1. When making a resistance measurement with a Test Cable other than the one supplied with the LKG 601 the user can zero the leads by adjusting the zero adjustment potentiometer This will not affect the internal calibration of the unit Annual calibration is recommended If the LKG 601 is returned to Netech for recalibration before the first year warranty expires Netech will provide a second year warranty The warranty is void if the LKG 601 is serviced by anyone other than Netech or if the warranty seal is broken 16 THEORY OF OPERATION The LKG 601 circuitry can be divided into main functional blocks as shown in Figure 3 Each performs a key role in the operation of the instrument AAMI or RMS TO gt IEC 601 DC gt Leps CURRENT AMP E SOURCE Figure 3 Functional block diagram of the LKG 601 BLOCK DIAGRAM DESCRIPTION ips S Z Q 2 G Z AV 1dSIC n fesl ra i Q O fzo T TEST RECEPTACLE Test Load The input test load is user selectable to AAMI ESI 1993 or to IEC 601 RMS Converter and Amplifier The input amplifier buffers the input The RMS converter generates a DC voltage equal to the RMS value at its input The gain Amplifier sets the gain for the measurements When the MODE switch selects leakage current measurement the AAMI LOAD is connected in front of the INPUT AMPLIFIER and the voltage developed across it from the leakage current is measured 17 Current Source an
2. surrounded by at least two inches of suitable shock absorbing material 4 Include a copy of the purchase order or reference your purchase order on your packing slip 5 Include the completed Service Form 6 Ship the instrument to the following address NETECH CORPORATION 110 Toledo Street Farmingdale NY 11735 ATTN SERVICE DEPT For the latest Revisions and updates visit www netech org 22
3. 1 lbs 45kg ENVIRONMENTAL Operating range 59 to 95 F 15 to 40 C Storage Temperature 0 to 122 F 20 to 60 C Relative Humidity 90 max at temperatures STANDARD ACCESSORIES Description Part Number User Manual 510 MANUAL Test Cable 503 OPTIONAL ACCESSORIES Description Part Number Soft Carrying Case 510 Case 20 INSTRUCTIONS FOR RETURN Damaged in Transit All shipments are carefully examined by NETECH and carefully packaged for shipment They are insured in the customers name with the carrier On receipt if the shipping container appears to have been damaged during shipment the instrument should be thoroughly inspected The delivering carriers paper should be signed noting the apparent damage If the instrument is damaged beyond use a new order should be placed with NETECH while awaiting reimbursement from the carrier for the damaged instruments Malfunction Please follow these steps for the repair or recalibration of your NETECH instrument 1 Obtain a Service Repair Form from our website at www Netech org Or from Customer Service Department Phone 631 531 0100 Fax 631 531 0101 Toll Free 800 547 6557 21 2 Complete the Service Form indicating the instruments model number serial number shipping and billing information Include the purchase order number or credit card information for the cost of the repair and or recalibration 3 Securely package your instrument in a strong box
4. LKG 601 includes a test lead for measuring Cord Resistance and Case Leakage No other cables are required to perform measurements Connect the LKG 601 to a 110 VAC outlet Or 220 VAC for 220 Volt Units Before proceeding with any measurement become familiar with the measurements and the function selector switch During the initial setup the last digit of the LCD will change gradually due to the time constant of the RMS to DC converter The display will stabilize to zero in a few seconds CAUTION Make sure that the power requirements of the Device Under Test are within the power ratings of the LKG 601 15 amps at 110 volts and 10 amps at 230 volts CAUTION Do not leave the Device Under Test continuously turned on and connected to the LKG601 11 OPERATING INSTRUCTIONS CORD RESISTANCE MEASUREMENT See Figure 2 e Set the polarity switch to the off center position e Move the MODE switch to CORD RESISTANCE Plug the Device Under Test into the test outlet e Using the test lead provided connect the chassis of the Device Under Test to the Case Ground jack on the LKG 601 e The display will indicate the Power Cord Resistance in milliohms Device Under Test Case Ground Power Cable DUT Figure 2 Block Diagram for Cord Resistance measurement and Case Leakage measurement external 12 2 CASE LEAKAGE CURRENT MEASUREMENTS Move the Mode switch to CASE LEAKAGE See Figure 2 The Earth L
5. INSTRUCTION MANUAL LKG 601 Electrical Safety Analyzer d NETECH Ww fi 110 Toledo Street lt Farmingdale NY 11735 lt USA http Awww Netech or 510 USER Manual Rev3 10 29 2007 Dear User We appreciate your purchase of the LKG 601 Electrical Safety Analyzer Properly used the LKG 601 will deliver years of high performance and accuracy This instruction manual has been designed as a tool to assist you in getting the most from your LKG 601 In order to properly use and to maintain this instrument please read the manual carefully Netech is recognized as an innovative designer and manufacturer of advanced biomedical and industrial test instruments We are ISO 9001 2000 Certified and fully committed to a continuous improvement process Further we guarantee absolute satisfaction with our products Your business is important to us and we are dedicated to providing you with the best customer and technical service possible Please contact us should you have any questions or concerns regarding your instrument We hope you will consider us again when you have a requirement for accurate reliable affordable test instruments Sincerely NETECH CORPORATION Innovative Biomedical Test Instruments Copyright Copyright 2002 by Netech Corporation All rights reserved No part of this publication may be reproduced or transmitted in any form other than for the purchaser s personal use without written permission from Ne
6. Operating Instructions Maintenance Theory of Operation Specifications Instructions for Return Figure 1 Figure 2 Figure 3 LIST OF FIGURES Front view LKG 601 Block Diagram Cord Resistance and Case Leakage Measurements Functional Block Diagram 11 12 16 17 19 21 12 GENERAL DESCRIPTION The LKG 601 is a compact low cost Electrical Safety Analyzer designed to evaluate the basic electrical safety of all electrical systems including medical devices and physiological instrumentation The LKG 601 measures Electrical Leakage Current Power Cord Ground Resistance and Device Current The LKG 601 is simple to operate A three position switch selects the test to be performed The user may select either the AAMI ESI 1993 or the IEC 601 1 test load to compensate for high frequency components in the measurement Accurate resistance measurements are made via a simple method using a single conductor cable No special cables are required A calibrated output is provided through a test jack to verify the performance of the LKG 601 Key Features e Compact e Easy to Use e Measures Device Current Cord Ground Resistance Ground Leakage Current Case Chassis Leakage Current Test Jack for Calibrated Outputs True RMS Measurement e AAMI or IEC 601 Test Load Selector 7 Figure 1 Front view of the LKG 601 CONTROLS AND INDICATORS DISPLAY A 3 Digit LCD indicates the result of the measu
7. d Amplifier Resistance is measured by connecting the current source to the resistance to be measured and measuring the voltage with the input amplifier Test Receptacle This supplies power to the unit under test 110 VAC rated at 15 Amps 220 VAC The rocker polarity switch selects NORMAL OFF and REVERSES polarity in the test receptacle Display All measurements are displayed on a 3 1 2 digit LED display Over range is indicated by showing a 1 in the most significant digit 18 SPECIFICATION S METER 3 1 2 Digit LED display LEAKAGE CURRENT 0 1999 Micro Amps All Current measurements are made through the AAMI ANSI ES1 1985 or IEC 601 1 test load The meter readings correspond to the true RMS value of the current Accuracy 1 of reading 1 LSD DC to 1 KHz 2 5 of reading 1 LSD 1 Khz to 100 KHz 5 of reading 1 LSD 100 KHz to 1 MHz RESISTANCE 0 TO 1999 Milli Ohms Accuracy 1 of reading 1 LSD DEVICE CURRENT 0 to 19 99 Amps Accuracy 2 Full Scale 1 LSD TEST RECEPTACLE Hospital Grade 110 VAC 15 Amp or 220 VAC 10 Amp The rocker polarity switch selects NORMAL OFF and REVERSE Polarity to the test receptacle and a momentary Neutral switch will open the neutral line to the test receptacle POWER REQUIREMENTS 110 VAC 50 60Hz 15 Amps or 220 VAC 10 Amps maximum rated for the test outlet and 0 1 Amp rated for the unit 19 PHYSICAL DIMENSIONS Size 5 5x 3 25 x 2 5 Inches Weight
8. eakage or Ground Leakage Current is measured through the ground conductor of the Device Under Test This is only applicable to devices with a three conductor power cord The leakage measurement is made internally when the Ground switch is momentarily opened The LCD will display the measurement in Micro Amps There is no other connection required To measure the Case Leakage of devices with plastic enclosures wrap aluminum foil around the case and connect the test lead to the foil The Case Leakage Current is measured through the test lead cable Connect the test Lead to the case of the DUT Device under test The leakage measurement is made when the Ground switch is momentarily opened The LCD will display the measurement in Micro Amps Both Earth and Case Leakage Current measurements should be made in all power switch combinations with the Polarity switch in Normal and Reverse and the Neutral switch Open and with the DUT On and Off 13 CAUTION When switching from Normal to Reverse or vice versa make sure to pause the rocker switch in the off middle position Power to the outlet will be OFF in the open neutral position 3 DEVICE CURRENT MEASUREMENT Move the Mode switch to DEVICE CURRENT Set the polarity switch to normal Turn the DUT On The LCD will display the Device Current in Amps CAUTION Do not attempt to open the unit There are no user serviceable parts inside Further the warranty will be void i
9. f the unit is opened by other than Netech trained personnel PERFORMANCE CHECK 1 Connect the LKG 601 to a live 110 VAC or 220 VAC for the LKG 601 220 outlet Display will turn on and read 0 1 when the Mode switch is in the Case Leakage position In the CORD RESISTANCE position the display will show the numeral one 1 in the left position indicating that the cord circuit is open 14 2 Move the Mode selector switch to all three positions and test the LED indication 3 Check the continuity between receptacle ground and case ground using the test cable If necessary adjust the display to zero When the open ground switch is pressed down the connection will be open 4 Connect the Test Lead to the TEST JACK Move the selector switch to CORD RESISTANCE The reading should be 1010 millionms 5 5 Move the switch to CASE LEAKAGE The display will indicate 200 Micro Amps 2 These tests will confirm that the LKG 601 is working properly 15 MAINTENANCE The LKG 601 requires only minimum maintenance Periodically check the inlet plug outlet receptacle and wiring for any damage or wear like cracks cuts or other defects Also check the case for any damage or cracks In order to ensure the accuracy it is recommended that the unit be checked periodically A calibrated output is provided at the TEST JACK for Case Leakage and Cord Resistance functions This feature is useful to check the integrity of the unit
10. misuse of the unit by the purchaser his employees agents or customers In no event shall Netech Corporation be liable for consequential damages Trademarks Netech and LKG 601 are trademarks of Netech Corporation Any other trademark names used in this manual are only for editorial purposes and the benefit of the respective trademark owner with no intention of improperly using that trademark Safety Considerations General Before operation the instrument and related documentation must be reviewed for familiarization with all safety markings and instructions Safety Symbols WARNING The WARNING sign denotes a hazard It calls attention to a procedure practice or the like which if not correctly performed or adhered to could result in personal injury Do not proceed beyond a WARNING sign until the indicated conditions are fully understood and met CAUTION The CAUTION sign denotes a hazard It calls attention to a procedure practice or the like which if not correctly performed or adhered to could result in damage to or destruction of part or the entire instrument Do not proceed beyond a CAUTION sign until the indicated conditions are fully understood and met The symbol to the left is the operator s manual symbol When you see this symbol on the instrument refer to the operator s manual 5 TABLE OF CONTENTS General Description Controls and Indicators Preparation for use
11. rement being made LEDS 3 red LEDS indicate the selected test mode MODE SWITCH Three position slide switch to select the test measurement A lit red LED indicates the selection GROUND SWITCH A switch to momentarily open the ground connection to the test receptacle POLARITY SWITCH A three position switch to select the normal and reverse polarity of the test receptacle The center position shuts the power off to the test receptacle NEUTRAL SWITCH A switch to momentarily open the neutral line to the test receptacle LOAD SELECTOR SWITCH A slide switch on the right side of the instrument to select either the AAMI or the IEC 601 1 test load TEST RECEPTACLE A 20 AMP power receptacle for the Device Under Test POWER CORD The power cord supplies power to the analyzer and to the Device Under Test 10 CASE GROUND A red jack is provided for connecting the test lead to the case chassis of the Device Under Test The Cord Resistance and Case Leakage are measured using this cable 11 TEST JACKS A calibrated output is provided at the test jack for resistance and current functions This performs a self check of the LKG 601 12 ZERO ADJUSTMENT This recessed adjustment allows for the zeroing of the display to eliminate test cable resistance variations 13 FUSE 15 Ampere fuse is in an accessible fuse holder for 110 Volt model The 230 Volt model uses a 10 Ampere fuse 10 PREPARATION FOR USE The
12. tech Corporation Quality Assurance Netech Corporation is ISO 9001 2000 Certified This instrument was thoroughly tested and inspected according to Netech s ISO 9001 2000 quality standards and test procedures and was found to meet those specifications when it was shipped from the factory Warranty Netech warranties the LKG 601 against defects in materials and workmanship for one year from the date of original purchase The standard warranty is extended for a second year if the instrument is returned to Netech for its recommended yearly recalibration During the warranty period we will repair or at our option replace at no charge a product that proves to be defective provided you return the product shipping prepaid to Netech Corporation This warranty does not apply if the product has been damaged by accident or misuse or as the result of service or modification by other than Netech Corporation or if its serial number is defaced or removed Netech reserves the right to discontinue the LKG 601 at any time and change its specifications price or design without notice and without incurring any obligation Netech guarantees availability of service parts for 5 years after the manufacture of the unit is discontinued The warranty is void if you elect to have the unit serviced and or calibrated by someone other than Netech The purchaser assumes all liability for any damages or bodily injury that may result from the use or
Download Pdf Manuals
Related Search
Related Contents
Sony VPL-PX41 User's Manual "service manual" Life Fitness Muti Jungle Handle Accessory Rack MJHAR User's Manual Solar—Alarrnanlage Ajuste - Triumph Needle Ltd ETCR Current Sensor Manual Mace CAM-73 User's Manual DeLOCK WLAN SMA antenna connection for installation Sunny Boy SB 2500 und SB 3000 String-Wechselrichter BOMBA HELICOIDAL KIBER KSF enológica Copyright © All rights reserved.