1. Chart SeQual Technologies Inc 5 Part No 7040 ESSEQUAL ud Name amp Address of Manufacturer OXYGEN CONCENTRATOR INFORMATION SeQual s Integra E Z Oxygen Concentrators are quiet reliable and low maintenance sources of oxygen The Integra E Z is designed and constructed to consistently provide high concentrations of oxygen in a lightweight compact cabinet that is easy to maneuver Using Oxygen Concentrators Warning Don t change the flow rate without consulting your physician Warning Avoid alcohol and sedatives when using oxygen they can slow your breathing DANGER Never use petroleum based products Use water based lubricants to moisten lips or nostrils if needed No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Materials that burn in air and even some material that will not burn in air are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty DE Recommended Operating Environment Guidelines 50 F to 104 F 10 C to 40 C Electrical Use no extension cords or electrical outlets controlled by a switch No closer than 3 inches 7 6 cm from t
2. SS SEQUAL WARNINGS AND CAUTIONS Warning and Caution notices used in this manual apply to hazards or unsafe practices that could result in personal injury or property damage Federal law U S A restricts this device to sale by or on the order of a physician SeQual Chart SeQual Technologies Inc Read Instruction Manual Do Not operate this oxygen concentrator without first reading and understanding this manual If you are unable to understand the warnings and instructions contact your local SeQual dealer or authorized service representative before attempting to use this oxygen concentrator Caution or Attention In certain circumstances oxygen therapy can be hazardous therefore it is advisable to seek medical advice before using the oxygen concentrator Do Not deviate from prescribed flow rates Adjust the flow setting only to your prescribed flow rate unless otherwise instructed by your physician Warning No Smoking or Naked Flames The use of oxygen therapy requires that A special care be taken to reduce the risk of fire Material that burns in air and even some Q material that will not burn in air are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can perman
3. eSEQUuAL INTEGRA Oxygen Concentrator INTEGRA E Z INSTRUCTION MANUAL Model 6323A OM 10 Model 6323A 10 P N 7040 Rev B March 2011 SEQUAL TABLE OF CONTENTS Warnings and GE VIUIO ERE EE ER 3 l e es tos TOR EEE EE EE REE m 4 Cont NI 4 IDEO SUICIDE EEE EE EE EE E DEN MM ML LEAD MM 4 TegioT9 Bez auis Eee UCO RE EE EEE EE NE SE EN RE re 4 Symbols Used in Instruction Manual and on Oxygen Concentrator sisnnssrsesssrrrrrsrrrrrrrrrresrrererrrrrrrrrrrrr rer renen nan 5 ENN NNN 6 Using Oxygen EGNA eide 6 Recommended Operating Environment Guidelines c mnssssersssrrrrssrrrrrsrrrrrsrrrrrrrrrrrrrrrrrrrrrrr rer r rss rer rara 6 VPT 7 Before Operating Your Oxygen CONNOR 7 Operating Your OG JANE ONE Oase se rene blancs ERE RR KERNE ERR mach den REEL tdi orat 8 FREEZE OFEN fed po ERE NE NS RE uM pecado e EN 9 SS Geli EO OR ETT T 9 MENNENE 9 OMO EE EE EN EE 9 NNN 9 OR 9 POWER Management Teste seats tim tanti apa tca tradu eR IRE Potete ata onte RU E noA MAR tht aL S ioa eat daher ltd 10 PEN EON MR MERERETUR 10 COnToLPanel Di Olay EEE ONE SEMPER 10 Ber or OD hat Se OSE ER NE ud p D P Rao eae 11 VEE 11 KO ANE TE ER EE NE EN NE EE RE 11 MAS INTER 0 FE EEE EEE Seager Melos Sareea SNES SAS 12 Trouble Shooting Guide ANN 13 FP 14 Electromagnetic Compatibility ua ae Est Lis 15 Disposal ortEdgulpimebitandACCeSSOLG6S 0 6 piren Seok eto aeo dle ecco a occa estat n Mm usd GER 15 Chart SeQual Technologies Inc 2 Part No 7040
4. off target flow Ing if EUN Unit is Unitisoverheating Compressor Hlasking N73 Running Power Power interruption Internal Fuse Blown Fuse Blown On or On or These conditions are displayed only on units equipped with oxygen monitor 1 Call the Service Provider if abnormal conditions are detected and cannot be remedied During System Malfunction condition the Green and Yellow lights may be On or Off 3 Usage Use recommended oxygen delivery accessories Warning Use of certain administration accessories not specified for use with this oxygen concentrator may impair the performance 4 Non Usage When not using the oxygen concentrator press the On Off button to turn it off Chart SeQual Technologies Inc 11 Part No 7040 ESSEQUAL MAINTENANCE Routine Maintenance Routine maintenance consists of cleaning the air inlet filter at regular intervals to keep the unit functioning for years The user can easily achieve this maintenance When other than routine maintenance is needed contact your SeQual dealer provider or authorized service representative Warning Unplug the oxygen concentrator from the power outlet before cleaning or routine maintenance Use a damp cloth for cleaning Do Not directly apply liquid spray cleaners or aerosol cleaners e Cabinet and Power Cord The cabinet and power cord can be cleaned with a mild detergent solution applied with a damp cloth and towel dried Air Inlet F
5. Part No 7040 ESSEQUAL ELECTROMAGNETIC COMPATIBILITY This equipment has been tested and found to comply with the limits for medical devices to the IEC 60601 1 2 1994 Electromagnetic Compatibility standard These limits are designed to provide reasonable protection against harmful interference in a typical medical installation This equipment generates uses and can radiate radio frequency energy and if not installed according with the instructions may cause harmful interference to other devices in the vicinity However there is no guarantee that interference will not occur in a particular installation If this equipment does cause harmful interference to other devices which can be determined by turning the equipment off and on the user is encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving device e Increase the separation distance between the equipment e Connect the equipment into an outlet on a circuit different from that which the other device s are connected e Consult the manufacturer or service technician for help DISPOSAL OF EQUIPMENT AND ACCESSORIES Follow local governing ordinances and recycling plans regarding disposal of device components ul REP Medical Product Services GmbH Chart SeQual Technologies Inc Borngasse 20 2200 Airport Industrial Drive Suite 500 35619 Braunfels Germany Ball Ground GA 30107 USA CE Marked For Model
6. This manual serves as a supplemental reference to help you operate and maintain the unit If you have any further questions or problems call your SeQual dealer or authorized service representative General Precautions Retain the product s safety and operating instructions for future reference Follow all operating and usage instructions Observe all warnings on the product and in the operating instructions To reduce the risk of fire bodily injuries and damage to the oxygen concentrator observe all safety precautions Do Not connect to an extension cord or electrical outlet controlled by a switch No other appliance should be plugged into the wall outlet Voltage requirements are marked on the label on back of unit Important Read Important Safety Guidelines section of this manual before plugging in the power cord Do Not use the unit if the power cord is damaged Warning Improper use of the power cord and plug may result in a fire or electrical shock hazard e Check to see if the power cord is plugged in and there is power to the outlet e Check that the oxygen concentrator is not placed near a heat register Important SeQual s products operate safely when used according to their marked electrical ratings and product usage instructions The power cord used with the Integra E Z Oxygen Concentrator must meet the requirements of the country where the product is used Chart SeQual Technologies Inc 8 Part No 7040 SEQUAL
7. 6323A OM 10 Europe C E 0029 6323A 10 220 240V 50Hz Chart SeQual Technologies Inc 15 Part No 7040
8. INSTALLATION INSTRUCTIONS 1 Select a Location Caution Locate oxygen concentrator in a well ventilated space that provides adequate airflow and check if air is being prevented from entering the oxygen concentrator by furniture draperies or clothing 2 Inspect Inlet Air Filter Air filter cleaning instructions are located in the maintenance section of this m Caution Always check to see if the inlet air filter is clean 3 ON OFF Button Insert the plug into the electrical wall outlet Push the On Off button the green light will illuminate to indicate the unit is running 4 Warm up Period After initially turning on the unit allow three 3 to ten 10 minutes for the unit to reach its specified performance If the unit is equipped with an oxygen concentration monitor OM models only the red yellow and green light will intermittently illuminate while the unit is warming up For details on warm up refer to the Display of Operating Conditions table on Page 11 Note After an initial startup or after a power interruption the unit may automatically reduce the flow rate briefly before returning to the set flow rate 5 Flow Rate Adjustment Adjust the flow rate to prescribed setting supplied by your physician Press the Plus button to increase or the Minus button to decrease the flow rate until the proper level is displayed N Warning 7he prescribed flow rate setting should only be adjusted under the advice of a physici
9. an Chart SeQual Technologies Inc 9 Part No 7040 ESSEQUAL POWER MANAGEMENT FEATURE In the event of a Power Interruption N Warning Because the duration of a power interruption is unknown seek an alternate supply of oxygen during the power interruption Resume use of the Integra E Z when power is restored If a power outage occurs the Integra E Z will activate both an audible and a visual red light alarm to inform you that the Integra E Z is not operating The Integra E Z must be located where the alarms can be seen and heard If the duration of the interruption is shorter than twenty seconds and the unit has not been turned off the Integra E Z will resume operation twenty seconds after the initial interruption Important During the power interruption e The user can press the ON OFF button to silence the alarms and turn the unit off e If the unit was not turned off the Integra E Z will automatically resume operation once power is restored e If the unit was turned off during the power interruption the user must press the ON OFF button again after the power is restored to turn the unit on Operating Conditions l Control Panel Display If the concentrator is an OM Model equipped with an oxygen monitor the oxygen concentration is being continuously monitored while the Integra E Z is running and oxygen is being produced If the concentration level falls below preset levels an indicator light will activate Operatin
10. ecified for use with this oxygen concentrator may impair the performance e Alarm System The oxygen concentrator has an alarm system with a red visual alarm indicator light and buzzer The alarm system will activate in the event of an internal fault or power outage The red and or yellow light may stay on for up to ten 10 minutes after start up OM models only This is a normal condition during the initial warm up period f the oxygen concentrator is equipped with an oxygen concentration monitor OM model this monitor indicates if the oxygen concentrator is performing to the correct specifications At normal therapeutic oxygen concentration levels the green light is activated At lower than normal therapeutic oxygen levels the yellow light is activated The red light activates and the unit alarms if the system malfunctions or when service is needed The oxygen concentrator will continue to run at lower levels of therapeutic oxygen but may shut down under other system malfunctions If the red light turns on you should discontinue use of the oxygen concentrator and refer to the troubleshooting section of this Chart SeQual Technologies Inc 7 Part No 7040 ESSEQUAL manual or contact your SeQual dealer provider or authorized service representative to schedule a service call OPERATING YOUR OXYGEN CONCENTRATOR Your SeQual dealer provider or authorized service representative will demonstrate how to operate the oxygen concentrator
11. ently damage the device and void the warranty Use No Oil or Grease A spontaneous and violent ignition may occur if oil grease or SN other petroleum substances come into contact with oxygen under pressure Keep these substances away from the oxygen concentrator tubing and connections and any other CN oxygen source Do Not use any petroleum based or other lubricants unless recommended by SeQual Technologies Inc e Keep oxygen concentrator and cord away from hot surfaces Do Not use the oxygen concentrator if it has a damaged cord or plug contact qualified service personnel for examination and repair e Air intake of the oxygen concentrator should be located in a well ventilated area e Oxygen concentrator should be located so as to avoid pollutants and fumes e If oxygen concentrator has been dropped damaged or exposed to water contact qualified service personnel for examination or repair e Do Not drop or insert any object into any opening e Do Not block the air opening of the oxygen concentrator or place it on a soft surface such as a bed or couch where the bottom air opening may become blocked e Do Not cover unit with a blanket towel quilt etc e Do Not place objects pillow blanket under unit Maintain at least 72 inch clearance under unit e Do Not overfill the humidifier if used Fill with distilled water only to the level shown by the manufacturer e Warning Do Not remove the cover Cover
12. g conditions are described in the table on the following page Green Light Yellow Light Red Light LCD Flow Indicator LPM Increase Flow Button On Off Button Decrease Flow Button om MANPEN lest ru Serre No Sersohoare of Make Furre H dd rohu bor Mus wd Lote gur Bring Liv aho re to sko Soe prt rm burr de Tur Ch red cover wih a teres o pho obw ty set aker vet Lorem d power cord So s tt Del not oontrotue by ee vedr red prescribed oryge acoestonves to Oo ouf as Grected P2 button 10 start or shop device Poena amp amp UU button 10 set prescribed Cu flow Clean fe ar rigke Ver woetly or as needed haha Jevne oque abr Formas Waren raczy Cal for service and ornas ute Gewese bebe aum rad os tee Nr varer F de am UGA ow Pre l ne amp saw ly o on Tu oo dme Syr Chart SeQual Technologies Inc 10 Part No 7040 SS SEQUAL 2 Display of Operating Conditions Flow Rate Green Yellow Red Audible Unit EE Shown Light Light Light Alarm State p g Off Unit not running Flow Tr Indicator Display Not Lit SEE Unit not running Flow J Indicator Display Lit seconds Running Periodic KEENE er Periodie nit is warming up Beep J Normal Operating epe e e e fem ge Periodic Unit turned Beep n On Off On Quickl ee pel Concentration ee pel Normal Level Flashing Running O Flow Rate Not Correct 10 off target flow Gr Concentration Flashing Running roma Level O Flow Rate Abnormal 30
13. he wall draperies and Smoke pollutant and fume free Relative Humidity 1096 95 Operating Time Up to 24 hours per day Flow Rate From 0 5 LPM to 10 0 LPM 40 L r 7 7 7 7 P Co oO I I I J N iN HE oe I i QQ e I I I Temperature C 4 Bp RE Mr nct dct I I I mm eee mo a I I I e e e m b e e e e m m 1 m I I I 0 55 60 65 70 75 80 85 90 95 Relative Humidity Important For rated performance at 10 LPM do not operate the Oxygen concentrator above the indicated boundaries Operating at flow rates less than 10 LPM will tolerate higher temperature humidity conditions Chart SeQual Technologies Inc 6 Part No 7040 SS SEQUAL THE PROPER LOCATION Usage in environments other than those described on the previous page may result in the need for increased oxygen concentrator maintenance Select a location in a well ventilated space that avoids fumes and pollutants and allows the oxygen concentrator to draw in room air through the intake filter on the top of the cabinet without being restricted Make sure the oxygen concentrator is at least three to six 3 6 inches 7 6 15 2 cm away from walls furniture and especially curtains or draperies that can impede or block adequate airflow to the ox
14. ilter Room air is drawn into the oxygen concentrator through the air inlet filter located on the top backside of the cabinet Cleaning the air inlet filter is the most important activity you will perform to keep your oxygen concentrator performing properly Check this air filter daily and clean it at least once a week 1 Remove the filter from the cabinet 2 Wash the filter in warm water using a mild detergent solution 3 Rinse the filter thoroughly and squeeze out the excess water 4 Allow the filter to air dry Caution The filter should be free of liquid water and dry to the touch before reinstallation as excess moisture may impair the proper operation of the oxygen concentrator 5 Reinsert the filter in the cabinet Caution The air inlet filter should be replaced as needed If the oxygen concentrator is used in a dusty environment the filter may need to be replaced more often Do Not operate the oxygen concentrator for more than 30 minutes without a filter installed e Oxygen Delivery Accessories Follow the instructions supplied by the manufacturer Chart SeQual Technologies Inc 12 Part No 7040 ESSEQUAL TROUBLE SHOOTING GUIDE Table 1 Concentrator is not Running Probable Cause Cannot Start Unit Power cord not plugged into wall i Back Light Off Internal Fuse Blown Power cord pulled from wall outlet Plug in power cord Push On Off Button to turn off unit Audible Alarm Power outage Push On Off Button t
15. level Green Light Normal Operation Oxygen Concentration Indicator Yellow Light Flashing Below Normal Operation 85 Red Light Flashing Abnormal Operation 70 Oxygen Outlet Pressure Electrical Power 26 5 in 66 cm H x 14 7in 37 cm W x 19 5 in 50 cm D Power Interruption Power Surge simulated circuit breaker Compressor Malfunction Audible Alarm Indicators Excessive Internal Temperature Outlet Pressure 35 36 psig 241 2 248 1 kPa relief valve Abnormal Oxygen Flow Rate gt 30 target flow Low Therapeutic Oxygen Output 7096 ox Back up Alarm Power 9V Battery Cabinet Air Inlet Compressor Inlet and HEPA Device Classification IEC Class II Type BF Applied Part Continuous Operation Temperature 4 F 20 C to 140 F 60 C Iransperpstoragesneguimsments Humidity Up to 95 Non Condensing Back pressure Effect on 0 psig 0 kPa 10 0 LPM Flow Indicator 1 psig 7 kPa 10 0 LPM Nominal Temperature 50 F to 104 F 10 C to 40 C Independent of OSCI Operation Atmospheric Pressure 85 396 and 70 5 oxygen concentration measurement accurac Where equipped Subject to 3 system concentration accuracy and for backpressures up to 7kPa Where equipped Subject to 5 system concentration accuracy and for backpressures up to 7kPa Outlet pressure may decline to 5 0 psig 34 5 kPa nominal at 10 000 ft altitude The system performance may decline above 10 000 ft 3048m Chart SeQual Technologies Inc 14
16. o restart Blinking Red Light unit when power comes back Flow Indicator Blank on Back Light Off Ensure the unit is not plugged into a wall switched outlet No power at wall outlet Check house circuit breaker or fuse reset or replace as needed Remove any obstruction s from Audible Alarm Unit overheating air openings and refer to page Blinking Red Light 6 The Proper Location Flow Indicator Blank Back Light On Compressor malfunction Service required Table 2 Concentrator Running Humidifier or tubing blocked or En RA os humidifier kinked and or tu Ing No oxygen flow Remove tube kink see Alarm Indications in Air Inlet Filter blocked Clean air inlet filter Table on Page 11 Internal component malfunction Restriction in humidifier or tubing Repair or replace as required Low oxygen flow ae Air Inlet Filter restricted Clean air inlet filter see Alarm Indications in Table on Page 11 Internal component worn or quie Service required malfunctioning Yellow Light On Button is stuck Flow Rate not visible on Audible Alarm Red Light On Flow Indicator Blank Back Light On Note If abnormal operating conditions are detected and cannot be remedied contact your Service Provider Remove power cord from wall System malfunction outlet Service required Chart SeQual Technologies Inc 13 Part No 7040 ESSEQUAL TECHNICAL DATA Flow Rate 0 5 to 10 0 LPM Oxygen Concentration 1 0 to 10 0 LPM 91 3 at sea
17. ra E Z Oxygen Concentrator is an electronically operated oxygen concentrator that separates oxygen from room air It does not remove oxygen from a room any more than by simply breathing the same room air You breathe the added oxygen through nasal prongs of a cannula small soft plastic tubes that fit just inside your nostrils Your physician has prescribed a specific oxygen flow rate setting to meet your individual needs This flow rate is determined by the amount of oxygen needed Recommended oxygen flow rates should be adjusted only under the advice of a physician This oxygen concentrator is not intended for life supporting or life sustaining applications As with any electrically powered device the user may experience periods of non operation as the result of electrical power interruption or the need to have the oxygen concentrator serviced by a technician This instruction manual will inform you about the use and care of the Integra E Z Oxygen Concentrator and will serve as a reference as you use your oxygen concentrator Important Safety Instructions are shown under the following headings Warning Important safety information for hazards that might cause serious injury Caution Information for preventing damage to the oxygen concentrator Do Not Information to which you should pay special attention Important Clarifying information specific instructions commentary sidelights or interesting points Chart SeQual Technologie
18. resence of an electric shock hazard Warning To reduce the risk of electric shock do not remove cover There are no user serviceable parts inside Refer servicing to a SeQual dealer or authorized service representative Audible Alarm An audible indicator is provided to announce in a tone distinctly different from the sounds generated by the oxygen concentrator during normal operation a reduction in oxygen concentration a large unintentional change in flow or when the mains power has been interrupted Normal Operation Green light is on indicating the concentrator is running under normal operating conditions ON OFF condition Push button starts concentrator running Flow Rate Adjustment The controlled quantity of oxygen increases or decreases in increments of 0 5 LPM with each press of the and flow rate adjustment buttons respectively Warning The prescribed flow rate setting should only be adjusted under the advice of a physician Type BF Equipment F type applied part complying with the specified requirements of IEC 60601 1 to provide a higher degree of protection against electrical shock than provided by type B applied part Class Il Equipment Equipment in which protection against electrical shock does not rely on basic insulation only but in which additional safety precautions are provided Amperes Alternating Current eB cc I OPE Authorized European Representative
19. rt of the humidifier When no humidifier is prescribed the oxygen tubing can be attached directly to the oxygen outlet port Connect the other end of the tube to the nasal cannula or facemask if not already attached Important The tube fitting may be tight Do Not use oil or grease to lubricate To ensure there is enough pressure to deliver oxygen the length of the tubing should not exceed 50 feet 15 2m Maximum extension Ka tubing length is dependent upon environmental conditions tubing type liter flow and any accessories used between DISS fitting and nasal cannula mask Shorter tubing lengths will be required in some cases The internal diameter of the tubing should be no less than 3 16 inch 0 48 cm A maximum suitable 50 extension is Salter Labs Tubing ref 2050 Tubing not specified for use with this oxygen concentrator may impair the performance e Cannula or facemask Your physician will have prescribed a cannula or facemask to deliver oxygen In most cases they are already attached to the oxygen tubing If not follow the instructions included with the cannula or facemask to attach them to the oxygen tubing Use accessories recommended by your SeQual dealer provider or authorized service representative A cannula suitable for use with the 10L Integra E Z is the Salter Labs High Flow Cannula ref 1600HF A facemask suitable for use with the 10L Integra E Z is the Salter Labs Facemask ref 8110 Use of cannulas or facemasks not sp
20. s Inc 4 Part No 7040 ESSEQUAL SYMBOLS USED IN INSTRUCTION MANUAL AND ON OXYGEN CONCENTRATOR Refer to this section for an explanation of the symbols and the warnings that accompany them Read Instructions This symbol when used alone or in conjunction with any of the following symbols indicates the need to consult the operating instructions provided with the product A potential risk exists if the operating instructions are not followed Attention or Caution Consult accompanying documents This symbol indicates important information that requires the user to take special precautions when certain conditions are present No Smoking or Naked Flames The use of oxygen therapy requires that special care be taken to reduce the risk of fire Materials that burn in air and even some material that will not burn in air are easily ignited and burn rapidly in high concentrations of oxygen For safety concerns all possible sources of ignition must be kept away from the oxygen concentrator and preferably out of the room in which it is being used Smoking in the proximity of an operating oxygen concentrator is dangerous and can permanently damage the device and void the warranty Use No Oil or Grease Indicates important information about the possibility of spontaneous and violent ignition that may occur if oil grease or other petroleum based substances come into contact with oxygen under pressure Electrical Shock Indicates the p
21. should only be removed by qualified service personnel e Warning An alternate source of supplemental oxygen is recommended in the event of a power outage alarm condition or mechanical failure Chart SeQual Technologies Inc 3 Part No 7040 ESSEQUAL INDICATIONS FOR USE The SeQual Integra E Z Oxygen Concentrator is intended to provide supplemental oxygen It is not intended for life supporting or life sustaining applications nor does it provide any patient monitoring capabilities CONTRAINDICATIONS An oxygen concentrator is not appropriate for any patient who would experience adverse health consequences as the result of such temporary interruption In certain circumstances oxygen therapy can be hazardous It is advisable to seek medical advice before using this oxygen concentrator Warning Federal law restricts this device to sale by or on the order of a physician INTRODUCTION Your physician has prescribed supplemental oxygen as part of a treatment plan Long term oxygen therapy is now widely accepted as the standard treatment for chronic hypoxia caused by chronic obstructive pulmonary disease Oxygen therapy has become more versatile in recent years due to the development of oxygen concentrators The new devices make it easier for physicians to select a mode of oxygen therapy that suits the patient s lifestyle An oxygen concentrator has been prescribed to provide you with a source for supplemental oxygen SeQual s Integ
22. ygen concentrator Do Not cover unit with a blanket towel quilt etc Do Not place objects under unit Maintain at least inch clearance under unit Keep the oxygen concentrator at least five 5 feet 1 5 m away from hot sparking objects or open sources of flame Do Not locate the unit near any heat source or heat register Before Operating Your Oxygen Concentrator There are many different types of oxygen tubing cannulas or facemasks and humidifiers that can be used with this oxygen concentrator You should contact your local SeQual dealer provider or authorized service representative for recommendations on which of these oxygen accessories will best fit your needs They will also be able to give you advice on the proper maintenance and cleaning of these accessories e Humidifier If your physician has prescribed a humidifier follow the manufacturer s instructions for attachment Add distilled water to the level suggested in the manufacturer s instructions and attach the humidifier to the oxygen outlet port of the oxygen concentrator Do Not overfill Only use a humidifier recommended by your SeQual dealer provider or authorized service representative A humidifier suitable for use with the Integra Oxygen Concentrator is the Salter Labs High Flow Bubble Humidifier ref 7900 Use of certain humidifiers not specified for use with this oxygen concentrator may impair performance e Oxygen Tubing Connect the oxygen tubing to the outlet po
Download Pdf Manuals
Related Search
Related Contents
Genelec 8040A User's Manual Manual De Utilização Transcend JetRam Speicher 4GB Pelco PT506-24A User's Manual InLine 33397I Mode d`emploi Module de mesure Protos 3400(X) DF-RS12 取扱説明書 MEMORY BELT Manual de usuario grabadores digitales EHD104 Sony MEX-GS610BT Operating Instructions Copyright © All rights reserved.