Home
        Akta FPLC instructions ENG
         Contents
1.                         5 Maintenance  5 4 Component maintenance    The system documentation also includes a spare part list to be used to find common  spare parts and their code numbers for ordering  This list can also be found online at  www gelifesciences com AKTA     5 5 Disassembly and assembly of components and    consumables    The operator must carefully read and understand the instructions supplied for each  component before disassembly and assembly of the component  When replacing  consumables  such as tubing and tubing connectors  all neccessary safety precautions  must be taken  Contact your local GE Healthcare representative if further information  or help is needed     WARNING    Disconnect power  Always disconnect power from the instrument  before replacing any component on the instrument  unless stated  otherwise in the user documentation     WARNING    Before disassembly  check that there is no pressure in the piping  system     WARNING    After assembly  the piping system must be tested for leakage at  maximum pressure for continued protection against injury risks  due to fluid jets  burst pipes or explosive atmosphere         gt   gt   gt        5 6 Replacement of fuses    WARNING    Disconnect power  Always disconnect power from the instrument  before replacing fuses         gt        Refer to Section 7 1 Specifications  on page 59 for information about the fuse type and  rating  If a fuse repeatedly blows  switch off the system mains switch and contact yo
2.                   Figure 1 2  Basic flow path  AKTAexplorer is shown as an example     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 11                                                                         Description    Buffers from container pass through Buffer valve V6 and or the  switch valve     Sample pass through sample valve V5 which selects a sample  depending on the setting in the control software     Pump A pumps buffer through the system   The buffers pass through Mixer M where they are mixed     This is where the sample is added to the flow path  The sample  can be added manually with a suitable syringe or pumped in via  the sample pump p1 from a sample chosen via sample valve V5  or an autosampler     The flow direction valve V7 is optional in some systems and used  to select the direction of the flow through column     Column selection valves V2 and V3 direct flow through a specified  column     Liquid returns to flow direction valve V7 and redirects flow to  outlet valve via pH  UV and Condictivity monitors     The outlet valve directs the flow either to waste  to fraction col   lection containers or to a fraction collector device such as Frac   950     Part    pA    V1  pl    V7    V2  V3    V7  UV  C     pH    V4    Step    The flow path is likely to include air sensors  flow restrictors  online filters  sample loops  and so on which vary in number and application depending on the system and its  strategy     Fore more detai
3.           OD xxxxx    Sample  Code No  XXXXXXXX Voltage   Serial No  XXXXXXX Frequency   Mfg Year  2009 Power max   Fuse        Qoy C     C ULL LLLI  Made in Sweden GE Healthcare Bio Sciences AB Or  9 751 84 Uppsala Sweden       Symbols used in safety labels    Label Meaning    Warning  Read the user documentation before using the system  Do not   N open any covers or replace parts unless specifically stated in the user  documentation     The system complies with the requirements for electromagnetic compliance        EMC  in Australia and New Zealand     The system complies with applicable European directives              22   KTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                      2 Safety instructions  2 2 Labels    Labels concerning hazardous    substances  Label Meaning    This symbol indicates that the waste of electrical and electronic equipment  must not be disposed as unsorted municipal waste and must be collected  separately  Please contact an authorized representative of the manufac    mummm turer for information concerning the decommissioning of equipment     This symbol indicates that the product contains hazardous materials in  o excess of the limits established by the Chinese standard SJ T11363 2006   Requirements for Concentration Limits for Certain Hazardous Substances  in Electronics              2 5 Emergency procedures    This section describes how to do an emergency shutdown of the AKTAexplorer   AKTApurifier and A
4.          4 Operation  4 5 Setting up a run    10 In the Save As dialog     e Enter method name in the Method name field and select folder to save the  method in     e Ifyou have more than one system connected to the computer  select system in  the For System area for which the method is intended       Select technique in the Technique area for which the method was written   11 Click OK     The method is saved  It can now be started from the System Control module     4 6 Performing a run    1 Select method  a In System Control  select File Run     b Select the required Method from the list     2 Specify variables  Enter identification names for the samples via the keyboard     3 Editresult file location and names    If required  edit the folder path and file names of the result files to be created     4 Preparations completed     Ensure that the preparations according to Section 4 4 Preparations before start  on  page 32 has been performed     5 Check the flow path   Make sure that    e there is enough buffer available   e the correct inlet is placed in each buffer   e the outlets are placed in correct bottles   e the columns are placed in correct positions   e the chosen fraction collector rack is filled and is in correct place   6 Prepare the samples    The samples should have been prepared and clarified using centrifugation and or  filtration through a 0 45 um filtert     1 fusing HisTrap    FF crude  clarification is not needed     AKTAexplorer  AKTApurifier and AKTAmicro 
5.       2 Safety instructions  2 1 Safety precautions    WARNING    Spread of biological agents  The operator has to take all necessary  actions to avoid spreading hazardous biological agents in the  vicinity of the instrument  The facility should comply with the  national code of practice for biosafety     WARNING    High pressure  The AKTAexplorer  AKTApurifier and AKTAmicro  instrument operates under high pressure  Wear protective glasses  at all times           Installing and moving the  instrument    WARNING    Supply voltage  Make sure that the supply voltage at the wall outlet  corresponds to the marking on the instrument  before connecting  the power cord     WARNING    Protective ground  The AKTAexplorer  AKTApurifier and AKTAmicro  instrument must always be connected to a grounded power outlet     WARNING    Power cord  Only use power cords delivered or approved by GE  Healthcare           AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 17                                     WARNING    Access to power switch and power cord  Do not block the rear  and side panel of the instrument  The Power switch must always  be easy to access  The power cord must always be easy to  disconnect     WARNING    Installing the computer  The computer should be installed and  used according to the instructions provided by the manufacturer  of the computer     CAUTION    Heavy object  Use suitable lifting equipment when moving the  systems  Three people are requ
6.      Using flammable liquids   e Personal protection     Installing and moving the instrument   e System operation    e Maintenance    General precautions    Always follow these General precautions to avoid injury when using the AKTAexplorer   AKTApurifier and AKTAmicro instrument     WARNING    Do not operate the AKTAexplorer  AKTApurifier and AKTAmicro  instrument in any other way than described in the AKTAexplorer   AKTApurifier and AKTAmicro and UNICORN manuals              14 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA             2 Safety instructions  2 1 Safety precautions    WARNING    Operation and user maintenance of the AKTAexplorer  AKTApurifier  and AKTAmicro instrument should be performed by properly trained  personnel only     WARNING    Before connecting a column to the AKTAexplorer  AKTApurifier and  AKTAmicro instrument  read the instructions for use of the column   To avoid exposing the column to excessive pressure  make sure  that the pressure limit is set to the specified maximum pressure of  the column     WARNING    Do not use any accessories not supplied or recommended by GE  Healthcare     WARNING    Do not use the AKTAexplorer  AKTApurifier and AKTAmicro  instrument if it is not working properly  nor if it has suffered any  damage  for example     PP  gt   gt     e damage to the power cord or its plug  e damage caused by dropping the equipment    e damage caused by splashing liquid onto it    CAUTION    Waste tubes 
7.     The AKTAexplorer  AKTApurifier and AKTAmicro systems are intended for research use  only  and shall not be used in any clinical procedures  or for diagnostic purposes     Safety notices    This user documentation contains WARNINGS  CAUTIONS and NOTICES concerning the  safe use of the product  See definitions below     Warnings    WARNING    WARNING indicates a hazardous situation which  if not avoided   could result in death or serious injury  It is important not to proceed  until all stated conditions are met and clearly understood           6 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                1 Introduction  1 1 Important user information    Cautions  CAUTION  CAUTION indicates a hazardous situation which  if not avoided   could result in minor or moderate injury  It is important not to  proceed until all stated conditions are met and clearly understood   Notices    NOTICE  9 NOTICE indicates instructions that must be followed to avoid    damage to the product or other equipment           Notes and tips    Note  A Note is used to indicate information that is important for trouble free  and optimal use of the product    TIP  A tip contains useful information that can improve or optimize your  procedures     Typographical conventions    Software items are identified in the text by bold italic text  A colon separates menu levels   thus File Open refers to the Open command in the File menu  Hardware items are  identified in the text
8.    2 Safety instructions  2 1 Safety precautions       20                                        2 Safety instructions  2 1 Safety precautions    WARNING    After assembly  the piping system must be tested for leakage at  maximum pressure for continued protection against injury risks  due to fluid jets  burst pipes or explosive atmosphere     WARNING    Before disassembly  check that there is no pressure in the piping  system     WARNING    Disconnect power  Always disconnect power from the instrument  before replacing fuses     WARNING    Decontaminate the equipment before decommissioning to ensure  that hazardous residues are removed     NOTICE    Cleaning  Keep the instrument dry and clean  Wipe regularly with  a soft damp tissue and  if necessary  a mild cleaning agent  Let the  instrument dry completely before use     o  gt   lt   gt   gt           2 2 Labels    This section describes the safety labels and labels concerning hazardous substances  that are attached to the AKTAexplorer  AKTApurifier and AKTAmicro instruments  For  information about marking of the computer equipment  refer to the manufacturer   s  instructions     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 21                                     2 Safety instructions  2 2 Labels    Labels on the instrument    The illustration below shows an example of the identification label that is attached to  the AKTAexplorer  AKTApurifier and AKTAmicro instruments        S              
9.   AKTAmicro User Manuals        53          Possible cause    Aging UV lamp    UV lamp in wrong po   sition    The theoretical extinc   tion coefficient too  low    6 2 Conductivity curve problems    Possible cause    Air in the pump or the  flow cell    Leaking tube connec   tions    Incorrect mixer func   tion    Dirty conductivity cell    Column not equilibrat   ed      KTAexplorer    KTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA          Error symptom    Low sensitivity    Error symptom    Baseline drift or noisy  signal                                  Corrective action    Clean the flow cell according to  procedure in Monitor User Manual     Use a temperature compensation  factor  See Monitor User Manual     Check that the pump is operating  and is programmed correctly     Check that the mixing chamber is  free from dirt or particles     Change to a larger mixing cham   ber volume if necessary     Check the motor operation  Place  a hand on the mixer and start it by  starting the pump at a low flow  rate  You should both hear and feel  the mixer motor and stirrer when  they are spinning     Check for loose tubing connec   tions  Use the flow restrictor     Wash the tubing and check pump  is operating properly     Change to smaller mixer volume        AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA       Possible cause    Dirty flow cell    Decrease in ambient  temperature    Incorrect pump func   tion    Dirty mixing chambe
10.   three people to lift and move it unless a suitable lifting device is used     AKTAexplorer 66 kg  AKTApurifier 41 kg  AKTAmicro 55 kg             The instrument can be transported on a trolley or a suitable lifting device capable of  supporting the weight of the instrument     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    26                                  3 Installation  3 2 Transport    NOTICE  o Lift the system in the upright position  Do not use the front panel    bar as a lifting handle           Before moving the system     e disconnect all cables and tubing connected to peripheral components and liquid  containers       remove all items from the top of the system   e close the valve door completely  only for AKTAexplorer      e grasp the system firmly by placing the fingers in the gap between the swivel platform  and the base of the main unit and lift     For more information on Transport  see AKTAexplorer  AKTApurifier and AKTAmicro  Installation Guides     5 5 Unpacking    e Check the equipment for damage before starting assembly and installation     e Document any damage and contact your local GE Healthcare representative     Remove straps and packing material and stand equipment upright on swivel foot before  starting installation     5 4 Assembly    The following parts must be added to the AKTAexplorer  AKTApurifier and AKTAmicro  instruments before they can be used       pH electrode  optional      Waste tube   e  CU 950 Cont
11.  B2 Teflon 1700   16 3 4 x 10   Switch valve   pump A3  B3 Teflon 150 1 6 302  Pump   Mixer G1  G2 PEEK 330 0 50 65  Mixer   On line filter G3 PEEK 200 0 50 39  Filter   V1  Inj  Valve  G4 PEEK 180 0 50 35  V1   UV  bypass  G5 PEEK 400 0 50 79  UV   Cond  Cell G6 PEEK 160 0 25 8  Cond  Cell   Flow restrictor G7 PEEK 80 0 25 4  Flow restrictor   V4 G8 PEEK 140 0 25 7  V4   Fraction collector G9 PEEK 600 0 25 29  V4   Flow through F3 PEEK 1000 0 50 196  V1 V4   Waste W1 W2 W3   Tefzel 100 0 75 442   AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 1                                                    6 Tubing    Table 6 4  Tubing specifications for AKTApurifier 100    Use Label Material Length   I D  Volume   mm   mm   ul    Inlet A1 Al Teflon 1700 29 112 x 10    Inlet A2 A2 Teflon 1900   9 12 5 x 103   Inlet B1 B1 Teflon 1500   29 9 9 x 10    Inlet B2 B2 Teflon 1700 29 112 x 10    Switch valve   pump A3  B3 Teflon 150 2 9 991   Pump   Mixer G1  G2 PEEK 330 0 75 146   Mixer   On line filter G3 PEEK 200 0 75 88   Filter   V1  Inj  Valve  G4 PEEK 180 0 75 80   V1   UV  bypass  G5 PEEK 400 0 75 177   UV   Cond  Cell G6 Tefzel 160 0 75 71   Cond  Cell   Flow restrictor G7 Tefzel 80 0 75 35   Flow restrictor   V4 G8 Tefzel 140 0 75 62   V4   Fraction collector G9 Tefzel 600 0 75 265   V4   Flow through F3 Tefzel 1000 1 0 785   V1 V4   Waste W1 W2 W3   Tefzel 100 0 75 442                         Tubing specifications for  AKTAmicro    Names in the Tubing i
12.  Replace the membranes and piston  seals according to the User manual                 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 45             Instructions reference    Disconnect the flow restrictor     Connect a capillary  approx  1 m  i d  1 mm   to a free port in the injection valve  Set the  valve manually to this port  Put the open  end in a waste container     Run the pump at 10 ml min with water   Note the back pressure  Bp1  on the pump  display  or in the Run Data window     Connect the flow restrictor to the open end  of the capillary  observe the IN marking    Putthe flow restrictor in the waste contain   er     Run the pump at 10 ml min with water   Note the back pressure  Bp2  on the pump  display  or in the Run Data window     Calculate the back pressure generated by  the flow restrictor  Replace it if it is not  within limit        Check that flow restrictor generates the follow   ing back pressure     FR 904  0 4  0 05 MPa  FR 902  0 2 x0 05 MPa    Check the back pressure as follows     1  2       Action    Flow restrictor    AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    5 Maintenance    5 2 User maintenance schedule    Interval    Monthly          46    5 Maintenance  5 2 User maintenance schedule    Interval Action Instructions reference    1 Disconnect the flow restrictor     2 Connecta capillary  approx  1 m  i d  1 mm   to the waste port  port 5  on the injection  valve  Set the injec
13.  User Manual     Ifthe problem remains  replace the  pH electrode        55          Possible cause    Loose connection of  conductivity flow ca   ble    Incorrect pump and  valves function    Incorrect temperature  compensation factor    Dirty or incorrectly  equilibrated column    Incorrect mixer func   tion    Possible cause    Faulty electrode con   nection    Damaged electrode    Incorrectly connected  pH monitor    Dirty pH electrode      KTAexplorer    KTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA          Error symptom    Incorrect or unstable  reading    6 3 pH curve problems    Error symptom    No response to pH  changes    Small response to pH  changes                                     6 Troubleshooting  6 3 pH curve problems    Error symptom Possible cause Corrective action   Slow pH responseor   Contaminated elec  Check the electrode glass mem    Calibration impossible   trode glass mem  brane  If it is contaminated  clean  brane the electrode following the instruc     tions in Monitor pH C 900 User  Manudl or UPC 900 User Manudl     Membrane has dried   Ifthe membrane has dried out  the  out electrode may be restored by  soaking it in buffer overnight                 56 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA          6 Troubleshooting  6 5 pH curve problems    Error symptom Possible cause Corrective action  Incorrect or unstable   Problem with elec  Check that the electrode cable is  pH reading trode conn
14.  by bold text  e g   Power switch      1 2 Regulatory information    This section describes the directives and standards that are fulfilled by the AKTAexplorer   AKTApurifier and AKTAmicro systems     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 7                         1 Introduction  1 2 Regulatory information    Manufacturing information    The Declaration of Conformity includes the following information     Requirement Content    Name and address of manufacturer GE Healthcare Bio Sciences AB    Bj  rkgatan 30  SE 751 84 Uppsala  Swe   den    Name and address of person responsible   Peter L  wendahl  Bj  rkgatan 30    for Technical File SE 751 84 Uppsala  Sweden  Name and ID of notified body INTERTEK SEMKO AB  NB 0413  Place and date of declaration Uppsala  Sweden  May 2009    Identity of person authorizedto sign DoC   Peter L  wendahl  See EC Declaration of Conformity             CE Conformity    This product complies with the European directives listed in the table  by fulfilling the  corresponding harmonized standards  A copy of the Declaration of Conformity is available    on request   2006 42 EC Machinery Directive  MD   2006 95 EC Low Voltage Directive  LVD   2004 108 EC ElectroMagnetic Compatibility  EMC  Directive             International standards    This product fulfills the requirements of the following standards        Standard Description   EN 61010 1  IEC Safety requirements for electrical  61010 1  UL 61010 1    equipment f
15.  d  column in Table refer to tubing labels in the liquid flow path  connection diagram  see Liquid flow path for AKTAmicro  on page 68     Tubing Material Color Max  Volume Connection points    o d  pressure of 10cm       0 35 mm 1 6 Titanium Grey   35MPa 9 6 ul From Pump P 905 to Mixer   G1  G2  mm    72 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                                    6 Tubing       Tubing Material Color Max  Volume Connection points  o d  pressure   of 10 cm  Union    PEEK Black  gt 35 MPa   Between Mixer and On line  m m filter   Between UV cell and Conduc   tivity cell  0 15 mm 1 16  PEEK Violet  gt 35MPa   18yl From On line filter to outlet   G3 to  or to fraction collector MS   G5   optional       PEEK tubing i d  0 15 mm is  installed at delivery     0 15 mm 375 Fused sili    Brown      35 MPa   18yl From On line filter to outlet   Di  0 15  um ca  or to fraction collector MS   optional       Tubing kit 0 15 is supplied  with the system     0 10 mm 200 Fused sili    Brown      35 MPa   0 8 ul From On line filter to outlet   Di  0 10  um ca  or to fraction collector MS   optional       Tubing kit 0 10 is supplied  with the system     0 75 mm 1 16  Tefzel Clear 7 MPa 44 2 ul Waste tubing    w1  W2    1 6 mm 1 8  Teflon Clear 3 4 MPa 201 1 ul Inlet tubing to Pump P 905   A1 to A3    B1 to B3                             1l 025and0 5 mm i d  tubing is supplied for non analytical applications     AKTAexplorer  A
16.  degree of contamination depends on  the sample and the chromatographic conditions employed  These should be considered  when designing a cleaning protocol     Routine cleaning should be performed at intervals aimed at prevention rather than  cleaning the system from growth or contamination     WARNING    Make sure that the piping system is completely leakage free before  performing any CIP on the system        Make sure that the process control method for cleaning flushes all possible flow paths  in the system  After cleaning  rinse the entire system with water or suitable liquid until  the piping tubing system is completely free from the CIP solution  monitors in the system  can be used as detectors   Do not leave NaOH or other cleaning agents in the system  for long periods        WARNING    Hazardous chemicals during maintenance  When using hazardous  chemicals for system or column cleaning  wash the system or  columns with a neutral solution in the last phase or step     WARNING    NaOH is corrosive and therefore dangerous to health  When using  hazardous chemicals  avoid spillage and wear protective glasses  and other suitable personal protective equipment           See also Section 5 8 Storage  on page 50     5 4 Component maintenance    Maintenance and preventive replacement of parts of the major components are described  in the respective manuals included in the system documentation     48 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 
17.  of operations and length of intervals necessary to maintain system function  and safety     Table 5 1  User maintenance schedule    Instructions reference    Visually inspect the system for leaks   1 Forcleaning the flow path  see Cleaning In   Place  on page 48     2 For leaving the system for a few days  see  Section 5 8 Storage  on page 50     Calibrate the pH electrode  if applicable  accord   ing to Monitor pH C 900 User Manual           Action  Leak inspection    Wash the system flow  path    Calibrate pH electrode   optional     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    Interval    Daily          44                               5 Maintenance  5 2 User maintenance schedule    Interval Action Instructions reference  Weekly Check inlet filters Check the inlet filters visually and replace them  if necessary     Replace on line filter   Replace the on line filter    if applicable     Change pump rinsing   Change rinsing solution  Always use 20   solution ethanol with 10 mM NaOH as rinsing solution     If the volume of rinsing solution in the storage  bottle has increased  it can be an indication of  internal pump leakage  Replace the piston seals  according to the User manual     If the volume of rinsing solution in the storage  bottle has decreased significantly  check if the  rinsing system connectors are mounted proper   ly    If the rinsing system connectors are not leaking   the rinsing membranes or piston seals may be  leaking 
18. 0  on page 65 and  Liquid flow path for AKTApurifier 10 and 100  on page 67  respectively     Table C 1  Tubing specifications for AKTAexplorer 10                   Use Label Material Length          Volume   mm   mm   ul    Inlets A11 to A18 A11 18 Teflon    1250 1 6 2 5 x 10    Inlet A1 1 Teflon 750 1 6 1 5 x 10    Inlet A2 A2 Teflon 2000 1 6 4 0 x 10    Inlet 81 B1 Teflon 1800   16 3 6 x 10    Inlet B2 B2 Teflon 1800   16 3 6 x 10    Switch valve   pump A3  B3 Teflon 150 1 6 302   Pump   Mixer 61  G2 PEEK 300 0 50 59   Mixer   On line filter G3 PEEK 150 0 50 29   Filter   V1  Inj  Valve  G4 PEEK 460 0 50 90   V1 V2 G5 PEEK 270 0 25 13   V2   V3  bypass  G6 PEEK 620 0 50 122   V3   UV G7 PEEK 550 0 25 27   UV   Cond  Cell G8 PEEK 160 0 25 8   Cond  Cell   Flow restrictor G9 PEEK 450 0 25 22   Flow restrictor   V4 G10 PEEK 120 0 25 6   V4   Fraction collector G11 PEEK 600 0 25 29   V4   Flow through F3 PEEK 1000 0 50 196   V1   Waste W1  W2 Tefzel    1300 0 75 574   AkTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 69                                                                6 Tubing    Material Length   I D      mm   mm        V4   Waste W3 Tefzel 1000 0 75 442    Table C 2  Tubing specifications for AKTAexplorer 100       Material Length   1 D    mm   mm    Inlets A11 to A18 A11 18 Teflon 1250 2 9 8 3 x 10   Inlet A1 Al Teflon 750 29 4 9 x 103  Inlet A2 A2 Teflon 2000   9 13 2 x 103  Inlet B1 B1 Teflon 1800   29 11 9 x 103  Inlet B2 B2 Tefl
19. 2125 02 9    7783 20 2    29340 81 6    67 68 5    75 08 1       Exposure    up to 2  months    OK    OK    Avoid    OK    OK    OK    OK  OK    OK  OK  OK    Avoid    OK  OK    Avoid    Avoid    OK    AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA       7 Reference information    Exposure     gt  1day    OK    OK    OK    OK    OK    OK    OK  OK    OK  OK  OK    OK    OK  OK    Avoid    Avoid    OK    7 2 Chemical resistance    Chemical    Acetic acid  70     Acetonitrile    Acetone  10     Ammonia  30     Ammonium chlo   ride    Ammonium bicar   bonate    Ammonium nitrate    Ammonium sul   phate    1 Butanol  2 Butanol  Citric acid    Chloroform    Cyclohexane  Detergents    Dimethyl sulphox   ide    1  4 Dioxane    Ethanol  100           60                                                       7 Reference information  7 2 Chemical resistance    Comments       Silicone not resis   tant  Pressure limit  for PEEK decreases     Silicone not resis   tant     Silicone not resis   tant  Pressure limit  for PEEK decreases     Silicone not resis   tant        Silicone not resis   tant  Titanium is af   fected by long term  use     Silicone not resis   tant     Elgiloy is affected  by long term use     Titanium  alumini   um oxide and glass  are affected by  long term use           61    27 35    2 7 1    75    7    777  2 7 6    2 2    7 53    228       Chemical Exposure Exposure      1 day up to 2   months   Ethyl acetate OK Avoid  Ethylene glycol  
20. Electrical shock hazard  All repairs should be done by service  personnel authorized by GE Healthcare  Do not open any covers  or replace parts unless specifically stated in the user  documentation     System operation    o  gt  b b       Maintenance       AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                        WARNING    Disconnect power  Always disconnect power from the instrument  before replacing any component on the instrument  unless stated  otherwise in the user documentation     WARNING    Hazardous chemicals during maintenance  When using hazardous  chemicals for system or column cleaning  wash the system or  columns with a neutral solution in the last phase or step     WARNING    Do not perform any type of maintenance work while the system is  powered electrically or when the piping system is pressurized  Note  that the piping system can be pressurized even when the system  is closed down     WARNING    Only spare parts that are approved or supplied by GE Healthcare  may be used for maintaining or servicing the system     WARNING    Make sure that the piping system is completely leakage free before  performing any CIP on the system     WARNING    NaOH is corrosive and therefore dangerous to health  When using  hazardous chemicals  avoid spillage and wear protective glasses  and other suitable personal protective equipment     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    
21. GE Healthcare    AKTAexplorer     AKTApurifier     and AKTAmicro       Operating Instructions    Original instructions       0   KTA    Table of Contents    Table of Contents                     1                         5  11 Important user information    m 6  1 2 Regulatory information          7  13 Instrument           a 10  T4    Control SOFtWOFe     aceite ette eec d b tees tr p o OR HR ER 12  2  Safety WAST DONE auo adstat ioo ticc tortas ka lae tael cha aho iuda latius ctos TERETE Sacs 14  2 1 Safety precautions    a 14  22       us 21  2 3 Emergency procedutes sui e baee ti e better oor cU cH 23  24                    24  LE LC S    ce ene ene nee sone Mn nan E er See ee re 26  3 1 SILETEGUIMEMMEMIUS ennha ect tu 00 26  3 2        E AE TES 26        AE        27  3 4 Assembly             27  3 5 Connections            sss m 28  3 6 Spare parts and aecessories     us eie ie I Hei hei po 29  4 Operation en 30  4 1 Operation overview 30  4 2 Starting the instrument         30  4 5 Starting the control system      31  A   Preparations before SEGFt siae ten tetris retreat eret retia erue 32  4 5             7 35  4 6   Performing       39  4 7 Procedures after a run       ttti ttt ttt         42  5          6 0 16       43  51            43  5 2 User maintenance schedule                44  5  anes 47  5 4 Component maintenance m 48  5 5 Disassembly and assembly of components and consumables                      ss 49  56  Replacement OF fusis enpuer t et RR bu asia N RC e
22. KTAmicro systems  The section also describes the result in the event  of power failure     Emergency shutdown    In an emergency situation  do as follows to stop the run     Step Action    1 To pause the run from UNICORN  click the Pause button in System Control   2 If required  switch off power to the instrument by pressing the Main power    Switch to the 0 position     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 23             2 Safety instructions  2 5 Emergency procedures    Power failure  The result of a power failure depends on which unit that is affected     Power failure to    will result in       AKTAexplorer  AKTApuri      The runis interrupted immediately  in an undefined  fier and AKTAmicro sys  state    tem   gt  The data collected up to the time of the power failure    is available in UNICORN    Computer e The UNICORN computer shuts down in an undefined  state       e The run continues  but data cannot be saved in UNI   CORN              2 4 Recycling information    The equipment shall be decontaminated before decommissioning and all local regulations  shall be followed with regard to scrapping of the equipment     Disposal  general instructions    When taking AKTAexplorer  AKTApurifier and AKTAmicro systems out of service  the  different materials must be separated and recycled according to national and local  environmental regulations     Recycling of hazardous  substances    AKTAexplorer  AKTApurifier and AKTAmicro instrument
23. KTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 73                      GE  imagination at work and GE monogram are trademarks of General Electric  Company     Drop Design  HisTrap  Superloop  UNICORN  AKTA  AKTAexplorer  AKTAmicro and  AKTApurifier are trademarks of GE Healthcare companies     All third party trademarks are the property of their respective owners        2009 General Electric Company   All rights reserved   First published Jul  2009    All goods and services are sold subject to the terms and conditions of sale of  the company within GE Healthcare which supplies them  A copy of these terms  and conditions is available on request  Contact your local GE Healthcare  representative for the most current information     UNICORN  Any use of this software is subject to GE Healthcare Standard Software  End User License Agreement for Life Sciences Software Products     GE Healthcare Europe GmbH  Munzinger Strasse 5  D 79111 Freiburg  Germany    GE Healthcare UK Limited  Amersham Place  Little Chalfont  Buckinghamshire  HP7 9NA  UK       GE Healthcare Bio Sciences Corp   800 Centennial Avenue  PO  Box 1327  Piscataway  NJ 08855 1327  USA       GE Healthcare Japan Corporation  Sanken Bldg 3 25 1  Hyakunincho Shinjuku ku  Tokyo 169 0073  Japan    28 9597 57 AA 11 2009    For local office contact information  visit  www gelifesciences com contact    GE Healthcare Bio Sciences AB  Bj  rkgatan 30   751 84 Uppsala   Sweden    www gelifesciences com AKTA    imagi
24. M NaOH or 2096 ethanol  solution     Columns and media shall be stored according to their respective instructions     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    50                               5 Maintenance  5 8 Storage    Storage conditions  The following conditions shall be maintained while the system is in storage     Temperature   2 to  30  C  preferably room temperature       Relative humidity  0 to 95   non condensing  preferably low humidity      After storage  clean the system  calibrate all monitors  and perform a leakage test before  using the system     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 51       Corrective action    Clean the system  Make sure air is  removed     Clean the column according to the  column instructions     Check the mixer function by plac   ing a stirrer bar on top of the mixer  housing  The stirrer bar should ro   tate when the system is in Run  mode  The mixer function can also  be checked by running the installa   tion test     Check the connections of the UV  cell optical fiber  Replace if neces   sary     Clean the UV cell by flushing De   cone 90  Deconex 11 or equivalent     Check if the signal is still noisy with  water     Purge the pump according to  Pump User Manual  Run a system  wash with buffer        AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA       6 Troubleshooting    6 Troubleshooting    Possible cause    Dirt or resi
25. OK OK 107 21 1  100   Formic acid  100    OK OK 64 18 6  Glycerol  100  OK OK 56 81 5  Guanidinium hy  OK OK  drochloride  Hexane OK Avoid  Hydrochloric acid    OK OK 7647 01 0  0 1 M  Hydrochloric acid    OK Avoid   gt 0 1M  Isopropanol  100    OK OK 67 63 0  Methanol  100  OK OK 74 93 1  Nitric acid  diluted OK Avoid  Nitric acid  30  Avoid Avoid  Phosphoric acid  OK Avoid 7664 38 2  10   Potassium carbon    OK OK 584 08 7  ate  Potassium chloride   OK OK 7447 40 7                  KTAexplorer    KTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                                 Exposure Comments    up to 2  months       Avoid ETFE  PP and PE not  resistant     OK  OK    OK  OK  OK  OK 7647 14 5 231 598 3    Avoid 11 2 2 5 PVDF and borosili   cate glass are af   fected by long term    use    OK 7757 82 6 231 820 9   Avoid PEEK and titanium  are affected by  long term use    Avoid   Avoid Silicone  PP and PE  are not resistant    Avoid ETFE  CTFE  PP and  PE are not resistant    Avoid Pressure limit for  PEEK decreases    OK 76 03 9 200 927 2   OK 176 05 1 200 929 3   OK 57 13 6 200 315 5                AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA       7 Reference information    Exposure     gt  1day    Avoid    OK  OK    OK  OK  OK  OK  OK    OK    OK    Avoid    Avoid    Avoid    OK    OK    OK    OK       7 2 Chemical resistance    Chemical    Pyridine    Sodium acetate    Sodium bicarbon   ate    Sodium bisulph
26. Operating Instructions 28 9597 57 AA 39       4 Operation  4 6 Performing a run    a Place the sample in chosen liquid container  vials for the autosampler or fill the  capillary loop or the Superloop with sample depending on chosen method     Make sure that no air enters the tubing  Place the tubing close to the bottom of  the liquid container but not too tight against the bottom     b Secure the tubing     7 Final check    Perform a final check that tubing  columns  solutions and buffers are placed according    to the method   8 Start the run    Click START to start the run on the selected systems     Viewing the run    The progress of the run can be viewed in detail in the System Control module     Method   q      System1 Wonbytare Capto Capto QYObM bsmax dead volume m01    Q  System Control 1   System 1       Result   q      System1 Wonbytare Capto Capto QV bsmax001       fF     Temp    EN             p           6     Controlled By  Administrator       SamplePres    No watch             Acc  Time Block Time          Pure  sample              File View Manual System Help                   Run Hold  Pause  Continue End        0            Instruments pn Run Status I Volume Block Volume  mm uv     Cond  and         Conc     Pressure        Flow                   TI  nze       s Port  amp  Port 8 Port S Port  Dy y v  OL OF OMG  9  Column Coltmn Q  t  et Buffer      Salve    Galve    Valve             corr Gradient TUTAB  TUTT    8 00 ml Gradient 0 0 B  0 00 ml    321 ml Paus
27. RN Manager window  see Figure 4 1     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 35          4 Operation  4 5 Setting up a run    The New Method dialog opens     New Method       2 If several systems are available  select which system to use in the For system list  box     3 Select Wizard to create a method using the Method Wizard  Click OK     36 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    4 Operation  4 5 Setting up a run    The Method Wizard dialog opens     Method Wizard for System  AA E10S  Main Selection  Column and BufferPrep    Main Selection     n a                        Column  Any    Column Position  Position  Bypass     Flexible Flow Rates NOTE  Different Flow Rates are Set on Variable Page  C Flow Regulation of the System Pump    BufferPrep       4 Inthe Method Wizard dialog     e Select the appropriate chromatographic technique in the Main Selection list box   for example Affinity       Select the column you intend to use in the Column list box  The correct column  volume  the recommended flow rate  and the correct pressure limit for that  column will then be automatically implemented in the method       Select column position in the Column Position list box     e If required  select Flexible Flow Rates and or Flow Regulation of the System  Pump and or BufferPrep     5 Click Next     6 Oneach new page in the Method Wizard  select the appropriate parameters and  click Next to continue     AK
28. TAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 37       4 Operation  4 5 Setting up a run    7 On the last page  click Finish   The Run Setup dialog opens with the Variables tab selected by default     Evaluation Procedures   Method Information   Start Protocol   Questions   Result Name    Frac 950 Variables Scouting   Notes    Gradent    BufferPrep Columns Reference Curves    Value Range  HiTrap  Protein  amp  HP  5 ml Global       penser Tt Cabin  Presse MPa  T NN              tart  Inst    pT vert 2    OFF    ma  _Sample_PressLimit          PressLimit Im an j  00   1 00    na lil   seme                                 3  eS  pese  Naini sna  ae ee                          C Show details  CI Show unused variables   v  Display tooltip for extended variable cells E dit Variable                   8 The method is represented by a number of blocks on the Variables tab  The blocks  are typical steps in a chromatographic run     Each block contains a number of method variables  If necessary  change the variables  to suit your application     9 Inthe Method Editor  select File Save As to save the method   The Save As dialog opens     Save As      c  SDefault       RL ETUXT2  Technique    ny e     B    Anion Exchange  Cation Exchange  RPC    HIC  Size Exclusion      Affinity  Method name      Chromatofocusing                                      8 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                               
29. a run    2 The New Method icon opens the Method Editor module and displays the New Method di   alog box    3 The System Control icon activates the System Control module and displays the Manual  instruction dialog box              Figure 4 1  The UNICORN Manager window     Control system in UNICORN    To open the System Control module in UNICORN  click the System Control icon in the  UNICORN Manager window  see Figure 4 1     4 4 Preparations before start    Prepare buffers  solutions and  inlets    1 Prepare buffers and solutions required for the run     32 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                   4 Operation  4 4 Preparations before start    2 Immerse all inlet tubing in the appropriate liquid containers as described in the  method     Purging the pump and inlet  tubing  Fill the pump and inlet tubing with liquid if small amounts of air need to be removed or  if the inlet tubing is empty   To fill the inlet tubing manually in System Control   1 Make sure that no method has been started   2 Setalow flow in System Control Manual Pump Flow  for example 0 5 ml min   3 Click Execute   4    Set the inlet valve to the appropriate position in System  Control Manual Flowpath InletValve  for example inlet A11 to A18 or A2     5 Connect a syringe to the purge valve        AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 33       4 Operation  4 4 Preparations before start    7 Slowly draw solution into t
30. and containers shall be secured and sealed to prevent  accidental spillage     CAUTION    Make sure waste container is dimensioned for maximum possible  volume when the instrument is left unattended             KTAexplorer    KTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 15                                           NOTICE    Avoid condensation by letting the unit equilibrate to ambient  temperature     2 Safety instructions  2 1 Safety precautions             Using flammable liquids    When using flammable liquids with the AKTAexplorer  AKTApurifier and AKTAmicro    instrument  follow these precautions to avoid any risk of fire or explosion     WARNING    Fire Hazard  Before starting the system  make sure that there is  no leakage     WARNING    A fume hood or similar ventilation system shall be installed when  flammable or noxious substances are used        WARNING    Always use appropriate personal protective equipment during  operation and maintenance of AKTAexplorer  AKTApurifier and  AKTAmicro system     WARNING    Hazardous substances  When using hazardous chemical and  biological agents  take all suitable protective measures  such as  wearing protective glasses and gloves resistant to the substances  used  Follow local and or national regulations for safe operation  and maintenance of the system        AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA             Personal protection       16                                  
31. anufacturer s  instructions  Wait for the computer to start up     Verify that the power indicator on the CU 950 is on when the computer has been  turned on     Log on to Windows     Start UNICORN by double clicking on the UNICORN shortcut icon on the Windows  desktop     Ab  a7    UNICORN    In the Logon dialog  select a user from the User name list and enter the password   If you log on for the very first time  select user default and enter the password default   Click OK     IINE          e UNICORN logon    User name   default Y       UNICORN starts and the UNICORN Manager window opens  see Figure 4 1     Note  See the UNICORN user documentation for instructions about how  to create new users     1    AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 31    4 Operation  4 3 Starting the control system                IE Methods E E Results  CA    default CA    default    Size Type Modified Created Name  Prev Folder             E  AC 1 sample ETT 248KB Method File 02 19 2009 09 22 02 20 2009 13 39  Boz162009   5  6 4 samples ETT 261KB Method File 02 24 2009 10 51 02 24 2009 10 51 C3o2172009  iz  AC 5 samples    ETT 265KB Method File 02 24 2009 13 14 02 24 2009 10 51  03102009  is Immunitettesti ETT 259KB Method File 05 05 2009 17 16 05 06 2009 13 37  03112009        EIT 241KB Method File 05 06 2009 13 47 03 11 2009 17 31  05062009        05122009    No  Description   1 The Instant Run icon immediately starts the system control wizard used to start 
32. arameter       IP20    100 120 220 240  V    50 to 60 Hz    600 VA    T6 3 AL 250V    450 x 490 x 610  mm    41kg     4 to  40   C    10 to 95     84 to 106 kPa  840  to 1060 mbar           IP20    100 120 220 240  V    50 to 60 Hz    370 VA    T 6 3 AL 250 V    450 x 480 x 610  mm    55 kg     4 to  40   C    10 to 95     84 to 106 kPa  840  to 1060 mbar        IP20    100 120 220 240  V    50 to 60 Hz    600 VA    T6 3 AL 250V    450 x 480 x 610  mm    66 8 kg     4 to  40   C    10 to 95     84 to 106 kPa  840  to 1060 mbar        Ingression protec   tion    Supply Voltage    Power consump   tion    Fuse specification    Dimensions  H x W  x D     Weight    Ambient tempera   ture    Relative humidity  tolerance  non   condensing     Atmospheric pres   sure       7 2 Chemical resistance                                  Chemical Exposure Exposure CAS no  Comments   gt  1 day up to 2  months  Acetaldehyde OK OK  Acetic acid   lt  5  OK OK                         AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 59          Comments       FFKM  PP and PE  swell     PVDF is affected by  long term use     Silicone is affected  by long term use     Kalrez     CTFE  PP  and PE are affected  by long term use     PVDF is affected by  long term use     ETFE  PP  PE and  PVDF are affected  by long term use           200 580 7    200 835 2    231 635 3    235 186 4    231 984 1    249 576 7    200 664 3    200 837 3       64 19 7    75 05 8    7664 41 7    1
33. ate  Sodium borate  Sodium carbonate  Sodium chloride    Sodium hydroxide   2M    Sodium sulphate    Sulphuric acid  dilut   ed    Sulphuric acid   medium concentra   tion    Tetrachloroethy   lene    Tetrahydrofuran    Toluene    Trichloroacetic  acid  1     Trifluoroacetic acid   1     Urea  8M       62                                                       7 Reference information  7 2 Chemical resistance       Chemical Exposure Exposure   Comments   gt  1 day upto 2  months  o Xylene and p Xy    OK Avoid PP and PE are af   lene fected by long term  use     7 3 System recommendations    Refer to AKTAexplorer  AKTApurifier and AKTAmicro User Manuals  or contact your local  GE Healthcare representative for the most current information     7 4 Literature    For further information regarding the AKTAexplorer  AKTApurifier and AKTAmicro system   refer to the following     e  AKTAexplorer  AKTApurifier and AKTAmicro User Manuals   e    AKTAexplorer  AKTApurifier and AKTAmicro Safety Handbooks  e  AKTAexplorer  AKTApurifier and AKTAmicro Installation Guides  e UNICORN User Manual    7 5 Ordering information    For ordering information visit www gelifesciences com AKTA     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 63       A Electrical and communication connections    Appendix A Electrical and  communication connections                                                 No  Description No  Description   1 Fraction collector  optional  5 CU 950   2 Ma
34. ation    This chapter provides instructions for the use of AKTAexplorer  AKTApurifier and  AKTAmicro     4 1 Operation overview    Workflow    The typical workflow in AKTAexplorer  AKTApurifier and AKTAmicro  after turning on the  system and connecting it to UNICORN  can be divided into a number of steps     Step Action Section    Create a method Create a method  on page 35  Prepare the system for a run Section 4 4 Preparations before start  on page 32  Start a run using a method Section 4 6 Performing a run  on page 39    During a run   view and change parameters   Viewing the run  on page 40  Procedures after a run Section 4 7 Procedures after a run  on page 42    Evaluate the results See UNICORN user documentation              1    2    3       Liquid flow path    See Appendix B Connection diagram   Liquid flow path  on page 65 for an illustration of  the liquid flow path in   KTAexplorer  AKTApurifier and AKTAmicro     4 2 Starting the instrument    Ensure that the waste container and needed buffer bottles are correctly connected   Check that all tubing connections are properly tightened and that all valves are connected  to a tube or termination     Turn on the Power switch on the instrument     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    30                         4 Operation  4 3 Starting the control system    4 3 Starting the control system    Starting UNICORN    Turn on the monitor  computer and optional printer according to the m
35. covers  or replace parts unless specifically stated in the user  documentation     WARNING    Disconnect power  Always disconnect power from the instrument  before replacing any component on the instrument  unless stated  otherwise in the user documentation     WARNING    Hazardous chemicals during maintenance  When using hazardous  chemicals for system or column cleaning  wash the system or  columns with a neutral solution in the last phase or step           AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 43                         WARNING    Do not perform any type of maintenance work while the system is  powered electrically or when the piping system is pressurized  Note  that the piping system can be pressurized even when the system  is closed down     NOTICE    When using hazardous chemicals  take all suitable protective  measures  such as wearing protective glasses and gloves resistant  to the chemicals used  Follow local regulations and instructions for  safe operation and maintenance of the system     NOTICE   Cleaning  Keep the instrument dry and clean  Wipe regularly with  a soft damp tissue and  if necessary  a mild cleaning agent  Let the  instrument dry completely before use                    5 Maintenance  5 1 General    5 2 User maintenance schedule    Table 5 1 provides a guide to maintenance operations and intervals at which these  operations should be performed by the user  The user is however responsible for deciding  the type
36. dues in the  flow path from previ   ous runs  Air in the  eluents     Residue in the column  from previous runs    Incorrect mixer func   tion    Bad UV fiber connec   tions    Dirty UV cell    Impure buffer    Air in the pump or in  the UV cell       6 1 UV curve problems    Error symptom    Ghost peak    Noisy UV signal  sig   nal drift or instability       52                         6 Troubleshooting  6 1 UV curve problems    Corrective action    Check the lamp run time according  to and replace if necessary  Refer  to AKTAexplorer  AKTApurifier and  AKTAmicro User Manuals     Check that the lamp position and  the filter position are both set to  the wavelength to be used  280 nm  or 254 nm  Refer to AKTApurifier  and AKTAmicro User Manuals   Does not apply to AKTAexplorer  and AKTApurifier without UPC     Calculate the theoretical extinction  coefficient of the protein  If it is ze   ro or very low at 280 nm  the pro   tein cannot be detected        Corrective action    Use a flow restrictor after the flow  cell     Tighten the connectors  If neces   sary  replace the connectors     Check the mixer function by plac   ing a stirrer bar on top of the mixer  housing  The stirrer bar should ro   tate when the system is in Run  mode  The mixer function can also  be checked by running the installa   tion test     Clean the conductivity cell by  flushing 1M NaOH or 20  ethanol     Equilibrate the column  If neces   sary  clean the column  Refer to  AKTAexplorer  AKTApurifier and
37. e  V1  14 Outlet Valve  V4   7 Restrictor 15 To Fraction Collector  optional   8 Sample Valve  V5                    66 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                  B Connection diagram   Liquid flow path    Liquid flow path for AKTApurifier  10 and 100    G7  H7  d       LI          LI          a       LI  L  LI  LI  LI          No  Description No  Description   1 Pump A 7 Bypass   2 Pump B 8 Column   3 Mixer 9 Flow restrictor   4 On line filter 10 To Fraction Collector  optional    5 Injection Valve  V1  11 Cond  optional   Connected after  the UV cell    6 Sample loop 12 pH  optional   Connected after  the conductivity cell           67       AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                            B Connection diagram   Liquid flow path    Liquid flow path for AKTAmicro  G3       9              5  UV COND 10  A2 W1   No  Description No  Description  1 Pump A 6 Sample loop  2 Pump B 7 Bypass  3 Mixer 8 Column  4 On line filter 9 Flow restrictor  supplied   5 Injection Valve  V1  10 To Fraction Collector  optional                    68 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                   6 Tubing    Appendix 6 Tubing    Tubing specifications for  AKTAexplorer    Names in the Label column in Table C 1 and Table C 2 refer to tubing labels in the liquid  flow path connection diagram  see Liquid flow path for AKTAexplorer 1
38. e 2009 07 08  08 28 23 Central European Standard Time  Manual    3 21 ml Continue 2009 07 08  08 28 46 Central European Standard Time  Manual    2 00 ml AutozeroU V    1 00 ml Gradient 100 0 B  0 00 ml    0 90 ml Gradient 0 0 B  0 00 ml   0 00 ml InjectionMark    4 00 ml End Block   4 00 ml Block Pure sample   4 00 ml Base SameAsMain   4 00 ml New Chromatogram Absmax   4 00 ml Gradient 100 0  B  0 00 ml     For Help  press F1   Pun    40 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                              4 Operation  4 6 Performing a run    Up to four view panes  Run Data  Curves  Flow scheme and Logbook can be displayed  showing different aspects of the run in real time       The Run Data view pane displays the current values for selected run parameters   e The Curves view pane displays the monitor signal values graphically       The Flow scheme view pane displays a graphical representation of the  chromatography system that shows the current status of the run  During a run  the  flow scheme shows open flow path s  in color and monitor signals with numerical  displays     e The Logbook view pane shows the actions as the run proceeds  All actions and  unexpected conditons are logged  with date  time and current user name  The log  book provides a complete history of the run and is saved in the result file     Customize the view panes    To customize the view panes  right click in the respective view pane and select Proper
39. ected properly     Checkthat the electrode is correct   ly inserted in the flow cell and  if  necessary  hand tighten the nut     Check that the pH electrode is not  broken     Calibrate the pH electrode     Clean the pH electrode if required   see Monitor pH C 900 User Manu   alor UPC 900 User Manual     Compare the response of the pH  electrode with that of another pH  electrode  If the response differ  greatly  the electrode may require  cleaning or replacement     In organic solvents such as  ethanol  methanol and acetonitrile   stable pH measurements are not  possible since dehydration of the  membrane will occur  It is recom   mended that the pH electrode is  not used in applications using or   ganic solvents  Mount the dummy  electrode instead     Incorrect pump or Check that the pump and valves  valve operation operate correctly   Air in the flow cell If air in the flow cell is suspected     tap the flow cell carefully or tilt it  to remove the air  Alternatively   flush the cell with buffer at 20  ml min  AKTAexplorer 100 and AK   TApurifier 100  or 10 ml min  AKTA   explorer 10 and AKTApurifier 10   or 0 5 ml min  AKTAmicro  for 1 2  min  Use the flow restrictor FR 902  after the pH electrode     Static interference                AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 57             Corrective action    There may be interference from  static fields  Connect the pH flow  cell and the rear panel of the  monitor using a standard 
40. et 49           SealiBRtlonimeebeedemiibereereed idunt 50  5 8       eusanecnauasteariacensaa ersereaaanoureer esate EE 50  b          latest uates ior Tireni Eh UY Cave E EE Teni enaiis 52  61          CUPVS                       52  6 2 Conductivity curve problems        ecsssesssessssssccssssssssssssesssssccssssssssnssssssessssseccssssnsssssessecseccessssssseaeeses 53  6 5   DphHic  rve problems  snae eret Parere ere be RR i derer efe 55  6 4   Pressure curve problems       58    AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 3    Table of Contents       7 Reference         59  A1     ce adam leoi qd ced 59  7 2 Chemical fesIStarice etre ice ter tet b ette re ert eoe Harb d ttende Ioue 59  GS System recommendations         63  74 Literature              sss   63  7 5 Ordering information    63  A Electrical and communication connections                        eere enne 64  B Connection diagram  gt  Liquid flow path   nece rrretrrrerrettn cene o rho reioas 65      TUDNO  gt  69    4 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    1 Introduction    1 Introduction    Purpose of the Operating  Instructions    The Operating Instructions provides you with the instructions needed to handle the  AKTAexplorer  AKTApurifier and AKTAmicro systems in a safe way     Prerequisites    In order to operate the systems in the way they are intended  the following pre requisites  must be fulfilled     e You should have a general 
41. he syringe  When fluid starts to enter the syringe  continue  to draw a few milliliters before closing the purge valve  Check that there is no visible  air left in the tubing        8 Repeat for the other purge valve   9 To fill inlet B1 and B2     a In System Control Manual Pump Gradient  select Target 10096 B and inlet B1  to fill B1 or inlet B2 to fill B2  Wait for the valve to turn  a clicking sound  before  starting the purging procedure     b When all inlets are filled  click End     Connect columns and  Superloop       For column positions  see the method     Remove air before connecting columns    Air remaining in the system may be removed by purging the pump and by selecting  Pump Wash and System Wash     1 Immerse A1 tubing in the buffer to be used   2 Select System Control Manual Pump PumpWash     Connecting tubing to columns  Refer to column manufacturer s instructions     34 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA       4 Operation  4 4 Preparations before start    Column attachment drop to drop          Attach columns manually by starting a low flow  see Purging the pump and inlet tubing   on page 33  and selecting System Control Manual Flowpath ColumnPosition     Preparing the fraction collector    Place the rack chosen in the method in the fraction collector and fill it with appropriate  tubes and or deep well plates     4 5 Setting up a run    Create a method  To create a method     1 Clickthe New Method icon in the UNICO
42. ins supply socket 6 UniNet 1 cable   3 UniNet 1 cable 7 USB cable  to computer   4 Power cord 8 Power converter                   Figure A 1  Electrical and communication connections for AKTAexplorer  AKTApurifier and  AKTAmicro    64 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                     B Connection diagram   Liquid flow path    Appendix B Connection diagram   Liquid  flow path    Liquid flow path for  AKTAexplorer 10             A11 A12 A13 A14 A15 A16 A17 A18    No  Description No  Description   1 Buffer Valve  V6  7 Column Selection Valve  V2    2 Pump A 8 Column Selection Valve  V3    3 Pump B 9 Flow restrictor   4 Mixer 10 Outlet Valve  V4    5 On line filter 11 To Fraction Collector  optional   6 Injection Valve  V1                    AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 65                            8 Connection diagram   Liquid flow path                Liquid flow path for  AKTAexplorer 100  65  63 z    I 2  Gi2 1  6   ETA O uS   KA  62 COND     UV   TN   9 T  w T 5  7 15  8 UYU  9 60 we w  619  8 ES 14  YIN     As  NOS Nay  UU VYOUVVVVVV U                  A11 A12 A13 A14 A15 A16 A17 A18 S1 S2 S3 S4 S5 S6 S7 S8 W3     F4 FS F6 F7 FB   No  Description No  Description  1 Buffer Valve  V6  9 Sample Pump  2 Pump A 10 Flow Direction Valve  V7   3 Pump B 11 Column Selection Valve  V2   4 Mixer 12 Restrictor  5 On line filter 13 Column Selection Valve  V3   6 Injection Valv
43. ired to lift the system safely     NOTICE    Disconnect power  To prevent equipment damage  always  disconnect power from the AKTAexplorer  AKTApurifier and  AKTAmicro instrument before an instrument module is removed  or installed  or a cable is connected or disconnected     NOTICE    AKTAexplorer  AKTApurifier and AKTAmicro shall be installed and  prepared by GE Healthcare personnel        AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    2 Safety instructions    o OP P b    2 1 Safety precautions       18                                     2 Safety instructions  2 1 Safety precautions          19    WARNING    Hazardous chemicals during run  When using hazardous  chemicals  run System CIP and Column CIP to flush the entire  system tubing with distilled water  before service and maintenance     WARNING    Hazardous biological agents during run  When using hazardous  biological agents  run System CIP and Column CIP to flush the  entire system tubing with bacteriostatic solution  e g  NaOH   followed by a neutral buffer and finally distilled water  before service  and maintenance     CAUTION    Hazardous chemicals in UV flow cell  Make sure that the entire  flow cell has been flushed thoroughly with bacteriostatic solution   for example NaOH  and distilled water  before service and  maintenance     NOTICE    Do not run Column CIP if using silica based packing material or  RPC columns  Remove the column from the system during CIP     WARNING    
44. ked     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                   AKTAexplorer  AKTApurifier and AKTAmicro are high pressure liquid chromatography  systems for use in laboratory scale production of biomolecules                    Figure 1 1  The main parts of the instruments  AKTAexplorer is shown as an example     Part Function    Monitor pH C 900   Monitor UV 900   Pump P 900  P 901 alt  P 903   Power switch   Switch valve  SV 903   Conductivity cell    UV cell       AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA          1 Introduction  1 3 Instrument    1 5 Instrument    Product description    1  2  3       10                         1 Introduction  1 5 Instrument    Part Function    8 Mixer M 925   9 Sample pump P 960  This is only standard in some versions    10 Valve door   11 Column selection valve  V3  PV 908    12 Sample valve  V5  PV 908    13 Injection valve  V1  INV 907    14 Column selection valve  V2  PV 908    15 Flow direction valve  V7  INV 907   This is only standard for AKTAExplorer  100 and AKTAExplorer 100 Air   16 Outlet valve  V4  PV 908    17 Buffer valve  V6  INV 908    18 Box 900             Detailed information on the components included in each system can be found in their  respective User Manual     Basic flow path                                                                                                                      T  T ii    5 WAY    W           Ln               
45. laborato   ry 4mm    banana plug    cable     Replace the pH electrode        Corrective action    Check all connections for leaks     Check that there is sufficient eluent  present in the reservoirs     Use degassed solutions   Purge the pump     Follow the instructions in Pump P   900 User Manual     Clean the valves in according to  Pump P 900 User Manual     Replace the piston seal according  to the instructions in Pump P 900  User Manual     Flush through to clear blockage   If necessary  replace tubing     Check inlet tubing filter  It can be   come clogged if unfiltered buffers  or samples are applied  See instruc   tions for flushing through at the  end of the run in Pump P 900 User  Manual        AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA          Possible cause    Problem with the  electrode    6 4 Pressure curve problems    Possible cause    Air bubbles passing  through or trapped in  the pump    Inlet or outlet check  valves not functioning  correctly    Piston seal leaking    Blockage or part  blockage of flowpath          6 Troubleshooting  6 3 pH curve problems    Error symptom    pH values vary with  varied back pressure    Error symptom    Erratic flow  noisy  baseline signal  irregu   lar pressure trace                            7 Reference information    7 Reference information    This chapter contains technical data  regulatory and other information     AKTApurifier    AKTAmicro    AKTAexplorer    7 1 Specifications    P
46. ls on liquid flow path  see Appendix B Connection diagram   Liquid flow  path  on page 65    UNICORN is a complete software for control and supervision of AKTAexplorer  AKTApurifier  and AKTAmicro  The software runs under Microsoft    Windows operating system     UNICORN is supplied with a method wizard which provides easy creation of methods  for purification     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA    1 Introduction  1 3 Instrument    1 4 Control software    UNICORN    control software    12          1 Introduction  1 4 Control software    For more information about UNICORN control system  see the UNICORN user manuals  supplied     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 13       2 Safety instructions    2 Safety instructions    This chapter describes safety compliance  safety labels  general safety precautions   emergency procedures  power failure and recycling of AKTAexplorer  AKTApurifier and  AKTAmicro     2 1 Safety precautions    Introduction    The AKTAexplorer  AKTApurifier and AKTAmicro instruments are powered by mains  voltage and handles pressurized liquids that may be hazardous  Before installing   operating or maintaining the system  you must be aware of the hazards described in  this manual  Follow the instructions provided to avoid personal injury or damage to  the equipment     The safety precautions in this section are grouped into the following categories     General precautions
47. n Section 4 6 Performing a run  on page 39     Cleaning columns    When running different types of purification methods and different samples after each  other  the columns should be cleaned between the runs according to the column  instructions  This will remove unspecific bound proteins and prevent column clogging     1 Inthe Method Editor module in UNICORN  create a method for column cleaning in  place  CIP      2 Immerse the tubing ends to be used in the correct containers according to the method  for the chosen run     3 Run the cleaning in place method as described in Section 4 6 Performing a run  on  page 39     42 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                5 Maintenance    5 Maintenance    This chapter provides instructions for routine component maintenance and a  maintenance schedule     5 1 General    Regular maintenance is important for safe and trouble free operation of your instrument   The user should perform daily and monthly maintenance  Preventive maintenance should  be performed on a yearly basis by qualified service personnel     For maintenance of a specific component  carefully read the component manual and  follow the instructions  To avoid personal injury when performing maintenance on the  AKTAexplorer  AKTApurifier and AKTAmicro instrument  follow the instructions below     WARNING    Electrical shock hazard  All repairs should be done by service  personnel authorized by GE Healthcare  Do not open any 
48. nation at work    
49. on 1800   9 11 9 x 103  Switch valve   pump A3  B3 Teflon 150 2 9 991  Pump   Mixer G1  G2 PEEK 300 0 75 133  Mixer   On line filter G3 PEEK 150 0 75 66  Filter   V1  Inj  Valve  G4 PEEK 460 0 75 203  V1 V7 G5 PEEK 470 0 75 208  V7   V2 G6 PEEK 410 0 75 181  Bypass G7 PEEK 620 0 75 274  V3   V7 G8 PEEK 470 0 75 208  V7 loop G9 PEEK 180 0 75 80  V7   UV G10 PEEK 370 0 75 163  UV   Cond  Cell G11 Tefzel 160 0 75 71  Cond  Cell   pH Cell G12 Tefzel 450 0 75 199  pH Cell   Flow restrictor   G13 Tefzel 110 0 75 49  Flow restrictor   1 G14 Tefzel 120 0 75 53  V4   Fraction collector G15 Tefzel 600 0 75 265  Flow through F3 Tefzel 1000 1 0 785  V1   Waste W1 Tefzel 1300 0 75 574                         70   KTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                                                                         6 Tubing       Material Length          Volume   mm   mm   ul   Sample pump   Waste w2 Teflon 1300 1 6 2 6 x 10   V4   Waste W3 Tefzel 1000 1 0 785    1 Not mounted at factory    Tubing specifications for  AKTApurifier  Names in the Label column in Table C 3 and Table C 4 refer to tubing labels in the liquid    flow path connection diagram  see Liquid flow path for AKTApurifier 10 and 100  on  page 67     Table C 3  Tubing specifications for AKTApurifier 10                            Material Volume    ul   Inlet A1 Al Teflon 1700   16 3 4 x 10   Inlet A2 A2 Teflon 1900   16 3 8 x 103  Inlet B1 B1 Teflon 1500   16 3 0 x 10   Inlet B2
50. or measurement  control   CAN CSA C22 2 and laboratory use   No  61010 1    8   KTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                               1 Introduction    1 2 Regulatory information    Notes    Harmonized with  2004 108 EC    Harmonized with  2006 42 EC    Harmonized with  2006 42 EC          Description    EMC emissions and immunity require   ments for electrical equipment for  measurement  control and laboratory  use    Safety of machinery   Basic concepts   general principles for design    Safety of machinery   Principles of  risk assessment       Standard    EN 61326 1    EN ISO 12100 1   12100 2    EN ISO 14121 1   14121 2       The CE marking and the corresponding Declaration of conformity is valid for the    instrument when it is     e used as a stand alone unit  or      connected to other CE marked instruments  or    e connected to other products recommended or described in the user documentation     and    e usedinthe same state as it was delivered from GE Healthcare  except for alterations  described in the user documentation     The Declaration of conformity is valid only for systems that are marked with the    CE marking     CE    CE Marking    Regulatory compliance of  connected equipment    Any equipment connected to the AKTAexplorer  AKTApurifier and AKTAmicro systems  should meet the safety requirements of EN 61010 1 IEC 61010 1  or relevant harmonized    standards  Within EU  connected equipment must be CE mar
51. r     2 connect a termination plug to the empty UniNet 1 socket  Frac 950 only      Flow path  All parts and tubing are mounted on the AKTAexplorer and AKTAmicro systems at delivery     AKTApurifier has no preconnected tubing  It is recommended that the mounting is done  by GE Healthcare service engineers  For more information regarding installation  see  AKTApurifier User Manual     Setting the delay volume in UNICORN    The delay volume is volume of liquid in the flow path from the UV sensor that identifies  the peakto the fraction collector  The length of tubes affects the delay volume that needs  to be changed in UNICORN     To change the delay volume in UNICORN   1 Select System Settings in System Control   2 Select Specials and then FracParameters     3 Enterthe delay volume and click OK     Note  To prevent bacterial growth  the system flow path is filled with 2096  ethanol at delivery     e Connect tubes for reagents  solvents and sample collection to the correct inlet and  outlet connections on the system  For more information  see AKTAexplorer   AKTApurifier and AKTAmicro User Manuals     Electrical power    Connect the power cord to a grounded power outlet specified in Section 5 1 Site  requirements  on page 26     5 6 Spare parts and accessories    For correct up to date information on spare parts and accessories visit   www gelifesciences com AKTA    AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 29             4 Operation    4 Oper
52. r  Insufficient mixing    chamber volume    Incorrect motor func   tion    Air in the flow cell    Dirty tubing    Incorrect mixer vol   ume       6 Troubleshooting    6 2 Conductivity curve problems    Error symptom    Conductivity measure   ment with the same  buffer appears to de   crease over time    Waves on the gradi   ent    Ghost peaks appear  in the gradient profile    Unlinear gradients or  slow response to  B  changes       54                               6 Troubleshooting  6 2 Conductivity curve problems    Corrective action    Check that the conductivity flow  cell cable is connected properly     Check that the pump and valves  operate correctly     If temperature compensation is  being used  check that the temper   ature sensor is calibrated  and that  the correct temperature compen   sation factor is in use     Check that the column is equilibrat   ed  If necessary clean the column     Check the operation of the mixer   The mixer function is checked by  placing a stirrer bar on top of the  mixer housing  The stirrer bar  should rotate when the system is  in Run mode  The mixer function  can also be checked by running  the installation test        Corrective action    Check that the electrode cable is  connected properly     The electrode glass membrane  may be cracked  Replace the elec   trode     Check that the pH monitor is cor   rectly connected according to the  User Manual     Clean the pH electrode as detailed  in Monitor pH C 900 User Manual  or UPC 900
53. rol unit between unit and computer  e Various buffer and sample bottles    Installing the pH electrode    Install the pH electrode in the flow cell according to the picture below     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 27          3 Installation  3 4 Assembly       Description 5 Description  1 To instrument rear  pH probe  5 Flow cell holder  2 Electrode 6 Flow cell  3 Nut 7 End cover  4 Dummy                   The flow cell holder can be placed on either the optical unit  for AKTApurifier  or on the  outside of the valve door  for AKTAexplorer   For more information on installation  see  AKTApurifier User Manual and AKTAexplorer Installation Guide     3 5 Connections    Communication    Connect the network  signal cables and computer according to the electrical drawings  in Appendix A Electrical and communication connections  on page 64     Ensure that UNICORN control software is installed on the computer   Installing Controller unit CU 950    Hang the CU 950 on the left side of the system by inserting the hooks on the front of  CU 950 into the channel on the side of the UV 900     28 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                3 Installation  3 5 Connections    Connect according to diagram in Appendix A Electrical and communication connections   on page 64    When using a fraction collector     1 connecta UniNet 1 cable between Monitor UPC 900 Monitor UV 900 and the fraction  collecto
54. s contain hazardous substances   Detailed information is available from your GE Healthcare representative     24 AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA                      2 Safety instructions  2 4 Recycling information    Disposal of electrical  components    Waste of electrical and electronic equipment must not be disposed as unsorted municipal  waste and must be collected separately  Please contact an authorized representative  of the manufacturer for information concerning the decommissioning of equipment     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 25       3 Installation    3 Installation    NOTICE  9 AKTAexplorer  AKTApurifier and AKTAmicro shall be installed and    prepared by GE Healthcare personnel           AKTAexplorer  AKTApurifier and AKTAmicro are delivered in protective packing material  and shall be unpacked with great care     Any equipment connected to AKTAexplorer  AKTApurifier and AKTAmicro must fulfill  applicable standards and local regulations    For detailed information on Installation  see AKTAexplorer  AKTApurifier and AKTAmicro  Installation Guides     3 1 Site requirements    Parameter Requirement             Electrical power 100 240 V  50 60 Hz   Ambient temperature 4 C to 40 C   Placement Stable laboratory bench min  200 x 80 cm  Humidity 20 to 9596  non condensing    3 2 Transport    The equipment weights are specified in the table below  Each system requires at least
55. ties   For more information about customizing the view panes  see the UNICORN user  documentation     Ending the run  To stop the run on a system before it is finished   Click End above the Run data view pane     Status indicator colors  The status indicator is located at the bottom of System Control   The table below shows how the indicator colors relate to the run status     Indicator color Run status    White End   Green Run or Manual  Yellow Hold   Red Pause               KTAexplorer    KTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 41                      4 Operation  4 6 Performing a run    Error indication    When a warning or an alarm is issued from a system  an error code is displayed  See  AKTAexplorer  AKTApurifier and AKTAmicro User Manuals for guidance     Evaluate the results    See AKTAexplorer  AKTApurifier and AKTAmicro User Manuals and UNICORN user  documentation for how to evaluate the results     4 7 Procedures after a run    Cleaning system    To keep the system in good shape  it is important to clean both the tubing and the outside  of the system regularly     1 Inthe Method Editor module in UNICORN  create a method for cleaning the system   2 Wash the outside of the inlet tubings with water and or ethanol    3 Immerse the tubing ends to be used in the container with cleaning solution   4    If the column valve is to be cleaned  remove the columns and reconnect the tubings  to the column valves     5 Run the cleaning method as described i
56. tion valve manually to  Waste position  Put the open end in a waste  container     3 Run the pump manually at 10 ml min with  water  Note the back pressure  Bp1  on the  pump display  or in the Run Data window     4 Setthe system to Pause and connect the  flow restrictor to the open end of the capil   lary lobserve the IN marking   Put the flow  restrictor in the waste container     5 Press Continue so that the pump run at 10  ml min with water  Note the back pressure   Bp2  on the pump display  or in the Run  Data window     6 Calculate the back pressure generated by  the flow restrictor  Bp2 Bp1   Replace it if it  is not within limit     Yearly Valve inspection Check for external or internal leakage  Replace  channel plate and distribution plate yearly or  when required  Refer to the relevant valve in   struction sheet                 5 5 Cleaning    Cleaning before  maintenance service    Before maintenance service is performed  the system owner must first clean the system  and complete a Decontamination Report  Contact GE Healthcare for further information     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 47             5 Maintenance  5 3 Cleaning    Cleaning In Place  All components in the system are designed for ease of CIP     After repeated separation cycles  contaminating material might progressively build up  in the system and on the column  This material may not have been removed by the  cleaning step described above  The nature and
57. understanding of how the PC and Windows    works   e You should understand the concepts of liquid chromatography   e You must read and understand the Safety Instructions     e The instrument and software should be installed  configured and calibrated according  to the Installation Guide     In this chapter    This chapter contains important user information  and a general description of the  AKTAexplorer  AKTApurifier and AKTAmicro systems and their intended use     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 5             1 Introduction  1 1 Important user information    1 1 Important user information    Read this before using the  AKTAexplorer  AKTApurifier and  AKTAmicro system     G     All users must read the safety instructions in the   KTAexplorer    KTApurifier and  AKTAmicro User Documentation to fully understand the safe use of the systems  before  installing  using  or maintaining the systems    Do not operate the AKTAexplorer  AKTApurifier and AKTAmicro systems in any other way  than described in the user documentation  Otherwise  you may be exposed to hazards  that can lead to personal injury  and you may cause damage to the equipment     Intended use    AKTAexplorer  AKTApurifier and AKTAmicro are liquid chromatography systems intended  for protein purification within method development and drug discovery  The systems  can be used to screen for optimal choice of columns  media and running parameters to  purify selected proteins 
58. ur  local GE Healthcare representative     AKTAexplorer  AKTApurifier and AKTAmicro Operating Instructions 28 9597 57 AA 49                            5 Maintenance    5 6 Replacement of fuses    WARNING    For continued protection from fire hazard  replace only with same  type and rating of fuse           5 7 Calibration    The table below lists the type and frequency of calibrations that can be done on the  instrument  Refer to UNICORN user documentation and to the individual component  User Manuals and Instructions for descriptions of how to perform these calibrations   The calibrations are performed from UNICORN by selecting System Calibrate in System    Control    Component How often   pH monitor  if applicable  Every day    Pump  if applicable  Whenever the running conditions are changed   e g  viscosity of sample or buffer  temperature   backpressure etc  If the sample pump is not used  frequently it should be calibrated before use    Pressure reading When required     Conductivity Cell constant   Only necessary if specific conductivity with high  flow cell accuracy is measured  Cond Calib      Temperature Must be done when changing the conductivity  flow cell  Temp      Entering a Must be done when changing the conductivity  new cellcon    flow cell  Cond Cell    stant                5 8 Storage    General recommendation    For storage  the system must first be cleaned as described in Cleaning In Place  on  page 48  After cleaning  the system must be filled with 0 01 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Guida dell`utente della fotocamera  HB31 (BG08) - Manuel d`installation  Gorenje R6192FX refrigerator  中 " ー 一 g。x型  lijadora excéntrica de aire comprimido / smerigliatrice eccentrica ad    OvisLink PCM-RDSI ISDN access devices  Office de tourisme de Cognac  OBJECTIFS Monter le DRC et formuler une réclamation dans les  BEDIENUNGSANLEITUNG PIZZA - TEIGFORMER    Copyright © All rights reserved. 
   Failed to retrieve file