Home

User manual MiniLift160 EN INT

image

Contents

1. The lift must undergo thorough inspection at least once per year Inspection must be performed by authorized personnel and in accordance with Handicare s service manual Repairs and maintenance may only be done by authorized personnel using original spare parts Y Spent batteries are to be left at the nearest recycling station Spent batteries can also be returned to Handicare or a Handicare dealer for recycling Cleaning disinfection If necessary clean the lift with warm water or a soap solution and check that the castors are free from dirt and hair Do not use cleaning agents containing phenol or chlorine as this could damage the aluminium and the plastic materials If disinfection is needed 70 ethanol 45 isopropanol or similar should be used Storage and transportation If the lift is not to be used for some time or e g during transport we recommend that the emergency stop button be pressed in The lift should be transported and stored in 10 C to 50 C and in normal humidity 30 75 The air pressure should be between 700 and 1060 hPa Let the lift reach room temperature before the batteries are charged or the lift is used 1060 hPa 75 OA 50 C 10 C 700 hPa 30 Service agreements Handicare offers the possibility of service agreements for maintenance and regular testing of your mobile lift Contact your local Handicare representative SystemRoMedic MANUAL 13 Symbols Read user manual May
2. cal emergency use the down button on the control box Mechanical emergency lowering For manual emergency lowering turn the round plastic knob on the actuator clockwise 8 MANUAL SystemRoMedic Using MiniLift160 Place an appropriately sized ThoraxSling behind and below the user s hips see user manual for ThoraxSling When the user has placed his her feet on the footplate move the lift forward towards the user Place the lift as close as possible to the user Take care to ensure that the user s shins are resting securely against the leg support and then adjust the leg support so that it does not press against the user s kneecaps about 2 cm lower Raise the locking handle on the lift arm and pull the lift arm out as far as possible towards the user Select the sling loop that reaches the hook on the boom shortest loop alternative Hook the lift loops to the underside of the sling bar SystemRoMedic MANUAL Using MiniLift160 Raise the locking handle and draw the boom back so that the lift straps on the sling are stretched and then lower the locking handle back into position Instruct the user to lean back and allow himself herself to be supported behind the back during raising Raise the lift to the desired height and en sure that the user is standing with both feet on the footplate and is grasping the sling bar handles with both hands to avoid being pinched between the slingbar and li
3. ce MiniLift160EE MiniLift160EM SystemRoMedic 60300010 60300011 60300012 60300013 IO Bd f SWL 160 kg 350 Ibs Manual nr 450 EN ver 3 150622 NS ha ndicare Table of contents Description Functional inspection Always read the manual Assembly of MiniLift160 Final inspection Using the product Important information Safe working load Charging batteries Hand controll Emergency stop Emergency lowering Instructions for use Using MiniLift160 Trouble shooting Accessories Maintenance Symbols Technical information Dimensions 6 11 9 10 11 12 13 14 15 16 MANUAL SystemRoMedic MiniLift160EE MiniLift160EM SystemRoMedic Lift arm Hooks for lift vest sling Hand control Leg support Footplate Rear castos with brakes Battery Emergency stop Adjustment of leg support 10 Electrical emergency lowering 11 Charging lamp 12 Front castors 13 Locking mechanism for adjustment of lift arm Sern Poff SY m 14 Base width adjustment MiniLift160 is a mobile sit to stand lift which has been developed to as gently as possible assist the user when rising from a sitting to a standing position When MiniLift160 is combined with the appropriate lifting accessories the user gets support under the feet for the front of the lower legs and behind the back which provides for a safe and secure sit to stand procedure The construction safely m
4. fers solutions and support to increase the independence of disabled or elderly people as well as to improve the convenience of those who are caring for them The Handicare Group is one of the leading healthcare companies in Europe with its own manufacturing organizations and sales companies in Norway Sweden Denmark Germany the Netherlands Great Britain France China Canada and the USA Handicare s products are also distributed by partners in more than 40 countries worldwide Our wide range of high quality products includes various manual and power wheelchairs seating systems scooters a complete easy transfer system and other patient handling aids stairlifts car adaptations rise and recline chairs and bathing and toileting products handicare Handicare AB Maskinvagen 17 SE 972 54 Lulea SWEDEN Tel 46 0 8 557 62 200 Fax 46 0 8 557 62 299 E mail info handicare se www handicare com Handicare AB is quality and environmentcertified in accordance with ISO 9001 IS014001 and ISO 3485
5. ft loops Helpful hint Use a longer loop alternative for short users and for users that cannot be fully raised Use a longer loop alternative for taller users and pull the boom out to allow the user to rise to a standing position 10 MANUAL SystemRoMedic Trouble shooting If the lift adjustment cannot be activated check the following That the emergency stop button is not activated That all cables are properly and securely connected Pull out the contact and plug it in again firmly That battery charging is not in progress That the battery is charged If the lift is not working properly contact your dealer If the lift makes unusual noises Try to determine the source of the sound Take the lift out of operation and contact your dealer SystemRoMedic MANUAL 11 Accessories ThoraxSling polyester XS XXL article no 45500003 009 ThoraxSling wipeable S XXL article no 45500004 009 ThoraxSling with seat support polyester S XL article no 45600004 008 ThoraxSling with seat support disposable non woven S XL article no 45690004 008 CalfStrap article no 70200033 SlingBarWrap MiniLift wipeable article no 70200012 For MiniLift160EE 60300012 600300013 Hand control HB33 6 6 buttons with service battery status and overload indicators article no 7O200089 Maas SS MAN oe zu E didddidh MANUAL SystemRoMedic Maintenance
6. ght between the footplate and the floor Ensure that the user s feet do not get caught between the footplate and the base when adjusting the base width e MiniLift1 60 is not intended for transporting users It may only be used for short transfers e Never move the lift by pulling on the actuator O Do not push e The lift must not be used or immersed in or under water including showering e The lift should not be left stored in a high humidity environment e The lift must not be charged in a wet room e To ensure optimal function the lift must be inspected regularly See section on Maintenance e Warranty applies only if repairs or alterations are done by an authorized technician e Using lifting accessories other than those approved by Handicare can entail a risk Safe working load Different products on the same lift system lift unit sling bar sling scales and other lifting accessories may have different allowable safe working loads The lowest allowable safe working load always determines the safe working load of the assembled system Always check the safe working loads for the lift and accessories before use Contact your dealer if you have any questions 6 MANUAL SystemRoMedic A tone sounding when using the lift indicates that the battery need recharging Charge the lift after use to ensure that the battery is always fully charged Lock the castor wheels when charging the battery 1 Connect the charging cable t
7. learance at base H Lifting height min Lifting height max J Hoisting range Ki Min distance from wall to CSP at max Height legs spread Minimum distance from wall to CSP at max reach legs spread K2 Min distance from wall to CSP at min height legs spread L Footplate clearance M Footplate height N Inside space between bars O Overall length P Reach from base with legs spread to 700 mm R Max reach from base Weight loose parts footplate Number of lifts per charge Footplate dimensions W x D h ll m Zz ie Tolerance 0 2 5 mm EE EM 36 2 919 mm 20 9 531 mm 30 4 773 mm 26 3 669 mm 37 8 959 mm 38 6 980 mm 7 4 190 mm 5 4 138 mm 4 8 123mm 2 8 fromm 37 940mm 55 3 1405 mm 18 3 466 mm 37 3 949 mm 8 1 208 mm 8 1 208 mm 19 Samm 2 3 59mm 25 5 650 mm 43 2 1098 mm 15 6 397 mm 29 1 741 mm 8 3 lbs 3 8 kg 20 25 A SAA EEL EML 35 5 902 mm 2 5 65 mm 0 7 20 mm 42 5 1080mm 15 9 x 13 8 406 x 392 mm SystemRoMedic Simple solutions for great results SystemRoMedic is the name of Handicare s unique easy transfer concept the market s widest and most comprehensive range of clever easy to use and safe transfer and lifting aids designed to make life easier both for the user and for the caregiver SystemRoMedic is a complete solution that provides for the majori
8. nghandles at the base of the mast 7 rm 4 Tighten locking handle 1 5 Check all mounting fixtures and locking handles 6 Deactivate the emergency stop button and perform a final inspection See Final N inspection Connect cables Hand control cord in HS cable to lift unit on mast in outlet M1 To collapse the lift follow the instructions below Lock castor brakes Change the base width pedal to horizontal Remove legsupport and footplate Run the lift arm to the lowest position and then activate in the emergency stop button Push the lift arm in as far as possible Loosen the locking handle at the base of the mast 1 You need not remove the locking handle Lift up on the mast and lower it towards the undercarriage NO I FR WB DP IN Warning risk of pinching between base and mast 8 Pull the mast back and tighten the locking handle 1 4 MANUAL SystemRoMedic Final inspection Check to ensure that no parts have been left in the packaging Inspect the lift for signs of wear and damage Check all four castor wheels and castor wheel locks Check all connections and fixtures including screws and bolts Check the emergency stop function by activating the emergency stop button and then pressing either the up or down button If nothing happens when the up or down buttons are pressed the emergency stop is functioning properly Grasp the hand control press the up button and run the lift a
9. not be discarded in domestic waste The product fulfills the demands stated in the MDD 93 42 EEC Type B according to the degree of protection against electric shock The device is intended for indoor use Class ll equipment A Always Read The User Manual EMC This equipment system is intended for use by healthcare professionals only This equipment system may cause radio interference or may disrupt the operation of nearby equipment It may be necessary to take mitigation measures such as reorienting or relocating the equipment system or shielding the location 14 MANUAL SystemRoMedic Technical Information Lifting speed 1 4 s 35 mm s without load Batteries Two 12V 2 9 Ah valve regulated sealed lead accumulator gel type batteries 400 mA DC 24V 5 5 Amp IP X4 Operationtime 10 at maximum continous operation of 2 minutes maximum 5 cycles per minute Push 4 OOON 89 3 lbs 40 5 kg Turning diameter 43 6 1109 mm IP X4 Operating forces buttons on 4N hand controll Max load 160kg 350 lbs Medical Device Class The product complies with the requirements of the Medical Devices Directive 93 42 EEC SystemRoMedic MANUAL 15 Dimensions Measurement A Length B Width inside min legspread Width indside max legspread C Width outside min legspread Width outside max legspread D Height E Maximum reach at 600 mm reference height F Height at base rear Height at base front G C
10. o a power outlet 2 Check to ensure that the lamps on the control box light up The green LED lamp indicates that the charger is receiving power and the yellow LED lamp indicates that the battery is charging 3 Charging stops automatically when the battery is fully charged Wall mounted charger 1 Remove the battery pack from the lift and place it in the wall mounted charger 2 Check to ensure that the LED lamp on the front of the charger lights up NOTE Before the lift is used for the first time it must be charged for at least 4 hours For maximum battery life charge batteries regularly We recommend daily charging when the lift is used daily The emergency stop must be deactivated during charging Battery information on the display The battery discharging will be shown in four stages Battery state 1 The battery is ok no need for charging 100 50 Battery state 2 Battery needs charging 50 25 Battery state 3 Battery needs charging Less than 25 A tone sounds when a button is pressed in this battery state Battery state 4 The battery needs charging 17V or lower At this stage some of the functionality of the lift is lost At this battery stage it is only possible to drive the lifting arm down Furthermore an audio signal will sound when a control button is activated The symbol will switch between the two pictures for 10 seconds The battery symbol is shown when the box is active until power d
11. oves the user forwards and upwards in a natural movement pattern and at the same time the leg muscles and balance is exercised MiniLift160 is available in four variants with electrical EE or manual EM base widening and with standard height or low legs for use with lower beds All variants are manufactured in steel for stability and strength and are suitable for users weighing up to 160 kg 350 lbs E Functional inspection Visual inspection Inspect lift functions regularly Check to ensure that material is free from damage Before use Ensure that the product is correctly assembled Check sling bar connection and safety latch function Check lift and base width movement Check to ensure that the actuator is correctly installed li Always read the user manual Always read the user manuals for all assistive devices used during a transfer Keep the user manual where it is accessible to users of the product The lift may only be used by persons who have received instruction in the operation of the lift SystemRoMedic MANUAL 3 Assembly of MiniLift160 Check to ensure that all components are included Lift unit control box and battery pack undercarriage footplate leg support hand control and cord user manual and charger The lift is collapsed when delivered Set up the lift according to the following instructions no tools required 1 Lock castor brakes 1 r 2 Place the mast in the base 3 Mount the lockni
12. own 2 minutes after use It is not possible to use other batttery types than BAJ1 BAJ2 The battery level is measured via voltage level This means that it is possible to experience e g that the battery switches from state 1 to state 2 and back to state 1 SystemRoMedic MANUAL iS Service information read out 12034 1257000 Basic service information can be read out on the display To get the service information on the 7 display please press the lifting arm up button for half a second The information will be shown for 90 360 minute or until other buttons are activated Total cycles done e Total work done e Total number of overloads e Days since last service Days between services Hand control Raising lowering the lift arm A Symbol indicate direction of travel Motion stops as soon as the button is released Electrical base width adjustment EE Markings on the buttons of the hand control indicate function Motion stops as soon as the buttons are released Manual base width adjustment EM Push down on the respective pedals at the back of the lift to widen and narrow the base Emergency stop To activate emergency stop Activate the red emergency stop button on the control box Resetting Turn the button in the direction of the arrows until the button pops out Electrical emergency Emergency lowering lowering Electrical emergency lowering For electri
13. rm all the way up Then press the down button and run the lift arm all the way down Test base width adjustment function MiniLift1 GOEE Press the button for base width adjustment to widen the base fully and then press the lt other button to narrow the base again Test lift function by lifting a person not a user using an approved sling At the same EM time check the emergency lowering function with someone on the lift See section on Emergency lowering ER YX If the lift is functioning correctly connect the charger and check to ensure that the charging lamp on the control box lights up NOTE Before the lift is used for the first time it must be charged for at least 4 hours See section on charging batteries Keep the user manual where it is accessible to users of the product SystemRoMedic MANUAL 5 Using the product IN Important Information e The lift must be assembled according to the assembly instructions provided with the lift e The lift may only be used indoors and on a level floor e Lifting accessories must be properly trial fitted and tested in relation to the user s needs and functional ability e Do not leave the user unattended during a transfer situation e Under no circumstances may max load be exceeded See section on Safe working load e The lift arm must be loaded throughout the entire lowering procedure to ensure a soft and smooth experience e Ensure that the user s feet do not get cau
14. ty of patient transfer or manual handling requirements From the simplest to the most complex scenarios from the lightest to the heaviest The concept encompasses assistive devices for four different categories of transfers e Transfer assistive devices for manual transfers of users between two locations Positioning assistive devices for manual repositioning of users within the same location e Support assistive devices for mobility support e g during sit to stand or gait training Lifting assistive devices for manual and mechanical lifting of users Improved work environment improved quality of care and cost savings The philosophy behind SystemRoMedic is focused on the prevention and reduction of occupational injuries while allowing users to experience a greater sense of independence and dignity Through a unique combination of training and a complete range of efficient transfer aids SystemRoMedic offers improvement of both work environment and quality of care and at the same time achieves significant cost savings Always make sure that you have the correct version of the manual The most recent version of all manuals are available for downloading at from our website www handicare com For questions about the product and its use Please contact your local Handicare and SystemRoMedic representative A complete list of all our partners with their contact details can be found on our website www handicare com Handicare of