1. GasGuard NH3 Sensor comes factory calibrated and should require only minimal adjustments after installation There are two pots on the preamp that are used for calibration see Figure 3 Note Never measure sensor output in mA Always use mVDC or VDC voltmeter settings Zero Calibration After the sensor is installed and has been powered up for a minimum of 8 hours the unit can be zero calibrated by the following Be sure the unit is in clean air e Adjust the zero pot until the sensor outputs 40 mV from Test to Test see Figure 3 Note To zero the sensor immediately after power up or cell replacement unplug cell from the transmitter and adjust signal to 40 mV Span Calibration DO NOT ADJUST THE SPAN POT WITHOUT CERTIFIED CALIBRATION GAS If span adjustment is required use the following procedure e Apply span gas at 0 8 L min span gas must be in air not nitrogen or other carrier e Sensor should react to gas within 15 seconds Once the output signal has peaked or two minutes maximum adjust the span pot until the correct output is achieved see Figure 3 With full scale span gas the calculated span value is 200 mVdc span gas sensor range 16 4 mA output Note Below are a few response characteristics which may be an indication that the gas sensor is at or near the end of its useful life If any of these are observed the cell should be replaced e Slow response to recovery from calibration ga
2. FOR A PARTICULAR PURPOSE CTI SHALL NOT BE LIABLE FOR ANY SPECIAL INDIRECT INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES INCLUDING LOSS OF DATA WHETHER ARISING FROM BREACH OF WARRANTY OR BASED ON CONTRACT TORT OR RELIANCE OR ANY OTHER THEORY 11 GG NH3 02 2013 SPECIALISTS IN Gas E
3. GASGUARD N H Ammonia Sensor OPERATING amp INSTALLATION MANUAL ALIBRAT f uH GasGuard NH3 Operating and Instruction Manual Table of Contents General description 4 4 Locating the sensor 4 Installation guidelines 5 WING ecto a a ua 6 Operation L us a kaq s ea eee eta 7 Sta UD 7 Calibration crieur inisinia 7 Maintenance au aaa iani 9 Specifications 10 Warrant ice uy 11 Calibration Technologies 866 394 5861 GasGuard NH3 Operating and Installation Manual General Description The GasGuard NH3 sensor is a 24 VDC three wire 4 20 mA sensor for ammonia which utilizes proven electro chemical sensor technology for fast and accurate leak detection The standard detection range of the GasGuard NH3 provides real time continuous monitoring of ammonia concentrations accurately down to 5 ppm without false alarms We typically recommend a 0 100 ppm range for all personnel and product protection areas Higher ranges 0 250 0 500 0 1000 are optional for higher alarm setpoint areas such as Engine Rooms The GasGuard provides an industry standard linear 4 20 mA output signal compatible with most gas detection systems and PLCs The output signal is not affected by drastic temperature and moisture variations t
4. ed high gas concentrations Alarm setpoints should not be lower than 10 of full scale range Wiring Guidelines e Always use three conductor insulated stranded shielded copper cable Do not pull sensor wiring with AC power cables This can cause electrical interference f cable runs cannot be made without a splice all splice connections should be soldered Ground the shield at the main control panel Connect the shield wire in the sensor terminal block labeled SHLD span zero Test 40 to 200 Figure 2 Wiring diagram GasGuard NH3 Operating and Instruction Manual Start up Sensor can be response tested and or span calibrated immediately after power up Allow 60 seconds for power up time delay of sensor to end green power LED will flash during power up Start Up Test 1 One person exposes each sensor to calibration gas 2 The second person stays at the control unit to determine that each sensor when exposed to the gas is connected to the proper input and responds causing appropriate alarm functions LED functions The green power LED blinks for 60 seconds during power up as the signal output is held at 4 mA The LED is then continuously lit during normal operation The red filter LED indicates EMI RFI by blinking once per second Extreme temperature changes are indicated by the LED being lit continuously During normal conditions the red LED is off Calibration The
5. hat occur during washdown defrost cycles etc The transmitter circuit board is sealed in potting compound protecting sensitive electronic components and copper tracing from corrosion The specially vented chemical resistant polycarbonate enclosure protects the sensor from accidental damage weather and direct hose hits from clean up crews Installation Locating the sensor One of the most important considerations when installing GasGuard NH3 sensors is that they must be easily accessible for calibration and maintenance For optimum personnel protection representative concentration reading that an employee would be exposed to mount the sensor at a height in the breathing zone of the employees It would typically be about five feet off the ground which also allows easy access As a general rule of thumb try to mount sensors within 30 feet of potential leak sources Caution Remove protective cap The sensor is shipped with a cap installed over the electrochemical cell to preserve cell life The cell will not detect ammonia vapors with this cap installed Remove cap and discard during installation GasGuard NH3 Operating and Instruction Manual Installation Guidelines e Remove and discard protective cell cap e Always mount the sensor vertically Must be easily accessible for calibration and 6 5 a maintenance e Mount the sensor close to the potential leak source 0 For optimum personnel protectio
6. hin the warranty period In no event shall liability hereunder exceed the purchase price actually paid by the buyer for the Product This warranty does not include a routine replacement of parts due to the normal wear and tear of the product arising from use b any product which opinion has been misused altered neglected or damaged by accident or abnormal conditions of operation handling or use c any damage or defects attributable to repair of the product by any person other than an authorized dealer or contractor or the installation of unapproved parts on the product The obligations set forth in this warranty are conditional on a proper storage installation calibration use maintenance and compliance with the product manual instructions and any other applicable recommendations of b the buyer promptly notifying CTI of any defect and if required promptly making the product available for correction No goods shall be returned to CTI until receipt by the buyer of shipping instructions from CTI and right of CTI to require that the buyer provide proof of purchase such as the original invoice bill of sale or packing slip to establish that the product is within the warranty period THE BUYER AGREES THAT THIS WARRANTY IS THE BUYER S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS
7. n mount sensor in the breathing zone 4 5 above floor Take air movement and ventilation patterns into account e To prevent electrical interference keep sensor and wire runs away from mercury vapor lights variable speed drives and radio repeaters Protect sensor from physical damage forklifts etc If mounting sensor outdoors consider prevailing 7 5 wind direction and proximity to the most likely source of leaks Protect the sensor from sun and rain as much as possible e Never mount the sensor in CA controlled atmosphere rooms because normal atmospheric MRI levels of oxygen are required for operation Inc e For highly critical locations more than one sensor should be installed in each room e Mount sensor enclosure through the mounting holes as shown in Figure 1 GasGuarp NH3 AMMONIA SENSOR 18 23 Figure 1 Mounting dimensions GasGuard NH3 Operating and Installation Manual Wiring Electrical wiring must comply with all applicable codes Electrical Power 24 VDC regulated 350 mA Output Linear 4 20 mA output Monitoring equipment may have a maximum input impedance of 700 ohms Cable Recommendation 20 3 shielded cable General Cable C2525A or equivalent Length of cable to sensor should be no greater than 1 500 feet Monitoring Monitoring equipment must be configured to indicate a fault if the signal is below 1 mA All signals over 20 mA must be consider
8. n gas accuracy and time since last calibration Zero Drift Less than 0 1 of full scale per month non cumulative Span Drift Application dependent but generally less than 3 per month Linearity 0 5 of full scale Repeatability 1 of full scale Green LED Blinks for 60 seconds during power up Stays on steady to indicate power Red LED Filter mode activated if blinking or on solid EMI RFI and extreme temperature changes trigger the filter function Wiring Connections 3 conductor shielded stranded 20 AWG cable General Cable C2525A or equivalent up to 1500 ft Enclosure Injection molded NEMA 4X polycarbonate sensor housing Captive screw in hinged lid For non classified areas Temperature Range 50 F to 140 F 46 C to 60 C Humidity Range 5 to 100 condensing Dimensions 7 7 high x 6 7 wide x 3 8 deep Weight 2 0 Ibs Limited Warranty amp Limitation of Liability Calibration Technologies Inc CTI warrants this product to be free from defects in material and workmanship under normal use and service for a period of one year gas sensor covered for six months beginning on the date of shipment to the buyer This warranty extends only to the sale of new and unused products to the original buyer warranty obligation is limited at CTI s option to refund of the purchase price repair or replacement of a defective product that is returned to a authorized service center wit
9. s e Failure of the output to reach 50 of the calibration gas value prior to span adjustment e Unable to achieve correct output during span GasGuard NH3 Operating and Installation Manual df UW SRE lnc 866 394 5861 Sensor cable plugs into socket here Green LED indicates power when lit Red LED indicates filter mode Span adjustment Zero adjustment if 40 to 200 40 200 mV ik Test if Figure 3 Sensor board components and zero span adjustments GasGuard NH3 Operating and Instruction Manual Maintenance The GasGuard NH3 was designed for long life and minimal maintenance For proper operation it is essential that the test and calibration schedule be adhered to Calibration Technologies recommends the following maintenance schedule Maintenance Guidelines e The sensor is shipped with a factory calibration Sensor should be calibrated 6 months from purchase date Calibrate the detector at least once every 6 months e Calibration should be performed with certified calibration gas Calibration kits and replacement cylinders are available from Calibration Technologies In highly critical areas a response test should be performed between calibrations to verify proper sensor response and alarm functions This can be done with calibration gas or ammonia water solution The response test is not required if multiple electrochemical sensors are installed in the same room All te
10. sts and calibrations must be logged Sensor Life These electrochemical cells are extremely reliable but several things can cause the cell chemicals to become depleted including aperiod of time exposure to high temperatures continuous long term exposure to ammonia When the cell becomes depleted the unit will give no indication of failure other than that the sensor will not respond For this reason it is absolutely essential that these sensors be calibrated on a regular basis Typical sensor life in a refrigerated area will be two to three years Typical life in a non refrigerated area will be one to two years When the cell becomes depleted a replacement cell can be obtained from Calibration Technologies Simply unplug the cell s ribbon cable from the transmitter pull the old cell from the spring clip discard the old cell and replace it with a new one The sensor can be calibrated after a 12 hour warm up period GasGuard NH3 Operating and Installation Manual Specifications Detection Principle Electrochemical Detection Method Diffusion Gases Ammonia NH3 Ranges 0 100 ppm standard 0 250 ppm 0 300 ppm 0 500 ppm 0 1000 ppm Custom ranges available Call for more information Output Signal Linear 4 20 mA max input impedance 700 Ohms Power Supply 24 VDC 350 mA Response Time Tso less than 30 seconds Too less than 60 seconds Accuracy 5 of value but dependent on calibratio