pressure when an alternating pressure time has been selected In Static Mode all air cushions in the mattress are maintained at constant pressure The Static LED is illuminated Green Adjusting the Minimum Pressure Setting A P Low
18. ng the System on page 29 3 Disconnect the air hose connector s from the control unit and allow air to vent from the mattress 4 Gently roll up the mattress with minimal handling and agitation 5 Ensure the cover surface is inside the roll 6 Store the unit keeping the mattress with the control unit Restore 5000 Restorative Innovations Inc 31 5 MAINTENANCE AND TROUBLESHOOTING 5 2 Troubleshooting PROBLEM CAUSE SOLUTION Mattress not inflating Not alternating Mattress hose disconnected Connect hose connectors and locle them in place Air hose kinked or split Unkink hose or replace split hose Major leak in the air cushions or overlay pad Replace leaking air cushions or overlay pad Kinked or split manifold Unkink manifold or replace split manifold Has power and fuse is good control unit does not come on Send control unit back to factory for repair Not alternating solenoid malfunction Send control unit back to factory for repair No air pump malfunction Send unit for repair No power Control unit off Check power source and turn on unit Power cord disconnected Connect power cord to the power source No power in the power source Power outage Check power source has power and turn it On Wait till the power source has power Blown fuse Restorative Innovations Inc 32 Replace blovvn fuse vvith an equivalent fu
19. rameters CONTROL UNIT DIMENSIONS LX VV XH WEIGHT Control unit and mattress have Latex Free com ponents Restore 5000 i 9 x 5 x 5 5 Ibs POWER CORD 10 14 Feet Long Hospital Grade CONNECTION Three 4 Flow Couplings PACKAGING piece box AIR FILTER Restorative Innovations Inc None Restore 5000 3 GENERAL GUIDELINES enera 3 1 General Guidelines WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any additional instructional material such as owner s manuals service manuals or instruction sheets supplied with this product or optional equipment If you are unable to understand the warnings cautions or instructions contact a healthcare professional dealer or technical personnel before attempting to use this equipment otherwise injury or damage may occur NOTICE THE INFORMATION CONTAINED IN THIS DOCUMENT IS SUBJECT TO CHANGE WITHOUT NOTICE Check all parts for shipping damage and test before using In case of damage DO NOT use Contraindications WARNING ALWAYS consult the patient s physician before using the Restore 5000 system Restore 5000 9 Restorative Innovations Inc 3 GENERAL GUIDELINES Installation VVARNING The Restore 5000 is recommended to be installed on medical bed frames with bed side or assist rails It i
20. rely Assist rails with dimensions different from the original equipment supplied or specified by the bed manufacturer may not be interchangeable and may result in entrapment or other injury Restorative Innovations Inc 12 Restore 5000 3 GENERAL GUIDELINES Electrical DANGER Electrical shock hazard DO NOT remove cover Refer to qualified service personnel WARNING Before performing any maintenance to the power unit disconnect the power cord from the wall outlet DO NOT insert items into any openings of the control unit Doing so may cause fire or electric shock by shorting the internal components The control unit MUST be kept away from all heat sources and radiators during operation Connect the equipment to properly grounded three prong wall outlet using 0 14 ft hospital grade power cord provided with the product Grounding reliability depends upon a properly grounded receptacle 3 prong Restore 5000 Restorative Innovations Inc 13 4 OPERATION peration AA CAUTION The control unit and mattress on the Restore 5000 series are designed to be used as a system DO NOT replace mattresses or control units with other models or other brands Otherwise damage to the system may occur Contact your supplier to get the correct replacement if needed Restorative Innovations Inc 14 Restore 5000 4 OPERATION 4 1 Installing the Mattress Mattress Replacement System Restore 5000 CAUTION 22 DO NOT strap the m
21. s preferred that the rails be in the raised position whenever a patient is on the bed Health care professionals assigned to each case should make the final determination whether side or assist rails are warranted after assessing patient risks of entrapment and falls in accordance with State patient restraint legislation or facility interpretation of such legislation Controls on the footboard may be obstructed by the power unit on a few bed frames It may be necessary to relocate the power unit Refer to Installing the Power Unit on page 17 Check that air hoses and power cord are clear of moving bed components before placing a patient on the bed Operate all bed frame motorized functions through their full range of motion to be certain that there is no pulling interference or pinching Do not strap the mattress to the bed frame at the head and foot ends Secure ALL mattress straps Secure the straps to the bed deck at the head and foot ends and to the frame at the center of the bed Otherwise damage to the mattress will occur when the head and foot ends are raised Refer to Installing the Mattress on page 14 Restorative Innovations Inc 10 Restore 5000 3 GENERAL GUIDELINES Fire Hazard DANGER Das Smolcing DO NOT SMOKE while using this device This system uses room air for circulation through the mattress A cigarette can burn a hole in the bed surface and cause damage to the mattress Also patient clothing bed sheets etc may be combust
22. se Restore 5000 Restorative Innovations Inc 54 8 Trooper Rd Norristown PA 19403 610 630 8400 Part No RES5000 Rev CM 2 13
23. stall a properly grounded 3 prong outlet Failure to use the correct plug and outlet can result in a potential safety hazard CAUTION Ensure that the power cord of the control unit is not pinched or has any objects placed on it and also ensure it is not located where it can be stepped on or tripped over Examine the hospital grade power cord supplied with the control unit 2 Perform one of the following If the plug is damaged Call your supplier for a replacement hospital grade cord If the plug is not damaged Plug the end of the supplied hospital grade power cord into the power outlet on the side of the control unit 3 Plug the other end of the plug into a properly grounded outlet on the wall Once the unit is plugged in an Amber led on the control unit is lit indicating that the system is in stand by mode Restore 5000 19 Restorative Innovations Inc 4 OPERATION 4 5 Using the Front Panel DETAIL MA55 FRONT PANEL Power Fail LED Low Pressure LED Fowler Button Mode Button Max Inflate Low Air Soft Button Loss Firm Button Alarm Silence Button LCD Display Lock Button Static Alternating Pressure Button Power Button FIGURE 5 Using the Front Panel Restorative Innovations Inc 20 Restore 5000 4 OPERATION Power Button l To turn the control unit on or off press and release the Power button Once the unit is plugged in an Amber led on the control unit is lit i
24. y cause injury or death Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment Follow the manufacturer s instructions Monitor patient frequently Read and understand the Owner s Operator s Manual prior to using this equipment ni 150708 Rev Restorative Innovations Inc 6 Restore 5000 2 2 Typical Product Parameters Electrical Parameters INPUT VOLTAGE AC 2 OVERVIEW Restore 5000 90 V INPUT FREQUENCY 60 Hz CURRENT MAXIMUM POWER CONSUMPTION 5 3W CIRCUIT PROTECTION Dual fused 250 V A fast blow fuses MODE OF OPERATION Environmental Parameters Continuous Restore 5000 OPERATING CONDITIONS AMBIENT TEMPERATURE RELATIVE HUMIDITY ATMOSPHERIC PRESSURE 50 104 F 30 75 Non Condensing 70 106 kPa STORAGE AND SHIPPING CONDITIONS AMBIENT TEMPERATURE RELATIVE HUMIDITY ATMOSPHERIC PRESSURE Restore 5000 40 158 F 10 100 Non Condensing 50 106 kPa Restorative Innovations Inc 2 OVERVIEVV Performance Parameters Restore 5000 VVEIGHT CAPACITY PRESSURE ZONE MAXIMUM FLOW MAXIMUM FLOW PRESSURE MAXIMUM FLOW TIMER SUPPORT SURFACE INFLATION TIME 350 Ibs 2 8 4LPM 35 5 mmHg 45 minutes 20 45 minutes PATIENT COMFORT CONTROL PRESSURES SOFT PRESSURE FIRM PRESSURE CYCLE TIME 8 4 mmHg 32 4 mmHg 5 10 15 20 minutes PATIENT CONTACT Mechanical Pa

Download Pdf Manuals

image

Related Search

Related Contents

Casio IQ-02 Owner's Manual  DP-900/DC-901  Einhell TH-AC 200/24 OF  HP LaserJet Pro CM1415fn printer user guide manual Operating  Ice-O-Matic ICE 1405 User's Manual  Pyle PLMCA31BT car kit  取扱説明書 - M  

Copyright © All rights reserved.
Failed to retrieve file