Home
        EMIS-MASS 260 User manual English v1.3.1
         Contents
1.                                                6 00000                           esos   gt                                                                                                       ZME 040000                                      Mc 1000      10 05 40                                                     ModBus  10 05 48                                                                                                                         10 05 48                                      Figure    3                                          Output signals  tab    In the group of parameters                                Current output  you can choose the measured  parameter for the current output                                     Mass flow rate                                       Volumetric flow rate                           Density   and also its value relating to maximum current of 20 mA     In the group of parameters                                       Pulse output  you can choose the measured  parameter for the current output                                     Mass flow rate                                       Volumetric flow rate                           Density   and also its value relating to 1 pulse on the pulse output                                                    Hardware monitoring  tab displays the state of the flowmeter and its  hardware parameters  such as measuring tubes oscillation frequency  voltage on the left and right coils  drive  coil 
2.                                              Automatic flow simulation  when the sequence of frequency  and current test signals appears on the flowmeter s outputs  see Menu item 44                                              Reset totalizers  see Menu item 18      The                                            Extended functions  tab contains the following parameters                                               The time delay between left and right coils taken as zero flow rate  see  Menu item 16                                                            Enable disable multipoint correction function  see Menu item 81   used for correction of the calibration coefficient in several points of flow rate  time delay  below     For different values of time delay between left and right signal coils and corresponding mass flow  rates DO   07                                  B            X  DX           the corresponding correction coefficients KO     K7                                               B            X  KX    are specified  Those coefficients are applied only of Multipoint  correction function is enabled     WWW flow nt rt ru                                    ID                  ID        500501502502                                                         _       Marcosi raco  a mew 0                   Kosdubuuusum                                 0                                       e mowe 1  IM frau                  g            L E    _ 1 44513   11 00000  1 1 44
3.                           Density calibration  contains 5 parameters                                                                    High density value corresponding to the medium of high  density  water                    BepxHero                 Mkc     Measuring tubes  oscillation frequency for the medium of high  density  water                                                                    Low density value corresponding to the medium of low  density  air                                                   MMC     Measuring tubes oscillation frequency for the medium of low  density  air                                                          99 100 C     Temperature correction coefficient for density as  percentage of measured density correction for temperature deviation of 100        WWW flow nt rt ru              EWis wass 260    On the                                      Output signals  tab parameters of the outputs are displayed     i                              V         7    9         6           _   ___ ____ _____ __ _      ___ _ _ d                                     3  xeo                                 Hepe                                                                                                                                                                                                                                                              ID        500501502502                           gt        5                                      
4.                   Connection  tab    WWW flow nt rt ru              EWis wass 260                                                                   APPENDIX C  Wiring diagrams  Transmitter Sensor   lt  L     Line     Line Function Color  1 1   Left signal coil   Brown  2 2   Left signal coil   Red  3 3   Right signal coil   Orange  4 4   Right signal coil   Yellow  5 5   Drive coil   Green  6 6   Drive coil   Blue  7 7   Temperature   Gray  8 8   Temperature   White  9 9   Temperature compensation Black                                                               H                    Power supply             lt                     Figure C 2     Power supply connection  WWW flow nt rt ru                      5         260 USER MANUAL       Output signal    Power supply         F  F I l          Power supply    AN reli B A F I  I  B A          Te an    F  F   l  B A    Figure C 4     Current output connection    Output signal          F  F l  1  B A    Power supply       Figure C 5     Digital output connection    WWW flow nt rt ru              EWis wass 260    APPENDIX D       Modbus register map version 2 xx  EMIS     Modbus register map version V2 17    Interface mode RS 485  asynchronous  half duplex  Baud rate 1200  2400  4800  9600 bit s    Data format 1 start bit  8 data bits  No parity bit and 1 or 2 stop bits  or 1 Odd Parity  or Even Parity  and 1 stop bit   Can be set in the Menu items 30  31   Default data format is 9600 8 N 1    Maximum length of data frame 1 byt
5.                  36     10 05 40                                                     ModBus     10 05 48                                                                                                                         10 05 48                                                   10 05 51                                                               36       Figure B 6                                            Working parameters  tab                                                     Low flow cutoff  see Menu item 17  in the units of kg h                                                 Flow direction options  see Menu item 19   If                   SIG  is chosen  then only direct flow will be accumulated  direction corresponding to the arrow on the sensor body   II                              BID  is chosen then flow will be counted on both direct and reverse directions                                         Reverse flow counting options  see Menu item 20   If                                INC  is chosen then the flow will be added to mass and volume totalizers  If                         DEC  is chosen  then the flow will be subtracted from mass and volume totalizers                                                    Enable disable Oil water analyze mode  see Menu item 52    The parameter group                   Procedures  allows the user to perform the following procedures                                       Zero calibration  see Menu item 16                    
6.              EMIS MASS 260    1 5 Explosion protection       The flowmeters of explosion proof modification  EMIS MASS 260 Ex   have a combined type  of protection  flameproof enclosure  complied with GOST R 51330 1  and the input and output   intrinsically safe  level   ib   complied with GOST R 51330 10  The sensor explosion proof  grade is showed in Table 1 7     Table 1 7     Sensor explosion proof grade    Temperature code Explosion proof grade          100    1Ex iblICT4X     200    1Ex I   IIC I 3X     350    1Ex I   IIC I 1X    Transmitter explosion proof grade is 1Exd ib lICT6X     Explosion proof grade is written on the name plates attached to the body of the sensor of  explosion proof modification and to the transmitter     The name plates  appearance is showed in the paragraph 1 6 Name plates      The  X  letter in the explosion proof grade means the special requirements       the measured medium temperature must not exceed the maximum temperature  according to the explosion proof grade temperature group       explosion protection is provided under pressure not exceeding the maximum allowable  pressure for the given modification       connection of external circuits to the flowmeter must be implemented through the  cable entries complied with GOST    51330 1       unused cable entry must be closed with the end cap supplied by the flowmeter s  manufacturer or other end cap complied with GOST R 51330 1      connection of the external devices to the pulse  current an
7.           upper limit of the full volumetric flow range  L h     Additional error of measurement of mass  volumetric  flow rate  caused by a change  of medium temperature is  0 05   of the maximum flow rate for every 10   C of  deviation from the zero calibration temperature     Additional error of measurement of mass  volumetric  flow rate  caused by a change  of pressure is  0 02   of the maximum flow rate for every 100 kPa of deviation from  the zero calibration pressure     The effect of changes in temperature and pressure can be adjusted by zero  calibration under the actual pressure and temperature  see paragraph 2 5 4 Zero  point adjustment       1 3 4 Power supply Electrical power of the flowmeters  depending on its voltage modification is provided  by an external DC 24V power supply or AC 220 V with frequency  50   1  Hz     Parameters of the power supply are presented in Table 1 3     Table 1 3     Power supply parameters    Nominal voltage Voltage range Maximum power consumption    DC 24 V 18   30 V 15 VA  AC 220 V 187   242 V 15 VA    1 3 5 Output signals The flowmeter has the following output signals     e pulse output   e current output   e digital output  RS 485 interface      To display the values of the mass flow  volume flow and other measured parameters  the flowmeter s transmitter has built in display     WWW flow nt rt ru                   EMIS MASS 260    1 3 5 1 Pulse output Pulse output signal is a periodical pulse signal with the frequency which is  p
8.          of the measured medium   Each tab has                                        Query period  field  This is the period of reading displayed  data from the flowmeter in the units of us  default value is 1000 us                                    PT                                                                                                                                                                                                                     NI                                                Mzuceenutit                                         100000                                             2250012500                                                             0 0000   LE  un 19527 34082                         zicu3 0 99596    Ey                            C 127  05553                     10 05 27                                                               36  10 05 40                                                     ModBus    10 05 48                                                                                                                                                                                           36    Figure B 1                               Monitoring  tab    WWW flow nt rt ru              EMis mass 260    On the         Calibration  tab the main calibration parameters are displayed                                         ID                            E   TORT                        paezada   gt           i I  ID hex  500501
9.      GOST    51330 0        GOST    51330 1        GOST    51330 10        Instruction BCH332 74 MMCC     Instructions for installation of electrical  equipment  power and lighting lines in hazardous areas          Other regulations in force within the enterprise     During installation you should pay attention to the specific operating conditions  mentioned in Section 1 5  Explosion protection      Before installation the flowmeter should be inspected  You should pay special  attention to explosion proof grade labels  warning labels  make sure that there is  no damage to the flameproof shell and flow sensor  Also make sure that  grounding terminal and the seals for cables and covers are available and in good  condition  also check the connecting cables condition     Upon completion of the electrical installation the electrical resistance of the  ground line should be checked  It must not exceed 1 Ohm  A copper wire with  section of at least 2 5 mm2  AWG13  should be used for grounding     Unused cable entry must be closed with the end cap supplied by the flowmeter s  manufacturer or other end cap complied with GOST R 51330 1     During installation  check the explosion proof surfaces of mounting parts  involved in providing explosion protection  Scratches  dents  chips on the  explosion proof surfaces of those parts  which are in the explosion path   are not  permitted     After the completion of the electrical connections it is necessary to close the  transmitter covers fir
10.     1ExiblIC T1 T4 X IP65   20      lt  Ta  lt   40             EMIS    CJSC  RUSSIA www emis kip ru Q       Figure 1 5     Sensor name plate    Sensor name plate contains the following data     Trademark of the manufacturer    Flowmeter name    Ex sign  for the flowmeters of  Ex  modification     Explosion protection grade  for the flowmeters of  Ex  modification    Ambient temperature range    IP grade     Number of explosion proof certificate and certification agency name  for  the flowmeters of  Ex  modification     Information about manufacturer     CAUTION     Before installation of the flowmeter  make sure that the information on the  name plates conforms to the data in the order        After verification procedure the flowmeter is sealed  Sealing is performed using  the seal and wire threaded through the holes in the front cover of the transmitter     Its also allowed to use a sealing tape attached to both the transmitter shell and  the front cover of the transmitter               EMis mass 260    1 7 Delivery set       The base delivery set and the accessories for the flowmeter are shown in Figures  1 6  1 7 and in Tables 1 12  1 13        Figure 1 6     Base delivery set       Table 1 12     Base delivery set    Item Base delivery Special       Coriolis Mass Flowmeter Cable    EMIS MASS 260   length  2 User manual    3 Passport of the flowmeter H  4 Packing box    5 Verification methods document    Available at the    Web site    WWW flow nt rt ru               EMIS
11.     2     Incorrect Register address            Data value is out of range of 0     65535  04H   Failed to write single register    Function 08  Diagnostics  supports only subfunction code 00   Quer    Slave address Subfunction Data value CRC           XX              Response     Slave address Subfunction Data value           XX         Error     WWW flow nt rt ru                 EWis wass 260    Exception code    01H   Function or subfunction is not supported           Incorrect Data value   04H   Failed to diagnose    Function 16  Write Multiple Registers    Query     Slave address Function Starting Registers Byte Data value CRC  address quantity count       _          oH              _    Response     Slave address Starting address Registers quantity   o  wu    j                1        Error     Exception code    01H   Function is not supported   02H   Incorrect Starting address or Registers quantity            Registers quantity      is out of range of 1     123 or Byte count is not equal to      2  04H   Failed to write multiple registers       Function 17  Report Slave ID  Query     Slave address CRC   o              f    Hesponse                                  A   FH               Error                      O    Exception code   01H   Function or subfunction is not supported  04H   Failed to fetch Slave ID    Minimal query period is 32 us  Recommended period is 125 us     WWW flow nt rt ru       11 EMIS MAsS 260    Example of queries     Read Reverse flow total cou
12.     Digital  3     Current  secondary   4     Discrete output  5     Discrete input       WWW flow nt rt ru                   5         260 USER MANUAL  Ades   Mode                               1167 Output channel B type  0     Current  primary   default   1     Pulse  2     Digital  3     Current  secondary   4     Discrete output  5     Discrete input  1168 Output channel C type  0     Current  primary   default   1     Pulse  2     Digital  3     Current  secondary   4     Discrete output  5     Discrete input      For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7              To write in registers the switches 2 4 6 8 must be in ON position     32 bit registers     single precision floating point values complied with IEEE 754 format  4 bytes  high bytes first  Functions  Read     04  Write     16    For example  value     1 5    corresponds to    OxBF          0x00 0x00  in the order from low to high address in  memory      Aes   Mode   Deschplon   Mentem          0100 Low density cutoff  g cm               Low density for density calibration                    0157 0158   RW   High density for density catoraton             0159 0160   RW   Period forlowdensity us      0161 0162 Period for high density  us  DEPT  0163 0164   RW   Temperature correction coefficient for density     100     6   Temperature averaging time  s            0193 0194  Low mass flow cutoff  kg h   Low volume flow cutoff  L h   5   5    0199 0200 
13.    EM 260 000   000 000 00 UM  27 11 2012  v1 3 1    Direct mass  measuring    High accuracy    Digital signal  processing    Straight  upsteam    downstream  piping not  required    High viscosity  liquids  measuring      EMIS    CJSC  Russia    CORIOLIS MASS FLOWMETER    EMIS MASS 260      USER MANUAL                   EMIS MASS 260       General This user manual provides basic technical specifications  instructions for usage   inf   storage and transportation  and other information necessary for the proper                     operation of the Coriolis Mass Flowmeters EMIS MASS 260  hereinafter   the     flowmeter  or       260       Modbus   is the abbreviation for Modicon Modbus Protocol and it is a registered  trademark of Modicon  Inc     ProLink   is a registered trademark of Micro Motion  Inc      EMIS   CJSC reserves the right to make changes in the design of the flowmeters   not deteriorated of their consumer characteristics  without prior notice  If you need  additional information on the EMIS equipment  please contact your local dealer or  the head office     Any use of the material of this publication  in whole or in part  without written  permission of the copyright holder is prohibited     CAUTION     Please contact    7 843 206 01 48  flow nt rt ru  www flow nt rt ru    CAUTION     This manual applies only to the Coriolis Mass Flowmeters EMIS MASS 260   This document does not cover other products of  EMIS   CJSC and the  products of other companies        www 
14.   GOST H 51330 1         electric load of anti spark circuit elements of the flowmeter does not exceed 2 3 of  their certified values     WWW flow nt rt ru               6 485 260        the values of parameters of the left right signal coils  drive coil and the temperature  sensor circuits do not exceed the limits in accordance with GOST    51330 10            spark safety barrier with Zener diodes is applied         electrical clearances and creepage distances comply with GOST    51330 10   Insulation resistance between the sensor outer shell and electrical circuits elements can  withstand the testing voltage of 500 V AC rms         internal capacity and inductance of the circuit do not accumulate energy  dangerous  for spark ignition gas mixtures of category IIC         current conducting elements and electronic components of the flowmeter circuit are  protected from the environment influence with the shell  which provides the protection degree  IP 65 according to GOST 14254     Block diagram of the flowmeter is shown in Figure 1 3     Explosion proof type Transmitter    Connection        Intrinsically safe type Senso    Inside connection for Integral type  Special cable for separate type    Safety    Fuse Power transformer    barrier       Signal output circuit    Optical  Coupler    Reliable Power transformer  Fuse    220VAC       Detect circuit         S    Figure 1 3     Block diagram of the flowmeter    Input and output parameters of intrinsically safe circuits are sh
15.  1 5     Current output signal parameters       Current output signal    Current value 4   20       Load resistance 250   600 Ohm    1 3 5 3 Digital output Digital interface complies with the requirements      EIA TIA 422 B         recommendations ITU V 11 and provides the opportunity of networking and  transferring of all measured parameters  The digital interface specifications are  presented in Table 1 6     Table 1 6     Digital interface specifications       Digital interface    Standard EIA RS 485   Data transfer protocol Modbus RTU   Data transfer baud rate 1200  2400  4800  9600 bit s   Maximum distance 300 m   Data format 8 data bits  1 start bit  no parity bit  2 stop bits  default      WWW flow nt rt ru               6 485 260    Data format can be changed in the menus 30  31        The following measured parameters can be transmitted through the digital interface   mass  volume  flow rate  mass  volume   density and temperature of the medium     Digital interface can also be used to calibrate and to configure the flowmeter     Flowmeter supports two versions of the Modbus register map           EMIS   register map version 2 xx is supported by the   EMIS Integrator    software and set as default at factory     e   ProLink   register map version 3 xx compatible with ProLink II software     Register map and description for   EMIS   version 2 xx are presented in the  Appendix D  for   ProLink   version 3 xx     in the Appendix E     Register map version switching can 
16.  20     OilDen g mL 20  C WaterDen g mL GasContainCorre ContainCorre i LowerDen h mL  OFF X XXX X XXX X XXX X XXX    14 452    13      Change Password EnterOldPassword EnterNewPassword NewPasswordAgain PWD Succeed  Press E to enter 000000 000000 INewPWD 000000    60         LI               D   Calibration FlowCal   g s us Temp Correction   H Period K2 us ero Correction  Press E to enter XXX X X XX XXXXXX X     00  0 68 168    6  Inner Para Freq xxxx Hz Drv xxx   Pred xxxxxx us PUL  xxxxx Hz                 XXXX XXXX  Press E to enter Time x xxx us L xx Ind CUR  xxxx mA i XXXX XXXX XXXX XXXX  7    76  E     Y  System Para Temperature Cal Temperature Zero No X TimeValue No X CorrectValue  Press E to enter X XXXX X XX   ENG OFF Press E to enter x XX XX XX XXX       Figure 2 14 Display menu structure               oss    Table 2 7     Displayed parameters       Menu        Button     Indication format one    0 EMIS Yinuo Ver 2 28  2012 03    Flow XXX XXX  Mass XXX XXX    Description               Measured parameters display    BEN Version and data of electronics  Showed on flowmeter s power up   Total mass and instant mass flow rate        Flow XXX XXX 3 54 Total mass and instant volumetric flow rate  This menu item is not displayed when Menu item 52          XXX XXX      OW Analyze Switch  if set to On    Oil XXX XXX A 51 Mass flow rate of pure oil and percentage of water in the oil water  This menu item is displayed only  2a Water XXX     if Menu item 52    OW Analyze Switch  
17.  4 to 20 mA in  Press   E   to enter steps of 0 5mA  frequency on the pulse output changes from 0 to 12000 Hz in steps of 375 Hz   Values change every 5 seconds        None          2  4                EMIS MASS 260         Flow Sim Start  47  12  Flow simulation mode starts by pressing   E    Simulated flow rate is specified in the Menu item 45     46 Press   E   to enter E     edit    Simulation stops after pressing                            a      Button Tet  5 fom Indication format TOREM Description                46     12  Input mass flow rate used in flow simulation  Signals on pulse and current outputs will correspond to  3 EN FlowSimulate t h     X  XX       x                Hold Select the menu item  which will be displayed if no button is pressed within 2 minutes  Select  Mass   First Menu Mass to move to Menu item 1   Volm  to move to Menu item 2  If  Hold  option is selected moving to other  Volm menu items will not occur     CheckSystem OFF  49     12     Gtate ON E _ edit Enable disable system check function        150   gt 12        Flow Filtering s X XXX E _ edit Flow rate averaging time from 0 5 to 10 s in steps of 0 5 s        47    49       A  Es       Flow compensation by percentage of gas in the oil water  It works only if oil water analyze function is                NO    53 20  C OilDen X XXX g mL    54 209   WaterDen X XXX g mL     56  13   E     edit     57  18  Flow compensation by percentage of pure water in the oil water  not realized in current 
18.  MASS 260 USER MANUAL       Figure 1 7 Accessories    Table 1 13   Accessories    Description    1 Connection kit  flanges  gaskets  stud bolts  nuts  washers   EMIS MASS 260 CK       2 Flowmeter s replacement mounting part EMIS VEKTA BT260    CAUTION     Upon receipt of the flowmeter it is necessary to       Check the packaging box for damage               sure that delivery set is complete   e Make sure the flowmeter model matches the order data     If the package is damaged  delivery set or flowmeter model doesn t match  the order you should draw up a statement        WWW flow nt rt ru              Enis mass 260    1 8 Model codes     EMIS MASS 260    flowmeter   s model codes are presented in Table 1 14     The example of the flowmeter   s modification for the order is showed below          BE urs    EMIS MASS 260  Ex   050     200   220     0 25            Table 1 14     Flowmeter model codes    Explosion protection        General type       Explosion proof type Ex iblIC T1 T4 X for sensor   Exd ib lICT6X for transmitter    X Special order    Ex       Nominal diameter  DN            10 mm  DN   15              25              40              50              80       100        100 mm  150        150 mm  200        200       X opecial order        Typical range in accordance with Table 1 2    X opecial order    Transmitter placement      Integral type          D Separate type with the connection cable length 3 m   DXX Separate type with specified connection cable length X
19.  Modbus register map version 3 xx  ProLink     WWW flow nt rt ru             EMIS MASS 260    1 PRODUCT DESCRIPTION AND TECHNICAL DATA       1 1 Applications    The flowmeter is designed to measure the mass and volume flow  density  mass  and volume of liquids  and use the gathered information for technological  purposes or commercial accounting     The flowmeter is used as the counter of gasoline  liquefied petroleum gas   kerosene  diesel fuel  oil  oil water and other liquids in the chemical   petrochemical  oil  food  pharmaceutical and other industries and public municipal  facilities     The flowmeter is used in technological processes automatic monitoring and  control systems in various industries  for stationary technological plants  land  mobile refueling and pumping equipment  and in commercial accounting systems     The flowmeter is designed for use in explosive safe and explosive environment   The flowmeter of explosion proof modification  EMIS MASS 260 Ex   has a  combined type of protection  flameproof enclosure  complied with  GOST R 51330 1  and the input and output  intrinsically safe  level   ib   complied  with GOST R 51330 10     CAUTION        The flowmeter is not intended for use at nuclear facilities     WWW flow nt rt ru              EMis mass 260    1 2 Principle of operation       The flowmeter consists of the following units  as shown in Figure 1 1    e Flow sensor  1    e Transmitter  2         Figure 1 1     Appearance of the flowmeter    Transmit
20.  of the flowmeter in conditions of pressure or    temperature exceeding their maximum allowable values is prohibited     Do not use the flowmeter with the cover opened  and also without the  chassis grounding        To prevent leakage of the gas filling the sensor s  outer shell  do not open the cover 1        WWW flow nt rt ru              EMis mass 260    2 3 Installation on pipeline    2 3 1 Installation When selecting the installation place you should follow these rules   location         There should be no strong vibration  high temperatures or strong magnetic  fields at the installation place  Therefore  do not install the flowmeter close  to transformers  power units and other mechanisms that create vibration  and electromagnetic interference          he flowmeter should not be installed in the strained section of the  pipeline and should not be used as a pillar for the pipeline        It is recommended to provide moisture protection for the flowmeter     e Flowmeter should be installed in easily accessible places  Free space  around the flowmeter should be provided for easy installation and  maintenance     e Display of the flowmeter should be installed at a place which is convenient  for the operator to read displayed data         he installation place of the flowmeter should be selected to ensure a  minimum temperature of the transmitters surface  In direct sunlight the  surface temperature may rise by up to 30 degrees compared to the  ambient temperature  so if ins
21.  order to ensure the required protection grade  after the installation or  maintenance of the flowmeter  the following requirements should be fulfilled  see  Figure 2 12      WWW flow nt rt ru          EMIS MASS 200    e Seals in the transmitter should not have dirt or damage  If necessary   clean or replace the seals  It is recommended to use the original sealing elements  from the manufacturer     e Electrical cables should fit the cable entry and must not be damaged   e Covers and the cable entries must be tightened firmly   e Unused cable entries must be plugged     e Just before the          entry the          should have  U shaped loop to prevent any liquid from going into transmitter when the liquid  flows down the cable        Do not install the transmitter with the cable entry directed vertically upwards        Figure 2 12 Recommendations for orientation of cables and cable entries    2 4 5 Grounding Transient processes induced by lightning  welding  big power consuming  equipment or switches may lead to distortion of the flowmeter s measuring or  damage it  For protection from transient processes one should provide a  connection of the ground terminals located on the transmitter body  see Figure  2 13   with the ground through a wire  designed for operation under high  currents     For grounding  it is recommended to use copper wire of at least 2 5 mm      AWG13  section  Ground wires should be as short as possible and have a  resistance of less than 1 Ohm     Trans
22.  point corresponding  to zero flow  Without actual flow there is a certain time delay between signal coils  anyway  that delay assumed as zero point  To perform this procedure  follow these  Steps    e Apply power to the flowmeter and let it warm up for at least 30 minutes    e Let the measuring medium flow through the flowmeter  as long as the  thermal balance is established    e Close the valve located after the flowmeter    e Make sure that the flowmeter s tubes are completely filled with fluid    e Close the valve located before the flowmeter    e Make sure that the fluid in the flowmeter is completely still   Staying in the root menu  see Figure 2 14  press     button several times  to move to the Menu item  Configure   Press Ej button and enter the password   default password is 000000   To enter the password press     to increment the  currently selected digit  to move to the next digit press button  After successful  password entering a message    Zero calibration    will appear  Press      to start zero  calibration procedure  A message  Zeroing   will appear  Press  E to confirm that   Zero calibration procedure lasts for 30 seconds  the remaining time is displayed     Zero calibration should also be performed if the flowmeter indicates some flow  when there is no actual flow at the moment     The flowmeter has a protection switch block  see Figure 2 15   which protects the  flowmeter   s calibration coefficients from unauthorized access  The protection  switch bloc
23.  to the ON the time needed to perform  and    System position  changes in those sections   Para    sections    A special LED indicator is provided for indication in case of some faults of the  flowmeter  LED is located over the display and its color and lighting period  depends on the kind of the flowmeter s fault according to Table 2 9     Table 2 9     LED diagnostics       Lighting period    Lights continuously right after power up Self diagnostics test failed    Lights continuously some time later after    Wrong zero calibration  power up    Flashing  The color is red Malfunction of the flowmeter    Flow rate is less than the lower  limit of the flow range for this size  of the flowmeter    Flashing  Lights within 3 4 of the period   dark within 1 4 of the period    To determine a fault in the sensor part of the flowmeter it is necessary to check  the resistance of the coils first  Resistance value should be in the ranges specified  in Table 2 10     Table 2 10     Coils diagnostics    Circuit Contact Resistance range        Left coil Brown  red 60     75  Right coil Orange  yellow 3 4 60     75  Drive coil Blue  green 5 6 6   30           Gray  white 7 8 75     175  sensor   Temporaire Gray  black 7 9 75     175  sensor    Reclamation act appearance and procedure of returning the faulty flowmeter and  its warranty repair are descripted in the passport of the flowmeter     WWW flow nt rt ru              EMis mass 260    3 TRANSPORTATION AND STORAGE       3 1 Transportatio
24.  unused cable entry       Connect the ground cable to the ground terminal  6    e Tighten the cover of the transmitter           Figure 2 11   Electrical connection of the flowmeter    Table 2 5     Notes for Figure 2 11    Back cover of the transmitter  Signal cable   Cable entries   Terminal block   Blind plug   Ground terminal     Jo      fF               Power cable    Maximum length of the power cable is 300 m with minimum wire section 0 8  mm   AWG18      Connecting to the current and pulse outputs should be performed with a twisted  pair wire with the maximum length of 150 m and minimal wire section 0 5 mm    AWG20      Sensor and transmitter of the flowmeter of separate type are connected through  a special 9 core shielded cable with maximum length of 300 m  Connection  diagram is shown in Figure C 1 of Appendix C     After mounting and electrical connection the zero point adjustment should be  performed  see paragraph 2 5 4  Zero point adjustment       CAUTION     When using the flowmeter in hazardous areas  apply the requirements for  explosion protection provided in paragraph 2 4 2  Installation with explosion  protection        WWW flow nt rt ru               6 485 260    2 4 2 Installation with Installation of flowmeters in explosive environment must be performed in  explosion protection accordance with requirements of         This manual        Section 3 4 of  Operational Code for Electrical Installations         Section 7 3 of  Electrical Installations Code    
25. 1  A 1 3 180 192 350 290 260 108 95 11 14  025 A 1 1  A 1 3 200 248 450 400 290 188 115 15 18  040 A 1 2  A 1 4 520 547 470 660 280 182 150 30 33  050 A 1 2  A 1 4 558 588 550 750 290 198 165 35 38  080 A 1 2  A 1 4 780 808 710 1040 320 227 205 80 83  100 A 1 2  A 1 4 920 946 860 1290 350 290 390 185 188  150 A 1 2  A 1 4 1100     1050 1600 380 280 440 320 323  200 A 1 2  A 1 4 1380     1160 1740 420 335 535 625 628      Overall width of the body  excluding transmitter    Table A 2     Flowmeter flange dimensions                d1  D6   2  mm MPa mm mm mm  10         1 6  2 5  4 40 60 90 12 2     4 14   m 6 4 8 35 41 70 100 16 4 3 4 14  1 6  2 5  4 15     46 65 95 12 2     4 14  id 6 4 11 6 40 46 19 105 16 4 3 4 14  1 6  2 5  4 27 3     65 85 115 13 3     4 14     6 4 24 8 58 65 100 140 20 4 3 4 18  1 6  2 554 41 1 E 85 110 150 13 3   4 18    6 4 37 76 84 125 170 22 4 3 4 22  1 6  2 5  4 52 3     99 125 165 18 2     4 18  idi 6 4 47 88 99 135 180 22 4 3 4 22  1 6  2 5  4 79 5 E 132 160 200 20 2    8 18  1 6 4 77 121 132 170 215 24 4 3 8 22  1 662 554 101 7     156 190 235 21 3     8 22    l 6 4 94 150 156 200 250 25 5 4 5 3 5 8 26  1 6  2 5  4 154   211 250 300 26 2    8 26      6 4 142 204 211 280 345 SED 4 5 3 5 8 33  1 6  2 5  4 200    285 320 375 35 3     12 30   gt  6 4 198 260 284 345 415 37 5 4 5 3 5 12 36    WWW flow nt rt ru          Table A 3     Connection kit flange dimensions and weight    010    015    025    040    050    080    100    150    200    EMIS MASS 2
26. 502502       Kan WE ET  2042402    13  00000    ps                             gt    _                                                          m 099654                                                             12190 93105                                                              000120                                                         10624 25094                         xosqibuuusum  257100 CL  4  31998                                     10 05 27                                                               36   10 05 40                                                     ModBus   10 05 48                                                                                                                       10 05 48                                                 10 05 51                            c                                 36       Figure B 2                               Calibration  tab    The group of parameters                                          Flow calibration  contains 2 parameters                                                          r c Mkc   is the flow calibration coefficient     which means the  flow rate  in g s  for 1 us of time delay between left and right coils signals                                                          996 100 C   sets the temperature correction coefficient for flow rate  as percentage of calibration coefficient correction for temperature deviation of 100         The group of parameters                       
27. 513   11 00000      1  44513  1 00000   1 44513  1 90000                                              zd  E                                                4        xeu  1 44513                                                           4  Kd 1 00000  _                               e            5  D5         1144513                                                              5  K5 11 00000      10 41 03                                                                                                                       10 41 03                                                10 41 06                                                                     Figure B 7                                                Extended functions  tab    On the                 Connection  tab the following parameters are displayed     ModBus                  Address of the flowmeter in Modbus network     ModBus                                   Data transfer baud rate in Modbus network                                               ID                  ID hex  500501502502                10 03 27                                                               36  10 05 40                                                     ModBus  10 05 48                                                                                                                       10 05 48                                                10 05 51                            c                                 35       Figure B 8   
28. 60    USER MANUAL    Figure A 2 Outline dimensions of connection kit flanges    1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4  1 6  2 5  4  6 4       10    15  11 6  27 3  24 8  41 1   87  52 3   47  79 5   7    101 7   94  154  142  200  198    40  34  46  39  65  57  85  75  99  87  132  120  156  149  211  203  285  259    60  70  65  75  85  100  110  125  125  135  160  170  190  200  250  280  320  345    90  100  95  105  115  140  150  170  165  180  200  215  235  250  300  345  375  415    12  16  12  16  13  20  13  22  18  22  20  24  21  25 5  26  31 5  35  3  5    h   mm        gt  N A  A N A    A Cc A N A N    4 5            Oo         o A A A A A A A Aa A  gt                 4 5        _                4 5    WWW flow nt rt ru          10 7  13 2  25 4  24 0  38 5           EWis wass 260    APPENDIX B  Working with  EMIS Integrator  software      EMIS Integrator   software is designed to query and configure flowmeters through digital interface  with a personal computer  hereinafter     the PC         To install   EMIS Integrator   it is necessary to run   EMISSoftware X X X exe   file  where X X X    means the version of the software  When installation started follow the instructions of the setup wizard     On the first start of the program it will automatically start searching the flowmeters connected to the  PC  In the log area a message about searching start wil
29. C and relative humidity up to 9596  non condens ing at 25 C         Flowmeters can be stored in a transport boxes stacking of up to 3 boxes in  height and also without package  For the long term storage it is recommended  to use the manufacturer s package     3 3 Utilization Flowmeters are free of harmful substances and components that are  hazardous to human health and the environment during and after the life of the  flowmeter and after utilization     Utilization of the flowmeter is performed separately for groups of materials   plastic parts  metal parts and fastening parts     4 VERIFICATION    Verification of the flowmeters is performed in accordance with the document   Instructions GSI  Coriolis Mass Flowmeter  EMIS MASS 260   Verification  procedure      WWW flow nt rt ru              EMis mass 260    APPENDIX A  Outline dimensions and weight                                                                                                                                                                                                                                                                                                    A  1 3 A  1 4  Figure A 1 Outline dimensions    Integral type     Figure A 1 1   A 1 2  Separate type     Figure A 1 3  A 1 4    WWW flow nt rt ru              EWis wass 260    Table   1     Outline dimensions and weight          mm 1 6 4 0 6 4 mm A 1 1  A 1  3   MPa MPa A 1 2 A 1 4    010 A 1 1  A 1 3 150 170 350 290 260 130 90 11 14  015 A 1 
30. Frequency of the signal on the pulse output   CUR  XX XX mA   Current value on the current output     Water X XXX g mL  73     6    Oil X XXX g mL  X X XXXX XXXX XXXX                                  MMM  d    N  c    Density of water in oil water under working conditions              Density of oil      oil water under working conditions     73 XXXX XXXX XXXX XXXX 174  56 Internal parameters for pressure compensation mode   EMIS EN kg CX XX 175  6 Menu language  EN   version of the display menu  X XX    74 9600  8      1 tY YY   Modbus data format  register map version              XXXXXX 1 168     6  XXXXXX 0   low density cutoff  g mL   bytes order for float point values  default value is 0      System Para  1  1  System parameters of the flowmeter     Press   E   to enter E76 Intended to use for the representatives of the manufacturer only     Display low volume flow cutoff  L h   totalizer mode  0     disabled  1     enabled      N                    77  1  Correction coefficient of the temperature Tk for simulation mode  in the range of 0 9     1 1    76                E     edit T  Tk xTdev   TO   Tdev is temperature deviation    77 Temperature Zero X XX    i  2 Zero temperature TO for simulation mode  in the range of  20C to  20         79      Correction coefficient of the density      for simulation mode  in the range of 0 9     1 1    78 DensityCal     E     edit p  pkx                            is density deviation    79 DensityZero X XX    d Zero density 00 for 
31. N position     16 bit registers     integer values  2 bytes  high byte first  Functions  Read     04  Write     06    0003 Register map version       1     EMIS  default   0016    2     ProLink  0039    Transmitter software version   Format XXX X      RO   sss  RW Mass flow rate unit 33  70     g s  73     kg s  74     kg min  77               75     kg h  default   78     t h    WWW flow nt rt ru          1 EMIS MASS 260 USER MANUAL    Ades   Mode                          36    0040 HW Density unit     91     g cm   default   96     kg L  92     kg m   RW    0041 Temperature unit  32       C  default   33     0042 RW Volumetric flow rate unit  0  mL s  24     L s  17     L min  29     m    day  138     L h  default   19     m  h  Pressure unit  0    MPa  7     Bar  default   12   kPa  Mass total unit  60          61     kg  default   62  t  Volume total unit  0    mL  41     L  default   43  m   Device type code  40     Core Processor  21 41 42     RFT9739 1700A 2700A    RW 28  Bytes sequence in floating point 32 bit registers  0     0 1 2 3  default   1     2 3 0 1  2     1 0 3 2  3     3 2 1 0  Output signals  0     None  1     Current   Pulse   RS485  default   2     Fieldbus  H1  or Profibus PA          CO  N    CO          0044    D       0045    CO         D       0046             gt        0120    0125  0126  0136  0313  0419  0420  0421  0422  0423  0424    0521 RW    N          1138    1166    Output channel A type  0     Current  primary   default   1     Pulse  2 
32. Upper limit of density range  g cm      0201 0203 RW Lower limit of density range  g cm       0231 0232        Zero point standard deviation  us         0233 0234 Zero point  us    0247 0248   RO   Mass flow rate      0249 0250 RO Densit    0251 0252  RO   Temperature                              1 o a  0253 0254   Ro Volumetric flow rate          0257 0258         Calculated pressure from external pressure sensor  kgf cm          WWW flow nt rt ru    4  62  65  63  9  0  17  75  8  9  16  1  3  3  2  3                 EWis wass 260     Ades   mos                        izer  i         0259 0260  0261 0262  0263 0264  0265 0266  0267 0268  0269 0270  0271 0272    Volume totalizer  inventory  Coefficient of pressure compensation for flow      PSI     XXX    Coefficient of pressure compensation for density      PSI     Flow calibration pressure  kgf cm    Pressure relating to 4 mA                    0273 0274    XXX    0275 0276 Pressure relating to 20 mA  kgf cm        mo    mo    mo    RO  NENNEN  0285 0286   RO   Measuring tubes oscillation frequency  Hz        0287 0288   RO  Leftcoilvotage    oo     0289 0290 RO  Rigmcolvdagem  fw  0291 0292   RO   Drive co o _    RO               MEN  NEN         Drive coil load  9  0293 0294 Live mass flow rate  zero point not counted   kg h  0451 0452    Input external pressure             ASCII registers    0072  Calibration coefficient  g s us            Format  XXXXXX Example     23 350      K 23 35 g s us  0075  0076 Temperature cor
33. X m  Max  length is 300 m   L Liquid  X Special order           Sensor material           Stainless steel    X opecial order    WWW flow nt rt ru                   EMIS MASS 260             1 6 Maximum pressure     1 6 MPa  2 5 Maximum pressure     2 5 MPa  4 0 Maximum pressure     4 0 MPa  6 4 Maximum pressure     6 4 MPa  X Special order       Medium temperature  100 Medium temperature  50    100   C  200 Medium temperature  50    200   C  separate type only   350 Medium temperature  50    350   C  separate type only   X opecial order                      _  24 24V DC  for DN lt  100 mm   220 220V AC  X Special order    Output signals    Pulse   digital RS 485 outputs    A Additional current output 4 20 mA  X Special order   0 15 Accuracy class 0 15   0 25 Accuracy class 0 25   0 5 Accuracy class 0 5  X opecial order           Display with basic functionality    X Special order       Connection to the pipeline        Flange type  X opecial order        Factory calibration  V Official verification    WWW flow nt rt ru               6 55 260    2 OPERATION AND MAINTENANCE       2 1 Model selection recommendations    One of the most important conditions for reliable operation of the flowmeter and to  obtain reliable measurement results is correspondence of the flowmeter s  modification to the process parameters  The list of the process parameters  required for optimal flowmeter modification selection is presented in Table 2 1     Table 2 1     Process parameters for modificati
34. ange of pressure is   0 015 kg m  for every 100 kPa of deviation from the density calibration pressure     WWW flow nt rt ru              EMis mass 260    Relative basic error of measurement of volumetric flow  volume  on pulse and digital  output signals  dy  calculated as    dy      do    Ap   p   100     Z   Qy   100     1  2    where          accuracy class  96    Ap     absolute basic error of measurement of medium density  kg m             measured medium density  kg m       Z     zero stability  according to Table 1 2   L h             measured volumetric flow rate  L h   Note     For the accuracy flow range  corresponding to a given accuracy class   according to Table 1 2   the value of Z is assumed to be O   Relative basic error of measurement of mass flow  mass  on current output signal           calculated as               0      0 2 154     4 16        Qumax     1 3   where          Relative basic error of measurement of mass flow  mass   96             20 mA     maximum value of current output signal            measured mass flow rate  kg h                  upper limit of the full mass flow range  kg h        Relative basic error of measurement of volumetric flow  volume  on current output  signal  dv  calculated as    dv     II6VI   0 2    Ias7  4 1 6 Qy   Qvmax     1 4     where         Relative basic error of measurement of volumetric flow  volume   96   Imax   20 mA     maximum value of current output signal   Qy     measured volumetric flow rate  L h        
35. be performed by the   EMIS Integrator   or  changing the value of the corresponding Modbus register  see Appendix D  using  third party serial port tools     1 3 6 Display The flowmeters OLED graphic display contains 4 lines with 16 symbols per line   The following measured parameters can be displayed     Mass flow rate   Volumetric flow rate   Medium density   Medium temperature   Total mass    Total volume     Display operation is provided by the means of three optical sensor buttons below  the display  Display operation described in the paragraph 2 5 3 Display operating      1 4 Pressure drop    The pressure drop on the flowmeter  AP  at maximum flow rate  pressure and  temperature is not more than 0 13 MPa for water medium     In the process of measurement flow of liquids it is necessary to consider the  cavitation effect  liquid boiling   which may occur in certain conditions of the flow   Cavitation may cause the flowmeter work abnormally  To prevent this  it is  necessary to keep the certain pressure at the distance of 5 DN after the flowmeter   That pressure must be not less than critical pressure  Pcr  calculated as    Pcr   2 9 AP   1 3 py  kPa  1 6     where AP     pressure drop on the flowmeter  kPa     pv     saturated steam pressure at working conditions  background  information   kPa     If the pressure calculated by this formula exceeds the actual pressure in the  pipeline  a safety valve should be installed to increase the pressure     WWW flow nt rt ru      
36. d digital outputs of the  flowmeters of  Ex  modification must be implemented in accordance with GOST R 51330 1     Explosion protection type of  flameproof enclosure  is implemented by putting the electrical  parts of the flowmeter into the flameproof enclosure in accordance with GOST    51330 1   which prevents the explosion from coming out of the flowmeter into the explosive environment   Explosion protection of the enclosure is ensured by the following means         the shell withstands the explosion test at the test pressure of 4 times the pressure of  the explosion         axial thread length and number of full turns of thread engagement comply with  GOST 8 51330 1 requirements         the gaps and lengths of flat and cylindrical flameproof joints comply with  GOST    51330 1 requirements         maximum flowmeter s surface temperature in working conditions must not exceed the  temperature range in accordance to GOST    51330 0 for the temperature groups       T4 for the flowmeters with temperature modification code   100       T3 for the flowmeters with temperature modification code   200       T1 for the flowmeters with temperature modification code   350       Explosion protection type of input and output  intrinsically safe  circuit level   ib   is  ensured by the following means         external power supply and connection of the external devices to the pulse  current and  digital outputs of the flowmeters of  Ex  modification must be implemented in accordance with
37. detectors  When there is no flow  the tubes don t  twist and there is no time delay between left and right detectors signals  Mass flow  rate Q is proportional to the time delay At between the detectors signals     Q   K  At 3 6   kg h  1 0     where        calibration coefficient  g s us   At     time delay between the detectors signals  us     The density of the medium is determined by measuring the period of oscillation of  the flow tubes  which is proportional to the density of the medium  The density  measuring channel is calibrated for two different kind of medium with exactly  known  measured by standard density meter  density  water and air   For each  medium the flowmeters shows the flow tubes oscillation period corresponding to  the medium density  The measured value of the oscillation period is shown in  Menu item 70  see the display menu   The medium density and the corresponding  oscillation period for both water and air are entered in the menu display  Menu  items 62     65  or through Modbus  Due to the linear dependency of the oscillation  period of the density the actual medium density can be determined by the  oscillation period measured by the flowmeter     Temperature measurement is performed using the built in platinum temperature  sensor Pt100  The measured temperature is used for automatic correction of the  flow and density when the medium temperature changes  Temperature correction  factors of flow and density are entered in the flowmeter memory at th
38. dification must be between  20 and  40          WWW flow nt rt ru              EMis mass 260    1 6 Name plates       Flowmeter s parameters are marked on the name plates attached to the top  side of the transmitter and to the body of the sensor  Flowmeter has the  following name plates     1  Main name plate with technical parameters  2  Sensor name plate with sensor explosion proof information     The main name plate is produced according to Figure 1 4 and contains data  presented in Table 1 11     CORIOLIS MASS FLOWMETER  15         EMIS MASS 260       4  Diameter Pmax MPa  14 Flow range      Accuracy LL           3   5      Twork Pulse weight    13 Date     Powervotage    V  6  15 Seil            tExd iblliICTeX IP65 7  A  O   EMIS    CJSC  RUSSIA www emis kip ru  11 10 9 8    Figure 1 4     Main name plate    Table 1 11     Main name plate data    1 Trademark of the manufacturer   2 Flowmeter name   3 Measuring approval sign   4 Maximum allowable pressure  Pmax    5                        5   6 Pulse weight   7 Power supply voltage   8 IP protection grade   9 Explosion protection grade and Ex sign  10 Information about manufacturer   11 Serial number of the flowmeter   12 Date of manufacturing   13 Temperature range of the measured medium  Twork   14 Full flow range  Q    15 Nominal diameter  DN     www flow nt rt ru          WWW flow nt rt ru    EMIS MASS 200       Sensor name plate is produced according to Figure 1 5     Q CORIOLIS MASS FLOWMETER Q       EMIS MASS 260 
39. e address   253 byte data   2byte CRC   256 Bytes    Frame structure   Function 01  Read Coil Status    Query   Function Starting address Coils quantity    Slave address  01H OOH 08H OOH 08H    Function Byte count Coil status    Exception code     01H   Function is not supported     2     Incorrect Starting address or Coils quantity  03H   Coils quantity is out of range of 1     200  04H   Failed to read coil       CR    Hesponse     Slave address CR    Error   Slave address    Function 04  Read Input Registers    Function Starting address Registers quantity             Hesponse     Slave address Function Byte count Data value          Error     Exception code     01H   Function is not supported   02H   Incorrect Starting address or Registers quantity           Registers quantity is out of range of 0     125  04H   Failed to read registers    Query     Slave address CR    CRC    WWW flow nt rt ru          EMIS MASS 260    Function 05  Write Single Coil       Query     Slave address Function Coil address Data value  0H     Hesponse     Slave address Function Coil address Data value  oH         o    Exception code     01H   Function is not supported   02H   Incorrect Coil address   03H   Data value is out of range of 0x0000     OxFFOO  04H   Failed to write single coil     1    Error   Slave address    i    Function 06  Write Single Register    Query    Lo                       Hesponse                           _    Error     Exception code     01H   Function is not supported 
40. e factory and  can be changed through the menu display or Modbus     Flowmeters have the manual correction of flow when the medium pressure  changes  When manual correction mode is enabled in the menu  the user should  specify the actual pressure value taken from the external pressure sensor  see  Menu items 39     41      The sensor generates the primary electrical signal containing information about  the time delay between the signals of the detectors  The primary signal is  transmitted to the electronic transmitter placed directly on the sensor or separate  from it  The transmitter with digital signal processor processes the primary signal   calculates mass and volume flow values  temperature correction  generates output  signals  and displays the information on the flowmeter s display     Features of this measurement principle    e direct measurement of mass flow rate in the pipeline without additional  errors        maintaining high accuracy and stability over long time   ability to measure the flow of high viscosity liquids  non Newtonian fluids   fluids containing solid or gas inclusions    e reliable operation under conditions of vibration and misalignment of the  pipeline  of the medium temperature and pressure changing        no moving parts inside the flowmeter  oscillation amplitude of the tubes is  very small and they can be considered fixed  and parts subject to wear   which extends the life time of the flowmeter   no obstructions to the flow inside the flowmete
41. e first  Functions  Read     04  Write     06    Description Menu item      Menu language     1     English  2     Russian  default  Register map version     1     EMIS  default   2     ProLink  Current output  0     Mass flow rate  default   1     Volumetric flow rate    Pulse output  0     Mass flow rate  default   1     Volumetric flow rate  2     Density       Modbus Baud rate  0     9600 bit s  default   1     4800 bit s  2     2400 bit s  3     1200 bit s    www flow nt rt ru      EMIS MASS 260 USER MANUAL       0015 67  006                      gt   gt      CO CO CO CO                 O Co A     O1 N O  CO    Flow direction    0016  RW  0     Single direction  default     1     Bidirectional  0018 RW Mass flow rate unit  0     g s  1     kg s       cm  3     t day  4     kg h  default   5     t h  0019 HW Density unit  0   g cm   default   1     kg L  2    kg m   Temperature unit  0      C  default   0     0021 HW Volumetric flow rate unit  0     mL s  1     L s  2     L min         m  day  4     L h  5     m h  default   0022 RW Mass total unit  0 9  1   kg  2     t  default   0023 HW Volume total unit  0   mL  1 L  2         default            0257         RO  Calculated pressure from external pressure sensor             Pressure relating to 4 mA  Bar       Pressure relating to 20 mA  Bar       0521 RW Bytes sequence in floating point 32 bit registers  0     0 1 2 3  default   1     2 3 0 1  mE      oi  3     3 2 1 0         For Menu items refer to display menu struc
42. ecautions    Installation  operation  maintenance of flowmeters should be performed by  persons studied this manual and safety instructions for working with electrical  devices        All operations on calibration and usage of flowmeters must comply with the  requirements for protection against static electricity     Installation of flowmeter in the pipeline and its removal from the pipeline should be  performed without pressure in the pipeline and with the power supply switched off   Electrical connection should also be performed only when the power supply is  switched off   During the installation  commissioning and maintenance shall be prohibited    e replacement of electronic components when the flowmeter is powered on     e connecting the flowmeter to the power supply with output voltage other than  specified in this manual     e using electrical devices and tools without protective grounding and also in  case of their malfunctions   During installation the hazardous factors are     e power supply voltage of 220V AC or higher  in case of the external power  supply located in close vicinity of the installation place      e excess pressure in the pipeline      high temperature of the medium     Operation of flowmeters of            modification must be performed in accordance  with the requirements of Chapter 7 3 of the  Electrical Installations Code  and  other regulations of using of electrical equipment in explosive environment     CAUTION     Installation and operation
43. flow nt rt ru            LMI5 M 59260  Table of contents    1 PRODUCT  DESCRIPTION AND  TECHNICAL DATA    2 OPERATION AND  MAINTENANCE    3 TRANSPORTATION  AND STORAGE    4 VERIFICATION          USER MANUAL    1 1 Applications   1 2 Principle of operation   1 3 Technical specifications  1 3 1 Technical parameters overview  1 3 2 Range of measurement  1 3 3 Accuracy of measurement  1 3 4 Power supply   1 3 5 Output signals   1 3 5 1 Pulse output   1 3 5 2 Current output   1 3 5 3 Digital output   1 3 6 Display   1 4 Pressure drop   1 5 Explosion protection   1 6 Name plates   1 7 Delivery set   1 8 Model codes    2 1 Model selection recommendations  2 2 Safety precautions   2 3 Installation on pipeline   2 3 1 Installation location   2 3 2 Orientation   2 3 3 Pipeline preparation   2 3 4 Installation   2 3 5 Thermal insulation   2 3 6 Cooling   2 3 7 Transmitter rotation   2 4 Wiring   2 4 1 Basic operations   2 4 2 Installation with explosion protection  2 4 3 Wiring recommendations  2 4 4 Waterproofing   2 4 5 Grounding   2 5 Operation and maintenance  2 5 1 Basic recommendations  2 5 2 Power Up   2 5 3 Display operating   2 5 4 Zero point adjustment  2 5 5 Protection switch   2 5 6 Maintenance   2 5 7 Troubleshooting    3 1 Transportation  3 2 Storage  3 9 Utilization    Appendix A     Outline dimensions and weight   Appendix B     Working with  EMIS Integrator  software  Appendix C     Wiring diagrams   Appendix D     Modbus register map version 2 xx  EMIS   Appendix E    
44. ible to change the units of measured  parameters which are displayed in the menu items 1  2  3     ERG                               x   35 hoxsc oo dU RO pe                                                         10 05 27                                                               36  10 05 40                                                     ModBus      10 05 48                                                                                                                       10 05 48                                                  10 05 51                                                               3    st       Figure B 5                                            Units  tab    WWW flow nt rt ru              685 260    On the                                        Working parameters  tab it is possible to set the following  parameters                                                                                                                                                                                                                                                                        ID                ID hex  510515025       mcewxa                             190 00000                                                   VOPHZHOSK   HVI                                                               CHZHOTOS                                                                                                    10 05 27                                              
45. if set to On        L1    Den XXX XXX      Temp XX XX Density and temperature of the measured medium     Configuration 15  1   Press   E   to enter E86             Press   E   button to enter the submenus for configuring the flowmeter                iN          Submenus of the    Configuration    menu    The following submenus can be accessed only after entering the 6 digit password  If the entered  E8 password is invalid a message  Wrong password  will be displayed  To return to the Menu item 4     Configuration    press          To enter the password again press                    Password  000000                     ESTE uns  op poe NM On QUI Mmi       18 Reset Totalizer 119  58 Press   E   to reset mass and volume totalizers  resettable totalizer    Press   E   to enter E   reset    Option of direct    SIG    or reverse    BID    direction of flow  If   SIG   is chosen then only direct flow  will be accumulated  direction corresponding to the arrow on the sensor body   If   BID   is chosen               16  17          11 51      MMM        SIG  BID        Flow Direction         then flow will be counted on both direct and reverse directions            EMIS MASS 260                       MMM          Menu           Button    it  m Indication format E  The way of counting reverse flow  whether it will be added    INC    or subtracted    DEC    from the    Reverse Flow Tol DEC 121  28   INC E     edit accumulated total mass and volume            X XXX   422 8     8 Accumula
46. ions  Select   Y   to accept  changes or   N   to cancel  then press   E   to quit edit mode     e  fthe menu item has numeric value its editing is performed by changing the    digits one by one and moving to the next digit on the right  The digit in currently  selected position is flashing  Press       to increment the digit in current position from  0 to 9  To move to the next digit press          Press   E   to finish editing  then select     Yes   to accept changes or   No   to cancel  then press   E   to quit edit mode     WWW flow nt rt ru           EMIS MASS 260    SILLIO MMM    DI                    Root menu  86    I  Mass xxx xxx 1 Password    Flow               t h 000000     E      16  2  4            21     16   8    Reset Totalizer Flow Direction Reverse Flow Tol Mass Inventory t VolmInventory m3  xis 58    Press E to enter SIG DEC X XXX X XXX  30 31     E  20mA Value Current Output MODBUS Address Baud Rate Parity Bit Stop Bits  XXX X Mass XX None One    37    Volume Unit VolumeFlow Unit Temperature Unit  m3 m3 h                     Flow Config Zero Calibration  Volm xxx xxx m3 Press E to enter X XXX  Flow xxx xxx m3 h  9 2  2a Pulse Weight g N                1 X XX  IWater               1011  xxx xxx t            Den   xxx xxx g mL  Temp xx xx   C    439      4 E FlowPressCompCoe      X XX  Press E to enter 48  Output AUTO Sim FlowSimulate t h Flow Sim Start 1  Press E to enter X XX Press E to enter    13  Addons Functions  Press E to enter    OW Analyse Switch
47. k is located at the front side of display board to the right from the  display  The protection switch block contains 8 switches     To access the protection switch block it is necessary to remove the sealing wire or  the sticker from the front cover of the transmitter  then remove the front cover        Figure 2 15 Protection switch block in default positions    At factory switches 2  4  6  8 are set to ON  see Figure 2 15   In this position  the  calibration parameters of the flowmeter cannot be changed  entering the menu  section 5  Calibration  and section 7  System Para  is not available  Such a state  of the switches is necessary to change of some Modbus registers through digital  interface  such as the Modbus register map version or display language  see  Appendices D and E      If any of switches 1 8 is set to ON  the entrance to the  Calibration  and  System  Para  sections will also be impossible     To enter the  Calibration  and  System Para  sections and change or just see the  calibration coefficients it is necessary to set all the switches to OFF positions  see  Figure 2 16      1             L  4      Figure 2 16 Protection switch block in OFF positions           Enis MAss 260    2 5 6 Maintenance    2 5 7 Troubleshooting          USER MANUAL    The flowmeter put into operation does not require special maintenance other than  periodic inspection to verify the following     e Compliance with the working conditions     e Whether the voltage and the other paramete
48. l appear  In the    Device list    the addresses and  descriptions of all found devices will be displayed  If no devices were found  it is recommended to check the  connection of the flowmeter to the PC  power supply and also check the program s settings  To change the  settings press button                         Settings  on the toolbar  In the appeared window specify the number of  COM port used and the baud rate the same as in flowmeter s settings   Default value is 9600 bit s   then click    OK       To work with one of the found flowmeters select it in the device list  Double click on its address and in  the main area of the program the following pages  tabs  will appear                                                                                                                                                                                                                                                                                    The active tab is                             Initially the active tab is                           Monitoring   it contains the following parameters  instant  mass flow rate                                                                    total mass                                       kr     instant  volumetric flow rate                                                                  total volume                                       II        density                                   and temperature                           
49. load  the values of signals on current and pulse outputs                                     Modbus    Modbus register map  switches the flowmeter to   ProLink   or    EMIS   Modbus register map     CAUTION     After switching to   ProLink   register map   EMIS Integrator   software will stop reading and  displaying parameters of the flowmeter  But                                 Modbus   field will still be  available to change     If the flowmeter was configured to use   ProLink   register map  then on the start of the program  a confirmation window will appear with the request to switch to the    EMIS    version of Modbus  registers for   EMIS Integrator   to work normally  Press   OK   to switch to   EMIS   version  or    Cancel   to keep the   ProLink   version of Modbus registers        WWW flow nt rt ru                           ID                      Ht 1 j   a dos4    i           L   Eu Moli                 gt  pun  113 50392  command 10 8490     4 00000 _  10 03701                                         IV       pezucmpose ModBus    10 05 27                                                               36  10 05 40                                                                   10 05 48                                                                                                                            Figure B 4                                                    Hardware monitoring  tab    On the                                        Units  tab it is poss
50. mitter can be grounded through the pipeline  if the pipeline provides the  ground        Figure 2 13 Grounding    CAUTION     There should be no potential applied or induced on the ground wire     Do not use the same wire for grounding of two or more devices     WWW flow nt rt ru               6 485 260    2 5 Operation and maintenance    2 5 1 Basic  recommendations    2 5 2 Power Up    2 5 3 Display operating          To ensure reliable operation of the flowmeter and maintain accuracy it is necessary  to satisfy the following requirements     e  o prevent damage to the measuring tubes of the flowmeter because of  hydroblow  opening   closing valves on the inlet pipe should be done gradually         Operation at flow rates close to the upper limit of the full flow should last no  more than 2 hours per day     Right after power up the flowmeter performs a self test  and if it succeeds  the  flowmeter begins to measure flow  to generate output signals and display the  measured values     Controlling of the flowmeter   s display is performed via three buttons of optical type   below the display  Thus  display operating doesn   t require opening the front cover  of the transmitter  which is important in explosive environment  in conditions of high  humidity or precipitations  and other conditions with a high probability of  contamination of the internal structural elements of transmitter or ingress of  moisture  liquids  foreign objects  etc     To  push  the optical button one 
51. mly and put the latch lock on the covers        2 4 3 Wiring During electrical connection of the flowmeter you should follow these  recommendations recommendations     e Wire cores must be cleaned and fixed on terminals to prevent a short  circuit between each other and the body of the flowmeter  ground      e  t is recommended to use separate power supplies or a multichannel  power supply with isolated channels to power the flowmeter and each of its  outputs     e  f it is necessary to calculate the load resistor value  the total load  impedance should be calculated as the sum of the cable resistance  external  load resistance  resistance of safety barriers  load resistance of the secondary  equipment     e  o minimize interference when transmitting of analog output signal 4 20  mA and digital signal it is recommended to use a shielded twisted pair cable   Grounding of the cable should be provided only on one side of the cable  near  the power supply      e  t is not recommended to lay the signal cables in the same conduit with  power wires  and also near sources of strong electromagnetic fields  If  necessary  grounding of signal wiring can be done at any point in the signal  circuit  For example  it is possible to ground the negative terminal of the power  supply  The electronics housing is grounded to the sensor housing     2 4 4 Waterproofing The flowmeter meets all the requirements for moisture protection grade specified  in the  Technical Specifications  section     In
52. n    While transportation of the flowmeter it is recommended to follow these  recommendations     e The flowmeter should be transported in a transport container  which  should protect the flowmeter from mechanical damage    e It is recommended to lay a waterproof paper inside the transport  container    e Transportation should be performed at ambient temperatures from  40  to  70     with relative humidity up to 100  non co ndensing at 35       e Protection against atmospheric precipitation must be provided    e Flowmeters can be transported by all kinds of roofed transport   including air transport in heated hermetic compartments in  accordance with the rules in force for this type of transport    e The requirements in the package handling marks should be fulfilled    e Method of stacking the boxes on the transport vehicle must prevent  their movement    e The boxes must not be the object to sharp blows while loading    unloading    e Period of stay      the transportation conditions must be not more than     months    e After transportation at a temperature below 0    container with the  flowmeter should be unpacked at least 12 hours after its storage in a  warm room     When the flowmeter is transported out of package it is necessary to follow  recommendations given in Figure 3 1        Figure 3 1 Out of package transportation rules    WWW flow nt rt ru              EWis wass 260    3 2 Storage Flowmeters can be stored in unheated rooms with air temperature of  20 to   70 
53. nt mode    Query   Slave address    41H    Function  01H    Hesponse   olave address    41H    Function  01H  Error   Slave address  41H    Function  81H    Starting address    OOH 08H    Byte count    01H    Error code  01H       USER MANUAL    Coils quantity CR    OOH 08H    CEH    UJ       g    Coil status  OOH    JJ         48H    4   gt         CRC  80H 44H    Result  Reverse flow is subtracted from the totalizer values    Reset totalizer  resettable     Query   Slave address Function    05H    JJ             I    Hesponse   Slave address Function    05H    um  I    Error   Slave address    41H 85H    Function    Perform zero calibration    Query   Slave address    41H 05H    Function    Response   Slave address    41H 05H    Function    Error   Slave address    41H 85H    Function       Coil address    00H 02H    Coil address    00H 02H    01H    Error code    Coil address    00H 04H    Coil address    00H 04H    Error code    01H    Data value  FFH OOH         CO  2                     00H 3AH  CRC  82H 84H    Data value CHR  FFH OOH C3 3BH         C       Data value CRC    CRC  82H 84H    WWW flow nt rt ru       EMIS MASS 260    Read mass flow rate       Query     Slave address Function Starting address Registers quantity CR    04 OOH A7H OOH 02H CEH E8H    2  I       Response     Slave address Function Byte count CRC    04H 04H 43H B4H 74H DOH         I    Error     Result  mass flow rate is 360 9126 kg s    Write mass unit of kg  Quer    Slave address Register addre
54. oid protrusion of the gasket into the pipeline opening     e Install the gasket between the other pair of flanges  slide the bolts through  the holes of the flanges  put washers and nuts  Nuts should not be finally tightened  yet     e Tighten the nuts in the sequence shown in Figure 2 6     When installing the flowmeter bending and torsional load on the joints  and also  mated flanges misalignment should be minimized     TW          Figure 2 5 Installation of the flowmeter in the pipeline    WWW flow nt rt ru              EMis mass 260    Table 2 3     Notes for Figure 2 5    Mounting part    1 Flowmeter flanges          Connection kit flanges    2   3 Gaskets  4 Nuts   5 Washers  6    Bolts  or stud bolts        Figure 2 6   The sequence of tightening the flange bolts    Transmitter of the separate type of the flowmeter can be mounted as shown in  Figure 2 7  Transmitter can be mounted with brackets or clamps to the rack  pipe or  wall        Figure 2 7     Installation of transmitter of separate type modification    WWW flow nt rt ru            EMIS MASS 260 USER MANUAL  Table 2 4     Notes for Figure 2 7       1 Transmitter base  2 Bolts   3 Clamp   4 Nuts    CAUTION        Do not install the transmitter with the cable entry directed vertically upwards     2 3 5 Thermal If thermal insulation of the pipeline and the flowmeter is necessary  see the  insulation recommendations in Figure 2 8        Figure 2 8   Recommendation for thermal insulation of the flowmeter    WWW flo
55. on selection    Item  Process parameter       1 Measured medium name   2 Composition and percentage of liquid components   3 Composition and percentage of solid impurities in fluid  4 Composition and percentage of gas inclusions in fluid  5 Measured medium density   6 Measured medium viscosity   7 Flow range   8 Required accuracy   9 Measured medium temperature   10 Pipeline pressure   11 Allowable pressure drop   12 Presence of regulation and control components in the system  13 Pipeline diameter   14 Pipeline orientation at the installation place   15 Ambient temperature   16 Explosion protection grade and requirements    CAUTION     To avoid erroneous self dependent selection of the modification of the flowmeter  please send the completed questionnaire to the nearest EMIS company  representative        Flowmeter size should be selected according to the actual flow rates in the  pipeline  which may differ from the calculated  design  values  Flowmeter size  should be chosen so that the actual flow rate of the medium was in the second  third of the flow range  Therefore  nominal diameter  DN  of the flowmeter can be  either equal or less than the nominal diameter of the pipeline     When mismatch pipeline diameter and the nominal diameter of the flowmeter  tapered transitions can be used  They can be made independently  at that to  ensure minimum loss of pressure  the central cone angle must not exceed 30  C     WWW flow nt rt ru                EMIS MASS 260    2 2 Safety pr
56. own in Tables 1 8 and 1 9     Table 1 8     Input parameters of intrinsically safe circuits    Parameter value       Parameter name  Left   right   Temperature                     signal coils Sensor    Maximum input voltage Ui  V 5 4 10 5 5 4  Maximum input current li  mA 72 70 72  Maximum input capacity Ci  pF 50 50 50  Maximum input inductance Li  mH 2 2 3 5 0 010  Maximum input power Pi  W 0 097 0 184 0 097    WWW flow nt rt ru                EMIS MASS 260       Table 1 9   Output parameters of intrinsically safe circuits       Parameter value for the circuit of    Parameter nam   i  ape Left   right coil Drive coil Temperature  power power sensor power       Maximum output voltage Uo  V 5 4 10 5 5 4  Maximum output current lo  mA 72 70 72  Maximum output capacity Co  uF 10 1 10  Maximum output inductance Lo  mH 5 4 5 5  Maximum output power Po  W 0 097 0 184 0 097    Parameters of the sensor coil windings are shown in Table 1 10     Table 1 10     Coil windings parameters    Left   right signal coils 0 13 20   0 5  Drive coil  DN10     DN40  0 13 300 11   0 5  Drive coil  DN50     DN200  0 27 300 8   0 5    Maximum length of the connection cable for the separate type flowmeter is 300 m     Drive coil power circuits are electrically isolated from other circuits by means of the undamaged  transformer according to GOST R 51330 10  Insulation between primary and secondary  windings can withstand voltage of at least 1 5 kV     Ambient temperature for the flowmeters of  Ex  mo
57. played in the menu items 1  2  3   Mass Unit g  kg  t    Select units for mass totalizer     Description       MassFlow Unit g s  kg s  kg min  134 19    select units for mass flow rate   t min  t h E     edit  4 Volume Unit mL  L  m  y uno Select units for volume totalizer         4   mL s  L s  L min   36      gt 10     5 VolumeFlow Unit m  min  L h  m  h Select units for volumetric flow rate   6  1  9    3   3   5    3 2     gt         P   5    4   4   1   4    Pressure compensation modes  Select   None   to disable pressure compensation  select   Manu    for manual compensation  when the actual pressure value is entered manually in the Menu item 41   PressComp Mode                      y     Auto select   Auto   for automatic pressure compensation  when actual pressure is measured by an   external pressure sensor  connected to the flowmeter s corresponding terminals     FlowPress X  XX 441   gt 11  Coefficient of pressure compensation for flow as percentage of calibration coefficient correction for  0                  E     edit pressure deviation of 1MPa  Default value is 0 01            139  11     1 InputPress MPa XX  XX E     edit Actual pressure value for manual compensation mode   Other 113  4                  Miscellaneous functions such as flow simulation  etc     Press   E   to start automatic flow simulation procedure  The sequence of frequency and current test  Output AUTO Sim signals appears on the flowmeter s outputs  Signal on the current output changes from
58. r    e ability to measure temperature and density of the liquid   no need for straight pipe sections before and after the flowmeter  as well  as flow conditioning plates  etc        1  EMIS MAsS 260       USER MANUAL    1 3 Technical specifications    1 3 1 Technical  parameters overview    Brief description of the technical specifications of the flowmeter is presented in  Table 1 1     Table 1 1     Technical specifications    Parameter Value       Nominal diameter  mm    10  15  25  40  50  80  100  150  200    Accuracy class 0 15  0 25  0 5  Gauge pressure Toa idi   Or up to 15 MPa  special order   Medium temperature  50    350         Explosion proof grade   modification EMIS MASS 260 Ex     Atmospheric pressure  Environment temperature    Relative humidity  96    Resistance to the external magnetic  field    Vibration resistance    1Exd ib lICT6X  transmitter   1ExiblIC T1 T4 X  sensor     84 0   106 7 kPa     40    55       or  50    70   C  special order     90   3 96  non condensing  at 25          Up to 40 A m  50 Hz    class V1 according to GOST R 52931    Enclosure protection IP65  Verification period 4 years  Temperature sensor Pt100    Service life    Outline dimensions    Materials used    not less than 12 years  See                      A    Sensor     stainless steel   Transmitter     aluminum alloy   Flowmeter does not contain precious    metals   Note  1  It is possible to produce flowmeters with special characteristics according to the  order   2  Temperatu
59. r source is  connected in a way that prevents the welding current to run through the    flowmeter   see Figure 2 3        Right Wrong  Figure 2 3 Connecting the power source for arc welding using the flowmeter  As a result  the installation place must appear as shown in Figure 2 4  where the  length L is the sum of the length of the flowmeter and thickness of two gaskets   M                ml         z                   Figure 2 4 Pipeline preparation for installation of the flowmeter    CAUTION     When using filters or gas separators  the length L should be enough for their  installation        WWW flow nt rt ru                   EMIS MASS 260    2 3 4 Installation Prior to installation  it is necessary   e Thoroughly clean the pipeline of cinder  sand  and other solid particles     e Inspect the flanges and the measuring tubes of the flowmeter and remove  solid mechanical and other particles from them     e Remove the preservative grease from the flowmeter  by passing through it  kerosene  gasoline or diesel fuel     To install the flowmeter in the pipeline do the following steps  see Figure 2 5      e Arrange the flowmeter so that the arrow on the sensor   s body corresponds  to the normal direction of flow     e Slide the bolts through the holes of one of the pipeline flanges and  flowmeter s flange  put washers and nuts  Nuts should not be finally tightened yet     e Put the gasket between the pipeline flange and flowmeter s flange and  align it  It is recommended to av
60. re range for OLED display is  40    70   C   1 3 2 Range of Flow ranges of the flowmeters with accuracy classes 0 15  0 25 and 0 5 are  measurement shown in Table 1 2  Normal operation of the flowmeter guaranteed within the full       flow range according to the Table 1 2     Operation of the flowmeter with the flow rate exceeding the upper limit of the full  flow range is not allowed     The lower limit of the full measuring flow range depends on the parameters of the  medium  and should be specified in the order     WWW flow nt rt ru                EMIS MASS 260    Table 1 2     Measuring mass and volumetric flow range for liquid         Nominal Full flow range  Accuracy flow range  kg h  L h  Zero stability   mm        E   10 20     1 000 70     1000 50     1 000 0 1  15 40     2 000 150     2 000 100     2 000 0 2  25 120   6 000 400   6 000 300     6 000 0 6  40 600     30 000 2 000     30 000 1 500   30 000 3   50 1000   50 000 3 500     50 000 2 500     50 000 5   80 2400    120 000 8 000     120 000 6 000     120 000 12  100 4 000     200 000 15 000   200 000 10 000     200 000 20  150 10 000     500 000 35 000     500 000 25 000     500 000 50  200 20 000     1 000 000 70 000     1 000 000 50 000     1 000 000 100    The flow ranges are presented for water at temperature of 20   25  C  pressure of  0 1   0 2 MPa and density of 1 000 kg m  under standard conditions  For liquids of  different density the volumetric flow range should be calculated by dividing these  flo
61. rection coefficient for flow Kt    100       Format  XXXX Example     5 00      Kt 5 0   100  C      For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7            Reserved for future modifications     WWW flow nt rt ru       
62. roportional to the measured value of the mass flow rate considering the damping  time specified in the menu 49        The pulse output can be configured to denote mass flow rate  volumetric flow rate or  density of the medium  Pulse output is active     The total number of pulses generated on the pulse output corresponds to the mass  or volume of the fluid passing through the flowmeter since the measurement start     Maximum frequency  fmax  of the pulse output signal calculated as  fmax   Qmax    3 6 m   Hz  1 5   where Qmax     upper limit of the full mass flow range  kg h   m     pulse weight  g pulse     Pulse output signal frequency range is 0   10000 Hz  Maximum frequency can be  increased up to 12000 Hz     The amplitude of the pulse output signal is 13 V   Default pulse weight is presented in Table 1 4     Table 1 4     Default pulse weight         so   re   2        50            100   150                     005             4   2   4   e   10  zo   ao    1 3 5 2 Current output The value of the current in the current output circuit ranges from 4 to 20 mA and is  proportional to the measured flow rate     The current value of 4 mA corresponds to zero flow rate  The current value of 20 mA  corresponds to the upper limit of the full flow range of the flowmeter  Qmax      The current output can be configured to denote mass flow rate  volumetric flow or  density of the medium     Current output is active   The current signal parameters are presented in Table 1 5     Table
63. rs of the power supply  conform to the requirements of parameter 1 3 4  Power supply      Visibility of the information on the nameplates and other labels     e Cleanness of the exterior surfaces of the flowmeter   e Tightness of connections of the flowmeter to the pipeline   e No damage to the flowmeters parts is visible     Examination period for the flowmeter depends on operation conditions and  defined by the maintenance organization in agreement with the operating    organization     In case of the flowmeter   s failure it is necessary to follow the instructions of the  paragraph 2 5 7  Troubleshooting      CAUTION     Working in inappropriate operating conditions may lead to failure of the       flowmeter or increase of the measuring error     Table 2 8   Troubleshooting    1 When the  flowmeter IS  powered there    is no indication  on display nor  any signal on  the flowmeter s  outputs    2 When the  flowmeter IS  powered the  measured  parameters are  displayed    correctly but  there are       signals on the  flowmeter s  outputs    3 When there is  a certain actual  flow the  flowmeter  shows zero flow  rate      the  outputs    Wrong connection of the  power wires to the  flowmeter     Power wires break     Power supply voltage does  not meet the requirements  of this manual     Wrong connection of the    output wires to the  flowmeter or secondary  devices     The actual flow rate is less  than the lower limit of the  measuring range for this  size of the flowme
64. s oscillation frequency for the medium of high density  water   Used for calibration for      E     edit density measuring   4   Density D1 X XXX a mL 465  5  Low density value corresponding to the medium of low density  air   Used for calibration for density         9 E     edit measuring     L Period      XXXXXX X us 166  55  Measuring tubes oscillation frequency for the medium of low density  air   Used for calibration for      E     edit density measuring     167  55  Temperature correction coefficient for density as percentage of measured density correction for  E     edit temperature deviation of 100         Density    TempComp         96     lt      C              00    1  8  8  8  9  6  6  6  6  6  6  6  6 Zero Correction       Correction of the zero flow point in the range of  19 to  19               EMIS MASS 260    Menu  gt   Button wee  item Indication format E Description    Inner Para Lisl  i i   6        s Oene Inner parameters of the flowmeter used for diagnostic purposes   Freq XX XX Hz 169     6 Measuring tubes oscillation frequency    68 Time XX XX us   Current time delay between left and right signal coils    Drv XXX 9o  YY 170     6 Drive coil load as percentage  XXX  and its standard deviation within 8 seconds as percentage          69 m XX mV    XX mV   Voltage on the left  L  and right      signal coils   t    L   Pred XXXXX XX us  71     6 Measuring tubes oscillation period  Ind XX X   C   Internal temperature of the flowmeter   PUL  XXXX X Hz 172  6 
65. should briefly put a finger or other opaque object  close to the display glass in the area of the button  At that time the LED next to the  display lights for a moment    Flowmeter can display the measured parameters shown in Table 2 6  To change  the display page  next pair of parameters   press the     button  Display sequence  corresponds to Table 2 6     Table 2 6     Displayed parameters sequence    Displayed parameter Display format       Mass flow rate Flow XXX XXX  Mass total Mass XXX XXX  Volume flow rate Flow XXX XXX  Volume total        XXX XXX  Density Den  XXX XXX  Temperature Temp XX X    If the displayed value becomes more than 999 999  the units are automatically  switched to bigger ones to be able to display that value     If no button is pressed within two minutes  flowmeter is automatically switched to  display mass or volume flow rate  Menu items 1 or 2   depending on the state of  the menu item 47  First Menu      The flowmeter can be configured via display menu  The structure of the menu is  shown in Figure 2 14  description of the menu items presented in Table 2 7     In the editable menu items press   E   button to switch to edit mode     e  f the menu item has several predefined options press       or         to move    between those options  The currently selected option is flashing  To accept selection  press   E    The message   Saved  Y      about saving changes will appear  Press                     to move between   Y    yes  and         no  opt
66. simulation mode  in the range of  0 1 g cm  to 40 1 g cm                    25      Menu NEP Button  tem Indication format acions    RUS 181   gt 7  Select display menu language  Switching from Russian language to English is possible only via    mE OFF  82      Enable disable multipoint correction function used for correction of the calibration coefficient in  MultipointCorrect           E     edit several points of flow rate  time delay      80        2 Restore Para 183  7  Restore of all coefficients and settings of the flowmeter  except for calibration parameters in the  8 Press   E   to enter E     edit menu items 60 67  to factory defaults     184     7   84 No X TimeValue YY YYY  85         Description                Correction Point X Number of correction coefficient  0     7   which values are specified in the menu items 84 and 85     185        The value of time delay between left and right signal coils  YY YYY us  for the correction coefficient  E     edit number X     476         E     edit    Correction coefficient  in the range of 0 9 to 1 1  applied to the calibration coefficient K for time delay    E     edit  No XCorrectValue                   point specified in the menu item 84           1I 5ILLI8 MMM           EWis wass 260    2 5 4 Zero point  adjustment    2 5 5 Protection switch    WWW flow nt rt ru           After installation and electrical connection a zero point adjustment  zero calibration   procedure should be performed  This procedure sets the base
67. ss       Response     Error   86H 82H   74H       Write calibration coefficient      1100 g s us    Function Starting Registers Byte Data value CRC  address quantity count    Query        Response     Error   41H 90H 8CH 14H    www flow nt rt ru              EMis mass 260    General information   The addresses in the tables below are decimal  the starting address is 1    For example  address 127 corresponds to hexadecimal address 0x007EH  126 decimal    Read Write mode  WO     write only  RO     read only  RW     read and write       Coils   Functions  Read   01  Write     05  Read  0           1   On   Write  0x0000     Off                    On          _    gt  gt  gt  gt        gt 9   0                Meta  0001 Output emulation 12   1     Start output emulation  0003 Reset totalizer  resettable    1     Reset totalizer    0004 WO Reset totalizer  inventory      1     Reset totalizer    0005 Zero calibration  1     Start zero calibration    0009 Heverse flow total count mode  0     Subtract from the totalizer  default   1     Add to the totalizer    0011 Oil water analyze  pure oil counting   0     Off  default   1           0013   Multipoint correction    0     Off  1   On    0082 Pressure compensation  0     Off  default   1              For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7           To write in registers the switches 2 4 6 8 must be in ON position     16 bit registers     integer values  2 bytes  high byt
68. tallation in the shade is not possible  it is  recommended to use a sunshade               Figure  2 1 Basic requirements for the installation place    CAUTION     If there is a strong vibration or pipeline strained at the installation place  it is  necessary to provide external support for the pipeline before and after the    flowmeter  The support base must be firm     In this case installation of the flowmeter in such places  including mobile units is  permitted        www flow nt rt ru             EMIS MASS 260 USER MANUAL    2 3 2 Orientation Flowmeter can be installed in horizontal  vertical or inclined sections of the pipeline   It is optimal to install the flowmeter in horizontal section     The flowmeter should be installed so that its measuring tubes are always filled with  fluid and the arrow on the sensor body coincided with the flow direction  In these  conditions  the flowmeter will operate properly in any orientation    Flowmeter does not require collateral straight pipeline sections before and after the  flowmeter  nor installation of additional equipment to level the flow profile  flow  conditioning plates  etc    But if two or more flowmeter are installed in the same  section of the pipeline  the distance between them should be at least 2 m     Installation recommendations are presented in Figure 2 2           Figure 2 2 Recommendations for installation    Table 2 2     Notes for Figure 2 2       Recommendations    It is recommended to install the flowmeter 
69. ted total mass  This value is also displayed in Menu item 1  Flow Mass      PY  voiminventory XXXX XX  16  28 Accumulated total volume  This value is also displayed in Menu item 2    Flow Volm      Output Config  10          9 Pi  ss   Fs OHIO Configuration of the outputs of the flowmeter   Pulse Weight XXX  XX g N      3 MF Input mass flow rate  volume flow rate or density corresponding to one pulse on the pulse output     The pulse output can be configured to denote mass flow rate   Mass    volumetric flow rate   Volm    or density   Dens   of the medium     Description                Pulse Output Volm    Dens  20mA Value XXX X       es Input mass flow rate  volume flow rate or density corresponding to 20mA on the current output   Mass    The current output can be configured to denote mass flow rate   Mass    volumetric flow rate                 or density     Dens     of the medium     MODBUS XXX 129  c Address of the flowmeter in Modbus network   Address E     edit    Baud Rate  e Data transfer baud rate in Modbus network     Current Output Volm               Parity control in data transfer protocol  Select   NO   for no parity    Even   for 1 even parity bit     Parity Bit   Odd   for 1 odd parity bit     Stop Bits     Select one or two stop bits in data transfer protocol in Modbus network           EMIS MASS 260          1I 5ILLI8 MMM                   Menu NEP Button  tem Indication format MICE    Unit Config  11  4         Dioses 5      Select units for parameters dis
70. ter     The actual flow rate is less  than the low flow cutoff  value set in the menu     Possible faults  their probable reasons and repairing ways are presented in Table  2 8     Probable reason Repairing way    Check the power cable  connection according to the  wiring diagram in Appendix C     Check the power wires and  replace them if needed     Check the power supply voltage  and set its value in accordance  with the requirements of this  manual     Check the output wires  connection according to the  wiring diagrams in Appendix C     Check the computer   s port used  to connect the flowmeter to the  computer through digital  interface  Make sure that the  same port number is set in the   Emis Integrator  settings     Fully open the valves to set the  flow rate within the measuring  flow range     Decrease the low flow cutoff  value to be less than the actual  flow rate     WWW flow nt rt ru         EMIS MASS 260 USER MANUAL  Probable reason Repairing way    Deviation of the zero point  because of temperature  and pressure deviation  from temperature and  pressure values on  previous zero calibration        4 When there is  no actual flow  the flowmeter  indicates a  certain flow    Perform zero calibration in  accordance with the paragraph  2 5 4    Zero point adjustment        5 In the menu it   is not possible At least one of the switches Set all the switches to OFF  to enter the of the protection switch positions  see Figure 2 15  for     Calibration    block is set
71. ter can be mounted right on the flow sensor  integral type  or separate  from it  separate type      The sensor is a measuring chamber with inlet and outlet flanges for mounting on a  pipeline  Inside the measuring chamber there are two parallel U shaped flow  tubes  which vibrate by means of an electromagnetic coil and a magnet     The principle of operation is based on the Coriolis effect     Figure 1 2 shows the forces affecting the flow tube through which the measured  liquid is flowing  during the half cycle fluctuations  when the tube moves up     At this time the liquid flowing into the tube creates resistance to its upward  movement and downward pressure on the tube  Absorbing vertical momentum by  driving around the tube s bend  the liquid  flowing out of the pipe  pushes the tube  up  This makes the tube twist  When the tube is moving down in the second half of  the oscillation cycle  it twists in the opposite direction  This twisting is called  Coriolis effect        Figure 1 2     Forces affecting the tube on the upward moving    WWW flow nt rt ru            EMIS MASS 260 USER MANUAL    WWW flow nt rt ru       Twist angle of the sensor tube is directly proportional to the amount of fluid  passing through the tube within the certain time  Electromagnetic speed detectors  situated on both inlet  left  and outlet  right  side of the tube  measure the tubes  oscillations frequency  Mass flow rate is determined by measuring the time delay  between the signals of those 
72. tion mode     0136 0137 Zero       for simulation mode         Zero density  for simulation mode          9   0138 0139   RW Time                                                 BONS      for point 0  DO  us    0140 0141  RW   Timedelayforpoint1 D1 us  i  at  ttt Time     point 1  D1  us    Time His dicot  Obs for point 2  D2  us    Time        for point 3  D3  us    Time His dirs  se O for point 4  D4  us    Time               for point 5  D5  us    Time                 for point 6  D6  us    Time Hin dys  OH  _  for point 7  D7  us          0154 0155   RW Correction coefficient for              Ko KO  0156 0157   RW Correction coefficient for                          Ld K1    0158 0159 Correction coefficient for point 2  K2  0160 0161 Correction coefficient for R      K3    0162 0163   RW  Sorecion onion sont RA                     coefficient for point 4  K4  0164 0165   RW                   d    i  coefficient for point 5  K5    0166 0167 Correction coefficient for point 6  K6    WWW flow nt rt ru              EWis wass 260     Address   Mode Desepion                     0168 0169                 Mass flow rate  kg h     CIT                                                   3 _                  M                                   _                                                   vaman  2 _           rr  RO                                                                            0182           RO   Frequenoy ofthe signal on the puse ouput     _  0184          RO   meas
73. tubes downward to improve      their filling with fluid and to prevent accumulation of gas in them     For vertical or inclined pipeline orientation it is recommended to install the  2 flowmeter in a pipeline section with the upward flow direction to improve  filling of the pipe with fluid     In the event of bending pipe it is recommended to install the flowmeter in the lower  section of the pipeline     Do not install the flowmeter on a horizontal pipe before the drainage section with free  flow  because in this case filling the flowmeter s tubes wit fluid is not guaranteed     2 3 3 Pipeline To prepare for the installation of the flowmeter  it is necessary     preparation e Check the completeness of the connection kit and conformity of all  mounting parts to the flowmeter s modification     e Cut the pipeline section of length Linst  Linst   Lm   2 Lg   2 Lf   2 1   where Lm     flowmeter length  see Appendix       Lg     gasket thickness   Lf     connection kit flange thickness minus depth of landing on the pipeline   e Install connection kit flanges on the pipeline     e Using the flowmeter s replacement mounting part  fix and center flanges  and weld them to the pipeline     WWW flow nt rt ru                 EMis mass 260    CAUTION   In the process of installation the flowmeter may be used instead of the  replacement mounting part only in the following cases     installation is carried out using a gas welding       when installing using an electric arc welding the powe
74. ture in Figure 2 14 and menu items description in Table 2 7      To write in registers the switches 2 4 6 8 must be in ON position         Reserved for future modifications     WWW flow nt rt ru              EMis mass 260    32 bit registers     single precision floating point values complied with IEEE 754 format  4 bytes  high bytes first  Functions  Read     04  Write     16    For example  value     1 5    corresponds to    OxBF          0x00 0x00  in the order from low to high address in  memory      Mess   Mode        menuitem     0100 0101 Calibration coefficient  g s us 6 1    0102 0103 Temperature correction coefficient for flow      100        le  High density for density calibration  g cm    0106 0107 Period for high density  us  63    FEN  RW Low density for density calibration  g cm     0112 0113        Temperature correction coefficient for density     100                                                          0118 0119   RW Low flow cutoff  kg h     0120 0121  Zeropointus  tw  0152 0125 Density of pure oil under standard conditions    g cm   ra    0124 0125 Density of pure water under standard conditions             0126 0127 Flow compensation by percentage of gas      the oil water  0128 0129 Flow compensation by percentage of pure water in the oil water 56        0130 0131   RW Correction coefficient of the temperature  for simulation mode     0132 0133   RW Zero temperature  for simulation mode     0134 0135 Correction coefficient of the density  for simula
75. uring tubes oscilaton trequeroy te  Je _                   RO              a  ITI EET mv                        0203                           _      For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7               Reserved for future modifications     WWW flow nt rt ru              EMis mass 260    APPENDIX E       Modbus register map version 3 xx  ProLink     Modbus register map version V3 02   General information   The addresses in the tables below are decimal  the starting address is 1    For example  address 127 corresponds to hexadecimal address 0x007EH  126 decimal    Read Write mode  WO     write only  RO     read only  RW     read and write    Coils   Functions  Read   01  Write     05  Read  0           1   On   Write  0x0000     Off                    On    Address   Mode   Description   menuitem      0     Stop totalizers  1     Start totalizers   0003 HW Reset totalizers  resettable  18  0     Abort  1     Reset totalizers   0004 RW Reset totalizers  inventory      0     Abort  1     Reset totalizers  0     Abort  1     Start zero calibration   0056 RW Reset mass totalizer  resettable   0     Abort  1     Reset mass totalizer   0057 RW Reset volume totalizer  resettable   0     Abort  1     Reset volume totalizer  0     Off  default   1             For Menu items refer to display menu structure in Figure 2 14 and menu items description in Table 2 7           To write in registers the switches 2 4 6 8 must be in O
76. versions of  E     edit flowmeters      55 GasContainCorre     X  XXX       enabled in the Menu item 52     56 WContainCorre X XXX        Z    OFF  58     13  Enable disable density range check  When enabled and II density of the medium is out of range  ON E     edit specified in menu items 58 and 59 then the volumetric flow will not be calculated     UpperDen X XXXX pd The upper limit of density range  see Menu item 57    LowerDen X XXXX iR The lower limit of density range  see Menu item 57      57 Density Range       58    59                      III  H SIui9 MMM         Menu NEP Button  tem Indication format MICE    E87 flowmeter     0         E ATI          te neon a             EnterN ow 209505 i e sit press          To enter the password again press                            ge  5  3  6  7    Description                        Change Password  Press   E   to enter          Calibration       2 1      Brass                  EGO Setting of calibration parameters provided for measuring flow  density  temperature   0 FlowCal X XXX g s us     n Calibration coefficient K   Temp Correction XXXV 162  55  Temperature correction coefficient for flow rate as percentage of calibration coefficient correction for     iic E     edit temperature deviation of 100       2 H Densitv 02 X XXX a mL 163  5  High density value corresponding to the medium of high density  water   Used for calibration for    y   9 E     edit density measuring   H Period   2 XXXXXX X us 164     5  Measuring tube
77. w nt rt ru               6 485 260    2 3 6 Cooling For high temperature modification of the flowmeter  with medium temperature  above  200     possibility of external cooling is p rovided  see Figure 2 9                                 Figure 2 9     External cooling supply    2 3 7 Transmitter If in the process of installation the front side of the transmitter appeared to be not in  rotation front of the user and display is not visible  it is possible to rotate the transmitter at  an angle of 90 or 1807 so that the display is fac ing the user  To do this  unscrew  the four bolts  1   see Figure 2 10   Then turn the transmitter  2  by 90 or 180      the desired direction and tighten those 4 bolts to ensure sealing           Figure 2 10     Rotation of the transmitter    CAUTION     To avoid excessive twisting of wires inside the flowmeter do not turn the  transmitter at an angle over 180                    its ini tial position        WWW flow nt rt ru                EMIS MASS 260    2 4 Wiring  2 4 1 Basic operations Electrical connections should be performed in the following sequence  see  Figure 2 11        Remove the back cover  1  of the transmitter   e  ead signal cables  2  and power cable  7  through the cable entries  3   e Loose the terminal block screws  4        Connect wires according to the wiring diagram shown in Appendix C   e Tighten the terminal block screws   e Tighten the clamp of the cable entry   e  f necessary  install the blind plug  5  instead of the
78. w range limits under standard conditions by actual density value     If the measured flow rate is less than low flow cutoff value  the flowmeter will indicate  zero flow and accumulation of mass and volume will pause  Low flow cutoff value is  set to 196 of the maximum flow rate  Cutoff value can be changed through the menu  display or through Modbus     Flowmeter can measure flow over 1  of the upper limit of the full flow range but  measurement error in the range of 196 to the lower limit of the full flow range  296  is  not standardized  However  this error can be estimated by the formula 1 1     Medium density measurement range is 200   3000 kg m      1 3 3 Accuracy of Relative basic error of measurement of mass flow  mass  on pulse and digital output  measurement signals        calculated as    M    do    Z          100     1 1     where        accuracy class  96   Z     zero stability  according to Table 1 2   kg h   Qm     measured mass flow rate  kg h     Note     For the accuracy flow range  corresponding to a given accuracy class   according to Table 1 2   the value of Z is assumed to be O     Absolute basic error of measurement of medium density  Ap  is   1 kg m      Absolute basic error of measurement of medium temperature is  1 C     Additional error of measurement of density  caused by a change of medium  temperature is  0 03 kg m  for every 10  C of deviation from the density calibration  temperature    Additional error of measurement of density  caused by a ch
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
AFEN 6/10_10-sprachig - produktinfo.conrad.com  Fujitsu Q702    Olivia Crib (W6901) Assembly and Operation Manual    Copyright © All rights reserved. 
   Failed to retrieve file