Home
        TMS 9000 Torque Measurement System
         Contents
1.     Tighten all bolts  in incremental steps  to the  bolt manufacturers rated torque specification   Use the respective sequence illustration shown    below depending on the number of bolts the  sensor requires  This bolting sequence applies  to both bolt circles of the torque sensor        Honeywell e Sensing and Control 3    TMS 9000 Torque Measurement System    November 2006  008 0688 00    Installation and Set up    Torque Sensor    The TMS 9000 series torque sensors may be  operated horizontally  vertically  or any angle in  between provided the load is applied through the  loading axis    All torque sensors in this series have bolt  patterns that mate directly to standard industrial  couplings  When mounted  one of the flanges  should be mated to a good quality double flex  coupling or a driveshaft arrangement that  incorporates universal joints at each end  This is  designed to compensate for angular and parallel  misalignment    Avoid applications that place extraneous loads  on the torque sensor     Caliper Coupling Module    The caliper coupling module must be firmly  mounted to a non rotating support structure  It  must be aligned with the epoxy glass annular  printed circuit board antenna so that the air gap  between the caliper and the antenna is  approximately equal on both sides  Care should  be taken to avoid any items touching one  another  and consideration should be given to  the effects of vibration as well as the free play in  any driveshaft sliding joint
2.   CalSteps 5    CalReset   Apply    100 Nm and set  CalValue1  100   Apply    50 Nm and set  CalValue2  50  Apply 0 Nm and set  CalValue3 0   Apply  50 Nm and set  CalValue4 50  Apply 100 Nm and set  CalValue5 100    e To obtain a frequency output of 5 KHz at 10  Nm and 15 kHz at 80 Nm then the  parameters would be  AnOutLow 10 and   AnOutHigh 80     The device will then be fully calibrated     Note that for best results and to conform to  accepted calibration practice  the unit under test  should be exercised three times at the full load  in the direction of loading prior to the setting of  calibration points  This is especially important  when calibrating in both the clockwise and the  counterclockwise directions  Please contact the  factory for a detailed description of calibration  practice and procedure     If an alternative analog output is selected at a  later date  or if different settings are chosen for  the  AnOutHigh Low parameters later  it is not  necessary to repeat the loading calibration  because all analog outputs are digitally driven     N ber 2006  TMS 9000 Torque Measurement System E    Troubleshooting          Problem Possible Solutions     Power On    light is not Check that 12VDC is being applied to the correct terminals  J1   showing and at the correct polarity     In certain cases  for example where the caliper coupling module  has been left in direct contact with a metal surface for some  time  the internal thermal protection circuit may have activat
3.  2 147 52  300 3600 406 73 300 2655 4 221 28  400 4800 542 30 400 3540 5 295 04  500 6000 677 88 500 4425 6 368 80  600 7200 813 45 600 5310 7 442 56  700 8400 949 03 700 6195 8 516 32  800 9600 1084 6 800 7081 0 590 08  900 10800 1220 2 900 7966 1 663 84  1000 12000   1355 8 1000 8851 2 737 60  2000 24000 2711 5 2000 17702 1475 2  3000 36000 4067 3 3000 26554 2212 8  4000 48000 5423 0 4000 35405 2950 4  5000 60000 6778 8 5000 44256 3688 0  6000 72000 8134 5 6000 53107 4425 6  7000 84000 9490 3 7000 61958 5163 2  8000 96000 10846 8000 70810 5900 8  9000 108000 12202 9000 79661 6638 4  10000 120000   13558 10000 88512 7376 0  20000 240000 27115 20000 177024 14752  30000 360000 40673 30000 265536 22128  40000 480000 54230 40000 354048 29504  50000 600000 67788 50000 442559 36880  60000 720000 81345 60000 531071 44256  70000 840000 94903 70000 619583 51632  80000 960000 108460 80000 708095 59008  90000 1080000 122018 90000 796607 66384  100000 1200000   135575 100000 885119 73760                   Honeywell e Sensing and Control 13    TMS 9000 Torque Measurement System    Specifications   Electronics    November 2006  008 0688 00       Power supply    Protection    Analog output signals     4 20 mA      10 VDC    10 kHz     5 kHz  60 kHz     20 kHz  zero torque   60 kHz     12 V DC     10   0 75A  9W maximum    Reverse polarity  connection or fault  condition will trip the  internal thermal fuse   self resetting      zero torque   12 mA    zero torque   0 V    zero torque   10 k
4.  default  select       JP2    Secondary RS232 RS485  default select       JP3          JP4             6 Honeywell e Sensing and Control    November 2006  008 0688 00    November 2006    TMS 9000 Torque Measurement System 008 0688 00       Command Set    Below is the list of parameters supported by the TMS 9000  corresponding to firmware version v1 38       Parameter  Name     AnOutHigh     AnOutLow    AuxOPType    BaudRate   CalCnts1    CalCnts9     CalPoints   CalReset     CalValue1  to   CalValue9     FastMode       Mode reporting access used by TMS Toolkit software to extract parameter information from the device  Read only    Sets or returns the value in engineering units applied to the input that will give 100   maximum positive full Read write  scale  output on the analog outputs  To invert the output polarity  enter the required negative full scale output  value     Sets or returns the value in engineering units applied to the input that will give 0   minimum negative full scale  Read write  output on the analog outputs  To invert the output polarity  enter the required positive full scale output value     Not yet supported   Not yet supported     Not yet supported  Read only    Internal calibration data  This is read only via TMS Toolkit and is viewable only in    CAL    level access  These Read only  parameters are exposed to enable the saving and loading of calibration data only     Sets the number of calibration points in use  Value must be between 2 and 9  Any ch
5.  events which are not of interest  A  safe value to use when setting up is 5000  which corresponds to 50  of FS  the reduce the threshold  later if required     Now consider an input change of 0 1 Nm  which is below the threshold and therefore subjected to filtering   The difference between the input and the output is 0 1 Nm so this is the change that will be used by the  filter comparator     If the FiltSteps parameter is set to 10  then the output of the TMS 9000 will be incremented towards the  input in 10 steps  as follows      1  2 0 1 Nm   1 3 0 1 Nm   1 4 0 1 Nm   1 5 0 1 Nm        1 9 0 1 Nm   1   10 0 1 Nm     The filter update rate being 1000 Hz gives an update period of 1 1000 0 001 seconds  therefore the filter  will settle to 63  of its final value in 0 01 seconds  being 0 001   FiltSteps      Using the formula described in the table above  it can be seen that the filter will settle to within 99 9  of  its final value in 0 07 seconds  being 0 001   FiltSteps   7     Honeywell e Sensing and Control C 3    N ber 2006  TMS 9000 Torque Measurement System E    Quick Look Up Table                                     to 39 9  of Final Value Equivalent to FiltSteps   seconds  Cut Off Frequency of  Hz  setting required  78 0 16 Hz 1000  3 58 0 32 Hz 500  1 45 0 79 Hz 200  0 7S 1 59 Hz 100  0 35 s 3 17 Hz 50  0 14s 7 94 Hz 20  0 07 s 15 87 Hz 10  0 035 s 31 75 Hz 5  0 014s 79 37 Hz 2  0 007 s 158 73 Hz 1                NOTE   The setting of FiltLevel is of great importance
6.  filter that behaves like an    RC    circuit     It has two user settings  the first being a    level     set by the parameter FiltLevel  and the second  being a filter    weight     set by the parameter  FiltSteps     The    level    works as a threshold  above which  the filter is reset to allow a fast response to a  event that exceeded the threshold  This is useful  in the case when well damped steady state data  is required  but when significant fast transients  and disturbances should not be filtered out     The    weight    of the filter is set by increasing the  number of filter steps  which in turn increases    the time constant of the RC filter  increasing the  damping effect     The settings of any of the TMS 9000 parameters  can be changed at any time via the RS232  communications link  Changing parameters  while the system is running will take effect  immediately  and in the case of filter setting  changes  will become effective as soon as the  filter flushes through     The TMS Toolkit software  supplied with the  TMS 9000 system  simplifies the task of  changing settings  although any character based  communications software could be used instead   e g  HyperTerminal         Filter Operation Detailed Description    Consider the input signal as being V  and the  output signal being V     In a steady state situation  V  will equal V     When V  changes  the extent of the change is  compared with the threshold  which is set as a  proportion of the full scale sensi
7.  of the input  and output  using floating point values for  convenience of the user  and the linearizing  routine can use up to 9 data points  user  selectable   so a significant amount of processor  power is consumed during these floating point  calculations     The next process is digital filtering  using a  parameter driven recursive algorithm that  performs output smoothing but also provides a  separate parameter that controls a filter bypass  in the event of a significant change in input  being required to be reflected through to the  output without delay     The filtered data is then converted to the  required analog output format or formats  the  TMS 9000 can drive the voltage or current loop  output at the same time as providing a  frequency output  using the output scaling  parameters that are independent from the input  calibration     The rate at which the microprocessor can  perform the separate linearizing and scaling  calculations is the limiting factor in determining  the available bandwidth of the TMS 9000     To provide a faster response for users that want  to analyze the dynamic data  a FASTMODE is  provided  and in this mode  the data is    piped     directly from the rotor to the analog voltage  output     The benefit of this mode is that the analog  voltage output is updated at the maximum data  rate  which is eight times faster than the normal  mode rate     The penalty of using fastmode is that the scaling  and linearizing stages are bypassed  so the  
8.  other warranties  expressed or implied  including  those of merchantability and fitness for a particular  purpose  In no event shall Honeywell be liable for  consequential  special  or indirect damages    While we provide application assistance personally   through our literature and the Honeywell web site  it is up  to the customer to determine the suitability of the product  in the application    Specifications may change without notice  The  information we supply is believed to be accurate and  reliable as of this printing  However  we assume no  responsibility for its use     Honeywell International Inc   Sensing  amp  Control   Sensotec     Lebow   2080 Arlingate Lane  Columbus  OH 43228  www honeywell com sensotec    Printed in USA  November 2006    SALES AND SERVICE   Honeywell serves its customers through a worldwide  network of sales offices  representatives and distributors   For application assistance  current specifications  pricing  or name of the nearest Authorized Distributor  contact  your local sales office or     E mail  sales  sensotec com  Internet  www honeywell com sensotec    Phone and Fax   Tel  614 850 5000  Fax  614 850 1111    Honeywell    Copyright    2006 Honeywell International Inc  All rights reserved        
9.  provides a quick reference to the data rates available in either of two available modes                       Mode Data rate  NORMAL When FILTSTEPS  gt 1  data rate is 1104  results per second  NORMAL When FILTSTEPS 1  data rate is 2207  results per second  FASTMODE 8828 results per second          Note that the analog voltage output channel should be only channel is use for any data rate above 1104  results per second  Therefore OPTYPE should be set to 1     Note that any traffic on the RS 232 port caused by TMS Toolkit or any other communications package will  disrupt the flow of data due to the interrupts that are generated by the external software     Honeywell e Sensing and Control D 4    N ber 2006  TMS 9000 Torque Measurement System E    The following examples assume that either the TMS Toolkit is available and running in CAL user mode  or  that a hard copy of the parameters list is available and is valid     Example 1   Theoretical determination of analog output value       The relationship between torque and digital counts can be determined by reference to the parameters  held in the TMS 9000  When in FASTMODE  the digital counts received from the rotor are simply piped  through to the analog voltage output channel  so the counts values can be used to determine the  expected analog output values  actual values may vary within the calibration accuracy of the analog  output channel  usually within 0 1 FS     Consider a sensor with a 2000Nm full scale torque measuring 
10.  set Read write  a threshold of 10  of the sensor rated capacity  then FiltLevel 1000  For a   step change in input  which is greater than  FiltLevel   10000   100    the new input value will be  passed immediately to the output  For a step change in input  which is below the threshold set by FiltLevel  the  output is filtered according to the setting of FiltSteps   NOTE   when FiltLevel is set to 0 or 1  digital filtering is disabled  Factory default value is 100  representing a  threshold of 1  of sensor rated capacity  refer to the factory calibration data sheet or the rating plate attached to  the sensor to confirm the rated capacity     Used to set the response time of the digital filter  Used in conjunction with FiltLevel to control the digital filtering Read write  behavior  Value range is 1 through 1000    Filtering takes the form of an RC equivalent where a change in input value  which is greater than the threshold   set by FiltLevel  causes the output value to be incremented in the number of steps set by FiltSteps  The filter   refresh rate is 1200 Hz    Factory default setting is 10  which in conjunction with the factory default setting of FiltLevel 100 provides for   an output increment in the form of  x 2  x 3  x 4  x 5  x 6      x 10  where x step change in input of more than    FiltLevel 10000  rated capacity   Given the filter update rate of 1200 Hz  the settling time to 63  final value   will be 8 3 ms  Update rate divided by FiltSteps  and to 1  final value 
11.  when adjusting the filter settings  Unexpected  torsional spikes and vibration noise can cause the frequent resetting of the filter  If in doubt  increase  FiltLevel  range 1 through 100000     When FiltLevel 1  the filter is bypassed and the torque value is delivered to the output processor at a  rate of 2207 Hz  thereby providing a 3dB cut off of 350 Hz    When FastMode 1  the scaling and linearizing algorithms are bypassed and the raw ADC count value is  delivered to the output processor at a rate of 8828 Hz  thereby providing a 3 dB cut off of 1400 Hz     Reference and Equations    To calculate the setting of FiltSteps required for a particular 3dB cut off frequency  use       FiltSteps   Update rate    Frequency   6 3     To calculate the 3dB cut off frequency for a particular setting of FiltSteps  use       Frequency     Update rate   FiltSteps     6 3    NOTE  the 3 dB point is also known as the half power point and occurs when the output voltage is equal  to 71  of the input or output power is 50  of the input power     OB    20 Log    Vou Vin   dB        10 Log    P P    oul in     Honeywell e Sensing and Control C 4    TMS 9000 Torque Measurement System    November 2006  008 0688 00    Sample Charts          Waiting for trigger q    1i CH2 2 01V   CH3E0 86V    De 1 1 DG 1 1    2004 06 21 12 45 01    2s div     NORM 500S s                               INPUT   Square Wave  0 1 Hz  OUPUT   Analog Voltage  FiltLevel 10000  FiltSteps 1000    Settling time to 99  is giv
12. Honeywell    Sensotec   Lebow       Operating Instructions for the Rev D  Nov  2006  TMS9000 Torque Measurement 008 0688 00    Sensing and Control    N ber 2006  TMS 9000 Torque Measurement System E    Intended Use    Rotating torque sensors are intended for use and a suitable load  They are also used to  between a power source and its load  They may measure the torque required to operate a given  be used to measure the power output of a drive load      such as an electric motor or gasoline engine     2 Honeywell e Sensing and Control    TMS 9000 Torque Measurement System    November 2006  008 0688 00    Operating Principles    Torque Sensor    Lebow Torque Sensors are designed  structures that perform in a predictable and  repeatable manner when a torque is applied   This torque is translated into a signal voltage by  the resistance change of strain gages  which are  attached to the torque sensor structure  The  change in resistance indicates the degree of  deformation  and in turn  the torque on the  structure    The strain gages are connected in a 4 arm  Wheatstone Bridge configuration which acts as  an adding and subtracting electrical network and  allows compensation for temperature effects as  well as cancellation of signals caused by  extraneous loading    When the torque sensor is rotating  a means  must be provided to transfer an excitation  voltage to the rotational element from a  stationary surface  and also to transfer the  torque signal from the rotational eleme
13. Hz     Output drive capability     e 4 20 mA  e    10 VDC  e Frequency    Digital resolution     e Normal mode  e Hi res mode    RF carrier frequency    Accuracy   e   Electronics  e Sensor  e System    Temperature Range     e Operating    e Compensated    e Zero temp stability  e Gain temp stability    500 Ohms max   2 k Ohms min   4 0 V p p   into 100 k Ohms  min load 1k Ohms    16 bit   0 01  FS   19 bit  0 001  FS     6 78 MHz    0 002  FS typical  0 050  FS typical  refer to system  calibration data sheet     40 to  85 C    40 to 185 F    10 to  50 C   14 to 122 F    lt 0 0025  FS C   lt 0 001  FS C    14 Honeywell e Sensing and Control    Frequency response    e Input sampling rate 17 656  samples sec  e Anti aliasing   filter  fixed   e Telemetry update  rate 8 828 kHz  e Fast mode data    4 1 kHz    throughput rate 8 828 kHz  e Normal mode data  throughput rate 1 104 kHz  e Analog output  bandwidth  max  3 kHz    3 db  e Group delay  typical   normal mode  2 5 ms  e Group delay  typical   fast mode  1 2 ms    Digital filtering     FIR mode 0 1 through 1000 Hz  IIR mode recursive algorithm  Cable length     Using RG58 cable 13 9 meters  45   6      or multiples thereof     max  10x standard     Using Belden 89907 17 6 meters  57 1        cable or multiples thereof  Enclosure rating  NEMA 4  IP65  EMC immunity  10V m     S30MHz 1 GHz     N 2  TMS 9000 Torque Measurement System UEA    Shunt Calibration    An electrical signal equivalent to that produced by a known load c
14. MA4 or IP65 environmental  part plug and socket connectors and the conditions  occasional water splash   The RF  connection details are shown in Figure 2  connection is made via the standard BNC   All cable connections should pass through the connector  and IP65 rated cable assemblies can  cable glands  which when properly assembled  be supplied upon request     provide adequate sealing to allow the module to    Figure 2                RS485 A 5o O e O    RS485 B                             Primary RS232 v    FS  port RS232 RXD    o  RS232 TXD    o   Y7  RS2320V    o  RS485 A  Secondary RS485 B   expansion  RS232 V       port RS232 RXD    2 Li mave      RS232 TXD    S J8  RS232 0V     o Expansion Connector J13          Power 0V    ne BNC connector to Calliper module  Via standard RG58 co axial cable   Length 13 9m  45 6 ft  or multiples thereof  Via standard Belden 89907 cable   Length 17 6m  57 ft  or multiples thereof     Honeywell e Sensing and Control 5    TMS 9000 Torque Measurement System    TABLE 1  Connector and Jumper Functions       Connector    Function       DC Power 12 V       Primary RS485 port       Secondary RS485 port       Current loop output       Voltage output       Frequency output       Primary RS232 port   NOTE   THIS IS THE  DEFAULT  COMMUNICATIONS PORT       J8    Secondary RS232 port       Factory use only       J10    Factory use only       J11       J12       J13    Expansion port       J14    Memory expansion port       JP1    Primary RS232 RS485
15. RMAL  BW   20MHz CH2   0 00V Zoom   10K Type   EDGE CH3    CH3   0 00V Delay   0 0ns  CH4   0 0   Hold Off   0 2us       Honeywell e Sensing and Control C 6    N ber 2006  TMS 9000 Torque Measurement System EU    APPENDIX D  Manual Supplement for TMS9000 SPM FASTMODE Operation and Settings    This supplement provides information related to v1 30 firmware and above     Intended Use    This supplement is intended for the purpose of Manual and the TMS Toolkit User Manual  both  describing the function and operation of the of which are supplied with a TMS 9000 Torque  FASTMODE feature that is included in the TMS Measuring System    9000 version 1 30 firmware and above  It should  be used in conjunction with the TMS 9000 User    Honeywell e Sensing and Control D 1    TMS 9000 Torque Measurement System    November 2006  008 0688 00    FASTMODE Operation General Description    The flow of data in the TMS 9000 is subjected to  various forms of processing as it passes from  input to output     This process is best described by use of a flow  chart as follows      Digitizing of  the strain gage  input signal    Modulation  and data  transmission    Demodulation  and data  recovery    Scaling and  linearizing    Filtering    Convert to  analog output       Normal mode    The data is transmitted from the rotor at the  maximum data rate but the rate has to be  slowed down for linearizing and scaling due to    the amount of processing required     the TMS  9000 features independent scaling
16. a new current Value of zero    Returns indication of 1 if the previous ZeroNow command was successful in setting the current Value to zero Read only  and returns 0 if the action was limited by  ZeroLimit    ZeroLimit The limit in engineering units at which the ZeroNow command will be allowed to operate  relative to the Read Write  computed Value at zero load that was stored during the calibration process  Therefore  ZeroLimit represents  the maximum allowable difference between the    calibration    zero and the    current    zero  If the ZeroNow  command is issued when the current Value is greater than  ZeroLimit  then the current Value will be moved to  the extent allowed by  ZeroLimit and the flag ZeroOK will be set to 0  Factory setting is 50  of the calibrated  range  Note that  ZeroLimit is a bipolar setting  so it will be applied to both directions    and    around the  calibration zero value      ZeroPVal Internal system zero data  This is read only via TMS Toolkit and is viewable only in    CAL    level access  This Read only  parameter is exposed to enable the saving and loading of calibration and zero data        Honeywell e Sensing and Control 9    TMS 9000 Torque Measurement System    November 2006  008 0688 00    System Calibration    The TMS 9000 features nine point  linearization and all calibration is achieved using  the following parameters      CalSteps    CalReset    CalValuet   CalValue2   CalValue3   CalValue4   CalValue5   CalValue6   CalValue7   CalVal
17. al feature operates by switching in a high precision shunt resistor across one of the arms of  the strain gage bride on the rotor  The change in output that occurs due to this shunting is repeatable and  is often used as a means of calibration  During factory calibration  the apparent change in torque output  due to shunt cal will have been recorded  and this value can be used to re calibrate the analog voltage  output when in FASTMODE     Consider a sensor with a 2250 Lbf in measuring range  The factory calibration certificate will include the  changes due to shunt cal as a list of effects such as follows      Shunt Calibration at Analog Output Settings      Lbfin   V   mA   Hz 10    Hz 60    High   Low   Units      Note   Hz 10  denotes 10 kHz range  10 kHz     5 kHz   Hz 60  denotes 60 kHz range  60 kHz     20 kHz        When in FASTMODE  the shunt cal values for voltage  current and frequency are invalid because the  scaling module is bypassed  so the only piece of information that remains valid and that we need to use  from this data table is the apparent change in TORQUE due to shunt cal  and in the case of this example   itis 1692 2 Lbf in   It follows that the change in analog voltage output when in FASTMODE will represent 1692 2 Lbf in   The exception to this case will be when a value has been set for  SCSCALE  This parameter allows the  effect of shunt cal to be varied  according to the value set  The default is 1 and any other value acts as a  multiplier     but on
18. an be obtained by activating the shunt  calibration function  The shunt calibration function is built in to the sensor itself  and it is therefore  necessary for the Rotor Active light to be showing before the function can be operated  By design  the  caliper coupling module is more sensitive to receiving data than it is to transmitting data  therefore it may  be necessary to adjust the caliper coupling module position to ensure good two way communications   prior to using the shunt cal function     The shunt calibration function is achieved by connecting a high precision resistor of know value  in  parallel  shunt  with one arm of the strain gage Wheatstone bridge  The connection is made by a solid  state switch  under the control of the microprocessor on the rotating sensor  when commanded by the  remote Signal Processing Module  This switch can be activated via the pushbutton on the face of the  signal processing module  The shunt calibration value is determined during factory calibration of the  torque sensor     The shunt calibration function is a very useful aid when setting up the system or when fault finding  In  applications where it is not possible nor practicable to perform dead weight system calibration  the shunt  calibration function can be used as an alternative  at the cost of some loss of calibration accuracy  To  provide for this eventuality  the shunt calibration value is factory set to represent between 50  and 95   of full scale  and is achieved by using 
19. ange in  CalPoints should Read write  be followed by a  CalReset command to clear the previous unwanted calibration data from the memories     NOTE   all calibration data and all analog output setting data will be cleared by  CalReset  It is recommended  to save the parameter list before invoking  CalReset     Resets all calibration information  When the reset command is issued  all calibration data and all settings of the Command  analog outputs   AnOutHigh and  AnOutLow  are cleared so no reliable output will be available until all of the  calibration points specified by the  CalPoints parameter and the required values of  AnOutHigh and   AnOutLow have been entered     NOTE   all calibration data and all analog output setting data will be cleared by  CalReset  It is recommended  to save the parameter list before invoking  CalReset     These values are written in engineering units when the appropriate load is applied  Each of the nine parameters Read write  can be written at any time  See Calibration section later in this document  NOTE     the values entered MUST be  in ascending order  starting with  CalValue1  negative values count as lower than positive values   The number  of calibration points entered must be equal to the number of calibration points activated by  CalPoints     NOTE   the existing calibration data is overwritten by any new input of  CalValue1 9  It is recommended to  save the parameter list before entering new values     Returns the raw A D counts valu
20. d Read write  value is the amount of zero offset being applied to the true Value  To zero the system  this parameter should be  set in engineering units to the value read when the system is supposed to be displaying zero  The action of  SysZero may be limited by  ZeroLimit as described above  in which case the flag ZeroOK will be set to 0  Note  that when any calibration parameter   CalValue1 9   CalReset  is changed  the value of SysZero is set to 0  and any zero offset is cancelled  This function  when used with  ZeroLimit  allows the current Value to be offset    to any desired level  remember to consider the dynamic loading range of the sensor itself when applying large  offsets        8 Honeywell e Sensing and Control    N ber 2006  TMS 9000 Torque Measurement System T T    Text memo field in which the name of the engineering units used for calibration can be stored for recall later  Read Write  Note that when reading some characters via a 7 segment display  TMS Toolkit uses a virtual 7 segment  display   some characters such as M will not display correctly   Returns the value of the applied torque in calibrated engineering units  Read only  Returns the software version Read only    Sets the current Value to zero unless limited by  ZeroLimit as described above  The action performed by Command  ZeroNow is to clear any previous zero offset then compare the true Value to  ZeroLimit  then  to the extent  allowable by  ZeroLimit  write the true Value to SysZero  resulting in 
21. e derived from the ADC on the rotating sensor  Read only    The ErrFlag parameter will indicate any errors that have occurred by returning a numeric value that is Read only  comprised of binary values representing the various error states  i e  the binary values for each error are added   together to produce the ErrFlag value    The error states are not retained between power cycles     Decimal Value Error Description  0 No error    1 Power cycled    2 Output clamped    4 Watchdog reset     Once the errors have been read they can be reset using the RstErrFlag command     Used to initiate raw throughput of data without scaling or filtering  Set this parameter to 1 to enable fast mode  Read write  This setting is volatile so the device will revert to normal mode after the next Reset or the next power up  Set to  zero to disable fast mode  When in fast mode  the internal raw A D counts results are fed directly to the analog    output  voltage or current  without any scaling or filtering  giving a data throughput rate of 8 8 kHz  when  FiltLevel 1     NOTE   the frequency output is not supported in fast mode  The fast mode can be scaled in the user   s data  acquisition system by using the shunt cal facility  and is intended to be used for dynamic measurements only     Honeywell e Sensing and Control 7    N ber 2006  TMS 9000 Torque Measurement System A ee    FiltLevel Used to set the threshold of operation of the digital filter  Values are set as parts per 10000  meaning that to
22. e the caliper coupling module to try and  achieve coupling and shunt cal functionality in an alternative  position    Check that there are no metal parts  flanges  covers  etc    within one and a half inches  40mm  of the caliper coupling  surfaces    Check that the power supply is actually 12 VDC when the  caliper coupling module is in the appropriate position  some  power supplies have built in protection circuits that cause a  reduction in supply voltage when current demands increase    Cannot communicate  Check all wiring    If using the RS232 port  check that the Rx pin of the host  computer is connected to the Tx pin of the TMS 9000 and vice  versa    Check that the communications cable being used is of high  quality or try a shorter length of cable  RS232 is sensitive to  cable length and grounding issues  especially when used with  laptop computers where grounding is uncertain                           Honeywell e Sensing and Control 11    N  TMS 9000 Torque Measurement System shined et       Cannot communicate  Check that the correct serial port is selected in the software or  TMS Toolkit  When using Windows  the serial port in use can  be found by using the CONTROL PANEL  SYSTEM   HARDWARE  DEVICE MANAGER  COM ports functions    On older desktop PC   s  the COM1 port is already in use for the  mouse  so a different COM port should be selected    If using a USB to Serial adapter  Windows assigns the COM  port designations dynamically so they may change whenever  the 
23. e the wires from the DAQ system through the conduit hole labeled    frequency    to the  desired output connector on the option board  Twisted pair 2 conductor shielded wire is  recommended for best performance  Strip off about 14    of insulation from the wire and tin  the ends with solder  Loosen the screws on the connector and slide the wires into the  connector next to the resistor leads  Remember  the ground  or common  wire must be  attached to the left side of the connector    7  Re attach the SPM cover and tighten the cover screws    8  Continue with system installation as described in the TMS9000 user manual     Honeywell e Sensing and Control A 1    N ber 2006  TMS 9000 Torque Measurement System Pe    APPENDIX B  Manual Supplement for TMS9000 SPM Remote Shunt Cal Option    This supplement provides information on the operation and specifications of the TMS9000 SPM with  the Remote Shunt Cal Option  P N 064 LW37039     Overview    The TMS9000 SPM with Remote Shunt Cal option allows the user to remotely activate and deactivate  the shunt cal mode via an external switch and cable     Setup    The Remote Shunt Cal option is installed and tested at the factory  A six pin circular connector is  mounted to the front panel of the SPM box as a connection point for the remote shunt cal switch  A  mating connector  023 LW181 034  is provided so the user can attach a cable between the SPM and the  customer supplied switch     1  Connect a two conductor cable between the remote s
24. ectronics module  The  data that is transmitted across the telemetry gap  consists of 8 828 results per second at a  resolution of 16 bit  and it is this data that is then  piped directly to the analog voltage output  whenever FASTMODE is turned on     The analog voltage output channel is a 16 bit  digital to analog converter with a bandwidth of  greater than 3 kHz  therefore the expected  analog output voltage for a full scale torque  measurement can be calculated by reference to  the calibration data tables held in the TMS 9000     Assuming that the factory calibrated  or user re   calibrated  data tables can be accessed using  the CAL user mode of the TMS Toolkit  the    output calibration can be determined using this  theoretical method  an example of which will be  given later     When the TMS Toolkit is not available  the user  will need to perform a physical system  calibration by placing a known torque on the  sensor and measuring the change in the analog  output voltage     In cases where the shunt calibration value is  known  the change in output due to shunt  calibration can be measured and the result  extrapolated to give a full scale equivalent  Note  that this result will need to be adjusted when the  shunt cal scaling feature has been used    SCSCALE is something other than 1   An  example of calculating the analog output voltage  by using shunt cal and the  SCSCALE  parameter is also given later     Normal mode and FASTMODE data update rates    The table below
25. ed   To reset this condition  remove power and wait ten minutes  before restoring power       Rotor Active    light is not Check that the RF cable is in good condition and is of the  showing correct length  look for damage to the outer sheath that may  indicate that the cable has been crushed at some time    Check that the caliper coupling module has been correctly  positioned in close proximity to the rotating antenna  Use the  positioning guide that was supplied with the system to confirm  the position  Move the caliper coupling module to try and  achieve coupling in an alternative position    Check that there are no metal parts  flanges  covers  etc    within one and a half inches  40mm  of the caliper coupling  surfaces    Check that the power supply is actually 12 VDC when the  caliper coupling module is in the appropriate position  some  power supplies have built in protection circuits that cause a  reduction in supply voltage when current demands increase    Shunt calibration does not Check that the    Rotor Active    light is showing  prior to using the  operate shunt calibration function    Check that the RF cable is in good condition and is of the  correct length  look for damage to the outer sheath that may  indicate that the cable has been crushed at some time    Check that the caliper coupling module has been correctly  positioned in close proximity to the rotating antenna  Use the  positioning guide that was supplied with the system to confirm  the position  Mov
26. en by  5   0 001   1000   5 seconds                3 ut     i   4   T1  Measure   Item Delay Time Range ALL Scan Cursor  OFF Setup   Setup T2 EXEC  2 8 kiv   Filter   Offset   Record Length   Trigger   Smoothing   OFF CH1  0 000V Main  10K Mode   NORMAL  BW   20MHz CH2   0 00V Zoom   10K Type   EDGE CH3    CH3   0 00   Delay   0 0ns  CH4   0 0V Hold Off   0 2us  Waiting for trigger q 2004 06 21 13 02 48    1i CH2 2 01      CH3 0 86V       1s  div  DC 1 1 DG 1 1          NORM 1kS s                                     2p  traced pai leisi  2 Wek ae  3    AUTO AT LYL PoR SRA SINGLE N SGL   Filter   Offset   Record Length   Trigger   Smoothing   OFF CH1   0 000V Main  10K Mode   NORMAL  BW   20MHz CH2   0 00V Zoom  10K Type   EDGE CH3    CH3   0 00   Delay   0 0ns  CH4   0 0   Hold Off   0 2us    Therefore overdamped response   only reaching 60  of full scale p p    INPUT   Square Wave  0 4 Hz  OUPUT   Analog Voltage  FiltLevel 10000  FiltSteps 1000    Settling time to 99  is given by  5   0 001   1000   5 seconds    value          Honeywell e Sensing and Control C 5    N ber 2006  TMS 9000 Torque Measurement System E    Sample Charts          q 2004 06 21 16 11 23  CH2 2 01V      gt  10ms div    INPUT   Square Wave  16 Hz  OUPUT   Analog Voltage  FiltLevel 10000    FiltSteps 38       Settling time to 63  is given by  0 001   38   0 038 seconds    Therefore overdamped response              Filter   Offset   Record Length   Trigger   Smoothing   OFF CH1  0 000V Main  10K Mode   NO
27. en using RG58 50 ohm cable  the  length must be maintained at 13 9 metres  45   6     feet  or a multiple thereof  When using Belden  89907 cable  length must be maintained at 17 6  metres  57 feet  or a multiple thereof  For cable  runs of less than 0 6 metre  2 feet   the cable  that is provided with the unit can be cut to the  required length  Otherwise  simply coil up any  excess length     Signal Processing Module    The receiver is mounted remotely with the  coaxial cable being the only connection between  it and the caliper coupling module  The receiver  has holes provided for permanent mounting   Request the appropriate certified drawing from  Lebow before making fixtures     When deciding where to locate the signal  processing module  consideration should be  given to the type of output that will be used  If  the analog voltage or current output is to be  used  then the signal processing module should  be mounted in an area of low electrical noise  and the connection between the module and the  data acquisition equipment should be as short  as possible and should be made up from double  screened twisted pair cable  If the frequency  output or the digital output is to be used  then  the signal processing module can be mounted in  the electrically noisy area provided that good  quality dual screened twisted pair cables are  used     N ber 2006  TMS 9000 Torque Measurement System E    Electrical Connections    The signal processing module features two  be operated in NE
28. equencies and with various filtering  characteristics     When using FASTMODE  the DAC is being updated at a rate of 8828 Hz therefore staircasing is  reduced as a result of the much faster update rate     Honeywell e Sensing and Control D 7       A WARNING    MISUSE OF DOCUMENTATION   e The information presented in this product sheet is  for reference only  Do not use this document as  product installation guide    e Complete installation  operation  and maintenance  information is provided in the instructions supplied  with each product    Failure to comply with these instructions could       A WARNING    PERSONAL INJURY   DO NOT USE these products as safety or emergency  stop devices or in any other application where failure  of the product could result in personal injury    Failure to comply with these instructions could       result in death or serious injury           result in death or serious injury        WARRANTY REMEDY   Honeywell warrants goods of its manufacture as being  free of defective materials and faulty workmanship   Honeywell s standard product warranty applies unless  agreed to otherwise by Honeywell in writing  please refer  to your order acknowledgement or consult your local  sales office for specific warranty details  If warranted  goods are returned to Honeywell during the period of  coverage  Honeywell will repair or replace  at its option   without charge those items it finds defective  The  foregoing is buyer   s sole remedy and is in lieu of all 
29. given by the following  table                      of Final Value Time to Settle  63   Filter Update rate   FiltSteps  99   Filter Update rate   FiltSteps   5  99 9   Filter Update rate   FiltSteps   7       Where the Update Rate is the period in seconds  1 f   fixed at 0 001s for firmware v1 30 and above  AND    Where the change in input signal magnitude is below the threshold set by FiltLevel     Honeywell e Sensing and Control C 2    N 2  TMS 9000 Torque Measurement System MUA    There follows an example of typical filter settings as applicable to a standard production TMS 9000 torque  measuring system     Consider a sensor with a 1000Nm full scale torque measuring range   Consider a test running at steady state torque of 550 Nm     Consider FiltLevel set at 1000  The unit of measure for FiltLevel approximates to 0 01 FS  where FS is  the full scale of the sensor   therefore a FiltLevel of 1000 sets a threshold of change of 10  of FS which  in this example is 100 Nm  above which the filter is reset and the output value becomes equal to the input  value again  Note that the threshold is with respect to the current value  and not with respect to zero  torque  Therefore  if the torque is fluctuating within a band of less than 100 Nm  the filter performance will  be determined by FiltSteps alone     In the case where torsional vibrations and spikes may be present in the input signal  then the threshold  set by FiltLevel should be raised so that the filter is not being reset by
30. high grade resistors that exhibit very low thermal sensitivity     Storage and Recalibration    This torque measurement system may be stored for an indefinite period in a dry place at room  temperature  Recalibration should follow your normal instrumentation certification schedule     Honeywell e Sensing and Control 15    November 2006  008 0688 00    APPENDIX    TMS 9000 Torque Measurement System    SUPPLEMENTARY INFORMATION  RELEVANT TO YOUR USER MANUAL  TMS 9000 TORQUE MEASUREMENT SYSTEM   Appendix A     Supplement for TMS 9000 SPM Square Wave Output Option  Appendix B     Supplement for TMS 9000 SPM Remote Shunt Cal Option    Appendix C     Supplement for TMS 9000 SPM Digital Filter Settings    Appendix D     Supplement for TMS 9000 FASTMODE Operation    16 Honeywell e Sensing and Control    N ber 2006  TMS 9000 Torque Measurement System oaao    APPENDIX A       Manual Supplement for TMS9000 SPM Square Wave Output Option    This supplement provides information on the operation and specifications of the TMS9000  SPM with the Square Wave Output Option  P N 064 LW37040     Overview    The Square Wave Output Option is a plug in module for the SPM that converts the sine wave  analog frequency output of the SPM to a square wave that is compatible with most RS 422 and  RS 485 type inputs to data acquisition systems and frequency pulse counters  The frequency of  the square wave is equal to the frequency of the standard sine wave output  The square wave  output is available in th
31. ly when the scaling module is operating     For FASTMODE operation  the scaling module is bypassed so the effect of shunt cal will be the original  effect  as manufactured  and may be significantly different from the calibration certificate value     To compensate for any value of  SCSCALE  calculate as follows    Certificated change in torque due to shunt cal is 1692 2 Lbf in  Value set for  SCSCALE is 1 5   Actual change due to shunt cal is 1692 2 1 5   1128 13 Lbf in    If  SCSCALE is 1  then no calculation is necessary     Honeywell e Sensing and Control D 6    N ber 2006  TMS 9000 Torque Measurement System E    Analog Output characteristics    The analog output channel is specified for a bandwidth of 3 kHz so there is no output filtering that  follows the digital to analog converter  DAC   This can lead to a    staircasing    effect when the  DAC is being updated at a relatively slow rate such as 1104Hz     For users that do not require wide bandwidth  this staircasing will not be a problem and can be  eliminated from the measurement by applying a suitable sampling rate at the data acquisition  end  Typically  a sampling rate of one quarter of the TMS 9000 DAC update rate  or less  would  be sufficient to solve this problem     For applications where the fidelity of the output waveform is of prime importance  the solution to  staircasing is to add a filter network across the analog voltage output terminals  Such networks  are available from Lebow in a range of cut off fr
32. nt back to  the stationary surface  This is accomplished  through the use of digital telemetry     Principle of Telemetry    The digital telemetry system consists of a  receiver transmitter module  a caliper style    coupling module  and a signal processing  module     The receiver transmitter module is an integral  part of the torque sensor and is connected to the  strain gauges and to the epoxy glass annular  printed circuit board that contains the rotating  antenna system  Within the receiver transmitter  module  the sensor signals are amplified   digitized  and are then used to modulate the  radio frequency carrier wave that is detected by  the antenna after being transmitted across the  air gap by the caliper coupling module  That  same carrier wave is rectified to provide power  to drive the strain gauges and the electronic  components in the module  which is managed  by a miniature microprocessor    The caliper coupling module connects to the  signal processing module via a simple co axial  cable  The detector circuitry in the signal  processing module recovers the digital  measurement data from the torque sensor and  passes it to the second microprocessor for  scaling and linearizing    The third microprocessor provides the drive to  the two analog outputs  as well as the standard  digital interfaces and the optional digital  interface modules  Extensive facilities are  provided in software for setup and configuration  of the complete system        Bolting Information
33. range    Calibration values most likely to have been used will be  approximately   2000  0 and  2000 Nm    The actual values used may have been adjusted to take account of local gravity and buoyancy and can  be seen from the parameters  CALVALUE1  2 and 3    Make a note of the actual values used and compare them to the values of  CALCNTS1  2 and 3    The  CALCNTSx values store the digital counts values that were output by the rotor for the load  conditions given by the relevant  CALVALUEx    The analog voltage output channel is 16 bit and it will generate an output of  10V when it is driven by a  counts value of 0  and will generate an output of  10V when driven by a COUNTS value of 65535   Therefore each count generates 0 0003052V starting from a base of  10V     The analog voltage output will be generated in direct relationship to the statement above  Therefore   using the following data  the analog voltage output will be      CALVALUE1  1998 699  CALCNTS1 21553   Therefore at a load of  1998 699 Nm  the analog Voltage will be  21553 0 0003052  10V    3 422V   CALVALUE2 0 000000  CALCNTS2 32700   Therefore at a load of 0 Nm  the analog Voltage will be  32700 0 0003052  10V    0 020V   CALVALUE3 1 998 500  CALCNTS3 43842    Therefore at a load of  1998 500 Nm  the analog Voltage will be  43842 0 0003052  10V    3 381V    Honeywell e Sensing and Control D 5    N ber 2006  TMS 9000 Torque Measurement System E    Example 2   Using the SHUNT CALIBRATION feature       The shunt c
34. ree formats  positive phase  TTL level   differential      5 VDC   and  negative phase  TTL level  180 degrees out of phase compared to positive phase      Setup    The Square Wave Output Option is installed and tested at the factory  A two pin green mating  connector and 120 ohm termination resistor is provided with the SPM to connect between the  output option board and the customer supplied data acquisition system  DAQ   counter or other  device  The following steps describe the process of connecting the SPM option board to the DAQ  or counter     1  Be sure AC DC power supply module is not connected to a power source    2  Place the SPM on a flat workbench or table  preferably with an ESD safe mat or cover to  dissipate electrostatic voltages  Wear an ESD safe grounded wrist strap while working  inside the SPM box    3  Carefully remove the four screws of the SPM cover and slowly lift the cover off the SPM   Take care not to damage the ribbon and ground wires connected to the inside of the  cover    4  The square wave option module is installed just to the right of J3 and J8 on the main  circuit board  A twisted pair jumper cable connects between J6 of the main board   frequency output  and the J1  frequency input  on the option board    5  Determine the desired output type  On the option board  the negative phase output is J2   differential output is J4  and positive phase is J5  The ground  common  point is on the  left side of the output connectors J2 and J5    6  Rout
35. relationship between input and output becomes  fixed  and the only way to calibrate the output  against the input is to calculate the expected  change in output value  by reference to the  calibration data stored in the TMS 9000   or  perform a physical system calibration or to use  the SHUNT CAL feature     Honeywell e Sensing and Control D 2    TMS 9000 Torque Measurement System    The following diagram shows the change in data  flow when using FASTMODE     Digitizing of  the strain gage  input signal    Modulation  and data  transmission    Demodulation  and data  recovery    Convert to  analog output       FAST mode    November 2006  008 0688 00    Because of the significant change in output  characteristics that takes place when  FASTMODE is selected  it is implemented as a  VOLATILE setting  therefore recycling the power  or performing a soft reset will return the TMS  9000 to NORMAL mode     As an indication to the user that FASTMODE is  in operation  the ROTOR ACTIVE light on the lid  of the TMS 9000 Signal processing Modules   SPM  is de activated     Honeywell e Sensing and Control D 3    TMS 9000 Torque Measurement System    November 2006  008 0688 00       FASTMODE Operation Detailed Description    The strain gage input value is digitized at a rate  of 17 656 samples per second with 24 bit  resolution  but this amount of data is in excess  of the capacity of the telemetry link  so it is  reduced by the simple averaging of every pair of  A D results at the rotor el
36. s     To assist in the process of aligning the caliper  and the antenna  a simple plastic alignment tool  is provided with each system  The tool is used to  hold the required clearance between the caliper  and the antenna while the caliper fixing bolts are  being tightened  and then is removed before the  sensor is rotated     The tolerances for end float  axial  are      4 5mm      3 16     and for run out  radial  are      1 0mm      1 16      For installations where run   out cannot be controlled within the specified  tolerance  the secondary coupling position can  be used  This is achieved by placing the edge of  the caliper in close proximity to the edge of the  antenna  In this position  the run out tolerance  can be at least doubled  at the expense of a  reduced signal to noise ratio caused by the  higher incidence of data drop outs  The axial    4 Honeywell e Sensing and Control    tolerance is limited by the distance between the  caliper sections     The caliper can also be mounted such that  only one side is in proximity to the antenna  if  the mounting arrangement does not allow for  placing of the antenna between the two sides of  the caliper     Successful positioning of the caliper can be  confirmed by the presence of the ROTOR  ACTIVE light on the signal processing module     The length of the RF cable connection  between the caliper coupling module and the  signal processing module is critical to system  performance  due to reflections and standing  waves   Wh
37. system is rebooted    The serial port settings are automatically modified by TMS  Toolkit so there is no need to change any of the settings in  Windows    The baud rate setting in TMS Toolkit should always be 38400  because that is the default baud rate of the TMS 9000    The    TMS ID    should be left blank because TMS Toolkit will  search the connected port for any TMS device and will  commence the communication automatically if present    Refer to the TMS Toolkit User Manual for more information                12 Honeywell e Sensing and Control    N ber 2006  TMS 9000 Torque Measurement System E    Conversion Table                                                                                                                                        Imperial Metric conversion Metric Imperial conversion  Foot pounds   inch pounds Nm Nm inch pounds foot pounds   1 12   1 35575 1 8 85119 0 73760   2 24 2 7115 2 17 702 1 4752   3 36 4 0673 3 26 554 2 2128   4 48 5 4230 4 35 405 2 9504   5 60 6 7788 5 44 256 3 6880   6 72 8 1345 6 53 107 4 4256   7 84 9 4903 7 61 958 5 1632   8 96 10 846 8 70 810 5 9008   9 108 12 202 9 79 661 6 6384  10 120 13 558 10 88 512 7 3760  20 240 27 115 20 177 02 14 752  30 360 40 673 30 265 54 22 128  40 480 54 230 40 354 05 29 504  50 600 67 788 50 442 56 36 880  60 720 81 345 60 531 07 44 256  70 840 94 903 70 619 58 51 632  80 960 108 46 80 708 10 59 008  90 1080 122 02 90 796 61 66 384  100 1200 135 58 100   885 12 73 760  200 2400 271 15 200 1770
38. tivity  by the  parameter FiltLevel     If the change exceeds the threshold  then the  new input value is passed immediately to the  output  thereby resetting the filter     If the change does not exceed the threshold   then the output value V  is updated by a  fractional amount of the new value V  until the  output value equals the input value again  The  number of steps set by FiltSteps determines the  number of fractional steps that are taken to  increment the output value  according to the  following series      1 2  1 3  1 4  1 5    ete  The output characteristic is therefore    exponential and behaves in a predictable  manner     Honeywell e Sensing and Control C 1    N 2  TMS 9000 Torque Measurement System UEA    To determine the settling time of the filter  the The table below provides a quick reference to  time taken to reach the V    V  condition   it is determine the filter characteristic   necessary to know both the filter update rate   and the number of fractional steps  The filter Note that this filter operates only when the  update rate is fixed at 1000 Hz in the firmware change in the input is below the threshold set by  v1 30 and above  although other filter update FiltLevel    rates can be made available upon request to the   factory     The cut off point  in Hz  is given by the  expression    Frequency   3dB       update rate number of steps  6 3    Filter Settling Time    The time required for the output to settle following a step change in input level is 
39. ue8   CalValue9    The minimum number of calibration points is  2  Calibration points can be created in any order  provided that the values they contain are in  ascending order starting with  CalValue1   Therefore  the lowest or the most negative   counter clockwise  calibration point should be  designated as  CalValue1     The number of calibration points that are in  use is set by the parameter     CalPoints     Any  change to the value of  CalPoints should be  followed by the issuance of a     CalReset     command  to clear the old calibration values  from the EEPROM memories     Calibration is achieved by applying known  loads at each of the calibration points that are  selected for use and then writing the engineering  units value to the appropriate  CalValuex  parameter     The analog outs are precalibrated in the  factory  so calibration of the input to the required  output range is automatic and is dependent on  the values entered for the parameters     AnOutHigh   AnOutLow     AnOutHigh and  AnOutLow are written to  using the engineering units value at which the  analog outputs are required to give the  maximum and minimum outputs     10 Honeywell e Sensing and Control    Available analog outputs are      Voltage range is    10 to  10 volts   Current output range is 4 to 20 mA   Frequency range is 5 kHz to 15 kHz or  alternatively  40 kHz to 80 kHz    Calibration Example     To calibrate from    100 to  100 Nm in five  steps of    100   50  0   50 and  100 Nm     Set
40. will be 41 7 ms  Update rate divided by   FiltSteps x 5   The settling time to 0 1  final value will be 58 3 ms  Update rate divided by FiltSteps x 7      Mode reporting access used by TMS Toolkit software to extract parameter information from the device  Read only  Returns the model name  TMS  Read only  string  OpType Sets or returns the currently selected analog output where O current  1 voltage  2 frequency 10 kHz  Read write  3 frequency 60 kHz  4 current and frequency 10 kHz  5 voltage and frequency 10 kHz  6 current and  frequency 60 kHz  7 voltage and frequency 60 kHz   Returns the number of parameters in the device   Set to the index number of the required parameter     ParaList Returns the information on the parameter indexed by Paraltem  Read  Format   index paraname type    a string  string  The value of type indicates the parameter   s properties by the addition of the following numerical values    Readable  Writeable  Command  String  Numeric  Boolean  Example     1 MODEL 33     In the above example where type   33 the parameter MODEL is a readable string     Returns the value of the applied torque in percentage terms  0 100  where this range is the selected range over Read only  which the analog outputs work and is set by  AnOutLow and  AnOutHigh     ete   Reset command to restart device and to implement parameter changes that require a reset     SysZero Allows manual setting of the current Value or querying of the current zero offset being applied  The returne
41. witch and the mating connector  Solder one  conductor to pin A of the mating connector and the other conductor to pin B  Attach the strain  relief to the connector    2  Attach the mating connector to the six pin connector on the SPM    3  After setting up the sensor and caliper module  power on the SPM and verify the Power LED and  the Rotor Active LED is lit on the top of the SPM  Turn on the remote shunt cal switch and verify  the Shunt Cal Mode LED is lit on the top of the SPM  Turn off the remote shunt cal switch and  verify the Shunt Cal Mode LED turns off     4  Setup of the Remote Shunt Cal Option is complete     Honeywell e Sensing and Control B 1    TMS 9000 Torque Measurement System    APPENDIX C    November 2006  008 0688 00       Manual Supplement for TMS9000 SPM Digital Filter Settings    This supplement provides information on the operation and specifications of the TMS9000 SPM with  the Digital Filter Settings  as they relate to v1 38 software     Intended Use    This supplement is intended for the purpose of  describing the function and operation of the  digital filtering algorithms that are included in the  TMS 9000 version 1 38 firmware  It should be  used in conjunction with the TMS 9000 User    Manual and the TMS Toolkit User Manual  both  of which are supplied with a TMS 9000 Torque  Measuring System     Filter Operation General Description    The digital filter algorithm in the v1 30 and later  firmware versions of the TMS 9000 is basically a  recursive
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Microsoft PowerPoint User Manual  Pyle PLW10RD  User Manual  Weed Eater 96132006300 Operator`s manual  Operating Instructions VEGACAP 27  Keys Fitness KF-45 User's Manual  ONE TOUCH T POP En  noncs D`INSTALLA non  ー HGD一006A HGD一。。7A 4/ 4  User`s Manual    Copyright © All rights reserved. 
   Failed to retrieve file