Home
INSTRUCTION MANUAL DIGIVAC Model 201 Digital Vacuum
Contents
1. Precise Practical Personal INSTRUCTION MANUAL DIGIVAC Model 201 Digital Vacuum Gauge and Vacuum Level Controller Revsion K11114 HW 11 11 YOU MUST READ THIS MANUAL BEFORE USE The DIGIVAC Company 105B Church Street Matawan NJ 07747 www DIGIVAC com 732 765 0900 732 765 1800 FAX CONTENTS 1 0 Description and Principle of Operation 2 0 Construction 3 0 Unpacking and Inspection 4 0 Installation 5 0 Operation 6 0 Servicing and Calibration 7 0 Notes on Calibration 8 0 Understanding Microns Torr and Absolute Pressure 9 0 Accessories and Modifications 10 0 Attachments and Illustrations 11 0 Terms amp Conditions DIGIVAC 201 series Instruction Manual Page 2 1 0 DESCRIPTION AND PRINCIPLE OF OPERATION The DIGIVAC Model 201 series is a highly configurable digital vacuum control instrument capable of driving and combining multiple sensors This device can save space money and complexity by combining many tasks into one device while enabling the user to have any level of accuracy desired for different ranges It is a highly configurable gauge that accepts different inputs based on driver board and sensor election and can have different outputs including Analog Recorder RS232 Ethernet SPDT relays and VLC outputs The 201 is configured with sensors based on the accuracy and range required to sense vacuum and display the pressure reading in user selectable units of Torr mBar or kilopascal The DIGIVAC Model 2
2. a screen that says tune then hit ENT e The first variable is S which is what you would use to adjust the set point for VLC control e Hit Sel to the desired variable either P I D or o e When the desired variable is displayed Press the A and vy to get to the desired value e Press ENT when completed Not pressing enter will result in no change of PID value PID overview PID control is largely used in industry and refers to the variables in the control equation proportional integral and derivative For a PID primer refer to http en wikipedia org wiki PID controller e P is implemented as a proportional gain not as a proportional band Larger values of P yield smaller error with less stability The range is 0 01 to 99 99 with units of e lis also again Larger values of will yield faster response with less stability The range is 0 00 to 99 99 with units resets minute e The D Range is 0 00 to 99 99 with units of minutes e QO is a feed forward term and is especially helpful with smaller vessels The range is from 0 99 where O 99 is traditional PID without the feed forward effect Here are the recommended PID tuning steps 1 To Start DIGIVAC 201 series Instruction Manual Page 8 a Start with P 0 14 0 08 d 0 S Set Point 1 Torr default O 0030 b Increase O in increments of 5 until the vacuum level is maintained at 1 2 Set P
3. be interactive with the readings of the instrument at higher pressures G The ATM adjustment is normally not necessary If necessary adjust the span with the ATM potentiometer DIGIVAC 201 series Instruction Manual Page 14 a For the 201H set the vacuum level to a steady 1980 millitorr and slowly turn the potentiometer on the right until the DIGIVAC gauge reads 1980 millitorr b For the 201V set the vacuum level to Atmosphere approximately 760 Torr and slowly turn the potentiometer on the right until the DIGIVAC gauge reads 760 Torr being careful not to over span c If you adjust the span recheck the zero then the span and the zero one last time 7 0 NOTES ON CALIBRATION The DIGIVAC is calibrated in nitrogen which has thermal properties virtually identical to air If you are using a thermocouple vacuum gauge other gasses will affect the readings by an amount proportional to the thermal conductivity of the gases In most cases the gases present in a vacuum system will be air nitrogen or oxygen and no appreciable errors will occur Certain other gases however have thermal conductivity significantly greater than air and will cause the instrument to read higher than the actual amount of pressure Examples of such gasses are water vapor fluorocarbon refrigerants and acetone Conversely other gasses have thermal conductivity significantly lower than air and will cause the instrument to read lower than actual pressure Examples
4. clause or sentence may be disregarded with the remainder of these terms and conditions valid and enforced In the event the unenforceable clause or sentence leaves a void in these terms and conditions a provision closely matching the intent of the unenforceable provision should be deemed inherent within these terms and conditions slightly modified to render such provision valid and enforceable GENERAL These terms and conditions are governed by the laws of the State of New Jersey USA You hereby consent to the exclusive jurisdiction and venue of the Courts of New Jersey in all disputes arising out of or relating to the use of this product Use of this product is unauthorized in any jurisdiction that does not give effect to all provisions of these terms and conditions MODIFICATION OF TERMS AND CONDITIONS DIGIVAC reserves the right to change the terms conditions and notices under which their products are offered DIGIVAC 201 series Instruction Manual Page 19
5. point value Normal run mode will resume 3 If you don t want the set points to actuate or the LEDs to illuminate at all set the set point for 000 One of the units LEDs to the right of the LCD will always be lit during normal operation to indicate which pressure range the display is indicating The Instrument has additional outputs which can be used e RS232 The instrument puts out a standard RS232 serial stream with settings 9600 8 N and 1 The unit transmits but does not receive and displays the current vacuum indication in the current units e Analog out This output reads from 0 to 5 Volts from a pressure of 1 micron all the way up to 5 Torr There is a graduation of 1 milliVolt per millitorr Therefore 10 milliVolts 10 millitorr and 4 Volts 4 Torr 201 VLC installation instructions DIGIVAC 201 series Instruction Manual Page 6 See above for proper installation of the vacuum gauge sensor Make sure the tube is installed on the actual vessel and is not in the pathway of the bleed air Connect the 2 position plug into the 201VLC 2 position vertical port Bleed outlet a Bleed from Vessel Bleed Valve gt nii lt Permantly 1m ke blocked orifice Tube that connects Connector that from the valve to the Goes into 2 pin vessel to be controlled vertical receptacle Make sure that the orifice closest to the wires of the valve remains blocked If this orifice becomes unblock
6. tubes AC and DC excitation are available Gauges have been provided for Hastings Varian Thermionics Veeco VRC and Fredericks gauge tubes DIGIVAC 201 series Instruction Manual Page 16 10 0 Attachments and Illustrations 201 types of expansion boards possible to populate a total of 4 slots Expansion Max Description CM 3 Expansion board for use with a capacitance manometer or any supported active gauge TC 3 Expansion board for use with a Varian 531 Thermocouple guage tube 2C 2 Expansion board with 2 SPDT relay outputs each output has Common NC and NO VLC 1 Expansion board for use to enable upstream vacuum level control Eth 1 Expansion board to enable Ethernet connectivity and web management Types of Sensors supported Expansion Sensor CM Total of 1 2 or 3 Setra or MKS capacitance manometers PCG 750 PVG500 or CDG 500 TC Total of 1 2 or 3 Varian 531 Thermocouple gauge tubes Sensor combinations Driver Max Description TC2 3 For use with a Varian 531 sensor If multiple are configured each slot can be displayed one at a time 1CP 3 For use with a CM sensor If multiple are configured each slot can be displayed one at a time 2CP 1 Combination where one value will me displayed based on the combination of a 0 1T and 10T CM or the combination of a 10T and 1000T CM 3CP 1 Combination where one value will me displayed based on the combination of a 0 1T and 10T and 1000T CM Slot and Driver combinations Slot name Driver Name Ga
7. vacuum system While threading the gauge tube in to the manifold the gauge tube cable should be disconnected to avoid damage In this way twisting of the cable and the octal socket on the tube is avoided Care should be exercised to install the tubes in a dry part of the system Since the instrument works on the principle of temperature rise the probes will not work if they become filled with a liquid such as vacuum or diffusion pump oil The gauge tube should be protected against oil and other contaminants by installing it in such a way to protect it A good practice is to mount the gauge tube in the most vertically distant place from oil and other contaminants as applicable The gauge tube should be mounted in the most stable pressure region of the vessel to be measured For example it would be better to install the gauge tube on a tank rather than on the pipe that is directly connected to a vacuum pump In the event of contamination see section 6 0 for gauge tube cleaning instructions If the gauge is used in a Neon sign processing facility the following is recommended to protect the gauge from damage from bombarding e The gauge tube should be isolated from the system with a stopcock The stopcock should be closed when bombarding e There should be at least 2 feet of tubing between the electrode and the DIGIVAC For best results the tubing should be metal e In extreme cases the gauge can be absolutely protected by installing a normally ope
8. 01 can either be panel mounted or sit on a bench top and can use thermocouple capacitance manometer PCG750 CDG500 or PVG500 sensors If in doubt about what gauge sensor you have consult the DIGIVAC packing list that came with your instrument for positive identification Consult the DIGIVAC website www DIGIVAC com for information about other DIGIVAC vacuum controllers and gauges The DIGIVAC Model configured with a Varian 531 Thermocouple gauge tube senses vacuum by measuring the temperature rise of an electrically heated thermocouple exposed to a vacuum As vacuum increases or more correctly as absolute pressure decreases fewer and fewer molecules of gas are available to cool the thermocouple With fewer molecules the air temperature rises and the thermocouple gauge thus senses the vacuum A precision reference inside the DIGIVAC in conjunction with an integrated circuit amplifier controls the electrical excitation of the sensor filament The DIGIVAC Model 201 configured with a capacitance manometer or other active gauge displays the vacuum level that corresponds to the sensor voltage output The DigiVac Model 201 has a total of 4 slots that can be populated by expansion boards Types of expansion boards Expansion Max Number Description CM 3 Expansion board for use with a capacitance manometer or any supported active gauge TC 3 Expansion board for use with a Varian 531 Thermocouple guage tube 2C 2 Expansion board with 2 SPDT relay outpu
9. 1D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 K11D21 K11D21 K11D21 K11D21 K11D21 K11D21 K11D21 K11D21 K11D21 K11D21 K11D21 Torr Sensor Config Vessel Size Ballast Vol Hose I D Pump CFM LPM S P l D O 531 TC2 C TC 0 245 ml 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 4 0 02 0 02 0 2 531 TC2 C TC 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 5 0 06 0 02 0 40 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 5 7 0 02 0 55 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 544 0 06 0 02 0 40 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 0 06 0 02 0 40 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 5 0 06 0 02 0 40 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 3 0 06 0 02 0 40 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 5 0 07 0 02 O 67 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 8 0 07 0 02 O 67 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 0 07 0 02 0 67 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 2 0 07 0 02 0 67 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 5 0 06 0 02 0 20 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 2 1 0 2 0 2 0 99 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 3 0 2 0 2 0 99 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 5 0 2 0 2 0 9
10. 9 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 10 0 2 0 2 0 99 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 1 0 05 0 05 O 99 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 07 0 05 0 05 O 99 CM 10T CM 0 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 05 0 05 0 05 O 99 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 5 0 08 0 02 O 75 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 08 0 04 0 02 0 75 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 0 05 0 05 0 99 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 0 06 0 02 0 45 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 2 0 06 0 02 0 45 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 1 5 0 06 0 02 O 45 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 1 0 05 0 05 O 99 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 2 0 05 0 05 O 99 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 07 0 05 0 05 O 99 531 TC2 C TC 1 5 Gallon 1 5 Gallon 0 5 inch Leybold Trivac 1 59 45 0 06 0 05 0 05 O 99 531 TC2 C TC 7 Gallon 0 0 5 inch Welch 1405 3 2 60 1 0 06 0 02 0 99 531 TC2 C TC 7 Gallon 0 0 5 inch Welch 1405 3 2 60 0 5 0 06 0 02 0 99 531 TC2 C TC 7 Gallon 0 0 5 inch Welch 1405 3 2 60 0 2 0 06 0 02 0 99 531 TC2 C TC 7 Gallon 0 0 5 inch Welch 1405 3 2 60 0 1 1 2 0 75 531 TC2 C TC 7 Gall
11. AC PRODUCT IS PROVIDED AS IS EXCLUSIVE OF ANY WARRANTY INCLUDING WITHOUT LIMITATION ANY IMPLIED WARRANTY OF MERCHANTABILITY FITNESS FOR A PARTICULAR PURPOSE NON INFRINGEMENT OR ANY OTHER WARRANTY EXPRESSED OR IMPLIED LIMITATION OF LIABILITY You agree and acknowledge DIGIVAC shall have no liability to you whatsoever for any direct indirect punitive incidental special consequential damages arising out of or connected with the use or misuse of its products In particular given the nature of DIGIVAC products you agree and acknowledge under no circumstances whatsoever shall DIGIVAC be liable to you for any consequential damages for damage to any non DIGIVAC product or service arising from the failure use or misuse of a DIGIVAC product including but not limited to any vacuum system engine vehicle factory or the like In the event a court of law with proper jurisdiction finds DIGIVAC liable to you for any purpose you agree and acknowledge DIGIVAC s maximum liability shall not exceed the purchase price of one unit of product giving rise to such liability or 250 00 whichever is greater ENTIRE OBLIGATION These terms and conditions express the entire obligation of DIGIVAC with respect to its products If any part of these terms and conditions are deemed void invalid unenforceable or illegal including but not limited to the warranty disclaimers liability disclaimers and liability limitations set forth above then the unenforceable
12. any purchase sale agreement between you and DIGIVAC EXCLUSIVE OBLIGATION The DIGIVAC product you are purchasing has been designed for a specific use within a set of suitable operating conditions as set forth in its User Manual or as indicated otherwise by DIGIVAC Any use of the DIGIVAC Product for any purpose or under any conditions other than those specified shall render any limited warranty void and shall expressly invalidate any liability of DIGIVAC for damages as a result of such misuse USER LIMITATION You may not modify copy distribute transmit display perform reproduce publish license create derivative works from transfer or sell any information software products or services obtained from or created by DIGIVAC to any third party without the express written consent of DIGIVAC to do otherwise Any violation of this provision shall give rise to an indemnification of DIGIVAC by you for any third party claims arising out of such violation THIRTY 30 DAY LIMITED WARRANTY All DIGIVAC products are warranted against any manufactured defect for a period of thirty 30 days from date of purchase unless such product is a custom work for you and not a standard DIGIVAC product Any product qualifying as a custom work shall not be warranted against any defects for any purpose and your acceptance of such custom work shall relieve DIGIVAC of any liability for any purpose WITH THE EXCEPTION OF THE LIMITED WARRRANTY ABOVE YOU AGREE ANY DIGIV
13. but in describing negative pressure the atmosphere is only a satisfactory reference if we are dealing with values of vacuum down to about 27 inches of mercury Below that it is much more useful to talk in terms of absolute pressure starting from absolute zero The DIGIVAC and all similar instruments do just this One TORR a commonly used unit is an absolute pressure of one millimeter of mercury A millitorr is equal to one thousandth of a TORR A MICRON is the same as a millitorr The full scale reading of a DIGIVAC is 1999 microns and is equivalent to 1 999 TORR of approximately 2 760 of atmospheric pressure This is less than 1 inches of mercury and less than 05 PSI DIGIVAC 201 series Instruction Manual Page 15 9 0 ACCESSORIES AND MODIFICATIONS The following are offered as accessory equipment or field installed modifications Padded shoulder strap Case with Velcro closure For instruments that will be used in the field particularly in cryogenic applications a padded shoulder strap case is available This case holds a DIGIVAC 100tc battery powered gauge in the optimal reading position The operator can open the Velcro cover pull out the gauge tube cable plug it into the tube on the equipment and see the reading lt was developed to assist in field service of cryogenic tank farms and vacuum jacketed piping COMPATIBILITY WITH OTHER GAUGE TUBES On special order DIGIVAC Instruments can be provided to use with most other vacuum gauge
14. ed the bleed valve will not operate properly If you require using a gas other than ambient air attach the backfill gas to the bleed outlet at no more than 1psi above ambient air pressure Additionally if you don t require to bleed back an particular gas the valve may be piped simply by putting _ ll ale es the vessel connection on the end sAN Next connect the bleed from vessel tube to a hose barb that accepts 1 16 inner diameter hose we use a 1 8 barb Below is a pictorial of the connections of the entire system DIGIVAC 201 series Instruction Manual Page 7 SE 201 VLC Vacuum Level Bleed valve Controller RS232 or Ethernet PC running optional 201 VLC applet Vacuum Pump Vacuum Vessel 201 VLC Tuning instructions The 201 can operate as a vacuum level controller when so configured The solenoid valve is mounted as close to the vessel as appropriate with the wires for the solenoid valve terminating on pins 1 amp 2 of the 2 position VLC connector Set the VLC set point by setting S in the tune menu to the desired vacuum level to be maintained The 201 has been tested on vessels from 0 2 Liters to 50 Liters with good success In many cases you may not need to touch the PID tuning variables In the event that the control out of the box doesn t work well you can change the P and D variables to obtain a more desirable result To get to the PID tuning variables press the SEL button until you get to
15. l connections to sensors relay outputs and VLC valves with the power disconnected NEVER DISCONNECT SENSOR OR OUTPUT WIRES WHILE UNIT IS POWERED UP In cases where the system has contaminants as is often the case with metalizing and coating equipment it is often effective to isolate the gauge tube with a solenoid or manual valve during periods when contamination is most active _ i AEn ded Model 201 SP1 Torr mbar kPa SP2 The DIGIVAC controller can be easily set to the desired units on the fly 1 Press the sel key three times during normal operation The currently selected units will blink 2 Press the A and v to get to the desired unit 3 Press Ent to complete your selection The DIGIVAC 201 has 2 set points that can be used to actuate external equipment These 2 set points can be adjusted from the front of the gauge in your currently selected units 1 To change SP1 a Press the sel key once to enter in set point 1 The SP1 LED should now blink b Press the A and v to get to the desired set point value Note the set point units are in millitorr For example a set point of 1000 is equal to one Torr c Press enter to accept the new set point value Normal run mode will resume 2 To change SP2 a Press the sel key twice to enter in set point 2 The SP1 LED should now blink b Press the A and v to get to the desired set point value c Press enter to accept the new set
16. n solenoid valve between the gauge tube and the system The solenoid valve coil should be in parallel with the bombarding transformer In this way the solenoid will be closed and the gauge tube will be positively protected whenever bombarding is done DIGIVAC 201 series Instruction Manual Page 4 e fa gauge is damaged by bombarding it can generally be brought back to operating condition by replacing the Op amp which controls the gauge tube current Consult DIGIVAC oe SP2 l For VLC The set point connections are i set point 1 and the bottom row of pins is for set point 2 The top 3 pins are in the order 1 common The common line of a switch 2 N C Normally closed This means that above the set point value there is a current path between the common and the N C terminal Put another way the switch is ON between these 2 terminals At the set point value and below higher vacuum lower pressure the connection is open Put another way the switch is OFF between the common and the N C connection at higher vacuum a lower vacuum reading 3 N O Normally open This means that above the set point value there is no current path between the common and N O connection Put another way the switch is OFF between these 2 terminals When the vacuum indication goes below the set point value higher vacuum lower pressure the current path closes Put another way the switch is ON between the N C and N O c
17. of such gasses include helium oxygen and to a lesser extent CO2 When interpreting readings using gasses other than air it should be borne in mind that the DIGIVAC reads Torr which is a measure of absolute pressure that is the opposite of vacuum Thus a lower numerical reading actually is a higher level of vacuum For more information refer to section 8 0 Note that isolated transducers like capacitance manometers are independent of gas type 8 0 UNDERSTANDING TORR The DIGIVAC and many similar instruments are calibrated in microns or millitorr It is appropriate to discuss what microns are and to relate microns to other measures of pressure and vacuum Microns are not really a measure of vacuum at all but rather of absolute pressure It will be recalled that the pressure of the atmosphere is 14 696 or approximately 14 7 pounds per square inch at sea level This pressure is due to the weight of all of the air in the earth s atmosphere above any particular square inch This 14 696 psi is equivalent to the pressure produced by a mercury column of approximately 29 92 inches high or 76 meters about 3 4 of a yard or 760 millimeters of mercury Atmospheric pressure varies greatly with altitude It decreases approximately 1 inch of mercury per thousand feet of altitude It also varies widely with local weather conditions Variations of one half inch in a single day are common The word vacuum means pressure lower than atmospheric or suction
18. oint c Increase P until oscillations observed at about 10 of average reading not Set Point the reading will likely still be lower than set point d Set P P 2 half the oscillation value of P obtained above e Start with l P 4 at this point the vacuum level should be approaching the set point 2 Tuning a If oscillations are greater then desired decrease P 10 20 at a time b If Vacuum Level is less than the set point increase in increments of 20 until convergence at the set point 3 Rules of thumb a Ifthe vacuum level is below the desired set point with maximum values of P amp I then increase O b If greater then preferred oscillations are occurring when I gt P try setting l P c If greater then preferred oscillations are occurring about a set point and P lt 1 reduce O Below are combinations of pumps vessel sizes and PID variables that were found to show good control This might be a good place to start Product 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC 201VLC Version K11D13 K11D13 K11D13 K11D13 K11D13 K11D13 K11D13 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k11D29 k1
19. on 0 0 5 inch Welch 1405 3 2 60 3 0 1 0 1 0 99 531 TC2 C TC 7 Gallon 0 0 5 inch Welch 1405 3 2 60 5 0 1 0 1 0 99 531 TC2 C TC 7 Gallon 0 0 5 inch Welch 1405 3 2 60 10 0 1 0 1 0 99 531 TC2 C TC 11 2 Gallon 0 0 5 inch Welch 1405 3 2 60 1 0 06 0 02 0 99 531 TC2 C TC 11 2 Gallon 0 0 5 inch Welch 1405 3 2 60 10 0 05 0 05 O 99 531 TC2 C TC 11 2 Gallon 0 0 5 inch Welch 1405 3 2 60 5 0 05 0 05 0O 99 531 TC2 C TC 11 2 Gallon 0 0 5 inch Welch 1405 3 2 60 1 0 05 0 05 0O 99 Model 201 RS232 command line command set DIGIVAC 201 series Instruction Manual Page 9 The 201 can be remotely connected via a DB9 RS2382 cable by setting your HyperTerminal to 9600 baud 8 none and 1 The unit can be set at the front panel using the membrane buttons as described above If you d prefer to control with RS232 use the command set below DIGIVAC 201 series Instruction Manual Page 10 MODEL 201 Cheat Sheet VAC Get vacuum reading Vacl Get vacuum reading of sensor in first slot ex slot A Vac2 Get vacuum reading of sensor in second slot ex slot B Vac3 Get vacuum reading of sensor in third slot ex slot C M Get mode setting Returns M A for automatic mode Returns M M for manual mode M Set mode to automatic M Set mode to manual M D Set to diagnostic mode T Get timing setting This is how frequently we will burp out data in automatic mode in seconds T 25 Set data burp rate to Four times per second T 30 Set data burp rate to
20. onnections at absolute vacuum readings below the set point value Take care in insuring that the wire connections are made fast and the voltage and current does not exceed 250V or 7A If you need to control a device that draws more power consider another relay in between the DIGIVAC output and the device to be controlled The Analog output is located in the center of the back panel and should be connected to a high impedance input The output impedance is 1KQ The RS232 connection can be made to a PLC or computer via a male DB9 cable connection to the female jack on the DIGIVAC The DIGIVAC acts as a DCE so a straight serial connection is appropriate Please use the supplied 5V AC adapter with your Instrument This adapter provides clean short protected power to protect and insure accuracy of the internal circuitry 5 0 OPERATION After installation the DIGIVAC is ready for immediate operation The unit will normally provide accurate readings immediately when used with a thermocouple gauge however occasionally a gauge tube will have absorbed material during storage and may require as much as 24 hours of operation before accurate readings are attained It is recommended that the DIGIVAC be energized continuously during vacuum system operation In this way the hot filament will not allow contaminants to condense DIGIVAC 201 series Instruction Manual Page 5 Only connect and disconnect cables with the power to the unit unplugged Make al
21. ring shipment a claim should immediately be made to the common carrier or the postal service as applicable The DIGIVAC warranty pertains only to the instrument and does not cover losses in shipping Each 201 should come with o 201 User Manual this document Display controller black box with blue buttons Mounting Brackets Feet connectors Power supply Phihong with international adapter sensor driver cards and cables Configured Sensors VLC optional o A Valve with inside port closest to body blocked off o 1 of very thin silicon 1 16 I D tubing to be put on valve o 2 feet of thin silicon 3 32 I D tubing to be connected to very thin tubing on valve and vessel o RS282 optional o R8232 Port o 5 RS232 Cable phono to Female DB9 o Ethernet optional includes RS232 o Ethernet Port Null modem RS232 converter 2 Ethernet Cable Null Modem Ethernet Cable Configuration CD O O O OOO O O O 0 4 0 INSTALLATION The instrument should be located in a clean dry environment for best results The unit can be panel mounted with the hardware provided in a 1 8 DIN panel cutout 3 64 x 1 78 92mm x 45 mm Alternatively the unit can be placed on a desktop by placing the 4 rubber feet included with your gauge on the underside of the unit The gauge tube cable should be identified by wire tags or markings specific to your environment Thermocouple gauge tubes must be installed in a thread down orientation in a clean dry
22. to be made for thermocouple vacuum gauges proceed as follows A Remove the Instrument from the panel B Remove the front panel cover of the instrument and locate the two calibration potentiometers The Zero potentiometer is to the right as you look at the LED Display the ATM or span adjust is to the left closest to the outside of the board C Precalibrate a Set the device on a table top with the tube plugged in b Place a voltmeter lead on ground center pin below display c Place the positive voltmeter lead on the lower side of the beige capacitor on the top of the board to the left of the display d Adjust the Zero POT to read 1 77 Volts e Place the Positive voltmeter lead in the via that is located on the left side of the board towards the edge f Adjust the span pot so the voltage is 1 185 Volts D Operate the vacuum system at the lowest attainable pressure and allow the system and the gauge tube to stabilize for several minutes Factory zero setting is done at a pressure of 1 millitorr 1 micron or less E Adjust the zero setting potentiometer so the unit reads zero Make sure not to under span Allow the measurement standard to rise to 1 millitorr and make sure the gauge reading also reads 1 millitorr Note the POTS are counter intuitive clockwise makes the value go down and vice versa and are very sensitive F Check the operation of the gauge at other pressures Normally adjustment of the zero will not
23. ts each output has Common NC and NO VLC 1 Expansion board for use to enable upstream vacuum level control Eth 1 Expansion board to enable Ethernet connectivity and web management Types of Sensors supported Expansion Sensor CM Total of 1 2 or 3 Setra Capacitance manometer MKS capacitance manometer PCG 750 PVG500 or CDG 500 TC Total of 1 2 or 3 Varian 531 Thermocouple gauge tubes 2 0 CONSTRUCTION The DIGIVAC consists of the indicating and controlling instrument the sensor configured the sensor cable the output interfaces and an international AC power adapter The instrument is housed in a rugged free standing plastic enclosure It can either be placed ona Suitable surface or can be mounted in a 1 8 DIN panel cutout The gauge tube houses the various thermocouple sensing heating and compensating elements and terminates in an octal connector On this model the connector wiring terminates at the instrument with a 6 position RJ24 Regulating circuitry in the DIGIVAC provides constant current for gauge tube excitation and thus compensates for resistance in the probe leads 3 0 UNPACKING AND INSPECTION DIGIVAC 201 series Instruction Manual Page 3 After the instrument is received it should be carefully unpacked and inspected for damage during shipment and for completeness The package should contain as a minimum the instrument the sensor and cable ordered an AC power adapter and an instruction manual In the event of a loss du
24. twice per minute T 60 Set data burp rate to once per minute T 600 Set data burp rate to once per 10 minutes Accepted range for T is 25 to 600 U Get units setting Responds appropriately with U 0 U 1 or U 2 U 0 Set units to Torr U 1 Set units to mBar U 2 Set units to kPa P Get P term P 2 35 Set P term I Get I term I 16 789 Set I term D Get D term D 8 087 Set D term 0O Get O term O 99 Set O term S7 Get S term S 02 Set S term SP1 Get SP1 setting SP1 123 456 Set SP1 SP2 Get SP2 setting SP2 34 506 Set SP2 V Get the Version such as K11D13 If you purchased the Ethernet option please refer to the model 200 Ethernet Addendum for initial setup Your 201 should be set by default with the IP address 192 168 0 200 Surf over to http 192 168 0 200 DIGIVAC 201 series Instruction Manual Page 11 or whatever your IP address was set to and say yes to security screens authorizing a DIGIVAC applet to run in your browser You should see a screen that looks like Y Vacuum Network by DigiVac Mozilla Firefox Edit wiew History Bookmarks Tools Help Connection Yes Logging Yes Units Torr 0 06723 TACTE 0 01517f 3 60m 00 09 46 00 37 25 01 09 46 Feel free to connect to the IP address 10001 with HyperTerminal putty or your favorite comm
25. uge Device TC2 TC2 531 2 CO 1 1 2 3 CP 0 1 Torr CM C1 0 1CP 1 0 Torr CM C10 0 1 2 3 CP 10 Torr CM C100 0 1CP 100 Torr CM C1000 0 1 2 3 CP 1000 Torr CM F70 F70 Varian FRG 700 F72 F72 Varian FRG 720 P75 P75 Varian PCG 750 PV5 PV5 Varian PVG 500 A29 A29 Alcatel ACC 2009 ETH Ethernet 2C Relays FET VLC Control DIGIVAC 201 series Instruction Manual Page 17 Below is the typical response of a thermocouple vacuum gauge tube DIGIVAC gauges which use pressure transducers are linear Milli Milli Torr volts 0 10 Digivac 500 Series Gauge Tubes 5 9 57 10 9 32 20 8 81 25 8 54 35 8 11 50 7 46 19 6 63 100 5 96 150 5 04 200 4 38 300 3 56 500 2 1 2 750 2 24 1000 1 97 1500 1 71 1900 1 59 50 100 200 500 1000 1900 Pressure Millitorr DIGIVAC 201 series Instruction Manual Page 18 11 0 TERMS AND CONDITIONS TERMS OF USE LIMITED WARRANTY amp LIABILITY WAIVER THE DIGIVAC COMPANY DIGIVAC offers all of its products with the following terms and conditions and notices as follows By accepting and or using a DIGIVAC product you hereby acknowledge and agree to the following terms and conditions and acceptance of these terms and conditions are a condition precedent to
26. unication tool to communicate directly with the device To use the 201 web terminal capability highlight Display Terminal DIGIVAC 201 series Instruction Manual Page 12 Vacuum Network by DigiVac Mozilla Firefox EE File Edit wiew History Bookmarks Tools Help l Display Product Help Connection Yes Logging Yes Units Torr Vacuum Network by DigiVac 0 08836 Tan Rate l E i ai ip r52 e o Seconds e 0 01000 0 01000 0 09000 Domain H a E Le n J Range ama Digivac Terminal VAC 759 9996 SP1 0 052 SP2 2 000 Press carriage return or enter to submit the query or change the set point Note that while the web terminal is open the main 201 screen will not update If at any time the web terminal or the main 201 applet window becomes unresponsive close the browser not just reloading the tab or closing the tab and reconnect DIGIVAC 201 series Instruction Manual Page 13 6 0 SERVICING GAUGE TUBE CLEANING In many cases a gauge tube may become fouled with oil or other foreign matter It is often possible to restore the functionality of contaminated probes with cleaning If the contaminant is Known the tube should be filled with a fluid that is known to be a solvent to that contaminant As an example ether is often effective in removing residues of some oils Commercial carburetor cleaners are very powerful sol
27. vents and are highly effective against some contaminants After cleaning with solvents the gauge tube should be completely dried or flushed with a volatile solvent to assure that it is dry prior to re installing it If this is not done contamination of the system may result 6 1 FACTORY REPAIR AND CALIBRATION The vacuum gauge assembly is designed to provide years of trouble free service and the liberal internal use of plug in components make it easily repairable No field servicing of the unit is recommended other than replacement of the gauge tube but factory servicing and calibration are available at a nominal cost and turn around times of 24 hours are typical 6 2 FIELD CALIBRATION Each DIGIVAC vacuum gauge controller is calibrated to the particular vacuum gauge sensor that is shipped with the unit While changing the gauge tube is possible it will result in a slightly different reading as all gauge tubes are not created equal Although it is preferable that all calibration be performed at DIGIVAC field calibration can be accomplished Before re calibrating the instrument it should be ascertained that the instrument is in fact incorrect In many cases the problem will be with a tube that is fouled or a system that is operating improperly It is recommended that a spare tube be kept on hand and stored in a clean dry place Then in cases of suspect readings the tube should be changed before proceeding further If adjustments are
Download Pdf Manuals
Related Search
Related Contents
USER MANUAL FOR SUCTION UNIT MOD.VACUTRON AS Trixie Belinda User's Manual Troubleshooting Guide applicable to: Philips SJA7180 Cell phone Mini Black Pouch Samsung SV-DVD3E دليل المستخدم Mise en page 1 Manual de Usuario Universal Remote Control MRF-200 User's Manual Operating Instructions Type 8045 Copyright © All rights reserved.
Failed to retrieve file