Standard Operating Procedure for Instruments

image

Contents

1. Document History Modified Document A ie Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Dissolution 2011 Dr Azza apparatus 1 OBJECTIVE The following document describes the procedure on the operation of Dissolution apparatus 2 SCOPE Dissolution apparatus is used in the pharmaceutical industry to determine the rate at which pure active pharmaceutical ingredients dissolve 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure King Faisal University College of Clinical Pharmacy 4 REFERENCES Instruction Manual Operation Guide Erweka company for Dissolution apparatus 5 DEFINITIONS Dissolution is defined as the process by which a known amount of drug substance goes into solution per unit of time under standardized conditions The bioavailability and bioequivalence data obtained as a result of dissolution testing can be used to guide the development of a new formulation and product development 6 PRECAUTIONS No part of the assembly including the environment in which the assembly is placed contributes significant motion agitation or vibration beyond that due to the smoothly
2. Set the DSC to the starting temperature in your run osc Control Panel y Click on this button to go to the i starting temp FAEPE te After the DSC status indicates that is At Temp look at the Heat Flow reading and wait until the Heat Flow is stable to 0 001 mW King Faisal University College of Clinical Pharmacy At Temp ie B 7196 j The Heat Flow should stabilize to within 0 001 mW before you start the run A Start the run by clicking on the Start Run button ee Sone tde After the run has started click on the Instrument Viewer to monitor the progress of the run Note that the default setting for the graph is to have Time displayed on the X axis If you want to change this to Temperature click on the Display menu and click on Rescale X and you will see an option to change the X axis to Temperature Please print out your blank spectrum and put your name on it and place it in the tray next to the computer monitor 82 Data Analysis blank_11 11 09 041111130407 dsd Heat FlowEndo Up mv 200 Temperature C King Faisal University College of Clinical Pharmacy gt To run an actual sample you must first encapsulate your sample in a special aluminum sample pan For instructions on how to do this please see the Appendix on sample preparation Place your sample pan into the middle of the
3. Flip power switch located on the side to turn off the machine Important Please go to the UV VIS instrument keypad and push the Mode button to return local control to the spectrometer This allows the instrument to turn off the deuterium lamp which will make the lamp last longer End of Document King Faisal University College of Clinical Pharmacy 18 SOP FOR SOXTEC EXTRACTION UNIT Prepared by Document Number Name Signature Date CCP018 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia Shariff Asst Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document F let Section Details of Amendments Date by Initials SOPs 2011 Dr Harsha First Draft on SOP for the operation of Soxtec 2045 1 OBJECTIVE To determine the concentration of the solute in the raw material by using Soxtec 2045 2 SCOPE King Faisal University College of Clinical Pharmacy Extraction is used to isolate soluble matter such as crude fat additives pesticides and critical minor constituents from complex materials The patented design of the Soxtec System 2045 makes it possible to perform extractions using a wide range of solvents in a quicker safer and more economical way compared to Soxhlet
4. King Faisal University College of Clinical Pharmacy 7 2 Running the samples e Select Start g Start Process Analyse Publish View Manage My own tab Open Dataset QFindDataset Create Dataset Paste Dataset Read Pars a m An t 4 _ a a e Click on Create data set e Following window opens J New Prepare for a new experiment by creating a new data set and initializing its NMR parameters according to the selected experiment type NAME COURSE l EXPNO PROCNO DIR USER Solvent decatur Experiment DIEE e Enter name and experiment number Keep the PROCNO 1 only e Select the solvent that you are using from the drop down menu e Experiment use current params e Title Enter title and press OK e New window opens To Read parameters Select Proton for 1H NMR C13CPD for 13C NMR DEPT First perform C13CPD than DEPT 135p OR first DEPT 90 than DEPT135p 2D NMR COSYGPSW for H H COSY Select HSQCGP for C H COSY e Click Read Pars e rpar window opens Select all and click OK Data Acquiring King Faisal University College of Clinical Pharmacy Preparation of NMR sample Prepare the sample in the required solvent and fill about 6 cm 4 cm to 8cm 6 cm is ideal Insert the sample tube in to the spinner Keep the spinner into the sample depth gauge press the sample tube till it touches the white base The bottom level sho
5. 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Shimadzu GCMS QP2010 system user s Guide Shimadzu GCMS Direct sample inlet device DI 2010 instruction manual Shimadzu auto injector auto sampler AOC 20 user manual 5 DEFINITIONS GCMS Gas Chromatography Mass Spectrometer 6 PRECAUTIONS GC column temperature should be below 280 oc Don t run the samples having more than 300 C boiling point During direct injection method use very less sample to avoid saturation of ionization chamber 7 OPERATING THE GCMS SPECTROPHOTOMETER 7 1 Turning on the Instrument Open the regulator of helium gas cylinder which is outside the research lab Than Open the helium gas pipe regulator which is located below the AAS instrument Set the Pressure between 7 9kg Switch on the Auto sampler power controller Behind the GCMS Turn on the Mass Spectrometer using the switch located on the lower left back panel of the instrument Turn on the GC system using the switch located on the lower right front panel of the instrument King Faisal University College of Clinical Pharmacy 7 2 Operation of GCMS software GCMS solution Real Ti ll Cl
6. a 4 Reference furnace is on the right side Sample furnace gt is on the left side King Faisal University College of Clinical Pharmacy After you have checked that both furnaces are empty put their platinum lids on close the swing arm cover and put the dome on top Dome cover is shown in place over the furnace assembly Enter sample ID information here gt Click on Browse to navigate to the folder you want to put the For the blank run you do not need to data in and to give enter anything for the sample weight ita filename For an actual sample you must enter its weight if you want to measure AH College of Clinical Pharmacy King Faisal University For the blank run set the final to400C gt win fra For the blank run set the fan f heating rate to 40 C mm gt Set the DSC to the starting temperature in your run Set the DSC to the starting temperature in your run gt Set the DSC to the starting temperature in your run iad F ad f y King Faisal University College of Clinical Pharmacy Swing arm cover is shown in place over the fumaces Push down on the lever to lock it in place Platinum lids are shown in place on top of each aati Wait tweezers deform the platinum lids to set them in place This will affect the flatness of the baseline Please do not bend or
7. 1 OBJECTIVE The following document describes the procedure on the operation of Zetasizer 2 SCOPE The Zetasizer measures particle and molecular size from below a nanometer to several microns zeta potential electrophoretic mobility and molecular weight Zetasizer is used for 2 KS Determination of optimum dispersion conditions for a wide range of applications 2 Stabilization destabilization of suspensions dispersions and emulsions via electrostatic stabilization mechanisms 7 KS Determination of critical coagulation concentrations sensitivity to the concentration of additives such as salts surfactants polymers 7 Estimation of adsorbed polymer layer thickness 7 Optimizing surface coatings effect of surface modifications KS Biological applications characterization of erythrocytes and blood platelets characterization of cells bacteria etc King Faisal University College of Clinical Pharmacy 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Instruction Manual Operation Guide Malvern company for Zetasizer MAN0317ISUE 2 1 JULY2004 5 DEFINITIONS The particle size measured in a DLS instrument is the diameter of
8. 7 KS screen 7 3 Adding product to the manifold for freeze drying King Faisal University College of Clinical Pharmacy gt KSA 7 7 7 7 KS 7 7 4 Make sure that the product is frozen in the flask before attaching to the freeze dryer When freezing product in flasks the flask should be tilted on its side If the flask is not tilted the expansion of freezing will break the flask Attach the flask to the vacuum valve using a 3 2 to 1 2 adapter Make sure that the adapter is greased Slowly open the valve to apply vacuum to the flask Wait for the READY FOR FREEZE DRYING indication before adding additional product Removing product from the system To remove the product from the system you must remove the vacuum in the flask vial container without removing the vacuum in the freeze dryer Simply turn the valve stem past closed to release the vacuum The vacuum is released through a small opening in the valve stem 7 5 Defrost When a sufficient amount of material accumulates on the condenser the frozen material will obstruct the vapor flow At this point the system must be defrosted Remove any unfinished product and keep it frozen The Opti dry offers hot gas defrost to speed the defrosting process 2 2 Add approximately 1 liter of warm water to the condensing chamber making sure not to fill above the vacuum port at the
9. bameter Bo Hardness lo Correction value N Diameter setting mm operating Manual NOTE You may test not more than 100 tablets during one test run If you enter a value out of the range of 1 100 the following message will appear King Faisal University College of Clinical Pharmacy Input range 0 0 100 0 Check your entry Check that the information you have provided so far is correct and then tap the Inspection tab The following window appears MB Pocket_PC File Zoom Tools Help Measurement Weight Thickness Diameter Hardness e Place the product sample to measure weight and tap Start Wait till the measurement is done The value will appear in the Weight field Now the window can look as follows King Faisal University E Pocket _PC File Zoom Tools Help AT ERWEKA H Test Measurement Weight 100 0 Thickness Diameter Hardness NA eo 14 51 Wait for Thickness College of Clinical Pharmacy INFO If no measuring value for weight or thickness measurements has been transferred within the request period the EE error message appears and you will be prompted to repeat the measurement e Place the product sample to measure thickness Wait till the measurement is done The value will appear in the Thickness field Now the window can look as follows King Faisal University College of Clinical Pharmacy MH Pocket_PC SEE File Zoom Tools Help J ERWEK
10. 2 SCOPE This document applies to the procedures to run NMR experiments on the Bruker 400 MHz NMR 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Bruker NMR Instruction Manual Operator s Manual Programming Procedure for Bruker NMR King Faisal University College of Clinical Pharmacy 5 DEFINITIONS NMR Nuclear Magnetic Resonance DEPT Distortion less Enhancement by Polarization Transfer COSY Correlation Spectroscopy NOESY Nuclear Overhauser effect spectroscopy 6 PRECAUTIONS ALWAYS SET DEPTH with depth gauge Failure to set depth will break your sample and or the instrument 7 PROCEDURE FOR OPERATING THE Bruker 400MHz NMR Spectrometer Turning on the TOPSPIN software e Switch on the PC e Wait until login screen appears e Switch on the Console e Wait at least 3 minutes e Login as NMR User and the PW is topspin e Start Topspin 3 0 e Ignore all error messages e Type edhead in the command line at the bottom of the screen e Ignore the error message e Edhead window opens Click on exit e Channel Routing window opens click on close e Type ii restart in the command line The BRUKER 400 MHz NMR is ready to use
11. College of Clinical Pharmacy King Faisal University College of Clinical Pharmacy 1 SOP FOR ATOMIC ABSORPTION SPECTROPHOTOMETER Prepared by Document Number Name Signature Date CCP001 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document let Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh Atomic Absorption Spectrophotometer 1 OBJECTIVE The following document describes the procedure on the operation of Shimadzu Atomic Absorption Spectrophotometer AAS 2 SCOPE Atomic absorption spectroscopy AAS is a spectroanalytical procedure for the qualitative and quantitative determination of chemical elements employing the absorption of optical radiation light by free atoms in the gaseous state In analytical chemistry the technique is used for determining the concentration of a particular element the analyte in a sample to be analyzed AAS can be used to determine over 70 different elements in solution or directly in solid samples 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students o
12. ie Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Buchi Nano 2011 Dr Harsha Spray Dryer B 90 1 OBJECTIVE Spray Drying is a unique method to convert a solution suspension or emulsion into a solid powder in one single process step by using Nano Spray Dryer B 90 2 SCOPE The latest application trends in pharmaceutical formulation and nanotechnology demonstrate the need for fine particles in small powder quantities and with very high yields The Nano Spray Dryer B 90 is BUCHI s 4th generation lab scale spray dryer and is particularly designed to evaluate spray drying during the early stages of product development The modular and flexible glass design makes it easy to spray dry a whole variety of applications and is ideal for feasibility studies in R amp D laboratories where only milligrams of powder need to be dried King Faisal University College of Clinical Pharmacy The output particle size of the Nano Spray Dryer B 90 is between 300 nm 5 um Thus the instrument is ideally suited to encapsulate particles and nano particles lt 100 nm from a solution or suspension 3 RESPONSIBILITIES The Nano Spray Dryer B 90 is only to be used by laboratory and specialist personnel and other persons who because of their training or professional experience are able to assess the dangers which can result while operating the equipme
13. Enter file name e Text Box Description of sample Executing Background Scan Empty the sample holder Click Bkg Start Button To stop the background scan click stop During the background scan do not use the close button X King Faisal University College of Clinical Pharmacy Sample Set Up Executing Sample Scan Prepare sample KBr pellet as per the above procedure 7 1 Place the KBr pellet in a pellet holder and put it into the sample beam of an IR Click sample start The scan can be paused by pressing stop The scan can be resumed by pressing continue The number of scans is displayed in the bottom left corner Saving Reading the Spectrum e To save the spectrum with a new file name click file then save as 7 3 Data Analysis The spectrum is displayed automatically Enlarge an area by clicking and dragging a box around the area to be enlarged Zoom In Left mouse click Zoom Out Double right mouse click Select the Graph drop down menu and click auto scale to return to the original spectrum Displaying Peaks e Open display box by e Selecting the calculation drop down menu and clicking peak table e Press the peak table button e Under graphic section press the zoom button to display the peaks Peak Detection Box e The default threshold for absorbance is 0 1 e The default threshold for transmittance is 80 Threshold e Absorbance will search for peaks above 0 1 e Transmittance will search for pe
14. Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History eee Details of Amendments Date ee SOPs First Draft on SOP for the operation of TC 412 Dr Emeka Thermal Cycler PCR 01 12 2011 1 OBJECTIVE The following document describes the procedure on the operation of TC 412 Thermal Cycle PCR 2 SCOPE The following procedure applies to designated personnel in Research laboratory PCR is a technique that allows Research Scientists to create or amplify a vast quantity of a specific sequence of DNA e 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure It is important that only trained personnel are allowed to operate this equipment King Faisal University College of Clinical Pharmacy 4 REFERENCES TC 412 Instruction Manual Operator s Manual Programming Procedure for TC 412 Thermal Cycler for use 5 DEFINITIONS K2 PCR Polymerase Chain Reaction DNA Deoxyribonucleic Acid 6 PRECAUTIONS Lab coat and hand gloves must be worn when preparing samples for the PCR When using the machine hand gloves must be
15. King Faisal University College of Clinical Pharmacy Death or serious poisoning by inhalation or incorporation of dried particles during spray process Wear safety goggles Wear safety gloves Wear a suitable protective mask Wear a laboratory coat Check for proper sealing before use Do not inhale dried particles Stop drying gas flow before opening the drying circuit Starting a spray drying process Switch on the instrument After the automatic self check the operating mode is activated 2 KS Switch on all other devices if installed in the current configuration 7 Set the drying gas rate to a reasonable flow of about 70 150 litre min with the o gt external gas regulator when using compressed gas or adjust the frequency controller when using the Aspirator set a rotational frequency between 30 40 Hz 7 Read the relative inside pressure typical pressure range 30 mbar to 60 mbar KS Select the desired inlet temperature typical setting between 80 C to 120 C Sd Switch on the heater by pressing the Heat button and wait until the system has K reached stationary temperature conditions typically after 5 to 10 min Switch on the peristaltic pump by pressing the Pump button Select the relative spray rate typically 100 to achieve maximum throughput Press the Spray button to start the spray process The spray process starts and the electr
16. REFERENCES King Faisal University College of Clinical Pharmacy Instruction Manual Operation Guide Tapped Density Tester SVM 202 Erweka Germany 5 DEFINITIONS The Tapped Density Tester provides a standardized reproducible method for measuring the tapped or bulk density of powdered granulated or flaked materials 6 PRECAUTIONS The Tapped Density Tester produces excessive noise levels During operation consider wearing approved hearing protection devices or operating the unit in the optional acoustic cabinet Operation of a Tapped Density Tester involves the use of solids and powders Unskilled improper or careless use of this instrument can create shock hazards 7 PROCEDURE FOR OPERATING THE AA SPECTROPHOTOMETER Dual platform Tapped Density Tester r 77 Check that all the connections of the instrument are proper KS 7 Kosa Weigh the sample amp fill in a measuring cylinder of Tapped Density Tester S Fix the measuring cylinder into the holder provided for holding the measuring cylinder on the instrument Power ON the instrument o gt Select the mode of operation either timer or counter by pressing TIME COUNT This key toggles the unit between the two operating modes The Timer LED King Faisal University College of Clinical Pharmacy illuminates to indicate the timer option is selected The Counter LED illuminates to indicate the count
17. e The air pump is automatically started when there is 3 min left of the analysis The last traces of solvent will now be collected in the condenser The Control Unit starts the buzzer signal when the countdown is finished If you want to use the pump for a longer time more than three minutes you must start it manually This may be necessary for high boiling solvents like Toluene or Xylene As soon as all cups are dry lift the cups from the hot plate e Remove the thimbles e If another extraction is to be performed immediately insert a new batch of thimbles and cups 5 2 7 Extraction on page 5 3 in USER MANUAL and open the condenser valves Note The recovered solvent volume can be adjusted by adding solvent at the top of the condenser Drying Step Continue Heating e It is possible to let the controller maintain the temperature of the hot plate for up to 59 minutes after the recovery step is finished e When setting up the program set the fourth step on the controller to a desired value from 0 to 59 minutes Resetting the Over temperature Protection King Faisal University College of Clinical Pharmacy Wait until the temperature of the plate is below the over temperature limit Turn the power off Turn the power on Turn the over temperature key to unlocked Turn the over temperature key to locked Changing of Solvent A WARMING Always be careful when changing solvent You should for example check the ignition
18. temperature of the new solvent used and make sure that the temperature of the hot plate does not exceed the ignition temperature Instructions First empty the condensers from solvent Run the system without samples and with a small amount 10 20 ml of new solvent in the cups Run a short program at the recommended temperature with the sample positioning rods in the rinse position for 10 minutes followed by recovery for 15 minutes Collect the solvent from the cups Discard the recovered solvent as it will be contaminated by the solvent previously used Repeat if you find it necessary Temperature Table Note The temperatures listed in the table below are given for guidance only they may have to be adjusted for factors like room temperature water temperature heat transfer from hot plate to cup etc A suitable temperature is when the drop rate is approx 3 5 drops per second King Faisal University College of Clinical Pharmacy Table 3 not recommended The figures m Table 3 refer to alummiwm and glass cups OT over temperature stands for the recommended settmzg on the over temmperatwe control pad on the Con trol Unit A WARMING The over temperature setting should always be checked for each solvent used Our recommendations are selected according to the ignition point To be sure you are advised to check the specific safety data sheet for the solvent to be used Closing Down e Switch the MAIN s
19. 12 and 14 psi Turn on the AAS Power switch is on the front lower right corner of the instrument From instrument drop down menu select the Connect Send Parameters button The program will proceed through an initialization checklist When the program asks if you would like to test the safety parameters click no and then ok You will need to do this 3 times We run though all of the safety checks once a week Instrument check list for Flame analysis will open Tick all the boxes and press OK King Faisal University College of Clinical Pharmacy Igniting flame Simultaneously press two buttons ignite and Purge The burner head will ignite 7 2 Soft ware Operation L Mg D Fine Mon ON READY Element selection A Use this dropdown box Parameters to select which element you are working with From the Parameters drop down menu select Edit Parameters and then click on Line Search This usually takes a few minutes When the line search finishes a spectrum will appear with a red line crossing at the maximum If the red line is not at the maximum run the line search again When this finishes hit Close and press OK on Edit parameter window 7 3 Data Collection AUTO ZERO BLANK AND COLLECT DATA Place the suction tube in DI water Watch the Absorbance vs time plot in the upper left corner of the screen B When the Abs level has stabilized cl
20. C2 7 Ww 1 Me 7 PROCEDURE FOR OPERATING THE AA SPECTROPHOTOMETER Thickness Tester Balance King Faisal University College of Clinical Pharmacy General view of Handheld Control Unit HCU E Pocket PC s BR START menu File Zoom Tools Help MJ ERWEKA H Test 7 4 09 33 Main window Options menu Task bar Action End E Calibration Audit Trailicon Keyboard icon window reminder King Faisal University College of Clinical Pharmacy Measuring Gauge Movable Test Jaw Height Gauge 0 5 mm Parallel printer Thickness RS232 Balance watch PC New measurements e A new measurement is a test you conduct on a product that has not been tested on the instrument yet nor has it been created in the database see Operating Manual Product Management King Faisal University College of Clinical Pharmacy e Thus you have to first enter the information about the product and only then perform needed measurement tests Switching on the instrument Power up the instrument by means of the mains switch on the rear side of the instrument When you log in to the instrument for the first time use the following details user admin and password abc123 To perform a new measurement 1 Log in to the instrument 2 Tap Action Measurement New The following window appears E Pocket_PC Sele File Zoom Tools Help HF ERWEKA E 461342 Measurement Weight 0 Thickness j0
21. First switch on the nitrogen generator Operation pressure should be between 80 100 Then turn on the mass spectrometer Then HPLC instrument from down to up i e DAD pump degasser Open the chem Station by clicking on the online chem Station icon King Faisal University College of Clinical Pharmacy 7 2 How to Run the LC MS amp Before running the LC MS you should first check the solvent reservoir to make sure that there is enough solvent available for the number of samples that you want to analyze o To create a new method select new method from the method drop down menu Set the injection volume by clicking on the injection icon Set the flow rate of the mobile phase run time clicking on the column icon C KS Click on the DAD and set the wavelength s required for detection of your analyte Click on MSD and setup parameters SCAN SIM mode m z range etc 2 KS 0 Sd Than save the method 0 Alternatively to create new method select Edit entire method from Method drop down menu and follow the steps 7 3 For single run KS Select sample info from the Run method dropdown menu Fill up the required information vial no sample name etc kS P Torun press Run Method OG To view the DAD and MSD graph select signal 1 and signal 2 from the view and online signal dropdown menu 7 4 For sequence run Select new sequen
22. Help 5 Carrier Flow LIT ml min Carrier Pressure 9kPa PurgeFlow 80 ml min TED agoa K Front Fumace 990 C Rear Fumace C GCOven IH C gt Performing the Leak test Click on the Instrument menu from the menu bar King Faisal University College of Clinical Pharmacy Select Leak test from the drop down list Click on the START leak test Button gt e amp The result is displayed in the label reporting passed or failed You must not proceed until the leak test has passed Setting Auto run Window Click on Analysis on the menu bar Select New autorun from the drop down menu Fill the sample table Autorun L Cystine Organic Div 0001016 Sample Table Method Report Configuration Sample Name Parameter Applied Set to CHNS Configuration 1 L Cystine 1 Std 0 5610 gt Ea 2 cystine Sta 1 0560 Eanes ee Additional Information 40005 it 80 2 LCystine 3 std 1 4520 Fage be oe era 4 Sample Smp 07010 Biggen ee 5 Sample5 cng 0 8360 APO a ES 6 Sample 6 Smp 1 2210 Oxidation time 87 sec oF Sample Delay 10 sec 10 Run time 320 sec 320 FrontFumace 1020 C 1020 j Print options zj RearFumace OFF C OFF Default reports F 115 Custom Oo Rows 6 Incomplete 0 Complete 6 Calibrations 1 CAL GC Oven 115 C pus C Allreports a me Delete Ro
23. Initials SOPs First Draft on SOP for the operation of Differential 2011 Dr Harsha Scanning Calorimetry DSC 1 OBJECTIVE Differential scanning calorimetry DSC monitors heat effects associated with phase transitions and chemical reactions as a function of temperature In a DSC the difference in heat flow to the sample and a reference at the same temperature is recorded as a function of temperature The sample is sealed in an aluminum pan The reference is an inert material such as alumina or just an empty aluminum pan The temperature of both the sample and reference are increased at a constant rate 2 SCOPE The Perkin Elmer DSC 7 differential scanning calorimeter is used to accurately measure melting points glass transitions and heats of fusion 3 RESPONSIBILITIES King Faisal University College of Clinical Pharmacy It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Differential Scanning Calorimetry DSC Instruction Manual Operation Guide 5 DEFINITIONS Nil 6 PRECAUTIONS e Never touch the outer and inner furnace lids The temperature of the furnace can reach as high as 450 C Use tweezers to remove the lids e Do not touch the inside of the furnace it m
24. QUITE DANGEROUS End of Document King Faisal University College of Clinical Pharmacy 2 SOP FOR CHNS ANALYZER Prepared by Document Number Name Signature Date CCP002 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document i Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Eurovector 2011 Dr Mahesh CHNS analyzer 1 OBJECTIVE The following document describes the procedure on the operation of CHNS analyzer 2 SCOPE CHN Analyzer is a scientific instrument which can determine the elemental composition of a samples such as organic and inorganic compounds This instrument is used for qualitative and quantitative analysis of elements CHNS in pharmaceuticals 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES King Faisal University College of Clinical Pharmacy EuroEA 3000 series installation Proced
25. extractions Thus the technique normally reduces extraction times to less than 20 of Soxhlet times and recovers 60 70 of the solvent 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES A Soxtec System 2045 Extraction Unit User Manual 1001 1992 Rev 1 3 5 DEFINITIONS Thimble Pure Cellulose Extraction Cup Holder For inserting and removing of extraction cups the cup holder is used Thimble Adapter For attaching the thimbles in the Extraction Unit Thimble Stand The thimbles are placed in the thimble stand during drying 6 PRECAUTIONS 2 KS Please read these operating instructions carefully and act accordingly 7 KS Due to the explosion risk when handling organic solvents the operator should avoid exposure of solvents to open fire static electricity etc Laboratory coats and clothing made of certain synthetic fibres are prone to generate static electricity and should be avoided The over temperature setting should always be checked for each solvent used Always use a damp cloth when wiping off surfaces on the extraction unit 7 OPERATING INSTRUCTIONS King Faisal University College of Clinical Pharmacy These captions are tra
26. in progress Stop any data acquisition Close the data acquisition program Closing this program will log you off so that you will not be billed for any more instrument time Close any other software such as the GCMS Postrun analysis program Remove your samples from the autosampler carousel End of the Document King Faisal University College of Clinical Pharmacy 10 SOP FOR HIGH PERFORMANCE LIQUID CHROMATOGRAPHY Prepared by Document Number Name Signature Date CCP010 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document oa Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh Spectrophotometer 1 Purpose The following document describes the procedure on the Operation of the Shimadzu HPLC 2 Scope This instrument is used to separate and quantify organic compounds in pharmaceutical formulations and biological fluids 3 Responsibilities It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the respo
27. in the Autorun window press Send to instrument When monitor window bar shows READY press START The acquisition window appears and acquisition starts Viewing the acquired Chromatograms and Results e Click the acquired samples button in the Acquisition window i Click the desired sample Table line The results window reports the element feet orate t CpstaveUrgernn creel OCG amp Sante bh x BIE lest tile ie E E E FT Catanncre town tie Sve K PUER ags Saated sangi Po85 Bangle ip Kann L Ojdire Orgar Ore 1018 Acid Emea Chdbuatin Type L Eherserts Andyeot MEHS gt Shutting down the instrument Click on Operator on the main menu Select Exit Callidus Than press OK Turn off instrument King Faisal University College of Clinical Pharmacy Close the gas cylinder valves End of Document King Faisal University College of Clinical Pharmacy 3 SOP FOR DIFFERENTIAL SCANNING CALORIMETRY DSC Prepared by Document Number Name Signature Date CCP003 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia Shariff Asst Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document i me Section Details of Amendments Date by
28. input and menu selection while a red round button can interrupt the running process PROCEDURE FOR OPERATING THE MAESTRO 101 e Turn on the MAESTRO 101 e Prepare the migration tank by filling it with 30 mL of buffer solution e Remove the reagent tray out of its position and fill each tank with 175 mL of corresponding solution i e first with Fixative solution second with Staining King Faisal University College of Clinical Pharmacy solution third and fourth with distaining solutions and place them back in corresponding positions to the contents e Select the predefined program sequence e Take the gel and out of its package and place it on table with gel side up and dry it with gel blotter e Place the application template on marking at cathode side and press gently to remove the bubbles between template and gel e Fill each slid with 1 uL of diluted serum and wait for about three minutes to let the serum intrude completely e Place the small paper wick on application slides and place the gel in gel bridge with gel side up and place the gel bridge in the tank e Close the lid of the migration tank e Select Learn parameter and then select migration tank on the controls e Confirm the voltage 70 volts and the migration time 25 minutes and then launch migration e After migration remove the gel from Gel Bridge and place insert it into gel frame e Place the gel frame into the moving arm and launch the developing proced
29. parameters for the gas chromatograph e The third tab has parameters for the mass spectrometer King Faisal University College of Clinical Pharmacy Sampler cy GC a Ms Inj Port SPL1 Inj Heat Port INJ1 Column Oven Temp 300 o y ks Injection Temp 20 0 ny Bs Injection Mode Splitless gt Sampling Time 050 min m Carrier Gas He Prim Press 500 900 Program Column Oven Temperature v Flow Control Mode Linear Velocity v eee Pressure 45 6 kPa Total Flow 24 0 mL min i Column Flow 1 00 mL min a Linear Velocity 35 9 EWE Total Program Time 13 80 min Purge Flow 2 0 mL min Column Spit Hato 20 0 Name DB 5 ms Thickness 0 25 um Length 30 0 m Diameter 025mm _ 8t r Detail of Injection Port H Ee ligh Press Injection ammer as aver Splitter Hold Far GC Program plit Ratio Program Prerun Program lime Program Sampler fg cc amp ms GCMS QP2010 lon Source Temp Ne Interface Temp 250 C Detector Voltage Relative to the Tuning Result C Absolute Solvent Cut Time Bs min jo kV Micro Scan Width O u Threshold hoo I Use MS Program _set GC Program Time 13 80 min Group 1 Event 1 1 4 00 13 80 Scan 2 0 00 0 00 Scan Fill out the required parameters as per your sample analysis and Save the method King Faisal University College of Clinical Pharma
30. patient centered care to meet the needs of the community PROGRAM GOALS Prepare active learners and future experts in Clinical Pharmacy with required knowledge skills and abilities Educate for management of Pharmacy services and practice Provide foundation for basic and clinical research in the area of drug development and usage PROGRAM OUTCOMES After completion of the progam the graduate will be able to l Apply fundamental scientific knowledge and principles as basis of Pharmaceutical and Clinical Sciences Apply knowledge of biomedical sciences for understanding of disease process and its diagnosis Develop patient data base from patient interview review hospital record and communication with other health professionals Apply knowledge of clinical Sciences in designing patient specific therapeutic plan based on best evidence and counseling the patient regarding it Retrieve interpret and report drug information from pharmaceutical and biomedical science recourses and apply information to specific patient care situation Recognize basic principles of drug development formulation and bioequivalence studies to contribute in pharmaceutical industry and in research institutes Appreciate necessary knowledge in organizational and administrative domains for management of Pharmaceutical services and practice Recognize basic principles to design implement and conduct research studies in different fields of pharmacy practice
31. resolution Peak assignment A chemical fingerprint for the compound of interest is generated ensuring correct peak assignment in the presence of complex matrices King Faisal University College of Clinical Pharmacy Molecular weight information Confirmation and identification of known and unknown compounds Structural information Controlled fragmentation enables structural elucidation Rapid method development Provides easy identification of eluted analytes without retention time validation Sample matrix adaptability Decreases sample preparation time 2 KS Quantitation Quantitative and qualitative data can be obtained easily with limited instrument optimization 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Agilent LC MSD Quick start guide Agilent Chem station foe LC 3D systems Understanding your spectra module Agilent 6100 series Single Quad LC MS system User s Manual 5 DEFINITIONS MSD Mass spectrometer detector LC Liquid chromatography 6 PRECAUTIONS Nitrogen gas generator should read less than 1 in the operating meter 7 PROCEDURE FOR OPERATING THE AGILENT LCMS 7 1 How to switch on the instrument LC MS
32. rotating stirring element 7 PROCEDURE FOR OPERATING THE DISSOLUTION APPARATUS Stirring Element Height Set the distance between the bottom of the vessel s inner surface and the lower part of the stirring element Use height gauges to check the distance between the bottom of the vessel and stirring element Measure the distance for each vessel position Some test assemblies allow the height of the stirring element to be set before beginning the test All values must be within 23 0 and 27 0 mm Baskets Place one tablet in each dry basket Attach the basket to its shaft The test 7 KS is considered to start when the basket is immersed in the medium at the prescribed height Immediately start the stirring 2 Paddles Allow the tablet to fall into the vessel in a standardized manner e g along the vessel wall or at the center of the vessel along the paddle shaft The test is considered to start when the tablet comes to rest at the bottom of the vessel The most reproducible conditions will exist with the sample settled directly under the paddle shaft e g on the vessel axis Immediately start the stirring 7 KSA Timing Record the start time of the dissolution test using a calibrated timing device The type of starting procedure must allow for sample collection and filtration of all positions within the current 2 tolerance limit for time 36 seconds for a 30 minute run The time needed to sample each vess
33. used automated sampling methods and filter probes should be validated against the manual sampling method Variables include carry over from system surfaces exposed to test solution dwell time for on line spectrophotometric analysis and filter interference End of Document King Faisal University College of Clinical Pharmacy 6 SOP FOR FREEZE DRYER BENCH TOP Prepared by Document Number Name Signature Date CCP006 Dr Azza Ali Hasan Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document A ie Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Millrock 2011 Dr Azza Bench Top Freeze Dryer 1 OBJECTIVE The following document describes the procedure on the operation of Millrock Bench Top Freeze Dryer 2 SCOPE Freeze Drying or lyophilization as it is referred to in the Pharmaceutical and Diagnostic industries is a dehydration technique which enables liquid or slurry products which have previously been frozen to be dried under a vacuum The applications of freeze drying are numerous such as Preservation of temperature sensitive products particularly those of biological origin such as enzy
34. 00 ea level 600m Rot release 3000ms Key sound No English Items in menu Configuration Description available parameters Units Temperature in C K or F Volume flow in l min m3 h ft8 h e Pressure in mbar Torr hPa LCD Contrast Select a screen contrast between 0 and 100 Sea level Select the sea level in meters over sea level Rot release Defines the time interval in which the rotary knob is active after pressing the selection buttons for Pump Spray and Temp The default time is 3 seconds This is a feature to avoid unintended inputs Key sound If YES is selected a sound is audible when pressing a function key To suppress the acoustic confirmation select NO Language Choose the language of the user interface between Chinese English French German Italian Japanese or Spanish Firmware calibration menu e To enter the Menu press the corresponding multifunctional button below the screen Inside the menu use the multifunctional buttons for navigation to Calibration Stop the gas flow and confirm by pressing the OK button to recalibrate the internal flow sensor of the Nano Spray Dryer B 90 King Faisal University College of Clinical Pharmacy Calibration Flow calibration Flow calibration Be sure that flow is zero Note The calibration should be performed periodically to ensure best spray drying results Starting a spray drying process
35. A H Test q 14 45 Measurement Weight 100 0 mg Thickness 3 00 mm Diameter Hardness NA N Wait for Hardness e Place the product sample to measure diameter Wait till the measurement is done The value will appear in the Diameter field Now the window can look as follows e Place the product sample to measure hardness Wait till the measurement is done The value will appear in the Hardness field After all the measurements have been done the window may look as follows E Pocket_PC DE File Zoom Tools Help J ERWEKA H Test q 14 56 Measurement Weight 100 0 mg Thickness 3 00 mm Diameter 10 00 mm Hardness NA N Wait for Hardness King Faisal University College of Clinical Pharmacy E Pocket_PC e Then a message appears asking you whether you want to print the results Tap OK to print Tap No to reject INFO To stop the test at any time tap the Stop button in the window You can continue the test later If you want to abort the test tap Cancel INFO If the nominal value for the measuring parameter diameter is set too low NA appears in the protocol after the measuring result INFO You can print measurement test results after the measurement See Reports for more details Cleaning e To guarantee the disturbance free operation of the instrument it is recommended to clean the instrument daily e To avoid adulterated measuring results clean the fixed and movable test ja
36. CES King Faisal University College of Clinical Pharmacy MAESTRO 101 04 2005 Manual Instruction manual for Biotec Fischer Serum Protein Electrophoresis Operating Manual for Serum Protein Electrophoeresison MAESTRO 101 with Biotec Fischer SPE Agarose gel kit code 370105 5 DEFINITIONS Proteins Serum proteins can be classified into Albumin Alpha Globulins Beta Globulins and Gamma Globulins Specimen Freshly collected serum or a solution containing various proteins 6 PRECAUTIONS Use the items reagents and products provided by the manufacturer Buffer contains Sodiumazide so it should not be mixed with acid and proper disposal flushing with abundant water is required Hemolysis may cause false elevation of alph 2 and beta fraction of the proteins Inaccurate results may be obtained due to specimen left uncovered 7 INTRODUCTION MAESTRO 101 consists of two parts a migration tank with an integrated power supply and a control unit for sequence programs with developing unit for migrated gels Display elements and operation elements are located on developing unit e Three green yellow and red squires inform about the status of the unit where green light area means ready to work yellow light area means active work in progress and red light area means fault in the status e LCD display is a 4 line display that shows status of commands e The operation elements include a knob used for data
37. H Dr 250WTD for weight variation 2011 Harsha 1 OBJECTIVE A tablet is designed to contain a specific amount of drug in a specific amount of tablet formula As a check that a tablet contains the correct amount of drug tablet weight thickness hardness is routinely measured Inconsistent tablet weights are caused by mechanical problems with the tablet press such as incorrect operation of the punches improper feed of granules into the die cavity and incorrect tablet press speed 2 SCOPE TBH 250WTD Applicable to determination of weight diameter hardness and thickness of a tablet during in process checking King Faisal University College of Clinical Pharmacy 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Instruction Manual Tablet Hardness Tester Type TBH 250 5 DEFINITIONS Nil 6 PRECAUTIONS 7 During the test parts of the body like fingers etc can be squeezed between the movable test jaw and fixed jaw Please put in the samples by tweezers or a similar tool To remove remaining tablet debris use the supplied brush During breaking the test samples fragments of the samples can spring off Use only protective glasses 7
38. INITIONS Tempered suppository The suppository whose temperature has been adjusted reset to measuring temperature Thermostat A device that monitors temperature and automatically maintains it at certain levels 6 PRECAUTIONS Before operating the thermostat the water bath must be filled with water to the marked filling level Otherwise the thermostat can be damaged Do not touch the dry heating rods on the thermostat You may burn yourself Make sure that the water bath has reached the necessary operational temperature before you start a test In case of any failure the housing should only be opened by a qualified staff Before opening the housing pull the mains plug 7 PROCEDURE FOR OPERATING THE ERWEKA SUPPOSITORY HARDNESS TESTER SBT 2 SAMPLE PREPARATION The suppositories to be tested should be rested for at least 24hrs at the required measuring temperature before starting the test Usually this temperature is at 25 C e Measure the actual temperature in the measuring chamber using a thermometer If required adjust the temperature in the measuring chamber using the thermostat PERFORMING TESTS Set the tester temperature to the required value by using the thermostat e Put the tempered suppository into the holding bracket that is made of red plastic with a tip pointing up Carefully add the suspension rod assembly King Faisal University 7 KS Place the testing a
39. Kingdom of Saudi Arabia Ministry of Higher Education King Faisal University College of Clinical Pharmacy p j aanas j i ia Y S DLAI E 2012 King Faisal University College of Clinical Pharmacy PO Box 400 Al Ahsa 31982 Kingdom of Saudi Arabia King Faisal University College of Clinical Pharmacy Kingdom of Saudi Arabia Ministry of Higher Education King Faisal University College of Clinical Pharmacy King Faisal University College of Clinical Pharmacy SI Particulars Page No Number TSO CUCU iy dr sipetce seater incundiascareudinndescavtateameqenmyeutivniyaaasaue 3 2 College Vision Mission Values and Goals ceeeeeeee eens 4 3 Program Mission Goals and outcomes ecceeeeeeee eee eeeee 5 4 Standard Operating Procedure cece eee e ence ence tees eeneees 6 134 i Atomic absorption Spectrophotometer 0cccccceeeeeeneeenees 6 il CHN AN GIN Ze sachin ss oihcunewepslanantatadendade lt iesyaesexeatens ieee 10 iii Differential Scanning Calorimeter 15 iv Disintegration Apparatus 66 ccc ccc cece ene tenet eens 28 v Dissolution Apparatus eeeeeeeeseesesessss sesers 32 vi Freeze Diye seients A EEE E 35 vii Fourier Transform Infrared SPeCtrOSCOPY cc cece evee ee nnnes 39 Vill Flying spot Scanning Densitometer 0000 c ccc cence eens 44 ix Gas Chromatography Mass sp
40. MER LED is on then timer mode is selected If the TIMER LED is off amp Timer display shows then manual mode is selected TIMER key toggles between TIMER mode and MANUAL MODE In timer mode two time ranges are possible 1 sec to 99 min amp 59 sec OG min to 9 hrs amp 59 min if this range is selected the forth left most digit of timer display will show For changing the timer range press SET key first and the ENTER key Once the test is started range cannot be altered After range is selected now set the desired time as mentioned Step No 8 Timer settings can be done only in timer mode To set the timer follow the instructions given below Press SET key SET LED will start glowing Press TIMER key Timer display shows last set value with right most digit flashing To set the digit go to step c Press A key Flashing digit value will show increment for every key press OR Press W key Flashing digit value will show decrement for every key press re Set the value for flashing digit 7 KS Press DISP SEL next significant digit will flash Now repeat step c to set the value Repeat step d to set all digits Press ENTER to accept the value 2 KSA Press ENTER key to complete the timer setting Set LED goes off Start the test using START STOP key TIMER display shows the elapsed time starting from 00 000 if the timer range is min
41. Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document re Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of BX41 2011 Dr Tasleem System Microscope 1 OBJECTIVE The following document describes the procedure on the operation of BX41 System Microscope 2 SCOPE Microscopes are used to study living organisms and their fundamental processes 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Olympus Instructions Instruction manual for BX41 System Microscope AX7474 5 DEFINITIONS King Faisal University College of Clinical Pharmacy The resolving power of a microscope is the distance two objects must be apart and still be seen as separate and distinct 6 PRECAUTIONS Do not place the microscope on a flexible surface as this could result in blocking gt the air vents and cause overheating or a fire To prevent obstruction of the natural convection based air cooling of the microscope make sure to leave at least 10 cm of free space between walls
42. PT 2 Titrator allows access to the manual control dialogues of the MPT 2 Autotitrator This menu item will be greyed out if the MPT 2 is not connected Engineering allows Malvern authorized service engineers to perform maintenance tasks The engineering screens are password protected End of Document King Faisal University College of Clinical Pharmacy PO Box 400 Al Ahsa 31982 Phone 00966 3 5817175 00966 3 5800000 Extn 1783 Fax 00966 3 5817174 E mail clinicalpharmacy kfu edu sa
43. aceutical Sciences College of Clinical Pharmacy KFU Document History Modified poeumens Details of Amendments Date by Section Initials SOPs First Draft on SOP for the operation of TBH 2011 Dr Harsha 250WTD for thickness 1 OBJECTIVE Thickness is one of the main control features of a tablet press the other is weight control and speed A tablet that is too thick may not fit into the counting device a tablet that is too thin may break Tablet thickness is a critical attribute and should be controlled with accuracy 2 SCOPE Thickness is an important physical attribute On the press thickness variation may mean there is a variation in weight and hardness or both 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure King Faisal University College of Clinical Pharmacy 4 REFERENCES Nil 5 DEFINITIONS Nil 6 PRECAUTIONS Nil 7 PROCEDURE FOR OPERATING THE TABLET THICKNESS TESTER AVERAGE WEIGHT OF A TABLET e Check the cleanliness of the balance e Collect Sample of 20 tablets e Weigh the tablets record the weight in BMR e Calculate the average weight of tablet e Calculate the mean average weight of tablets of average wei
44. aks below 80 Noise Level e Default absorbance is 0 01 e Default transmittance is 2 Recalculate generates a new peak table Range Set specifies the range in which the peaks will be generated Selecting a Single Peak e Place cursor over the peak of interest e Click on the peak King Faisal University College of Clinical Pharmacy 7 4 Shutting down the instrument Close the Hyper IR 1 51 Software Remove the KBr pellet holder Turn off the computer End of Document King Faisal University College of Clinical Pharmacy 8 SOP FOR FLYING SPOT SCANNING DENSITOMETER Prepared by Document Number Name Signature Date CCP008 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document Secon Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh Flying Spot Scanning Densitometer 1 Purpose The following document describes the procedure on the Operation of the Shimadzu Flying Spot Scanning Densitometer 2 Scope This instrument is used for qualitative and quantitative analysis of DNA Protein Pharmaceuticals etc 3 R
45. ame settings This is useful if the parameters of the measurement record are not already saved in the SOP directory View Menu This selects which reports will be shown in the measurement file window and which Toolbars are to be displayed Workspaces will select the workspace toolbar and which measurement workspace is to be displayed The reports available are those selected when creating the workspace with the Configure Workspace dialogues The reports available will change to match the workspace chosen Select Status bar to toggle the status bar from being displayed or not Configure Menu Use this menu to create or edit the measurement settings in an SOP prior to performing the measurement New SOP opens the SOP creation wizard while Existing SOP allows a previously created SOP to be altered Measure Menu Select this menu when ready to perform a measurement There is a choice of using an existing measurement SOP or manually setting up the measurement and sample details Once the measurement details have been entered or an SOP has been chosen the Measurement display will appear Tools Menu With this tool estimates of the Molecular weight and particle shape can be generated Instrument allows access to the manual controls of the instrument or each connected accessory Serial port allows the USB communications port to which the instrument is King Faisal University College of Clinical Pharmacy connected to be chosen M
46. arams Printing a Spectrum M2 Click boxes 3 and 4 under quadrants The boxes are on the right side of the menu Select preview to display what will be printed Select Print Shutting down the instrument gt gt gt Remove cuvette Close PC 1501 program Turn off instrument End of Document King Faisal University College of Clinical Pharmacy 17 SOP FOR SPECTROPHOTOMETER Prepared by Document Number Name Signature Date CCP017 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document A ie Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh Spectrophotometer 1 Purpose The following document describes the procedure on the Operation of the Shimadzu UV Visible Spectrophotometer 2 Scope This instrument is used to measure how much light of a given wavelength is absorbed by a liquid sample 3 Responsibilities It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of the students technician
47. armacy e Open panel door and place the cuvettes with blank solution in the cuvette holders e Make sure to use the appropriate orientation for the cuvettes you re using e Also make sure that the cuvettes used for the auto zeroing are the same cuvette e Press the Baseline 3 Instrument 1 550 000 nm 0 002 Abs 100 Auto Zero WE Baseline GoTo wt Start sSPisconnect Starts a measurement on the current active instrument e When the Auto Zero is complete open the panel door and remove the front cuvette e Do not replace cuvette in rear holder e Using the same cuvette style fill an empty cuvette with about 2 ml of the sample e Clean the cuvette with a lab tissue e Place in front cuvette holder using the appropriate orientation and close the panel door e Press START to take a reading Saving the Spectrum e Select spectrum from the drop down menu e Press save as and give name and save 2 KSA Open panel door and remove test sample from front cuvette holder To test additional samples Place cuvettes in front holder and press start for a reading Press RETURN to bring you back to step Note This will erase your old data Remove cuvettes remaining in holders Then click on the Disconnect button near the bottom of the screen Exit from the UV Probe software King Faisal University College of Clinical Pharmacy Push the mode button on the UV VIS instrument keypad
48. ations or problems that may occur while performing the procedure King Faisal University College of Clinical Pharmacy 4 References Shimadzu instruction manual IRPresetige 21 FTIR 8400S Operation guide Shimadzu Instructional manual FTIR 8400S User s system Guide 5 Definitions FTIR Fourier Transform Infrared KBr Pellet Absorbance of IR by Solid samples are measured in KBr pellets Which is prepared by pressing mixture of 1 100 solid sample KBr 6 Precautions Regularly Check the desiccant and change if more than 30 has converted in to pink color 7 Procedure for operation of FTIR 7 1 Preparation of KBr Pellet for solid samples 2 Dry IR grade KBr in a drying oven for at least 1 hour 0 Allow the KBr to cool in a desiccator Weigh out 300 mg and place in a marble mortar and pestle and grind to fully powder it C KS Carefully weigh 3 mg of the finely ground solid sample and place in the marble mortar and pestle with the KBr 0 Sd Grind for 60 seconds to thoroughly mix 0 Sd Place the KBr compound mixture into an evacuable die on a hydraulic laboratory press Press in vacuo at 15 000 pounds of pressure for 6 minutes Release the pressure remove the die from the press disassemble the die and remove the KBr pellet Place the KBr pellet in a pellet holder and put it into the sample beam of an IR 7 2 Operation of Software Turn on the FTIR by pressing O
49. button to switch the heater ON OFF Pump button to adjust the pump rate Spray button to switch spray function ON OFF Spray button to change the relative spray rate Pump button to switch the pump ON OFF Temp button to change the inlet temperature 9 Selection knob to change process parameters and Menu button to enter the firmware menu structure instrument configuration Peristaltic Pump selection Four digits 1 4 are selectable 1 Clockwise pump rotation at standard speed 2 Clockwise pump rotation at double speed 3 Pump rotation direction alternates every 3 seconds 4 Pump rotation direction alternates every 6 seconds Spray selection Select a spray head output between 0 and 100 spray rate This directly effects the throughput of the spray head Temperature selection Set the drying gas inlet temperature between 0 C to 120 C Note The lowest possible drying gas inlet temperature in open loop mode depends on the local room temperature Operating controls King Faisal University College of Clinical Pharmacy Multifunctional buttons The multifunctional buttons below the screen are programmable Their actual use is displayed on the screen directly above the buttons Configuration gt Calibration Info gt 1 Go back one step Escape 2 Move cursor upwards 3 Move cursor downwards 4 Enter next menu level OK confirm selection Rotary knob The rotary
50. can be recovered carefully with a spatula or a brush Recommended set of tools for particle recovery Buchi particle scraper order no 11055338 Buchi weighing paper in A4 format order no 11055339 Particle recovery procedure Place a blank sheet of weighing paper on a free area inside a fume hood Remove the collector cylinder from the bottom basement of the particle collector Place the collector cylinder on the sheet of paper King Faisal University College of Clinical Pharmacy Use a particle scraper to collect particles from the cylinder Lift the cylinder from the weighing paper and put it aside Fill the powder from the sheet of paper into a storage vessel Depending on the substance a particle recovery rate of up to 90 can be achieved Note Act carefully e g avoid shaking movements in order to prevent loss of particles Clean all contaminated parts thoroughly to avoid hazardous situations and cross contamination For information about cleaning see section 7 6 USER MANUAL Using the Nano Spray Dryer Records software on a PC The PC software enables online data monitoring and storage of experimental runs in a virtual library Functionality Documentation of process data of experimental runs gt Export of process parameters for further data analysis Connecting the instrument to the PC Connect the Nano Spray Dryer B 90 to the PC by means of the USB cable Insert the CD int
51. ce from the sequence drop down menu Create new sequence table by filling the required information Than save the sequence table Than press run method 7 5 Calibration Open offline chem Station software Select load signal from open sequence file drop down menu Select MSD Or DAD signal and new calibration table and press OK To add second point select load signal from Press add the level and feed the required information King Faisal University College of Clinical Pharmacy To calculate the unknown concentration load the unknown concentration signal and click report 7 6 Tuning of MSD Once in a month Select ATunes from View MSD tune dropdown menu Tune autotune with dual polarity 7 7 Shut down Save the required information and click off instrument To switch off the MSD Select MSD vent form the View Diagnosis Maintenance Note Switch off the MSD only if you are not using more than 15 days kS Than close the chem Station software Switch off the HPLC system up to down degasser pump auto sampler DAD o gt End of Document King Faisal University College of Clinical Pharmacy 12 SOP FOR MICROSCOPE Prepared by Document Number Name Signature Date CCP012 Dr Sahibzada Tasleem Rasool Assistant Professor Department of Biomedical Sciences College of Clinical Pharmacy KFU Revision
52. ctive you can click on it to start your run King Faisal University College of Clinical Pharmacy 7 5 Data Analysis When the data acquisition is complete click on Top icon on the left hand e Real Time side to move up one level Then click on the Data Analysis icon Button to launch the data Analysis program The last data file acquired automatically comes up If you wish to view a Data Acquisition different file just double click the one in the list you want C KS Click and drag a box around peaks you want to zoom in on C2 KS Alternatively you can open the data file by clicking on GCMS Postrun analysis Which is on the desktop 0 Sd From the file dropdown menu open data file display the file you want Processing C To display a mass spectrum of a chromatographic peak right click the mouse anywhere in the chromatogram to obtain a menu Then left click on Average Spectrum to get a cursor Left click and drag the cursor across the chromatographic peak and its mass spectrum will be displayed 7 6 How to acquire batch runs of large numbers of samples optional To set up the batch table Click on the Batch Processing Icon After the batch table has been set up put your samples into the auto sampler carousel and click on the Start icon to start the batch 7 7 How to shut down the instrument when you are finished Make sure that no data acquisition is
53. cy 7 4 Data acquisition Acquisition 4 Be sure to remember to put your vial in the autosampler carousel You need to use a 2 ml autosampler vial with a Teflon lined cap such as Fisher Scientific part 03 391 15 The minimum volume of solution should sufficient so that the solution level will be high enough for the syringe needle to enter it p Z When you put the sample in the carousel please note the number of the well in which you place the vial p Make sure that the solvent rinse vials contain the solvent you are using p Z On the left hand side click on the Sample Login icon to enter sample information Sample Login C GCMSsolution Data Project1 Jim_Splitless qgm x Linet Acquisition Information Sample Name GCMS test solution Sample ID GCMS test solution Data File GEMS test 5 17 2011 qad Ss Baseline Data a Data Description P Method Detail id Snapshot Sampler Vial H f Injection Volume 2 uL Syringe Capacity Multi Inj Times j Tuning File Ic GCMSsolution System T une _default qat Li Advanced Cancel Help e The tuning file should remain as the default 4 Guide e Specifies where the vial is located in the autosampler e Enter a data file name and a location for the file Click on the Standby icon to download your acquisition method parameters to the instrument When the Start icon becomes a
54. d to 1 2 l min to prevent solvent evaporation from the condensers Extraction e Attach the thimbles to the adapter Fig 5 2 Fig 5 2 Fig 5 3 e Weigh the sample and transfer quantitatively to the thimble e Put a defatted cotton wool plug on top of the sample e Move the knobs to the boiling position and insert the thimbles Fig 5 3 Lower the knobs to rinsing position the thimbles will now hang just below the condenser valve e Press down heating plate handle and insert the extraction cups tare with glass beads 5mm and loaded with 50 70 ml of the solvent to be used Fig 5 4 The cups are now clamped into the condensers s Fig 5 4 King Faisal University College of Clinical Pharmacy e Start the program on the Control Unit by pressing the key The buzzer signal starts when the set temperature minus 5 C is reached Press the timer key and move the extraction mode knobs to the boiling position Thimbles will now be immersed in the solvent e Boil with the thimbles immersed Time depending on sample and solvent type Make sure that the condenser valves are open Fig 5 4 e Move the extraction mode knobs to the RINSING position when the buzzer signal starts The thimbles will now hang above the solvent surface Press the timer key l Solvent Recovery Removal of Thimbles and Extraction Cups e After rinsing close the condenser valves by turning a quarter turn Fig 5 4
55. damaging it Do not attempt to use organic solvents to clean the microscope components other than the glass components To clean them use a lint free soft cloth slightly moistened with a diluted neutral detergent Do not disassemble any part of the microscope as this could result in malfunction or reduced performance King Faisal University College of Clinical Pharmacy When not using the microscope keep it covered with a dust cover 10 PRECAUSTIONS If the microscope is used in a manner not specified by this manual the safety of the user may be imperiled In addition the equipment may also be damaged Always use the equipment as outlined in this instruction manual End of Document King Faisal University College of Clinical Pharmacy 13 SOP FOR NMR SPECTROMETER Prepared by Document Number Name Signature Date CCP013 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document i let Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Bruker 14 3 2012 Dr Mahesh 400MHz NMR 1 OBJECTIVE The following document describes the procedure on the operation of Bruker 400MHz NMR
56. e and create a new program Use the arrows to set the conditions for running your samples Use the End key to save your settings Running samples 2 Reaction volumes between 20 200u1 must be used and tubes used must be able to withstand a pressure of 1 atmosphere at 100 C All reagents and samples must be in ice before and after cycling Use arrow keys to select the program you want to run Duration of run will show on the screen Machine will turn off automatically at the end of selected no of cycles If you need to run samples overnight please inform the Technologist in charge To turn off the instrument 7 4 1 Switch off after use by switching to O at the back If you experience any problems during the use of this equipment please report to the Technologist in charge immediately End of Document King Faisal University College of Clinical Pharmacy 25 SOP FOR ZETASIZER Prepared by Document Number Name Signature Date CCP025 Dr Azza Ali Hasan Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document le Section Details of Amendments Date by Initials SOPs 2011 Dr Azza First Draft on SOP for the operation of Zetasizer
57. e employs a UIS Universal Infinity System optical design and should be used only with UIS eyepieces objectives and condensers for the BX2 series Some of the modules designed for the BX series are also usable For details please consult Olympus or the catalogue Less than optimum performance may result if inappropriate accessories are used 8 PROCEDURE FOR OPERATING THE BX41 SYSTEM MICROSCOPE Stabilize the BX41 System Microscope following all precautionary measures explained above K Sd Set the main switch to I ON and adjust the brightness C2 Select the light path trinocular tube C2 Place the specimen on the stage 0 Sd Engage the 10X objective in the light path 0 Sd Bring the specimen in focus S27 Kd Adjust the interpupillary distance C KS Adjust the diopter 0 Sd Adjust the light axis 0 BX Adjust the aperture iris and field iris diaphragms e Engage the desired objective in the light path and bring the specimen in focus C2 Insert the required filters 0 6 Adjust the brightness 0 Sd Start observation 9 MAINTENANCE AND STORAGE Clean all glass components by wiping gently with gauze To remove fingerprints or oil smudges wipe with gauze slightly moistened with a mixture of ether 70 and alcohol 30 The filter mount lens on the base is made of plastic Do not rub it with a strong force to prevent
58. e of Clinical Pharmacy 7 2 Creating a Method Click the Method drop down menu Click LC Set Up Make sure the instrument is isocratic mode Click the Mode drop down menu Make sure Isocratic is checked Time Length of your method Changes solvent system ratios when needed Adjust flow rate Percent Concentration Adjusts the graph limits Click parameters Adjust flow rate and Adjust pressure Click OK Click the File drop down menu Click Exit 7 3 Saving the Method Click the file drop down menu Click save as Enter desired name of the method Click Save 7 4 Acquisition Set Up Click the set up icon Select Sample Frequency Ensure your frequency is 1 unit greater than your sample requires Enter Run Time Default is 5 minutes s Save the method again 7 5 Running the sample Select the run single icon Has a blue arrow pointing to the right amp Enter sample identification name Select the desired method Select from browse button Select the file destination Select from browse button ka Click start 7 6 Injecting the sample A blank should be run first Make sure the knob is turned to load Insert sample using a blunt tip syringe 7 7 Quickly turn the knob up to inject the sample King Faisal University College of Clinical Pharmacy 7 8 Select analyze icon This will display retention times and peak areas 7 9 Shutting down the inst
59. ectrometer 0 c0cce cee ne ees 47 x High Performance Liquid Chromatography 000 0000e0e 53 xi Liquid Chromatography Mass spectrometer 00c 00eeeee 57 xii MCT OSCOME xc disincsecmuiudundiescmment ine n reni ia 6l xiii NMR spectrometer cathe cinhsvGubvatawpvenieaineiecieeee 65 xiv Nano Spray Dry e iccisiipriirsi taranii iieri tirari aanne A 70 XV Serum Protein Electrophoresis 82 xvi Spectrofluorophotometer ox se clenidoveraesdcauiewesasbhagnivseasesaaod 85 xvii SDECIFOPNONOMELCT conics ueacdsneredaciantsdaredaceadtsdeuensneasepeaaes 89 xviii Soxtec Extraction LHI cc voc cslcaacusemncewtacedsmscbieeeuaraeaaseoast 93 xix Suppository Hardness Tester 101 XX Tablet Abrasion and Friability Tester 0 0000 c cece cece eeees 104 xxi Tablet Thickness Tester orton evens sucnasesq ep taeeneeetanphereseserans 111 xxii Tablet Thickness Weight Diameter and Hardness Tester 114 xxiii Tapped Density TOSIEl soit donvaveupced acne eesanesecenteesaeaaveatacss 123 xiv Thermal CVE essere ee cena vesesk canes E RE ETEA 126 XXV ZElASIZET recois nii iei aTa EEEE E A ba Red Seka EER bake 129 King Faisal University College of Clinical Pharmacy INTRODUCTION The Standard Operating Procedures SOPs for instruments include the standard protocol to be followed for the proper and efficient functioning of instruments during the operation by students faculty lab technicians and other researchers of the
60. el may be King Faisal University College of Clinical Pharmacy accommodated with a staggered start allowing a minimal start time interval between vessels and is recommended if possible 7 Observations Carefully record any visual observations of the dissolution test such as basket wobble air bubble formation or the condition or motion of the disintegrating tablet particles 7 KS Sampling Withdrawal and filtration of the sample aliquot conclude the test interval After 30 minutes of testing and with rotation continuing withdraw a portion of the dissolution medium from each vessel about 30 mL Sample from a zone midway between the surface of the dissolution medium and the top of the basket or top of the paddle blade not less than 1 cm from the vessel wall 7 hS Filter each sample immediately through a syringe filter 0 45 um PVDF type or equivalent discarding the first 5 mL portion of filtrate Note Equivalent filters will provide filtrate with analytical response within 1 of the unfiltered solution Discard volume is a function of the filter and may not be identical to that recommended in this document Cool the filtered dissolution samples to ambient temperature prior to UV analysis sample and standard solutions are at the same temperature Do not centrifuge the sample A separate unused filter cannula and clean syringe should be used for each vessel 7 KSA Automated Sampling 7 If
61. er option is selected 2 KS Enter either the number of taps up to 999 999 or the time duration in hh mm ss format as applicable and press ENTER Once entered the test duration is saved by the unit and does not need to be reset unless you want to change it To change the duration enter the new value and press ENTER Pass the product through a number 18 sieve 1 00 mm sieve opening if appropriate and weigh the product for each cylinder separately Step 5 Load the product into the cylinders and level it without compacting if necessary Press START STOP to begin The time duration or number of taps count down on the display screen and the test stops automatically when 0 is reached The display automatically resets to the original set value preparing the instrument for the next test KS To calculate the tapped density use the following equation tapped density initial weight final volume Stopping a Test When running a test on a Tapped Density Tester you can stop the test by pressing START STOP The cylinders stop immediately and the display returns to the original test length setting END OF DOCUMENT King Faisal University College of Clinical Pharmacy 24 SOP FOR THERMAL CYCLER PCR Prepared by Document Number Name Signature Date CCP024 Dr Promise Emeka Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr
62. es College of Clinical Pharmacy KFU Document History Modified Document A ie Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of 2011 Dr Azza Disintegration apparatus 1 OBJECTIVE The following document describes the procedure on the operation of Disintegration apparatus 2 SCOPE The Disintegration apparatus used to explore disintegration mechanism and times of rapidly disintegrating tablets 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES King Faisal University College of Clinical Pharmacy Instruction Manual Operation Guide Erweka Company for disintegration apparatus 5 DEFINITIONS The Erweka disintegration tester is a 2 basket system 6 PRECAUTIONS If the equipment is used in a manner not specified by the manufacturer the protection provided by the equipment may be impaired 7 PROCEDURE FOR OPERATING THE DISINTEGRATION APPARATUS Ensure the calibration status is valid and make sure that the instrument is clean and free from dust Switch on the main switch and put on the ON OFF switch at the rear panel to ON position Follow the proced
63. esponsibilities It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of the students technicians to follow the SOP as described and to inform the instructor about any deviations or problems that may occur while performing the procedure 4 References Shimadzu Flying Spot Scanning Densitometer CS 9301PC instruction manual 5 Definitions King Faisal University College of Clinical Pharmacy TLC Tin layer Chromatography 6 Precautions N A 7 Procedure for operation of Densitometer Turn on the scanner ON by flipping the yellow switch on the side of the machine Automatically initialization starts and takes about 10 seconds Start CS 9301PC data processing software by double clicking the CS 9301PC icon on the desktop When communication between the scanner and the PC is established set back up parameters in the scanner to display the current coordinates wavelength and absorbance Click on the lt lambda gt pushbutton dialog box will open Reset the wavelength to required wavelength and click on OK Click on lt X go to gt will open dialog box Reset the X axis coordinate to 50 and click OK Use the same procedure to move the Y axis stage as well Specify measurement result output by clicking on lt config gt dropdown menu and selecting scan configura
64. f integral press left button and drag to define Right click under integral to select Adjust selected integral tails to be parallel with baseline by scrolling bias and then slope e To calibrate click under integral and select calibrate Click save return e Click pick peaks Make sure define region is selected With left button drag the box to include the top of all peaks to be picked To add a peak click and select peak with cursor and left click Click save return e To save Select Publish from the main menu and save by selecting PDF e To print Go to print on left hand e Options select Print with layout start plot editor e Press OK 7 4 To turn off the instrument e First eject sample by clicking on the acquire main menu and selecting Turn on sample lift air ej e Remove the sample and inject again by selecting Turn off sample lift air ij e Click on left corner icon yellow e Press terminate TopSpin and press OK End of Document King Faisal University College of Clinical Pharmacy 14 SOP FOR NANO SPRAY DRYER Prepared by Document Number Name Signature Date CCP014 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia Shariff Asst Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document
65. ght of each in Process check CHECKING OF INDIVIDUAL WEIGHT VARIATION e Collect sample of 20 tablets e Weigh the tablets record the weight in BMR Calculate the average weight of tablet e Weigh 20 individual tablets one by one and record the observations in the BMR individually King Faisal University College of Clinical Pharmacy e Check the lower weight and upper weight for determination of lower and upper percentage deviation e Upper deviation Average weight of 20 tablets Upper weight x 100 Average weight of 20 tablets e Lower deviation Average weight of 20 tablets Lower weight x 100 Average weight of 20 tablets e After checking initial parameters if it is within the limit compression machine can be started ABBREVIATIONS e BMR Batch Manufacturing Record END OF THE DOCUMENT King Faisal University College of Clinical Pharmacy 22 SOP FOR TABLET THICKNESS WEIGHT DIAMETER AND HARDNESS TESTER Prepared by Document Number Name Signature Date CCP022 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia Shariff Asst Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document Details of Amendments Date byt Section Initials SOPs First Draft on SOP for the operation of TB
66. hould be discarded The seals themselves are somewhat more forgiving in that a defective seal reduces the vacuum However defective seals should still be removed from service A greater risk is present in seals that connect two pieces of glass All seals of this type place rubber between joined pieces of glass If these seals are used improperly the glass pieces come into contact and scratch each other during installation These scratches compromise the integrity of the glass thus creating a potentially serious implosion hazard Install any seals of this type carefully to avoid this if you are uncertain of how to do this do not proceed until you have received further training from one of the contact people given above 7 2 Most vacuum pumps are supplied with a gas ballast adjustment It is recommended that the gas ballast be closed during the operation of the Freeze Dryer Push the RUN button 2X The first time you press the run button the displays a reminder to check the system 2 KS The second press activates the freeze drying run The Condenser will get below 40 C CTVAC within 10min CTtim 2 KS 2 KS After 10 minutes of operation the vacuum pump will start Vacuum will go to 500mT VACRDY within 10min VACtim If NO alarm will sound and VACU gt 500mT after 10min will flash on the 7 KS 7 KS lower part of the screen If YES then READY FOR FREEZEDRYING will flash at the top of the
67. ical field in the particle collector is turned on King Faisal University College of Clinical Pharmacy Optimizing parameters The relevant parameters for the spray drying process depend on each other List of spray drying parameters drying gas flow rate relative inside pressure inlet temperature feed rate The properties of the spray dried solution determines the maximum process temperature The maximum thermal load of the solid particles is caused by the spray head temperature or by the maximum inlet drying gas temperature To reduce the thermal load on the feed the sample vessel maybe cooled The achievable droplet sizes depend on the installed spray cap Mean values for water applications are shown in the following table Table 6 1 Spray nuts Spray cap Mean water Approx water Span hole size droplet size D throughput 2x Pw w 40m Bum 20mh lt 14 o 55m 15pm 6omhn lt 16 7Oum 21pm 150ml h lt 1 6 End of spray drying process A Caution Risk of minor or moderate burning when handling hot parts e Do not touch hot parts e Let the system cool down for some minutes after use Finish a spray drying process Switch off the heater by pressing the Heat button and let the system cool down Depending on the environmental temperature this may take several minutes Switch off the spray head by pressing the Spray button Stop the gas flow at an inlet temperature below 70 C Read the internal p
68. ick on the Auto Zero button F King Faisal University College of Clinical Pharmacy 7 4 Place the suction tube in your 0 concentration blank standard When the Abs level has stabilized click the blank button Measure the absorbance of standards and unknowns according to the rows To do this select STD in the first column and press ok and place the suction nozzle in one of the solutions and aspirate Once the Abs level has stabilized click on the Start button The instrument will wait 5 seconds and then collect data for 5 seconds C The results will then be displayed in the Measurement Results Table E MRT Repeat step 4 for all the std solutions Check the calibration curve displayed on the right upper of the window D To measure the sample in the MRT select UNK and than repeat step 4 pa Note down the concentration of unknown solution from MRT To turn off the instrument Press the Extinguish button on the front of the instrument Close the main tank valve on the acetylene tank Purge the line of acetylene by pressing the Purge button Close the air valve which is below the bench Make sure that both the meters are showing zero reading One on the acetylene gas cylinder and one below the bench of AA instrument Od Turn off the power to the AAS Exit out of the software NOTE DO NOT LEAVE WITHOUT CLOSING THE ACETYLENE TANK ACETYLENE IS EXPLOSIVE AND
69. ick the GCMS Real time Analysis icon 2 KS C2 KS jik Log in window will open and press OK no password to start Lab solution Data software Acquisition After you login the data acquisition software will automatically come up C Click the vacuum control Icon to open the vacuum control windows Data Analysis C KS Click the auto start function to start vacuum system 0 Sd Wait for at least four hours to get good vacuum 0 KS Select the peak monitor window and check the MS back ground Processing C2 After good vacuum is achieved go for Auto tuning And click auto tuning icon cg on the assistant bar System Configuration k Tuning 0 Sd Check the Auto tunning result If all ok you are ready for analysis 7 3 Creating the method Already created by you You can change or edit the Method file to suit your u experiment If you change the method please save it as your method by Vacuum Control clicking on File and Save Method File As After the data acquisition software comes up click on File and Open Method w in order to choose and open a data acquisition method From method drop down menu select system selection and tick DI for direct injection operation and GC for GCMS operation To create new method There are 3 tabs for entering information in the method e The first tab has parameters for the auto injector Sampler e The second tab has
70. ight be hot and internal pollution can cause loss of performance 7 PROCEDURE FOR OPERATING THE AA SPECTROPHOTOMETER gt Turn on the power switch on the front of the DSC TILA Turn ON the power switch for the DSC After Windows comes up double click the LOG ON icon The LOG ON window will come up Type in your username amp password Log into Windows King Faisal University College of Clinical Pharmacy x Pyris LogOn User Name Password pee After you login the Pyris software will load You will see a menu bar at the top of the screen which will indicate that the DSC 7 is Offline Click on the DSC 7 icon to bring it online The DSC 7 window will come up Duato be sved ae See eB Ht i You should see temperature status information for the TGA 7 if everything is working properly King Faisal University College of Clinical Pharmacy DSC psy 25 000 C e Fill the DSC with ice water with the metal funnel in place gt First we need to run a blank spectrum to verify that the DSC was found in good condition After you run your blank spectrum print it out and put your name on it and place it in the tray by the computer monitor Before you run the blank check to make sure that both the sample and reference furnaces are empty Make sure that both furnaces are empty before running the blank
71. knob is used to select values out of lists within sub menus e Turning the knob clockwise moves the cursor downwards e Turning the knob anticlockwise moves the cursor upwards View parameters menu e To enter the Menu press the corresponding multifunctional button below the screen Inside the menu use the multifunctional buttons for navigation e The View parameters menu and its sub menus provide information about the actual system status e It is accessible at any time and can supply the user with useful information about ongoing processes King Faisal University College of Clinical Pharmacy e No changes can be made here Parameters nlet temperature Jutiet temperature Spray head temp a concerti 35 mbar ov 0 ua V Volage V Current Parameters 100 timin O mbar closed as flows el pressure 2 he Valye Os Vay amp Glass detector vet vokage closed 230 VAC Note in closed loop mode the N and O valve open and close dapanding on the O content in the ges bop to maintsin the ssiactad O concentration lt 4 or lt 1 Configuration menu To enter the Menu press the corresponding multifunctional button below the screen Inside the menu use the multifunctional buttons for navigation to Configuration and confirm the selection King Faisal University College of Clinical Pharmacy Menu iew parameters Configuration 1
72. l settings in the input field move to the appropriate position with the lt CL gt key All data inputs are supported by a beep tone The beep tone confirms that the e command has been accepted and carried out Test Run Time e Two modes of operation are available either based on a time hours minutes and seconds or b number of drum revolutions Maximum time 9 hours 59 minutes 59 seconds Maximum revolutions 99999 e The pre set time is permanently retained in the memory until a new time setting is input e Press key lt F2 gt for G T to adjust time mode or revolution mode King Faisal University College of Clinical Pharmacy run time ih 15m 00s UPM 40 START GIT UPM QOO e Enter a test run time by means of the numeric keys starting with hours e g test run time is 1 hour 15 minute Press key lt 1 gt and confirm with lt ENTER gt e The current position being set is indicated by blinking Enter 15 and confirm with lt ENTER gt enter the seconds and confirm with lt ENTER gt run time UPM 40 START G T QOOO e Number of revolutions are entered by means of the numeric keys Confirm the settings with lt ENTER gt RPM e Driving speed is infinitely adjustable from 20 to 40 min e Press key lt F3 gt to adjust a speed value e Enter the requested speed by means of the numeric keys and press lt ENTER gt to confirm The instrument returns t
73. le You must then click on File and Save Data in order to save the calculated results with To convert a spectrum to ASCII format click on the Edit menu and click on Copy Then go to a text program such as Word and click on Edit and Paste The sample data points and the acquisition parameters will be displayed in text format and can then be saved How to shut down the DSC King Faisal University College of Clinical Pharmacy A First run another blank spectrum to show that you have left the DSC in good condition Please print out the blank graph put your name on it and put it in the tray near the computer monitor KA Click on the Start Pyris button and then click on Close All This will close the software A Log out of Windows Turn off the DSC TAC Thermal Analysis Controler 7 KSA a FOR THE DSC Tum off ne zo the power switch on the DSC TAC Turn off the power switch on the DSC mmm ose switch on the DSC Turn off the nitrogen gas macd End of Document King Faisal University College of Clinical Pharmacy 4 SOP FOR DISINTEGRATION APPARATUS Prepared by Document Number Name Signature Date CCP004 Dr Azza Ali Hasan Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Scienc
74. lets corresponding as near as possible to 6 5g is to be used whilst for tablets with a unit mass greater than 650mg a sample of 10 whole tablets is to be used e According to the British Pharmacopoeia 100 rotations are considered to be the standard number of rotations to be used for each test e According to the British Pharmacopoeia a maximum loss of mass not greater than 1 0 per cent is considered acceptable for most types of tablets 7 PROCEDURE FOR OPERATING THE AA SPECTROPHOTOMETER King Faisal University College of Clinical Pharmacy Tablet Friability Abrasion Testing Friability Drum left 4 JU Knurled Nut Foil Keypad Collecting Vessel Mains Switch Fuse Mains Input Socket Switching on King Faisal University College of Clinical Pharmacy e Switch the instrument on by means of the on off switch located on the rear panel e Before operation the instrument must be at the correct operation temperature Switch on the instrument approx 10 minutes before starting a test Touch sensitive Keypad ERWEKA TAR 1 0 ALPHA NUMERIC ERWEKA e Operation will be done by an alpha numeric touch sensitive foil keypad with LCD display located in the front of the instrument To enter figures following keys are available 000 090 The F1 F2 and F3 keys have the functions as indicated on the LCD Display A data input or confirmation will be completed with the UEnter key To correct actua
75. linical Pharmacy King Faisal University College of Clinical Pharmacy 20 SOP FOR TABLET ABRASION AND FRIABILITY TESTER Prepared by Document Number Name Signature Date CCP020 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia shariff Asst Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document me Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Tablet 2011 Dr Harsha Abrasion Friability Tester 1 OBJECTIVE The mechanical strength of tablet can be determined by its hardness and through friability test In friability test the tablets are prone to abrasion hence enabling us to check for the tablet strength under application of force in different manner 2 SCOPE Friction and shock are the forces that most often cause tablets to chip cap or break The friability test is closely related to tablet hardness and is designed to evaluate the ability of the tablet to withstand abrasion in packaging handling and shipping 3 RESPONSIBILITIES King Faisal University College of Clinical Pharmacy e It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to
76. macy practice education and research through commitment to community engagement MISSION OF THE COLLEGE To excel in pharmacy education patient centered care community engagement and research VALUES Based on Islamic Principles the College of Clinical Pharmacy promotes an environment of mutual respect and collaboration where we value v v v v v Excellence Pursue exceptional quality and performance in all that we do Responsiveness to community Lifelong self directed learning Creativity Integrity at all levels of teaching training research and patient care COLLEGE STRATEGIC GOALS 1 Prepare Pharm D graduates through excellence in Pharmacy education and training 2 Recruit develop and retain internationally recognized competent faculty members in all the desciplines 3 Continue to develop administration that will provide leadership for achieving strategic goals 4 Maintain and enhance the quality assurance management verified through accreditation by National and International Accriditting agencies in Pharmacy Education 5 Engage the local national and international communities for mutual benefits through expanding the opportunities for collaboration continuing professional development and postgraduate education 6 Conduct exceptional research in basic and clinical sciences King Faisal University PROGRAM MISSION Pharm D To Prepare Pharm D graduates who will provide
77. mes blood plasma vaccines etc To achieve a chemical balance such as for biological reagents To provide a practical solution for certain delivery problems for example the packaging of constituents that cannot be mixed in the liquid state but which are solidified in successive stages and then freeze dried To implement an important stage of a product such as concentration To improve storage life and improved marketing of the pharmaceutical product King Faisal University College of Clinical Pharmacy 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES Instruction Manual Operation Guide Copyright 2009 Millrock Technology for Millrock Bench Top Freeze Dryer 5 DEFINITIONS Biological materials typically must be dried to stabilize them for storage preservation or shipping Many times drying causes damage and some loss of cellular or protein activity Lyophilization called freeze drying is a method of sample preservation that significantly reduces damage to biological samples 6 PRECAUTIONS Wear Appropriate Eye Protection At All Times When Working With Or Anywhere Near A Lyophilizer For Medical Emergencies Fires Or Explosions Call emergency
78. n Off button which is in front of the Instrument Click on the FTIR icon 0 KS e A dialogue box will appear e Press OK e Double clicking the FTIR Box located in the bottom left corner Select Measuring Mode e Transmittance T used for qualitative IR analysis King Faisal University College of Clinical Pharmacy e Absorbance abs used for quantitative IR analysis The absorbance is directly proportional to the sample concentration e Interferogram IFG used for special purposes e Power Spectrum used for special purposes Select Resolution e The recommended resolution for liquids and solids is 4 or 8 e The lower the resolution the longer the run time of the sample Select Range e The desired range is 4000 cm to 500 e A range between 400cm and 350cm will display undesired water peaks Select Number of Scans e The number of scans ranges between and 4000 e Use more scans is the sample has a low concentration is a thin film or if the FTIR accessory decreases light to the detector e 20 scans at a resolution of 4 with a mirror speed of 2 8 only takes 20 seconds e The more scans the better s n ratio C View Scans e Leave on Auto e Controls the real time display of the spectra when scanning Scan Drop Down Menu e Initialize Initialized optical compartment e Sample Scan Starts sample measurement e Back Scan Starts background measurement File Name e
79. n SOP measurement select Measure Start SOP from the menu bar and select an SOP to use With an SOP chosen the Measurement display will appear below The measurement will begin by pressing the Start _ button 7 2 Manual Measurement Close the lid and Turn on the instrument and wait 30 minutes for the laser to stabilize Start the Zetasizer software Prepare the sample following the sample preparation guidelines Choose the cell s appropriate for the sample and measurement type Fill the cell s with the prepared sample Make an SOP measurement If necessary Open or create a new measurement file Select Measure Start SOP from the Zetasizer software Select the SOP required and select Open Follow any onscreen instructions that appear The Measurement display will now be shown When requested insert the cell into the instrument and wait for the temperature to stabilize Click Start _ The measurement will be made and the results displayed and saved to the open measurement file 7 3 Soft ware Operation Menu bar The menu bar contains the main menu headings for all software functions Items which end with a row of dots will cause dialogue boxes to appear Similarly any items which end with an arrow _ will cause a second menu to appear Items that are shown in grey indicate that they are not available Grey items indicate that some security setting has been activa
80. n this procedure and to ensure adherence to this procedure King Faisal University College of Clinical Pharmacy It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES AA 6300 Instruction Manual Operation Guide for Shimadzu Atomic Absorption Spectrophotometer Installation Procedure of WizAArd control software for Shimadzu Atomic Absorption Spectrophotometer 5 DEFINITIONS Blank solution Solution having only reagents used for eliminating absorption by reagents STD Standard solutions 6 PRECAUTIONS Don t open the flame window when flame is on Do not leave without closing the acetylene tank when not in use acetylene is explosive and quite dangerous 7 PROCEDURE FOR OPERATING THE AA SPECTROPHOTOMETER 7 1 Turning on the AAS Turn on the computer Double click on the WizAArd shortcut on the desktop The Operation tab should be highlighted Click on the picture of the AAS Enter admin for the user id and hit ok No password The program opens with wizard selection Select the element selection Icon and press OK Element selection window will appear First click on the select Element Than load parameters will appear In this select the required element eg K potassium and press OK and then finish Make sure that the Acetylene gas line pressure should be between
81. nsibility of the students technicians to follow the SOP as described and to inform the instructor about any deviations or problems that may occur while performing the procedure King Faisal University College of Clinical Pharmacy 4 References Shimadzu solvent delivery module LC 20AT prominence Shimadzu HPLC user s manual 5 Definitions HPLC High performance Liquid Chromatography 6 Precautions gt Always use guard column to avoid clogging of HPLC analytical column gt Filter the mobile phase with 0 45um filter paper 7 Procedure Degassing Solvents Ensure there are enough degassed solvents before running a sample Get a 500 mL or 1000 mL Erlenmeyer filter flask Place your solvent and a stir bar into the flask and sit it on a hot plate Hook up a vacuum line and stopper the top of the flask Turn on the vacuum and the stir plate DO NOT HEAT The solvent is degassed when bubbles no longer appear Place parafilm over the filtering arm and parafilm the top of the flask after you put the tubing in 7 1 Starting the Instrument Turning on the Instrument e Turn on the Column oven CTO 10A e Turn on the Liquid Chromatographs LC 10AS e Turn on the UV Detector SPD 10AV Starting the Computer e If the computer is off long in to the system e Open the program by Clicking on the HPLC Icon e Click OK No password e Click Shimadzu LC Icon King Faisal University Colleg
82. nslated Erie Se Be Canine Coli Lo i G G 1 Qvertem perature setting 2 Set for overtemperature 3 Program number 4 Temperature setting 5 Display 6 Hot Plate temperature 7 Plus 8 Set for program a Minus 10 Hot Plate close down go 4 11 Start Stop program 12 Timer 13 Air 14 Manual heating 15 Time for recovery step 16 Time for rinsing step 17 Time for boiling step 18 Locking device for change of overtem perature setting Fig 5 1 The display on the Control Unit 2045 USER MANUAL 8 PROCEDURE FOR OPERATING THE SOXTEC 2045 King Faisal University College of Clinical Pharmacy Control Unit Settings Press the POWER key on the Control Unit switch should light up and the displaywill show the version number of the internal software before going to standby e g showing the previously run program Select a program 1 to 9 for analysis by pressing Alor Check that the over temperature setting is correct for the solvent used in the Application Sub Note Note If the over temperature setting is not correct or if there is no recommendation in the table check with the laboratory supervisor that the correct setting is used If the application is being run for the first time and it is necessary to set up a new program see next section Setting Up a New Program Use the relevant Application Sub Note and Table 3 page 16 to set the correct temperatures and extraction times fo
83. nstruction Manual of Shimadzu Spectrofluorophotometer RF 5301PC series Installation Procedure of control software for Shimadzu Spectrofluorophotometer 5 DEFINITIONS Excitation wavelength Emission wavelength 6 PRECAUTIONS N A 7 PROCEDURE FOR OPERATING THE AA SPECTROFLUOROPHOTOMETER Starting the Instrument Turn on the computer Turn on the instrument The on off switch is located on the lower right corner Starting the software Double click the icon PC 150X Soft ware Operation To open the menu Select acquire mode drop down menu Ensure spectrum is checked Select configure drop down menu Select spectrum parameter Spectrum Parameters Spectrum Type C Excitation EX Wavelength nm EM Wavelength Range Start End Sensitivity High C Low Recording Range Slit Width nm Low EX 10 C20 High 50 000 EM 10 20 Response Time Repeat Scan Auto File Cancel Select spectrum type Scan Speed Super Fast C Medium C Slow When excitation is selected the emission grating is in a fixed position When emission is selected the excitation grating is in a fixed position Enter desired parameters King Faisal University College of Clinical Pharmacy amp Excitation Emission wavelength 0 220 900nm Excitation Emission wavelength range 220 900nm The minimum scanning range is 25 Sensitivity can be set to high o
84. nt Personnel without this training or persons who are in training require thorough instruction These operating instructions are the basis 4 REFERENCES Operation Manual Nano Spray Dryer B 90 Version A 5 DEFINITIONS Dehumidifier Air humidity changes occurring between different seasons and regions affect the spray drying process By using the Dehumidifier B 296 the drying conditions are exactly defined and made reproducible by removing the water content with a cold trap at 5 C The integrated hygrometer displays the air humidity Aspirator The drying air is sucked or blown through the device by the aspirator motor creating under pressure conditions By regulating the aspirator speed the amount of heated drying air can be increased or decreased 6 PRECAUTIONS Please read these operating instructions carefully and act accordingly The wearing of personal safety equipment such as safety glasses protective clothing is necessary The glass parts are to be visually checked regularly for possible damage stars cracks The following uses are expressly forbidden Use of gases with unknown chemical composition Spray drying of biohazardous materials such as viruses or bacteria Spray drying of organic solvents in open loop and without the Inert Loop B 295 King Faisal University College of Clinical Pharmacy Installation or use of the instrument in rooms which require ex protected instr
85. number Immediately And Report The Details Of The Incident If solvents are part of the product being freeze dried they may be flammable or hazardous Proper precautions should be taken Acids and bases may cause damage to the stainless steel and vacuum pump If the gas ballast is left open for extended periods the oil can be pumped out the exhaust causing the pump to fail Do not fill the chamber above the vacuum port as water will enter the vacuum tubing and drain into the vacuum pump Do not chip the ice from the condenser or chamber If acid or base products are collected immediately neutralize and dispose of properly 7 PROCEDURE FOR OPERATING THE Freeze dryer 7 1 Inspection of glassware amp seals King Faisal University College of Clinical Pharmacy Use only glassware designed for the lyophilizer The lyophilizer when in good working order should have a pressure of 5 50 millitorr normal atmospheric pressure is 760 torr as indicated on the LED screen If the pressure is outside of this range contact one of the people listed above This means that all glassware attached and under vacuum on the lyophilizer represents a significant implosion hazard All glassware used on the lyophilizer must be free of any visible defect cracks chips or scratches no matter how seemingly minor Any glassware that is defective in this way must not be used under any circumstances If such glassware is found it s
86. o the PC and follow the descriptions on the CD for system requirements and software installation End of Document King Faisal University College of Clinical Pharmacy 15 SOP FOR SERUM PROTEIN ELECTROPHORESIS Prepared by Document Number Name Signature Date CCP015 Dr Sahibzada Tasleem Rasool Assistant Professor Department of Biomedical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document re Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Biotec 2011 Dr Tasleem Fischer Serum Protein Electrophoresis 1 OBJECTIVE The following document describes the procedure on the operation of Biotec Fischer Serum Protein Electrophoresis 2 SCOPE Serum Protein Electrophoresis examines specific proteins in the blood called globulins Serum protein electrophoresis is used to identify patients with multiple myeloma and other serum protein disorders 3 RESPONSIBILITIES e It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFEREN
87. o the main menu King Faisal University College of Clinical Pharmacy Start run time 1h 15m 00s UPM 30 START G T UPM OOO e Press key lt F1 gt to start a test run nominal ih 10m 00s actual Oh 15m 10s UPM 29 9 STOP OOO e Pre set test run time nominal elapsed test run time actual and actual speed value are indicated on the display e Press key lt F2 gt to stop a test run TEST END e On completion of the test run time the instrument stops King Faisal University College of Clinical Pharmacy e Press key lt FI gt to start the emptying function whereupon the drum runs three revolutions in reverse automatically emptying the contents of the drum into a collecting vessel Maintenance e Routinely check drums for signs of cracks or chips e Inspect electrical connections for signs of corrosion and damage e Immediately wipe any spills from the exterior of the equipment using a moistened clean cloth e Clean drums thoroughly after each use using deionised water and dry with a soft cloth END OF DOCUMENT King Faisal University College of Clinical Pharmacy 21 SOP FOR TABLET THICKNESS TESTER Prepared by Document Number Name Signature Date CCP021 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia Shariff Asst Professor Department of Pharm
88. or other objects and the left right and rear sides of the microscope and the lamp socket when installing the microscope To avoid potential shock hazards and burns when replacing the light bulb set the main switch at OFF then disconnect the power cord from the wall outlet in advance Make sure the voltage selector switch 3 located on the rear side of the microscope frame is set to conform with the local supply voltage When shipped from the factory the switch is set to 220 240 V Always use the power cord provided by Olympus If no power cord is provided please select the proper power cord by referring to the section PROPER SELECTION OF THE POWER SUPPLY CORD at the end of this instruction manual If the proper power cord is not used product safety and performance cannot be guaranteed Always ensure that the grounding terminal of the microscope and that of the wall outlet are properly connected If the equipment is not grounded Olympus can no longer warrant the electrical safety and performance of the equipment Never insert metallic objects into the air vents of the microscope frame as this could result in electrical shock personal injury and equipment damage The power cord may melt by the heat of lamp socket if the cord comes in contact with it Distribute the power cord at an enough distance from the lamp socket 7 INTRODUCTION King Faisal University College of Clinical Pharmacy This microscop
89. r low Recording range limit of the graph allowable range is 100 to 1000 Repeat Scan allows automatic consecutive scanning from 2 100 scans Press OK Running an Experiment 7 Loading a Sample KS Flip door down 2 Insert a cuvette Close the door Click on the start button 2 When the scan is completed the file name dialogue box appears 7 We KS 2e Type in a file name o gt 7 Maximum of 8 characters 7 KS A comment may be typed into the dialogue box Select presentation drop down menu Press graph Press radar Both axes Zoom In Click a drag a box around the area you wish to enlarge 7 To return to full screen 7 KS Right mouse click ae Select limits ae Press reset 2 Press ok King Faisal University College of Clinical Pharmacy Cross Hair Coordinates 2 2 KS 2 KS 7 2 KSA SC PS SC 7 2 2 kS Right mouse click Select cross hair Select display To remove the cross hair coordinates press display again Select presentation drop down menu Select plot to display the plot layout window Under the graph column Click on the first box in row If it is already checked uncheck the box then recheck the box The default for plot data is screen this will print what appears on the screen Select Ok Select box under P
90. r the method First select a program with the keys Then use the key and the Al keys to set temperature and times for the different extraction steps Q When finished wait for the display to stop blinking or press the key until you have passed the last extraction step The recovery time should be set according to experience of the time needed for recovery of the solvent to the collection vessel A typical value is between 5 and 10 minutes depending on solvent used and temperature The fourth time can be used to keep the hot plate warm for up to 59 minutes if more analyses are to be performed later Cooling Water Cooling water must be supplied to the system before the analysis starts Adjust the flow rate to 1 2 l min 20 C Pre heating The hot plate can be pre heated prior to the analysis by pressing the s key The hot plate will warm up to the set temperature for the program shown in the display Preparation of the Sample King Faisal University College of Clinical Pharmacy To get correct results the pre preparation of the sample is very important For further information see Application Note 301 USER MANUAL Some types of samples e g meat should be mixed with sand or celite Follow the same procedure as for traditional Soxhlet Starting Up e Press the MAINS button switch lamp should light up e Open the cold water tap for the reflux condensers The flow should be adjuste
91. rear of the chamber Press the DEFROST button The compressors will direct hot gas to the condenser coil until it reaches 40 C then cycle off If the presence of cold water is detected the controller will repeat the process until all ice is melted End of Document King Faisal University College of Clinical Pharmacy 7 SOP FOR FOURIER TRANSFORM INFRARED SPECTROSCOPY Prepared by Document Number Name Signature Date CCP007 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document A le Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh FTIR 8400S 1 Purpose The following document describes the procedure on the Operation of the Shimadzu FTIR 2 Scope This instrument is used to determine the functional groups present in the organic compounds 3 Responsibilities It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of the students technicians to follow the SOP as described and to inform the instructor about any devi
92. removed during the operation of the machine All chemicals used as reagents in PCR reaction must have safety storage procedure There is a risk of burns if the PCR block is inadvertently touched when hot Clean machine after use if there has been sample spillages by wiping with a damp soapy cloth While cleaning care should be exercised to prevent water from entering the inside the unit Before cleaning always disconnect the power supply and allow machine to cool below 50 C Do not use abrasive cleaners The key pad must not be pressed with sharp objects such as pens pencils and finger nails as will cause damage 7 PROCEDURE FOR OPERATING THE TC 412 THERMAL CYCLER PCR MACHINE Turning on the PCR machine Switch on the machine at the rear with symbols I mains switch ON and O mains switch OFF ka The machine will then display the opening four line display King Faisal University College of Clinical Pharmacy The top three lines are information or program lines while the bottom line is the prompt line The prompt figure in the line will flash The arrow keys will indicate selection of f while lt indicates delete This arrow key J indicates enter and is used to enter operation or edit mode It is alsgused to accept a field or program change The symbol is the end key To into programs which will be flashing press Enter This will show a screen for new program with Templates Copy a templat
93. ressure of the Nano Spray Dryer B 90 No overpressure should be present Take out the pump inlet tube of the sample solution The pump will now transport air displacing product deposits in the tubing and spray head Turn OFF the peristaltic King Faisal University College of Clinical Pharmacy pump afterwards and lower the pump bed to release the mechanical load of the pulleys at the tube e See section 6 8 USER MANUAL before open the glass assembly in safe environment prior to any particle recovery Collecting the particles from the collector cylinder Death or serious poisoning by inhalation or incorporation of dried particles at recovery Wear safety goggles Wear safety gloves Wear a suitable protective mask Wear a laboratory coat Do not inhale dried particles Stop drying gas flow before opening the drying circuit Only recover particles in sufficiently ventilated flue or glove box Do not disperse the dried particles Do not clean dusty parts with compressed air General information about particle recovery To recover the particles from the wall of the collecting electrode a particle scraper and collecting paper can be used It allows recovering the particles manually and deposits the particles on a sheet of weighing paper The electrostatic particle collector has very high separation efficiency Nevertheless some particles may be deposited on the glass cylinder wall and the inner star electrode These particles
94. rument K KS C2 KS Wash the column with 50 water and 50 methanol or Acetonitrile for 10 min To Stop the pump change the flow rate to 0 ml min Click on the pump icon Turn off the Column oven Turn off the Liquid Chromatographs Turn off the UV Detector Exit from the HPLC software Shutdown the computer End of Document King Faisal University College of Clinical Pharmacy 11 SOP FOR LIQUID CHROMATOGRAPHY MASS SPECTROMETER Prepared by Document Number Name Signature Date CCP011 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Document EER 5 Modified Seton Details of Amendments ate by Initials SOPs First Draft on SOP for the operation of Shimadzu NOIL Dr Mahesh Spectrofluorophotometer 1 OBJECTIVE The following document describes the procedure on the operation of Agilent LCMS 2 SCOPE Liquid chromatography coupled with mass spectrometry detection LC MS is one of the most powerful analytical tools for organic compound analysis The key advantages of using LC MS methods over HPLC methods include Selectivity Co eluting peaks can be isolated by mass selectivity and are not constrained by chromatographic
95. s amp heating rate If you want to add a cooling step click on Add a step and click on Temperature Scan Tgm e B e E0 E thrin Dua Sapling Opions Select value frosnar A eC Tell the DSC to go to your starting temperature DSC Control Panel ut Click on the Go to Temp button o o God tlt When it is At Temp monitor Fie Edt Wew Toots Window Help lalelo olele slaj x5 we At Temp 25 000 C Heat Flow should be stable to 0 001 mW the Heat Flow Wait until the Heat Flow is stable to 0 001 mW Then click on the Start Run button King Faisal University College of Clinical Pharmacy Click on the Start Run button To monitor the progress of the run you can click on the Instrument Viewer window DSC Data Analysis To calculate the onset temperature DSC Lea 50 0 c tel peak area and enthalpy click on the Calc menu and click on Peak Area Click on the option to include Onset Position the X s at the start of the peak and at the end of the peak Then click on Calculate Click on Include Click and drag the X s to the start of the peak u and the end of the peak ii After you click on Calculate the Onset temp peak top area and AH are calculated and displayed You can click and drag the results to position them anywhere on the graph was 1s 155 gt the raw data fi
96. s to follow the SOP as described and to inform the instructor about any deviations or problems that may occur while performing the procedure 4 References King Faisal University College of Clinical Pharmacy Shimadzu UV Visible Spectrophotometer instruction manual PharmaSpec UV 1700 series Operation guide Shimadzu UV Visible Spectrophotometer instruction manual PharmaSpec UV 1700 series User s system Guide Shimadzu Instruction Manual UV probe tutorial 5 Definitions Cuvette A small transparent quartz vessel This is what a sample is put into for testing in the UV Visible Spectrophotometer 6 Precautions N A 7 Procedure for operation of Spectrophotometer Open panel door and make sure cuvette holders are empty then close the panel door Flip open the display unit and Turn spectrophotometer ON by flipping the yellow switch on the side of the machine The machine will automatically initialize Connecting UV spectrophotometer to software e Turn on the system e Click the CPS1670 icon e On the instrument press F4 e Press connect in the system Instrument will automatically connect to computer software e O oe e XNE N 2 10l Auto Zero Wi Baseline gt Go To vw K Connect SDisconnect For Help press F1 lActive Spe e Hold the cuvette from the top to prevent tampering with the measurements and wipe the sides with a lab tissue King Faisal University College of Clinical Ph
97. sample furnace and then put an empty aluminum pan into the reference furnace Sample should be centered in the sample furnace Va We An empty aluminum pan which is on the left side gt j should be centered in the reference fumace which is on the right side Carefully put the platinum lids on top of each furnace then secure the swing arm cover over the furnaces and then put the dome lid on top Platinum lids are shown in place Use tweezers to gently put them in place Please do not bend or deform the lids Swing amm cover is shown in place Push gently down on the lever to lock it in place King Faisal University College of Clinical Pharmacy Dome coveris shown in place Click on i Enter sample ID Browse to navigate information here gt LJ to the folder you want to put the j data in and to give A it a filename King Faisal University College of Clinical Pharmacy Method Editar Uniitied Se Initial Statetab gt n Oaa tobe saveda nam 1110S Set your starting ie ee a foie a ad Ea foo ge a a gue the level for the Y Yiri 1000 ft lasera foo baseline in mW ana Traa E l po gw AeA Aaa oe Osace Gergen Sirisd Sanglelrio baalSute Progam Method File Hame Urttled 1D Inun standsd Sorrel By clicking on each L ns step you can highlight the step to change its temperature
98. sec 0 00 if the timer range is hrs min In any mode to START or to STOP the test press the START STOP key Warm distilled water to 37 2 C and take in 1000ml beaker The volume of the water should be such that the wire mesh of the moving basket at its highest point is at least 25mm below the surface of the liquid and at its lowest point is at least 25mm above the bottom of the beaker King Faisal University College of Clinical Pharmacy gt KeS 7 2 KS 7 Place the beaker in the disintegration test apparatus Suspend the basket assembly in to the beaker by fixing it in to the moving shaft Introduce one tablet in each of the six tubes of the basket of the assembly and place the disc in each Press START STOP button and note the time Run the instrument for required period and switch off Lift the basket and observe for completion of disintegration Cleaning Empty the beaker and clean Remove the discs and clean the disintegration test apparatus End of Document King Faisal University College of Clinical Pharmacy 5 SOP FOR DISSOLUTION APPARATUS Prepared by Document Number Name Signature Date CCP005 Dr Azza Ali Hasan Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU
99. ssembly into the testing chamber and start the time counting using a stop watch The measurement starts The initial load determined by the weight of the complete rod assembly is 600g After one full minute of testing add an additional weight of 200g one disc weight After one minute add the next weight Continue the procedure until the suppository breaks under the corresponding total weight Check the time of the test run on your stopwatch EVALUATION OF TEST RESULTS If the suppository collapses under the initial weight of the suspension rod assembly i e 600 grams it has a strength value of less than 600 grams implying that the suppository is too soft and therefore not usable When the suppository collapses within the range of added weights the strength value is calculated as the total of the suspension rod assembly 600 grams plus the value of the added weights 200 grams each The value of the last weight added before the suppository collapses depends on the moment at which the collapse occurs If the collapse occurs within 20 seconds from the moment the weight is applied then the value of the weight is rated as zero If the collapse occurs within 20 to 40 seconds from the appliance of the weight the value of this weight is counted as 100 grams If the collapse occurs after 40 seconds from the weight application then the full value of 200 grams is included into the total applied load sum End of Document College of C
100. t and press OK oe To perform peak analysis select edit peak param from peak drop down menu Select CH1 and click OK To get peak information select peak list from peak drop down menu gt wy End of Document King Faisal University College of Clinical Pharmacy 9 SOP FOR GAS CHROMATOGRAPHY MASS SPECTROMETER Prepared by Document Number Name Signature Date CCP009 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document H Seti Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh Spectrofluorophotometer 1 OBJECTIVE The following document describes the procedure on the operation of Shimadzu GCMS QP2010 Spectrometer 2 SCOPE GC MS analysis where the effluent to the GC instrument is the feed to the MS instrument is in wide use for confirmation testing of substances Drug testing manufacturing quality control and environmental testing are some typical uses K2 compounds This is instrument is useful for qualitative and quantitative analysis of volatile organic King Faisal University College of Clinical Pharmacy
101. ted or that the item is not relevant to the system connected File Menu The File menu is used to Open or create a New measurement file The measurement file is where all the measurement records results will be stored Select Save as to store a measurement file under a different file name Once a measurement file has been created use Export to export the measurement details to other software packages such as Excel or Wordpad Select Batch print to simultaneously print a number of measurement records King Faisal University College of Clinical Pharmacy Create PDF is only enabled if the 21 CFR part 11 feature key has been installed As a shortcut a list of the most recently used measurement files are shown at the bottom of the menu for immediate opening Edit Menu Allows movement and manipulation of records in the Measurement file window s Records can be cut copied pasted and deleted into their own or other measurement files Edit result allows an existing measurement record to be reanalyzed using different dispersant and particle properties Comments on the editing reason added The edited measurement will then be added to the bottom of the Records view listing To view the settings for any particular measurement record select the record and then Extract SOP The SOP dialogues will appear showing the original measurement settings These can then be saved as an SOP so measurements can be made again with the s
102. the sphere that diffuses at the same speed as the particle being measured The Zetasizer system determines the size by first measuring the Brownian motion of the particles in a sample using Dynamic Light Scattering DLS and then interpreting a size from this using established theories Brownian motion is defined as The random movement of particles in a liquid due to the bombardment 6 PRECAUTIONS Failure to follow these guidelines could result in exposure to hazardous voltages and laser radiation The system is capable of heating the cell to high temperatures Care should be taken when removing the cells if a measurement has been performed at high temperatures It is recommended that the cell area is allowed to cool before removing the cell A warning triangle is provided in the cell basin Due to the risk of melting polystyrene cuvettes must not be used for measurements above 50 C 7 PROCEDURE FOR OPERATING of Zetasizer 7 1 SOP measurements A Standard Operating Procedure SOP is like a template that pre defines all the measurement settings This ensures that measurements made on the same type of sample are made in a consistent way SOPs are ideal if the same type of sample is regularly measured inputting the same parameters each time a measurement is made is tedious and runs the risk of making errors SOPs can be created or modified as required King Faisal University College of Clinical Pharmacy To perform a
103. this procedure e It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES e Instruction Manual Operation Guide Tablet Friability Abrasion Testing TAR 200 Erweka Germany e British Pharmacopoeia 2009 Vol 4 London The Stationary Office 2009 P A438 A439 5 DEFINITIONS e Abrasion Drum Consists of standardized internal plastic baffles used to test the abrasion strength of the tablets e Friability Abrasion Tester Consists of a drive unit that rotates a transparent plastic drum at a standard rpm rotation per minute e Friability Test Provides a standardized reproducible method for measuring the tendency of a tablet to chip or powder when subjected to the mechanical stresses of the manufacturing packaging and transportation processes of the tablet 6 PRECAUTIONS e Please pay attention that no article of the clothing or long hair will come into the rotating drum or the drum shaft e Do not open the electrical compartment due to the risk of shock and allow only trained personnel to do so e When tightening the locking nut ensure that it is not over tightened since this can cause damage to the drum e Do not use cleaning compounds containing ammonia or abrasive cleaners to clean the plastic drum since such compounds may damage plastic causing it to crack and fracture e For tablets with a unit mass equal to or less than 650mg a sample of tab
104. tion 7 1 Execute as single lane scan Click on lt scanner gt in the main menu and select parameters Click on lt change gt from the submenu and click on control parameters Make sure that SINGLE is selected for lane Press START button o gt OG Channel selection dialog box opens select CH1 2 When file save dialog box opens save the file with file name eight alphanumeric characters 7 2 Execute auto lane scan Click on lt scanner gt in the main menu and select parameters o gt Click on lt change gt from the submenu and click on control parameters Make sure that AUTO is selected for lane Press START button Channel selection dialog box opens select CH1 When the autolane scan is ended use a submenu on the dialog box to save the file with file name 7 3 Display King Faisal University College of Clinical Pharmacy 7 4 Select lt File gt from the drop down menu Press on lt File selection gt Select the required file and click on lt LOAD gt Selection dialog box opens Click on the CH1 radio button and press OK Click on lt Present gt in the main menu select radar from the dropdown menu to optimize the display Read the data using cursor by clicking the right mouse button Peak Analysis Click on lt Peak gt and select Find Peaks gt KSA C Processing parameters dialog box open Set the width Slop and Drif
105. uld be set to 2 cm from the bottom White base ALWAYS SET DEPTH with depth gauge Failure to set depth will break your sample and or the instrument e Select Turn on sample lift air ej from Acquire main menu e Drop the sample tube on the red sample holder Top of the instrument e Select Turn off sample lift air ij e Wait till you hear tick sound Lock e Click on lock 7 e Solvents table will open e Select the solvent and press OK Tune e Click Tune on yellow and select Tune match probhead automatically e Wait till the message atma finished 2 KS Spin e Spin is required for 1D for 2D 3D no need to spin e Turn sample rotation on e Wait till red arrow appear on sample Shim e Click on shim yellow button e Wait till topshim complete appears Prosol e click on prosol Gain e Click on gain and wait till finish e Check the scan time if required change For 1H NMR set scan in multiples of 4 usually 16 scans GO King Faisal University College of Clinical Pharmacy e To start scan click GO e Click on Acqu from the inner window to see the FID 7 3 Data Process Saving and Printing e Click on Process main menu and select Proc spectrum Cale Spectrum AdjustPhases Caib Axis tPickPeakse f integrates Advanced Click Adjust phase Using left button scroll and to adjust phase Click save return Click integrate Click to define beginning and end o
106. uments Spray drying of substances which might explode or ignite due to the processing and the selected parameters of the Nano Spray Dryer B 90 or installed sub systems Use of samples which might produce oxygen during the processing 7 OPERATING INSTRUCTIONS Watch this video link for reference before starting http wn com spray_drying PROCEDURE FOR OPERATING THE NANO SPRAY DRYER B 90 Check that the voltage of the socket is the same as that given on the apparatus plate The apparatus is supplied with an power cord which is directly plugged to the apparatus Starting up the instrument 7 KSA Make sure the Nano Spray Dryer B 90 is properly connected to the mains supply Carry out a final installation check see section 5 9 REFER MANUAL before every spray drying process Switch on the Nano Spray Dryer B 90 at the main switch Switch on all other electrical devices of the actual configuration at their main switch Main screen kS kS 2 King Faisal University College of Clinical Pharmacy After the internal system check at bootup finished successfully the main screen view below appears at the Nano Spray Dryer B 90 Gas flow co 1 00 lzmin Inlet temp 120 Outlet temp Head temp 48 Pressure 3D rwar e amp dd gDc Nano Spray Dryer B 90 Display showing the current instrument status Heat
107. university college The following elements are included in the SOPs 1 Objective Describe the purpose of the process or procedure 2 Scope Express the applications of the process or procedure and any organization or regulatory requirements as well as any limits to the use of the procedure 3 Responsibilities Denotes the individual s responsibility and minimum experience the user should have to use the instrument satisfactorily 4 References Cited references used for preparing these SOP s and for further detailed information 5 Definitions Any acronyms abbreviations or specialized terms used are defined in this segment 6 Precautions Indicate the activities which could result in equipment damage degradation of sample or possible invalidation of results and critical steps in the procedure 7 Procedure All pertinent steps to be followed are described in sequence along with the materials required to accomplish the goal Each student faculty lab technician or any researcher handling the instruments is expected to read and be familiar with each section of the SOP This is intended to be a dynamic document and will be revised periodically Mention of trade names or commercial products does not constitute endorsement or recommendation for use King Faisal University College of Clinical Pharmacy VISION OF THE COLLEGE To be recognized nationally and internationally for preparing role models in Phar
108. ure e Calculate the gel with Turboscan densitometer NOTE DO NOT LEAVE WITHOUT PROPERLY SWITCHING OFF THE UNIT AND CLEANING THE GEL MOUNTS End of Document King Faisal University College of Clinical Pharmacy 16 SOP FOR SPECTROFLUOROPHOTOMETER Prepared by Document Number Name Signature Date CCP016 Dr Mahesh Attimarad Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Ibrahim Alhaider Dean Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document i Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Shimadzu 2011 Dr Mahesh Spectrofluorophotometer 1 OBJECTIVE The following document describes the procedure on the operation of Shimadzu Spectrofluorophotometer 2 SCOPE Spectrofluorophotometer is used for qualitative and quantitative analysis of fluorescent organic and inorganic compounds 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure King Faisal University College of Clinical Pharmacy 4 REFERENCES I
109. ure for CHNS Eurovector instruments and software Callidus Software Interface Version 5 1 5 DEFINITIONS Bypass Un weighed standard samples are introduced to clean and condition the entire analytical circuit prior to sample analysis 6 PRECAUTIONS Each time leak test has to be performed You must not proceed data acquisition until the leak test has passed 7 PROCEDURE FOR OPERATING THE AA SPECTROFLUOROPHOTOMETER Starting the Instrument Open the valves of the Helium and Oxygen gas cylinder and set the pressure to 4 bar 60 psi Turn on the instrument The on off switch is located at the rear of the instrument Turn on the computer and to Start the software Double click the icon Callidus software gt Soft ware Operation From the main menu click Operator login Enter user name EVR and Password EA3000 and click on the OK button Create method for CHNS From the main menu select Instrument Setting In the method frame type the instrument parameters as reported in the Analytical Protocol received with EA amp Press the save as button to store the parameters in the methods library Click on the send button to load the new parameters on the instrument The front furness and GC oven will star to heat up immediately h Wait for the instrument to condition until the instrument status shows READY EuroEA 3000 Status Ready Instrument Analysis Sample Manager Operator Password
110. ure to set the temperature For setting the temperature select the BATH probe by pressing PROBE key This key toggles between BATH probe EXTSEL Press SET key LED to the right of the key will glow Press TEMP key Temperature display shows previous set temperature with right most digit flashing 7 Press A key Flashing digit value will show increment for every key press OR Press W key Flashing digit value will show decrement for every key press Set the value for flashing digit Press DISP key Next significant digit flashes Repeat the above step Repeat the step c amp d till the desired temperature is set Press ENTER key Set temperature value will be registered and SET LED will be off To confirm the set value of temperature press SET key first and TEMP key next See the set value Press ENTER key to come out from set mode NOTE After pressing the SET key one should start the settings within 10 seconds otherwise the set mode will be terminated automatically Press the TEMP key to start temperature controller The LED above TEMP key will start glowing Select the bath temp indicator using PROBE SEL key Once the bath temperature reaches the set value you are ready to start the test by pressing START STOP key 8 There are two modes of operation King Faisal University College of Clinical Pharmacy Timer mode b Manual mode a Select mode by pressing TIMER key if the TI
111. w Auto Fill Clear Table T Save changes to Library Method C No reports LAS Insert Row Copy Sel Balance pari EET Configure Data Processing Autorun m Calibration Type m Integration Parameters C KFactor Linear i P i L Start mee Peak Thr Peak Width Min Area Baseline een toad m Component Table 1 u 1000 Valley Element SWS 26s Sree sen Time Event Functions Retentiontime s 33 50 114 aot Window s 3 10 46 40 Start interval komt lede BEE Stand by at the end of Autorun Balance Interface automatic transfer of weights Connect the balance cable to COM part 2 of your computer Prep are the sample table without entering weights for standards and samples Click the balance button below the sample table Click on the sample table the row of the position to which the weight must be transferred Weigh the sample and wait for balance stability than press the transfer button of the balance The transferred weight appears in the sample table at the active cell amp The process is repeated with each subsequent standard and sample King Faisal University College of Clinical Pharmacy gt Naming Auto run C2 KS Click on the save button in the autorun frame Creat new autorun name window opens Click add type Type new name and click OK Running an Experiment When all the settings have been made
112. witch off e Turn off the cold water tap e Check that the condensers are free from solvent End of Document King Faisal University College of Clinical Pharmacy 19 SOP FOR SUPPOSITORY HARDNESS TESTER Prepared by Document Number Name Signature Date CCP019 Dr Arshia Shariff Assistant Professor Dept of Pharmaceutics College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Sreeharsha N Assistant Professor Dept of Pharmaceutics College of Clinical Pharmacy KFU Document History Modified Document Iei Section Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of ERWEKA 06 12 2011 Suppository Hardness Tester SBT 2 1 OBJECTIVE The following document describes the procedure on the operation of ERWEKA Suppository Hardness Tester Type SBT 2 2 SCOPE The following procedure applies to designated personnel in Research laboratory 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure 4 REFERENCES King Faisal University College of Clinical Pharmacy Instruction manual for operation of ERWEKA Suppository Hardness Tester Type SBT 2 5 DEF
113. ws from remaining tablet pieces Do not use any detergents END OF DOCUMENT King Faisal University College of Clinical Pharmacy 23 SOP FOR TAPPED DENSITY TESTER Prepared by Document Number Name Signature Date CCP023 Dr Sree Harsha Assistant Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Revision Number Approved by RO Dr Arshia Shariff Asst Professor Department of Pharmaceutical Sciences College of Clinical Pharmacy KFU Document History Modified Document A le Sacto Details of Amendments Date by Initials SOPs First Draft on SOP for the operation of Tapped 2011 Dr Harsha Density Tester SVM 202 1 OBJECTIVE The tapping action is generated by a rotating eccentric camshaft which lifts the platform and allows it to drop back to its original position 2 SCOPE Applicable to the Tapped Density Tester in Quality Control Department 3 RESPONSIBILITIES It is the responsibility of designated personnel in Research lab to train staff and students on this procedure and to ensure adherence to this procedure It is the responsibility of designated personnel staff or Student to follow the instructions of this procedure Each side of the dual Tapped Density Tester is calibrated to the corresponding platform Ensure the correct platform is placed into the proper sleeve left and right 4