1.     9  If the insulation resistance of the EUT is within 2000M Q  thus   the panel will display the insulation resistance value of the EUT    and the testing is kept going until the RESET key is pressed    If the test on the EUT is disapproval  the FAIL indicator on the   panel will be lighted up and the buzzer will blare out a warning     10    wm    Now remove the high pot probe from the EUT  and press  RESET to turn off the buzzer and FAIL light  then the  instrument will be back to the initial status     Artisan Technology Group   Quality Instrumentation     Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER  USER MANUAL          6  MAINTENANCE    The following instructions are used by qualified person only to avoid  electrical shock  do not perform any service other than contained in the  operation instructions unless you are qualified to do so     6 1  Fuse Rating and type  If the fuse blows  the product will not operate  Try to determine and  correct the cause of the blown fuse  then replace the fuse with  correct rating and type shown as below     WARNING  For continued fire protection  replace only    with 250V fuse of the specified type and rating  and  disconnect the power cord before proceeding fuse    replacement     6 2  Cleaning  To keep the instrument clean  wipe the case with a damp cloth and  detergent  Do not use abrasives or solvents     17    Artisan Technology Group   Quality Instrumentation        Guaranteed    888  88 SOURCE  
2.    A rtisan   Artisan Technology Group is your source for quality  TecmoogyGroup   new and certified used pre owned equipment          FAST SHIPPING AND SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT   DELIVERY Experienced engineers and technicians on staff Sell your excess  underutilized  and idle used equipment   TENS OF THOUSANDS OF at our full service  in house repair center We also offer credit for buy backs and trade ins   IN STOCK ITEMS www artisantg com WeBuyEquipment 7   EQUIPMENT DEMOS   HUNDREDS OF InstraV ea    REMOTE INSPECTION LOOKING FOR MORE INFORMATION    MANUFACTURERS Remotely inspect equipment before purchasing with Visit us on the web at www artisantg com 7 for more  our interactive website at www instraview com   information on price quotations  drivers  technical   LEASING MONTHLY specifications  manuals  and documentation    RENTALS    ITAR CERTIFIED    D a gaa tia Contact us   888  88 SOURCE   sales artisantg com   www artisantg com             ELECTRICAL SAFETY TESTER    USER MANUAL          WARNING  MANUFACTURER ASSUMES NO LIABILITY  IF UNIT OPERATED IN AN UNSAFE  MANNER     WARNING    THIS INSTRUMENT GENERATES AND  DELIVERS A  HAZARDOUSLY HIGH  VOLTAGE  5kV   BE EXTREMELY  CAREFUL WHEN USING THIS  INSTRUMENT  BE SURE TO READ  MANUAL SECTION 3 PRECAUTIONS    ELECTRICAL SAFETY TESTER  USER MANUAL          EC Declaration of Conformity    We  GOOD WILL INSTRUMENT CO   LTD   No  95 11  Pao Chung Rd   Hsin Tien City  Taipei Hsien  Taiwan   GOOD WILL INSTRUM
3.  Electrical Safety Testers comply with the requirement of the  electrical equipment  amp  appliance control ordinances and JIS  CSA  UL   BS and other overseas standards as well  The testers can be used for  withstanding voltage test of the various types of electrical and equipment  and components    The GPT GPI 800 series are based on the family of GW  withstanding voltage tester including AC Withstanding Voltage test  DC  Withstanding Voltage test and Insulation Resistance test  Please refer to    the table as follows                                   Function  Model AC DC IR  GPT 805  500VA  Vv  GPT 815  500VA  Vv Vv  GPI 825  S00VA  Vv V  GPI 826  100VA  Vv V  1    ELECTRICAL SAFETY TESTER  USER MANUAL          1 2  Features  The GPT GPI 800 series offer several other features   1  Current and output voltage setting  A safe way to set the current and output voltage without high  voltage activated   2  Easily and quickly setting with front panel control  A user friendly interface provides user an easy and quick way to  set all parameters   3  Testing time  Use micro processor to control testing time   4  Adjustable ARC detect level  The ARC detect level can be adjusted with front panel control   5  Adjustable output voltage during test  The output voltage can be adjusted during testing that add an  operation flexibility   6  An alert indicator for high voltage  A flashing red LED indicates dangerous situation during high  voltage output is activated   7  Remote I O control
4.  Quality Instrumentation     Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER          USER MANUAL  CONTENTS PAGE  1  PRODUCT INTRODUCTION              cssssssssssssesceesseseeees 1  1 1  Description              ccccceccscccsccccresccssccssscsceseseess 1  1 2 FeaT ese cise ces cccecssacedsesesececseidssesedecacesseeesesatacens 2  2  SPECIFICATION              ccceccscceccscecscecsccecscnceeseees 3  3  PRECAUTIONS BEFORE OPERATION                000 6  3 1 Unpacking the instrument                csccsccsseecsceceees 6  3 2 Checking the Line Voltage              sccscscscssssssceeeees 6  3 3 Environment          sssssossoosssossessoossossesoseosesssoeses 8  4  PANEL INTRODUCTION                 cccscccsssscscsceceeees 9  4 1 Front Pamel              ccccccscscccecececcscccecescsesseceseees 11  4 2 Rear Panel               csccceccsececcscecsceccccscesceesceseces 12  5  OPERATION METHOD               ccccccecscsceceececeeeces 13  5 1 AC Withstanding Test              ccceseccsccesccceceeceeee 13  5 2 DC Withstanding Test              cscccssecsccscccscsceeeees 14  5 3  Insulation Test            ssscccscsscecsccssceccececcsssescscsees 16  6  MAINTENANCE           ccccececcscscsccceccsescecscecscessecnes 17  6 1 Fuse rating and type            ccccececcscecceccsecsceeceeeees 17  6 2  Cleaning           scccsccssseccccccccsccseccscececccseesesceseess 17  iii    ELECTRICAL SAFETY TESTER  USER MANUAL          SAFETY TERMS AND SYMBOLS    These terms may a
5.  www artisantg com       A rtisan   Artisan Technology Group is your source for quality  TecmoogyGroup   new and certified used pre owned equipment          FAST SHIPPING AND SERVICE CENTER REPAIRS WE BUY USED EQUIPMENT   DELIVERY Experienced engineers and technicians on staff Sell your excess  underutilized  and idle used equipment   TENS OF THOUSANDS OF at our full service  in house repair center We also offer credit for buy backs and trade ins   IN STOCK ITEMS www artisantg com WeBuyEquipment 7   EQUIPMENT DEMOS   HUNDREDS OF InstraV ea    REMOTE INSPECTION LOOKING FOR MORE INFORMATION    MANUFACTURERS Remotely inspect equipment before purchasing with Visit us on the web at www artisantg com 7 for more  our interactive website at www instraview com   information on price quotations  drivers  technical   LEASING MONTHLY specifications  manuals  and documentation    RENTALS    ITAR CERTIFIED    D a gaa tia Contact us   888  88 SOURCE   sales artisantg com   www artisantg com             
6. ENT  SUZHOU  CO   LTD    No  69 Lushan Road  Suzhou New District Jiangsu  China   declares that the below mentioned product   GPT 805 815  GPI 825 826   are herewith confirmed to comply with the requirements set out in the Council  Directive on the Approximation of the Law of Member States relating to  Electromagnetic Compatibility  89 336 EEC 92 31 EEC  and Low Voltage  Equipment Directive  73 23 EEC  93 68 EEC    For the evaluation regarding the Electromagnetic Compatibility and Low Voltage  Equipment Directive  the following standards were applied     EN 61326 1  Electrical equipment for measurement  control and laboratory use      EMC requirements  1997 A1  1998 A2 2001        Conducted and Radiated Emissions Electrostatic Discharge  EN 55011 Group I class A  1998 EN 61000 4 2  1995 A1  1998       Current Harmonic Radiated Immunity   EN 61000 3 2  2000 EN 61000 4 3  1996 A1  1998  Voltage Fluctuation Electrical Fast Transients   EN 61000 3 3  1995 EN 61000 4 4  1995   Surge Immunity   EN 61000 4 5  1995  Conducted Susceptibility   EN 61000 4 6  1996   Power Frequency Magnetic Field  TEC 61000 4 8  1993   Voltage Dips  Interrupts   EN 61000 4 11  1994    Safety Requirements IEC EN 61010 1  2001   Remark  Also complied with Continuity of Protective Bonding Tester  Insulation  Resistance Test  Voltage Test  and Residual Voltage Test in accordance with  the Sub Clauses 19 2  19 3  19 4 and 19 5 of EN 60204 1  1997                               ii    Artisan Technology Group  
7. R   BL  AC withstand voltage test   J   DC withstand voltage test  only for GPT 815   or Insulation Test  only for GPI 825 826    16   500V 1000V EL  500V  BL  1000V for GPI 825 826 only   17   High Voltage Output High voltage output terminal  Seat  18   RETURN terminal  The test return terminal   19   Power switch Power on to start test           11             Current Monitor       The cutoff current can be directly monitored by  disconnecting the shorting bar from these terminals  and connecting a milliammeter  DC AC  between  them  The milliammeter should be capable of  measuring the cutoff current  Be sure to connect the  shorting bar when no milliammeter is connected  between these terminals        Artisan Technology Group   Quality Instrumentation     Guaranteed    888  88 SOURCE   www artisantg com       ELECTRICAL SAFETY TESTER  USER MANUAL          5  OPERATION  5 1  AC Withstanding Test    Take GPT 805 as example  if want to output SOOOVAC  set the   cut off current to 100mA and ARC to 50mA  During the 60s test   time  the instrument will keep on testing even when the ARC  appears    1  Connect the Withstanding Insulation Tester to the ground    2  Adjust the    Voltage Knob    counter clockwise to the lowest  voltage position    3  Set power switch to ONL position    4  Set CURRENT ARC to position    5  Set Range to 100mA range    6  Set to I Set position by pressing  8  key  then adjust the  Current VR until the value of 100 0 is displayed on the panel    7  Set 
8. Technology Group   Quality Instrumentation    ELECTRICAL SAFETY TESTER  USER MANUAL          The wire which is coloured Brown must be connected to the  terminal marked with the letter L or P or coloured Brown or  Red    If in doubt  consult the instructions provided with the equipment  or contact the supplier     This cable appliance should be protected by a suitably rated and  approved HBC mains fuse  refer to the rating information on  the equipment and or user instructions for details  As a guide   cable of 0 75mm  should be protected by a 3A or 5A fuse  Larger  conductors would normally require 13A types  depending on the  connection method used    Any moulded mains connector that requires removal   replacement must be destroyed by removal of any fuse  amp  fuse  carrier and disposed of immediately  as a plug with bared wires  is hazardous if a engaged in live socket  Any re wiring must be  carried out in accordance with the information detailed on this  label         Guaranteed    888  88 SOURCE   www artisantg com       ELECTRICAL SAFETY TESTER  USER MANUAL          1  PRODUCT INTRODUCTION    1 1  Description    The GPT GPI 800 Electrical Safety Testers  EST  are designed for  AD DC Withstanding Voltage test and Insulation Resistance test  IR  in  order to provide a safe and accurate test environment for the operator   With thoughtful design described in 1 2  Feature insures a safe operation  environment of high voltage test to protect user from hazardous impact     The
9. elow        T 6 3A ET 3 15A  115V 105 125V 250V 230V 207 250V 250V    WARNING  To avoid personal injury  disconnect the  power cord before removing the fuse holder     Operator   s Precaution    1  With immense high output voltage and current of the puncture  tester  only qualified person can operate the tester in order to  avoid fatal electric shock     2  On job training is required for operator to better use the tester  smoothly and safely     3  The operator is prohibited to dress with metal ornaments or wear  metal decoration in order to avoid electric shock     4  The person with cardiac or wear a pacemaker must not to operate  the tester     Artisan Technology Group   Quality Instrumentation        ELECTRICAL SAFETY TESTER    USER MANUAL          Secure Testing   Never operate the tester in the place with electric circuit device  around    The earth lead should be well connected in accordance with  instruction  The Return Lead has to be connected to the tested object  first before linking up test probe  Do not plug the high voltage test  probe to the high voltage output terminal before doing the testing   Also  do not touch the electric conductor of test probe and the  operator has to fully control the power on off by using switch or  remote control  which should not be lay aside carelessly     WARNING  During the testing  do not touch the  tested object or any other connected objects     3 3  Environment    The normal ambient temperature range of this instrument is fr
10. he EUT is approval  the PASS indicator on the  panel will be lighted up    17  During testing  if the ARC is occurred  both the lights of ARC  and FAIL will be on  and the buzzer will blare out a warning   now remove the high pot probe from the EUT  and press RESET  to turn off the buzzer  ARC and FAIL lights  then the instrument  will be back to the initial status    18    wm    If the test is over the jump out current  the FAIL indicator on the  panel will be lighted up and the buzzer will blare out a warning   Now remove the high pot probe from the EUT  and press  RESET to turn off the buzzer and FAIL light  then the  instrument will be back to the initial status     15    ELECTRICAL SAFETY TESTER    USER MANUAL          5 3  Insulation Test    Take GPT 825 as example  set output to 1000V  but without   specifying the test time    1  Connect the Withstanding Insulation Tester to the ground    2  Adjust the    Voltage Knob    counter clockwise to the lowest  voltage position    3  Set power switch to ONE position    4  Set the key  15  to IR T position    5  Set the Timer to OFF Ef position    6  Connect a test lead from the Return terminal  18  to the EUT   equipment under test     7  Connect a red high pot probe from    High voltage output  terminal  17     to the EUT  equipment under test     8  Press RESET key once to appear STBY  Stand by  on the  display  then  press START key to proceed testing  During  testing  the warning indicator  3  will be flashing continuously
11. ler  In addition to the 9 pin remote I O controller to control  START RESET  it also provides PASS FAIL TEST signal for  professional use     Artisan Technology Group   Quality Instrumentation     Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER  USER MANUAL          2  SPECIFICATION  15  C  35  C RHS75    500VA          ELECTRICAL SAFETY TESTER                   1  AC Hi Pot Specifications  only for GPT 805 815  GPI 825                          Voltage Regulation 15    Voltage Range 0 100 5 00kV   Voltage Accuracy  3  of reading  3 counts  Current cut off range 0 3mA 100mA   Current Accuracy  3  of reading  2 counts                USER MANUAL  100VA  3  AC Hi Pot Specifications  only for GPI 826   Voltage Regulation 15   Voltage Range 0 100 5 00kV  Voltage Accuracy  3  of reading  3 counts       Current cut off range    0 3mA 20mA          Current Accuracy        3  of reading  2 counts          Warning  The test time by using 5000V 100mA is within  a 180s  it needs 15 minutes interval after 180s of continuous    testing  If proceed continuity test  the test voltage and  current must be at 5000V 50mA        2  DC Hi Pot Specifications  only for GPT 815        A    Warning  The test time by using 5000V 20mA is within  180s and it needs 15 minutes interval after 180s of  continuous testing  If proceed continuity test  the test  voltage and current must be at 5000V 10mA        4  Insulation Resistance Specifications  only for GPI 825 826        DC Vo
12. ltage    500V 1000V                         Voltage Regulation 20    Voltage Range 0 100 5 00kV   Voltage Accuracy  3  of reading  3 counts  Current cut off range 0 3mA 10mA   Current Accuracy  3  of reading  2 counts    Resistance Range    1 2000M Q       Resistance Accuracy          1 500M Q   5  of reading  501 2000M Q   10  of reading                Warning  The test time by using 5000V 10mA is within 180s   it needs 15 minutes interval after 180s of continuous testing   If proceed continuity test  the test voltage and current must  be at 5000V 5mA     A    Artisan Technology Group   Quality Instrumentation        A  A    Warning  The main purpose provided by the series of the  instruments is for Puncture Testing     Warning  Unlike power supply which can be used on the  extreme low load range  when the series are used on the  AC voltage  the suggestive load should be more than  500VA at 50kQ and 100VA at 250k Q  while used on the  DC voltage  more than 500k Q load is suggested     Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER  USER MANUAL             5  ARC Detect       Detect Current    SOOVA   0 3 100mA       100VA   0 3 20mA       6  Interface       Terminal type    Terminal seat       7  General Specification       Power Source    AC115V  230V 10   50 60Hz       Operation  Environment    Indoor use  altitude up to 2000m   Relative Humidity 80   Maximum    Installation category II   Pollution Degree 2       Operation Temperature  amp   H
13. om 0    to 40  C  32   to 104  F   To operate the instrument over this specific  temperature range may cause damage to the circuits   Do not use the instrument in a place where strong magnetic or  electric field exists as it may disturb the measurement    WARNING  This is a Class A product  In a domestic     environment this product may cause radio    interference in which case the user may be required  to take adequate measures     Guaranteed    888  88 SOURCE   www artisantg com    USER MANUAL    ELECTRICAL SAFETY TESTER    USER MANUAL    ELECTRICAL SAFETY TESTER       4  PANEL INTRODUCTION    357d DU NOV    SOAR NANNU PU MOHI 7  NONDA S UIO BAL BAN    70 1 aL0W3a8                                  Front Panel    Rear Panel    10    Guaranteed    888  88 SOURCE   www artisantg com    Artisan Technology Group   Quality Instrumentation        ELECTRICAL SAFETY TESTER    USER MANUAL          4 1  Front Panel                                                 ELECTRICAL SAFETY TESTER    USER MANUAL            Rear Panel         Fuse Holder with    Voltage Selector    To change AC source voltage  pull the fuse holder and  replace with an adequate fuse to connect with AC  power cord           Ground Terminal    Connect Ground terminal to the earth ground           Remote I O Control    Remote I O output joint point    Test signal  Two test holes has to be shorted when  getting into test mode by pressing Start  key  and it won   t return to open status  until the test result of PASS o
14. ppear in this manual or on the product     WARNING  Warning statements identify condition or     practices that could result in injury or loss of life     or practices that could result in damage to this    CAUTION  Caution statements identify conditions  product or other property     The equipment shall not be used for measurements  within category II  III and IV     The following symbols may appear in this manual or on    AA     t ah    DANGER ATTENTION Protective Earth  ground  Frame or    High Voltage refer to Manual Conductor Terminal Chassis  Terminal Terminal  iv    Artisan Technology Group   Quality Instrumentation     Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER  USER MANUAL          FOR UNITED KINGDOM ONLY    NOTE  This lead appliance must only be wired by competent    persons    WARNING  THIS APPLIANCE MUST BE EARTHED    IMPORTANT  The wires in this lead are coloured in accordance  with the following code     Green  Yellow  Earth  Blue  Neutral  Brown  Live  Phase        As the colours of the wires in main leads may not correspond  with the colours marking identified in your plug appliance   proceed as follows     The wire which is coloured Green  amp  Yellow must be connected  to the Earth terminal marked with the letter E or by the earth    symbol    or coloured Green or Green  amp  Yellow     The wire which is coloured Blue must be connected to the  terminal which is marked with the letter N or coloured Blue or  Black     Artisan 
15. r FAIL  comes available  or the RESET key is  pressed    PASS signal  At the time when the test result comes  out to be PASS  two PASS holes has to  be shorted  and return to open until the  RESET key is pressed    FAIL signal  At the time when the test result comes  out to be FAIL  two FAIL holes has to  be shorted  and return to open until the  RESET key is pressed    RESET  The function is same as the RESET key on   the front panel    START  The function is same as the START key on   the front panel                             1  Model Number Model number and description  2   Display Indicate all messages about test procedure   3   CAUTION During test the red LED will flash to indicate  Indicator LED dangerous   4  ISTART Button Press the green button to start a test procedure   5   RESET Button Press the red button to reset stop a test procedure   6   ARC VR Control ARC range   7   Current VR Control the range of Cut off current   8   I Set ARC Set m   1 Set  W  ARC Set  9   ARC Con  Stop E  A signal for stop test  when the ARC is in use   the test will be stopped     A signal for continuous test   10   Range Switch over the range   11   Time VR Control test time   12  TIMER ON OFF EL  Power on and the test time can be adjusted   HL   Power off and the test time can not be adjusted   13   Test Voltage Test voltage adjustment   14   Setting  Current   W   Can adjust Current ARC VR to set current and  ARC   amp  Test  ARC   T  Stand by status for testing   15   AC DC or ACA
16. test   time  the instrument will stop testing when the ARC appears    1  Connect the Withstanding Insulation Tester to the ground    2  Adjust the    Voltage Knob    counter clockwise to the lowest  voltage position    3  Set power switch to ONE position    4  Set the key  15  to DCE position    5  Set CURRENT ARC of SETTING to IL position    6  Set Range to 10mA range    7  Set to I SetW position by pressing  8  key  then adjust the  Current VR until the value of 10 0 is displayed on the panel    8  Set to ARC Set IL by pressing  8  key  then adjust ARC VR  until the value of 10 0 is displayed on the panel    9  Set to Test BL by pressing  14  key    10  Set Timer to ON m position  adjust the knob until the timer  value is at 180s         Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER    USER MANUAL          11  Connect a test lead from the Return terminal  18  to the EUT   equipment under test     12  Connect a red high pot probe from    High voltage output  terminal  17     to the EUT  equipment under test     13  Press RESET key once to appear STBY  Stand by  on the  display  then adjust  13  the voltage knob until the value of 5 00  is display on the panel    14  Then  press START key to proceed testing  during testing  the  warning indicator  3  will be flashing continuously    15  If the value displayed on the panel drops because of the load  adding test  adjust    Voltage knob  13     to maintain the value at  5 00    16  If the test on t
17. to ARC Set EL by pressing  8  key  then adjust ARC VR  until the value of 50 0 is displayed on the panel    8  Set to Test W by pressing  14  key    9  Set Timer to ON   position  adjust the knob until the timer  value is at 60s    10  Connect a test lead from the Return terminal  18  to the EUT   equipment under test     11  Connect a red high pot probe from    High voltage output  terminal     17  to the EUT  equipment under test     12  Press RESET key once to appear STBY Stand by  on the  display  then adjust  13  the voltage knob until the value of 5 00  is display on the panel    13  Then  press START key to proceed testing  during testing  the  warning indicator 3  will be flashing continuously     13    Artisan Technology Group   Quality Instrumentation    ELECTRICAL SAFETY TESTER    USER MANUAL          14  If the value displayed on the panel drops because of the load  adding test  adjust    Voltage knob    to maintain the value at 5 00    15  If the test on the EUT is approval  the PASS indicator on the  panel will be lighted up    16  If the test on the EUT is disapproval  the FAIL indicator on the  panel will be lighted up and the buzzer will blare out a warning   Now remove the high pot probe from the EUT  and press  RESET to turn off the buzzer and FAIL light  then the  instrument will be back to the initial status       DC Withstanding Test    Take GPT 815 as example  if want to output SOOOVDC  set the   cut off current to 10mA and ARC to 10mA  During the 180s 
18. umidity    OC   40  C   lt 70        Storage temperature  amp   Humidity     10  C   70  C   lt 70                 Power cord xl   Accessories Test lead  GHT 105A  x 1    Instruction manual xl  Dimension 446 L  x 330 W  x 149 H   m m     GPT GPI 8X5  20kgs  GPI 826  13kgs approx           ELECTRICAL SAFETY TESTER  USER MANUAL          3  PRECAUTIONS BEFORE OPERATION    3 1  Unpacking the instrument  The product has been fully inspected and tested before shipping  from the factory  Upon receiving the instrument  please unpack and  inspect it to check if there is any damage caused during  transportation  If any sign of damage is found  notify the bearer  and or the dealer immediately     3 2  Checking the Line Voltage  The instruments can be applied with any kind of line voltage shown  in the table below  Before connecting the power plug to an AC line  outlet  make sure the voltage selector of the rear panel is set to the  correct position corresponding to the line voltage  It might be  damaged the instrument if connected to the wrong AC line voltage     WARNING  To avoid electrical shock the power  cord protective grounding conductor must be  connected to ground      N The equipment shall not be used for    measurements within category II  III and IV     Artisan Technology Group   Quality Instrumentation     Guaranteed    888  88 SOURCE   www artisantg com    ELECTRICAL SAFETY TESTER    USER MANUAL          When line voltages are changed  replace the required fuses shown as  b
    