Home

Pq100v7.1.1 - Environmental Instruments

image

Contents

1. 7 5 0 ACCESSORIES ss e i eerte ane a v E 7 5 l Cables sooo ota ane tho dede dui edel debe da debe dt R odis 7 5 2 Inlets Adapters and Filter Holders 8 5 3 Batteries and Chargers reais 8 6 0 OPERATION DETAILS onines oa sena ete nete mu ne ah ce ea cais anger am adaga 9 6 1 Turning ON the PQIOO eres 9 6 2 Stopping the Run oed eoi mo ois doi gi la a ic ai n d 9 6 3 Power Off osos o eo owes een Pd EG REN od S RON d de 10 6 4 Download the Data sssssse Henn 10 6 5 Programming the Run sss nnns 12 6 6 Select Delay ooo dee nex pesce exec uic ne he du 12 6 7 Selecting a Delay Start Time eem 12 6 8 Setup via Computer odis ed exe du Rua oe ease ete a Fa aaa 15 6 9 Set Fens Cal eios di eer End E EURO UR En 17 7 0 CALIBRATION et prado ot pnt efe bdo oL Tope de RET de h Ta pde Nt ane eo bl ed 62 ed Du da 20 7 1 Calibrate Flow Rate sss 20 7 2 Manual Calibration ceeduie ende rd i OC E ER CU ERN ER iai REPRE TET U 21 8 0 9 0 10 0 11 0 12 0 13 0 14 0 15 0 16 0 17 0 18 0 19 0 20 0 AC POWER SUPPLY CHARGER irren MAINTENANCE AND SERVICE eee 24 9 1 Replacement Component List sss 24 9 2 Rebuilding the Pump after 5000 Hours 26 WARRANTY INFORMATI ON ttenni rrek kett EAEE EEEE
2. 2 0 GETTING STARTED CHECKLIST 2 1 Included with the PQ100 PQ100 Main Unit PQ101 Battery Charger 120 240 VAC CQ2 PC Communication Adapter Cable Download Software also avail at http www bgiusa com aam portable htm User Manual also avail at http www bgiusa com aam portable htm PQ102 Hose Adapter QUU CU IE 2 2 Additional items included with the PQ 167 PM Spec 1 Rigid stand with legs 2 SSI2 5 PM 10 EPA FRM Louvered Inlet 3 F20 47mm Filter Holder for BGI 16 7 4 F21 Filter Cassette for use with F20 above 2 3 User Supplied Items 1 ADeltaCal or TetraCal formally triCal Calibrator Note Dry Calibrators and Rotameters are not recommended 2 Additional Filter Holder F20 with Filter Cassettes F21 2 3 Power Input RS 232 Adapter P N 3679 Note Permanent damage can occur if the PQ100 is operated without a filter in series with the Inlet If the unit is operated without a SSI head with filter and holder a suitable filter must be used to prevent damage to the pump and sensors The instrument will not run for more than a few seconds if it detects insufficient Static Pressure SP 2 4 ASSEMBLING THE PQ167R PM10 AIR SAMPLING SYSTEM Included with your PQ167R PM10 Air Sampler System are the following items Item s are indicated on the appended drawings located on pages 35 and 36 Page 36 14 36 34 36 12 36 11A 36 160 36 01 36 3 35 4 5 9 35 6 7 8 36 162 36 161 1 1 1 1 1
3. The BGI PQ100 is an Intelligent Air Pump that can monitor it s own airflow rate and thereby adjust the pump speed to compensate for changes in load pressure and or other forces which would otherwise hamper the flow of air through a filter or sample collector The PQ100 can be programmed to begin it s sampling job at a specific date and time and stop sampling after the user defined run time is completed A Liquid Crystal Display provides the operator with a readout of flow rate Barometric pressure ambient temperature Date Time and calibration functions The PQ100 operates at flow rates ranging from 5 to 20 liters per minute It may be Operated at EPA Standard conditions of 16 67 lpm The EPA standard conditions are a Barometric Pressure of 760 mm of Hg and a Standard Temperature of 25 C This is also commonly referred to as mass flow The instrument may also be set to run at actual conditions wherein the selected flow rate is maintained at the actual Barometric pressure and ambient temperature Certain default values have been built into this instrument to reflect EPA style air sampling procedures The default Run Time is 24 hours The default Start Date and Time is midnight tomorrow Principle of Operation The operating principle of the PQ100 can be appreciated by referring to the block diagram below Air is drawn by the pump through a size selective inlet device and or filter It then passes inside the instrument hou
4. 1 Scroll to Select Duration then press Enter button Selecting a Run Duration Time 1 The 000 Hrs selection should be flashing To change the number of hours use the Up or Down buttons then press the Enter button This moves the gt to 00 Min 2 To change the number of minutes use the Up or Down buttons then press the Enter button This moves the gt to Done 3 To accept the Sample Duration or Run Time press the Enter button This returns you to the Program menu where you will be prompted to start the run 14 Start Run 1 Press the Enter button to start the run 6 8 Setup via computer Once a file has been configured you now have the choice of setting up the instrument by using the buttons on the device or it may be set up by wire using the computer If the computer method is desired click on Set Up Run and the screen at right will appear Immediately click on the Sync Date Time button and wait for the acknowledgement to appear o Units Date Time Synchronized with Computer Having computer and instrument time in agreement is an important and often overlooked requirement of effective data capture The balance of the Set Up Run pane is intuitively set For EPA type 24 hour sampling select the Midnight and 24 hour buttons Your selections will immediately be displayed in the Start Date Time and Run Length window areas If some other dates times and run lengths are desired simp
5. PQ100 Firmware Version 6 0 or 1 0M and higher Air Sampler INSTRUCTION MANUAL PMi9 REFERENCE SAMPLER DESIGNATION NO RFPS 1298 124 BGI Incorporated 58 Guinan Street Waltham MA 02154 Tel 781 891 9380 Fax 781 891 8151 e mail info O bgiusa com www bgiusa com Manual Version 7 1 1 April 2013 PREFACE There are now two separate manuals governing the set up operation and maintenance of the PQ100 Air Sampler There have been detail improvements in the electronic and firmware functionality such that the addition of further appendices would render a single manual unwieldy Since Version 1 0M of the PQ 100 adds an improved button and screen layout as well as a choice of Standard Mass or Volumetric flow rate the time has arrived to split the manuals into two distinct versions Both manuals are current and will be maintained and updated as necessary In addition to the other improvements there is now a built in Language option which will shift all the display screens to Spanish En adision a otros progresos hay ahora una opcion de lenquaje la cual cambiara toda la pantaya a Espanol The correct manual for your instrument may be determined by the front cover The original manual is marked as follows PQ100 Air Sampler INSTRUCTION MANUAL Firmware Version 5 X and less The modernized manual is marked as follows PQ100 Firmware Version 1 0M and higher Air Sampler INSTRUCTION MANUAL This instrument has been specif
6. 1 1 1 3 1 1 1 2 1 1 Item Quantity Part Number PQ100 PQ101 PQ102 CQ1 CQ2 PQMAN HS5 TP100R A1634 SSI 2 5 J R3035 F20 F21 A1741 A1904 Description Air Sampler Main Unit Battery Charger Auxiliary Power Supply Hose Adapter Charger External Battery Cable PC Communication Adapter Cable Manual 1 foot Rubber Hose Tripod Frame Legs Inlet Water Jar Filter Holder Requires F21 Cassette 47mm Filter Adapter Cassette w Screen Long Down Tube Filter Adapter For complete system assemblies refer to Figures 1 2 3 4 5 6 and 7 located on pages 32 thru 38 3 0 HOW TO USE THE PQ100 IMMEDIATELY If you have already set up the PQ 100 to be placed in service the instrument may be operated immediately Referring to Figure 1 3 press the button labeled ON OFF When the main idle display screen appears press Enter The instrument will immediately begin running at the default flow rate of 16 67 lpm To stop the PQ 100 press Enter The complete details of the run will be displayed on the screen as shown in Figure 2 3 To cancel this information press Escape The PQ100 has been designed to be a highly interactive instrument Investigators familiar with air sampling using microprocessor controlled hardware will have little difficulty following the critical paths without further instruction PQ100 af N BGI mu Waltham Mass U S A Figure 1 3 PQ 100 Keypad File About Screen 4 m PQ
7. 9 Exploded Diagram of Inlet with Filter Holder 51 4014 Figure 10 Initial Disassembly of Inlet 52 The jet may also be removed from the top of the inlet as shown in Figure 11 4013 Figure 11 Jet Removed for Cleaning or Size Change 53 Field Data Sheet BGI Model PQ100 Field Test Site Identification Field Operator BGI PQ100 Sampler SetUp Date M TU W TH F SA SU FRM Inlet Serial RUN Run Date M TU W TH F SA SU Run Date Post Sampling Information Record sampler and site conditions below Sample data is available from the Review Last Run Data and Conditions screen Collected Filter ID Filter Condition O Normal O Otner Elapsed Time ET Total Volume m3 Error Flags Circle PQ F T M AVG Flow Rate Q Lpm CV Tmax C Tmin C Tavg C BPmax mmHg BPmin mmHg BPavg mmHg Comments FRM Service Information A ARA O ADO AL ee e Confirm performed tasks and record all data Work shall be performed as per the Field Test Schedule The calibration standard shall be DeltaCal S N 505 Visually inspect the sampler o rings filter holder housing electrical connections and the internal water trap Empty and clean the water collection jar as needed Note any abnormalities External Leak YES NO Results PASS FAIL Initial cmH20 CA dd Final cmH20 Internal Leak YES NO Results PASS FAIL Initial cmH20 Ss Ed Final emH20 Verified Ambient YES NO Res
8. No representative of ours has the authority to change or modify this warranty in any respect 27 11 0 TROUBLESHOOTING Problem Answer Problem Answer Problem Answer Problem Answer Pressing the Enter key causes the unit to shutdown This indicates that the PQ100 is now powered off and ready to begin a sample run at the designated Start Date and Time Be sure that the Start Date and Time are set correctly The Actual Flow Rate is not what is indicated in the Display If operation is set for Standard conditions variations in Atmospheric Conditions BP and Temp will cause slight variations in flow as displayed on the Calibration Device If operation is set for Actual conditions and the flow is incorrect check and if necessary correct the Temp and BP calibration and then recalibrate the air flow hear a popping metal sound as the PQ100 load pressure increases When running the PQ100 at very high load pressures i e 150 centimeters of water or better a strange metal popping sound may be heard by the user Do not be concerned This is merely the internal pulsation dampner Plenum adjusting its chamber volume The flow rate does not seem to hold when additional pressure is added to the load usually tested using a valve This is usually caused by a leak somewhere between the PQ100 Inlet and the calibration measuring device Make sure that the hose adapter is firmly tightened in the PQ100 inlet check t
9. Assurance Handbook In brief the filter should have the following characteristics Size Circular 46 2 mm diameter V 0 25 mm Medium Polytetrafluoroethylene PTFE Teflon with integral support ring Support ring Polymethylpentene PMP or equivalent inert material 0 38 V 0 04 mm thick outer diameter 46 2 mm V 0 25 mm and a width of 3 68 mm 0 00 mm 0 51mm Pore size 2um as measured by ASTM F316 94 Filter thickness 30 to 50 um Maximum pressure drop clean filter 30 cm H20 column Q9 16 67 Lpm clean air flow Maximum moisture pickup Not more than 10 ug weight increase after 24 hour exposure to air of 40 percent relative humidity relative to weight after 24 hour exposure to air of 35 percent relative humidity Collection efficiency Greater than 99 7 percent as measured by the DOP test ASTM D 2986 91 with 0 3 um particles at the sampler s operating face velocity Alkalinity Less than 25 microequivalents gram of filter as measured by the guidance given in reference 2 in section 13 0 of this appendix 39 18 2 Filter Handling Filters should be handled delicately using non serrated forceps never using fingers even in laboratory gloves to touch any part of the filter When not in use filters should be stored in protective cartons in conditions of moderated temperature and relative humidity Filters should from the always be transported from the laboratory to the sampling location in the filter cassette whic
10. Description 162 5 A1751 PQ 100 Down Tube 163 5 A2245 Stabilizer Assembly 165 1 B2923 Venturi 166 1 10001 3110 32 x 4 Socket Cap Screw 2 167 1 PQ10210 Venturi Hose 2 168 3 SC0103 44 40 x 1 Pan Head Screw 169 3 10212 Radiation Shield Assembly 170 3 1529 Inlet Gasket 171 7 Bunal38 O Ring 172 1 3476 Battery Clamp Rods 173 1 3382 Battery Clamp 174 1 NT10229 Battery Clamp Nuts Rebuilding the Pump after 5000 Hours To determine Pump Cumulative Time you must use the PQ100 download software PUMP cumulative time is the number of actual service hours of the dual diaphragm pump and is shown when either printing or screen viewing a download When this time exceeds 5000 hours the pump should be rebuilt The rebuild is a relatively easy task and requires the replacement of the diaphragms valves and bearing A kit of parts is available from the factory and includes instructions Service can also be performed by the factory Pump rebuilding instructions are online at http www bgiusa com aam pumpman 3 1 pdf Kit Description X014 Pump Rebuild Kit Includes valves diaphragms and aoe rings X015 Advanced Pump Rebuild Kit Includes X014 with bearing X016 Total Pump Rebuild Kit Includes X014 with bearing and motor X017 Pump Diaphragms Only 26 10 0 WARRANTY INFORMATION BGI Incorporated warrants equipment of its manufacture and bearing its nameplate to be free from defects in workmanship and material We make no warranty expr
11. US EPA Standard conditions for PM10 are 25 C and 760 mm of Hg This system is also referred to as Mass Flow gt Select F unit 19 7 0 CALIBRATION 7 1 Calibrate Flow rate The preferred way to calibrate the PQ100 is to use the tetraCal Direct Cal mode The tetraCal Direct Cal works as follows The tetraCal puts out a continuous stream of flow rate information in ascii format When the tetraCal Direct Cal mode is selected on the pump menu D cal the pump is instructed to look for the stream of flow rate data It then compares the tetraCal flow rate data to it s own flow rate information and calculates an offset and then automatically adjusts the pump motor speed to match the data coming from the tetraCal At the Setup and Calibration Menu 1 Scroll using the Up and Down buttons to the Cal Flow Rate position Press the Enter button to accept At D cal menu Connect the Pump and calibrator using tubing and filter 1 Turn the tetraCal ON and allow it to zero itself 2 Using the Up and Down buttons scroll to the D Cal position and press the Enter button The pump will automatically begin to run At this point the pump instantaneously compares its data to the tetraCal data and calculates an offset 3 When the flow readings on the PQ100 is stable press the Enter button 20 7 2 Manual Calibration Qa p TUBING FILTER HOLDER T PQ 100 1 Using the Up an
12. data to disappear Pressing ESC will cause it to reappear Run information is not lost until overwritten by a new run Total volume is displayed as Actual volume or as Standard volume corrected to sea level and Standard temperature The information will alternately be displayed on the last line 6 3 Power Off 1 Press the 1 O button to turn OFF the PQ 100 and frito 77 save the data for downloading later 2 At the Power OFF screen select yes and press the Enter button to turn OFF the unit When the Omni is turned ON the next time the data will be available for download or it can be recalled from the main idle display screen screen 2 by pressing ESC 6 4 Download the data x E il 1 Connect a standard 9 pin RS232 serial cable between the serial port on the PQ100 and either a USB port or serial port on the device being used for downloading 2 Click on the BGILink Icon on your desktop This will bring up the download and wire operating program a cia ta You will be presented with the screen shown below If this is the first time that an air sampler is to be downloaded to this computer program select a a sta Setup Configuration o Connect to he Omni unt uang the configued conmuncations poft and ietieve the dala hon the LLL etui Pa ee o vias select communscations pot and ngos Baliar Sehr Set he ype d battery the unt uang Dead heb ard explanatonz Dove and Et
13. of the display When the battery has reached full charge 16 hours typically the PQ100 will then display Charged in the lower right display THE PQ100 SHOULD ALWAYS BE CONNECTED TO THE CHARGER WHEN NOT IN USE THIS WILL PROVIDE MAXIMUM RUN TIME WHENEVER NEEDED AND WILL NOT HARM THE INTERNAL BATTERY 23 9 0 MAINTENANCE AND SERVICE 9 1 Replacement Component List The following is a list of components for replacement and or servicing the PQ100 and may be ordered from the factory Refer to drawing figures 1 thru 7 for identification by Item 1 5 4 4 A1752 5 4 BUNA135 6 4 Al727 L27 7 4 F1 U 8 4 A1729 L29 9 4 B1425 lla 5 B1740 14 5 PQ100 18 1 PQ103 19a 1 SC0101 190 1 WS0100 20 1 B1412 21 1 HS3 22 1 PM3026 23 1 B1413 24 1 SC1013 32 1 PQFP 33 1 3 SC0104 34 5 PQ102 35 2 B1292 36 2 B1293 37 2 A1288 38 2 A1281 39 2 A1289 40 2 B1283 41 2 A1287 42 2 DDRI 6632 43 2 A1294 44 2 SC0105 45 2 SC0106 46 2 A1290 47 2 PQMOTOR 48 2 BUNAO10 Description SSI 2 5Dicot Inlet Filter Holder Top O Rings Cassette Top Screen Cassette Bottom Filter Holder Bottom Stand Plate Main Unit Main Battery 6 32x1 4 Screw w washer 6x 372 Flat Washer Pulsation Dampner Plenum Hose Pump Assembly Chassis 8 x 5 8 Sheet Metal Screws Front Panel Assembly 6 32 x 3 8 Flat Head Screws Hose Adapter Valve Upper Plate Valve Lower Plate Diaphragm Retainer Diaphragm Diaphragm Retainer Bushing
14. set in a sunny location or the actual field sampling site and not run for 10 days The solar panel will fully charge both batteries Troubleshooting Battery is not maintaining at least 5096 charge B caused by inclement weather or excessive current drain Excessive current drain is caused by an excessively dirty filter or a worn out pump After installing a new filter if problem persists check for worn pump valves or diaphragms If either battery is in excess of two years old B replace Reference 1 Stand Alone Photovoltaic Systems A Handbook of Recommended Design Practice Available from National Technical Information Service US Department of Commerce 5285 Port Royal Road Springfield VA 22161 Document No SAND87 7023 43 19 5 Solar Panel Parts List Quantity Figure Part Number Description 1 11 SP 21 Solar panel assembly 1 14 A1920 PQ100 adapter cable TABLE I1 Clear Sky Insolation Data from Northernmost to Southernmost U S cities Months wherein full charge can be maintained City N Continuous Every other Third day Sixth day Latitude Sampling day Sampling Sampling Sampling Caribou ME 46E 52 NOTE This approximation is based upon 5 Kwh M received as being necessary to fully restore the PQ100 system whilst drawing 500 MA typical 44 ANGLE ADJUSTABLE STAND SOLAR PANEL 1 X 10 X 36 PINE BOARD DRIVE STAKE INTO GROUND 105 AH BATTERY MARINE TROLLING CLAMP SUPPORT TO PQ 100 REA
15. weighing the type and size of filters specified and have a readability of Vlug The balance should be calibrated as specified by the manufacturer at installation and should be recalibrated immediately prior to each weighing session 40 19 0 SOLAR PANEL POWER SUPPLY 19 1 Introduction The SP21 solar panel kit is intended to permit the PQ100 to run for extended or indefinite periods of time depending on the available sunlight solar radiation at a given location The solar panel may only be used as the sole source of power for a U S EPA designated instrument if sampling is not being performed every day i e continuously Because of the low current draw of the instruments they are highly amenable to this technique Given sufficient sunlight they may be deployed in locations where no line power is available The basic components of the solar kit are 1 32 watt solar panel with mounting brackets 2 Built in voltage regulator 3 100 amp hour approx ballast battery User supplied The purpose of the external high capacity battery is to provide back up power on days when there is little or no sunlight The recommended battery capacity will provide 7 8 days run time with little or no sunlight It will recharge almost completely after one days use during a day of full sunlight while the instrument is non operational Complete recharging of a fully depleted system would require 10 days This type of system should not be considered
16. 100 af a BGI m a Waltham Mass U S A Figure 2 3 Image of a Completed Run 4 0 SPECI FI CATIONS 4 1 4 2 4 3 4 4 Flow Rate Precision and Accuracy Flow Rate Selections 5 to 20 Lpm in 0 01 Lpm increments Flow Rate Accuracy 0 5 When calibrated with a DeltaCal or TetraCal formally triCal Flow Rate Precision 1 Run Times for Various Applications Flow Rate Media Type Pressure Drop Run Time 16 7 Lpm 47mm Teflo 10 8 cm H20 33 6 Hours 16 7 Lpm 47mm Fiberglass 25 4 cm H20 32 4 Hours 16 7 Lpm 47mm Quartz 33 0 cm H20 28 4 Hours 14 0 Lpm 37mm MCE 0 8 112 0 cm H20 13 0 Hours 12 0 Lpm 37mm MCE 0 8 94 0 cm H20 13 5 Hours 12 0 Lpm 25mm MCE 0 8 170 0 cm H20 11 6 Hours 10 0 Lpm 37mm MCE 0 8 6 0 cm H20 19 8 Hours Dimensions and Weights PQ100 Main Unit Only Dimensions Height 5 3 13 46cm Width 9 9 25 15cm Depth 9 7 24 64cm Weight 19 Lbs 8 63Kg Electrical Specifications Main Battery 12V 12Ah Short Circuit Protected Clock Battery 3 6V Lithium Cell Clock Battery Life 2 Years Motor Drive High Efficiency Pulse Width Modulator Flow Rate Measurement Flow Sensor Serial Data Signals Outputs 9V Nom Inputs 30V Max Charging System 100 250VAC 50 60 Hz Note The PQ100 can run from the Charger with or without the internal battery 4 5 Hardware Requirements for PQ Software All programs and utilities for the PQ100 Version 6 0 and higher require a computer running Windows XP or Windows
17. 7 with sufficient RAM memory for efficient performance Available software is furnished on a CD with new instruments and is available online at no charge from the BGI web site http www bgiusa com aam portable htm BP Barometric pressure with selectable units mm of Hg Millibars or in of Hg Ta Ambient Temperature Qa Actual flow rate at prevailing BP and Ta Qs Standard flow rate at standard conditions of 760 mm of Hg and 25 C 4 6 Dimensions Weights and Flow Specifications for SSI PQ167R Inlet Kit Including Rigid Tripod Assembly Flow Rate 16 7 Lpm Weight 16 Lbs 7 3Kg Tripod Diameter Footprint 51 1 29M Tripod Inlet Overall Height 76 1 93M 5 0 ACCESSORIES 5 1 Cables CQ2 Connects the PQ100 to a computer for downloading Communication Adapter Cable CQ2A optional or instead of CQ2 Connects the PQ100 to a computer for downloading while the Communication Adapter Plug charger or solar panel is connected CQ4 Use to connect the PQ100 to a large external Battery typically External High Capacity Battery Cable an 80 Amp Hour for greater than 48 hour run time 5 2 Inlets Adapters and Filter Holders PQ102 Hose Adapter For attachment of a rubber hose to the PQ100 PM10 Size Selective Inlet for seperation and collection of SSI 2 5 Inlet particles with an aerodynamic equivalent of 10 micrometers or less Connects to SSI 2 5 Inlet and holds Filter Media also used as F20 47mm Filter Hold
18. Always operate the PQ100 in a normal upright position The legs should be bolted down to prevent tipping in conditions of high winds Do not operate the PQ100 if any of the parts are defective damaged or missing iii CAUTION Moisture Entry into Sampling System The normal operation of the PQ100 is for outdoor ambient particulate sampling therefore it is exposed to changing environmental weather conditions BGI has built into the PQ100 several systems to minimize the entry of water The first of these is the design of the Inlets The 45 degree down turned pie pan shape of the inlets PQTSP SSI2 5 prevents rain from entering the system Additionally a drain and water jar on the side of the inlet is intended to remove any wind blown rain that might enter the system However fog and moisture from use in tropical or mountain sampling sites can enter the unit and a secondary water trap has been built into the PQ100 A purge valve is mounted on the underside of the PQ100 Tripod Stand Any water entering the system is collected in this water trap and the user must periodically press the valve and allow any water to drip out of the sampling system Itis a prudent idea to do this manual step between each sample event if the sampling site is located in a high humidity tropical or mountain site Reference the following drawing to locate the water trap and valve Simply press upward and the valve spring will open allowing the water to escape Note Do
19. NTRATION TOTAL SAMPLE VOLUME m m Where CONTAMINANT WEIGHT is the difference between initial filter weight and final filter weight TOTAL SAMPLE VOLUME is the Volume of air passed through the filter in cubic meters CONCENTRATION is the quantity of particulate matter in micrograms per cubic meter 3l 17 0 FIGURES Figure 1 Exploded Assembly View 32 86 4 PER SIDE 49 2 PER SIDE 46 2 anne S 41 48 4 PER SIDE o 38 2 2 0 6 36 2 Q 4 l i j CONCAVE SIDE Figure 2 Exploded View of Pump 35 2 33 67 170 55 2 168 3 52 4 53 4 57 E b A 2025 Figure 3 Exploded View of Front of PQ100 34 7 FILTER MEDIA CASSETTE ASSEMBLY ee E a 2418 Figure 4 Exploded View of Filter Cassette and Filter Holder 35 un 34 m a HOLDER 4 8 EX Ff YA a 11A 163 o Figure 5 Exploded View of PQ167R Stand 36 4022 1999 Figure 6 PQ167R Rigid Mounting Stand 37 153 1791 Figure 7 Exploded View of PM10 Inlet 38 18 0 FILTER HANDLING AND WEI GHI NG It is recommended that these guidelines be followed for PM sampling using the NO The following guidelines are based on the regulations developed for sampling PM2 5 18 1 Filter Specifications For exact compliance with EPA procedures for PM2 5 refer to 40 CFR Part 50 Appendix L and Section 2 12 of EPA s Quality
20. Pump Housing Follower Yoke Bearing as assembly only Eccentric Set Screw 2 5x8mm CheeseHead Screw Motor Mount Motor O Rings 24 Item Fig PN Description 49 2 A1402 Locating Pins 50 2 B1408 Valves 51 3 SC4014 4 40x1 2 Phillips Pan Head 52 3 NT3006 4 40 Hex Nut w Captive Washer 53 3 NT9088 312 56 Hex Nut 54a 3 SC3014 8 32 Phillips Pan Head 54b 3 LW4001 8 Internal Tooth Lock Washer 55 3 SC3028 10 32x1 2 Slotted 56 3 7 16 32 Jam Nut 57 3 ICA001 Utility Connector Cable Assembly 58 3 OM10109 Display 59 3 C2717 P C Assembly 60 3 B2816 Front Panel 61 3 B2905 Graphic Faceplate 62 3 10211 Handle 62A 3 2188 Handle Standoffs 63 3 A1419 A1420 Exhaust Port 64 3 A1418 Inlet Receptacle 65 3 CCC001 Cap and Chain 66 3 A1483 Connector Gasket 67 3 BUNA116 O Ring 71 1 BP001 Rubber Bumper 86 2 SC1014 8x1 Sheet Metal Screws 87 2 A1403 Inlet Outlet Tube 88 2 SC1013 8 x 5 8 Sheet Metal Screws 146 7 SC3030 6 32 x 3 8 Philips pan head screw 147 7 1705 L5 10 micron inlet top 148 7 1709 L9 Spacer 149 7 1709 L8 Screen 150 7 1711 L11 10 micron inlet sub top 151 7 1712 L12 10 micron inlet body 152 7 Viton 036 O ring 153 7 DI 101310 micron inlet tube 155 7 SC3032 1 4 NPT nipple 156 7 C3033Jar top 158 7 J R3035Glass jar 159 7 Buna 026 O ring A3258 Utility Port Cable A3259 Internal Battery Cable A3260 Temperature Cable A3261 Pump Cable 160 5 A1634 PQ Leg 25 9 2 161 5 F20 PQ 100 Filter Holder Items Fig PN
21. R LEG SEE ENLARGED VIEW BELOW HEIGHT ADJUSTING ROD REAR LEG WING NUT Figure I1 Setup of Solar Panel 45 NORTH SOLAR PANEL SOUTH Figure 12 Orientation 46 VERTICAL MEASUREMENT BEFORE USING THE CHART BELOW DETERMINE WHETHER YOU HAVE THE LONG STYLE 13 WIDE x 51 LONG OR THE SHORT STYLE 21 WIDE x 25 LONG SOLAR PANEL VERTICAL MEASUREMENT TILT ANGLE SHORT STYLE 21 x25 LONGSTYLE 13 x51 E ei qoe ox os 08 Dos 33 x2 84 3 Cas se 42 93 4 5 Figure 13 Setting Tilt Angle 47 12 GROUND S CONTRO BOX PQ 200 CPC PLUG BACK OF CPC SOLAR PANEL RECEPTICALE BATTERY 80 100 AH PQ 100 SOLAR PANEL ADAPTER PQ 100 CPC PLUG 1980 Figure 14 Wiring Connections 48 20 0 PQ167 High Altitude miniPM Inlet IMPORTANT FLOW RATE MUST BE SET FOR 5 lpm 20 1 Size selective inlet The size selective inlet will be familiar to all who have had experience with the Standard EPA Louvered Inlet in its original 16 7 lpm configuration The only two differences are that it has been scaled down to 1 3 its original dimensions and the acceleration jet in the impactor is changeable over five size ranges An exploded diagram of the inlet is shown in Figure 9 with all parts identified 20 2 Jet Differential Function TSP PM 10 PM 4 0 PM 2 5 PM 1 0 If a Size Selective Jet SSJ other than PM10 was ordered fu
22. SP Air Sampler System are the following items Item s are indicated on the appended drawings Page Item Quantity Part Number Description 36 14 1 PQ100 Air Sampler Main Unit 1 PQ101 Battery Charger Auxiliary Power Supply 36 34 1 PQ102 Hose Adapter 1 CQ1 Charger External Battery Cable 1 CQ2 PC Communication Adapter Cable 1 PQMAN Manual 36 12 1 HS5 1 foot Rubber Hose 36 11A 1 TP100R Tripod Frame 36 160 3 A1634 Legs 36 01 1 SSI 2 5 Inlet 36 3 1 J R3035 Water J ar 35 4 5 9 1 F20 Filter Holder Requires F21 Cassette 35 6 7 8 2 F21 47mm Filter Adapter Cassette w Screen 36 162 1 A1741 Long Down Tube 36 161 1 A1904 Filter Adapter For complete system assembly refer to Figures 1 2 3 4 5 6 and 7 located on pages 32 thru 38 30 15 0 INSTALLING FILTER MEDIA Referring to Figure 4 the filter media is placed on screen 7 and placed on the inner lip of cassette base 8 Cassette top 6 is then inserted into cassette base when pin of base is aligned with locating hole on cassette top Holder base 9 and holder top 4 are supplied with O Rings 5 already installed Loaded cassette is then inserted into base 9 with the screen side down and holder top 4 is then screwed into base 9 firmly as to insure against leaks 16 0 CONCENTRATION EQUATION For reference the following is the formula for Concentration used by the PQ Software CONTAMINANT WEIGHT mg x 1000 micrograms nen ne ene eee nee nee CONCE
23. d Down buttons scroll to the Manual position and press the Enter button The pump will automatically begin to run gt Adjusting Flow 1 Adjust the flow reading on the tetraCal or any other calibration device to match the reading on the PQ100 using the Up and Down buttons One button push is approximately equivalent to a change of 0 1 lpm Either button may be held down to effect large changes 2 Press the Enter button to accept the Calibration 21 gt Selecting Barometric Pressure Units 1 Scroll using the Up and Down buttons to the Select BP unit position then press the Enter button to accept gt Setting the Barometric Pressure Units 1 Move the using the Up and Down buttons to select the barometric pressure units that are correct for you 2 Press the Enter button to accept the units 22 8 0 A C POWER SUPPLY CHARGER The PQ100 should be Recharged before use to allow for the greatest available Run Time Connect the Charger to the PQ100 Utility Adapter socket located on the front panel Plug the Charger into an appropriate power source BE SURE TO CHECK CHARGER SPECIFICATIONS FOR CORRECT POWER SOURCE INPUT Charging is activated only when AC power is actually applied to the charger This allows the system to be used in a long run time application where the battery is to be used for power loss protection While the Charger is engaged the PQ100 will display DCin in the lower right corner
24. eene nnns 27 TROUBLI NG SHOOTING titre n ter a a ec vae i t p E 28 SYSTEM BATTERI ES ern ntn m xd gr em e od Ken I egi itd NO Ve VETE US 28 SHUTDOWN MESSAGES 2 2 pror ot veter ed ti be tu RE TED Tv FER be Pe PETERE e ET De Te RE Deo eA ee pos 29 ASSEMBLING THE BGI TSP AIR SAMPLING SYSTEM 30 INSTALLING FILTER MEDIA sssssmm menn 31 CONCENTRATION EQUATION Heer 31 FIGURES m a a Ea taa ea 32 FILTER HANDLI NG AND WEI GHI NG 39 ESSE canon E A E ETT 39 a K E EE E E E EE 40 DES eto eee Sn a Eq E 40 IBAMA rapa DR RI UT PESOS Ta 40 SOLAR PANEL POWER SUPPLY EEEE 41 19 1 Introductio RE SORBONNE AR RE RR voee acte RR ta dp bue t etudes 41 19 2 Operational Considerations ssssssse en 41 19 3 c eaten hatte MA Rp tins OO ERES du Udo o RR 42 TOA uoi oti edic ad E veta CRUEL Bd vibe EROR IO D E Y Cc a 43 TOS MM MU ULM DEED EARLIER IEEE IUE Ind 44 PQ167 HIGH ALTITUDE m niPM e menn 49 20 1 Size Selective Inlet iot ocio tee mot te pa ve oa Doo Do DE De Pa iE Ce PATET O eae ut eis 49 20 2 et Differential fe es eh eg he DSc oO VO Mr PO e E 49 20 3 Maintenance saio deser eadein e nim e aa ci RAT RU a a a de NF adia 49 20 4 Summary of Maintenance Items ssssssssseeemnn 50 PQ100 Sample Data Sheet ee enne 54 PQ100 Instruction Manual Revision History ver 6 0 and Higher 55 vi 1 0 INTRODUCTION
25. er with PQ TSP Inlet F21 2 Filter Cassette for use with F20 above Plastic cassette containing a screen and filter media PQ TSP Universal TSP Sampling Inlet Replaces SSI 2 5 Inlet for TSP sampling 5 3 Batteries and Chargers PQ101 Universal Battery Used to recharge system batteries and to power the PQ100 for Charger Auxiliary Power long run times when AC power is available PQ103 Replacement Battery 6 0 OPERATION DETAILS 6 1 Turning ON the PQ100 1 Press the I O button An initiation screen will briefly appear and then advance to the Main Menu This screen will have the serial number and the code version number gt At the Main Menu 1 Scroll using the up and down buttons to make a selection Position the gt in front of the selection and press the enter button gt Run Now Initiates a sampling event gt Run Programmed Setup and initiate a programmed sampling event gt Set Fcns Cal Setup Time Units and calibration Functions Set Language Choice of English or Spanish Note Prior to using the PQ100 it is wise to set up the Date Time and Preferences Advance to the Set Preferences section of this manual 6 2 Stopping the Run 1 While the pump is running press the Enter button to stop the run 2 The final run data will be displayed on the LCD Note After the Run pressing ESC will cause the elapsed run
26. ess or implied except as set forth herein BGI s liability under this warranty extends for a period of one 1 year from the date of BGI s shipment It is expressly limited to repairing or replacing at the factory during this period and at BGI s option any device or part which shall within one year of delivery to the original purchaser be returned to the factory transportation prepaid and which on examination shall in fact be proved defective BGI assumes no liability for consequential damages of any kind The purchaser by acceptance of this equipment shall assume all liability for consequences of its misuse by the purchaser his employees or others This warranty will be void if the equipment is not handled installed or operated in accordance with our instructions If damage occurs during transportation to the purchaser BGI must be notified immediately upon arrival of the equipment Return transportation charges are collect A defective part in the meaning of this warranty shall not when such part is capable of being repaired or replaced constitute a reason for considering the complete equipment defective Acknowledgment and approval must be received from BGI prior to returning parts or equipment for credit BGI Incorporated makes engineering changes and improvements from time to time on instruments of its manufacture We are under no obligation to retrofit these improvements and or changes into instruments which have already been purchased
27. for latitudes higher than 45 50E N or S or particularly overcast regions 19 2 Operational Considerations While the use of solar power is highly desirable from the standard of utilizing a renewable energy source and being freed from the need to locate a source of power in difficult situations there are some preliminary considerations Clearly the PQ100 is not operating directly from the received energy of the sun but rather from a battery which has been charged by that energy If a PQ100 were to be run continuously from the internal and recommended external battery 7 to 8 days run time could be achieved However considering only EPA designated sampling conditions i e sampling from midnight to midnight then it would be possible to run on alternate days yielding one day to replenish the energy used Given that this is accomplished in full sunlight while the instrument is running the extra day recovery reduces the need for full sunlight by 5096 If the popular one in 3 days or one is 6 days schedule is utilized the probability of complete replenishment is greatly increased Experience has shown that on cloudless days in the Boston area 5 Kwh M insolation will replenish the energy used by a PQ100 In order to determine the suitability of the PQ100 solar system for a given location Appendix A of the cited reference gives the insolation index for 54 locations in the US and other places throughout the world Given a one in 6 day sampling sc
28. h should be protected within a metal canister These canisters may be ordered directly from BGI 18 3 Filter Cassette Handling The filter cassettes provided for use with the PQ100 have been designed with an interference fit to prevent the cassette from coming apart easily therefore some care must be exercised when opening and closing the cassette especially when a filter is inside Always maintain the cassette in an upright position especially if the filter has already been used to collect particles To open place a clean flat blade device knife edge screwdriver blade against the outside edge of the cassette between the upper and lower halves and gently wedge them apart DO NOT TWIST THEM APART this could tear the filter Set the upper half of the cassette aside To close the cassette place it the upper and lower halves together and gently press them together being careful not to twist them When closed the two halves should seat snugly together with the backing screen securely held in place between the two halves 18 4 Filter Weighing Because of the small amounts of material collected an extremely high quality microbalance and carefully temperature and humidity controlled filter weighing room are recommended For complete EPA recommended details consult 40 CFR 50 Appendix L8 0 Federal Register J uly 18 1997 and Section 2 12 of EPA s Quality Assurance Handbook The analytical balance used to weigh filters must be suitable for
29. he inlet mechanism and filter holder You can usually isolate the suspect device by starting at the PQ100 inlet and working your way out 12 0 SYSTEM BATTERIES MAIN INTERNAL BATTERY If the Internal Battery should ever require replacement use ONLY the proper BGI Battery PQ103 EXTERNAL BATTERY The External Battery is used to provide double the Run Time Capacity of the PQ100 CAUTION REVERSAL OF THE BATTERY LEADS CAN CAUSE IRREPARABLE DAMAGE TO THE PQ100 CLOCK BATTERY The Real Time Clock Calendar is backed up by a Lithium Cell that will require changing once every 2 years It is a commonly available coin Cell CR203 28 13 0 SHUTDOWN MESSAGES The following is a description of the various messages that can be displayed by the PQ100 to indicate the reason for sample job termination or current status of the PQ100 Select Reset and the following screen will appear Select yes and the instrument will reset to a value that is nearly correct and the calibration process may be resumed If a Direct calibration procedure is selected and does not function a screen will appear as a notice and suggest a cause If a run is attempted with a battery having insufficient charge for the programmed time the following message will appear BATTERY VOLTAGE TOO LOW Corrective action is to recharge replace battery No data will be lost 29 14 0 ASSEMBLING THE BGITSP AI R SAMPLING SYSTEM Included with your PQT
30. hedule only Fairbanks Alaska is unsuitable for solar application in the months of November December and January These are clear sky tables and seasonal overcast must be considered in individual locals Table 11 comprises locations at various US latitudes and indicates operational months vs sampling schedules 41 There are other factors which will reduce the energy replenishment of the system and make accurate performance predictions difficult Amongst these are 1 Dirt on the solar panel 2 Extreme cold weather affecting battery performance 3 Extremely high particulate loadings causing high filter resistance and consequent high current drain 4 Old Aused up batteries more than two years old While all the preceding factors are to be considered in the deployment of a solar powered PQ100 They are not easy to quantitate The effect of too little sunlight will be noticed on the Apercent charge remaininge on the PQ100 s main screen Given perfect replenishment it will always read 99 If at any time it falls below 50 it is well to consider replacing the large battery with a fully charged one At a minimum given winter gloom a fully charged battery and an every other day sampling schedule a one month operating period is achievable at any location below 45E latitude 19 3 Setting up Subsequent to unpacking a new unit it is attached to the rear leg of the PQ100 as shown in Figure 11 Itis important that the board provided be loca
31. ically designed to meet or exceed the operational requirements of a Reference Method sampling device under 40 CFR Part 50 Appendix J Reference Method for the Determination of Particulate Matter as PM 10 in the Atmosphere and was designated a Federal Reference Method Sampler Number RFPS 1298 124 in December 1998 SAFETY The PQ100 should only be operated as described and for its intended use Because the PQ100 runs primarily from battery power all of the typical hazards associated with high voltages and internal A C wiring have been reduced or eliminated Personal injury damage to the instrument or fire can occur if the following electrical precautions are not observed Caution should always be exercised when attaching the A C mains power connection Do not attempt to connect mains power if the plug or wire is cracked or frayed Do not attempt to connect mains power if the power cord leads or outlet are wet Do not immerse power cords in water or other liquids Place power cords away from traffic and do not allow anything to rest on them during operation Do not overload AC outlets Do not attach improperly wired external batteries solar panels or power sources Do not open the control panel or handle any other of the electrical parts while power is applied to the PQ100 Always disconnect the power supply first In addition personal injury or damage to the instrument could occur if the following precautions are not observed
32. ly edit the information in the windows Select the instrument being controlled in this case the PQ 100 15 The Flow Rate will automatically default to 16 67 lpm but any other flow from 5 to 20 lpm may be selected Pressing the Set button will send the information to the PQ 100 A glance at the screen will show that the countdown has commenced If the PQ100 is running off of internal battery power the screen will not appear as the unit will hibernate until it is time to start Nevertheless the screen can be verified bypressing the ON OFF button The display will briefly appear before resuming hibernation 16 6 9 Set Fcns Cal 1 Scroll to Set Fcns Cal then press Enter button Setup and Calibration Menu Date and Time Set current date and time 2 Cal Temp Adjust temperature to match a tetraCal temp reading Cal BP Adjust Barometric Pressure to match the tetraCal BP Reading Cal Flow rate Adjust flow rate to match a tetraCal Flow meter gt Select BP unit Set the units preference for Barometric Pressure mm of Hg is the default value Making a Selection 1 To select Date and Time scroll using the Up and Down buttons then press the Enter button to accept Setting the Date and Time 1 Move the gt using the Up and Down buttons 2 Press the Enter button to select the item The item will then flash 3 Use the Up and Down buttons to correct the numeric value Press and hold
33. nergy production the wintertime the angle should be the latitude plus 15 The maximum summertime production is obtained at latitude minus 15 The tilt angle and one method of setting is shown in Figure 13 It may also be set with a user supplied protractor bubble level Wiring Connections On the back of the solar panel is a rectangular box from which two cables exit as shown if Figure N4 The cable with 5 16 inch ring terminals is intended to connect to the external battery described in section N 2 The white wire is positive The black wire is negative and the green wire with the tinned end is ground A 12 inch ground spike and terminal is provided The other wire emanating from the box is equipped with a CPC connector This wire is installed on the PQ100 panel normally used for the power supply cable This wire is installed in place of the power supply cable when running on solar power 19 4 Overall Operation and Troubleshooting Prior to deploying a PQ100 with solar panel it is prudent to ensure that the internal battery is fully charged This is accomplished by plugging the PQ100 s power supply into a source of line current for 16 hours Full internal battery charge will be indicated on the main menu display as 9996 or charged when the battery is full charged and the power supply is disconnected The external solar panel battery may be initially charged from any automotive battery charger Alternatively the PQ100 may simply
34. pregam 10 Selecting Setup Configuration will open the pane shown on the right The first time it is opened it will appear blank as above right There are several opportunities available at this point for the operator to exert his preferences The com port is an auto select feature and the correct number will appear when the dropdown menu marker is selected An example of a filled in Configuration pane is shown at right Click on Save and the configuration will be saved as you directed Now at the conclusion of a run you may click on Download and the following pane will appear Pressing the Save button will send the run information to the J OBS file from which it may be retrieved as desired Notes may also be added at this time 11 6 5 Programming the Run 1 Scroll the gt to Run Programmed using the Up and Down buttons then press the Enter button Program Menu Select Delay The amount of time before the sample event begins Select Duration Sample Run Time gt Start Run Start the program 6 6 Select Delay 1 Scroll to Select Delay then press the Enter button 6 7 Selecting a delay start time The Delay start menu offers 2 options gt Start Midnight Select Delay Option 1 gt Start Midnight 1 Select Start Midnight then press the Enter button gt Selecting a run time duration offers 2 Options gt Run for 24 Ho
35. rnished with your instrument it was furnished as a separate item The individual jets are hand detachable and removed installed by screwing in and out A light grease should be applied to the threads to prevent seizure J ets manufactured prior to May 2005 were not marked As a guide to their functional size refer to the table of approximate internal dimensions below Later jets were color coded and their functional size can also be found in the following table P N 1 D In 1 D mm Color 2599 0 38 9 6 Clear 2616 0 26 6 6 Blue 2741 0 14 3 6 Green 2617 0 11 2 8 Red 2618 2 holes 2 holes Black 20 3 Maintenance Items which require cleaning and maintenance are common to all ambient air sampling devices fitted with size selective inlets The inlet and the sampler may be considered two separate items for cleaning and maintenance purposes Cleaning should occur once every 90 days or sooner in highly polluted environments Until such time as sufficient experience has been gathered the unit should be inspected once a month In order to perform an inspection it is only necessary after removing the inlet from the top of the filter holder to unscrew the top from the bullet as shown in Figure 10 Normal cleaning of air sampling inlets is generally best done with clean water and lint free wiping cloths If an ultrasonic cleaner is available it is the preferred device as it will remove dirt from deep corners and pockets avoiding the need for further di
36. sassembly After ultrasonic or any liquid cleaning be certain to dry thoroughly before reassembling and placing in service 49 20 4 Summary of Maintenance tems Frequency Every 5 sampling days Monthly Quarterly every 3 months Maintenance item 1 Service water collector bottle 1 Clean inlet surfaces 2 Check inlet screen for any clogging 1 Inspect O rings Remove and lightly coat them with Vacuum grease 2 Clean impaction surface Frequency may vary depending on climate amount of particulate matter in the air weather and so on 50 1 Detail Part Qty Description 1 2583 1 2583 NOZZLE ENTRY 2 1729129 1 1729429 CASSETTE LOWER SECTION 3 024BUNA 1 024 ORING 4 1728 128 1 17284128 FILTER SCREEN 5 OM10123 1 NIPPLE 6 015BUNA 2 015 ORING 7 2589 1 2589 SCREEN 8 2672 1 2672 UPPER FILTER HOLDER 9 2584 1 2584 TOP 10 2596 3 2596 RECEIVER TUBE 11 2597 1 2597 OUTER TUBE 12 2586 1 2586 LOWER PLATE 13 01425 1 1425FILTER HOLDER 14 2585 1 2585 WIND DEFLECTOR 15 2587 1 2587 RAIN DEFLECTOR 16 OM10120 1 JAR TOP 17 OC13 1 JAR 18 2598 1 2598 IMPACTOR NOZZLE 1 9 2617 1 2617 NOZZLE INSERT PM 2 5 20 2595 1 2595 TARGET PLATE 21 135BUNA 2 135 ORING 22 10002 8 440x1 4 PAN HEAD 23 OM10124 3 2 56x1 8PAN HEAD 24 1727427 1 1727427 CASSETTE UPPER SECTION 25 2588 4 2588 SPACER 26 2602 1 2602 EXIT ADAPTER 4015 Fig
37. sing to the Flow Sensor The signal generated by the sensor is then routed to a microprocessor which determines if the flow is at the set value and adjusts the pump speed to maintain the correct flow rate Because the flow sensor is sensitive and all pumps produce pulsation to some degree a pulsation damping volume has been introduced to control this effect The microprocessor not only controls the flow rate accurately and precisely to the set point but also performs several other functions These include turning the instrument on at a preselected time and running it for a defined interval The flow is maintained by the processor to a designated value which for EPA PM sampling is 16 67 Ipm 1 m hr A pulse width modulated signal is configured and sent to the pump motor in a constantly updated manner based on signal information received from the Flow Sensor The microprocessor also stores all parametric information generated during the run period and configures it for presentation on the visual display and downloading to the software provided with the instrument The system includes a 12 volt battery and external battery charger A C power supply The power supply function permits operation if desired with no battery whatsoever SIZE SELECTIVE INLET PULSATION DAMPING FLOW Ae SENSOR OLU PUMP EX MICRO PROCESSOR BATTERY CHARGER VISUAL DATA DISPLAY BATTERY A C POWER SUPPLY SCHEMATIC DIAGRAM OF PQ 100 SYSTEM 1910 7 0
38. ted as shown in the figure with the battery placed on top of it This serves to anchor the lower end of the panel to prevent its lifting during high winds Failure to do so could result in damage to the solar panel and the PQ100 NOTE In due consideration of weight shipping expense and ready local availability a battery is not furnished with the solar panel kit However the recommended battery is known as a Atrolling motore battery This is a marine type battery used for low speed electric outboard motors They are equipped with handles and 5 16 inch binding posts with wing nuts Because of their marine specifications they are a Adeep dischargee battery which is also the type recommended for solar panel applications Direction The direction of the solar panel will be with its long axis from the north to south with the foot low end of the panel to the south Inasmuch as the cell will be attached to the rear leg of the PQ100 this means that the back of the PQ100 is pointing due south and the front of the instrument is due North as illustrated in Figure I2 Tilt Angle The tilt angle is defined as the angle of inclination of a solar collector measured from the horizontal The reason for tilt angle is because of the suns elevation will vary over a range of 47 from winter solstice to summer solstice 42 For the greatest annual energy production the tilt angle should equal the latitude at the location of deployment For best e
39. this operation only when the sampler vacuum pump is not operating Make sure the valve resets itself so there is no leak in the system DRAIN VALVE LOCATED AT BOTTOM OF INLET COLUME UNDERNEATH STAND PLATE iv TABLE OF CONTENTS Preface on iW ik UP teksti cs E E beetle ba wate at DD dave Br dU Das ii Safety oit ern nr aa Ree ree dm DM ORE APSARA RETESET RESORT TRC CRETE iii LO INTRODUCTION ieicseetin tede Sx qot Ra t Fe temas Fut a et ge wt E vein di pe dv 1 Principle of Operation sssssse meer 1 2 0 GETTING STARTED CHECKLIST ieee 3 2 1 Included with the PQIOO ccc cee cease sseee memes 3 2 2 Additional Items included with the PQ167 PMyo Spec 3 2 3 User Supplied Items ees rece d nd ce de ed e AE d 3 2 4 Assembling The PQ167R PMy Air sampling System 4 3 0 HOW TO USE THE PQ100 IMMEDIATELY ssssss HH 5 4 0 SPECIFICATIONS sas tei etate itle eec Sa 6 4 1 Flow Rate Precision and Accuragy sosssseee mme 6 4 2 Run Times for Various Applications sssssee 6 4 3 Dimensions and Weights PQ100 Main Unit Only 6 4 4 Electrical Specifications sssssesee 6 4 5 Hardware Requirements for PQ Software ssesseeeene 7 4 6 Dimensions Weights and Flow Specifications for SSI
40. to accelerate the speed of the numeric change 4 Press the Enter button to accept the value and the 9 will automatically advance to the next item 5 Select Done to return to the Setup and Cal Menu 17 gt Calibrate Temperature 1 To select Cal Temp scroll using the Up and Down buttons then press the Enter button to accept gt Adjusting the Temperature 1 Compare the temperature reading from the PQ100 to a tetraCal or other standard 2 If they differ change the numeric value on the PQ100 using the Up and Down buttons 3 Press the Enter button to accept the value and return to the Setup and Cal Menu gt Calibrate Barometric Pressure 1 To select Cal BP scroll using the Up and Down buttons then press the Enter button to accept gt Adjusting the Barometric Pressure 1 Compare the Barometric Pressure reading from the PQ100 to a tetraCal or other standard 2 If they differ change the numeric value on the PQ100 using the Up and Down buttons 3 Press the Enter button to accept the value and return to the Setup and Cal Menu 18 gt Select Flow Rate Measurement Flow rate may be controlled in two modes either as Actual Flow which means the flow rate at the instantaneous Barometric Pressure and Ambient Temperature in which case it is known as Qu Alternatively Standard flow may be selected This is the flow rate at a set of standard conditions In the case of the
41. ults Sampler Reading C A Acceptance Criteria Ft al ad Standard Reading em e Verified Barometric YES NO Results Sampler Reading mmHg Acceptance Criteria dosi Standard Reading mmHg Tu NR Verified Sampler YES NO Results Sampler Reading Lpm Acceptance Criteria Wt Standard Reading Lpm AES AG SRO IM Verified Sampler YES NO Results Sampler Reading Acceptance Criteria oe ke Standard Reading eai Field Blank YES NO Filter ID If acceptance criteria are not Installed met calibration adjustment 1 40 Samples of the sensor s is required Sample Run Setup Information Record the following information regarding the subsequent sample run Use the Run Sampler w User Defined Start Stop screen to schedule the event Installed Filter ID Start Date amp Time Stop Date amp Time Signature PQ100 Instruction Manual Revision History for I nstruments Having Firmware Version 6 0 1 0M and Higher Version 7 0 Created New Manual Version 7 0 1 Updated Version 7 0 2 Updated Fig 1 Version 7 1 0 New Download instructions Version 7 1 1 Corrected P N in section 19 1 November 2008 April 2009 April 2010 May 2011 April 2013 55
42. urs Automatically sets the run time for 24 hours Select Duration Allows for setting a customized run 12 gt Run for 24 Hours 1 Select Run for 24 Hours then press the Enter button At this point the programmed run begins gt The Delay Start Data Screen provides information about the run and the time counts down until the start time is reached gt Stopping the Run before the Pump turns ON 1 Press the Escape Button The Display returns to the Main Menu gt Stopping the Run after the Pump turns ON 1 After the pump is running on a delay program press the following code to stop the run Up Down and Enter The run will be aborted and snap to the Run Interrupted screen 2 The final run data will be displayed on the LCD User Aborted The Run Interrupted screen displays information up to the moment the run was halted Pressing the Enter button will return to the main screen Delay Start Option 2 gt Selecting a custom delay start time 1 Scroll to Select Delay then press Enter button 13 gt Select Delay 1 If you need to change any item on this menu Use the Up or down buttons to scroll then press the Enter button 2 To change the number use the Up or Down buttons and press the Enter button to advance to the next item 3 Pressing the Enter button Done advances to the Select Duration screen Select a custom Run Duration time

Download Pdf Manuals

image

Related Search

Related Contents

Sea Gull Lighting 8720-15 Installation Guide  User- and installation manual Sirius  Konftel Conference Phone 50 User's Manual  Télécharger - Mairie de Villemotier  Flame Light 1,5m LED V1  Série ZSE30/ISE30  « Le Tikoun - Time is the Time ! »  Metra 99-5026 mounting kit  MANUEL D`UTILISATION  Equipo para auto Nokia CK-100 Manual del usuario y guía de  

Copyright © All rights reserved.
Failed to retrieve file