1. 5 Connect the mask system to the free end of the long air tubing The final assembly should look like this 6 Switch on the HUMIDAIRE and the AUTOSET CS2 WARNING Make sure that the power cords and plugs are in good condition and the equipment is not damaged 7 Perform the Air Circuit Learn function see page 26 For information about filling and maintaining your humidifier refer to the HumidAire User s Manual The AUTOSET CS2 is now ready for use with the HUMIDAIRE See Starting Treatment on page 21 16 CAUTION Be very careful not to allow water to enter the AUTOSET CS2 as this could affect therapy and or damage the device Do not tip the HUMIDAIRE while it is connected to the AUTOSET CS2 If water enters the AUTOSET CS2 device turn off the main power switch at the back of the unit and unplug it from the power outlet Return the unit to your equipment supplier for checking Setup the HUMIDAIRE and the AUTOSET CS2 unit so that the tubing to the mask runs higher than the machines This will prevent condensation build up ENGLISH f condensation appears in the mask turn the humidifier settings down RESLINK The RESLINK is a device that records data onto a SmartMedia card during your therapy The card can be returned to your clinician so that they can observe your progress and adjust flow generator parameters if necessary The RESLINK connects to the back of
2. THEALRMS EA ee ae a LOW MASK PRESSURE ALARM POWER FAIL ALARM HIGH PRESSURE ALARM TABLE OF CONTENTS 10 11 13 14 14 15 17 17 18 19 21 21 21 21 22 23 25 26 27 27 28 29 29 31 31 32 32 ENGLISH HIGH LEAK ALARM INSUFFICIENT PRESSURE SUPPORT ALARM FLOW BLOCKED ALARM CLEANING AND MAINTENANCE DAILY WEEKLY PERIODICALLY AIR FILTER HYPOALLERGENIC AIR FILTER SERVICING TROUBLESHOOTING 00 c ccc eee ec e eee eee nes GLOSSARY e CRE ELE ttt mint em ty hE TECHNICAL SPECIFICATIONS n air adi Meine etd LIMITED WARRANTY 0000 cece cee eee e a INTRODUCTION AUTOSET CS2 The AUTOSET CS 2 flow generator has been designed to be reliable comfortable and easy to use You can observe your treatment and make adjustments to settings Using a keypad and LCD screen USER OWNER RESPONSIBILITY The user or owner of this system shall have sole responsibility and liability for any injury to persons or damage to property resulting from ENGLISH operation which is not in accordance with the operating instructions supplied maintenance or modifications carried out unless in accordance with authorised instructions and by authorised persons Please read this manual carefully before use DEFINITIONS The WARNING heading alerts you to possible injury The CAUTION heading explains special measures for the safe and effe
3. Humidity operating storage and transportation 1076 9576 non condensing HOUSING CONSTRUCTION Injection moulded plastic AIR TUBING x 2m AIR FILTER Synthetic fibre PATIENT CONNECTION PORT 22mm taper compatible with EN 1281 1 1997 Anaesthetic amp Respiratory Equipment Conical Connectors POWER SUPPLY AC Input range 10 120 V and 220 240 V 50 60 Hz 60 VA DC 30V via ResMed DC DC converter PERFORMANCE Operating pressure range 4 20 cmH3O SOUND PRESSURE LEVEL lt 30 dB tested in accordance with the requirements of ISO 17510 1 BATTERY TYPE Lithium ALARM SPECIFICATIONS All alarms comply with EN 475 for Medium priority alarms 3 beeps 935 Hz repeated at intervals of 25 seconds yellow LED flashes IGH MAXIMUM MASK PRESSURE ALARM Stops air delivery at 25 cmH O gt 700 msec IGH MEAN MASK PRESSURE ALARI Stops air delivery when mean mask pressure exceeds 15 cmH O averaged over approx minute LOW MASK PRESSURE ALARM Mask pressure is less than 3 cmH3O for longer than 5 seconds while the patient is using the AUTOSET CS2 TECHNICAL SPECIFICATIONS 43 44 INSUFFICIENT PRESSURE SUPPORT ALARM s triggered if for 3 breaths in a row the maximum mask pressure achieved at inspiratory to expiratory transition is either lt 87 5 of the absolute mask pressure targetted at that point or lt absolute mask pressure targetted minus cmH O whichever is the low
4. Insufficient pressure support alarm on page 33 This menu has a series of screens showing information about your most recent treatment session If enabled by your clinician the following data may be viewed e Mask fit Average Pressure Usage USING THE AUTOSET CS2 ENGLISH 27 28 MASK FIT Gives a rating of how well the mask was fitting during your last session on the AUTOSET CS2 The stars indicate how good the fit was more stars better fit Star rating Definition Excellent Very good Good Xo Adjust mask Adjust mask HIGH LEAK Adjust mask AVERAGE PRESSURE Provides information on the average air pressure during your most recent session on the AUTOSET CS2 USAGE Tells you how long the AUTOSET CS2 was in use during your most recent session OPTIONS MENU Smart Data Backlight Language The Options menu allows you to change Auto Appear SMART DATA f the SMART Results menu f you want to BACKLIGHT DATA AUTO APPEAR option is set to ON the screens available under the Results menu will be displayed automatically for approximately 20 minutes after your sessi on is completed After this time the data will still be available in the turn off AUTO APPEAR select enter in the SMART DATA screen by pressing the Left key The AUTO APPEAR screen will be displayed Select change by pressing the Left key and you
5. LK 18L min RR 12 APNEA VT 428 Apnoea Hypopnoea if present Tidal volume Figure 18 Treatment Screen 2 Minute ventilation Target ventilation MV 6 2 TARG 5 8 exit Figure 19 Treatment Screen 3 NOTE TARG is not a relevant parameter in CPAP mode USING THE AUTOSET CS2 ENGLISH 29 30 THE ALARMS The AUTOSET CS2 unit is fitted with alarms to alert you to changes that will affect your treatment The most common reason for an alarm to sound is because the system has not been properly assembled Check that the air tubing and pressure sensor tube have been properly attached to the flow generator and mask and humidifier if used You can mute an alarm by pressing the Alarm Mute key once Unless the alarm requires you to turn the power off you can press any key again to remove the alarm message from the LCD screen If the problem is still present the alarm will sound and display again after one minute The Alarm LED will remain lit for as long as the problem s present ENGLISH All menus remain available during an alarm We recommend that you test the audible alarm once a week To do this hold down the Mute key while the AUTOSET CS2 is starting up If the alarm is functioning correctly it will beep twice LCD message Alarm type Action LOW PRESSURE Low mask pressure alarm Check tube connections Check circuit LCD turns off Power fail alarm Turn off
6. Power switch AC locking clip AC power cord DC locking clip DC power cord Plug the free end of the power cord into a power outlet CAUTION Be careful not to place the device where it can be bumped or where someone is likely to trip over the power cord The AUTOSET CS2 can also function when mounted on a wall unless the HumidAire 2i or HumidAire 2iC is attached The preferred position is with the air tubing at the bottom so that there are no bends in the tubing THE AUTOSET CS2 SYSTEM 11 12 2 Connect air tubing Connect the air tube firmly onto the air outlet at the front of the unit Air tubing connector Pressure sensor tube connector Figure 4 Front view of AutoSet CS2 To attach the pressure sensor tube to the unit screw the Luer connector onto the socket on the side of the unit Figure 5 Twisting the Luer lock connectors together WARNING Use care when handling the air tubing Be particularly careful when attaching the Luer connector to the AUTOSET CS2 If you kink or excessively twist the sensor tube it may affect the protection offered by the mask pressure alarm system 3 Attach the mask to the air tubing The mask system should already be assembled If it is not please refer to the user instructions supplied with the mask Connect the mask system to the free end of the air tubing ENGLISH 7 1 Tm Figure 6 Assembling the AutoSet CS2 system HUM
7. exit LOW PS ON 1 exit r RESULTS enter t exit MASK FIT good 4 exit USAGE 4 56hrs exit OPTIONS enter SMART DATA enter J exit AUTO APPEAR OFF BACKLIGHT AUTO 1 exit change exit LANGUAGE ENGLISH change exit SERVICING enter t SN 1234567890123 4567 exit PCB 123456789012 exit SW SX245 0117 1 exit Figure 16 Navigating the Patient menu These menus appear only if at least I Smart Data option has been enabled by AVG PRESS 10 4 M exit 1 CPAP MODE RAMP 20min ichange exit MASK FULL FACE change exit LEARN CIRCUIT yes exit SMARTSTOP OFF change t exit _ TON 7 USING THE AUTOSET CS2 ENGLISH 25 26 SETTINGS MENU You will only see the options relevant to your therapy In CPAP Continuous Positive Airway Pressure mode the clinician sets a pressure that will be constant throughout treatment In ASV CS mode AUTOSET CS2 continually measures your breathing and regulates the air pressure accordingly MASK Select the type of mask that you will be using you will only need to adjust this if you have more than one mask The following table shows the setting that should be selected for each mask type If your mask type is not mentioned below please contact your Clinician Not all mask types can be used with the AUTOSET CS2 Option in Settings menu Use setting wit
8. 27 learn circuit 26 Luer connector 12 M mask connection 13 indication of fit 28 recommended 10 select from menu 26 medical information 5 menus 23 how to use 24 navigation 25 O options menu 28 P power cord AC 11 connection 11 DC 11 locking clip 11 pressure sensor tube connecting 12 proximal cuff 10 ENGLISH Q quick setup guide 7 R ramp 26 27 ResLink 17 results menu 27 S settings menu 25 INDEX 53 54 Smart data 28 SmartStop 27 sound level 27 standby mode 21 starting treatment 21 stopping treatment 21 supplemental oxygen 19 System Error 7 message 39 T treatment screens 29 U user owner responsibility 5
9. AUTOSET CS2 as shown For full details on using the RESLINK please refer to the ResLink Users Manual ANTI BACTERIAL FILTER Your clinician may recommend that you use an anti bacterial filter A filter should be placed in the air tubing between the mask and the flow generator If using a humidifier place the filter between it and the flow generator see Figure 10 below Typically the filter should be replaced daily Check the filter instructions for full details Whenever a filter is added removed or replaced you must run the Air Circuit Learn function see Learn Circuit on page 26 Note Only hydrophobic anti bacterial filters should be used with a HUMIDAIRE 2i or HUMIDAIRE 2iC THE AUTOSET CS2 SYSTEM 17 18 Anti bacterial filters gt uh Figure 10 Inserting anti bacterial filters USING DC POWER TO RUN THE AUTOSET CS2 The AUTOSET CS2 can be run on DC current in case of power failure or for portable use ResMed supplies a convertor that allows a 2V battery to be connected to the DC input of the AUTOSET CS2 The converter can also be plugged in to a power source in a car or plane When using DC current power is turned on and off via the switch on the connector to the power source not the power switch at the back of the AUTOSET CS2 The DC converter is a separate accessory that can be purchased through your ResMed distributor The humidifer heater will be automatically disabled w
10. AutoSet CS2 at power switch HIGH PRESSURE High pressure alarm Turn off AutoSet CS2 at power switch Turn power off HIGH AVG PRESS High pressure alarm Turn off AutoSet CS2 at power switch Turn power off HIGH LEAK High leak alarm Adjust the mask Check circuit LOW PS Insufficient Pressure Check tubing Check circuit Support alarm Run LEARN CIRCUIT FLOW BLOCKED Flow Blocked alarm Turn off AutoSet CS2 Check air Turn power off circuit Turn on AutoSet CS2 LOW MASK PRESSURE ALARM TRIGGERED BY pressure sensor line is blocked or disconnected air pressure at the mask has fallen below a set level mask is removed and SmartStop has been disabled IF A LOW PRESSURE ALARM SOUNDS Check that the air tube and sensor tube are connected properly 2 Turn the AUTOSET CS2 off and on again at the power switch If the alarm persists return the unit to ResMed for servicing THE ALARMS 31 WILL CLEAR WHEN the low pressure condition is fixed treatment is stopped by pressing the Start Stop key POWER FAIL ALARM TRIGGERED BY power failure machine is disconnected or switched off while delivering treatment IF A POWER FAIL ALARM SOUNDS The flow generator stops delivering air pressure WILL STOP when the Alarm Mute key is pressed or after 2 minutes or when power is restored CAUTION Remove the mask from your face if the power fails HIGH PRESSURE ALARM NOTE TRIGGERED BY the
11. Max Pressure 10 BPM 15 BPM 20 BPM Volume mL 8 Pressure cmH2O NOTE The manufacturer reserves the right to change these specifications without notice GLOSSARY OF SYMBOLS Read this Operating Manual before using TECHNICAL SPECIFICATIONS 45 N Type CF equipment AC switch Class Il Environmental information WEEE 2002 96 EC is a European Directive that requires the proper disposal of Ol electrical and electronic equipment This device should be disposed of separately not E as unsorted municipal waste To dispose of your device you should use appropriate collection reuse and recycling systems avallable in your region The use of these collection reuse and recycling systems is designed to reduce pressure on natural resources and prevent hazardous substances from damaging the environment If you need information on these disposal systems please contact your local waste administration The crossed bin symbol invites you to use these disposal systems If you require information on collection and disposal of your ResMed device please contact your ResMed office local distributor or go to www resmed com environment GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC EMISSIONS AND IMMUNITY Guidance and manufacturer s declaration electromagnetic emissions The AutoSet CS2 is intended for use in the electromagnetic environment specified below T
12. according to the instructions in the manual Remove the air tubing and pressure sensor tube from the AUTOSET CS2 unit and the mask 2 Remove the pressure sensor tube and Proximal cuff from the air tubing 3 Wash the mask system according to the instructions supplied with it 4 Wash the air tubing and Proximal cuff in warm water using mild detergent Rinse thorough y hang and allow to dry 5 Before next use assemble the mask and headgear according to the mask user instructio ns 6 Reconnect the pressure sensor tube to the air tubing then connect both to the Proximal cuff Attach the Proximal cuff to the mask Air tubing Remove Proximal cuff amp Remove the pressure sensor tube from air tubing Tubing clips Figure 20 Undoing the air tubing assembly for cleaning CLEANING AND MAINTENANCE ENGLISH 35 CAUTION Do not use bleach chlorine alcohol or aromatic based solutions including all scented oils moisturising or antibacterial soaps to clean the cushion mask air tubing or the AUTOSET CS2 These solutions may cause hardening and reduce the life of the product Do not wash or dry the mask frame at a temperature above 80 C 176 F Exposure to higher temperatures may reduce the life of the product Do not hang the air tubing and pressure sensor tube in direct sunlight as the tubing may harden over time and eventually crack
13. of AutoSet CS2 The AUTOSET CS2 can run on AC or DC power See page to see how to connect the power cords and use the locking device See page 7 to connect a RESLINK See page 36 to replace the air filter THE AUTOSET CS2 SYSTEM Proximal cuff Tube clip Pressure sensor tube Luer connector Cr mtm Won nim mim mim D Connect this end Air tubing Connect this end to the mask to the flow generator Figure 3 Air tubing fully assembled All necessary air tubing components are packed with the AUTOSET CS2 2 metre length of air tubing e pressure sensor tube to measure pressure at the mask with Luer connectors tube clips to hold the pressure sensor tube onto the air tubing a Proximal cuff a special connector that attaches the pressure sensor tube into the main air tubing so that you only need to make one attachment to the mask The air tubing is supplied with the Proximal cuff connected Attach the tube clips evenly along the air tubing Press the pressure sensor tube into the tube clips and screw the Luer connectors together at the Proximal cuff Make sure the pressure sensor tube is not kinked obstructed or twisted Note ResMed recommends the following tubing products for use with the AutoSet CS2 replacement air hose only PN 14948 replacement air delivery system hose sensor line dips proximal cuff PN 26909 MASKS The following mask systems supplied sep
14. supplier for service Return the device to your equipment supplier for service TROUBLESHOOTING 39 40 GLOSSARY Alarms When more than one alarm is activated at the one time the highest priority alarm will be displayed Backlight The backlight turns off if no button has been pressed for two minutes It can also be kept permanently on by selecting Auto in the Backlight screen ofthe Options menu bpm breaths per minute L min litres per minute Mean The average value of the monitored parameter Minute ventilation A measure of the respiratory rate x tid Proximal cuff A connector between the air tubing and the sensor tubing attaches to the Proximal cuff so that pressure at detected and monitored SmartStart A feature that starts the AUTOSET CS2 automat into the mask al volume mask The pressure the mask can be cally when you breathe SmartStop A feature that stops the AUTOSET CS2 automatically when you remove the mask You can turn SmartStop on or off in the Settings menu Tidal volume The volume of air in litres breathed in or ou t in one breath GLOSSARY ENGLISH 41 42 TECHNICAL SPECIFICATIONS DIMENSIONS H X W x D 142 x 248 x 293mm 5 6 x 98 x 11 5 WEIGHT Flow generator with cover 3 7kg Flow generator with HUMIDAIRE 21 empty 4 2kg ENGLISH ENVIRONMENTAL CONDITIONS Operating temperature 5 C to 40 C Storage and transport temperature 20 C to 60 C
15. will switch from ON to OFF similarly you can turn it from OFF to ON Your clinician may also turn off AUTO APPEAR On Display a permanent backlight for the LCD Auto Backlight turns off if no button has been pressed for two minutes LANGUAGE Change the language of the LCD display Languages available are English French German Italian Spanish Portuguese Swedish Dutch SERVICING MENU This menu displays the following information SN the flow generator s serial number PCB the flow generator s PCB number printed circuit board SW the version of software currently installed on the flow generator TREATMENT SCREENS There are three Treatment screens displayed in the AUTOSET CS2 LCD The LCD will display the Treatment screens when treatment commences and during treatment if no key is pressed for 20 minutes Press the QuickView key at any point within a menu to display Treatment Screen You can return to your original screen if you press the QuickView key again within 20 minutes Press the Right key to move from a Treatment screen to the Standby screen If an alarm is triggered while you are in the Treatment screens the arrow will flash Press the Down key until you reach the alarm message Scroll through the screens in the usual way using the Up Down key Treatment mode Pressure ASV CS 5 0 15 0 LUI Y exit Figure 17 Treatment Screen 1 Leak Respiratory rate
16. IDIFIERS A humidifier may be required if you are experiencing dryness of the nose throat or mouth The following ResMed humidifiers are compatible with the AUTOSET CS2 HUMIDAIRE HUMIDAIRE 2i HUMIDAIRE 2iC heated humidifier heated humidifier passover humidifier HUMIDIFIER ACCESSORY HUMIDAIRE ONLY Medium air tubing 52cm Figure 7 Humidifiers compatible with the AutoSet CS2 When using a humidifier stop the AUTOSET CS2 by using the Start Stop key or by removing the mask to trigger the SMARTSTOP function Do not switch the AUTOSET CS2 off at the power switch or the power source while it is running as this could result in condensation in the motor THE AUTOSET CS2 SYSTEM 13 14 CONNECTING A HUMIDIFIER NOTE After connecting or removing a humidifier always perform the Air Circuit Learn function see page 26 WARNING When using a humidifier position it so that it is lower than your head will be during sleep and at the same level or lower than the AUTOSET CS2 HUMIDAIRE 2i AND HUMIDAIRE 2iC The HUMIDAIRE 2i provides heated humidification and the HUMIDAIRE 2iC provides passover humidification To attach the HUMIDAIRE 2i or HUMIDAIRE 2iC remove the AUTOSET CS2 front cover and attach the humidifier docking station and water chamber Please refer to the HUMIDAIRE 2i or HUMIDAIRE 2iC User s Manual for details No other accessories are required for use Once you have performed the Air Circui
17. PERIODICALLY The mask and air tubing are subject to normal wear and tear Inspect them regularly for damage 2 Open the power cord locking clip and remove the cord Clean the exterior of the flow generator with a damp cloth and mild detergent 3 Inspect the air filter to check if it is blocked by dirt or contains holes See full instructions below WARNING Beware of electric shock Do not immerse the flow generator or power cord in water Always unplug the flow generator before cleaning and be sure that it is dry before reconnecting CAUTION Do not attempt to open the AUTOSET CS2 There are no user serviceable parts inside Repairs and internal servicing should only be performed by an authorised service agent AIR FILTER Inspect the air filter every month and check if it is blocked by dirt or contains holes Air filter cover The filter should be replaced every 6 months or more often if the flow generator is operating in a dusty environment 36 NOTE WARNING Do not wash the air filter once it has become soiled The air filter is not washable or reusable FITTING THE FILTER Remove the filter cover at the back of the flow generator Remove and discard the old air filter Insert a new filter the blue tinted side facing towards you Match the cut corner of the filter with the shape of the casing Replace the air filter cover The blue tinted side of the filter should be facing out HYPOALLERGENIC AI
18. R FILTER The ResMed hypo allergenic filter is a standard filter to which an electrostatic filter has been bonded to capture very small particles The filter is intended for those users of ResMed flow generators who will benefit from enhanced filtering of the air delivered during treatment nspect the air filter every month and check if it is blocked by dirt or contains holes The filter should be replaced every three months or more often if the flow generator is operating in a dusty environment Fit as described above SERVICING This product AUTOSET CS2 should be inspected by an authorised ResMed Service Centre 5 years from the date of manufacture Prior to this the device is intended to provide safe and reliable operation provided that it is operated and maintained in accordance with the instructions provided by ResMed Applicable ResMed warranty details are provided with the device at the time of original supply Of course as with all electrical devices if any irregularity becomes apparent you should exercise caution and have the device inspected by an Authorised ResMed Service Centre CLEANING AND MAINTENANCE ENGLISH 37 38 TROUBLESHOOTING If there is a problem try the following suggest contact your equipment supplier or ResMed generator Problem Insufficient air delivered from flow generator Possible cause Air filter is dirty Air tubing is kinked or punctured ions If the problem cannot
19. RESMED AUTOSET CS 2 USER S MANUAL English ENGLISH ResMed Ltd 1 Elizabeth Macarthur Drive www Bella Vista NSW 2153 Australia Distributed by ResMed Corp 9001 Spectrum Center Boulevard San Diego CA 92123 USA EC REP ResMed UK Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire 0X14 4RW UK See Resmed com for other ResMed locations worldwide For patent information see Resmed com ip SmartMedia is a trademark of Toshiba Papillon is a trademark of MAP Medezin Technologie GmbH Activa AutoSet CS HumidAire HumidAire 2i HumidAire 2iC Mirage ResControl ResLink SmartStart Smart Data Ultra Mirage and Vista are trademarks of ResMed Ltd O 2015 ResMed Ltd 268333 1 2015 10 ResMed com TABLE OF CONTENTS INTRODUCTION nn vete xxv o a AUTOSET CS2 USER OWNER RESPONSIBILITY MEDICAL INFORMATION INTENDED USE CONTRAINDICATIONS WARNINGS CAUTIONS QUICK SETUP GUIDE cete eee bathe a V e ES THE AUTOSET CS2 SYSTEM ees AUTOSET CS2 COMPONENTS MASKS ASSEMBLING THE AUTOSET CS2 SYSTEM HUMIDIFIERS CONNECTING A HUMIDIFIER HUMIDAIRE 21 AND HUMIDAIRE 21C HUMIDAIRE RESLINK ANTI BACTERIAL FILTER USING DC POWER TO RUN THE AUTOSET CS2 USING SUPPLEMENTAL OXYGEN USING THE AUTOSET CS2 STARTING TREATMENT STOPPING TREATMENT STANDBY MODE CONTROL PANEL FUNCTIONS ADJUSTING SETTINGS MENUS SETTINGS MENU ALARMS MENU RESULTS MENU OPTIONS MENU SERVICING MENU TREATMENT SCREENS
20. arately are recommended for use with the AUTOSET CS2 system Please refer to the table on page 26 to set up your mask correctly for use with the AUTOSET CS2 ULTRA MIRAGE MIRAGE FULL FACE FuLL FACE MASK MASK SERIES 2 MIRAGE ACTIVA ULTRA MIRAGE Your clinician can explain the features of the masks available and will discuss the mask that best suits your needs 10 ASSEMBLING THE AUTOSET CS2 SYSTEM Complete the following steps to assemble the AUTOSET CS2 system Make sure the area around the unit is clean dust free and clear of bedding clothes and any other potential blockages To add a humidifier see page 14 WARNING Use only the air tubing system supplied with the AUTOSET CS2 system ENGLISH NOTE Connect the power cord Place the flow generator on a table near the head of your bed There are two sockets at the rear of the flow generator the top one for an AC standard mains electricity and the lower one for a DC backup power cord see page 18 ResMed recommends using the AC power cord supplied with the unit In the absence of this cord a standard AC power cord may be used The customised ResMed power cord can be held in place with the locking clips To insert the locking clip pinch the free ends together and fit the pins into the holes on both sides of the socket Insert the power cord into the socket Push the locking clip down so that the groove holds the power cord in place
21. at your ResMed product shall be free from defects in material and workmanship for the period specified below from the date of purchase by the initial consumer This warranty is not transferable ENGLISH Product Warranty Period ResMed humidifiers ResControl ResLink ResTraxx 1 Year ResMed flow generators 2 Years Accessories mask systems including mask frame cushion 90 Days headgear and tubing Excludes single use devices Note Some models are not available in all regions Ifthe product fails under conditions of normal use ResMed will repair or replace at its option the defective product or any of its components This Limited Warranty does not cover a any damage caused as a result of improper use abuse modification or alteration of the product b repairs carried out by any service organization that has not been expressly authorized by ResMed to perform such repairs c any damage or contamination due to cigarette pipe cigar or other smoke d any damage caused by water being spilled on or into a flow generator Warranty is void on product sold or resold outside the region of original purchase Warranty claims on defective product must be made by the initial consumer at the point of purchase This warranty is in lieu of all other express or implied warranties including any implied warranty of merchantability or fitness for a particular purpose Some regions or states do not allow li
22. be solved Do not attempt to open the flow Solution ENGLISH Replace air filter Straighten or replace tubing Flow generator operating erratically ie switching on and off pressure varying at random Water may have entered the air tubing Remove the water from the air tubing Pressure rises inappropriately You are talking coughing voluntarily breath holding eg while rolling over in bed or intentionally breathing in an unusual manner Severe mask leak or mouth leak greater than 30 L min Avoid talking with the mask on and breathe as normally as possible Correct your leak condition Alarm stays on continuously Blockage in air circuit Internal malfunction Ensure there is no barrier to the passage of air Return the device to your equipment supplier for service No LCD display Power not connected or switch at back is not on Ensure the power cable is connected and that the switch at the back of the unit is in the on position FAULT message displayed on LCD Internal malfunction Return the device to your equipment supplier for service System Error 7 message Device power on has been interrupted by breathing into the mask during warm up Internal malfunction Turn the flow generator off and on again Do not place mask on face during warm up 30 seconds Select LEARN CIRCUIT page 26 If the error is not cleared contact your equipment
23. ction indicated by the text above it in the LC RIGHT KEY RED Performs the function indicated by the text above it in the LCD eg exit Cancels operations Extended hold exits to the top level of the menu QUICKVIEW KEY Takes you immediately to the Treatment screens with information on your current atment Screens on page 29 therapy See Tre ALARM LEDS AND ALARM MUTE The Alarm LEDs are lights that indicate that an alarm has been triggered Alarms can be muted by pressing the Alarm mute key once The Alarm LE as long as the alarm is being triggered See The Alarms on page 31 ADJUSTING SETTINGS By using the Left Right and Up Down keys you can adjust some aspects of your treatment The settings can be seen in the LC The settings are arranged into five menus Table 1 Menus and adjustments for AutoSet CS2 D display D eg enter change ENGLISH Ds will remain lit for Settings Alarms Results Options Servicing Ramp CPAP Sound level Mask fit Smart data SN serial mode only LOW MED Auto appear number change time HIGH on off Mask type LOW PS Average Pressure Backlight Auto PCB printed change type on circuit board Learn circuit run Usage Language English French German Italian Spanish Swedish Portuguese Dutch SW software version SmartStop off on Leak alert off on These menus appear only if at least one S
24. ctive use of the device MEDICAL INFORMATION INTENDED USE The AUTOSET CS2 is indicated to stabilise the ventilation of adult patients exhibiting central sleep apnoea CSA mixed sleep apnoea and periodic breathing with or without obstructive sleep apnoea It is intended for home and hospital use CONTRAINDICATIONS The AUTOSET CS2 is not a life support ventilator and may stop operating with power failure or in the unlikely event of certain fault conditions ASV therapy is contraindicated in patients with chronic symptomatic heart failure NYHA 2 4 with reduced left ventricular ejection fraction LVEF lt 45 and moderate to severe predominant central sleep apnoea Before using the AUTOSET CS2 tell your doctor if you have any of the following conditions acute sinusitis or otitis media epistaxis severe nose bleeds causing a risk of pulmonary aspiration conditions predisposing to a risk of vomiting into mask impaired ability to clear secretions hypotension or significant intravascular volume depletion pneumothorax or pneumomediastinum recent cranial trauma or surgery Below are general warnings and cautions Further specific warnings cautions and notes appear next to the relevant instructions in the manual INTRODUCTION WARNINGS The AUTOSET CS2 is NOT a life support ventilator Before putting patients on ASV each patient should be assessed for heart failure In case of
25. er HIGH LEAK ALARM FLOW BLOCKED ALARM IEC 601 1 CLASSIFICATIONS Class Il Type CF Leak exceeds 30 L min for more than 20 seconds Flow is less than 6 L min for more than 20 seconds Class Il medical electrical equipment provides protection against electric shock either by double insulation or reinforced insulation and does not require a protective earth Table 3 Displayed values Display Value Range Accurac 8 y Resolution Proximal cuff Pressure 3 to 40 cmH2O 0 5 cmH2O 0 1 cmH2O Mass flow sensor Leak 0 99 litres per 12 litres per 1 litre per breath minute minute Respiratory rate 8 30 breaths per 10 1 breath per minute minute Tidal volume 0 4500 millilitres 10 1 millilitre Minute ventilation 0 40 litres per 10 0 1 litres per minute minute Target MV 0 40 litres per 10 0 1 litres per minute minute Table 4 Maximum low impedance flow at stated pressures Pressure cmH O Flow L min 4 2 217 8 6 218 13 0 219 PRESSURE VARIATION 16 4 eaa Max Pressure r Q 12 o 5 10 4 2 3 Max Pressure z o 8 5 94 1 3 Max Pressure E 4 10BPM 2 15BPM D 20BPM 0 0 1 2 3 Breaths PRESSURE VOLUME CURVE 1 3 Max Pressure 2 3 Max Pressure a E 9 L E os gt gt Ll 7 Pressure cmH5O Pressure cmH2O
26. esence of an open flame THE AUTOSET CS2 SYSTEM 19 20 USING THE AUTOSET CS2 WARNING Before starting therapy with a new mask select the correct mask type in the Settings menu page 26 When adding or removing a new component such as a mask humidifier or anti bacterial filter perform the Air Circuit Learn function page 26 ENGLISH STARTING TREATMENT When the system is fully assembled press the power switch at the back of the AUTOSET CS2 to ON The Welcome screen will be displayed while the flow generator is warming up Allow the unit to warm up for approximately 30 seconds before placing the mask on your face Figure 12 Welcome screen Once the AUTOSET CS2 has warmed up breathing into the mask should start therapy immediately If not press the Start Stop key once to commence treatment The AUTOSET CS2 restarts in the mode in which it stopped STOPPING TREATMENT If SMARTSTOP is enabled treatment will stop automatically when you remove your mask If the SMARTSTOP function has been disabled press the Start Stop key STANDBY MODE NOTE AUTOSET CS2 menu Figure 13 Standby screen Once you have stopped treatment the AUTOSET CS2 will be in standby mode In standby you can press the power switch at the back to OFF to turn off the flow generator commence treatment again by breathing into the mask or pressing the Start Stop key enter any of the menus by pressing the Left key When in standby mode a gent
27. h Vista VISTA Ultra ULTRA MIRAGE Full Face MIRAGE FULL FACE MASK SERIES II ULTRA MIRAGE FULL FACE MASK Activa ACTIVA Papillon Papillon Table 2 Selecting mask options in the Settings menu LEARN CIRCUIT The Air Circuit Learn function takes approximately 20 seconds to run It allows the unit to learn what pressure it needs to build in to its calculations based on the components in the system Select LEARN CIRCUIT every time you add or remove a component eg mask humidifier or move to a significantly different altitude To run this function Set up the entire AUTOSET CS2 system flow generator mask humidifier etc as you wish to use it 2 Make sure that the mask is unobstructed so air can flow from the mask to the flow generator Turn on the AUTOSET CS2 at the power switch 4 Move through the menus until you reach LEARN CIRCUIT and select yes by pressing the Left key 5 A message will remind you to check that the airflow from the mask is clear Select start with the Left key 6 A progress screen is displayed while the Air Circuit Learn function is running LEARNING CIRCUIT III WwW NOTE 7 When the function is complete a screen will appear saying that the circuit has been learned successfully The next menu item is then displayed Commence treatment as described in Starting Treatment on page 21 If there is a problem with the circuit a screen saying Invalid ci
28. he customer or the user of the AutoSet CS2 should assure that it is used in such an environment baa Electromagnetic environment Emissions test Compliance guidance RF emissions CISPR11 Group 1 The AutoSet CS2 uses RF energy only for its internal function Therefore its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment RF emissions CISPR 11 Class B The AutoSet CS2 is suitable for use in all establishments including domestic Harmonic Emissions Class A establishments and those directly IEC 61000 3 2 connected to the public low voltage network that supplies buildings used for Voltage Fluctuations Flicker Complies domestic purposes Emissions IEC 61000 3 3 Medical Electrical Equipment needs special precautions regard ing EMC and needs to be installed and put into service according to EMC information provided in this document Warnings The AutoSet CS2 should not be used adjacent to or stacked with other equipment If adjacent or stacked use is necessary the AutoSet CS2 should be observed to verify normal operation in the configuration in which it will be used The use of accessories eg Humidifiers other than those specified in this manual is not recommended They may result in increased emissions or immunity of the AutoSet CS2 TECHNICAL SPECIFICATIONS ENGLISH 47 Guidance and manufacturer s declaration electromagnetic immunit
29. hen the AUTOSET CS2 is operating from the DC supply Power consumption of the AUTOSET CS2 varies with treatment settings A typical current of between 2 5A and 3 0A will be drawn from a 12V battery during use Adequate battery type and sizing is required before operation on DC and it is recommended that ResMed s Technical Services division be contacted for information on your specific application CAUTION power using the ResMed DC DC convertor specified for this application The AUTOSET CS2 uses 30V DC power Only connect the AUTOSET CS2 to DC USING SUPPLEMENTAL OXYGEN If oxygen is being added at the mask the oxygen tube is attached to the mask port Refer to your mask manual for full instructions Up to 15 L min of oxygen can be added at the mask when using an AUTOSET CS2 system ENGLISH Oxygen tube Oxygen tube Figure 11 Oxygen tube attached to Mirage FFMS2 and Ultra Mirage masks WARNINGS e If oxygen is used with this device the oxygen flow must be turned off when the device is not operating Explanation When the device is not in operation and the oxygen flow is left on oxygen delivered into the ventilator tubing may accumulate within the device enclosure and create a risk of fire Always begin AUTOSET CS2 therapy before the oxygen supply is turned on Always turn the oxygen supply off before stopping AUTOSET CS2 therapy Oxygen supports combustion Oxygen should not be used while smoking or in the pr
30. le non therapeutic flow of air will still be present If the SMART DATA AUTO APPEAR option is set to ON the screens available under the Results menu will be displayed automatically for approximately 20 minutes after your session is completed USING THE AUTOSET CS2 21 22 CONTROL PANEL FUNCTIONS The control panel of the AUTOSET CS2 has a keypad that allows you to start or stop treatment adjust settings on the flow generator mute or stop alarms view treatment data QuickView key LCD display Alarm LEDs Alarm mute key Start Stop Left Up Down Right key key key key Figure 14 The AutoSet CS2 control panel LCD SCREEN Displays information about flow generator settings alarms and treatment The LCD screen is backlit whenever any key is pressed The lighting will turn off automatically when no button has been pressed for 2 minutes or can be set to stay on see Options menu on page 28 An alarm message will overwrite any other message on the screen The original screen will reappear when any key is pressed START STOP KEY Press the key once to start or stop treatment UP DOWN KEY Moves backwards and forwards within a menu and between menus e Increases and decreases values of parameters press once to adjust in single increments hold down to move quickly through the values LEFT KEY GREEN Enters a menu Confirms and applies settings Performs the fun
31. mart Data option has been enabled by the clinician To change a setting press the Up Down key until you arrive at the menu you require e Press the Left key to enter the menu then use the Up Down key again to scroll through the options USING THE AUTOSET CS2 23 24 Once you find the option you require press the Left key to select the function that is displayed in the bottom left of the screen eg change yes Use the Up Down key to alter the parameters and press the Left key when you have the setting you require Press the Right key to exit from the menu EXAMPLE You wish to change the sound level for the alarms SETTINGS enter Press down arrow on Up Down key to move to next menu ALARMS Press Left key to enter menu options enter A SOUND LEVEL LOW change exit 1 Press Left key to select change 2 Press Up Down key to reach setting 3 Press Left key to enter setting 4 Press Right key to exit RESULTS enter i OPTIONS enter SERVICING enter Figure 15 Using the keys to adjust settings MENUS The options avallable in the Settings menu will depend on the therapy you are receiving set by your clinician AUTOSET CS2 menu SETTINGS enter ASV CS MODE MASK FULL FACE change exit LEARN CIRCUIT yes 1 exit SMARTSTOP OFF change 1 exit LEAK ALERT OFF change 1 ALARMS enter 1 SOUND LEVEL LOW change
32. mask pressure exceeds the set trigger level 25 cmH3O for more than 700 milliseconds IF A HIGH PRESSURE ALARM SOUNDS The treatment will stop Turn power off Check that the air tube and sensor tubes are connected properly Turn power back on Remove mask and perform Air Circuit Learn function Try using the flow generator one more time If the high pressure alarm activates repeatedly discontinue use and return to ResMed for servicing If the alarm does not recur then continue to use as normal TH Os UD BwN The alarm system is designed to ignore coughing However if a cough is especially intense and prolonged it may trigger the high pressure alarm WILL STOP WHEN the AUTOSET CS2 is turned off HIGH LEAK ALARM TRIGGERED BY high mask leak greater than 30 L min for more than 20 seconds IF A HIGH LEAK ALARM SOUNDS Adjust the mask to minimise leak WILL STOP WHEN the mask leak has been rectified Note The High Le ak alarm can be turned off in the Leak Alert option in the Settings menu INSUFFICIENT PRESSURE SUPPORT ALARM TRIGGERED BY air pressure at the mask has failed to reach an expected level for 3 breaths the Air Circuit Learn function page 26 has not been run after adding a new ENGLISH component eg new mask humidifier to the system IF AN INSUFFICIENT PRESSURE SUPPORT ALARM SOUNDS e check that the pressure sensor tubing is not kinked e check that the pressure sensor t
33. mitations on how long an implied warranty lasts so the above limitation may not apply to you ResMed shall not be responsible for any incidental or consequential damages claimed to have occurred as a result of the sale installation or use of any ResMed product Some regions or states do not allow the exclusion or limitation of incidental or consequential damages so the above limitation may not apply to you This warranty gives you specific legal rights and you may also have other rights which vary from region to region For further information on your warranty rights contact your local ResMed dealer or ResMed office LIMITED WARRANTY R001 307 2 05 06 52 INDEX A air circuit learn function 26 air filter 36 hypoallergenic 37 air pressure monitoring 28 air tubing attaching to mask 13 cleaning 35 connecting 12 air tubing system 9 10 12 alarm flow blocked 33 high leak 32 high pressure 32 low mask pressure 31 power fail 32 alarm LED 23 alarms menu 27 sound level 27 anti bacterial filter 17 ASV CS mode 26 AutoSet CS2 9 usage 28 B backlight change setting 28 C cautions 6 contraindications 5 CPAP mode 26 D DC current 18 F Fault message 39 H HumidAire connection 15 HumidAire 2i 14 HumidAire 2iC 14 humidifier connecting 14 types 13 key alarm mute 23 31 left 23 right 23 start stop 22 up down 22 L language change setting 28 LCD display 39 leak alert
34. n distances between portable and mobile RF communications equipment and the AutoSet CS2 The AutoSet CS2 is intended for use in an environment in which radiated RF disturbances are controlled The customer or the user of the AutoSet CS2 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment transmitters and the AutoSet CS2 as recommended below according to the maximum output power of the communications equipment Separation distance according to frequency of transmitter m Rated maximum output power of transmitter 150kHz to 80MHz 80 MHz to 800 MHz 800MHz to 2 5 GHz w d 1 17 d 0 35 d 0 70 NP 0 01 0 17 0 04 0 07 0 1 0 37 0 11 0 22 1 1 17 0 35 0 70 10 3 69 1 11 2 21 100 11 70 3 50 7 00 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be determined using the equation applicable to the frequency of the transmitter where P is the ing of the transmitter in watts W according to the transmitter manufacturer maximum output power ra NOTE 1 At 80 MHz and 800 MHz the separation distance for the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people LIMITED WARRANTY ResMed warrants th
35. of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation distance in meters m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compliance level in each frequency range Interference may occur in the vicinity of equipment marked with the following symbol t NOTE 1 At 80 MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the AutoSet CS2 is used exceeds the applicable RF compliance level above the AutoSet CS2 should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the AutoSet CS2 P Over the frequency range 150 kHz to 80 MHz field strengths should be less than 10 V m TECHNICAL SPECIFICATIONS 49 50 Recommended separatio
36. rcuit will be displayed instead of the progress screen Check that all the connections between the pieces of equipment are secure You may need to return to the previous chapter of this manual to check that you have set up the system correctly SMARTSTOP Turn the SMARTSTOP function on or off When SMARTSTOP is on the AUTOSET CS2 will automatically turn itself off when the mask is removed The unit will automatically SmartStart whenever you start to breathe into the mask LEAK ALERT Turn Leak Alert on or off Leak Alert sounds an alarm when you have a high leak from your mask You can then adjust the mask on your face to remove the leak f Leak Alert is turned off an extra leak alert message will be displayed at the end of a session if high leak occurred The message will be displayed until a key is pressed RAMP CPAP MODE ONLY Ramp time is the length of time taken for the air pressure to build up to the maximum pressure set by your clinician This setting allows you to increase or decrease the ramp time in 5 minute intervals f you find that you are still awake when the air pressure starts to build you may wish to increase the Ramp time ALARMS MENU RESULTS MENU SOUND LEVEL You can change the sound level of the alarms in the Alarms menu Choose low medium MED or high level to suit your needs Low PS If the clinician has enabled the Insufficient Pressure Support alarm LOW PS will show in this menu item See
37. signs and symptoms of heart failure an objective assessment of LVEF should be performed The entire manual should be read before using the AUTOSET CS2 e Advice contained in this manual should not supersede instructions given by the prescribing physician The AUTOSET CS2 should be used with masks and accessories recommended by ResMed or the prescribing physician Use of incorrect masks and accessories may adversely affect the function of the AUTOSET CS2 The AUTOSET CS2 is designed for use with masks that allow exhaled gases to be flushed out through vent holes Exhaled gases will be rebreathed if the mask is worn with the machine turned off or the vent holes are occluded If this occurs over prolonged periods suffocation may occur In the event of power failure or machine malfunction remove the mask The air flow for breathing produced by this device can be as much as 6 C 11 F gher than the temperature of the room Caution should be exercised if the room temperature is warmer than 32 C 90 F The AUTOSET CS2 can be set to deliver pressures up to 20 cmH O In the unlikely event of certain fault conditions pressures of up to 40 cmH2O for up to 0 7 seconds are possible The AUTOSET CS2 is not suitable for use in the vicinity of flammable anaesthetics 1f oxygen is used with the AUTOSET CS2 the oxygen flow should be stopped when the device is not operating If oxygen flow continues when the device is not opera
38. ss the Start Stop key once to commence treatment page 21 NOTE You can observe treatment progress eg mask leak via the Treatment screens page 29 7 Remove your mask to stop treatment If SMARTSTOP is enabled treatment will stop automatically If the SMARTSTOP function has been disabled press the Start Stop key once QUICK SETUP GUIDE ENGLISH THE AUTOSET CS2 SYSTEM AUTOSET CS2 COMPONENTS The AUTOSET CS2 flow generator is supplied with air tubing and a power cord Masks are supplied separately as you and your clinician have to decide which mask is best for you Masks that can be used with AUTOSET CS2 are listed on page 10 Some people experience dryness of the nose mouth and throat while using the AUTOSET CS2 system especially during winter Others may experience sneezing and or a runny or blocked nose during the first few weeks of AUTOSET CS2 treatment In many cases these symptoms of nasal irritation can be resolved with a humidifier See page 14 to connect a humidifier to the AUTOSET CS2 ENGLISH LCD screen Control panel Carry handle Removable front cover Pressure sensor tube connector Air tubing connector Figure 1 AutoSet CS2 flow generator The control panel allows you to adjust settings on the flow generator and the LCD screen displays information about your treatment See Control Panel Functions on page 22 ResLink attachment point Air filter Power cord Figure 2 Rear
39. t Learn function the AUTOSET CS2 is ready for use See Starting Treatment on page 21 HumidAire 2i AutoSet CS2 HumidAire 2iC AutoSet CS2 docking station docking station HumidAire 2i water chamber HumidAire 2 water chamber Figure 8 HumidAire 2i and Humidaire 2iC attached to the AutoSet CS2 Figure 9 HumidAire 2i attached to AutoSet CS2 and mask HUMIDAIRE You will need a Medium size 52cm air tube to connect the AUTOSET CS2 unit to the HUMIDAIRE see Figure 7 ENGLISH 1 Make sure both the HUMIDAIRE and the AUTOSET CS2 are turned off Fill the HUMIDAIRE with water as described in the humidifier manual Place the filled water chamber inside the HUMIDAIRE 2 Gently detach enough of the tube clips to release the end of the pressure sensor tube from the AUTOSET CS2 air tubing Connect the AUTOSET CS2 air tubing to the left connector port on the humidifier and the medium 52cm air tubing to the right connector port Close the HUMIDAIRE lid HumidAire AutoSet CS2 air tubing Ne Jum 3 Place the AUTOSET CS2 on top of the HUMIDAIRE Do not place the AUTOSET CS2 unit underneath the humidifier This is to avoid water spilling into the unit THE AUTOSET CS2 SYSTEM 15 4 Connect the free end of the medium air tubing to the AUTOSET CS2 Connect the pressure sensor tube by twisting the Luer lock onto the AUTOSET CS2 see page 12
40. ting oxygen may accumulate within the device and create a risk of fire Do not use the AUTOSET CS2 if there are obvious external defects or unexplained changes in performance Do not open the AUTOSET CS2 case There are no user serviceable parts inside Repairs and internal servicing should only be performed by an authorised service agent 2 CAUTIONS You should report unusual chest pain severe headache or increased breathlessness to your physician An acute upper respiratory tract infection may require temporary discontinuation of treatment The following side effects may arise during a course of therapy with the AUTOSET CS2 drying of the nose mouth or throat bloating earor sinus discomfort eye irritation Skin rashes chest discomfort QUICK SETUP GUIDE 1 Attach the air tubing and pressure sensor tube to the AUTOSET CS2 and mask page ii 2 Attach a humidifier if required page 13 3 Switch on the AUTOSET CS2 Make sure the mask is unblocked 4 Select the mask setting page 26 5 Perform the Air Circuit Learn function page 26 at first time use or if you are adding or removing a component of the system eg a humidifier or anti bacterial filter 6 Fit the mask and commence breathing into it The SMARTSTART function should trigger treatment to commence within the first couple of breaths If not pre
41. typical commercial or hospital environment If the user of the AutoSet CS2 requires continued operation during power mains interruptions it is recommended that the AutoSet CS2 be powered from an uninterruptible power source Power frequency 50 60 Hz magnetic field IEC 61000 4 8 3 A m 3 Alm Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment NOTE Ut is the a c mains voltage prior to application of the test level Continued next page Guidance and manufacturer s declaration electromagnetic immunity continued The AutoSet CS2 is intended for use in the electromagnetic environment specified below The customer or the user of the AutoSet CS2 should assure that it is used in such an environment I SE i D Immunity test Inte CR ae Electromagnetic environment guidance test level level 2 Portable and mobile RF communications equipment should be used no closer to any part of the AutoSet CS2 including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter Recommended separation distance Conducted RF 3 Vrms 3 Vrms d 1 17 IEC 61000 4 6 150 kHz to 80 MHz Radiated RF 10 V m 10 V m d 0 35 NP 80 MHz to 800 MHz IEC 61000 4 3 80 MHz to 2 5 GHz d 0 70 NP 800 MHz to 2 5 GHz where P is the maximum output power rating
42. ubing is properly connected run LEARN CIRCUIT WILL STOP WHEN the pressure sensor tubing is unobstructed and properly connected to the flow generator and mask treatment is stopped Start Stop key or SMARTSTOP the AUTOSET CS2 is turned off FLOW BLOCKED ALARM TRIGGERED BY blockage in air circuit IF A FLOW BLOCKED ALARM SOUNDS The trea Remove nu M ResMed tment will stop Turn power off Check whether there is a blockage in the air circuit blockage Turn power back on If the Flow blocked alarm activates repeatedly discontinue use and return to for servicing If the alarm does not recur then continue to use as normal WILL STOP WHEN blockage is removed THE ALARMS 33 34 CLEANING AND MAINTENANCE You should regularly carry out the cleaning and maintenance described in this manual CAUTIO N Do not wash the pressure sensor tube If fluid enters the pressure sensor tube allow it to dry completely by hanging it in a clean place out of direct sunlight If the pressure sensor tube cannot be dried completely it should be replaced DAILY Disconnect the air tubing and pressure sensor tube and hang them in a clean dry place unti and crack next use Do not hang the air tubing in direct sunlight as it may harden over time 2 Clean the mask according to the mask user instructions 3 If you are WEEKLY using a humidifier clean it
43. y The AutoSet CS2 is intended for use in the elec romagnetic environment specified below The customer or the user of the AutoSet CS2 should assure that it is used in such an environment Immunity test IEC60601 1 2 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD 6 kV contact 6 kV contact Floors should be wood concrete or ceramic tile If IEC 61000 4 2 8 kV air 8 kV air floors are covered with synthetic material the relative humidity should be at least 3096 Electrical fast 2 kV for power 2 kV Mains power quality should transient burst supply lines be that of a typical IEC 61000 4 4 commercial or hospital 1 kV for input Not Applicable environment output lines Surge 1 kV differential 1 kV differential Mains power quality should IEC 61000 4 5 mode mode be that of a typical 2 kV common mode 2 kV common mode commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 lt 5 Ut gt 95 dip in Ut or 0 5 cycle 40 Ut 60 dip in Ut or 5 cycles 70 Ut 30 dip in Ut or 25 cycles lt 5 Ut gt 95 dip in Ut or 5 sec lt 12V gt 95 dip in 240V or 0 5 cycle I6V 60 dip in 240V or 5 cycles 168V 30 dip in 240V or 25 cycles lt 12V gt 95 dip in 240V or 5 sec Mains power quality should be that of a