1. Declaration of Conformity Name und Adresse des Herstellers Manufacture s name and address Nom et adresse du Fabricant ISM Deutschland GmbH Laubach 30 D 40822 Mettmann Germany Die ISM Deutschland GmbH bescheinigt die Konformitat fur das Produkt The ISM Deutschland GmbH declares conformity of the product ISM Deutschland GmbH derclare la conformite du produit Bezeichnung Product name SF6 Raumluftmonitor SF6 Roommonitor Typ Type Type SF6 AREACHECK P2 Mit den folgenden Bestimmungen with applicable regulations avec les directives suivantes EMV Richtlinie 89 336 EWG erganzt durch 91 263 EWG 92 31 EWG EMC Drective 89 336 EEC amended by 91 263 EWG 92 31 EEC Directive EMC 89 336 CEE amendee par 91 263 EWG 92 31 CEE Niederspannungsrichtlinie 73 23 EWG erganzt durch 93 68 EWG Low Voltage Equipment Directive 73 23 EEC amended by 93 68 EEC Directive des equipements basse tension 73 23 CEE amendee par 93 68 CEE Angewendete harmonisierte Normen Harmonized standards applied Normes harmonisees utilisaes Sicherheit Safety S curit EN 61010 1 1993 IEC CEI 1010 1 1990 A 1 1992 VDE 0411 1994 Uberspannungskategorie Overvoltage category Cat gorie de surtension II Verschmutzungsgrad Degree of pollution Degr de pollution 2 Elektromagnetische Vertraglichkeit Electromagnetic compatibility Compatibilit electromagn tique EN 50082 2 1995 VDE 0839 T82 2 ENV 50140 1993 IEC CEI 1
2. INNOVATION User Manual V2 48 SF6 AREACHECK P2 The SF6 Room Monitor SFG AREACHECK pa SMARTSENSOR P2 Parma 2100 0006 P2S00039 B Advanced Gas Sensing Technologies 9 z ISM User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 2 MESSTECHNIK f Table of Contents Declaration of Conformity ees 3 Installation sausnunnunnnnnnnnnnnnnnnnnnnunnnnnnnnnnnnnnnnnnnnnnnnnnnn A LOCO eE ese eee ee ee eee re a eee eee 4 Nae AA E cine ele ebm AA AA ee ee HO 5 Network diagram sceccecsecrcnceearcecnearcesnecececeearsesnenrseseeneseseensseteentanseens 6 Oppa NO aaa aaa Auto Zero ALINA AA 7 Measuring Interval cccscccsecseccsecsesrseceearceseenrcessenrseseenscusnenseuseensansanns 7 alto NONG ee 8 Lamp Test and Manual Reset 23 nennen 8 Configuration Via the PC interface 2222222222222224 D Connecting to the Host Computer eene 9 SER Paka AA Aa 9 Changing the Sensor Unit 10 02 SENSO aaa aaa aaa ll Functionality aaa 11 PhysicabdOCOHOl ee AA AA 11 DELO IPE ee AA AA AA 11 Sensor ird E ANA AA AA m UD 11 User INterfaCE AA AA AAP DEO 11 S AN 11 Technical Data erre aai 12 NOLBS ea een m Advanced Gas Sensing Technologies m Oy 23 User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 3 MESSTECHNIK INNOVATION
3. Scan period 2 Alarm level 3 Exit and resume operation Just type the corresponding number to invoke the desired option Any other inputs will be discarded After selecting 1 Scan period enter the desired time interval in minutes followed by a carriage return The allowed range is 1 99 minutes Upon selecting 2 Alarm level enter the number corresponding to the desired alarm level Option 3 will finish the setup and return the instrument to normal operation Since it will now perform a restart B Advanced Gas Sensing Technologies UE SU BUT 2000 A SE6 BUREAU VERITAS Md Sn IET m LI ve Art u d ION 3S E N SCIENC MESSTECHNIK User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 10 Changing the Sensor Unit For changing the sensor unit carry out the following Steps e Power down the P2 remove the fuse observing the hints below e Remove screws e Take off SmartSensor as indicated below e Pull the connector off the SmartSensor e Fit the replacement SmartSensor e Power up the P2 insert fuse observing the hints below Note To precisely determine SF6 concentration SF6 AREACHECK P2 requires a reference value representing clean air This reference value is taken and stored whenever the instrument is powered up Or reset You should therefore make sure that power up takes place under clean air conditions ATTENTION The fuse T1A is at mains voltage All works on it are to b
4. AREACHECK P2 requires a reference value representing clean air This reference value is taken and stored whenever the instrument is powered up or reset You should therefore make sure that power up takes place under clean air conditions A fresh reference value is also taken after manual reset via the key switch Therefore after each manual reset the instrument will temporarily only respond to SF6 concentrations exceeding the level that was present at the time of manual reset This will avoid the next measuring cycle to trigger an undesired new alarm The reference value will auto adapt to falling SF6 concentrations so after once meeting a clean air condition 5F6 AREACHECK P2 will automatically return to full sensitivity Since a reset is also performed when the instrument is taken into setup mode via the host computer interface the above also applies in this case After Power up or reset the first measuring cycle starts immediately beginning with the fan run time Measuring Interval Since SF6 AREACHECK P2 is designed to determine slowly changing concentrations it samples the air at a fixed time interval rather than measuring continuously This feature minimizes the amount of air per time that passes the instrument s intake system thus extending its service life Via the measuring time interval it is possible to perfectly adapt the instrument to the actual application For monitoring generally unattended spaces a long time interval sho
5. 001 4 3 1995 VDE 0847 T3 ENV 50141 1993 IEC CEI 1000 4 6 VDE 0843 6 ENV 61000 4 2 1995 IEC CEI 1000 4 2 1995 VDE 0847 T4 2 Prifscharfe Level Niveau 2 EN 61000 4 4 1995 TEC CET 1000 4 4 1995 VDE 0847 T4 4 Prifscharfe Level Niveau 3 EN 50081 1 1992 EN 55011 1991 CISPR11 1991 VDE0875 T11 1992 Gruppe group groupe 1 Klasse Class Classe B Datum Date Date Unterschrift Signature Signatur 01 08 2005 Se lt tnm Hr Becker Technical Director B Advanced Gas Sensing Technologies User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 4 MESSTECHNIK Installation Location SF6 AREACHECK P2 is designed for indoor wall mounting Since SF6 tends to sink towards the ground in undisturbed air we suggest to place it approx 50 60 cm above ground level EE LED Status Display Run Test Reset Key switch Audible Alarm Sample Sample Intake Exhaust SmartSensor Module LC LEBER UL HOHER ICC TCI GCE HE DEAE ETAGE TITI ABIRE PEI ATA ORAE TITA CPT DEDE UELAPE GSI HAO LAGA USE A LYYI DOE E ICE AA pa Ish Bn eee oss ee escthodeies User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 5 Wiring Mains power connects to the orange clamp block according to the designation printed on the PCB near each clamp For an optional audible and or visible alarm device a voltage free switchover relay output is supplied It is ava
6. e carried out by qualified personnel only observing appropriate precautions CES BB AS ERBEN ELA GCE DALA BALA ETTI PAA ABIRE PEI ATA ORAE TITA CPT JA CEDILAOAJGDAA HAO LAGA SIU EIS DOE E ICE AA pa Ish i GAS sensing 1ocinologies User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 11 MESSTECHNIK INNOVATION O2 Sensor Functionality The optionally available O2 sensor continuously measures the O2 content in the ambient air and reports it to the network controller if present An alarm condition on both the measuring station and network controller is raised whenever the oxygen level is outside the allowed range of 16 25 Physical location The O2 sensor is part of the SmartSensor it is found protruding downwards from it Sensor Type The O2 sensor is of galvanic cell type It exhibits low cross sensitivity to CO2 and other gases good Stability and linearity over its range Sensor Lifetime With a reasonable safety margin sensor life is fixed at one year no matter if shelfed or used this is due to the nature of galvanic cells in which the chemical reaction goes on at all times as long as oxygen is present When lifetime is over the SmartSensor must be replaced against a fresh readily calibrated one User Interface When a low oxygen condition is encountered the measuring station will issue an alarm involving the built in alarm buzzer the red alarm lamp and the alarm relay The low oxygen conditio
7. ilable on the blue 3 position clamp block in the middle Contacts are from left to right N O COM N C During normal operation COM connects to N O During an alarm or error condition COM disconnects from N O and switches over to N C The same applies in case of power failure and whenever the instrument is taken into setup mode via the host computer interface via the RS232 connector The leftmost 6 position clamp block is provided for network operation Physically the network follows RS485 standards and hence has three wires named A B and GND Their positions are marked on the PCB next to the clamps They must be wired to the corresponding pins on the SF6 NETWORKCONTROLLER P2NC Double clamps are provided for each wire for easy daisy chaining of several instruments LE Pr mE RS 485 daisy chain PIN B Supply Voltage N RS 485 daisy chain PIN A Supply Voltage L1 RS 485 daisy chain PIN GND Supply Voltage Earth RS 232 Connector Alarm Relay PIN Normally Closed Alarm Relay PIN Common Alarm Relay PIN Normally Open B Advanced Gas Sensing Technologies NG User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 6 MESSTECHNIK Network diagram wiring detail GND gt m Advanced Gas Sensing Technologies INNOVATION User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 7 MESSTECHNIK Operation Auto Zero at Startup To precisely determine SF6 concentration SF6
8. n will be reflected by the yellow 500 and 1000 ppm lamps flashing In case of a high oxygen condition the same applies except that the 1500 and 2000 ppm lamps will flash On the SF6 NETWORK CONTROLLER P2NC the O2 status will be marked as OK HI or LO respectively for each measuring station Safety Should the O2 sensor fail be missing or spent a low oxygen alarm will be issued B Advanced Gas Sensing Technologies INNOVATION User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 12 MESSTECHNIK Technical Data Detection principle SF NIC 9 O GC Galvanic Cell Range 0 2000 ppmv SF 0 30 O Resolution SFs 500 ppm O 1 me TU Noise level of audible alarm wee gagong ba Pwe T NNNM B Advanced Gas Sensing Technologies a Earifaries Na User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 13 MESSTECHNIK Notes B Advanced Gas Sensing Technologies BOAT 8 y ERFAN VERITAS Sy Girto aits T Li FE al fach
9. ost LEDs 1500 2000ppm will blink when an oxygen concentration above the upper alarm threshold of 25 is present Lamp Test and Manual Reset When the key switch is turned to the spring return TEST position all lamps the internal audible alarm and the optional external alarm device will turn on until the key switch is released When released a pending alarm condition is cleared NOTE An error condition will NOT be cleared When the key switch is turned and released twice within 2 seconds a system reset is performed which also clears any pending alarm and error conditions and logs a fresh reference value m Advanced Gas Sensing Technologies GU WE soso B fe BURFAD VERITAS M Sr Cinta aria m LI FE al fag User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 9 MESSTECHNIK Configuration via the PC interface Connecting to the Host Computer Parameter modification in stand alone operation is done exclusively through the host computer interface This has the clear advantage that any unintended or unauthorised fiddling is discouraged The user parameters described in the following are relevant for stand alone operation only In network operation parameters received from the network base unit will take precedence The host computer interface follows RS232 standards A special interface cable connects to the 3 position plug which is found between the two blue clamp blocks The lamp next to the plug indicates
10. the presence of valid RS232 levels from the host It will be lit when the host is on and correctly connected On the host computer a standard terminal program is used to talk to the SF6 AREACHECK P2 A suitable program called Hyper Termina is a standard Microsoft Windows component and by default present in most installations Optionally any other terminal software will work Set up the terminal program for the following parameters 9600 baud 8 data bits no parity 1 stop bit This is a standard configuration also known as 9600 8 N 1 Further select none for the handshake protocol and TTY as emulation if available The option append CR for received lines should be checked for a more readable screen output For your convenience we will be pleased to send you a predefined configuration file for Hyper Terminal which will perform all relevant settings for you Note While the PC interface is connected the instrument will talk to the host computer only and no longer communicate over the network When present an SF6 NETWORK CONTROLLER P2NC will see this station as OFFLINE and issue an alarm accordingly User Parameters User adjustable parameters are the time interval between each two samples Scan period and the alarm level To activate the setup from the terminal first send a line feed then the instruction TtS followed by another carriage return A choice menu with the following options will then be presented 1
11. uld be selected while in the working area a shorter interval is suggested SF6 AREACHECK P2 comes with a factory set measuring interval of 30 minutes which is the suggested starting point for most applications B Advanced Gas Sensing Technologies C Gatas User Manual SF6 AREACHECK P2 Room Monitor V2 48 Page 8 Indicator Lights 6 LEDs on the instrument s front panel represent its state as follows Placement of LEDs RUN This light acts as a life sign to confirm the SF6 AREACHECK P2 is in operation As long as no internal error condition is met it will blink at all times During the actual SF6 measurement normally this phase is very short it will blink at a faster rate Should this lamp stop blinking an internal error has occurred ALARM This light will be permanently on if the last measuring cycle has determined a SF6 concentration above the selected alarm threshold In case of an internal error state it will blink quickly The instrument s internal audible alarm works in parallel with this lamp 500 1000 1500 2000 ppm These 4 yellow lamps reflect the last sampling period s result in terms of alarm levels as designated On instruments equipped with the optional oxygen sensor these lights have the following additional functions e The two leftmost LEDs 500 1000ppm will blink when an oxygen concentration below the lower alarm threshold of 16 is present e The two rightm