Home
        21209885F
         Contents
1.                                         Figure 7   18  Replacing the Detector Assembly    Flow RTD Use the following procedure to replace the flow resistive temperature    Rep lacement diction  RTD   see Figure 7   19    Equipment Required     Sample flow RTD  Adjustable wrench    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1 Equipment Damage Some internal components can be damaged by small    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Refer to    Removing the Measurement Case Assembly and Lowering  the Partition Panel    in this chapter to lower the partition panel  then  proceed to the next step below     3  Loosen the kynar fittings using the wrench and slide out sample flow    RTD     7 36 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Servicing  Transformer Replacement    4  Unplug the RTD connector from the measurement interface board     flowtemp1        5  Carefully install the new RTD by following the previous steps in  reverse  Match the marked black line on the RTD to be flush with the  fitting     Note The black line is the pre measured depth for the RTD  4    6  Calibrate the instrument  Refer to the    Calibration    chapter in this  manual     Udjustable Wrench    Unplug RTD Connector       Figure 7   19  Replacing the RTD    Transformer Use the following procedure to replace the transformer  Figure 7   20    Replaceme n
2.               Water Collector             Collector Assembly             Figure 5 1  U S  EPA PMio Inlet    Once a month the sampler inlet should be dismantled and cleaned  Mark  each assembly point of the sampler inlet with a pen or pencil to provide  reference marks during reassembly     Disassemble the sample inlet unit according to Figure 5   1  If the assembly  screws are frozen  apply penetrating oil or commercial lubricant to make  removal easier  Clean all interior surfaces and the bug screen with a general  purpose cleaner or compressed air  paying particular attention to small  openings and cracks  Cotton swabs and or a small brush are helpful   Completely dry all components     Also monthly  check the O rings for distortion  cracks  fraying  or other  problems and replace as necessary  Apply small amounts of grease to the  rings before assembling the unit     Reassemble the unit in accordance with the previously scribed match  marks  Particular care should be taken to ensure that all O rings seals are  properly seated and that all screws are uniformly tightened     The same schedule should be used for the European PM  Inlet  e g   Leckel   at approximately a 1 2 month interval  The components of the  inlet should be inspected and cleaned as necessary  Figure 5   2 provided  below provides an image and design drawing of the inlet  with references to  parts     Model 50307 SHARP Instruction Manual 5 3    Preventive Maintenance  Cleaning the Inlets        Nozzle  8   
3.          0 10 20 30 40 50 60 70 80 90 100 110 120 130  Candidate 1  ug m               Figure 42  Results of the parallel measurements with the candidates SN 1   SN 2  measured  component PMpz5  test site Cologne  winter     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  TUVRheinland  Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 148 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F       Model 50301 SHARP  SN 1   SN 2  Bornheim  Summer  Raw data  130  120  110  100   90  80  70  60  50  40  30  20  10   0                  Measured values          Regression line               y x                      Candidate 2  ug m            fa  N     y                                              0 10 20 30 40 50 60 70 80 90 100 110 120 130  Candidate 1  ug m               Figure 43  Results of the parallel measurements with the candidates SN 1   SN 2  measured  component PM25  test site Bornheim  summer        Model 50301 SHARP  SN 1   SN 2  Bornheim  Winter  Raw data                   Measured values          Regression line                     y x                Candidate 2  ug m                                                         O 10 20 30 40 50 60 70 80 90 100 110 120 130    Candidate 1  ug m               Figure 44  Results of the parallel mea
4.        Instrument inputs and outputs          Outputs    Analogue 0 20 mA or 4 20 mA  optional   Digital RS232  RS485  TCP IP  10 status re   lays   Inputs    16 digital inputs   8 analogue inputs  optional        740300_2013_936_21209885F_EN docx    Page 49 of 424       A TUVRheinland      Precisely Right     Page 50 of 424    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F       Protocols    Communication with PC over Ethernet and  iPort software   C Link  MODBUS  Gesytec  ESM  Data  Streaming and NTP    Simultaneous connections from different loca   tions over Ethernet       Status signals   Error messages       Available  for an overview please refer to  Chapter 6 of the instruction manual             740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 51 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    4  Test programme    4 1 General  The suitability test was carried out with two instruments of identical design with the serial 
5.       16 67   000   002   16 74 16 60                                                 18 0 120  17 5 4 J        00                             100  5     17 0 4  80     2  8  a  r  2 aH n    WU Wi 1 oc        SN 1 Flow rate  l min      16 5 MI   A i   Ni N nW lh  ANAM   60          5  Limit of flow rate  e               SN 1 VAC  mmHg   5        SN1 SHARP  g m    O  16 0 4 L4    Q          15 5 4     20       eT  15 0     0  1 7 2011 0 00 2 26 2011 0 00 4 17 2011 0 00 6 6 2011 0 00  Figure 39  Flow of candidate SN 1  field   18 0 120  17 5                                         100  17 0     80              SN 2 Flow rate  l min              Pump vaccuum and SHARP concentration                      2 Fl   agi t Ei a itmi  z 16 5 KA m W ngit N wy j NT dA Yy L 60         5  Limit of flow rate  ir     SN 2 VAC  mmHg          SN2 SHARP  pg m    16 0 4 L 40  15 5   20  15 0 r r 0  1 7 2011 0 00 2 26 2011 0 00 4 17 2011 0 00 6 6 2011 0 00          Figure 40  Flow of candidate SN 2  field     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 141 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 4 8 Tightness of the measuring system    The complete measuring system shall be che
6.       cccccccceceececeeeeeeeeeeeeeeeeeeeeeeeeeeeees 146  Table 35  Overview equivalence test of Model 5030i SHARP for PMa5                 eeeeeee 154  Table 36  Uncertainty between the reference devices Uret for PMo5      ccccceeeeeeeeeeeeeeeeeees 156  Table 37  Summary of the results of the equivalence test  SN 1  amp  SN 2  measured   component Piles  raw data sich niemasnanisnaeisnaiiienanii teal 157  Table 38  Summary of the results of the equivalence check  SN 1  amp  SN 2    after SION GNC ON GGION vee cc ccgess vests cacteatcenansaisassttatbacebenesassute tecenedectasgenvveneceuneeauies 169  Table 39  Stability of standard weight                 cccceceeeee eee eeeeeetee settee ee eeeeenaaeeeeeeeeeeeeeetneee 211  Table 40  Stability of the Control filters           eee cece eee eeecnne eter eee eeeeeenaaeeeeeeeeeeeeeeteee 213  Table 41  Conditioning and weighing timeScales                cccceceeeeeeeeeeeeeeneeeeeeeeeeeeeeenaaees 214    740300_2013_936_21209885F_EN docx    A TUVRheinland      TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Precisely Right     Page 10 of 424    Figures    Figure 1   Figure 2   Figure 3   Figure 4   Figure 5   Figure 6     Figure 7   Figure 8   Figure 9     Figure 10   Figure 11   Figure 12     Figure 13   Figure 14   Figure 15   Figure 16     Figure 17   Figure 18   Figure 19     Figure 20   Figure 21   Figure 22     Figure 23   Figure 24   Figure 25   Figure 26     Figure 27     Figure 28   Figure 29   Figure 30   Figu
7.       to return to the Main Menu or  _ gt _  to return to the  Run screen        Model 50307 SHARP Instruction Manual 3 9    Operation  Range Menu       Range Menu The Range menu allows the operator to select the concentration units  PM    standard ranges  and to set the custom ranges     e Inthe Main Menu  choose Range       DH pikari  PME UY Sad    ihe EE EE          Concentration Units The Concentration Units screen defines how the particles concentration  readings are expressed  Concentration units of micrograms per cubic meter   pg m   and milligrams per cubic meter  mg m   are available     When switching the selected units  the instrument   s analog ranges all  default to the highest range  For example  when switching from mg m  to  ug m   the ranges default to 10000 ug m    Therefore  whenever you change  units  you should also check the range settings     Note If the units change  the instrument should be re calibrated  A display  warning will appear  A    e Inthe Main Menu  choose Range  gt  Conc Units     3 10 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Range Menu        ts             ROR E CHS  EUs EEE TE Saaad Bone a    SHARP Range The SHARP Range screen defines the concentration range of the analog  outputs  For example  a SHARP range of 0 100 pg m  sets the analog  output to concentrations between 0 and 100 pg m  by default     The display shows the current SHARP range  The next line of the display is    used to change the range
8.      Additionally  the Model 5030  proportional radiation detector is sensitive  to natural alpha activity of the aerosol passing through the sampling  system  The additional alpha count rate is used to measure the presence of  Radon gas and infer the daughter nuclides of Radon gas  These daughter  nuclides create a small beta emission  This externally sampled beta emission  is perceived by the detector to be a higher count rate and the higher the  count rate  the less apparent mass detected  Therefore  by measuring the  natural alpha activity due to Radon gas  the proportional beta interference  from the daughter nuclide emissions are subtracted from the count rate to  provide a refined mass measurement  This potential error from Radon gas  interference may be  in lower strength radioactive devices  e g    lt 100 pCi   as high as actual particle concentrations  especially after filter change or  when the natural activity concentration changes  The Model 5030  has  implemented this procedure to determine separately the natural activity in  the aerosol and to eliminate this error  This procedure further allows the  Model 5030  to provide a more stable reading at lower ambient particulate  concentrations compared to similar methods  Furthermore  from this  information the activity concentration of the natural noble gas Radon is  derived     Model 50307 SHARP Instruction Manual 1 7    Introduction  Specifications    Specifications    1 8 Model 50307 SHARP Instruction Manual    T
9.      Figure 52  Reference vs  candidate  SN 2  measured component PM25  Cologne  winter     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system    Model 5030  SHARP with PMz25 pre separator of the company Thermo Fisher    Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 161 of 424                                                                      100    90 4 a SN1 Bornheim Summer    WA  80 4 Z    70   L  D a  2 60   hk  8 50   ra  So f    404 f  a A  30   a A  20   aA A    Ae  10 4 ki    oA                  0 10 20 30 40 50 60 70 80 90 100  PM  5 Reference   yg m     Figure 53  Reference vs  candidate  SN 1  measured component PM25  Bornheim  summer   100  a  90   a SN2 Bornheim Summer   ZL  Fil  a0 4 A  S    70 4 Z  2    2 60   LZ         3 50    2  E a    40   y  S  30 4 ga  Fd  xy  20 4 a4  Ape  a lake  10 4 a  ad  0 10 20 30 40 50 60 70 80 90 100  PM     Reference   yg m     Figure 54  Reference vs  candidate  SN 2  measured component PM2 5  Bornheim  summer     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  TUVRheinland  Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 162 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  F
10.      Table 24   Table 25   Table 26   Table 27   Table 28   Table 29     Table 30     TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Description of the test sites              cece eceeeeeee cnet eeeeeeeeeeeeee eee eetenaaaeaaaaaeeeeeeeenenea    Description of the test site  additional campaign 2013               cccceeeeeeeeeeees  Technical data for Model 5030i SHARP  manufacturer   s specifications            Overview of software versions during suitability testing               eeeeeeees    Field test Sit S ennnen ANE    Ambient conditions at the field test sites  daily mean values               00ceeee  Evaluation of the Teddington  summer  campaign with rH adjusted   To reeuticap teh ca thence ete nceneiseanasaiee E eeteners  Ambient conditions at the additional site    Bornheim  Summer 2013       shown as daily mean values sessccicisierertesiasectisicnercestrssteirertnnmeeunene  Results of the equivalence test at the site Bornheim  Summer 2013    SN 1  58     amp  SN 2  50     component PM2 5  raw data              cccseeeeeeeeeeeees  Results of the equivalence test at the site Bornheim  summer 2013    SN 1  58     amp  SN 2  50     component PM2z5  slope correction 1 068            Results of the equivalence tes
11.     0 048  0 383     0 585 21 87 11 4    Orthogonal Regression Limit Value of 30 ug m3  SN2 Dataset    No s Slope  b      Up Intercept  a      Ug Wom       217 ug m3  Individual Datasets Bornheim Summer 2013   44 0 904 0 934     0 045  0 520     0 549 14 41 11 4       Assessment     1 The expanded uncertainty of the data corrected by the slope of 1 068 lies for SN 1   58    and SN 2  50    beneath the allowed 25      740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH      T  VRheinland      Air Pollution Control   s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 67 of 424   Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    Closing  the corrected data set for Bornheim  Summer 2013 was included in the original  equivalence evaluation from suitability testing  see Table 38  according to point    8 2 Suitabil   ity testing of CEN TS 16450  9  as additional 5th data set  to check whether the criteria for  equivalence test are still fulfilled     Table 11  Results of the equivalence test    suitability testing   Bornheim  summer 2013      SN 1  amp  SN 2  component PMz5  slope correction 1 068    PM2 5 5030i Sharp 25 9  2 17 ug m3 Orthogonal Regression Betw een Instrument Uncertainties  l  Slope  Correctea Slope  b      Up Intercept  a      Ua Reference Candidate    All Data 11 7 0 997   
12.     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUV Rheinland     Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 73 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Filter handling     Mass determination  The following filters were used for suitability testing           Table 14  Used filter materials  Measuring instrument Filter material  type Manufacturer  Reference instrument LVS3 Emfab       47 mm Pall                   The filter material Emfab      Teflon coated glass fibre filters  was used during the    Combined  MCERTS and TUV PM Equivalence Testing    programme by special request of our British  partners  because according to  8   they consider it most appropriate for the measurement  task    The handling of filters meets the requirements of EN 14907     The procedures for filter handling and weighing are described in detail in Appendix 2 of this  report     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 74 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   
13.     Filter  Tape Counter    Possible Cause    Loose reel nuts    Broken tape    Loose photo interrupt  cable    Troubleshooting  Troubleshooting Guides    Action    Tighten reel nuts     Replace tape     Verify connections        Instrument Alarms     Filter  Tape Change    Loose reel nuts    Broken tape    Tape motor nut turning    Tighten reel nuts     Replace tape     Verify connection to measurement  interface board     Replace tape motor        Instrument Alarms     Bench    Cam working not  functioning    Motor turning  cam not    Verify connector to measurement  interface board     Replace motor     Verify setscrews on cam coupling        Instrument Alarms      Motherboard Status    Instrument Alarms      Interface Status    Instrument Alarms     1 0  Exp Status    Internal cables not  connected properly    Board defective    Check that all internal cables are  connected properly  Re cycle AC  power to instrument  If still  alarming  change board     Replace board        Detector Alarms     Alpha    Detector    Settings    Re calibrate detector     Verify alarm settings are between 0  minimum and 100 maximum        Detector Alarms     Beta    Disconnected cable    Defective detector    Settings    Verify cable connections     Replace detector Beta counts  trending towards zero     Verify alarm settings are between  5000 minimum and 20000  maximum        Nephelometer Alarms      Board Status    Disconnected cable    Verify cable harness is connected to  the base    WARN
14.     Gesytec  Bayern Hessen  Protocol  Gesytec Commands    Sample Data Reply String in response to Data Query Command  DA      In response to a valid data query command  the instrument responds in the  following format      lt STX gt MD09 lt SP gt  lt address gt  lt SP gt  lt measured  valuel gt  lt SP gt  lt status gt  lt SP gt  lt SFKT gt  lt SP gt  lt address 1 gt  lt SP gt  lt measured  value2 gt  lt SP  gt  lt status gt  lt SP gt  lt SFKT gt  lt ETX gt  lt BCC gt     The response uses the same command terminators as used by the received  command  i e  if the received command was terminated with a  lt CR gt  then  the response is terminated with  lt CR gt   and if the command was terminated  with a  lt ETX gt  lt BCCs  then the response is terminated with lt ETX gt  and the  computed checksum  lt BCC gt      The 09 after the MD indicates that nine measurements are present in the  reply string  10 for ten measurements and so on  This will also determine  the length of the reply string      lt address gt  is the Instrument Address  Each subsequent measurement  attached to the response will have the  lt address   X gt   where X keeps  incrementing by 1 for each measurement included      lt measured value gt  is the concentration value in currently selected gas units   represented as exponential representation with 4 characters mantissa and 2  characters exponent  each with sign     Mantissa  sign and 4 digits  The decimal point is assumed to be after the  first digit and is n
15.     Manufacturer Thermo Fisher Scientific    Model 5030 SHARP    Type of instrument    Serial No  SN1 SN2    A TUVRheinland      Precisely Right     Page 183 of 424    Measured values from field test sites  related to actual conditions    PM2 5  Measured values in g m   ACT     Page 2 of 11       Ref  2  PM2 5   g m     Ref  1  PM10   pg m     Ref 2   PM10   pg m     Ratio  PM2 5 PM10         SN 1  PM2 5   pg m     SN 2 Remark  PM2 5   ug ms     Test site       4 7 2011  4 8 2011  4 9 2011  4 10 2011  4 11 2011  4 12 2011  4 13 2011  4 14 2011  4 15 2011  4 16 2011  4 17 2011  4 18 2011  4 19 2011  4 20 2011  4 21 2011  4 22 2011  4 23 2011  4 24 2011  4 25 2011  4 26 2011  4 27 2011  4 28 2011  4 29 2011  4 30 2011  5 1 2011  5 2 2011  5 3 2011  5 4 2011  5 5 2011  5 6 2011          23 2  34 9    23 4  31 3  18 0  24 4  32 0  31 9    24 2  34 8    22 3  31 5  17 1  23 2  31 5  31 9       Zero point  Zero point  Zero point    SN1 switched off involuntarily  Recovery after re start    SN1 switched off for short time    Outlier Ref  PM2 5          Cologne  Winter          740300_2013_936_21209885F_EN docx    T  V Rheinland D T  V Rheinland Energie und Umwelt GmbH    Precisely Right     Page 184 of 424    Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F          Annex 5 Mea
16.     Reference Detector Board         SHARP Board Assy Source Board    SHARP Board Bracket Assy    Scatter Detector Board    Optical Bench RH Board    Optical Bench Inlet    Figure 7   3  SHARP Optics Component Layout    7 10 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Servicing  Removing the SHARP Optics Assembly    Removing the SHARP   The SHARP optics assembly can be removed from the measurement case    assembly  Refer to the following steps when a procedure requires lowering  Optics Assembly the partition panel  Figure 7   4            lt    ___     Captive Screws  2     Electrical Connector    Figure 7   4  Removing the SHARP Optics Assembly  Equipment Required     Philips screwdriver    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    f Equipment Damage Some internal components can be damaged by small  1  Turn the instrument OFF and unplug the power cord   2  Ifthe instrument is mounted in a rack  remove it from the rack     3  Disconnect the electrical connector     4  Unfasten the two captive hardware screws     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 11    Servicing  Removing the Measurement Case Assembly and Lowering the Partition Panel    5  Pull the cover in an upward direction     6  Replace the SHARP optics assembly by following the previous steps in    reverse     Removing the The measurement case assembly can be removed and the partiti
17.    100       90 4 4 SN2 Teddington Summer Sam zi             80 4    70 4 7  60 5      o      40    30 4 Vv    20 4     0 io    PM  5 5030i Sharp   pg m   os             0 10 20 30 40 50 60 70 80 90 100  PM  5 Reference   yg m             Figure 58  Reference vs  candidate  SN 2  measured component PM25  Teddington  sum     mer     740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Page 164 of 424    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F                                                                         100  F    Z  90 5 Z  a SN1    Fii  y  80 5 7  x    70 4 xy  E Pi  g 60 4 ff     a   uv f k  S 50 4 r  fe x  3 f    40 4 E  a agta  30   ae  te  20   x 4  a  DEA     ra  o 4  E m    0 10 20 30 40 50 60 70 80 90 100  PM  Reference   yg m     Figure 59  Reference vs  candidate  SN 1  measured component PMz25  values 2 18 g m   100 7       Z  a SN2   ae  80   sf    704 ee  d ri    J  2 60 5 Sf  S 50 iv    S WA  a ye  z 404 z  x  30 4 a  rer a  20   a 2  fi  10    K  0 10 20 30 40 50 60 70 80 90 100  PM  5 Reference   yg m    Figure 60  Reference vs  candidate  SN 2  measured component PM2 5  values 2 18 g m     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH   
18.    2  Remove the cable from the bottom of the radiation shield assembly that  houses the ambient RH temperature assembly     3  Remove the two mounting nuts on the mounting clamp of the  radiation shield assembly  Remove the radiation shield assembly     4  Turn over the radiation shield assembly and remove the three thumb  screws     5  Remove the three spacers and mounting bracket   6  Slide out the ambient RH temperature assembly     7  Slide in the new ambient RH temperature assembly by following the  previous steps in reverse     8  Calibrate the instrument  Refer to the    Calibration    chapter in this  manual     Thermo Fisher Scientific    Servicing  Front Panel Board Replacement    Remove RH Temperature  Assembly                 lt _          Unfasten Thumb Screws  3     Remove Cable    Figure 7   22  Replacing the Ambient RH Temperature Assembly    Front Panel Board     Use the following procedure to replace the front panel board  Figure 7       Replacement       Equipment Required     Front panel board    f Equipment Damage Some internal components can be damaged by small    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Remove the three ribbon cables and the two wire connector from the  front panel board     3  Pop the board off of the two top mounting studs and remove the board  by lifting it up and 
19.    5030i Sharp   pg m   a  oO             0 10 20 30 40 50 60 70 80 90 100  PM  5 Reference   yg m            Figure 27    Comparison of reference and candidate SN 2  measured component PM25  Ted   dington  Summer   relative humidity threshold adjusted to 50      740300_2013_936_21209885F_EN docx    TUVR h 7      TUV Rheinland Energie und Umwelt GmbH   einland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 64 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Since these parameters differ from the ones determined during previous campaigns  the  possible causes for this discrepancy were analysed  In this context  the volatile content in fi   ne dust was reviewed in detail  The content of volatiles in suspended particulate matter had  been determined in measurements carried out in parallel with measuring systems of the type  TEOM FDMS  The results showed that the test site Teddington  Summer  is characterised by  a generally very low level of suspended particulate matter and a significantly high content of  volatile dust components  The average content of volatiles in suspended particulate matter  PM 5 amounted to approx  35   during the campaign carried out in Teddington  summer    compared to a content of approx  20     25   during the three previous campaigns     Based on si
20.    5V       LCD DATA J2    o o N O o A UO N  gt      gt   a    oO    Ground    25V   LCD Bias Voltage  Ground   KEYBOARD J3 1 Keypad Row 1 Input  Keypad Row 2 Input     a a a  A o N          Keypad Row 3 Input  Keypad Row 4 Input  Keypad Col 1 Select  Keypad Col 2 Select  Keypad Col 3 Select    oOo N OQO wo A WO N    Keypad Col 4 Select       LCD BACKLIGHT J4          5V Supply  2 NC  3 Ground    Table 6   6    0 Expansion Board  Optional  Connector Pin Descriptions    Connector Reference Pin Signal Description   Label Designator   EXPANSION I O J1 1 Analog Voltage Input 1  2 Analog Voltage Input 2  3 Analog Voltage Input 3    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 6 21    Troubleshooting  Connector Pin Descriptions    6 22 Model 50307 SHARP Instruction Manual          Connector Reference Pin Signal Description  Label Designator  4 Ground  5 Analog Voltage Input 4  6 Analog Voltage Input 5  7 Analog Voltage Input 6  8 Ground  9 Analog Voltage Input 7  10 Analog Voltage Input 8  11 Ground  12 NC  13 Current Output Return  14 Ground  15 Current Output 1  16 Current Output Return  17 Current Output 2  18 Current Output Return  19 Current Output 3  20 Current Output Return  21 Current Output 4  22 Current Output Return  23 Current Output 5  24 Current Output Return  25 Current Output 6  MOTHER BD J2 1  5V  2  24V  3  24V  4 Ground  5 Ground  6 Ground  7  RS485 to Motherboard  8  RS485 to Motherboard    Thermo Fisher Scientific    Troubleshooting  Connector 
21.    740300_2013_936_21209885F_EN docx    a 3    TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 124 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 3 14 Response time    The response time  rise  of the measuring system shall not exceed 180 s   The response time  fall  of the measuring system shall not exceed 180 s     The difference between the response time  rise  and response time  fall  of the meas   uring system shall not exceed 10   of response time  rise  or 10 s  whatever value is  larger     Note    This test is not relevant for particulate measuring systems   6 2 Equipment   Not applicable    6 3 Performance of test   Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 125 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 3 15 Differ
22.    Field Name  Function   Output Address Hi  Output Address Lo  Output Value Hi  Output Value Lo     Hex   05  00  05  FF  00     Hex   05  00  05  FF  00    Table C 1 through Table C   3 lists the MODBUS addresses supported for    the Model 50301     IMPORTANT NOTE The addresses in the following tables are Protocol  Data Unit  PDU  addresses  Verify the coil number on your MODBUS    master to ensure that it matches the coil number on the instrument  A    Note Coil status 1 indicates active state  A    Table C   1  Read Coils for 5030                      Coil Number Status   0 INVALID   1 LOCAL REMOTE  2 SERVICE   3 NOT USED   4 GEN ALARM   5 CONC ALARM   6 INST ALARMS    Thermo Fisher Scientific    Thermo Fisher Scientific    Coil Number    MODBUS Protocol  MODBUS Parameters Supported    Status       BETA DETECTOR ALARMS       NEPHELOMETER ALARMS       RH TEMP ALARMS       PRES VAC ALARMS                FLOW ALARMS  NOT USED       NOT USED       NOT USED       EXT ALARM 1       EXT ALARM 2           gt  a  gt         a  N lop  ol P ow N    EXT ALARM 3    IMPORTANT NOTE The addresses in the following tables are Protocol  Data Unit  PDU  addresses  Verify the register number on your  MODBUS master to ensure that it matches the register number on the    instrument  A    Note For additional information on how to read registers and interpret the  data  refer to the     0x03 0x04  Read Holding Registers Read Input  Registers    section in this appendix  A    Table C 2  Read Regis
23.    RH    JS sample relative humidity value  9 RH    m1 non filtered mass in ug or mg based on units selected   0 reports    offline    in service mode     online    if NOT in service mode       Thermo Fisher Scientific Model 50302 SHARP Instruction Manual E 1    ESM Protocol Commands  ESM Commands Supported    Command    Description       UA    beta count rate  1 s        U    beta zero count rate  1 s        U    alpha count rate  1 s        alpha zero count rate  1 s        U          B  C  UD  Q  S    U    radon EEC activity concentration  Bq m3     Read neph  coef factor        read device address   instrument ID       flags   need nephelometer alarms    Table E 2  Write Commands for 5030     Command  d4    Description    baud rate       d7    device address   instrument ID       K4    concentration factor       KB    filter change dust load in ug       KD  KG    filter change cycle in hours    external heating control  O off   1 RH control   2 TEMP control       KH    set point air flow regulation in l h       KN    O temp comp on pres comp off  1 temp comp off pres comp off       KS    nephelometer coef factor       KU    establishes the RH target value       KT    establishes the temp target value       K     norm Pressure       K     norm Temp       Y     write parameters to EEPROM    Table E 3  Control Commands for 5030     Command  A    E 2 Model 50307 SHARP Instruction Manual    Description    pump off    Thermo Fisher Scientific    ESM Protocol Commands          E
24.    Table 3   2 lists the standard ranges     e Inthe Main Menu  choose Range  gt  SHARP Range     Pr aT    DN EEE TE Sanat Bone       Table 3 2  Standard Ranges                         mg m  pg m   0 1 100  1 0 1000  2 0 2000  3 0 3000  5 0 5000  10 00 10000  C1 C1   C2 C2   C3 C3    C1  C2  and C3 are custom ranges  For more information about custom  ranges  see    Set Custom Ranges    below     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 11    Operation  Range Menu    Set Custom Ranges    3 12 Model 50307 SHARP Instruction Manual    The analog outputs are arranged on the rear panel connector as shown in  Figure 3   4  See Table 3   3 for channels and pin connections                        ecocooceococ00o00o  lelolololololelolelolelelelelelelels                        Figure 3   4  Pin Out of Rear Panel Connector  Table 3 3  Default Analog Outputs                      Channel Connector Pin 1 0 Terminal Pin Description   1 14 1 PM   2 33 3 Neph   3 15 5 Flow   4 34 7 Ambient Temperature   5 17 9 Flow Temperature   6 36 11 Ambient Relative  Humidity   Ground 16  18  19  35  37 2 4 6 8  10  12 Signal Ground    Note All channels are user definable  If any customization has been made  to the analog output configuration  the default selections may not apply  A    The Set Custom Ranges menu lists three custom ranges  which are user   defined  In the mg m  mode  any value between 0 1 mg m  and 10 00  mg m  can be specified as a range  In the pg m    mode  any val
25.    e The foils must not be wiped  otherwise a loss or gain of mass can occur  thereby biasing the mass calibration     e Foils sets can be returned to Thermo Fisher Scientific for recalibration  as necessary     e Separate foil sets are recommended for periodic QC checks  auditing   and calibration     e Should the user suspect the foil window s  has been damaged   scratched  or coated  these foil sets should be returned to Thermo  Fisher Scientific for re calibration     The null foil has the same approximate mass as a clean filter spot and the  span foil is a calibrated mass increase above the null foil  Therefore  the  foils come in a set and must not be mixed with other sets since a bias in  mass calibration will occur     Since the beta attenuation method is linear with mass increase  the Model  50302 need not be calibrated in the exact range of beta attenuation per filter  spot  What is important  is calibrating the corresponding beta count  reduction with an increase of calibrated mass     Thermo Fisher Scientific    Mass Coefficient    Auto Mass Coefficient    Thermo Fisher Scientific    Calibration  Calibration Procedure    Note To achieve the most accurate mass calibration  the flow temperature  should be as stable as possible and therefore the instrument should have the  heating system turned off  and the instrument should be allowed to  equilibrate to shelter or ambient temperature  If the Model 5030   Beta is  mounted within an ambient shelter  then the most stab
26.    set bkg pm value   This command is used to set PM backgrounds to user defined values to  value  where value is a floating point representation of the background in  current selected units  The following example sets the PM background to  5 2 pg m        Send  set bkg pm 5 2  Receive  set bkg pm 5 2 ok  bkg sharp    This command reports the current SHARP backgrounds  The following  example reports that the SHARP background is 4 pg m        Send  bkg sharp  Receive  bkg sharp 4 000E 00 pg m3  set bkg sharp value    This command is used to set SHARP backgrounds to user defined values to  value  where value is a floating point representation of the background in  current selected units  The following example sets the SHARP background  to 5 2 pg m      Send  set bkg sharp 5 2  Receive  set bkg sharp 5 2 ok    set cal baro pres span   This command automatically calculates the span coefficient based on the  entered target pressure value  The following example sets the calibration  pressure to 720 5 mmHg     Send  set cal baro pres span 720 5  Receive  set cal baro pres span 720 5 ok    set cal flow pres span value   This command automatically sets the flow pressure span based on the flow  pressure entered at the time of calibration  The following example  successfully sets the new value for flow span based on the flow pressure to  50 5 mmHg     Send  set cal flow pres span 50 5  Receive  set cal flow pres span 50 5 ok    set cal flow span value   This command automatically sets t
27.   0 011  0 059     0 189   lt  18 ug m3 12 1 1 040     0 027   0 251     0 267  2 18 ug m3 16 4 1 018     0 030   0 806     0 914    Orthogonal Regression Limit Value of 30 ug m3  Dataset  Nos r  Slope  b      Up Intercept  a      Ua Wem        gt  17 ugm3    Bornheim Winter 1 006     0 025   0 175     0 624  Cologne Winter 1 017     0 028   0 481     0 633  Individual Datasets Bornheim Summer 1 037     0 031     0 437    Teddington Summer 1 043     0 047   0 590     0 381 4 1    Bornheim Summer 2013 1 080     0 048  0 383     0 585 11 4  4 5     lt  18 ug m3 198 0 850 1 114     0 031    0 861     0 306 20 99  Combined Datasets 2 18 ug m3 55 0 951 1 035     0 032   1 069     0 959 17 45 100 0  All Data 253 0 962 1 016     0 013   0 104     0 209 13 58 25 3    Orthogonal Regression Limit Value of 30 ug m3  Dataset  Slope  b      Up Intercept  a      Ug Wem       2 17 ugm3    Bornheim Winter 1 033     0 029   Cologne Winter 0 990     0 025   Individual Datasets Bornheim Summer 0 885     0 028  Teddington Summer 0 947     0 048   Bornheim Summer 2013 0 934     0 045     lt  18 ug m3 207 0 838 1 004     0 028   0 009     0 282 11 34 4 8  Combined Datasets 218 ug m3 54 0 941 1 019     0 034   1 164     1 052 19 41 100 0  All Data 261 0 960 0 985     0 012  0 081     0 203 13 32 24 5       740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambie
28.   11 6 17 2013 12 7 1 7 13 3 23 3 29 7 1004 64 7 0 9 185 0 0  12 6 18 2013 17 6 2 3 13 0 27 2 34 8 1005 61 3 0 4 178 0 0  13 6 19 2013 32 3 4 6 14 3 26 9 35 0 1003 67 8 1 9 244 0 0  14 6 20 2013 9 4 3 0 32 3 20 5 25 1 1003 78 5 1 0 187 34 6  15 6 21 2013 4 3 1 1 24 6 19 0 23 4 1005 69 8 1 6 196 0 3  16 6 22 2013 3 7 1 3 35 0 19 0 23 7 1004 67 8 1 8 198 1 5  17 6 23 2013 3 0 0 4 13 6 16 2 19 2 1005 69 9 1 6 216 0 9  18 6 24 2013 8 4 1 0 12 2 14 2 17 4 1013 76 9 1 8 255 1 5  19 6 25 2013 6 5 0 9 14 6 13 4 16 8 1018 Ti 1 8 259 0 3  20 6 26 2013 9 3 1 9 20 8 13 9 16 7 1018 70 9 14 250 9 8  21 6 27 2013 9 7 2 1 21 1 13 2 17 1 1014 78 5 0 7 230 3 9  22 6 28 2013 8 8 21 24 3 14 1 16 7 1010 86 1 0 3 174 16 4  23 6 29 2013 5 9 0 7 12 4 14 8 18 8 1012 73 9 2 6 269 1 8  24 6 30 2013 7 2 1 3 18 3 t7 22 4 1012 66 4 0 6 198 0 0  25 7 1 2013 EA A 1 4 17 9 18 8 25 4 1008 74 9 0 7 215 21 0  26 7 2 2013 7 9 1 6 20 5 21 6 27 1 1003 62 7 0 6 183 0 3  27 7 3 2013 3 7 0 7 19 0 17 5 20 1 1004 85 6 0 2 213 16 0  28 7 4 2013 Tae 0 4 5 3 20 0 24 7 1014 71 4 0 9 232 0 0  29 7 5 2013 13 0 1 3 10 4 19 8 24 8 1020 74 4 0 3 222 0 0  30 7 6 2013 13 2 2 0 15 4 22 4 29 3 1020 65 4 1 0 191 0 0                                             volatile PM compounds  determined with parallel operated TEOM FDMS system    740300_2013_936_21209885F_EN docx             TUV Rheinland Energie und Umwelt GmbH A TUVR heinland      Luftreinhaltung      Precisely Right     Report on the suitability test of the ambient air quali
29.   7  After calibration  perform a leak test as described in the    Preventive  Maintenance    chapter     Pressure Board Use the following procedure to calibrate the pressure board assembly   Calibration Equipment Required     Manometer    1  From the Main Menu  choose Service  gt  Pres  Vacuum Calibration  gt   Baro Pres  Adjust the barometer pressure span by entering the actual  barometric pressure value     Note If Service Mode is not displayed  refer to    Accessing the Service  Mode    on page 7 4   then return to the beginning of this step  A    2  Return to the Pres Vacuum Calibration  choose Vac Flow  gt  Flow Pres    Span  Calibrate the flow pressure span according to the following     a  Place digital manometer on bench     7 20 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Analog Output Testing    Thermo Fisher Scientific    Servicing  Analog Output Testing    b  Turn on manometer and allow to self calibrate  Do not move  manometer while self calibrating     c  After zeroing the digital manometer  connect the         side of the  manometer to the Delta P A  port on the rear panel     d  Connect the           side of the manometer to the Delta P A  port on  the rear panel     e  Open both Delta P A  and A  toggle valves to open flow through  the manometer     f  Cal Flow Pres Span by entering the actual value read by the  manometer     g  Press to save value     h  Turn off the manometer  this meter does not have an automatic  turn off feature    
30.   7 IOut4 19 Analog_In6  8 Isolated ground 20 Ground   9 lOut5 21 Analog_In7  10 Isolated ground 22 Analog_In8  11 lOut6 23 Ground   12 Isolated ground 24 Ground       Model 50307 SHARP Instruction Manual 2 21    Detail    i    Installation  Startup    Power Connections    A    Startup    2 22 Model 5030  SHARP Instruction Manual    Before the power is connected  be sure that the front panel power switch is  OFF and be sure to verify that the local power matches the label on the  back of the analyzer     WARNING The Model 5030  is supplied with a three wire grounded    power cord  Under no circumstances should this grounding system be    defeated  A    Use the following procedure when starting the instrument     1  Turn the power ON and allow 12 hours for the beta detector to  stabilize     2  Although the instrument has been calibrated at the factory  if the  instrument has been equilibrated  temperature sensors may be checked    Assure that the heater is off and that the instrument is sampling room  air at a constant temperature for at least 1 2 hours with the cover    removed      3  After a 12 hour stabilization  the following steps should be performed     a  From the Main Menu  choose Instrument Controls  Under  Instrument Controls     i     ii     iii     vi     vii     Flow should be set to 16 67 LPM    Heater Control should be set to RH with a RH Threshold of  50 58   For lower RH threshold settings  please contact  technical support     Filter Tape mass limit should 
31.   Calibrate the instrument  Refer to the    Calibration    chapter in this    manual     Detector    3 Pin Data Ref Det Source      RH  A    2 Pin Female  not used     Figure 7   27  SHARP Interface Board Connections    Thermo Fisher Scientific    Servicing  Delrin Nut Replacement    w Loosen screws  ZA       Loosen screws    Figure 7 28  Replacing the SHARP Optics Board    Delrin Nut Use the following procedure to replace the Delrin nut  Figure 7 29    Replacement   Equipment Required   Delrin nut  Philips screwdriver  Set of standard hex drives    Adjustable wrench    amounts of static electricity  A properly grounded antistatic wrist strap    f Equipment Damage Some internal components can be damaged by small    must be worn while handling any internal component  A    1  Turn the Pump Flow OFF  From the Main Menu  choose Instrument  Controls  gt  Set Flow Pump  gt  Pump  Press to toggle and set the  pump flow to OFF     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 49    Servicing  Delrin Nut Replacement    2  Turn the Heater OFF  From the Main Menu  choose Instrument  Controls  gt  Set Heater  gt  Control  Use       until the value reads  OFF  and press         3  Loosen the two hex screws from the collar of the heater adapter cover   4  Loosen the two captive screws of the heater adaptor cover and slide up   5  Loosen the bottom 5 8 inch union nut and remove the tube adapter   6  Slide off the heater tube adapter down    7  Slide off ferrite bead upward
32.   Clean inside parts of pump  change     Clean air passage     Close valves        Check cable connection     Check cable connection        Check cable connection        Model 50307 SHARP Instruction Manual 6 3    Troubleshooting  Troubleshooting Guides    6 4 Model 50307 SHARP Instruction Manual    Malfunction    No SHARP  Concentration    Possible Cause    Disconnected Scat Det cable    Action    Check cable connection        No Zero Air Flow    Rear solenoid valve normally  closed    Clogged in line filter or  internal filter    Pump unplugged    Internal solenoid valve  normally closed    Check zero in bulkhead tubing  connection    Check pump exhaust  connection    Check internal    T    fitting for  plug or tubing not connected    Check rear panel connection     replace valve     Replace filter     Plug in pump     Check cable connection   replace  valve     Kinked tube   insert tube     Insert tube in    Y    fitting or pump     Insert blocking port plug connect  tubing                 Ref Det Voltage Low or high source current Check source current level Adjust  to 62 67 mA  Check min and max values LED  current 65mA   Neph Cal Bad Zero Re zero   Neph Temp High Low Check min and max values   Re calibrate   Neph RH Low High Check min and max values   Re calibrate   Led Current High Low Check min and max values Verify    65mA   5mA    Thermo Fisher Scientific    Thermo Fisher Scientific    Table 6 2  Troubleshooting   Alarm Messages    Alarm Message    Instrument Alarms 
33.   Figure    Board Replacement    Equipment Required     Photo interrupt board  Hex driver  3 32 inch  Philips head screwdriver   1    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    f Equipment Damage Some internal components can be damaged by small    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Disconnect cable     3  Remove mounting screws     7 32 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Servicing  Proportional Valve Replacement    4  Remove board     5  Re install the photo interrupt board by following the previous steps in  reverse     Unfasten screws  2         Unhook Cable                          Figure 7   16  Replacing the Photo Interrupt Board    Pro po rtional Valve     Use the following procedure to replace the proportional valve  Figure 7       Replacement       Equipment Required     Proportional valve assembly    Hex driver  3 32 inch    amounts of static electricity  A properly grounded antistatic wrist strap    f Equipment Damage Some internal components can be damaged by small    must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Refer to    Removing the Measurement Case Assembly and Lowering  the Partition Panel    in this chapter to lower the partition panel  then  proceed to the next step below     3  Disconnect connector from t
34.   Interface Board  Replacement    Thermo Fisher Scientific    Servicing  Measurement Interface Board Replacement    Use the following procedure to replace the measurement interface board     Figure 7 15    Equipment Required   Measurement interface board    Philips screwdriver    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Lower the partition panel  then proceed to the next step below  Refer to     Removing the Measurement Case Assembly and Lowering the  Partition Panel    in this chapter     2  Unplug all connectors from the measurement interface board  Note the  locations of the connectors to facilitate re connection     3  Unscrew the two screws at the top of the measurement interface board   Pop the measurement interface board off from the two bottom  mounting studs and remove the board     4  To install the measurement interface board  follow the previous steps in  reverse     5  Re install the measurement case assembly     6  Calibrate the instrument  Refer to the    Calibration    chapter in this  manual     Model 50307 SHARP Instruction Manual 7 31    Servicing  Photo Interrupt Board Replacement       Mounting Studs  2     Measurement Interface Board    Mounting Screws  2     Figure 7 15  Replacing the Measurement Interface Board    Photo Interru pt Use the following procedure to replace the photo interrupt board
35.   Introduction    The Model 50307 Synchronized Hybrid Ambient Real time Particulate   SHARP  Monitor is a hybrid nephelometric radiometric particulate mass  monitor capable of providing precise  real time measurements with a  superior detection limit  The SHARP Monitor measures the mass  concentration of ambient PMio  PM2 s  and PM    aerosol in real time  The  SHARP Monitor incorporates a high sensitivity light scattering photometer  whose output signal is continuously reference to the time averaged  measurements of an integral beta attenuation mass sensor  as developed on  the Model 5014  Beta  This system achieves heretofore unattained short   term precision and accuracy  The SHARP Monitor incorporates advanced  firmware to optimize the continuous mass calibration of the nephelometric  signal  ensuring that the measured mass concentration remains independent  of changes in the particle population being sampled     The SHARP Monitor incorporates a dynamic heating system  DHS   designed to maintain the relative humidity of the air passing through the  filter tape of the radiometric stage below the user defined threshold   typically below the point at which the collected particles accrete and retain  liquid water  This DHS system minimizes the internal temperature rise by  not heating when sample relative humidity is below the threshold to which  the heater is controlling  As the ambient RH increases above the threshold   the applied heating is optimized to maintain the RH th
36.   Model 50307 SHARP Instruction Manual 7 23    Servicing  Analog Output Calibration    7 24    Analog Output  Calibration    Model 50307 SHARP Instruction Manual                   Input Channel Pin   4 5   5 6   6 7   7 9   8 10   Ground 4 8 11 14    Use the following procedure to calibrate the analog outputs if a meter  reading in the    Analog Output Testing    procedure differed by more than  one percent or after replacing the optional I O expansion board     Equipment Required     Multimeter    Connect a meter to the channel to be adjusted and set to voltage or  current as appropriate  Figure 7   11 shows the analog output pins and  Table 7   4 identifies the associated channels     From the Main Menu  choose Service  gt  Analog Out Cal     The Analog Output Cal screen appears     Note If Service Mode is not displayed  refer to    Accessing the Service  Mode    on page 7 4   then return to the beginning of this step  A      At the Analog Output Cal menu  press  _    to scroll to the desired    voltage channel or current channel corresponding to the rear panel  terminal pin where the meter is connected  then press         With the cursor at Calibrate Zero  press       The Analog Output Cal line displays Zero     Note When calibrating the analog output  always calibrate zero first and  then calibrate full scale  A    5  Use        until the meter reads the value shown in the Set    Output To line  0 0 V  or 0 0 or 4 0 mA   then press to save the    value     Thermo Fisher S
37.   T  VR hei n land     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 165 of 424  Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    6 1 5 4 11 Application of correction factors and terms    If the highest resulting expanded uncertainty of the candidate method is larger than the  expanded relative uncertainty  which is defined in the requirements on the data quality  of ambient air measurements according to EU Guideline  7   the application of correc   tion factors or terms is permitted  The corrected values have to fulfil the requirements  according to point 9 5 3 2 et seqq  of the Guide    Demonstration of Equivalence of Am   bient Air Monitoring Methods        6 2 Equipment  No additional equipment required     6 3 Performance of test  Refer to module 5 4 10   6 4 Evaluation    If evaluation of the raw data according to module 5 4 10 leads to a case where Wem  gt  Wano   which means that the candidate systems is not regarded equivalent to the reference method   it is permitted to apply a correction factor or term resulting from the regression equation ob   tained from the full dataset  The corrected values shall satisfy the requirements for all da   tasets or subsets  refer to module 5 4 10   Moreover  a correction factor may be applied even  for Wem  lt  Wago in order to impro
38.   addr gw 192 168 1 1   set addr gw address    This command sets the default gateway address  where address consists of       9    four numbers ranging from 0 255 inclusive  separated by             Note This command cannot be used when DHCP is on  Refer to the  DHCP command that follows for additional information  A    Send  set addr gw 192 168 1 1   Receive  set addr gw 192 168 1 1 ok   addr ip   This command reports the IP address of the analyzer   Send  addr ip   Receive  addr ip 192 168 1 200   set addr ip address    This command sets the analyzer   s IP address  where address consists of four       9    numbers ranging from 0 255 inclusive  separated by             Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Communications Configuration    Note This command cannot be used when DHCP is on  Refer to the  DHCP command that follows for additional information  A    Send  set addr ip 192 168 1 200  Receive  set addr ip 192 168 1 200 ok  addr nm   This command reports the IP netmask    Send  addr nm   Receive  addr nm 255 255 252 0    set addr nm mask  This command sets the netmask mask  where mask consists of four numbers      9    ranging from 0 255 inclusive  separated by    Note This command cannot be used when DHCP is on  Refer to the  DHCP command that follows for additional information  A    Send  set addr nm 255 255 252 0  Receive  set addr nm 255 255 252 0 ok  addr ntp    This command reports the IP address for the network time pr
39.   alarm flow min 16 00 LPM    set alarm flow min value   set alarm flow max value   These commands set the flow alarm minimum and maximum values to  value  where value is a floating point number representing flow alarm limits  in liters per minute liters per hour  The following example sets the flow  alarm maximum value to 17 34 LPM     Send  set alarm flow max 17 34  Receive  set alarm flow max 17 34 ok       alarm flow pres min   alarm flow pres max   These commands report the current flow pressure alarm minimum and  maximum value settings  The following example reports that the flow  pressure alarm minimum value is  10 0 mmHg     Send  alarm flow pres min  Receive  alarm flow pres min  10 0 mmHg    set alarm flow pres min value   set alarm flow pres max value   These commands set the flow pressure alarm minimum and maximum  values to value  where value is a floating point number representing flow  pressure alarm limits in millimeters of mercury  The following example sets  the flow pressure alarm maximum value to 50 0 mmHg     Model 50307 SHARP Instruction Manual B 19    C Link Protocol Commands    Alarms    B 20    Model 50307 SHARP Instruction Manual    Send  set alarm flow pres max 50  Receive  set alarm flow pres max 50 ok    alarm flow temp min   alarm flow temp max   This command reports the current flow temperature alarm maximum value  setting  The following example reports that the flow temperature alarm  maximum value is 20 0   C     Send  alarm flow temp max  Rec
40.   in the case  of erec   For details on erec formatting  see the    Record Layout Definition     section at the end of this appendix  For details on how to decode the flag  fields within these records  see the  flags  command     Send  erec   Receive  erec  07 53 05 04 09 flags 110000 PM 0 000 1 24Hr Avg PM  0 000 1 Coef A 1 000 Bkg A 0 000 Range A 10000 000 Flow  A 0 000 6 5 3 1 1 1 O Temp Threshold A 30 000 Temp  Threshold A 35 000    list lrec   list srec   list stream   list sp   These commands report the list of current selections for lrec logging data   srec logging data  streaming data output  or the scratch pad  sp  list     The scratch pad is a temporary memory area which is used to set up lists of  selections for lrec  srec  or streaming data items  The user can copy any of  these lists to the scratch pad  modify individual elements in the list  then  save the scratch pad back to the original list  Refer to the    sp field     command for information on how to edit the scratch pad     The following example shows the list for streaming data output     Send  list stream   Receive  list stream  field index variable  X x time   lrec   srec    lrec rec num  srec rec num  lexy rec num  sexy rec num    lrec aa bb 00 pp qq yy  srec aa bb 00 pp qq yy    rec   The starting record index number  1 most recent    num   The number of records to return  1 to 10    x  0 1      Reply termination format  see    set format format     command    y  0 1  2     Output format  see    set 
41.   lay States neeesser a rN E E E R 3 64  Test val Oe OQUtP  tS at oe ie oe enon E R tetas 3 64  Set Analog O  tp  ts sne nueig fea Rak nii iR 3 64  Instrument Configuration      sesesseeeesesesesreessstsssestseseseseseseseseseses 3 65  Contact Infotmati   nes istesin eieiaeo iaaii 3 65  Alatmis M  nussicsenseriennnnu aen iaria 3 66  Instrument Alarmsnicceiira henii RE 3 66    x Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Contents    Filter Tape We tet an  2 ot extents tan tut caraanin tuted sf cc atcade tues e 3 67  Max Filter Tape Counter  iiicccieiaccapapetcetinndibsaneistulecannaiet 3 67  Detector Alaris ostiu r Suskind eh cSschelaep ca a a eee weed 3 67  Min and Max Alpha Alarms j c3 10 scotsestevstesferasedstaragderseeuseeosieuseess 3 68  Minand Max  Beta Alarms nm nanaonan aa 3 68  Wephelometer Alaris Sci cos a eae Sis RA i 3 68  Nephelometer Temperature ci ctsa sce sata cts ae og  3 69  Min and Max Nephelometer Temperature Limits    3 69  Nephelometer RH iscsiticucutiuecpacveknunnuwadiacdaukatiaiys 3 70  Min and Max Nephelometer RH Limits    3 70  LED  Cree nt acdsee ie sen educa cengaads odes nhn a RR N E a RR a 3 70  Min and Max LED Source Current Eimits  cesta se eee  3 71  Reference Detector Voltage         ssesssssssesessersesererrsrrrerrerrereesesreees 3 71  RH Temperature Alarms ccccts2enjoverscistedivbesveruiverietantavisverseceuneteins 3 72  Ampie RA fest rinor eaa at Ea A a AE N 3 72  Min and Max Ambient RH Lim
42.   optional    102645 00 Cable  DB37M to Open End Cable  Six Feet  optional    102646 00 Cable  DB37F to Open End  Six Feet  optional    102659 00 Cable  DB25M to Open End  Six Feet  optional    6279 Cable  RS 232  Six Feet  optional    102888 00 Terminal Board PCB Assembly  DB37F  standard    102891  00 Terminal Board PCB Assembly  DB37M  standard    103084 00 Terminal Board PCB Assembly  DB25M  included with optional 1 0  Expansion Board    105968 00 SHARP PCB Assembly   104100 00 RH Temperature PCB Assembly       7 8 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Servicing  External Device Connection Components    Primary Measurement Head Assy      Tape Radius Guide                         Detector Amp    Bottom Filter Tape Supply Reel    Radius  Spindle Tape Wheel Assy    SHARP Connection Bracket Assy    Inlet  External  Solenoid  Tube Mounting Block Assy        PE n   Counter Wheel Photo  Cam Final Assy o   Bop l Interrupt Board Assy    Cam Photo Interrupt    SHARP Connection Board    Bracket Assy    Pressure Board Proportional Valve    w fitting Assy        ibi Ball ma a  rarr S       1 0 Expansion Board  Opt     Front Panel Digital Output Board    Front Panel Board       Motor w  Photo Motherboard    Measurement Interface Interrupt Coupler Assy    Power Supply Board    Optional  Transformer    Figure 7 2  Model 5030  Component Layout    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 9    Servicing  External Device Connection Components
43.   represented by the two characters    A     amp     B         The checksum is referred to as  lt BCC gt  throughout this document     The following commands are supported by the Gesytec protocol   e Instrument Control Command  ST     e Data Sampling Data Query Command  DA     There is one control command supported by the Gesytec protocol     This  lt control commands is a single letter  which triggers an action in the  instrument  These commands are active only when service mode is inactive  and the zero span option is present     Command    N    initiates a filter change   The following are the different acceptable formats of the ST command    lt STX gt ST lt address gt  lt control command gt  lt ETX gt  lt BCC gt   OR   lt STX gt ST lt address gt  lt control command gt  lt CR gt   OR   lt STX gt ST lt address gt  lt SP gt  lt control command gt  lt CR gt   OR   lt STX gt ST lt address gt  lt SP gt  lt control command gt  lt ETX gt  lt BCC gt     The  lt address gt  is optional  which means it can be left out completely  The   lt address gt   if present  must match the Instrument Address  Additional space  can be present after the  lt address gt      Model 50307 SHARP Instruction Manual D 3    Gesytec  Bayern Hessen  Protocol  Gesytec Commands    If the received command does not satisfy the above formats  or if the   lt address gt  does not match the Instrument Address  the command is  ignored     This is a sample command to switch the instrument to initiate a filter  chang
44.   specific operation and service information     Chapter 1    Introduction    provides an overview of product features   describes the principle of operation  and lists the specifications     Chapter 2    Installation    describes how to unpack  setup  and startup  the instrument     Chapter 3    Operation    describes the front panel display screens  the  front panel pushbuttons  and the menu driven firmware     Chapter 4    Calibration    provides the procedures for calibrating the  analyzer and describes the required equipment     Chapter 5    Preventive Maintenance    provides maintenance  procedures to ensure reliable and consistent instrument operation     Chapter 6    Troubleshooting    presents guidelines for diagnosing  analyzer failures  isolating faults  and includes recommended actions  for restoring proper operation     Chapter 7    Servicing    presents safety alerts for technicians working  on the analyzer  step by step instructions for repairing and replacing  components  and a replacement parts list  It also includes contact  information for product support and technical information     Chapter 8    System Description    describes the function and location  of the system components  provides an overview of the firmware  structure  and includes a description of the system electronics and  input output connections     Chapter 9    Optional Equipment    describes the optional equipment  that can be used with this analyzer     Chapter 10    Model 5014i to 5030i U
45.   the measuring system resumes operation automati   cally  However  at least the first 12 h worth of data after voltage returns shall be discarded   since the system needs an equilibration phase after restart until the BETA detector adjusts to  ambient conditions  see point 6 1 4 1 4 Set up times and warm up times      6 5 Assessment    All instrument parameters are protected against loss through buffering  After the power sup   ply has returned  the measuring device goes back to failure free operational condition and  resumes measurement  However  at least the first 12 h worth of data after voltage returns  shall be discarded  since the system needs an equilibration phase after restart until the BETA  detector adjusts to ambient conditions     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 92 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    6 1 5 2 5 Operating states    The measuring system shall allow the control of important operating states by telemet   rically transmitted status signals     6 2 Equipment  PC for data acquisition   6 3 Performance of test   
46.   the user can retrieve measurement data and instrument information     change parameters and perform tests for controlling the functionality of the measuring sys   tem     Mg 7  m3    NEP SMFL Lae  to  RANGERRRAUGEEED I AGSEALARM    Figure 14  Main screen of the user interface       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 43 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    The main screen is on the top level of the user interface  It usually displays the SHARP con   centration  By scrolling down with the arrow buttons f   the screens for PM concentration     BETA   NEPH concentration  ambient air conditions  flow conditions and mass sensor data  are displayed     In addition  the main screen displays a status bar with the current time as well as different  status messages  e g  the current mode the nephelometer is set to   In some cases  the fol   lowing symbols can be observed     Status Bar Icons    The password  lock  icon indicates that Theal bell  icon indicat  no parameter changes can be made from        gt  G    he         he alarm   en   icon indicates  the front panel that an alarm is active     The service  wrench  icon indicates that the  instrument is in the servi
47.   used along with the board level connection diagrams to troubleshoot  board level faults        Motherboard Connector Pin Descriptions    on page 6 10     Measurement Interface Board Connector Pin Descriptions    on page 6 15     Front Panel Board Connector Pin Diagram    on page 6 20       I O Expansion Board  Optional  Connector Pin Descriptions    on page 6   21       Digital Output Board Connector Pin Descriptions    on page 6 23     SHARP Measurement Interface Board Pin Connections    on page 6 24    Table 6 3  Motherboard Connector Pin Descriptions    Connector Reference Pin Signal Description  Label Designator    INTF DATA J1 1 Ground   RS485 to Interface Board   RS485 to Interface Board    w nN       10 BASE T J2           Ethernet Output      Ethernet Output      Ethernet Input      NC   NC    on ee WO N    Thermo Fisher Scientific    Connector Reference Pin  Label Designator    co N Om    Troubleshooting  Connector Pin Descriptions    Signal Description    Ethernet Input      NC  NC       EXPANSIONBD J3    ba    oOo N oon A Wo N     5V    24V    24V   Ground   Ground   Ground    RS485 to Expansion Board   RS485 to Expansion Board       INTERFACE 24V J4    N       24V    Ground       1 0 J5    o o N own A OO N  gt        a    oO        gt    x  a    a      co co N o ol A ow N    N N    O    Thermo Fisher Scientific    Power Fail Relay N C  Contact  Digital Ground   TTL Input 1   TTL Input 2   Digital Ground   TTL Input 5   TTL Input 7   TTL Input 8   TTL Input 10   
48.  0  1 207 0 3  97 8 28 2011 12 15 6 19 7 1009 69 2 1 3 212 0 0  98 8 29 2011 6 1 0 9 15 0 14 5 18 4 1008 66 7 20 243 0 0  99 8 30 2011 8 6 1 5 17 3 13 6 18 4 1008 73 6 0 8 236 0 0  100 8 31 2011 14 2 2 9 20 7 14 8 21 4 1007 72 0 0 7 225 0 0  101 9 1 2011 17 9 4 3 23 8 16 4 23 4 1006 71 6 0 6 182 0 0  102 9 2 2011 14 9 22 14 5 21 2 29 4 1004 72 2 0 8 160 0 0  103 9 3 2011 23 24 5 30 9 1002 67 0 1 3 132 3 6  104 9 4 2011 8 1 1 0 12 3 20 2 27 4 1002 79 5 11 223 0 6  105 9 5 2011 4 9 0 4 8 3 16 6 21 4 1009 62 9 1 9 217 0 0  106 9 6 2011 5 4 Ae 32 2 17 4 20 6 1005 66 8 2 6 219 4 8  107 9 7 2011 5 9 1 3 21 4 14 9 18 1 1004 73 1 2 2 246 5 7  108 9 8 2011 1 0 14 7 16 2 1003 84 7 11 209 3 3  109 9 9 2011 6 9 1 0 13 9 19 0 21 6 1004 86 9 0 4 167 0 0  110 9 10 2011 ara 23 8 29 7 1001 73 0 1 5 155 0 0  111 9 11 2011 5 6 1 4 24 9 16 2 22 1 1003 86 0 0 7 165 16 2  112 9 12 2011 5 6 1 0 18 0 19 4 24 6 1004 71 1 1 7 204 0 0  113 9 13 2011 6 7 0 9 12 7 16 7 20 8 1006 67 3 1 6 219 0 0  114 9 14 2011 6 9 1 0 14 8 15 2 19 6 1011 65 1 1 5 224 0 0  115 9 15 2011 11 7 1 5 12 8 14 1 20 4 1013 75 3 0 6 207 0 0  116 9 16 2011 13 3 17 1 21 9 1006 72 6 1 4 145 0 0  117 9 17 2011 16 8 21 4 1001 70 6 1 0 207 3 6  118 9 18 2011 3 5 13 3 16 5 998 76 4 1 0 200 4 5  119 9 19 2011 8 0 13 6 18 1 1008 75 8 1 4 231 0 9  120 9 20 2011 11 15 6 18 5 1014 78 0 0 5 196 0 0                                                         volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_9
49.  1 000  2 000  3 000  5 000  10 000 g m        Minimum Detection Limit   lt 0 5 pg m   1 hour    2 o   lt 0 2 ug m   24 hour    2 o       Resolution  0 1 ug m3  not to be confused with precision        Precision   2 0 g m3  lt 80 g m3   5 g m3  gt 80 g m3  24 hour        Accuracy for Mass Measurement   5  using NIST traceable mass foil set  Data Averaging and Output    Real time Mass Conc Average  1 minute       Long Term Averaging  60 to 3 600 sec and 24 hour       Data Output Rate  every 1 second  Operating Range    The temperature of the sampled air may vary between  30 and 50   C  The 50307 units must be  weather protected within the range of 4   C to 50   C     An optional Complete Outdoor Enclosure provides complete weather protection        Non condensing   lt 95  RH inside 5030   Sample Flow    Volumetric flow control system uses a calibrated subsonic precision orifice and the measured  ambient temperature and pressure to maintain constant volumetric flow rates     Nominal Flow Rate  16 67 L min   Flow Precision  2  of measured value  Flow Accuracy  lt 5  of measured value  Data Storage    Internal data logging of user specified variables  capacity of 190 000 records     Model 50307 SHARP Instruction Manual 1 9    Introduction  Specifications    1 10 Model 50307 SHARP Instruction Manual    Filter Media   Sample Filter  Glass fiber filter type  width 40 mm  length  17 m  Whatman  Inc   Approximately 650 filter tape changes per roll    Inlet Accessories   European or U 
50.  1 1006 52 2 0 9 167 0 0  47 7 23 2013 13 8 1 8 12 8 25 5 33 7 1004 62 0 0 6 159 0 0  48 7 24 2013 20 1 5 0 24 9 21 7 27 6 1006 78 7 0 7 213 3 6  49 7 25 2013 11 6 4 5 38 4 22 5 28 1 1006 81 7 0 7 145 15 1                                                   volatile PM compounds  determined with parallel operated TEOM FDMS system    740300_2013_936_21209885F_EN docx    TUVR heinland   TUV Rheinland Energie und Umwelt GmbH  Precisely Right Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system   Page 208 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Appendix 2    Filter weighing procedure  A  German test sites  Cologne and Bornheim     A 1 Carrying out the weighing   All weightings are done in an air conditioned weighing room  Ambient conditions are 20   C   1   C and 50    5   relative humidity  which conforms to the requirements of Standard EN  14907     The filters used in the field test are weighed manually  In order to condition the filters  includ   ing control filters   they are placed on sieves to avoid overlap    The specifications for pre  and post weighing are specified beforehand and conform to the  Standard                          Before sampling   pre weighing After sampling   post weighing  Conditioning 48 h 2h Conditioning 48 h  2h  Filter weighing Filter weighing  Re conditioning 24 h  2 h Re c
51.  18 2012   11 18 2012  11 19 2012   11 19 2012  11 20 2012   11 20 2012  11 21 2012   11 21 2012  11 22 2012   11 22 2012  11 23 2012   11 23 2012  11 24 2012   11 24 2012  11 25 2012   11 25 2012  11 26 2012   11 26 2012  11 27 2012 11 27 2012    1  2  3  4  5  6  7  8                740300_2013_936_21209885F_EN docx    T  V Rheinland  TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung  Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 178 of 424    Annex 2 Dependence of zero point on ambient temperature  SHARP  Page 1 of 2    Manufacturer Thermo Fisher Scientifc    Standards Zero filter  Type Model 50301 SHARP    Serial No  SN1 SN2    KIME E AE e a     C   ug m    ug m    g m   ug m    ug m    ug m               740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVR heinland S  Luftreinhaltung    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5  Page 179 of 424  Report No   936 21209885 F    Annex 2 Dependence of zero point on ambient temperature  NEPH  Page 2 of 2  Manufacturer Thermo Fisher Scientifc  Standards Zero filter    Type Model 50301 SHA
52.  2012             Figure 21    Course of PM25 concentrations  reference  at the test site    Bornheim  motorway  parking lot  winter                            70  a 60  E        50       gt   S 40  a     9  a 30     a  o 20   lt   fo   O  10  0  o a a a o a a o o a a      E D           D    t             t                     o  A A X A A A X A A A X   N  N              e D D o     S S 3 S 3    3 3 3 3    N 15 al o    ite  ro    5 N Te   a A o     A ro      N                Figure 22  Course of PM2 5 concentrations  reference  at the test site    Teddington  summer       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland 9    Air Pollution Control i i  ir Pollution Contro Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM25pre separator of the company Thermo Fisher Page 59 of 424  Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The following figures show the measuring cabinet at the field test sites Cologne  parking lot    Bornheim  motorway parking lot  and Teddington     Figure 23           Figure 24  Field test site Bornheim  motorway parking lot    740300_2013_936_21209885F_EN docx    es x D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     9S Air Pollut    Air Pollution Control  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 60 of 424 Model 5030  SHARP 
53.  24 h may  not exceed 1   of this averag   ing time     Testing was carried out in compliance  with the minimum requirements stated  in Table 5 of Standard VDI 4202  Part    1  September 2010      The calculated mass is related to a  defined and actively regulated sam   pling volume  thus determining the  particulate mass concentration     Not applicable to PMz5 sampling sys   tems  Reference is made to module  5 4 10 of this report     Not applicable to PMz5 sampling sys   tems  Reference is made to module  5 4 10 of this report     A statistical correlation between the  reference measuring method and the  device reading could be demonstrat   ed     No deviation of the measured signal  from the nominal value  gt  0 5 g m   caused by interference of moisture in  the sample could be observed for  PM25  No negative influence on the  measured values at varying relative  humidity was detected during the field  test  Comparability of the candidates  with the reference method according  to Guide  Demonstration of Equiva   lence of Ambient Air Monitoring  Methods     4  is also given for days  with a relative humidity  gt  70       It is possible to form daily averages  based on the three measurement cy   cles using the described instrument  configuration and a cycle duration of 8  hours     740300_2013_936_21209885F_EN docx       m i    TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air qua
54.  3  Return to the Pres Vacuum Calibration  choose Vac Flow  gt  Vac Pres  Span  Calibrate the vacuum pressure span according to the following     a  Use an acceptable vacuum manometer  If using a blue Druck  manometer  place blue manometer on bench     b  Turn on manometer and let manometer self calibrate  Do not  move manometer while self calibrating     c  Connect the         side of the manometer to the Delta P A  port on  the rear panel     d  Open Delta P A  toggle vales to open flow through the  manometer     e  Cal Vac Pressure Span by entering the actual value from the  reference manometer     f  Press to save value     The analog outputs should be tested if the concentration value on the front  panel display disagrees with the analog outputs  To check the analog  outputs  connect a meter to an analog output channel  voltage or current   and compare the meter reading with the output value set on the Test  Analog Outputs screen     Equipment Required   Multimeter    Use the following procedure to test the analog outputs     Model 50307 SHARP Instruction Manual 7 21    Servicing  Analog Output Testing    1  Connect a meter to the channel to be tested  Figure 7   11 shows the  analog output pins and Table 7   4 identifies the associated channels     2  From the Main Menu  choose Diagnostics  gt  Test Analog Outputs   The Test Analog Outputs screen appears     3  Press to scroll to the desired channel corresponding to the rear  P 8  panel terminal pins where the meter is 
55.  3 SN1 defective  8 18 2012 6 9 SN1 defective  8 19 2012 7 8 SN1 defective  8 20 2012 9 1 SN1 defective  8 21 2012 6 5 SN1 defective  8 22 2012 5 7 SN1 defective  8 23 2012 4 5 SN1 defective  8 24 2012 3 5 SN1 defective  8 25 2012 3 9 SN1 defective  8 26 2012 5 4 SN1 defective  8 27 2012 3 2 SN1 defective  8 28 2012 6 4 SN1 defective  8 29 2012 4 1 SN1 defective                   740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVR heinland      Luftreinhaltung   i  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 191 of 424  Report No   936 21209885 F    Annex 5 Measured values from field test sites  related to actual conditions Page 10 of 11    Manufacturer Thermo Fisher Scientific    PM2 5  Type of instrument Model 50301 SHARP Measured values in g m   ACT     Serial No  SN1 SN2          Ref  1 Ref  2 Ref  1 Ref 2  Ratio SN 1 SN2 Remark Test site  PM2 5 PM2 5 PM10 PM10 PM2 5 PM10 PM2 5 PM2 5   ug m    ug m   g m   ug m       ug m   ug m   8 30 2012   SN 1 defective Teddington  Summer  8 31 2012   SN1 defective  9 1 2012   SN1 defective  9 2 2012   SN1 defective  9 3 2012 X   T SN1 defective  9 4 2012 t     SN1 defective  9 5 2012   SN1 defective  9 6 2012    Stop PM10 Reference  9 7 2012 Inlet   gt  Zero filter  9 8 2012 Zero point  9 9 2012 Zero point  9 
56.  32   e    Proportional Valve Replacement    on page 7 33   e    Detector Assembly Replacement    on page 7 34   e    Flow RTD Replacement    on page 7 36   e    Transformer Replacement    on page 7 37   e    Radius Tube Replacement    on page 7 38   e    Ambient RH Temperature Replacement    on page 7 39   e    Front Panel Board Replacement    on page 7 41   e    LCD Module Replacement    on page 7 42   e    Removing the SHARP Optics Cover    on page 7 44   e    External Solenoid Valve Replacement    on page 7 45   e    SHARP RH Temp Sensor Replacement    on page 7 46   e    SHARP Interface Board Replacement    on page 7 47   e    Delrin Nut Replacement    on page 7 49   e    Internal SHARP Cable Assembly Replacement    on page 7 51  e    Case Cable Replacement    on page 7 52   e    External Zeroing Filter    on page 7 53    e    Service Locations    on page 7 54    7 2 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Safety Precautions    Thermo Fisher Scientific    A    Servicing  Safety Precautions    Read the safety precautions before beginning any procedures in this  chapter     WARNING The service procedures in this manual are restricted to qualified  service representatives  A    If the equipment is operated in a manner not specified by the  manufacturer  the protection provided by the equipment may be  impaired  A    The detector assembly should be replaced by a qualified technician  knowledgeable in dealing with radiation precautions  By removing th
57.  4 3 1 14 4  0 9 3 9 1020 81 8 0 6 138 0 3  176 1 17 2012 33 0 3 9 11 8  0 6 4 7 1022 81 5 0 5 123 0 3  177 1 18 2012 19 2 4 7 24 7 3 1 ell 1018 84 3 1 5 126 6 6  178 1 19 2012 3 5 2 6 73 2 7 2 11 9 1006 84 8 4 1 253 12 8  179 1 20 2012 6 1 2 6 42 0 3 1 5 4 1008 85 5 3 1 239 5 1  180 1 21 2012 2 3 7 3 10 3 1000 81 9 5 3 261 6 9                                                volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx                TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Luftreinhaltung      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 199 of 424  Report No   936 21209885 F                                                                Annex 6 Ambient conditions at field test sites Page 7 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ug m   ug ms          C     C   hPa       m s   4  mm   181 1 22 2012 Bornheim  Winter 5 2 0 8 14 5 6 4 9 3 1003 77 0 4 7 271 1 8  182 1 23 2012 7 4 1 3 17 6 4 2 Tea 1008 83 7 3 0 268 3 6  183 1 24 2012 2T 2 3 5 5 1014 87 4 0 9 188 0 0  184 1 25 2012 19 9 5 0 25 1 2 5 4 2 1012 81 0 3 9 126 0 0  185 1 26 20
58.  4 8  Combined Datasets 218 ug m3 51 0 947 1 090     0 036   1 159     1 101 21 88 100 0  All Data 217 0 962 1 055     0 014  0 066     0 241 18 34 27 2       740300_2013_936_21209885F_EN docx    g   a TUV Rheinland Energie und Umwelt GmbH  TUVRheinland  Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 158 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F             Bornheim Winter           a Cologne Winter           Bornheim Summer        e Teddington Summer        iseecscenad  lt  18 ug m 3 Line              2 18 pg m 3 Line    PM    5030i Sharp   yg m        All Data Line                0 10 20 30 40 50 60 70 80 90 100  PM    Reference   ug m           Figure 48  Reference vs  candidate  SN 1  amp  SN 2  measured component PMps  all test sites    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 159 of 424          PM    5030i Sharp   pg m     100    90      80      70 4    60 5          a SN1                         50 60 70 80 90 100  PM  5
59.  4 bytes     There is an optional single digit d which may follow any of the numeric  fields  which indicates that after the field has been parsed out  the resulting  value is to be divided by 104d  Thus the 16 bit field OxFFC6 would be  interpreted with the format specifier  n3  as the number  0 058     The subsequent lines in the erec layout response describe the appearance of  the full panel  The full instrument panel as it appears on the screen has two  columns of lines  Each line is composed of three major components   1  a  text field   2  a value field  and  3  a button  None of these three  components is required  The text field contains statically displayed text     The value field displays values which are parsed out of the response to a  erec command  It also displays  though background changes  alarm status   The button  when pressed  triggers input from either a dialog box or a  selection list  There are five kinds of buttons  B  I  L  T  and N     Each line in the layout string corresponds to one line on the display  The  layout string describes each of the three major fields as well as translation  mechanisms and corresponding commands     The first field in the layout string is the text  It is delimited by a      The  string up to the first     will be read and inserted in the text field of the line     This is followed by a possible string enclosed in quotes that is used to place  a string into the value field     The value source  which is the item  or word 
60.  5    Report No   936 21209885 F    6 4 Evaluation    All instrument functions listed in the manual are available and can be activated  The current  system status is monitored continuously and problems are indicated by a series of different  warning messages     An external check of the zero point using a zero filter is possible at any time  A check of the  radiometric measurement using the    mass transfer standard    foil set can also be performed  anytime    It is not possible to produce a reproducible signal for the sensitivity of the nephelometer us   ing external test standards  However  the nephelometer is regularly flushed with zero air dur   ing operation  auto zero function as part of every filter change   thus determining and docu   menting the position of the zero point on a regular basis  Information on the current and pre   vious NEPH backgrounds can be retrieved from the instrument at all times  Since the nephe   lometer is continuously calibrated with a very stable radiometric measurement  a separate  examination of the stability of the nephelometer   s signal under operating conditions can be  omitted     6 5 Assessment    All system functions listed in the manual are available  can be activated and function well   The current system status is continuously monitored and displayed by a series of status  messages     The results of the external zero checks performed with a zero filter during the field tests as  well as the results of the regular mass coefficien
61.  6 0 8 7 10 9 1015 82 3 1 2 161 0 3  223 3 4 2012 30 8 9 7 31 4 8 6 11 5 1008 84 6 2 6 129 10 2  224 3 5 2012 5 4 3 6 67 6 5 6 7 8 1010 70 7 4 6 162 0 0  225 3 6 2012 14 0 3 8 27 3 4 8 10 3 1018 72 5 1 6 173 0 0  226 3 7 2012 15 0 5 4 36 4 5 2 7 8 1011 76 5 3 8 218 9 8  227 7 17 2012 Teddington  Summer 4 3 3 69 0 19 2 25 2 1001 76 1 0 3 213 0 5  228 7 18 2012 4 0 2 0 50 6 16 4 19 6 991 83 8 0 4 212 1 3  229 7 19 2012 4 6 1 6 34 3 16 4 21 2 994 TF 0 2 270 0 0  230 7 20 2012 5 0 0 8 14 9 16 2 20 0 1001 69 7 0 7 221 0 3  231 7 21 2012 6 0 1 3 22 4 16 9 22 8 1007 64 9 0 3 189 0 0  232 7 22 2012 6 9 1 5 21 0 18 6 23 8 1007 61 9 0 6 203 0 0  234 7 23 2012 8 6 2 0 22 7 21 4 29 4 1002 57 0 0 4 197 0 0  234 7 24 2012 11 4 2 0 17 8 22 8 30 8 997 56 3 0 3 213 0 0  235 7 25 2012 18 1 29 16 1 24 2 32 3 998 47 1 1 0 117 0 0  236 7 26 2012 21 2 4 6 21 8 21 8 28 3 996 62 0 Aci 125 0 0  237 7 27 2012 12 0 5 4 44 7 19 9 25 1 992 60 4 0 9 256 0 0  238 7 28 2012 1 8 17 1 23 8 992 55 5 0 3 236 0 0  239 7 29 2012 1 3 13 6 19 9 995 78 7 0 2 198 61  240 7 30 2012 49 ia 23 9 16 3 22 0 997 66 3 0 3 195 4 3                                                            volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx             TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Luftreinhaltung      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separa
62.  6 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ug ms   Lg m3          C     C   hPa       m s   q  mm   151 10 21 2011 Bornheim  Summer 17 8 4 2 23 6 5 2 11 4 1019 79 3 41 154 0 0  152 10 22 2011 5 4 6 8 11 9 1013 69 3 3 9 128 0 0  153 10 23 2011 23 1 6 1 26 3 7 4 13 6 1007 71 6 2 3 138 0 0  154 10 24 2011 15 4 4 0 25 7 9 8 12 4 999 67 2 3 8 132 0 0  155 10 25 2011 11 6 2 0 17 0 10 9 13 0 997 68 8 1 9 132 0 0  156 10 26 2011 6 3 1 6 24 5 9 4 15 7 1006 74 9 0 9 171 0 0  157 10 27 2011 18 7 3 5 18 9 10 1 13 4 1008 80 4 27 126 0 0  158 10 28 2011 17 9 3 2 18 0 14 0 20 1 1016 75 8 0 7 149 0 0  159 10 29 2011 3 1 13 5 17 2 1015 79 8 1 0 132 0 9  160 10 30 2011 11 9 2 6 22 1 13 5 18 2 1014 86 7 0 4 170 0 0  161 10 31 2011 4 4 11 9 Ta 1009 87 0 1 0 152 0 3  162 11 1 2011 22 4 4 8 21 4 12 2 15 6 1007 83 2 2 0 134 1 8  163 11 2 2011 17 2 5 3 30 5 12 5 14 7 1002 80 8 29 124 0 3  164 11 3 2011 10 0 3 2 32 2 13 7 17 3 995 69 7 3 8 140 0 0  165 11 4 2011 14 0 2 2 15 4  166 11 5 2011 1 6  167 11 6 2011 55 0 6 4 11 6  168 11 7 2011 34 4 4 8 13 9 P  169 11 8 2014 43 7 56 128 Outage of weather station  170 11 9 2011 23 7 5 4 22 8  171 11 10 2011 14 1 4 5 32 1  172 11 11 2011 27 4 4 5 16 3  173 11 12 2011 6 7 LZ 9 4 1024 81 6 4 1 135 0 0  174 11 13 2011 21 5 6 0 28 0 5 4 10 1 1023 83 7 2 4 131 0 0  175 1 16 2012 Bornheim  Winter 21
63.  62  0 30                      The graphical representation of the flow rates at the test site used as example  Cologne  win   ter  nominal  16 67 l min  shows that all values determined during sampling deviate less than   5   from the respective nominal value  The deviation of the 24 h averages for the total flow  rate of 16 67 l min is as well significantly smaller than the specified  3   of the nominal val   ue    All determined daily averages present deviations less than  3   from the nominal value  all  instantaneous values deviate less than  5       Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results    Table 31 and Table 32 show the determined flow parameters  Figure 39 and Figure 40 illus   trate the flow measurements for both candidates SN 1 and SN 2     740300_2013_936_21209885F_EN docx         T  V Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system  Page 140 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Table 31  Parameters for flow measurement  24 h average   SN 1  field     No  24h values Dev  from nominal       Std Dev   l min       16 67   000   002   17 20 16 56    Table 32  Parameters for flow measurement  24 h average   SN 2  field     No  24h values Dev  from nominal       Std  Dev   l min 
64.  7 19 2013 16 3 14 8  7 20 2013 17 3 16 0  7 21 2013 18 0 17 7  7 22 2013 23 2 22 0  7 23 2013 25 4 24 5  7 24 2013 31 5 30 3  7 25 2013 21 3 20 3                                     740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     Page 206 of 424    TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F                         Annex 8 Ambient conditions at the additional test site Bornheim  Summer 2013 Page 1 of 2   No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5   Mean ambient temperature   Max  ambient temperature  Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   pg m   pg m         C     C   hPa       m s   4  mm   1 6 7 2013 Bornheim 16 6 21 7 29 1 1010 55 5 1 0 Bii 0 0  2 6 8 2013 21 4 26 8 1005 62 3 21 243 0 0  3 6 9 2013 13 8 15 6 19 2 1001 78 7 1 8 273 4 5  4 6 10 2013 15 7 14 4 18 1 1005 75 9 1 2 253 0 6  5 6 11 2013 12 6 3 2 25 4 18 8 23 8 1008 61 5 0 6 198 0 0  6 6 12 2013 6 8 1 3 19 5 21 1 23 7 1008 67 1 1 0 181 0 0  7 6 13 2013 5 5 0 3 4 7 17 0 27 6 1007 77 9 1 3 209 22 5  8 6 14 2013 5 3 1 1 21 4 16 1 21 2 1009 65 4 0 6 181 0 0  9 6 15 2013 5 2 0 4 6 9 17 2 22 6 1005 63 1 1 4 209 0 0  10 6 16 2013 7 5 0 6 8 6 17 7 23 7 1007 63 9 0 7 226 0 0
65.  8 6 4  312 10 10 2012 13 8 13 9 12 8 12 6  313 10 11 2012 6 3 6 4 5 6 Te             740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Luftreinhaltung      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 193 of 424  Report No   936 21209885 F                      Annex 6 Ambient conditions at field test sites Page 1 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   pg m   pg m         C     C   hPa       m s  g  mm   f 3 8 2011 Cologne  Winter 31 4 6 4 20 5 7 9 14 4 1013 54 0 2 3 147 0 0  2 3 9 2011 18 9 5 2 27 7 7 1 10 6 1010 75 8 3 5 232 0 9  3 3 10 2011 9 2 10 1 1008 68 9 5 3 231 0 0  4 3 11 2011 16 4 3 1 19 0 8 1 11 4 1008 69 5 3 8 197 0 3  5 3 12 2011 3 5 12 1 16 4 998 61 6 3 3 147 0 3  6 3 13 2011 13 2 2 0 15 2 tie 14 1 1001 77 3 2 0 156 1 5  7 3 14 2011 19 1 9 8 13 8 1010 81 2 0 3 114 0 0  8 3 15 2011 37 6 7 2 19 2 12 3 19 6 1006 66 2 22 96 0 0  9 3 16 2011 56 4 9 3 16 6 9 5 16 9 1000 71 9 2 5 126 0 0  10 3 17 2011 50 1 8 8 17 6 5 7 7 2 1009 86 9 4 7 267 0 0  11 3 18 2011 28 2 10 2 36 3 6 0 7 6 1018 89 1 ist 135 11 1  12 3 19 2011 5 8 5 0 12 1 1027 59 5 1 2 123 0 0  13 3 20 
66.  A    e Inthe Main Menu  choose Service  gt  Detector Calibration        Auto Detector Calibration The Auto Detector Calibration screen allows the user to optimize the    detector performance  This should be done at least once per year and as  often as once per quarter     e Inthe Main Menu  choose Service  gt  Detector Calibration  gt  Auto        This procedure will take approximately 30 minutes until completed   Thereafter  a high voltage  HV   beta reference threshold and alpha    threshold will be assigned a new valve     4 12 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Nephelometer  Calibration    Nephelometer RH    Nephelometer  Temperature    Calibration  Calibration Procedure    The Nephelometer Calibration submenu is used to view and set the  nephelometer calibration  The nephelometer calibration screen is visible  only when the instrument is in service mode  For more information on the  service mode  see    Service Mode    earlier in the chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Nephelometer Calibration        aa BUES Pm Tit  BAPE at Lid    DN EEE TE Sanad Bone       The Nephelometer RH screen allows the user to view and calibrate the  nephelometer relative humidity     Note Calibration of the nephelometer RH must be done with the heater in  the    OFF    position  A    e Inthe Main Menu  choose Service  gt  Neph Calibra
67.  A PC was connected locally via Ethernet to the AMS and the transfer of data including sys   tem status was tested     Furthermore  it is possible to monitor and operate the instrument via serial interface     It is also easy to monitor and operate the AMS remotely with an appropriate router or mo   dem     6 4 Evaluation    The measuring system enables an extensive telemetric control and operation in a variety of  ways  Ethernet  RS232  RS485      6 5 Assessment    The measuring systems can be extensively monitored and operated from an external PC via  modem or router     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 93 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 2 6 Switch over    Switch over between measurement and functional check and or calibration shall be  possible telemetrically by computer control or manual intervention     6 2 Equipment  No additional equipment required   6 3 Performance of test    The AMS can be monitored and partly operated by the user directly on the instrument or via  telemetric remote control     Some functions  e g  performing a 
68.  B of  4   RSS is determined analogue to the  calculation in module 5 4 10     With respect to b     The value of the slope b may be used as a term to correct all input values y  according to the  following equation         Yi  Y icorr g b  The resulting values of Yi cor may then be used to calculate the following new terms by linear  regression   Y corr  C  dx   and  Ue el Voss   aie   u     x    F  c    d   1 x  F   x u  b      n   2     with u b    uncertainty of the original slope b  the value of which has been used to obtain  Yi corr    Algorithms for the calculation of intercepts as well as slopes and their variances by orthogo   nal regression are described in detail in annex B of  4   RSS is determined analogue to the  calculation in module 5 4 10     With respect to c   The values of the slope b and of the intercept a may be used as correction terms to correct  all input values y  according to the following equation     _yiza  Vicon b    The resulting values of Yi cor may then be used to calculate the following new terms by linear  regression        Y corr  C  dx     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control      Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 167 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F   
69.  BKG until the value is  saved     e Inthe Main Menu  choose Calibration Factors  gt  PM Bkg        ERED CH  perm    PPP L Ea ln BA i          PM Coefficient The PM Coefficient screen allows the user to manually adjust the PM  coefficient  The PM concentration based on the SET COEF TO is  displayed to facilitate setting  The PM concentration displayed in the Run  portion of the screen is NOT affected by the SET COEF until the value is    saved     e Inthe Main Menu  choose Calibration Factors  gt  PM Coef        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 15    Operation  Calibration Factors Menu    SHARP Background    SHARP Coefficient    Nephelometer Background  Values    3 16 Model 50307 SHARP Instruction Manual    The SHARP Background screen allows the user to manually adjust the  SHARP Intercept  The SHARP concentration based on the SET BKG TO  is displayed to facilitate the setting  The SHARP concentration displayed in  the Run portion of the screen is NOT affected by the SET BKG until the  value is saved     e Inthe Main Menu  choose Calibration Factors  gt  SHARP Bkg        The SHARP Coefficient screen allows the user to manually adjust the   SHARP coefficient  The SHARP concentration based on the SET COEF  TO is displayed to facilitate setting  The SHARP concentration displayed  in the Run portion of the screen is NOT affected by the SET COEF until    the value is saved     e Inthe Main Menu  choose Calibration Factors  gt  SHARP Coef        The N
70.  Cleaning    on page 5 5  e    Pump Rebuilding    on page 5 6  e    Leak Test    on page 5 6  e    Filter Tape Replacement    on page 5 7  e    Greasing the Cam    on page 5 8  e    External Pump Exhaust Filter    on page 5 8    e    Cleaning the SHARP Optics Assembly    on page 5 10    Read the safety precautions before beginning any procedures in this  chapter     WARNING If the equipment is operated in a manner not specified by the  manufacturer  the protection provided by the equipment may be  impaired  A    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  For more  information about appropriate safety precautions  see the    Servicing     chapter  A    Model 50307 SHARP Instruction Manual 5 1    Preventive Maintenance  Replacement Parts    Replacement  Parts    A    Cleaning the Outside  Case    A    Cleaning the Inlets    A    U S  EPA PM   Inlet    5 2 Model 50307 SHARP Instruction Manual    See the    Servicing    chapter for a list of replacement parts and the associated  replacement procedures     WARNING If the equipment is operated in a manner not specified by the  manufacturer  the protection provided by the equipment may be  impaired  A    Clean the outside case using a damp cloth being careful not to damage the  labels on the case     Equipment Damage Do not use solvents or other cleaning products to  clean the outside
71.  EEE of          1 0 Board Voltages The I O Board screen  read only  is used to display the voltage readings on  the I O expansion board  This menu is only displayed if the I O expansion  board option is installed     e In the Main Menu  choose Diagnostics  gt  Voltages  gt  I O Board        DHAT Py Tame p piiopi  a Lb ib  i          Detector Board Voltages The Detector Board screen  read only  is used to display the voltage  readings on the measurement interface board     e Inthe Main Menu  choose Diagnostics  gt  Voltages  gt  Detector Board        ERED CHA Pi Tam oH miiohel    BEEP umt Bone PME TY Saad Haat ake ME Scala  Pi EE EE       Nephelometer Board The Nehpelometer Board screen  read only  is used to display the voltage  Voltages readings on the nephelometer board     e Inthe Main Menu  choose Diagnostics  gt  Voltages  gt  Nephelometer  Board     3 60 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Diagnostics Menu       RH Temperature The RH Temperature screen  read only  displays the current relative  humidity and temperature readings  The board temperature is the air  temperature measured by a sensor located on the measurement interface    board     e Inthe Main Menu  choose Diagnostics  gt  RH Temperature        Pressure Vacuum The Pressure Vacuum screen  read only  displays the current pressure    readings  The pressure is measured by a pressure transducer  The vacuum is  the amount of vacuum under the filter tape  The flow pressure 
72.  Ground  PURGE 3 DC J25  24V  2 Spare1_Sol  VD1 J26 AIN_VD 1  2 Ground  VD2 J27 AIN_VD 2  2 Ground  BOARD TEMP J28 Board_Therm  2 Ground  SPARE FLOW J29  24V  2 Aout_Spare_Flow  3 Ground  PURGE 4 DC J30 1  24V  2 Spare2_Sol  PRESS BD J31 1  15V    Model 50307 SHARP Instruction Manual 6 17    Troubleshooting  Connector Pin Descriptions    6 18 Model 50307 SHARP Instruction Manual    Connector  Label    Reference  Designator    z9  5    Oo aoa N on A OO N    Signal Description    AIN_30PSIA  AIN_FLOW_1  AIN_VAC_2  AIN_FLOW_2  AIN_VAC_1   5V    15V       Ground       PRESS TEMP    J32         w N    Press_RID  Press_RID    Ground       CNTR WHEEL    J33           mo N OQO wo A WO N     5V  Ground  Ground  Ground  VOA3   5V  NC   NC       BETA 1    J34           o oO A WO N    Ground   RS485   RS485  Ground   5V    Ground       BETA 2    J35           o oO A WO N    Ground   RS485   RS485  Ground   5V    Ground       LOAD MOTOR    J36            24V  Load_Motor_On       SP PHOTO    J37     5V    Thermo Fisher Scientific    Connector  Label    Reference  Designator    Pin    on FP WO N    Troubleshooting  Connector Pin Descriptions    Signal Description    Ground  Ground   5V   VO_SP       CAM WHEEL    J38         YN oOo om FP WO N     5V  Ground  Ground  NC  Dig_Slit   5V   NC       HWELL 1    J39           FP WoO N    AIN_RTDP   5V  AIN_RHP_Hum    Ground       SAMPLE 2 TEMP    J40         wo N    Sample2A_RTD  Sample2B_RTD    Ground       SAMPLE 1 TEMP         wo N    Sample1A_R
73.  In the Main Menu  choose Diagnostics  gt  Analog Input Readings        The Analog Input Voltages screen  read only  displays the 8 raw analog  voltage readings  if the I O expansion board option is installed      In the Main Menu  choose Diagnostics  gt  Analog Input Voltages        The Digital Inputs screen  read only  displays the state of the 16 digital  inputs  Pull ups are provided on all of the inputs  so if nothing they will  read  1   if an input is brought to ground  it will read  0      In the Main Menu  choose Diagnostics  gt  Digital Inputs     Model 50307 SHARP Instruction Manual    3 63    Operation  Diagnostics Menu    Relay States    Test Analog Outputs    Set Analog Outputs    3 64 Model 50307 SHARP Instruction Manual       EH pikari  Ce a E S a E       The Relay States screen displays the state of the 10 digital outputs and  allows toggling of the state to either on  1  or off  0   The relays are  restored to their original states upon exiting this screen     e Inthe Main Menu  choose Diagnostics  gt  Relay States     e Press to toggle and set the relay state to open or closed for the  selected digital output        Pi Taal tH ithe    Haat abe EE Scala  Pi EE EE          The Test Analog Outputs menu is used to set each of the analog output  channels to zero or full scale  Channel choices include all analog outputs  6  voltage channels  and 6 current channels  if the I O expansion board  option is installed      e Inthe Main Menu  choose Diagnostics  gt  Tes
74.  Inputs    Thermo Fisher Scientific    Operation  Volumetric Conditions       Figure 3   5  Common Flags       Figure 3   6  Detector A Flags    The Analog Inputs screen allows the user to select the parameter  none or  analog inputs 1 8  to the selected record field  The selected item is shown    Model 50307 SHARP Instruction Manual 3 31    Operation  Volumetric Conditions    by     lt       after it  Note that at this point  pressing indicates that these  are proposed changes as opposed to implemented changes  To change the  selected record format and erase record log file data  see    Commit  Content    below     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Select Content  gt  select Field  gt  Analog Inputs        ip priori  PIL ERE i       Commit Content The Commit Content screen is used to save any changes that have been  made to any of the record fields for the selected record type  Saving changes  will erase record log file data for that record type  If no changes have been  made    NO CHANGES TO RECORD LIST  will appear  For more  information about selecting the content of logged data fields  see    Select  Content    above     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Commit Content        meh ET eT  ai    CHES Pi EH Chi  Sneek Enone ETET Sed h i  i    EHO pmp pri  Eo Eiee HH    Pihenni i          Reset to Default Content The Reset to Default Content screen is used to reset all of the datalogging  
75.  Instrument Controls Menu       Temperature Threshold The Temperature Threshold screen is used to change the temperature  threshold value     e Inthe Main Menu  choose Instrument Controls  gt  Set Heater  gt     Temp Threshold           Pr irog    CHEER    Bai PERT pihe frin i i          Filter Tape Control The Filter Tape Control menu allows the user to manually move or set the  next time period for tape control     e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape  Control        Manual The Manual screen allows the user to move or stop the filter tape and open  or close the measurement bench     e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape    Control  gt  Manual     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 21    Operation  Instrument Controls Menu       DH vided  i aa E SO a EA          Mass Limit The Mass Limit screen allows the user to set the mass limit  The default is  1 500 pg  or 1 5 mg  and is specified as part of any U S  EPA and TUV  approvals  However  the limit can be increased as much as 5 000 pg for  non regulatory monitoring     e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape  Control  gt  Mass Limit        DH vided  POPLAR          NextTime The Next Time screen is used to view and set the initial date and time  24   hour format  for the next filter tape change     e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape  Control  gt  Next Time           RRS E CHT mito    Wi rnd  E PURE
76.  J1    Pin         FP WoO N    Signal Description    GNDA   5V  Temp  RH       DET SRC    J2         on FP WwW N    ADC1  IRED  PHD A  PHD K  NC       DETECTOR    J3    GNDA  Det Out   5V    Thermo Fisher Scientific    Thermo Fisher Scientific    Troubleshooting  Service Locations                Locations       Connector Reference Pin Signal Description  Label Designator  4  5V  DATA RS485 J4 1 Ground  2  RS485  3  RS485  VALVE DRIVE 1 J7  24V DC  2 Drive 1  VALVE DRIVE 2 J8  24V DC  2 Drive 2  24V IN J9  24V DC  2 Ground    1 866 282 0430 Toll Free    1 508 520 0430 International    For additional assistance  service is available from exclusive distributors  worldwide  Contact one of the phone numbers below for product support  and technical information or visit us on the web at www thermo com aqi     Model 50307 SHARP Instruction Manual 6 25    Thermo Fisher Scientific    Chapter 7  Servicing    This chapter explains how to replace the Model 50307 subassemblies  It  assumes that a subassembly has been identified as defective and needs to be  replaced  or is an    expendable    item  not covered under warranty   Expendable items are indicated with an asterisk     in the    Model 5030i  Replacement Parts    table     For fault location information  refer to the    Preventive Maintenance     chapter and the    Troubleshooting    chapter in this manual     The service mode section in the    Operation    chapter also includes  parameters and functions that are useful when making
77.  Mass Calibration  gt  Mass  gt   Manual        Alpha Efficiency The Alpha Efficiency Coefficient screen allows the user to view and set the  Coefficient alpha efficiency value     3 94 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu    Note This value is set in the factory and should not be adjusted  A    e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Alpha Eff  Coefficient     aa ee  et Taic       Detector Calibration The Detector Calibration submenu is used to view and set the detector  calibration  The detector calibration screen is visible only when the  instrument is in service mode  For more information on the service mode   see    Service Mode    earlier in the chapter     Note This adjustment should only be performed by an instrument service  technician on a defined service interval and at least once per year  or if the  detector is replaced  A    e Inthe Main Menu  choose Service  gt  Detector Calibration        ili          Auto Detector Calibration The Auto Detector Calibration screen allows the user to conduct a detector  optimization routine that will establish the optimal plateau for detector  performance  This process takes approximately 30 minutes and will set a  high voltage and thresholds for alpha and beta reference counts     e Inthe Main Menu  choose Service  gt  Detector Calibration  gt  Auto     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 95    Operation  Service Menu    Manual 
78.  Oe ee 3 105  Unlock Unser ments iinne scr cece eds EREE i 3 105  Chapter 4 Calibration aaaea sacks cc ct teagan cnet aadaki naonana adnada nanginin 4 1  Frequency of Calibration        esssssssessssiseseseseersrsrsrsrsrsrsrsrsrstseseseseseseseses 4 1  Equipment Required     s s sssesssseisesesesesrsesrsesrsrsrsrstsrstststststsestsesesesesereses 4 1  Pre CalibratioN enisi sansa ie ioia RA Erge Ba AEEA NSSE 4 2  Calibration Procedure  s jecsi detent dead 4 2  RH Temperature Calibration iis  taba nadia  eich dt ecg lua ee 4 3  Ambient T  emperat  re 23 cate ceases sees e R 4 3  Ambient Relative Um Gt yeas et cre enone vaG attest aiutccaeaentecee 4 3  Flow Temp  tatirecorireintnen hed eke ode natal nes bal eetact as 4 4  Pressure Vacuum Calibration iacci2iscvecsseit  a iesdiaciiaeetitinetiiniecndie  4 4  Barometer Pressure    ali ratio marie vos del ccechetet ater tem edvecd Meade Seapets 4 4  Calibrate Vacuum Flow Zeb 5 2 eiesccadatasceeesasececanaseeensentaciedeseaad 4 5  Calibrate Vacuum Pressure Spats d ioe hie etesae ns onuaeoseds 4 5  Calibrate Flow Pressure S paiva Gio6 eee sec tee dane eee seed dseeeees 4 6  Flow Calibration sienien sea iaa a dae 4 7  Auto  Flow Calibtati  n osioon kot i E Aa a as 4 7  Mass Calibratio Nesser A E 4 7  Mass Coefficient ance Gs cet a EEE E acacia 4 9  Auto Mass MCOCTICICN i  ataverdisestone Soest mses iaudedstacncesteangene eek 4 9  Detector Calibration einn tea Matsa ites tas cara teen tu kt Ge 4 12  Auto Detector Calibrate ties scented sa
79.  PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    cepts    of October 2001 and further specified in VDI 4203  Part 3    Testing of automated  measuring systems   Test procedures for point related ambient air measuring systems for  gaseous and particulate air pollutants    of August 2004     Standard VDI 4202  Part 1 and VDI 4203  Part 3 were republished after an extensive revision  with latest update September 2010  Unfortunately  after this revision uncertainties and con   tradictions remain regarding the concrete minimum requirements on the one hand  and the  general relevance of the test points on the other hand when testing particulate ambient air    measuring devices  The following test points require clarification     6 1 5 3 2 Repeatability standard deviation at zero point   no minimum requirement defined   6 1 5 3 3 Repeatability standard deviation at reference point  not relevant for particulate devices   6 1 5 3 4 Linearity  lack of fit    not relevant for particulate devices   6 1 5 3 7 Sensitivity coefficient of surrounding temperature  no minimum requirement defined   6 1 5 3 8 Sensitivity coefficient of supply voltage   no minimum requirement defined   6 1 5 3 11 Standard deviation from paired measurement   no minimum requirement defined   6 1 5 3 12 Long term drift   no minimum requirement defined   6 1 5 3 13 Short term drift   not relevant for particulate device
80.  Pressure User set alarm min value User set alarm max value  Ambient RH User set alarm min value User set alarm max value  Sample RH User set alarm min value User set alarm max value  Ambient Temp User set alarm min value User set alarm max value  Board Temp User set alarm min value User set alarm max value  Flow Temp User set alarm min value User set alarm max value  Flow Vol User set alarm min value User set alarm max value  Beta User set alarm min value User set alarm max value  Beta Raw User set alarm min value User set alarm max value  Alpha Raw User set alarm min value User set alarm max value  Mass  200 5 000   Beta Zero 0 65 000       Thermo Fisher Scientific    Operation  Volumetric Conditions                         Output Zero  0   Value Full Scale 100  Value  External Alarms 0 7 000   Neph Temp User set alarm min value User set alarm max value  Neph RH User set alarm min value User set alarm max value  Neph IRED User set alarm min value User set alarm max value  Neph Ref 350 2000   Beta Ref 0 65 000   Mass Comp 0 100    Choose Signal to Output The Choose Signal to Output screen displays a list of the analog output  signal choices  Choices are grouped into three categories  Concentrations   Other Measurements  and Analog Inputs  if the I O expansion board  option is installed   This allows the user to select the output signal that will  be assigned to each analog channel  The Concentrations screen is shown  below  See Table 3 5 for a list of items for each signal g
81.  Reference   yg m          Figure 49     Reference vs  candidate  SN 1  measured component PMps  all test sites          PM    5030i Sharp   yg m      100    90 4    80 4    30 4    20 4    70 4    60 4    50 4    40 J                                  A      a SN2     a  We  i  V  a     f  a F    7      J s     a  ay  T        0  r r r r r r r r  0 10 20 30 40 50 60 70 80 90 100    PM    Reference   ug m           Figure 50     Reference vs  candidate  SN 2  measured component PMos  all test sites    740300_2013_936_21209885F_EN docx    Precisely Right     Page 160 of 424    A TUVRheinland      TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F                                                                      100 7  90   a SN1 Cologne Winter     Pa     Pa  80 4 7    70 4     2 f  B 60   Sf  3 50 4 7    40   A  a  30 4 Vaa  20   a  afl  10 5 a  K  a E E E E E ee  0 10 20 30 40 50 60 70 80 90 100  PM    Reference   yg m   Figure 51  Reference vs  candidate  SN 1  measured component PM25  Cologne  winter   100 j  V  90 J    SN2 Cologne Winter   a  ae  80      Pa    70 4 i  2 v  2 60   a    Fa  ba A  8 50   7    40   A     30   ri  R  20   on   vA  10   af  a     i  0 10 20 30 40 50 60 70 80 90 100  PM     Reference   yg m
82.  Report on the suitability test of the ambient air quality measuring system   Page 214 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    B  UK test site  Teddington     B 1 Implementation of Weighing Protocols    NPL  National Physical Laboratory  were subcontracted to weigh filters manually for the field  study  In line with EN14907 filters were kept in the weighing room for less than 28 days  the  glove box used for weighing was maintained at  20   1    C and  50   5     and filters were  weighed twice before and after sampling  Table 41 summarises the conditioning and weigh   ing timescales utilised     Table 41  Conditioning and weighing timescales    Pre Sampling Post Sampling    Condition minimum of 48 hours Condition 48 hours    Weigh Filters Weigh Filters  Condition 24 hours Condition 24 hours  Weigh Filters Weigh Filters       At the start of each weighing session the balance was exercised to remove mechanical stiff   ness  and then calibrated  At the start and end of each batch of filters  a 50 and 200 mg  check weight were weighed  In line with the recommendations of the UK PM Equivalence  Report  8   filters were weighed relative to a 100 mg check weight  and not a tare filter  as the  latter was shown to lose mass over time  Four filters were weighed between check weights   as the balance drift over time had been shown to be small     740
83.  SN 1  and 217  SN 2  valid paired values      The parameters of the calibration function   y m x b  were determined by orthogonal regression  The analysis function is the reversal of the cali   bration function  It reads    x   1 m   y     b m  The slope m of the regression line characterises the sensitivity of the measuring system  the  ordinate intercept b characterises the zero point   6 4 Evaluation    This results in the following parameters     Table 26  Parameters of the calibration and analysis function  measured component PM25    System No  Calibration function Analysis function    SO E    System 1 1 087  0 408 0 920  0 375   SN 1    System 2 1 055 0 066 0 948 0 063   SN 2        740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 133 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 5 Assessment    A statistical correlation between the reference measuring method and the device reading  could be demonstrated     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  See module 5 4 10     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability te
84.  SN 1 and SN 2   The candidates were set up as follows     Component threshold for heater regulation  SN 1 PM25 58    SN 2 PM25 50      The following test program was performed    e comparison campaign with a minmum of 40 valid paired values reference vs candi   date system   e determination of the uncertainty between candidate systems Ups according to Guide   e determination of the expanded uncertainty of the candidate systems according to  Guide   e application of the correction factors  terms determined in chapter 6 1 5 4 10 Cal   culation of the expanded uncertainty of the instruments    e recalculation of equivalence of the 4 data sets from the existing suitability test and the  additional data set    Bornheim Summer 2013 with different thresholds for the heater  regulation according to the approach from point    8 2 Suitability testing    der CEN TS  16450  9     The additional campaign was performed between 07 June 2013 and 25 July 2013  The am   bient conditions during the campaign were as follows     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system    Model 5030  SHARP with PMz25 pre separator of the company Thermo Fisher    Scientific for the component suspended particulate matter PMz 5     Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 65 of 424    Table 8  Ambient conditions at the additional site    Bornheim 
85.  Settings  gt  NTP Svr        egl    ep prr  fant Pii fini i          1 0 Configuration The I O Configuration menu deals with configuration of the analyzer   s  I O system  The analog input configuration is displayed only if the I O  expansion board option is installed     e Inthe Main Menu  choose Instrument Controls  gt  I O  Configuration      O EH OE Bel  PIL    Eo Miket       Output Relay Settings The Output Relay Settings menu displays a list of the 10 digital output  relays available  and allows the user to select the logic state and instrument  parameter for the relay selected     3 44 Model 5030  SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions    Note The digital outputs may take up to one second after the assigned  state occurs to show up on the outputs  A    e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Output Relay Settings        Logic State The Logic State menu item is used to change the selected I O relay to  either normally open or normally closed  The default state is open  which  indicates that a relay connected between the digital output pin and ground  is normally open and closes to trigger the digital output action     e Press to toggle and set the logic state to normally open or    normally closed        WH hid          Instrument State The Instrument State submenu allows the user to select the instrument  state that is assigned to the selected relay output  A submenu lists signal  types 
86.  Settings  gt  Serial Settings  gt  Data Bits        DH Guided  PLP          The Parity screen is used to select the parity bit for the serial port to None   default   Even  or Odd     In the Main Menu  choose Instrument Controls  gt  Communication    Settings  gt  Serial Settings  gt  Parity        DH minded  Piet          The Stop Bits screen is used to set the number of stop bits for the serial  port to 1  default  or 2     In the Main Menu  choose Instrument Controls  gt  Communication    Settings  gt  Serial Settings  gt  Stop Bits     Thermo Fisher Scientific    Operation  Volumetric Conditions       RS 232 RS 485 Selection The RS 232 RS 485 Selection screen allows the user to choose between the    RS 232 or RS 485 specification for serial communication     Equipment Damage Disconnect the serial cable before changing the RS   232 and RS 485 selection to prevent damage to any equipment currently  connected to the analyzer  A    e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Serial Settings  gt  RS 232 485 Selection        Instrument ID The Instrument ID screen allows the operator to edit the instrument ID  number  The ID is used to identify the instrument when using the C Link  or MODBUS protocols to control the instrument or collect data  It may be  necessary to edit the ID number if two or more instruments of the same  model are connected to one computer  Valid instrument ID numbers are  from 0 to 127  The Model 50302 has a default inst
87.  Summer 2013     shown as    daily mean values                                  Bornheim   motorway parking lot  summer   2013  No  of paired values 44   reference   Ratio PMz25 PM1o      Range 39 1     75 1  Average 59 9  Ratio Volatiles PM25       Range 4 7     38 4  Average 17 4  Air temperature    C   Range 13 2     27 6  Average 19 8  Air pressure  hPa   Range 1001     1020  Average 1010  Rel  humidity      Range 52 2     86 1  Average 67 8  Wind velocity  m s   Range 0 2  3 5  Average 1 2  Precipitation  mm d   Range 0 0     34 6  Average 32             All values are to be found in Appendix 7  PM measuring values  and 8  ambient conditions      In total 45 paired values were determined  A significant outlier according to Grubbs is the  paired value for the reference PM25 of 18 July 2013  Ref1  11 3yg m   Ref2  9 0 g m    These values were discarded from the data pool  Therefore 44 paired values are available    for evaluation     The evaluation of the comparison measurements according to the Guide  4  leads to the fol     lowing result     740300_2013_936_21209885F_EN docx    Y 7    TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 66 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Table 9  Results of the equivalence t
88.  TE Aad fuese PEO Fed bo bi Feed Eo beet FD ETE    DH mini   i fimi nE g        a a DHAN Popop  Rmi mii Bot PE     Period The Period screen allows the user to set the period  The next filter tape  change would be scheduled after the number of hours the period is set for   The official tape change for TUV and U S  EPA for PM2 5 and PM is set    for 8 hours     3 22 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Instrument Controls Menu    e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape  Control  gt  Period        Counter The Counter screen allows the user to record the current filter tape count  and reset the tape count to zero     e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape  Control  gt  Counter           Poh fen  Mm  Tore  H  d hi          Tape Zero Ratio The Tape Zero Ratio screen allows the user to set the occurrence of  nephelometer zeroing after X filter changes  For PM2  X   42 and for PMio  X   14 based on a 2 week zeroing cycle  A conservative ratio of 1 1 will  zero the nephelometer with every filter tape change     e Inthe Main Menu  choose Instrument Controls  gt  Filter Tape  Control  gt  Tape Zero Ratio        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 23    Operation  Volumetric Conditions    Volumetric The Volumetric Conditions menu allows the user to turn  Conditions temperature pressure compensation on and off and to set the standard    pressure and temperature values in vario
89.  Table 21  Dependence of the zero point  NEPH  on surrounding temperature  deviation in  ug m     average of three measurements    Temperature Deviation    TNE   SNEINERH    a    gt  CC  Oo mod       The results of the three single measurements are presented in annexes 2 and 3 of the ap   pendix     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control      Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 111 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 3 8 Sensitivity coefficient of supply voltage    The sensitivity coefficient of supply voltage shall not exceed the requirements of Table  2 of Standard VDI 4202 Part 1  September 2010   A value c  at 70   to 80   of the up   per limit of the certification range shall be used as reference point     Note     This item cannot be evaluated according to the current versions of Standards VDI 4202 Part  1  September 2010  and VDI 4203 Part 3  September 2010  when testing particulate meas   uring systems  due to the fact that a minimum requirement has not been defined  By resolu   tion of the competent German body  see module 5 3 1   reference is made to the following  requirement in the previous version of Standard VDI 4202 Part 1  June 2002      The change in the measured values 
90.  The following example reports that the ncal is functional     Send  alarm ncal status  Receive  alarm ncal status ok    alarm neph led cur min   alarm neph led cur max   These commands report the current nephelometer LED alarm minimum  and maximum value settings  The following example reports that the  nephelometer LED alarm minimum value is 50 mA     Send  alarm neph led cur min  Receive  alarm neph led cur min 50 0 mA    set alarm neph led cur min value   set alarm neph led cur max value   These commands set the nephelometer LED alarm minimum and  maximum values to value  where value is a floating point number  representing nephelometer LED alarm limits in mA  The following  example sets the nephelometer LED alarm maximum value to 75 0 mA     Send  set alarm neph led cur min 75  Receive  set alarm neph led cur min 75 ok    alarm neph rh min   alarm neph rh max   These commands report the current nephelometer relative humidity alarm  minimum and maximum value settings  The following example reports  that the nephelometer relative humidity alarm minimum value is 5 0      Send  alarm neph rh min  Receive  alarm neph rh min 5 0      set alarm neph rh min value   set alarm neph rh max value   These commands set the nephelometer relative humidity alarm minimum  and maximum values to value  where value is a floating point number  representing nephelometer relative humidity alarm limits in percent  The  following example sets the nephelometer relative humidity alarm maximum    val
91.  Thermo Fisher Scientific    Chapter 1    Chapter 2    Chapter 3    Contents    AVE OCU CNG EN OND esaisscced a sasssctencSatddecsnndnnsanndnentaotaninmondunte bebe dunmdnassntanauisauananiaatidy 1 1  P  inciple Ol Operation a  blonde essa ERENS 1 3  SPECICA LIONS to  ethene vacate eea cated ra hus aieiaa eae nie tease 1 8   Installation   eiis raisini asuadbvbaidatncasuedicdieussuassuaaeaicsassincanas 2 1  P  ckaging and Transport neirinne RR t 2 1  PIFRE en ive aed wthaa aaa EA E EN 2 2  Unpacking and Inspection    ssssssseeesssseieeeeeeetsestersssesrsrsrsrsisesesesesesesesrses 2 2  Acceptance Testing and Startup Procedures 00    cceseseeseeseeeeeeseeeeneeee 2 3   Acceptance Testei a a i a 2 4  One Point Temperature Verification  ae c  ccc taeandgestanclancenace 2 7  One Point RH Sensor Verification        ss essesesseseseeeesesesseseseseesesesseses 2 7  One Point Barometric Pressure Verification         s sssesesesessesesreeeseses 2 8  One Point Volumetric Flow Rate Verification      s ssssssssssesseseseseeseses 2 8  Leak Check Procedure ists vac cata ae castes E N a 2 9  Nephelometer Source Verification       e ssesesesesesessssesesreeesesseseresesese 2 9  Nephelometer eTOCs belie al al ca ie ea ate Coa 2 9  Sepan an a a tes atissencenacsales ana beds tustestesassslens 2 10  Inlet  Systemi er e E E E T 2 10  Heater iv tiencaientt dassedansanueal aayia uawlscdiwuah odes tube ictalad aystn sven aled dd pasetabuncste 2 11  Sample Tube Extension  Lengths 1 02   cccstesccsceti
92.  These limits are designed to provide reasonable protection against harmful  interference when the equipment is operated in a commercial environment   This equipment generates  uses  and can radiate radio frequency energy  and  if not installed and used in accordance with the instruction manual   may cause harmful interference to radio communications  Operation of this  equipment in a residential area is likely to cause harmful interference  in  which case the user will be required to correct the interference at his or her  Own expense  A    iv Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    About This Manual  WEEE Symbol    WEEE Symbol The following symbol and description identify the WEEE marking used on    the instrument and in the associated documentation     Symbol Description    Marking of electrical and electronic equipment which applies to waste  electrical and electronic equipment falling under the Directive 2002 96 EC        WEEE  and the equipment that has been put on the market after 13 August  2005   amp     Where to Get Help Service is available from exclusive distributors worldwide  Contact one of  the phone numbers below for product support and technical information  or visit us on the web at www thermo com aqi     1 866 282 0430 Toll Free  1 508 520 0430 International    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual v    About This Manual  Where to Get Help    vi Model 50307 SHARP Instruction Manual Thermo Fisher Scientific   
93.  This command reports whether pressure compensation is for actual or  standard conditions  The following example shows the pressure  compensation is standard     Send  pres comp  Receive  pres comp std    set pres comp selection  selection     act   std      Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Measurement Configuration    These commands turn the pressure compensation to either actual or  standard conditions  The following example turns the pressure  compensation to actual     Send  set pres comp act  Receive  set pres comp act ok  pres std    This command reports the standard pressure  The following example  reports that the standard pressure is 760 mmHg     Send  pres std  Receive  pres std 760 mmHg    set pres std value   This command sets the standard pressure to value  where value is a floating   point number representation of standard pressure between 0 to 760  mmHg  The following example sets the standard pressure to 730 mmHg     Send  set pres std 730  Receive  set pres std 730 ok  range    This command reports the current PM range  If the mode is incorrect  the  instrument responds with    can   t  wrong settings     The following example  reports that the PM range is set to 5 mg m     according to Table B 5     Send  range  Receive  range 4  5 000E 00 mg m3    set range selection  This command selects the PM ranges  according to Table B   5  The  following example sets the PM range to 10 mg m        Send  set range 5  Rec
94.  This offset   ZERO  should now be entered in the screen below        Be sure to save the entry and compare the values once more  If it appears  that the temperature has shifted into the wrong direction  change the sign  of your offset value     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 4 3    Calibration  Calibration Procedure    Flow Temperature Assuming the instrument cover has been removed  the heater has been  turned off and removed from the instrument and the instrument has been  sampling room temperature air for 1 hour  this calibration can now  proceed     Using a NIST traceable thermometer as a reference collocated next to the  small sample tube inlet on top of the instrument  measure and compare  three individual readings between both the reference and the 50302  response  Taking an average of both sets of readings  calculate the average  difference between the two readings and record that as your offset  This  offset  ZERO  should now be entered in the screen below        Be sure to save the entry and compare the values once more  If it appears  that the temperature has shifted into the wrong direction  change the sign  of your offset value     Pressure Vacuum There are three pressure sensors that can be calibrated  however  the  Calibration primary sensor to be calibrated is the barometric pressure sensor  Both the  vacuum sensor and pressure flow sensors are re zeroed automatically with  every filter tape change     Barometer Pressure    eran 
95.  a a R 8 5  EEE AN SOLOMON waa  son snauenp annant i a a a a 8 5  Pump  Intake T Si aul ut ann e a a ease NSE 8 5  FirmWat    tiasutssiiqctetut yaeatedeeagtoditiagedadsladdatatGhettideidddasst sadatedsbaddotakdhyanlartig 8 5  Instrument Control  sncssaissstteninwsvatnichilesndsieneiededastienvasicebdanntees 8 5  Monitoring Signals menaient iii 8 6  Output Comm  nication sinni aiar 8 6  Electr  nics arien TETE E E E E E 8 6  Motherboard  einir fsada a e a a a a a 8 6  Nephelometer Interface Board    ssssssssssssesessssseseeesessrrrsterrreserrssesese 8 7  Nephelometer interface board connectors   ssssssssssssseseseeeseesesesessses 8 7  Measurement Interface Board   s ssssssssesesesesesesesersrsrsrsrersrersesesesesesessse 8 7  Measurement Interface Board Connectors       essesseseseseeeeseseeeressses 8 8  Flow Sensor OVS te aad scar aa aa oh Gas Saco R eek r cade neat eee 8 8  Pressure Sensor SSCHIOLY  Secrets in us cs hal ace Sasa nat eaGicaeaevyar ee 8 8  Detector Amplifier  Assemiblys fe  cect Wixteanscesset ete area recente 8 8  Digital Output Board js2ccecieee desi eeiitieeeoiiewuetiacdaginlincucn 8 8  I O Expansion Board  Optional      dssst1ic nsacetitiacidsctitiendcontes 8 9  Front Panel Connector Board oca 2jccheetietioe eel a teceed ects nataceae ated 8 9  VO Components en siete cisstiasd e a a ivoheatederice 8 9    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual xv    Contents    xvi    Chapter 9    Chapter 10    Appendix A    Appendix B    Model 50307 S
96.  accident  disaster or event of force majeure   iii  misuse  fault  or negligence of or by Buyer   iv  use of the Products in a manner for which    Model 50307 SHARP Instruction Manual A 1    Warranty  Warranty    they were not designed   v  causes external to the Products such as  but not  limited to  power failure or electrical power surges   vi  improper storage  and handling of the Products or  vii  use of the Products in combination  with equipment or software not supplied by Seller  If Seller determines  that Products for which Buyer has requested warranty services are not  covered by the warranty hereunder  Buyer shall pay or reimburse Seller for  all costs of investigating and responding to such request at Seller s then  prevailing time and materials rates  If Seller provides repair services or  replacement parts that are not covered by the warranty provided in this  warranty  Buyer shall pay Seller therefor at Seller s then prevailing time and  materials rates  ANY INSTALLATION  MAINTENANCE  REPAIR   SERVICE  RELOCATION OR ALTERATION TO OR OF  OR  OTHER TAMPERING WITH  THE PRODUCTS PERFORMED BY  ANY PERSON OR ENTITY OTHER THAN SELLER WITHOUT  SELLER S PRIOR WRITTEN APPROVAL  OR ANY USE OF  REPLACEMENT PARTS NOT SUPPLIED BY SELLER  SHALL  IMMEDIATELY VOID AND CANCEL ALL WARRANTIES WITH  RESPECT TO THE AFFECTED PRODUCTS     THE OBLIGATIONS CREATED BY THIS WARRANTY  STATEMENT TO REPAIR OR REPLACE A DEFECTIVE PRODUCT  SHALL BE THE SOLE REMEDY OF BUYER IN THE EVENT OF A  
97.  adjustments or  diagnosing problems     For additional service assistance  see    Service Locations    at the end of this  chapter     This chapter includes the following parts information and component  replacement procedures     e    Safety Precautions    on page 7 3   e    Firmware Updates    on page 7 4   e    Accessing the Service Mode    on page 7 4   e    Replacement Parts List    on page 7 5   e    Cable List    on page 7 6   e    External Device Connection Components    on page 7 8    e    Removing the Measurement Case Assembly and Lowering the  Partition Panel    on page 7 12    e    Fuse Replacement    on page 7 14   e    Fan Filter Replacement    on page 7 14   e    Detector Amplifier Assembly Replacement    on page 7 15  e    Detector Amplifier Board Calibration    on page 7 17   e    External Pump Replacement    on page 7 17    e    Pressure Board Replacement    on page 7 18    Model 50307 SHARP Instruction Manual 7 1    Servicing  Service Locations    e    Pressure Board Calibration    on page 7 20   e    Analog Output Testing    on page 7 21   e    Analog Output Calibration    on page 7 24   e    Analog Input Calibration    on page 7 25   e    Themistor Replacement    on page 7 26   e    I O Expansion Board  Optional  Replacement    on page 7 27  e    Digital Output Board Replacement    on page 7 29   e    Motherboard Replacement    on page 7 30   e    Measurement Interface Board Replacement    on page 7 31  e    Photo Interrupt Board Replacement    on page 7
98.  and      _ RSS  US s Yicor     n   2     with u b    uncertainty of the original slope b  the value of which has been used to obtain  Yicorr and with u a    uncertainty of the original intercept a  the value of which has been used  to obtain yi corr    Algorithms for the calculation of intercepts as well as slopes and their variances by orthogo   nal regression are described in detail in annex B of  4   RSS is determined analogue to the  calculation in module 5 4 10     The values for Uc scor are used for the calculation of the combined relative uncertainty of the  candidate systems after correction according to the following equation        u  x    c    d   1 x  F   x2u  b   u   a        us scor  Yi   W  omoor Vi    a  ta    For the corrected dataset  uncertainty We cmcorr IS calculated at the daily limit value by taking yi  as the concentration at the limit value     The expanded relative uncertainty Wemcor is Calculated according to the following equation   Wewcorr   k W cmcorr  In practice  k 2 for large number of available experimental results    The highest resulting uncertainty Wemcorr is compared and assessed with the requirements  on data quality of ambient air measurements according to EU Standard  7   Two results are  possible     1  Wom  lt  Wago     Candidate method is accepted as equivalent to the standard method    2  Wom  gt  Wago     Candidate method is not accepted as equivalent to the standard method   The specified expanded relative uncertainty Waqo for p
99.  be cycled after any of these parameters  have been changed for the change to take effect  A    Model 50307 SHARP Instruction Manual 3 41    Operation  Volumetric Conditions    e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP Settings     i 7  J ae ee  Ties EEE i       Use DHCP The Use DHCP screen is used to specify whether to use Dynamic Host  Configuration Protocol  DHCP  or not  When DHCP is enabled  the  network dynamically provides an IP address for the instrument     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP Settings  gt  Use DHCP     carte Smee Ti  k T hebe a i    ERED    fi  EPOPEE Eunet Bone PME TY Saad Haat ade EE mi eae  Pi EE EE       IP Address The IP Address screen is used to edit the IP address  The IP address can  only be changed when DHCP is off  If DHCP is on  the instrument will  respond with    NOT SETTABLE IF DHCP IS ON     For more  information on DHCP  see    Use DHCP    above     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP Settings  gt  IP Addr        3 42 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions    Netmask The Netmask screen is used to edit the netmask  The netmask is used to  determine the subnet on which the instrument can directly communicate  to other devices  The netmask can only be changed when DHCP is off  If  DHCP is on  the instrument will respond with    NOT SET
100.  by a space     Send  det status  Receive  det status 1 9657 9600 4623  diag volt det    This command reports the diagnostic voltage measurements on the  detector board  The voltage is positive 5     Send  diag volt det  Receive  diag volt det 4 9  diag volt iob    This command reports the diagnostic voltage measurements on the I O  expansion board  The sequence of voltages is  Positive 24  positive 5   positive 3 3  and negative 3 3  Each voltage value is separated by a space     Send  diag volt iob  Receive  diag volt iob 24 10 4 90 3 20  3 20  diag volt mb    This command reports the diagnostic voltage measurements on the  motherboard  The sequence of voltages is  Positive 24  positive 15  positive  5  positive 3 3  and negative 3 3  Each voltage value is separated by a space     Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Diagnostics    Send  diag volt mb  Receive  diag volt mb 24 10 14 90 4 90 3 20  3 20    diag volt mib   This command reports the diagnostic voltage measurements on the  measurement interface board  The sequence of voltages is  Positive 24   positive 15  negative 15  positive 5  and positive 3 3  Each voltage value is  separated by a space     Send  diag volt mib  Receive  diag volt mib 24 98 14 80  14 90 4 96 3 20  diag volt neph    This command reports the current voltage readings of the nephelometer  interface board  The following example reports the nephelometer interface  board values     Send  diag volt neph  Re
101.  by the number of records or by date and time  Note that  both types of records cannot be viewed at the same time  only the selected    record type     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  View Logged Data     3 26 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions       WH rid    E eo iF a morr  mrii mu Le       Number of Records The Number of Records screen is used to select the number of records to  view  ending with the most recent  It also shows the total number of  records that have been logged for the selected record type        The Record Display screen  read only  displays the selected records        Date and Time The Date and Time screen is used to set a start date and time for which to  view logged data  For example  if    20 Jan 2009 10 00    is entered  then the  first logged data record that is displayed is the first record after this time  If  set to one minute logging  this would be at    20 Jan 2009 10 01           Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 27    Operation  Volumetric Conditions    The Record Display screen  read only  displays the selected records               gt        EPOPEE Emt Bone BUEN Sad i    hT    to PUREE SE i          Erase Log The Erase Log screen is used to erase all saved data for the selected record  type only  not both srecs and lrecs      e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Er
102.  can be omitted  The potential  dependence of the nephelometer   s signal on surrounding temperature is thus examined sole   ly at the zero point     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 108 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    The devices were exposed to the following surrounding temperature sequence in the climatic  chamber    20   C  5   C     20        40 C 20       After an equilibration time of approx  24 h per temperature step  the recording of measured    values at zero point  3 x 24 h per temperature step  and at reference point  3 x per tempera   ture step  started     6 4 Evaluation  Zero point     The values measured at the concentration of the respective 24 h single measurements were  read out and evaluated  The absolute deviation in g m  per temperature step was examined  in relation to the default value at 20   C     Reference value  Bo   2 g m3  Reference point     The percentage change of the determined measured values for the mass coefficient was ex   amined at each temperature step in relation to the default value at 20   C     It should be noted that concentration values could not be simulated by checking the mass  coefficient  
103.  case  A    The inlet assemblies can include a TSP inlet  a U S  EPA compatible PMio  inlet  or a Digital PMio or PM2s5 inlet  Cleaning schedules and routine  maintenance of assembly components should be in accordance with a good  quality assurance plan     In general  it is recommended that the inlet assemblies be cleaned on a  quarterly interval with mild soap solution  a thorough rinsing  and dried  with a lint free cloth  If any impaction surfaces require impactor grease  it  should be applied prior to reassembly  Furthermore  a general inspection   and replacement as necessary  of O rings should be done     Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  For more  information about appropriate safety precautions  see the    Servicing     chapter  A    The water collector bottle located on the PMi inlet should be inspected at  least every five sampling days  Figure 5   1   Remove any accumulated  water  clean the interior of the bottle  inspect the seals  and replace the    bottle in the holder     Thermo Fisher Scientific    European PMio PM25    Thermo Fisher Scientific    Inlet    Preventive Maintenance  Cleaning the Inlets       Top Plate with  f    NCT Deflector Cone    Screen         Spacer        P fi Lower Plate with  crew Rain Deflector        Acceleration Assembly             Ht                                         
104.  commercial at sign     with a source indicator and a starting bit indicator   All alarm information is presumed to be two bits long  low and high   The  bitfield extraction is performed on the integer part of the source  Typical  alarm information would appear as   6 4      Then  there appears an optional translation table within braces       This is  a string of words separated by spaces  An example translation table would  be   Code_0 Code_1 Code_2 Code_3    The value  once extracted  is used  as a zero based index into the translation table to determine the string to    display     Then there appears an optional selection table within parentheses           This is a string of numbers separated by spaces   0 1    The selection table  lists the translation table entries which the user may select from when  setting the parameter  This is not necessarily the same as the entries which  may be displayed     Then there appears an optional button designator  This will be one of  B       fes T   T     or  N    B  Indicates a button which pops up an input dialog prompting the    user for a new value using the designated input format  The input  format is specified from the  B  through the subsequent semicolon     Thermo Fisher Scientific    Thermo Fisher Scientific    Examples    C Link Protocol Commands  Record Layout Definition    I   Indicates a button which pops up a selection list with input  translation  That is  the values read are translated before they are  compared to the
105.  component location   Table 7 1  Model 5030  Replacement Parts  Part Number Description  100480 00 Front Panel Pushbutton Board  110570 00 Processor Board  new   Port and   cramfs required   100533 00 Motherboard  100539 00 Digital Output Board  100542 00 1 0 Expansion Board  optional   102340 00 Front Panel Connector Board  102496 00 Front Panel Display  106926 00 Transformer  Step Down  220 240 VAC  optional   104290 00 Measurement Interface Board  105869 00 Detector Assembly  LND4335   105938 00 Photo Interrupt Board Assembly  10661 1 00 Rear Flow RTD Assembly  106536 00 RH Temperature Assembly  106535 00 Heater Tube Assembly  Three Feet  106037 00 Load Motor Assembly  106032 00 Tape Motor Assembly  106470 00 Proportional Valve Assembly  106946 00 Pressure Board Assembly  w tubing assembly   425454008 Detector Amp Assembly  106923 00 Pressure Valves  106540 00 Extension tubes w fitting  101055 00 AC Receptacle Assembly  101681 00 Power Supply Assembly  24 VDC  with base plate and screws   101688 00 Ambient Temperature Thermistor  with connector   100907 00 Fan  24 VDC  8630 Filter Guard Assembly  with foam    109612 00 Fuse  250 VAC  10 0 Amp  SlowBlow  for 110 VAC and 115 VAC  models    109613 00 Fuse  250 VAC  5 0 Amp  SlowBlow  for 220 240 VAC models            Model 50307 SHARP Instruction Manual 7 5    Servicing  Replacement Parts List                                                                                  Part Number Description   10 001403 External Pump Ass
106.  eE EEEE REER E E 134  6 1 5 4 6 Averaging CMO vcrccrcacds dea coactentiecus since coreedueaesaiaiacuneeeseeieawancapaddacaiecenerateneszenien 137  6 1 5 4 7 Constancy of sample volumetric flOW              cccceeeeeeeeeeeeeeeeneneeeeeeeeeeeeeeeneeeeaees 138  6 1 5 4 8 Tightness of the measuring SYSTOM               eeeeeeeceeeee eee eeeeeeeeeeeeeeeeeeteeeenaaeeeaaaes 141  6 1 Methodology of the equivalence check  modules 5 4 9     5 4 11        eeeeeeeeeeees 143  6 1 5 4 9 Determination of uncertainty between systems under teSt Upg          ccceeeeeeeeees 144  6 1 5 4 10 Calculation of the expanded uncertainty of the instruments                 0008 151  6 1 5 4 11 Application of correction factors ANd terMS         cccccceceeeeteeeeeeteeeeeeeeeeteeeenaaees 165  6 1 5 5 Requirements on multiple component measuring systems            ceeeeeees 170  7  RECOMMENDATIONS FOR PRACTICAL USE              cccceceeeeeeeeeeeeeeeneeeeeneeeeseeeeesees 171  S  MIE Ue cade appaaatveennnas expan E waaaadrqaantareanatsaevenneaan 173  95 APPENDIX Sacaatseppsratvcensiass iacancencnian E mean caemunasiasygatsat maaaterqaaniananatsaavenneian 174    740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     Page 8 of 424    Tables    Table 1   Table 2   Table 3   Table 4   Table 5   Table 6   Table 7     Table 8     Table 9     Table 10   Table 11   Table 12   Table 13   Table 14   Table 15   Table 16   Table 17   Table 18   Table 19   Table 20   Table 21     Table 22   Table 23
107.  following example clears srecs     Send  set clr srecs  Receive  set clr srecs ok    set copy sp to lrec   set copy sp to srec   set copy sp to stream   These commands copy the current selections in scratch pad  sp  into the  lrec  srec  or streaming data list     The scratch pad is a temporary memory area which is used to set up lists of  selections for lrec  srec  or streaming data items  The user can copy any of  these lists to the scratch pad  modify individual elements in the list  then  save the scratch pad back to the original list  For more information on how  to edit the scratch pad  see the    sp field    command     The following example copies the current list in scratch pad into the lrecs  list     Send  set copy sp to lrec  Receive  set copy sp to lrec ok    Model 50307 SHARP Instruction Manual B 27    C Link Protocol Commands  Datalogging    B 28    Model 50307 SHARP Instruction Manual    set copy lrec to sp   set copy srec to sp   set copy stream to sp   These commands copy the current contents of the lrec  srec  or streaming  data list into the scratch pad  sp   These commands are useful in easy  modification of current lrec  srec  or streaming data lists     The scratch pad is a temporary memory area which is used to set up lists of  selections for lrec  srec  or streaming data items  The user can copy any of  these lists to the scratch pad  modify individual elements in the list  then  save the scratch pad back to the original list  For more information on
108.  from paired measurement  no minimum requirement defined   6 1 5 3 12 Long term drift   no minimum requirement defined   6 1 5 3 13 Short term drift   not relevant for particulate devices   6 1 5 3 18 Overall uncertainty   not relevant for particulate devices    For this reason an official request to the competent German body was made  to define a co   ordinated procedure for dealing with inconsistencies in the Standard     740300_2013_936_21209885F_EN docx    7      TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system  Page 100 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F  The following procedure was suggested     Test points 5 3 2  5 3 7  5 3 8  5 3 11 and 5 3 12 are evaluated as before basing on the min   imum requirements stated in VDI 4202 Part 1 of 2002  i e  using the reference values Bo  B   and B3      The performance of the test points 5 3 3  5 3 4  5 3 13 and 5 3 18 is omitted  as they are not  relevant for particulate measuring devices     The competent German body agreed with the proposed procedure by decisions of 27 June  2011 and 07 October 2011   6 5 Assessment    The test is based on the minimum requirements of Standard VDI 4202  Part 1  September  2010   Test points 5 3 2  5 3 7  5 3 8  5 3 11 and 5 3 12 are evaluated 
109.  gt  RH Temperature Alarms  gt   Board Temp  gt  Min or Max     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 75    Operation  Alarms Menu       Pressure Vacuum The Pressure Vacuum Alarms submenu allows the user to view and set the  Alarms pressure alarm parameters     e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms        Barometric Pressure The Barometric Pressure screen displays the current pressure reading and    sets the minimum and maximum alarm limits  Acceptable alarm limits  range from 400 to 800 mmHg  If the barometric pressure reading goes  beyond the minimum or maximum limit  an alarm is activated and the    alarm  bell  icon appears in the status bar on the Run screen and in the  Main Menu     e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms  gt   Baro Pres        Min and Max Barometric The Minimum Barometric Pressure alarm limit screen is used to change the  Pressure Limits minimum barometric pressure alarm limit  The minimum and maximum  barometric pressure screens function the same way     e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms  gt  Baro  Pres  gt  Min or Max     3 76 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu       Vacuum The Vacuum screen displays the current vacuum reading and sets the  minimum and maximum alarm limits  Acceptable alarm limits range from   5 mmHg to 250 mmHg  If the vacuum reading goes beyond the  minimum or maximum limit  an alarm 
110.  have occurred   If no alarms are detected  the number zero is displayed     If the item being monitored goes outside the lower or upper limit  the  status of that item will go from    OK    to either    LOW    or    HIGH      respectively  If the alarm is not a level alarm  the status will go from    OK     to    FAIL     The number of alarms detected is displayed to indicate how  many alarms have occurred  If no alarms are detected  the number zero is    displayed     Items displayed are determined by the options installed  To see the actual  reading of an item and its minimum and maximum limits  move the cursor  to the item and press     If the readings go beyond either the minimum  or maximum limit  an alarm is activated and the alarm  bell  icon appears  in the status bar on the Run screen and in the Main Menu     If any alarms are active  the alarm  bell  icon is displayed on the right side  of the status bar     e Inthe Main Menu  choose Alarms     ip priri  Piet       Instrument Alarms The Instrument Alarms submenu allows the user to view a set of alarms for  different boards  The filter tape change  heater power  and bench items   motherboard  interface board and I O expansion board  are read only   The motherboard status  interface board status  and I O expansion board  status  if installed  indicate that the power supplies are working and  connections are successful  There are no setting screens for these alarms     e Inthe Main Menu  choose Alarms  gt  Instrument 
111.  heating system     The ambient particulate is then drawn down a vertically mounted heated  sample tube  The heat applied to the sampe is adjustable to either a fixed  temperature or a RH threshold at the filter spot  The purpose of heating  the sample is to reduce particle bound water and to decrease the relative  humidity  RH  of the sample stream  This is done to reduce positive  artifact measurement that can potentially occur as a result of condensation  on the filter tape or conditions of elevated humidity  The heater is  controlled by the CPU and the power applied is proportional to the  conditions     Downstream of the heated tube  is the nephelometer assembly  The  particulate travels through the insulated nephelometer in a laminar pathway  and is introduced to the radial tube upstream of the beta attenuation head   The nephelometer assembly contains an 880 nanometer pulsed light  source  a reference detector for illumination output control  a hybrid silicon  photodetector  temperature sensor  relative humidity sensor  programmable  PCB  and communications cable     The nephelometer assembly has six ranges that are automatically selected to  provide the most precise reading possible  During scheduled auto zeroing    Thermo Fisher Scientific    Thermo Fisher Scientific    Introduction  Principle of Operation    events  a rear panel mounted zeroing solenoid is activated and HEPA   filtered air is delivered to the nehpelometer  whereby each range is zeroed   The results 
112.  how  to edit the scratch pad  see the    sp field    command     The following example copies the current list of lrecs into the scratch pad     Send  set copy lrec to sp  Receive  set copy lrec to sp ok    data treatment lrec   data treatment srec   These commands report the current selection of data treatment for  concentrations in the lrecs or srecs  The following example reports the data  treatment for concentrations in lrec as minimum     Send  data treatment lrec  Receive  data treatment lrec min    set data treatment lrec string  set data treatment srec string  string     cur   avg   min   max      These commands set the data treatment to current  average  minimum  or  maximum for the concentration values recorded in the lrecs or srecs  The  following example sets the data treatment for concentrations in lrec as  minimum     Send  set data treatment lrec min   Receive  set data treatment lrec min ok   erec   erxy   x  0 1        Reply termination format  see    set format format     command     y   0 1   2    Output format  see    set erec format format    command     These commands return a snapshot of the main operating conditions   measurements and status  at the time the command is issued  The  following example shows a typical response     Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Datalogging    The format is defined within the command  in the case of erxy  by the  current settings of the  format  and  erec format  commands
113.  is possible to form daily averages based on the three measurement cycles using the de   scribed instrument configuration and a cycle duration of 8 hours     The time required for changing the filter tape amounts to a maximum of 0 21    thus lying  well below the permissible 1       Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 138 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 4 7 Constancy of sample volumetric flow    The sample volumetric flow averaged over the sampling time shall be constant within    3   of the rated value  All instantaneous values of the sample volumetric flow shall be  within a range of   5   of the rated value during sampling    6 2 Equipment   A flow meter was used for testing this requirement  as indicated in Section 4     6 3 Performance of test    The sample volumetric flow was calibrated before testing at the first field test site  checked  with a dry gas meter or a mass flow meter before testing at the other field test sites  and if  necessary readjusted     The constancy of the sample volumetric flow is monitored inter
114.  its cause  Trou   bleshooting was also delayed due to vacation of the staff and the associated communication  problems  For this reason  the measuring system could only be brought back to normal oper   ation on 6 September 2012  The cause of the failure was a faulty pressure sensor    The long outage period cannot only be attributed to the measuring system itself  since the  identification and troubleshooting of such a problem by the staff is usually performed within a  week  Therefore  this outage was taken up in the evaluation with a duration of 8 days    SN 2    SN 2 presented a malfunction of the nephelometer at the test site Bornheim  summer   The  system had to be repaired  for which the values measured from 28 July 2011 to 4 August  2011 had to be discarded    On 9 August 2011  a failure of the filter tape was observed  The cause was not identified   Errors in the stabilisation of the SHARP signal were observed from 11 September 2012 to 13  September 2012 at the test site Teddington  summer   Again  the cause could not be identi   fied    No further malfunctions of the instruments were observed    The customary maintenance works  not including zero filter operation   e g  maintenance of  the sampling inlet and check of flow rate   leak tightness  generally do not take more than  approx  1 h per day  For this reason  the pertaining daily averages were not discarded     740300_2013_936_21209885F_EN docx    a      TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland 0 A
115.  m      Model 50307 SHARP Instruction Manual B 17    C Link Protocol Commands  Alarms    B 18 Model 50307 SHARP Instruction Manual    Send  alarm conc inst neph min  Receive  alarm conc inst neph min 0 000E 00 pg m3    set alarm conc inst neph min value   set alarm conc inst neph max value   These commands set the 24 hour instant nephelometer concentration  alarm minimum and maximum values to value  where value is a floating   point representation of the concentration alarm limits  Values must be in  the units that are currently set for use  The following example sets the  instant nephelometer concentration alarm maximum value to 10000  pg m      Send  set alarm conc inst neph max 10000  Receive  set alarm conc inst neph max 10000 ok    alarm conc inst pm min   alarm conc inst pm max   These commands report the current 24 hour instant PM concentration  alarm minimum and maximum values setting  The following example  reports that the instant PM concentration minimum is 100 pg m        Send  alarm conc inst pm min  Receive  alarm conc inst pm min 1 000E 02 pg m3    set alarm conc inst pm min value   set alarm conc inst pm max value   These commands set the 24 hour instant PM concentration alarm  minimum and maximum values to value  where value is a floating point  representation of the concentration alarm limits  Values must be in the  units that are currently set for use  The following example sets the instant  PM concentration alarm maximum value to 10000 pg m        Send  se
116.  matter PM2 5   Report No   936 21209885 F    filled    The sensitivity coefficient of the  sample gas pressure at refer   ence point shall not exceed the  specifications of Table 2 of  Standard VDI 4202 Part 1   September 2010      The sensitivity coefficient of the  sample gas temperature at ref   erence point shall not exceed  the specifications of Table 2 of  Standard VDI 4202 Part 1   September 2010      The sensitivity coefficient of the  surrounding temperature at ze   ro and reference point shall not  exceed the specifications of  Table 2 of Standard VDI 4202  Part 1  September 2010      For PM     Zero point value for ATu of  15K between  5   C and  20    C or 20 K between  20   C  and  40   C shall not exceed  Bo    The measurement value in the  range of B  shall not exceed    5   for ATu of 15 K between   5   C and  20   C or for 20 K  between  20   C and  40   C     The sensitivity coefficient of the  electric voltage at reference  point shall not exceed the  specifications made in Table 2  of Standard VDI 4202 Part 1   September 2010      For PM     Change in measured value at  By  maximum Bo within the volt   age interval  230  15  20  V     The change in the measured  value caused by interfering  components in the sample gas  shall not exceed the require   ments of Table 2 of Standard  VDI 4202 Part 1  September  2010  at zero and reference  point     740300_2013_936_21209885F_EN docx    Not applicable     Not applicable     The maximum dependence on sur   yes 
117.  matter PMz 5   Report No   936 21209885 F    6 1 5 3 5 Sensitivity coefficient of sample gas pressure    The sensitivity coefficient of sample gas pressure at reference point shall not exceed  the requirements of Table 2 of Standard VDI 4202 Part 1  September 2010   A value c   at 70   to 80   of the upper limit of the certification range shall be used as reference  point     Note    This test is not relevant for particulate measuring systems   6 2 Equipment   Not applicable    6 3 Performance of test   Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 106 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 3 6 Sensitivity coefficient of sample gas temperature    The sensitivity coefficient of sample gas temperature at reference point shall not ex   ceed the requirements of Table 2 of Standard VDI 4202 Part 1  September 2010   A  value c  at 70   to 80   of the upper limit of the certification range shall be used as  reference point     Note    This test is not relevant for particula
118.  measured  component PM 3   all test sites  values 2 18 g m8    740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  TUVRheinland  Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 150 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F       Model 5030i SHARP  SN 1   SN 2  All test sites   lt 18 g m     Raw data  130                  Measured values          Regression line               y x                      Candidate 2  ug m                                                         O 10 20 30 40 50 60 70 80 90 100 110 120 130  Candidate 1  ug m               Figure 47  Results of the parallel measurements with the candidates SN 1   SN 2  measured  component PM 3   all test sites  values  lt  18 g m     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 151 of 424  Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    6 1 5 4 10 Calculation of the expanded uncertainty of the instruments    For the test of PM25  measuring systems the equivalency with the reference met
119.  meter under applica   tion of the adapter     proved to be ap   propriate parameters for monitoring  the instrument s tightness  The labor   atory tests resulted in a maximum  leakage  lt  1   of the nominal flow  rate of 16 67 l min for both candi   dates     The tightness check must only be per   formed with the tightness check  adapter and the described procedure   It can otherwise damage the instru   ment     The uncertainty between the candi   dates Ups is with a maximum of 1 68  g m  for PM25 below the required  value of 2 5 g m8     The determined uncertainties WCM  without application of correction fac   tors lay for all observed datasets be   low the defined expanded relative un   certainty Wdgo of 25   for fine dust        TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    A TUVRheinland      Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 27 of 424  Scientific for the component suspended particulate matter PMz 5     Report No   936 21209885 F    filled    5 4 11 Application of  correction factors  and terms    Requirements on  multiple   component  measuring sys   tems    If the maximum expanded un   certainty of the systems under  test exceeds the data quality  objectives according to the Eu   ropean Directive on ambient air  quality  7  for the test of PMz5  measuring systems  the appli   cation of correction factors and  te
120.  number in the DATA EREC  response  appears next  This is followed by an optional bitfield designator   The datum identified by the value source can be printed as a string  s    hexadecimal  x   decimal  d   floating point  f  or binary  b  number     Model 50307 SHARP Instruction Manual B 67    C Link Protocol Commands  Record Layout Definition    Alarm Information    Translation Table    Selection Table    Button Designator    B 68 Model 50307 SHARP Instruction Manual    Typically  bitfield extractions are only done for decimal or hexadecimal  numbers     Floating point numbers can be followed with an optional precision  specifier which will be used as an argument to printf s  f format  for  example  a field of  4  would be translated into the printf command of     3f    Alternately  the special character     specifier  this causes an indirection on the precision specifier  which now  becomes a field number        can precede the precision    This is useful when formatting  for example  numbers which have varying  precision depending on the mode of the instrument     Binary numbers can also have an optional precision specifier which is used  to determine how many bits to print  For example  the specifier  b4  will  print the lowest four bits of the parsed number     There are serious restrictions on where an  s  field may appear  currently  sources 1 and 2 must be  s   and no others may be  s        The value source is followed by optional alarm information  indicated by a 
121.  numbers SN 1 and SN 2     The test started in the year 2009 with the software version V00 05 41 114  The software was  steadily enhanced and optimised during the course of testing     At the beginning of the field test in January 2011  software version V01 00 01 197 was im   plemented  This version was used during the whole field test    The software was updated to version V01 00 03 225 after conclusion of the regular field test  and before starting the remaining laboratory tests  This software version was as well used for  the additional campaign    Bornheim  Summer 2013       In the meantime the software version was updated another time due to a discontinuation of a  processor board to version V02 00 00 232      The following software modifications were implemented              Table 4  Overview of software versions during suitability testing   Version Description of change Status  during suitability test   V00 05 41 114   Initial version Installed during la   boratory test  part 1   V01 00 00 163 1  Integration of support for Model 5030i Not installed   2  Improvement of algorithms for controlling RH  and TEMP    3  Addition of    Carriage return    to ESM protocol  4  Replacement of term    Zero    with    Offset     5  Adjustment of digital filter for    Mass    and       PM    to 60 min   6  Elimination of sample temperature from user  interface   7  Raise of limit for Temp offset values to      10   C    8  Implementation of mass determination with   out flow in servi
122.  of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Pre Spread  Sp     Post Spread  Sp s   and Blank Spread  Ssiank  were calculated using the  following equations     Spre   RM anpi     RM ang E A 4    S Post T RM End  _ RM End2 E A 5         CM sna    CM pend CM anger   CM anfi  S blank m 2 2    As with the UK PM Equivalence Report  8   it was not possible to weigh all filters within the  15 day timeframe suggested in EN14907     However  as filters were removed immediately from the reference samplers and placed in the  refrigerator  it was not necessary to determine if Tambien  exceeded 23   C  It is felt that as 15  days was impractical for a relatively small scale field study  it is less likely to be attainable if  this methodology were adopted by a National or Regional network  and as such  the meth   odology employed herein is representative of how the reference samplers would be operated  in practice     A 2 Analysis of Protocols Employed    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM 5   Report No   936 21209885 F    A TUVRheinland      Genau  Richtig     Page 217 of 424    The distributions of pre and post weight for all EMFAB filters w
123.  off the heater tube adapter    8  Slide off delrin nut upward off the heater tube adapter     9  Replace the delrin nut by following the previous steps in reverse          z    _    lt  Union Nut               Heater Tube Adapter       lt        Ferrite Bead     e Dern Nut    Figure 7 29  Replacing the Delrin Nut    7 50 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Internal SHARP  Cable Assembly  Replacement    Thermo Fisher Scientific    A    Servicing  Internal SHARP Cable Assembly Replacement    Use the following procedure to replace the case cable  Figure 7   30    Equipment Required   Internal SHARP Cable Assembly    Philips screwdriver    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF and unplug the power cord   2  Ifthe instrument is mounted in a rack  remove it from the rack     3  Refer to    Removing the Measurement Case Assembly and Lowering  the Partition Panel    in this chapter to lower the partition panel  then  proceed to the next step below     4  Disconnect the electrical connector labeled    SHARP 1      5  Unfasten the four screws     6  Pull connect up and out and install electrical connector by following  the previous steps in reverse     Model 50307 SHARP Instruction Manual 7 51    Servicing  Case Cable Replacement       SHARP 1           Screws  4     Figu
124.  offset  This command reports the current ambient RH offset  in percent  The  following example reports that the ambient RH offset is 0 6      Send  amb rh offset  Receive  amb rh offset 0 6      set amb rh offset value   This command sets the ambient RH offset to value  where value is a  floating point number representing ambient RH offset in percent  The  following example sets the ambient RH offset to 7      Send  set amb rh offset 7  Receive  set amb rh offset 7 ok    amb temp offset  This command reports the current ambient temperature offset  in degrees    C  The following example reports that the ambient temperature offset is 5  aC     Send  amb temp offset  Receive  amb temp offset 5 0 degC    set amb temp offset value   This command sets the ambient temperature offset to value  where value is  a floating point number representing ambient temperature offset in degrees  C  The following example sets the ambient temperature offset to 3   C     Send  set amb temp offset 3  Receive  set amb temp offset 3 ok    baro mass coef  This command reports the current barometric mass coefficient  The  following example reports that the barometric mass coefficient is 0 000540     Send  baro mass coef  Receive  baro mass coef 0 000540    Model 50307 SHARP Instruction Manual B 35    C Link Protocol Commands  Calibration    B 36 Model 50307 SHARP Instruction Manual    set baro mass coef value   This command sets the barometric mass coefficient to user defined values  to value  where va
125.  quality measuring sys     tem Model 50301 SHARP for suspended particles     Table 3     Technical data for Model 5030i SHARP  manufacturer s specifications        Dimensions   Weight    Model 50301 SHARP       Measuring system    425 x 584 x 219 mm   19 kg  without pump        Sampling tube    Approx  0 9 m long  isolated heated tube     1 8 m extension tube       Sampling inlet    Depends on manufacturer  during suitability  testing  Digitel DPM2 5 01 00 16       Power requirements    Power input    Analyser  100 240 V  50 60 Hz  Pump  115 V  50 60 Hz    Approximate maximum of 880 W       Ambient conditions       Temperature     5 to  40   C  during suitability testing        Humidity    Non condensing       Sample flow rate  Inlet     16 67 I min   1 m3 h                Nephelometer Source IR LED  6 mW  880 nm  Detector Hybrid silicone amplifier   Radiometry Source 146   lt 3 7 MBq   lt  100 uCi   Detector Proportional radiation detector       Temporal resolution SHARP    1 min  at data output rate of 1 s        Parameter  Filter change       Filter change  time  cycle     0h  100 h  gt Default  8 h  3 changes per day        Filter change  particle mass    0     9999 ug  gt  Default  1500 ug       Parameter  Sample heater       Max  heating temperature    30   C  during suitability testing        Nominal relative humidity    50     58    during suitability testing        Data storage capacity  internal     Max  190 000 datasets   gt  650 d at storage in   tervals of 5 min 
126.  re   calibration     Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  For more  information about appropriate safety precautions  see the    Servicing     chapter  A    Thermo Fisher Scientific    Preventive Maintenance  External Pump Exhaust Filter    1  Turn the Heater OFF  From the Main Menu  choose Instrument  Controls  gt  Set Heater  gt  Control  Use  t     until the value  reads OFF  and press         2  Disconnect RH temperature connector     3  From the SHARP MIB Board J1  remove the five socket heads from  the top end cap and lift off     4  Remove sealing gasket  note  may be stuck to bottomside of top end  plate      5  Carefully vacuum out optic chamber     6  Re assemble in reverse order   note careful placement of the sealing    gasket        Figure 5   6  Replacing the External Pump Exhaust Filter    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 5 9    Preventive Maintenance  Cleaning the SHARP Optics Assembly    Cleaning the  SHARP Optics  Assembly    A    5 10 Model 50307 SHARP Instruction Manual    Use the following procedure to inspect and clean the SHARPoptics  assembly  On an annual basis the SHARP optics assembly should be  removed and cleaned     Equipment Required   Philips screwdriver  Adjustable wrench    Hex wrench    Equipment Damage Some internal components can be damaged by small  amoun
127.  resolution of 11 bits     e 0 20mA  e 4 20mA    The user can calibrate each analog output zero and span point through the  firmware  At least 5  of full scale over and under range are also supported   but may be overridden in the firmware  if required     The analog outputs may be assigned to any measurement or diagnostic  channel with a user defined range in the units of the selected parameter   The current outputs are independent of the voltage outputs  The current  outputs are isolated from the instrument power and ground  but they share  a common return line  Isolated GND      The optional I O expansion board includes eight analog voltage inputs   These inputs are used to gather measurement data from third party devices  such as meteorological equipment  The user may assign a label  unit  and a  conversion table  2 to 10 points   Each point in the conversion table  consists of an analog input voltage value  0 10 5 V  and a corresponding  user defined reading value  Only two points are necessary for linear inputs   however  a larger number of points may be used to approximate non linear  inputs  All voltage inputs have a resolution of 12 bits over the range of 0 to  10 volts     The instrument includes one power fail relay on the motherboard and ten  digital output relays on the digital output board  These are reed relays rated  for at least 500 mA   200 VDC     The power fail relay is Form C  both normally opened and normally closed  contacts   All other relays are Form 
128.  rounding temperature in the range 5     C to 40   C was 0 2 g m  at zero    No deviation  gt   2 6   of the default   value at 20   C could be observed at  reference point     No deviations  gt   1 4   in relation to  the default value at 230 V caused by  voltage changes could be detected for  PMes     Not applicable  i       TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 23 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    filled    5 3 10 Averaging effect   For gaseous components the   Not applicable   measuring system shall allow  the formation of hourly averag   es  The averaging effect shall  not exceed the requirements of  Table 2 of Standard VDI 4202  Part 1  September 2010      5 3 11 Standard devia   The standard deviation from  The reproducibility for PM2 5 was 10  tion from paired   parallel determinations shall be   for the complete dataset of the field  measurement done with two identical meas    test    uring devices during field test   It shall not exceed the specifi   cations stated in Table 2 of  Standard VDI 4202 Part 1   September 2010      For PM   RD   10 related to B4     5 3 12 Long term drift The long term drift at zero point   The maximum deviation at zero point  and reference point s
129.  s  s 41x  f SF  f SF  f  f  f  f SF  lx   1x    t D L fftttttff A A  flags pm baro vac pflow ambrh srh ambtemp stemp fvol  dflg aflg    lrec mem size   srec mem size   These commands report the number of Irecs and srecs that can be stored  with the current settings and the number of blocks reserved for lrecs and  srecs  The following example shows that 1075 blocks were reserved for lrecs  and the maximum number of Irecs that can be stored in memory is  241979  Memory allocation can be changed using the    malloc    command     Send  lrec mem size   Receive  lrec mem size 241979 recs  1075 blocks  lrec per   srec per    Model 50307 SHARP Instruction Manual B 31    C Link Protocol Commands  Datalogging    B 32 Model 50307 SHARP Instruction Manual    These commands report the lrecs and srecs logging period  The following  example shows that the srec logging period is 5 minutes     Send  srec per  Receive  srec per 5 min    set lrec per value  set srec per value    value   1 5 15 30 60     These commands set the lrecs and srecs logging period to value in minutes   The following example sets the lrec logging period to 15 minutes     Send  set lrec per 15  Receive  set lrec per 15 ok  malloc lrec   malloc srec    These commands report the currently set memory allocation for lrecs and  srecs in percent of total memory     Send  malloc lrec  Receive  malloc lrec 10      set malloc lrec value  set malloc srec value  value   0 to 100    These commands set the percent of memory space a
130.  selection list options     L   Indicates a button which pops up a selection list without any  translation  The output value is the number of the selected option     T   Indicates a button which pops up a selection list with output  translation  The number of the option selected is used as an index into  the translation table to generate an output string     N   Indicates a button which only sends the subsequent command to  the instrument  No user prompting happens     The following string through an optional         or the end of the line is the  command which is to be sent to the instrument upon the completion of the  button selection  The command string should normally contain print style  formatting to include the user input  Ifa         is present  it indicates a  command which is sent to the instrument upon successful completion of  the button command to update the value field     Some examples    n  is the C syntax for an end of line character       Concentrations n   This is a single text only line   1  n 1    This is a single blank line       NO 3s n     This is a line which appears slightly indented  The text field is  NO   the  value is taken from the third element of the data response  and interpreted  as a string       NO 18sBd ddd set no coef  s n     This is a line which also appears slightly indented  The next field is also   NO   but the value is taken from the eighteenth element of the data  response  again interpreted as a string  A button appears on this l
131.  shows the user interface with the current concentration values     0       NEFT SMPL 12  to  RANGERRBAUGE       DLAGSBALARM    Figure 32  Display of the measured SHARP concentration    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 77 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 4 1 2 Easy maintenance    Necessary maintenance for the measuring systems should be possible without larger  effort  if possible from outside     6 2 Equipment  No additional equipment required   6 3 Performance of test    Necessary regular maintenance works were carried out according to the instructions of the  manual     6 4 Evaluation    The following maintenance works must be performed by the user     1  Check of system status    The system status can be monitored and controlled directly on the system itself or  online    2  In general  the sampling inlet must be cleaned according to the manufacturer   s in   structions while taking into account the local suspended particle concentrations  dur   ing suitability test every 4 weeks     3  Inspection of filter tape supply     one filter tape is enough for approx  650 filter tape  changes per roll  approx  215 days at cycles of 8 hours   S
132.  status lines and to  output analog voltages that represent the measurement data  Connectors  located on the motherboard include     Thermo Fisher Scientific    Nephelometer  Interface Board    Nephelometer interface  board connectors    Measurement Interface  Board    Thermo Fisher Scientific    System Description  Electronics    External connectors include   e External Accessory  e RS 232 485 Communications  two connectors   e Ethernet Communications    e I O connector with Power Fail Relay  16 Digital Inputs  and 6  Analog Voltage Outputs     Internal connectors include   e Function Key Panel and Display  e Measurement Interface Board  e I O Expansion Board  e Digital Output Board  e AC Distribution    The nephelometer interface board serves as the circuitry for the  nephelometer measurement  an internal valve control  and on board  processing of zero and particle concentration by photometry     Connectors located on the nephelometer interface board include   e RS485  e 24 VDC Supply  e LED Source Current  e Reference Detector  e Scattering Detector    e 1 RH Sensor    The measurement interface board serves as a central connection area for all  measurement electronics in the instrument  It contains power supplies and  interface circuitry for sensors and control devices in the measurement  system  It sends status data to the motherboard and receives control signals  from the motherboard     Model 50307 SHARP Instruction Manual 8 7    System Description  Electronics    Measu
133.  the results of the zero and span drift test described in point 6 1 5 3 12 Long term drift  were taken into account for the determination of the maintenance interval     6 4 Evaluation    No drifts beyond the permissible limits were detected during the entire field test period   Therefore  the maintenance interval is determined by the specified maintenance procedures   see module 4 1 2      During operation  the maintenance works can be limited to checks of contamination  plausi   bility and status error messages     6 5 Assessment    The maintenance interval is defined by the necessary maintenance procedures and has  been set to 1 month     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results    The necessary maintenance works can be found in module 4 1 2 of this report and in chapter  5 of the instruction manual     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 95 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 2 8 Availability    The availability of the measuring system shall be determined during the field test and  shall be at least 95       6 2 Equipment  No additional equipment required   6 3 Performance of test    The start and end t
134.  the specifica   tions of the guide    PMbs  30 g m8    Reference is made to test point 6 1 5 4 10 Calculation of the expanded uncertainty of  the instruments of this report        Evaluation period         6 5 Assessment    It is possible to assess the measuring system in the range of the relevant limit values   Minimum requirement fulfilled  yes    6 6 Detailed presentation of test results    Please refer to test point6 1 5 4 10 Calculation of the expanded uncertainty of the in   struments of this report     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 89 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 2 2 Measuring range    The upper limit of the measuring range shall be greater or equal to the upper limit of  the certification range     6 2 Equipment  No additional equipment required   6 3 Performance of test    It was tested  whether the upper limit of the measuring range of the measuring system is  greater or equal to the upper limit of the certification range     6 4 Evaluation    As a standard  a measuring range of 0     10 000 g m    is adjusted on the measuring system     As appropriate default setting of the analogue output for European conditions a measur
135.  the status bar on the  Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Inst PM        Min and Max Instant PM The Minimum Instant PM Concentration alarm limit screen is used to    Concentration Limits    3 82  Model 50307 SHARP Instruction Manual    change the minimum instant PM concentration alarm limit  The    Thermo Fisher Scientific    Operation  Alarms Menu    minimum and maximum instant PM concentration screens function the  same way     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Inst PM  gt  Min  or Max           Cp bebe       Instant SHARP The Instant SHARP screen displays the current instant SHARP  concentration and sets the minimum and maximum alarm limits   Acceptable alarm limits range from  10 to 10000 g m  or   01 to 10  mg m     If the instant SHARP concentration goes beyond either the  minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Inst SHARP        Min and Max Instant The Minimum Instant SHARP Concentration alarm limit screen is used to  SHARP Concentration change the minimum instant SHARP concentration alarm limit  The  Limits minimum and maximum instant SHARP concentration screens function  the same way     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Inst SHARP  gt   Min or Max     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 
136.  threshold   or temp threshold   The following example reports that the heater is RH  threshold     Send  ht control  Receive  ht control RH    set ht control selection  selection     off   rh   temp      Model 50307 SHARP Instruction Manual B 47    C Link Protocol Commands  Measurement Configuration    B 48 Model 50307 SHARP Instruction Manual    This command sets the heater to off  RH threshold  or temp threshold   The following example sets the heater to temp threshold     Send  set ht control temp  Receive  set ht control temp ok  int time    This command reports the current integration time of the calculated mass  and concentration  The following example reports the integration time is  15 minutes     Send  int time  Receive  int time 15 min    set int time value   This command sets the integration time to value  where value can be 15   20  30  40  45  or 60 minutes  The following example sets the integration  time to 20 minutes     Send  set int time 20  Receive  set int time 20 ok  mass limit    This command reports the current filter mass limit  The following example  reports that the mass limit is 1 500 mg     Send  mass limit  Receive  mass limit 1500    set mass limit value   This command sets the filter mass limit to value  where value is a floating   point number representation of mass limit between 0 5 to 5 0 mg or 500 to  5 000 pg  The following example sets the mass limit to 1 0 mg m        Send  set mass limit 1 0  Receive  set mass limit 1 0 ok  pres comp   
137.  to the RH sensor  Furthermore  try to avoid RH comparisons   lt 30  RH and  gt 80  RH  a    Model 50307 SHARP Instruction Manual 2 7    Installation  Acceptance Testing and Startup Procedures    If the RH sensors are slightly out of tolerance   5  RH  the acceptance test  should be classified as marginal  If the sensor performance is outside this  range  please perform a RH sensor calibration     For more information about the RH sensor calibration  see the     Calibration    chapter     It is recommended that the NIST traceable hygrometer should also  compare well with the RH measurement used within a gravimetric  laboratory that is part of a compliance program     One Point Barometric Record the Model 5030 barometric pressure sensor reading  This value is   Pressure Verification in wnits of mmHg  If necessary  using the pressure unit conversion  Table  2 1  to convert your NIST traceable measurement to units of mmHg for  an appropriate comparison  The Model 5030  barometric sensor should  compare within  10 0 mmHg tolerance of your NIST traceable barometer   If the Model 50302 sensor performance is within this tolerance  the  acceptance test has passed     If the sensor is slightly out of tolerance   12 mmHg  the acceptance test  should be classified as marginal  If the sensor performance is outside this  range  please perform a sensor calibration     For more information about the barometric pressure sensor calibration  see  the    Calibration    chapter     One Point Vol
138.  using the controlled variable rH       rel   ambient humidity of the sample     gt  calculated from the rel  humidity of ambient air  ambient  temperature and sample temperature  and the filter temperature sensor  The heating condi   tions can be adjusted according to the requirements of each measurement network     The parameters were adjusted as follows during suitability testing   Rel  air humidity threshold  50   Filter temperature sensor  max  30   C    This means that the heater turns on as soon as the relative humidity rH rises above 50     However  the filter temperature does not exceed 30   C  The relative humidity threshold value  of 50   was specified by the manufacturer  since it represents the filter conditioning value of  the gravimetric reference method     None of the slopes or offsets determined from the comparison with the reference method  were significantly visible during the comparison campaigns at the test sites Cologne  winter    Bornheim  summer  and Bornheim  winter   All expanded measurement uncertainties calcu   lated according to the Guide    Demonstration of Equivalence of Ambient Air Monitoring Meth   ods     4  lay below the maximum permissible value of 25   without application of correction  factors  After concluding the measurements in Germany  the final comparison campaign  started in summer of 2012 at the test site Teddington  During this campaign  an underestima   tion of the suspended particulate content measured by the candidates was obs
139.  visible only when the instrument is in service mode   For more information on the service mode  see    Service Mode    earlier in  this chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Analog Out Calibration        The Analog Output Calibrate Zero screen allows the user to calibrate the  zero state of the selected analog output  The operator must connect a volt  meter to the output and adjust the output until it reads 0 0 V on the meter  for a voltage channel  or either 0 or 4 mA for a current channel  depending  on the selected range   See the    set output to     field on the display     e Inthe Main Menu  choose Service  gt  Analog Out Cal  gt  select    Channel  gt  Calibrate Zero     Model 50307 SHARP Instruction Manual 3 99    Operation  Service Menu       ip pror  PIL          Analog Output Calibrate The Analog Output Calibrate Full Scale screen allows the user to calibrate  Full Scale the full scale state of the selected analog output  The operator must connect  a volt meter to the output and adjust the output until the meter reads the  value shown in the    set output to     field  in either V or mA  depending on  the selected output channel     e Inthe Main Menu  choose Service  gt  Analog Out Cal  gt  select  Channel  gt  Calibrate Full Scale        Cu ied  Piket iini i          Analog Input Calibration The Analog Input Calibration menu is used to calibrate the 8 analog inpu
140.  zero occur     e Inthe Main Menu  choose Calibration Factors  gt  Neph Bkg  gt   Restore Prev Values     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 17    Operation  Instrument Controls Menu    BETES E Saeed Beane    Nephelometer  Coefficient    Instrument Controls  Menu    3 18 Model 50307 SHARP Instruction Manual    a RMP E SP e  4ribbo ob      OEH pikari  A a E a E       ERED hy Pa Tate ip kobe    The Nephelometer Coefficient screen allows the user to manually adjust  the Nephelometer coefficient  The Nephelometer concentration based on  the SET COEF TO is displayed to facilitate setting  The Nephelometer  concentration displayed in the Run portion of the screen is NOT affected  by the SET COEF until the value is saved     e Inthe Main Menu  choose Calibration Factors  gt  Neph Coef        DH Guided  fot PEL PEPE i          The Instrument Controls menu contains a number of items that may be  selected to control various instrument operational parameters  The  firmware controls listed in this menu enable control of the listed  instrument functions     e Inthe Main Menu  choose Instrument Controls     Thermo Fisher Scientific    Operation  Instrument Controls Menu       prek prnpr Poh dm PP i  PPP PLES Piel bab i    WH nid       Set Flow Pump The Set Flow Pump menu allows the user to manually adjust the flow  values  and to change the operating state of the pump to ON or OFF     e Inthe Main Menu  choose Instrument Controls  gt  Set Flow Pump        WH mh
141. 0    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 11 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    Figure 40  Flow of candidate SN 2  field  2 0 2 2       ceeeeccceceeeeeeeeeeeeeeneeeeeeeeeeeeeeenseaaaaeeeeeeeees 140  Figure 41  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PMos  all test SITES          ccc cccecccccceeeeeeeeeeeeeeeeeeeeeeeeeeees 147  Figure 42  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PM25  test site Cologne  winter                    ceeeeeeeeeees 147  Figure 43  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PMos  test site Bornheim  SUMMe                ccccceeeeeeees 148  Figure 44  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PMos  test site Bornheim  winter                cccccccceeeeees 148  Figure 45  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PMbs  test site Teddington  summer                  eeee 149  Figure 46  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PMs    all test sites  v
142. 0 10 1 10  Beta Cnts Alarm 0  0 0 0 0 1 0 J0  Neph Brd Alarm 0  0 0 J0 1  0 0  0  Detector Board Status Alarm 0 10 0 l1 0  0 0  0  Motherboard Status Alarm 0  0 T 0 0  0 00  Flow Alarm 0 J1 0 J0 0 10 0 J0  Nephelometer Reference Voltage   1 0 0 10 0 0 0 10  Alarm                                     Thermo Fisher Scientific    Appendix E  ESM Protocol Commands    This appendix provides a description of the ESM Protocol commands   from the prior FH62 platform  that can be used to remotely control a  Model 50303 instrument using a host device such as a PC or a datalogger   ESM protocol may be used over RS 232  RS 485 or over Ethernet  ESM    functions can be accessed over Ethernet using TCP port 9884     Up to three simultaneous connections per protocol may be made over  Ethernet     ESM Commands fable F  through Table E   3 lists the ESM commands supported for the    Supported Model 50303   Table E   1  Read Commands for 50307                                                    Command Description   C3 PM conc always always jig m3   C2 Neph conc g m    C SHARP conc pg m      H3 Avg PM conc always yg m3   H2 Avg Neph conc yg m3   HT Avg SHARP conc ug ms   JB temperature sampling head   ambient temp   JD temperature inside of the flow meter orifice   flow temp  JF different pressure of the air flow measuring module in Pa  JG low pressure suction chamber   JH barometer in hPa   Jl air flow sampling head volumetric flow   JJ norm air flow  273K  1013hPa    JR relative humidity value
143. 0 4 26 2011 17 3 1 9 11 0 16 7 23 4 1011 51 5 1 8 166 2 1  51 4 27 2011 7 5 10 8 12 5 1010 90 4 0 7 213 8 9  52 4 28 2011 17 0 5 4 31 5 14 2 20 5 1005 77 6 0 7 176 0 3  53 4 29 2011 19 2 2 8 14 7 ieee 24 9 1002 56 8 EF 112 3 0  54 4 30 2011 13 1 2 5 19 0 16 9 24 2 1002 47 4 17 141 0 0  55 5 1 2011 6 8 2 3 32 9 14 8 22 4 1002 44 5 1 6 111 0 0  56 5 2 2011 9 3 21 23 0 11 0 17 8 1004 53 3 2 0 116 0 0  57 5 3 2011 9 2 22 24 4 10 0 17 2 1011 49 4 1 0 164 0 0  58 5 4 2011 11 4 29 25 2 9 7 16 2 1016 61 5 1 3 168 0 0  59 5 5 2011 3 8 14 1 19 8 1015 46 9 2 2 119 0 0  60 5 6 2011 13 6 2 8 20 3 18 6 24 8 1012 41 1 2 5 110 0 0                                                   volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx             TUV Rheinland Energie und Umwelt GmbH A TUVR heinland      Luftreinhaltung      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 195 of 424  Report No   936 21209885 F                            Annex 6 Ambient conditions at field test sites Page 3 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation  m5  ym3   A  C  C  hPal A  m s mm   61 5 7 2011 Colog
144. 00 pg m      The suitability test was carried out in accordance with the current standards for suitability  tests while taking into account the latest developments   The test was performed in consideration of the following standards     e Standard VDI 4202 Part 1     Performance criteria for performance tests of automated  ambient air measuring systems     Point related measurement methods for gaseous  and particulate air pollutants     September 2010 or June 2002  1     e Standard VDI 4203 Part 3     Testing of automated measuring systems     Test proce   dures for point related ambient air measuring systems for gaseous and particulate air  pollutants     September 2010 or August 2004  2     e Standard EN 14907     Ambient air quality   Standard gravimetric measurement meth   od for the determination of the PM25 mass fraction of suspended particulate matter      German version EN 14907  2005  3     e Guide    Demonstration of Equivalence of Ambient Air Monitoring Methods     English  version of January 2010  4     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 29 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    3  Description of the AMS tested    3 1 Measuring principle    T
145. 01    Operation  Service Menu    Display Pixel Test    The Display Pixel Test screen is used to test the LCD display  by toggling  between all pixels on and all pixels off to ensure that they are functioning  properly  The display pixel test screen is visible only when the instrument is  in service mode  For more information on the service mode  see    Service  Mode    earlier in this chapter     e Inthe Main Menu  choose Service  gt  Display Pixel Test                    Restore User Defaults The Restore User Defaults screen is used to reset the user calibration and    configuration values to factory defaults  The restore user defaults screen is  visible only when the instrument is in service mode  For more information  on the service mode  see    Service Mode    earlier in this chapter     e Inthe Main Menu  choose Service  gt  Restore User Defaults        Restore Factory The Restore User Defaults screen is used to reset the user calibration and  Defaults configuration values to factory defaults  The restore user defaults screen is    3 102 Model 50307 SHARP Instruction Manual    visible only when the instrument is in service mode  For more information  on the service mode  see    Service Mode    earlier in this chapter     e Inthe Main Menu  choose Service  gt  Restore User Defaults     Thermo Fisher Scientific    Operation  Password Menu       Password Menu The Password menu allows the user to configure password protection  If  the instrument is    locked     none of the 
146. 06 4 7  87 8 18 2011 8 8 2 8 31 5 22 5 31 3 1004 76 6 1 2 174 20 9  88 8 19 2011 6 4 1 8 21 16 8 21 2 1011 80 0 1 5 235 3 3  89 8 20 2011 1 7 20 7 28 1 1011 66 6 0 8 157 0 0  90 8 21 2011 10 3 1 2 11 9 23 2 31 4 1007 74 8 1 0 184 0 3                                                         volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx    ss      TUV Rheinland Energie und Umwelt GmbH  A TUV Rheinland    Luftreinhaltung    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 196 of 424 Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F                                                                      Annex 6 Ambient conditions at field test sites Page 4 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ug m3   ug m3          C     C   hPa       m s  ul  mm   91 8 22 2011 Bornheim  Summer 10 9 3 2 29 2 20 4 24 4 1009 76 5 1 2 253 0 0  92 8 23 2011 19 1 2 9 15 4 22 6 27 8 1005 78 4 0 9 206 0 0  93 8 24 2011 GA 1 8 25 3 20 1 27 2 1007 76 6 0 7 192 0 6  94 8 25 2011 11 8 3 1 26 0 20 8 27 6 1003 83 4 1 0 176 21  95 8 26 2011 5 2 1 7 32 1 19 4 30 2 999 83 7 1 5 195 29 1  96 8 27 2011 0 7 15 3 20 0 1007 77
147. 1    C Link Protocol Commands  Hardware Configuration    B 52 Model 50307 SHARP Instruction Manual                                                    Level Contrast Level  5 25   6 30   7 35   8 40   9 45   10 50   11 55   12 60   13 65   14 70   15 75   16 80   17 85   18 90   19 95   20 100   date    This command reports the current date  The following example reports the    date as April 1  2009     Send  date  Receive  date 04 01 09    set date mm dd yy  mm   month    dd   day   yy   year   This command sets the date of the analyzer   s internal clock  The following  example sets the date to May 1  2009     Send  set date 05 01 09  Receive  set date 05 01 09 ok    set default params    This command sets all the parameters to their default values  This does not  affect the factory calibrated parameters     Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Hardware Configuration    Send  set default params  Receive  set default params ok  det    This command reports the status of the detector board   sharp beta map none   The following example reports that the detector is  beta     Send  det  Receive  det beta  pump    This command reports the current status of pump as on or off  The  following example reports that the pump is set to off     Send  pump  Receive  pump off    set pump onoff  onoff     on   off      This command sets the pump on or off  The following example sets the  pump to on     Send  set pump on  Receive  set pump on ok  sa
148. 1 22 4 21 2  185 1 26 2012 22 7 22 3 29 3 27 6 79 0 24 3 24 3  186 1 27 2012 17 0 16 0 24 1 24 3 68 2 16 0 16 4  187 1 28 2012 18 1 19 2  188 1 29 2012 64 0 62 4 69 2 69 8 90 9 69 8 71 5  189 1 30 2012 61 2 60 2 71 2 71 6 85 0 66 5 67 4  190 1 31 2012 36 5 36 6 44 2 42 8 84 0 38 3 38 4  191 2 1 2012 25 3 25 1 31 6 31 4 80 0 26 0 27 6  192 2 2 2012 20 1 20 2 25 6 25 3 79 2 23 1 21 8  193 2 3 2012 29 0 28 9 37 6 36 7 77 9 27 9 28 3  194 2 4 2012 34 1 34 8  195 2 5 2012 24 1 25 6 31 3 30 1 81 1 26 9 27 8  196 2 6 2012 31 9 32 4 41 3 40 4 78 7 38 2 38 3  197 2 7 2012 25 2 25 7 37 6 35 5 69 7 28 8 28 6  198 2 8 2012 33 4 34 4 43 6 42 7 78 6 37 7 40 1  199 2 9 2012 30 1 32 6 38 7 37 7 82 0 33 2 34 6  200 2 10 2012 Inlet   gt  Zero point  201 2 11 2012 Zero point  202 2 12 2012 Zero point  203 2 13 2012 39 6 36 0 41 9 41 0 91 1 37 2 37 4  204 2 14 2012 10 3 9 8 16 6 15 7 62 3 11 4 11 2  205 2 15 2012 TA 6 7 14 7 15 0 46 6 8 5 8 4  206 2 16 2012 15 6 14 2 20 8 21 9 69 7 17 1 15 9  207 2 17 2012 9 7 9 1 11 3 11 9 80 8 10 3 8 6  208 2 18 2012 3 4 3 6  209 2 19 2012 9 9 10 4  210 2 20 2012 10 3 10 6 16 6 17 6 61 0 9 9 9 9  740300_2013_936_21209885F_EN docx       TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25     Report No   936 21209885 F    A TUVRheinland      Pre
149. 10 2012 Audits  9 11 2012       SN2 stabilisation problems SHARP  9 12 2012   SN2 stabilisation problems SHARP  9 13 2012   2   SN2 stabilisation problems SHARP  9 14 2012  9 15 2012  9 16 2012  9 17 2012  9 18 2012  9 19 2012  9 20 2012  9 21 2012  9 22 2012  9 23 2012  9 24 2012  9 25 2012  9 26 2012  9 27 2012  9 28 2012                         740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    TUVRheinland      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Page 192 of 424    Annex 5 Measured values from field test sites  related to actual conditions Page 11 of 11       Manufacturer    Type of instrument    Thermo Fisher Scientific    Model 5030i SHARP    PM2 5    Measured values in g m   ACT                                         Serial No  SN1 SN2  No  Date Ref  1 Ref 2  Ratio SN 1 SN2 Remark Test site  PM10 PM10 PM2 5 PM10 PM2 5 PM2 5   m   m  gt  g ms  m   301 9 29 2012 8 3 8 2 8 4 6 8 Teddington  Summer  302 9 30 2012 4 8 4 6 5 0 5 2  303 10 1 2012 4 0 3 9 3 9 4 2  304 10 2 2012 3 7 3 7 3 8 4 2  305 10 3 2012 3 6 3 4 2 4 2 6  306 10 4 2012 3 5 3 5 2 8 4 2  307 10 5 2012 3 3 3 3 4 3 4 8  308 10 6 2012 141 5 11 2 12 3 13 3  309 10 7 2012 16 7 16 2 21 1 16 7  310 10 8 2012 15 4 15 4 17 4 17 5  311 10 9 2012 8 3 8 5 5
150. 116 9 16 2011 13 0 13 7 23 0 25 1 55 4 Inlet   gt  Zero filter   117 9 17 2011 Zero point   118 9 18 2011 3 2 3 9 7 0 7A 50 4 Zero point   119 9 19 2011 7 8 8 2 12 5 11 6 66 2 Zero filter   gt  Inlet   120 9 20 2011 3 7 6 9                               740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     Page 186 of 424    Annex 5    Measured values from field test sites  related to actual conditions    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Page 5 of 11       Manufacturer  Type of instrument    Serial No     Thermo Fisher Scientific    Model 50301 SHARP    SN1 SN2       PM2 5  Measured values in g m   ACT              No  Date Ref  1 Ref  2 Ref  1 Ref 2  Ratio SN 1 SN 2 Remark Test site   PM2 5 PM2 5 PM10 PM10 PM2 5 PM10 PM2 5 PM2 5   m   m3  m3  m3    m3  m    121 9 21 2011 12 4 12 3 6 5 Bornheim  Summer   122 9 22 2011 6 4 7 8 19 2 18 9 37 3 9 8   123 9 23 2011 12 2 13 4 26 1 26 2 49 1 16 5   124 9 24 2011   125 9 25 2011 15 7 14 5 21 3 21 7 70 0   126 9 26 2011 18 8 20 6   127 9 27 2011 38 3 39 8   128 9 28 2011   129 9 29 2011 17 1 16 0   130 9 30 2011 12 4 11 8 23 4 24 5 50 6   131 10 1 2011   132 10 2 2011   133 10 3 2011 13 5 14 8   134 10 4 2011 9 8 9 8 15 9 16 3 60 9   135 10 5 2011 4 
151. 12 22 5 6 2 27 7 2 6 4 0 1010 78 6 4 1 127 0 6  186 1 27 2012 16 5 4 0 24 5 2 3 7 4 1016 85 6 1 1 124 0 0  187 1 28 2012 3 2 1 6 4 4 1021 81 0 0 9 166 0 3  188 1 29 2012 63 2 6 6 10 4  0 4 0 3 1020 81 2 1 4 238 0 0  189 1 30 2012 60 7 6 9 11 4  0 1 0 9 1017 75 8 1 5 110 0 0  190 1 31 2012 36 5 5 8 15 9  2 7 0 9 1018 62 8 2 0 116 0 0  191 2 1 2012 25 2 5 2 20 7  5 2 0 7 1023 53 6 22 144 0 0  192 2 2 2012 20 2 3 8 18 6  7 0  0 8 1026 50 2 1 5 186 0 0  193 2 3 2012 28 9 5 9 20 3  8 5  4 0 1031 69 0 1 0 252 0 0  194 2 4 2012 8 2  8 5  3 7 1031 69 8 1 0 182 0 0  195 2 5 2012 24 9 4 9 19 8  7 9  3 6 1027 57 9 tT 142 0 0  196 2 6 2012 32 2 4 8 15 0  8 9  3 6 1029 57 8 1 1 152 0 0  197 2 7 2012 25 5 4 3 16 8  7 6  4 3 1031 61 6 1 8 148 0 0  198 2 8 2012 33 9 5 4 15 8  5 7 0 0 1030 68 4 0 8 187 0 0  199 2 9 2012 31 3 6 7 21 4  5 4  2 2 1030 81 1 1 6 259 0 0  200 2 10 2012  5 7  0 3 1031 64 9 ti 116 0 6  201 2 11 2012  7 6  2 1 1028 67 2 tet 198 0 0  202 2 12 2012  4 2  1 5 1024 69 3 0 9 233 0 0  203 2 13 2012 37 8 8 7 23 1 0 4 1 9 1012 89 7 0 7 208 21  204 2 14 2012 10 1 3 1 5 0 1006 88 5 4 4 280 4 2  205 2 15 2012 6 9 3 5 5 3 1008 81 1 5 7 303 0 6  206 2 16 2012 14 9 4 2 5 3 1016 90 4 1 2 246 cam  207 2 17 2012 9 4 5 6 TA 1013 87 7 At 229 12  208 2 18 2012 1 9 6 5 9 4 1005 81 8 2 5 213 1 5  209 2 19 2012 0 4 1 5 48 1017 84 7 2 9 270 0 9  210 2 20 2012 10 5 2 3 21 5 1 0 5 1 1026 76 4 0 8 179 0 0                                                         volatile PM compounds  determined 
152. 2  Compress the collar of the quick connect fitting and remove the tubing  from fittings     3  Install filter assembly by following the previous steps in reverse     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 53    Servicing  Service Locations         lt  Push in to release    Figure 7 32  Replacing the External Zeroing Filter    Service Locations For additional assistance  service is available from exclusive distributors  worldwide  Contact one of the phone numbers below for product support  and technical information or visit us on the web at www thermo com aqi     1 866 282 0430 Toll Free  1 508 520 0430 International    7 54 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Chapter 8  System Description    This chapter describes the function and location of the system components   provides an overview of the firmware structure  and includes a description  of the system electronics and input output connections and functions as  follows     e    Hardware    on page 8 1  e    Firmware    on page 8 5  e    Electronics    on page 8 6    e    I O Components    on page 8 9    Hardware Model 5030  hardware  Figure 8 1  includes   e Nehpelometer assembly  e Primary measurement head  e Detector amplifier  e Cam photo interrupt board assembly  e Counter wheel interrupt board assembly  e Proportional valve  e Cam motor  e Tape motor  e Pressure board  e External solenoid    e Pump intake T fitting    Thermo Fisher Scientific Model 50307 SHARP Instr
153. 20002 10  11 12 2008 4 0 20002 0  17 12 2008 5 0 20003 10  18 12 2008 6 0 20002  10  07 01 2009 7 J  10  08 01 2009 8 0  14 01 2009 9 0 20000  10  15 01 2009 10 10  21 01 2009 11 0  22 01 2009 12 0  29 01 2009 13 0  30 01 2009 14 0 20000  10  04 02 2008 15 10  05 02 2009 16 0  11 02 2009 17 0  12 02 2009 18 0 20000  10  18 02 2009 19 0 20000 0  19 02 2009 20 0 20000 0  26 02 2009 21 0 20000 0  27 02 2009 22 0 19999  10    Marked in yellow   average value  Marked in green   lowest value  Marked in blue   highest value    740300_2013_936_21209885F_EN docx    TUVR heinland   TUV Rheinland Energie und Umwelt GmbH  Precisely Right Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system   Page 212 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F                                              Stability control filter Emfab   0 09280   0 09260 LESSER wwsator lt estos   0 09240   TMi    0 09220 A TM2    0 09200    TM3    0 09180 x Mean TM1    0 09160 a ee Oe ee agt 6 ee eee ee o Mean TM2   0 09140 e Mean TM3   0 09120 i dd   0 09100   0 5 10 15 20 25  No  of weighing             Figure 62  Stability of the control filters    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH H        Luftreinhaltung A TUVRheinland    sani E      Genau  Richtig   Report on the suitability test of the ambient air quality 
154. 2011 20 4 5 5 27 1 5 3 13 2 1027 57 7 0 9 150 0 0  14 3 21 2011 22 3 5 2 23 3 6 9 16 1 1029 56 5 1 0 166 0 0  15 3 22 2011 41 6 10 7 25 7 9 4 17 2 1031 62 7 Hal 184 0 0  16 3 23 2011 20 4 7 6 37 1 10 7 18 6 1030 66 8 1 2 161 0 0  17 3 24 2011 19 4 7 1 36 7 10 9 18 6 1021 67 2 1 0 174 0 0  18 3 25 2011 27 5 9 4 34 0 11 8 18 0 1010 59 4 1 6 183 0 0  19 3 26 2011 5 0 7 7 11 6 1010 64 8 1 5 105 0 0  20 3 27 2011 24 7 6 6 26 5 9 3 16 3 1006 60 9 tt 196 0 0  21 3 28 2011 20 6 6 7 32 4 7 2 13 7 1009 60 2 1 9 172 0 0  22 3 29 2011 44 4 13 9 31 4 9 6 18 5 1007 62 1 V1 168 0 0  23 3 30 2011 15 6 6 9 44 6 12 6 15 9 1008 66 7 2 4 170 0 0  24 3 31 2011 5 5 2 5 44 8 13 8 15 6 1011 78 2 3 7 230 6 5  25 4 1 2011 8 1 3 3 40 7 13 9 18 8 1014 78 1 2 3 175 0 0  26 4 2 2011 4 3 17 6 24 3 1006 62 2 2 6 159 0 0  27 4 3 2011 14 2 4 8 34 0 10 9 15 8 1009 85 3 2 0 251 87  28 4 4 2011 8 9 3 3 37 3 10 0 15 0 1017 65 3 aT 214 0 0  29 4 5 2011 11 2 3 8 34 4 11 8 15 1 1020 71 9 21 173 0 9  30 4 6 2011 13 0 4 4 34 0 16 2 23 0 1019 73 9 1 8 196 0 0                                                            volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx    A TUVRheinland      Page 194 of 424    Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 212098
155. 24 hour  averaging time     e Inthe Main Menu  choose 24HR Averages        DH piker  E E et    EO Mike       24 Hour Average The 24 Hour Average Start Time screen allows the user to adjust the 24   Start Time hour averaging start time   e Inthe Main Menu  choose 24HR Averages  gt  Avg Time     EH pikari  Ce a E SO a EA       Calibration Factors Calibration factors are determined during automatic and manual  Menu calibration and are used to correct the concentration readings  The  Calibration Factors menu displays the calibration factors as shown below   The instrument can also be calibrated manually using this menu  For more  information about calibration  see Chapter 4     Calibration        The calibration factors for the SHARP BKG  offset  and SHARP COEF   slope  are set to 0 0 and 1 0 by default  Should regional settings require an  adjustment to these values that DO NOT compromise regulatory  approvals for this instrument  those values can be adjusted at this point     e Inthe Main Menu  choose Calibration Factors     3 14 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Calibration Factors Menu       RRS E CHS mito    WH mpini  E PURE TE Aad Bees PY Fad be bo Bedi beet T4       PM Background The PM Background screen allows the user to manually adjust the PM  Intercept  The PM concentration based on the SET BKG TO is displayed  to facilitate the setting  The PM concentration displayed in the Run  portion of the screen is NOT affected by the SET
156. 260 000 pflow  52 000 ambrh 0 000 srh 0 000 ambtemp    32 000 stemp  41 000 fvol 0 000 cflg 00001F86 aflg  OOOOFCOC   08 32 05 04 09 flags 00110000 pm 0 000 baro 0 000 vac  260 000 pflow  52 000 ambrh 0 000 srh 0 000 ambtemp    32 000 stemp  41 000 fvol 0 000 cflg 00001F86 aflg  OOOOFCOC    Irec format   srec format   erec format   These commands report the output format for lrecs and srecs  and erec data  in various formats such as ASCII without text  ASCII with text  or binary   The following example shows the output format for lrecs is ASCII with  text  according to Table B 3     Send  lrec format  Receive  lrec format 1    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Datalogging    set lrec format format  set srec format format  set erec format format  These commands set the output format for lrecs and srecs  and erec data     according to Table B   3  The following example sets the lrec output format  to ASCII with text     Send  set lrec format 1  Receive  set lrec format 1 ok    Table B 3  Record Output Formats       Format Output Format  0 ASCII no text   1 ASCII with text  2 Binary data    lrec layout   srec layout   erec layout   These commands report the layout  string indicating the data formats  for  data that is sent out in response to the lrec  srec  erec  and related  commands  For details on how to interpret the strings  see    Record Layout  Definition    later in this appendix     Send  lrec layout  Receive  lrec layout 
157. 3  137 10 7 2011 3 4 0 5 14 1 9 9 13 4 1005 81 6 3 6 272 5 7  138 10 8 2011 0 8 8 7 11 4 1009 85 5 24 258 6 0  139 10 9 2011 6 2 1 9 30 6 12 2 16 3 1011 84 5 1 4 190 5 4  140 10 10 2011 5 0 0 7 14 3 17 7 21 3 1009 74 4 3 2 261 0 3  141 10 11 2011 1 6 0 8 50 2 16 3 18 0 1010 77 4 3 6 251 0 0  142 10 12 2011 2 5 0 0  0 7 12 5 15 8 1012 91 1 0 9 226 17 9  143 10 13 2011 4 2 0 5 11 8 9 9 15 8 1022 76 3 0 6 209 0 0  144 10 14 2011 7 2 1 7 23 1 8 7 15 9 1024 69 6 1 0 151 0 0  145 10 15 2011 2 0 7 8 14 8 1020 68 8 152 162 0 0  146 10 16 2011 12 4 2 3 18 4 8 5 14 8 1016 73 8 1 5 157 0 0  147 10 17 2011 19 4 4 9 25 2 10 5 16 6 1011 78 5 0 8 163 0 0  148 10 18 2011 5 5 2 2 39 9 9 2 15 0 1003 82 0 1 2 197 3 0  149 10 19 2011 45 0 6 13 6 8 1 14 0 1010 74 4 1 5 225 0 0  150 10 20 2011 9 4 1 9 20 7 5 6 10 6 1018 79 8 1 0 223 0 0                                                                     volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx    A TUVRheinland      Page 198 of 424    Precisely Right     TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F                                                                               Annex 6 Ambient conditions at field test sites Page
158. 3  2   gt          a  E   e   E   gt   O  A       f  O     0   X aw  Foilkit damaged   observed on Nov 14        Figure 37  Drift of the measured value SN 1  measured component PM2 5       Span point drift PM2 5 SN 2       Dev  from previous value  ug m                  Figure 38  Drift of the measured value SN 2  measured component PM2 5    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 123 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 3 13 Short term drift    The short term drift at zero point and reference point shall not exceed the requirements  of Table 2 of Standard VDI 4202 Part 1  September 2010  within 12 h  for benzene  24 h  in the laboratory test and within 24 h in the field test  A value c  at 70   to 80   of  the upper limit of the certification range shall be used as reference point     Note     By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems     6 2 Equipment   Not applicable    6 3 Performance of test  Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable  
159. 3  Table 6 8  SHARP Measurement Interface Board Pin Connections                 6 24  Table 7   1  Model 50307 Replacement Parts    ccccccceceseceseesceteeseetenenees 7 5  Table 7 2  Model 50307 Cables    cccccccccsseseesseseeseseesecsestesteseeseeeeesteseesees 7 7  Table 7 3  External Device Connection Component             c cccccccseeceeeeneeeeees 7 8  Table 7   4  Analog Output Channels and Rear Panel Pin Connections             7 23  Table 7 5  Analog Input Channels and Rear Panel Pin Connectione                7 23  Table 8 1  RS 232 DB9 Connector Pin Configuration 0 0 0 0    ccc 8 12  Table 8 2  RS 485 DB9 Connector Pin Configuration 00    cece 8 12  Table 9   1  Cable WMS ccets  xc ccteterettte cates aca has arate stelted thetic a  9 2  Table 9 2  Color Codes for 25 Pin and 37 Pin Cables    cccccceeeteeeeeee 9 3  Table 9 3  Mounting Options sec xcs8 sh ence eelaedel vosu sual lt nc weae ned 9 3  Table B   1  Error Response Messages           c cccececcescesesesescetesesescetesesestetesesesees B 3  Table B 2  C Link Protocol Commands ou    cc ceccccececessesteseeseeeetesteeesteseeseesee B 3  Table B   3  Record Output Formats 0    cccccccccccscescscesestesesteseeteseeeseseseeeeees B 31  Table B   4  Stream Time Values oo    cecceccccccscesssteseeseeseeseeesteseeseeseeseeneseeseens B 34  Table B   5  Standard Ranges    cccccccccecscessseseseseeseseetesestesesteseseseseeees B 49  Table B   6  Contrast evel ssc  sino  dia iss ces caste ua bccacks cimdeasabctvant hia nade 
160. 3 8 17 3 989 76 1 0 5 198 46   307 10 5 2012 3 3 1 6 47 8 11 0 16 7 992 87 0 0 4 226 3 6   308 10 6 2012 11 4 1 6 14 3 8 9 16 2 999 81 2 0 3 231 2 3   309 10 7 2012 16 4 4 4 26 9 11 9 14 4 1003 81 3 0 2 160 1 8   310 10 8 2012 15 4 47 30 2 11 5 12 9 996 94 6 0 5 138 2 3   311 10 9 2012 8 4 3 1 37 1 12 0 15 0 999 71 6 0 7 82 1 3   312 10 10 2012 13 8 3 3 24 1 12 4 16 0 994 68 2 0 7 98 1 3   313 10 11 2012 6 4 3 4 53 8 13 4 15 6 983 90 0 0 5 171 1 3                                                                     volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    TUVRheinland      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Page 204 of 424    Annex 7 Measured values from additional test site Bornheim  Summer 2013  related to actual conditions    Page 1 of 2                                                                  740300_2013_936_21209885F_EN docx                                                                                   Manufacturer Thermo Fisher Scientific  PM2 5  Type of instrument Model 5030i SHARP Measured values in pg m   ACT   Serial No  SN1 SN2  Date Ref  1 Ref  2 Ref  1 Ref 2  Ratio Remark Test site  aa PM2 5 PM2 5 PM10 
161. 3 83    Operation  Alarms Menu       Instant Nephelometer The Instant Nephelometer screen displays the current instant nephelometer  concentration and sets the minimum and maximum alarm limits   Acceptable alarm limits range from  10 to 10000 ug m   or   01 to 10  mg m  If the instant nephelometer concentration goes beyond either the  minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Inst Neph        DH Guided  PLP          Min and Max Instant The Minimum Instant Nephelometer Concentration alarm limit screen is  Nephelometer used to change the minimum instant nephelometer concentration alarm  Concentration Limits limit  The minimum and maximum instant nephelometer concentration  screens function the same way     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Inst Neph  gt   Min or Max        3 84 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu    Service Menu The Service menu appears only when the instrument is in the service mode   When the service mode is active  the service  wrench  icon is displayed on  the right side of the status bar  To put the instrument into the service  mode     e Inthe Main Menu  choose Instrument Controls  gt  Service Mode     Advanced diagnostic functions are included in the service mode     Meaningful data should not be collected when the instrument i
162. 3 96  Beta Ref boresho lexi sssaiaetssiaastvs Ni deceitiavesadiiart iG ntucnah dence tiaaieas 3 97  Alpha Threshold sistre ieietu aeiiae s 3 97  Nephelometer Calibration       sssssseseseseesisisisisesrserrsrrrsrssessee 3 97  Nephelometer RE nran E a 3 98  Nephelometer Temperature       ssssssssssessseseseseresrtesrtrrrrrrerererseseeres 3 98  Nephelometer Source Level sicisicssetivscessteeasaietinnatictasstondvetatevates 3 98  Analog Output Calibration 25sec les heielecctesecs ba  ei cacteedens yccteoupeetaes 3 99  Analog Output Calibrate Zero cate cusnveciertlsaknuwolncatnes 3 99  Analog Output Calibrate Full Scale ccc consntatu ca cccuacs earn 3 100  Analog Input Caran  nstaue  easaees tacstecassaa sesdegs tacanggenttonegunneaueads 3 100  Analog Input Calibrate Zero weiter iasera ti ated cieserl   3 101  Analog Input Calibrate Full Scale ics sssicelecgec atti scvecsdeecseedastes 3 101  Display Pixel Test   stcstiepeevs ita otiedeeddainiits deiiidetie 3 102  Restore User Detail ts ricci tideadacvast ei tocbcre aun narnd eA eectats bape oia  3 102    Thermo Fisher Scientific    Contents    Restore Factory Drala sctea tata setetisteccesnes sites st coeee teed cnctnecatataticte 3 102  Password Men  s e aa AA ENRE 3 103  Ser Passwords ssir etisi eoir aaar a a 3 103  Lock Instrument iassa iskin neue nae 3 104  Lock Unlock and Local Remote Operation         ceeeeeseeseeeereeee 3 104  Cisne Password i Ns cs teas a a tae NG te 3 104     Ucn coal Fe ook 10  cs  cer een ae re Dirt ooo
163. 30  SHARP with PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 3 18 Overall uncertainty    The expanded uncertainty of the measuring system shall be determined  The value de   termined shall not exceed the corresponding data quality objectives in the applicable  EU Directives on air quality listed in Annex A  Table A1 of Standard VDI 4202 Part 1   September 2010      Note     By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems  Reference is made to module 5 4 10     6 2 Equipment    By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems  Reference is made to module 5 4 10     6 3 Performance of test    By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems  Reference is made to module 5 4 10     6 4 Evaluation    By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems  Reference is made to module 5 4 10     6 5 Assessment    By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems  Reference is made to module 5 4 10     Minimum requirement fulfilled     6 6 Detailed presentation of test results    By resolution of the competent Germa
164. 300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung A TUVRheinland      os ae Genau  Richtig   Report on the suitability test of the ambient air quality measuring system    Model 5030  SHARP with PMz5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PMs 5  Page 215 of 424  Report No   936 21209885 F    The Check weight Mass  CM  of the filter was calculated for each weighing session using  E A 1 below     CM      eshte Pana  E A 1       Where   Mcheck bet   Mass of check weight weighed immediately prior to sample filter   Meheck att   Mass of check weight weighed immediately after sample filter     The Relative Mass  RM  of the filter was calculated for each weighing session using E A 2  below     RM   pne      CM E A 2  Where   Miter   Mass of sample filter    Particulate Mass  PM  is calculated using the following equation in accordance with  EN14907        E A 3    PM    ea     RM rnan    ee   RM 5 09 l    2 2    Where   Beg1 denotes weighing session 1 prior to sampling  Beg2 denotes weighing session 2 prior to sampling  End1 denotes weighing session 1 after sampling    End2 denotes weighing session 2 after sampling    740300_2013_936_21209885F_EN docx    TU VRheinland   TUV Rheinland Energie und Umwelt GmbH    f Luftreinhaltung  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 216 of 424 Model 5030  SHARP with PM2s5 pre separator
165. 307 SHARP Instruction Manual B 61    C Link Protocol Commands  1 0 Configuration    B 62 Model 50307 SHARP Instruction Manual    Send  analog vout range 2  Receive  analog vout range 2 3    set analog vout range channel range   This command sets analog voltage output channel to the range  where  channel is between 1 and 6 inclusive  and range is set according to Table B   12  The following example sets channel 2 to the 0 10 V range     Send  set analog vout range 2 3  Receive  set analog vout range 2 3 ok    Table B 12  Analog Voltage Output Range Values                Range Output Range   1 0 1 V   2 0 100 mV   3 0 10 V   4 0 5 V   0  cannot be set to this  but may report  Undefined  dig in    This command reports the status of the digital inputs as a 4 digit  hexadecimal string with the most significant bit  MSB  being input 16     Send  dig in  Receive  dig in oxffff  din channel    This command reports the action assigned to the digital input channel and  the index number of the corresponding active state  The following example  reports input 1 to be assigned an index number 3 corresponding to action  of    PM mode    with the active state being high     Send  din 1  Receive  din 1 3 PM MODE high    set din channel index state   This command assigns digital input channel  1 16  to activate the action  indicated by index  1 12   when the input transitions to the designated state   high or low   Use the    list var din    command to obtain the list of  supported index value
166. 36_21209885F_EN docx          TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Luftreinhaltung      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 197 of 424  Report No   936 21209885 F                                                                                     Annex 6 Ambient conditions at field test sites Page 5 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ugm   ug m3        c     C   hPa       m s   4  mm   121 9 21 2011 Bornheim  Summer 1 0 16 9 20 5 1011 69 5 0 8 204 0 0  122 9 22 2011 hell tt 15 1 15 2 18 7 1011 72 2 1 2 231 0 0  123 9 23 2011 12 8 1 6 12 4  124 9 24 2011 1 4  125 9 25 2011 15 1 Bal  Fal   126 9 26 2011 Outage of weather station  127 9 27 2011 3 1  128 9 28 2011  129 9 29 2011 16 6 1 7 10 0  130 9 30 2011 12 1 11 8 9 18 4 26 8 1017 68 3 12 155 0 0  131 10 1 2011 1 2 18 1 28 0 1018 70 6 0 5 176 0 0  132 10 2 2011 3 5 17 8 26 8 1016 75 4 0 3 213 0 0  134 10 3 2011 14 2 3 4 24 0 18 8 26 5 1013 65 9 0 8 168 0 0  134 10 4 2011 9 8 1 4 14 1 17 8 20 5 1013 72 4 1 6 214 0 0  135 10 5 2011 3 5 1 9 53 3 17 5 20 0 1011 70 8 1a 199 0 0  136 10 6 2011 4 8 1 1 23 0 13 2 19 3 1001 71 2 2 3 213 0 
167. 4          2011 Thermo Fisher Scientific Inc  All rights reserved     Specifications  terms and pricing are subject to change  Not all products are available in all countries  Please  consult your local sales representative for details     Thermo Fisher Scientific  Air Quality Instruments  27 Forge Parkway  Franklin  MA 02038  1 508 520 0430    www thermo com aqi    Thermo Fisher Scientific    WEEE Compliance    This product is required to comply with the European Union   s Waste  Electrical  amp  Electronic Equipment  WEEE  Directive 2002 96 EC  It is  marked with the following symbol     Thermo Fisher Scientific has contracted with one or more  recycling disposal companies in each EU Member State  and this product  should be disposed of or recycled through them  Further information on  Thermo Fisher Scientific   s compliance with these Directives  the recyclers  in your country  and information on Thermo Fisher Scientific products  which may assist the detection of substances subject to the RoHS Directive  are available at  www thermo com WEEERoHS     WEEE Compliance    Thermo Fisher Scientific    About This Manual    This manual provides information about installing  operating  maintaining   and servicing the Model 5030 Synchronized Hybrid Ambient Real time  Particulate  SHARP  Monitor  It also contains important alerts to ensure  safe operation and prevent equipment damage  The manual is organized    into the following chapters and appendices to provide direct access to  
168. 4  Measurement Interface Board Connector Pin Descriptions                   Connector Reference Pin Signal Description  Label Designator  PUMP J1 1 AC   2 Pump_AC   3 AC Ground  4 NC  HEATER 2 J2 1 AC   2 Heater2_AC   3 AC Ground  HEATER 1 J3 1 AC   Heater1_AC   3 AC Ground  WDT DISABLE J4 Input  2 Ground  AUX PWR 120V J5 AC   2 AC   PURGE 2 J6 AC        Model 50307 SHARP Instruction Manual 6 15    Troubleshooting  Connector Pin Descriptions    6 16 Model 50307 SHARP Instruction Manual    Connector  Label    Reference  Designator    Pin    Signal Description    Purge2_AC        PURGE 1    J7    AC   Purge1_AC        PUMP 1    J8     24V  Pump1_On Off       JTAG INTF    J9       Oo a N on A OO N    a        a  a a  A w N    TDO          FLOW TEMP 2       Flow2A_RTD  Flow2B_RTD    Ground       AUX HT 1    AUX HT 2    AC   AuxHT_1_AC   AC   AuxHT_2_AC        TT MOTOR    AC   TT Motor_AC        PUMP 2     24V  Pump2_On Off       FLOW TEMP 1          Flow1A_RTD  Flow1B_RTD    Ground       SHARP 1     24V    Thermo Fisher Scientific    Thermo Fisher Scientific    Troubleshooting  Connector Pin Descriptions                                                 Connector Reference Pin Signal Description  Label Designator  2  RS485  3  RS485  4 Ground  PURGE 2 DC J18 1  24V  2 Purge2_DC  PUMP FLOW 1 J19 1  24V  2 Aout_Pump1_Flow  3 Ground  SHARP 2 J20 1  24V  2  RS485  3  RS485  4 Ground  FAN J21  24V  2 Fan_Power  PURGE 1 DC J22  24V  2 Purge1_DC  PUMP FLOW 2 J23  24V  2 Aout_Pump2_Flow  3
169. 5 2 5 7 0 6 5 51 8   136 10 6 2011 5 5 4 1 10 8 10 4 45 6   137 10 7 2011 3 6 3 1 8 1 7 6 42 9   138 10 8 2011   139 10 9 2011 6 0 6 4 10 1 10 4 60 7   140 10 10 2011 4 7 5 2 11 9 12 5 40 9   141 10 11 2011 1 3 2 0 5 5 5 0 31 2   142 10 12 2011 1 9 3 2 5 3 5 0 49 3   143 10 13 2011 4 2 4 2 11 5 11 7 36 3   144 10 14 2011 5 9 8 5 14 8 14 0 50 2   145 10 15 2011   146 10 16 2011 11 1 13 7 17 0 16 7 73 7   147 10 17 2011 18 6 20 1 28 0 27 3 70 0   148 10 18 2011 4 3 6 7 11 6 11 5 47 7   149 10 19 2011 3 8 5 2 Outlier Ref  PM10   150 10 20 2011 9 3 9 5 11 9 15 8                                     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25     Report No   936 21209885 F    Annex 5         Manufacturer  Type of instrument    Serial No     Thermo Fisher Scientific    Model 5030i SHARP    SN1 SN2    Measured values from field test sites  related to actual conditions    A TUVRheinland      Precisely Right     Page 187 of 424    PM2 5    Measured values in pg m   ACT     Page 6 of 11       No                Date Ref  1  PM2 5  m3       10 21 2011  10 22 2011  10 23 2011  10 24 2011  10 25 2011  10 26 2011  10 27 2011  10 28 2011  10 29 2011  10 30 2011  10 31 2011  11 1 2011  11 2 2011  11 3 2011  11 4 2011  11 5 2011  11 
170. 5 8 S S  Tubing  5 8 Union  w Teflon Ferrule    3 Ft Universal Heater  lt     _   _   __ HEPA Filter    3 Way Valve Exhaust    Figure 2   5  Model 5030  Vertical View    2 12 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Installation  Setup    Sample Tube There is one standard length of heated sample tube at one meter  Sample  Extension Lengths tube extensions are used for added height     Rack Mounting One limiting factor when siting the Model 5030  is placement within a   Option standard 19 inch rack mounting  Future planning should be made to  reserve the topmost rack mounting position available due to the vertical  positioning of the sample tube directly from the roof into the top of the  Model 50302  In addition  modification to the rack cabinet to  accommodate the vertical tubing connection also needs to be considered  prior to installation  For more information  see Chapter 9    Optional  Equipment        Review the following installation steps prior to field installation     Monitor Installation Once all height requirements are established  the following steps are usually  taken to complete the installation     1  Remove cover to the 50307 base instrument and inspect for proper  filter tape installation     2  Place 50302 unit in place of intended operation     3  Attach the insulated heater tube assembly to the 5 8 inch OD tube on  top of the 50302  Figure 2 6      Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 2 13    Installation  
171. 53  Decimal Places iiaia ice doi la A EE S 3 54  Number of Table Points      e sessessssesssssseeeerersesrsrssrsrsrerrrrsrsesrsresese 3 54  Table Points nnna lace aarti ee 3 55  NO OU ac IE a gal Bite el ie Da ay 3 55  NSE PANS seals ach cantatas  datatenage tun EE 3 55  Screen Contr  st saroien a a 3 56  Service Mod  s rait iieis iss aan a ei i E esis 3 56  Date  Tim   tnise ane a aan 3 57  BENA E save des sagt te spenees te tontosi cate auuaete 3 57  Diagnostics Meine 6 52 ses a ocississs cess ated seen RGA 3 58  Programi Versions osetia osita Ka EE R outta taa nares P 3 58  Voles ee o ects RRA RE Tocca a e iS 3 59  Motherboard Voltages      ssessesesesesesesesesesessesessesrsesesesesesesesseses 3 59  Interface Board Voltages       ssssssessessssesssssssssesesesesreeeteerrereesrsrsrees 3 59  I O Board  Voltages irrin a 3 60  Detector Board Voltages eini enaa n eee s 3 60  Nephelometer Board Voltages  covsdessanectvssstesasetunsedeashersdessutvet ass 3 60  RH Temperatute norinni aiia eiiis 3 61  Press  re Vactumunnun t iinosnen anaa aA E EE 3 61  FloWa anren an E a E E E E s 3 61  Detector Statuts ra e ir e cana e cared 3 62  Nephelome  ter Statusi aieri asitin inesi js seca tangs ree ei assi 3 62  CR iessen costes age oes E EAA AV EEA IEN 3 62  Analog Input REARS see ies cscedssvavicds Gadera sense Msaddrauessuveee acedetsas   3 63  Analog Input Voltages c  sc2isntitesaivarcstteareletiageantnaniala dates 3 63  Digital Inputs oiio oh took er lea Ka e er AE E cue nets 3 63  R
172. 6 2011  11 7 2011  11 8 2011  11 9 2011  11 10 2011  11 11 2011  11 12 2011  11 13 2011        1 16 2012  1 17 2012  1 18 2012  1 19 2012  1 20 2012  1 21 2012       32 7  19 5  3 5  6 2    Ref  2  PM2 5    33 2  18 9  3 5  5 9           44 3   28 9  4 1  8 6       SN1 SN2       Ratio  PM2 5 PM10                  Remark       Inlet   gt  Zero filter  Zero point   Outlier Ref PM10  Zero point  Zero filter   gt  Inlet                               Zero point  Zero point  Zero point    Bornheim  Summer             Test site                     Bornheim  Winter                740300_2013_936_21209885F_EN docx    TUV Rh        TUV Rheinland Energie und Umwelt GmbH   einland Luftreinhaltung  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 188 of 424 Model 5030  SHARP with PM2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F                                              Annex 5 Measured values from field test sites  related to actual conditions Page 7 of 11  Manufacturer Thermo Fisher Scientific  PM2 5  Type of instrument Model 50301 SHARP Measured values in pg m   ACT   Serial No  SN1 SN2  Ref  1 Ref  2 Ratio Test site  PM2 5 PM2 5 PM2 5 PM10   m  gt   m  gt  A  181 1 22 2012 5 4 5 1 11 4 12 9 43 1 7 7 8 4 Bornheim  Winter  182 1 23 2012 7 0 7 9 11 7 13 5 59 0 9 1 10 3  183 1 24 2012 20 9 21 5 17 3 16 8  184 1 25 2012 19 9 20 0 28 3 25 5 74 
173. 60  E          50     eT  o 40  cc     o  a 30     a  o 20  c   e   oO  10  0              ba            baa   ia x ia    jo  jo  fo  o o o fo  o jo   N N N N N N N N N  ise  oO oO m Tt a a a re      ire  N D re  N  e gt   Ko   se      m N N e    oa N                Figure 19  Course of PM25 concentrations  reference  at the test site    Cologne  parking lot                          ee E  i   winter   70   a 60        ES      50   S   ary   S 40   oc        19   a 30       ao   5 20   c   fe    O   10   0      e ere er e er e re e ere e r re r r r    r ro fe fF          SF    SF SF r r r r  D O O O Q O O O O QG O O O O O O  N A NA NNN NAN NNN NN NN NON  Bn oF oOo O O O Q0 Q0 Q Q O O0 O Q0 O0    Oo  Qo Sr O O O  O e One m m  ee ao FE  O Try O O A Q O A Q O MO O N SF re N  A    O O    A AN O  gt  e A O  gt   e A 0 O             Figure 20  Course of PM25 concentrations  reference  at the test site    Bornheim  motorway  parking lot  summer       740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  TUVRheinland  Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 58 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F                      Conc  PM2 5  Ref   in g m           16 01 2012  23 01 2012  30 01 2012  06 02 2012  13 02 2012  20 02 2012  27 02 2012  05 03
174. 72 h post weighing   Dust   corrected dust mass of the filter   This shows that the method becomes independent from weighing room conditions due to the  corrective calculation  Influence due to the water content of the filter mass between virgin  and loaded filter can be controlled and do not change the dust content of sampled filters     Hence  point 9 3 2 5 of EN 14907 is fulfilled     The example of the standard weight between November 2008 and February 2009 shows that  the permissible difference of max  20 ug from the previous measurement is not exceeded        Stability of standard weight between Nov 08 and Feb 09                                  0 20010 fo   0 20005       2   oo                   0    0             Standard weight     0 20000      CRAG    3     Mean standard weight      0 19995   0 19990      0 5 10 15 20 25    No  of weighing             Figure 61  Stability of standard weight    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    A TUVRheinland      Genau  Richtig     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM 5     Report No   936 21209885 F    Table 39     Stability of standard weight                   Difference to    Weighing Standard the previous  Date No  weight weighing  g Hg  12 11 2008 1 0 20002  13 11 2008 2 0 20001  10  10 12 2008 3 0 
175. 75     0 205     0 667     0 900     0 975    086     0 534     0 676     110     0 530     0 466    0 407    157 0 847  1 120     0 035   0 611     0 342 23 38 4 5  52 0 952 1 111     0 034   1 326     1 050 22 93 100 0  209 0 967  1 087     0 014   0 408     0 240 20 28 28 2    Limit Value of 30 pg m 3     117 4   0 656              Orthogonal Regression       Slope  b      ub   Intercept  a      ua WCM      217 ug m3       Individual Datasets    Combined Datasets    The check of the five criteria from chapter 6 1    Bornheim Winter  Cologne Winter    Bornheim Summer    Teddington Summer     lt  18 ug m3  218 ug m3  All Data        104     0 031   0 840     0 778 22 60 56 1    058     0 027  0 394     0 592 18 90 53 5      0 030  1 099     0 427 14 77 15 7      0 051  0 207 0 433 11 83             ules 5 4 9     5 4 11  resulted as follows    More than 20   of the data are greater than 17 g m8   The in between uncertainty of the candidates is smaller than 2 5 g m8   The in between uncertainty of the reference is smaller than 2 0 g m8     Criterion 1   Criterion 2   Criterion 3     740300_2013_936_21209885F_EN docx    KEY   Criterion 1  Criterion 2  Criterion 3  Criterion 4  Criterion 5    L        L  C     Other    Methodology of the equivalence check  mod     TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pr
176. 8 12 2  235 7 25 2012 17 8 18 5 26 5 26 6 68 4 18 3 19 5  236 7 26 2012 21 0 21 4 18 2 19 4 Outlier Ref  PM10  237 7 27 2012 12 3 11 8 20 0 20 3 59 7 14 7 13 7  238 7 28 2012 5 6 4 6  239 7 29 2012 4 3 3 3  240 7 30 2012 5 0 4 8 10 0 9 8 49 2 4 3 4 5                               740300_2013_936_21209885F_EN docx    ss      TUV Rheinland Energie und Umwelt GmbH  TUVRheinland  Luftreinhaltung    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 190 of 424 Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Annex 5 Measured values from field test sites  related to actual conditions Page 9 of 11                Manufacturer Thermo Fisher Scientific       PM2 5  Type of instrument Model 50301 SHARP Measured values in g m   ACT               Serial No  SN1 SN2                                                                                                                                  Ratio Test site   PM2 5 PM10  7 31 2012 10 9 Teddington  Summer  8 1 2012  8 2 2012    8 3 2012 4 7 4 3 Outlier Ref  PM2 5  8 4 2012  8 5 2012  8 6 2012  8 7 2012 g  8 8 2012 11 2 SN1 malfunction  8 9 2012 12 2 SN1 defective  8 10 2012 13 9 SN1 defective  8 11 2012 12 8 SN1 defective  8 12 2012 5 6 SN1 defective  8 13 2012 5 2 SN1 defective  8 14 2012 6 8 SN1 defective  8 15 2012 10 3 SN1 defective  8 16 2012 7 4 SN1 defective  8 17 2012 5
177. 8 3 993 55 8 0 8 180 0 0  259 8 18 2012 6 9 2 4 34 5 24 2 32 2 995 67 6 0 2 214 0 0  260 8 19 2012 6 4 2 3 35 7 2LF 30 8 997 68 7 0 4 208 0 0  261 8 20 2012 7 4 2 8 37 4 19 7 27 8 1001 77 3 0 2 215 0 0  262 8 21 2012 3 0 TE 23 3 998 76 0 0 1 207 0 3  263 8 22 2012 20 17 6 23 9 998 69 3 0 2 210 0 0  264 8 23 2012 2 3 17 4 22 7 992 69 0 0 2 190 0 0  265 8 24 2012 2 6 17 1 20 6 983 81 7 0 8 174 2 8  266 8 25 2012 1 9 16 7 22 5 987 81 9 0 5 243 27 9  267 8 26 2012 1 5 16 9 23 3 999 69 4 0 4 206 0 0  268 8 27 2012 21 18 3 21 4 992 74 9 0 8 195 0 8  269 8 28 2012 2 0 18 0 23 7 995 74 2 0 5 190 0 3  270 8 29 2012 1 5 15 0 20 6 993 85 2 0 6 184 4 1                                                         volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx    A TUVRheinland      Page 202 of 424    Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung                                                       Annex 6 Ambient conditions at field test sites Page 10 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ug m
178. 80 9 0 5 190 10 2  297 9 25 2012 2 6 14 43 7 12 6 16 2 971 80 1 0 7 158 41  298 9 26 2012 5 7 0 8 14 7 12 4 16 2 980 88 7 0 6 222 2 5  299 9 27 2012 4 6 1 3 27 2 12 6 18 8 993 77 0 0 4 227 0 0  300 9 28 2012 3 4 1 4 41 1 12 3 18 3 997 84 5 0 4 193 1 5                                             volatile PM compounds  determined with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx          TUV Rheinland Energie und Umwelt GmbH TUVRheinland    Luftreinhaltung      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM2 5  Page 203 of 424  Report No   936 21209885 F                                                                   Annex 6 Ambient conditions at field test sites Page 11 of 11   No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ugm   ugm         C     C   hPa       m s   4  mm    301 9 29 2012 Teddington  Summer 8 3 Lf 20 7 10 4 17 5 1003 75 1 0 4 233 0 0   302 9 30 2012 47 2 2 47 4 14 9 17 9 997 79 1 0 7 186 2 5   303 10 1 2012 3 9 20 50 8 13 0 17 0 993 92 0 0 2 194 4 8   304 10 2 2012 3 7 2 5 66 7 13 8 18 3 989 84 5 0 7 186 2 8   305 10 3 2012 3 5 1 5 44 5 10 5 15 0 988 81 8 0 2 232 0 5   306 10 4 2012 3 5 1 8 51 2 1
179. 847 1 043     0 033   0 520     0 320 12 76 4 5    Combined Datasets 1 039    218 ug m3 0 952        0 032   1 195     0 983 17 53 100 0    All Data 209 0 967 1 017     0 013   0 367     0 224 13 22 28 2    Orthogonal Regression  Dataset    Slope  b      Up    Bornheim Winter 1 033     0 029   0 763    Cologne Winter 0 990     0 025  0 386    Individual Datasets    Bornheim Summer 0 885     0 028    Teddington Summer 0 947     0 048    Intercept  a      Ug    Limit Value of 30 ug m3    Wem       217 ug m3      0 729        0 554    1 052     0 400  0 234     0 406     lt  18 ug m3 166 0 817 0 983     0 033  0 176     0 323 12 08 4 8    Combined Datasets       218 ug m3 0 947 1 019     0 033   1 033     1 032 18 45 100 0    All Data 217 0 962 0 987     0 013  0 079     0 226 13 68 27 2    740300_2013_936_21209885F_EN docx    a 3    TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 170 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 5 Requirements on multiple component measuring systems    Multiple component measuring systems shall comply with the requirements set for  each component  also in the case of simultaneous operation of all measuring channels     6 2 Equipment   Not applicable    6 3 Performance of test  Not 
180. 85 F    6 5 Assessment    No deviations  gt   1 4   in relation to the default value at 230 V caused by voltage changes  could be detected for PM25     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Table 22 shows a summary of the test results     Table 22  Dependence of the measured values on mains voltage  deviation in      Mains voltage    Deviation    Cv f v f u   u       The single values are presented in annex 4 of the appendix     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 113 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 3 9 Cross sensitivity    The change in the measured value caused by interfering components in the sample  gas shall not exceed the requirements of Table 2 of Standard VDI 4202 Part 1  Sep   tember 2010  at zero and reference point     Note     This test is not relevant for particulate measuring systems  Instead  minimum requirement  5 4 5 is valid  The results of this test can be found in module 5 4 5     6 2 Equipment   Not applicable    6 3 Performance of test  Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of 
181. 85 F    TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung                                                                      Annex 6 Ambient conditions at field test sites Page 2 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ug ms   ug m3          C     C   hPa       m s   4  mm   31 4 7 2011 Cologne  Winter 13 4 5 5 40 7 13 8 21 6 1019 67 2 3 0 245 0 0  32 4 8 2011 19 4 12 9 18 2 1018 64 7 2 9 255 0 0  34 4 9 2011 11 3 18 9 1018 59 8 1 3 183 0 0  34 4 10 2011 11 4 14 0 23 1 1016 60 2 1 1 191 0 0  35 4 11 2011 15 1 16 0 25 0 1012 58 8 3 9 244 27  36 4 12 2011 8 6 2 9 33 9 7 7 12 4 1018 66 7 4 1 257 0 9  37 4 13 2011 12 3 3 2 26 2 10 1 14 3 1013 57 1 2 0 203 0 0  38 4 14 2011 19 5 4 4 22 4 8 0 14 2 1013 65 4 0 6 159 0 0  39 4 15 2011 12 4 3 1 24 7 10 4 17 0 1014 53 6 1 2 169 0 0  40 4 16 2011 2 2 11 9 16 9 1017 51 7 0 9 166 0 0  4 4 17 2011 2 5 11 4 19 2 1017 53 7 1 2 139 0 0  42 4 18 2011 17 1 3 0 17 8 14 3 21 2 1011 48 6 1 9 149 0 0  43 4 19 2011 17 7 2 8 15 8 15 5 25 5 1009 52 4 1 2 146 0 0  44 4 20 2011 20 4 2 9 14 2 16 6 25 7 1008 51 3 ti 154 0 0  45 4 21 2011 3 0 17 8 26 3 1006 54 1 0 7 180 0 0  46 4 22 2011 2 6 20 0 27 8 1003 51 8 1 3 146 8 3  47 4 23 2011 2 5 18 0 28 9 1005 58 0 0 7 152 0 0  48 4 24 2011 2 3 18 1 28 0 1011 51 7 1 0 172 0 0  49 4 25 2011 20 2 2 4 12 0 16 8 26 0 1013 50 3 Ta 153 0 0  5
182. 885 F    Reference Detector Board    SHARP Board Assy        Source Board  SHARP Board Bracket Assy    Scatter Detector Board    Optical Bench    Optical Bench Inlet    Figure 2  Hardware components Model 50301 SHARP     Optics module    Primary Measurement Head Assy      Tape Radius Guide                   Detector Amp    Bottom Filter Tape Supply Reel  Radius  Spindle Tape Wheel Assy    SHARP Connection Bracket Assy    Inlet       External    Solenoid    Counter Wheel Photo  Tube a Ay Interrupt Board Assy    Cam Photo Interrupt  SHARP Connection D       Board  Bracket Assy 7    Pressure Board    Proportional Valve  w fitting Assy    Front Pane               gt  p 1 0 Expansion Board  Opt     Digital Output Board  Front Panel Board       Motor w  Photo    Measurement Interface Interrupt Coupler Assy    Power Supply Board    Motherboard    Optional  Transformer    Figure 3  Hardware components Model 50301 SHARP     Beta module    740300_2013_936_21209885F_EN docx    TUVRheinland    Precisely Right     Page 34 of 424    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F       Figure 4     740300_2013_936_21209885F_EN docx    Picture of hardware components Model 50301 SHARP     Beta module    TUV Rheinland Energie und Umwelt 
183. A  normally opened contacts  and are  used to provide alarm status and mode information from the analyzer  as  well as remote control to other devices  such as for controlling valves during  calibration  The user may select what information is sent out from each  relay and whether the active state is opened or closed     Thermo Fisher Scientific    Digital Inputs    Serial Ports    RS 232 Connection    Thermo Fisher Scientific    System Description  1 0 Components    Sixteen digital inputs are available  which may be programmed to signal  instrument modes and special conditions including     e Filter tape change  e Turn pump off  e Turn pump on  e Analog outputs to zero  e Analog outputs to full scale  The actual use of these inputs will vary based on analyzer configuration     The digital inputs are TTL level compatible and are pulled up within the    analyzer  The active state can be user defined in firmware     Two serial ports allow daisy chaining so that multiple analyzers may be  linked using one PC serial port     The standard bi directional serial interface can be configured for either RS   232 or RS 485  The serial baud rate is user selectable in the firmware for  standard speeds from 1200 to 115200 baud  The user can also set the data  bits  parity  and stop bits  The following protocols are supported     e C Link   e MODBUS Slave   e Gesytec  Bayern Hessen   e Streaming Data   e ESM    The Streaming Data protocol transmits user selected measurement data via  the se
184. APE CNTRL  108 PUMP CNTRL ON OFF  109 TEMP COMP ON OFF  110 PRES COMP ON OFF  111 HEATER ON OFF   112 HTR RH TEMP CNTRL  113 EXT ALARM 1   114 EXT ALARM 2   115 EXT ALARM 3    Thermo Fisher Scientific    Serial Communication  Parameters    Thermo Fisher Scientific    Appendix D  Gesytec  Bayern Hessen  Protocol    This appendix provides a description of the Gesytec  Bayern Hessen or  BH  Protocol Interface and is supported both over RS 232 485 as well as  TCP IP over Ethernet     The Gesytec commands that are implemented are explained in detail in this  document  The Gesytec protocol support for the iSeries enables the user to  perform the functions of reading the various concentrations and to trigger  the instrument to be in sample zero span mode if valid for that instrument   This is achieved by using the supported Gesytec commands listed below     Up to three simultaneous connections are supported over Ethernet     For details of the Model 5030  Gesytec Protocol specification  see the  following topics     e    Serial Communication Parameters    on page D 1  e    TCP Communication Parameters    on page D 2  e    Instrument Address    on page D 2   e  Abbreviations Used    on page D 2   e    Basic Command Structure    on page D 2   e    Block Checksum    on page D 3   e    Gesytec Commands    on page D 3    The following are the communication parameters that are used to configure  the serial port of the iSeries to support Gesytec protocol     Number of Data bits  7 or 8   N
185. ARP Instruction Manual 3 89    Operation  Service Menu    under the filter tape from the positive AP port on the rear panel of the  instrument and enter the value on this screen before calibrating  A    e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration  gt   Vac Flow  gt  Vac Pres Span        Calibrate Flow Pressure The Flow Pressure Span screen allows the user to view and set the flow  Span sensor calibration span point     Note The operator should use an independent manometer  capable of 30  mmHg AP  to measure the differential pressure across the orifice using the  positive and negative AP ports on the rear panel  and enter the value on this  screen before calibrating  A    e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration  gt   Vac Flow  gt  Flow Pres Span     DH vided  Piet       Restore Default The Restore Default Calibration screen allows the user to reset the default  Calibration values     Note Be sure to close the toggle switch after calibration is complete  A  e Inthe Main Menu  choose Service  gt  Pres  Vacuum Calibration  gt     Vac Flow  gt  Set Defaults     3 90 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu       Flow Calibration The Flow Calibration submenu is used to view and set the flow calibration  flow point  The flow calibration screen is visible only when the instrument  is in service mode  For more information on the service mode  see    Service  Mode    earlier in the chapter     Not
186. Alarms     3 66 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu       PRES CH  Pi Tit    Wi mid  PS PHP ELSES Py AG Lf bi       Filter Tape Counter The Filter Tape Counter screen displays the current filter tape counter  value and sets the maximum alarm limit  If the filter tape counter reading  goes beyond the maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Instrument Alarms  gt  Filter  Tape Counter        Max Filter Tape Counter The Maximum Filter Tape Counter screen is used to change the maximum  filter tape counter alarm limit  480   90  usage  10  remaining      e Inthe Main Menu  choose Alarms  gt  Instrument Alarms  gt  Filter Tape  Counter  gt  Max           CHEE Eel    Bast    ae POE Sek come  EE fee PEE EE          Detector Alarms The Detector Alarms submenu allows the user to view a set of alarms for  detector boards  Board status indicates that the power supplies are working  and connections are successful  There are no setting screens for this alarm     e Inthe Main Menu  choose Alarms  gt  Detector Alarms     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 67    Operation  Alarms Menu       Min and Max Alpha The Alpha Alarms screen allows the user to view the minimum and  Alarms maximum alarm parameters  Acceptable alarm limits range from 0 to 100   If the alpha count reading goes beyo
187. Align the radius tube adapter with the open port on the bottom of the  optics assembly as per Figure 10   9 and gently push down and then  screw down the two tabs using the captive hardware        Figure 10 9  Optics Installation    3  Connect the cable assembly from the rear panel of the Optical assembly  and secure it to the optics connector that was installed in Step 3 of this  procedure and as shown in Figure 10 10        Figure 10   10  Connecting the Optics to the Optics Connector    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 10 9    Model 5014i to 5030i Upgrade Procedures  Upgrade of the Heater Tube Assembly    Upg rade of the Usethe following procedure for upgrade of the heater tube assembly   Heater Tube  Assembly    1  The heater assembly needs to be revised for the 5030  SHARP to  include a ferrite bead and a small faraday cage  The four steps shown  below in Figure 10 11 and Figure 10 12 should be used to modify the  heater assembly     Note The sample downtube should be slightly loose at the roof line to  allow vertical play while working on this assembly  If the tubing height has  changed and needs to be cut back  please cut and clean the tubing before  attaching to the optical assembly  A      Collar    i       Figure 10 11  Upgrading 5030  Heater Assembly    10 10 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Model 5014i to 50301 Upgrade Procedures  Model Label Change       Figure 10   12  Connecting Heater to Optics Assem
188. Baffle Plate       Safety Snap Locks  Tube  d   16 14     Figure 5 2  European PM   Inlet    PM size selective inlet  1 m3 h  for 50302 Beta     Same construction as the EN reference inlet  but with 1 m3 h  With special surface protection using an aluminium EMATAL surface    With special water separator and easy exchangeable impactor baffle  plate    Standard tube connection  d   16 mm for connecting a continuous  monitor    Easy converting to PM25 using the PM25 nozzle plate  8 nozzles     Easy converting to PM  using the PM  nozzle plate  8 nozzles     Cyclone Ifa Sharp Cut Cyclone  SCC  or a BGI Very Sharp Cut Cyclone is being  Maintenance  used for PM2s sampling  Figure 5 3   once every 2 3 weeks the dust pot  should be unscrewed  emptied  and cleaned with a lint free cloth  Once  every three months  the entire SCC assembly should be inspected and  cleaned  O rings checked for distortion  cracks  fraying  or other problems   Replace O rings as necessary     5 4 Model 50307 SHARP Instruction Manual    Thermo Fisher Scientific    Heater and Sample  Tube    Weather Proofing    Fan Filter Inspection  and Cleaning    Thermo Fisher Scientific    Preventive Maintenance  Fan Filter Inspection and Cleaning       Figure 5 3  PM25 Sharp Cut Cyclone    On at least an annual basis  more frequently in heavily polluted  environments  the sample tube that attaches to the inlet and to the Model  5030 Beta module should be removed and cleaned  The use of a bottle  brush and string will all
189. C Link Protocol Commands  Measurement Configuration    form to save time in transmission  It is sent as a type  5  binary C Link  response with no checksum     The RLE encoding consists of a 0 followed by an 8 bit count of  consecutive OxFF bytes  The following  c  code will expand the incoming  data     void       inte  05   KF  unsigned char far  sc4bpp   sc2bpp   screen   ptr     unpackDisplay   void far  tdib  unsigned char far  rlescreen      ptr   screen    unsigned char far   malloc 19200      RLE decode the screen   for  i 0  i lt 19200  amp  amp   ptr   screen   lt  19200  i             ptr        rlescreen   i    if    rlescreen   i     0      unsigned char rlecount      unsigned char     rlescreen     i      while  rlecount        ptrt      0   rlecount             else if    rlescreen   i            Oxff   unsigned char rlecount      unsigned char     rlescreen     i      while  rlecount         ptrt      Oxff   rlecount         To convert this data into a BMP for use with Windows  it needs to be  saved as a 4 bit per pixel gray scale image  Also note that BMP files are  upside down relative to this data  that is  the top display line is the last line  in the BMP     sc   screen   These commands are meant for backward compatibility with the C series   Screen information is instead reported using the    iscreen    command above     Send  screen   Receive  screen This is an iSeries instrument   Screen information not available    conc unit    This command reports th
190. Configuration Protocol  DHCP  as on or off  DHCP is used to assign an  IP address to the instrument automatically  The following example shows  that DHCP is on     Send  dhcp  Receive  dhcp on    set dhcp onoff  onoff     on   off      This command enables  on  and disables  off  the DHCP service  When  DHCP is set to on  the instrument gets the IP address  the netmask  address  and the gateway address from a DHCP server  When DHCP is set  to off  the instrument gets these addresses from system memory  The  following example sets the DHCP service to on     Note If DHCP is changed from on to off and then the IP address  the  netmask address  or the gateway address is changed  you must cycle power  to the instrument before the change takes effect  Until you cycle the power   the address assigned by the DHCP server will still be used and reported as  the current address  A    Send  set dhcp on  Receive  set dhcp on ok  format    This command reports the current reply termination format  The following  example shows that the reply format is 00  which means reply with no  checksum  according to Table B 8     Send  format  Receive  format 00    Model 50307 SHARP Instruction Manual B 57    C Link Protocol Commands  Communications Configuration    B 58 Model 50307 SHARP Instruction Manual    set format format   This command sets the reply termination format  where format is set  according to Table B   8  The following example sets the reply termination  format to checksum     Send  s
191. D  qua uog jNeJaq 0  Jesey             HY HdAN UaJUOD yPwwog  dwe1 HdaN jua Uuod Jo9 eg  snjejs JajowWojayday 607 esesg  Joy 8g eyeq peb607 mai   40D 89g OAY VOUS 1999S  gg sBuijes Buibbojejyeq  eudiy   INSS  Id  ssey aunyesodua    SN e S 10 99 0q suonpuog IAJaWIN OA  SMO 4 oey o1azede    MO 4 Jajuno  winnoe   ponad  amawoeg aul  eyeg HBS  winnoe  ainssaid OWI 1X8N  dwa  peog ywq SSe  due  moj4 odes  dwa   ajdwes youag  dwe  juaiquiy enuey  HY   jdwes 10 u09   deL 4   l 4 sonjeA Add 910 S  H  Hy wW  qwuy pjoysary  dwa  sonjeA  sainyesad we 1  HY ployseuuL HY ueds  pueog Jajawojaydeny dwaj 189  pueog 10 99 0q HY ajduies  peog O I HO Aapis  pueog eoRja ul 101 u09 apow yd  N  pueoqueujoyy 49 83 H 18S Byg yd  N s    uey wojsng 13S  s    eyoA duind Mo 4 18S 4209 dYYHS ous Bay     uey dyVHS  SUOISIdA WeIBOlg apo u  w  nse  N PJA dyVHS dHVHS Bay suun 23u09  sonsoubeig S 0UOD JUSUINASU  S10 0ey uOneIqieD s  bei  ny IHbe SUIT    UONeIbajU  abuey             nua uey  su    195 UNY    uaa  1S9  yo    uael0S  dn 18mog       Figure 3 3  Flowchart of Menu Driven Firmware    3 5    Model 50307 SHARP Instruction Manual    Thermo Fisher Scientific    Operation  Firmware Overview    Power Up Screen The Power Up screen is displayed when power is applied to the Model    5030    This screen is displayed while the internal components are warming  up and diagnostic checks are being performed     Thermo    IEN TI    vies       Run Screens The Run screens display the SHARP concentration  ambient cond
192. DEFECTIVE PRODUCT  EXCEPT AS EXPRESSLY PROVIDED IN  THIS WARRANTY STATEMENT  SELLER DISCLAIMS ALL  OTHER WARRANTIES  WHETHER EXPRESS OR IMPLIED  ORAL  OR WRITTEN  WITH RESPECT TO THE PRODUCTS   INCLUDING WITHOUT LIMITATION ALL IMPLIED  WARRANTIES OF MERCHANTABILITY OR FITNESS FOR ANY  PARTICULAR PURPOSE  SELLER DOES NOT WARRANT THAT  THE PRODUCTS ARE ERROR FREE OR WILL ACCOMPLISH  ANY PARTICULAR RESULT     A 2 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Instrument    Identification    Thermo Fisher Scientific    Number    Appendix B  C Link Protocol Commands    This appendix provides a description of the C Link protocol commands  that can be used to remotely control a Model 5030  analyzer using a host  device such as a PC or a datalogger  C Link protocol may be used over RS   232  RS 485  or Ethernet  C Link functions can be accessed over Ethernet  using TCP port 9880     Streaming data is sent out the serial port or the Ethernet port on a user   defined periodic basis  Streaming data over Ethernet is only generated when  a connection is made on TCP port 9881     Up to three simultaneous connections per protocol may be made over  Ethernet     For details  see the following topics    e    Instrument Identification Number    on page B 1  e    Commands    on page B 2   e    Measurements    on page B 11   e    Alarms    on page B 14   e    Diagnostics    on page B 23   e    Datalogging    on page B 27   e    Calibration    on page B 34   e    Keys Display    
193. Detector  Calibration    Beta Ref Alpha Counts    High Voltage    3 96 Model 50307 SHARP Instruction Manual                tO Mihei i    The Mass Coefficient submenu allows the user to perform a mass foil    calibration using a zero  null  and span foil set     In the Main Menu  choose Service  gt  Detector Calibration  gt  Manual                 tO Mihei i    The Beta Ref Alpha Counts screen allows the user to view the real time    alpha    and beta counts     In the Main Menu  choose Service  gt  Detector Calibration  gt  Manual     gt  Beta Ref Alpha Counts                 OPE a EEE of    The High Voltage screen allows the user to view and set the high voltage    value     In the Main Menu  choose Service  gt  Detector Calibration  gt  Manual     gt  High Voltage     Thermo Fisher Scientific    Operation  Service Menu       Beta Ref Threshold The Beta Ref Threshold screen allows the user to view and set the detector  operating voltage manually     e Inthe Main Menu  choose Service  gt  Detector Calibration  gt  Manual     gt  Beta Ref Threshold        Alpha Threshold The Alpha Threshold screen allows the user to view and set the alpha  threshold manually     e Inthe Main Menu  choose Service  gt  Detector Calibration  gt  Manual     gt  Alpha Threshold        CRHTE ENE  DN EEE TE Sanad Bone fet tLe bt          Nephelometer The Nephelometer Calibration submenu is used to view and set the  Calibration nepbelometer calibration  The nephelometer calibration screen is visible  
194. EEE  i          Decimal Places The Decimal Places screen allows the user to select how many digits are  displayed to the right of the decimal point  from 0 to 6  with a default of 2     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  Decimal Places        Number of Table Points The Number of Table Points screen allows the user to select how many  points are used in the analog input conversion table for the selected  channel  The instrument uses linear interpolation between the points in  this table to determine what the reading value is based on the analog input  voltage  Each point in the table consists of an analog input voltage value   0 10 5 V  and a corresponding reading value  Only two points are  necessary for linear inputs  however a larger number of points may be used  to approximate non linear inputs  The points range from 2 to 10  with a  default of 2     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  Table Points        3 54 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions    Table Points The Table Point submenu allows the user to set up an individual table  point     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  Point 1 10        ERED CH  MP Tit    WH nied  PS PHP ESE Pi AG ane       Volts The Volts 
195. FF  unplug the power cord  and remove the  cover     2  Remove the fan guard and filter from the fan by unsnapping it     3  If the fan is not being replaced  install the new filter  and snap it back  into place  and skip the remaining steps     4  Disconnect the fan power cable from the fan  Do not lose the four  nuts     Thermo Fisher Scientific    Servicing  Detector Amplifier Assembly Replacement    5  Remove the four fan mounting screws and remove the fan     6  Install a new fan following the previous steps in reverse        BLE INSERTED THIS WAY  EXTERIOR SIDE OF REAR PANEL    STICKER  AIRFLOW     REFER TO  FOR POSITIONING OF FAN    Mounting Screws  4     Filter    PARS P yy i J  je p IN  P Z  i EDY Fan Guard  ZY    Keps Nuts  4        Figure 7   6  Replacing the Fan    Detector Amplifier Use the following procedure to replace the detector preamplifier assembly  Assembly  Figure 7 7      Replacement Equipment Required     Detector amplifier assembly  Philips screwdriver    Adjustable wrench    WARNING The detector pre amp assembly has a power backup built in      DO NOT OPEN    caution handeling  Backup power will dissipate after 8    A hours  A    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF and unplug the power cord     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 15    
196. Figure 4   1  Differential Pressure and Vacuum Calibration Ports               0  4 6  Figure 5 1  U S  EPA PM  o INI6t es os so ee ot ots ee es 5 3  Figure 5   2  European PM ig Inlet    ccccecesesceceseseseetesesesestesesesesteeseseees 5 4  Figure 5 3  PMzs5 Sharp Cut Cyclone o oo    cceccccccccscssesceseseeseeceseseseeseeteseeeeseaess 5 5  Figure 5   4  Inspecting and Cleaning the Fan    5 6  Figure 5   5  Replacing the Filter Tape oo    ccc cceescscecesesestetesesesteteteteseseees 5 8  Figure 5   6  Replacing the External Pump Exhaust Filter    cece 5 9  Figure 6   1  Board Level Connection Diagram   Common Electronics                6 8  Figure 6 2  Board Level Connection Diagram   Measurement System             6 9  Figure 6 3  Board Level Connection Diagram   SHARP              ccccccceeeeeeeee 6 10  Figure 7   1  Properly Grounded Antistatic Wrist Strap occas 7 4  Figure 7   2  Model 50307 Component Layout    cece cesceteseeteeteeeees 7 9  Figure 7   3  SHARP Optics Component Layout    ccccccccccececeteseeteeeeeeee 7 10  Figure 7   4  Removing the SHARP Optics Assembly               ccccccceecseeeeeeeee 7 11  Figure 7   5  Removing the Measurement Case Assembly and Lowering the   Partition Rane cx sci crs e et wot Spm etre Set Selatan E ten cae 7 12  Figure 7   6  Replacing the Fan oo    ccc ccccecesesesceesesescetesesestetesesesteeseseaes 7 15    Model 50307 SHARP Instruction Manual xix    Figures    xx Model 50307 SHARP Instruction Manual    Figure 7 7  Replacing th
197. Fisher Scientific    Operation  Volumetric Conditions    a ed nF es Patio pH mith  et   mi i mii       Communication The Communication Settings menu is used for communications control  Settings and configuration     e In the Main Menu  choose Instrument Controls  gt  Communication  Settings        Eppuri pri pr ares Pe Tit op mote    ee Pii E E    EN PUE TE ueek Bone ETET Sei Boat    i       Serial Settings The Serial Setting submenu is used for serial communications control and  configuration     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Serial Settings        ER pr CH  perm  Bet EPP aif ly in    CH bo bet  Bs EEE TE Sanat Bone EY Fane Bast    ae 5 i iNi i       Baud Rate The Baud Rate screen is used to set the RS 232 RS 485 interface baud rate   Baud rates of 1200  2400  4800  9600  19200  38400  57600  and 115200  are available  The analyzer   s default baud rate is set to 9600 to provide  backwards compatibility with the older C series analyzers     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Serial Settings  gt  Baud Rate     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 35    Operation  Volumetric Conditions    Data Bits    Parity    Stop Bits    3 36 Model 50307 SHARP Instruction Manual       DH Guided  PLP          The Data Bits screen is used to set the number of serial data bits to either 7  or 8  default      In the Main Menu  choose Instrument Controls  gt  Communication   
198. GmbH  Air Pollution Control    A TUVRheinland      Precisely Right   Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 35 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F                                           HEATER  HEPA     L NEPHELOMETER     Y _ DETECTOR  j men TAPE  Prea   B SOURCE  Gr  o   tea  ORIFICE        PUMP dt  ap   EXHUAST t   SY    L PROPORTIONAL  T VALVE  AA    VACUUM  e   PUMP J  4  MAKE UP AIR    Figure 5  Diagram     Set up of Model 5030i SHARP    740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 36 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The nephelometer assembly is downstream the heated tube  The fine dust follows a laminar  path by the isolated nephelometer and reaches then the radial tube above the radiometric  assembly  The nephelometer consists of a photometer based on light scattering with a  pulsed near IR LED operating at a central wavelength of 880 nm  The level of scattering is  determined in a measured volume of 1 cm  in forward direction at an angular range of 60   80    The design o
199. Ground   TTL Input 13   TTL Input 15   Digital Ground  Analog Voltage 1  Analog Voltage 3  Analog Ground  Analog Voltage 5  Analog Ground  Analog Ground   Power Fail Relay COM  Power Fail Relay N O  Contact    Model 50307 SHARP Instruction Manual 6 11    Troubleshooting  Connector Pin Descriptions    6 12 Model 50307 SHARP Instruction Manual                         Connector Reference Pin Signal Description  Label Designator  22 Digital Ground  23 TTL Input 3  24 TTL Input 4  25 TTL Input 6  26 Digital Ground  27 TTL Input 9  28 TTL Input 11  29 TTL Input 12  30 TTL Input 14  31 TTL Input 16  32 Digital Ground  33 Analog Voltage 2  34 Analog Voltage 4  35 Analog Ground  36 Analog Voltage 6  37 Analog Ground  VOLTAGE TEST J6 1  5V  2  3 3V   15V   15V  SER EN J7 Serial Enable Jumper  2  3 3V  24V IN J10  24V  2 Ground  RESET PROC J12 Reset Proc  2 Ground  DIGITAL 1 0 J14  5V  2  24V  3  24V  4 Ground  5 Ground  6 Ground  7 SPI Reset  8 SPI Input    Thermo Fisher Scientific    Troubleshooting  Connector Pin Descriptions    Connector Reference Pin Signal Description  Label Designator    9 SPI Output   10 SPI Board Select   11 SPI Clock    RS485 to Rear Panel   RS485 to Rear Panel   5V    5V    5V       EXT  RS485 J15         Ground  Ground  Ground  NC   NC   24   24   24   24   24   24 MONITOR J17 1 24V Power Monitor  2 Ground    FRONT PANEL J18 1 Ground  BD        co loo  N o  Sz p ow N    oO     a  a  a a  ol  gt  w N             Ground   LCLK    LCD Signal  Ground   Ground   
200. HARP Instruction Manual    A  alog Vo  lt  ge OUt puts  cas n sesucas istered cdeanesarsives nineden idn 8 9  Analog Current Outputs  Optional  oo    cee eeeeseeceeeeeeeeeeeeeeneeeeees 8 10  Analog Voltage Inputs  Optional  2   0 istei ceapeted cesses tenadetnnsencten 8 10  Digital Relay Outputs 2 013  suiecussacterdecusteeesenstieusdeanveusduatdewetsteetlne 8 10  Digital Inputs inate ot heath nae eat Seta AER 8 11  Seral TORS ss SiS ae i es aS Ee eee a RRR 8 11  RS 252 Comme ction sssaaa ei iaeiiai a a i a 8 11  RS 485 Connection ensesine iseer a ER inise 8 12  Ethernet Connection      e ssseseessesesesesrsrersssrsesesrsrsrsesrsesetrerrrseerersreee 8 12  External Accessory Connector        ssssssssssssseesesetetesrtetetetererererersrereres 8 12  Optional  Equipment siin 9 1  Inlet Ass  m blies niar kn nodet a a i Ei 9 1  Sampling Tube Extensions     sssssssseseseseseseseseresrsesrseseststsestsseseseses 9 1  I O Expansion Board Assembly       essseeseeeeeseseessssrsrsesrrerrrrrerereess 9 1  25 Pin Terminal Board Assembly        s sssessesesseseseeeesesesseseseseesesessesesesee 9 1  Terminal Block and Cable Kits     s s sssessessssesessseesesessesesesesseseseeeesesessese 9 2  Caples arcas A ta an Sc A E E eal ia tace td teat 9 2  IMaunting Opti  ns ia ius contisnctieiedncieedhtas iA E 9 3  Roof Flange Assembly     s sssssessssesesesesisesesesesesrsrsrsrsrsrsrsrststststseseseseseses 9 8  Model 5014  to 50307 Upgrade Procedures               ssesessssseseseeeeeees 10 1  Instrume
201. ING If disconnected  turn  off AC power to the base before re   connecting        Nephelometer Alarms      Neph Sample    Bad zero    Re zero     Model 50307 SHARP Instruction Manual 6 5    Troubleshooting  Troubleshooting Guides    6 6 Model 50307 SHARP Instruction Manual       Alarm Message Possible Cause Action  Settings Verify settings   Nephelometer Alarms     Bad zero Re zero   Neph Cal  No flow Verify values        Nephelometer Alarms      Neph Temp    Nephelometer Alarms      Neph RH    Disconnected cable    Error in calibrating    Defective sensor    Verify cable connections     Re calibrate     Replace board           Nephelometer Alarms      LED Current    Current settings too  low    Disconnected cable  from base    Disconnected cable  from SHARP optics  board    Set LED current to 65 m    Verify cable connection     Verify cable connection        Nephelometer Alarms      Ref Det Volt    Current settings too  low    Disconnected cable  from base    Disconnected cable  from SHARP optics  board    Set LED current to 65 mA     Verify cable connection     Verify cable connection        RH Temperature Alarms        Ambient RH    Disconnect cable   external  on rear panel    Disconnect cable   internal     Verify cable connection     Verify cable connection        RH Temperature Alarms        Sample RH    RH Temperature Alarms        Ambient Temp    Disconnected heater    Disconnect cable   external  on rear panel    Disconnect cable   internal     Verify cable connecti
202. IOLET   15 RED WHITE 33 LIGHT BLUE BLACK   16 ORANGE WHITE 34 GRAY GREEN   17 GREEN WHITE 35 GRAY RED   18 BLUE WHITE 36 GRAY VIOLET   19 VIOLET WHITE oy LIGHT GREEN BLACK          Mounting Options The analyzer can be installed in the configuration described in Table 9 3  and shown in Figure 9 1 through Figure 9   4     Thermo Fisher Scientific    Table 9 3  Mounting Options          Mounting Type Description   Bench Positioned on bench  includes mounting feet and front panel side   trim handles    EIA rack  top position  Mounted in an ElA style rack  includes mounting slides and front  panel ElA rack mounting handles    Retrofit rack  top Mounted in an EIA style rack  includes mounting slides and front   position  panel EIA rack mounting handles  This configuration is intended    for direct replacement of a C series instrument in an existing  rack  The rail mounting location is lower on the case and the  front mounting screw slots have non standard EIA locations     Model 50307 SHARP Instruction Manual 9 3    Optional Equipment  Mounting Options       z    16 75 REF    1 56     ra    13 75    Figure 9   1  Bench Mounting    9 4 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Optional Equipment  Mounting Options       1 48    Figure 9 2  EIA Rack Mounting    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 9 5    Optional Equipment  Mounting Options       2 61    Figure 9 3  Retrofit Rack Mounting    9 6 Model 50307 SHARP Instruction Manual Therm
203. It was therefore not possible to examine the range of B4     6 5 Assessment    The maximum dependence on surrounding temperature in the range 5   C to 40   C was 0 2  g m    at zero    No deviation  gt   2 6   of the default value at 20   C could be observed at reference point   Minimum requirement fulfilled  yes    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH TUVR heinland    Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher Page 109 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 6 Detailed presentation of test results    Table 19  Dependence of the zero point  SHARP  on surrounding temperature  deviation in  g m     average of three measurements    Temperature Deviation    e l e f ooe g e       Table 20  Dependence of sensitivity  mass coefficient  on surrounding temperature  devia   tion in    average of three measurements    Temperature Deviation          740300_2013_936_21209885F_EN docx    m      TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 110 of 424 
204. LLP    LCD Signal  LFLM     LCD Signal  LD4   LCD Signal  LDO     LCD Signal  10 LD5      LCD Signal  11 LD1     LCD Signal  12 LD6     LCD Signal  13 LD2     LCD Signal    o a N O o A O N    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 6 13    Troubleshooting  Connector Pin Descriptions    6 14 Model 50307 SHARP Instruction Manual             Connector Reference Pin Signal Description  Label Designator  14 LD7    LCD Signal  15 LD3     LCD Signal  16 LCD Bias Voltage   17  5V  18 Ground  19 Ground  20 LCD_ONOFF     LCD Signal  21 Keypad Row 2 Input  22 Keypad Row 1 Input  23 Keypad Row 4 Input  24 Keypad Row 3 Input  25 Keypad Col 2 Select  26 Keypad Col 1 Select  27 Keypad Col 4 Select  28 Keypad Col 3 Select  29 Ground  30 Ground  31 Ground  32 Ground  33  24V  34  24V  RS232 RS485 A P1 A 1 NC  2 Serial Port 1 RX   RS485 IN   3 Serial Port 1 TX   RS485 OUT   4 NC  5 Ground  6 NC  7 Serial Port 1 RTS   RS485 OUT   8 Serial Port 1 CTS   RS485 IN   9 NC  RS232 RS485 B P1 B 1 NC  2 Serial Port 2 RX   RS485 IN   3 Serial Port 2 TX   RS485 OUT   4 NC    Thermo Fisher Scientific    Thermo Fisher Scientific    Troubleshooting  Connector Pin Descriptions             Connector Reference Pin Signal Description  Label Designator  5 Ground  6 NC  7 Serial Port 2 RTS   RS485 OUT   8 Serial Port 2 CTS   RS485 IN   9 NC  AC IN PJ1 1 AC HOT  2 AC NEUT  3 AC Ground  AC 24VPWR PJ2 1 AC HOT  2 AC NEUT  3 AC Ground  AC INTF BD PJ3 1 AC HOT  2 AC NEUT  3 AC Ground    Table 6   
205. LO  MOLY snyeg sogowopydoy  uopeuqeg mo Bopeuy dwe  adweg Py Beg  1007 OUS Uden dwe  wanuy 409 88  dwa  Uden Hy adwes sg  Hy Uday Hy warquay ewy  uopwequeg ydoy swej onzeodwo Hy ssev  penueyy WA 0 py snyng 1090300  oy weung G37 Smola  UOREIgED 104900 Hi Uden Mold  WeDIpedD 3 eudi dwa  uden wnroe    UODg 00D SSR IO Udon agowaeg  uaap owg HALS L  N wnnoepyunssoid  upeo 9 UNNIBA STyeIS peog dwe  peog  WPO puey  suuepy ydeon dwe  moli  VOREIGHED sseni eee dua1 adwes  mney eydy dua  pumiguy  oy Sres peog Hy ores  uopeqieg mold swuwejy 1033000 Hy pugu  NOY OBA  SES dX3 ON seimwedwes Hy   801 OB SNBIS SORRY PJBOR JHOWOIEUGEN  UOUE WNNoEA salg SRS PROMOLBOW peog 309800  pomssed YOUN ae   Moli 4ouag weog ON  PIOMSSEY CACWOY dwe  adweg euep ede  204 pseog GOR Ley  piomesey o Buey o dwe  pagu JHUNOD ede  Jai PsBOMeLBOYY  uau MUZSU  4907 He ugua suuejy puau rnugsuj seBeyon  PIOMSSed eS  VOREIGHED dWIUHY Perooeg suuejy SUOIRIOA wE Bod  omewa wines Tay ooh Ta    euozewjy   ou iewg    ouo  epo eo1es   4seuog uous  Or  jouumy J JURUNI  9b PUUEYJ IPA  SyeueyD jueuna Iy  SRUUBUD BRIA TY   53u00 pano Boreu    oe  suges mdu yey   Ot   s6ugps   ePy indino  Byuog OF  DUROS di dOL  Bguog meg buueang    WUBUOD YNE OF ESOL    wawog nwwog   wawy paas   60  oses3   Beg pebio  ma    DSW VOINS Pees  sSumes Bu bloj   ES  d   anyewa    SUORIPUCD OLZEWIN OA    pew  laguog ode soyid  poesy  dwa   prousevul HY  dw     H    40  t03u09  40708  0S  duind imo  40S  epon Wewansmey    Wao Weunnsu    seSuey Ws IPS  
206. Link Protocol Commands  Commands    B 10 Model 50307 SHARP Instruction Manual                                                                                     Command Description Page  no or srec Reports number of srecs stored in memory B 32  pm Reports the measured PM concentration B 13  pm 24avg Reports the 24 hour average PM concentration B 13  power up mode Reports sets the power up mode as local or remote B 59  pres comp Reports sets pressure compensation to actual or standard B 48  pres std Reports sets the standard pressure B 49  program no Reports analyzer program number B 60  pump Reports sets pump on or off B 53  push Simulates pressing a key on the front panel B 44  range Reports sets current PM range B 49  relay Sets relay logic status for the designated relay s  B 65  relay stat Reports relay logic status for all relay s  B 65  reset tape counter Resets the tape count to zero B 26  rh th Reports sets the current relative humidity threshold for the B 50  detector  ri  right  Simulates pressing right pushbutton B 44  ru  run  Simulates pressing run pushbutton B 44  sample rh Reports the sample relative humidity B 26  save Stores parameters in FLASH B 53  save params Stores parameters in FLASH B 53  sc  screen  C series legacy command that reports a generic response B 45   Use iscreen instead   sharp Reports the 1 minute running average SHARP monitor B 14  concentration  sharp24avg Reads the last calculated 24 hour average SHARP monitor B 14  concentration  sp fi
207. MODBUS   502    Here are the MODBUS ADU  Application Data Unit  formats over serial  and TCP IP        Serial  Slave Address   Function Code   Data   Error Check          TCP IP    MBAP Header   Function Code   Data                      The MODBUS slave address is a single byte in length  This is the same as  the instrument ID used for C Link commands and can be between 1 and  127 decimal  i e  0x01 hex to 0x7F hex   This address is only used for  MODBUS RTU over serial connections     Note Device ID    0     used for broadcast MODBUS commands  is not  supported  Device IDs 128 through 247  i e  0x80 hex to OxF7 hex  are not  supported because of limitations imposed by C Link  A    In MODBUS over TCP IP  a MODBUS Application Protocol Header   MBAP  is used to identify the message  This header consists of the  following components     Transaction Identifier 2 Bytes 0x0000 to OxFFFF  Passed back in response   Protocol Identifier 2 Bytes 0x00  MODBUS protocol    Length 2 Bytes 0x0000 to OxFFFF  Number of following bytes   Unit Identifier 1 Byte 0x00 to OxFF  Passed back in response     Thermo Fisher Scientific    Function Code    Data    Error Check    Function Codes     0x01 0x02  Read  Coils Read Inputs    Thermo Fisher Scientific    MODBUS Protocol  Function Codes    The function code is a single byte in length  The following function codes  are supported by the instrument     Read Coils 3 0x01  Read Inputs   0x02  Read Holding Registers   0x03  Read Input Registers   0x04  F
208. Manual Thermo Fisher Scientific    Hardware  Configuration    Thermo Fisher Scientific    C Link Protocol Commands  Hardware Configuration    set temp std value   This command sets the standard temperature to value  where value is a  floating point number representation of standard temperature between 0 to  25   C  The following example sets the standard temperature to 15   C     Send  set temp std  Receive  set temp std 15 ok  temp th    This command reports the heater temperature threshold for the sample   The following example reports that the temperature threshold is 30   C     Send  temp th  Receive  temp th 30 degC    set temp th value  This command sets the heater temperature threshold for the sample to  value  where value is a floating point number representation of temperature    threshold between 0 to 70   C  The following example sets the temperature  threshold to 35   C     Send  set temp th 35  Receive  set temp th 35 ok  contrast    This command reports the screen   s level of contrast  The following example  shows the screen contrast is 55   according to Table B   6     Send  contrast  Receive  contrast 11  55     set contrast  evel  This command sets the screen   s  evel of contrast  according to Table B   6   The following example sets the contrast level to 50      Send  set contrast 10  Receive  set contrast 10 ok    Table B 6  Contrast Levels                Level Contrast Level  0 0    1 5    2 10    3 15    4 20     Model 50307 SHARP Instruction Manual B 5
209. Minimum requirement fulfilled  yes    The following Table 35 shows an overview of the results of the equivalence check for the  candidate Model 5030i SHARP for PM2 5  In case that a criterion is fulfilled or not the text is  represented in green or red  Furthermore  the five criteria from chapter 6 1 Methodology of  the equivalence check  modules 5 4 9     5 4 11  are taken into account  the related cells are  highlighted in colour     740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     Page 154 of 424    Table 35     TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Overview equivalence test of Model 5030i SHARP for PM25       PM2 5 5030i Sharp       All Data   lt  18 ug m3  218 ug m3    SN1 Dataset    Individual Datasets    Combined Datasets    29 1  2 17 ug m3    WCM    Slope  b      ub    Bornheim Winter  Cologne Winter  Bornheim Summer  Teddington Summer   lt  18 yg m3  218 yg m3  All Data    Dataset    Orthogonal Regression Betw een Instrument Uncertainties    Intercept  a      ua Reference Candidate    148 0 865  1 066     0 032   0 040     0 317    51 0 959  1 090     0 032    Orthogonal Regression Limit Value of 30 pg m3  Slope  b      ub   Intercept  a      ua WCM      2 17 ug m3     0
210. Model 50307 SHARP Instruction Manual                                                                                        Command Description Page  erec layout Reports current layout of erec data B 31  filter period Reports sets the number of hours to be added to filter next B 46  time for the filter tape change  filter time Reports sets the next date and time of the filter tape change B 47  flags Reports current active measurement status flags in HEX B 12  flow Reports the current sample flow B 25  flow pres Reports the current flow pressure B 25  flow pres span Reports sets the flow pressure span B 39  flow pres offset Reports the flow pressure offset B 39  flow span Reports sets the current flow span B 39  flow target Reports sets the current flow B 47  flow temp Reports the current flow temperature B 25  flow temp offset Reports sets offset for flow temperature B 39  format Reports sets current reply termination format B 57  he  help  Simulates pressing help pushbutton B 44  high volt Reports sets the current high voltage for the detector B 40  host name Reports sets host name string B 58  ht control Reports sets the current status of the heater B 47  instr name Reports instrument name B 58  instrument id Reports sets instrument id B 58  int time Reports sets mass and concentration integration time B 48  io bd Reports sets the presense of the I O expansion board B 25  isc  iscreen  Retrieves framebuffer data used for the display B 44  layout ack Reports disables stale l
211. N E ATE R S 3 20  RH Threshold 2i   ctrctinvedndliianwinwaiiatudy nR 3 20  Temperature Threshold otic sevetiastooddest Wut ictotetehnieindade 3 21   Filter Tape Contro lden ienn cote exalt de secaldansenl ria a cal atk cat eats 3 21  Manual asa US ogee aea tae eat eae eae Sata ah ak Gals ae olde 3 21  Mass Limitinin fs Galea heen a a A A aS  3 22  Nert  TIME ni a a AE 3 22  Periodont kroi iee n E R E Ea AE EEES 3 22  COUNTER eleni at a a E T a 3 23  Tape Zero Rati  neasasi jini aai a iiie 3 23   Volumetric Conditions    s ssseseseessesesesesesrsesrsssrsrsrststsrstststsesesesesesesesese 3 24   Temperature Compensation  ssiticeciesoass ised casgedials iicnctstosncesd ets 3 24  Standard Temperature esa cis ny asset enceshesediedaigeeceandt ve esudsbadead bane 3 24  Pressure Compensation is vasicis destscsevetssstdiasssieutessdedadawivtasiuansse 3 25  Staridard Press techies ioe cient joacahe codes teaeend cA eats a cokecaest  3 25    viii Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Contents    Datalopging SettiNgSs seiss eesriie sk aaea e aai 3 25  SANE A EETA E E E T T 3 26  View bogged Dataset ei ea a i ea 3 26  Number of Recordsen a a ta 3 27  Dat eae BARN EA A ARE EEE E AE EEA EEEE E 3 27  Erase Logone aens a A er AA NaS E e as 3 28  Select Content       sssesesereseessseessisssrresreresetssrtsseressesssereseessressreesees 3 28  Choose Field Dataran a ae AA AEEA 3 29  Eo a ETAT  E E E ee E 3 29  Other Measurements       ss sessessseeereeeserssrssreserees
212. No additional equipment required   6 3 Performance of test    The measuring system is operated using the front panel pushbuttons and LCD display     A change of the parameters or the adjustment of sensors is only possible by pressing several  key sequences     The measuring system can also be password protected  If a password protection is activated   lock icon   it is still possible to see data  parameters and settings on the front panel user in   terface  It is  however  not possible to perform any changes     As an outside installation of the measuring device is not possible  additional protection is giv   en by installation at locations  to which unauthorised people have no access  e g  locked  measuring cabinet      6 4 Evaluation    Unintended and unauthorised adjustment of instrument parameters can be avoided by pass   word protection  Even without password protection  the adjustment of sensors and change of  the parameters can only be done by pressing several key sequences  Moreover  additional  protection against unauthorised intervention is given by installing the system in a locked  measuring cabinet     6 5 Assessment    The AMS is protected against unauthorised and unintended adjustment  In addition  the AMS  shall be locked up in a measuring cabinet     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUV Rheinland     Air Pol
213. O 4  1S9  jaxiq Aejdsig wnnoe    Q    JauUeYd Indu  S  ld eg  uoneaqieg nduj Bojeuy sunejy WwNNILA ANSSAd  g   JauueyD jueuND due   pueog  9   JauueYUD abelo dw  moj4  uoneiqued Ino Bojeuy dwe  adwes  19497 DHS Uden dw   jusiquiy  dwe  udey HY ejdwes  Hy Yuden HY Juaiquiy  uoneiqieg yd  y swuejy eunjesedwas HY  enuen HOA 18d 34  ony queuing G37  UOHeIGIeD 10 99 9q HY uden  qualoiyja0d 43 eydy dwie 1 yd  N  leo uden  14909 oeg   jdwe   yd  N  4U919144909 WNN  A snes peog    ua1914909 jeway L sunejy yden  uoneiqieg SSe gag  jenuey  eudiy  ony snes peog  uoneiques Mmoja swuejy 10 99 9q  MO 4 98A sn e S dX O I  ssalg eg snes    ep  u   uoneiqieg wnnseA Sad Snes pseoqieyjow  piomssegd 490 UN dwa moj youeg  piomssed sAoway dwe  ejdwes eBueyg ede  1314  piomssed Huey dw  juaiquy Jayunog ade  JayI4  juawinsjsu  4907 Hy aiqwy swuejy  uaWINJ sU     PAOMSSed 19S uoHeiqied diN3L HY       piomsseq CEEI    pa 29   q sunejy    sunely    9   JeuueyD ebeyoA  TIV  s ndyno Boyeuy so   Ok  L nding  s     e s   ej  y  91 1 Indu  syndy jeubiq  8 4  ul Boyeuy  soabeyo  indy Bojeuy  8    NI  sBulpeay indu  Bojeuy  u49  10 99 9q JOY  queuing qa      uoz    wIL    wu  L   3eq   yo u0  apo  ad1A1eg   SB1 UOD UBBlDS  9 4 jauueyg Juang  9 4 jauueyo abeyoA  S QUUBYD JUELIND II  sj  uueyo   LA Iv  Byuog ndino Boyeuy  JJ  s6umes indy jey 61q  OL L  sBues Aejay indino  6yuo9 O I  sbulnes dl d  L  Byuog eyeq Bulweays   oo0 O1q4 uolTeoUNWWOD  al  uawnysu   sBulnes  ees  sBuijes uonesunwwo9  Bul6Gojeyeq ainByuo
214. P Instruction Manual 7 29    Servicing  Motherboard Replacement    Motherboard  Replacement    A    7 30 Model 50307 SHARP Instruction Manual    5  Pop the digital output board off of the mounting studs and remove the  board     6  To install the digital output board  follow the previous steps in reverse     Use the following procedure to replace the motherboard  Figure 7   13    Equipment Required    Motherboard   Philips screwdriver    Nut driver  3 16 inch    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Remove the I O expansion board  optional   if used  See the    I O  Expansion Board  Optional  Replacement    procedure in this chapter     3  Remove the digital output board  See the    Digital Output Board  Replacement    procedure in this chapter     4  Unplug all connectors from the motherboard  Note connector locations  to facilitate re connection     5  Using the nut driver  remove the six standoffs securing the board to the  rear panel  Figure 7   14      6  Pop the motherboard off of the support bracket  and remove the  motherboard     7  To install the motherboard  follow the previous steps in reverse     8  Calibrate the analog voltage outputs as defined earlier in this chapter   all ranges      Thermo Fisher Scientific    Measurement
215. PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 4 Evaluation    An evaluation of the drift at zero point and of the measured value in 24 h is not possible     The zero point drift was evaluated on the basis of the results of the external zero point  measurements performed on a regular basis  For this  the values were compared with the     measured values    of the previous test and with the    measured values    of the first test     The reference point drift was evaluated on the basis of the results of the mass coefficient de   termination  Here again  the values were compared with the    measured values    of the previ   ous test and with the    measured values    of the first test     The large deviation found for SN 1 during the span foil measurements carried out from 12  May 2011 to 19 July 2011 was caused by a damaged filter foil set  This was only detected on  14 November 2011  after another unusually large deviation was found  The test was then re   peated with a new set of filter foils and the system passed showing no abnormalities  The  deviations determined by comparison with the previous values  see yellow marking in Table  25  are therefore not attributable to the measuring system     It should be noted that concentration values could not be simulated by checking the mass  coefficient  It was therefore not possible to examine the range of B4   6 5 Assessment    T
216. PM10 PM2 5 PM10  g ms g m3    1 6 7 2013 56 6 Bornheim  2 6 8 2013  3 6 9 2013 66 9  4 6 10 2013 59 1  5 6 11 2013 60 7  6 6 12 2013 47 4  7 6 13 2013 42 1  8 6 14 2013 49 3  6 15 2013 50 0  6 16 2013 44 8  6 17 2013 60 3  6 18 2013 60 9  6 19 2013 66 5  6 20 2013 62 1  6 21 2013 62 2  6 22 2013 63 8  6 23 2013 59 8  6 24 2013 61 8  6 25 2013 50 4  6 26 2013 63 5  6 27 2013 69 2  6 28 2013 60 4  6 29 2013  6 30 2013  7 1 2013  7 2 2013  7 3 2013  7 4 2013  7 5 2013  7 6 2013          TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 205 of 424                         Annex 7 Measured values from additional test site Bornheim  Summer 2013  related to actual conditions Page 2 of 2  Manufacturer Thermo Fisher Scientific  PM2 5  Type of instrument Model 5030i SHARP Measured values in pg m   ACT   Serial No  SN1 SN2  Ref  1 Ref 2  Ratio Remark Test site  PM10 PM10 PM2 5 PM10   Lg m    Lg m        7 7 2013 14 9 14 4 Bornheim  7 8 2013 16 3 16 1  7 9 2013 24 9 22 6  7 10 2013 19 1 17 5  7 11 2013 26 6 24 9  7 12 2013 Outlier Reference PM10  7 13 2013 20 4 20 7  7 14 2013 22 2 21 5  7 15 2013 26 2 25 6  7 16 2013 30 2 28 9  7 17 2013 18 7 18 0  7 18 2013 17 0 17 3 Outlier reference PM2 5 
217. Pin Descriptions    Table 6 7  Digital Output Board Connector Pin Descriptions    Connector Reference Pin  Label Designator    MOTHER BD J1         Oo a N on A OQO N    Signal Description     5V    24V   24V  Ground  Ground  Ground  SPI Reset  SPI Input  SPI Output  SPI Board Select  SPI Clock       DIGITAL J2 1  OUTPUTS    oOo a N oo A WW N      X        oO       a N N   n  a   N  co loo  N lop  ol p w N    N  oO    Thermo Fisher Scientific    Relay 1 Contact a    Relay 2 Contact a   Relay 3 Contact a   Relay 4 Contact a   Relay 5 Contact a   Relay 6 Contact a   Relay 7 Contact a   Relay 8 Contact a   Relay 9 Contact a   Relay 10 Contact a   NC   Solenoid Drive Output 1  Solenoid Drive Output 2  Solenoid Drive Output 3  Solenoid Drive Output 4  Solenoid Drive Output 5  Solenoid Drive Output 6  Solenoid Drive Output 7  Solenoid Drive Output 8  Relay 1 Contact b    Model 50307 SHARP Instruction Manual 6 23    Troubleshooting  Connector Pin Descriptions    6 24 Model 50307 SHARP Instruction Manual    Connector  Label    Reference  Designator    Pin    22  23  24  25  26  27  28  29  30    32  33  34  35  36  37    Signal Description    Relay 2 Contact b  Relay 3 Contact b  Relay 4 Contact b  Relay 5 Contact b  Relay 6 Contact b  Relay 7 Contact b  Relay 8 Contact b  Relay 9 Contact b  Relay 10 Contact b   24V    24V    24V    24V    24V    24V    24V    24V    Table 6 8  SHARP Measurement Interface Board Pin Connections    Connector  Label    TEMP RH    Reference  Designator   
218. Power Supply 24 V Output       115 VAC Supply to Measurement Interface Board       RS 485 Data       AC Power Switch to Motherboard    DC Power Supply Status Monitor       Motherboard to Front Panel Board    DC Power Supply AC Input       Main AC from Receptacle Assembly       AC to Power Switch          oO  Oo   oOo lele   Se  SaS Nea a  S  5    Se   Sa   eS Se eS   SS  2  g  B  2        Fan Power Assembly       106028 00    RH Temperature Connector Assembly       106030 00    Heater Cable to Heater Assembly       106029 00    Pump Cable to Pump Plug Assembly       106844 00    Rear Solenoid Cable Assembly       106027 00    Pressure to Interrupt Board Cable Assembly       105952 00  106189 00    Cam Photo Interrupt Cable Assembly  Counter Wheel Photo Interrupt Cable Assembly       106400 00    External RH Temperature Cable Assembly       108156 00    SHARP to Measurement Interface Board Cable Assembly          108170 00    Thermo Fisher Scientific    SHARP Case Cable Assembly    Model 5030  SHARP Instruction Manual 7 7    Servicing  External Device Connection Components    External Device    able 7 3 lists the standard and optional cables and components used for    connecting external devices such as PCs and dataloggers to an iSeries  Connection instrument     Components Table 7 3  External Device Connection Components                            Part Number Description   102562 00 Terminal Block and Cable Kit  DB25   optional    102556 00 Terminal Block and Cable Kit  DB37 
219. Precisely Right    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 145 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F  Furthermore the evaluation of the following datasets is done for    e Each test site and comparison individually    e 1 dataset with measured values  lt  18 g m  for PM25  Basis  averages of reference  measurement     The uncertainty between the candidates Ups is calculated from the differences of all 24 hour  results of the simultaneously operated candidate systems according to the following equa   tion   yr   Yio      i 1  2n  results of the parallel measurements of individual 24 h values i  No  of 24 h values    2 d  Ubs a    with Vit and yi 2  n    6 5 Assessment    The uncertainty between the candidates Ups is with a maximum of 1 68 g m  for PM25 below  the required value of 2 5 g m8     Minimum requirement fulfilled  yes    740300_2013_936_21209885F_EN docx    y      T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM  5   Report No   936 21209885 F    Page 146 of 424    6 6 Detailed presentation of test results    Table 34 shows the values calcu
220. RH Termperature Alarms  gt   Sample RH        Wi rbd          Min and Max Sample RH The Minimum Sample RH alarm limit screen is used to change the  Limits minimum ambient RH alarm limit  The minimum and maximum sample  RH screens function the same way     e Inthe Main Menu  choose Alarms  gt  RH Temperature Alarms  gt   Sample RH  gt  Min or Max           u ompi    bed PERRO Pihet PRET          Ambient Temperature The Ambient Temperature screen displays the current ambient temperature  value and sets the minimum and maximum alarm limits  Acceptable alarm  limits range from  30   C to 60   C  If the ambient temperature reading goes  beyond the minimum or maximum limit  an alarm is activated and the    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 73    Operation  Alarms Menu    alarm  bell  icon appears in the status bar on the Run screen and in the  Main Menu     e Inthe Main Menu  choose Alarms  gt  RH Termperature Alarms  gt     Ambient Temp        gum Ty  EUO iad fat          Min and Max Ambient   The Minimum Ambient Temperature alarm limit screen is used to change  Temperature Limits the minimum ambient temperature alarm limit  The minimum and  maximum ambient temperature screens function the same way     e Inthe Main Menu  choose Alarms  gt  RH Temperature Alarms  gt   Ambient Temp  gt  Min or Max        Flow Temperature The Flow Temperature screen displays the current ambient temperature  value and sets the minimum and maximum alarm limits  Acceptable 
221. RP    Serial No    SN 1  NEPH   SN 2  NEPH     SN 1  NEPH  ne pce value ea value pases value  Ta mee m aes Ta ae      0 0       a  ug m    ug m    g m  ua md  ug m   rm mf             740300_2013_936_21209885F_EN docx    T  V Rheinland  TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung  Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Page 180 of 424    Annex 3 Dependence of measured value on ambient temperature Page 1 of 1    Manufacturer Thermo Fisher Scientific    Standards Reference foils  Type Model 50301 SHARP    Serial No  SN1 SN2    ky ni e             740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    7 AX TOvrheinland    Luftreinhaltung 3    Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5  Page 181 of 424  Report No   936 21209885 F    Annex 4 Dependence of measured value on mains voltage Page 1 of 1    Manufacturer Thermo Fisher Scientific    Standards Reference foils  Type Model 50301 SHARP    Serial No  SN1 SN2    O Mains voltage Dee value Dee value   value   V  a ae             740300_2013_936_21209885F_EN docx    TUVRheinland      Preci
222. RP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu       Vacuum Mass Coefficient The Vacuum Mass coefficient screen allows the user to view and set the  vacuum coefficient using manual calibration mode   Factory set      e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Vacuum  Coefficient              BOHMEE gue fr          Barometer Mass The Barometer Mass Coefficient screen allows the user to view and set the  Coefficient barometer coefficient using manual calibration mode   Factory set      e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Baro  Coefficient           WH mid          Mass Coefficient The Mass Coefficient submenu allows the user to view and set the mass  coefficient using manual or auto mode   Factory set      e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Mass  Coefficient     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 93    Operation  Service Menu       Auto Mass Coefficient The Auto Mass Coefficient screen allows the user to conduct a mass foil  calibration     Note After pressing a series of screens will walk you through the  mass foil calibration procedure  For details  see    Auto Mass Coefficient    on  page 4 9  A    e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Mass  gt   Auto        Manual Mass Coefficient The Manual Mass Coefficient screen allows the user to adjust the mass  coefficient used in beta attenuation     e Inthe Main Menu  choose Service  gt 
223. Report No   936 21209885 F    5  Reference measuring methods    In accordance with the specifications of EN 14907  the following instruments were used dur   ing the field test     1  Reference equipment for PM2 5     Low Volume Sampler LVS3   Manufacturer  Ingenieurb  ro Sven Leckel  LeberstraBe 63  Berlin   Germany   Date of manufacture  2007   PM25 Sampling inlet  Two PM2 5 reference instruments were operated simultaneously during the test  each with a  controlled volume flow of 2 3 m  h  The accuracy of the volume flow control is  lt 1   of the  nominal volume flow under real operating conditions   The sampling air for the LVS3 is sucked in through the sampling inlet by a rotary vane vacu   um pump  The sampling air volume flow is measured between the filter and the vacuum  pump with a measuring orifice  The suctioned air streams then out of the pump passing  through a separator for abrasion of the rotary vane towards the air outlet   The electronic measuring equipment of the LVS3 displays the inlet sampling air volume in  standard or operating m  as soon as the sampling is complete   To determine the PM25 concentration  the laboratory performed a gravimetric determination  of the amount of suspended particulate matter on the respective filters  The obtained result  was then divided by the respective volume of sampling air in operating m8     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control   s  Precisely Ri
224. Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 153 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F     Point 9 5 4  The combined uncertainty of the candidates w  cy is calculated for each dataset  by combining the contributions from 9 5 3 1 and 9 5 3 2 according to the following equation     2  u i  Weom Yi    vam    For each dataset  the uncertainty Wem is calculated at the level of y    30 g m  for PM2s      Point 9 5 5  The expanded relative uncertainty of the results of the candidates is calculated  for each dataset by multiplication of Wom with a coverage factor k according to the following  equation    Wom   K  Wow    In practice  k 2 for large n     Point 9 6  The highest resulting uncertainty Wcy is compared and assessed with the re   quirements on data quality of ambient air measurements according to EU Standard  7    Two results are possible     1  Wom S  Waqo     Candidate method is accepted as equivalent to the standard method    2  Wom  gt  Wago     Candidate method is not accepted as equivalent to the standard method   The specified expanded relative uncertainty Wago for particulate matter is 25    7     6 5 Assessment   The determined uncertainties Wcm without application of correction factors lay for all ob   served datasets below the defined expanded relative uncertainty Waqo of 25   for fine dust   
225. S  EPA PM10 Inlets       European or U S  EPA PM2 5 Inlet Configurations       PM2 5 VSSC          PM1 0 SCC       Tripod  Data Output and Input    iPort software to view and change system operation from PC       Outputs  Selectable Voltage  RS232 RS485  TCP IP  10 Status Relays  and Power Fail  Indication  standard   0 20 or 4 20 mA Isolated Current Output  optional        Inputs  16 Digital Inputs  standard   Eight 0 to 10 VDC Analog Inputs  optional  8 User Defined  Analog Outputs  0  1 or 0  5 VDC        Protocols  C Link  MODBUS  Geystitech  Bayern Hessen   ESM Protocol  streaming data  and  NTP  Network Time Protocol  protocols  Simultaneous connections from different locations  over Ethernet        2 User Defined Contact Closure Alarm Circuits       4 Averaged Analog Inputs  0   5 VDC  with user defined conversion to engineering units  Data Storage    Maximum of 190 000 Concentration records with date  time  and flags       Up to 32 types of logged data parameters  increased parameters reduces number of records  Power Requirements  Model 50307  100 240 VAC  50 60 Hz  to be specified upon order        Pump  115 VAC  50 60 Hz  4 25 A       Maximum Power  805 Watts  115V   880 Watts  220 240V   Physical Dimensions  W  16 75     42 5 cm  x D  23     58 4 cm  x H  8 62     21 9 cm        Weight  40 Ibs   19 kg   Optional Accessories    6 foot extension tubes       Tripod       Outdoor Ambient Shelter with HVAC    Thermo Fisher Scientific    Chapter 2  Installation    The fo
226. SCII Responses        sssesseeeseseieieeererersrersesreres B 66  Format Specifier for Binary Responses       eeeeeeeererrseseeee B 66  Format Specifier for Erec Layout      eeeeeeeerserrssrseee B 67  Tte ae a a a e a aa B 67  Value SNE aereis erei eee TSE EA OE cut beast TESS ina B 67  Value Sourc aa a a a A atta tia as oat B 67  Alarm Informations issernida aan ia Aa ai B 68  Tratislati  n Table nne ds tterd lenn nnan RN RREAN i B 68  Selection lable ssssaisiiienceeccas dena Mandl A A NN B 68  Bitton Designat  t  rerrundnannnno nn eaaa ania B 68  Examples jicccc scutes n a AEE EE B 69  Appendix C MODBUS Protocols sisia ae aaas C 1  Serial Communication Parameters             ssssssccsssssssssssssssesssssnesenss C 1  TCP Communication Parairetersa  ic2sasnscencttiaaaia tesadhietwautevebeael tts C 2  Application Data Unit Definition viasacitinecciud niewdeienaineih C 2  Slive Addres Sannen a cee qusirond ond aaar aa C 2  MBAP Head  k annn sann naa TNN S C 2  F  nction Codema a E A E E S C 3  MVR AL cesta aida sare iad Ae dea es ues he Casta dase cee sada A Ee C 3  PRETO G NCG Kecareiteeterk ls A AS foal C 3  PU CGH OOS E TEE T E A E T TEE C 3   0x01 0x02  Read Coils Read Inputs        cceceesseeseeeeseeseeeeseeseneeaees C 3   0x03 0x04  Read Holding Registers Read Input Registers               C 5   0x05  Force  Write  Siig l    oil az  5 es ars sear tes stared uaitbe  cussccedduetiated C 7  MODBUS Parameters Supported wars  3 eer ee cise teeta eas C 8  Appendix D Gesytec  Bay
227. SHARP MIB  s p  BETA 2 DETECTOR  PJ17 Fy Pin LI  PJ20     WJ39    4 Pin  MJ33               TJCOUNTER WHEEL  138   gt    s    CICAM WHEEL  W31 bo PRESSURE BOARD  MI28  5 BOARD TEMP THERM     PJ3 m IEATER 1_120 VAC  a N       2 FIN      P  S  jo      ir  a ee PURGE 2120 VAC  PA r          L a TT MOTOR 120 VAC                                                                                                       PUS at AUX POWER 120 VAC  Zh    PJ45 ais         J21     FAN  Pi 2 Pin  aditi AUXHT  2 rin    PJ12 sa Bin   JAUXHT 2  PJ     PUMP  2 Pin  PJ14 PUMP 2  2 Pin  eee    JZ IO    2 Pir    PJ30   M   2 Pir    PJ36      C  A  TIF ER  MOD  Py a      Pin  x   J44 PJ43 5 Q PI IRCE 4 C  2 Pin a J18   x 4VD                                                                                                             AC  R a  MJ42     5   JTEST RTD   i SREY     MJ41     5    SAMPLE 1 TEMP  a In   yar   J40     JSAMPLE 2 TEMP  MJ19        ph PUMP 3 Pn      MJ23         IPUMP FLOW 2 S   e pin eae A  3 P S J10           44 TEMP 2  L o CISPARE FLO 3 Pin a  N I rae J      L_ zp O 1  V SoD 2  2 Pin  EAT   gt  a       Figure 6 2  Board Level Connection Diagram   Measurement System    Thermo Fisher Scientific Model 5030  SHARP Instruction Manual 6 9    Troubleshooting  Connector Pin Descriptions    Connector Pin  Descriptions    6 10 Model 50307 SHARP Instruction Manual    Figure 6 3  Board Level Connection Diagram   SHARP    The connector pin descriptions in Table 6   3 through Table 6 7 can be
228. SM Commands Supported  Command Description  E pump on  F filter change    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual E 3    
229. Send  mass coef  Receive  mass coef 7100 0    set mass coef value   This command sets the mass coefficient to user defined values to value   where value is a floating point representation of the coefficient  The  following example sets the mass coefficient to 7000     Send  set mass coef 7000  Receive  set mass coef 7000 ok    thermal mass coef  This command reports the current thermal mass coefficient  The following  example reports that the thermal mass coefficient is 19 000000     Send  thermal mass coef  Receive  thermal mass coef 19 000000    set thermal mass coef value   This command sets the thermal mass coefficient to user defined values to  value  where value is a floating point representation of the coefficient  The  following example sets the thermal mass coefficient to 20 5     Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Calibration    Send  set thermal mass coef 20 5  Receive  set thermal mass coef 20 5 ok    vac mass coef  This command reports the current vacuum mass coefficient  The following  example reports that the vacuum mass coefficient is 2 0     Send  vac mass coef  Receive  vac mass coef 2 000000    set vac mass coef value   This command sets the vacuum mass coefficient to user defined values to  value  where value is a floating point representation of the coefficient  The  following example sets the vacuum mass coefficient to 3 1     Send  set vac mass coef 3 1  Receive  set vac mass coef 3 1 ok  neph cal bkg    Th
230. Servicing  Detector Amplifier Assembly Replacement    2  Remove the SHARP optics cover assembly and the measurement case  assembly  as described earlier     3  Disconnect the cable from the detector amplifier assembly connector  on the measurement interface board and disconnect the SHV  connector from the detector amplifier     4  Remove the one screw holding the detector amplifier mounting bracket  and slide out the detector amplifier assembly     Unhook SHV connector    Unfasten screw and slide  detector amplifier out    Slide Out       Unhook connector from  Interface Board    Figure 7 7  Replacing the Detector Amplifier Assembly    5  Install the new detector amplifier assembly by following the previous  steps in reverse     6  Calibrate the detector amplifier board  Refer to the    Detector Amplifier  Board Calibration    procedure that follows     7 16 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Detector Amplifier  Board Calibration    External Pump  Replacement    Thermo Fisher Scientific    A    Servicing  Detector Amplifier Board Calibration    Note If Service Mode is not displayed  refer to    Accessing the Service  Mode    on page 7 4  then return to the beginning of this step  A    7  Calibrate the instrument  Refer to the    Calibration    chapter in this  manual     Calibrate the detector amplifier after replacing the detector amplifier board     Note This adjustment should only be performed by an instrument service  technician  A    1  Let 
231. Settings na reesen asea raa E aE EEEE 3 41  Us DHCP sec aceasta a a a a a ON i 3 42  PAd A L A E TE 3 42  PSG aS AAE E EEE EE E E achcuen Met edan tela ee tae 3 43  D  fa  lt Gateway ioi a oist as EEr E E R aS 3 43  Host Name o A NTN 3 43  Network Time Protocol Server        cccccccessccssssecesssecssssecesseeesseeeeeas 3 44   T O Configuratio enseing ern pe eee ney ort e ee cretrrr omen terest etre 3 44  Output Relay  Settings inise ete teas aa tenet ses 3 44  L  pie Stateside ae ae oiii 3 45  Tastriment State urena ea e eteei tase 3 45  eea EEEE E EAA E A E 3 46    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual ix    Contents    TNO ATA gps nn a a A NNR 3 46  Digital Input Settings       sesesesssesesseseseeesessrsesrsrtttrsrerrererereesesrseees 3 46  EE e T E abestolge tchutel dents eadao de ndaitbaaes 3 47  TiSte tiene ACHON sassscesiaivendeustsasinmsSiesuascal veut entenienn ese  3 47  Analog Output Configuration  Select Channel           cee 3 48  Allow Over Under Ratices sac5iiii celiac ak Goede a dieccdee dae seas  3 48  Analog Output Configuration  Select Action     cesses 3 49  Select ATG norio nd secs score cme e a Aeee EEVEE EERSTE 3 49  Minimum and Maximum Value   ssssssssseseseeeeseeeserrrrrrrererersrsesres 3 49  Choose Signal to Output   s sssssssssssesssesssesesesesrsrersrsrerrerrrereesesrseses 3 51  Analog Input Configuration      sesesesssesesesesesesesesesesrsrrrrrrerererersees 3 52  Descriptor n a a a a aN 3 53  LOE IEA E A TEE EE ET E EE 3 
232. Setup           _ Tubing Connector       Fasten Screws  2     Fasten Captive    To Rear Panel  Hardware  2     50307  H1             Fasten Nut    Figure 2   6  Heater Assembly  4  Attach a second tubing connector to the top of the heater tube     5  Carefully push additional 5 8 inch OD stainless steel tubing through  the roof flange and then drop into the tubing union on top of the  heater     6  Any exposed 5 8 inch OD stainless steel tubing should be insulated  with additional pipe insulation inside the shelter     7  Sleeve any additional water tight capping over the tubing that now  extends above the roof line     8  Sleeve the white Delrin plastic tripod adapter over the 5 8 inch OD  tubing to an appropriate height to accommodate the tripod and    2 14 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Installation  Setup    ambient T RH sensor assembly  Tighten this adapter against the  stainless steel tubing using the two setscrews     9  Sleeve the tripod over the tripod adapter to a proper height and tighten   If using a tripod  using the adjustable legs of the tripod  plumb the inlet  tube assembly     10  Attach the white ambient T RH sensor radiation shield assembly onto  a portion of the exposed tripod adapter     11  Attach and tighten the 5 8 inch to 3 8 inch reducing T fitting to the  5 8 inch stainless steel tubing     12  Place the PMio or PM2 s inlet into the 5 8 inch stainless T fitting     13  Feed one end of the black 25 foot T RH cable t
233. TABLE IF  DHCP IS ON     For more information on DHCP  see    Use DHCP     above     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP Settings  gt  Netmask        aa ee  a         Default Gateway The Default Gateway screen is used to edit the gateway address  The  default gateway can only be changed when DHCP is off  If DHCP is on   the instrument will respond with    NOT SETTABLE IF DHCP IS ON      For more information on DHCP  see    Use DHCP    above  Any traffic to  addresses that are not on the local subnet will be routed through this    address     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP Settings  gt  Gateway        Host Name The Host Name screen is used to edit the host name  When DHCP is  enabled  this name is reported to the DHCP server     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP Settings  gt  Host Name     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 43    Operation  Volumetric Conditions       Network Time Protocol The Network Time Protocol  NTP  Server screen is used to edit the IP  Server address of the NTP server  An NTP server may be used to periodically  synchronize the instrument   s real time clock with a standard  More  information about the NTP servers and a list of public servers may be  found at http   www ntp org        e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  TCP IP
234. TD  Sample1B_RTD    Ground          TEST RTD    AC IN    J42    J43     gt     n      w N    SpareA_RTD  SpareB_RTD    Ground       AC   AC   AC Ground       24V IN    J44    a     24V    Ground       DATA    Thermo Fisher Scientific    J45    Ground   RS485   RS485    Model 50307 SHARP Instruction Manual 6 19    Troubleshooting  Connector Pin Descriptions    Table 6   5  Front Panel Board Connector Pin Diagram       Connector Reference Pin Signal Description  Label Designator  MOTHERBOARD J1 1 Ground  2 Ground  3 LCLK    LCD Signal  4 Ground  5 Ground  6 LLP     LCD Signal  7 LFLM     LCD Signal  8 LD4     LCD Signal  9 LDO     LCD Signal  10 LD5     LCD Signal  11 LD1     LCD Signal  12 LD6     LCD Signal  13 LD2     LCD Signal  14 LD7     LCD Signal  15 LD3     LCD Signal  16 LCD Bias Voltagel  17  5V  18 Ground  19 Ground  20 LCD_ONOFF     LCD Signal  21 Keypad Row 2 Input  22 Keypad Row 1 Input  23 Keypad Row 4 Input  24 Keypad Row 3 Input  25 Keypad Col 2 Select  26 Keypad Col 1 Select  27 Keypad Col 4 Select  28 Keypad Col 3 Select  29 Ground  30 Ground  31 Ground  32 Ground  33  24V    6 20 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Troubleshooting  Connector Pin Descriptions    Connector Reference Pin Signal Description  Label Designator    w  BS     24V   LDO_5V     LCD Signal  LD1_5V     LCD Signal  LD2_5V     LCD Signal  LD3_5V     LCD Signal  LCD_ONOFF_5V    LCD Signal  LFLM_5V    LCD Signal  NC   LLP_5V     LCD Signal  LCLK_5V    LCD Signal
235. TUV RHEINLAND  ENERGIE UND UMWELT GMBH       Dakks    Deutsche   Akkreditierungsstelle   D PL 11120 02 00  Report on the suitability test of the ambient air  quality measuring system Model 5030  SHARP  with PM25 pre separator of the company Thermo  Fisher Scientific for the component suspended  particulate matter PM  5    TUV Report  936 21209885 F  Cologne  20 September 2013    www umwelt tuv de       teu service de tuv com    The department of Environmental Protection of TUV Rheinland Energie und Umwelt GmbH  is accredited for the following work areas     Determination of air quality and emissions of air pollution and odour substances     Inspection of correct installation  function and calibration of continuously operating emission measuring instru   ments  including data evaluation and remote emission monitoring systems     Combustion chamber measurements     Performance testing of measuring systems for continuous monitoring of emissions and ambient air  and of elec   tronic data evaluation and remote emission monitoring systems     Determination of stack height and air quality projections for hazardous and odour substances     Determination of noise and vibration emissions and pollution  determination of sound power levels and execu   tion of sound measurements at wind energy plants    according to EN ISO IEC 17025   The accreditation is valid up to 22 01 2018  DAkkS register number  D PL 1 1120 02 00     Reproduction of extracts from this test report is subject to writt
236. Using a NIST traceable barometer  measure the barometric pressure  and  Calibration    convert as necessary  in units of millimetres of mercury  mmHg   Use the  SPAN feature from the barometric calibration submenu        4 4 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Calibrate Vacuum Flow    Zero    Calibrate Vacuum    Thermo Fisher Scientific    Pressure Span    Calibration  Calibration Procedure       Be sure to save the entry and compare the values once more  Repeat as  necessary to within 2 mmHg     The Calibrate Vacuum Flow Zero screen calibrates the vacuum flow sensor  at zero value  This is done automatically with every filter tape change   However  if for any reason this needs to be done  proceed to the following  screen below to execute a zeroing filter tape change        The Calibrate Vacuum Pressure Span screen allows the user to view and set  the vacuum sensor calibration span point     To calibrate the vacuum sensor  zero a digital manometer  capable of  measuring up to 100 mmHg  and attach to the  APA port on the rear  panel  Then push the toggle switch on the right inward to open  see Figure  4   1   The manometer reading can now be used to calibrate the vacuum  sensor     Model 50307 SHARP Instruction Manual 4 5    Calibration  Calibration Procedure    Calibrate Flow Pressure  Span    4 6 Model 50307 SHARP Instruction Manual      RH   temp    7  O VACUUM   y             ame  eon Oe    Figure 4 1  Differential Pressure and Vacuum Calibrati
237. a cece tt dtesadd acatstore 7 18  Pressure Board Calibtatiany iiccdscssess sd iishussaddatdolvandentias dle disuedeeus 7 20  Analog O  tput Testigo uisa ienne a a A RA RR 7 21  Analog Output Calibration s  2 4 4 c00vicl nde enneaseia 7 24  Analog Input Calibration   s ssseseseeeseeeeesesesesssssssssssrsrsisisiseresesesesesrseses 7 25  Calibrating the Input Channels to Zero Volts   sssessseseeeeeeesssse 7 25  Calibrating the Input Channels to Full Scale wo    cee eee 7 26  Themistor Re pA Ce eN ts cata and spies oie cede aceaia Jase tand ese b cevecdangt tunica  7 26  I O Expansion Board  Optional  Replacement   0 0     eeseseeeereeeeeeees 7 27  Digital Output Board Replacement    cesseseseseeseeeeeeeeeeseeeeseeneeee 7 29  Motherboard Replacement  sivss2exis decstaccsadtei antec atest accteanes 7 30  Measurement Interface Board Replacement         sceeeseseeeereeeeeeneesees 7 31  Photo Interrupt Board Replactinentic 42 6 2ees ean ees 7 32  Proportional Valve Replacement    0 6 00 ay aseeta aie eta  7 33  Detector Assembly Replacement 2as 202 20s  200 4a s2ad  ty nea cieaiteaedeeeceeatess 7 34    Thermo Fisher Scientific    Contents    Flow RTO Rep acer tig senngsssnnisecni iadenin  7 36  Transformer Replacement   s sssesesssesesssssssssssesssssrsessststseststerteesesesrsees 7 37  Radius Tube Replacement   s s sssssesesesesessssssssssrsrsrsrsrsestseststsesesesesesessse 7 38  Ambient RH Temperature Replacement      sssssseeeeeeeeererererrrerrrrrsese 7 39  Front Panel Board Repla
238. adapter and the de   scribed procedure  It can otherwise damage the instrument     It is recommended to check the tightness of the measuring system on a quarterly basis ac   cording to the described procedure    6 4 Evaluation    The tightness was checked with the adapter as part of the laboratory test     The criteria specified by the manufacturer upon which the tightness check is considered  passed     maximum difference of 0 42 l min    2 5   of the flow rate  between the flow rates  measured by the instrument and by the external flow meter under application of the adapter      proved to be appropriate parameters for monitoring the instrument   s tightness  The labora   tory tests resulted in a maximum leakage  lt  1   of the nominal flow rate of 16 67 l min for  both candidates     The tightness check must only be performed with the tightness check adapter and the de   scribed procedure  It can otherwise damage the instrument     740300_2013_936_21209885F_EN docx    TUVR heinland   TUV Rheinland Energie und Umwelt GmbH   eintan Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 142 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    6 5 Assessment    The criteria specified by the manufacturer upon which the tightness check is considered  passed     maximum difference of  0 42 
239. age 31 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Both C  and Cys values are averaged in the same way over a variable time span  The aver   aging period depends on the variation coefficient of the mass concentration determined by  the nephelometer itself  This coefficient is calculated using a mathematical algorithm  in  which both the current nephelometer concentration and the concentration measured by the  nephelometer one second before are integrated  The coefficient is calculated continuously  and is averaged by a digital filter  one minute time constant   The resulting factor is then mul   tiplied by the maximum time constant of 720 min  In case of small and stable concentrations   the averaging period    time constant  of the radiometric measurement is automatically ex   tended to up to 12 h in order to increase the accuracy of mass concentration determination   For highly variable and large concentrations the process is exactly the opposite  i e  the av   eraging time is reduced to up to 20 min  In practical use  however  the time constant is sel   dom less than 4 hours     The C    Cre ratio of the equation above is the so called  auto  calibration factor  which ap   plies a real time calibration to the one minute average of the nephelometer signal Cy  In or   der to optimise this continuous mass calibration of the nephelometer signal  software has  been implemented in the Model 5030i SHARP measuring system  
240. al                                                                                  Command Description Page  alpha eff Reports sets the current value of alpha efficiency for the B 34  detector  alpha th Reports sets the current value of alpha threshold for the B 34  detector  amb rh Reports the ambient relative humidity B 23  amb rh offset Reports sets the current ambient relative humidity offset B 35  amb temp Reports the current ambient temperature B 23  amb temp offset Reports sets the current ambient temperature offset B 35  analog iout range Reports sets analog current output range per channel B 61  analog vin Retrieves analog voltage input data per channel B 61  analog vout range Reports sets analog voltage output range per channel B 61  avg24 time Reports sets 24 hour averaging time to start averaging the B 11  concentration  baro mass coef Reports sets the current barometric mass coefficient B 35  baro pres Reports the current barometric pressure B 24  baro span Reports sets the current barometric pressure span B 36  baud Reports sets current baud rate B 56  beta th Reports sets the current reference beta threshold for the B 36  detector  bkg pm Reports sets current PM background B 36  bkg sharp Reports sets current SHARP background B 37  board temp Reports the current board temperature B 24  cal baro pres span Automatically calculates the span coefficient B 37  cal flow pres span Automatically sets the flow pressure span B 37  cal flow span Automatically sets the fl
241. alarm  limits range from 5 to 60   C  If the flow temperature reading goes beyond  the minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  RH Termperature Alarms  gt   Flow Temp        rip gies   Piet          3 74 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu    Min and Max Flow The Minimum Flow Temperature alarm limit screen is used to change the  Temperature Limits minimum flow temperature alarm limit  The minimum and maximum  flow temperature screens function the same way     e Inthe Main Menu  choose Alarms  gt  RH Temperature Alarms  gt   Flow Temp  gt  Min or Max        Board Temperature The Board Temperature screen displays the current board temperature  value and sets the minimum and maximum alarm limits  Acceptable alarm  limits range from 5 to 60   C  If the board temperature reading goes beyond  the minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  RH Termperature Alarms  gt     Board Temp        Wi mb           Min and Max Board The Minimum Board Temperature alarm limit screen is used to change the  Temperature Limits minimum board temperature alarm limit  The minimum and maximum  board temperature screens function the same way     e Inthe Main Menu  choose Alarms 
242. allows the user to assign one of the  concentrations to the selected record field  The currently selected item is  shown by     lt       after it  Note that at this point  pressing indicates  that these are proposed changes as opposed to implemented changes  To  change the selected record format and erase record log file data  see     Commit Content    below  The display shows the current heater for the  mode selected  PM      e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Select Content  gt  select Field  gt  Concentrations        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 29    Operation  Volumetric Conditions    Other Measurements The Other Measurements screen allows the user to assign one of the other  available measurement types to the selected record field  The currently  selected item is shown by     lt       after it  Items displayed are determined by  the options installed  For Common Flags and Detector A Flags  see Figure  3   5 and Figure 3   6 for data descriptions  Note that at this point  pressing  indicates that these are proposed changes as opposed to  implemented changes  To change the selected record format and erase  record log file data  see    Commit Content    below     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Select Content  gt  select Field  gt  Other Measurements        DH vided  PLP i    3 30 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Analog
243. alues 2 18 UG MS          eee 149  Figure 47  Results of the parallel measurements with the candidates SN 1   SN 2   measured component PM    all test sites  values  lt  18 UG M       ee 150  Figure 48  Reference vs  candidate  SN 1  amp  SN 2  measured component PM25   UNM VUES Se acca tet a cena eee nen tea chen acetoacetate a ne serene eemateeaenenanaaae 158    Figure 49  Reference vs  candidate  SN 1  measured component PMps  all test sites     159  Figure 50  Reference vs  candidate  SN 2  measured component PMzps  all test sites     159  Figure 51  Reference vs  candidate  SN 1  measured component PMz2 5     Cologne  WINIET  saxtvecortserecacerstiveentoeterenedeinde anaE AAEE EOE aA EEA EKE EE anaes 160  Figure 52  Reference vs  candidate  SN 2  measured component PM25    Cologne TL  ac ccacte aus ostel cece aan anagsscdeacecengsadsduacdystecceceucesmessaceaaeasae  160  Figure 53  Reference vs  candidate  SN 1  measured component PM  5    Bornheim  summer   lt tc ierdee Sitidcam dese iiieieelisd tinea oe 161  Figure 54  Reference vs  candidate  SN 2  measured component PM  5    Bornheim  SUIMIMEN  sessirnar eee en neni A ERNE AEN RR 161  Figure 55  Reference vs  candidate  SN 1  measured component PM2s5    Bornheim  winter  a tceccukonecsatiertineieaniceoiccencea seam enoiereeiaxeumeisetetreiaiete 162  Figure 56  Reference vs  candidate  SN 2  measured component PMz2 5    Bornheim  winter  eneeenneer cere errr ee eer ee re eee ee nee ee ee eee 162  Figure 57  Reference 
244. and 3  measurement sites respectively     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 6 Detailed presentation of test results    A TUVRheinland      Precisely Right     Page 169 of 424    Table 38 shows the results of the evaluations of the equivalence check after the application    of the correction factor for the slope on the complete dataset     Table 38   rection    Summary of the results of the equivalence check  SN 1  amp  SN 2  after slope cor        29 1  217 ug m3    Twas  ee    PM2 5 5030i Sharp Orthogonal Regression    Slope Corrected    All Data 0 999     lt  18 ug m3 0 994        0 030  0 006    218 ug m3 1 020     0 030   0 803    Orthogonal Regression  Dataset    Slope  b      Up    Bornheim Winter 1 006     0 025   0 175    Cologne Winter 1 017     0 028   0 481    Individual Datasets    Bornheim Summer 1 037     0 031    0 469    Teddington Summer 1 043     0 047    0 590    Intercept  a      ua    Betw een Instrument Uncertainties    Slope  b      Up Intercept  a      Ug Reference Candidate        0 012   0 084     0 210        0 297      0 913   Limit Value of 30 ug m3  Wem       217 ug m3      0 624       0 633      0 437        0 381     lt  18 ug m3 157 0 
245. and TUV Rhein   land   It comprises the testing of the latest series of suspended particle measuring systems  by different manufacturers in the laboratory and at field test sites in the UK and in Germany     The minimum requirements were fulfilled during suitability testing     TUV Rheinland Energie und Umwelt GmbH therefore suggests publication as a suitability   tested measuring system for continuous monitoring of suspended particulate matter PM25 in    ambient air     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 18 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    1 2 Certification proposal    Due to the positive results achieved  the following recommendation is put forward for the no   tification of the AMS as a suitability tested measuring system     AMS designation   Model 5030  SHARP with PM25pre separator for suspended particulate matter PM25    Manufacturer   Thermo Fisher Scientific  Franklin  USA    Field of application     Continuous monitoring of suspended particulate matter of the PM25 fraction in ambient air   stationary operation     Measuring ranges during suitability test        Component Certification range Unit  PMo5 0     1000 g m                       Sof
246. and remove the  cover     2  Loosen the two screws of the radius mounting block using a 9 16 inch  hex driver     3  Remove radius tube SHARP adapter     4  Loosen the two mounting screws on the measurement head assembly   Remove tube by slightly turning side to side  while pulling     5  Verify the sealing o ring is in position before sliding the new radius  tube assembly into the measurement head                    6  Install the new radius tube assembly by following the previous steps in  reverse   k    Radius Tube Adapter   T  I UI i       o  6 fo  a  4     Unfasten Screws  2                          4      ti   __ Insulation  Pull Out Tube                                                          Unfasten Screws  2                                                                          Figure 7 21  Replacing the Radius Tube    Ambient Use the following procedure to replace the ambient RH temperature    RH Temperature 9  s  m gt Y    iss 7 22     Replacement  Pent Required     Ambient RH temperature assembly  Adjustable wrench    Pliers    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 39    Servicing  Ambient RH Temperature Replacement    7 40    Model 50307 SHARP Instruction Manual    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover  
247. andard VDI 4202 Part 1  September 2010      For all other certification ranges the group averages of measured values about the cal   ibration function shall not exceed 5   of the upper limit of the corresponding certifica   tion range     Note     By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems  For PM25 particulate measuring systems  this test shall be  performed according to minimum requirement 5 4 10    Calculation of the expanded uncertain   ty of the instruments        6 2 Equipment  See module 5 4 10  PMz5   6 3 Performance of test    For PM25 particulate measuring systems  this test shall be performed according to minimum  requirement 5 4 10    Calculation of the expanded uncertainty of the instruments       6 4 Evaluation  See module 5 4 10  PMz25   6 5 Assessment    For PM2 5 particulate measuring systems  this test shall be performed according to minimum  requirement 5 4 10    Calculation of the expanded uncertainty of the instruments        Minimum requirement fulfilled     6 6 Detailed presentation of test results  See module 5 4 10  PMz5     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 105 of 424  Scientific for the component suspended particulate
248. anuary 2010  4  the following 5 criteria must be  fulfilled to proof the equivalence     1  Of the full dataset at least 20   of the results obtained using the standard method shall  be greater than the upper assessment threshold specified in 2008 50 EC for annual  limit values        28 g m  for PM  and currently 17 g m    for PMo 5     2  The in between uncertainty of the candidate must be less than 2 5 yg m   for all data  and for two sub datasets corresponding to all the data split greater than or equal to and  lower than 30 ug m  or 18 g m  for PM  and PM25 respectively     The in between uncertainty of the reference method must be less than 2 0 g m8     4    The expanded uncertainty  Wc  is calculated at 50 ug m  for PM  and 30 g m8 for  PMz2 5 for each individual candidate instrument against the average results of the refer   ence method  For each of the following permutations  the expanded uncertainty must  be less than 25       e Full dataset     e Datasets representing PM concentrations greater than or equal to 30 g m  for  PMio  or concentrations greater than or equal to 18 g m  for PM25  provided that  the subset contains 40 or more valid data pairs     e Datasets for each individual test site     5  Preconditions for acceptance of the full dataset are that  the slope b is insignificantly  different from 1   b 1  lt  2 u b   and the intercept a is insignificantly different from 0    a   lt  2 u a   If these preconditions are not met  the candidate method may be ca
249. applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     TUVRheinland      Air Pollution Control 3    Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 171 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    7  Recommendations for practical use    Works in the maintenance interval  4 weeks    The following procedures are required to be undertaken at regular intervals    Regular visual inspection   telemetric monitoring   Check of system status   Check of error messages   Contamination check   Check of the instrument functions according to the instructions of the manufacturer  Maintenance of the sampling inlet according to the instructions of the manufacturer    For the rest  the manufacturer s directions and recommendations shall be followed     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Precisely Right     Air Pollution Control    Report on the suitability test of the ambient air quality measuring system    Page 172 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo    Fisher Scientific for the component suspended p
250. aring first   0   gt  local  1   gt  remote  etc    The selection list shows that only the first  two values  in this case  are to be shown to the user when the button is  pressed  The  T  button indicates full translation  input code to string  and  user selection number to output string       xC   This is a line that starts a new column  the  xC or  L      Comp 6 11x off on Tset temp comp  s n     This shows that the bitfield end  the second part of a bitfield specification   is optional  The bitfield will be one bit long  starting in this case at the  eleventh bit      Background  7f 8Bd ddd set 03 bkg  s n     This shows the use of indirect precision specifiers for floating point  displays  The background value is taken from the 7th element  and the  precision specifier is taken from the 8th  If the asterisk were not present  it  would indicate instead that 8 digits after the decimal point should be  displayed     Thermo Fisher Scientific    Serial Communication  Parameters    Thermo Fisher Scientific    Appendix C  MODBUS Protocol    This appendix provides a description of the MODBUS Protocol Interface  and is supported both over RS 232 485  RTU protocol  as well as TCP IP    over Ethernet     The MODBUS commands that are implemented are explained in detail in  this document  The MODBUS protocol support for the iSeries enables the  user to perform the functions of reading the various concentrations and  other measurement values  read the status of the digital outputs of 
251. articulate matter PM2 5   Report No   936 21209885 F    Further maintenance works  The following works are necessary in addition to the regular works in the maintenance inter     val     Inspection of filter tape supply     one filter tape is enough for approx  650 filter tape  changes per roll  approx  215 days at cycles of 8 hours   Semi annual filter tape  changes are expected to be sufficient under normal conditions  no unscheduled filter  changes due to significantly high dust concentrations   The    Filter Tape Counter    func   tion allows for monitoring the current number of filter tape changes as well as configur   ing alarm thresholds which remind the user of upcoming changes    An inspection of the sensors for ambient temperature and ambient pressure shall be  carried out every three months according to the manufacturer   s specifications    A flow rate check must be performed every three months according to the manufac   turer   s specifications    The system   s leak tightness shall be inspected every three months according to the  manufacturer s specifications    The external pump exhaust filter shall be exchanged every six months    As part of an annual basic maintenance  the carbon vanes of the vacuum pump shall  be exchanged and the SHARP optics module shall be cleaned once a year  The  cleaning of the SHARP optics module is ideally performed by an experienced service  technician  Furthermore  it is recommended to carry out a yearly inspection and  if re   qu
252. articulate matter is 25    7     6 5 Assessment   The candidate systems fulfil the requirements on data quality of ambient air quality meas   urements during the test even without application of correction factors  A correction of the    slope leads nevertheless to a significant improvement of the expanded measuring uncertain   ty of the full data comparison     Minimum requirement fulfilled  yes    740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 168 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The evaluation of the full dataset for both candidates delivers  nevertheless  a significant  slope  see Table 35      The slope for the full dataset is 1 068 with an uncertainty of the slope of merely 0 013     The 2006 UK Equivalence Report  8  highlighted that this was a flaw in the mathematics re   quired for equivalence as per the November 2005 version of The Guidance as it penalised  instruments that were more accurate  Appendix E Section 4 2 therein   This same flaw is  copied in the January 2010 version  It is the opinion of TUV Rheinland and their UK partners  that the Model 5030i SHARP measuring system for PM2 5 is indeed being penalised by the  mathematics for being accurat
253. as  Parameters of the calibration and analysis function  measured   component PMs s eisisssi enie r ee ed ect dees  Deviation between reference measurement and candidate on days   with a relative humidity  gt  70    measured component PMos           ccccceeeeeeeee  Comparison candidate SN 1 with reference instrument  rel  humidity    gt  70    all test sites  measured component PMo65              cccceeeeeeeeeeeeeteeeeeeees  Comparison candidate SN 2 with reference instrument  rel  humidity    gt  70    all test sites  measured component PMo5               0 csesseseseeeesesseseeees  Results of flow rate GNOCKS iiscsvioinienicciveeitedinnisaneada tivsded anbelsdiaeidadeuetinedeaeabees    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 9 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    Table 31  Parameters for flow measurement  24 h average   SN 1  field              008 140  Table 32  Parameters for flow measurement  24 h average   SN 2  field             006 140  Table 33  Results Of the tightness CHECK  ccisiccissoiictdevtinvel sscl esi eictiasatniniedebeateeieees 142  Table 34  Uncertainty between systems under test Up  for candidates SN 1   and SN 2  measured Component PMas      
254. as before basing on  the minimum requirements stated in VDI 4202 Part 1 of 2002  i e  using the reference values  Bo  B  and Bz   The performance of the test points 5 3 3  5 3 4  5 3 13 and 5 3 18 is omitted   as they are not relevant for particulate measuring devices     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 101 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 3 2 Repeatability standard deviation at zero point    The repeatability standard deviation at zero point shall not exceed the requirements of  Table 2 in the certification range according to Table 1 of Standard VDI 4202 Part 1   September 2010     In case of deviating certification ranges  the repeatability standard deviation at zero  point shall not exceed 2  of the upper limit of this certification range     Note     This item cannot be evaluated according to the current versions of Standards VDI 4202 Part  1  September 2010  and VDI 4203 Part 3  September 2010  when testing particulate meas   uring systems  due to the fact that a minimum requirement has not been defined  By resolu   tion of the comp
255. ase Log     Cae RG 0 PATCH gy mpi CREEL TTI g guppi    EN EEE Eut Bone ETET Fad Bast    abe POE Sek aet PPh EE E EAA a E i L E ELLS   P   Eba i EE EE       Select Content The Select Content submenu displays a list of 32 record fields to use to  configure the data to be logged  When a field is selected  a submenu list  appears of the items to choose from to assign to those fields  Item types are  Concentrations  Other Measurements  and Analog Inputs  if the I O  expansion board is installed   These record fields comprise a temporary list  of items for the selected record type that must be committed via the  datalogging menu before the changes will apply  Note that committing any  changes to this list will erase all currently logged data for the selected record  type only  as the format of the stored data is changed     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Select Content     3 28 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions         ee Clie lac  et i   mF het i PS       Choose Field Data The Choose Field Data submenu displays a list of the types of data that can  be logged for the current field  Choices are Concentrations  Other  Measurements  and Analog Inputs  if the I O expansion board is installed      e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Select Content  gt  Field 1 32        Wi hiv   A a E PEPE       Concentrations The Concentrations screen 
256. asuring range of 0     10 000  g m  is set by default  It is possible  to adjust other measuring ranges     The upper limit of the measuring  range of the measuring system is  greater than the respective upper limit  of the certification range     Negative measured signals are dis   played directly on the AMS and are  transmitted correctly via the respec   tive signal outputs of the measuring  system     All instrument parameters are pro   tected against loss through buffering   After the power supply has returned   the measuring device goes back to  failure free operational condition and  resumes measurement  However  at  least the first 12 h worth of data after  voltage returns shall be discarded   since the system needs an equilibra   tion phase after restart until the BETA  detector adjusts to ambient condi   tions     The measuring systems can be ex   tensively monitored and operated  from an external PC via modem or  router     In principle  all necessary operations  for performing a functional check can  be monitored directly on the system  or via telemetric remote control     The maintenance interval is defined  by the necessary maintenance proce   dures and has been set to 1 month        TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control   s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 21 of 424   Scientific 
257. at reference value B  caused by normal changes  in the mains voltage in the interval  230  15  20  V shall not exceed Bo     6 2 Equipment  Adjustable isolating transformer and span foil set for testing the reference point   6 3 Performance of test    In order to test the dependence of the measured signals on supply voltage  the voltage was  reduced from 230 V to 210 V and then increased with an intermediate step of 230 V to 245  V     The reference point test comprised a check of the mass coefficient in order to test the stabil   ity of the sensitivity of both candidates SN 1 and SN 2     Since the AMS is not intended for mobile application  a separate check of the dependence of  the measured signals on the mains frequency was omitted     6 4 Evaluation    The percentage change of the determined measured values for the mass coefficient was ex   amined at each voltage step in relation to the default value at 230 V     It should be noted that concentration values could not be simulated by checking the mass  coefficient  It was therefore not possible to examine the range of B4     740300_2013_936_21209885F_EN docx    a 3    TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland  gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 112 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 212098
258. ate matter PM2 5     Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 185 of 424                Annex 5 Measured values from field test sites  related to actual conditions Page 4 of 11  Manufacturer Thermo Fisher Scientific  PM2 5   Type of instrument Model 50301 SHARP Measured values in g m     ACT   Serial No  SN1 SN2   No  Date SN 1 SN 2 Remark Test site   PM2 5 PM2 5   m l  m    91 8 22 2011 10 9 10 8 19 7 19 3 55 8 12 4 11 8 Bornheim  Summer   92 8 23 2011 19 2 19 1 29 9 30 1 63 7 23 0 20 5   93 8 24 2011 8 7 7 4 16 9 16 7 41 9 6 2 5 7   94 8 25 2011 11 5 12 1 18 8 18 6 63 4 11 4 9 2   95 8 26 2011 4 9 5 5 10 7 10 7 48 8 5 1 4 4   96 8 27 2011 2 1 1 9   97 8 28 2011 Fi 7 6 2 9 3 4   98 8 29 2011 5 9 6 2 11 4 11 5 53 0 6 1 5 6   99 8 30 2011 9 1 8 1 17 1 16 6 51 1 9 3 8 0   100 8 31 2011 14 5 13 9 26 0 23 6 57 2 14 5 14 2   101 9 1 2011 17 7 18 2 27 5 26 1 66 9 14 3 15 1   102 9 2 2011 14 9 15 0 25 1 24 1 60 6 11 7 12 0   103 9 3 2011 14 1 14 3   104 9 4 2011 8 2 8 0 12 7 12 1 65 3 6 1 9 9   105 9 5 2011 4 8 5 0 9 2 9 1 53 8 3 4 2 8   106 9 6 2011 5 2 5 6 11 1 10 6 49 8 2 9 3 3   107 9 7 2011 6 1 5 8 12 5 13 2 46 1 5 5 6 0   108 9 8 2011 3 0 3 3   109 9 9 2011 6 8 7A 12 1 11 8 57 9 6 2 6 5   110 9 10 2011 7 6 8 1   111 9 11 2011 5 7 55 9 4 9 3 59 5 4 9 5 5   112 9 12 2011 5 2 6 1 11 6 11 6 48 5 3 6 4 5   113 9 13 2011 6 3 7 2 16 3 16 8 40 6 4 4 6 0   114 9 14 2011 6 7 7 2 15 3 15 6 44 9 4 8 7 4   115 9 15 2011 11 1 12 2 24 3 24 9 47 4 9 9 10 3   
259. ature  thermistor  Figure 7   12      Equipment Required     Thermistor assembly    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    Thermo Fisher Scientific    Servicing  1 0 Expansion Board  Optional  Replacement    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Squeeze the thermistor latch and pull the thermistor assembly from the  AMB TEMP connector on the measurement interface board     3  Snap the new thermistor into the AMB TEMP connector        Thermistor    Figure 7   12  Replacing the Thermistor    1 0 Expansion Board Use the following procedure to replace the optional I O expansion board   Optional   Figure 7 13      Replacement   Pment Required     I O expansion board  Nut driver  3 16 inch    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    f   Equipment Damage Some internal components can be damaged by small    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 27    Servicing  1 0 Expansion Board  Optional  Replacement    2  Unplug the I O expansion board cable from the EXPANSION BD    connector on the motherboard     3  Remove the two standoffs holding the I O expansion board connector  to the rear panel  Figur
260. autions  see the    Servicing       chapter  A   1  Remove left side panel    2  From the main menu choose Instrument Controls  gt  Filter Tape  Control  gt  Manual  gt  Bench s   Press  e     3  Loosen both reel nuts on tape spindler  Remove used filter tape and  empty tape spool from tape spindles    4  Insert new filter tape on left tape spindle and tighten reel nut    5  Insert new blank pick up spool on right tape spindle    6  Route tape according to Figure 5 5    7  Attach end of tape to blank tape spool on right tape spindle with 2 inch  of tape    8  Turn blank tape spool until there is two complete wrappings of tape    9  Tighten right reel nut  From Filter Tape Control  gt  Manual  gt  choose  Tape and press to MOVE tape    10  From the main menu choose Alarms  gt  Instrument Alarms to verify  that there are no filter tape alarms    11  Replace left side cover     Model 50302 SHARP Instruction Manual 5 7    Preventive Maintenance  Greasing the Cam       Greasing the  Cam    External Pump  Exhaust Filter    A    5 8 Model 50307 SHARP Instruction Manual    Figure 5   5  Replacing the Filter Tape    Use the following procedure to grease the cam   Equipment Required   Dow Corning Molykote G N    1  Apply metal assembly paste around the parameter of the cam     It is recommended to replace the external pump exhaust filter every six  months  On an annual basis  remove top plate of optics and vacuum out  optic chamber per the following procedure or send it for service and
261. ay no 1 logic to normally open     Note If the command is sent without an appended relay number  then all  the relays are assigned the set logic of normally open closed   amp     Model 50307 SHARP Instruction Manual B 65    C Link Protocol Commands  Record Layout Definition    Record Layout  Definition    Format Specifier for  ASCII Responses    Format Specifier for  Binary Responses    B 66 Model 50307 SHARP Instruction Manual    Send  set relay open 1  Receive  set relay open 1 ok    The erec  lrec  and srec layouts contain the following    e A format specifier for parsing ASCII responses   e A format specifier for parsing binary responses   In addition to these  the erec layout contains    e A format specifier for producing the front panel display screens    Values are read using either the ASCII or binary format specifiers and  converted to uniform internal representations  32 bit floats or 32 bit  integers   These values are converted into text for display on the screen  using the format specifier for the front panel display  Normally  the  specifier used to parse a particular datum from the input stream will be  strongly related to the specifier used to display it  such as  all of the floating  point inputs will be displayed with an  f output specifier  and all of the  integer inputs will be displayed with a  d  specifier      The first line of the layout response is the scanf like parameter list for  parsing the fields from an ASCII erec response  Parameters are separa
262. ayout layout changed indicator           B 59  le  left  Simulates pressing left pushbutton B 44  list din Lists current selection for digital inputs B 64  list dout Lists current selection for digital outputs B 64  list Irec Lists current selection Irec logging data B 29  list sp Lists current selection in the scratchpad list B 29  list srec Lists current selection srec logging data B 29  list stream Lists current selection streaming data output B 29  list var aout Reports list of analog output  index numbers  and variables B 64  list var din Reports list of digital input  index numbers  and variables B 64  list var dout Reports list of digital output  index numbers  and variables B 64       Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands                                                                                        Commands   Command Description Page  list var log Reports a list of datalogging index numbers and variables B 64  Ir Reports the last Irec stored B 29  Irec Reports maximum number of Irecs B 29  Irec format Reports sets output format for Irecs  ASCII or binary  B 30  Irec layout Reports current layout of Irec data B 31  Irec mem size Reports maximum number of Irecs that can be stored B 31  Irec per Reports sets Irec logging period B 31  malloc Irec Reports sets memory allocation for lrecs B 32  malloc srec Reports sets memory allocation for srecs B 32  mass Reports the presence of mass B 26  mass coef Reports sets the curr
263. be omitted     A special adapter for connecting an external flow measuring instrument is mounted to the in   let in order to check the leak tightness of the measuring system  The leak test adapter pro   duces a high vacuum in the system  The leak tightness of the measuring system can be  evaluated by comparing the flow rate measured by the system  at high vacuum  and the flow  rate at the inlet    external flow measuring instrument   If the difference exceeds the manu   facturer   s specification of 420 ml min    approx  2 5   of the nominal flow rate   the system  shall be tested for leaks     740300_2013_936_21209885F_EN docx    TUVRheinland      Precisely Right     Page 48 of 424    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F       Figure 18  Leak test adapter    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    TUVRheinland      Precisely Right     Report on the suitability test of the ambient air quality measuring system    Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher    Scientific for the component suspended particulate matter PM25     Report No   936 21209885 F    Table 3 contains a list of important technical data of the ambient air
264. be set to 1 500 ug      Filter Tape   Next Time  this should have a start date      Filter Tape Period should be set to 8 hours for PM2 s5and PMoo     U S  EPA applications may adjust this value as high as 24 hours  for PMio only  However  8 hours is recommended for  consistent performance     Volumetric Conditions should be set to ACT for both Temp    and Press    Lrecs and Srecs should be chosen for datalogging and all  parameters should be selected in the order they appear from     Concentrations    to    Other Measurements     Default Lrecs are    Thermo Fisher Scientific    Data Content    Shutdown    Important Tips    Thermo Fisher Scientific    Installation  Data Content    set to 60 minute logging with 30  memory and Srecs are 5   minute logging with 70  memory allocated  If there are any  changes  be sure to Commit Content     b  From the Main Menu  choose Service  Under Service     Note If Service Mode is not displayed  refer to    Accessing the Service  Mode    on page 7 4  then return to the beginning of this step  A    m       Check Calibrate Ambient Temperature  Nephelometer    Temperature  and Flow Temperature    Check Calibrate Nephelometer Ambient RH    ne    ii          iii  Check Calibrate Barometric Pressure Span  iv  Check Calibrate Flow Calibration    v  Perform a Beta Auto Detector Calibration  30 minutes     Perform a Mass Foil Calibration  factory calibrated     me     vi     vii  Mass Coefficients are set at the factory          viii  Perform a Nephe
265. between pump intake and the  vacuum port on rear panel and include the push to connect    T     fitting with the 0 040    thru hole on the T side in this line     b  From the valve assembly that is now attached to the rear panel   connect the threaded fitting to the pump exhaust     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 10 7    Model 5014i to 5030i Upgrade Procedures  SHARP External Valve Installation    c  Connect the 3 8    tubing with inline HEPA filter to the available  port  Port  1  on the external valve assembly using the available  push connect fitting     Note This step can also be delayed until after the zero air delivery tube and  T fitting are installed at the sample intake  a    d  At the sample inlet  attach the stainless steel T fitting to the sample  downtube and tighten the swagelok fitting with Teflon ferrules     do not overtighten these ferrules     e  Snake the green vacuum tubing downwards towards the rear valve  assembly and attach to Port  1 of the valve  If this is a field  installation  proper water sealants should be used to avoid any  leakage into the instrument        Figure 10 8  Plumbing Diagram    10 8 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Model 50141 to 50301 Upgrade Procedures  SHARP Optics Assembly Installation    SHARP Optics Use the following procedure for SHARP external valve installation   Assembly    Installation 1  Remove the SHARP Optics Assembly port protection caps and plugs     2  
266. bly Cover    Model Label Use the following procedure for model label change   Change    1  Remove the 50147 label and replace with the new 50302 SHARP Label    on the front panel Front Panel in accordance with Figure 10 13 below        FRONT FRAME  LOWER FIXING HOLE LABEL   REFERENCE   101614 00 JIG    Figure 10 13  Front Panel Label Replacement    Power On Upgrade is complete   Instrument    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 10 11    Thermo Fisher Scientific    Appendix A  Warranty    Seller warrants that the Products will operate or perform substantially in  conformance with Seller s published specifications and be free from defects  in material and workmanship  when subjected to normal  proper and  intended usage by properly trained personnel  for the period of time set  forth in the product documentation  published specifications or package  inserts  Ifa period of time is not specified in Seller   s product  documentation  published specifications or package inserts  the warranty  period shall be one  1  year from the date of shipment to Buyer for  equipment and ninety  90  days for all other products  the  Warranty  Period    Seller agrees during the Warranty Period  to repair or replace  at  Seller s option  defective Products so as to cause the same to operate in  substantial conformance with said published specifications  provided that   a  Buyer shall promptly notify Seller in writing upon the discovery of any  defect  which notice shall i
267. cal systems with the serial numbers   System 1  SN 1  System 2  SN 2    The following performance criteria were tested in the field     e Comparability of the candidates according to the Guide    Demonstration of Equiva   lence of Ambient Air Monitoring Methods      e Comparability of the candidates with the reference methods according to the Guide      Demonstration of Equivalence of Ambient Air Monitoring Methods      Calibratability  analytical function   Reproducibility   Zero drift and sensitivity   Dependence of the measured values on sample humidity   Maintenance interval   Availability   Constancy of sample volumetric flow   Overall uncertainty of the test candidates    The following auxiliary devices were used during the field test     e Measuring cabinet provided by T  V Rheinland  air conditioned to approx  20   C   e Measuring cabinet for low volume samplers provided by DEFRA  UK   air conditioned  to approx  20   C   e Weather station  WS 500 manufactured by ELV Elektronik AG for use in Germany   and MK III Series manufactured by Rainwise  US  for use in the UK  for the determi    nation of meteorological characteristics such as air temperature  air pressure  air hu    midity  wind velocity  wind direction and precipitation   2 reference samplers LVS3 for PM25 in accordance with point 5   1 gas meter  dry   1 mass flow meter Model 4303  manufactured by TSI    1 mass flow meter type BIOS volumeter  manufactured by BIOS  for use in the UK   1 mass flow meter type T
268. ce mode     The main menu for operating the measuring system can be reached by pressing Ca   This button always leads to the previous sub menu     The four keys directly below the LCD display are programmable soft keys     by default  the  keys lead to the sub menus    RANGE        AVG        DIAGS    and    ALARM        The sub menu    Range    allows for setting the unit the concentration values are to be output in  and adjusting standard as well as customised measuring ranges     In the sub menu    Integration Time    it is possible to adjust the averaging period of the radio   metric measurement within the range 15     60 min  by default set to 20 min      The sub menu    24 Hour Average    allows the user to define the start time of the 24 hour av   eraging period of the concentration measurement     The    Calibration Factors    sub menu displays the calibration factors adjusted in the measuring  system as well as those determined during manual and or automatic calibrations  It is also  possible to adjust them using this menu     The    Instrument Controls    sub menu may be selected to control various instrument opera   tional parameters  It allows for parameterising the flow rate  turning the pump on and off   configuring the sample heater DHS  adjusting filter tape control operations  moving the tape  manually  adjusting a reliable mass limit  gt  during test  1500 ug  setting a time and interval  for the next filter tape change  etc    determining volumetric conditi
269. ce mode for improving the  calibration of mass measurement    9  Implementation of the    Infinite Mass Accumu   lation Method                   740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     Page 52 of 424    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F                V01 00 01 197 1  Release of Model 5030i code Installed during field  2  Integration time variable for mass and con    test  centration determination  range 15     60 min   added  3  Necessary code adjustments for the amplifi   er  due to the change from batteries to ca   pacitor for buffering high voltage supply for  up to 2 h during power outages  4  Increase of filter tape movements per roll by  approx  50    5  Change of filter tape after three failed at   tempts no longer possible  6  Elimination of minor bugs  V01 00 02 199 1  Calibration coefficient is frozen at c  BETA    Not installed   lt  2 5 g m  and or c  NEPH   lt  1 yg m  in   stead of c  BETA   lt  2 5 ug m   and or c   SHARP   lt  1 g m   2  Flow temp replaced by sample temp in Bay   ern Hessen protocol  3  Increase of accuracy of    diag volts    com   mand  V01 00 03 225 1  Modification of key variables and soft keys   Installed during la   for USB keypa
270. cecsceseeteeteteseeteeeeseeee 10 3  Figure 10 3  Filter Tape Installation PathWay        0  ccccccceceesceteeesteteseseees 10 4  Figure 10   4  Connecting Connection Bracket Wiring to Measurement Interface   ate Fe  2 eRe eae RRS ne E EE E nae aM Pen E 10 5  Figure 10 5  Radius Tube Adapter Install 0    ceceesesceteesesteneseseees 10 6    Thermo Fisher Scientific    Thermo Fisher Scientific    Figures    Figure 10   6  Radius Tube Adapter Alignment Check    ccc 10 6  Figure 10 7  Rear Valve Assembly Installation     0     ccccccececcscecesesesteteseseees 10 7  Figure 10 8  Plumbing Diagram 0 0    cecccccccc cece cesses eeeseeseseesesteeeneeeeeseeees 10 8  Figure 10   9  Optics Installation    ceceeseesseceseessesseeseeseeseeeseeseeseseeseneeneeees 10 9  Figure 10   10  Connecting the Optics to the Optics Connector           0     0 10 9  Figure 10 11  Upgrading 50307 Heater Assembly               c cceeeceeeeeeeeeeeeees 10 10  Figure 10 12  Connecting Heater to Optics Assembly Cover             ccee 10 11  Figure 10 13  Front Panel Label Replacement    ccceccceceesceteeseees 10 11  Figure B  1  Flags Field ccunseetensiec anew Sawees ne ana See nes B 12    Model 50307 SHARP Instruction Manual xxi    Figures    xxii Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Tables    Table 1 1  Model 50307 Specifications oo    c ccccccccccccscscscesestesesteseetesesteseeseees 1 8  Table 2 1  Pressure Units COnVErSsiON         ccccccccsessestese
271. ceive  diag volt mib 3 3 3 3 5 0 7 2  5 0  7  flow    This command reports the current sample flow  The following example  reports that the sample flow is 16 69 LPM     Send  sample flow  Receive  sample flow 16 69 LPM  flow pres    This command reports the current flow pressure for the detector  in  mmHg  The following example reports that the flow pressure is 24 1  mmHg     Send  flow pres  Receive  flow pres 24 1 mmHg  flow temp    This command reports the current flow temperature for the detector  in  degrees C  The following example reports that the flow temperature is 22 3    C    Send  flow temp  Receive  flow temp 22 3 deg  io bd    This command reports the presence of the I O expansion board  yes no    The following example reports the I O expansion board is present  yes      Model 50307 SHARP Instruction Manual B 25    C Link Protocol Commands  Diagnostics    B 26    Model 50307 SHARP Instruction Manual    Send  io bd  Receive  io bd yes  mass    This command reports the presence of mass  The following example  reports that the mass is 0 1 mg m        Send  mass  Receive  mass 0 100E 00  neph bkg    This command reports the current nephelometer background values stored  on the nephelometer interface board  The following example reports the  nephelometer auto ranging background values     Send  neph bkg   Receive  neph bkg1 0 00272 0 00272 0 01099 0 04351 0 17430  0 65535   neph mode    This command reports the current operating mode of the nephelometer   The followi
272. cements ic25  eel  hind ite ermal aoe 7 41  LCD Module Replacement  lt cacste 8 6 Gliese Ge cad 7 42  Removing the SHARP Optics Covet siic isascvtsesiccsssgndeciontderaentavess 7 44  External Solenoid Valve Replacement  lt cissie ccestesecccnscestca cesses 7 45  SHARP RH Temp Sensor Replacement  ssisecisciussiaaicesltacctenercaie 7 46  SHARP Interface Board Replacement ioc  iteccascect sezavecndes a sonteteatete 7 47  Delrin Nut Replacements iesse cet REI a 7 49  Internal SHARP Cable Assembly Replacement    sssssssseseseeseeeeeeeeeese 7 51  Case Cable Repl  cemient nnmnnn ia aie 7 52  External Zeroing Filter       eeeeeseeeseseseseeeseseseeesrsrsesrsrsrsrertrtrererrerererseses 7 53  Service OC ALIO MS nn nei a cpt a a A R E 7 54  Chapter 8 System Description            ssscsssssccssscsssssessssceesseesssscnesseecsessnessesenessnesseecersanesseees 8 1  TLE wy ae ree teu tcp vcesrtca sais sonne vast Cimireteyeediat tnt cusostnseenatnianguatatt i iansatie  8 1  Nephelometer Assembly  2  c0   gcc tisercstnwe jedi acca tiactiiagiecndihe 8 4  Primary Measurement Head 5 ccdedsictstetoicacsensat calens tieedeasveh anedeheneahvoes 8 4  Detector AMPUTET nats sac ots eden oe ce cae aeiee eon EE E cote  8 4  Cam Photo Interrupt Board gases eat he tesaen nua eee 8 4  Counter Wheel Interrupt Board Assembly          ee eeeeseeecreeteeeeneeeeeees 8 4  Proportional Valves  orasini senarai sheans 8 4  Cam MOO a oreen E EEE E EATE D O A e 8 4  EE OI E E 8 4  Pressure Board arsson erpen oann n a
273. cient time to equilibrate to room temperature  due to temperature variations during shipping and or storage  A    2  Following a sufficient equilibration period  connect the power cord to  the rear of the instrument and to a properly grounded power supply     Thermo Fisher Scientific    Thermo Fisher Scientific    Installation  Acceptance Testing and Startup Procedures    Note If the Model 50307 turns ON at this point  set the power switch to  the OFF position and then proceed  A    3  Connect the pump intake to the vacuum port mounted on the rear  panel of the instrument using the 3 8 inch vacuum tubing assembly  with the plastic T fitting  Connect the pump exhaust tube to Port  2  on the 3 way solenoid on the rear panel     4  Connect Port  1 of the rear panel 3 way solenoid to 3 8 inch  compression fitting on the 5 8 inch x 3 8 inch reducing T fitting using  the 3 8 inch vacuum tubing assembly with HEPA filter     5  Connect the vacuum pump power cord to the white circular grounded  power supply  labeled PUMP  on the rear panel of the instrument     6  Connect the 4 pin Temperature RH cable to the 4 pin connector  labeled RH TEMP on the rear of the monitor     7  Be sure that the pressure sensor calibration port toggle switches on the  rear panel are pushed outward and away from the barbed     Delta P     8  Set monitor power switch to the ON position     During startup  the   Series splash screen will be shown followed by a  mechanical filter tape change and a cycling of 
274. cientific    Analog Input  Calibration    Calibrating the Input  Channels to Zero Volts    Thermo Fisher Scientific    Servicing  Analog Input Calibration    6  Press        to return to the previous screen   7  Press         to select Calibrate Full Scale     8  Use   t    until the meter reads the value shown in the Set  Output To line  then press to save the value     Use the following procedures to calibrate the analog inputs after replacing  the optional I O expansion board  These procedures include selecting  analog input channels  calibrating them to zero volts  and then calibrating  them to full scale using a known voltage source     Use the following procedure to calibrate the input channels to zero volts     1  From the Main Menu  choose Service  gt  Analog Input Cal   The Analog Input Cal screen appears     Note If Service Mode is not displayed  refer to    Accessing the Service  Mode    on page 7 4  then return to the beginning of this step  A    2  At the Analog Input Cal screen  press  _    to scroll to a channel  and  press   J     3  With the cursor at Calibrate Zero  press ce    The screen displays the input voltage for the selected channel     4  Make sure that nothing is connected to the channel input pins and  press to calibrate the input voltage on the selected channel to    zero volts     The screen displays 0 00 V as the voltage setting     5  Press  a   gt      to return to the Analog Input Cal screen and  repeat Steps 2 through 4 to calibrate other in
275. cisely Right     Page 189 of 424                                          Annex 5 Measured values from field test sites  related to actual conditions Page 8 of 11  Manufacturer Thermo Fisher Scientific  PM2 5  Type of instrument Model 50301 SHARP Measured values in pg m   ACT   Serial No  SN1 SN2  No  Date Ratio SN 1 SN 2 Remark Test site  PM2 5 PM10 PM2 5 PM2 5   m gt   m   211 2 21 2012 9 9 10 5 16 3 17 4 60 4 11 6 11 0 Bornheim  Winter  212 2 22 2012 13 3 14 4 20 8 22 2 64 5 15 4 16 0  213 2 23 2012 8 3 8 6 9 1 10 2 87 3 11 6 8 6  214 2 24 2012 12 1 11 6 14 7 14 9 80 1 13 0 12 3  215 2 25 2012 15 7 14 5  216 2 26 2012 26 5 26 4 38 6 39 6 67 7 29 3 27 6  217 2 27 2012 19 6 18 9 26 0 25 6 74 6 20 6 17 7  218 2 28 2012 18 7 16 0 21 8 21 7 79 8 16 1 16 0  219 2 29 2012 27 6 26 5 34 3 33 7 79 5 20 7 21 8  220 3 1 2012 19 6 18 0 29 1 29 8 63 7 18 8 17 3  221 3 2 2012 15 9 13 3 18 9 18 0 78 8 12 9 13 3  222 3 3 2012 23 4 21 1  223 3 4 2012 31 5 30 0 42 8 41 3 73 2 26 8 24 1  224 3 5 2012 6 1 4 6 8 1 TA 70 6 5 8 6 1  225 3 6 2012 15 2 12 8 21 5 20 5 66 7 14 6 15 0  226 3 7 2012 15 5 14 5 24 8 22 4 63 4 17 1 15 3  227 7 17 2012 4 2 4 5 8 5 8 9 50 1 2 6 6 9 Teddington  Summer  228 7 18 2012 3 8 4 2 8 9 9 4 44 2 2 8 9 9  229 7 19 2012 4 6 4 6 9 1 8 9 51 2 4 9 3 8  230 7 20 2012 5 1 49 10 3 10 8 47 7 4 8 6 4  231 7 21 2012 5 8 6 1 10 9 10 7 55 0 ZT 7A  232 7 22 2012 7A 6 8 10 6 10 6 65 3 7 7 8 1  233 7 23 2012 8 5 8 8 12 7 12 8 67 8 10 9 9 4  234 7 24 2012 11 4 11 5 17 7 17 7 64 6 11 
276. cked for tightness  Leakage shall not ex   ceed 1   of the sample volume sucked     6 2 Equipment  The test required an adapter for tightness check and a mass flow meter   6 3 Performance of test    The Model 50301 SHARP can perform a tightness check using an adapter     For testing the leak tightness  the instrument is switched to service mode  First  an external  flow meter is installed at the inlet  followed by a recording of the flow rates indicated on the  instrument itself and by the external measurement as well as of the displayed vacuum  VAC  in mmHg   The instrument flow rate is compared with the flow rate of the external flow meter  and adjusted  Then  the tightness check adapter is inserted into the inlet as connection be   tween inlet and external flow meter  The adapter creates a significantly higher vacuum in the  instrument for it reduces the size of the air sampling opening  The flow rates indicated by the  instrument and by the external flow meter are compared again  Ideally  i e  at absolute tight   ness   there should be no difference between both flow rates  The manufacturer specifies a  maximum permissible difference of 0 42 I min    2 5   of the flow rate   If greater deviations  were to be found  a leak test must be performed taking special care of checking the many fit   tings above the radiometric measuring head  If the cause cannot be found  the manufacturer  should be contacted     The tightness check must only be performed with the tightness check 
277. ckseite des Deckblatts und der folgenden Anlage mit insgesamt 33 Seiten     Registrierungsnummer der Urkunde  D PL 11120 02 00    Im Auftrag    PIA ATIN    Berlin  23 01 2013 Andrea Valbuena  Abteilungsleiterin    Siehe Hinweise auf der R  ckseite    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    A TUVRheinland   Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 176 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Deutsche Akkreditierungsstelle GmbH    Standort Berlin Standort Frankfurt am Main Standort Braunschweig  Spittelmarkt 10 Gartenstra  e 6 Bundesallee 100  10117 Berlin 60594 Frankfurt am Main 38116 Braunschweig    Die auszugsweise Ver  ffentlichung der Akkreditierungsurkunde bedarf der vorherigen schriftlichen  Zustimmung der Deutsche Akkreditierungsstelle GmbH  DAkkS   Ausgenommen davon ist die separate  Weiterverbreitung des Deckblattes durch die umseitig genannte Konformitatsbewertungsstelle in  unveranderter Form     Es darf nicht der Anschein erweckt werden  dass sich die Akkreditierung auch auf Bereiche erstreckt   die   ber den durch die DAkkS best  tigten Akkreditierungsbereich hinausgehen     Die Akkreditierung erfolgte gem     des Gesetzes   ber die Akkreditierungsstelle  AkkStelleG  vom   31  Juli 2009  BGBI    S  2625  sowie 
278. company Thermo Fisher Page 79 of 424  Scientific for the component suspended particulate matter PMs 5    Report No   936 21209885 F    6 1 4 1 3 Functional check    If the operation or the functional check of the measuring system requires particular in   struments  they shall be considered as part of the measuring system and be applied in  the corresponding sub tests and included in the assessment    Test gas units included in the measuring system shall indicate their operational readi   ness to the measuring system by a status signal and shall provide direct as well as re   mote control via the measuring system     6 2 Equipment  Manual  zero filter and span foil set   6 3 Performance of test    The system status is constantly monitored and problems are indicated by a series of different  warning messages  Parameters which are relevant for a proper functioning of the measuring  system  e g  NEPH zero adjustment  flow rates  pump vacuum  can also be reviewed on the  instrument itself and or recorded continuously by a data logger     Furthermore  it is possible to check the zero point of the measuring system externally  For  this  a zero filter is installed at the device inlet  The use of this filter allows the provision of  particle free air     During the test  the position of the zero point was determined approx  every 4 weeks using a  zero filter     In order to check the stability of the beta measurement  the    mass transfer standard    foil set  is required  This set co
279. connected  and press        The Set Analog Outputs screen appears     4  Press     to set the output to zero   The Output Set To line displays Zero     5  Check that the meter is displaying the zero value  If the meter reading  differs by more than one percent of the full scale output  the analog  outputs should be adjusted  Refer to the    Analog Output Calibration     procedure that follows     6  Press      to set the output to full scale   The Output Set To line displays Full Scale     7  Check that the meter is displaying a full scale value  If the meter  reading differs by more than one percent of the full scale output  the  analog outputs should be adjusted  Refer to the    Analog Output  Calibration    procedure that follows     8  Press to reset the analog outputs to normal     7 22 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Servicing  Analog Output Testing       10  EXPANSION    E  a Current Outputs             Analog Voltage Inputs    Analog Voltage Outputs       ___         Figure 7   11  Rear Panel Analog Input and Output Pins    Table 7   4  Analog Output Channels and Rear Panel Pin Connections                      Voltage Pin Current Pin   Channel Channel   1 14 1 15   2 33 2 17   3 15 3 19   4 34 4 21   5 17 5 23   6 36 6 25   Ground 16  18  19  35 37 Current Output Return 13  16  18  20  22  24       Table 7   5  Analog Input Channels and Rear Panel Pin Connections    Input Channel Pin       1 1  2 2  3 3     
280. creen displays the current nephelometer source current  value and sets the minimum and maximum alarm limits  Acceptable alarm  limits range from 50 to 70 mA  If the LED current reading goes beyond    3 70 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu    the minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  LED Current        ERE    Wi moi   PS PHP L          Min and Max LED Source The Minimum LED Source Current alarm limit screen is used to change  Current Limits the minimum LED source current alarm limit  The minimum and  maximum LED Source Current screens function the same way     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  LED Current  gt   Min or Max        Reference Detector The Reference Detector Voltage screen displays the current nephelometer  Voltage reference detector voltage value and the minimum and maximum alarm  limits  Acceptable alarm limits range 350 to 2000 mV  If the reference  detector voltage reading goes beyond the minimum or maximum limit  an  alarm is activated and the alarm  bell  icon appears in the status bar on the  Run screen and in the Main Menu  There is no    set    mode for this value     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  Ref Det Volt        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 71    Operation  Ala
281. d also includes signal  buffers for the graphics display control signals and a high voltage power  supply for the graphics display backlight     External I O is driven from a generic bus that is capable of controlling the  following devices     e Analog output  voltage and current   e Analog input  voltage    e Digital output  TTL levels    e Digital input  TTL levels     Note The instrument has spare solenoid valve drivers and I O support for  future expansion  A    The instrument provides six analog voltage outputs  Each may be firmware  configured for any one of the following ranges  while maintaining a  minimum resolution of 12 bits     e 0 100 mV  e 0 1V  e 0 5V  e 0 10 V    The user can calibrate each analog output zero and span point through the  firmware  At least 5  of full scale over and under range are also supported   but may be overridden in the firmware  if required     Model 50307 SHARP Instruction Manual 8 9    System Description  1 0 Components    Analog Current  Outputs  Optional     Analog Voltage  Inputs  Optional     Digital Relay  Outputs    8 10 Model 50307 SHARP Instruction Manual    The analog outputs may be assigned to any measurement or diagnostic  channel with a user defined range in the units of the selected parameter   The voltage outputs are independent of the current outputs     The optional I O expansion board includes six isolated current outputs   These are firmware configured for any one of the following ranges  while  maintaining a minimum
282. d as refer   ence point    In case of deviating certification ranges  the repeatability standard deviation at refer   ence point shall not exceed 2   of the upper limit of this certification range  In this case  a value c  at 70   to 80   of the upper limit of this certification range shall be used as  reference point     Note     By resolution of the competent German body  see module 5 3 1   this test is not relevant for  particulate measuring systems     6 2 Equipment   Not applicable    6 3 Performance of test  Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 104 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 3 4 Linearity  lack of fit     The analytical function describing the relationship between the output signal and the  value of the air quality characteristic shall be linear     Reliable linearity is given  if deviations of the group averages of measured values about  the calibration function meet the requirements of Table 2 of in the certification range  according to Table 1 of St
283. d concentrations that becomes a dynamic correction  factor for the real time nephelometer  and the product of this correction  factor and nephelometer becomes the SHARP Monitor concentrations     Table 1   1 lists the specifications for the Model 50303   Table 1 1  Model 50307 Specifications    Safety Electrical Designations   Designed to meet    CE  EN 61326 1997   A1 1998   A2 2001   A3 2003  EN 61010 1  UL  61010 1 2004   CSA  C22 2 No  61010 1 2004   FCC  Part 15 Subpart B  Class B   Source and Detector    IR LED  6mW  880 nm Silicon Hybrid Amplifier Photodetector  B Source  Carbon 14   lt 3 7 MBq   lt 100 uCi  Sealed Source Device    Proportional Radiation Detector  Standard System Configuration    Menu driven software for user interaction via 320 x 420 pixels graphics display       Thermo Fisher Scientific    Thermo Fisher Scientific    Introduction  Specifications    Connecting and Interface Cables  and Vacuum Pump       3 foot heated sample tube       Ambient T RH Assembly       Consumables for average first year s operation  ambient        iPort Software for Local or Remote Communication  Sample Heater    3 foot insulated heater tube       Three Modes of Heater Control    OFF     Heater is turned off in non condensing environment  TEMP     fixed sample temperature  RH     Dynamic Heating of sample at set sample RH  threshold  e g   50  RH    Instrument Performance  16 67 l min  1s  stable conditions     Measurement Range  0 to 1 0  2 0  3 0  5 0  10 0 mg m  0 to 100 
284. d support boratory test  part  2  Alarms added to the list of digital outputs 2  as well as the  3  Additional parameters for recording added additional campaign  T i Bornheim  Sum   4  Changes to timing procedure after filter ex    mer 2013     change     no calculations in the first 15 min  following pump stop  5  Update of algorithms for calculating ambient  temperature  ambient humidity and flow  temperature in order to provide a more pre   cise sample humidity value  sample rH   6  Recording of amplification and mode for  NEPH added  7  Update of pressure calculations smoothing  in order to improve flow control  8  Update of heating control  PID control opti   mised   V02 00 00 232  1  Modification to new i series processor board   Not installed  exam   110570 ined during regular  2  Menu navigation    Servise Pres Vacuum Cal    production audit   ibration Vac Flow    changed  3  Diagnostics Voltages detector Board menu  no longer locks up  4  Title line on    Calibration factors Neph BKG  Values restore Prev Values    verschoben             740300_2013_936_21209885F_EN docx          TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 53 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    5     range    Clink comma
285. der Verordnung  EG  Nr  765 2008 des Europ  ischen Parlaments  und des Rates vom 9  Juli 2008 Uber die Vorschriften fiir die Akkreditierung und Marktiiberwachung  im Zusammenhang mit der Vermarktung von Produkten  Abl  L 218 vom 9  Juli 2008  S  30     Die DAkkS ist Unterzeichnerin der Multilateralen Abkommen zur gegenseitigen Anerkennung der  European co operation for Accreditation  EA   des International Accreditation Forum  IAF  und   der International Laboratory Accreditation Cooperation  ILAC   Die Unterzeichner dieser Abkommen  erkennen ihre Akkreditierungen gegenseitig an     Der aktuelle Stand der Mitgliedschaft kann folgenden Webseiten entnommen werden   EA  www european accreditation org   ILAC  www ilac org   IAF  www iaf nu    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVR heinland      Luftreinhaltun 3    j Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 177 of 424  Report No   936 21209885 F    Annex 1 Detection limit Page 1 of 1  Manufacturer Thermo Fisher Scientific    Type Model 50301 SHARP Standards ZP Zero filter    Serial No  SN 1 SN2    Date Measured values  g m3  Date Measured values  ug m    Sn 1     2    11 13 2012 11 13 2012  11 14 2012 E 11 14 2012  11 15 2012   11 15 2012  11 16 2012   11 16 2012  11 17 2012   11 17 2012  11
286. ditions    Not all of the I O available in the instrument are brought out on the  supplied terminal board  If more I O is desired  an alternative means of  connection is required   See optional    Terminal Block and Cable Kits      A    e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Digital Input Settings        Logic State The Logic State menu item is used to change the selected I O relay to  either normally open or normally closed  The default state is open  which  indicates that a relay connected between the digital input pin and ground is  normally open and closes to trigger the digital input action     e Press to toggle and set the logic state to normally open or  normally closed        Instrument Action The Instrument Action submenu allows the user to choose the instrument  action that is assigned to the selected digital input     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Digital Input Settings  gt  select Relay  gt  Instrument Action     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 47    Operation  Volumetric Conditions        DH pikari  Ce a E S a EA    Analog Output Configuration The Analog Output Configuration menu displays a list of the analog   Select Channel  output channels available for configuration  Channel choices include all    Allow Over Under Range    3 48 Model 50307 SHARP Instruction Manual    voltage channels  all current channels  individual voltage channels 1   6  and  in
287. dius Tube Mounting bracket screws     2  Slide back bracket downwards     3  Slowly remove the existing Radius Face Seal by pulling straight up with  a twisting action     4  Slowly insert the Radius Tube Adapter onto the Radius Tube and push  down with a slight twist until it is seated     5  Tighten the radius tube mounting bracket screws as shown in Figure    10 5     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 10 5    Model 5014i to 50301 Upgrade Procedures  Radius Tube Adapter Installation       Figure 10   5  Radius Tube Adapter Install    6  Reassemble the two covers of the instrument in reverse order and check  alignment of radius tube adapter through the hole in the cover as  shown in Figure 10   6  Be sure to remove the old dust cover cap and  the large plastic hole cover        Figure 10 6  Radius Tube Adapter Alignment Check    10 6 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Model 5014i to 50301 Upgrade Procedures  SHARP External Valve Installation    SHARP External Use the following procedure for SHARP external valve installation   Valve Installation    1  Attach the SHARP External Valve Assembly  which includes the to the  Rear Panel using the captive hardware on bracket and attach the 3 pin  connector to the rear panel labeled valve  Figure 10 7         Figure 10   7  Rear Valve Assembly Installation    2  Follow the following plumbing procedure in accordance with Figure  10 8 below     a  Attach the 3 8    vacuum tubing 
288. dividual current channels 1 6  if the I O expansion board option is    installed      e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration     gt  Analog Output Config        Eike EE EE          The Allow Over Under Range screen is used to select whether or not the  analog outputs are allowed to exceed the maximum selected value of 100  mV  1 V  5 V  10 V  or 20 mA or the minimum selected value of 0 V  0  mA  or 4 mA  By default  this parameter is set to on and 5  over and  under range is allowed for all analog output channels     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration     gt  Analog Output Config  gt  Allow Over Under Range     Thermo Fisher Scientific    Operation  Volumetric Conditions       ERED CH  perm    WH mid  Pepi L Pi AG Lf bi       Analog Output Configuration The Analog Output Configuration menu displays a list of the analog   Select Action  output configuration choices  from which the user selects the parameter to  adjust for the selected output channel  Configuration choices include  selecting range  setting minimum maximum values  and choosing the  signal to output     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Output Config  gt  All Voltage Channels  All Current  Channels  Voltage Channel 1   6 or Current Channel 1 6        ERE CH  Pm Tit    Wi mid  PS PHP L Pi AG Lb i       Select Range The Select Range screen is used to select the hardware range for the selected  analog outp
289. docx    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    A TUVRheinland      Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5     Report No   936 21209885 F    Requirements on measuring systems for particulate air pollutants    Page 25 of 424    filled    5 4 1 General    Equivalency of  the sampling  system    Reproducibility of  the sampling  systems    Calibration    5 4 5 Cross sensitivity    5 4 6 Averaging effect    Test according to the minimum  requirement stated in Table 5  of Standard VDI 4202  Part 1     Furthermore  the particle mass  concentration shall be related  to a defined volume     The equivalency to the refer   ence method according to EN  12341  T2  shall be demon   strated     This shall be demonstrated in  the field test for two identical  systems according to EN  12341  T2      The systems under test shall  be calibrated in the field test by  comparison measurements  with the reference method ac   cording to EN 12341 and  EN 14907  Here  the relation   ship between the output signal  and the gravimetrically deter   mined reference concentration  shall the determined as a  steady function     Shall not exceed 10   of the  limit value     The measuring system shall al   low for formation of 24 h aver   ages  The time of the sum of all  filter changes within
290. ducibility at B4  e U   Uncertainty  e Bi   25 g m    for PM25  e Sp   Standard deviation from paired measurements  e n   No  of paired measurements  e tno 95    Students factor at confidence level of 95    e Xij   Measured signal of system 1  e g  SN 1  at i    concentration   e Xj   Measured signal of system 2  e g  SN 2  at i    concentration     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    A TUVRheinland   Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 116 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F  6 5 Assessment    The reproducibility for PM2  5 was 10 for the complete dataset of the field test   Minimum requirement fulfilled  yes    6 6 Detailed presentation of test results    The test results are summarised in Table 23  The graphical representation for PM25 is shown  in Figure 41 to Figure 45   Note  The determined uncertainties are applied to reference value B  for each test site     Table 23  Concentration averages  standard deviation  uncertainty range and reproducibility  in the field  measured component PM25    Test site No  of  ae  SN 1                 ver 20 4 20 3 0 827 2 007 1 66   winter    Bornheim 91 11 8 11 7 11 7 1 354 1 986 2 69   summer    Bornheim 49 21 3 21 3 21 3 1 033 2 010 2 08 12   winter    a  0 956  2 009    
291. duduspoactuieadhactadedetatateasetadsroatanss 4 12  Nephelometer Calibration  2i  acihe danviailineniwene denies 4 13  Nephelomicter RIA siesta  scored teats she EE 4 13  Nephelometer I emiperattret  ciia eee ee eh ae 4 13  Nephelometer Source Levelt nt  002 cscg taste ea etc gd 4 14  Chapter 5 Preventive Maintenance                  s ssssssssssssssssssssesssessseeesnssneseesesssenssnsaneseens 5 1  Sate ty Precast ee ea us teu anaes act arcu aucun EE 5 1  Replacement Parts   teccctir siae oa Jesse Geant hut  RIE 5 2  Cleaning the Outside Casen issuri atin steak et area loin Sette 5 2  Cleaning the Tiletsii tiso sosirii Peles tesel att Ghee karat ahateenactet 5 2  U S EPA PM ig Valet x aiaiicansguesviagdecnduandieshauniasddeasncedvasticntvendany aeons 5 2  European PMio PM3s5 Ti ets sf acee es secest ctvacncc tls aic rence sewlegned econ 5 3  Cyclone Wiainite manic 2a acs ties ceed eee Ge eee cide oes eet eae ce 5 4  Heaterand Sample W uber  0c tea enii 5 5  Weather Proofingssanna ena ieee iter tte RRs 5 5    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual xiii    Contents    xiv    Chapter 6    Chapter 7    Model 50307 SHARP Instruction Manual    Fan Filter Inspection and Cleaning ct 45istcccecnce tuto ol cata adehuae  can cares ncnh 5 5  Pump Rebuilding   ycj cciecncitiea cies screed pateedtitiaeceleeideae bane ncees 5 6  bedk Testa ste catch ct sages hve nedied Spats Mos leur ves dele aiae andi Metin eed esetet 5 6  Filter Tape Replacement s i sccuizscsecusze
292. e  It is proposed that the same pragmatic approach is taken  here that was previously undertaken in earlier studies     The slope for the full dataset is 1 068  For this reason  a slope correction of the entire dataset  was performed and evaluated again with the corrected values  All datasets fulfil the require   ments on data quality after the correction  The measurement uncertainty was improved sig   nificantly  only at the test sites    Bornheim  summer     and    Teddington  summer     a slight  worsening could be observed for SN 2  see Table 38 in comparison to Table 35      The version of the Guide of January 2010 requires that when operating in networks  a  candidate method needs to be tested annually at a number of sites corresponding to the  highest expanded uncertainty found during equivalence testing  These criteria are banded in  5   steps  Guide  4   point 9 9 2  Table 6   We have to bear in mind that the highest  determined expanded uncertainty lies in the range 10   to 15   after the correction of the  slope     The respective realisation of the above mentioned requirement on on going QA QC in net   works is the responsibility of the network operator or of the responsible authority of the mem   ber state  However  TUV Rheinland and their UK partners recommend that the expanded  uncertainty for the full dataset is referred to for this  namely 18 0    uncorrected dataset  and  12 0    dataset after slope correction   which again would require an annual test at 4 
293. e  detector assembly  the C 14 radioactive source is partially exposed  Safety  glassed must be worn during this replacement procedure  The amount of  C 14 meets the U S  NRC regulations as an exempt amount of radioactive  source  lt 100 pCi  A    The detector pre amp assembly has a power backup built in     DO NOT  OPEN    caution handeling  Backup power will dissipate after 8 hours  A    CAUTION The detector window is very fragile  Handle with great care and  do not wipe or touch the window  Furthermore  do not touch the solder  cable connection  electrode  with your bare fingers  The oils from your skin  can damage the detector  A    Safety glasses must be worn while replacing the detector assembly  A    If the LCD panel breaks  do not let the liquid crystal contact your skin or  clothes  If the liquid crystal contacts your skin or clothes  wash it off  immediately using soap and water  A    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  Figure 7   1   If an  antistatic wrist strap is not available  be sure to touch the instrument chassis  before touching any internal components  When the instrument is  unplugged  the chassis is not at earth ground  A    Do not remove the LCD panel or frame from the LCD module  A    Model 50307 SHARP Instruction Manual 7 3    Servicing  Firmware Updates    Firmware Updates    Accessing the  Serv
294. e  instrument address 14      lt STX gt ST014 lt SP gt N lt CR gt     Data Sampling Data This command initiates a data transfer from the instrument  The  Query Command instrument responds with measurement data  which depends on the range   DA  mode and is listed in    Measurements reported in response to DA  command    below     The command structure for a data query command is as follows    lt STX gt DA lt address gt  lt ETX gt  lt BCC gt     The  lt address gt  is optional  which means it can be left out completely  The   lt address gt   if present  must match the Instrument Address  Additional space  can be present after the  lt address gt      If the  lt address gt  is left out  then no space is allowed in the query string   A command with no address is also a valid command     The following are the different acceptable formats of the DA command  with Instrument Address 14      lt STX gt DA lt CR gt    lt STX gt DA014 lt CR gt    lt STX gt DA lt SP gt  lt 14 lt ETX gt  lt BCC gt    lt STX gt DA lt ETX gt  lt BCC gt     The data query string is valid and will be answered with data transmission  only if the command starts with  lt STX gt   which is followed by the  characters DA  and the  lt address gt   if present  matches the Instrument  Address  and the command is terminated with either  lt CR gt  with no  checksum or  lt ETX gt   followed by the correct checksum  lt BCC gt      D 4 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific
295. e 36  Zero point drift SN 2  measured component PM25    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control 3    Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PMz25 pre separator of the company Thermo Fisher   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    Page 121 of 424    Table 25  Sensitivity drift SN 1  amp  SN 2                                           SN1 SN2  Measured Deviation from Deviation from Measured Deviation from Deviation from  Date Value previous value start value Date Value previous value start value            1 27 2011 7170 9   7178 5   5 12 2011 7194 2 E 0 3 5 12 2011 7255 4   7 19 2011 7270 6    11 14 2011 7189 0 0 3 7215 0      3 8 2012 7332 3 2 0 2 3 3 8 2012 7006 9  2 9  2 4  6 18 2012  10 19 2012 7214 0 0 2 0 6 7119 2  1 0  0 8       740300_2013_936_21209885F_EN docx    s 7    TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 122 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Span point drift PM2 5 SN 1    D   e     a          N          o  ro     ro         A         D   E  E  a  Le   c             gt   2   5
296. e 7   14      4  Pop the board off of the mounting studs and remove the board   5  To install the I O expansion board  follow the previous steps in reverse     6  Calibrate the analog current outputs and analog voltage inputs as  defined earlier in this chapter          Motherboard    Digital Output Board  1 0 Expansion Board  Opt           Figure 7   13  Replacing the I O Expansion Board  Optional     7 28 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Servicing  Digital Output Board Replacement          lt           _  0 Expansion Board    Motherboard       Digital Output Board       Figure 7   14  Rear Panel Board Connectors    Dig ital Outp ut Board Use the following procedure to replace the digital output board  Figure 7       Replacement       Equipment Required     Digital output board  Nut driver  3 16 inch    amounts of static electricity  A properly grounded antistatic wrist strap    f Equipment Damage Some internal components can be damaged by small    must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Remove the I O expansion board  optional   if used  See the    I O  Expansion Board  Optional  Replacement    procedure in this chapter     3  Disconnect the digital output board ribbon cable from the  motherboard     4  Using the nut driver  remove the two standoffs securing the board to  the rear panel  Figure 7   14      Thermo Fisher Scientific Model 50307 SHAR
297. e Detector Amplifier Assembly           cccccccceceeeeeee 7 16  Figure 7 8  Replacing the PUM uuu    cece cccecesesesceesescsteseseseeteseseseacenesesees 7 18  Figure 7 9  Replacing the Pressure Board ASS MDIY            ccecccececesesceteteseeee 7 19  Figure 7   10  Replacing the Pressure Board Assembly Plumbing Connection   7 20  Figure 7 11  Rear Panel Analog Input and Output Pins        0 0  cece 7 23  Figure 7   12  Replacing the Thermistor           ccccccccccecesssesceteesestetesesesceteseseees 7 27  Figure 7 13  Replacing the I O Expansion Board  Optional          0   0c ccecee  7 28  Figure 7   14  Rear Panel Board Connectors           cccecccccccscccecsseetesestesteteeeeeeee 7 29  Figure 7   15  Replacing the Measurement Interface Board             cccceeceee 7 32  Figure 7   16  Replacing the Photo Interrupt Board 0    ccc ceteeseeee 7 33  Figure 7 17  Replacing the Proportional Valve occ cecccceceestetesesesteteseseees 7 34  Figure 7 18  Replacing the Detector Assembly o oo    cece cece ce teteeeteteseseees 7 36  Figure 7 19  Replacing the RTD ees   osc  ce eae cecseer es centers var teeters eee tents 7 37  Figure 7   20  Replacing the Transformer         7 38  Figure 7   21  Replacing the Radius Tube ou    ccc cccecsceceesestetesesesteteseseees 7 39  Figure 7   22  Replacing the Ambient RH Temperature Assembly                      7 41  Figure 7 23  Replacing the Front Panel Board and the LCD Module                7 42  Figure 7   24  Removing the SHARP Optics Co
298. e PM  concentration goes beyond either the minimum or maximum limit  an  alarm is activated and the alarm  bell  icon appears in the status bar on the  Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Avg PM        Min and Max Average The Minimum Average PM Concentration alarm limit screen is used to  PM Concentration Limits change the minimum average PM concentration alarm limit  The    minimum and maximum average PM concentration screens function the  same way     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Avg PM  gt  Min  or Max        DH Guided  PME TY aad    PME Eee EE EE          3 80 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Average SHARP    Min and Max Average  SHARP Concentration  Limits    Average Nephelometer    Thermo Fisher Scientific    Operation  Alarms Menu    The Average SHARP screen displays the current average SHARP  concentration and sets the minimum and maximum alarm limits   Acceptable alarm limits range from 0 to 10000 g m  or 0 to 10 mg m     If  the average SHARP concentration goes beyond either the minimum or  maximum limit  an alarm is activated and the alarm  bell  icon appears in  the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Avg SHARP        Wi mnie           The Minimum Average SHARP Concentration alarm limit screen is used  to change the minimum average SHARP concentration alarm limit  The  minimum and 
299. e Packing Material 0 00 00    ccccccecesesestetesesesteteceteseseees 2 3  Figure 2 2  Model 50307 Sampling Room Air Under Stable Conditions            2 7  Figure 2 3  Model 5030  Side View   inches  millimeters            0 0ccccccee  2 10  Figure 2   4  Model 50307 Top View   inches  millimeters            00 0cccccceee 2 11  Figure 2   5  Model 50307 Vertical View 0    ccc cececeesceteesesteteseseeteteseseees 2 12  Figure 2   6  Heater Assembly ou    ccc cece ceceteseseeceseseseeteseseseeteseseseecesesesaes 2 14  Figure 2 7  Rear Panel Plumbing Schematic        0  0 ccccccccccccesestesesteeeteeeeeeees 2 16  Figure 2 8    O Terminal Board Views             cccccccscsccscssestesesteseeteseseseeeseeseeees 2 18  Figure 2 9  Pin Out of Rear Panel Connectol          ccc ccccccccccceteseetesesteeeeeeee 2 19  Figure 2 10  D O Terminal Board ViQWS          ccccccccccsccestesestesestesestesteeseeseeees 2 20  Figure 2 11  25 Pin Terminal Board ViGWS          c ccccccceccsescetesesesteteseseeteteseseees 2 21  Figure 3 1  Front Panel Dis Ways  ser  ecce ces vente eto entanet velar de cuenta eer  3 2  Figure 3 2  Front Panel PUSNDUtIONS        ccc cecescseeteesesestesesesesteeseseees 3 2  Figure 3 3  Flowchart of Menu Driven Firmware           ccccecccescsescetetesesteteseseees 3 5  Figure 3   4  Pin Out of Rear Panel Connector    3 12  Figure 3   5  Common Fags ieu sesso ee ss a ae 3 31  Figure 3   6  Detector A Flags    cccccccccccecesesesesestesesesesteesesesteesssesteeseseaes 3 31  
300. e SHARP coefficient is 1 05     Send  coef sharp  Receive  coef sharp 1 050    set coef sharp value   This command sets the SHARP coefficient to user defined values to value   where value is a floating point representation of the coefficient  The  following example sets the SHARP coefficient to 1 04     Send  set coef sharp 1 04  Receive  set coef sharp 1 04 ok    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Calibration    flow pres span  This command reports the flow pressure span recorded at the time of  calibration  The following example reports that the flow span is 1 2     Send  flow pres span  Receive  flow pres span 1 200    set flow pres span value   This command sets the flow pressure span to value  where value is a  floating point representation of the pressure span specified at the time of  calibration  The following example successfully sets the new value for flow  pressure to 1 5     Send  set flow pres span 1 5  Receive  set flow pres span 1 5 ok    flow pres offset  This command reports the flow pressure offset for the detector at the time  of calibration  The following example reports that the flow pressure offset is    3 0     Send  flow pres offset  Receive  flow pres offset 3 0  flow span    This command reports the current flow span at the time of calibration  The  following example reports that the flow at calibration is 1 00     Send  flow span  Receive  flow span 1 000    set flow span value   This command sets the c
301. e This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Flow Calibration        prek ee meh dt Erm  PPP L mie LG ae       Auto Flow Calibration The Auto Flow Calibration screen allows the user to view and set the  correct flow rate  The span value is automatically calculated     e Inthe Main Menu  choose Service  gt  Flow Calibration  gt  Auto        ERED aks PP Tite  7 METEEN Lid    mith  a   mii       Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 91    Operation  Service Menu    Manual Flow Calibration The Manual Flow Calibration screen allows the user to view and  incrementally set the flow sensor calibration span factor     e Inthe Main Menu  choose Service  gt  Flow Calibration  gt  Manual        Mass Calibration The Mass Calibration submenu is used to view and set the mass calibration  point  The mass calibration screen is visible only when the instrument is in  service mode  For more information on the service mode  see    Service  Mode    earlier in the chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Mass Calibration        Thermal Mass Coefficient The Thermal Mass Coefficient screen allows the user to view and set the  thermal coefficient using manual calibration mode   Factory set      e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Thermal  Coefficient     3 92 Model 50307 SHA
302. e a letter was detected    data not valid    when the value should be numeric     Supplied value is not acceptable for entered command       can t  wrong settings    Command not allowed for current measurement mode       can t  mode is service    Command not allowed while instrument is in service mode       feature not enabled    1 0 expansion board is not detected       flags no alarm active    No measurement alarms are active    The    save    and    set save params    commands  duplicated for backward  compatibility  store parameters in FLASH memory  It is important that  this command be sent each time instrument parameters are changed  If    changes are not saved  they will be lost in the event of a power failure     Table B 2 lists the 50307 C Link protocol commands  The interface will  respond to the command strings outlined below     Table B 2  C Link Protocol Commands                               Command Description Page  1 Simulates pressing soft key 1 pushbutton B 44  2 Simulates pressing soft key 2 pushbutton B 44  3 Simulates pressing soft key 3 pushbutton B 44  4 Simulates pressing soft key 4 pushbutton B 44  addr dns Reports sets domain name server address for Ethernet port B 54  addr gw Reports sets default gateway address for Ethernet port B 54  addr ip Reports sets IP address for Ethernet port B 54  addr nm Reports sets netmask address for Ethernet port B 55  addr ntp Reports sets IP address for network time protocol server B 55  alarm alpha count Reports al
303. e case   Remove one screw from the top front of the partition panel     While holding the case securely  loosen the captive screw at the rear of  the measurement case assembly  and pull the measurement case  assembly from the rear of the case     Remove the screw at the top rear of the partition panel that secures the  top of the panel to the measurement case assembly  and lower the  panel  being careful not to put excessive tension on the cables     Replace the measurement case assembly by following the previous steps  in reverse     Model 50307 SHARP Instruction Manual 7 13    Servicing  Fuse Replacement    Fuse Replacement    Fan Filter  Replacement    A    7 14 Model 50307 SHARP Instruction Manual    Use the following procedure to replace the fuse   Equipment Required     Replacement fuses  refer to the    Replacement Parts List    in this    chapter    1  Turn the instrument OFF and unplug the power cord   2  Remove the fuse drawer  located on the AC power connector   3  Ifeither fuse is blown  replace both fuses     4  Insert fuse drawer and re connect power cord     Use the following procedure to replace the fan and the fan filter  Figure 7     6      Equipment Required   Fan  Fan filter  Philips screwdriver    Cable cutter  for tie wrap removal     Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument O
304. e current concentration units  mg m  or pg m       The following example reports that the concentration units are set to  mg m      Model 50307 SHARP Instruction Manual B 45    C Link Protocol Commands  Measurement Configuration    B 46 Model 50307 SHARP Instruction Manual    Send  conc unit  Receive  conc unit pg m3    set conc unit unit  unit     g m    mg m       This command sets the concentration units to mg m  or pg m   The  following example sets the concentration units to mg m        Send  set conc unit mg m3  Receive  set conc unit mg m3 ok    custom range  range   1 2 3     This command reports the user defined value of custom range 1  2  or 3   The following example reports that custom range 1 is defined to 5 50  mg m     Send  custom 1  Receive  custom 1 5 500E 00 mg m3    set custom range range value   set custom 1 value   set custom 2 value   set custom 3 value   These commands are used to set the maximum concentration for any of the  three custom ranges 1  2  or 3 to range value  where value is a floating point  number representing concentration in the currently selected units  The  following example sets the custom 1 range to 100 5 ug m       Send  set custom 1 100 5 ug m3  Receive  set custom 1 100 5 g m3 ok  filter period    This command reports the number of hours to be added to filter next time  for the filter tape change  The following example reports that the filter  period is eight hours     Send  filter period  Receive  filter period 8 HRS    set filte
305. e different Run screens           Menu The     is used to display the Main Menu when in the  Run screen  or back up one level in the menu system  For    more information about the Main Menu  see    Main Menu     later in this chapter      2     Help The  2    is context sensitive  that is  it provides    additional information about the screen that is being  displayed  Press  2    for a brief explanation about the  current screen or menu  Help messages are displayed using  lower case letters to easilv distinguish them from the  operating screens  Press   gt    to return to the Run  screen  or any other key to exit a help screen            J up  Down The four arrow pushbuttons   _ JCs    jana           gt    Left  Right   gt  _   move the cursor up  down  left  and right or  change values and states in specific screens       Enter The is used to select a menu item  accept set save    a change  and or toggle on off functions                          The soft keys are multi functional keys that use part of the display to  identify their function  The function of the soft keys is to provide a  shortcut to the most often used menus and screens  They are located  directly underneath the display  and user defined labels in the lower part of  the display indicate the function of each key at that time     To change a soft key  place the menu cursor     gt     on the item of the selected  menu or screen you wish to set  Press   gt    followed by the selected soft  key within 1 second o
306. e following equation to calculate the percent  difference     Qm Qi    l     D  100x    If  D is within  4   then the Model 5030   volumetric flow rate  acceptance test has passed  If the Model 5030i volumetric flow rate is  slightly out of tolerance   5   the acceptance test should be classified as  marginal  If the volumetric flow rate performance is outside this range   please perform a volumetric flow calibration     For more information about the volumetric flow rate calibration  see the     Calibration    chapter     The leak check of the 50307 SHARP is conducted by first measuring the  volumetric flow rate at the inlet adapter  Next  place the factory supplied  leak check adapter on top of the inlet adapter and then install the flow   meter on top of the leak check adapter  Compare the volumetric flow rate  of the 50307 to the flow meter and calculate the difference  If the difference  is less than 0 80 L min  80 ml min  then the leak check passes  If it is  greater than 80 ml min  check for leaks in the system by checking the  various fittings upstream of the measurement head  If the source of the leak  cannot be found  the instrument requires service by a trained technician     The nephelometer light source has been factory set to a current of  approximately 65 milliamps  mA   Should this valve be outside a range of  60 70 mA  the source current should be adjusted in the calibration menu     The 5030  has an auto zero function and it is common during setup to  potent
307. e is the index number of a variable in the  analog output variable list  Available variables and their corresponding  index numbers may be obtained using the command    list var aout     The     set sp field    command is used to create a list of variables which can then  be transferred into the Irec  srec  or streaming data lists  using the    set copy  sp to lrec        set copy sp to srec     or    set copy sp to stream    commands     Send  set sp field 1 34  Receive  set sp field 1 34 ok  stream per    This command reports the current time interval in seconds for streaming  data  The following example reports the streaming period is set to 10    seconds   Send  stream per  Receive  stream per 10 sec    set stream per number value  number value     1   2  5  10  20  30   60   90  120   180   240   300      This command sets the time interval between two consecutive streaming  data strings to number value in seconds  The following example sets the  number value to 10 seconds     Send  set stream per 10  Receive  set stream per 10 ok    Model 50307 SHARP Instruction Manual B 33    C Link Protocol Commands  Calibration    Calibration    B 34 Model 50307 SHARP Instruction Manual    stream time   This command reports if the streaming data string will have a time stamp  attached to it or not  according to Table B   4  The following example  reports that the streaming data shall not include a time stamp     Send  stream time  Receive  stream time 0    set stream time value   Thi
308. e reference  This  average observed flow rate from your reference meter should now be  entered in the screen below        Errori pri pr ey yes Enmen  ni METERA Lid    CH ninii  Rri Taic bm iii       Be sure to save the entry and compare the values once more  Repeat as  necessary to within     2       Perform the    Leak Test    procedure on page 5 6 to verify no leaks     The Model 5030 SHARP is calibrated for mass in the factory using a series  of null and span foils  The mass transfer standard foil sets are available from  Thermo Fisher Scientific  and should be used as part of a QA Program for  performing a QC check on the mass measurements  This mass foil  calibration procedure can be used for QC checks  auditing  and calibration     Model 5030  SHARP Instruction Manual 4 7    Calibration  Calibration Procedure    4 8 Model 50307 SHARP Instruction Manual    The Mass Calibration submenu is used to view and set the mass calibration  point  The mass calibration screen is visible only when the instrument is in  service mode  For more information on the service mode  see    Service  Mode    earlier in the chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Mass Calibration gt Mass  Coefficient        It is recommended to perform an annual Mass Coefficient on the Model  50307 Beta Monitor     e The foil sets must be kept in a clean container  Do not touch the foil  window with your fingers  
309. e separator of the company Thermo Fisher Page 155 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    Criterion 4  All of the expanded uncertainties are below 25       Criterion 5  The slopes used for evaluation of the complete dataset are significantly great   er than the permissible values for SN 1 and SN 2     Other  The evaluation of the complete dataset for both candidates together shows  that the AMS demonstrates a very good correlation with the reference method  with a slope of 1 068 and an intercept of  0 103 at an expanded total uncer   tainty of 18      The January 2010 version of The Guidance is ambiguous with respect to which slope and in    tercept should be used to correct a candidate should it fail the test of equivalence  After   communication with the convenor of the EC working group  which is responsible for setting  up the Guide  Mr Theo Hafkenscheid   it was decided that the requirement of the November   2005 version of the Guidance are still valid  and that the slope and intercept from the orthog    onal regression of all the paired data should be used  These are shaded gold and marked      other    in the key on the above Table 35     The 2006 UK Equivalence Report  8  highlighted that this was a flaw in the mathematics re   quired for equivalence as per the November 2005 version of The Guidance as it penalised  instruments that were more accurate  Appendix E Section 4 2 therein   This same flaw is  co
310. e signal outputs of the measuring system     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH      TUVRheinland      Air Pollution Control   s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 91 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 2 4 Failure in the mains voltage    In case of malfunction of the measuring system or failure in the mains voltage for a pe   riod of up to 72 h  uncontrolled emission of operation and calibration gas shall be  avoided  The instrument parameters shall be secured by buffering against loss caused  by failure in the mains voltage  When mains voltage returns  the instrument shall auto   matically reach the operation mode and start the measurement according to the operat   ing instructions     6 2 Equipment  Not additional equipment required   6 3 Performance of test    A failure in the mains voltage was simulated and it was tested  whether the instrument re   mains undamaged and is ready for measurement after restart of the power supply     6 4 Evaluation    Since the measuring systems need neither operation nor calibration gases an uncontrolled  escape of gases is not possible     In case of a failure in the mains voltage
311. eatures include RS 232 485 and Ethernet    C Link  MODBUS  Geysitech  Bayern Hessen   ESM Protocol   streaming data  and NTP  Network Time Protocol  protocols   Simultaneous connections from different locations over Ethernet     Radon gas activity measurement  Rn 222  noble gas  and mass  refinement    Flexible data storage configurations    Volumetric air flow rate control with actual standard conditions  concentration data output    Processor controlled calibration of all sensors  880 nm near forward scattering nephelometer  Beta Detector Life  10 years    C 14 activity below USA authorized limit values  shipped as non   hazardous material under Code UN2911  easy handling of the source  and instrument  No license is needed in most countries     Thermo Fisher Scientific    Introduction  Principle of Operation    For details of the analyzer   s principle of operation and product  specifications  see the following topics     e    Principle of Operation    on page 1 3  e    Specifications    on page 1 8    Thermo Fisher Scientific is pleased to supply this continuous ambient  particulate monitor  We are committed to the manufacture of instruments  exhibiting high standards of quality  performance  and workmanship   Thermo service personnel are available for assistance with any questions or  problems that may arise in the use of this instrument  For more  information on servicing  see the    Servicing    chapter     Principle Of The Model 5030  is based on the principles of aero
312. ecting and Cleaning the Fan    Rebuild the sample pump once every 12   18 months  or as necessary  depending on the environment is is used in  The pump repair kit contains  instructions for rebuilding the pump  See the    Servicing    chapter for a list  of replacement parts  Perform the    Leak Test    procedure that follows     This leak test procedure uses a volumetric flow meter and a custom leak  check adapter  Use the following procedure to verify no leaks     1  Place a reference volumetric flow meter  e g   BGI Delta Cal  onto the  inlet adapter and calibrate the 50307 SHARP so that the reference flow  meter and the 50302 SHARP Monitor read the same flow rate     2  Install the custom leak check adapter onto the inlet adapter and then  place the reference flow meter onto the leak check adapter     Thermo Fisher Scientific    Filter Tape    Replacement    Thermo Fisher Scientific    A    3     Preventive Maintenance  Filter Tape Replacement    Record the reference volumetric flow meter reading and the instant  flow reading  If the difference between both readings is less than 0 42  L min   2 5    the leak check passes     Use the following procedure to replace the filter tape in case of breaks or if    the tape runs out     Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap    must be worn while handling any internal component  For more    information about appropriate safety prec
313. ed    6 3 Performance of test   The test was carried out at four different comparisons during field test  Different seasons and  varying concentrations for PM2 5 were taken into consideration     Of the complete dataset  at least 20   of the concentration values determined with the refer   ence method shall be greater than the upper assessment threshold according to 2008 50 EC   7   For PMz5 the upper assessment threshold is at 17 g m8     At each comparison campaign at least 40 valid data pairs were determined  Of the complete  dataset  4 test sites  209 valid data pairs for SN 1  217 valid data pairs for SN 2  a total of  29 1   of the measured values are above the upper assessment threshold of 17 g m  for  PM 5  The measured concentrations were referred to ambient conditions     6 4 Evaluation  According to point 9 5 3 1 of the Guide  Demonstration of Equivalence of Ambient Air Moni   toring Methods    applies     The uncertainty between the candidates Ups must be  lt  2 5 g m     An uncertainty  gt  2 5 g m   between the two candidates is an indication that the performance of one or both systems is  not sufficient and the equivalence cannot be declared     The uncertainty is determined for   e All test sites and comparisons together  full dataset     e 1 dataset with measured values 2 18 g m  for PM2 5  Basis  averages of reference  measurement     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control      
314. edenes E 13  1 2  Certification Proposal wid  mentinnehminnnnebnnenneiaineekhauabnen Eman 18  1 3 Summary of test CS UMD Soest eg tee eee ceed edge eee 19  2  TASK DERINITION ere eee ene reer ee ar ore ee ee ee oe eee eee eee are 28  21 gt  Nature    tesien nnan a eee a eee meee eee ee reer 28  22  CDI OG UV Erran E E a A AEE EN EERE 28  3 DESCRIPTION OF THE AMS TESTED vinin nara 29  3 1 Measuring PINGING 228 sce Seta tao aunt atic sa ess ko secede ole coi 29  3 2 Functionality of the measuring system            cece eee eeeteeee teeter eeeeetetaaeeeeeeeeeeteeeteeenaaees 32  3 3 AMS SCONE and SOU Die cesticsasdssterardcuhaceiaasvoroderatersVeesendss reuetcentseenserd Mlepaedeuvexarbienere 37  4y TEST PROGRAMME  skiemeniniai Eaa EEn E RENEE REA REEERE AEREA 51  41 General eee eee eee eee E E ee eee eee 51  4 2 Laboratory VS Uae arcs ca ee ae ecm eta egmeneeeeee dee 54  43 Field TOG cossccretinseaselenenupebengeseulanssunebeupiumetasasnnenesoutendaysaepiasecebennepiuassundencouseiungeusedeneiape 55  5  REFERENCE MEASURING METHODS siscs2cs ced dectstsiscisdetciestieeds odie didi edieel 74  6  TEST RESULTS Zr ee ne e ee eee ee er rere eT 75  6 1 4 1 1 Measured value display 0 0 2 0    cceceeeeecceeee cette eeeeeeeceaeeeee tees etea aaa aaaaaaaeeeeeeeenenaa 75  6 1 4 1 2 Easy maintenance acs cecgteszacs sec tect seactais eecectescardassyiscoatteneig teatageasdect siateaieecaeuass 77  6 1 4 1 3 Functional check         ssssenesseessoeseenrrneresorrrrtrrnnrrstrrrenrrnnntsorrserrrenr
315. eeseees 3 81  Min and Max Average Nephelometer Concentration Limits        3 82  Ja  st  nt PM fs cad cag sabes tues danecans zens  an paa aa a RR 3 82  Min and Max Instant PM Concentration Limits    eee 3 82   iistain GSR Psi ccnaip tec aitesovbvccdsecetaaah aoon aa 3 83  Min and Max Instant SHARP Concentration Limits               3 83  Instant Nephelometer tacet co cacei cdccla petasenutcateegecect decane tates 3 84    Model 50307 SHARP Instruction Manual xi    Contents    xii    Model 50307 SHARP Instruction Manual    Min and Max Instant Nephelometer Concentration Limits          3 84  Device MVCN Ge tae ctah eda EE la hii ai let ane Boel 3 85  RH Temperature    al ibrati git sentra see telsleaheled fond toiiee dents 3 85  Ambient Temperature  veicciiccevetionsesssssherveedi deageitnctictudenicase 3 86  Fow Temperature oa taka seid Sate ete Dineen een as 3 86  Ambient Relative Humidity geese ene cent Pea eee ay av eD ante mr een arr 3 86  Pressure Vacuum Calibration         ccccccscscsessccesscecsscesessecssseucseseacnse 3 87  Barometer Pressure Caltbrationtiaud ca concave casatdsacenavdecceeivinaocd teers 3 87  Calibrate Barometer Pressure Offset            c ccccsscscssscecsssseesssceesees 3 88  Calibrate Barometer Pressure Span tcrcsntesiersdeaceonaisaceiedeed sonbetonds 3 88  Restore  Default Galibratiofic dsstusnsidsca aera asnaetiasas 3 88  Vacuum Flow Calibration nose iaetiaictad haa iseadell aciwcanematiiats 3 89  Calibrate Vacuum Flow Offfset           ccccccssss
316. eighed relative to the tare  filter and check weight are shown in    50  20  D   Z o  baas  ge          5   100  Y            ow  50    0  50  20  D   a  baas  p               _     jos  WY wo       12   o  ou  50  0  O O O O O O OOU O O O O O OOG O O O O OOOHOO  SBEBSSSFSSFRSRSFSEE RRS RSRSFSSESEBS  ceeeat  evpvnaesezeeQnpkB 88 8eeaeszeszraogpa sg  EegeSRR SSP RR KT SRRESBSEBEEEE  o     r       s    Range of Spread   ug 5    Figure 63  If filters lose relative mass between weightings  then the distribution will be shifted    to the right  whereas if there is a gain in the relative mass the distribution will shift to the left     EN14907 states that unsampled filters should be rejected if the difference between the    masses of the two pre weightings is greater than 40 ug  Similarly  EN14907 states that sam     pled filters should be rejected if the difference between the masses of the two post weight     740300_2013_936_21209885F_EN docx    TUVRheinland TUV Rheinland Energie und Umwelt GmbH  Precisely Right Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system   Page 218 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    ings is greater than 60 ug  Filters were not rejected based on these criteria  The observed  distributions of repeat mass measurements are considered unlikely to have had a significant  effect o
317. eitenmeneoconienenneoeceinaire 41  Vacuum pump  type  GAST Model 75R647  PN 110836 00                    seeeeee 41  Main screen of the user interface              ccccccccecccceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeaeeeaeeeees 42  Flow chart     Overview of menu assiSted firmware           cccccceeeeeeeeeeeeeeeeeeeeeeeees 45  Zero filter for supplying air free of suspended particles and its connection  adaple 6 cersuaciteneetae deine etait oon ae aa sens ati 46  Set with foil holder and test foils 2 2 0 0    ceeeceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeees 47  Leak test adapter eer eee er ene ee eee eee ere ree eee eer ee 48  Course of PM2 5 concentrations  reference  at the test site    Cologne    parking lot  WINIET siccreccecremitesrctemmetn weed ment eran mar 57  Course of PM2 5 concentrations  reference  at the test site    Bornheim    motorway parking lot  SUMME                 cceeeeeeee eect eeeeeetee eset tree eeeeeaaeeeaaeeeeeeeeeneeee 57  Course of PM25 concentrations  reference  at the test site    Bornheim    motorway parking lot  WWE scecsiets eee aie ates 58  Course of PM  5 concentrations  reference  at the test site    Teddington   SUMMON    scvea enact ae Acutdetienct ted eapetastbae seas teuccsuss acewurseustaniguce idtedsdateiahe nce tesueeesddaas 58  Field test site Cologne  parking lOt             cccceeeeeeesseeceeeeeeeeeeeeeseneaeeeeeeeeeeeenenenal 59  Field test site Bornheim  motorway parking lot             ccceeeeeesseeeeeeeeeeeeeeenaaees 59  Field 
318. eive  alarm flow temp max 20 0 degC    set alarm flow temp min value   set alarm flow temp max value   This command sets the flow temperature alarm maximum value to value   where value is a floating point number representing flow temperature alarm  limits in   C  The following example sets the flow temperature alarm  maximum value to 70   C     Send  set alarm flow temp max 70  Receive  set alarm flow temp max 70 ok    alarm ib status  This command reports the status of the interface board as FAIL or OK   The following example reports that the interface board status is OK     Send  alarm ib status  Receive  alarm ib status ok    alarm io status   This command reports the status of the I O expansion board as FAIL or  OK  The following example reports that the I O expansion board status is  OK     Send  alarm io status  Receive  alarm io status ok    alarm mb status  This command reports the status of the motherboard as FAIL or OK  The  following example reports that the motherboard status is OK     Send  alarm mb status  Receive  alarm mb status ok    alarm nb status   This command reports the status of the nephelometer interface board as  FAIL or OK  The following example reports that the nephelometer  interface board status is OK     Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Alarms    Send  alarm nb status  Receive  alarm nb status ok    alarm ncal status  This command reports the status of the last nephelometer auto zero  calibration 
319. eive  set range 5 ok    Table B 5  Standard Ranges                   Selection ng m  mg m   0 100 0 1   1 1000 1 0   2 2000 2 0   3 3000 3 0   4 5000 5 0   5 10000 10 0       Model 50307 SHARP Instruction Manual B 49    C Link Protocol Commands  Measurement Configuration             Selection g m  mg m   6 C1 C1  7 C2 C2  8 C3 C3  rh th    This command reports the current heater RH threshold for the sample   The following example reports that the RH threshold is 30      Send  rh th  Receive  rh th 30      set rh th value  This command sets the heater RH threshold for the sample to value  where  value is a floating point number representation of RH threshold between    25 to 90   The following example sets the RH threshold to 35      Send  set rh th 35  Receive  set rh th 35 ok    temp comp   This command reports whether temperature compensation is for actual or  standard conditions  The following example shows the temperature  compensation is standard     Send  temp comp  Receive  temp comp std    set temp comp selection  selection     act   std      These commands turn the temperature compensation to either actual or  standard conditions  The following example turns the temperature  compensation to actual     Send  set temp comp act  Receive  set temp comp act ok  temp std    This command reports the standard temperature  The following example  reports that the standard temperature is 18   C     Send  temp std  Receive  temp std 18 degC    B 50 Model 50307 SHARP Instruction 
320. el 50307 SHARP Instruction Manual    The inner box that the instrument resides in is an ISTA 1A certified  packaging and is appropriate for transporting the instrument on local paved  roads to the monitoring site  Precautions should be taken to secure this  package from shifting during local transport     Using the ISTA 2A packaging is strongly recommended for transporting  the instrument over poor roads or on highways     When lifting the instrument  use a procedure appropriate to lifting a heavy  object  such as bending at the knees while keeping your back straight and  upright  Grasp the instrument at the bottom in the front and at the rear of  the unit  Although one person may lift the unit  it is desirable to have two  persons lifting  one by grasping the bottom in the front and the other by  grasping the bottom in the rear     Equipment Damage Do not attempt to lift the instrument by the cover or  other external fittings  A    If there is obvious damage to the shipping container when the instrument is  received  notify the carrier immediately and hold for inspection  The carrier  is responsible for any damage incurred during shipment     Use the following procedure to unpack and inspect the instrument     1  Remove the instrument from its shipping container s  and set it on a  table or bench that allows easy access to both the front and rear     2  Remove the cover to expose the internal components     3  Remove the packing material in the analyzer  Figure 2   1      Th
321. eld Reports sets item number and name in scratch pad list B 33  sr Reports the last srec stored B 29  srec Reports maximum number of srecs B 29  srec format Reports sets output format for srecs  ASCII or binary  B 30  srec layout Reports current layout of srec data B 31  srec mem size Reports maximum number of srecs that can be stored B 31  srec per Reports sets srec logging period B 31  stream per Reports sets current time interval for streaming data B 33  stream time Reports sets a time stamp to streaming data or not B 34       Thermo Fisher Scientific    Measurements    Thermo Fisher Scientific    C Link Protocol Commands                                           Measurements   Command Description Page  tape counter Reports the number of counts corresponding to the filter tape B 27   movement  tape to ncal ratio Reports sets the ratio of filter tape changes per B 43   nephelometer auto zero  temp comp Reports sets temperature compensation to actual or standard B 50  temp std Reports sets the standard temperature B 50  temp th Reports sets the temperature threshold for the detector B 51  thermal mass coef Reports sets the current thermal mass coefficient B 40  time Reports sets current time  24 hour format  B 53  tz Reports sets the timezone string for the NTP server B 60  up Simulates pressing up pushbutton B 44  vac mass coef Reports sets the current vacuum mass coefficient B 41  vac pres Reports the current vacuum pressure B 27  vac pres span Reports sets the vacuum pres
322. em     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Streaming Data Config  gt  select Item  gt  Concentrations        Other Measurements The Other Measurements screen allows the user to assign one of the other  available measurement types to the selected streaming data item  The  currently selected item is shown by     lt       after it  Once an item is selected   pressing will save the selected streaming data item  For Common  Flags and Detector A Flags  see Figure 3   5 and Figure 3   6 for data  descriptions     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Streaming Data Config  gt  select Item  gt  Other  Measurements     3 40 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Operation  Volumetric Conditions       ERED CH  Ps Tit    Wi hid  PS PHP ESE Pi AG Lf bie       Analog Inputs The Analog Inputs screen allows the user to assign an analog input signal   none or analog inputs 1 8  to the selected streaming data item  The  currently selected item is shown by     lt       after it  Once an item is selected     pressing will save the selected streaming data item     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Streaming Data Config  gt  select Item  gt  Analog Inputs        TCP IP Settings The TCP IP Settings menu is used for defining parameters that are  required for Ethernet communications     Note The instrument power must
323. embly   59 008630 Pump Rebuild Kit   24 000483 Pump Tubing  External 3 8 inch  106994 00 Tubing  Internal 3 8 inch    104401 00 Tape Guides   106443 00 Filter Tape   108183 00 Tape Hardware Kit   112298 00 Replacement O ring Kit    106531 00 Cover Assembly   106988 00 Handles  front  cover    FH125C14 Calibration Kit  Cal Foils    10 000447 Modified Shoe for Tripod   106445 00 Metal Assembly Paste   107000 00 Fitting  5 8 Union w Nylon Ferrules  107001 00 Tubing  5 8 Stainless Steel   109583 00 Roof Flange Assembly   57 002758 Large Bypass Filter Assembly    110504 00 SHARP Radius Tube Adapter Final Assembly  108154 00 SHARP Connector Bracket Assembly  109408 00 Tubing Assembly  Zero In 3 8 inch   108151 00 SHARP Assembly   110489 00 Final SHARP Cover Assembly w Insulation  109489 00 Inlet Dust Cover   110084 00 Delrin Nut   104961 00 Gasket Extrusion Chamber   104955 00 RH Temp Board Gasket   110481 00 SHARP 3 way Ext Solenoid Valve Final Assembly  110086 00 Heater Adapter CE Kit   110082 00 Custom Gasket for Inlet        Expendable item  not covered by warranty    7 6 Model 50307 SHARP Instruction Manual    Thermo Fisher Scientific    Servicing  Cable List    Cable List Table 7   2 lists the Model 50303 cables  See the    Troubleshooting    chapter  g P    for associated connection diagrams and board connector pin descriptions     Table 7 2  Model 50307 Cables    Part Number  101349 00    Description  AC Power Cable  115 VAC  US        AC Power Cable  220 VAC  EU        DC 
324. emi annual filter tape  changes are expected to be sufficient under normal conditions  no unscheduled filter  changes due to significantly high dust concentrations   The    Filter Tape Counter     function allows for monitoring the current number of filter tape changes as well as  configuring alarm thresholds which remind the user of upcoming changes    4  An inspection of the sensors for ambient temperature and ambient pressure shall be  carried out every three months according to the manufacturer   s specifications    5  A flow rate check must be performed every three months according to the manufac   turer   s specifications    6  The system   s leak tightness shall be inspected every three months according to the  manufacturer s specifications    7  The external pump exhaust filter shall be exchanged every six months    8  As part of an annual basic maintenance  the carbon vanes of the vacuum pump shall  be exchanged and the SHARP optics module shall be cleaned once a year  The  cleaning of the SHARP optics module is ideally performed by an experienced service  technician  Furthermore  it is recommended to carry out a yearly inspection and  if re   quired  a re calibration of the radiometric measurement using the foil set    9  The cleaning of the sampling tube shall be paid attention to during the annual basic  maintenance     The instructions of the manual  chapters 4 and 5  shall be observed when performing  maintenance works  In general  all works can be carried out w
325. emperature  A    Note The optimal contrast will change from one LCD to another  If the  LCD is replaced  the contrast may need to be reset  A    Note If the display contrast is not optimal  but the content on the screen is  visible  select Instrument Controls  gt  Screen Contrast and adjust the screen  contrast  If the content on the screen is not visible  use the    set contrast 10     C Link command to set the screen contrast to mid range  then optimize  the contrast  See    contrast levels    in Appendix B     C Link Protocol  Commands    for more information on this command   amp     e Inthe Main Menu  choose Instrument Controls  gt  Screen Contrast        rip gies  PIL ERE i          The Service Mode screen is used to turn the service mode on or off   Turning service mode    ON    locks out any remote actions and allows access  to parameters and functions that are useful when making adjustments or  diagnosing the Model 5030    The service  wrench  icon on the status bar is  shown when service mode is on  For more information about the service  mode  see    Service Menu    later in this chapter     Thermo Fisher Scientific    Operation  Volumetric Conditions    Note The service mode should be turned off when finished  as it prevents  remote operation  A    e Inthe Main Menu  choose Instrument Controls  gt  Service Mode        Date Time The Date Time screen allows the user to view and change the system date  and time  24 hour format   The internal clock is powered by its o
326. en consent     TUV Rheinland Energie und Umwelt GmbH  D   51105 Cologne  Am Grauen Stein  Tel   49 221 806 5200  Fax   49 221 806 1349    740300_2013_936_21209885F_EN docx    i      TUV Rheinland Energie und Umwelt GmbH  Lah i ne mang Air Pollution Control  recisely Right     Report on the suitability test of the ambient air quality measuring system   Page 2 of 424 Model 5030  SHARP with PM2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    Acknowledgement    TUV Rheinland Energie und Umwelt GmbH would like to acknowledge the involve   ment of Bureau Veritas and the National Physical Laboratory as our UK partners in  this project for their involvement in collecting the UK field data  project management   and data processing     NPL    BUREAU National Physical Laboratory       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control       Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    Page 3 of 424       DAkks    Deutsche  Akkreditierungsstelle  D PL 11120 02 00    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2   pre separator of the company Thermo  Fisher Scient
327. ence  average in g m     Deviation from   reference 4 1 4 1  average in    Deviation from         The single values are presented in annexes 5 and 6 of the appendix     The presentation and the assessment of the measuring uncertainties Wom of days with a rel   ative humidity  gt  70   are shown in Table 28 and Table 29  The single values are presented  in annexes 5 and 6 of the appendix     740300_2013_936_21209885F_EN docx    y      T  V Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system  Page 136 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM  5   Report No   936 21209885 F    Table 28  Comparison candidate SN 1 with reference instrument  rel  humidity  gt  70    all  test sites  measured component PM25       Comparison candidate with reference according to   Guide  Demonstration of Equivalence Of Ambient Air Monitoring Methods   January 2010   Candidate Modell 50301 SHARP SN SN 1   Test site All test sites  rH gt 70  Limit value 30 g m    Status of measured values Raw data Allowed uncertainty 25     Slope b significant  Uncertainty of b  Ordinate intercept a not significant  Uncertainty of a  Deviation at limit value  Uncertainty Uc s at limit value  Combined measurement uncertainty wom  Expanded measurement uncertainty Wom  Status equivalence test       Table 29  Co
328. ence between sample and calibration port    The difference between the measured values obtained by feeding gas at the sample  and calibration port shall not exceed the requirements of Table 2 of Standard VDI 4202  Part 1  September 2010   A value c  at 70   to 80   of the upper limit of the certifica   tion range shall be used as reference point     Note   This test is not relevant for particulate measuring systems     6 2 Equipment   Not applicable    6 3 Performance of test  Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 126 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 3 16 Converter efficiency  In the case of measuring systems with a converter  the converter efficiency shall be at  least 98       Note    This test is not relevant for particulate measuring systems   6 2 Equipment   Not applicable    6 3 Performance of test   Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation 
329. end  set alarm board temp max 38  Receive  set alarm board temp max 38 ok    alarm conc avg neph min   alarm conc avg neph max   These commands report the current 24 hour average nephelometer  concentration alarm minimum and maximum values setting  The following  example reports that the average nephelometer concentration minimum is  0 ug m        Send  alarm conc avg neph min  Receive  alarm conc avg neph min 0 000E 00 pg m3    set alarm conc avg neph min value   set alarm conc avg neph max value   These commands set the 24 hour average nephelometer concentration  alarm minimum and maximum values to value  where value is a floating   point representation of the concentration alarm limits  Values must be in  the units that are currently set for use  The following example sets the  average nephelometer concentration alarm maximum value to 10000  pg m      Send  set alarm conc avg neph max 10000  Receive  set alarm conc avg neph max 10000 ok    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Alarms    alarm conc avg pm min   alarm conc avg pm max   These commands report the current 24 hour average PM concentration  alarm minimum and maximum values setting  The following example  reports that the average PM concentration minimum is 100 pg m        Send  alarm conc avg pm min  Receive  alarm conc avg pm min 1 000E 02 pg m3    set alarm conc avg pm min value   set alarm conc avg pm max value   These commands set the 24 hour average PM concentratio
330. eneration is tied to the instrument   s real time clock  For example   if the logging period for srecs is set to 30 minutes  a new srec will be  generated on every hour and every half hour  10 00  10 30  11 00        Lrecs and srecs function independent of each other  So  for example  an  srec containing just the current SHARP level could be generated every 5  minutes while an lrec containing a full set of diagnostic data could be  generated once every hour     The analyzer s computer system includes three megabytes of flash memory  which is enough to store a full lrec containing 32 data items and a full srec    Model 50307 SHARP Instruction Manual 3 25    Operation  Volumetric Conditions    containing 32 items once each minute for a week   gt 20 000 total records    If logging is limited to the minimum content of date  time  SHARP  concentration and error flags  the analyzer can store data once each minute  for 4 months   gt 190 000 total records      The Datalogging Settings menu allows the user flexibility in how data is  stored and recorded     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings        Select Srec Lrec The Select Srec Lrec screen is used to select the log record type for other  operations in this menu     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Select Srec Lrec        DH Guided  Piet          View Logged Data The View Logged Data screen is used to select the starting point to view  the logged data
331. ent mass coefficient B 40  mass limit Reports sets the current filter mass limit B 48  me  menu  Simulates pressing menu pushbutton B 44  neph Reports the 1 minute running average nephelometer B 12   concentration  neph 24avg Reports the last calculated 24 hour average nephelometer B 12   concentration  neph bkg Reports the current nephelometer background B 26  neph cal bkg Sets the nephelometer into an auto zero calibration routine B 41  neph coef Reports sets the nephelometer span coefficient B 41  neph led current Reports the nephelometer LED current B 13  neph mode Reports the current operating mode of the nephelometer B 26  neph prev bkg Reports the previous nephelometer background B 26  neph raw Reports the raw  unaveraged  nephelometer concentration B 13  neph ref det Reports the nephelometer reference detector voltage B 13  neph reset bkg Resets measured background to zero B 41  neph restore bkg Restores background to the previous background values B 42  neph rh Reports the nephelometer relative humidity B 13  neph rh offset Reports sets the nephelometer relative humidity sensor B 42   offset  neph src level Reports the nephelometer source level B 42  neph temp Reports the nephelometer temperature sensor B 13  neph temp offset Reports sets the nephelometer temperature sensor offset B 42  neph thermal coef Reports sets the nephelometer thermal coefficient B 43  no of Irec Reports number of Irecs stored in memory B 32       Model 50307 SHARP Instruction Manual B 9    C 
332. entific for the component suspended particulate matter PMs 5     Page 209 of 424  Report No   936 21209885 F    Weighed filters are packed in separate polystyrene jars for transport and storage  These jars  remain closed until the filter is placed in  Virgin filters can be stored in the weighing room for  up to 28 days before sampling  Another pre weighing is carried out if this period is exceeded     Sampled filters can be stored for not more than 15 days at a temperature of 23   C or less   The filters are stored at 7   C in a refrigerator     740300_2013_936_21209885F_EN docx    TU V R h ein   an d    TUV Rheinland Energie und Umwelt GmbH    z   Luftreinhaltung  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 210 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    A2 Filter evaluation    The filters are evaluated with the help of a corrective term in order to minimise relative mass  changes caused by the weighing room conditions     Equation     Dust   MF post       Mrara X   MKonpost   MKonpre      F1     MKonpre   average mass of the 3 control filters after 48 h and 72 h pre weighing   MKoNpost   average mass of the 3 control filters after 48 h and 72 h post weighing   Mrara  average mass of the filter after 48 h and 72 h pre weighing   MF post   average mass of the loaded filter after 48 h and 
333. eph temp offset 2 5 ok  neph thermal coef    This command reports the nephelometer thermal coefficient set at the  factory  The following example reports that the nephelometer thermal  coefficient is 0 00058     Send  neph thermal coef  Receive  neph thermal coef 0 00058    set neph thermal coef value   This command sets the nephelometer thermal coefficient to user defined  values to value  where value is a floating point representation of the  coefficient  The following example sets the nephelometer thermal  coefficient to 0 001     Send  set neph thermal coef 0 001  Receive  set neph thermal coef 0 001 ok    tape to ncal ratio  This command reports the ratio of filter tape changes per nephelometer  auto zero  The following example reports this value     Send  tape to ncal ratio  Receive  tape to ncal ratio 1 1    set tape to ncal ratio   This command sets the filter tape change per nephelometer auto zero ratio  to a user defined value  where value is a floating point representation of the  ratio  The following example sets the nephelometer filter tape change to  auto zero to 2 1     Send  set tape to ncal ratio 2 1  Receive  set tape to ncal ratio 2 1 ok    vac pres span   This command reports the vacuum span for the detector at the time of  calibration  The following example reports that the vacuum pressure span is  1 0     Send  vac pres span  Receive  vac pres span 1 000    Model 50307 SHARP Instruction Manual B 43    C Link Protocol Commands  Keys Display    Keys Displa
334. ephelometer Backgrounds are determined during automatic and  manual calibration and are used to maintain a stable zero  The  nephelometer background menu is shown below  The instrument can also  be calibrated manually using this menu  For more information about  calibration  see Chapter 4     Calibration        e Inthe Main Menu  choose Calibration Factors  gt  Neph Bkg Values     Thermo Fisher Scientific    Operation  Calibration Factors Menu       CREE RE Te    Wi rid  E PURE TE Peak Bases PME Oe hei E E oo uei E beet HH       Nephelometer Mode The Nephelometer Mode screen allows the user to place the nephelometer  into one of three modes  Available modes are Sample  Zero  and Stdby   The sample mode begins photometric detection  The zero mode begins a  zeroing process that will begin with a filter tape change and the standby  mode  Stdby  mode will stall any measurement by the nephelometer     e Inthe Main Menu  choose Calibration Factors  gt  Neph Bkg  gt  Neph  Mode        Values The Values screen displays the optional background for each range of the  nephelometer  These values can be control charted by the user to help  evaluate any optional drift     e Inthe Main Menu  choose Calibration Factors  gt  Neph Bkg  gt   Values        WH mrhivt  Pin Piri i       a prin pn PELIR E  Bel LEbdbert pe    EN EEE TE Sanat Bone EY anal Bat    ade POE S ei n       Restore Previous Values The Restore Previous Values screen is used to restore the prior background  values should a poor
335. eports the 24 hour averaging PM concentration  The  following example reports the 24 hour PM concentration value     Model 50307 SHARP Instruction Manual B 13    C Link Protocol Commands  Alarms    Alarms    B 14 Model 50307 SHARP Instruction Manual    Send  pm 24avg  Receive  pm 24avg 10 000E 00 ug m3  sharp    This command reports the 1 minute running average SHARP  concentration  The following example reports the 1 minute running  average value     Send  sharp  Receive  sharp 1 187E 01 pg m3  sharp 24avg    This command reports the last calculated 24 hour average SHARP  concentration  which defaults to 9 999 pg m  if 24 hours has not passed      Send  sharp 24avg  Receive  sharp 24avg 9 999E 03 pg m3    alarm alpha count min   alarm alpha count max   These commands report the current alpha count alarm minimum and  maximum value settings  The following example reports that the alpha  count alarm minimum value is 200     Send  alarm alpha count min  Receive  alarm alpha count min 200    alarm amb rh min   alarm amb rh max   These commands report the current ambient relative humidity alarm  minimum and maximum value settings  The following example reports the  ambient relative humidity alarm minimum value is 35 0      Send  alarm amb rh min  Receive  alarm amb rh min 35 0      set alarm amb rh min value   set alarm amb rh max value   These commands set the ambient relative humidity alarm minimum and  maximum values to value  where value is a floating point number  representin
336. er Pressure Span screen allows the user to view and  Pressure Span pressure sensor calibration span point     Note The operator should use an independent barometer to measure the  ambient pressure and enter the value on this screen before calibrating  a    e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration  gt   Baro Pres Calibration  gt  Span     ip priori  Piet       Restore Default The Restore Default Calibration screen allows the user to reset the default  Calibration values     e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration  gt   Baro Pres Calibration  gt  Set Defaults     3 88 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu       Vacuum Flow Calibration The Vacuum Flow Calibration submenu is used to calibrate the  vacuum flow offset  vacuum flow span  or reset default values     e Inthe Main Menu  choose Service  gt  Pres  Vacuum Calibration  gt   Vac  Flow        Calibrate Vacuum Flow The Calibrate Vacuum Flow Offset screen calibrates the vacuum flow  Offset sensor offset value     e Inthe Main Menu  choose Service  gt  Pres  Vacuum Calibration  gt   Vac Flow  gt  Vac Flow Offset        Calibrate Vacuum The Calibrate Vacuum Pressure Span screen allows the user to view and set  Pressure Span the flow sensor calibration span point     Note The operator should use an independent manometer capable of  measuring approximately 100 mmHg of vacuum  Measure the vacuum    Thermo Fisher Scientific Model 50307 SH
337. er at or  to the right of the cursor one place to the left  PAGE is used to change the  keyboard character page  For the English language  this switches between  upper and lower case alphabetic characters  SAVE stores the string from the  entry line into the parameter  Alternately  if the active cursor is moved to  the entry line  may be pressed to store the string from the entry line    into the parameter        DH iid  PHL EO    The Model 50302 utilizes the menu driven firmware as illustrated by the  flowchart in Figure 3   3  The Power Up screen  shown at the top of the  flowchart  is displayed each time the instrument is turned on  This screen is  displayed while the instrument is warming up and performing self checks   After the warm up period  the Run screen is automatically displayed  The  Run screen normally displays the current SHARP concentrations  but also  has additional Run screens for ambient conditions  sample conditions  and  mass sensor data  From the Run screen  the Main Menu can be displayed  by pressing  _      The Main Menu contains a list of submenus  Each    Thermo Fisher Scientific    Operation  Firmware Overview      This chapter describes each    submenu and screen in detail  Refer to the appropriate sections for more    information     instrument settings    submenu contains related       Oyu JOe UOD  g Jo 09 0q  Y 10 99 9q   pueog uoisuedxg O I    uoneinByuoD puawnysu     swuejy 29u09  synejag Aojoey 310 S  H swunejy MO 4  synez  q 13SN 310 S  H4 M
338. erature rise ensuring negligible loss of semi volatiles from the col   lected sample when the ambient relative humidity is below the threshold to which the heater  is controlling  As the ambient relative humidity increases above the threshold  the applied  heating is optimised to maintain the relative humidity threshold above the beta attenuation fil   ter tape of the radiometric stage  The regulation of the DHS is carried out using the control  variable    relative sample humidity    rH  here  nominal value  lt  50     58    see Section 4 3  Field test  and the filter temperature sensor  here  maximum value of 30   C   The heating  conditions can be adjusted subject to the requirements of the measurement network     Figure 1 and Figure 5 show the path a particle sample travels within the Model 5030i SHARP  measuring system after passing the sampling inlet and sampling tube  including the DHS     Conditioned Aerosol Sample        SHARP Board    SHARP Box shown  with Cover Off    Beta Attenuation    Nephelometer    Figure 1  Sample path within the Model 5030i SHARP measuring system    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    UV Rhe A TUVRheinland    Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 33 of 424  Scientific for the component suspended particulate matter PMz 5     Report No   936 21209
339. ere tested in the laboratory     Laboratory test  part 1  2009      e Description of system functions  e Determination of dependence of sensitivity on ambient temperature  e Determination of dependence of sensitivity on mains voltage    Laboratory test  part 2  2012 2013      Detection limit   Determination of the dependence of zero point   sensitivity on ambient temperature  Determination of the dependence of zero point   sensitivity on the mains voltage  Tightness of the sampling system    The following auxiliary devices were used to determine the system characteristics during the  laboratory test     Climatic chamber  temperature range from    20   C to  50   C  accuracy  lt  1   C   Adjustable isolating transformer   Zero filter for external zero point checks   Span foil sets    The measured values were recorded internally by the instruments  Once stored in the sys   tems  they were transmitted via Ethernet to the iPort software for read out     The laboratory tests and results are presented in Section 6     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 55 of 424  Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    4 3 Field test    The field test was carried out with two identi
340. ermo Fisher Scientific    Installation  Acceptance Testing and Startup Procedures             Remove Packing   2 pieces           Remove Packing   2 pieces                                                                          Units without Optional I O Board Units with Optional 1 0 Board    Acceptance Testing  and Startup  Procedures    Thermo Fisher Scientific    Figure 2 1  Remove the Packing Material   4  Check for possible damage during shipment    5  Check that all connectors and circuit boards are firmly attached   6  Re install the cover    7  Remove any protective plastic material from the case exterior     8  Remove the external pump from its shipping container and place next  to the instrument     The Model 50307 has been bench tested and calibrated at the factory prior  to shipping  The mass sensors  RH sensors  flowmeter temperature sensor   external ambient temperature sensor  barometric pressure  and volumetric  flow rate have been calibrated to traceable standards  For a quick start   proceed immediately to the    Startup    procedures and skip the    Bench  Acceptance Test        To assure the best quality data  it is recommended that you perform an  acceptance test  Furthermore  as part of a quality assurance program  acceptance testing will be conducted prior to field installation  This is an  excellent opportunity to compare the monitor to the primary and transfer  standards utilized by the end user within their specific monitoring    Model 50307 SHARP I
341. ern Hessen  Protocol                  ssssssssssessssesesseesseeseestesteenteenees D 1  Serial Communication Parameters             ces ssecssecsssscsssessesssstessseeenes D 1  TCP Communication Parameters seis tus  cistshsevava nesters desaneeavansseciansteaves D 2  lnstt  ment Add tess  n eto fo0 kal hoe ats Sone at aes alia 22 D 2  Abbreviations Used inaa a RAAS D 2  Basic Command Structure  ssesssssesssessssserssersossersseesoesessesesersesssersessesres D 2  Block Checksum Characters  lt BCC gt        cccccscsesseesssecssscessceseeeeseeesseeens D 3  Gesytec COLTS ct oh Js asian nie iss eair Eis D 3  Instrument Control Command  ST       ccccccccceessssecessssseceesssseceessnees D 3  Data Sampling Data Query Command  DA     ci eeeeseeeseereeeeeeeees D 4  Measurements reported in response to DA command           sees D 7  Operating and Error S tatuisicts  54k teacs Gade ane Gecnrineneene D 8  Appendix E ESM Protocol CommandS           s sssssssssssssssusnnnsssnusnnnnnnessunnnnnnnnnnnnnnnnnnnnnennnnnnnne E 1  ESM Commands Supported   s s sssssssesssesesssssssrsrsssrsrseseseseseseseseseseseseses E 1    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual xvii    Contents    xviii Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Figures    Figure 1 1  SHARP Monitor Sample Path    cc cccececceceeceteseeseeteteeteneeees 1 3  Figure 1 2  Nephelometer Sample and Zero Flow Schematic            0    0cc0  1 5  Figure 2 1  Remove th
342. erved in com   parison with the reference method  The evaluation of the campaign led to the following re   sults     Table 7  Evaluation of the Teddington  summer  campaign with rH adjusted to 50      Orthogonal Regression Limit Value of 30 ug m3  PM2 5 5030i Sharp Dataset  No s r Slope  b      Up Intercept  a      ua Wem       217 ug m3    Teddington Summer w ith 50   RH Setting  38 0 856 0 902     0 058  0 305     0 493 20 03 5 3  Teddington Summer w ith 50   RH Setting  38 0 850 0 819     0 054  0 965     0 459 31 07 5 3       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMs 5    Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 63 of 424       100          90 4 a SN1 Teddington Summer with 50   RH Setting                 80 4  70    60 5    so Al    40 5    PM    5030i Sharp   pg m     30 4    20 5    10 4              gt       Ne    0    r r r r r r r  0 10 20 30 40 50 60 70 80 90 100  PM    Reference   ug m            Figure 26  Comparison of reference and candidate SN 1  measured component PM25  Ted   dington  summer   relative humidity threshold adjusted to 50         100          90 4 4 SN2 Teddington Summer with 50   RH Setting                       80 4 fo    70 4 ra    60 4 x    PM 
343. eseesectesteseeseeseetesteseeseesee 2 6  Table 2 2    O Terminal Board Pin Descriptions       0  0 ccccccceceseeeseeesteeeeees 2 18  Table 2 3  Default Analog OUtpUtS o oo  ccc ce cecescstesestesesteseeteseeteseeeeees 2 19  Table 2   4  D O Terminal Board Pin Descriptions         0 0 cccccccscesestesesteeeteees 2 20  Table 2   5  25 Pin Terminal Board Pin DeSCriptions            cccccececeeesesee tence 2 21  Table 3 1  Front Panel Pushbuttons   2    cece cece cece eeteseeeseeseeeeeeseeseseeeeeteneees 3 3  Table 3 2  Standard Ranges ace  ceestcscteee sort astuercactteentassusncast tetera etvendatuel    3 11  Table 3 3  Default Analog Outputs   0    ccccececsessesecsesesseceesesseeseeseceeseess 3 12  Table 3 4  Analog Output Zero to Full Scale oo    ccc cccecececeeetesesteteseeeees 3 50  Table 3   5  Signal Type Group Choices    ceeececccsessestetesteseeseeseeesteseeseens 3 51  Table 6 1  Troubleshooting   General Guide oo    cs tees teseeteseeeees 6 2  Table 6 2  Troubleshooting   Alarm Messages          c cccececcsescecetesesteteseseetenenees 6 5  Table 6 3  Motherboard Connector Pin Descriptions          2  120011105111051111011111  6 10  Table 6   4  Measurement Interface Board Connector Pin Descriptions           6 15  Table 6   5  Front Panel Board Connector Pin Diagram        0   ccccccceceseeeeeeee 6 20  Table 6   6    O Expansion Board  Optional  Connector Pin Descriptions          6 21  Table 6 7  Digital Output Board Connector Pin Descriptions             0 cccce 6 2
344. est at the site Bornheim  Summer 2013  SN 1  58     amp  SN 2  50    component PM25  raw data    Orthogonal Regression Limit Value of 30 ug m3  Dataset  Slope  b      Up Intercept  a      Ug Wem       217 ug m3  Individual Datasets Bornheim Summer 2013   44 0 917 1 157     0 051  0 372     0 625 35 91 11 4    Orthogonal Regression Limit Value of 30 ug m3  Dataset    Individual Datasets Bornheim Summer 2013   44 0 904    1 001    0 048       0 048   0 518    0 586      0 586 11 75 11 4       Assessment     1  The uncertainty between the two candidate systems is 1 53 g m  and thus is below  the permissible 2 5 yg m     2  The expanded uncertainty for the raw data for SN 1  58   is higher as the acceptable  25   the expanded uncertainty for the raw data for SN 2  50   is below the permis   sible 25      The correction factors  terms detected during suitability testing in 6 1 5 4 10 Calculation  of the expanded uncertainty of the instruments are afterwards applied to the raw data sets     Therefore a correction of the data sets for SN 1 and SN 2 with a slope of 1 068  uncertainty  of slope 0 013  is carried out  This results to the following evaluation     Table 10  Results of the equivalence test at the site Bornheim  summer 2013  SN 1  58      amp  SN 2  50     component PM25  slope correction 1 068    Orthogonal Regression Limit Value of 30 ug m3  Dataset  No s Slope  b      Up Intercept  a      Ug Wom       217 ug m3  Individual Datasets Bornheim Summer 2013   44 0 917 1 080 
345. et format 01  Receive  set format 01 ok    Table B 8  Reply Termination Formats       Format Reply Termination  00  lt CR gt   01  lt NL gt  sum xxxx  lt CR gt     where xxxx   4 hexadecimal digits that represent the sum of all the  characters  bytes  in the message    host name  This command reports the host name string  The following example  reports the host name is set to Series     Send  host name  Receive  host name iSeries    set host name string  This command sets the host name string  where string is 1 13 alphanumeric  characters  The following example sets the host name to analyzer01     Send  set host name analyzero1  Receive  set host name analyzero1 ok  instr name    This command reports the instrument name  The following example  reports the instrument name for the Model 50302     Send  instr name  Receive  instr name  SHARP    Particle Analyzer    instrument id  This command reports the instrument identification  ID   The following  example reports the current setting of the instrument ID     Send  instrument id  Receive  instrument id 14    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Communications Configuration    set instrument id value   This command sets the instrument ID to value  where value is an integer  between 0 and 127 inclusive  The following example shows the instrument  ID changed to 12     Note Sending this command via RS 232 or RS 485 will require the host to  use the new ID for subsequent commands  A    Se
346. etent German body  see module 5 3 1   reference is made to the following  requirement in the previous version of Standard VDI 4202 Part 1  June 2002    The detection limit of the measuring system shall be smaller or equal to reference val   ue Bo  The detection limit shall be determined in the field     6 2 Equipment  Zero filter for testing the zero point   6 3 Performance of test    The detection limit is determined for candidates SN 1 and SN 2 by operating the devices with  the respective zero filters installed at the inlets of both measuring systems  Sampling air free  of suspended particulate matter was offered over a period of 15 days  for a duration of 24ha  day  The determination of the detection limit was done in the laboratory  since the provision  of particulate free air for such a long time was not possible under field conditions     6 4 Evaluation    The detection limit X is determined from the standard deviation s   of the SHARP measured  values by applying particulate free sampling air to both candidates  It corresponds to the  standard deviation of the average Xo of the measured values Xo of each candidate multiplied  with the Students factor     1    X   th 10 95   So where  S o   facet  X    Reference value  Bo   2 g m8       740300_2013_936_21209885F_EN docx    a 3    TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP w
347. etraCal venturi  manufactured by BGI   for use in the UK   Power consumption measuring device type Metratester 5  manufactured by Gosen   Metrawatt    e Zero filter for external zero point checks   e Span foil sets    Two Model 5030i SHARP systems and two PM2 5 reference instruments were each operated  simultaneously for 24 h during the field test  The reference equipment worked discontinuous   ly  i e  the filter had to be changed manually after each sample drawing     740300_2013_936_21209885F_EN docx    a      TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland j    Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 56 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The impaction plates of the PM25 sampling inlets of the reference devices were cleaned ap   prox  every 2 weeks and lubricated with silicone grease in order to guarantee a safe separa   tion and deposition of the particulate  The PMzs5 sampling inlets of the candidates were  cleaned approx  every 4 weeks  The inlet was cleaned according to the manufacturer   s in   struction while taking into account the local suspended particle concentrations     The flow rate was tested on each candidate and reference device prior to and after each    change of location with a dry gas meter and a mass flow meter connected wi
348. ewed but not changed using the front panel interface  and the remote     Set    commands are active     If the instrument keyboard is unlocked via the front panel using Password   gt  Unlock Instrument  the instrument reports being in Local mode  the  front panel interface is unlocked  and settings can be changed from the  front panel     Refer to    C Link Protocol Commands    appendix for detailed information  about    mode        allow mode     and    power up mode    commands     The Change Password screen is used to change the password used to  unlock the instrument   s front panel  The change password screen is shown  if the instrument is unlocked     e Inthe Main Menu  choose Password  gt  Change Password       Thermo Fisher Scientific    Operation  Password Menu    Remove Password The Remove Password screen is used to erase the current password and  disable password protection  The remove password screen is shown if the  instrument is unlocked and the password is set     e Inthe Main Menu  choose Password  gt  Remove Password       Unlock Instrument The Unlock Instrument screen is used to enter the password to unlock the  front panel  The unlock instrument screen is shown if the instrument is    locked     e Inthe Main Menu  choose Password  gt  Unlock Instrument       Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 105    Frequency of    Thermo Fisher Scientific    Calibration    Equipment  Required    Chapter 4  Calibration    The Model 50302 is an i
349. exceed the  reference value By in 24 h and in the maintenance interval   The temporal change in the measured value in the range of the reference value B  shall  not exceed 5   of B  in 24 h and in the maintenance interval     6 2 Equipment  Zero filter for testing the zero point and span foil set for testing the reference point   6 3 Performance of test    The test was performed as part of the field test over a period 20 months     As part of the monthly regular checks  including those at the beginning and at the end of  tests at each field test site   the measuring systems were operated with a zero filter at the  device inlet over a period of at least 24 h each  The values obtained at zero concentration  were evaluated    Furthermore  the stability of the mass coefficient was tested and evaluated as reference point  check at the beginning and at the end of the tests at each field test site    It is not possible to produce a reproducible signal for the sensitivity of the nephelometer us   ing external test standards  Since the nephelometer is continuously calibrated with a very  stable radiometric measurement  a separate examination of the stability of the nephelome   ter   s signal at the reference point can be omitted     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 118 of 424 Model 5030  SHARP with 
350. f pressing the right arrow key  The    edit soft key  prompt    will be displayed for configuration of the new label     Note Not all menu items may be assigned to soft keys  If a particular menu  or screen item cannot be assigned  the key assignment screen will not come  up upon entering right arrow soft key combinations  All items under the  Service menu  including the menu itself  cannot be assigned soft keys  A    Model 50307 SHARP Instruction Manual 3 3    Operation  Firmware Overview    Alphanumeric Entry  Screen    Entry Line         gt     Keyboard                pE  Rri Taic    Firmware  Overview    3 4 Model 50307 SHARP Instruction Manual          Programmable Soft Key Labels    The alphanumeric entry screen is used to enter strings consisting of letters   numbers  and other characters  The cursor may be positioned within the  entry line using the    Jand    gt   keys  Ifa character is entered over an  existing character  that character will be overwritten  Use the     and         keys to switch between the entry line and the keyboard as well as to  move within the keyboard  To select a character to add to the string  use  the cursor keys to position the cursor over the desired character  and then    press the key to add that character to the entry line     On the right side of the keyboard are special functions  BKSP is used to  move the cursor in the entry line one place to the left  deleting the  character that was to the left of the cursor and moving any charact
351. f the measuring inlet ensures a low internal background produced by opti   cal reflection  The analogue raw signal is detected in cycles of one second  digitised for fur   ther processing and the minute average is calculated  The nephelometer module offers six  automatically chosen measuring ranges which ensure that measured values as precise as  possible are obtained  During the periodic zero adjustment procedures  performed as part of  every filter change   the zero adjustment magnetic valve on the rear side is activated and  HEPA filtered air    free of suspended particles  is admitted into the nephelometer  Every  range is then zeroed again  The results of the zero adjustment can be read on the display  and are as C link command available     At the location at which the nephelometer is attached to the enclosure of measuring system   an isolated radial tube hits the sampling tube  The nephelometer module can thereby be eas   ily detached from the actual measuring system  This makes it possible to reconfigure the  Model 5030i SHARP measuring system  combination of nephelometric and radiometric  measurements  into the Model 50141 BETA measuring system  see test report  936 21209885 H      After passing the nephelometer  the particles of the particle sample are deposited on the  glass fibre filter tape for radiometric measurement  The particles are collected on the filter  tape up to a specified threshold  during test  fast forward of filter tape after 8 h or when mass  is  
352. ference  recognised as signif   icant outliers and discarded at each test site     Table 12  Results of Grubbs outlier test     Reference PM25    Graph Number Site Sampler Number of data    Maximum Number Number identified   Number Deleted Number of data   pairs that can be deleted pairs remaining    Bornheim Winter PM  s Reference  Cologne Winter PM  s Reference    Bornheim Summer   PM  Reference       Teddington Summer   PM  5 Reference    740300_2013_936_21209885F_EN docx    i 7    TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 70 of 424    The following paired values were discarded                                         Table 13  Discarded reference PM  paired values according to Grubbs  Test site Date Reference 1  ug m       Reference 2  g m3   Cologne  Winter  05 May 2011 11 2 14 8  Teddington  summer  03 Aug 2012 5 5 4 6   100    90 4   Outlier Deleted       Outlier Not Deleted       1 to 1 Line          80         z aouaIajay Wd                90 100    PM   5 Reference 1             Figure 28  Grubbs test results for the PM25 reference method  Cologne  winter     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land       Air Pollutio
353. ffset value   This command sets the nephelometer relative humidity sensor offset to  user defined values to value  where value is a floating point representation  of the offset  The following example sets the nephelometer relative  humidity sensor offset to 3      Send  set neph rh offset 3  Receive  set neph rh offset 3 ok    neph src level  This command reports the nephelometer source level set at the factory  The  following example reports that the nephelometer source level is 44      Send  neph src level  Receive  neph src level 44      set neph src level value   This command sets the nephelometer source level to user defined values to  value  where value is a floating point representation of the level  The  following example sets the nephelometer source level to 50      Send  set neph src level 50  Receive  set neph src level 50 ok    neph temp offset  This command reports the nephelometer temperature offset set at the  factory  The following example reports that the nephelometer temperature    offset is 2 6   C     Send  neph temp offset  Receive  neph temp offset 2 6 degC    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Calibration    set neph temp offset value   This command sets the nephelometer temperature offset to user defined  values to value  where value is a floating point representation of the offset   The following example sets the nephelometer temperature offset to 2 5   C     Send  set neph temp offset 2 5  Receive  set n
354. for the component suspended particulate matter PMz 5    Report No   936 21209885 F    filled    5 2 8 Availability Minimum 95    The availability was 97 4   for SN 1  and 96 1   for SN 2 without test   related outages  or 90 2   for SN 1  and 89 0   for SN 2 including test   related outages     5 2 9 Instrument soft    The version of the instrument   The version of the instrument soft   ware software to be tested shall be   ware is shown on the display  Chang    displayed during switch on of es to the instrument software are  the measuring system  The test   communicated to the test institute   institute shall be informed on  changes in the instrument  software  which have influence  on the performance of the  measuring system     5 3 Requirements on measuring systems for gaseous air pollutants    5 3 1 General Minimum requirement accord    The test is based on the minimum re   ing VDI 4202 Part 1  quirements of Standard VDI 4202   Part 1  September 2010   Test points  5 3 2  5 3 7  5 3 8  5 3 11 and 5 3 12  are evaluated as before basing on the  minimum requirements stated in VDI  4202 Part 1 of 2002  i e  using the  reference values Bo  B  and B2   The  performance of the test points 5 3 3   5 3 4  5 3 13 and 5 3 18 is omitted  as  they are not relevant for particulate  measuring devices     5 3 2 Repeatability The repeatability standard de    The investigations result in detection  standard devia    viation at zero point shall not   limits of 0 11 g m8 for system 1  SN  ti
355. from  Date Value previous value start value Date Value previous value start value  pg m  pg m  pg m  pg m  pg m  pg m   3 5 2011 0 0  3 6 2011 3 6 2011 0 0  3 7 2011 3 7 2011  4 8 2011 4 8 2011  4 9 2011 4 9 2011  4 10 2011 4 10 2011  5 10 2011 5 10 2011  5 11 2011 5 11 2011  7 22 2011 7 22 2011  7 23 2011 7 23 2011  7 24 2011 7 24 2011  9 17 2011 9 17 2011  9 18 2011 9 18 2011  10 22 2011 10 22 2011  10 23 2011 0 0 0 2  11 11 2011 11 11 2011 11 1 0  11 12 2011 11 12 2011 0 0  1 1  11 13 2011  2 11 2012  0 1  0 2  2 12 2012 2 0 2 0 1 9 2 12 2012 2 3 2 4    3 9 2012  3 10 2012 0 0 0 1 0 0 0 0  9 8 2012 0 0  0 1  0 1 9 8 2012  0 1  9 9 2012 0 0 0 0  0 1 0 0  10 12 2012 0 0 0 0 0 0 0 0  10 13 2012 0 0 0 0  0 1 10 13 2012 0 7          740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  TUVRheinland       Air Pollution Control    Precisely Right     Report on the suitability test of the ambient air quality measuring system    Page 120 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F          Measured value  ug m    IANO   O O OOGO  7   O  0     Zero point drift PM2 5 SN 1    9 N N N QS Y        N N N N N K N N  O O O O O O O  K D v yw oe w we we oe        N      X D    N RN N  Time          Figure 35  Zero point drift SN 1  measured component PM25       Measured value  ug m            Zero point drift PM2 5 SN 2          Figur
356. g ambient relative humidity alarm limits in percent  The  following example sets the ambient relative humidity alarm maximum  value to 55      Send  set alarm amb rh max 55  Receive  set alarm amb rh max 55 ok    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Alarms    alarm amb temp min   alarm amb temp max   These commands report the current ambient temperature alarm minimum  and maximum value settings  The following example reports that the  ambient temperature alarm minimum value is 4 0   C     Send  alarm amb temp min  Receive  alarm amb temp min 4 0 degC    set alarm amb temp min value   set alarm amb temp max value   These commands set the ambient temperature alarm minimum and  maximum values to value  where value is a floating point number  representing ambient temperature alarm limits in   C  The following  example sets the ambient temperature alarm maximum value to 38   C     Send  set alarm amb temp max 38  Receive  set alarm amb temp max 38 ok    alarm baro pres min   alarm baro pres max   These commands report the current barometric pressure alarm minimum  and maximum value settings  The following example reports that the  barometric pressure alarm minimum value is 500 0 mmHg     Send  alarm baro pres min  Receive  alarm baro pres min 500 0 mmHg    set alarm baro pres min value   set alarm baro pres max value   These commands set the barometric pressure alarm minimum and  maximum values to value  where value is a floating point 
357. g example sets the sample relative humidity alarm maximum value    to 40      Send  set alarm sample rh min 40  Receive  set alarm sample rh min 40 ok    alarm status det  This command reports the board status of the detector as FAIL or OK  The  following example reports that the detector board status is OK     Send  alarm status det  Receive  alarm status det ok    Thermo Fisher Scientific    Diagnostics    Thermo Fisher Scientific    C Link Protocol Commands  Diagnostics    alarm tape counter max   This command reports the current filter tape count alarm maximum value  setting  The following example reports that the filter tape count alarm  maximum value is 480     Send  alarm tape counter max  Receive  alarm tape counter max 480    set alarm tape counter max value   This command sets the filter tape count alarm maximum value  where value  is an integer number representing filter tape alarm limits  The following  example sets the filter tape count alarm maximum value to 490     Send  set alarm tape counter max 490  Receive  set alarm tape counter max 490 ok    alarm vac pres min   alarm vac pres max   These commands report the current vacuum pressure alarm minimum and  maximum value settings  The following example reports that the vacuum  pressure alarm minimum value is  20 0 mmHg     Send  alarm vac pres min  Receive  alarm vac pres min  20 0 mmHg    set alarm vac pres min value   set alarm vac pres max value   These commands set the vacuum pressure alarm minimum and maxim
358. ght   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 75 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6  Test results    6 1 4 1 1 Measured value display  The measuring system shall be fitted with a measured value display   6 2 Equipment  No additional equipment is required   6 3 Performance of test  It was checked whether the measuring system has a display for measured values   6 4 Evaluation    The measuring system has a display for measured values  The main screen displays the cur   rent SHARP concentration at all times  updated every second   From here  it is easy to select  the screens displaying the PM concentration    BETA   the NEPH concentration  the ambient  conditions  the flow conditions and the mass sensor data using the arrow keys 7       6 5 Assessment    The measuring system is fitted with a measured value display   Minimum requirement fulfilled  yes    740300_2013_936_21209885F_EN docx    A TUVRheinland      TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 76 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F  6 6    Detailed presentation of test results    Figure 32
359. gt  is abbreviation for space  ASCII code 0x20     The following is the basic structure of a Gesytec command    lt STX gt Command text lt ETX gt  lt BCC gt    OR    lt STX gt Command text lt CR gt     Each command is framed by control characters   lt STX gt  at the start and  terminated with either  lt ETX gt  or  lt CR gt      If a command is terminated with  lt ETXs gt   then two additional characters     lt BCC gt  is attached after  lt ETX gt   This is the block checksum     Block Checksum Characters  lt BCC gt  may be added to the command to  prevent processing invalid commands     Thermo Fisher Scientific    Block Checksum  Characters  lt BCC gt     Gesytec Commands    Instrument Control Command    Thermo Fisher Scientific     ST     Gesytec  Bayern Hessen  Protocol  Block Checksum Characters  lt BCC gt     The Block Checksum Characters are calculated beginning with a seed value  of 00000000  binary  0x00   and bitwise exclusive ORing with each of the  characters of the command string  or response  including the framing  characters  lt STX gt  and  lt ETX gt   The checksum works as an error check  The  command terminator determines the presence or absence of  lt BCC gt      If a command is terminated by  lt ETX gt  then the next two characters are the  checksum  if the command is terminated with  lt CR gt  no checksum is  attached     The block checksum is represented by two characters  which represent a 2  digit hex number  1byte   e g  1 byte OxAB hex checksum will be
360. gt 1500 ug on the filter tape or when the deviation of the current flow rate from the nominal  flow rate is more than 2 5        The filter tape can be found between the proportional detector and the 14C beta source  The  beta ray is transmitted from the bottom upwards passing through the filter tape and the ac   cumulated particle layer  The intensity of the beta ray is attenuated by the increasing mass  load  thus reducing its intensity  The reduced beta intensity is then measured by the propor   tional detector  The mass on the filter tape is calculated from the continuous integrated count  rate  The determined count rate is continuously corrected with regard to changes in the air  density     Furthermore  the radiometric mass determination is optimised by the simultaneous determi   nation and consideration of the natural a activity of aerosols  This additional a count rate  serves the purpose of measuring radon  which causes undesirable interference in the radio   metric mass determination of particles and must therefore be compensated     In order to keep the sampling flow rate at a constant nominal value  the continuous meas   urement of the flow rate as well as its adjustment is carried out by means of a proportional  valve     The PM concentrations are shown on the display on the front side of the measuring system  as SHARP    hybrid values   PM    radiometrically measured values  gt  analogous to Model  50141 BETA  and NEPH    values measured with scattered light   T
361. hall not   was 2 4 g m  in relation to the previ   exceed the requirements of   ous value and 2 2 g m  in relation to  Table 2 in the field test of  the start value  and is therefore within  Standard VDI 4202 Part 1   the permissible limits of Bo   2 ug m       September 2010  in the field   The values obtained during the sensi   test  tivity drift tests amount to a maximum  For PM  of  2 9   in relation to the previous    Zero point  within 24 h and   value for PMas   within the maintenance interval  a maximum of BO     As reference point  within 24 h  and within the maintenance in   terval a maximum 5   of B1     5 3 13 Short term drift  The short term drift at zero   Not applicable    point and reference point shall  not exceed the requirements of  Table 2 of Standard VDI 4202  Part 1  September 2010  within  12 h  for benzene 24 h  in the  laboratory test and within 24 h  in the field test        740300_2013_936_21209885F_EN docx    T aa VRh 7      TUV Rheinland Energie und Umwelt GmbH   U einland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 24 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    filled    5 3 14 Response time   The response time  rise  of the   Not applicable   measuring system shall not ex   ceed 180 s    The response time  fall  of the  measuring syste
362. he ambient air measuring system Model 5030i SHARP operates combining the principles of  particle light scattering  nephelometry  and beta attenuation  The designation SHARP stands  for    Synchronised Hybrid Ambient Real time Particulate     The clever combination of a highly  sensitive but potentially inaccurate measuring technique  NEPH  and a non sensitive but  highly accurate real time measurement  BETA  aims at providing a precise and  at the same  time  sensitive real time measurement of fine dust     Nephelometry is based on the measurement of light radiation intensity  which is scattered by  a totality of particles passing through a measured volume defined by the intersection of the il   lumination beam and the field of view of the detection optics  In contrast to particle counters   the measured volume shall be large in terms of inverse function of the concentration of parti   cle numbers  The resulting signal is linearly proportional to the particle volume concentration  for an aerosol with constant optical characteristics  particle size  form and refractive index    provided that the conditions of independent and single scattering that shall be maintained for  the concentration range in focus are met     A nephelometer   s response R is defined by the following formula as the relation of the detec   tion signal for light scattering to the mass concentration of the particles        R    ff f  a  j al    i    i   f A F O dd  dado    sar  where   lo   irradiation inte
363. he beta measurement chamber houses an RTD temperature sensor  The  T sensor directly measures the temperature of the incoming sample stream  for a continuous air mass compensation  and the flow system  The flow and  ambient sensor measurements are combined within a firmware algorithm to  maintain an accurate measurement and control of volumetric flow at the  inlet and through the differential orifice     The Model 50303 also includes three pressure sensors to measure the  differential pressure across a subsonic orifice  vacuum under the filter tape   and barometric pressure  The flow P sensor is solely used for measurement  at the orifice for flow rate control  The vacuum sensor is used for air mass  compensation  as an indicator for clogging of the system and is included in  the volumetric flow control  The barometric P sensor is used for air mass  compensation and for volumetric flow rate regulation     The volumetric flow system is built on a calibrated feedback signal from the  CPU to the proportional valve signal  By using this flow control  flow  stability is easily maintained     With regard to the hybrid  50307 SHARP Monitor methodology  the  nephelometer and beta attenuation concentrations are calculated and both  values are digitally filtered  e g  averaged  with an elastic time constant that  is a function of the real time concentrations  The time constant of this  dynamic digital filter will vary with concentrations  It is the ratio of those  two digitally filtere
364. he high level firmware and loaded  on power up if a different version is detected     Each board has a specific address that is used to identify to the firmware  what functions are supported on that board  This address is also used for  the communications between the low level processors and the high level  processor     Every tenth of a second  the frequency counters  analog I O  and digital  I O are read and written to by the low level processors  The counters are  accumulated over the past second and the analog inputs are averaged over  that second  The high level processor polls the low level processors once  per second to exchange the measurement and control data     Model 50307 SHARP Instruction Manual 8 5    System Description  Electronics    Monitoring Signals    Output  Communication    Electronics    Motherboard    8 6 Model 50307 SHARP Instruction Manual    Signals are gathered from the low level processors once per second  and  then processed by the high level processor to produce the final  measurement values  The one second accumulated counts are accumulated  and reported for the user specified averaging time  If this averaging time is  greater than ten seconds  the measurement is still reported every 10  seconds  The one second average of the other analog inputs are reported  directly  no additional signal conditioning is performed by the high level  processor      The front panel display  serial and Ethernet data ports  and analog outputs  are the means of com
365. he highly sensitive but potentially inaccurate  measurement technique  NEPH  is continuously calibrated by the non sensitive but very pre   cise measurement technique  BETA   The mass collected on the filter tape is determined by  the radiometric measurement  The calculated mass is then related to a defined and actively  regulated sampling volume  thus determining the particulate mass concentration     6 5 Assessment    Testing was carried out in compliance with the minimum requirements stated in Table 5 of  Standard VDI 4202  Part 1  September 2010      The calculated mass is related to a defined and actively regulated sampling volume  thus de   termining the particulate mass concentration     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland  oe ait Pollut    Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 130 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    6 1 5 4 2 Equivalency of the sampling system  The equivalency between the PM  9 sampling system and the reference method accord   ing to EN 12341  T5  shall be demonstrated     Note    Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 10 of th
366. he inlet of the instrument  By reviewing the top view  drawing of the instrument  Figure 2   3   the roof flange must be centered  in direct alignment with the 5 8 inch OD tube entering the 5030  instrument     Thermo Fisher Scientific    Installation  Setup    17 3  4404         168  4255            i J so  1273   i T    E    223  568 0   25 0  635 0     Figure 2   4  Model 5030  Top View   inches  millimeters     Heater Once the x y coordinates of the instrument and roof flange system are  established and a waterproof flange system has been installed  perform a dry  run by attaching the heated sample tube on top of the 5030  base  If the  heated tube protrudes through the roof flange  either the instrument should  be lowered or an elongated PVC sleeve and cap system should cover the  heater while maintaining a waterproof seal  However  for the remainder of  this description  it is assumed the heater will fit within the ceiling height of  the air monitoring station     Based on the vertical height requirements  use additional 5 8 inch OD  stainless tubing and connectors to make up the difference between the    desired inlet height and the top of the heater     Additional routing is necessary for the ambient T RH sensor assembly     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 2 11    Installation  Setup                 Ambient Inlet   PM or PMs     Ambient T RH    T Fitting Sensor Assembly    Tripod Adapter  T RH Sensor Cable    Tripod                       
367. he limit value for the annual average     Thermo Fisher Scientific has commissioned TUV Rheinland Energie und Umwelt GmbH to  perform of a suitability test of Model 5030  SHARP measuring system for component sus   pended particulate matter PMo 5s     e Standard VDI 4202 Part 1     Performance criteria for performance tests of automated  ambient air measuring systems     Point related measurement methods for gaseous    and particulate air pollutants     September 2010 or June 2002  1     e Standard VDI 4203 Part 3     Testing of automated measuring systems   Test proce   dures for point related ambient air measuring systems for gaseous and particulate air  pollutants     September 2010 or August 2004  2     e Standard EN 14907     Ambient air quality   Standard gravimetric measurement meth   od for the determination of the PM25 mass fraction of suspended particulate matter      German version EN 14907  2005  3     e Guide    Demonstration of Equivalence of Ambient Air Monitoring Methods     English  version of January 2010  4     740300_2013_936_21209885F_EN docx    a      TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland j    Air Pollution Control  Precisely Right   Report on the suitability test of the ambient air quality measuring system    Page 16 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PMzs    Report No   936 21209885 F   The measuring system Model 5030i SHARP is a synchr
368. he maximum deviation at zero point was 2 4 g m  in relation to the previous value and 2 2  g m    in relation to the start value  and is therefore within the permissible limits of Bo    2 ug ms    The values obtained during the sensitivity drift tests amount to a maximum of  2 9   in rela   tion to the previous value for PMo 5     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results    Table 24 presents the values obtained at zero concentration and the calculated deviations in  relation to the previous value and to the start value in g m  Figure 35 to Figure 36 presents  a graph of the zero point drift during the entire test     Table 25 shows the deviations of the measured values from the respective previous value in     Figure 37 and Figure 38 presents a graph of the drift of the measured values  in relation  to the previous value      740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5     Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 119 of 424                                                                         Table 24  Zero point drift SN 1  amp  SN 2  test with zero filter  SN1 SN2  Measured Deviation from Deviation from Measured Deviation from Deviation 
369. he measured values can  be obtained through different output paths  analogue  digital  Ethernet      740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control A     Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 37 of 424    3 3 AMS scope and set up    The measuring system comprises the PM25 sampling inlet  the heated sampling tube  dy   namic heating system DHS   the  optional  extension tube  the ambient air sensor  including  radiation protection shield   the vacuum pump  the nephelometer module    SHARP optics  module   the central unit     SHARP beta module  identical in design to the 5014i measuring  system  see test report 936 21209885 H  including glass fibre filter tape  the respective con   nection lines  cables and adapters  roof bushing with flange  and a manual in German        Figure 6  Model 5030i SHARP  SHARP optics modules  top    SHARP beta module  bot   tom      740300_2013_936_21209885F_EN docx    A TUVRheinland    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 38 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the compo
370. he measurement interface board     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 33    Servicing  Detector Assembly Replacement    4  Disconnect plumbing     5  Remove four mounting screws and remove proportional valve  assembly     6  Replace the proportional valve assembly by following the previous steps  in reverse     7  Perform a leak test as described in the    Preventive Maintenance     chapter     Unhook Tubing from  Measurement Head    Unfasten Screws  4     Unhook Tubing from  Rear Panel       Unhook Connector from SS  Measurement Interface a  Board    Beta 1       Figure 7   17  Replacing the Proportional Valve    Detector Assembly Use the following procedure to replace the detector assembly  Figure 7       Replacement       Equipment Required     Detector assembly  Hex driver  4 mm    Safety glasses    7 34 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Servicing  Detector Assembly Replacement    WARNING The detector assembly should be replaced by a qualified  technician knowledgeable in dealing with radiation precautions  By  removing the detector assembly  the C 14 radioactive source is partially  exposed  Safety glassed must be worn during this replacement procedure   The amount of C 14 meets the USNRC regulations as an exempt amount  of radioactive source  lt 100 pCi  A    The service procedures in this manual are restricted to qualified  representatives  A    If the equipment is operated in a man
371. he new value for flow span based on the  corrected flow entered  The following example sets the new value for flow  span based on the corrected flow to 17 2 LPM     Model 50307 SHARP Instruction Manual B 37    C Link Protocol Commands  Calibration    B 38 Model 50307 SHARP Instruction Manual    Send  set cal flow span 17 2  Receive  set cal flow span 17 2 ok    set cal vac pres span value   This command automatically sets the vacuum span based on the vacuum  pressure entered at the time of calibration  The following example  successfully sets the new value for vacuum span based on the vacuum  pressure to 120 5 mmHg     Send  set cal vac pres span 120 5  Receive  set cal vac pres span 120 5 ok    set cal vac flow offset   This command automatically triggers the filter change and sets the new  value for vacuum and flow offset  The following example successfully sets  the vacuum and flow offset     Send  set cal vac flow offset  Receive  set cal vac flow offset ok  coef pm    This command reports PM coefficient  The following example reports that  the PM coefficient is 1 200     Send  coef pm  Receive  coef pm 1 200    set coef pm value   This command sets the PM coefficient to user defined values to value   where value is a floating point representation of the coefficient  The  following example sets the PM coefficient to 1 200     Send  set coef pm 1 200  Receive  set coef pm 1 200 ok  coef sharp    This command reports SHARP coefficient  The following example reports  that th
372. head during a filter tape change     After the measurement head is opened  this motor is used to advance the  filter tape a fixed amount prior to initiating another beta attenuation event  on a fixed sample spot     Thermo Fisher Scientific    Pressure Board    External Solenoid    Pump Intake T   Fitting    Firmware    Instrument Control    Thermo Fisher Scientific    System Description  Firmware    This PCB assembly is mounted on the floor plate and has three pressure  sensors attached  The pressure board is used to measure the barometric  pressure  differential pressure across the orifice and vacuum under the filter  tap  All tubing attached to the pressure board is numbered in case of the  need to swap out this board     The external solenoid is used to permit the pump to either exhaust  normally  or to deliver filtered air to the instrument during the  nephelometer zero     The pump intake T fitting provides excess air specifically for periods of  nephelometer zeroing     The processor firmware tasks are organized into four areas   e Instrument control  e Monitoring signals  e Measurement calculations    e Output communication    Low level embedded processors are used to control the various functions on  the boards  such as analog and digital I O  These processors are controlled  over a serial interface with a single high level processor that also controls  the front panel user interface  The low level processors all run a common  piece of firmware that is bundled with t
373. hermo Fisher Scientific    Optional Equipment  Roof Flange Assembly    The flange design also has four holes that can be used to secure the flange  to the roof surface by the use of carriage bolts  washers and nuts  The  length of the carriage bolts depends on the thickness of the roofline  Figure  9   6 shows a picture of the roof flange that is used with this assembly        Figure 9   6  Securing the Roof Flange    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 9 9    Instrument  Preparation    Thermo Fisher Scientific    Chapter 10  Model 5014  to 5030  Upgrade  Procedures    This chapter describes the 50147 to 50307 upgrade procedures  For details   see the following topics        Instrument Preparation    on page 10 1      SHARP Connection Bracket Assembly Installation    on page 10 2     Radius Tube Adapter Installation    on page 10 5      SHARP External Valve Installation    on page 10 7      SHARP Optics Assembly Installation    on page 10 9      Upgrade of the Heater Tube Assembly    on page 10 10      Model Label Change    on page 10 11       Power On Instrument    on page 10 11    Use the following procedure to prepare the instrument     If upgrading at an installed site  take the appropriate measures to secure  the sample downtube and remove the instrument heater assembly and  set aside     From Instrument Panel  select Main Menu  select Instrument Controls  and then Service Mode  Toggle to    Service Mode On        From Instrument Panel  select F
374. his replacement procedure  The  amount of C 14 meets the USNRC regulations as an  exempt amount of radioactive source  lt 100 pCi  A    The detector pre amp assembly has a power backup built  in     DO NOT OPEN    caution handeling  Backup power  will dissipate after 8 hours  A    The detector window is very fragile  Handle with great  care and do not wipe or touch the window  Furthermore   do not touch the solder cable connection  electrode  with  your bare fingers  The oils from your skin can damage the  detector  A    Safety glasses must be worn while replacing the detector  assembly   amp     If the LCD panel breaks  do not let the liquid crystal  contact your skin or clothes  If the liquid crystal contacts  your skin or clothes  wash it off immediately using soap  and water  A    Do not attempt to lift the analyzer by the cover or other  external fittings  A    Some internal components can be damaged by small  amounts of static electricity  A properly grounded  antistatic wrist strap must be worn while handling any  internal component   amp     Disconnect the serial cable before changing the RS 232  and RS 485 selection to prevent damage to any  equipment currently connected to the analyzer  A    Do not use solvents or other cleaning products to clean  the outside case  A    Do not remove the LCD panel or frame from the LCD  module  A    The LCD polarizing plate is very fragile  handle it  carefully   amp     Do not wipe the LCD polarizing plate with a dry cloth  as  it 
375. hod  shall be demonstrated according to chapter 9 5 3 2 to chapter 9 6 of the guidance doc   ument    Demonstration of Equivalence of Ambient Air Monitoring Methods    in the field  test at least at four sampling sites representative of the future application  The maxi   mum expanded uncertainty of the systems under test shall be compared with data  quality objectives according to Annex A of Standard VDI 4202 Part 1  September  2010      6 2 Equipment   Auxiliary equipment described in Section 5    6 3 Performance of test   The test was carried out at four different comparisons during field test  Different seasons and  varying concentrations for PM25 were taken into consideration     Of the complete dataset  at least 20   of the concentration values determined with the refer   ence method shall be greater than the upper assessment threshold according to 2008 50 EC   7   For PM25 the upper assessment threshold is at 17 g m       At each comparison campaign at least 40 valid data pairs were determined  Of the complete  dataset  4 test sites  209 valid data pairs for SN 1  217 valid data pairs for SN 2  a total of  29 1   of the measured values are above the upper assessment threshold of 17 g m  for  PM 5  The measured concentrations were referred to ambient conditions     6 4 Evaluation   Point 9 5 3 2  The calculation of the in between uncertainty of the reference devices Uret is  carried out prior to the calculation of the expanded uncertainty of the candidates     The in bet
376. hogonal regression    u x    random uncertainty of the reference procedure if value Ups  which is calcu   lated for using the candidates  can be used in this test  refer to point 6 1  5 4 9 Determination of uncertainty between systems under test ubs     Algorithms for the calculation of ordinate intercept a as well as slope b and its variances by  orthogonal regression are described in detail in annex B of  4      The sum of the  relative  residuals RSS is calculated by the following equation   RSS   yy  a bx      Uncertainty ucr is calculated for   e All test sites and comparisons together  e Each test site and comparison individually    e  1 dataset with measured values PM2 5 2 18 g m     Basis  average values of the refer   ence measurement     Preconditions for acceptance of the full dataset are that   e The slope b is insignificantly different from 1   b  1   lt  2 u b   and  e The intercept a is insignificantly different from 0   a   lt  2 u a     Where u b  and u a  are the standard uncertainties of the slope and intercept  respectively  calculated as the square root of their variances  If these preconditions are not met  the can   didate method may be calibrated according to point 9 7 of the Guide  refer to 6 1 5 4 11   Application of correction factors and terms  The calibration shall only be applied to the full  dataset     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control      Precisely Right    
377. ht    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 99 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 3 1 General    The test is based on the minimum requirements of Standard VDI 4202 Part 1  Septem   ber 2010      6 2 Equipment  No additional equipment required   6 3 Performance of test    The test is based on the minimum requirements of Standard VDI 4202  Part 1  September  2010      6 4 Evaluation    Standard VDI 4202  Part 1 and VDI 4203  Part 3 were republished after an extensive revision  with latest update September 2010  Unfortunately  after this revision uncertainties and con   tradictions remain regarding the concrete minimum requirements on the one hand  and the  general relevance of the test points on the other hand when testing particulate ambient air  measuring devices  The following test points require clarification     6 1 5 3 2 Repeatability standard deviation at zero point  no minimum requirement defined  6 1 5 3 3 Repeatability standard deviation at reference point    not relevant for particulate devices   6 1 5 3 4 Linearity  lack of fit    not relevant for particulate devices   6 1 5 3 7 Sensitivity coefficient of surrounding temperature  no minimum requirement defined   6 1 5 3 8 Sensitivity coefficient of supply voltage   no minimum requirement defined   6 1 5 3 11 Standard deviation
378. ially disturb the zero  To evaluate the zero of the 5030i  place a  HEPA filter at the inlet for approximately five minutes  or until the  concentration has stabilized  The nephelometer concentration should be  1  ug m    If this is not the case  the HEPA filter should be removed  and a    Model 50307 SHARP Instruction Manual 2 9    Installation  Setup    Setup    Inlet System    nephelometer zero should be conducted  After re zeroing the nephelometer   repeat the zero check steps above     Setting up this monitor includes siting the monitor per local government  agency requirements  creating a waterproof inlet system installation with  vertical support  connecting the inlet system to the instrument through the  dynamic heater system  connecting vacuum lines  establishing datalogging  setting  and making the power connections     When siting the inlet system  it is best to first establish the instrument  location and height within the air monitoring station  From whatever  chosen surface the instrument is to rest on  rack or table top   add  approximately 24 inches to that height to allow for overall height of the  50307 monitor     Next allow approximately 40 inches to attach the heated sample tube to the  top of the 5030  via 5 8 inch OD tubing unions that use Teflon ferrules        2 10 Model 50307 SHARP Instruction Manual    Figure 2 3  Model 50307 Side View   inches  millimeters     It is also important to locate the roof flange installation in vertical  alignment with t
379. ic Model 50307 SHARP Instruction Manual 7 45    Servicing  SHARP RH Temp Sensor Replacement    SHARP RH T EMP     Use the following procedure to replace the SHARP RH Temp sensor  Sensor  Figure 7 26    Replacement    T Equipment Damage Some internal components can be damaged by small    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  nephelometer from the base instrument     2  Remove the two side cover screws  and remove cover     3  Disconnect the RH Temp connector from the SHARP MIB board     TEMP RH        4  Remove the two socket head screws from the RH Temp board  assembly from the upper optic chamber endplate     5  Remove the RH Temp gasket from the old RH Temp board assembly  and add it to the new RH Temp board assembly     6  Slide in the new RH Temp assembly by following the previous steps     7  Calibrate the RH Temp sensor  Refer to the    Calibration    chapter in    this manual     7 46 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    SHARP Interface    Board    Replacement    Thermo Fisher Scientific    A    Servicing  SHARP Interface Board Replacement    TEMP RH       Socket head  Screws  2     Figure 7   26  Disconnecting the RH Temp Connector    Use the following procedure to replace the SHARP interface board  Figure  7 27 and Figure 7 28      Equipment Required   SHARP interface board  Philip
380. ic electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    A Equipment Damage Some internal components can be damaged by small    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Disconnect plumbing from the pressure board assembly  Figure 7   10    Note the plumbing connections to facilitate reconnection     7 18 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Servicing  Pressure Board Replacement    3  Disconnect the pressure board from the floor plate on the measurement  interface board     4  Remove the four pressure board assembly retaining screws and remove  the pressure board assembly  Figure 7   9          Unhook Pressure Cable from  Interface Board    Press    Board    Remove Plumbing  Remove Screws  4     Figure 7   9  Replacing the Pressure Board Assembly    5  To install the pressure board assembly  follow the previous steps in  reverse     Model 50307 SHARP Instruction Manual 7 19    Servicing  Pressure Board Calibration    Eevee        PRESSURE  SENSOR BD    VAC FLOW AMB                OPEN FORT OPEN PORT    TO REAR PANEL       DELTA PA    PORT 1 TO REAR PANEL         DELTA PA    SWITCH PORT 1    TO PRIMARY MHEAD  PORT 1    TO PRIMARY MHEAD  PORT 2    Figure 7   10  Replacing the Pressure Board Assembly Plumbing Connection    6  Calibrate the pressure sensor  Refer to the    Pressure Board Calibration     procedure that follows   
381. ice Mode    7 4 Model 50307 SHARP Instruction Manual    The LCD polarizing plate is very fragile  handle it carefully  a    Do not wipe the LCD polarizing plate with a dry cloth  as it may easily  scratch the plate  A    Do not use alcohol  acetone  MEK or other Ketone based or aromatic  solvents to clean the LCD module  but rather use a soft cloth moistened  with a naphtha cleaning solvent  A    Do not place the LCD module near organic solvents or corrosive gases  A    Do not shake or jolt the LCD module  a    2     Pa    2 2    Figure 7   1  Properly Grounded Antistatic Wrist Strap       The firmware can be updated by the user in the field via the serial port or  over the Ethernet  This includes both the main processor firmware and the  firmware in all low level processors  Refer to the   Port manual for the  firmware update procedure     If the Service menu is not displayed on the Main Menu  use the following  procedure to display it     1  From the Main Menu  choose Instrument Controls  gt  Service Mode     The Service Mode screen appears   2  Press to toggle the Service Mode to ON   3  Press        gt        toreturn to the Main Menu     4  Return to the procedure     Thermo Fisher Scientific    Replacement Parts    Thermo Fisher Scientific    Servicing  Replacement Parts List    Table 7 1 lists the replacement parts for the Model 5030  major                                                                            List subassemblies  Refer to Figure 7 2 to identify the
382. ich have influence on the performance of the measuring system     6 2 Equipment  No additional equipment required   6 3 Performance of test    It was tested  whether the instrument software version can be displayed on the instrument   The manufacturer was made aware that any change to the instrument software must be re   ported to the test institute     6 4 Evaluation    The current software version is not displayed on the instrument during switch on  It can   however  be reviewed at any time by selecting the menu    Diagnostics     item    Program Ver   sions       The test started in the year 2009 with the software version V00 05 41 114  The software was  steadily enhanced and optimised during the course of testing     At the beginning of the field test in January 2011  software version V01 00 01 197 was im   plemented  This version was used during the whole field test     The software was updated to version V01 00 03 225 after conclusion of the regular field test  and before starting the remaining laboratory tests  This software version was as well installed  during the additional comparison campaign    Bornheim  summer 2013     In the meantime the software version was set to the current version V02 00 00 232  due toa  discontinuation of a processor board     An overview of software changes can be found in Section 4 1General     The reliability of the AMS was consistently increased with every change  The changes intro   duced up to version V02 00 00 232  are not expected t
383. ield index     The following example reports  the list of analog outputs  index numbers  and variables     Send  list var aout  Receive  list var aout  index variable  O none  1 pm  3 baro  4 vac  6 pflow  8 ambrt  9 srh  11 ambtemp  12 brdtemp  13 ftemp  15 stemp    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  1 0 Configuration    17 frol  19 beta  28 braw  29 araw  30 avgpm  32 mass  34 bzero  45 exfg  50 bref  52 cflg  53 aflg    relay stat   This command reports the current relay logic as normally    open    or  normally    closed     if all the relays are set to same state  that is all open or all  closed  The following example shows that the status of all the relays    logic is  set to normally    open        Send  relay stat  Receive  relay stat open    Note If individual relays have been assigned different logic  then the  response would be a 4 digit hexadecimal string with the least significant  byte  LSB  being relay no 1  A    For example     Receive  relay stat 0x0001   indicates relay no 1 is set to normally open logic  all others are normally    closed     Receive  relay stat 0x0005   indicates relay no 1 and 3 are set to be normally open logic  all others are  normally closed     set relay open   set relay open value   set relay closed   set relay closed value   These commands set the relay logic to normally open or closed for relay  number value  where value is the relay between 1 and 16  The following  example sets the rel
384. ifferent measurement sites with manageable ef   fort  The set up time is approximately 1 hour and the warm up time takes a minimum of ap   prox  12 hours after the measuring system has adjusted to ambient conditions     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land     Air Pollution Control s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 83 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 4 1 5 Instrument design    The instruction manual shall include specifications of the manufacturer regarding the  design of the measuring system  The main elements are    Instrument shape  e g  bench mounting  rack mounting  free mounting    Mounting position  e g  horizontal or vertical mounting    Safety requirements   Dimensions   Weight   Power consumption    6 2 Equipment  A power consumption measuring instrument and a scale were used for this test   6 3 Performance of test    The installation of the delivered instruments was compared with the description given in the  manuals  The specified power consumption was determined for a period of 24 h under nor   mal operating conditions during field test     6 4 Evaluation    The measuring system mus
385. ific for the component suspended particulate matter PM25    Instrument tested  Model 5030  SHARP with PM2 5 pre separator    Manufacturer  Thermo Fisher Scientific  27 Forge Parkway  Franklin  Ma 02038    USA  Test period  June 2009 to September 2013  Date of report  20 September 2013  Report number  936 21209885 F  Editor  Dipl  Ing  Karsten Pletscher    Tel    49 221 806 2592  karsten pletscher de tuv com    Scope of report  Report  173 Pages  Annex Page 174  Manual Page 220  Manual Of 204 Pages  Total 424 Pages    740300_2013_936_21209885F_EN docx    W A    T  V Rheinland Energie und Umwelt GmbH  Li i de land Air Pollution Control  recisely Right     Report on the suitability test of the ambient air quality measuring system   Page 4 of 424 Model 5030  SHARP with PM25 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Blank page    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control 2    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 5 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Content   1  SUMMARY AND PROPOSAL FOR DECLARATION OF SUITABILITY       ssosseseeeeeesn  13  LI SUMMA sssecererajenesdpatenereosiectadiedenssepeteneraee resmiewieconorer
386. ilter Tape  and then Manual  and select    the BENCH to open     From Instrument Panel  select Diagnostics  Instrument Configuration     and change the DETECTOR configuration from BETA to SHARP     Turn instrument power off with the measurement head in the open  position     Model 50307 SHARP Instruction Manual 10 1    Model 5014  to 50301 Upgrade Procedures  SHARP Connection Bracket Assembly Installation    6  Unplug instrument from AC power source        SHARP Connection Use the following procedure for SHARP connection bracket assembly  Bracket Assembly installation   Installation      Locate the Connection Bracket  Figure 10 1  and remove two screws  and set aside         4 pin SHARP   Connector             2 pin connector  not used                Figure 10   1  SHARP Optics Connection Bracket    2  Remove both sides of the Cover  Remove the tape pick up spool   Figure 10 2  by removing the thumb screw  washer  filter tape   spacers  o rings and rear filter support guide plate     10 2 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Model 5014i to 50301 Upgrade Procedures  SHARP Connection Bracket Assembly Installation       STEPS            1   UNFASTEN THUMB NUT   2   SLIDE TAPE PICKUP SPOOL OFF   3   SLIDE TAPE ASSEMBLY OUT TO GAIN ACCESS TO SCREW AREA  4   4   FASTEN 2X SCREWS TO CONNECTION BRACKET THRU WALL PLATE  5   MAKE SURE CONNECTOR ORIENTATION CLOSER TO THIS EDGE             Figure 10   2  Tape Pick up Spool Removal    3  Remove the two Philips 
387. imes of the availability tests were defined by the start and end of the field  tests at each test site  All measurement interruptions  e g  due to system failures or mainte   nance works  were taken into account     6 4 Evaluation    Table 16 and Table 17 show a summary of the operation  maintenance and malfunction  times  The measuring systems were operated over a period of 313 measurement days  see  annex 5   This period includes a total of 22 days of zero filter operation  audits and days   which had to be rejected due to the change to zero filter  see annex 5      Outages caused by external influences  which cannot be attributed to the devices them   selves  were recorded on 10 March 2011 and 14 March 2011  power outage   Since the AMS  needs some hours for stabilising  one day following the outages had to be discarded as well   Furthermore  SN 1 was switched off inadvertently on 26 April 2011  Therefore  both the 26  and 27 April 2011 could not be evaluated  An unintended switch off of SN 1 was also record   ed on 3 May 2011  This time  the system was down for a very short time  so that the data of  the following day did not have to be discarded  On this account  the total operating time is  reduced to 306 days for SN 1 and 309 days for SN 2     The systems presented the following malfunctions   SN 1     A malfunction of SN 1 was observed on 8 August 2012 at the test site Teddington  summer    It was  however  not noticed directly and the staff on site could not identify
388. ine which   when pressed  pops up an input dialog which will state  Please enter a new  value for NO using a d ddd format     The string entered by the user is used  to construct the output command  If the user enters  for example   1 234    the constructed command will be  set no coef 1 234      Model 50307 SHARP Instruction Manual B 69    C Link Protocol Commands  Record Layout Definition    B 70 Model 50307 SHARP Instruction Manual      NO 21f Code_0 Code 1 Code 2 Code 3 Code 4 Code 5 Code 6  Code_7 Code _8 Code_9 Code 10 Code 11 Lset range no  d n     This is a line which appears slightly indented  the title is again  NO   and  the value is the twenty first element of the data response  interpreted as a  floating point number  There is a no translation button which creates a  selection list of twelve  Code nn  options  The number of the user  selection is used to create the output command      Mode 6 12 13x local remote service service  0 1 Tset mode   s n     This is a line which has a title of  Mode  and value taken from the sixth  field of the data response  There is a bitfield extraction of bits 12 through  13 from the source  the value type is not important here because the value  is being translated to an output string   Once the bits have been extracted   they are shifted down to the bit zero position  Thus  the possible values of  this example will be 0 through 3  The translation list shows the words  which correspond to each input value  the zero index value appe
389. ing  range of 0     200 or 0     1000 g m  is recommended      Recommended  measuring range  0     200 or 0     1000 ug m8  Upper limit of the certification range  PM25  50 g m8  6 5 Assessment    A measuring range of 0     10 000 g m  is set by default  It is possible to adjust other meas   uring ranges     The upper limit of the measuring range of the measuring system is greater than the respec   tive upper limit of the certification range     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 90 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 2 3 Negative output signals  Negative output signals or measured values may not be suppressed  life zero    6 2 Equipment  No additional equipment required   6 3 Performance of test    The AMS was tested on its ability to display negative output signals in the laboratory and in  the field     6 4 Evaluation  The AMS can display negative values both on the front panel and over data outputs   6 5 Assessment    Negative measured signals are displayed directly on the AMS and are transmitted correctly  via the respectiv
390. instrument is set to power up in the local remote mode  the keypad  can be used to make changes via the front panel  If the instrument is set to  power up in the remote locked mode  changes can not be made from the  front panel  The following example sets the instrument to power up in  remote locked mode     Send  set power up mode 1  Receive  set power up mode 1 ok    Table B   10  Power Up Mode Values       Value Power up Mode  0 Local Unlocked Mode  default   1 Remote Locked Mode    program no  This command reports the analyzer   s model information and program  version number     Send  program no  Receive  program no iSeries 50301 00 05 37 093  tz    This command reports the    tz    timezone string for the NTP server  See     Network Time Protocol Server    in the    Communications Settings     section of the    Operation    chapter for more information     Send  tz  Receive  tz EST 5EDT    set tz string   This command sets the timezone string for the instrument for use with the  NTP server  where string is a standard timezone string  Common strings are  listed in the    Timezone    screen description in Chapter 3     Send  set tz EST 5EDT  Receive  set tz EST 5EDT ok    Thermo Fisher Scientific    1 0 Configuration    Thermo Fisher Scientific    C Link Protocol Commands  1 0 Configuration    analog iout range channel   This command reports the analog current output range setting for channel   where channel must be between 1 and 6  inclusive  The following example  repor
391. ion  gt   Ambient RH     3 86 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu       aa pe Phin Py Tit  Rri Taic mF i i        Pressure Vacuum The Pressure Vacuum Calibration submenu allows the user to view and  Calibration calibrate the pressure sensor and vacuum flow sensor  The pressure vacuum  calibration submenu is visible only when the instrument is in service mode   For more information on the service mode  see    Service Mode    earlier in  this chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration        Barometer Pressure The Barometer Pressure Calibration submenu is used to calibrate the  Calibration barometer pressure offset  span or reset default values     e Inthe Main Menu  choose Service  gt  Pres  Vacuum Calibration  gt   Baro Pres           rt   fool  emj  what          Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 87    Operation  Service Menu    Calibrate Barometer The Calibrate Barometer Pressure Offset screen allows the user to calibrate  Pressure Offset the pressure sensor at offset pressure     Note The operator should use an independent barometer to measure the  ambient pressure and enter the value on this screen before calibrating  a    e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration  gt   Baro Pres Calibration  gt  Offset        Calibrate Barometer The Calibrate Baromet
392. ion Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F       Figure 10     Model 50301 SHARP measuring system     Optics module       Figure 11     740300_2013_936_21209885F_EN docx    Model 50301 SHARP measuring system     Beta module    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland 9    Air Pollution Control g    Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher   Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    Page 41 of 424       Figure 12  Model 5030i SHARP measuring system  Systems 1 and 2 from right  at measur   ing station       Figure 13  Vacuum pump  type  GAST Model 75R647  PN 110836 00     740300_2013_936_21209885F_EN docx    A TUVRheinland   TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system  Page 42 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F  The measuring system is operated through pushbuttons and an LCD display on the front side    of the instrument  Here
393. ir Pollut    Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 96 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    6 5 Assessment    The availability was 97 4   for SN 1 and 96 1   for SN 2 without test related outages  or  90 2   for SN 1 and 89 0   for SN 2 including test related outages     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Table 16  Determination of availability  without test related outages     Sen Sime NT  operano me iP a ft  owein ipa fe    Maintenance Jo 0       Table 17  Determination of availability  including test related outages   R N a  Pa time 309    Maintenance incl  zero fil   ter operation       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 97 of 424  Scientific for the component suspended particulate matter PMz5    Report No   936 21209885 F    6 1 5 2 9 Instrument software    The version of the instrument software to be tested shall be displayed during switch on  of the measuring system  The test institute shall be informed on changes in the instru   ment software  wh
394. ir Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 86 of 424 Model 5030  SHARP with PM25 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F  6 6 Detailed presentation of test results    Figure 33 shows a view of the rear side of the instrument with the respective measured value  outputs        Figure 33     Rear view of Model 5030i SHARP    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH      T  VRheinland      Air Pollution Control  gt  s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 87 of 424   Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F    6 1 5 1 General    The manufacturer s specifications in the instruction manual shall be by no means better  than the results of the performance test     6 2 Equipment   No additional equipment required    6 3 Performance of test   The test results were compared with the specifications of the manual    6 4 Evaluation   Deviations found on the first draft of the manual from the actual design were cleared   6 5 Assessment    No differences between the instrument design and the descriptions given in the manual were  detected     Minimum requirement fulfilled  yes  6 6 Detailed presenta
395. ired  a re calibration of the radiometric measurement using the foil set    The cleaning of the sampling tube shall be paid attention to during the annual basic  maintenance     Further details are provided in the user manual     Department of Environmental Protection Air Pollution Control       Pr aie Kuido Tog  V4  Dipl  Ing Karsten Pletscher Dipl   Ing Guido Baum    Cologne  20 September 2013  936 21209885 F    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 173 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    8  Literature     1  Standard VDI 4202  Part 1     Performance criteria for performance tests of automated  ambient air measuring systems     Point related measurement methods for gaseous and  particulate air pollutants     June 2002  amp  September 2010    2  Standard VDI 4203  Part 3     Testing of automated measuring systems     Test proce   dures for point related ambient air measuring systems for gaseous and particulate air  pollutants     August 2004  amp  September 2010    3  Standard EN 14907     Ambient air quality     Standard gravimetric measurement method  for the determination of the PM25 mass fraction of suspended particulate matter     Ger   man ver
396. is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms  gt   Vacuum        ERED    EN EEE TE Sanat Beee          Min and Max Vacuum The Minimum Vacuum alarm limit screen is used to change the minimum  Limits vacuum alarm limit  The minimum and maximum vacuum screens  function the same way     Note The maximum vacuum value should not be changes  as this is also  used as a protective feature for the instrument  A    e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms  gt   Vacuum  gt  Min or Max        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 77    Operation  Alarms Menu    Flow The Flow screen displays the current flow reading and sets the minimum  and maximum alarm limits  Acceptable alarm limits range from  5 mmHg  to 40 mmHg  If the flow reading goes beyond the minimum or maximum  limit  an alarm is activated and the alarm  bell  icon appears in the status  bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms  gt   Flow        ip priori  PIL          Min and Max Flow Limits The Minimum Flow alarm limit screen is used to change the minimum  flow alarm limit  The minimum and maximum flow screens function the  same way     e Inthe Main Menu  choose Alarms  gt  Pressure Vacuum Alarms  gt   Flow  gt  Min or Max        ip priri  Piet          Flow Alarms The Flow Alarms submenu all
397. is command sets the nephelometer into an auto zero calibration routine   The following example set this command     Send  set neph cal bkg  Receive  set neph cal bkg ok  neph coef    This command reports the nephelometer span coefficient set at the factory   The following example reports that the nephelometer span coefficient is    388 000     Send  neph coef  Receive  neph coef 388 000    set neph coef value   This command sets the nephelometer span coefficient to user defined  values to value  where value is a floating point representation of the  coefficient  The following example sets the nephelometer span coefficient  to 250 00    Send  set neph coef 250  Receive  set neph coef 250 ok  neph reset bkg    This command will reset all measured background values to zero  The  following example sets this command     Send  set neph reset bkg  Receive  set neph reset bkg ok    Model 50307 SHARP Instruction Manual B 41    C Link Protocol Commands  Calibration    B 42 Model 50307 SHARP Instruction Manual    neph restore bkg  This command will restore the background values to the previous  background values  The following example sets this command     Send  set neph restore bkg  Receive  set neph restore bkg ok  neph rh offset    This command reports the nephelometer relative humidity sensor offset set  at the factory  The following example reports that the nephelometer relative  humidity sensor offset is 7 3      Send  neph rh offset  Receive  neph rh offset 7 3      set neph rh o
398. is report   6 2 Equipment   Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 10 of this report   6 3 Performance of test   Not applicable to PM25 sampling systems  Reference is made to module 5 4 10 of this report   6 4 Evaluation   Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 10 of this report   6 5 Assessment    Not applicable to PM25 sampling systems  Reference is made to module 5 4 10 of this report   Minimum requirement fulfilled            6 6 Detailed presentation of test results    Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 10 of this report     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     TUVRheinland      Air Pollution Control      Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 131 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 4 3 Reproducibility of the sampling systems    The PM  sampling systems of two identical systems under test shall be reproducible  among themselves according to EN 12341  T5   This shall be demonstrated in the field  test     Note    Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 9 of this report   6 2 Equipment   Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 9 of this report   6 3 Pe
399. is shown as the byte value 0xCD  or binary  1100 1101  Output 10 is the MSB of this byte  and output 2 is the LSB   By convention  bits within a byte are shown with the MSB to the left and  the LSB to the right  Thus  the outputs in the first byte are    10 through 2      from left to right  In the last data byte  the status of outputs 15 11 is shown  as the byte value 0x0A  or binary 0000 1010  Output 15 is in the fifth bit  position from the left  and output 11 is the LSB of this byte  The four  remaining high order bits are zero filled     Read Holding Input Registers reads the measurement data from the  instrument  Issuing either of these function codes will generate the same  response  These functions read the contents of one or more contiguous  registers     These registers are 16 bits each and are organized as shown below  All of  the values are reported as 32 bit IEEE standard 754 floating point format   This uses 2 sequential registers  least significant 16 bits first     The request specifies the starting register address and the number of  registers  Registers are addressed starting at zero  Therefore  registers  numbered 1   16 are addressed as 0 15  The register data in the response  message are packed as two bytes per register  with the binary contents right  justified within each byte  For each register  the first byte contains the high  order bits and the second contains the low order bits     Model 50307 SHARP Instruction Manual C 5    MODBUS Protocol  Functi
400. is the  pressure differential across the orifice     e Inthe Main Menu  choose Diagnostics  gt  Pressure Vacuum        Flow The Flow screen  read only  displays the current sample flow reading  The  flow is measured by the internal pressure board sensors     e Inthe Main Menu  choose Diagnostics  gt  Flow     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 61    Operation  Diagnostics Menu       tOo Miker i          Detector Status The Detector Status screen  read only  displays the current sample mass  reading  alpha counts  and beta counts     Note A mass value will only be calculated when the flowrate is within  acceptable limits  However  when in Service Mode  a mass value will be  calculated if the pump is turned OFF  A    e Inthe Main Menu  choose Diagnostics  gt  Detector Status        PU ba EE EE       Nephelometer The Nephelometer Status screen  read only  displays the current  Status  2ephelometer reading     e Inthe Main Menu  choose Diagnostics  gt  Nephelometer Status        OPER EEE          CRn The CRn screen  read only  displays the current CRn reading   e Inthe Main Menu  choose Diagnostics  gt  CRn     3 62 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Analog Input  Readings    Analog Input  Voltages    Digital Inputs    Thermo Fisher Scientific    Operation  Diagnostics Menu          The Analog Input Readings screen  read only  displays the 8 user scaled  analog readings  if the I O expansion board option is installed     
401. is to collocate the inlet at the same height as other similar   use inlets  The    Installation    chapter discusses specific siting criteria  In  addition to using the optional inlets  extending the sample downtube  length is necessary using a rigid sample tube     The I O expansion board provides six analog current output channels  0   20 mA or 4 20 mA  and eight analog voltage inputs  0 10 V   The DB25  connector on the rear panel provides the interface for these inputs and  outputs     The 25 pin terminal board assembly is included with the optional I O  expansion board  Refer to    Terminal Board PCB Assemblies    in the     Installation    chapter for information on attaching the cable to the    Model 50302 SHARP Instruction Manual 9 1    Optional Equipment  Terminal Block and Cable Kits    Terminal Block and  Cable Kits    Cables    9 2 Model 50307 SHARP Instruction Manual    connector board  For associated part numbers  refer to the    Servicing     chapter     The optional terminal block and cable kits provide a convenient way to  connect devices to the instrument  These kits break out the signals on the  rear panel connector to individual numbered terminals     Two types of terminal block and cable kits are available  One kit is for the  DB37 connectors and can be used for either the analog output connector  or the relay output connector  The other kit is for the DB25 connector and  can be used for the optional I O expansion board  For associated part  numbers  refe
402. isher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F          100       90 4 a SN1 Bornheim Winter     f             80 4 A  70 4 af  60 4 A    50 4 F    40 4 Z    PM  5 5030i Sharp   pg m         a  s    20   K             0 10 20 30 40 50 60 70 80 90 100  PM    Reference   yg m            Figure 55  Reference vs  candidate  SN 1  measured component PM25  Bornheim  winter              90 4 a SN2 Bornheim Winter     A             80 4 f  70 4 u   60 4 f    50 4 ri    PM  5 5030i Sharp   ug m    S    40   af  30   AL    20 5             0 10 20 30 40 50 60 70 80 90 100  PM    Reference   yg m             Figure 56  Reference vs  candidate  SN 2  measured component PM25  Bornheim  winter     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 163 of 424       100          90   a SN1 Teddington Summer                        80         70 4 s    60 4      50   of    PM    5030i Sharp   yg m     40 4 Af  30   fp    a  20 5 rare             0 10 20 30 40 50 60 70 80 90 100  PM   Reference   yg m             Figure 57  Reference vs  candidate  SN 1  measured component PM25  Teddington  sum     mer        
403. ital inputs in the instrument  which triggers the respective action     This function code is used to set a single action to either ON or OFF  The  request specifies the address of the action to be forced  Actions are  addressed starting at zero  Therefore  action number 1 is addressed as 0   The requested ON OFF state is specified by a constant in the request data  field  A value of 0xFF00 requests the action to be ON  A value of 0x0000  requests it to be OFF  All other values are illegal and will not affect the  output  The normal response is an echo of the request  returned after the  state has been written     Note This function will not work if the instrument is in service mode  A    Request   Function Code 1 Byte 0x05   Starting Address 2 Bytes 0x0000 to maximum allowed by instrument  Output Value 2 Bytes 0x0000 or OxFFOO   Response   Function Code 1 Byte 0x05   Starting Address 2 Bytes 0x0000 to maximum allowed by instrument  Output Value 2 Bytes 0x0000 or OxFFOO   Error Response   Function Code 1 Byte Function code   0x80   Exception Code 1 Byte 01 IIlegal Function  02 IIlegal Address     03 Illegal Data  04 Slave Device Failure    Model 50307 SHARP Instruction Manual C 7    MODBUS Protocol    MODBUS Parameters Supported    MODBUS Parameters    Supported    C 8 Model 50307 SHARP Instruction Manual    Here is an example of a request to write Coil 5 ON     Request   Field Name  Function   Output Address Hi  Output Address Lo  Output Value Hi  Output Value Lo    Response
404. ith PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 102 of 424    6 5 Assessment    The investigations result in detection limits of 0 11 g m3 for system 1  SN 1  and 0 07 ug m8  for system 2  SN 2      Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Table 18  Detection limit PM25                         System SN 1 System SN 2  Number of values n 15 15  Mean of the zero values x  pg m  0 02 0 03  Standard deviation of values s   g m  0 05 0 03  Students Factor tn 1 0 95 2 14 2 14  Detection limit X g m8 0 11 0 07             Single values from the determination of the detection limit can be looked up in annex 1 of the  appendix     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 103 of 424  Scientific for the component suspended particulate matter PMs 5    Report No   936 21209885 F    6 1 5 3 3 Repeatability standard deviation at reference point    The repeatability standard deviation at reference point shall not exceed the require   ments of Table 2 in the certification range according to Table 1 of Standard VDI 4202  Part 1  September 2010   The limit value or the alert threshold shall be use
405. ith commonly available tools     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 78 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 5 Assessment    Maintenance works can be carried out with commonly available tools taking reasonable time  and effort  The semi annual filter tape change and the works described in points 7ff are to be  performed while the system is on standstill only  These works come up every six or twelve  months  In the meantime  maintenance works are limited to the check of contaminations   plausibility checks and possible status error messages     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results    The maintenance works were carried out during the test in accordance with the instructions  given in the manual  No problems were noticed while following the described procedures  All  maintenance works could be performed smoothly with customary tools     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei nland     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the 
406. itions   sample conditions  and mass sensor data  The status bar displays the time   24 hour format   the password  lock  icon  service  wrench  icon  alarm   bell  icon  and optional zero span sample solenoid valve status  if installed     Status Bar Icons    The password  lock  icon indicates that    no parameter changes can be made from        gt  B    4  E The alarm  bell ion indicates    the front panel    that an alarm is active     The service  wrench  icon indicates that the  instrument is in the service mode     3 6 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Firmware Overview    The Run screen normally displays the current SHARP concentrations  but  also has additional Run screens for ambient conditions  sample conditions     and mass sensor data  Pressing      and __  _  will scroll through the    different Run screens        PU ER ERR ET eh T hE  RAGES EEE f PRETE EE LEY    Status Bar                  Concentration   lt     Mass Units    Status Bar           ee Sa 6 J5   lt  Status Bar Icons       Ambient Conditions Run The Ambient Conditions Run screen displays the real time volumetric flow  Screen rate at the inlet  the ambient temperature  the barometric pressure  and the  ambient relative humidity     Status Bar                  Status Bar                   Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 7    Operation  Firmware Overview    Sample Conditions Run The Ambient Conditions Run screen displays the samp
407. its    sesssesesesesesesseseereeerersrreresesse 3 72  Sample RA a ri yea or is a a E en 3 73  Min and Max Sample RH Limits    ssesesesesesesesesssssererereeeesrrerrseseses 3 73  Ambient Temperat  re makonen e o at 3 73  Min and Max Ambient Temperature Limits   sssssesseseseseeeeeeeeeee 3 74  Flow Temperature  norinni aises ieai iaaii aa 3 74  Min and Max Flow Temperature Limits    n    3 75  Board  Temperaturer aregon a a A RAS 3 75  Min and Max Board Temperature Limits    3 75  Pressure Vacuum Alarms    ssesessesssssssesrsssesssesssesssisesrsesrseseerersesreres 3 76  Barometric Pressure  coiilascsisagicvsetiveds tussnweiesavbarssanahauesadeasegedanereer 3 76  Min and Max Barometric Pressure Limits 0 0    seeeeseeseneeeeeeees 3 76  Vacuum suerte tceuteciatee a a a n a ete teense 3 77  Min and Max Vacuum Limits    3 77  FO a a ad alae Oa sea 3 78  Min and Max Flow Wirinttss 5 ahs ata cat ether  3 78  Flow Lae Stace aistnde tela cacared doe cde ccbei ates Rel acne ded noes tec ee 3 78  FloW iii asic Gagner tate 3 79  Min and Max Flow Limits s isce dive coeds anil hier tndedeect AudeieSeaee 3 79  Concentration Alaris  223  ses Padceohe  Nace caedeaen eee  seca Manger etice lan ceaer a  3 79  Ayerage PMen siie n ea a E a a 3 80  Min and Max Average PM Concentration Limits   0 0 0    3 80  Ayerage SHARP nsise e A caneauntawtunuetdes 3 81  Min and Max Average SHARP Concentration Limits                 3 81  Average Nephelometer     ssssseseeesesessesesesrsesesesrsrsrsrsrerrrerrerer
408. ity test of the ambient air quality measuring system   Page 62 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Parameterisation of the relative humidity threshold for the Dynamic Heating System DHS    The Model 5030i SHARP measuring system uses a heated sampling system in order to pre   vent the effects of condensation in the system  especially in the nephelometer and filter  tape   The heater must also be designed in such a way that the potential risk of volatile parti   cle loss is minimised  The Dynamic Heating System DHS implemented in the Model 5030i  SHARP measuring system ensures that the relative humidity of the air passing through the  filter tape at the radiometric stage is kept well below the point at which the collected particles  adhere together  thus producing disrupting condensation effects  Furthermore  the relative  humidity in the nephelometer is adjusted below the specified limit  The system minimises the  increase of internal temperature  This ensures a negligible loss of semi volatile particle com   ponents  provided that the relative ambient humidity is maintained below the limit to which  the heater is adjusted  As that the relative ambient humidity rises above the threshold  the  applied heating is optimised to maintain the relative humidity threshold above the filter tape  of the radiometric stage  The DH System is adjusted
409. l min     2 5   of the flow rate  between the flow  rates measured by the instrument and by the external flow meter under application of the  adapter     proved to be appropriate parameters for monitoring the instrument   s tightness  The  laboratory tests resulted in a maximum leakage  lt  1   of the nominal flow rate of 16 67 I min  for both candidates    The tightness check must only be performed with the tightness check adapter and the de   scribed procedure  It can otherwise damage the instrument     Minimum requirement fulfilled  yes   6 6 Detailed presentation of test results   Table 33 contains the values determined during the tightness check   Table 33  Results of the tightness check    System   s Pump vacu  System   s Pump vacu  Ext  flow me    Difference of Related to  measured um  without measured um  with ter   s measured flow rates  nominal flow  value  with  adapter value  with adapter value  with with adapter rate  out adapter adapter adapter       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control      Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 143 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 Methodology of the equivalence check  modules 5 4 9     5 4 11   According to the version of the Guide from J
410. lated for the uncertainty between systems under test Ups   The graphical representation is displayed in Figure 41 to Figure 47     Table 34  Uncertainty between systems under test Ups for candidates SN 1 and SN 2   measured component PM25    Candidate Test site No  of Uncertainty Ups  values    SN  SN 1 SN2 All test sites 242    Single test sites    52  91  49  50    mer   82  winter   a      Classification over reference value       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVR heinland    Air Pollution Control   x  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 147 of 424  Scientific for the component suspended particulate matter PM25    Report No   936 21209885 F       Model 5030i SHARP  SN 1   SN 2  All test sites  Raw data                  Measured values         Regression line                     y x                Candidate 2  yg m                                                         0 10 20 30 40 50 60 70 80 90 100 110 120 130    Candidate 1  ug m               Figure 41  Results of the parallel measurements with the candidates SN 1   SN 2  measured  component PMzs  all test sites       Model 5030i SHARP  SN 1   SN 2  Cologne  Winter  Raw data                o Measured values           Regression line                  y x                   Candidate 2  ug m                                                
411. le 2   4  D O Terminal Board Pin Descriptions    Detail    B                                        Screw Signal Description Screw Signal Description  1 Relay1_ContactA 13 Relay7_ContactA  2 Relay1_ContactB 14 Relay7_ContactB  3 Relay2_ContactA 15 Relay8_ContactA  4 Relay2_ContactB 16 Relay8_ContactB  5 Relay3_ContactA 17 Relay9_ContactA  6 Relay3_ContactB 18 Relay9_ContactB  7 Relay4_ContactA 19 Relay10_ContactA  8 Relay4_ContactB 20 Relay10_ContactB  9 Relay5_ContactA 21  not used    10 Relay5_ContactB 22  24V   11 Relay6_ContactA 23  not used    12 Relay6_ContactB 24  24V    25 Pin Terminal Board    2 20 Model 50307 SHARP Instruction Manual       The 25 pin terminal board is included with the optional I O Expansion  Board  Figure 2   11 shows the recommended method for attaching the  cable  customer supplied  to the terminal board using the included tie     Thermo Fisher Scientific    Thermo Fisher Scientific    Installation  Connecting External Devices    down and spacer  Table 2   5 identifies the connector pins and associated    signals     See Detail    A         See Detail    B          Assembled Connector    Figure 2 11  25 Pin Terminal Board Views       Detail    A       Table 2 5  25 Pin Terminal Board Pin Descriptions                                     Screw Signal Description Screw Signal Description  1 lOut1 13 Analog_In1  2 Isolated ground 14 Analog_In2  3 lOut2 15 Analog_In3  4 Isolated ground 16 Ground   5 lOut3 17 Analog_In4  6 Isolated ground 18 Analog_In5
412. le filter  Screen temperature  the sample filter relative humidity  the differential pressure  across the orifice  and the vacuum under the filter tape     Title Bar                   Status Bar              lt       Status Bar Icons    m g       Mass Sensor Data Run The Mass Sensor Data Run screen displays the real time beta counts during  Screen attenuation and the initial beta counts from a fresh filter spot     Title Bar                     Status Bar                agp       3 8 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Firmware Overview    Hybrid Data Run Screen   The Hybrid Data Run screen displays the mass concentration values from  the rea time beta attenuation method  PM  and the real time nephelometer  method  NEPH   These values are entered into the dynamic digital  filtering algorithm to calibrate the nephelometer in real time    Title Bar    Thermo Fisher Scientific    Status Bar                Main Menu    rip gui pi    PME ba EE NEE       The Main Menu contains a number of submenus  Instrument parameters  and settings can be read and modified within the submenus according to  their function  The concentration appears above the Main Menu and  submenus in every screen  The Service menu is visible only when the  instrument is in service mode  For more information on the service mode   see    Service Mode    later in this chapter        e Use  t Jand     to move the cursor up and down     e Press to make a selection     e Press   
413. le time of day to  perform a mass calibration would be either early morning or late   afternoon  A    The preferred method of mass coefficient calibration is the Auto Mode     e Inthe Main Menu  choose Service  gt  Mass Calibration  gt  Mass  Coefficient        ERED CH  PT it    H WH mht  PSPS Py AG Lf bi       The Auto Mass Coefficient screen allows the user to conduct a mass  calibration of the measurement head and will walk you through the  process     Use the following procedure to conduct a mass calibration   1  In the Main Menu  choose Service  gt  Mass Calibration  gt  Mass  gt  Auto     2  Enter the SPAN foil value  e g   1 328 pg  using the keypad and press  to save the foil value and to open the bench  The beta    attenuation chamber should now open        WH nied       Model 50307 SHARP Instruction Manual 4 9    Calibration  Calibration Procedure    3  Cut break the filter tape and remove from the bench  Then insert the  Foil Holder from the calibration kit and press while maintaining  a slight sideward pressure on the filter holder        4  Insert the Null Zero foil with the label side up and press to  proceed with the foil calibration  or press to stop the foil            Wi mh ivd    DH mii   Pinar Ai Lafin rri          The auto zero procedure of mass will begin and last approximately 270  seconds  When the zero is complete  the beta attenuation measurement  will remain closed and an average Beta count rate is displayed        5  Remove the zero null foil and i
414. librat     ed using the values obtained for slope and or intercept of all paired instruments togeth   er     The fulfilment of the 5 criteria is checked in the following chapters     In chapter 6 1 5 4 9 Determination of uncertainty between systems under test Up  criteria 1  and 2 will be checked    In chapter 6 1 5 4 10 Calculation of the expanded uncertainty of the instruments criteria  3  4 and 5 will be checked    In chapter 6 1 5 4 11 Application of correction factors and terms  there is an exemplary    evaluation for the case  that criterion 5 cannot be fulfilled without the application of correction  factors or terms     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland  oe ait Pollut    Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 144 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    6 1 5 4 9 Determination of uncertainty between systems under test ups    For the test of PM25 measuring systems the uncertainty between the systems under  test shall be determined according to chapter 9 5 3 1 of the guidance document    Demonstration of Equivalence of Ambient Air Monitoring Methods    in the field test at  least at four sampling sites representative of the future application     6 2 Equipment   No additional equipment requir
415. lips screwdriver    f Equipment Damage Some internal components can be damaged by small    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF and unplug the power cord    2  Ifthe instrument is mounted on a rack  remove it from the rack   3  Unfasten the two screws    4  Lift cover in an upward direction     5  Replace the SHARP optics cover by following the previous steps in    reverse     7 44 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Servicing  External Solenoid Valve Replacement    External     Use the following procedure to replace the external solenoid valve  Figure    Solenoid Valve            Replacement    ment Reauiret    External solenoid valve    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A       Equipment Damage Some internal components can be damaged by small  1  Turn the instrument OFF  unplug the power cord     2  Disonnect all vacuum tubing from the solenoid     3  Disconnect the solenoid electrical connector from the rear panel  connector     4  Unfasten the two captive hardware screws from the rear panel     5  Replace the solenoid valve by following the previous steps in reverse     To Outdoor Roof Flange         Captive  Hardware  2     To Pump  Exhaust    Connector    Figure 7 25  Replacing the External Solenoid Valve    Thermo Fisher Scientif
416. lity Assurance  Handbook for Air Pollution Measurement Systems  published by the U S   EPA  Research Triangle Park  NC 27711  can be consulted for detailed    quality assurance guidelines     Optional mass transfer standards are available from Thermo Fisher  Scientific for performing an annual mass calibration  However  additional  specialized equipment is needed to calibrate the Model 5030  for  temperature  relative humidity  barometric pressure  and volumetric flow   The following equipment should be used     e A thermistor or thermocouple thermometer capable of measuring  ambient temperatures in a range of    20 to 50   C  readable to the    Model 50307 SHARP Instruction Manual 4 1    Calibration  Pre Calibration    Pre Calibration    Calibration  Procedure    4 2 Model 50307 SHARP Instruction Manual    nearest 0 1   C  This thermometer should be referenced to within an  accuracy of  0 5   C to NIST traceable precision thermometers   Multiple thermometers may be used to cover the temperature range as  long as each thermometer meets the accuracy and readability  specifications described above     e A barometer capable of measuring barometric pressure over a range of  600 to 800 mmHg  80 to 106 kilo Pascals  kPa   and readable to the  nearest 1 mmHg  At least once a year  this barometer should be  calibrated to within  5 mmHg of a NIST traceable barometer of  known accuracy     e Flow rate Transfer Standard  FTS  measurement equipment capable of  calibrating  or verifying  
417. lity measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    filled    Page 26 of 424    5 4 7 Constancy of  sample volumet   ric flow    Tightness of the  measuring sys   tem    Determination of  uncertainty be   tween systems  under test ubs    5 4 10 Calculation of  the expanded  uncertainty of the  instruments      3   of the rated value  All in   stantaneous values of the  sample volumetric flow shall be  within a range of   5   of the  rated value during sampling     Leakage shall not exceed 1    of the sample volume sucked     Shall be determined according  to chapter 9 5 2 1 of the guid   ance document    Demonstration  of Equivalence of Ambient Air  Monitoring Methods    in the field  test for at least two identical  systems     Determination of the expanded  uncertainty of the candidates  according to chapter 9 5 2 2ff  of the guidance document    Demonstration of Equivalence  of Ambient Air Monitoring  Methods        740300_2013_936_21209885F_EN docx    All determined daily averages present  deviations less than  3   from the  nominal value  all instantaneous val   ues deviate less than  5       The criteria specified by the manufac   turer upon which the tightness check  is considered passed     maximum dif   ference of  0 42 I min     2 5   of  the flow rate  between the flow rates  measured by the instrument and by  the external flow
418. llocated for lrecs and  sreco to value  where value is a floating point number representing percent   The following example sets the memory allocation for lrecs to 10      Note Issuing these commands will clear all the logged data in memory  All  the existing records should be retrieved using appropriate commands  if    required  A   Send  set malloc lrec 10  Receive  set malloc lrec 10 ok  no of lrec   no of srec    These commands report the number of Irec and srecs stored memory  The  following example shows that 50 Irecs have been stored in the memory     Send  no of lrec  Receive  no of lrec 50 recs    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Datalogging    sp field number  This command reports the variable number and name stored at the index in  the scratch pad list     The scratch pad is a temporary memory area which is used to set up lists of  selections for lrec  srec  or streaming data items  The user can copy any of  these lists to the scratch pad  modify individual elements in the list  then  save the scratch pad back to the original list     The following example shows that field 1 in the scratch pad is set to index  number 3  which is for the variable PM concentration pressure     Send  sp field 1  Receive  sp field 1 3 PM    set sp field number value  number   1 32 for lrec and srec lists  1 8 for streaming data lists    This command sets the scratch pad field number  item number in scratch  pad list  to value  where valu
419. llowing installation procedures for the Model 5030  describes  packaging  lifting the instrument  unpacking the instrument  setup and  installation  including the sample tube and pump  and attaching the analog  and or digital outputs to a recording device  The installation should always  be followed by instrument calibration as described in the    Calibration     chapter of this manual     This chapter provides the following recommendations and procedures for  installing the instrument     e    Packaging and Transport    on page 2 1   e    Lifting    on page 2 2   e    Unpacking and Inspection    on page 2 2   e    Acceptance Testing and Startup Procedures    on page 2 3  e    Setup    on page 2 10   e    Connecting External Devices    on page 2 17   e    Startup    on page 2 22   e    Data Content    on page 2 23   e    Shutdown    on page 2 23   e    Important Tips    on page 2 23    Packaging and The Model 5030 instrument  power cord  and operator manual are  Transp ort shipped in an ISTA 2A certified packaging and all other items accessories  are shipped separately  The ISTA 2A certified packaging is comprised of  the instrument within an inner box which is secured within an outer box  using foam end caps and corner bracing  In this configuration  the  packaging is ready for shipping by carrier domestically and internationally     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 2 1    Installation    Lifting    Lifting    Unpacking and  Inspection    2 2 Mod
420. log3 17 TTL_Input2   6 Analog ground 18 TTL_Input3   7 Analog4 19 TTL_Input4   8 Analog ground 20 Digital ground   9 Analog5 21 TTL_Input5   10 Analog ground 22 TTL_Input6   11 Analog6 23 TTL_Input7   12 Analog ground 24 Digital ground       Thermo Fisher Scientific    D O Terminal Board    Thermo Fisher Scientific    Installation  Connecting External Devices    The analog outputs are arranged on the rear panel connector as shown in  Figure 2 9  See Table 2   3 for channels and pin connections                O        0000000000000000 00  oO000000000000000000                      Figure 2 9  Pin Out of Rear Panel Connector  Table 2 3  Default Analog Outputs                   Channel Connector Pin 1 0 Terminal Pin Description  1 14 1 PM   2 33 3 AVG PM   3 15 5 NEPH   4 34 7 AVG NEPH   5 17 9 SHARP   6 36 11 AVG SHARP  Ground 16  18  19  35  37 2  4  6  8  10  12 Signal Ground       Note All channels are user definable  If any customization has been made  to the analog output configuration  the default selections may not apply  4    Figure 2   10 shows the recommended method for attaching the cable   customer supplied  to the terminal board using the included tie down and  spacer  Table 2   4 identifies the connector pins and associated signals     Model 50307 SHARP Instruction Manual 2 19    Installation  Connecting External Devices    See Detail    A         See Detail    B          rj  Assembled Connector       Detail    A       Figure 2   10  D O Terminal Board Views    Tab
421. lometer Zero Calibration  5 minutes     Please use   Port to access data download from the instrument     Use the following procedure when shutting down the instrument   1  Turn the power OFF     2  Cover instrument ports when not in use     DO NOT  e Do not pressurize the monitor   e Do not allow liquids to enter the sample port     e Do not expose the instrument to excessive vibration or magnetic  interference     e Do not remove adjust the beta attenuation bench of the 50307     e Do not remove the C 14 radioactive source at any time     Model 50307 SHARP Instruction Manual 2 23    Installation  Important Tips    DO    e Operate the instrument in a climate controlled environment     4 50   C    e Set DHCP to ON when changing the IP address     2 24 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Chapter 3  Operation    This chapter describes the front panel display screens  front panel  pushbuttons  and menu driven firmware  For details  see the following  topics     e    Display    on page 3 1   e    Pushbuttons    on page 3 2   e    Firmware Overview    on page 3 4   e    Range Menu    on page 3 10   e    Integration Time    on page 3 13   e    24 Hour Average    on page 3 14   e    Calibration Factors Menu    on page 3 14  e    Instrument Controls Menu    on page 3 18  e    Diagnostics Menu    on page 3 57   e    Alarms Menu    on page 3 66   e    Service Menu    on page 3 85    e    Password Menu    on page 3 103    Display The 320 x 240 pixels graphic
422. lrec srec format format     command    aa   hours  01 to 23     Model 50307 SHARP Instruction Manual B 29    C Link Protocol Commands  Datalogging    B 30 Model 50307 SHARP Instruction Manual    bb   minutes  01 to 59   oo   month  01 to 12   pp   day  01 to 31    qq   year    These commands output lrec or srec logged data  The output format is  determined by the    set lrec format     and    set srec format    commands  The  logging time is determined by the    set lrec per    and    set srec per     commands     In the following example  assume there are 740 lrecs currently stored in  memory  When the command    lrec 100 5    is sent  the instrument counts  back 100 records from the last record collected  record 740   and then  returns 5 records  640  641  642  643  and 644  For details on how to    decode the flag fields within these records  see the    flags    command     Send  lrec 5  Receive  lrec 100 5   08 28 05 04 09 flags 00110000 pm 0 000 baro 0 000 vac  260 000 pflow  52 000 ambrh 0 000 srh 0 000 ambtemp    32 000 stemp  41 000 fvol 0 000 cflg 00001F86 aflg  OOOOFCOC   08 29 05 04 09 flags 00110000 pm 0 000 baro 0 000 vac  260 000 pflow  52 000 ambrh 0 000 srh 0 000 ambtemp    32 000 stemp  41 000 fvol 0 000 cflg 00001F86 aflg  OOOOFCOC   08 30 05 04 09 flags 00110000 pm 0 000 baro 0 000 vac  260 000 pflow  52 000 ambrh 0 000 srh 0 000 ambtemp    32 000 stemp  41 000 fvol 0 000 cflg 00001F86 aflg  OOOOFCOC   08 31 05 04 09 flags 00110000 pm 0 000 baro 0 000 vac  
423. lter for supplying air free of suspended particles and its connection adapter    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRhein land 2    Air Pollution Control       Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 47 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    In order to check and  if required  calibrate the radiometric measurement  the manufacturer  provides a set of mass transfer standard foils  This set comprises one null foil and two span  foils  The null foil has the same approximate mass as a clean filter spot and the span foil is a  calibrated mass increase above the null foil  Therefore  the foils come in a set and must not  be mixed with other sets        Figure 17  Set with foil holder and test foils       It is not possible to produce a reproducible signal for the sensitivity of the nephelometer us   ing external test standards  However  the nephelometer is regularly flushed with zero air dur   ing operation  auto zero function as part of every filter change   thus determining and docu   menting the position of the zero point on a regular basis  Since the nephelometer is continu   ously calibrated with a very stable radiometric measurement  a separate examination of the  stability of the nephelometer   s signal under operating conditions can 
424. lue is a floating point representation of the coefficient   The following example sets the barometric mass coefficient to 0 000630     Send  set baro mass coef 0 000630  Receive  set baro mass coef 0 000630 ok  baro span    This command reports the current barometric pressure span recorded at the  time of calibration  The following example reports that the barometric span  is 1 0     Send  baro span  Receive  baro span 1 0000    set baro span value   This command requires to manually specifying the span value for the  barometric pressure span  where value is a floating point representation of  barometric pressure span  The following example successfully sets the  barometric span to 1 023     Send  set baro span 1 023  Receive  set baro span 1 023 ok  beta th    This command reports the current reference beta threshold for the  detector  The following example reports that the beta threshold is 440     Send  beta th  Receive  beta th 440    set beta th value   This command sets the reference beta threshold for the detector to value   where value is a floating point representation of the beta threshold  The  following example sets the beta threshold to 500     Send  set beta th 500  Receive  set beta th 500 ok  bkg pm    This command reports the current PM backgrounds  The following  example reports that the PM background is 4 g m        Send  bkg pm  Receive  bkg pm 4 000E 00 pg m3    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Calibration 
425. lution Control s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 85 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 4 1 7 Data output    The output signals shall be provided digitally  e g  RS232  and or as analogue signals   e g  4 mA to 20 mA      6 2 Equipment  PC with    iPort    software   6 3 Performance of test    The test was carried out using a PC with the software    iPort    via Ethernet    The AMS was connected to a PC via Ethernet and the data were downloaded     The AMS also offers the possibility of outputting analogue signals and measured signals or  communicating via serial interface RS 232 RS 485     6 4 Evaluation  The measured signals are offered as follows at the rear side of the instrument   Analogue  0 0 1  0 1  0 5 and 0 10 V selectable concentration range  0 20 or 4 20 mA selectable concentration range  Digital  Ethernet with    iPort  software  RS 232   RS 485 interface  6 5 Assessment    Measured signals are output as analogue signals  in V or mA  and digitally  via Ethernet   RS 232  RS 485      The connection of additional measuring and peripheral devices to the respective ports of the  devices is possible  e g  analogue inputs      Minimum requirement fulfilled  yes    740300_2013_936_21209885F_EN docx    A T  VRheinland   T  V Rheinland Energie und Umwelt GmbH    A
426. lyhalogenierten Dibenzo p Dioxinen und Dibenzofuranen bei Emissionen und Immissionen  Probenahme  von faserf  rmigen Partikeln bel Emissionen und Immissionen  Ermittlung von gas  oder partikelf  rmigen  Luftinhaltsstoffen mit kontinuierlich arbeitenden Messger  ten  Bestimmung von Geruchsstoffen in Luft   Kalibrierungen und Funktionspr  fungen kontinuierlich arbeitender Messger  te f  r Luftinhaltsstoffe ein   schlie  lich Systemen zur Datenauswertung und Emissionsfern  berwachung  Eignungspr  fungen von auto   matisch arbeitenden Emissions  und Immissionsmesseinrichtungen einschlie  lich Systemen zur Datenaus   wertung und Emissionsfern  berwachung  Feuerraummessungen  Ermittlung der Emissionen und Immissionen  von Ger  uschen  Ermittlung von Ger  uschen und Vibrationen am Arbeitsplatz  Akustische und schwingungs   technische Messungen Im Eisenbahnwesen  Bestimmung von Schallleistungspegeln von zur Verwendung im  Freien vorgesehenen Ger  ten und Maschinen nach Richtlinie 2000 14 EG und Konformit  tsbewertungsver   fahren  Schallmessungen an Windenergieanlagen  Ger  uschemission  Gerduschimmission   Immissionsprognose  auf der Grundlage der Technischen Anleitung zur Reinhaltung der Luft und der Geruchsimmissions Richtlinie  und der VDI 3783 Blatt 13    Modul Immissionsschutz    Die Akkreditierungsurkunde gilt nur in Verbindung mit dem Bescheid vom 23 01 2013 mit der  Akkreditierungsnummer D PL 11120 02 und ist g  ltig bis 22 01 2018  Sie besteht aus diesem Deckblatt   der R  
427. m  summer 2013  Annex 8  Ambient conditions during the additional campaign Bornheim  summer    2013  Appendix 2 Filter weighing procedure    Appendix 3 Manuals    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVR hein land 2    Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 175 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F       Dakks  A Deutsche    Akkreditierungsstelle    Deutsche Akkreditierungsstelle GmbH    Beliehene gem        8 Absatz 1 AkkStelleG i V m     1 Absatz 1 AkkStelleGBV  Unterzeichnerin der Multilateralen Abkommen  von EA  ILAC und IAF zur gegenseitigen Anerkennung    Akkreditierung I    Die Deutsche Akkreditierungsstelle GmbH best  tigt hiermit  dass das Priiflaboratorium    TUV Rheinland Energie und Umwelt GmbH  mit seiner    Messstelle fiir Immissionsschutz  Environmental Protection   Am Grauen Stein  51105 K  ln    und seiner unselbstandigen Messstelle  Robert Koch StraBe 27  55129 Mainz    die Kompetenz nach DIN EN ISO IEC 17025 2005 und CEN TS 15675 2007 besitzt   Pr  fungen in folgenden Bereichen durchzuf  hren     Bestimmung  Probenahme und Analytik  von anorganischen und organischen gas  oder partikelf  rmigen  Luftinhaltsstoffen im Rahmen von Emissions  und Immissionsmessungen  Probenahme von luftgetragenen  po
428. m limit  an  alarm is activated and the alarm  bell  icon appears in the status bar on the  Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  Neph Temp        RRS E       hi  ii          Min and Max The Minimum Nephelometer Temperature alarm limit screen is used to  Nephelometer change the minimum nephelometer temperature alarm limit  The  Temperature Limits minimum and maximum nephelometer temperature screens function the  same way     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  Neph Temp  gt   Min or Max     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 69    Operation  Alarms Menu    ip gies  PIL       Nephelometer RH The Nephelometer RH screen displays the current nephelometer relative  humidity value and sets the minimum and maximum alarm limits   Acceptable alarm limits range from 5 to 95   If the nephelometer RH  reading goes beyond the minimum or maximum limit  an alarm is activated    and the alarm  bell  icon appears in the status bar on the Run screen and in  the Main Menu     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  Neph RH        Min and Max The Minimum Nephelometer RH alarm limit screen is used to change the  Nephelometer RH Limits minimum nephelometer RH alarm limit  The minimum and maximum  nephelometer RH screens function the same way     e Inthe Main Menu  choose Alarms  gt  Neph Alarms  gt  Neph RH  gt   Min or Max        ip priori  PLP          LED Current The LED Current s
429. m paired measurements under field conditions shall be de   termined with two identical measuring systems by paired measurements in the field  test  It shall not exceed the requirements of Table 2 of Standard VDI 4202 Part 1  Sep   tember 2010      Note     This item cannot be evaluated according to the current versions of Standards VDI 4202 Part  1  September 2010  and VDI 4203 Part 3  September 2010  when testing particulate meas   uring systems  due to the fact that a minimum requirement has not been defined  By resolu   tion of the competent German body  see module 5 3 1   reference is made to the following  requirement in the previous version of Standard VDI 4202 Part 1  June 2002    The    Reproduzierbarkeit     reproducibility  Rp of the measuring system shall be deter   mined by parallel measurements with two identical measuring systems and shall be at  least equal to 10  B  shall be used as the reference value     6 2 Equipment  The reproducibility was determined with the auxiliary equipment described in Section 5   6 3 Performance of test    Reproducibility is defined as the maximum difference between two randomly chosen single  values obtained under equal conditions  This test was carried out with two identical systems  operated simultaneously during the field test  All measurement data obtained during the field  test was evaluated     6 4 Evaluation  The reproducibility is calculated as follows        n  R  1210 where U   t8p  tinogs  and sp   exe  i    e R   Repro
430. m shall not ex   ceed 180 s    The difference between the re   sponse time  rise  and re   sponse time  fall  of the meas   uring system shall not exceed  10  of response time  rise  or  10 s  whatever value is larger     5 3 15 Difference be  The difference between the  Not applicable   tween sample measured values obtained by  and calibration feeding gas at the sample and  port calibration port shall not ex   ceed the requirements of Table  2 of Standard VDI 4202 Part 1   September 2010      5 3 16 Converter effi  In the case of measuring sys    Not applicable   ciency tems with a converter  the effi   ciency of the converter shall be  at least 98       5 3 17 Increase of NO2   In case of NOx measuring sys    Not applicable   concentration tems  the increase of NO2 con   due to residence   centration due to residence in  in the measuring   the measuring system shall not  system exceed the requirements of  Table 2 of Standard VDI 4202  Part 1  September 2010      5 3 18 Overall uncer  The expanded uncertainty of By resolution of the competent Ger   tainty the measuring system shall be   man body  see module 5 3 1   this   determined  The value deter  test is not relevant for particulate  mined shall not exceed the cor    measuring systems  Reference is  responding data quality objec    made to module 5 4 10   tives in the applicable EU Di   rectives on air quality listed in  Annex A  Table A1 of Standard  VDI 4202 Part 1  September  2010         740300_2013_936_21209885F_EN 
431. maximum average SHARP concentration screens function  the same way     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Avg SHARP  gt   Min or Max        WH mri           The Average Nephelometer screen displays the current average  Nephelometer concentration and sets the minimum and maximum alarm  limits  Acceptable alarm limits range from 0 to 10000 ug m   or 0 to 10  mg m  If the average nephelometer concentration goes beyond either the  minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Avg Neph     Model 50307 SHARP Instruction Manual 3 81    Operation  Alarms Menu    Min and Max Average  Nephelometer  Concentration Limits    Instant PM       CHA Pit  Hels pid          The Minimum Average Nephelometer Concentration alarm limit screen is  used to change the minimum average nephelometer concentration alarm  limit  The minimum and maximum average nephelometer concentration  screens function the same way     e Inthe Main Menu  choose Alarms  gt  Conc Alarms  gt  Avg Neph  gt   Min or Max        The Instant PM screen displays the current instant PM concentration and  sets the minimum and maximum alarm limits  Acceptable alarm limits  range from  10 to 10000 pg m  or   01 to 10 mg m     If the instant PM  concentration goes beyond either the minimum or maximum limit  an  alarm is activated and the alarm  bell  icon appears in
432. may easily scratch the plate  A    Do not use alcohol  acetone  MEK or other ketone based  or aromatic solvent to clean the LCD module  but rather    Model 50307 SHARP Instruction Manual iii    About This Manual  Electrical Safety Certifications    Alert Description  use a soft cloth moistened with a naphtha cleaning  solvent  A    Do not place the LCD module near organic solvents or  corrosive gases  A    Do not shake or jolt the LCD module  A    Electrical  S afety The product has been tested and has been documented to be in compliance  Certifications with the following U S  and Canadian safety standards     UL Standard 61010 1 2004 2nd Edition  ites CAN CSA C22 2 No  1010 1 2004 2nd Edition    Eh C     IEC 61326 2 3  2006  c US Thermo Fisher Scientific certifies that this product operates in compliance    Py with the EC Directive 89 336 EEC in reference to electrical emissions and  immunity  Specifically  the equipment meets the requirements of EN  61326 1 1998 for Immunity and Emissions  In addition  the hardware has  been tested for personal or fire safety hazards in accordance with EN61010   1 2001  Safety  in fulfillment of EC Directive 73 23 EEC     FCC Compliance Changes or modifications to this unit not expressly approved by the party  responsible for compliance could void the user   s authority to operate the  equipment     Note This equipment has been tested and found to comply within the  limits for a Class A digital device  pursuant to Part 15 of the FCC Rules  
433. measuring system    Model 5030  SHARP with PM2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25  Page 213 of 424  Report No   936 21209885 F    Table 40  Stability of the control filters       Control filter No                                                              Weighing No  TM1 TM2 TM3  1 0 09257 0 09155 0 09110  2 0 09258 0 09155 0 09113  3 0 09260 0 09155 0 09115  4 0 09260 0 09157 0 09116  5 0 09262 0 09156 0 09117  6 0 09264 0 09157 0 09116  7 0 09262 0 09154 0 09114  8 0 09260 0 09156 0 09116  9 0 09262 0 09156 0 09113  10 0 09263 0 09160 0 09117  11 0 09263 0 09158 0 09118  12 0 09263 0 09158   0 09117  13 0 09267 0 09160 0 09118  14 0 09265 0 09157 0 09116  15 0 09266 0 09159 0 09119  16 0 09269 0 09162 0 09122  17 0 09268 0 09162 0 09121  18 0 09267 0 09161 0 09121  19 0 09266 0 09161 0 09118  20 0 09268 0 09160 0 09120  21 0 09264   0 09161 0 09117  22 0 09264   0 09159 0 09116          Average 0 09264 0 09158   0 09117  Standard de   viation 3 2911E 05   2 4937E 05   2 8558E 05  rel   standard de   viation 0 036 0 027 0 031             Median 0 09264 0 09158 0 09117  lowest value   0 09257 0 09154 0 09110  highest value   0 09269 0 09162 0 09122                               Marked in yellow   average value  Marked in green   lowest value  Marked in blue   highest value    740300_2013_936_21209885F_EN docx    TUVR heinland   TUV Rheinland Energie und Umwelt GmbH  Precisely Right Luftreinhaltung   
434. mentation  The Calibration submenus allow the user to view and  calibrate the instrument  All calibration submenus are visible only when the  instrument is in service mode  For more information on the service mode   see    Service Mode    earlier in Chapter 3     Thermo Fisher Scientific    Calibration  Calibration Procedure    RH Temperature All screens that appear in this chapter have been referenced in the previous  Calibration       lt hapter     Ambient Temperature Using a NIST traceable thermometer as a reference collocated next to the  ambient RH temperature sensor assembly  measure and compare three  individual readings between both the reference and the 5030  response   Taking an average of both sets of readings  calculate the average difference    between the two readings and record that as your offset  This offset   ZERO  should now be entered in the screen below        ERE CH  perm    WH mnie   PS PHP ELSES Pi AG Lf bie       Be sure to save the entry and compare the values once more  If it appears  that the temperature has shifted into the wrong direction  change the sign  of your offset value     Ambient Relative Using a NIST traceable hygrometer as a reference collocated next to the  Humidity ambient RH temperature sensor assembly  measure and compare three  individual readings between both the reference and the 5030  response   Taking an average of both sets of readings  calculate the average difference  between the two readings and record that as your offset 
435. milar experiences during measurement campaigns in China  the manufacturer  suggested a change in the parameterisation of the relative air humidity threshold from 50    to 58   to solve the problem  This would activate the heater only when the relative humidity  exceeds 58   instead of 50   as in the previous campaigns  The Teddington  summer   comparison campaign was subsequently repeated using the new parameters  average vola   tile content during this campaign  34 3     No problems demonstrating the equivalence dur   ing this measurement campaign were detected after applying this change  The results are  presented in detail in Section 6 1 5 4 10 Calculation of the expanded uncertainty of the  instruments     Table 6 and Annex 6 of this report present the results of the measurement campaigns with  regard to content of volatiles in suspended particulate matter PM2 5     To qualify the necessary change of the threshold for the relative humidity to control the sam   ple heating from originally 50  to 58   necessary due to the results of the campaign in Ted   dington UK   an additional campaign at one of the German sites with two candidate systems  with settings of each 50   and 58   was performed  decision of the 32  Expert discussion     test reports    of May 14 15 201  Table 2 gives an overview on the additional campaign  The  results of this campaign are shown in chapter 4 3 Field test from page 55     The tests were performed at the site Bornheim with the candidate systems
436. mparison candidate SN 2 with reference instrument  rel  humidity  gt  70    all  test sites  measured component PM25       Comparison candidate with reference according to   Guide  Demonstration of Equivalence Of Ambient Air Monitoring Methods   January 2010   Candidate Modell 50301 SHARP SN SN 2   Test site All test sites  rH gt 70  Limit value 30 g m    Status of measured values Raw data Allowed uncertainty 25     Slope b significant  Uncertainty of b  Ordinate intercept a not significant  Uncertainty of a  Deviation at limit value  Uncertainty Uc s at limit value    Combined measurement uncertainty wom    Expanded measurement uncertainty Wom  Status equivalence test       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 137 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 4 6 Averaging effect    The measuring system shall allow for formation of 24 h averages   The time of the sum of all filter changes within 24 h may not exceed 1   of this averag   ing time     6 2 Equipment   A clock was required for this test    6 3 Performance of test   It was tested  whether the AMS allows the formation of daily averages   6 4 Evaluation    By default  the measuring system saves the meas
437. mperature RH cable assembly   Please be sure the side panel has been removed for at least 60 minutes  A    2 6 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific       One Point Temperature    One Point RH Sensor    Thermo Fisher Scientific    Verification    Verification    Installation  Acceptance Testing and Startup Procedures       Figure 2 2  Model 5030  Sampling Room Air Under Stable Conditions    Record the temperature sensor readings from the Model 50307 and  compare to your NIST traceable thermometer  Each of these measurements  should be within  2   C tolerance of your NIST traceable thermometer   Within this tolerance  the temperature sensors have passed the acceptance  test     If the sensors are slightly out of tolerance   3   C   the acceptance test  should be classified as marginal  If the sensor performance is outside this  range  please perform a temperature sensor calibration     For more information about the temperature sensor calibration  see the     Calibration    chapter     Record the ambient RH sensor and nephelometer RH sensor readings from  the Model 50307 and compare to your NIST traceable Hygrometer  The  Model 50302 RH sensors should compare within  2  RH tolerance of  your NIST traceable hygrometer  If the RH sensor performance is within  this tolerance  the acceptance test has passed     Note The ambient temperature verification should be completed prior to  performing the RH sensor verification due to a thermal compensation  applied
438. mprises one null foil and two span foils  The null foil has the same ap   proximate mass as a clean filter soot and the span foil is a calibrated mass increase above  the null foil  Therefore  the foils come in a set and must not be mixed with other sets  The foil  set is used for determining mass only     A comparison campaign was carried out once at the beginning and once at the end of the  test in order to determine the mass coefficient     It is not possible to produce a reproducible signal for the sensitivity of the nephelometer us   ing external test standards  However  the nephelometer is regularly flushed with zero air dur   ing operation  auto zero function as part of every filter change   thus determining and docu   menting the position of the zero point on a regular basis  Information on the current and pre   vious NEPH backgrounds can be retrieved from the instrument at all times  Since the nephe   lometer is continuously calibrated with a very stable radiometric measurement  a separate  examination of the stability of the nephelometer   s signal under operating conditions can be  omitted     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 80 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2
439. municating the results of the above calculations  The   front panel display presents the concentrations simultaneously  The display  is updated every 1 10 seconds  depending on the averaging time     The analog output ranges are user selectable via firmware  The analog  outputs are defaulted based on the measurement range  Negative  concentrations can be represented as long as they are within  5  of full   scale  The zero and span values may be set by the user to any desired value     All electronics operate from a universal switching supply  which is capable  of auto sensing the input voltage and working over the all specified ranges     External pumps and heaters all operate on 110 VAC  An optional  transformer is required if operating on the 210 250 range     An on off switch controls all power to the analyzer and is accessible on the  front panel     The motherboard contains the main processor  power supplies  and a sub   processor  and serves as the communication hub for the instrument  The  motherboard receives operator inputs from the front panel function keys  and or over I O connections on the rear panel  The motherboard sends  commands to the other boards to control the functions of the instrument  and to collect measurement and diagnostic information  The motherboard  outputs instrument status and measurement data to the graphics display  and to the rear panel I O  The motherboard also contains I O circuitry and  the associated connector to monitor external digital
440. n Control   7  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 71 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F             100  e Not Outlier  90 5   Outlier Deleted     Outlier Not Deleted  80 4 1 to 1 Line                zZ aouaIajay Wd                PM  5 Reference 1             Figure 29  Grubbs test results for the PM2 5 reference method  Bornheim  summer              100    Not Outlier  90 5   Outlier Deleted     Outlier Not Deleted  80 4 1 to 1 Line                z souaIajaY FIN                PM     Reference 1             Figure 30  Grubbs test results for the PM2 5 reference method  Bornheim  winter     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  TUVRheinland  Air Pollution Control    Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz 5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 72 of 424                   100    Not Outlier  90       Outlier Deleted     Outlier Not Deleted  80 1 to 1 Line          Z souaIajay Wd                0 10 20 30 40 50 60 70 80 90 100    PM     Reference 1             Figure 31  Grubbs test results for the PM25 reference method  Teddington  summer 
441. n alarm  minimum and maximum values to value  where value is a floating point  representation of the concentration alarm limits  Values must be in the  units that are currently set for use  The following example sets the average  PM concentration alarm maximum value to 10000 mg m        Send  set alarm conc avg pm max 10000  Receive  set alarm conc avg pm max 10000 ok    alarm conc avg sharp min   alarm conc avg sharp max   These commands report the current 24 hour average SHARP  concentration alarm minimum and maximum values setting  The following  example reports that the average SHARP concentration minimum is 0 0  pg m      Send  alarm conc avg sharp min  Receive  alarm conc avg sharp min 0 000E 00 pg m3    set alarm conc avg sharp min value   set alarm conc avg sharp max value   These commands set the 24 hour average SHARP concentration alarm  minimum and maximum values to value  where value is a floating point  representation of the concentration alarm limits  Values must be in the  units that are currently set for use  The following example sets the average  SHARP concentration alarm maximum value to 10 000 pg m        Send  set alarm conc avg sharp max 10000  Receive  set alarm conc avg sharp max 10000 ok    alarm conc inst neph min   alarm conc inst neph max   These commands report the current 24 hour instant nephelometer  concentration alarm minimum and maximum values setting  The following  example reports that the instant nephelometer concentration minimum is 0  pg
442. n body  see module 5 3 1   this test is not relevant for  particulate measuring systems  Reference is made to module 5 4 10     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 129 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 4 1 General    Testing of measuring systems for particulate air pollutants is carried out in compliance  with the minimum requirement stated in Table 5 of Standard VDI 4202  Part 1   Furthermore  the particle mass concentration shall be related to a defined volume  The  relation to volume with respect to pressure and temperature shall be comprehensively  described     6 2 Equipment  No additional equipment required   6 3 Performance of test    Testing was carried out in compliance with the minimum requirements stated in Table 5 of  Standard VDI 4202  Part 1  September 2010     It was tested  whether the measured particulate mass concentrations are related to a defined  volume    6 4 Evaluation   Testing was carried out in compliance with the minimum requirements stated in Table 5 of  Standard VDI 4202  Part 1  September 2010      The Model 5030i SHARP measuring system combines two different methods of measure   ment  nephelometry   radiometry   Here  t
443. n the results     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25     A TUVRheinland      Genau  Richtig     Page 219 of 424  Report No   936 21209885 F  20  20  D     150       pe      O  oi 100  9            a  50    0  50  20  D            Oo 150      O       Q   N wo       7   O  a  50  0  BBSBSBBSBEFEBSSRBRSSSREBERVPBBSBEFRBSBRBSSSSB  ye Cp epaeaeaeaeeesetfsaeaesee2e2nnns  BPanpBRBAL SES       9 c  BeS88R8R BR 9 HR  FT SRAEFBBR BEE EE  8         se   g  Range of Spread   ug 5  Figure 63  Distribution for Emfab filters of  A  Pre spread weighed relative to the check    weight and  B  Post spread weighed relative to the check weight     740300_2013_936_21209885F_EN docx    TUVRheinland    Precisely Right     Page 220 of 424    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung  Report on the suitability test of the ambient air quality measuring system    Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Report No   936 21209885 F    Appendix 3    Manuals    Model 5030  SHARP    Instruction Manual   Synchronized Hybrid Ambient Real time  Particulate Monitor   Part Number 108148 00   25Mar201
444. nally with a reliable deviation  from the nominal value of  2 5    If this criterion is not met during operation  the measuring  system performs a filter change    To determine the constancy of the sample volumetric flow  the internal data regarding flow  rate and pump vacuum    information on filter load  were recorded for the test site Cologne   winter   characterised partly by high concentrations  and the flow rates were evaluated on a  24 h basis    6 4 Evaluation    Average values  standard deviations as well as maximum and minimum values were deter   mined from the flow rate measured values     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 139 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 5 Assessment    The results of the flow rate checks performed before testing at each field test site are pre   sented in Table 30                    Table 30  Results of flow rate checks  Flow rate check before SN 1 SN 2  testing at     l min  Deviation  l min  Deviation  Test site  from nominal from nominal   value     value       Cologne  winter 16 33  2 04 16 44  1 38  Bornheim  summer 16 48  1 14 16 51  0 96  Bornheim  winter 16 66  0 06 16 68 0 06  Teddington  summer 16 80 0 78 16
445. nc alarm setting    Re zero    Verify conc alarm settings     Model 50307 SHARP Instruction Manual 6 7    Troubleshooting  Board Level Connection Diagrams                                                                                                                      Board Level Figure 6 1 and Figure 6   2 are board level connection diagrams for the  Connection             electronics and measurement system  These illustrations can be  used along with the connector pin descriptions in Table 6 3 through Table  D lagrams 6 7 to troubleshoot board level faults   VOEXPANSION  BOARD  EPANSICNIO J1  MEASUREMENT  DIGITAL OUTPUT  TO SYSTEM  BOARD  DOTA  ag 2  11Pn 8Pn  14  PLA i 14Pin LOD  PSAP  Hia oe DISPLAY  RLPS PIB Jig               PANEL  9Fin     ARn 3Fin  BOARD    B  1OBASET 2 Sa L_ KEYPANEL  EXPANSION RS485       J15  a MOTHER BOARD Hoy SPAREDATA  ANALOGOUIPUTS   DIGITAL INPUTS 37 pn     J alla  gt  3Fin  3Fin  s 2  n  Pe POWER GOOD  oe dia  PJI RB Re  3Pn 3Pn   3Pin  24VDC  REARPANEL     AC POWER 24VDC  ANUT     30m SUPPLY i  AC  FRONT PANEL  FONERSN S O    eS eet ae ce we  gt  3Pn    1 1 a  I TRANGFORVER 1 Ue  ee ee 1 k inie     QOVAC or 240VAC Chior    6 8 Model 50307 SHARP Instruction Manual    Figure 6   1  Board Level Connection Diagram   Common Electronics    Thermo Fisher Scientific    Troubleshooting  Board Level Connection Diagrams                                                                                  PJ34 ep   BETA 1 DETECTOR  PJ35   BETA 2 DETECTOR 
446. nclude the product model and serial number  if  applicable  and details of the warranty claim   b  after Seller   s review  Seller  will provide Buyer with service data and or a Return Material  Authorization     RMA      which may include biohazard decontamination  procedures and other product specific handling instructions  and  c  then   if applicable  Buyer may return the defective Products to Seller with all  costs prepaid by Buyer  Replacement parts may be new or refurbished  at  the election of Seller  All replaced parts shall become the property of Seller   Shipment to Buyer of repaired or replacement Products shall be made in  accordance with the Delivery provisions of the Seller   s Terms and  Conditions of Sale  Consumables  including but not limited to lamps   fuses  batteries  bulbs and other such expendable items  are expressly  excluded from the warranty under this warranty     Notwithstanding the foregoing  Products supplied by Seller that are  obtained by Seller from an original manufacturer or third party supplier are  not warranted by Seller  but Seller agrees to assign to Buyer any warranty  rights in such Product that Seller may have from the original manufacturer  or third party supplier  to the extent such assignment is allowed by such  original manufacturer or third party supplier     In no event shall Seller have any obligation to make repairs  replacements  or corrections required  in whole or in part  as the result of  i  normal wear  and tear   ii 
447. nd  set instrument id 12  Receive  set instrument id 12 ok  layout ack    This command reports the stale layout layout change indicator     that is  attached to each response if the erec layout has changed since the last time  erec layout was requested  according to Table B   9  The following example  reports that the instrument is configured to do nothing     Send  layout ack  Receive  layout ack 0    set layout ack value   This command disables the stale layout layout change indicator     that is  attached to each response if the erec layout has changed since the last time  erec layout was requested  according to Table B   9     Send  set layout ack  Receive  set layout ack ok    Table B 9  Set Layout Ack Values       Value Function  0 Do nothing  default   1 Append          power up mode   This command reports the current power up mode setting to either 0    local unlocked or 1   remote locked  according to Table B   10  The default  value is 0  power up in local unlocked mode  The following example shows  that the instrument is configured to power up in the remote locked mode     Send  power up mode  Receive  power up mode 1    Model 50307 SHARP Instruction Manual B 59    C Link Protocol Commands  Communications Configuration    B 60 Model 50307 SHARP Instruction Manual    set power up mode value   This command is used to configure the instrument to power up in the  local unlocked mode  value   0  or the remote locked mode  value   1    according to Table B 10     If the 
448. nd Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control    s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 69 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    Sampling period   EN 14907 defines a sampling period of 24h  1 h    The sampling period was set to 24 h for all instruments during the entire field test  10 00      10 00 in Teddington and Cologne  and 7 00     7 00 in Bornheim     Data handling    All paired reference values determined during the field tests were subject to statistical testing  according to Grubbs  99    to prevent influences of clearly implausible data on the meas   urement results  Paired values identified as significant outliers may be discarded from the da   ta pool as long as the critical value of the test statistic does not fall below the target  The  January 2010 version of Guide  4  stipulates that only 2 5   of the paired values may be  identified as outliers and discarded    During the    Combined MCERTS and T  V PM Equivalence Testing    programme we agreed  with our British partners not to discard any value measured by the candidates  unless the im   plausibility is attributed to technical causes  Thus  no values measured by the candidates  were discarded during the entire testing period    Table 12 shows an overview of the amount of paired values  re
449. nd hybrid silicon photodetector to measure the intensity of scattered light  in the presence of aerosol  This assembly houses a nephelometer   temperature sensor  relative humidity sensor  and PCB interface     The primary measurement head uses a beta source and proportional  alpha beta detector to measure the attenuation of beta particles by sampled  aerosol across a glass fiber filter and to discriminate against any naturally  occurring beta interferences from the daughter nuclides of alpha decay  products     This cylindrical tube houses a printed circuit board  provides power to the  proportional detector and amplifies the beta and alpha counts prior to  being sent to the 50307 measurement interface board     This small printed circuit board includes a photo interrupter to detect the  closed position of the measurement head  This is connected to a motor  assembly used to open and close the measurement head     This small printed circuit board is mounted onto a chopper wheel that  measures the amount of filter tape advanced during a filter tape change     The proportional valve is a proportional solenoid valve that is used as a  means of flow control  In response to the measured flow across a subsonic  orifice  the proportional valve will either partially open or close to make a  flow adjustment in real time     This motor is used to drive the cam to rotate  By rotating the cam  the  measurement head lever arm is raised and lowered for opening and closing  the measurement 
450. nd now deliberately  writes a response in every case   6  Several parameters are no longer displayed  when the unit is password locked   7  Remove blanket rejection of    set    for the    sp  field    Clink command                The reliability of the AMS was consistently increased with every change  The changes intro   duced up to version V02 00 00 232  are not expected to affect the performance of the AMS  significantly    The suitability test comprised a laboratory test for determining the system   s performance  characteristics and a field test carried out over several months at different locations    All determined concentrations are given as g m   operating conditions      The report in hand provides a description of the minimum requirement for each test item in  accordance with the number and wording used in standards  1  2  3  4      740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 54 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    4 2 Laboratory test    The laboratory test was carried out with two identical instruments of the type Model 5030i  SHARP with serial numbers SN 1 and SN 2  In conformity with standards  1  2   the following  performance criteria w
451. nd risks of equipment damage  Refer to the following types of  alerts you may see in this manual     Safety and Equipment Damage Alert Descriptions    Description    DANGER A hazard is present that could result in death or serious  personal injury if the warning is ignored  A    Alert  WARNING A hazard or unsafe practice could result in serious  personal injury if the warning is ignored  A  CAUTION A hazard or unsafe practice could result in minor to  moderate personal injury if the warning is ignored  A  f Equipment Damage A hazard or unsafe practice could result in property    damage if the warning is ignored  A    Safety and Equipment Damage Alerts in this Manual    Alert Description  WARNING The Model 5030  is supplied with a three wire grounded  power cord  Under no circumstances should this    grounding system be defeated  A    Thermo Fisher Scientific    Thermo Fisher Scientific    Alert    A CAUTION    f Equipment Damage    About This Manual  Safety and Equipment Damage Alerts    Description    If the equipment is operated in a manner not specified by  the manufacturer  the protection provided by the  equipment may be impaired  A    The service procedures in this manual are restricted to  qualified service personnel only  A    The detector assembly should be replaced by a qualified  technician knowledgeable in dealing with radiation  precautions  By removing the detector assembly  the C 14  radioactive source is partially exposed  Safety glasses  must be worn during t
452. nd the maximum limit  an alarm is    activated and the alarm  bell  icon appears in the status bar on the Run  screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Detector Alarms  gt  Alpha        Min and Max Beta The Beta Alarms screen allows the user to view the minimum and  Alarms maximum alarm parameters  Acceptable alarm limits range from 5000 to  20000  If the beta count reading goes beyond the maximum limit  an alarm    is activated and the alarm  bell  icon appears in the status bar on the Run  screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Detector Alarms  gt  Beta        amga    PME TY Saad fat          ip gies   PLP          Nephelometer The Nephelometer Alarms submenu allows the user to view and set the  Alarms nephelometer alarm parameters  Board status indicates that the power  supplies are working and connections are successful  The nephelometer    3 68 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu    temperature  nephelometer RH  LED current  and reference detector  voltage all have adjustable alarms pre set in the factory     e Inthe Main Menu  choose Alarms  gt  Neph Alarms        Wi moh           Nephelometer The Nephelometer Temperature screen displays the current nephelometer  Temperature temperature value and sets the minimum and maximum alarm limits   Acceptable alarm limits range from 0 to 60   C  If the nephelometer  temperature reading goes beyond the minimum or maximu
453. ne  Winter 18 4 2 5 13 8 21 9 28 3 1011 37 0 3 4 109 0 0  62 5 8 2011 12 3 Tr 13 7 22 1 28 6 1013 34 7 4 1 97 0 0  63 7 25 2011 Bornheim  Summer 11 6 172 22 8 1001 73 8 0 8 193 1a  64 7 26 2011 11 8 2 9 24 1 17 0 20 6 1006 78 4 11 259 0 0  65 7 27 2011 14 0 4 2 29 9 17 3 24 3 1010 84 8 0 7 229 51 7  66 7 28 2011 18 2 5 1 27 9 17 7 24 5 1011 85 6 0 6 228 5 9  67 7 29 2011 9 0 3 5 39 0 16 9 20 2 1011 76 2 2 9 299 0 0  68 7 30 2011 1 6 14 5 16 1 1010 80 4 2 3 288 0 0  69 7 31 2011 9 7 1 2 12 6 13 4 16 6 1009 76 2 0 9 244 0 0  70 8 1 2011 12 1  r 13 8 19 1 26 1 1008 68 1 0 9 177 0 0  71 8 2 2011 10 2 1 4 13 4 23 2 30 0 1007 60 8 1 5 143 0 0  72 8 3 2011 16 8 2 6 15 8 19 7 24 1 1006 82 9 LS 195 5 4  73 8 4 2011 8 6 3 2 36 8 22 3 29 2 1005 71 6 0 8 204 3 6  74 8 5 2011 9 6 1 8 18 3 20 6 25 0 1004 75 8 i Pal 221 0 6  75 8 6 2011 A  19 3 25 4 996 85 0 1 4 171 Tale  76 8 7 2011 3 0 1 3 43 5 17 8 22 4 998 64 0 1 5 209 1 8  7 8 8 2011 3 4  0 2  6 8 15 5 18 4 1000 74 7 25 219 6 2  78 8 9 2011 3 0  0 1  4 8 13 8 17 8 1012 76 8 27 265 14 8  79 8 10 2011 5 8 2 3 38 9 18 0 22 4 1013 57 5 1 3 224 0 0  80 8 11 2011 5 3 1 3 25 3 20 9 25 5 1004 53 8 1 2 220 0 3  81 8 12 2011 3 4 1 0 29 7 18 5 2st 1003 78 9 a Pea  228 21  82 8 13 2011 2 0 20 1 23 7 1001 77 0 0 7 185 0 0  83 8 14 2011 3 1 0 1 3 8 17 4 19 9 1000 86 2 11 219 17 4  84 8 15 2011 5 7 0 9 15 7 17 9 22 4 1009 71 8 1 2  230 0 0  85 8 16 2011 6 0 1 6 27 6 19 1 23 9 1010 69 0 0 7 190 0 6  86 8 17 2011 14 3 3 1 21 6 RIA  27 8 1007 73 8 0 7 2
454. nent suspended particulate matter PM2 5    Report No   936 21209885 F    As sampling inlet  a PMz25 inlet  manufacturer  Digitel  type  DPM2 5 01 00 16  which serves  as pre separator for suspended particles sucked in from outside air  is used  The instruments  are operated at a constant volume flow rate adjusted to 16 67 l min   1 0 m  h  It is also pos   sible to use PM  and PM  sampling inlets instead           Nozzles  8   Baffle Plate    Safety Snap Lock   Tube  d   16 14     Figure 7   European  PM25 sampling inlet for Model 5030i SHARP    The heated sampling tube  approx  0 9 m long  and the  optional  extension tube  approx   1 8 m long  connect the sampling inlet to the actual measuring system  The adjustment of the    sample heater  dynamic heating system DHS  is carried out as described in Section 3 2  Functionality of the measuring system     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control A    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 39 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F       Figure 8  Heated sampling tube  DHS        Figure 9  Model 5030i SHARP measuring system    740300_2013_936_21209885F_EN docx    TUVRheinland      Precisely Right     Page 40 of 424    TUV Rheinland Energie und Umwelt GmbH  Air Pollut
455. ner not specified by the  manufacturer  the protection provided by the equipment may be  impaired  A    CAUTION The detector window is very fragile  Handle with great care and  do not wipe or touch the window  Furthermore  do not touch the solder  cable connection  electrode  with your bare fingers  The oils from your skin  can damage the detector  A    Safety glasses must be worn while replacing the detector assembly  A    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Disconnect the SHV connector from the detector amplifier assembly    and slide throught the wall plate     3  Loosen both detector mounting screws and slowly lift the detector  assembly     4  Install the new detector assembly by following the previous steps in  reverse  while taking special care not to puncture the mylar     5  Calibrate the detector assembly  Refer to the    Detector Calibration     procedure on page 4 12     Model 50307 SHARP Instruction Manual 7 35    Servicing  Flow RTD Replacement    Unhook SHV Connector                                                                                                                                                                                                                               ll    o o Unfasten Screws  2          
456. ng example shows that the nephelometer is in sample mode     Send  neph mode  Receive  neph mode sample  neph prev bkg    This command reports the previous nephelometer background values  stored on the nephelometer interface board  The following example reports  the previous nephelometer auto ranging background values     Send  neph prev bkg  Receive  neph prev bkg 0 00272 0 00272 0 01094 0 04357 0 17422  0 65535    set reset tape counter  This command resets the tape count to zero     Send  set reset tape counter  Receive  set reset tape counter ok  sample rh    This command reports the sample RH  in percent  The following example  reports that the sample RH is 20      Send  sample rh  Receive  sample rh 20 0      Thermo Fisher Scientific    Datalogging    Thermo Fisher Scientific    C Link Protocol Commands  Datalogging    tape counter  This command reports the number counts which corresponds to the filter  tape movement  The following example reports that the tape count is 250     Send  tape counter  Receive  tape counter 250  vac pres    This command reports the current vacuum pressure for the detector  in  mmHg  The following example reports that the vacuum pressure is 100  mmHg     Send  vac pres  Receive  vac pres 100 0 mmHg    clr records  This command will clear all lrecs and srecs that have been saved     Send  clr records  Receive  clr records ok    set clr lrecs   set clr srecs   These commands will clear only the lrecs or only the srecs that have been  saved  The
457. ng term drifin ashe ste oles late Base thee ieee eae 117  5 3 13 Short term AON ccs ccascacranctecenenaciesenacaimieteniimess Sa ietsuseienimaeteaadersteneiecetmieaeameacpeny 123  5 3 14 Response time aaj sas sch a scceradecca soci ebacedarecegsachaucatuan heer eseaatausassaieaciuceeeiauieiaactice 124  5 3 15 Difference between sample and calibration Port             c  ccceeeeeeeeeeeeetteeeeeees 125  5 3 16 Converter efficiency siccncccccncccarsenseursndeenendeuaenstenineeenceenentenis dxeeonereceneceeenescnnbete 126    5 3 17 Increase of NO  concentration due to residence in the measuring system    127    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland    Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 7 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    6 1 5 3 18 Overall uncertainty           ssssseeseeesssesererrrrresstererrrrrtrsserttttrrrrrnntresetnttrnnnneseennnee 128  6 1 AA General oree e a E a a EE E 129  6 1 5 4 2 Equivalency of the sampling system                   ccccceeeeeseeeeeneeeeeeeeeeetesensaaaeaaeaes 130  6 1 5 4 3 Reproducibility of the sampling systems               ccceeeeeeeeeteeeeeeeeeeeteeeeeneeeeeees 131  6 1 Oe    6  e212    eee meee eee eee ee ee eee eee eer ee re ee 132  6 1 5 AS CrOSS SOriSitivity aicese
458. nrrrnnrrsseresrrennnrnnnrnserrenne 91  5 2 5 Operating States   iieiea EE E EAEEREN E NA EEE EE ENE ENAERE E 92  5 2 6 Switch Over Senne ee ee mee ee ere ree eee eee eee ae 93  5 2 7 Maintenance interval ia secs cteadentest cnet caan ceases nsne dean cianteatnviettncecntndeaniantucnamctass 94  5 28 Availability annn a a E A 95  5 2 9 Instrument software           ssssseesseesseseetrrrrtresttttrrnnnrtsttrssttttttrnnnnesetnnttnn nense nenne 97  5S3 CARING ial rne AEE E E E E E S 99  5 3 2 Repeatability standard deviation at zero point          sssssseresessseserrrrrresssrrrrrnn 101  5 3 3 Repeatability standard deviation at reference POINt               ccceeeeeeetettereeeees 103  5 3 4 Linearity  lack of fit  oiciacerctcetneiesereateuideee ds dreaaetertatneeddedanld Geceneeeandneaeenrets 104  5 3 5 Sensitivity coefficient of sample gas pressure           ssssssresesssererrrrrrrsserrrrrrrn 105  5 3 6 Sensitivity coefficient of sample gas temperature             c  ccceeeeeeeeeeeeetteeeeeeees 106  5 3 7 Sensitivity coefficient of surrounding temperature                 cceeeeeeeeesnteeeeeeees 107  5 3 8 Sensitivity coefficient of supply voltage              cceceeeeeeeeeeeesteeeeeeeeeetteteteeeenaaees 111  5 3 9 Cross sensitivity aa ths ees te tc eet ee eee 113  5 3 10 Averaging effect cecicqaccterereexeietenencietdnmraceileceneueneiereinnicetel el eeemeieuetenciers 114  5 3 11 Standard deviation from paired measureMent              cccceeeeeeeeeeeeeeeeetteeeeeeees 115  5 312 Lo
459. nsert with the Span foil  e g   1328 ug   and press to begin the span calibration  or press      to stop  the foil     4 10 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Calibration  Calibration Procedure       The span foil calibration procedure is now active and after another 270  seconds the calibration procedure should end  whereby a new Mass  Coef value is shown and the percent difference is shown from the last  foil calibration to the current calibration  This percent difference can  also be used to directly audit the mass coefficient value since a percent  difference in a mass coefficient is the same percent mass difference  overall        6  After the span calibration  the screen will ask you to press to  accept the new calibration  Otherwise press         to keep the old Mass  Coef value  Remove the span foil and press   4    The bench will now  open  Remove the filter holder  replace the filter tape and press    to confirm the calibration routine has now been competed        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 4 11    Calibration  Calibration Procedure       Detector Calibration The Detector Calibration submenu is used to view and set the detector  calibration  The detector calibration screen is visible only when the  instrument is in service mode  For more information on the service mode   see    Service Mode    earlier in the chapter     Note This adjustment should only be performed by an instrument service  technician 
460. nsity N   particle number concentration  Cm   mass concentration f d     particle size distribution function  A   illumination wavelength r   distance of scattered particles to detector    inio   Lorenz Mie scattering functions  alternately dependent on the relation of particle size  to wavelength and on the refractive index    f A    combined dependence of wavelengths on emitter and detector  F 0    angular scattering geometry of the optical system    The high sensitivity of the nephelometer is thus accompanied by an overall dependence of  the measured signal on the size of particles  This means that nephelometric measuring sys   tems are  in contrast to gravimetric reference measurements  prone to have difficulties  caused by shifts in response     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 30 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    The nephelometer is therefore calibrated on site with a HEPA filter  zero point  and test aer   osols for reference  with PM25 Sharp Cut Cyclone connected upstream   Zero point checks  can be performed anytime with help of the HEPA filter at the instrument inlet  Furthermore   the sensitivity of the nephelometer is continuousl
461. nstruction Manual 2 3    Installation  Acceptance Testing and Startup Procedures    Acceptance Test    2 4 Model 50307 SHARP Instruction Manual    program  Furthermore  it is an opportunity to assure that the monitor is  operating according to the manufacturer specifications     After acceptance testing  a completed monitoring installation will require  final volumetric flow rate verification     Prior to installing the Model 50302  you should perform the acceptance  testing procedures  These tests are conducted to evaluate the out of box  performance of the instrument  perform any necessary calibrations prior to  final site installation  and familiarize the user with the menu structure     Use the following procedure to perform the bench acceptance test   Equipment Required   Model 50302  Power Cord  Ambient Temperature RH Cable Assembly  Vacuum Pump Assembly  Flow Adapter Assembly  NIST traceable Thermometer  NIST traceable Hygrometer  NIST traceable Barometer  NIST traceable Manometer s   NIST traceable Volumetric Flow Transfer Standard           Do not use the heated sample tube assembly at this point      1  After unpacking the instrument place the Model 50303 and accessories  onto a table or bench located within a stable indoor environment   Remove the side panel cover and inspect the filter tape  Leave cover  OFF until after acceptance testing has been completed     Note Prior to continuing with acceptance testing  the instrumentation and  accessories should have suffi
462. nstrument that can make accurate mass  concentration measurements of PM  PM25 and PM o  However  with all  electronic instruments  the accuracy of the measurements depends on  proper calibration     In general terms  calibration  also called standardization  is the process that  establishes the relationship between sensor output signals and the  parameter the operator is attempting to measure  The 5030  instrument  uses sensors that measure temperature  humidity  pressure  and radiation   This chapter describes the procedures for performing the necessary sensor  calibrations  This chapter is also a follow up to the Acceptance Testing  outlined in the    Installation    chapter and a continuation of the menu  display descriptions in the    Operation    chapter     Each instrument is calibrated and tested for accuracy at the factory  The  mass calibration is conducted with a set of NIST traceable mass calibration  foils  The radiation detector is calibrated against certified check sources for  beta and alpha emissions  The temperature  relative humidity  pressure and  flow rate are all calibrated against NIST traceable standards     It is recommended that the instrument be calibrated once per year and  encounter quality checks on a periodic basis in accordance with the users  specific quality assurance project plan  For example  volumetric flow checks  on a two week basis are common in some air monitoring agencies and with  full scale quarterly audits for flow and mass  The Qua
463. nt Pre DaTAliO Mixa caksics seavecensens tet snastyssnrenvdes sosinrbenccared vise tenn 10 1  SHARP Connection Bracket Assembly Installation           cceeeeeeeee 10 2  Radius Tube Adapter Installation s i   9 asvaiioheteaenenveine 10 5  SHARP External Valve  Installation 20 55  4  Ace acae eee 10 7  SHARP Optics Assembly Installation  j 22 1024  canines 10 9  Upgrade of the Heater Tube Assembly    10 10  Model Label Change iiinis 10 11  Power On INSU MICH sicions ioaea 10 11  VCC  es aa A 1  C Link Protocol CommandS               nnnnn1rrrreereerrrnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnrrrrrrrrrnrne B 1  Instrument Identification Number ses4 0it  ioe eae elec B 1  Commands sste i oaeaeei aian dened ah SS etl B 2  Commands List  sonans eiea rere a aN EE EEA S A a B 3  Meas  tementS oieta o A a EES B 11  Aarm S A n O ae R ae B 14  Diagnostics seesi iiris iiaii iiai iii B 23  VAAL TN Oe enganes eate Ee coe cae E a a s ES B 27  Calibrations awteciieivivatiiacey Nesa SRE Era K EE aa EATE ENEE EREE NE B 34  Keys Displays ona ea saad iee R EEEE aeaa B 44  Measurement Configuration sissiesiaicacstasstiectiineischeeaneeeneatassdacdastieede B 45  Hardware Configuration ssn tees 228 ee ENE lene atone B 51    Thermo Fisher Scientific    Contents    Communications Configuration cstepdcstcccaenetiess dectangetunctiercaneeace cs B 54  W O Configuration  sssini iad aini a Naaa B 61  Record Layout Definition      s sssssssesssssesssesssseststersesssestsrtersesesseseseeee B 66  Format Specifier for A
464. nt air quality measuring system   Page 68 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    Assessment summary    To verify the equivalence of the data between both candidates with the threshold set for the  heating of 50   respectively 58   a comparison campaign was performed at one of the  German sites  The gathered data sets  SN 1 with 58  setting  SN 2 with 50   setting  were  subjected to an equivalence test  The assessment led to the following results     1  The uncertainty between the two candidate systems is 1 53 g m    and thus is below  the permissible 2 5 ug m gt     2  The expanded uncertainty of the data corrected by the slope of 1 068 is for SN 1   58    and SN 2  50    below the permissible 25    3  The combined evaluation of the 4 original data sets of suitability testing with the addi     tional 5    data set Bornheim  summer 2013 leads as well to a compliance with the  equivalence criteria according to the guide  4     Hence the verification of the equivalence of the candidate systems with a threshold for the  rel  humidity of 58   was proved also for a German site     Therefore  the Model 5030i SHARP measuring system can alternatively as well be operated  with a relative humidity threshold of 58    in particular for any site where a significantly high  volatile content is expected     740300_2013_936_21209885F_EN docx    TUV Rheinla
465. number  representing barometric pressure alarm limits in millimeters of mercury   The following example sets the barometric pressure alarm maximum value  to 800 0 mmHg     Send  set alarm baro pres max 800  Receive  set alarm baro pres max 800 ok    alarm bench status  This command reports the current status of the bench as FAIL or OK  The  following example reports that the bench status is OK     Send  alarm bench status  Receive  alarm bench status ok    alarm beta count min  alarm beta count max    Model 50307 SHARP Instruction Manual B 15    C Link Protocol Commands  Alarms    B 16 Model 50307 SHARP Instruction Manual    These commands report the current beta count alarm minimum and  maximum value settings  The following example reports that the beta count  alarm minimum value is 5000     Send  alarm beta count min  Receive  alarm beta count min 5000    alarm board temp min   alarm board temp max   These commands report the current board temperature alarm minimum  and maximum value settings  The following example reports that the board  temperature alarm minimum value is 0   C     Send  alarm board temp min  Receive  alarm board temp min 0 0 degC    set alarm board temp min value   set alarm board temp max value   These commands set the board temperature alarm minimum and  maximum values to value  where value is a floating point number  representing board temperature alarm limits in   C  The following example  sets the board temperature alarm maximum value to 38   C     S
466. o Fisher Scientific    Optional Equipment  Mounting Options             EXTENDER BRACKET  OPTION 2111                   343    iN    g  car BAR NUT SHOULD BE  17a    ASSEMBLED WITH    OFF   CENTERED  g TAPPED HOLE AWAY  FROM INSTRUMENT AS  SHOWN        Ah f  Rgs  RACK MTG RAIL So    REAR MIG BRACKET INSTRUMENT CASE    FOR SUDES       RACK SUDE  P N 5225    Figure 9   4  Rack Mount Option Assembly    Thermo Fisher Scientific Model 5030  SHARP Instruction Manual 9 7    Optional Equipment  Roof Flange Assembly    Roof Flange An optional roof flange assembly can be used with flat roof applications  as   shown in Figure 9 5  This roof flange assembly uses 3    PVC tubing   Assembly fittings and roof flange with gasketing  The installer must assure a water   tight installation when using a flange of this design  A 2 1 2    circular hole  should be drilled through the roofline of the the enclosure  or building  if it  has a flat roof  The roof surface should be clean and swept of all debris  The  flange should be centered over the hole that has been drilled  A generous  bead of RTV or silicone sealent should be applied to the circumfernce of  the ponts shown in Figure 9 5     t    7 OD S5 5ampke hbng            Towards niet    Liquid tight compression fing                       pe a N  F          Sampl tube   ambient sensor aceway    Feed amben serso cable through here    Roof fange eet   Flange gasket    Figure 9 5  Roof Flange Assembly    9 8 Model 50307 SHARP Instruction Manual T
467. o affect the performance of the AMS  significantly     6 5 Assessment    The version of the instrument software is shown on the display  Changes to the instrument  software are communicated to the test institute     Minimum requirement fulfilled  yes    740300_2013_936_21209885F_EN docx    TUVRh        TUV Rheinland Energie und Umwelt GmbH  einland Air Pollution Control   Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 98 of 424    6 6 Detailed presentation of test results    PROGRAM VERSIONS   PRODUCT  MODEL 5830i  VERSION  62 06 06  232    FIRMWARE  arlo hzl  B AMPL   03 00 31  01 15 18       Figure 34  Software version  here V02 00 00 232   displayed by selection of the menu    Di   agnostics   Program Versions       Note     The subordinate software versions FIRMWARE  general parameters of the i series   B AMPL   beta   amplifier  and NEPH  nephelometer  listed in the software display above are fixed  main software components  Changes to the subordinate programmes FIRMWARE  general  parameters of the i series   B AMPL  beta   amplifier  and NEPH  nephelometer  are neces   sarily bound to a change in the main software     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control   s  Precisely Rig
468. o the reference method  In that event the results achieved by that meth   od must be corrected to produce results equivalent to those that would have been achieved  by using the reference method   2008 50 EC  Annex VI  B      The Guidance    Demonstration of Equivalence of Ambient Air Monitoring Methods     4   which  was developed by an ad hoc EC task group  Source     http   ec europa eu environment air quality legislation pdf equivalence pdf  Version January  2010  describes an equivalence check method for non standard measurement methods     The specifications of the Guide for the equivalence testing were included in the last revision  of the VDI Standards VDI 4202  Part 1  1  as well as VDI 4203  Part 3  2      The following limits were applied during suitability testing     PM25  Daily limit value  24 h  not defined  Annual limit value  1 a  25 ug m       as well as for the calculations according to the Guide  4   PM25  Limit value 30 g m8    Standard VDI 4202  Part 1 of 2002 describes all    Performance criteria for performance tests  of automated ambient air measuring systems     General parameters for the related tests are  given in Standard VDI 4203  Part 1    Testing of automated measuring systems   General con     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 14 of 424 Model 5030  SHARP with
469. oburyY dYvHS  sun WED    SWL VORS  Suy       740300_2013_936_21209885F_EN docx    Flow chart     Overview of menu assisted firmware    Figure 15       z D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland      gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 46 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    In addition to the possibility of direct communication via pushbuttons display  the measuring  system may be completely operated and parameterised over RS232 RSRS485 and Ethernet      both directly from a PC and indirectly via modem or network  During suitability testing  the  measuring system was accessed over Ethernet and the terminal software iPort  especially for  downloading the measurement data saved internally  The nature and scope of internal data  logging  logging interval  recorded parameters  can be configured on the    Instrument Con   trols    menu  This remote operation of the measuring system contains the same features as  direct operation     The measuring system supports C Link  MODBUS  Gesytec  Bayern Hessen   ESM   Streaming Data and NTP protocols     To perform external zero checks of the measuring system  a zero filter is mounted to the  AMS inlet  This filter ensures the supply of air free of suspended particles        Figure 16  Zero fi
470. odule  but rather use a soft cloth moistened  with a naphtha cleaning solvent  A    Do not place the LCD module near organic solvents or corrosive gases  A  Do not shake or jolt the LCD module  a    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Disconnect the ribbon cable and the two wire connector from the front    panel board   3  Remove the four screws at the corners of the LCD module   4  Slide the LCD module out towards the center of the instrument   5  Replace the LCD module by following the previous steps in reverse     Note The optimal contrast will change from one LCD screen to another   After replacing the LCD screen  the contrast may need to be reset  If the  content on the screen is visible  select Instrument Controls  gt  Screen  Contrast and adjust the screen contrast  If the content on the screen is not  visible  use the    set contrast 10    C Link command to set screen constrast to  mid range  then optimize the contrast  See the    C Link Protocol  Commands    appendix for more information on this command   amp     Model 50307 SHARP Instruction Manual 7 43    Servicing  Removing the SHARP Optics Cover    Removing the SHARP The SHARP optics cover can be removed from the SHARP optics  Refer to  P P  the following steps when a procedure requires access to the SHARP optics    Optics Cover assembly  Figure 7   24      Remove Screw          Remove Screw    Figure 7   24  Removing the SHARP Optics Cover    Equipment Required     Phi
471. of a measuring cabinet  A non stationary  application is therefore only to be undertaken together with the associated peripheral devic   es    The following steps are strictly necessary for the installation of the measuring device     Unpacking and installation of the measuring system  in a rack or on a bench   Installation of the sampling tube  dynamic heating system and PM25 sampling inlet  Connection of the pump   Fitting of the ambient temperature sensor  near the sampling inlet    Set up of all connecting and control lines   Connection of power supply   Optional  Connection of peripheral recording and control systems  data logger  PC  with iPort  to the respective ports   e Switch on of the measuring system   e Set up of filter tape    The time it takes to perform these works makes up the set up time  In this case  it amounts to  approx  1 hour     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 82 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The warm up time is the time between the start of operation of the measuring system and the  point when it is ready for measurement    After switching on the system  the system is in the equilibration phase until the BETA detec   t
472. of either alarms or non alarm to choose from     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Output Relay Settings  gt  select Relay  gt  Instrument State        Perea    Wi nist  ENE          Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 45    Operation  Volumetric Conditions    Alarms The Alarms screen allows the user to select an alarm status for the selected  relay output  The currently selected item is shown by     lt       after it  The  I O board status alarm is only present if the I O expansion board is  installed     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Output Relay Settings  gt  select Relay  gt  Instrument State  gt  Alarms        DH vided  Piet       Non Alarm The Non Alarm screen allows the user to select a non alarm status for the  selected relay output  The currently selected item is shown by     lt       after it     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration     gt  Output Relay Settings  gt  select Relay  gt  Instrument State  gt  Non   Alarm        Digital Input Settings The Digital Input Settings menu displays a list of the 16 digital inputs  available  and allows the user to select the logic state and instrument  parameter for the relay selected     Note The digital inputs must be asserted for at least one second for the  action to be activated  A    3 46 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Con
473. of test results  Not applicable     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 127 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 1 5 3 17 Increase of NO  concentration due to residence in the measuring system    In case of NOx measuring systems the increase of NOz concentration due to residence  in the measuring system shall not exceed the requirements of Table 2 of Standard VDI  4202 Part 1  September 2010     The requirements of Table 2 apply to certification ranges according to Table 1 of  Standard VDI 4202 Part 1  September 2010   For deviating certification ranges the re   guirements shall be proportionally converted     Note    This test is not relevant for particulate measuring systems   6 2 Equipment   Not applicable    6 3 Performance of test   Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 128 of 424 Model 50
474. of the zeroing are available via display and C Link read  commands     At the location where the nephelometer attaches to the 50304 instrument  case  a radial insulated tube joins the sampling tube to the collection  chamber  The particulate is drawn onto the glass fiber filter tape between  the proportional detector and the C beta source  Here  the beta beam  passes upwards through the filter tape and accumulated particulate layer   The intensity of the beta beam is attenuated with the increasing mass load  resulting in a reduced beta intensity measured by the proportional detector   From a continuously integrated count rate the mass on the filter tape is  calculated  During sampling  the mass on the filter spot is always known  and the rate of particulate accumulation is observed     A continuous measurement of the flow rate determines the sample volume   A proportional valve is used to maintain the sample flow rate at its nominal  value  From both integrated sample volume and mass detection the  particulate concentration is calculated  At any given time the output signal  delivers the concentration of particulate in actual or standard conditions     The Model 50307 outputs particulate matter concentrations to the front  panel display as SHARP  PM  and NEPH  which are the hybrid  beta  attenuation and nephelometer derived particulate concentrations   respectively  Displayed concentrations  analog outputs  also makes the data  available over the serial or Ethernet connection
475. off the slotted bottom support     4  Replace the front panel board by following the previous steps in  reverse     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 41    Servicing  LCD Module Replacement    Mounting Studs  2     Ribbon Cables  3      lt _     LCD Module  Replacement    A  A    7 42 Model 50307 SHARP Instruction Manual    LCD Module    LCD Module Screws  2   Retaining Screw        Front Panel Board       Two Wire Connector    a          Bottom Support  Figure 7 23  Replacing the Front Panel Board and the LCD Module  Use the following procedure to replace the LCD module  Figure 7   23    Equipment Required     LCD module    Philips screwdriver    CAUTION If the LCD panel breaks  do not let the liquid crystal contact  your skin or clothes  If the liquid crystal contacts your skin or clothes  wash  it off immediately using soap and water  A    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    Do not remove the LCD panel or frame from the LCD module  A    The LCD polarizing plate is very fragile  handle it carefully  a    Thermo Fisher Scientific    Thermo Fisher Scientific    Servicing  LCD Module Replacement    Do not wipe the LCD polarizing plate with a dry cloth  as it may easily  scratch the plate  A    Do not use alcohol  acetone  MEK or other Ketone based or aromatic  solvents to clean the LCD m
476. on        Verify cable connection     Verify cable connection        RH Temperature Alarms        Flow Temp    Disconnected cable    Sensor defective    Verify connection to measurement  interface board     Replace sensor        RH Temperature Alarms        Board Temp    Disconnected cable    Sensor defective    Verify connection to measurement  interface board     Replace sensor        Thermo Fisher Scientific    Thermo Fisher Scientific    Alarm Message    Pressure Vacuum Alarms      Barometer Pressure    Possible Cause    Disconnected cable    Troubleshooting  Troubleshooting Guides    Action    Verify cable connections   Replace board        Pressure Vacuum Alarms      Vacuum    Disconnected plumbing    Disconnected cable    Sensor defective    Verify plumbing connections     Verify cable connection     Replace pressure board        Pressure Vacuum Alarms      Flow    Disconnected plumbing    Disconnected cable  Sensor defective    AP valve open    Verify plumbing connections    Verify cable connection   Replace pressure board     Verify valve is in close position        Flow Alarms     Flow    Low flow    No flow    Check that pump is connected   Verify that both AP valves are in the  closed position    Verify proportional valve plumbing     Verify connection to measurement  interface board        Conc Alarms     PM    Conc negative    Conc alarm setting    Check heater and Beta counts     Verify conc alarm settings        Conc Alarms     AVG SHARP    Conc negative    Co
477. on Codes    C 6 Model 50307 SHARP Instruction Manual    Request   Function Code  Starting Address  Quantity of Registers    Response  Function Code  Byte Count    Register Value     N   Quantity of Registers    Error Response  Function Code    Exception Code    1 Byte  2 Bytes  2 Bytes    1 Byte  1 Byte  N  x 2 Bytes    1 Byte  1 Byte    0x03 or 0x04    0x0000 to maximum allowed by instrument    1 to maximum allowed by instrument    0x03 or 0x04  2 x N   N N or N     Function code   0x80    01 IIlegal Function  02 IIlegal Address     03 Illegal Data  04 Slave Device Failure    Here is an example of a request and response to read registers 10 13     Request   Field Name  Function   Starting Address Hi  Starting Address Lo  No  of Registers Hi    No  of Registers Lo    Response   Field Name   Function   Byte Count   Register Value Hi  10   Register Value Lo  10   Register Value Hi  11   Register Value Lo  11      Hex   0x03  0x00  0x09  0x00  0x04     Hex   0x03  0x06  0x02  0x2B  0x00  0x00    Thermo Fisher Scientific     0x05  Force  Write     Thermo Fisher Scientific    Single Coil    MODBUS Protocol  Function Codes    Register Value Hi  12  0x00  Register Value Lo  12  0x64  Register Value Hi  13  0x00  Register Value Lo  13  0x64    The contents of register 10 are shown as the two byte values of 0x02 0x2B   Then contents of registers 11   13 are 0x00 0x00  0x00 0x64 and 0x00  0x64  respectively     The Force  Write  Single Coil function simulates the activation of the  dig
478. on Ports           A PB          e Inthe Main Menu  choose Service  gt  Pres  Vacuum Calibration  gt   Vac Flow  gt  Vac Pres Span        The Flow Pressure Span screen allows the user to view and set the flow  sensor calibration span point     To calibrate the flow pressure sensor span point  the instrument must be  pulling an active flow through the filter tape  By use of a manometer   connect the respective     ports of your NIST traceable manometer to the  respective rear panel     ports   Please be certain to zero any digital  manometers prior to connecting   Open the ports by pressing the toggle  switches inward  Calculate an average reading from the reference  manometer in units of mmHg  Enter this span value into the screen below     Thermo Fisher Scientific    Flow Calibration    Auto Flow Calibration    Mass Calibration    Thermo Fisher Scientific    Calibration  Calibration Procedure    e Inthe Main Menu  choose Service  gt  Pres Vacuum Calibration  gt   Vac Flow  gt  Flow Pres Span        Wi hid  A a E PEPE          Be sure to save the entry and compare the values once more  Repeat as  necessary to within 2 mmHg     The preferred method of flow rate calibration is the Auto Flow Calibration     Place a NIST traceable volumetric flow meter on top of the instrument  after complete installation  reassembled heater tube  sample tubes  and  inlets   Allow approximately 1 minute for the flow to stabilize  Once stable   measure and average three individual readings from th
479. on at zero point   exceed the requirements of  1  and 0 07 g m  for system 2  SN   Table 2 in the certification   2    range according to Table 1 of  Standard VDI 4202 Part 1   September 2010      For PM  Max  Bo     5 3 3 Repeatability The repeatability standard de    Not applicable   standard devia    viation at reference point shall  tion at reference   not exceed the requirements of  point Table 2 in the certification  range according to Table 1 of  Standard VDI 4202 Part 1   September 2010      5 3 4 Linearity  lack of   The analytical function describ    For PM2 5 particulate measuring sys   fit  ing the relationship between   tems  this test shall be performed ac   the output signal and the value   cording to minimum requirement  of the air quality characteristic   5 4 10    Calculation of the expanded  shall be linear  uncertainty of the instruments           740300_2013_936_21209885F_EN docx    A TUVRheinland      Precisely Right     Page 22 of 424    5 3 5 Sensitivity coeffi   cient of sample  gas pressure    5 3 6 Sensitivity coeffi   cient of sample  gas temperature    Sensitivity coeffi   cient of sur   rounding tem   perature    Sensitivity coeffi   cient of supply  voltage    5 3 9 Cross sensitivity    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate
480. on page B 44   e    Measurement Configuration    on page B 45   e    Hardware Configuration    on page B 51   e    Communications Configuration    on page B 54  e    I O Configuration    on page B 61   e    Record Layout Definition    on page B 66    Each command sent to the analyzer over the serial port must begin with the  American Standard Code for Information Interchange  ASCII  symbol or  byte value equivalent to the instrument   s identification number plus 128     Model 50307 SHARP Instruction Manual B 1    C Link Protocol Commands  Commands    Commands    B 2 Model 50307 SHARP Instruction Manual    For example  if the instrument ID is set to 25  then each command must  begin with the ACSII character code 153 decimal  The analyzer ignores any  command that does not begin with its instrument identification number  If  the instrument ID is set to 0  then this byte is not required  For more  information on changing Instrument ID  see Chapter 3     Operation        The analyzer must be in the remote mode and service mode not active in  order to change instrument parameters remotely  However  the command     set mode remote    can be sent to the analyzer to put it in the remote mode   Report commands  commands that don   t begin with    set     can be issued  either in the remote or local mode  regardless of the service mode setting   For information on changing modes  see Chapter 3     Operation        The commands are not case sensitive  Each command must begin with the  
481. on panel  can be lowered to improve access to connectors and components  Refer to  Measurement Case the following steps when a procedure requires lowering the partition panel  Assembly and Lowering  gure 7 5    the Partition Panel    REMOVE SCREWS Remove SHARP Optics    Pass Through Cover Assembly  Connectors             Loosen  Captive  Screw        Remove Screws  Remove Screw        gt      Figure 7   5  Removing the Measurement Case Assembly and Lowering the  Partition Panel    Equipment Required     Philips screwdriver    7 12 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Servicing  Removing the Measurement Case Assembly and Lowering the Partition Panel    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap    must be worn while handling any internal component  A    10     11     Turn the instrument OFF and unplug the power cord   If the instrument is mounted in a rack  remove it from the rack     Remove the SHARP optics cover assembly  then proceed to the next  step below  Refer to    Removing the SHARP Optics Assembly     described previously     Disconnect the plumbing connections at the rear of the measurement  case assembly     Disconnect the connectors that pass through the center of the partition  panel     Remove two screws from the left side rear of the case  viewed from the  front      Remove one screw from the bottom rear of th
482. onditioning 24 h  2 h  Filter weighing and immediate packaging Filter weighing       The balance is always ready for use  An internal calibration process is started prior to each  weighing series  The standard weight of 200 mg is weighed as reference and the boundary  conditions are noted down if nothing out of the ordinary results from the calibration process   Deviations of prior weighings conform to the Standard and do not exceed 20 pg  refer to Fig   ure 61   All six control filters are weighed afterwards and a warning is displayed for control fil   ters with deviations  gt  40 ug during evaluation  These control filters are not used for post   weighing  Instead  the first three acceptable control filters are used while the others remain in  the protective jar in order to replace a defective or deviating filter  if necessary  Figure 62  shows an exemplary process over a period of more than four months     All filters which display a difference of more than 40 ug between the first and second weigh     ing are excluded from the pre weighing process  Filters exhibiting deviations of more than 60  ug are not considered for evaluation after post weighing  as conforming to standards     740300_2013_936_21209885F_EN docx       TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung      TUVRheinland      os ae Genau  Richtig   Report on the suitability test of the ambient air quality measuring system    Model 5030  SHARP with PMz5 pre separator of the company Thermo  Fisher Sci
483. one record type will automatically change  the other     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 33    Operation  Volumetric Conditions    e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Configure Datalogging  gt  Memory Allocation           te    ERED CHa  perm    Wi mid  PS PHP ELSES Pi AG Lf bie       uo pH EE  at a E a E        Data Treatment The Data Treatment screen is used to select the data type for the selected  record type  whether the data should be averaged over the interval  the  minimum or maximum measured during the interval  or the current value   last value measured   Data treatment doesn   t apply to all data  just to the  concentration measurement  All other data points log the current value at  the end of the interval     Note This feature is found in all   Series instruments  but it is  recommended that the data type be set to ONLY the current value  cur   as  the datalogging averaging is done in addition to the normal concentration  averaging  A    e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Configure Datalogging  gt  Data Treatment        DH Guided  PLP          Flag Status Data The Flag Status Data screen is used to set the flag status data to either ON  or OFF for the selected record type     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Configure Datalogging  gt  Flag Status Data     3 34 Model 50307 SHARP Instruction Manual Thermo 
484. onised hybrid real time measuring  instrument for particulate matter which determines the particulate concentration using a  combined nephelometric radiometric method of measurement  With the aid of a pump  ambi   ent air is sucked at 16 67 l min via a PM25 pre separator  The dust laden sampling air passes  through a light scattering photometer  nephelometer  and is then collected on a filter tape   The determination of particulate mass collected on the filter tape is performed continuously  during sampling by means of the radiometric measuring principle of beta absorption  The ac   tual reading of the measuring system Model 5030i SHARP is determined from the values    measured by the nephelometer and radiometer by means of a calculation algorithm     The tests took place in the laboratory and for several months in the field     The following test sites were chosen for the field test according to Table 1                 Table 1  Description of the test sites  Cologne  Bornheim  Bornheim  Teddington  UK   parking lot  motorway parking motorway parking summer  winter lot  lot   summer winter  Period 03 2011     05 2011 07 2011     11 2011 01 2012     03 2012   07 2012     10 2012  No  of paired values   Candidates 52 91 49 50  Characteristics Urban background Rural structure   Rural structure   Urban background  motorway motorway  Rank of pollution Average to high Low to average Average Average                      To qualify the necessary change of the threshold for the relati
485. only when the instrument is in service mode  For more information on the  service mode  see    Service Mode    earlier in the chapter     Note This adjustment should only be performed by an instrument service  technician  A    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 97    Operation  Service Menu    Nephelometer RH    Nephelometer  Temperature    Nephelometer Source  Level    3 98 Model 50307 SHARP Instruction Manual    In the Main Menu  choose Service  gt  Nephelometer Calibration        The Nephelometer RH screen allows the user to view and calibrate the    nephelometer relative humidity     In the Main Menu  choose Service  gt  Neph Calibration  gt  Neph RH        The Nephelometer Temperature screen allows the user to view and    calibrate the nephelometer temperature sensor calibration     In the Main Menu  choose Service  gt  Neph Calibration  gt  Neph    Temp                 EHO guided  Piet    The Nephelometer Source Level screen allows the user to view and    calibrate the source level     In the Main Menu  choose Service  gt  Neph Calibration  gt  Neph Src    Level     Thermo Fisher Scientific    Analog Output Calibration    Analog Output Calibrate    Thermo Fisher Scientific    Zero    Operation  Service Menu       The Analog Output Calibration menu provides access to the 6 voltage  channels and 6 current channels for calibration  Current channels are  visible only if the I O expansion board is installed  The analog output  calibration menu is
486. ons  configuring the inter   nal data logging and communication settings  adjusting the screen contrast  turning the ser   vice mode on and off  wrench icon appears on the main screen when service mode is on  as  well as setting up the date and time     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 44 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The    Diagnostics    sub menu provides access to diagnostic information and functions  e g   software version  current reading of the different sensors  radiometric detector and nephe   lometer status  status of the analogue and digital inputs and outputs  etc  This information is  especially useful when troubleshooting the instrument     The    Alarms    sub menu displays the status of the measuring system with regard to potential   ly upcoming alarm messages  The bell icon on the main screen indicates that one  or more   alarm s  are active     The    Service    sub menu  available only when the service mode is activated on the    Instru   ment Controls    sub menu  gt  if on  the wrench icon appears on the main screen  provides ac   cess to advanced diagnostic functions  These include  for example  the adjustment of 
487. or has adjusted to the ambient conditions  This phase has a duration of at least 12 hours  Af     ter the equilibration phase  the following actions are necessary for starting operation the first  time     e Check of system settings with regard to flow  heating  filter tape parameters  record   ing of measured values  date and time   e Check and  if required  adjustment of the temperature sensors  ambient temperature   nephelometer temperature  flow temperature     e Check and  if required  adjustment of rel  air humidity of the nephelometer   e Check and  if required  adjustment of the ambient air pressure measurement   e Check and  if required  adjustment of the flow rate   e Leak test   e Check and  if required  adjustment of the nephelometer   s zero point with a HEPA filter  at the inlet   e  f necessary  check of the radiometric mass determination     calibration with mass  foils    Duration  approx  1 hour    In case of restarting operation after a short downtime  e g  after a power outage  the above  mentioned steps can be omitted with exception of a check of the system parameters  a plau   sibility test of the sensor values and an inspection of possible status error messages     If required  potential changes in the basic parameterisation of the measuring instrument can  likewise be performed in just a few minutes by staff familiarised with the instruments     6 5 Assessment    The set up and warm up times were determined     The measuring system can be operated at d
488. or of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5     Report No   936 21209885 F    Table 6     Ambient conditions at the field test sites  daily mean values    TUVRheinland      Precisely Right     Page 61 of 424                                  Cologne  Bornheim  Bornheim  Teddington  UK    parking lot  motorway parking   motorway parking summer  winter lot  summer lot  winter   No  of paired values 48 87 41 65   reference   Ratio PM25 PMio      Range 38 8     86 3 27 1     88 1 43 1  91 1 36 2     70 5  Average 61 7 57 3 73 0 53 9  Ratio Volatiles PM2 5       Range 11 0     44 8 0 0   53 3 10 4     73 2 10 8     70 8  Average 26 2 21 6 25 2 34 3  Air temperature    C   Range 5 0     22 1 5 2     24 5  8 9     10 4 8 9     24 2  Average 12 3 15 5 1 9 16 4  Air pressure  hPa   Range 998     1031 995     1024 1003     1031 971     1010  Average 1012 1008 1018 996  Rel  humidity      Range 34 7     90 4 53 8     91 1 50 2     90 5 47 1     94 6  Average 61 8 75 1 77 7 72 7  Wind velocity  m s   Range 0 3 5 3 0 3     4 1 0 6  5 7 0 1 1 9  Average 1 9 1 4 2 1 0 5  Precipitation  mm d   Range 0 0  11 1 0 0   51 7 0 0     12 8 0 0     30 0  Average 0 9 2 7 1 5 1 6                        The content of volatiles was measured with a TEOM FDMS system operated simultaneously    740300_2013_936_21209885F_EN docx    TUVR h 7      TUV Rheinland Energie und Umwelt GmbH   einland Air Pollution Control  Precisely Right    Report on the suitabil
489. or the full dataset is referred to for this   i e  18    which again would require an annual test at 4 measurement sites  Guide  4   chap   ter 9 9 2  Table 6      740300_2013_936_21209885F_EN docx    a 3    TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 156 of 424    6 6 Detailed presentation of test results    Table 36 presents an overview on the uncertainties between the reference devices Ue from  the field tests  In Table 37  the results of the equivalence test including the determined ex   panded measuring uncertainties Woy from the field test are summarised     Table 36  Uncertainty between the reference devices Ure for PM25    Reference Test site Amount Uncertainty Ups  eo values  Ooo T O o    1     2 Cologne  winter 48 0 58       The uncertainty between the reference devices Ure is  lt  2 g m  for all test sites     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH      TUVRheinland      Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 157 of 424  Scientific for the component suspended particulate ma
490. orce  Write  Single Coil   0x05  Read Exception Status   0x06    If a function code is received that is not in this list  an invalid function  exception is returned     The data field varies depending on the function  For more description of  these data fields  see    Function Codes    below     In MODBUS over serial  an error check is included in the message  This is  not necessary in MODBUS over TCP IP because the higher level protocols  ensure error free transmission  The error check is a two byte  16 bit  CRC   value     This section describes the various function codes that are supported by the    Model 5030i     Read Coils Inputs read the status of the digital outputs  relays  in the  instrument  Issuing either of these function codes will generate the same  response     These requests specify the starting address  i e  the address of the first  output specified  and the number of outputs  The outputs are addressed  starting at zero  Therefore  outputs numbered 1   16 are addressed as 0 15     The outputs in the response message are packed as one per bit of the data  field  Status is indicated as 1   Active  on  and 0   Inactive  off   The LSB  of the first data byte contains the output addressed in the query  The other    Model 50307 SHARP Instruction Manual C 3    MODBUS Protocol  Function Codes    C 4 Model 50307 SHARP Instruction Manual    outputs follow toward the high end of this byte  and from low order to  high order in subsequent bytes  If the returned output q
491. ot transmitted     Exponent  sign and 2 digits    Example     5384000 0 is represented as  5384 06   0 04567 is represented as  4567 02     lt status gt   is formed by  lt  operating status  gt  and  lt  error status  gt  and  separated by a space  i e       lt operating status gt  lt SP gt  lt error status gt     Each of the two   lt operating status gt  and  lt error status gt   are formed by two  characters  each representing a 2 digit hex number which is one byte  8  Bits  operation status and one byte  8 Bits  error status     These two bytes contain the information about the main operating  conditions of the instrument at that instant  For details on how to interpret  the status bytes  refer to Table D 1 and Table D 2 below      lt SFKT gt   is the space provided for future use for special function  It  currently contains a string of ten 0   s  i e   lt 0000000000 gt      Model 50307 SHARP Instruction Manual D 5    Gesytec  Bayern Hessen  Protocol  Gesytec Commands    D 6 Model 50307 SHARP Instruction Manual    The Gesytec serial number defaults to zero  To set the Gesytec serial  number select Main Menu  gt  Instrument Controls  gt  Communication    Settings  gt  Gesytec Serial No     Example of response to DA command from an instrument with Gesytec  serial number set to 000  The Gesytec serial number is bold in the  example     Gesytec Protocol with transmission of three concentrations  Instrument ID  is 1  Operation Status is 03  Error Status is 04      Data Query St
492. otocol server   See    Network Time Protocol Server    in the Communications Settings     section of the    Operation    chapter for more information     Send  addr ntp  Receive  addr ntp 10 209 43 237  set addr ntp address    This command sets the network time protocol server address  where address      9    consists of four numbers ranging from 0 255 inclusive  separated by             Send  set addr ntp 10 209 43 237  Receive  set addr ntp 10 209 43 237 ok    allow mode cmd   This command reports the current allow mode setting  1   allow    set mode  local    and    set mode remote    commands  0   ignore    set mode local    or     set mode remote   commands  according to Table B   7  The default value is  0  ignore the commands  The following example shows that the instrument  is configured to ignore    set mode local    or    set mode remote    commands     Send  allow mode cmd  Receive  allow mode cmd O    Model 50307 SHARP Instruction Manual B 55    C Link Protocol Commands  Communications Configuration    B 56 Model 50307 SHARP Instruction Manual    set allow mode cmd value   This command is used to configure the instrument to value  where value is  either 1   accept or 0   ignore the    set mode local    and    set mode remote     commands  according to Table B   7     If the instrument is set to accept the commands  value   1   the    set mode  local    command will unlock the instrument and the keypad can be used to  make changes via the front panel     If the ins
493. ow span B 37  cal vac flow offset Automatically triggers the filter change and sets the vacuum B 38  and flow offset  cal vac pres span Automatically sets the vacuum pressure span B 38  clr recs Clears away only Irecs that have been saved B 27  clr records Clears away all logging records that have been saved B 27  clr srecs Clears away only srecs that have been saved B 27  coef pm Reports sets current PM coefficient B 39  coef sharp Reports sets current SHARP coefficient B 38  conc unit Reports sets current concentration units B 45  contrast Reports sets current screen contrast B 51       Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands                                                                                  Commands  Command Description Page  copy Irec to sp Sets copies current Irec selection into the scratch pad B 28  copy sp to Irec Sets copies current selections in scratch pad into Irec list B 27  copy sp to srec Sets copies current selections in scratch pad into srec list B 27  copy sp to stream Sets copies current selections in scratch pad into stream list B 27  copy srec to sp Sets copies current srec selection into the scratch pad B 28  copy stream to sp Sets copies current streaming data selection into the scratch B 28  pad  crn Reports the current radon concentration B 24  custom Reports sets defined custom range concentration B 46  data treatment Irec  Reports sets data treatment for concentration values in lrecs B 28  data t
494. ow you to remove any deposits within the sample  tube  This same procedure also applies to the heater tube     The    Installation    chapter covers detailed installation instructions  After  proper installation  it is recommended to check the weather proof  installation at all interfaces exposed to ambient conditions  Ensure that the  condition of the roof flange and silicone caulking will prevent any  precipitation from entering the shelter and possibly damaging the  instrument s electronics     Use the following procedure to inspect and clean the fan filter  Figure 5     4      1  Remove the fan guard from the fan and remove the filter     2  Flush the filter with warm water and let dry  a clean  oil free purge will  help the drying process  or blow the filter clean with compressed air     3  Re install the filter and fan guard     Model 50307 SHARP Instruction Manual 5 5    Preventive Maintenance  Pump Rebuilding    A    REFER TO STICKER  AIRFLOW   FOR POSITIONING OF FAN    Pump Rebuilding    Leak Test    5 6 Model 50307 SHARP Instruction Manual    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  For more  information about appropriate safety precautions  see the    Servicing     chapter  A       ABLE INSERTED THIS WAY       EXTERIOR SIDE OF REAR PANEL    Mounting Screws  4     7 Filter     S    Fan Guard       Figure 5   4  Insp
495. ows the user to view and set the flow alarm  parameters     e Inthe Main Menu  choose Alarms  gt  Flow Alarms     3 78 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu          Flow The Flow screen displays the current flow and sets the minimum and  maximum alarm limits  Acceptable alarm limits range from 16 00 to 17 34  LPM  If the flow reading goes beyond the minimum or maximum limit  an  alarm is activated and the alarm  bell  icon appears in the status bar on the  Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  Flow Alarms  gt  Flow        Min and Max Flow Limits The Minimum Flow alarm limit screen is used to change the minimum    flow alarm limit  The minimum and maximum flow screens function the  same way     e Inthe Main Menu  choose Alarms  gt  Flow Alarms  gt  Flow  gt  Min or    Max        mye WH rnd    Saal wows  EE fee PEE aE E          Concentration Alarms The Concentration Alarms submenu allows the user to view and set the 24     hour average concentration and the instant concentration for PM  SHARP   and nephelometer alarm parameters     e Inthe Main Menu  choose Alarms  gt  Conc Alarms     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 79    Operation  Alarms Menu       Average PM The Average PM screen displays the current average PM concentration and  sets the minimum and maximum alarm limits  Acceptable alarm limits  range from 0 to 10000 pg m  or 0 to 10 mg m     If the averag
496. pendence of the measured value at zero concentration shall not  exceed the reference value Bo if ambient temperature is changed by 15 K in the range  of  5    to  20    or by 20 K in the range of  20    to  40        The temperature dependence of the measured value in the range of the reference val   ue B  shall not exceed 5   of the measured value if ambient temperature is changed  by 15 K in the range of  5    and  20    or by 20 K in the range of  20   C and  40         6 2 Equipment    Climatic chamber for a temperature range of  5 to  40   C  zero filter for testing the zero point  and span foil set for testing the span point     6 3 Performance of test    In order to test the dependence of zero point and measured values on the surrounding tem   perature  the complete measuring systems were operated in a climatic chamber     For the zero point test zero filters were installed at the inlets of both measuring systems SN 1  and SN 2  to which particle free sampling air was applied     The reference point test comprised a check of the mass coefficient in order to test the stabil   ity of the sensitivity of both candidates SN 1 and SN 2     It is not possible to produce a reproducible signal for the sensitivity of the nephelometer us   ing external test standards  Since the nephelometer is continuously calibrated with a very  stable radiometric measurement  a separate examination of the stability of the nephelome   ters signal at the reference point under operating conditions
497. peration  lowers the vacuum chamber plate   advances the filter tape a fixed length  raises the vacuum chamber plate  and  initiates the pump operation and a zeroing of the sample spot  Immediately  after a filter change  a new measurement cycle is initiated with an  automatic zero adjustment of the mass signal  The filter change interval and  time of day can be user adjusted  Additionally  automatic filter changes will  occur if the mass on the filter exceeds 1 500 pig since the last filter change  or if the flow rate through the filter tape is reduced by more than 5  due  to potentially restrictive particle deposition     Using the example of PM25 monitoring  ambient air is drawn into a 10  micron  pm  size selective inlet which discriminates against 50  of the  particles with an equivalent aerodynamic diameter above 10 pm  This inlet  has been approved previously through prior U S  EPA designations  RFPS   0699 130  131  and 132  and is also listed within 40 CFR Part 50   Appendix L as a 10 MICRON INLET  including model SA246b   Further  particle discrimination is achieved by inserting a PM2  cyclone between the  10 micron inlet and sample tube  European style inlets are also available  with the 5030    The volumetric flow rate at the inlet entrance is 16 67 liters  per minute  1 m  hour   In addition  the ambient temperature and relative  humidity is measured via a naturally aspirated radiation shield and standard  sensor as part of the volumetric flow control and dynamic
498. pgrade Procedures    describes  the 5014 to 50302 upgrade procedures     Appendix A    Warranty    is a copy of the warranty statement     Model 50307 SHARP Instruction Manual    About This Manual  Safety    Safety    Safety and Equipment    Damage Alerts    Model 50307 SHARP Instruction Manual    e Appendix B    C Link Protocol Commands    provides a description of  the C Link protocol commands that can be used to remotely control  an analyzer using a host device such as a PC or datalogger     e Appendix C    MODBUS Protocol    provides a description of the  MODBUS Protocol Interface and is supported both over RS   232 485  RTU protocol  as well as TCP IP over Ethernet     e Appendix D    Gesytec  Bayern Hessen  Protocol    provides a  description of the Gesytec  Bayern Hessen or BH  Protocol Interface  and is supported both over RS 232 485 as well as TCP IP over  Ethernet     e Appendix E    ESM Protocol Commands    provides a description of  the ESM Protocol commands  from the prior 5030 platform  that  can be used to remotely control an analyzer using a host device such  as a PC or datalogger     Review the following safety information carefully before using the analyzer   This manual provides specific information on how to operate the analyzer   however  if the analyzer is used in a manner not specified by the   manufacturer  the protection provided by the equipment may be impaired     This manual contains important information to alert you to potential safety  hazards a
499. pha count alarm maximum value B 14  max   alarm alpha count Reports alpha count alarm minimum value B 14  min   alarm amb rh max Reports sets ambient relative humidity alarm maximum value B 14       Model 5030  SHARP Instruction Manual B 3    C Link Protocol Commands  Commands    B 4 Model 50307 SHARP Instruction Manual                                                                      Command Description Page  alarm amb rh min Reports sets ambient relative humidity alarm minimum value B 14  alarm amb temp Reports sets ambient temperature alarm maximum value B 15  max   alarm amb temp Reports sets ambient temperature alarm minimum value B 15  min   alarm baro pres max Reports sets barometric pressure alarm maximum value B 15  alarm baro pres min Reports sets barometric pressure alarm minimum value B 15  alarm bench status Reports the bench status B 15  alarm beta count Reports beta count alarm maximum value B 15  max   alarm beta count Reports beta count alarm minimum value B 15  min   alarm board temp Reports sets board temperature alarm maximum value B 16  max   alarm board temp Reports sets board temperature alarm minimum value B 16  min   alarm conc avg Reports sets average nephelometer concentration alarm B 16  neph max maximum value   alarm conc avg Reports sets average nephelometer concentration alarm B 16  neph min minimum value   alarm conc avg pm Reports sets average PM concentration alarm maximum B 17  max value   alarm conc avg pm Reports sets average PM concent
500. pied in the January 2010 version  It is the opinion of TUV Rheinland and their UK partners  that the Model 5030i SHARP measuring system for PM25 is indeed being penalised by the  mathematics for being accurate  It is proposed that the same pragmatic approach is taken  here that was previously undertaken in earlier studies     Therefore  according to Table 35  the slope offset should be corrected for PM25 due to its  significance  Nonetheless it should be noted that the determined uncertainties Won lie below  the specified expanded relative uncertainty Waqo of 25   for particulate matter for all datasets  considered without application of correction factors     In this particular case the slope for the complete dataset comparison is 1 068  Therefore  un   der point 6 1 5 4 11 Application of correction factors and terms  an additional evalua   tion using the respective calibration factor for the datasets is carried out     The intercept for the complete dataset comparison is  0 103     The reworked version of the Guide of January 2010 requires that when operating in net   works  a candidate method needs to be tested annually at a number of sites and that the  number of the instruments to be tested is dependent on the expanded measurement uncer   tainty of the device  The respective realisation is the responsibility of the network operator or  of the responsible authority of the member state  However  TUV Rheinland and their UK  partners recommend that the expanded uncertainty f
501. port No   936 21209885 F    Blank page    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control   s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 13 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    1  Summary and proposal for declaration of suitability    1 1 Summary   According to Directive 2008 50 EC of 21 May 2008  7  on    Air quality and cleaner air for Eu   rope     replaces Council Directive 96 62 EC of 27 September 1996 on ambient air quality as   sessment and management including its daughter directives 1999 30 EC  2000 69 EC   2002 3 EC and the Council Decision 97 101 EC   the methods described in Standard EN  12341    Air quality   Determination of the PM   fraction of suspended particulate matter  Ref   erence method and field test procedure to demonstrate reference equivalence of measure   ment methods    and Standard EN 14907    Ambient air quality   Standard gravimetric meas   urement method for the determination of the PM25 mass fraction of suspended particulate  matter    serve as reference methods for suspended particle measurement of the respective  mass fraction  However  EC member states are free to use any other method in the case of  particulate matter for which the Member State concerned can demonstrate displays a con   sistent relationship t
502. proper instrument identification number  ASCII  character  The command  in the following example begins with the ASCII character code 153  decimal  which directs the command to the Model 50303  and is  terminated by a carriage return    CR     ASCII character code 13 decimal         Many of the commands have two forms  One form reads parameter from  the instrument   s memory  and the other writes  or updates  a parameter   The syntax for a write command adds the word    set    in front of the  command and provides an argument  Command responses are generally  echoed with a data element appended     Note If the Service Mode is active  C Link    set    commands are not  allowed  This is to prevent parameters from being changed remotely while  the unit is being serviced locally  A    If an incorrect command is sent  an error message is generated  The list of  error responses is shown in Table B   1  The following example sends the  incorrect command    set unit mg m3    instead of the correct command    set  conc unit mg m3        Send  set unit mg m3  Receive  set unit mg m3 bad cmd    Thermo Fisher Scientific    Commands List    Thermo Fisher Scientific    C Link Protocol Commands    Table B 1  Error Response Messages    Command Response    Description    Commands             bad cmd Command is not recognized   too high Supplied value is higher than the upper limit   too low Supplied value is lower than the lower limit   invalid string Supplied string invalid  typically becaus
503. put channels to zero as  necessary     6  Continue with the    Calibrating the Input Channels to Full Scale     procedure that follows     Model 50307 SHARP Instruction Manual 7 25    Servicing  Themistor Replacement    7 26    Calibrating the Input  Channels to Full Scale    Themistor  Replacement    A    Model 50307 SHARP Instruction Manual    Use the following procedure to calibrate the input channels to full scale by  applying a known voltage to the channels     Equipment Required     DC voltage source  greater than 0 volts and less than 10 volts     Connect the known DC voltage source to the input channel  1 8  to be  calibrated  Figure 7   11 shows the analog input pins and Table 7   5  identifies the associated channels     From the Main Menu  choose Service  gt  Analog Input Cal   The Analog Input Cal screen displays input channels 1 8       At the Analog Input Cal screen  press  _ 4 _  to scroll to the channel    selected in Step 1  and press        Press     to scroll to Calibrate Full scale  and press           The screen displays the input voltage for the selected channel     Use        gt  Jand   t     to enter the source voltage  and press    e Jto calibrate the input voltage for the selected channel connected  to the source voltage        Press        gt       J to return to the input channels display and repeat  Steps 3 5 to calibrate other input channels to the source voltage as  necessary     Use the following procedure to replace the optional ambient temper
504. r  change    Reel nuts not tight    No more filter tape    Filter not properly adhered or  broken    Filter tape transport drive  motor defective    Optical tape counter defective    Tighten reel nuts     Replace with new filter tape     Adhere filter tape to take up reel  and wind over itself once     Replace motor     Replace optical tape transport  sensor        Pressure sensor s   board do not hold  calibration    Valves closed    Loose plumbing    Pressure board defective    Verify AP valves on rear panel are  in the closed position     Verify all plumbing connections     Replace pressure printed circuit  board        Highly variable  concentration    Recorder noise  Sample concentration varying  Foreign material in optical    bench    Digital electronics defective    Replace or repair recorder and or  leads     Run Model 5030  on a zero filter      if quiet  there is no malfunction     Clean optical bench     Replace board with a spare  board        Instrument does not  calibrate properly    System leak  Pressure or temperture out of    calibration    Digital electronics defective    Flow    Find and repair leak     Re calibrate pressure or  temperature     Replace one board at a time with  a spare board to isolate the  defective board     Verify AP valves on rear panel are  in the closed position     Re calibrate        Display is off    6 2 Model 50307 SHARP Instruction Manual    Wrong contrast setting    Adjust contrast setting     Thermo Fisher Scientific    Trouble
505. r Measurements  and Analog Inputs  if  the I O expansion board option is installed      e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Streaming Data Config        pranih pruni Ph dm PP  PPP LES ited Beh P    WE mid       Streaming Data Interval The Streaming Data Interval screen is used to adjust how frequently a new  record will be generated  The following interval times are available  1  2  5   10  20  30  60  90  120  180  240  and 300 seconds     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Streaming Data Config  gt  Interval     Pra    Bs EEE TE Sanat Bone       Note Add Labels  Prepend Timestamp  and Add Flags are toggle items that  change between yes or no when selected  A    Choose Stream Data The Choose Stream Data screen displays a list of the types of data that can  be sent via streaming data  Choices are Concentrations  Other  Measurements  and Analog Inputs  if the I O expansion board is installed      e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Streaming Data Config  gt  Item 1 8     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 39    Operation  Volumetric Conditions       Concentrations The Concentrations screen allows the user to assign one of the  concentrations to the selected streaming data item  The currently selected    item is shown by     lt       after it  Once an item is selected  pressing  will save the selected streaming data it
506. r Variable   91  amp  92 EXT ALARMS   93  amp  94 NEPH TEMP   95  amp  96 NEPH RH   97  amp  98 NEPH IRED   99  amp  100 NEPH REF   101  amp  102 BETA REF   103  amp  104 NOT USED   105  amp  106 COMMON FLAGS  107  amp  108 BETA DETECTOR FLAGS  109  amp  110 NOT USED   111   amp  112 COMP MASS  113  amp  114 NOT USED   115  amp  116 AVG NEPH          IMPORTANT NOTE The addresses in the following tables are Protocol  Data Unit  PDU  addresses  Verify the coil number on your MODBUS    master to ensure that it matches the coil number on the instrument  A    Note Writing 1 to the coil number shown in the following table will  initiate the    action triggered    listed in the table  This state must be held for  at least 1 second to ensure the instrument detects the change and triggers  the appropriate action  A    Note The coils within each coil group in the following table are mutually  exclusive and will not be triggered if there is conflict  Before you assert  1   one coil in a group  make sure the other coils in the group are de asserted     0   4    Table C 3  Write Coils for 50307                   Coil Number Action Triggered  100 INVALID   101 NOT USED   102 NOT USED   103 NOT USED   104 AOUTS TO ZERO  105 AOUTS TO FS          Model 50307 SHARP Instruction Manual C 11    MODBUS Protocol  MODBUS Parameters Supported    C 12 Model 50307 SHARP Instruction Manual                                     Coil Number Action Triggered  106 FILTER TAPE COUNT RESET TO ZERO  107 FILTER T
507. r period value   This command sets the number of hours to be added to filter next time for  the filter tape change to value  where value is a floating point representation  of the filter period in hours  The following example sets the filter period to  10 hours     Send  set filter period 10  Receive  set filter period 10 ok    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Measurement Configuration    filter time  This command reports the next date and time of the filter tape change  The  following example reports that the date and time is January 1  2008  6 12    pm     Send  filter time  Receive  filter time 01Jano8 18 12    set filter time dd mm yyyy hh mm ss  dd   day  mm   month    yy   year  hh   hours    mm   minutes  ss   seconds    This command sets the next filter tape change  The following example sets  the filter time to October 1  2008  11 32 pm     Send  set filter time 10 01 2008 23 32  Receive  set filter time 10 01 2008 23 32 ok  flow target    This command reports the current flow  The following example reports  that the flow is 16 0 LPM     Send  flow target  Receive  flow target 16 00    set flow target value   This command sets the flow to value  where value is a floating point  number representation of flow between 1 to 20 LPM  The following  example sets the flow to 16 67 LPM     Send  set flow target 16 67  Receive  set flow target 16 67 ok  ht control    This command reports the current status of the heater  off  RH
508. r to    External Device Connection Components    on page 7 8     Each kit consists of    e one six foot cable   e one terminal block   e one snap track  Note Supporting all of the connections on units with the optional I O  expansion board requires    e two DB37 kits   e one DB25 kit    Table 9 1 identifies the optional individual cables that are available for the  instrument and Table 9 2 provides the cable color codes  For associated  part numbers  refer to    External Device Connection Components    on page    7 8     Note Table 9   2 provides the color coding for both 25 pin cables and 37   pin cables  Color codes for pins 1 25 are for 25 pin cables  color codes for    pins 1 37 are for 37 pin cables  A    Table 9 1  Cable Options             Description Cable Length  DB37M to open end Six feet  DB37F to open end Six feet  DB25M to open end Six feet  RS 232 Six feet    Thermo Fisher Scientific    Optional Equipment  Mounting Options    Table 9 2  Color Codes for 25 Pin and 37 Pin Cables                                                       Pin Color Pin Color   1 BLACK 20 RED BLACK   2 BROWN 21 ORANGE BLACK   3 RED 22 YELLOW BLACK   4 ORANGE 23 GREEN BLACK   5 YELLOW 24 GRAY BLACK   6 GREEN 25 PINK BLACK   7 BLUE End color codes for 25 pin cables  continue for 37 pin cables    8 VIOLET 26 PINK GREEN   9 GRAY 27 PINK RED   19 WHITE 28 PINK VIOLET   11 PINK 29 LIGHT BLUE   12 LIGHT GREEN 30 LIGHT BLUE BROWN   13 BLACK WHITE 31 LIGHT BLUE RED   14 BROWN WHITE 32 LIGHT BLUE V
509. ration alarm minimum B 17  min value   alarm conc avg Reports sets average SHARP concentration alarm maximum B 17  sharp max value   alarm conc avg Reports sets average SHARP concentration alarm minimum B 17  sharp min value   alarm conc inst Reports sets instant nephelometer concentration alarm B 17  neph max maximum value   alarm conc inst Reports sets instant nephelometer concentration alarm B 17  neph min minimum value   alarmconcinstpm     Reports sets instant PM concentration alarm maximum value B 18  max   alarm conc inst pm Reports sets instant PM concentration alarm minimum value B 18  min   alarm conc inst Reports sets instant sharp concentration alarm maximum B 18  sharp max value   alarm conc inst Reports sets instant sharp concentration alarm minimum B 18  sharp min value       Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands                                                                                        Commands  Command Description Page  alarm filter tape Reports the filter tape status B 19  status  alarm flow max Reports sets flow alarm maximum value B 19  alarm flow min Reports sets flow alarm minimum value B 19  alarm flow pres max Reports sets flow pressure alarm maximum value B 19  alarm flow pres min Reports sets flow pressure alarm minimum value B 19  alarm flow temp Reports sets flow temperature alarm maximum value B 20  max  alarm flow temp Reports sets flow temperature alarm minimum value B 20  min  alarm ib stat
510. re 31   Figure 32   Figure 33   Figure 34     Figure 35   Figure 36   Figure 37   Figure 38   Figure 39     Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Sample path within the Model 50301 SHARP measuring systeM                32  Hardware components Model 50301 SHARP     Optics module                 0008 33  Hardware components Model 5030i SHARP     Beta module                   ccccee 33  Picture of hardware components Model 50301 SHARP     Beta module              34  Diagram     Set up of Model 50301 SHARP            ccccccccccecceeeeeeeeeeeeeeeeeeeeeeeseeeeess 35  Model 50301 SHARP  SHARP optics modules  top    SHARP beta   module  OOHONN   sseserce createed ete ceay a csten siaceentten poasteeneteneseretensdeeentnenmetieas 37   European  PM25 sampling inlet for Model 50301 SHARP    onssa 38  Heated sampling tube  DAS  ssssccicicieitissacssdanticidensseiaintet etn dedeiens 39  Model 50301 SHARP measuring system             eeeeeeeeeeeee teeter ee eeectaeeeeeeeeeee erent 39  Model 50301 SHARP measuring system     Optics MOdUIe              cccccceeeeeeteeees 40  Model 50301 SHARP measuring system     Beta moOdule                ccccccceceeeeeeees 40  Model 50301 SHARP measuring system  Systems 1 and 2 from right  at  measuring Station sec cct cetersconerccarcrencieeecaecstete
511. re 7   30  Replacing the Internal SHARP Cable Assembly    Case Cable Use the following procedure to replace the case cable  Figure 7 31    Replacement Equipment Required   Case Cable  Philips screwdriver    Adjustable wrench    amounts of static electricity  A properly grounded antistatic wrist strap    f Equipment Damage Some internal components can be damaged by small    must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  remove the  electrical connector  loosen the two captive screws on the SHARP  optics base plate  and remove the SHARP assembly from the base unit     2  Remove the two SHARP cover screws and remove the cover     3  Remove the hex nut from the cable clamp     7 52 Model 5030  SHARP Instruction Manual Thermo Fisher Scientific    Servicing  External Zeroing Filter    4  Replace the case cable assembly by following the previous steps in  reverse     Unplug        Hex Nut       Cover Screw    Cover Screw    Figure 7 31  Replacing the Case Cable    External Ze roing Use the following procedure to replace the external zeroing filter  Figure 7         2    Filter       Equipment Required     Adjustable wrench    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    f Equipment Damage Some internal components can be damaged by small    1  Note the orientation of the arrow on the filter assembly  See Figure 2   7  for orientation     
512. reatment srec  Reports sets data treatment for concentration values in srecs B 28  date Reports sets current date B 52  default params Sets parameters to default values B 52  det Reports sets the status of the detector board B 53  det status Reports the current mass for the detector B 24  dhcp Reports sets state of use of Dynamic Host Configuration B 57  Protocol  DHCP   diag volt det Reports diagnostic voltage levels on the detector board B 24  diag volt iob Reports diagnostic voltage levels on the I O expansion board B 24  diag volt mb Reports diagnostic voltage levels on the motherboard B 24  diag volt mib Reports diagnostic voltage levels on the measurement B 25  interface board  diag volt neph Reports diagnostic voltage levels on the nephelometer B 25  interface board  dig in Reports status of the digital inputs B 62  din Reports sets digital input channel and active state B 62  do  down  Simulates pressing down pushbutton B 44  dout Reports sets digital output channel and active state B 63  dtoa Reports outputs of the digital to analog converters per B 63  channel  en  enter  Simulates pressing enter pushbutton B 44  er Returns a brief description of the main operating conditions B 28  in the format specified in the commands  erec Returns a snapshot of the main operating conditions B 28   measurements and status  in the specified format  erec format Reports sets erec format B 30       Model 50307 SHARP Instruction Manual B 7    C Link Protocol Commands  Commands    B 8 
513. record field items to default values for the selected record type  For more    information about selecting the content of logged data fields  see    Select  Content    above     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Reset to Default Content     3 32 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions    PE PTT   PEE ETE bel  tO PEL a EA       Configure Datalogging The Configure Datalogging menu deals with datalogging configuration for  the currently selected record type     e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Configure Datalogging        RCRSE CHS mito    Wi mht  E PORE TE Saad foose i r Sel be hb Bedi beet Bhi       Logging Period Min The Logging Period Min screen is used to select the logging period in  minutes for the selected record format  srec or lrec   List of choices include     off  1  5  15  30  and 60 minutes  default      e Inthe Main Menu  choose Instrument Controls  gt  Datalogging  Settings  gt  Configure Datalogging  gt  Logging Period Min     a ed ne Patio pH mith  Marian mF mii       Memory Allocation The Memory Allocation Percent screen is used to select the percentage of  Percent total memory that may be used by the selected record type  lrecs or srecs    Percentages between 0 and 100  are available in increments of 10   Changing this value results in log erasure for both types of records  and    changing the percent allocated to 
514. reesstessreesrresereeseee 3 30  Analog Inputs se crc cae Rate R a R R Ekas 3 31  Commit Contente a na a she R A teens betas eed 3 32  Reset to Default Content    ssssesssseesrsssrrssressreessressreesseessressrresreeseee 3 32  Configure Datalogging       seesesesesesesssesesrsesrsrsesetesrereeresesesserssse 3 33  Logging Period Minecisalsisnninannenia ni i otis 3 33  Memory Allocation Percent       ssessesssssessssessesessersesesseseseseesesessese 3 33  Data Treatments eoan E E E an 3 34  Flag Status Data siota a a a D 3 34   C  mm  nication Set MES assor aiia ia 3 35  Serial Seting Soni drt ouon dirt bianstvbetushasat anata Duan tuat at yah eeetestas 3 35  Bait RAG EE AE E E TETE 3 35  Data EI OSE EE TE E N E E EEE EA ETE 3 36  PALLY e se ade Ae a a Ge ee ee eee 3 36  Stop Bits sssislsdeats onei aia eiea N aii a i ai 3 36  RS 232 RS 485 Selection      eeseeeeeeeeeeeeesessesssressersseessressrresereesses 3 37  Instrument IDa A N A a 3 37  Gesytec Serial Nunb  r x sansa saevsscsveesvnsbatcorsarege coteassecdadsvansytabenen sear 3 38  Communication P totecol ai  sanieti vdieo hi went tlasas 3 38  Streaming Data Configuration veaase tala etic esas 3 38  Streaming Data Interval sisine rniii anea E Goatees a 3 39  Choose Stream Data      sssssessesssseeseeesersseessreeseresereeseessrresrreeereeseee 3 39  Concentrations nern i a E E E R 3 40  Other Measurements        sssssesssseereeesersserssressrersrerssressrresrreesreessee 3 40  Analog I PUES seit osare ai kr a aeeai 3 41  TEPIP 
515. rement Interface  Board Connectors    Flow Sensor System    Pressure Sensor  Assembly    Detector Amplifier  Assembly    Digital Output Board    8 8 Model 50307 SHARP Instruction Manual    Connectors located on the measurement interface board include   Data communication with the motherboard  e 24 Vand 120 VAC power supply inputs  e Fan and solenoid outputs  e 120 VAC output and thermistor input from the sample heater  e Flow and pressure sensor inputs  e Proportional valve output  e Optical pickup input  e Pre amp board  e Ambient temperature RTD and RH  e FlowRTDs    The flow sensor system consists of a subsonic orifice inside the primary  measurement head  a differential pressure and vacuum sensors  and a flow  temperature RTD  The flow system output is produced by measuring the  pressure difference across a precision orifice  This unit is used for  measuring the flow of sample gas in the measurement system     The pressure sensor assembly consists of a board containing three pressure  transducers with tube connections  The pressure transducer output is  produced by measuring the pressure difference between the sample gas  pressure and ambient air pressure     The detector amplifier assembly amplifies the signal from the proportional  detector that receives beta particle emissions from the C 14 source through  the sample and filter tape  The output of the detector amplifier is fed to the  measurement interface board  A coaxial cable from the detector amplifier  supplies po
516. reshold at the  sample filter tape  The purpose of this heating system is specifically  designed to force the continuous mass monitor to agree with the  gravimetric reference method and the relative humidity conditions to  which the reference filter samples are conditioned  Furthermore  sufficient  flexibility is provided within the firmware to configure the heating  conditions to satisfy global monitoring protocols     The measuring range of the Model 5030  is 0 10 000 pg m        A complete particulate measuring assembly consists of the following main  parts     e SHARP Optics Module  e SHARP Beta Module  e Ambient Inlet Sampling System  PMio  PM2 5  PM1 0     e Dynamic Heater    Model 50307 SHARP Instruction Manual 1 1    Introduction  Principle of Operation    1 2 Model 50307 SHARP Instruction Manual    Diaphragm Vacuum Pump   External Ambient T RH Sensor Assembly   Internal Data Recording Unit and External Communication Ports  Optional 6 Foot Sample Extension Tubes   Optional Tripod Support   Optional Roof Flange Kit    The Model 50307 combines proven technology  easy to use menu driven  firmware  and advanced diagnostics to offer unsurpassed flexibility and  reliability  The Model 50302 has the following features     320 x 240 pixels graphics display   Menu driven firmware   Field programmable ranges   Multiple user defined analog outputs   Analog input options   Linearity through all ranges   User selectable digital input output capabilities   Standard communications f
517. rformance of test   Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 9 of this report   6 4 Evaluation   Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 9 of this report   6 5 Assessment    Not applicable to PM25 sampling systems  Reference is made to module 5 4 9 of this report   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable to PM2 5 sampling systems  Reference is made to module 5 4 9 of this report     740300_2013_936_21209885F_EN docx    a 3    TUV Rheinland Energie und Umwelt GmbH  TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system  Page 132 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo    Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    6 1 5 4 4 Calibration    The systems under test shall be calibrated in the field test by comparison measure   ments with the reference method  Here  the relationship between the output signal and  the gravimetrically determined reference concentration shall the determined as a  steady function     6 2 Equipment  See module 5 4 10   6 3 Performance of test    For PM3 s5     The comparability of the measuring systems according to module 5 4 10 was demonstrated  during the test     The determination of the calibration and analysis function was carried out using the complete  dataset  209 
518. rial port in real time for capture by a serial printer  datalogger  or PC     A null modem  crossed  cable is required when connecting the analyzer to  an IBM compatible PC  However  a straight cable  one to one  may be  required when connecting the analyzer to other remote devices  As a  general rule  when the connector of the host remote device is female  a  straight cable is required and when the connector is male  a null modem  cable is required     Data Format     1200  2400  4800  9600  19200  38400  57600  or 115200 BAUD    Model 50307 SHARP Instruction Manual 8 11    System Description  1 0 Components    RS 485 Connection    Ethernet Connection    External Accessory  Connector    8 12 Model 50307 SHARP Instruction Manual    7 or 8 data bits    1 or 2 stop bits    No  odd  or even parity    All responses are terminated with a carriage return  hex 0D     Refer to Table 8   1 for the DB9 connector pin configuration   Table 8 1  RS 232 DB9 Connector Pin Configuration                DB9 Pin Function  2 RX   3 TX   7 RTS   8 CTS   5 Ground    The instrument uses a four wire RS 485 configuration with automatic flow  control  SD   Refer to Table 8 2 for the DB9 connector pin configuration     Table 8 2  RS 485 DB9 Connector Pin Configuration             DB9 Pin Function  2   receive  8   receive  7   transmit  3   transmit  5 ground    An RJ45 connector is used for the 10Mbs Ethernet connection supporting  TCP IP communications via standard IPV4 addressing  The IP address  ma
519. ring       lt STX gt DA lt CR gt     Reply String    lt STX gt MDO3 lt SP gt 001 lt SP gt  2578 01 lt SP gt 03  lt SP gt 04 lt SP gt OO00000000  lt SP gt 002  lt SP gt   t T t  Address First Concentration E format  25 78 Address 1     5681 00 lt SP gt 03 lt SP gt 04 lt SP gt 0000000000 lt SP gt 003 lt SP gt  1175 01 lt SP gt 03 lt SP gt 04 lt SP    d t T  Second Concentration   5 681 Address 2 Third Concentration 11 75  0000000000 lt SP gt  lt CR gt     Example of response to DA command from an instrument with Gesytec  serial number set to 123  The Gesytec serial number is bold in the  example     Gesytec Protocol with transmission of three concentrations  Instrument ID  is 1  Operation Status is 03  Error Status is 04      Data Query String   lt STX gt DA lt CR gt     Reply String    lt STX gt MD03 lt SP gt 001 lt SP gt  2578 01 lt SP gt 03  lt SP gt 04 lt SP gt 1230000000  lt SP gt 002  lt SP gt   t T t  Address First Concentration E format  25 78 Address 1     56814 00 lt SP gt 03 lt SP gt 04 lt SP gt 0000000000 lt SP gt 003 lt SP gt  1175 01 lt SP gt 03 lt SP gt 04 lt SP    T t d  Second Concentration   5 681 Address 2 Third Concentration 11 75  1230000000 lt SP gt  lt CR gt     The attached concentrations are in the selected gas units  The  measurements that are attached to the response  if not valid in a particular  mode  will report a value of 0 0     Thermo Fisher Scientific    Measurements reported in  response to DA command    Thermo Fisher Scientific    Gesytec  Ba
520. rmine whether the comparability of both candidates  with the reference method according to Guide  Demonstration of Equivalence of Ambient Air  Monitoring Methods     4  is also given for days with more than 70   relative humidity     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control       Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher Page 135 of 424  Scientific for the component suspended particulate matter PMz 5   Report No   936 21209885 F    6 5 Assessment    No deviation of the measured signal from the nominal value  gt  0 5 ug m  caused by interfer   ence of moisture in the sample could be observed for PM25  No negative influence on the  measured values at varying relative humidity was detected during the field test  Comparabil   ity of the candidates with the reference method according to Guide  Demonstration of  Equivalence of Ambient Air Monitoring Methods     4  is also given for days with a relative hu   midity  gt  70       Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  A summary of the results is presented in Table 27     Table 27  Deviation between reference measurement and candidate on days with a relative  humidity  gt  70    measured component PM  5    Field test  days with a relative humidity  gt 70      pT tare  oT ae    Deviation from om  refer
521. rms Menu    RH Temperature The RH Temperature Alarms submenu allows the user to view and set the  P  Alarms relative humidity or temperature alarm parameters     e Inthe Main Menu  choose Alarms  gt  RH Temperature Alarms        Ambient RH The Ambient RH screen displays the current ambient relative humidity  value and sets the minimum and maximum alarm limits  Acceptable alarm  limits range from 5 to 95   If the ambient RH reading goes beyond the  minimum or maximum limit  an alarm is activated and the alarm  bell   icon appears in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  RH Termperature Alarms  gt   Ambient RH        ep gies   Piet          Min and Max AmbientRH The Minimum Ambient RH alarm limit screen is used to change the  Limits minimum ambient RH alarm limit  The minimum and maximum ambient  RH screens function the same way     e Inthe Main Menu  choose Alarms  gt  RH Temperature Alarms  gt     Ambient RH  gt  Min or Max     3 72 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Alarms Menu       Sample RH The Sample RH screen displays the current sample relative humidity value  and sets the minimum and maximum alarm limits  Acceptable alarm limits  range from 5 to 95   If the sample RH reading goes beyond the minimum  or maximum limit  an alarm is activated and the alarm  bell  icon appears  in the status bar on the Run screen and in the Main Menu     e Inthe Main Menu  choose Alarms  gt  
522. rms is allowed  Values cor   rected shall meet the require   ments of chapter 9 5 2 2 ff  of  the guidance document    Demonstration of Equivalence  of Ambient Air Monitoring  Methods        Shall comply with the require   ments set for each component  also in the case of simultane   ous operation of all measuring  channels        The candidate systems fulfil the re   quirements on data quality of ambient  air quality measurements during the  test even without application of cor   rection factors  A correction of the  slope leads nevertheless to a signifi   cant improvement of the expanded  measuring uncertainty of the full data  comparison     Not applicable     740300_2013_936_21209885F_EN docx    a   D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 28 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    2  Task definition    2 1 Nature of test    Thermo Fisher Scientific has commissioned TUV Rheinland Energie und Umwelt GmbH to  carry out suitability testing of the measuring system Model 5030  SHARP with PM25 pre   separator  The test was performed as a complete suitability test     2 2 Objective    The AMS shall determine the content of PMz5 fine dust in ambient air within the concentra   tion range of 0   10
523. rough he roof flange  assembly and connect to the bottom of the ambient T RH assembly     14  Connect the other end of the T RH cable to the back of the instrument   RH Temp      15  Attach the heater power connector to HEATER 1     16  Using the available 3 8 inch green vacuum tubing  connect the intake  of the pump to the vacuum fitting at the rear panel of the 50302     17  Plug the pump power chord into the AC PUMP receptacle at the rear  of the 50302 instrument     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 2 15    Installation  Setup           lt        RH Temperature  Final Assembly                 SHARP Electric  Connection    Exhaust    Local PC    Zero Air  Make up    Figure 2   7  Rear Panel Plumbing Schematic    2 16 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Connecting  External Devices    Terminal Board PCB  Assemblies    1 0 Terminal Board    Thermo Fisher Scientific    Installation  Connecting External Devices    Several components are available for connecting external devices to iSeries  instruments     These connection options include   e Individual terminal board PCB assemblies  standard   e Terminal block and cable kits  optional   e Individual cables  optional     For detailed information on the optional connection components  refer to  the    Optional Equipment    chapter  For associated part numbers  refer to     External Device Connection Components    on page 7 8     The terminal board PCB assemblies are circui
524. roup choice     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Output Config  gt  select Channel  gt  Choose Signal to  Output        RCPS E CHS mito  DN EEE TE Sanat Bone ETET Aad Bast    ade           ae i       aa pe  Pri ii    ny es perry  T had ie oie       Table 3   5  Signal Type Group Choices       Concentrations Other Measurements Analog Inputs  None None None  PM Barometer Pressure Analog Input 1  if the I O expansion       Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 51    Operation  Volumetric Conditions    Concentrations Other Measurements Analog Inputs       board is installed                                            AVG PM Vacuum Analog Input 2  if the I O expansion  board is installed   Neph Flow Pressure Analog Input 3  if the I O expansion  board is installed   SHARP Ambient RH Analog Input 4  if the 1 0 expansion  board is installed   AVG SHARP Sample RH Analog Input 5  if the 1 0 expansion  board is installed   AVG Neph Ambient Temp Analog Input 6  if the I O expansion  board is installed   Board Temp Analog Input 7  if the I O expansion  board is installed   Flow Temp Analog Input 8  if the 1 0 expansion  board is installed   Flow Vol  Beta  Beta Raw  Alpha Raw  Mass  Beta Zero       External Alarms       Neph Temp  Neph RH  Neph IRED  Neph Ref  Beta Ref                   Common Flags       Det Flags       Mass Comp    Analog Input Configuration The Analog Input Configuration menu displays a list of 
525. rument ID of 14  For  more information about the instrument ID  see Appendix B    C Link    Protocol    or Appendix C    MODBUS Protocol        e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Instrument ID     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 37    Operation  Volumetric Conditions       Gesytec Serial Nunber The Gesytec Serial Number screen is used to set the serial number for the  Gesytec protocol  The Gesytec serial number defaults to zero  For more  information about the serial number  see Appendix D     Gesytec  BH   Protocol        e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Gesytec Serial No        DH Guided  Piet          Communication Protocol The Communication Protocol screen is used to change the protocol for    serial communications  Possible choices include  C Link  MODBUS   Gesytec  and Streaming Data     e Inthe Main Menu  choose Instrument Controls  gt  Communication  Settings  gt  Communication Protocol        Streaming Data Configuration The Streaming Data Configuration menu is used to allow for configuration  of the 8 streaming data output items  streaming interval  current data  format  and current timestamp setting  The Choose Stream Data submenu  displays a list of the analog output signal group choices to choose from     3 38 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Volumetric Conditions    Choices are Concentrations  Othe
526. s   6 1 5 3 18 Overall uncertainty   not relevant for particulate devices  covered by 5 4 10    For this reason an official request to the competent German body was made  to define a co   ordinated procedure for dealing with inconsistencies in the Standard     The following procedure was suggested     Test points 5 3 2  5 3 7  5 3 8  5 3 11 and 5 3 12 are evaluated as before basing on the min   imum requirements stated in VDI 4202 Part 1 of 2002  i e  using the reference values Bo  B   and B3      The performance of the test points 5 3 3  5 3 4  5 3 13 and 5 3 18 is omitted  as they are not  relevant for particulate measuring devices     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control   s  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 15 of 424   Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    The competent German body agreed with the proposed procedure by decisions of 27 June  2011 and 07 October 2011     Since according to the applied standards the reference values to be used are explicitly ad   justed to component PMio  for component PMz5 the following reference values have been  agreed upon     PMz2 5 PMio  for comparison   Bo 2 g m 2 g m   B  25 g m  40 g m   B2 200 g m  200 g m     B  shall be merely adjusted to the level of t
527. s   Lg m3          C     C   hPa       m s   4  mm   271 8 30 2012 Teddington  Summer 0 9 12 6 17 7 1005 70 1 1 5 302 0 0  272 8 31 2012 1 2 15 3 19 4 1010 60 5 0 7 289 0 0  273 9 1 2012 1 8 16 6 22 7 1006 75 4 0 2 210 0 0  274 9 2 2012 2 7 16 8 19 5 1007 88 3 0 1 238 0 0  275 9 3 2012 3 3 18 4 26 9 1006 74 1 0 2 239 0 0  276 9 4 2012 8 8 3 0 33 8 18 9 26 6 1005 64 6 tA 295 0 0  277 9 5 2012 10 8 1 2 10 8 14 7 21 9 1010 63 9 1 0 144 0 0  278 9 6 2012 11 3 2 5 22 1 16 3 23 2 1008 67 9 0 2 198 0 0  279 9 7 2012 2 4 17 5 28 4 1005 63 9 0 1 219 0 0  280 9 8 2012 2 6 tA 28 3 998 64 8 0 1 208 0 0  281 9 9 2012 1 9 20 1 27 4 991 66 5 0 8 191 0 0  282 9 10 2012 2 5 17 5 21 0 991 75 3 0 9 192 0 0  283 9 11 2012 4 4 1 3 30 5 12 8 19 4 997 63 4 0 4 252 0 0  284 9 12 2012 1 5 13 4 20 4 998 72 4 0 5 267 2 5  285 9 13 2012 4 7 2 0 43 2 16 3 21 4 998 63 9 0 3 205 0 0  286 9 14 2012 5 2 1 8 34 2 14 3 21 3 997 70 8 0 7 257 0 0  287 9 15 2012 8 7 2A 24 4 17 0 24 3 997 71 9 0 2 204 0 0  288 9 16 2012 3 4 1 6 46 7 15 6 19 6 994 75 9 0 2 207 0 0  289 9 17 2012 5 9 2 0 33 7 15 5 20 7 994 67 9 0 2 207 0 0  290 9 18 2012 6 4 1 0 14 9 11 7 19 3 1000 64 2 0 7 278 0 0  291 9 19 2012 8 2 1 4 17 3 12 2 18 3 1004 65 9 0 4 260 0 0  292 9 20 2012 5 7 1 9 32 9 13 3 19 3 1001 70 9 0 2 203 0 0  293 9 21 2012 5 3 2 1 39 9 12 1 17 3 998 77 3 0 4 273 1 3  294 9 22 2012 7 5 1 8 24 6 12 5 17 2 1000 60 6 0 8 89 0 0  295 9 23 2012 12 4 6 3 50 7 11 8 14 7 981 93 8 1 3 80 30 0  296 9 24 2012 3 2 2 3 70 8 12 2 16 6 971 
528. s and corresponding actions  The following example  sets the digital input channel 1 to 3 on a low to high transition     Send  set din 1 3 high  Receive  set din 1 3 high ok    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  1 0 Configuration    dout channel   This command reports the index number  output variable and the active  state assigned to digital output channel  The following example reports  output 4 to be assigned an index number 11 corresponding to action of     flowl alarms        Send  dout 4  Receive  dout 4 11 FLOW ALARMS open    set dout channel index state   This command assigns digital output channel to be assigned to the action  associated with index  and assigns it an active state of state  open or closed    Use the    list var dout    command to obtain the list of supported index  values and corresponding state  The following example sets the digital  output channel 4 to state 11     Send  set dout 4 11 open  Receive  set dout 4 11 open ok    dtoa channel   This command reports the outputs of the 6 or 12 Digital to Analog  converters  according to Table B 13  The following example shows that the  D A  1 is 97 7  full scale     Send  dtoa 1  Receive  dtoa 1 97 7     Note If the instrument is in a mode which does not provide a particular  output  and that output is selected  the value will be 0 0  a    All channel ranges are user definable  If any customization has been made  to the analog output configuration  the defaul
529. s command enables value  where value is to attach or disable time stamp  to streaming data string  according to Table B   4  The following example  attaches a time stamp to streaming data     Send  set stream time 1  Receive  set stream time 1 ok    Table B   4  Stream Time Values    Value Stream Time   0 Disables time stamp to streaming data string   1 Attaches time stamp to streaming data string  alpha eff    This command reports the current value of alpha efficiency for the  detector  The following example reports that the alpha efficiency is 0 12     Send  alpha eff  Receive  alpha eff 0 120  set alpha eff value    This command sets the alpha efficiency for the detector to value  where  value is a floating point number representing the alpha efficiency for the  detector  The following example sets the alpha efficiency to 0 15     Send  set alpha eff 0 15  Receive  set alpha eff 0 15 ok  alpha th    This command reports the current value of alpha threshold for the  detector  The following example reports that the alpha threshold is 1220     Send  alpha th  Receive  alpha th 1220    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Calibration    set alpha th value   This command sets the alpha threshold for the detector to value  where  value is a floating point number representing the alpha threshold for the  detector  The following example sets the alpha threshold to 1250     Send  set alpha th 1250  Receive  set alpha th 1250 ok    amb rh
530. s in the  service mode     e Inthe Main Menu  choose Service        RH T emperature The RH Temperature Calibration submenu allows the user to view and set  Calibration tbe relative humidity and temperature sensor calibration  The  RH temperature calibration submenu is visible only when the instrument is  in service mode  For more information on the service mode  see    Service  Mode    earlier in this chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  RH Temp Calibration     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 85    Operation  Service Menu       Ambient Temperature The Ambient Temperature screen allows the user to view and calibrate  ambient temperature  Adjust the zero offset until the ambient temperature  agrees with your treaceable standard     e Inthe Main Menu  choose Service  gt  RH Temp Calibration  gt   Ambient Temp        Flow Temperature The Flow Temperature screen allows the user to view and calibrate flow  temperature  Be sure to have the cover removed and equilibrated to room  temperature     e Inthe Main Menu  choose Service  gt  RH Temp Calibration  gt  Flow  Temp        EHO guided  Piet          Ambient Relative The Ambient Relative Humidity screen allows the user to view and  Humidity calibrate relative humidity  Adjust the zero offset until the RH agrees with  your treaceable standard     e Inthe Main Menu  choose Service  gt  RH Temp Calibrat
531. s liquid crystal display  LCD  shows the  sample concentrations  instrument parameters  instrument controls  help   and error messages  Some menus contain more items than can be displayed  at one time  For these menus  use  _  _  and      to move the cursor up  and down to each item     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 1    Operation  Pushbuttons              3  0       Figure 3 1  Front Panel Display    CAUTION If the LCD panel breaks  do not let the liquid crystal contact  your skin or clothes  If the liquid crystal contacts your skin or clothes  wash  it off immediately using soap and water  A    Pushbutto NS The Pushbuttons allow the user to traverse the various screens menus        Figure 3 2  Front Panel Pushbuttons    3 2 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    Soft Keys    Operation    Pushbuttons  Table 3   1 lists the front panel pushbuttons and their functions   Table 3 1  Front Panel Pushbuttons  Key Name Function       Soft Keys The     soft keys  are used to provide shortcuts that    allow the user to jump to user selectable menu screens  For  more information on processing soft keys  see    Soft Keys     below       gt      Run The   gt   is used to display the Run screen  The Run    screen normally displays the current SHARP concentrations   but also has additional Run screens for ambient conditions   sample conditions  and mass sensor data  Press   and     to scroll through th
532. s screwdriver   2    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Document the Nephelometer Coefficient  From the Main Menu   choose Calibration Factors  gt  Neph Coef  Record the nephelometer  coefficient value     2  Document the Nephelometer Calibration  From the Main Menu   choose Service  gt  Neph Calibration  Record the nephelometer RH  value  the nephelometer temperature value  and the nephelometer  source level value     3  Turn the instrument OFF  unplug the power cord  and remove the  SHARP otpics cover     Model 50307 SHARP Instruction Manual 7 47    Servicing  SHARP Interface Board Replacement    2 Pin Male   not used     7 48 Model 50307 SHARP Instruction Manual       Unplug all connectors from the SHARP interface board  Note the    locations of the connectors to facilitate re connection     Loosen the two screws from the bottom of the rear board panel and  remove     Unfasten the two screws at the top of the SHARP interface board  Slide  the SHARP interface board up from the two bottom mounting studs  and remove the board     To install the SHARP interface board  follow the previous steps in    reverse   Restore the nephelometer coefficient value previously documented     Restore the nephelometer RH value  the nephelometer temperature  value  and the nephelometer source level value previously documented     
533. sacteentencdietbenstsciencvents 2 13  Rack Mounting Options  neten aaa en sal Cae es 2 13  Monitor Installation sere tessa ase  Gsncnacscccee cal acta he acsstar  2 13  Connecting External Devices 528s aise eA ca eon Se eu 2 17  Terminal Board PCB Assenib tess   i casstavssasdoasconayscesaebsaavensdstecesbirs 2 17  TO Retinal Board sa cessicssiiscsitteasisiceavaveasiaseaikwiecntiacienunenns 2 17  D O Terminal Boatd assii sant vesbuacosstrege annsin ikse 2 19  25 Pin Terminal Board    s s ssssssssesseseseeessesesseseseseesesesesresesesseseseeee 2 20  Power Connections nnenernena i ene a a e es 2 22  Stat   prea a O A E desta aE AAEE a ia 2 22  Data Content seiiet ier s aa aN ee ee eds 2 23  Shutdown seirinin ne ur e a a E a 2 23  Imp  rtant Tipsene gn a e a a aai 2 23   Oper  ti  i izina aa aaa aaa a araia 3 1  Displayen annia T da 3 1  Push Dutton Seii ut aao ne e Rro ES a EErEE EEES Are aides 3 2   DOU E A E E AE 3 3  Alphanumeric Entry  Screen aeinn haraa oeaan 3 4    Model 50307 SHARP Instruction Manual vii    Contents    Firmware COVELVICW  sa cacesacasies nish cnetands seta testeaeaasssvases aiaa 3 4  Power  Up  Screen saiisine tiisin siaina ninesi 3 6  R  n Screens uasai iape aip a h aie EA Maxi ave a 3 6   Ambient Conditions Run Screenis ciciccscccevsesctiecctientevtenersidetiaenss 3 7  Sample Conditions Run Screen    ssseeeeeeerrrrrsesrsesses 3 8  Mass Sensor Data Run Screen   sssssesssesessessrresesrsrrsresreresresrreresrsreees 3 8  Hybrid Data Run Screen       sesss
534. screen allows the user to set the input voltage for the selected  table point in the conversion table  from 0 00 to 10 50  The default table is  a two point table with point 1  0 00 V   000 0 U and point 2  10 00 V    10 0 U  where U is the previously entered unit of measure     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  select Point  gt  Volts         ae i       ao ee Poh Fly PPT  PS Pi ELSES Pi AG sae a ae Le    CH mtb       User Value The User Value screen allows the user to set the output value for the  corresponding input voltage for the selected table point in the conversion  table  from  9999999 to 99999999  The default table is a two point table  with point 1  0 00 V   000 0 U and point 2  10 00 V   10 0 U  where U is  the previously entered unit of measure     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  select Point  gt  User Value     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 55    Operation  Volumetric Conditions    Screen Contrast    Service Mode    3 56 Model 50307 SHARP Instruction Manual       The Screen Contrast screen is used to change the contrast of the display   Intensities between 0 and 100  in increments of 5 are available  Changing  the screen contrast may be necessary if the instrument is operated at  extreme temperatures     Note The optimal contrast will change with changes in t
535. screws from the connection bracket and align  the bracket onto the wall plate  as shown above in Step 4  and insert the  two screws through the wall plate and tighten     4  Reinstall the tape pickup spool in the reverse order of removal in Step 2  by sliding all the spacers  O rings and rear support guide plate into its  original assembly order     5  Install the filter tape following the filter tape path shown in Figure 10     3  Use adhesive tape to adhere the filter tape to the take up spool and  wind the tape over itself at least once     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 10 3    Model 5014i to 50301 Upgrade Procedures  SHARP Connection Bracket Assembly Installation       Figure 10 3  Filter Tape Installation Pathway    6  Complete the installation of the SHARP Optics Connection Bracket by  connecting the 4 pin connector to the    SHARP 1    connector on the  measurement interface board as shown below in Figure 10 4     Note It should not be necessary to remove the electronics from the  instrument case to achieve this connection and the spare 2 pin connector is  not used  A    10 4 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Model 5014i to 50301 Upgrade Procedures  Radius Tube Adapter Installation    ote    P    atl       Figure 10   4  Connecting Connection Bracket Wiring to Measurement Interface  Board    Radius Tube Use the following procedure for radius tube adapter installation   Adapter Installation  1  Loosen the Ra
536. scssssccssscecsssseesseeeesees 3 89  Calibrate Vacuum  Pressure Span    3 89  Calibrate Flow Pressure Span  cictcc2ceecetaceeeiiecohr di captea eet eden  3 90  Restore Default Calibtatiotiicsaicicidctgccatiavacsett azine nleatenc ease ides 3 90  Flow Calibration yam seman cet natn ene SNe TI Rte CPN AR er at ener mroe 3 91  Auto Flow Calibration esccasaciessishcactsagshenicasstanssacneees ateacadseeettecebaas 3 91  Manual Flow Calibration      sssseeeseessseeeseeereeeseessressrresrresereesserssress 3 92  Mass Calibrations aert e an cabin a A 3 92  Thermal Mass Coefficient sss ntiaise nsiaion oinadiatuntsuateanteniantoaSes 3 92  Vacuum Mass Coefficient s5 c1cdcasssctoencitadeistnadvcantvvorctodieeieaneats 3 93  Barometer Mass Coefficient       ccccccccccccssssscscssssscssssscssssscssssscoseess 3 93  Mass Coefficient      ssesessseeeseesseeeseeesreesrressrersseresetssresseressreesserssrese 3 93  Auto Mass CO cticie ni Gics  avsacerssasataanleaseane ce iacdees tacaudsaastoandaeaeaeacets 3 94  Manual Mass Coefficient sesnies odd iunas sot euccraslagthinantettauasaatnases 3 94  Alpha Efficiency Coemicient sic cccecsancbetdcsennce cciasencdatvacatonteteuts 3 94  Detector Calibration  isicwwecn del a a a a ss 3 95  Auto Detector Calibiatl O fives ian intoncddots nats sadelaawieetenntes 3 95  Manual Detector Calibration                 cc ccscscsssssssssssssssscassesenses 3 96  Beta Ref Alpha Re Oe 5 esc cc a ae le hel as 3 96  PSV Ota eats site  enina aE ode cheater RR race nce ee  
537. sdicusdsateavscialsuatesneaateanteaeiesteeaeetet 5 7  Greasing the Gani cas one n A a a n Sk 5 8  External Pump Exhaust Filter  site Goedel thse eee 5 8  Cleaning the SHARP Optics Assembly    s s s 5 10  Troubleshooting  irsidir ea a a a a a 6 1  Salty  Veet INS v urnie a oina ea 6 1  Troubleshooting Guides     sssesseeeeeeessseessssrsssssesrsssisrsrsesrsrsesesesesesesee 6 1  Board Level Connection Diagrams     ssessesseseeesesesrsesrrrsereressseseseseses 6 8  Connector Pin Descriptions        seesssseseerssresesreresresterrssrsrreresreresreseerese 6 10  Seryic   L  cati  n Siia a a a a a a 6 25  STIATE  sts E a E E E vnatestanceraiaeovaanaes 7 1  Safety Precautions renan a a R R ANRT 7 3  Fitmware Updatesioioueniis radai A a a 2  7 4  Accessing the Service Mode ssnin onae ana 7 4  Replacement Parts LIsCresnerie ninani To a E E acl 7 5  Cable ASE a con est daencce ate a aa E A NS 7 7  External Device Connection Components    s s ss sesesesessssssessersreeseeeseseses 7 8  Removing the SHARP Optics Assembly    s s 7 11  Removing the Measurement Case Assembly and Lowering the Partition  Panel dman an a a a e NR 7 12  Fuse Replacement chit Giese atest EAS 7 14  Fan Filter Rep cements Atle aa a aE EE 7 14  Detector Amplifier Assembly Replacement     ssssssseseseeeeereereerrsesese 7 15  Detector Amplifier Board Calibration     ssseeeeeeerrsssese 7 17  External Pump Replacement 4 2  2 20 Giuctintiicsoinuemnaciedtincieente dS 7 17  Pressure Board Replacement  lt 422entacee ccseues th
538. sely Right     Page 182 of 424    Annex 5  Manufacturer  Type of instrument    Serial No     Thermo Fisher Scientific    Model 50301 SHARP    SN1 SN2    TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F    Measured values from field test sites  related to actual conditions Page 1 of 11    PM2 5  Measured values in pg m   ACT        Ref  1 Ref 2  Ratio Test site    PM10 PM10 PM2 5 PM10   ugm    _ ug ms           ONOahWNM         740300_2013_936_21209885F_EN docx    3 8 2011  3 9 2011  3 10 2011  3 11 2011  3 12 2011  3 13 2011  3 14 2011  3 15 2011  3 16 2011  3 17 2011  3 18 2011  3 19 2011  3 20 2011  3 21 2011  3 22 2011  3 23 2011  3 24 2011  3 25 2011  3 26 2011  3 27 2011  3 28 2011  3 29 2011  3 30 2011  3 31 2011  4 1 2011  4 2 2011  4 3 2011  4 4 2011  4 5 2011  4 6 2011                   Cologne  Winter    Complete power loss  Recovery after re start    Power loss  only Thermo  Recovery after re start                   TUV Rheinland Energie und Umwelt GmbH  Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Annex 5
539. seseseesssessssesrsstsesesessestsetersesesseseseees 3 9  Main Ment mkeo atie iaer o aAA EATE SE N A EEE enO 3 9   Range Men   hinh taa A E A Saat 3 10  Concentraci  n UNIS sisas ers tiisin ani a aeia seas 3 10  SHARP ATIC Cec 2  sstsatscgsi ennnen e i itii 3 11  Set Custom  RAR SES s cscop dels iostecseaccastieod national aa e N 3 12   Custom RANGES sc fsdee ei  A a RA ATA Ni 3 13   Integration Timed  siaciduetucs dete a a aa 3 13   24 Hour Averigue atit A a E 3 14  24 Hour Average Start TiMe    ssseeessseseseseeresrsreseserreresrsrrsesesreresrees 3 14   Calibration Factors Menu 4c chet tee eld Ca asa a Tec acta ac 3 14  PM Background zeerste naisessa otri a ai aa 3 15  PM Coefficients anie aati Sale ete TS 3 15  SHARP Backgrounds osiers stae cker iaip t 3 16  SHARP Coefficient ics zescfevvevas asda cisascsalesiserat eiTe NE E Eai 3 16  Nephelometer Background Values    s seseseseseseseseseeseeseeesssssessssse 3 16   INephelomieter Mode rror esos Gk Ra 3 17  Vallese E a S Shea A A 3 17  Restore Previous Values       seseseesessssesesesrsssesssisrsisesesesrseeesesrseses 3 17  Nephelometer Coefficient     s sseeessssesssesesersereereisisrsesesrsesesesesesesee 3 18   Instrument Controls Menu secs sca adie sc  peaca fonstytoucaresiontiteceanudeeseeonibenenes 3 18   Set Flow  Pump   4 2 s0  dniiu identi EEE aS 3 19  1 Esp eee eee tee eam Re See tee hae N E A een 3 19  PPO Ps srie a a comedagaedirea ses a seaweeds Aaaa aii 3 19   Set Heaters riii i i 3 20  Sve i n caged tre de ete 
540. settings may be changed via the  front panel user interface  but they can still be changed via remote  operation  When the front panel is locked  the user can still navigate the  menu and view data  instrument parameters  and settings  but nothing can  be changed  The password  lock  icon on the right side of the status bar  indicates that the password lock is active  The items visible under the  Password menu are determined by the instrument   s password status     e Inthe Main Menu  choose Password        Set Password The Set Password screen is used to set the password that is used to unlock  the front panel  and is shown if the instrument is unlocked and the  password is not set     e Inthe Main Menu  choose Password  gt  Set Password             Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 103    Operation  Password Menu    Lock Instrument    Lock Unlock and  Local Remote Operation    Change Password    3 104 Model 50307 SHARP Instruction Manual    The Lock Instrument screen is used to lock the instrument   s front panel so  users cannot change any settings from the front panel  The lock instrument  screen is shown if the instrument is unlocked and the password is set     e Inthe Main Menu  choose Password  gt  Lock Instrument       If the instrument is locked via the front panel using Password  gt  Lock  Instrument  the instrument reports being in Remote mode  In this mode   the front panel is    locked     where data can be viewed  settings can be  vi
541. shooting  Troubleshooting Guides       Malfunction Possible Cause Action  LCD cable loose Check connection and cable  integrity   LCD defective Replace display   The change of Keyboard not enabled Enable keyboard through menu     parameters is not  possible       No mass value    ADC defect  Mains are broken  No high voltage value    Detector defect    Change the main circuit board   Check and repair   Check HV     Change detector        No automatic filter  change    Mechanical defect    If pump is not running    If the filter change parameter  reaches the FC value for the  second time  then  the monitor  stops    Check filter transport motor and  mechanics  Make sure tape is  adhered to take up reel and  intact     Check pump power supply and  wires     Adjust filter change parameters        LCD out of function or  dark    Wrong contrast    LCD defect    Connection cables from the  main board are broken    Adjust contrast     Change LCD   Replace display ribbon cable        Thermo Fisher Scientific    Air flow rate  disturbance or air flow  is too low    Neph Board Failures    Low Source Current    Air flow parameters wrong    No connection with plug  25    gump     Pump do not work properly    Very high pollution or blocking  in the air passage    Check AP calibration valve are  in closed position    External cable not connected  Internal cable not connected     Disconnected Source Ref  cable    Check air flow rate parameters     Put in the plug  pump regulation  input    
542. sion EN 14907  2005    4  Guide    Demonstration of Equivalence of Ambient Air Monitoring Methods      English version of January 2010     5  Operating manual Modell 5030i SHARP  12 March 2013    6  Operating manual LVS3  Stand 2000    7  Directive 2008 50 EC of the European Parliament and of the Council of 21 May 2008  on ambient air quality and cleaner air for Europe    8  Report    UK Equivalence Programme for Monitoring of Particulate Matter      Report No   BV AQ AD202209 DH 2396 of 5 June 2006     9  Technical specification CEN TS 16450     Ambient air     Automated measuring systems  for the measurement of the concentration of particulate matter  PM10  PM2 5      English  Version of May 2013    740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 174 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    9  Appendix  Appendix 1 Measured and calculated values  Annex 1  Detection limit  Annex 2  Temperature dependence of zero point  Annex 3  Temperature dependence of the sensitivity  Annex 4  Dependence on supply voltage  Annex 5  Measured values at the field test sites  Annex 6  Ambient conditions at the field test sites  Annex 7  Measuring Values from the additional campaign Bornhei
543. sion numbers of the  programs installed  Prior to contacting the factory with any questions    regarding the instrument  please note the product model name and the  program version numbers     e Inthe Main Menu  choose Diagnostics  gt  Program Versions     3 58 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Diagnostics Menu       CREE CRE Te    Wi mri   E PURE TE Aaad Fooss PUG Fad LT FO bed HH       Voltages The Voltages menu displays the current diagnostic voltage readings  This  screen enables the power supply to be quickly read for low or fluctuating  voltages without having to use a voltage meter  The I O board item is only    displayed if the I O expansion board option is installed     e In the Main Menu  choose Diagnostics  gt  Voltages        CREED CE pre    Wi mpini  E PURE E Aad feoeo PEO Saad Eee dedi bet HH       Motherboard Voltages The Motherboard screen  read only  is used to display the voltage readings  on the motherboard     e Inthe Main Menu  choose Diagnostics  gt  Voltages  gt  Motherboard        ERED CH  perm  Be Ebb Eeh bias ini    CH io bet  Bis EEE TE Sanat Bone TOP Fae Bast    ade 5 i iNi i       Interface Board Voltages The Interface Board screen  read only  is used to display the voltage  readings on the measurement interface board     e Inthe Main Menu  choose Diagnostics  gt  Voltages  gt  Interface  Board     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 59    Operation  Diagnostics Menu       E a a
544. sol light scattering     nephelometer  and beta attenuation  Model 5014   Beta  to measure  Operation precise and accurate ambient aerosol concentrations  The Model 50302  SHARP aerosol sample pathway is shown in Figure 1 1     Conditioned Aerosol Sample              SHARP Board    SHARP Box shown  with Cover Off    Beta Attenuation    Nephelometer    Figure 1   1  SHARP Monitor Sample Path    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 1 3    Introduction  Principle of Operation    1 4 Model 50307 SHARP Instruction Manual    The SHARP optical assembly senses the light scattered by the aerosol  passing through an 880 nm illumination beam  The nephelometry response  is linear with aerosol concentration  independent of sample flow rate  and a  running one minute average and dynamic average are continuously  calculated  A relative humidity  RH  measurement is located at the sample  filter tape assuring a representative measurement of the aerosol  conditioning prior to real time mass determination     Thereafter  the aerosol is deposited onto a glass fiber filter tape  The filter  tape will accumulate an aerosol sample towards a threshold value   whereupon the filter tape will automatically advance prior to reaching  saturation  During the collection of aerosol onto the filter tape the SHARP  Monitor uses the radiometric principle of beta attenuation through a  known sample area to continuously collect and detect the deposited mass   Additionally  the beta atten
545. span foil test for checking the mass coefficient of the radi   ometric measurement  can only be activated directly on the instrument     6 4 Evaluation    All operating procedures that do not require practical work on site can be monitored by the  user directly on the instrument as well as by telemetric remote control     6 5 Assessment    In principle  all necessary operations for performing a functional check can be monitored di   rectly on the system or via telemetric remote control     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 94 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 2 7 Maintenance interval    The maintenance interval of the measuring system shall be determined during the field  test and specified  The maintenance interval should be three months  if possible  but at  least two weeks     6 2 Equipment  No additional equipment required   6 3 Performance of test    This test was carried out in order to determine  which maintenance procedures are required  at which period to maintain the correct functionality of the measuring system  Furthermore  
546. st of the ambient air quality measuring system   Page 134 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 4 5 Cross sensitivity    The interference caused by moisture in the sample may not exceed 10   of the limit  value in the range of the limit value     6 2 Equipment   No additional equipment required    6 3 Performance of test   The determination of interference caused by moisture in the sample was carried out under  field conditions     For this purpose  the differences between the determined reference value    nominal value   and the measured value of the respective candidate were calculated for days with more than  70   relative humidity during the field test  The average difference was set as conservative  estimate of the interference caused by moisture in the sample     In addition  the reference equivalence functions of both candidates were determined from the  field investigations for days with more than 70   relative humidity   6 4 Evaluation    The average difference between the determined reference value    nominal value  and the  measured value of the respective candidate was calculated for days with more than 70    relative humidity during the field test  The relative deviation from the average concentration  was calculated     Annual limit value PM25   25 g m8 10   of ALV   2 5 g m     Further investigations were made to dete
547. summer      Allsites e  ae 1 970       SN 1   Concentration averages  system SN 1   SN 2   Concentration averages  system SN 2  Concentration averages of systems SN 1  amp  SN 2       e  Ol OI Ol    ges      The single values are presented in annex 5 of the appendix     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control      Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 117 of 424  Scientific for the component suspended particulate matter PMs 5   Report No   936 21209885 F    6 1 5 3 12 Long term drift    The long term drift at zero point and reference point shall not exceed the requirements  of Table 2 of Standard VDI 4202 Part 1  September 2010  in the field test  A value c  at  70   to 80   of the upper limit of the certification range shall be used as reference  point     Note     This item cannot be evaluated according to the current versions of Standards VDI 4202 Part  1  September 2010  and VDI 4203 Part 3  September 2010  when testing particulate meas   uring systems  due to the fact that a minimum requirement has not been defined  By resolu   tion of the competent German body  see module 5 3 1   reference is made to the following  requirement in the previous version of Standard VDI 4202 Part 1  June 2002    The temporal change in the measured value at zero concentration shall not 
548. sure span B 43  vac pres offset Reports the vacuum pressure offset B 44  vf pres cal def Sets vacuum and flow pressure calibration defaults B 44    avg24 time    Send   Receive     The following example shows that the averaging time is set to 18 30     avg24 time  avg24 time 18 30    set avg24 time hh mm  hh   hours  01 to 23     mm   minutes  01 to 59     This command reports the 24 hour averaging time in hours and minutes     This command sets the 24 hour averaging time to start averaging the  concentration from h mm until next 24 hours  It is then set to next 24  hours automatically  The following example sets the 24 hour averaging  time to 17 50     Send   Receive     set avg24 time 17 50  set avg24 time 17 50 ok    Model 50307 SHARP Instruction Manual    B 11    C Link Protocol Commands  Measurements    B 12 Model 50307 SHARP Instruction Manual    flags   This reports 8 hexadecimal digits  or flags  that represent the status of the  mode  pressure and temperature compensation status  concentration units   concentration modes and alarms  To decode the flags  each hexadecimal  digit is converted to binary as shown in the Figure B   1  It is the binary  digits that define the status of each parameter     Send  flags  Receive  flags 00000000    Figure B 1  Flags Field    neph   This command reports the 1 minute running average nephelometer  concentration  The following example reports the 1 minute running  average value     Send  neph  Receive  neph 1 405E 01 pg m3  neph 24a
549. sured value display     4 1 2 Easy mainte    Necessary maintenance of the   Maintenance works can be carried out  nance measuring systems should be   with commonly available tools taking  possible without larger effort  if   reasonable time and effort   possible from outside     4 1 3 Functional check   If the operation or the function    All system functions listed in the  al check of the measuring sys    manual are available  can be activat   tem require particular instru    ed and function well  The current sys     corresponding sub tests and  included in the assessment     4 1 5 Instrument de  Shall be specified in the in   The instrument design specifications  sign struction manual  listed in the operating manual are  complete and correct     4 1 6 Unintended ad   justment  ing system against illicit or un   intended adjustment during   up in a measuring cabinet   operation     4 1 7 Data output    alogue signals  5  Performance requirements    5 1 General The manufacturers specifica   No differences between the instru   tions in the instruction manual   ment design and the descriptions giv   shall be by no means better   en in the manual were detected   than the results of the perfor   mance test        740300_2013_936_21209885F_EN docx    A TUVRheinland      TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the compan
550. sured values from field test sites  related to actual conditions Page 3 of 11  Manufacturer Thermo Fisher Scientific   PM2 5  Type of instrument Model 50301 SHARP Measured values in pg m   ACT     Serial No     SN1 SN2       Test site    5 7 2011 19 1   k Cologne  Winter  5 8 2011 12 3                7 25 2011 12 1     17 8 65 4 Recovery after zero point Bornheim  Summer  7 26 2011 11 8   i 19 1 61 2  7 27 2011 14 3 k i 20 9 65 9  7 28 2011 17 5 5   25 2 70 6 SN2 NEPH malfunction  7 29 2011 10 2     16 6 54 2   SN2 in repair   7 30 2011 SN2 in repair   7 31 2011 9 8 F 14 2 70 4 SN2 in repair   8 1 2011 12 7 19 2 64 8 SN2 in repair   8 2 2011 10 3   20 1 52 1 SN2 in repair   8 3 2011 17 0 26 1 66 5 SN2 in repair   8 4 2011 8 7 14 7 60 2   SN2 in repair   8 5 2011 8 6   18 0 54 7  8 6 2011  8 7 2011 2 8 i   6 4 49 5  8 8 2011 2 8 a a 7 5 44 1  8 9 2011 2 7 k 11 2 27 1 f SN 2 filter tape error  8 10 2011 5 3   12 7 47 0  8 11 2011 5 5 x 1141 47 3  8 12 2011 3 0   x 7 0 50 2  8 13 2011  8 14 2011 5 f   6 7 45 5  8 15 2011      13 9 41 7  8 16 2011 12 9 45 1  8 17 2011 25 0 56 3  8 18 2011    5 15 9 54 1  8 19 2011 12 8 48 8  8 20 2011                         8 21 2011        17 2 60 4       740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Luftreinhaltung    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PMz2 5 pre separator of the company Thermo  Fisher Scientific for the component suspended particul
551. surements with the candidates SN 1   SN 2  measured  component PM25  test site Bornheim  winter     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher  Scientific for the component suspended particulate matter PMz 5     Report No   936 21209885 F    A TUVRheinland      Precisely Right     Page 149 of 424       130  120  110  100  90  80  70  60  50  40  30  20  10  0    Candidate 2  yg m         Model 5030i SHARP  SN 1   SN 2  Teddington  Summer  Raw data                                        0                                           10 20 30 40 50 60 70 80 90 100 110 120 130    Candidate 1  ug m              Measured values      Regression line               y x             Figure 45  Results of the parallel measurements with the candidates SN 1   SN 2  measured  component PMz 5  test site Teddington  summer           Candidate 2  ug m    a  oO    Model 5030i SHARP  SN 1   SN 2  All test sites  218 g m   Raw data          77                p  7                                                                LEE PT ETE                                           10 20 30 40 50 60 70 80 90 100 110 120 130  Candidate 1  ug m            o Measured values      Regression line    o       y x                Figure 46  Results of the parallel measurements with the candidates SN 1   SN 2 
552. t    suitability testing   Bornheim    summer 2013     SN 1  amp  SN 2  component PMz25  slope correction 1 068          Results of Grubbs outlier test     Reference PMoa5        ccccccceeeeeeeeeeeeeeeeeeeeeeeeeess  Discarded reference PM2 5 paired values according to Grubbs               0008  Used filter materials  sristi me m i a i aa  Certification ranges dene ee mee ee ene meee eee ey Pere ee  Determination of availability  without test related outages      Determination of availability  including test related outages              ee  Detection limit PMo5        cccccccccccccseecceceseeeeeeeesceeseeeeeeeeeeeeeeseeneeeeeueeeeeseneeeeees  Dependence of the zero point  SHARP  on surrounding temperature   deviation in g m     average of three measurements    nenene  Dependence of sensitivity  mass coefficient  on surrounding temperature   deviation in    average of three MeasureMens                    seeseeeeseeeeeseeeeeees  Dependence of the zero point  NEPH  on surrounding temperature   deviation in g m     average of three measurements    eene  Dependence of the measured values on mains voltage  deviation in             Concentration averages  standard deviation  uncertainty range and  reproducibility in the field  measured Component PMas              ccccccceeceeeeeeeees  Zero point drift SN 1  amp  SN 2  test with Zero filter                     sseseeseeeeeseeeeeeeees  Sensitivity drift SN 1  amp  SN 2 sce ec tennxedaacbuca ciinespansdansstacsesepanddesndoasenSeausass
553. t  Pin Piri i    ERE CHa  wee  Be Ebb Eesi bias ini    EN ETE EE Sanat Bone TON Fane Bast    ade P       Flow The Flow screen is used to view and adjust the flow     e Inthe Main Menu  choose Instrument Controls  gt  Set Flow Pump  gt   Flow     set    ERED CH  Pa  Tit ob popin pei    PPP LES Pied Bei PERI PLP PRE       Pump The Pump screen allows the user to toggle the pump to either ON or OFF     e Inthe Main Menu  choose Instrument Controls  gt  Set Flow Pump  gt   Pump     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 19    Operation  Instrument Controls Menu       DH pikari  Ce a E S a EA          Set Heater The Set Heater menu allows the user to set the status of the heater  RH  threshold  and temperature threshold in various modes     e Inthe Main Menu  choose Instrument Controls  gt  Set Heater        DH pikari  E a E SO EA          Control The Control screen is used to set the status of the heater to either  ON OFF  RH threshold  or TEMP threshold     e Inthe Main Menu  choose Instrument Controls  gt  Set Heater  gt   Control           Cachet    EEE EE uat Bone BME UY Saad    OPE T EEE of          RH Threshold The RH Threshold screen is used to change the RH threshold value in  percent  The RH threshold is typically compatible with the gravimetric  method being used for comparison     e Inthe Main Menu  choose Instrument Controls  gt  Set Heater  gt  RH    Threshold     3 20 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation 
554. t  channels at both zero and full scale  The analog input calibration menu is  visible only when the I O expansion board is installed and when the  instrument is in service mode  For more information on the service mode   see    Service Mode    earlier in this chapter     Note This adjustment should only be performed by an instrument service  technician  A    e Inthe Main Menu  choose Service  gt  Analog Input Calibration     3 100 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Operation  Service Menu       ERED CH  perm  if    EN EUH TE Sanat Bone ETET Aad Bast    Wi mid          PRES CH  mT  Bet Ebb Eesi HERTE i    EH ninii  BPs EEE TE ueet Bone EEY Fad Bast    ade P i iNi i       Analog Input Calibrate The Analog Input Calibrate Zero screen allows the user to calibrate the  Zero zero state of the selected analog input     e Inthe Main Menu  choose Service  gt  Analog Input Cal  gt  select    Channel  gt  Calibrate Zero     CRIED Agi  THT gy mpi       PPP L Pil Bei PERI PLP PRE    Analog Input Calibrate The Analog Input Calibrate Full Scale screen allows the user to calibrate  Full Scale the full scale state of the selected analog input     e Inthe Main Menu  choose Service  gt  Analog Input Cal  gt  select  Channel  gt  Calibrate Full Scale   Hook up a voltage source of 10 V  to the analog input channel             Sead Bast    abe POE Sek come  EE fee PEE SEE    EUs EEE TE Sanat Beee       Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 1
555. t Analog Outputs        PME ieee EE i       The Set Analog Outputs screen contains three choices  Set to full scale  set  to zero  or reset to normal  Full scale sets the analog outputs to the full   scale voltage  zero sets the analog outputs to 0 volts  and reset returns the  output to normal operation  The analog outputs are returned to normal    Thermo Fisher Scientific    Operation  Diagnostics Menu    operation upon exiting this screen  The following example shows the screen  when all analog outputs are set to    normal    operating mode     e Inthe Main Menu  choose Diagnostics  gt  Test Analog Outputs  gt   ALL  Voltage Channel 1   6  or Current Channel 1 6     Parma       Instrument Configuration The Instrument Configuration screen displays information on the    hardware configuration of the instrument     Note If the analyzer is in service mode  pressing will toggle the    selected item between yes and no  A    e Inthe Main Menu  choose Diagnostics  gt  Instrument  Configuration        Contact Information The Contact Information screen displays the customer service information     e Inthe Main Menu  choose Diagnostics  gt  Contact Information        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 65    Operation  Alarms Menu    Alarms Menu The alarms menu allows the user to view a set of alarms for different boards  and displays a list of items that are monitored by the analyzer  The number  of alarms detected is displayed to indicate how many alarms
556. t Equipment Required   Transformer assembly    Philips screwdriver   2    f Equipment Damage Some internal components can be damaged by small    amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 7 37    Servicing  Radius Tube Replacement    Radius Tube  Replacement    A    7 38 Model 50307 SHARP Instruction Manual    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Disconnect cables from the measurement interface board and  motherboard     3  Unscrew the four mounting screws and lift out the transformer  assembly     4  Install the new transformer by following the previous steps in reverse     Unhook Short Cable from  Transformer from Common  Electronic Board    AC INT BD    Unhook Long Cable from  Transformer to back of  Unfasten Screws  4  Measurement Interface Board          Figure 7 20  Replacing the Transformer    Use the following procedure to replace the radius tube  Figure 7   21    Equipment Required    Radius tube assembly   Hex driver  9 16 inch   Hex driver  7 64 inch    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    Thermo Fisher Scientific    Servicing  Ambient RH Temperature Replacement    1  Turn the instrument OFF  unplug the power cord  
557. t alarm conc inst pm max 10000  Receive  set alarm conc inst pm max 10000 ok    alarm conc inst sharp min   alarm conc inst sharp max   These commands report the current 24 hour instant SHARP concentration  alarm minimum and maximum values setting  The following example  reports that the instant SHARP concentration minimum is 0 pg m        Send  alarm conc inst sharp min  Receive  alarm conc inst sharp min 0 000E 00 pg m3    set alarm conc inst sharp min value   set alarm conc inst sharp max value   These commands set the 24 hour instant SHARP concentration alarm  minimum and maximum values to value  where value is a floating point    Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Alarms    representation of the concentration alarm limits  Values must be in the  units that are currently set for use  The following example sets the instant  SHARP concentration alarm maximum value to 10000 pg m        Send  set alarm conc inst sharp max 10000  Receive  set alarm conc inst sharp max 10000 ok    alarm filter tape status  This command reports the status of the filter tape as FAIL or OK  The  following example reports that the filter tape status is OK     Send  alarm filter tape status  Receive  alarm filter tape status ok    alarm flow min   alarm flow max   These commands report the current flow alarm minimum and maximum  value settings  The following example reports that the flow alarm minimum  value is 16 0 LPM     Send  alarm flow min  Receive
558. t be installed in horizontal mounting position  e g  on a bench or a  rack  protected from weather inconsistencies  The temperature at the installation site shall be  within the range of 5   C to 40   C with a relatively low amount of temperature fluctuation  no  direct sunlight  no direct exposure to heater or air conditioner      The dimensions and weights of the measuring system conform to the specifications in the  manual     The power consumption of the measuring system and pump is specified by the manufacturer  as a maximum of approx  880 W  During a 24 hour test  these specifications were checked   At no time the mentioned value was exceeded during these investigations     6 5 Assessment    The instrument design specifications listed in the operating manual are complete and correct   Minimum requirement fulfilled  yes    6 6 Detailed presentation of test results  Not required for this test     740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 84 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 4 1 6 Unintended adjustment    It shall be possible to secure the adjustment of the measuring system against illicit or  unintended adjustment during operation     6 2 Equipment  
559. t boards with a D Sub  connector on one side and a series of screw terminals on the other side   These assemblies provide a convenient mechanism for connecting wires  from a data system to the analyzer   s I O connectors     The following terminal board PCB assemblies are available for iSeries  instruments     e T O terminal board PCB assembly  37 pin  standard   e D O terminal board PCB assembly  37 pin  standard     e 25 pin terminal board PCB assembly   included with optional I O  expansion board     Figure 2   8 shows the recommended method for attaching the cable   customer supplied  to the terminal board using the included tie down and  spacer  Table 2   2 identifies the connector pins and associated signals     Note Not all of the I O available in the instrument is brought out on the  supplied terminal board  If more I O is desired  an alternative means of  connection is required  See optional    Terminal Block and Cable Kits     A    Model 50307 SHARP Instruction Manual 2 17    Installation  Connecting External Devices    See Detail    A       See Detail    B          Assembled Connector    2 18 Model 50307 SHARP Instruction Manual       Figure 2 8    O Terminal Board Views    Table 2 2    0 Terminal Board Pin Descriptions       Detail    A    Detail    B                                     Screw Signal Description Screw Signal Description  1 Analog1 13 Power_Fail_NC   2 Analog ground 14 Power_Fail_COM  3 Analog2 15 Power_Fail_NO  4 Analog ground 16 TTL_Input1   5 Ana
560. t checks can be found in Section 6 1  5 3 12 Long term drift of this report     Minimum requirement fulfilled  yes  6 6 Detailed presentation of test results  Please refer to Section 6 1 5 3 12 Long term drift    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land     Air Pollution Control   s  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 81 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 4 1 4 Set up times and warm up times  The set up times and warm up times shall be specified in the instruction manual   6 2 Equipment  A clock was required for this test   6 3 Performance of test    The AMS were put into operation in accordance with the manufacturer   s description  The  necessary set up and warm up times were recorded separately    Constructional measures undertaken prior to the measurement  such as the opening of an in   let through the cabinet roof  were not taken into account for evaluation     6 4 Evaluation    The set up time comprises the time needed for all necessary works from system installation  to start up    The measuring system must be protected from weather inconsistencies  e g  in an air   conditioned measuring cabinet  Moreover  extensive on site construction work is required in  order to lead the sampling tube through the roof 
561. t of problems  See    Service Locations    on page 6   25 for contact information  In any correspondence with the factory  please  note both the serial number and program number of the instrument     This chapter provides the following troubleshooting and service support  information     e    Safety Precautions    on page 6 1   e    Troubleshooting Guides    on page 6 1   e    Board Level Connection Diagrams    on page 6 8  e    Connector Pin Descriptions    on page 6 10    e    Service Locations    on page 6 25    Read the safety precautions in the Preface and    Servicing    chapter before  performing any actions listed in this chapter     The troubleshooting guides presented in this chapter are designed to help  isolate and identify instrument problems     Table 6 1 provides general troubleshooting information and indicates the  checks that you should perform if you experience an instrument problem     Table 6 2 lists all the alarm messages you may see on the graphics display  and provides recommendations about how to resolve the alarm condition     Model 50307 SHARP Instruction Manual 6 1    Troubleshooting  Troubleshooting Guides    Table 6 1  Troubleshooting   General Guide    Malfunction    No power    Possible Cause    Main fuses are blown are  missing    Digital electronics defective    Action   Check voltages from power  supply    Check that all boards and  connectors are seated properly     Replace with spare boards to  isolate the problem        No automatic filte
562. t selections may not apply  A    Table B 13  Default Analog Output Channel Assignments                      DtoA Function Range   1 Voltage Output PM   2 Voltage Output AVG PM   3 Voltage Output Flow   4 Voltage Output Ambient Temperature   5 Voltage Output Flow Temperature   6 Voltage Output Ambient Relative Humidity  7 Current Output PM       Model 50307 SHARP Instruction Manual B 63    C Link Protocol Commands  1 0 Configuration    B 64 Model 50307 SHARP Instruction Manual                   DtoA Function Range   8 Current Output AVG PM   9 Current Output Flow   10 Current Output Ambient Temperature   11 Current Output Flow Temperature   12 Current Output Ambient Relative Humidity  list din  list dout    These commands report the current selection for the digital inputs or the  digital outputs in the format  Output no Index number variable name  active state  The active state for digital outputs is open or closed  The active  state for digital inputs is high or low     Send  list dout  Receive  list dout  output index variable state  1 5 CONC ALARM open  2 1 LOCAL REMOTE open  3 3 UNITS open    list var aout   list var dout   list var din   list var log   These commands report the list of possible index numbers  and the  variables  associated with that index number  available for selection in the  current mode for analog outputs  digital outputs  digital inputs and  datalogging  The index number is used to insert the variable in a field  location in a list using    set sp f
563. taee wes B 51  Table B 7  Allow Mode Command Values             cccccecceccesceseeeeseeeetesteseeseens B 56    Model 50307 SHARP Instruction Manual xxiii    Tables    xxiv Model 50307 SHARP Instruction Manual    Table B 8  Reply Termination Formats 00 0 0    cccecccececeeseecetesesesteteseseetetenees B 58  Table B   9  Set Layout Ack Va lU S      cccccecccccccccsscstesesteseseseeseseseseseseseseeneees B 59  Table B   10  Power Up Mode Values    ccccceccccsceceesestetesesesteteseseseenenens B 60  Table B 11  Analog Current Output Range Values 0 0    cece ceeeceneeeens B 61  Table B 12  Analog Voltage Output Range Values oo    cece cecseeteeeteees B 62  Table B 13  Default Analog Output Channel Assignments            0  cccccee B 63  Table C 1  Read Coils for S030 0 secctncncentetwadaente ernie C 8  Table C 2  Read Registers for 50308     ecsccscccseseceesccseccssesccsenccsescesenessere C 9  Table C   3  Write Coils for 50300 oo    cecseeceesteseeseeseesestesteseeseenseteseeseens C 11  Table D 1  Operating Status for Model 50300    ccccccceceteseeteeeeteeee D 8  Table D 2  Error Status for Model 50303    cceccecceeeectesteseeseeseeesteseeseenee D 8  Table E 1  Read Commands for 50302 0 0    cecccccccsceseeseeeetestesteseeeeeetesteseesee E 1  Table E   2  Write Commands for 50303 occ eeseceeteetesteseeseeeeteeteseesee E 2  Table E   3  Control Commands for 50307    cccecceeeseeseeteeteseeseeeeteetesteseesee E 2    Thermo Fisher Scientific    Thermo Fisher Scientific    Chapter 1
564. te measuring systems   6 2 Equipment   Not applicable    6 3 Performance of test   Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VRheinland      Air Pollution Control  gt     Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 107 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 3 7 Sensitivity coefficient of surrounding temperature    The sensitivity coefficient of surrounding temperature at zero and reference point shall  not exceed the requirements of Table 2 of Standard VDI 4202 Part 1  September  2010   A value c  at 70   to 80   of the upper limit of the certification range shall be  used as reference point     Note     This item cannot be evaluated according to the current versions of Standards VDI 4202 Part  1  September 2010  and VDI 4203 Part 3  September 2010  when testing particulate meas   uring systems  due to the fact that a minimum requirement has not been defined  By resolu   tion of the competent German body  see module 5 3 1   reference is made to the following  requirement in the previous version of Standard VDI 4202 Part 1  June 2002      The temperature de
565. ted by  spaces and the line is terminated by a     n     the normal line separator    character   Valid fields are      s   parse a string    d   parse a decimal number    ld   parse a long  32 bit  decimal number   f   parse a floating point number    x   parse a hexadecimal number    lx   parse a long  32 bit  hex number        ignore the field    Note Signed versus unsigned for the integer values does not matter  it is  handled automatically  A    The second line of the layout response is the binary parameter list for  parsing the fields from a binary response  Parameters MUST be separated  by spaces  and the line is terminated by a   n   Valid fields are     t   parse a time specifier  2 bytes    D   parse a date specifier  3 bytes    i   ignore one 8 bit character  1 byte    e   parse a 24 bit floating point number  3 bytes  n x     Thermo Fisher Scientific    Format Specifier for    Thermo Fisher Scientific    Erec Layout    Text    Value String    Value Source    C Link Protocol Commands  Record Layout Definition    E   parse a 24 bit floating point number  3 bytes  N x   f   parse a 32 bit floating point number  4 bytes     c   parse an 8 bit signed number  1 byte    C   parse an 8 bit unsigned number  1 byte   n   parse a 16 bit signed number  2 bytes    N   parse a 16 bit unsigned number  2 bytes   m   parse a 24 bit signed number  3 bytes   M   parse a 24 bit unsigned number  3 bytes       parse a 32 bit signed number  4 bytes    L   parse a 32 bit unsigned number 
566. ters for 5030                                      Register Number Variable   0 INVALID  1 amp 2 PMa   3 amp 4 NOT USED  5 amp 6 BARO PRES  7 amp 8 VACUUM  9 amp 10 NOT USED  11 amp 12 FLOW PRES  13 amp 14 NOT USED  15 amp 16 AMB RH  17 amp 18 SAMP RH  19  amp  20 NOT USED  21 amp 22 AMB TEMP    Model 50307 SHARP Instruction Manual C 9    MODBUS Protocol  MODBUS Parameters Supported                                                                                                                Register Number Variable   23  amp  24 BOARD TEMP  25  amp  26 FLOW TEMP  27  amp  28 NOT USED  29  amp  30 NOT USED   31  amp  32 NOT USED  33  amp  34 FLOW VOL  35  amp  36 NOT USED  37  amp  38 BETA   39  amp  40 ANALOG IN 1  41  amp  42 ANALOG IN 2  43  amp  44 ANALOG IN 3  45  amp  46 ANALOG IN 4  47  amp  48 ANALOG IN 5  49  amp  50 ANALOG IN 6  51  amp  52 ANALOG IN 7  53  amp  54 ANALOG IN 8  55  amp  56 BETA RAW  57  amp  58 ALPHA RAW  59  amp  60 AVG PM   61  amp  62 NOT USED  63  amp  64 MASS   65  amp  66 NOT USED  67  amp  68 BETA ZERO  69  amp  70 NOT USED   71  amp 72 NEPH   73  amp 74 NOT USED  75  amp  76 NOT USED  77  amp 78 NOT USED  79  amp  80 NOT USED   81  amp  82 NOT USED  83  amp  84 SHARP   85  amp  86 NOT USED  87  amp  88 AVG SHARP  89  amp  90 NOT USED             C 10 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Thermo Fisher Scientific    MODBUS Protocol  MODBUS Parameters Supported                                           Register Numbe
567. test results  Not applicable     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland     gt     Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 114 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 3 10 Averaging effect    For gaseous components the measuring system shall allow the formation of hourly av   erages     The averaging effect shall not exceed the requirements of Table 2 of Standard VDI  4202 Part 1  September 2010      Note    This test is not relevant for particulate measuring systems   6 2 Equipment   Not applicable    6 3 Performance of test   Not applicable    6 4 Evaluation   Not applicable    6 5 Assessment    Not applicable   Minimum requirement fulfilled       6 6 Detailed presentation of test results  Not applicable     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH     T  VR hei n land     Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 115 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    6 1 5 3 11 Standard deviation from paired measurement    The standard deviation fro
568. test site Teddington 2 2   ci aca etieaendiie eee ais dope ee dietieeeenie 60    Comparison of reference and candidate SN 1  measured component  PM25  Teddington  summer   relative humidity threshold adjusted to 50        63  Comparison of reference and candidate SN 2  measured component  PM25  Teddington  summer   relative humidity threshold adjusted to 50        63    Grubbs test results for the PM25 reference method  Cologne  winter                70  Grubbs test results for the PM  5 reference method  Bornheim  Summer          71  Grubbs test results for the PM  5 reference method  Bornheim  winter              71  Grubbs test results for the PM25 reference method  Teddington  summer       72  Display of the measured SHARP concentration           ccccccceeeeeeeeeeeeeeeeeeeeeeeeeeees 76  Rear view of Model 50301 SHARP            cccccccccecceeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeeess 86  Software version  here V02 00 00 232   displayed by selection of the   menu    Diagnostics   Program Versions              cccccceeeeeeeeeeeeeeeeeeeeenneeeeeeeeeeeeees 98  Zero point drift SN 1  measured Component PMo5              ccceeeeeeeeeeeeetteeeeeeees 120  Zero point drift SN 2  measured Component PMo5              ccceeeeeeeeeeeeeteeeeeeeees 120  Drift of the measured value SN 1  measured component PMos          ccccccceee 122  Drift of the measured value SN 2  measured component PMos           cccccceeee 122  Flow of candidate SN 1  field  siicieicsstietievanie Maaeaniiee eae 14
569. th a hose to the  air inlet     Test sites and arrangement of the measuring systems    During the field test  the measuring systems were arranged in such a way that the sampling  inlets were positioned outside the measuring cabinet above its roof  The central units of both  candidates were located inside the air conditioned measuring cabinet  The entire reference  equipment  LVS3  was placed outdoors on the roof of the cabinet     The field test was carried out at the following test sites                    Table 5  Field test sites  No  Measurement site Period Characteristics  1 Cologne  parking 03 2011     05 2011 Urban background  lot  winter  2 Bornheim  motor  07 2011     11 2011 Rural structure   traffic  way parking lot  influence  summer  3 Bornheim  motor  01 2012     03 2012 Rural structure   traffic  way parking lot  influence  winter  4 Teddington  07 2012     10 2012 Urban background  summer                      740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland      Air Pollution Control      Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 57 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    Figure 19 to Figure 22 show the course of PM25 concentrations at the field test sites obtained  with the reference systems                             70  a 
570. the  analyzer  and to trigger or simulate the activation of a digital input to the  instrument  This is achieved by using the supported MODBUS parameters  listed below     Up to three simultaneous connections are supported over Ethernet     For details of the Model 5030i MODBUS Protocol specification  see the  following topics     e    Serial Communication Parameters    on page C 1  e    TCP Communication Parameters    on page C 2  e    Application Data Unit Definition    on page C 2  e    Function Codes    on page C 3   e    MODBUS Parameters Supported    on page C 8    Additional information on the MODBUS protocol can be obtained at  http   www modbus org  References are from MODBUS Application  Protocol Specification V1 1a MODBUS IDA June 4  2004     The following are the communication parameters that are used to configure    the serial port of the iSeries to support MODBUS RTU protocol     Number of Data bits  7 or 8   Number of Stop bits  lor2   Parity   None  Odd  or Even   Data rate   1200 to 115200 Baud  9600 is default     Model 5030  SHARP Instruction Manual C 1    MODBUS Protocol  TCP Communication Parameters    TCP Communication  Parameters    Application Data  Unit Definition    Slave Address    MBAP Header    C 2 Model 50307 SHARP Instruction Manual    iSeries instruments support the MODBUS TCP protocol  The register  definition is the same as for the serial interface  Up to three simultaneous  connections are supported over Ethernet     TCP connection port for 
571. the  sensors with regard to ambient air temperature  flow temperature  ambient humidity  air  pressure and flow  Furthermore  it allows for checking   adjusting the radiometric determina   tion   mass coefficient  calibrating the detector  to be performed by a service technician of the  manufacturer only  and nephelometer as well as setting up the analogue inputs and outputs     The    Password    sub menu allows the user to configure and activate password protection for  the instrument  the lock icon on the main screen indicates that password protection is active      740300_2013_936_21209885F_EN docx    Precisely Right   Page 45 of 424    A TUVRheinland      Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher  Figure 15 provides an overview of the menu assisted firmware     Scientific for the component suspended particulate matter PM2 5     Report on the suitability test of the ambient air quality measuring system  Report No   936 21209885 F    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    opu yOHRUCD  a209   ypa  pubog vosuedg OA  uogunS yuo g weUINs SU  Di PULE IDLA  TW  sndyno Bowuy ise   Or NaNO  sons Ajoy  9 4 pdu  synduj pey Big   1 ul Gopuy  sedey on dul Sojeuy  swuey 2009 Si NI  synyog Hope j aozsoy suey Moy sSulpeey ynduj Bojeuy  synepq sn aozsoy NO uyo  390  joxig   Bjdsia WNNOEA 20508900  JOY  B  OUD mdu sax oeg qweun9 037  voRne ges ndu  Gojeuy suue WNnoByyelnsseld Hai HAIN  Si puu waung dwa   peog dwal HAIN  Ot PWO HUN AL
572. the 8 analog input    3 52 Model 50307 SHARP Instruction Manual    channels available for configuration  This screen is only displayed if the I O  expansion board option is installed  Configuration includes entering the  Descriptor  Units  Decimal Places  choice of 1 10 points in the table  and  corresponding number of points selected     Thermo Fisher Scientific    Operation  Volumetric Conditions    e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration     gt  Analog Input Config        Descriptor The Descriptor screen allows the user to enter the descriptor  or name  for  the selected analog input channel  The descriptor is used in datalogging  and streaming data to report what data is being sent out  The descriptor    may be from 1 to 3 characters in length  and defaults to IN1 to IN8  user  input channel number      e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  Descriptor        Units The Units screen allows the user to enter the units for the selected analog  input channel  The units are displayed on the diagnostic screen and in    datalogging and streaming data  The units may be from 1 to 3 characters in  length  and defaults to V  volts      e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Input Config  gt  select Channel  gt  Units     Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 53    Operation  Volumetric Conditions       PGi 
573. the instrument sample zero air for about 90 minutes     2  From the Main Menu  choose Service  gt  Detector Calibration  gt  Auto     The Auto Detector Calibration screen appears     Note If Service Mode is not displayed  refer to    Accessing the Service  Mode    on page 7 4  then return to the beginning of this step  A    3  At the Detector Calibration screen  press to start the auto    detector calibration     Use the following procedure to replace the pump  Figure 7   8    Equipment Required    Pump   Adjustable wrench    Equipment Damage Some internal components can be damaged by small  amounts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  A    1  Turn the instrument OFF  unplug the power cord  and remove the  cover     2  Disconnect the pump power cable from the rear panel     3  Remove both lines from the pump     Model 50307 SHARP Instruction Manual 7 17    Servicing  Pressure Board Replacement    4  Install the new pump by following the previous steps in reverse     5  Perform a leak test as described in the    Preventive Maintenance     chapter        Connect SHARP Valve  Port 1        Connect to Rear Panel Elec Pump    Connect to Rear                    gt   Panel    VACUUM       Figure 7 8  Replacing the Pump    Pressure Board Use the following procedure to replace the pressure board  Figure 7   9    Replacement Equipment Required   Pressure board    Philips screwdriver   2    amounts of stat
574. the pump power     To continue with the acceptance testing     e From the Main Menu select Diagnostics  gt  RH Temperature              premik prn in Ph dm    PrP LES E bii i    pHo mnie        The RH Temperatures screen  read only  displays the current relative  humidity in percent and the temperature readings in   C  The board  temperature is the air temperature measured by a sensor located on the  measurement interface board     Model 50307 SHARP Instruction Manual 2 5    Installation  Acceptance Testing and Startup Procedures    e Press        to return to the Diagnostics menu  and choose  Pressure Vacuum        rp Guided  Pii ERE i          The Pressure Vacuum screen  read only  displays the barometric  vacuum   and flow pressure readings in mmHg  The pressure is measured by a  pressure transducer  The vacuum is the amount of vacuum under the filter  tape  The flow pressure is the pressure differential across the orifice     Table 2 1  Pressure Units Conversion  Pressure Conversions  mmHg   atm x 0 75006  mmHg   inHg x 25 4  mmHg   hPa x 0 75006       Assuming adequate time has passed for thermal equilibration   approximately 1 hour   and the vacuum pump has been drawing room air  into the instrument  compare the ambient temperature  sample  temperature  and flow temperature to your NIST traceable thermometer  placed at the inlet tube  see Figure 2   2      Note Please be sure that the pump exhaust or any other heat source is not  influencing the sensor reading of the te
575. the volumetric flow rate measurement with an  accuracy of  2 percent  This flow rate standard must be a separate   stand alone device  It must have its own certification and be traceable  to a NIST primary standard for volume or flow rate  Dry piston meters  and bubble flow meters should only be used under controlled  laboratory conditions  Ambient field measurements should use a NIST   traceable low pressure drop orifice venturi flow meter  such as  BGI  Delta Cal  or a Streamline Pro    Model SX Kit  2 25 L min      e A relative humidity  RH  standard capable of measuring in the range of  35 75  RH  readable to the nearest 0 5  RH and accurate to within  2  RH     Prior to calibration  be sure the Model 5030  is operating properly  The  Model 50307   s internal diagnostics makes this a quick and simple process   Turn on the instrument and allow it to stabilize for one hour prior to  calibrating  If you are receiving this instrument for the first time  the  optional acceptance testing in the    Installation    chapter should also be  reviewed  If the internal flow temperature sensor requires calibration  the  dynamic heating system should be turned OFF  the cover should be  removed  and the instrument should sample room temperature air at least  one hour prior to calibrating     The order of calibration described below is preferred for optimal  performance of the 5030i and comes from many years of experience  working with and developing ambient particulate monitoring  instru
576. tion  gt  Neph RH        a pe  a         The Nephelometer Temperature screen allows the user to view and  calibrate the nephelometer temperature sensor calibration     Note Calibration of the nephelometer temperature must be done with the  heater in the    OFF    position  A    Model 50307 SHARP Instruction Manual 4 13    Calibration  Calibration Procedure    e Inthe Main Menu  choose Service  gt  Neph Calibration  gt  Neph Temp        Nephelometer Source The Nephelometer Source Level screen allows the user to view and  Level calibrate the source level  This should be within 60 70 mA  If it is outside  of this range  please adjust the   SRC Level until the IRED reads 65 mA    e Inthe Main Menu  choose Service  gt  Neph Calibration  gt  Neph Src  Level        4 14 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Safety    Precautions    Thermo Fisher Scientific    A  A    Chapter 5  Preventive Maintenance    This chapter describes the periodic maintenance procedures that should be  performed on the instrument to ensure proper operation  Since usage and  environmental conditions vary greatly  you should inspect the components  frequently until an appropriate maintenance schedule is determined     This chapter includes the following preventive maintenance information   e    Safety Precautions    on page 5 1  e    Replacement Parts    on page 5 2  e    Cleaning the Outside Case    on page 5 2  e    Cleaning the Inlets    on page 5 2  e    Fan Filter Inspection and
577. tion of test results  Please refer to Section 6 4 of this module     740300_2013_936_21209885F_EN docx    a      TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland j    Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 88 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    6 1 5 2 1 Certification ranges  The certification range intended for testing shall be determined   6 2 Equipment  No additional equipment required   6 3 Performance of test  The certification range intended for testing shall be determined   6 4 Evaluation    Standard VDI 4202  Part 1 contains the following minimum requirements on certification  ranges for systems measuring suspended particulate matter in ambient air     Table 15  Certification ranges       Measured com    Lower limit   Upper limit   Limit value  ponent of CR of CR    in pg m  in pg m  in pg m     The certification ranges are related to the limit value with the shortest assessment period and  used for the assessment of the measuring system in the range of the limit value  This as   sessment of the measuring system in the range of the limit value is performed as part of the  determination of the expanded uncertainty of the candidates according to Guide  4   For this  purpose  the following values are used as reference values in accordance with
578. tnnneseeerenennn ennet 79  6 1 4 1 4 Set up times and warm up tiMeS           ssssseeseeessserenrrrrtrsstrrtrrnnnesstrrrnernnnnsnneeeet 81  6 1 4 1 5 Instrument design eee arte cess ee ctecee pata eens a Bcesertnetceackacteaneu na pactettenmeigess 83  6 1 4 1 6 Unintended adjustment sicasi2sedixctiviatesisseSincidetaieetee de scaeteeie pe asad cee 84  6 1 4 1 7 Data output ase crecctacestenenesad macecnniectsmiadeendstieunebeadmeenaiananiorenciecenuataaseumenbiedemtetaane 85    740300_2013_936_21209885F_EN docx    a      TUV Rheinland Energie und Umwelt GmbH  A TUVRheinland Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system    Page 6 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    6 1    Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    RE 210   2  ee   eemeeeree aera ver om eer Er ee ee ee ee 87  5 2 1 Certification ranges see cto asd dened aia det eementeeeedsecsatia ys oe end eciee se aeneneeecaas 88  B22 Measuring ANOS scrire eaa Ee EEA E EN EAEE AEREAS 89  5 2 3 Negative output signals             sseesseseenererersserettrrrrtrtssrrtnnrrnnrrnnnresertnnnrnnnnnseeeenne 90  5 2 4 Failure in the mains voltage            sssessneneeeessesreerrrnreessrre
579. tor of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 201 of 424  Report No   936 21209885 F                      Annex 6 Ambient conditions at field test sites Page 9 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation  m5  ym3   A  C  C  hPaj A  m s mm   241 7 31 2012 Teddington  Summer 9 3 2 3 25 0 17 2 21 0 994 84 5 0 4 170 0 5  242 8 1 2012 5 9 2 0 33 9 18 4 23 3 989 75 5 0 7 185 0 0  243 8 2 2012 5 4 1 5 28 4 17 7 22 6 993 73 1 0 8 195 0 0  244 8 3 2012 1 7 17 8 22 7 993 71 6 0 8 180 2 3  245 8 4 2012 5 1 2 3 46 1 16 4 21 0 990 80 7 0 6 200 1 8  246 8 5 2012 47 22 45 7 17 3 22 4 988 77 6 0 4 210 2 3  247 8 6 2012 3 5 0 5 15 5 16 0 23 3 995 77 5 0 2 190 0 8  248 8 7 2012 3 4 1 5 44 9 17 2 21 1 1003 80 7 0 3 201 0 8  249 8 8 2012 8 1 2 3 28 8 19 0 25 5 1005 73 0 0 2 263 0 0  250 8 9 2012 11 5 2 7 23 5 21 0 28 3 1007 62 7 0 4 202 0 0  251 8 10 2012 13 3 4 4 32 8 21 4 29 4 1005 59 5 0 5 142 0 0  252 8 11 2012 15 1 5 0 33 4 19 8 24 4 997 63 9 1 9 78 0 0  253 8 12 2012 72 3 2 44 6 19 8 26 1 992 70 6 0 8 171 0 0  254 8 13 2012 47 2 5 53 9 19 2 23 2 992 81 2 0 7 184 1 5  255 8 14 2012 aa 3 0 38 6 21 1 26 1 991 66 3 0 9 139 0 0  256 8 15 2012 hell 21 29 8 19 0 23 3 989 71 8 1 0 187 0 5  257 8 16 2012 6 6 2 0 31 1 19 8 22 9 996 65 3 11 160 0 0  258 8 17 2012 6 7 2 2 32 2 23 4 2
580. trument is set to ignore the commands  value   0   the instrument  will respond with    ok    as if the command has been accepted and acted  upon  but will not change the instrument lock status  this is for  compatibility with systems expecting and    ok    response      Note The instrument will always respond to the command    mode    with  the status of the password lock as    mode local    or    mode remote      regardless of the above setting  A    The following example sets the instrument to accept the    set mode local     and    set mode remote    commands     Send  set allow mode cmd 1  Receive  set allow mode cmd 1 ok    Table B 7  Allow Mode Command Values       Value Allow Mode Command  0 Ignore  default   1 Accept   baud    This command reports the current baud rate for the serial port   RS232 RS485   The following example reports that the current baud rate  is 9600 baud     Send  baud  Receive  baud 9600    set baud rate  rate     1200   2400   4800   9600   19200   38400   57600   115200      Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Communications Configuration    This command sets the instrument baud rate to rate  The following  example sets the instrument   s baud rate to 9600     Note After the command is sent  the baud rate of the sending device must  be changed to agree with the instrument  A    Send  set baud 9600  Receive  set baud 9600 ok  dhcp    This command reports the current state of use of the Dynamic Host  
581. ts current output channel 4 is in the 4 20 mA range  according to  Table B 11  This command responds with    feature not enabled    if the I O  expansion board is not detected     Send  analog iout range 4  Receive  analog iout range 4 2    set analog iout range channel range   This command sets analog current output channel to the range where  channel is between 1 and 6 inclusive  and range is set according to Table B   11  The following example sets current output channel 4 to the 0 20 mA  range  This command responds with    feature not enabled    if the I O  expansion board is not detected     Send  set analog iout range 4 1  Receive  set analog iout range 4 1 ok    Table B 11  Analog Current Output Range Values          Range Output Range  1 0 20 mA   2 4 20 mA   0  cannot be set to this  but may report  Undefined    analog vin channel   This command retrieves the analog voltage input channel data  both the  calculated value and the actual voltage  In the following example  the     calculated    value of channel 1 is 75 325 degrees F  volts are 2 796  This  command responds with    feature not enabled    if the I O expansion board  is not detected     Send  analog vin 1  Receive  analog vin 1 75 325  2 796 V    analog vout range channel   This command reports the analog voltage output channel range  where  channel is between 1 and 6 inclusive  according to Table B 12  The  following example reports that analog voltage output channel 2 is set to 3     0 10 V      Model 50
582. ts of static electricity  A properly grounded antistatic wrist strap  must be worn while handling any internal component  For more  information about appropriate safety precautions  see the    Servicing     chapter  A    1  Turn the instrument OFF  unplug the power cord  and remove the  SHARP optics assembly  Refer to    Removing the SHARP Optics  Assembly    on page 7 11     2  Remove the two philips screws and remove the SHARP optics cover   Refer to    Removing the SHARP Optics Cover    on page 7 44     3  Disonnect the RH Temp connector from the SHARP interface board   4  Remove the eight socket head screws from the upper end plate   5  Gently remove the end plate  Carefully remove the sealing gasket     6  Carefully vacuum out the optics chamber using caution not to scrap  the interior of the optics chamber     7  Re assemble following the previous steps  Take extra caution in re   aligning the sealing gasket to avoid air leaks     Thermo Fisher Scientific    Safety Precautions    Troubleshooting    Thermo Fisher Scientific    Guides    Chapter 6  Troubleshooting    This instrument has been designed to achieve a high level of reliability  In  the event of problems or failure  the troubleshooting guidelines  board level  connection diagrams  connector pin descriptions  and testing procedures  presented in this chapter should be helpful in isolating and identifying  problems     The Technical Support Department at Thermo Fisher Scientific can also  be consulted in the even
583. tter PMz 5    Report No   936 21209885 F    Table 37  Summary of the results of the equivalence test  SN 1  amp  SN 2  measured compo   nent PMz5  raw data    29 1  2 17 ug m3 Orthogonal Regression Betw een Instrument Uncertainties    Slope  b      Up Intercept  a      ua Reference Candidate    All Data 1 068     0 013   0 103     0 225    PM2 5 5030i Sharp     lt  18 pg m3 1 066    0 032   0 040     0 317   gt 18 ug m3 1 090    0 032   0 900     0 975    Orthogonal Regression Limit Value of 30 ug m3    Slope  b      Up Intercept  a      ua Wem       217 ug m3    Bornheim Winter 1 075     0 027    0 205     0 667 56 1    Dataset    Cologne Winter 1 086     0 030   0 534     0 676 53 7  Individual Datasets    Bornheim Summer 1 110     0 033   0 530     0 466 15 4    Teddington Summer 1 117     0 050   0 656     0 407 4     1    lt  18 ug m3 157 0 847 1 120     0 035   0 611     0 342 23 38 4 5  Combined Datasets 218 ug m3 52 0 952 1 111     0 034   1 326     1 050 22 93 100 0  All Data 209 0 967 1 087     0 014   0 408     0 240 20 28 28 2    Orthogonal Regression Limit Value of 30 ug m3  Dataset  Slope  b      Up Intercept  a      ua Wem       2 17 ug m3    C S  Bornheim Winter 41 1 104     0 031   0 840     0 778 56 1    Cologne Winter 1 058     0 027   0 394     0 592 53 5  Individual Datasets  Bornheim Summer 0 947     0 030  1 099     0 427 15 7    Teddington Summer 63 1 016     0 051  0 207     0 433 3    2    lt  18 ug m3 166 0 817 1 057     0 035  0 123     0 344 17 55
584. tware version   V02 00 00 232     Restrictions   None    Notes    1  The requirements according to guide    Demonstration of Equivalence of Ambient Air Moni    toring Methods    are fulfilled for the measured component PMo 5    The measuring system shall be operated within a lockable measuring cabinet    3  The measuring system is to be calibrated on site at regular intervals by application of the  gravimetric PM25 reference method according to EN 14907    4  It is recommended that the measuring system is operated with a threshold for the relative  humidity of 58   in particular at sites with significantly high ratio of volatiles in the sus   pended particulate matter    5  The test report on the suitability test is available online under www qgal1 de     N    Test report   T  V Rheinland Energie und Umwelt GmbH  Cologne  Germany  Report No   936 21209885 F of 20 September 2013    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland      Air Pollution Control 3    Precisely Right     Report on the suitability test of the ambient air quality measuring system   Model 5030  SHARP with PM25 pre separator of the company Thermo Fisher Page 19 of 424  Scientific for the component suspended particulate matter PMz 5    Report No   936 21209885 F    1 3 Summary of test results    filled    Requirements on the instrument design    4 1 General requirements    4 1 1 Measured value   Shall be available  The measuring system is fitted with a  display mea
585. ty measuring system   Model 5030  SHARP with PMz2 5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5  Page 207 of 424  Report No   936 21209885 F                      Annex 8 Ambient conditions at the additional test site Bornheim  Summer 2013 Page 2 of 2  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5   Mean ambient temperature   Max  ambient temperature  Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   pg m   pg m         C     C   hPa       m s  g  mm   31 7 7 2013 Bornheim 11 0 1 1 10 0 23 1 29 7 1020 58 8 1 2 218 0 0  32 7 8 2013 11 0 1 4 12 9 23 0 29 8 1019 59 6 1 4 214 0 0  33 7 9 2013 14 4 27 19 2 23 4 29 9 1014 59 4 1 4 237 0 0  34 7 10 2013 10 0 1 8 18 1 19 5 24 2 1012 62 6 3 5 261 0 0  35 7 11 2013 14 0 1 4 10 0 15 7 19 7 1013 70 1 1 7 215 0 0  36 7 12 2013 16 7 3 4 20 4 16 5 21 9 1013 70 8 1 2 250 0 0  37 7 13 2013 15 3 28 18 5 17 7 22 9 1014 68 3 1 1 241 0 0  38 7 14 2013 14 5 3 7 25 5 18 9 24 2 1014 69 1 17 249 0 0  39 7 15 2013 17 4 2 0 11 3 21 3 27 8 1013 62 9 0 8 188 0 0  40 7 16 2013 19 4 27 14 2 22 5 29 2 1013 58 8 0 8 184 0 0  41 7 17 2013 13 4 2 0 14 8 23 2 29 3 1014 59 0 1 2 218 0 0  42 7 18 2013 2 0 24 5 31 6 1014 56 8 1 7 224 0 0  43 7 19 2013 8 2 17 21 3 23 5 29 1 1013 58 3 2 3 241 0 0  44 7 20 2013 9 8 1 7 17 5 21 1 26 6 1011 68 5 1 3 226 0 0  45 7 21 2013 12 6 1 8 14 6 25 3 32 3 1009 57 4 1 1 155 0 0  46 7 22 2013 15 2 2 2 14 4 27 6 35
586. uantity is not a  multiple of eight  the remaining bits in the final data byte will be padded  with zeros  toward the high order end of the byte   The Byte Count field  specifies the quantity of complete bytes of data     Note The values reported may not reflect the state of the actual relays in  the instrument  as the user may program these outputs for either active  closed or open  A    Request   Function Code 1 Byte 0x01 or 0x02   Starting Address 2 Bytes 0x0000 to maximum allowed by instrument  Quantity of Outputs 2 Bytes 1 to maximum allowed by instrument   Unit Identifier 1 Byte 0x00 to OxFF  Passed back in response   Response   Function Code 1 Byte 0x01 or 0x02   Byte Count 1 Byte N    Output Status N Byte N NorN 1     N   Quantity of Outputs   8  if the remainder not equal to zero  then N N 1    Error Response  Function Code 1 Byte 0x01 or 0x02  Exception Code 1 Byte 01 IIlegal Function  02 lllegal Address     03 Illegal Data  04 Slave Device Failure    Thermo Fisher Scientific     0x03 0x04  Read Holding  Registers Read Input    Thermo Fisher Scientific    Registers    MODBUS Protocol  Function Codes    Here is an example of a request and response to read outputs 2 15     Request   Field Name  Hex   Function 0x01   Starting Address Hi 0x00  Starting Address Lo 0x02  Quantity of Outputs Hi 0x00  Quantity of Outputs Lo 0x0D  Response   Field Name  Hex   Function 0x01   Byte Count 0x03  Output Status 2   10 OxCD  Output Status 11 15 Ox0A    The status of outputs 2   10 
587. uation chamber measures alpha emissions from  the accumulated aerosol and excludes negative mass artifacts from the  presence of daughter nuclides from radon gas decay to achieve a    refined  mass    measurement  Simultaneous refined mass measurement and sample  volume measurement through a calibrated orifice provide a continuous  concentration measurement of the ambient mass concentration     Thermo Fisher Scientific       Ta             Introduction  Principle of Operation    RHa INLET                                                                                                                                                                                                             Pa HEATER  HEPA  FILTER  4  NEPHELOMETER  T  3 WAY  NC VALVE  Y o DETECTOR  E      B SOURCE  GF  o VAC  ORIFICE iI  PUMP R  EXHUAST 1 G         PROPORTIONAL  ice    4  VACUUM  PUMP  i  MAKE UP AIR    Figure 1 2  Nephelometer Sample and Zero Flow Schematic    Thermo Fisher Scientific    Model 50307 SHARP Instruction Manual 1 5    Introduction  Principle of Operation    1 6 Model 50307 SHARP Instruction Manual    A clean filter spot is introduced by an automatic filter change to the  combined sampling and detection chamber at least every 24 hours  usually  centered at midnight  However  the official filter tape change frequency for  TUV and U S  EPA testing is for both PM25and PM   is every 8 hours   The automatic filter changes are controlled by the instrument firmware  which halts the pump o
588. uction Manual 8 1    System Description  Hardware    Primary Measurement Head Assy      Tape Radius Guide    Detector Amp                      Bottom Filter Tape Supply Reel  Spindle Tape Wheel Assy    SHARP Connection Bracket Assy    Inlet  External  Solenoid  Tube Mounting Block Assy  h A   Counter Wheel Photo  Cam Final Assy o  m E   Interrupt Board Assy    Cam Photo Interrupt    SHARP Connection Board    Bracket Assy    Pressure Board Proportional Valve    dikei Mall nad  ney S w fitting Assy     ai iF     1 0 Expansion Board  Opt     Front Panel              gt  Digital Output Board    Front Panel Board    Motor w  Photo  Power Supply eel Interface Interrupt Coupler Assy Meinerbgara  Optional  Transformer    Figure 8 1  Model 50307 Hardware Components    8 2 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    System Description    Reference Detector Board         SHARP Board Assy Source Board    SHARP Board Bracket Assy    Scatter Detector Board    Optical Bench RH Board    Optical Bench Inlet    Figure 8   2  SHARP Optics Component    Hardware    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 8 3    System Description  Hardware    Nephelometer  Assembly    Primary  Measurement Head    Detector Amplifier    Cam Photo Interrupt    Board    Counter Wheel Interrupt  Board Assembly    Proportional Valve    Cam Motor    Tape Motor    8 4 Model 50307 SHARP Instruction Manual    The nephelometer assembly uses an 880 nanometer incident light source  a
589. ue between  100 pg m  and 10000 ug m  can be specified as a range     Thermo Fisher Scientific    Operation  Integration Time    e Inthe Main Menu  choose Range  gt  Set Custom Ranges        Custom Ranges The Custom Ranges screen is used to define the custom ranges     The display shows the current custom range  The next line of the display is    used to set the range  For more information about selecting ranges  see     SHARP Range    above     e Inthe Main Menu  choose Range  gt  Set Custom Ranges  gt  Custom  Range 1  2  or 3        Integration Time The Integration Time screen allows the user to adjust the compounded  averaging of the beta attenuation mass and concentration  The default  value should be set to 20 minutes and may be adjusted between 15   60  minutes  The display shows the current integration time is set to 15  The  next line of the display is used to change the integration time     e Inthe Main Menu  choose Integration Time        Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 13    Operation  24 Hour Average    24 Hour Ave rage The 24 Hour Average screen defines a start time during which  concentration measurement is averaged over a daily 24 hours period  The  average concentration of the primary readings are calculated for that time  period and averaged out  A new 24 hour average concentration is displayed  once per day  The display shows the current 24 hour average SHARP  concentration  The next line of the display is used to change the 
590. ue to 95      Send  set alarm neph rh min 95  Receive  set alarm neph rh min 95 ok    Model 50307 SHARP Instruction Manual B 21    C Link Protocol Commands  Alarms    B 22 Model 50307 SHARP Instruction Manual    alarm neph temp min   alarm neph temp max   This command reports the current nephelometer temperature alarm  maximum value setting  The following example reports that the  nephelometer temperature alarm maximum value is 5 0   C     Send  alarm neph temp max  Receive  alarm neph temp max 5 0 degC    set alarm neph temp min value   set alarm neph temp max value   This command sets the nephelometer temperature alarm maximum value  to value  where value is a floating point number representing nephelometer  temperature alarm limits in   C  The following example sets the  nephelometer temperature alarm maximum value to 40 0   C     Send  set alarm neph temp max 40  Receive  set alarm neph temp max 40 ok    alarm sample rh min   alarm sample rh max   These commands report the current sample relative humidity alarm  minimum and maximum value settings  The following example reports  that the sample relative humidity alarm minimum value is 5      Send  alarm sample rh min  Receive  alarm sample rh min 5 0      set alarm sample rh min value   set alarm sample rh max value   These commands set the sample relative humidity alarm minimum and  maximum values to value  where value is a floating point number  representing sample relative humidity alarm limits in percent  The  followin
591. um  values to value  where value is a floating point number representing  vacuum pressure alarm limits in millimeters of mercury  The following  example sets the vacuum pressure alarm maximum value to 250 0 mmHg     Send  set alarm vac pres max 250  Receive  set alarm vac pres max 250 ok  amb rh    This command reports the ambient RH  in percent  The following example  reports that the ambient RH is 30      Send  amb rh  Receive  amb rh 30 0    amb temp    This command reports the current ambient temperature  in degrees C  The  following example reports that the ambient temperature is 15   C     Send  amb temp  Receive  amb temp 15 0 degC    Model 50307 SHARP Instruction Manual B 23    C Link Protocol Commands  Diagnostics    B 24 Model 50307 SHARP Instruction Manual    baro pres  This command reports the current barometric pressure  in mmHg  The  following example reports that the barometric pressure is 722 mmHg     Send  baro pres  Receive  baro pres 722 0 mmHg  board temp    This command reports the current board temperature  in degrees C  The  following example reports that the board temperature is 15   C     Send  board temp  Receive  board temp 15 0 degC  crn    This command reports the current radon concentration for the detector   The following example reports that the radon concentration is 1 2 Bq m        Send  crn  Receive  crn 1 20 Bg m3  det status    This command reports the current mass for the detector  The sequence is   a  B  Bc  Brer  Each value is separated
592. umber of Stop bits  lor2   Parity   None  Odd  or Even   Data rate   1200 to 115200 Baud  9600 is default     Model 50307 SHARP Instruction Manual D 1    Gesytec  Bayern Hessen  Protocol  TCP Communication Parameters    TCP Communication  Parameters    Instrument Address    Abbreviations Used    Basic Command  Structure    D 2 Model 50307 SHARP Instruction Manual    iSeries instruments support the Gesytec protocol over TCP IP  The register  definition is the same as for the serial interface  Up to three simultaneous  connections are supported over Ethernet     TCP connection port for Gesytec  9882    The Gesytec instrument address has a value between 0 and 127 and is  represented by a 3 digit ASCII number with leading zeros or leading spaces  if required  e g  instrument address of 1 is represented as 001 or   lt SP gt  lt SP gt 1   The default instrument address is 14     The instrument address is the same as the Instrument ID used for C Link  and MODBUS commands  This can be set via the front panel     The instrument address is represented by  lt address gt  in the examples  throughout this document     Note Device IDs 128 through 247 are not supported because of  limitations imposed by the C Link protocol  A    The following is a list of abbreviations used in this document    lt CR gt  is abbreviation for Carriage Return  ASCII code 0x0D    lt STX gt  is abbreviation for Start of Text  ASCII code 0x02    lt ETX gt  is abbreviation for End of Text  ASCII code 0x03    lt SP 
593. umetric The flow rate of the Model 5030  should be verified during this acceptance  Flow Rate Verification test  Prior to this test it is important for the previous temperature and  pressure acceptance tests to be completed  Should the ambient temperature   flow temperature or barametric sensors require calibration  this should be  done prior to the flow rate verification     e Press        to return to the Diagnostics menu  and choose Flows        EMUIRETED CHAT pyg CH coir  i i i 5  L    EPOPEE Emet Booe EOT Fe fat          The Flow screen  read only  displays the volumetric flow rate in liters per  minute  The flow is measured by the internal pressure board sensors     2 8 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Leak Check  Procedure    Nephelometer Source    Verification    Nephelometer Zero    Thermo Fisher Scientific    Check    Installation  Acceptance Testing and Startup Procedures    Attach the small sample tube adapter to the Model 50302 inlet  Depending  upon your NIST traceable volumetric flow transfer standard  FTS   you  may either attach your FTS now or add the flow audit adapter to  accommodate any flexible tubing that your FTS may require  Allow 60  seconds to stabilize the flow after connecting     Record the Model 50302 Volumetric Flow Rate as Qm and the FTS flow  rate as Qi  assuring that both are being recorded in units of actual liters per  minute  Take three readings each from Qm and Qi and average the  respective values  Use th
594. ured values as 5 minute averages or hour   ly averages  configuration used during suitability testing  in the internal storage  These val   ues can  for example  be read out on the iPort software and processed as required with a  spread sheet analysis  Furthermore  it is also possible to parameterise the AMS in a way that  daily averages are calculated internally and output as such    The filter tape transport itself takes approx  1 min  After changing the filter tape  both the ra   diometric and nephelometric measurements are zeroed  The zeroing of the nephelometric  measurement is carried out within 10 min after the pump has stopped  the radiometric meas   urement zeroes within 15 minutes after pump stop  It should be noted that the measuring  system does not output new SHARP values during the first 10 min after filter tape change   After completing the zeroing of the nephelometer  the SHARP signal can be determined  again  since it is calculated from the nephelometer concentration and the auto calibration fac   tor  Nevertheless  the auto calibration factor is frozen until the end of the zeroing phase of  the radiometric measurement and is only updated upon its completion    The time required for filter tape transport is 1 min  It corresponds to approx  0 21   lag time  for 24 h measurements with three tape changes a day    The available time of the SHARP signal at cycles of 8 h is 470 min   480 min     10 min   cov   ering 97 9   of the total time     6 5 Assessment    It
595. urrent flow span to the value  where value is a  floating point representation of the flow span specified at the time of  calibration  The following example sets the flow span to 1 5     Send  set flow span 1 5  Receive  set flow span 1 5 ok    flow temp offset  This command reports offset for flow temperature  in degrees C  The  following example reports that the flow temperature offset is 5   C     Send  flow temp offset  Receive  flow temp offset 5 0 degC    Model 50307 SHARP Instruction Manual B 39    C Link Protocol Commands  Calibration    B 40 Model 50307 SHARP Instruction Manual    set flow temp offset value   This command sets the offset value for flow temperature  where value is a  floating point number representing flow temperature offset in degrees C   The following example sets the flow temperature offset to 7   C     Send  set flow temp offset 7  Receive  set flow temp offset 7 ok  high volt    This command reports the current high voltage for the detector  The  following example reports that the high voltage is 1400 V     Send  high volt  Receive  high volt 1400    set high volt value   This command sets the high voltage for the detector to value  where value is  a floating point representation of the high voltage  The following example  sets the high voltage to 1445 V     Send  set high volt 1445  Receive  set high volt 1445 ok  mass coef    This command reports the current mass coefficient  The following example  reports that the mass coefficient is 7100     
596. us Reports the interface board status B 20  alarm io status Reports the I O expansion board status B 20  alarm mb status Reports the motherboard status B 20  alarm nb status Reports the status of the nephelometer interface board B 20  alarm ncal status Reports the status of the last nephelometer auto zero B 21  calibration  alarm neph led cur Reports sets current nephelometer LED alarm maximum B 21  max value  alarm neph led cur Reports sets current nephelometer LED alarm minimum value B 21  min  alarm neph rh max Reports sets nephelometer relative humidity alarm maximum B 21  value  alarm neph rh min Reports sets nephelometer relative humidity alarm minimum B 21  value  alarm neph temp Reports sets nephelometer temperature alarm maximum B 22  max value  alarm neph temp Reports sets nephelometer temperature alarm minimum B 22  min value  alarm sample rh Reports sets sample relative humidity alarm maximum value B 22  max  alarm sample rh min Reports sets sample relative humidity alarm minimum value B 22  alarm status det Reports the detector status B 22  alarm tape counter   Reports sets tape counter alarm maximum value B 23  max  alarm vac pres max Reports sets vacuum pressure alarm maximum value B 23  alarm vac pres min     Reports sets vacuum pressure alarm minimum value B 23  allow mode cmd Reports sets the current    set    allow mode command B 55       Model 50307 SHARP Instruction Manual B 5    C Link Protocol Commands  Commands    B 6 Model 50307 SHARP Instruction Manu
597. us modes     e Inthe Main Menu  choose Instrument Controls  gt  Volumetric  Conditions        an  BEY Saad fat    ad          Temperature The Temperature screen is used to turn the temperature correction to  Compensation actual or standard     e Press to toggle and set the temperature compensation to actual  or standard        Standard Temperature The Standard Temperature screen is used to set the standard temperature     e Inthe Main Menu  choose Instrument Controls  gt  Volumetric  Conditions  gt  Temperature  gt  Standard        3 24 Model 50307 SHARP Instruction Manual Thermo Fisher Scientific    Pressure Compensation    Standard Pressure    Datalogging  Settings    Thermo Fisher Scientific    Operation  Volumetric Conditions    The Pressure Compensation screen is used to turn the pressure correction  to actual or standard     e Press to toggle and set the pressure compensation to actual or  standard        The Standard Pressure screen is used to set the standard pressure     e Inthe Main Menu  choose Instrument Controls  gt  Volumetric  Conditions  gt  Pressure  gt  Standard        Pare irog          The   Series instruments include a built in datalogging capability as a  standard feature  The operator is allowed to create two different types of  records  which for historical reasons  are named lrecs and srecs  Each record  can contain up to 32 different fields  or data items  and records can be  created at user defined intervals ranging from 1 to 60 minutes     Record g
598. ut channel  Possible ranges for the voltage outputs are  0 100  mV  0 1  0 5  and 0 10 V  Possible ranges for the current outputs are  0 20  mA and 4 20 mA     e Inthe Main Menu  choose Instrument Controls  gt  I O Configuration   gt  Analog Output Config  gt  select Channel  gt  Select Range     ETRE EE Saeed baose Pra Aad       Minimum and Maximum The Minimum and Maximum Value screens are used to edit the zero  0   Value and full scale  100  values  respectively  in percentages for the selected    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 49    Operation  Volumetric Conditions    analog output channel  See Table 3   4 for a list of choices  The minimum  and maximum output value screens function the same way  The following  example shows the set minimum value screen     e Inthe Main Menu  choose Instrument Controls  gt  IO Configuration    3 50 Model 50307 SHARP Instruction Manual     gt  Analog Output Config  gt  select Channel  gt  Set Minimum or  Maximum Value        Table 3   4  Analog Output Zero to Full Scale                                                                               Output Zero  0   Value Full Scale 100  Value  PM Zero  0  Range Setting   AVG PM Zero  0  Range Setting   Neph Zero  0  Range Setting   SHARP Zero  0  Range Setting   AVG SHARP Zero  0  Range Setting   AVG Neph Zero  0  Range Setting   Barometer Pressure User set alarm min value User set alarm max value  Vacuum User set alarm min value User set alarm max value  Flow
599. ve    set save params   These commands store all current parameters in FLASH memory  It is  important that each time instrument parameters are changed  that this  command be sent  If changes are not saved  they will be lost in the event of  a power failure  The following example saves the parameters to FLASH    memory    Send  set save params  Receive  set save params ok  time    This command reports the current time  24 hour format   The following  example reports that the internal time is 2 15 30 pm     Send  time  Receive  time 14 15 30    set time hh mm ss  hh   hours   mm   minutes   ss   seconds    Model 50307 SHARP Instruction Manual B 53    C Link Protocol Commands  Communications Configuration    Communications  Configuration    B 54 Model 50307 SHARP Instruction Manual    This command sets the internal clock  24 hour format   The following  example sets the internal time to 2 15 pm     Note If seconds are omitted  the seconds default to 00  A    Send  set time 14 15   Receive  set time 14 15 ok   addr dns   This command reports the TCP IP address for the domain name server   Send  addr dns   Receive  addr dns 192 168 1 1   set addr dns address    This command sets the domain name server address  where address consists      lt a gt      of four numbers ranging from 0 255 inclusive  separated by             Send  set addr dns 192 168 1 1   Receive  set addr dns 192 168 1 1 ok   addr gw   This command reports the default TCP IP gateway address   Send  addr gw   Receive
600. ve humidity to control the sam   ple heating from originally 50   to 58    necessary due to the results of the campaign in  Teddington UK   an additional campaign at one of the German sites with two candidate sys   tems with settings of each 50   and 58   was performed  decision of the 32   Expert meet   ing  test reports    of May 14 15 2013   Table 2 gives an overview on the additional campaign   The results of this campaign are shown in chapter 4 3 Field test from page 55     740300_2013_936_21209885F_EN docx       TUV Rheinland Energie und Umwelt GmbH A TUV Rheinland     Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher Page 17 of 424  Scientific for the component suspended particulate matter PM25   Report No   936 21209885 F                   Table 2  Description of the test site  additional campaign 2013   Bornheim   motorway parking lot   summer  Period 06 2013     07 2013  No  of paired values  Candidates  45  Characteristics Rural structure   motorway  Rank of pollution Low to average             The complete test was performed within the scope of the test programme    Combined  MCERTS and TUV PM Equivalence Testing Programme     In the context of European har   monisation  the test programme was developed and carried out by British and German test  institutes  Bureau Veritas UK  amp  Ireland  National Physical Laboratory NPL 
601. ve the accuracy of the candidate systems     Three different cases may occur   a  Slope b not significantly different from 1   b     1  lt  2u b       intercept a significantly different from 0  la  gt  2u a     b  Slope b significantly different from 1   b     1   gt  2u b    intercept a not significantly different from 0   a   lt  2u a     c  Slope b significantly different from 1   b  1   gt  2u b        intercept a significantly different from 0  la  gt  2u a     With respect to a     The value of the intercept a may be used as a correction term to correct all input values y   according to the following equation     Y corr   yj  a    740300_2013_936_21209885F_EN docx    aa 7    TUV Rheinland Energie und Umwelt GmbH   TUVRheinland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 166 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    The resulting values of yicor  may then be used to calculate the following new terms by linear       regression   Vicor  C   OX   and  Us_s  Vicor    _    u  x     c    d   1 x  P  u  a   Ea  n 2   with u a    uncertainty of the original intercept a  the value of which has been used to obtain  Yi corr     Algorithms for the calculation of intercepts as well as slopes and their variances by orthogo   nal regression are described in detail in annex
602. ver    ccccccccceceteceteeeteeee 7 44  Figure 7 25  Replacing the External Solenoid Valve          ccccccccceceteeeteeee 7 45  Figure 7 26  Disconnecting the RH Temp Connector    7 47  Figure 7 27  SHARP Interface Board Connections            c cccccccecceceeeteeeteeee 7 48  Figure 7 28  Replacing the SHARP Optics Board    cece 7 49  Figure 7   29  Replacing the Delrin Nut      s    7 50  Figure 7 30  Replacing the Internal SHARP Cable Assembly             0 0  000  7 52  Figure 7 31  Replacing the Case Cable  o      c cccccccccccscescecescetesteteeteteseeeeee 7 53  Figure 7   32  Replacing the External Zeroing Filter oo    ccc ceceeeceteeseeee 7 54  Figure 8   1  Model 50307 Hardware Component              ccccsccscesesceesceteeeeeees 8 2  Figure 8   2  SHARP Optics Component         cece cece ccccseesesteseeeeseeteseeteseeseseseesenes 8 3  Figure 9   1  Bench WMOUnting 252s aie lat a ae as os aa te 9 4  Figure 9 2  EIA Rack Mounting oo    cece cc cecesescecesesescetesesesteesesesteeseseateneness 9 5  Figure 9 3  Retrofit Rack MOUNTING    ccc cece escetesesescetesesesteteseseeteenens 9 6  Figure 9   4  Rack Mount Option Assembly      0    c c  cece cece ceesceesteseeteseeteseeeeees 9 7  Figure 9   5  Roof Flange Assembly ou    ccc cccesceceesesceteseseseeteseseseetesesesteenens 9 8  Figure 9 6  Securing the Roof Flange    9 9  Figure 10 1  SHARP Optics Connection Bracket         0 0 ccccccccecseceteeeteeee 10 2  Figure 10 2  Tape Pick up Spool Removal          c ccceccccecce
603. vg    This command reports the last calculated 24 hour average nephelometer  concentration  which defaults to 9 999 pg m  if 24 hours has not passed      Thermo Fisher Scientific    Thermo Fisher Scientific    C Link Protocol Commands  Measurements    Send  neph 24avg  Receive  neph 24avg 9 999E 03 pg m3       neph led current  This command reports the nephelometer LED current  The following  example shows that the LED current is 65 7 mA     Send  neph led current  Receive  neph led current 65 7 mA  neph raw    This command reports the raw  unaveraged  nephelometer concentration   The following example reports the raw  unaveraged concentration value     Send  neph raw  Receive  neph raw 1 331E 01 pg m3  neph ref det    This command reports the nephelometer reference detector voltage  The  following example reports that the reference detector voltage is 1096 mV     Send  neph ref det  Receive  neph ref det 1096 mV  neph rh    This command reports the nephelometer relative humidity  The following  example reports that the relative humidity is 27 2      Send  neph rh  Receive  neph rh 27 2    neph temp    This command reports the nephelometer temperature sensor  The  following example reports that the temperature sensor is 27 2   C     Send  neph temp  Receive  neph temp 27 2 degC  pm    This command reports the measured PM concentration  The following  example reports that the PM concentration is 14 1 mg m        Send  pm  Receive  pm 1 410E 01 mg m3  pm 24avg    This command r
604. vs  candidate  SN 1  measured component PM  5s    Teddington  summer  ovseccccctvesecenes cases liteadeeagsaagccneredeass ives  aagessemnsermaueelseeerees 163  Figure 58  Reference vs  candidate  SN 2  measured component PM2s5    Teddington  SUMIMEN  oreceimicedecrscunciadencrmcetecerenteecnsredenenersasnautersinctmatenenpen 163  Figure 59  Reference vs  candidate  SN 1  measured component PM  5    Valles 2 73 GMM  ors a R E a E E 164  Figure 60  Reference vs  candidate  SN 2  measured component PM2s5    yalues  2 18 WU MN a esa iieri ter sai dies nua ka as Sich nnd ada rara N ans Ea i aa aaae 164  Figure 61  Stability of standard weight             sssssseeeeesssessnrrnrrrsserrtrrnrtrnsrrrrnttrnrrnnnnnseernnee 210  Figure 62  Stability of the Control filterS           esssnnnneeeeseeeennrrnnnnsseerrnrrrnnrssertrrnrrnrrnnnneseernnee 212    Figure 63  Distribution for Emfab filters of  A  Pre spread weighed  relative to the check weight and  B  Post spread weighed  relative to the check weight  ci susocotesasesanventvletwtanersansaeueanecanpeacsnusvcaniesonumerenees 219    740300_2013_936_21209885F_EN docx    TUVRheinland    Precisely Right     Page 12 of 424    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH  Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5     Re
605. ween uncertainty of the reference devices Ure is determined analogous to the un   certainty between candidates and shall be  lt  2 ug m      The evaluated results are illustrated in    Detailed presentation of test results    of this test item     A linear correlation y    a   bx  is assumed between the results of both methods in order to  evaluate the comparability of the candidates y and the reference procedure x  The correlation  between the average values of the reference devices and the candidates is established by  orthogonal regression     Regression is calculated for   e All test sites and comparisons together  e Each test site and comparison separately    e 1 dataset with measured values PM25 2 18 g m   Basis  average value of reference  measurement     740300_2013_936_21209885F_EN docx    a 3 D T  V Rheinland Energie und Umwelt GmbH  A TUVRheinland    0 Air Pollut    Air Pollution Control  Precisely Right     Report on the suitability test of the ambient air quality measuring system   Page 152 of 424 Model 5030  SHARP with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    For further evaluation  the results of the uncertainty us s of the candidates compared with the  reference method are described with the following equation  which describes Ucr as a func   tion of the PM concentration xi     uay   p UO    fa   01  F    With RSS   Sum of the  relative  residuals from ort
606. wer and grounding to the detector     The digital output board connects to the motherboard and provides  solenoid driver outputs and relay contact outputs to a connector located on  the rear panel of the instrument  Ten relay contacts  normally open  with  power off   are provided  which are electrically isolated from each other   Eight solenoid driver outputs  open collector  are provided along with a  corresponding  24 VDC supply pin on the connector     Thermo Fisher Scientific    1 0 Expansion Board   Optional     Front Panel  Connector Board    1 0 Components    Analog Voltage  Outputs    Thermo Fisher Scientific    System Description  1 0 Components    The I O expansion board connects to the motherboard and adds the  capability to input external analog voltage signals and to output analog  currents via a connector located on the rear panel of the instrument  It  contains local power supplies  a DC DC isolator supply  a sub processor  and analog circuits  Eight analog voltage inputs are provided with an input  voltage range of OV tol0 VDC  Six current outputs are provided with a  normal operating range of 0 to 20 mA     The front panel connector board interfaces between the motherboard and  the function key panel and graphics display  It serves as a central location to  tie the three connectors required for the function key panel  the graphics  display control lines  and the graphics display backlight to a single ribbon  cable extending back to the motherboard  This boar
607. which ensures that the  measured mass concentration remains unaffected by changes in the total particle population     740300_2013_936_21209885F_EN docx    T i VRheinl    TUV Rheinland Energie und Umwelt GmbH   U einland Air Pollution Control  Precisely Right    Report on the suitability test of the ambient air quality measuring system   Page 32 of 424 Model 5030  SHARP with PM2s5 pre separator of the company Thermo   Fisher Scientific for the component suspended particulate matter PM2 5    Report No   936 21209885 F    3 2 Functionality of the measuring system    The Model 5030i SHARP measuring system combines the principles of aerosol light scatter   ing  nephelometry  and beta attenuation  analogous to Model 5014i BETA  see test report  936 21209985 H      The particle sample passes through the PM  25 sampling inlet with a flow rate of 1 m   h    16 67 I min  and reaches via the heated sampling tube  DHS   dynamic heating system   the actual Model 5030i SHARP measuring instrument     The use of the dynamic heating system DHS ensures that the relative humidity of the air  passing through the filter tape of the radiometric stage stays well below the point at which the  collected particles accrete and condensation effects that may distort the measured value ap   pear  It also allows for maintaining the relative humidity below a specified threshold in the  nephelometer  set value during suitability testing  50   and 58   rel  humidity   This system  minimises the internal temp
608. with PM  5 pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F         JN    s          Figure 25  Field test site Teddington    In addition to the measuring systems used for determining the content of suspended particu   late matter in ambient air  a data acquisition system was installed at the cabi   net measurement site in order to determine the meteorological characteristics  Air tempera   ture  pressure and humidity as well as wind velocity  direction and precipitation were continu   ously assessed and their half hourly averages recorded     The installation of the cabinet itself and the arrangement of the sampling probes are de   scribed as follows     e Height of cabinet roof  2 50 m   e Height of the sampling point for testing 1 10 m   0 51 m above the cabinet roof  e Reference instrument 3 61   3 01 m above ground   e Height of weather vane  4 5 m above ground  Germany     2 5 m above ground  Uk     Table 6 contains  therefore  not only an overview of the most important meteorological fac   tors determined during measurement at the four test sites  but also an overview of the PM ra   tio during the entire test  The individual values are presented in annexes 5 and 6     740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH    Air Pollution Control    Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM25 pre separat
609. with parallel operated TEOM FDMS systems    740300_2013_936_21209885F_EN docx    ss      TUV Rheinland Energie und Umwelt GmbH  A TUV Rheinland    Luftreinhaltung    Precisely Right   Report on the suitability test of the ambient air quality measuring system  Page 200 of 424 Model 5030  SHARP with PM  pre separator of the company Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F                                                                            Annex 6 Ambient conditions at field test sites Page 8 of 11  No  Date Test site Mean PM2 5 conc  PM2 5 volatiles Ratio volatiles   PM2 5 Mean ambient temperature Max  ambient temperature Ambient pressure Rel  humidity Wind velocity Wind direction Precipitation   ug m3   ug m3          C     C   hPa       m s   4  mm   211 2 21 2012 Bornheim  Winter 10 2 1 3 13 1 4 4 8 5 1023 70 8 1 8 176 0 0  212 2 22 2012 13 9 6 3 11 2 1017 69 6 1 6 156 3 0  213 2 23 2012 8 4 2 4 29 0 8 2 10 3 1016 86 2 1 0 191 0 0  214 2 24 2012 11 9 3 6 30 5 9 9 13 7 1017 86 4 2 8 256 1 2  215 2 25 2012 6 3 7 0 10 9 1016 78 1 3 4 269 0 3  216 2 26 2012 26 5 6 4 24 2 5 2 6 9 1019 87 2 1 9 244 0 6  217 2 27 2012 19 2 5 9 30 9 7 7 10 3 1016 77 4 1 1 211 0 6  218 2 28 2012 17 4 5 1 29 3 9 4 10 7 1016 86 5 1 8 252 0 0  219 2 29 2012 27 0 7 0 26 0 10 4 13 3 1017 87 9 0 6 142 0 0  220 3 1 2012 18 8 4 6 24 7 8 8 10 6 1018 90 5 0 8 178 0 0  221 3 2 2012 14 6 4 6 31 6 9 0 11 8 1016 87 0 22 151 0 0  222 3 3 2012
610. wn  battery when instrument power is off     e Inthe Main Menu  choose Instrument Controls  gt  Date Time        Timezone The Timezone screen is used to set the timezone for the Network Time  Protocol  NTP  server  This should be set to the timezone that the  instrument is located in  If the exact timezone is not shown in the list  it  may be entered via the C LINK    tz    command  see Appendix B   The  selections are  UTC  GMT   EST  GMT 5   CST  GMT 6   MST   GMT 7   PST  GMT 8   YST  GMT 9   HST  GMT 10   NST   GMT 11   DLW  GMT 12   CET  GMT 1   EET  GMT 2   BST   GMT 3   DLT  GMT 4   ECH  GMT 5   FOX  GMT 6   GLF  GMT   7   CCT  GMT 8   JST  GMT 9   GST  GMT 10   LMA  GMT 11    DLE  GMT 12   EDT  GMT 5 4   CDT  GMT 6 5   MDT   GMT 7 6   and PDT  GMT 8 7      Note The current timezone may say NULL before the timezone is set for  the first time  or if the timezone was cleared with a C Link command  A    Thermo Fisher Scientific Model 50307 SHARP Instruction Manual 3 57    Operation  Diagnostics Menu    e Inthe Main Menu  choose Instrument Controls  gt  Timezone        Cu gu  PL a i    Diagnostics Menu The Diagnostics menu provides access to diagnostic information and  functions  This menu is useful when troubleshooting the instrument  The    analog input readings and analog input voltages are only displayed if the  I O expansion board option is installed     e Inthe Main Menu  choose Diagnostics        Program Versions The Program Versions screen  read only  shows the ver
611. y    B 44 Model 50307 SHARP Instruction Manual    set vac pres span value   This command sets the vacuum span to the value specified at the time of  calibration  where value is a floating point representation of vacuum span   The following example successfully sets the new value for vacuum pressure  span to 1 002     Send  set vac pres span 1 002  Receive  set vac pres span 1 002 ok    vac pres offset  This command reports the vacuum offset for the detector at the time of  calibration  The following example reports that the vacuum pressure offset    is 0 3     Send  vac pres zero  Receive  vac pres zero 0 3  set vf pres cal def    This command resets the vacuum and flow pressure calibration to their  default values     Send  set vf pres cal def  Receive  set vf pres cal def ok  push button   do me 1  down menu 2  en ri 3  enter right 4  he ru   help run   le up   left    button     do   down   en   enter   he   help   le   left   me   menu   ri   right    ru run up 1 2 3 4     These commands simulate pressing the front panel pushbuttons  The  numbers represent the front panel soft keys  from left to right     Send  push enter  Receive  push enter ok  isc   iscreen    These commands retrieve the framebuffer data used for the display on the  iSeries instrument  It is 19200 bytes in size  2 bits per pixel  4 pixels per  byte arranged as 320 by 240 characters  The data is sent in RLE encoded    Thermo Fisher Scientific    Measurement  Configuration    Thermo Fisher Scientific    
612. y Thermo  Fisher Scientific for the component suspended particulate matter PM2 5   Report No   936 21209885 F    Page 20 of 424    filled    General requirements on measuring systems    5 2 1 Certification    ranges    Measuring range    Negative output  signals    Failure in the  mains voltage    Operating states    Switch over    Maintenance in   terval    Have to comply with the re   quirements of Table 1 of the  Standard VDI 4202 Part 1     The upper limit of measure   ment of the measuring systems  shall be greater or equal to the  upper limit of the certification  range     Negative output signals or  measured values may not be  suppressed  life zero      Uncontrolled emission of oper   ation and calibration gas shall  be avoided  The instrument pa   rameters shall be secured by  buffering against loss caused  by failure in the mains voltage   When mains voltage returns   the instrument shall automati   cally reach the operation mode  and start the measurement ac   cording to the operating in   structions     The measuring system shall al   low the control of important op   erating states by telemetrically  transmitted status signals     Switch over between meas   urement and functional check  and or calibration shall be pos   sible telemetrically by computer  control or manual intervention     If possible 3 months  minimum  2 weeks     740300_2013_936_21209885F_EN docx    It is possible to assess the measuring  system in the range of the relevant  limit values     A me
613. y be configured for static addressing or dynamic addressing  set using a    DHCP server      Any serial port protocols may be accessed over Ethernet in addition to the    serial port  Up to three simultaneous connections are allowed per protocol     The external accessory connector is not used in the Model 50303 analyzer     This port is used in other models to communicate with smart external    devices that may be mounted hundreds of feet from the analyzer using an  RS 485 electrical interface     Thermo Fisher Scientific    Inlet Assemblies    Sampling Tube  Extensions    1 0 Expansion Board  Assembly    25 Pin Terminal  Board Assembly    Thermo Fisher Scientific    Chapter 9  Optional Equipment    The Model 5030  is available with the following options   e    Inlet Assemblies    on page 9 1  e    Sampling Tube Extensions    on page 9 1  e    I O Expansion Board Assembly    on page 9 1  e    25 Pin Terminal Board Assembly    on page 9 1  e    Terminal Block and Cable Kits    on page 9 2  e    Cables    on page 9 2  e    Mounting Options    on page 9 3  e    Roof Flange Assembly    on page 9 8    The Model 5030  can be used for different applications  Using different  inlet configurations  PMio  PM2 s  PM  Inlets   several particle size fractions  can be monitored  The main application is the measurement of PMio and  PM_s for ambient air quality monitoring and health effect studies     Often the most compatible way to install the Model 50302 at air quality  monitoring sites 
614. y being re calibrated by the radiometric  measurement during operation    The principle of radiometric mass determination is based on the physics law of beta attenua   tion as effect of passing through a thin layer of matter     This relation is described as follows             A f      C   In           My Ot J  J  where   Cs    particle mass concentration A   particle collection area  filter spot   Um   mass attenuation coefficient  dependent on the energy of the beta source   Q   sampling flow rate t   sampling time  fo   initial beta count rate  tare  fi   final beta count rate    The radiometric mass determination is calibrated on site with help of null and span foils  In  using the    mass transfer standard    foil set    span foil set   the radiometric measurement can  be checked anytime for quality assurance purposes and  if required  calibrated     The output mass concentration Cm of Model 5030i SHARP is calculated as follows     where    Cm   mass concentration SHARP   Cn  nephelometer signal averaged over 1 minute   C     radiometric mass concentration  dynamically filtered over a variable time span  Cn2  nephelometric concentration  dynamically filtered over a variable time span    740300_2013_936_21209885F_EN docx    TUV Rheinland Energie und Umwelt GmbH A TUVRheinland     Air Pollution Control   s  Precisely Right   Report on the suitability test of the ambient air quality measuring system  Model 5030  SHARP with PM2 5 pre separator of the company Thermo Fisher P
615. yern Hessen  Protocol  Gesytec Commands    The following 14 measurements reported in response to DA command for    the Model 5030  include   e SHARP  e PM  e Nephelometer  e Volumetric Flow  e Ambient Temperature  e Sample Temperature  e Ambient RH  e Sample RH  e Flow Pressure  e Vacuum Pressure  e Barometric Pressure  e Average SHARP  e Average PM  e Average Neph    Model 50307 SHARP Instruction Manual D 7    Gesytec  Bayern Hessen  Protocol  Gesytec Commands    Operating and Error  Status    D 8 Model 50307 SHARP Instruction Manual    See Table D 1 for operating status and Table D 2 for error status for the    Model 50302     Table D 1  Operating Status for Model 5030                                                                                                                             D7   D6   D5   D4 D3   D2   D1   DO   gt  Bit 8  7 6  5 4  3 2  1   gt  Hex value 80   40 20   10 08   04   02   01   MSB LSB   Operating status    Service Mode  On  0   0 0   0 0 10 0 l1  No Password Protect 0  0 0 10 0 10 1 10  Pump  Off  0  0 0 J0 0 J1 0  0  Heater  Off  0  0 0  0 1 10 0  0  Filter Change 0  0 0 J1 0  0 0  0  Nephelometer Zero 0  0 1 0 0  0 0  0  Nephelometer Standby 0 l1 0 10 0 10 0 10  Not Used 1 10 0 J0 0  0 0 J0   Table D 2  Error Status for Model 5030    D7   D6   D5   D4 D3   D2   D1   DO   gt  Bit 8  7 6  5 4  3 2  1   gt  Hex value 80   40 20   10 08   04   02   01   MSB LSB   Error status    Filter Tape Cntr Alarm 0 10 0 10 0 10 0 l1  Filter Change 0 J0 0  0 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
1150.D.01.0004_4 R3 Installation Manual French.cdr  ワンタッチャブル カタログダウンロード(PDF)    取り扱い説明書 CB400SF/CB1000SF(国内仕様)共に∼93  RESISTANCE/RESISTIVITY PROBE  Acronis True Image Home 2010    Copyright © All rights reserved. 
   Failed to retrieve file