Planmeca Sovereign™ Classic

image

Contents

1. P User s Manual Planmeca Sovereign Classic 121 MAINTENANCE PARTS 122 Planmeca Sovereign Classic User s Manual 23 UNIT DISPOSAL Frame and covers metal Aluminium galvanized steel plastic PVC PUR other plastics rubber glass Component boards Cables u transformers steel UNIT DISPOSAL In order to reduce the environmental load over the products entire lifecycle Planmeca products are designed to be as safe as possible to manufacture use and dispose of Parts which can be recycled should always be taken to the appropriate processing centres after hazardous waste has been removed Disposal of obsolete units is the responsibility of the waste possessor All parts and components containing hazardous materials must be disposed in accordance with waste legislation and instructions issued by the environmental authorities The risks involved and the necessary precautions must be taken into account when handling waste products Batteries must be disposed of following the requirements of Directive 2006 66 EEC and in accordance with waste legislation and instructions issued by the environmental authorities Waste Hazardous Main materials Recyclable waste for disposal material disposal site separate collection OM m X Amalgam separator filters collectors Packing Wood cardboard _ Other parts Refer to the instructions suppl
2. The duration of the signal can be programmed see section 19 7 Door open assistant call on page 82 When the assistant call function is enabled the Assistant call button is displayed on the control panel and you can call for the assistant from the control panel or the foot control You can select whether the door open function or the assistant call function is enabled Both of them can not be enabled at the same time For instructions see section 19 7 Door open assistant call on page 82 Control panel and Flexy holder Press Assistant call to call for the assistant You will hear a short signal tone when the function starts Hold your finger on the Assistant call button for more than 0 5 seconds to continue the function until the button is released Foot control When all the instruments are in their holders you can push the pedal to the left to call for the assistant If the pedal is pushed for longer than 0 5 seconds the function continues for as long as the pedal is pushed Alternatively the foot control can be configured so that you call for the assistant when you push the left side or right side knob up Contact your Planmeca dealer The duration of the signal can be programmed see section 19 7 Door open assistant call on page 82 User s Manual OPERATING DENTAL UNIT 17 7 Planmeca SingLED operating light User s Manual Es O B e NOTE NOTE NOTE Control panel and
3. CAUTION 50 Planmeca Sovereign Classic NOTE NOTE NOTE NOTE Do not charge the foot control battery while treating a patient The battery must be charged outside the patient area The battery charging area must be dry Do not expose the charger to liquids The power supply is marked and specified as a part of Planmeca Sovereign Classic The batteries must be replaced only by a qualified Planmeca service technician The foot control batteries must be removed whenever the foot control is stored for a longer period The batteries must be removed only by a qualified Planmeca service technician The standard foot control contains a radio device FCC YII002 and IC 9050A 002 and the wireless foot control contains a radio device FCC YII001 and IC 9050A 001 See also section 24 2 FCC Class B Notice for wireless foot control on page 136 User s Manual SWITCHING UNIT ON AND OFF 14 SWITCHING UNIT ON AND OFF SC UM 11 eps O Y A ELLI NOTE NOTE NOTE User s Manual The ON OFF switch 1 is located at the base of the Planmeca Sovereign Classic cuspidor next to the suction arm Press the switch once to switch the unit on Press the switch a second time to switch it off When Planmeca Sovereign Classic is switched on the ON OFF switch light is on The control panel briefly shows the software version number If your unit is equipped with a water heater fill the cu
4. NOTE NOTE NOTE NOTE NOTE NOTE NOTE NOTE NOTE NOTE NOTE FOR YOUR SAFETY National regulations concerning the quality of dental water and dental air must be followed when using the Planmeca Sovereign Classic dental unit The air used by the Planmeca Sovereign Classic dental unit instruments must be dry clean and oil free The water used by the Planmeca Sovereign Classic dental unit instruments and cup fill are to be used for rinsing only For more information please contact your local Planmeca dealer The main water feed must be turned off when the dental unit is not in use Before switching on the dental unit make sure that the main water feed air pressure and suction motor are turned on The user must monitor the microbial load of the water used by the dental unit Biotest Plus by Alpro Medical GmbH or equivalent test can be used for this purpose If a method such as electro chemical activation is used to disinfect the dental unit water and waterlines disinfection with Planmeca PlanClear is not necessary For more information please contact your local Planmeca dealer EMC requirements have to be considered and the equipment must be installed and put into service according to the specific EMC information provided in the accompanying documents Portable and mobile RF communications equipment can affect the Planmeca Sovereign Classic dental unit The Planmeca Sovereign Classic dental unit
5. The Panasonic intraoral camera is delivered with the following manual e User s and installation manual For dental care professionals and service personnel Describes the Panasonic intraoral camera as well as instructs how to operate and clean the camera Also describes how to install the Panasonic intraoral camera to the dental unit Before using the dental unit water and waterline disinfectant Planmeca PlanClear read the disinfectant s material safety data sheet Before using the suction system disinfectant D rr Orotol Plus read the disinfectant s material safety data sheet Before using an instrument read the instrument s user s manual For a full list of accessories refer to the Planmeca product price list User s Manual TRAINING 3 TRAINING A hands on user s training is given in connection with the dental unit installation 4 PREVENTIVE MAINTENANCE To guarantee the dental units proper operation the dental unit must be checked and serviced once a year by a qualified Planmeca service technician User s Manual Planmeca Sovereign Classic 3 SYMBOLS ON PRODUCT LABELS 5 SYMBOLS ON PRODUCT LABELS Packaging symbols Standard ISO 780 and ISO 15223 1 This side up Fragile 1 2 3 4 5 6 Reep dry Temperature limitation Air pressure limitation E E E en Humidity limitation Packaging symbol The number of stacked packages is limited to 2 due to the weight of the package Type B equipment Stan
6. 12 COURE terres nee ee engen 41 1222 Altaching control panel iin 41 12 2 3 Detaching controlpanel zen ae en 42 1224 TOUCH SOGE essi 42 12 3 Control panel on Flexy holder anstehen 44 13 FOOT CONTROL iaa ie 45 Sl ZOVEINIEW ee ee ee ee eine 45 192 KN0Ds and pedal ass e a e 45 183 Wireless TOO CONTOL ee ae As 49 14 SWITCHING UNIT ON AND OFF oocccoccocconccnnnoncnnnnnnnnnnnonanonnnnnnnnnanonanananzaazanas 51 15 CHECKING SOFTWARE VERSI N Cosina 52 16 OPERATING PATIENT CHAIR susana ad 53 16 1 A A oone ennaa 53 16 2 LAUNOMAlIG OD CT AUON asien nee 55 16 3 Trendelenburg POSION siria os 56 16 4 AMS O DOS MO RN sa ndo 56 17 OPERATING DENTAL UNIT 2 2 2 220 2200 2000020000n00nn0nan0n nun nun nnnnnnnnnnnnnnnn nun 57 17 1 Language zes i ea cietat 57 17 2 BOWN SC Teenie 58 o COCO ets tec ee EB ee her ee 58 174 Ale ee een 59 17 3 DOORLODEN ers eier 59 17 0 AS SIGUA CUM ac see ee ee Ne ee Me 60 17 7 Planmeca SingLED operating light u02220022000000000 nenne nnnnn nenne nnnnn nenne nennen 61 18 OPERATING INSTRUMENTS 2 2 2220220022000 n00nn0nnn0nnnnnnnnnnnunnnnnnnnnnnnnnunnnnn 62 18 1 AASTUNMENCIO rene nee ee een are 62 18 2 NICLOMOLOR man ae E nai enementy tees 63 1853 MU Be ee Beeren 66 18 4 Scaler winner a ee 70 189 IROIVIMECLIS CU ON ION rs da dote 72 18 07 Mirasa camera na eier eis miele 73 19 PROGRAMMING nata eiii 74 191 IAKOQUELON ces mes see oceans oe
7. 5 Assistant call or alternatively Door open User s Manual 02 04 2014 CP4 Display idle eps 6 Chair buttons 7 OP light brightness display only 8 Wireless foot control battery display only 9 Unit buttons 10 Program button Planmeca Sovereign Classic 43 CONTROL PANEL 12 3 Control panel on Flexy holder The control panel on the Flexy holder can be used for controlling the dental unit and the chair When you press the Flexy button the chair will be driven to the rinsing position For more information see section 16 4 Rinsing position on page 56 flexy_panel eps 1 Chair buttons 2 Unit buttons 3 Flexy button 44 Planmeca Sovereign Classic User s Manual 13 FOOT CONTROL 13 1 Overview NOTE CAUTION CAUTION 13 2 Knobs and pedal User s Manual FOOT CONTROL The Planmeca Sovereign Classic dental unit has one integrated foot control that operates the instruments the unit and the chair The foot control is available as a wireless and a standard version All the foot control functions except instrument drive and manual chip blow can be performed on the control panel too The foot control is a precision instrument Do not stand on or apply unnecessary force to the foot control and its knobs Do not use the foot control in areas where liquids are likely to be present on the floor _3 pS NewFootctrl_3 e 1 Handle 3 Centre knob 2 Left side knob 4 Ri
8. Flexy holder Press Operating light to switch the operating light on or off Foot control Push the left side knob up to switch the operating light on off This is the default factory setting The intensity of the operating light can be programmed see section 19 6 Intensity of operating light on page 81 The intensity can also be adjusted by holding down the Operating light button or keeping your hand in front of the infrared light sensor of the operating light for longer than one second After one second the intensity will increase until you release the button switch When you press and hold the button switch again the intensity will start to decrease The intensity starts to decrease increase after the maximum minimum value has been reached When the limit value is reached you will hear a short signal tone The operating light intensity will decrease slightly when the chair is moving When the polymerisation light is taken from the holder and returned to it within 2 seconds the operating light will dim This feature is configurable contact your Planmeca dealer You can change the operation mode of the light by pressing the on off switch for more than 15 seconds you will hear two signal tones one after 3 and one after 15 seconds 1 The operating light can be switched on off and the brightness can be adjusted from both the light and the dental unit 2 Brightness adjustment can be done only from the operating light
9. If this equipment does cause harmful interference to radio or television reception which can be determined by turning the equipment off and on the user iS encouraged to try to correct the interference by one or more of the following measures e Reorient or relocate the receiving antenna e Increase the separation between the equipment and receiver e Connect the equipment into an outlet on a circuit different from that to which the receiver is connected e Consult the dealer or an experienced radio television technician for help Modifications Any modifications made to this device that are not approved by Planmeca may void the authority granted to the user by the FCC to operate this equipment 136 Planmeca Sovereign Classic User s Manual PLANMECA Planmeca Oy Asentajankatu 6 00880 Helsinki Finland tel 358 20 7795 500 fax 358 20 7795 555 sales planmeca com www planmeca com PLANMECA Oy s PLANMECA Oy s Environmen tal Quality System Management System Certificate awarded in 1994 Certificate avvarded in 2000 y Det Norske Veritas by Det e Veritas by Del ki fulfils the highest acknowledged s i cknowledged international standards international standards
10. Unit controlled polymerisa tion light 48 Planmeca Sovereign Classic User s Manual FOOT CONTROL 13 3 Wireless foot control NOTE The wireless foot control is an optional feature The standard foot control and the wireless foot control can not be used simultaneously Should such a situation occur the standard foot control overrides the wireless foot control When the wireless foot control has been idle for 30 minutes default value it goes into sleep mode You can wake the foot control by pressing the foot control handle You can also put the wireless foot control into sleep mode by pressing the foot control handle Before using the wireless foot control check the power level of its battery The power level is indicated by a battery symbol on the control panel NOTE The battery symbol can be hidden from the control panel Contact your Planmeca dealer Sleep mode lt 40 full lt 20 full When the wireless foot control battery is running low you need to recharge the battery To charge the foot control battery connect the foot control to the power outlet using the provided cable and power adapter The LEDs on the foot control blink green while the battery is being charged When the wireless foot control battery is full and the foot control is connected to the power outlet the LEDs on the foot control are a steady green User s Manual Planmeca Sovereign Classic 49 FOOT CONTROL CAUTION CAUTION
11. instrument is not recognised Software upgrade may be needed This instrument is not recognised Connect only supported instruments HE2003 Unsupported assistant instrument This instrument is not recognised Connect only supported instruments HE3015 HE3020 HE3021 Collision risk Upward chair movement prevented Downward chair movement prevented The distance from the backrest to the cuspidor or bowl is too small Move the bowl to rest position or move the chair away from the cus pidor The chair cannot be driven up because a safety switch is activated Make sure that nothing blocks the movement The chair cannot be driven down because a safety switch is activated Make sure that nothing blocks the movement HE3030 Seat sensor is not moving The seat sensor is not moving with the desired speed HE3031 HE3040 Backrest sensor is not moving Armrests are svvivelled out The backrest sensor is not moving with the desired speed Chair movements are prevented Turn the armrests to the rest position HE3050 Chair durability test active Chair movements are used with timed control HE3051 HE4004 Upvvard chair movement limited VVater lea sensor active The bowl position prevents chair upward movements Return the bowl to home posi tion Water leak detected The water container is not filled but instruments can be used as long as there is water in the co
12. number 10032652 Revision 2 Released 04 April 2014 iv Planmeca Sovereign Classic User s Manual 1 INTRODUCTION NOTE NOTE NOTE NOTE CE INTRODUCTION The Planmeca Sovereign Classic dental unit is an electrically controlled dental device that consists of a patient chair cuspidor delivery arm dental instruments and an operating light The Planmeca Sovereign Classic dental unit is meant to be used for dental treatment by dental care professionals This manual describes the Planmeca Sovereign Classic dental unit and how to use it Depending on the configuration of your dental unit this manual may contain parts that do not apply to your dental unit Please read this manual carefully before using the unit The use of the Planmeca Sovereign Classic dental unit is allowed only under the supervision of the dental care professional This manual is valid for Sovereign Classic software version 1 0 1 or newer In error situations this manual is the primary source of information For information on OEM products please refer to OEM documentation The Planmeca Sovereign Classic dental unit fulfills the requirements of Directive 93 42 EEC Classification according to European Council Directive 93 42 EEC Class IIA The settings and values shown in this guide are only examples and should not be interpreted as recommended values unless otherwise stated 2 ASSOCIATED DOCUMENTATION User s Manual Planmeca So
13. of these buttons again To start the polymerisation cycle with the foot control first select the desired curing mode by pressing the navigation button on the handpiece Then push the foot control pedal to the right or to the left or down to start the polymerisation cycle The polymerisation cycle can be interrupted by pushing the foot control pedal to the right or to the left or down The length of the cycle is programmable see section 19 3 5 Planmeca Lumion polymerisation light on page 78 When the polymerisation light is activated the length of the polymerisation cycle is displayed on the control panel When you start the cycle you will hear a signal tone This signal tone is repeated every 10 seconds and also at 5 seconds The progress of the polymerisation cycle is displayed on the control panel User s Manual 18 6 Intraoral camera User s Manual NOTE OPERATING INSTRUMENTS Independent mode The Planmeca Lumion polymerisation light can also be operated in the independent mode The independent mode is typically used when the Planmeca Lumion polymerisation light is attached to the assistant element but the independent mode can also be configured to be used on the dentist side the configuration is performed by a Planmeca service technician When the Planmeca Lumion polymerisation light is in the independent mode it can only be controlled by the buttons on the instrument itself not over the control pan
14. openings in the flushing holder DCI syringe Remove the instrument cover and use an insert before placing the syringe in the flushing holder Lock the syringe into place with a rotating movement when the syringe is in the holder Luzzani syringe The metallic syringe cover must be removed before placing the syringe into the flushing holder Note the orientation of the syringe It does not require an insert and must be placed in the holder as shown below ps SC UM 52 e Planmeca Sovereign Classic 103 MAINTENANCE AND CLEANING 20 5 Cleaning instruments 20 6 Cleaning dental unit surfaces NOTE 104 Planmeca Sovereign Classic 6 Balanced instrument arms Start the extensive flushing by bending the instrument arms of all water consuming instruments to an angle of at least 90 and simultaneously selecting Extensive flushing on the control panel Hanging tube instruments Start the extensive flushing by selecting Extensive flushing on the control panel dy Release the instrument arms if your dental unit is equipped with balanced instrument arms 8 Follow the messages on the control panel that inform you of the cleaning cycle progress 9 Once the flushing cycle is finished remove the instruments from the flushing holder and place them in the instrument console Close the cuspidor door The dental unit is now ready for normal operation Clean and service the instruments according to the information suppli
15. pedal can be released and then pushed again and the time continues to increase Continue until the cup is filled to the desired level Press Program 1 Press Program 2 Press Operating light A pop up window opens 3 Change the intensity of the operating light with the plus and minus buttons The minimum value is 70 of the maximum intensity and the maximum 100 and the adjustment step is 2 4 Press OK The pop up window closes and the new value is displayed on the control panel The light intensity can also be adjusted as described in section 17 7 Planmeca SingLED operating light on page 61 Planmeca Sovereign Classic 61 PROGRAMMING 19 7 Door open assistant call 82 Planmeca Sovereign Classic 1 NOTE Press Program Press Door open or Assistant call depending on which function is enabled A pop up window opens Program Door open ON Nurse call N Door open Open time 5s E A grey button in the pop up window means that the function is disabled and a blue button that the function is enabled Optionally enable the function that is currently disabled by pressing the grey button of the function you wish to enable The button turns blue For example instead of being able to open the door from the control panel you want to call for the assistant In the programming pop up window press the grey Assistant call button The button turns blue which means that the a
16. shall only be connected to a trusted private network and not for example the Internet Never place heavy objects or containers of liquid on any part of the unit or hang objects from the units arm structures Electromagnetic interference between the equipment and other devices can occur in very extreme conditions Do not use the equipment in close conjunction with sensitive devices or devices creating high electromagnetic disturbances Care should be taken when other movable equipment is used in conjunction with the Sovereign Classic dental unit External equipment intended for connection to signal input signal output or other connectors shall comply with relevant IEC standard e g IEC 60950 for IT equipment and the IEC 60601 series for medical electrical equipment In addition all such combinations systems shall comply with the IEC 60601 1 standard Equipment not complying to IEC 60601 1 shall be kept outside the patient area see figure Any person who connects external equipment to signal input signal output or other connectors has formed a system and is therefore responsible for the system to comply with the requirements of IEC 60601 1 If in doubt contact a qualified technician or your local representative Planmeca Sovereign Classic 7 FOR YOUR SAFETY CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION 8 Planmeca Sovereign Cla
17. shown in the figure 1 Attach the lightning dock to the instrument console by rotating the lightning dock arm counterclockwise when the lightning dock is placed on the right side of the console or clockwise when the lightning dock is placed on the left side of the console until you hear a click 2 Planmeca Sovereign Classic 23 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 8 2 Detaching lightning dock La ps SC_UM_51 e 7 9 Planmeca ProX X ray unit NOTE CAUTION CAUTION CAUTION CAUTION 24 Planmeca Sovereign Classic To make the detachment easier turn the iPad screen outward see figure Press the locking button located underneath the lightning dock arm 1 and turn the arm approximately 30 clockwise 2 when the lightning dock is placed on the right side of the console or counterclockwise when the lightning dock is placed on the left side of the console Detach the lightning dock by pulling the arm out from the instrument console 3 The Planmeca ProX X ray unit can be mounted to the Planmeca Sovereign Classic dental unit pylon The Planmeca ProX X ray unit is an option and is not available in all market areas Drive the chair carefully when positioning the ProX X ray unit Do not drive the chair during the exposure Do not touch the external PC and the patient at the same time Move the ProX X ray unit behind the pylon when it is not being used For more information about Planmeca P
18. that inform you of the cleaning cycle progress NES Flushing water lines Time remaining 1 40 The system identifies the instruments and fills all waterlines with Planmeca PlanClear disinfection solution Planmeca Sovereign Classic 101 MAINTENANCE AND CLEANING NOTE 10 11 12 UNN F UN NS If you have not inserted all water consuming instruments in the instrument flushing holder help message HE5000 is displayed HE5000 Required instruments are not selected when starting instrument flushing or waterline cleaning Insert required instruments into instrument flushing holder Shut down the dental unit for the night Also turn off the main water feed for the night Planmeca PlanClear disinfection solution must not be left in the dental unit waterlines for longer than one night not for example over the weekend In the morning turn the main water feed and the unit back on Waterline flushing starts automatically When the procedure is completed a message is displayed on the control panel When the cleaning program is completed help message HE5003 is displayed on the control panel Remove the instruments from the flushing holder and place them in the instrument console Close the cuspidor door HE5003 Waterline cleaning program completed Return instruments to the instrument console Test if the water is clean For instructions see section 20 4 3 Af
19. the cleaning programs the progress of the cleaning cycle is displayed on the control panel The instrument and suction tube positions on the control panel correspond to the their positions on the instrument console and in the Flexy holder A selected item is displayed in blue an item that is being cleaned is displayed in a blinking green and an item that has already been cleaned is displayed in a steady green lf the cleaning procedure for some reason is interrupted or not successful for example there is no water flow the item is displayed in yellow To enter the Maintenance mode press the Maintenance button Select the cleaning program in the window that appears D Maintenance ll N Short flushing NS CS N Long flushing Be Suction cleaning o NES Waterline cleaning P S 0 6 P While performing the flushing cleaning follow the instructions displayed on the control panel User s Manual MAINTENANCE AND CLEANING The follovving table lists buttons that are common for the cleaning programs Displays additional information about the cleaning program when you press the button during the cleaning cycle Closes the pop up window Minimises the pop up window Cancels the selected action Indicates that the touch screen is unlocked Press to lock Useful for example when cleaning the touch screen or when using a table top instrument Indicates that the touch screen is locked Pre
20. the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The type of chip blow can be programmed see section 19 3 3 Chip blow on page 77 Standard pedal You can temporarily activate the manual chip blow by pushing and holding down the pedal Wide pedal You can temporarily activate the manual chip blow by pushing and holding the pedal to the right The flow of air will continue until you remove your foot from the pedal The type of manual chip blow does not depend on the type of the automatic chip blow but is always dry The manual chip blow can be set to be dependent on programming contact your Planmeca dealer Press Fibre optic light to switch the turbine s fibre optic light on off When the fibre optic light is on the indicator light is green Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The fibre optic light intensity can be programmed see section 19 3 4 Fibre optic light LED light on page 78 Planmeca Sovereign Classic 69 OPERATING INSTRUMENTS 18 4 Scaler NOTE CAUTION Speed power NOTE CAUTION Instrument spray 70 Planmeca Sovereign Classic A scaler requires additional electronics and the scaler type can not be changed without changing electronics Do not use the scaler on patients with cardiac pacemakers The scaler can cause dis
21. 000 mm x 2700 mm H x D x W User s Manual Planmeca Sovereign Classic 131 TECHNICAL SPECIFICATIONS Side view CA i Ss m RA N m co al a A e LO a O Si M Q lo gt N O O I CO ol 218 Si N DO O N CO lt O N CIS n 2 AN e O v o D de O N c SIG S O 5 ae I E 5 Slo y 0 Sys 132 Planmeca Sovereign Classic N li NN N J E A Py A J Max Horizontal 980 7 1050 41 3 Se Min Entry 385 15 2 1940 76 4 User s Manual TECHNICAL SPECIFICATIONS Top view chair centerline chair centerline for right hand unit for left hand unit 364 14 3 a i 2 RA 240 chair swivel i turning of suction arm O DIES O XY O 7 ee zZ NI Y A a 55 5 xu DD Ga gt N left right turning left right turning of unit 120 of unit 120 NU 550 21 7 640 25 2 centerline for left right outermost unit conversion cuspidor position User s Manual Planmeca Sovereign Classic 133 TECHNICAL SPECIFICATIONS chair centerline chair centerline for right hand unit for left hand unit 364 14 3 left right turning of 240 chair re PR vo Sue l i installed to Planmeca Sovereign E installation site u I left right turning a of unit 120
22. 1 1 ed 3 requirements Use only Planmeca specified devices inside the patient area The floor of the patient area must be dry User s Manual TECHNICAL SPECIFICATIONS Inside patient area Outside patient area 1 Dental unit 7 Planmeca ProSensor PoE port and power cable 2 Planmeca monitor 8 Planmeca ProX generator assembly 3 Planmeca ProX X ray tube head and arm assembly 9 Foot control battery charger 4 Planmeca SingLED operating light 10 External PC 5 Foot control Use only IEC 60601 1 approved power source supplied by Planmeca 6 Planmeca ProSensor control box User s Manual Planmeca Sovereign Classic 129 TECHNICAL SPECIFICATIONS Cuspidor positions Cuspidor at 4 o clock Cuspidor at 6 o clock Common nght handed care situation Common left handed care situation Cuspidor at 2 o clock Cuspidor at 10 o clock Maximum swivel nght Maximum swivel left 130 Planmeca Sovereign Classic User s Manual TECHNICAL SPECIFICATIONS Optimal space requirements 500 REFERENCE POINT AO 59 1 YES qua REFERENCE POINT Centre point for unit swivel r Chair centerline S Dee ge left handed position Dee mirror line for Erg TUNING un 1390 REFERENCE POINT 2180 Chair centerline for right handed position Centre point for chair swivel 500 Minimum space requirements 2200 mm x 2
23. ANCE AND CLEANING SC_UM_060 eps 20 4 7 Waterline cleaning Introduction Cleaning interval 98 Planmeca Sovereign Classic 6 When the cleaning program is completed help message HE5006 is displayed on the control panel Return the suction tubes to the Flexy holder and close the cuspidor door Suction cleaning program completed Return suction tubes to the suction arm The dental unit is now ready for normal operation Before attaching the handpieces back on the suction tubes wipe the identification bushings 1 with Durr FD 333 FD 322 The dental units water system is fully isolated from the inlet waterline according to the requirements of EN1717 standard and the specific regulations of the German water authorities Waterlines are a breeding ground for biofilm Biofilm may contain bacteria that are harmful to the patient and to the dental team This is why the waterlines of the dental unit should be cleaned regularly with a waterline disinfectant The cleaning interval depends on whether continuous cleaning is enabled or not e Continuous cleaning enabled Perform waterline cleaning once a month after the working day but not over the weekena e Continuous cleaning disabled Perform waterline cleaning once a week after the working day but not over the weekend The disinfectant is left in the unit overnight and the waterlines of the unit are flushed with water the next morning The disinfec
24. ANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 7 Unit swivel The Planmeca Sovereign Classic dental unit can be swivelled manually 120 o P A gt N 120 i 120 Y 7 gt N s SC UM 36 3 eps Pull the lever on the unit base to unlocR the base and swivel the dental unit to the desired position Lock the base by pushing the handle back in AL o For an overview of how the cuspidor can be positioned see section Cuspidor positions on page 130 For detailed measurements on the unit swivel area see section Top view on page 133 Also the chair can be swivelled see section 11 3 Chair swivel on page 38 22 Planmeca Sovereign Classic User s Manual 7 8 iPad lightning dock CAUTION 7 8 1 Attaching lightning dock es ps SC_UM_50 e User s Manual NOTE PLANMECA SOVEREIGN CLASSIC DENTAL UNIT A lightning dock for your iPad where you can charge its battery can be connected to either side of the instrument console The lightning dock is only to be used together with the iPad so if you remove the iPad always remove the lightning dock too The iPad may break if it is dropped on the floor The lightning dock supports a limited range of iPad models For details please contact your local Planmeca dealer Push the lightning dock arm into the attachment opening in a position where the arm is rotated approx 30 forward as
25. NOTE Fibre optic light NOTE User s Manual OPERATING INSTRUMENTS The amount of water and the spray water number 1 or 2 are displayed on the control panel The amount of water that runs through the scaler can be set separately for water 1 and 2 see section 19 3 2 Instrument spray on page 77 The LM scaler has three operation modes low 0 40 medium 0 70 and full O 100 When the scaler is the active instrument the operation mode can be changed by pressing Power Check the operation mode symbol on the control panel to verify that pushing the left side knob down activates the correct function The EMS No Pain scaler has three operation modes e endodontics power range 1 50 e scaling power range 1 100 e restoration power range 50 100 The power range for all three operation modes is displayed as 1 100 for example in the restoration mode power range 50 is displayed as 1 When the scaler is the active instrument the operation mode can be changed by pressing Power Check the operation mode symbol on the control panel to verify that pushing the left side knob down activates the correct function Press Fibre optic light to switch the scaler s fibre optic light on off When the fibre optic light is on the indicator light is green Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The fibre optic ligh
26. NTENANCE AND CLEANING The oil collector underneath the instrument console has to be checked weekly and emptied when full You can check how full the oil collector is without removing it completely Pull the oil collector downward from the attachment knob 1 and check the oil level in the narrower chamber 2 If it has started to fill it is time to empty the oil collector N S215 eps To remove the oil collector from the instrument console unscrew the oil collector attachment knob and carefully pull the collector out User s Manual Planmeca Sovereign Classic 107 MAINTENANCE AND CLEANING 20 8 Cleaning cuspidor 20 8 1 Bowl a 108 Planmeca Sovereign Classic NOTE NOTE The glass bowl should be cleaned after every patient and after the working day Clean the bowl as follows 1 2 Clean the bowl with a soft brush Rinse the bowl by pressing the Bow rinse button on the control panel or by using the foot control The outside of the bowl can be wiped clean or disinfected with a cloth If required the bowl can be removed for rinsing under tap water or washing at 65 C 1 2 Remove the bowl filter and the cover cap of the filter Turn the bowl out from the cuspidor and push the bowl fastening clip carefully down Rotate the bowl counterclockwise and slightly upward You can now remove the bowl by pulling it horizontally away from the cuspidor Empty and clean the bowl filter after
27. PLANMECA Z Planmeca Sovereign Classic z J users manual TABLE OF CONTENTS 9 a A QO N 10 11 INTRODUCTION zu een 1 ASSOCIATED DOCUMENTATION zu2u022002000n00nu0nnunnunnn nun nun nun n nun nunnnunnunnnn 1 TRAINING 2 0 nenne 3 PREVENTIVE MAINTENANCE u uuue iin 3 SYMBOLS ON PRODUCT LABELS usina nia 4 FOR YOUR SAFETY sis 7 6 1 SAC SWIICHES dee aora 11 6 2 Chai MOVEMENT FESITICIONS uti sk 12 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 22 22 224 200220002002 13 7 1 Dental UNIECOMIQUIA ION Srta ls 13 7 1 1 Over the patient delivery with balanced instrument arms oocccccnccncccncnnccncnncnnncnnnnnons 13 7 1 2 Over the patient delivery with hanging tube instruments cccccoccnccnccnnccnconccnnnnnnnnnos 14 e Detachable pans o le 15 7 3 APPDICA PANS tt ote 18 le OWN Geiucin soa aades cee ale tiiames tast 19 HO MONO ito 20 6 USB COANCCIVINY astas ara aos 21 EF IMEI il das 22 AS IRAN OOCK an eek 23 7 8 1 Attaching onning OOCR une 23 7 8 2 DETACHING onmino Jack un nee a tbbaneenebieasttuaie 24 L9 Planmeca PrOX X Tay Ulla 24 7 10 Zeiss OPMI pico MICFOSCOPE ccccoccccncnccncnccnccnonncnnnnonnnnonnnnnnnnonnnnnnnnnnnnnnnnnnnnnrnnnnnnnnaninnnnes 25 OPERATING Lina ia 26 INSTRUMENT SYSTEM cin cms 27 9 1 Over the patient OP delivery arm ccooccccccnccncncconccnonocnnnnnnnnonanonncnonnnnonnnonnnnonaronancnnns 27 92 Instrument Console uni da 28 9 2 1 Balance
28. Program to activate the programming mode The button turns blue Select the desired function from the control panel A pop up window opens where you can adjust the values with the plus and minus buttons Press OK to store the new setting into memory The pop up window closes To check a function setting other than automatic position without programming it press Program and the button of the desired function The setting is displayed on the control panel Press OK to close the window without changing the setting When you press Program you will also see the following symbols on the control panel service mode software update wireless foot control if installed These are reserved for service situations only 74 Planmeca Sovereign Classic User s Manual 19 2 Automatic chair positions NOTE Ye O 9 0 3 w A Y x 4 PROGRAMMING Move the chair to the required position by using the chair movement buttons on the control panel or by using the foot control The chair height cannot be programmed near the upper limit The help code HE 08 is displayed on the control panel if the chair position is not allowed When needed the upper limit can be altered Contact your Planmeca dealer Press Program Press the desired chair position where to store the chair s position rinsing position A B C or D If you want the operating light to be on off in this position turn it on off P
29. The light can be switched on off from both the operating light and the dental unit Planmeca Sovereign Classic 61 OPERATING INSTRUMENTS 18 OPERATING INSTRUMENTS 18 1 Instrument logic 62 Planmeca Sovereign Classic NOTE Especially in four handed dentistry it is important to enable smooth cleaning and exchange of instruments between the dentist and the assistant The Planmeca Sovereign Classic dental unit contains an instrument logic with the principles described below The active instrument can be operated with the foot control and its settings are displayed on the control panel and they can be changed 1 The last instrument picked up from the instrument console can be driven with the foot control pedal a short push to the right left or down is enough After activating one instrument the others can be picked up from the console without changing the active instrument Only after the active instrument has been returned to the console another instrument can be activated by picking it up from the instrument console After this it can be driven with the foot control as above Other instruments can be away from the console but they can not be driven The memory operation of the logic settings makes it possible to store parameters and settings for up to eight instruments Changing the place of the instrument on the console does not change the parameters and settings of that instrument The instrument logic does not contro
30. X uu22420022020 nenn nnnnn nenn nnnnnnnnnnn nenne neuen nnnnnennnnnnnnnne nennen 113 20 11 Cleaning Planmeca ProSensor Genis herein 114 20 12 Cleaning extemal PO a ee re aio 114 20 13 Cleaning Zeiss OPM DICO rasen ea 114 21 HELP amp ERROR MESSAGES u 115 2 lc ACCT CVG se eier ee 115 21 2 Hep Messages u idos 116 PAO ENORIMESS ADOS an iiden 118 22 MAINTENANCE PARTS iniciara ia 121 239 UNI DISPOSAL uni rd 123 24 TECHNICAL SPECIFICATIONS 22 222022200 2000a000nnnnnanunnnnnnanunnnnnnnnnnnnen 124 24 1 DIMENSIONS sail i 127 User s Manual Planmeca Sovereign Classic iii TABLE OF CONTENTS 242 FCC Class B Notice for wireless foot control oo occcocococcccocococccconoconannonononananonnnnanono 136 The manufacturer assembler and importer are responsible for the safety reliability and performance of the unit only if installation calibration modification and repairs are carried out by qualified and authorized personnel electrical installations are carried out according to the appropriate requirements such as IEC 60364 equipment is used according to the operating instructions Planmeca pursues a policy of continual product development Although every effort is made to produce up to date product documentation this publication should not be regarded as an infallible guide to current specifications We reserve the right to make changes without prior notice COPYRIGHT PLANMECA Publication
31. a Sovereign Classic User s Manual PATIENT CHAIR 11 5 Headrest 11 5 1 Adjusting headrest height The height of the headrest can be adjusted by sliding it NOTE The headrest can only be pulled out as far as to the MAX mark 11 5 2 Adjusting headrest angle To adjust the angle of the headrest press the bar on the side of the headrest support to release the locking mechanism Manually set the headrest to the required angle and release the bar When adjusting the headrest should be supported by hand User s Manual Planmeca Sovereign Classic 39 PATIENT CHAIR 11 5 3 Adjusting headrest for children or short patients 40 Planmeca Sovereign Classic The headrest can be turned around and repositioned for better head support for children or short patients Pull the headrest out and turn it around so that the cushion faces backwards Push the headrest back into the chair Turn the cushion around 180 counterclockwise Press the bar on the side of the headrest support to release the locking mechanism and position the headrest at the top of the chair The headrest is now repositioned To adjust the angle of the headrest press the locking bar Manually set the headrest into the required position and release the bar When adjusting support the headrest with your other hand User s Manual 12 CONTROL PANEL 12 1 Two control panels CONTROL PANEL The Planmeca Sovereign Classic dental unit has two control p
32. aches the desired position release the centre knob You can steer the chair only in one direction at a time 54 Planmeca Sovereign Classic User s Manual OPERATING PATIENT CHAIR 16 2 Automatic operation e SC_UM_17 eps User s Manual Automatic chair positions may be stored into memory Press briefly the chair position button where the required position for the selected user is stored The chair will move automatically to the preprogrammed position To stop the chair from moving either press any of the chair positioning buttons push the foot control pedal in any direction or push the foot control centre knob in any direction Alternatively you can use the foot control to drive the chair to the automatic positions Push the centre knob briefly to the position A B C or D where the preprogrammed chair position is stored The chair will move automatically to the preprogrammed position To stop the chair from moving before it reaches the preprogrammed position push the foot control pedal or the centre knob in any direction The chair movement stops also when the stop plate 2 is pushed or when the backrest 1 is pressed upward The chair can be driven normally after the possible obstruction has been removed When the chair has reached its preprogrammed position the position can be adjusted To drive
33. age between the instrument and connector causes leakage air to drift into the hose lining A scaler requires additional electronics and the scaler type can not be changed without changing electronics User s Manual INSTRUMENT SYSTEM 9 4 Tray tables 9 4 1 Tray table for balanced arm instrument console The tray table is attached to the mounting arm with a magnetic connector and can easily be attached and detached You can rotate the table 360 to the desired position The maximum weight limit on the tray is 2 kilograms 4 4 Ibs The tray mounting arm is attached to the instrument console with a quick connector The tray assembly can be removed from the instrument console as follows Pull the N ring of the locking mechanism outwards 1 and pull the N tray arm away from its position 2 0 SC_UM_23 ep The mounting arm can be attached to the instrument console by pushing it to its position ps SC UM 24 e User s Manual Planmeca Sovereign Classic 31 INSTRUMENT SYSTEM 9 4 2 Tray table for hanging tube instrument console 32 Planmeca Sovereign Classic S203 2 eps S202 2 eps The tray mounting arm is attached to the instrument console with a quick connector enabling an easy attachment and detachment of the tray table You can rotate the table 360 to the desired position The maximum w
34. ainer level sensor error Orotol container level sensor error Water container level sensor error Main air pressure too low Main air pressure sensor error VS A water level Suction system error User s Manual Planmeca Sovereign Classic 119 HELP amp ERROR MESSAGES Foot control EXPLANATION Foot control calibration error Foot control calibration error Control panel EXPLANATION Control panel connection error 120 Planmeca Sovereign Classic User s Manual MAINTENANCE PARTS 22 MAINTENANCE PARTS The user can change certain detachable parts of the dental unit a Or mate Matas Saliva suction nozzle 1 00221015 PU ABS Suction tip Universal Cannula Protect D rr 5pcs 10034007 PBT 2 Suction tip Prophylaxis Cannula D rr 4 pcs 3 10034009 Saliva suction nozzle Cattani 4 10020902 Polypropylene homopolymere 25 mineral filled Saliva suction handpiece D rr 5 101007949 High volume suction adapter Cattani 6 00006936 Polypropylene homopolymere 25 mineral filled High volume suction adapter D rr 7 10011158 PP D rr coarse filter 0725 041 00 1 piece 8 00221013 D rr coarse filter 0725 041 00 12 pcs 10005741 PP NOTE 2 filters needed Filters retain solid particles with a diameter of 2 2 mm Handle of Planmeca SingLED operating light 1 10020763 PSU piece 9 NOTE 2 handles needed Cover cap for bowl filter 16 10005746
35. ally by pushing the pedal briefly down If the pedal is pushed for longer than 1 second water flows for as long as the pedal is pushed and the bowl is not rinsed The cup fill is not activated unless the cup is positioned in its place under the cup fill tube Alternatively an automatic cup fill feature can be configured When the cup fill is automatic the cup will fill automatically when you place the cup underneath the tap To configure this feature please contact a qualified Planmeca service technician User s Manual OPERATING DENTAL UNIT NOTE 17 4 Timer NOTE NOTE 17 5 Door open NOTE User s Manual The flow rate of the cup fill can be adjusted by turning the black knob inside the unit see section 19 5 1 Adjusting cup fill and bowl rinse flow rates on page 79 The duration of the cup fill can be programmed see section 19 5 3 Duration of cup filling on page 80 The water in the dental unit is intended for rinsing only not for drinking The timer memory contains six preprogrammed time settings Follow the steps below to activate a timer 1 Press Timerto open a list of available timers 2 Select a timer from the list The duration of the timer is displayed on the control panel The countdown begins immediately but the selected time is displayed for five seconds before the actual countdown is displayed If you wish to close the list without selecting a timer press Close W
36. anel can be used to drive the chair When driving the chair near the upper limit make sure that the console arm does not press or hit the patient When driving the backrest up make sure that the patient s hand or arm does not get squeezed between the armrest and the bachrest To drive the backrest down hold your finger on the Backrest down button until the chair reaches the required position To drive the backrest up hold your finger on the Backrest up button until the chair reaches the required position To drive the chair up hold your finger on the Chair up button until the chair reaches the required position Make sure that the bowl is not above the patient chair when driving the chair upward Planmeca Sovereign Classic 53 OPERATING PATIENT CHAIR To drive the chair down hold your finger on the Chair down button until the chair reaches the required position NOTE In case the chair is equipped with the chair mounted left right suction arm make sure the Flexy holder is not above the cuspidor when driving the chair down If the chair does not move downward and HE 3021 is displayed on the control panel check that the chair mounted arm is not in the upmost position This arm position prevents the chair from moving downward Alternatively you can adjust the position of the chair with the foot control Push and hold the centre knob in the desired direction see picture below When the chair re
37. anels The main control panel is located on the instrument console and the other one on the Flexy holder 12 2 Control panel on instrument console 12 2 1 Overview CAUTION 12 2 2 Attaching control panel ps SC UM 9 e User s Manual The control panel on the instrument console features a touch screen and it can be used to control and program the instruments the dental unit and the chair Also maintenance procedures can be started from the control panel The control panel is attached by the quick locks to the instrument console The control panel may break if it is dropped on the floor To make the attachment easier turn the control panel away from the attachment arm Push the attachment arm into the attachment opening in a position where the attachment arm is rotated approx 30 forward as shown in the figure 1 Attach the control panel to the instrument console by rotating the attachment arm counterclockwise when the control panel is placed on the right side of the console or clockwise when the control panel is placed on the left side of the console until you hear a click 2 CAUTION Ensure that the dental unit works correctly after attaching the control panel If needed restart the dental unit Planmeca Sovereign Classic 41 CONTROL PANEL 12 2 3 Detaching control panel ps SC UM 10 e 12 2 4 Touch screen 42 Planmeca Sovereign Classic To make the detachment easier turn the contro
38. art the waterline cleaning see section Starting cleaning cycle on page 100 NOTE The Continuous cleaning section lets you enable disable continuous cleaning see section 20 3 Continuous cleaning on page 87 If you want to close the window without resetting the counter press Cancel Planmeca Sovereign Classic 99 MAINTENANCE AND CLEANING Starting cleaning cycle NOTE NOTE x NOTE NOTE CAUTION CAUTION A 100 Planmeca Sovereign Classic Adjust the water flow of the cup fill tube so that the water does not splash into the bowl The duration of waterline cleaning is lt 10minutes in the evening an affect time of min 8 hours lt 30 minutes in the morning The flushing cycle can be interrupted by pressing Cancel After the interruption you must flush the waterlines before the dental unit can be used normally Immediately remove any Planmeca PlanClear splashes from the dental unit Only Planmeca PlanClear disinfectant must be used Planmeca does not guarantee the suitability of and is not liable for damages caused by other disinfectants The flushing cycle is performed using domestic water which might reduce the effect of Planmeca PlanClear if the domestic water is not clean WARNING Even though every effort has been taken to ensure patient safety even in case of malfunction or misuse always make sure that the unit is properly flushed before taking it into use For instruction
39. ations the dental unit may also restrict the water flow to prevent leakage O YA ps q I eps X N SC UM 53 e FA 20 10 Cleaning Planmeca ProX For information on how to clean the Planmeca ProX X ray unit see the Planmeca ProX user s manual User s Manual Planmeca Sovereign Classic 113 MAINTENANCE AND CLEANING 20 11 Cleaning Planmeca ProSensor 20 12 Cleaning external PC CAUTION 20 13 Cleaning Zeiss OPMI pico 114 Planmeca Sovereign Classic For information on how to clean the Planmeca ProSensor sensor see the Planmeca ProSensor user s manual The external PC together with its mouse and keyboard can be wiped with a dry cloth or according to the manufacturer s instructions When cleaning the external PC always disconnect the PC from the mains electricity supply For cleaning instructions refer to the manufacturers documentation User s Manual HELP amp ERROR MESSAGES 21 HELP amp ERROR MESSAGES 21 1 Overview NOTE User s Manual The dental unit displays three types of safety messages notifications help messages and error messages Notifications are typically issued for guidance in maintenance situations for example when instruments are flushed Notifications are blue in colour Help messages are green and provide helpful information for the user They are displayed for example if you are using the unit or instrument incorrectly or if the fun
40. cleaning programs as follows Short instrument After every patient Duration flushing 30 seconds instrument ANN AB Long instrument In the morning and after the Duration flushing working day 9 minutes default can be programmed ANN F UN WNN Suction tube clean After the working day Duration ing lt 5 minutes 88 Planmeca Sovereign Classic User s Manual MAINTENANCE AND CLEANING Continuous Once a month Cleans the dental units vvater cleaning after the work lines Waterline cleaning Extensive flushing 20 4 2 Before you start User s Manual enabled Continuous Once a week cleaning after the work disabled ing day but ing day but The amount of disinfectant not over the concentrate used in a cleaning weekend cycle depends on whether con tinuous cleaning is enabled or disabled For more information see Water and waterline dis infection PlanClear on page 126 not over the weekend Duration lt 10 minutes in the evening min 8 hours affect time lt 30 minutes in the morning When prompted by the dental Extensive flushing rinses unit NOTE NOTE NOTE NOTE NOTE NOTE NOTE excess disinfection solution from the waterlines The pro gram should be used for example when waterline cleaning has been interrupted because of a power failure or when turning off continuous cleaning The dental unit tells you when you need
41. ction is not allowed for some reason The help message disappears automatically when the situation is corrected Error messages warn of a fault in the dental unit and require actions from the user To correct the error situation and thus to close the error message follow the instructions in the message The level of detail displayed in the error messages is configurable Contact your Planmeca dealer When a help or error message is issued it is displayed as a semi transparent bar at the top of the control panel NE N i 7 ki ee E Fi 2 Tr perm a Dee ww D AS 12 12 26 28 02 2014 Planmeca Sovereign Classic 115 HELP amp ERROR MESSAGES 21 2 Help messages HELP CODE Help messages provide helpful information for the user They are displayed for example if you are using the unit or instrument incorrectly or if the function is not allowed for some reason The following table lists the help messages MEANING ACTION OR EXPLANATION HE1000 Error messages disabled The unit operates normally but no errors are reported exhibition use Use service mode 9100 to turn error reporting back on See ser vice mode 9100 in chapter Service mode Technical Manual for details HE2000 Wrong scaler module installed This scaler does not have the correct module installed Install the correct scaler module HE2001 HE2002 Unsupported instrument Unsupported assistant syringe This
42. d arm instrument console 0222002000 2000 0nn0 Benno nnne nenn nenne nenn nenne anne nnnenennennn 28 9 2 2 Hanging tube instrument console zu220s0200nunnnnnnnnnnnnnnnnnnnnennnnennnnennnnnnnnnnnnnnnennene 29 9 3 QUICK CONNEC ROSES aii 30 94 WAV tables seele 31 9 4 1 Tray table for balanced arm instrument console u02400204020n0 onen nnne nenn ada 31 9 4 2 Tray table for hanging tube instrument console uuu uuus40ssnnenne nenne nenne nnnne nennen 32 9 5 014 8 02 9 ame is 33 SUC HON SYSTEM comidas 34 10 1 SUCTION MANGDICCES a ita sien 34 10 2 Suction arm with Flexy holder oocccoocccocccococcnnocococoncnnonoroconnnnonarnnnannnonanononnnnnnnnos 34 PATIENT CHAIR 2 22 2 nee 37 142 General aus asien NO Bates 37 11 2 Trendelenburg DOSINON eiserne 37 PES Chair SWYE Si bc 38 114 AMES an 38 UES AAA o a een Er 39 11 5 1 Adjusting headrest height z2s0220202200002n0nnonnnonnnn nenne neuen neuen nnnnnnnnnnnnnnne nennen 39 1152 Adjusting headrest angle Aa ik 39 Users Manual Planmeca Sovereign Classic i TABLE OF CONTENTS 11 5 3 Adjusting headrest for children or short patients oocccocccccncccnnoconocononocnnnanonononons 40 12 CONTROL PANEL sn nun 41 12 1 IWoConlrol Panels cetacis bee sales ee ea ei Heuss 41 12 2 Control panel on instrument console 02220022000220 00000 nano nenn nenn nenne nenne nennen 41
43. dard IEC 60878 Type B applied parts for specified dental instruments Standard IEC 60878 Type BF applied part for specified dental instruments Standard IEC 60878 Alternating current Standard IEC 60878 On off switch Attention consult accompanying documents Standard ISO 7010 General warning Standard ISO 7010 Warning crushing hazard hand Standard IEC 60878 Warning hot surface Standard ISO 7010 4 Planmeca Sovereign Classic User s Manual User s Manual Planmeca PlanClear A SYMBOLS ON PRODUCT LABELS To avoid risk of electric shock this equipment must only be connected to a supply mains with protective earth Chemical hazard Irritant as defined by Directive 67 548 EEC Planmeca PlanClear is a disinfectant for the water and waterlines Only put Planmeca PlanClear in the container The container must be filled with Orotol Plus which is a disinfectant for the suction tubes and suction lines Protected against dripping water Standard IEC 60529 Disposable item Do not reuse Standard ISO 7000 Separate collection for electrical and electronic equipment according to Directive 2002 96 EC WEEE Date of manufacture Standard IEC 60878 Protective earth Standard IEC 60878 Radio certification Japanese Radio Law certification symbol Indicates the direction of rotation for increasing decreasing the water flow for the bowl rinse and cup fill Planmeca Sovereig
44. e chair allowing the dentist and the assistant to move their legs freely while sitting down Integrating the chair with the unit allows the chair to be raised high for a good working position When you drive the backrest down the seat follows the movements of the backrest and tilts accordingly This ensures that the patients head stays in the correct position throughout the movement Make sure no one sits on the legrest or the backrest The Trendelenburg position is used in the event of a patient collapsing The chair is driven to a position where the patient s feet are higher than the head For information on how to drive the chair to the Trendelenburg position see section 16 3 Trendelenburg position on page 56 Planmeca Sovereign Classic 37 PATIENT CHAIR 11 3 Chair swivel The patient chair can be swivelled manually Press either of the buttons on the chair underside marked with arrows in picture to release the lock and turn the chair to the desired position u ps SC_UM 27 e You can swivel the chair within the movement limits shown in the picture The chairs home position is shown with solid black outlines ps SC UM 28 2 e 11 4 Armrests Both armrests can be moved 90 outwards Before you move the armrest horizontally you must lift it slightly to unlock it The armrest can be locked into the positions presented in the picture SC_UM_34 eps 38 Planmec
45. e must be placed in the second instrument position from the right Labels mark the positions of the syringe and the USB device The other instruments can be positioned according to your own preferences Both the control panel and the tray table can be right or left mounted LS ps SC UM 22 e The instrument holders can be removed by pulling them out from their openings for example for cleaning The holders are replaced simply by pushing them firmly into their places The angle of the instrument holder can be slightly adjusted a AN instrument console you can make the adjustment of the A instrument holders more rigid By tightening the screws D By slightly tightening the screws 1 2 under the IN A SA A a IS completely you can lock the angle of the instrument SIG RS holder Note that you have to tighten both screws equally otherwise you may no longer be able to place the O instrument holders to the console UL Similarly you can untighten the screws to unlock the UU Uy angle or loosen the adjustment of the instrument holders User s Manual Planmeca Sovereign Classic 29 INSTRUMENT SYSTEM 9 3 Quick connector hoses sov 009 30 Planmeca Sovereign Classic sov_008 NOTE NOTE NOTE NOTE The instruments are equipped with quick connect
46. each patient and replace it when necessary The cover cap of the filter makes the removal of the filter easier but the filter can also be used without the cover cap Do not empty the bowl filter to the drain User s Manual 20 8 2 Water container ps S SC UM 43 e KO Nes ZA U 3 N AS N CAUTION 20 8 3 Instrument flushing holder ps SC UM 31 e User s Manual MAINTENANCE AND CLEANING If continuous cleaning is disabled wash the water container in a washer disinfector once a week at 93 C Otherwise wash the water container when needed Lift the water container upward to remove it from the cuspidor Remove the cap including the hose attached to it from the container before placing them in the washer disinfector Be careful not to splash the contents of the water container when removing it as it may contain traces of Planmeca PlanClear disinfectant Immediately remove any Planmeca PlanClear splashes from the dental unit When you place the water container back in the cuspidor make sure to place the three adjacent openings in the container carefully over the three pipes in the cuspidor Also make sure you place the cap back properly so that it is tightly closed Clean the instrument flushing holder once a month in a wash
47. eca Sovereign Classic dental unit is that the syringe can be right or left mounted A USB device must be placed in the second instrument position from the right Labels mark the positions of the syringe and the USB device The other instruments can be positioned according to your own preferences The control panel can be right or left mounted The instrument arms can be removed by pulling them out from their holders for example for cleaning The arms are replaced simply by pushing them firmly into their places When placing the instrument hose to the roller bend the hook of the hose guide carefully and pass the hose over the roller The balance of the instrument arms can be changed according to the weight of the instrument and personal preferences Their flexibility can be adjusted as follows 1 Pull out the roller 2 Adjust the balance of the instrument arm by moving the roller to the desired position Note that by positioning the roller higher up the arm is lighter to bend 3 Push in the roller to lock its position When balancing adjusting the instrument arms bear in mind that the instruments shall under no circumstances fall over the patient User s Manual INSTRUMENT SYSTEM 9 2 2 Hanging tube instrument console The console can be equipped with up to six dynamic instruments A special feature of the Sovereign Classic dental unit is that the syringe can be right or left mounted A USB devic
48. ed with the instrument Intraoral camera Clean the intraoral camera according to the information supplied with the instrument Throw away the disposable hygiene sleeve after use All exposed surfaces should be periodically wiped clean with a non abrasive cleaning agent designed for use with dental units To avoid damage the splashed cleaning solution must be wiped immediately from the surfaces of the dental unit User s Manual MAINTENANCE AND CLEANING The following tables describe how to clean the dental unit surfaces NOTE The parts must be washed before autoclaving them Alternative cleaning method Cleaning solution Washer Autoci v disinfector 134 C 93 C After every Headrest metal D rr FD 333 patient and after parts only D rr FD 322 the working day Armrests metal D rr FD 333 parts only D rr FD 322 Protective fascia of D rr FD 333 operating light D rr FD 322 Handle ofoperating D rr FD 333 X X light D rr FD 322 Patient mirror of Mild soap and operating light water solution Instrument console D rr FD 333 D rr FD 322 Hygienic mem D rr FD 333 X X brane D rr FD 322 Instrument hoses D rr FD 333 D rr FD 322 Balanced instru D rr FD 333 ment arms D rr FD 322 Hanging tube D rr FD 333 X X instrument holders D rr FD 322 Control panel D rr FD 333 D rr FD 322 Cup fill tube D rr FD 333 D rr FD 322 Bowl Mild soap and X water solution Cus
49. eight limit on the tray is 2 kilograms 4 4 Ibs The quick connectors make it easy to for example change from left handed to right handed operation To detach the tray table press the locking button located underneath the tray mounting arm 1 and turn the tray mounting arm approximately 30 clockwise 2 when the tray table is placed on the left side of the console or counterclockwise when the tray table is placed on the right side of the console Detach the tray table by pulling the arm out from the instrument console 3 To attach the tray table push the tray mounting arm into the attachment opening in a position where the tray mounting arm is rotated approx 30 backward as shown in the figure 1 Attach the tray mounting arm to the instrument console by rotating the tray mounting arm counterclockwise when the tray table is placed on the left side of the console or clockwise when the tray table is placed on the right side of the console until you hear a click 2 User s Manual 9 5 instruments User s Manual INSTRUMENT SYSTEM The instrument console has six instrument positions The following instruments are available for Planmeca Sovereign Classic e Syringe e Turbine e Micromotor e Scaler e Polymerisation light e Intraoral camera The syringe can be placed in the leftmost or rightmost position The syringe can be used simultaneously with other instruments Otherwise there are no restr
50. el or the foot control To start the polymerisation cycle in the independent mode first select the desired curing mode by pressing the navigation button on the handpiece Then briefly press the start button on the handpiece to start a 10 second exposure or press and hold the start button for 2 seconds to start a 20 second exposure The polymerisation cycle can be interrupted by pressing the start button In the 10 second cycle a signal tone is heard at the beginning and the end of the cycle In the 20 second cycle a signal tone is heard as the cycle starts after 2 seconds to confirm the start of the 20 second cycle after 10 seconds and at the end of the cycle When the Planmeca Lumion polymerisation light is in the independent mode it can not be controlled over the foot control Refer to the Planmeca Lumion manual of use and maintenance The intraoral camera can only be operated from the buttons on the handpiece For detailed information see the user s and installation manual for the Panasonic intraoral camera Planmeca Sovereign Classic 73 PROGRAMMING 19 PROGRAMMING 19 1 Introduction Most of the programming follows the same pattern NOTE NOTE NOTE If you are programming an instrument pick up the instrument from the instrument console If the instrument is operated during programming the changes in settings are seen immediately The air water flow of the syringe can not be programmed Press
51. er disinfector at 93 C or autoclave it at 134 C To remove the instrument flushing holder follow the steps below 1 Open the cuspidor door 2 Lift up the instrument flushing holder from the cuspidor Planmeca Sovereign Classic 109 MAINTENANCE AND CLEANING 20 8 4 Coarse filters ps Empty the coarse filters daily or when they are full Clean or replace them monthly SC UM 41 e NOTE The coarse filter should be emptied into a separate amalgam container 20 8 5 Suction tube cleaning holder Clean the suction tube cleaning holder in a washer disinfector once a month at 93 C To remove the suction tube cleaning holder follow the steps below 1 Open the cuspidor door 2 Pull out the suction tube cleaning holder The plug can be detached from the cleaning holder prior to washing but you can just as well wash the cleaning holder with the plug still attached ps SC_UM_32 e NOTE When you place the suction tube cleaning holder back in the cuspidor after washing make sure it is pushed firmly into its position in the cuspidor 110 Planmeca Sovereign Classic User s Manual 20 9 Cleaning suction system NOTE MAINTENANCE AND CLEANING Dry the suction tube holders and the tube bushings properly after cleaning Wet surfaces might disturb suction tube recognition 20 9 1 Inthe morning 20 9 2 Aft
52. er each patient NOTE 20 9 3 After the working day SC UM 059 eps User s Manual Rinse the suction tubes with 1 litre of water by aspirating water and air through the suction handpieces Wipe the suction handpieces with D rr FD 333 FD 322 Remove the used aspirating tips For hygienic and operational reasons empty one glass of water 100 200 ml with each suction tube by aspirating water and air through the suction handpiece Do this even if only the saliva suction tube has been used Clean the suction handpieces and wipe them with D rr FD 333 FD 322 or autoclave them Wash or disinfect the Flexy holder with D rr FD 333 FD 322 Do not use a spray disinfection solution for the suction arm Disinfect the suction system by running Suction cleaning see section When the suction system has been rinsed autoclave the suction handpieces Flexy holder Disinfect the supplementary holders suction tube holders including rollers and instrument holder in a washer disinfector at 93 C Planmeca Sovereign Classic 111 MAINTENANCE AND CLEANING 20 9 4 Weekly cleaning procedures Ww s 20 9 5 Cleaning suction handpieces 112 Planmeca Sovereign Classic If your dental unit is equipped with a D rr amalgam separator or D rr VS A separator open the bowl valve by pressing the VS A button in the Maintenance main menu The suction starts and the valve closes automatically after about 15 seconds or
53. f all water consuming instruments to an angle of at least 90 and simultaneously selecting Long flushing on the control panel Hanging tube instruments Start the flushing cycle by selecting the Long flushing on the control panel Release the instrument arms if your dental unit is equipped with balanced instrument arms Follow the messages on the control panel that inform you of the cleaning cycle progress The system flushes the water container the instruments the cup fill tube and bowl The total flushing time is displayed on the control panel ER Flushing container Time remaining 1 40 Planmeca Sovereign Classic 95 MAINTENANCE AND CLEANING 20 4 6 Suction cleaning NOTE NOTE 96 Planmeca Sovereign Classic If you have not inserted all water consuming instruments in the instrument flushing holder help message HE5000 is displayed Required instruments are not selected when starting instrument flushing or waterline cleaning Insert required instruments into instrument flushing holder 9 When the flushing cycle is completed help message HE5001 is displayed on the control panel Remove the instruments from the flushing holder and place them in the instrument console Close the cuspidor door Instrument flushing completed Return instruments to the instrument console The dental unit is now ready for normal operation Insert a plug in the empty holder s in the cleaning holder
54. fection cleaning orotolr plus suction unit disinfectant 126 Planmeca Sovereign Classic User s Manual TECHNICAL SPECIFICATIONS 24 1 Dimensions Positioning of patient dentist and assistant The following picture shows an example of how to position the dentist 1 and assistant 2 during treatment The patient must always be positioned in the patient chair with arms and legs resting on the upholstery The dentist and assistant can move within the areas marked in the picture below NOTE The following picture is an example only and presents on0e possible scenario The actual positioning of dentist and assistant depends on the used working method treatment situation region etc and can therefore not be explicitly stated in this manual User s Manual Planmeca Sovereign Classic 127 TECHNICAL SPECIFICATIONS Patient area NOTE NOTE NOTE CAUTION CAUTION 128 Planmeca Sovereign Classic The patient area is 1 5 m 59 17 in each direction from the dental unit The external PC its Reyboard and mouse as vvell as Planmeca ProSensors power supply and the Planmeca ProX generator assembly must be located outside the patient area The dentist assistant and patient must not touch the equipment outside the patient area during treatment Connect only Planmeca specified devices to the dental unit The external PC must be protectively earthed and IEC 60950 approved CE marked The monitor must fulfill IEC 6060
55. ff and to adjust the intensity For more information see section 17 7 Planmeca SingLED operating light on page 61 Planmeca SingLED can also be switched on off by pressing the on off switch at the back of the operating light Do not allow the patient to grab the Planmeca SingLED operating light or its arm when getting seated or getting up from the patient chair User s Manual INSTRUMENT SYSTEM 9 INSTRUMENT SYSTEM 9 1 Over the patient OP delivery arm The OP delivery arm is attached to the top of the dental unit and swings over the chair CAUTION Do not lean on the OP delivery arm CAUTION Do not allow the patient to grab the OP delivery arm when getting seated or up from the patient chair The instruments can be positioned using the handles on the instrument console The rotation area of the delivery arm instrument console and the control panel are shown in the illustrations below The following presents the OP delivery arm with a balanced arm instrument console The following presents the OP delivery arm with a hanging tube instrument console SC UM 58 eps User s Manual Planmeca Sovereign Classic 27 INSTRUMENT SYSTEM 9 2 Instrument console 9 2 1 Balanced arm instrument console 3C UM 21 eps Adj D eps 28 Planmeca Sovereign Classic o isp o gt o 7 NOTE The console can be equipped with up to six dynamic instruments A special feature of the Planm
56. g instrument flushing and test the water again If the hydrogen peroxide concentration is still not lt 250 ppm perform long instrument flushing once more Instrument flushing for both the dentists and the assistant s instruments are started at the same time The duration of short instrument flushing is 30 seconds instrument The flushing cycle can be interrupted by pressing Cancel After the interruption the dental unit can be used normally 1 Turn the cup fill tube so that it is above the bowl and not above the cup holder 2 Press Maintenance to go into maintenance mode 3 Remove the instrument handpieces from the instruments and clean them according to the manufacturer s instructions 4 Open the cuspidor door 5 Place the instruments into the openings in the flushing holder DCI syringe Remove the instrument cover and use an insert before placing the syringe in the flushing holder Lock the syringe into place with a rotating movement when the syringe is in the holder Luzzani syringe The metallic syringe cover must be removed before placing the syringe into the flushing holder Note the orientation of the syringe It does not require an insert and must be placed in the holder as shown below _ ps SC UM 52 e User s Manual MAINTENANCE AND CLEANING 6 Balanced instrument arms Start the flushing cycle by bending the instrument arms of all water consuming instruments to an angle of at least 90 a
57. ght side knob 5 Pedal Planmeca Sovereign Classic 45 FOOT CONTROL 46 Planmeca Sovereign Classic The factory default functions of the foot control when the dental unit is equipped with a normal headrest a standard pedal and no instrument is activated are presented below long push N ALS M Chair up Backrest down A NA a El X Backrest up w aN NO y Chair down ON a operation ESA on off Toggle USB camera on off NE NE Spit position Nurse A J I an En tay ana The functions that are activated by pushing the left side or right side knob up are configurable The function for left side knob down can be configured separately for a selection of instruments and the function depends on the selected instrument Note that for the left side knob down function you can only define one function per instrument position Contact your Planmeca dealer User s Manual _04 pS SC UM 54 e FOOT CONTROL Tvvo types of pedal Tvvo foot control pedals are available the standard pedal 1 and a slightly wider pedal 2 that functions like a gas pedal when you drive the instrument That is the further down you push the pedal the higher the instrument speed is The functional differences between the standard pedal and the wide pedal only apply to the micromotor turbine and scaler and are described in the table below Increase instrument speed Push pedal to left right Push pedal down Change instr
58. hen the timer reaches zero you will hear two signal tones The function can be cancelled by pressing Timer again The preset timers can be reprogrammed see section 19 4 Timer settings on page 79 Other functionalities can be used while the timer is running The timer can not be run from the foot control When the door open function is enabled the Door open button is displayed on the control panel and you can open the door from the control panel or the foot control You can select whether the door open function or the assistant call function is enabled Both of them can not be enabled at the same time For instructions see section 19 7 Door open assistant call on page 82 Control panel Press Door open to open the door You will hear a short signal tone when the function starts Hold your finger on the Door open button for more than 0 5 seconds to continue the function until the button is released Planmeca Sovereign Classic 59 OPERATING DENTAL UNIT 17 6 Assistant call 60 Planmeca Sovereign Classic TIP NOTE TIP Foot control When all the instruments are in their holders you can push the pedal to the left to open the door If the pedal is pushed for longer than 0 5 seconds the function continues for as long as the pedal is pushed Alternatively the foot control can be configured so that the door is opened when you push the left side or right side knob up Contact your Planmeca dealer
59. ic 63 OPERATING INSTRUMENTS Speed power reduction Instrument spray NOTE NOTE NOTE NOTE 100 64 Planmeca Sovereign Classic Press Instrument speed power reduction to reduce the micromotors speed or power to a preset level If the preset value is for example 50 the foot control pedal movement controls the speed power between 10 50 instead of the normal range of 10 100 The preset maximum value is displayed on the control panel when the micromotor is picked up from the instrument console Press Instrument speed power reduction again to switch the power reduction off When the micromotor s speed or power is reduced the indicator light is green Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The low speed function can be selected in case you need to reduce the starting speed power level The function is not available for brushless micromotors Contact your Planmeca dealer to enable disable the low speed function of the unit The instrument speed power reduction level can be programmed See section 19 3 1 Instrument speed power reduction on page 76 Note that the speed power level depends on the instrument The instrument cooling spray setting can be selected to be water amp air air or no spray The amounts of water and air for the currently used spray type are displayed on the control panel Press Instru
60. ically by pressing Bowl rinse cup fill again The flow rate of the bowl rinse can be adjusted by turning the black knob inside the unit see section 19 5 1 Adjusting cup fill and bowl rinse flow rates on page 79 The duration of bowl rinsing can be programmed see section 19 5 2 Duration of bowl rinsing on page 80 Control panel Press Cup fill and the unit will automatically fill the cup and then rinse the bowl Cup filling can be stopped before it stops automatically by pressing Cup fill again If Cup fill is pressed and held for longer than 1 second water flows for as long as the button is pressed and the bowl is not rinsed Flexy holder Press Bow rinse cup filltwice or press the button once until the cup filling starts 0 5 1 sec and the unit will automatically fill the cup and then rinse the bowl Cup filling can be stopped before it stops automatically by pressing Bowl rinse cup fill again If Bowl rinse cup fillis pressed and held for longer than 1 second water flows for as long as the button is pressed and the bowl is not rinsed Foot control When all the instruments are in their holders you can fill the cup either by briefly pushing the pedal down twice or by pushing down the pedal once until the filling starts 0 5 1 sec The unit will automatically fill the cup and then rinse the bowl Cup filling stops automatically after a preset time Cup filling can be stopped before it stops automatic
61. ictions concerning the placing of the instruments on the console The control panel will give you detailed guidance while using the instruments Planmeca Sovereign Classic 33 SUCTION SYSTEM 10 SUCTION SYSTEM 10 1 Suction handpieces When a saliva or high volume suction handpiece is lifted from its holder the suction will automatically start When the handpieces are returned the suction will stop When you are using the suction handpiece the suction can be controlled by sliding the adjuster up or down T CIS es CSS E Z J E Y Y Y Y Y Y Y w Y Nu a 7 ANA 10 2 Suction arm with Flexy holder The suction handpieces are placed in the Flexy holder The Flexy holder is a suction holder with integrated control panels that enable you to control selected functions of the dental unit For more information on the control panel see section 12 3 Control panel on Flexy holder on page 44 The Flexy holder is attached to the suction arm which in its turn is attached to the cuspidor The Flexy holder has three openings In one you place the assistants syringe and the suction handpieces are placed in the two remaining openings In addition one or two supplementary holders can be placed on either side of the Flexy holder The supplementary holder can be equipped with a USB intraoral camera or a polymerisation light The picture shows the default positions for the suction handp
62. ieces and syringe To change the positions please contact a qualified Planmeca service technician High volume suction handpiece Saliva suction handpiece Syringe 34 Planmeca Sovereign Classic User s Manual User s Manual SUCTION SYSTEM The suction tube instrument and supplementary holders can be removed from the Flexy holder for example for cleaning Remove the suction tube holder by pulling it downward from the Flexy holder To replace it push it firmly into its place Remove the roller from the suction tube holder by pulling it away from the holder To replace it push it firmly into its place Planmeca Sovereign Classic 35 SUCTION SYSTEM Remove the instrument holder by squeezing it from the bottom and at the same time lifting it upward To replace it squeeze the holder and insert it into its place Remove the supplementary holder by pulling it out from the Flexy holder To replace it push it firmly into its place For cleaning instructions see section 20 9 6 D rr CAS1 deposit cup on page 113 36 Planmeca Sovereign Classic User s Manual 11 PATIENT CHAIR 11 1 General CAUTION 11 2 Trendelenburg position EJ al en SC UM 8 2 eps UN User s Manual VR DAN PATIENT CHAIR The Planmeca Sovereign Classic patient chair is integrated with the floor mounted unit which gives plenty of room under th
63. ied by the manufacturer User s Manual Planmeca Sovereign Classic 123 TECHNICAL SPECIFICATIONS 24 TECHNICAL SPECIFICATIONS Original manufacturer PLANMECA Oy Asentajankatu 6 00880 Helsinki FINLAND phone 35 20 77950 500 fax 358 20 7795 555 www planmeca com Colours Painted parts RAL 9016 Upholstery colours Please consult your dealer for availability Mechanical dimensions Weight Installed H x D x VV 1930 mm x 1175 mm x 1930 mm see template for details 300 kg 661 lbs Environmental conditions Unit operating conditions Temperatures 15 C to 35 C 59 F to 95 F Relative humidity Non condensing humidity 5 RH to 95 RH Air pressure 800 hPa to 1020 hPa 12 psi to 15 psi Altitude lt 2000 m less than 1 25 miles Storage conditions Temperatures 5 C to 60 C 23 F to 140 F Relative humidity Non condensing humidity 5 RH to 95 RH Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi If the unit has been stored at temperatures below 10 C 50 F for more than few hours time must be allowed for the unit to reach the room temperature in its own package before connecting the dental unit to mains voltage Transportation conditions Temperatures 20 C to 60 C 4 F to 140 F Relative humidity Non condensing humidity 5 RH to 95 RH Air pressure 700 hPa to 1060 hPa 10 psi to 15 psi Mains voltage and frequency Mains voltage setting 100V 240V Fuse rat
64. in backflow prevention system which protects the water used in the instrument from contamination A power cut will shut down the software controlled backflow prevention system lf you are using a turbine without a built in backflow prevention system contaminated water could be released in the patient s mouth in the event of a power cul User s Manual Speed power CAUTION Quickstart Speed power reduction User s Manual NOTE NOTE NOTE NOTE NOTE NOTE OPERATING INSTRUMENTS Standard pedal To drive the turbine push the foot control pedal to the right or to the left Wide pedal To drive the turbine push the foot control pedal down Pushing the pedal further to the right left or down will increase the speed or power of the turbine As you push the pedal the power output is displayed on the control panel The normal range is 5 100 To stop the turbine allow the pedal to return to the rest position The speed power level depends on the instrument Ensure that the pedal is in rest position when you activate the turbine The turbine air motor will start with maximum speed if the quickstart is enabled When the instrument is active but not operated press Quickstart to enable disable the quickstart When quickstart is enabled the indicator light is green Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function Press Ins
65. iner is almost full Amalgam container is full Activation of suction tubes is prohibited Replace or empty the container as soon as possible Replace or empty the container immediately The suction system cannot be used before the container is emptied or replaced Return suction tubes to holders HE4058 Use of suction tubes is prohibited The suction tubes contain Orotol Plus Empty the suction tubes by selecting Suction clean ing from the Maintenance page HE5000 HE5001 Required instruments are not selected vvhen starting instrument flushing or vvater line cleaning Instrument flushing completed Insert required instruments into instrument flushing holder Return instruments to the instrument console HE5002 Instrument flushing has been interrupted Restart the instrument flush or return instru ments to the instrument console HE5003 HE5004 VVaterline cleaning program completed VVaterline cleaning program has been inter rupted Return instruments to the instrument console VVaterlines must be flushed before the instru ments can be used Press continue to perform flushing HE5005 Required suction tubes are not selected when starting suction cleaning Insert the required suction tubes into the suc tion tube cleaning holder HE5006 Suction cleaning program completed Return suction tubes to the suction arm HE5007 Suction cleaning program has bee
66. ing amp type FF 10A 500V 5mm x 20mm type 0001 1014 Schurter Mains frequency 50 60 Hz 124 Planmeca Sovereign Classic User s Manual Povver consumption Idle unit Typical average Maximum Maximum load of optional MSO Electrical classification IP classification TECHNICAL SPECIFICATIONS 100 VV unit not in use OP light turned on 150 W during patient treatment 1000 W 135 kg patient both chair motors are running and chair and backrest are driven upwards 300 VA Class I type B chair B instruments BF intraoral camera IPX1 protected against falling water foot control Operation of chair lift and backrest motors Intermittent operation ED 6 25 sec ON 400 sec OFF Maximum lifting capacity of chair excluding weight of unit Water supply Pressure range Flow rate Quality Connection Air supply Pressure range Flow rate Air quality Connection Suction connection Vacuum Flow rate Connection Drain connection Capacity Connection Foot control power supply Type User s Manual 135 kg 297 6 lbs min 180 kPa 26 psi max 900 kPa 130 psi lt 4 1 min maximum consumption at any instance hardness lt 8 dH 1 dH 20mg Ca 3 litre water 1 4 min 550 kPa 80 psi max 900 kPa 130 psi gt 55 litres minute maximum consumption at any instance medical grade dry and oil free 1 4 150 170 mbar 550 800 I min 50 46
67. ing the minimum value the light is switched off Press OK The pop up window closes and the new values are displayed on the control panel 19 3 5 Planmeca Lumion polymerisation light NOTE The duration of the polymerisation cycle can be programmed only for the Planmeca Lumion polymerisation light 2 78 Planmeca Sovereign Classic 1 Press Program Press Polymerisation cycle A pop up window opens Change the length of the polymerisation cycle with the plus and minus buttons The value range is 5 100 seconds The adjustment step is 5 sec and the default value is 10 sec Press OK The pop up window closes and the new values are displayed on the control panel User s Manual 19 4 Timer settings 19 5 Bowl rinse and cup fill PROGRAMMING Up to six timer settings can be programmed into the timer memory 1 Press Program 2 Press Timer A pop up window opens 3 Press the timer that you want to adjust A new pop up window opens 4 Change the length of the timer with the plus and minus buttons The minimum value is 5 sec and the maximum 20 min The adjustment step is 5 sec 5 Press OK The pop up window closes 19 5 1 Adjusting cup fill and bowl rinse flow rates SC_UM_46 eps User s Manual The flow rates of the bowl rinse and cup fill can be adjusted with the two black knobs located inside the dental unit When you have adjusted the flow rates you might need
68. ith the plus and minus buttons Scaler Adjust the flow rates for water 1 and water 2 with the plus and minus buttons The minimum value of all parameters is O no flow and maximum 100 The adjustment step is 1 in the value range O 25 and 5 in the value range 25 100 5 Press OK The pop up window closes and the new values are displayed on the control panel 19 3 3 Chip blow 1 Activate the instrument 2 Press Program Press Chip blow A pop up window opens 4 Adjust the flow rates for water and air with the plus and minus buttons The minimum value of both parameters is O no flow and maximum 100 The adjustment step is 1 in the value range O 25 and 5 in the value range 25 100 5 Press OK The pop up window closes and the new values are displayed on the control panel Le NOTE The type of manual chip blow does not depend on the type of the automatic chip blow but is dry as a factory default The manual chip blow can be set to be dependent on programming contact your Planmeca dealer User s Manual Planmeca Sovereign Classic 77 PROGRAMMING 19 3 4 Fibre optic light LED light 2 co Activate the instrument Press Program Press Fibre optic light A pop up window opens Change the light intensity with the plus and minus buttons The minimum value is 70 of the maximum intensity and the maximum 100 and the adjustment step is 2 After reach
69. l panel Contact your Planmeca dealer Press Chip blow to switch the automatic chip blow on or off When the indicator light is green the automatic chip blow is on and two short blows of water air or both will occur after the micromotor is stopped Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The type of chip blow can be programmed see section 19 3 3 Chip blow on page 77 Standard pedal You can temporarily activate the manual chip blow by pushing and holding down the pedal Wide pedal You can temporarily activate the manual chip blow by pushing and holding the pedal to the right The flow of air will continue until you remove your foot from the pedal The type of manual chip blow does not depend on the type of the automatic chip blow but is always dry Planmeca Sovereign Classic 65 OPERATING INSTRUMENTS Fibre optic light NOTE 18 3 Turbine CAUTION 66 Planmeca Sovereign Classic Press Fibre optic light to switch the micromotor s fibre optic light on off When the fibre optic light is on the indicator light is green Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The fibre optic light intensity can be programmed see section 19 3 4 Fibre optic light LED light on page 78 100 The turbines supplied by Planmeca have a built
70. l panel outward see figure Press the locking button located underneath the attachment arm 1 and turn the attachment arm approximately 30 clockwise 2 when the control panel is placed on the right side of the console or counterclockwise when the control panel is placed on the left side of the console Detach the control panel by pulling the arm out from the instrument console 3 The touch screen shows information related to the current operation and changes accordingly Some of the buttons have indicator lights that show the status of that specific function When the unit is in programming mode the Program button is blue In the programming mode the setting to be changed is displayed on the touch screen see section 19 PROGRAMMING on page 74 In the programming mode a disabled or unselected function is displayed in grey To enable or select the function press the grey button and it will turn blue A blue button means that the function is enabled or selected In case of malfunction an error code or a help message is displayed on the touch screen see section 21 HELP amp ERROR MESSAGES on page 115 User s Manual CONTROL PANEL The picture below shows the main functions of the control panel when the dental unit is in idle state and no instrument is activated 8 100 11 04 28 1 Icons for cleaning programs 2 Time display only 3 Date optional display only 4 Maintenance button
71. l the syringe which can be used at any time If you use similar instruments at the same time keep them in their own places The memory operation does not recognise the parameters and settings of the similar instruments if their places are changed User s Manual 18 2 Micromotor Speed power Reverse rotation User s Manual NOTE CAUTION NOTE OPERATING INSTRUMENTS amp 100 Standard pedal To drive the micromotor push the foot control pedal to the right or to the left Wide pedal To drive the micromotor push the foot control pedal down Pushing the pedal further to the right left or down will increase the speed or power of the micromotor As you push the pedal the power output is displayed on the control panel The normal range is 10 100 To stop the micromotor allow the pedal to return to the rest position The speed power level depends on the instrument Ensure that the pedal is in rest position when you activate the micromotor In normal operation the micromotor rotates in a clockwise direction To reverse the direction of rotation press Reverse Reversing is possible only when the micromotor is picked up from the instrument console but not operated When the micromotor is set to rotate in the reverse direction the indicator light is amber Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function Planmeca Sovereign Class
72. me to switch the air spray on A yellow indicator light shows that the air spray is switched on Press Instrument spray a third time to switch off the instrument cooling spray The indicator light goes out Alternatively you can select the spray setting with the foot control The indicator lights on the control panel are lit accordingly Standard pedal Push the pedal down briefly to select the spray setting Wide pedal Push the pedal briefly to the left to select the spray setting Push the pedal once to switch on the water amp air spray Push the pedal again to switch on the air spray Push the pedal a third time to switch off the instrument spray Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The spray must be switched off when using an instrument without a waterline inside the handpiece The instrument spray operated with the foot control can be disabled in the service mode in which case you can switch the spray on and off only from the control panel Contact your Planmeca dealer User s Manual Automatic chip blow Manual chip blow NOTE Fibre optic light User s Manual NOTE NOTE OPERATING INSTRUMENTS Press Chip blow to switch the automatic chip blow on or off When the indicator light is green the automatic chip blow is on and two short blows of water air or both will occur after the turbine is stopped Check
73. ment spray once to switch on the water amp air spray A green indicator light shows that the water amp air spray is switched on Press Instrument spray a second time to switch the air spray on A yellow indicator light shows that the air spray is switched on Press Instrument spray a third time to switch off the instrument cooling spray The indicator light goes out The spray flow rate can be programmed See section 19 3 2 Instrument spray on page 77 Alternatively you can select the spray setting with the foot control The indicator lights on the control panel are lit accordingly User s Manual Automatic chip blow Manual chip blow NOTE NOTE NOTE NOTE User s Manual NOTE OPERATING INSTRUMENTS Standard pedal Push the pedal down briefly to select the spray setting Wide pedal Push the pedal briefly to the left to select the spray setting Push the pedal once to switch on the water amp air spray Push the pedal again to switch on the air spray Push the pedal a third time to switch off the instrument spray Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The spray must be switched off when using an instrument without a waterline inside the handpiece The instrument spray operated with the foot control can be disabled in the service mode in which case you can switch the spray on and off only from the contro
74. mm 2 0 1 8 min 10 l min 50 46 mm 2 0 1 8 or copper 35 32 mm 1 4 1 3 MENB1010A0903F01 Planmeca Sovereign Classic 125 TECHNICAL SPECIFICATIONS VVireless foot control Battery size AA LR6 Battery type NiMH 1 2V min 2050mAh low self discharge UL listed USA Canada only Planmeca ProSensor power supply optional Type PSA16U 480 POE Water and waterline disinfection PlanClear Procedure Disinfection concentrate Affecttime Continuous disinfectant feed Hydrogen peroxide concentra Continuous Feed can also be disabled tion is 0 02 in normal opera tion the concentration should be between 150 and 250 ppm Dental unit waterline cleaning Hydrogen peroxide concentra Min 8h when continuous cleaning is tion is 0 25 monthly cycle enabled Dental unit waterline cleaning Hydrogen peroxide concentra Min 8h when continuous cleaning is tion is 2 5 weekly cycle disabled Cleaning of primary water lines Hydrogen peroxide concentra Min 8h Bowl rinse suction rinse water tion is 7 5 undiluted Plan monthly cycle feed line to water container Clear Suction system cleaning Orotol Plus Procedure Disinfection concentrate Affect time Suction system cleaning Orotol Plus concentration is Min 2 minutes 3 6 can be adjusted More information on Orotol Plus can be found at http www duerrdental com en products suction disin
75. n Classic 5 SYMBOLS ON PRODUCT LABELS VVARNING MAINS VOLTAGE MAX 300 VA MEDICAL A O DEVICES ONLY LBL 10035304 B Turn the main switch off before servicing the unit CAUTION 6 Planmeca Sovereign Classic LBL G 012 A A The multiple socket outlet MSO is reserved for medical devices that comply with the IEC 60601 1 standard The MSO is available as an option Note that the mains voltage is always present at the mains terminal under the cover when the unit is switched on Do NOT open the cover Only connect equipment approved by Planmeca to the multiple socket outlet WARNING Do not connect a multiple socket outlet or extension cord to the system WARNING The unit contains live mains voltage parts Always first turn off the power to the unit externally before attempting to service the lift motor backrest motor or any parts inside the electronics control box The power supply must be disconnected externally from for example a fuse or a mains switch The fuse or main switch must be lockable into off position Turning off the unit from its own mains switch DOES NOT cut off the mains volt age from all internal nodes The power supply for the Planmeca ProX X ray unit Planmeca ProSensor PoE port and external PC must be disconnected externally by removing the plug from the socket outlet User s Manual 6 FOR YOUR SAFETY User s Manual patient area eps NOTE NOTE NOTE
76. n inter rupted Restart the suction cleaning or return suction tubes to the suction arm HE6000 User s Manual Foot control calibration 1 6 top left Move the foot control pedal to the top left posi tion then push and release one of the foot control knobs Planmeca Sovereign Classic 117 HELP amp ERROR MESSAGES HELP CODE MEANING ACTION OR EXPLANATION HE6001 Foot control calibration 2 6 top centre Move the foot control pedal to the top centre position then push and release one of the foot control knobs HE6002 Foot control calibration 3 6 top right Move the foot control pedal to the top right position then push and release one of the foot control knobs HE6003 Foot control calibration 4 6 bottom right Move the foot control pedal to the bottom right position then push and release one of the foot control knobs HE6004 Foot control calibration 5 6 bottom centre Move the foot control pedal to the bottom cen tre position then push and release one of the foot control knobs HE6005 Foot control calibration 6 6 bottom left Move the foot control pedal to the bottom left position then push and release one of the foot control knobs HE6006 Foot control calibration successful The foot control was calibrated successfully and is ready for use HE9001 Update halted because wireless foot con Software update halted Activate the wireless trol is sleeping foot control b
77. n nenne nenne nnnne nennen 88 20 42 ICICLE VOM Salsa tod 89 204 3 Alter ClEAaNINg coit tintes mota ciu eis est 91 20 4 4 Short instrument flushing ccooccccococnnccocnncococonononononnonononononononnnnonannnononnnnonananenanens 92 20 45 Long Instrument TIUSAING za A A eos 94 20 4 6 SUCHONCIESMIMO ca as ica 96 20 4 7 Waferline clean pi Seales erred eee ont Sean cn eas 98 20 4 8 TEXIENSIVEHTUSHING asii diia iia tios nt 103 20 5 Cleaning ins UES recta 104 20 6 Cleaning dental unit surfaces oocccoocccnnccnccncoconncnconcncnnncnonnronnnnnnonnrnnnnnnnonnronnnnnnonannnnos 104 20 7 Cleaning instrument console cccoocccccccnccncnccccnncnnnnconnnnnnnnnconnnnononnnonnnnonnnnnonnnenannnonas 107 20 731 OIC lec ai a 107 20 8 Cleaning CUSPIAOR eine neh en Laer 108 20 81 BOU sic ee ee ee 108 20 82 Walter container na a items 109 20 8 3 INSELUMERETUSHING ROO ars re atent asil ener 109 20 84 Coarse MES ea een 110 20 8 5 Suction tube cleaning holder aaa ee a di iia 110 299 Cleaning SUCTION System ansiada 111 20 9 1 MTS MOMIA rennen 111 20 32 Aftereach Patent nn ae Dictats ales ii 111 20 9 3 AITerING WOFKING Cay ia te ai si 111 20 9 4 Weekly cleaning procedures uusssssenssnenennnnnnnnnnnnnennnnennnennenennennnnnnnennnnnnn nennen 112 20 95 Cleaning suction handpieces nee nnd aa 112 20 9 6 DUECAST deposit CUD ar ea eigen 113 20 97 VoOLARCONIAINER aaa 113 20 10 Cleaning Planmeca Pro
78. nd simultaneously selecting Short flushing on the control panel Hanging tube instruments Start the flushing cycle by selecting Short flushing on the control panel 7 Release the instrument arms if your dental unit is equipped with balanced instrument arms 8 Follow the messages on the control panel that inform you of the cleaning cycle progress The system will first identify the instruments and then each instrument hose is flushed for 30 seconds in its turn A Flushing instruments Time remaining 1 40 If you have not inserted all water consuming instruments in the instrument flushing holder help message HE5000 is displayed 8 HE5000 Required instruments are not selected vvhen starting instrument flushing or vvaterline cleaning Insert required instruments into instrument flushing holder User s Manual Planmeca Sovereign Classic 93 MAINTENANCE AND CLEANING 20 4 5 Long instrument flushing 94 Planmeca Sovereign Classic O 35 Es NOTE NOTE NOTE NOTE 9 When the flushing cycle is completed help message HE5001 is displayed on the control panel Remove the instruments from the flushing holder and place them in the instrument console Close the cuspidor door Instrument flushing completed Return instruments to the instrument console The dental unit is now ready for normal operation Instrument flushing for both the dentists and the assistant s instruments are star
79. ng 3 Close the cap of the PlanClear container Orotol Plus disinfection solution is used for cleaning the suction tubes and suction system When the Orotol Plus level is low help message HE4015 is displayed and you must fill the container The fill interval is approximately four weeks depending on how often the suction tubes and system are cleaned The Orotol Plus container must be filled with Orotol Plus only Wear protective gloves and glasses when you fill the container The Orotol Plus container is marked with a label see adjacent picture Fill the container as follows 1 Open the cap of the Orotol Plus container 2 Fill the container with Orotol Plus up to the maximum marking 3 Close the cap of the Orotol Plus container User s Manual 20 3 Continuous cleaning User s Manual JO 5 EA NOTE NOTE MAINTENANCE AND CLEANING When continuous cleaning is enabled the dental units waterlines are kept clean by continuously feeding a disinfectant solution from the water container to the waterlines and instruments The disinfectant solution is mixed in the water container A small amount of Planmeca PlanClear disinfectant concentrate is regularly dosed to the water container where it is mixed with water and fed further to the waterlines the instruments and the cup fill For information on the disinfection concentration see section 24 TECHNICAL SPECIFICATIONS on page 124 When continuo
80. ntainer Close the main water to prevent any damage to the property and contact service HE4010 No cup in cup holder 116 Planmeca Sovereign Classic Cup not detected while using cup fill function Place a cup in the cup holder User s Manual HELP CODE MEANING HELP amp ERROR MESSAGES ACTION OR EXPLANATION HE4011 PlanClear container almost empty Fill the container with PlanClear waterline dis infectant HE4012 Turn off the unit Turn off the unit for the disinfection effect time HE4013 Operation not supported Enable bowl rinsing amp cup filling functionality before usage HE4014 Water container almost empty Unable to fill the water container Make sure the main water is on If necessary fill the con tainer manually HE4015 Orotol container almost empty Fill the container with Orotol Plus suction line disinfectant HE4016 Cup in cup holder The hygiene procedure requires that the cup holder is empty Remove the cup from the holder HE4017 Too strong content in water container Perform long instrument flushing to change the water in the waterlines HE4018 Water use is prohibited The vvaterlines contain disinfectant Perform vvaterline flushing HE4019 VVater container content changed To ensure the validity of the waterline content perform long instrument flushing HE4040 HE4041 HE4050 Amalgam conta
81. or 8 Patient chair 2 Suction arm 9 Operating light 3 Foot control 10 Over the patient OP delivery arm 4 Hanging tube instruments 11 Monitor 5 Control panel 12 ProX X ray unit optional not available in all market areas 6 Instrument console 13 Bowl 7 Tray table 14 Planmeca Sovereign Classic User s Manual PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 2 Detachable parts The following detachable components are marked with a manufacturer trademark Do not perform dental treatment when either or both of these are detached Balanced instrument arms e Headrest The follovving detachable components are not critical to operation e Hygienic membrane User s Manual Planmeca Sovereign Classic 15 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT Armrests ps 2 N gt D O 09 ps SC_UM_25 e Ls ps SC_UM_29 e 16 Planmeca Sovereign Classic User s Manual PLANMECA SOVEREIGN CLASSIC DENTAL UNIT e Bowl filter and filter cap nn Dentist s instruments and hoses ps SC UM 33 e SC UM 44 eps e Suction tubes User s Manual Planmeca Sovereign Classic 17 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT Assistant s instruments and hoses Ll ps SC UM 47 e E ps SC UM 48 e 7 3 Applied parts Applied parts are parts of the dental unit that in normal treatment situations come into contact with the patient The applied parts of the Planmeca Sove
82. or hoses to be connected to the instrument console Insert the hose to its place so that the locking mechanism is facing down and pull down the locking bar 1 Remember to pull up the locking bar 1 before removing the quick connector hose When disconnecting the syringe empty the water and air from the hose before opening the connector The instrument location can be changed simply by removing the instrument with its hose and reconnecting it to the new place The old instrument settings will remain the same despite the new location of the instrument hose The instrument selection can also be interchanged The settings of the previously used instruments are stored in the memory and are recalled when the instrument is reconnected The syringe can be placed in the leftmost or rightmost position The syringe can be used simultaneously with other instruments Otherwise there are no restrictions concerning the placing of the instruments on the console Make sure that the locking bar is properly locked to the console to avoid any leakage Always make sure that the instrument hose is correct for the used instrument The control system identifies the instrument hose not the instrument The control system does not recognise the instrument change for example replacement of a turbine with an air motor The instrument sealings must be correct and unbroken and the instrument must be attached properly to the hose connector The leak
83. ould be taken when driving the patient chair from one position to another Obstructions in the patient chair s line of movement activate safety switches that stop the motorised movements The safety switches and their functions are described below 1 Backrest An obstruction between the backrest and the floor when driving down the chair and or the backrest stops downwards chair and backrest movements Remove the obstruction to resume normal operation Chair An obstruction between the chair and the floor stops downwards chair and backrest movements Remove the obstruction to resume normal operation Legrest An obstruction between the legrest and the floor stops downwards chair and backrest movements Remove the obstruction to resume normal operation OP delivery arm joint An obstruction between the OP delivery arm and the chair when driving up the chair stops upwards chair movements Remove the obstruction to resume normal operation OP delivery arm An obstruction between the OP delivery arm and the chair when driving up down the chair stops up downwards chair movements Remove the obstruction to resume normal operation ADS Planmeca Sovereign Classic 11 FOR YOUR SAFETY 6 2 Chair movement restrictors SC UM 20 2 eps 12 Planmeca Sovereign Classic Some dental unit parts are equipped with sensors that recognise when the part is in its home position When the part is not in the home position chair m
84. ovements are disabled or restricted 1 Bowl The bowl is above the patient chair and stops upward chair movements Return the bowl to its home position to resume normal operation The home position is presented in section 7 4 Bowl on page 19 Armrests When the armrests are turned outwards all chair movements are disabled Turn the armrest inwards so that it is aligned with the chair to resume normal operation Chair swivel Depending on the chair swivel angle the up downwards movements of the chair and backrest might be restricted Swivel the chair back to its home position to resume normal operation The home position is presented in section 11 3 Chair swivel on page 38 je User s Manual PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 1 Dental unit configurations 7 1 1 Over the patient delivery vvith balanced instrument arms Y ae 1 Cuspidor 8 Balanced instrument arms 2 Suction arm 9 Operating light 3 Foot control 10 Over the patient OP delivery arm 4 Patient chair 11 Monitor 5 Tray table 12 ProX X ray unit optional not available in all market areas 6 Control panel 13 Bowl 7 Instrument console User s Manual Planmeca Sovereign Classic 13 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 1 2 Over the patient delivery with hanging tube instruments IK I SC OP HT eps 1 Cuspid
85. p once after switching the unit on to start the heating Notice that when using the water heater the water temperature can rise up to 49 C At start up with the exception of start up during maintenance procedures instruments and suction tubes have to be in their holders After the unit is switched on it will take a few seconds until the Planmeca Sovereign Classic dental unit is ready for use Planmeca Sovereign Classic 51 CHECKING SOFTWARE VERSION 15 CHECKING SOFTWARE VERSION Check the software version as follows gt 1 Press Program 2 Press About this unit The following is displayed NOTE The following is an example only and does not necessarily portray the actual situation About this unit Unit Type Sovereign Classic Unit SW Version 1 0 1 85 R gt GUI Hardware Rev n a GUI Serial Number GUI Diagnostics Licences Designed and Assembled by Planmeca in Finland 3 Press OK to close the window 52 Planmeca Sovereign Classic User s Manual OPERATING PATIENT CHAIR 16 OPERATING PATIENT CHAIR 16 1 Manual operation User s Manual NOTE NOTE NOTE CAUTION CAUTION All chair movements are blocked when an instrument excluding syringe is operated The operating light intensity will decrease slightly when the chair is moving As an alternative to the control panel buttons described below the chair buttons on the Flexy p
86. pidor Durr FD 333 Durr FD 322 Tray tables Durr FD 333 Durr FD 322 Flexy holder and Durr FD 333 suction arm Durr FD 322 Suction handpieces Durr FD 333 X Durr FD 322 Monitor Durr FD 333 Durr FD 322 After every Upholstery Mild soap and patient water solution Bowl filter Mild soap and water solution User s Manual Planmeca Sovereign Classic 105 s MAINTENANCE AND CLEANING Alternative cleaning method Cleaning solution l Washer Rotecna disinfector 134 C a 93 C After the Upholstery D rr FD 312 DS DS HERE working day On Flexy holder Suction tube hold ers incl rollers instrument holder and supplementary holders Weekly Upholstery Treatment with D rr FD 360 Wipe away any excess oil after treatment Foot control Non abrasive cleaning agent designed for dental units Cover of operating D rr FD 333 D rr FD 322 light When needed polish cover with a dry and soft lint free cloth Monthly Instrument flushing Washer disinfector 93 C or autoclave 134 C holder See also section 4 8 3 Instrument ah holder on page 109 Suction tube clean Washer disinfector 93 C ing holder See also section 20 8 5 Suction tube cleaning holder on page 110 Coarse filters Rinse under running water or replace them See also section 20 8 4 Coarse filters on page 110 106 Planmeca Sovereign Classic User s Manual 20 7 Cleaning instrument console 20 7 1 Oil collector MAI
87. r Shut down the unit and contact your property s caretaker and a qualified Planmeca service technician When a water leak is detected help message HE 4004 is displayed Turn off the clinic s main water tap and contact a qualified Planmeca service technician A power cut will shut down the software controlled backflow prevention system If you are using a turbine without a built in backflow prevention system contaminated water could be released in the patient s mouth in the event of a power cut WARNING No modification of this dental unit is allowed WARNING Only instruments or equipment approved by Planmeca may be connected to this dental unit WARNING Do not simultaneously touch the patient and the PC Planmeca Sovereign Classic 9 FOR YOUR SAFETY A A WARNING Do not touch the patient when the cuspi dor door is open WARNING The patient must not be in contact with instruments when the patient is resusci tated with a defibrillator WARNING The worm screw securing nut in the chair lift motor must always be attached to the worm screw and must not be removed If the securing nut is faulty or displaced immediately stop using the dental unit and contact a qualified Planmeca service technician The worm screw nut is marked with an arrow in the picture below IO Planmeca Sovereign Classic User s Manual FOR YOUR SAFETY 6 1 Safety switches SC_UM_20 eps User s Manual Care sh
88. reign Classic dental unit include the instruments the patient chair with upholstery the armrests the outside of the cuspidor including bowl and the instrument console 18 Planmeca Sovereign Classic User s Manual PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 4 Bowl The glass bowl is attached to the top of the cuspidor It can be rotated 370 around its axis according to the figure below NOTE Always place the bowl in home position as illustrated in the picture below NOTE Make sure that the bowl is not above the patient chair when you drive the chair upward SC_UM_3 eps CAUTION Do not allow the patient to grab the bowl when getting seated or getting up from the patient chair User s Manual Planmeca Sovereign Classic 19 PLANMECA SOVEREIGN CLASSIC DENTAL UNIT If the bowl assembly for some reason comes off the cuspidor you can put it back as follows 1 Place the ring to the underside of the bowl assembly 2 Make sure that the peg goes into the slot as shown in the picture below gt IMG_18150 JPG 3 Push the bowl downwards at the same time making sure that no cables are squeezed 4 Turn the bowl assembly into place You will know that it is in place when you hear the clicking sound of the micro switch 7 5 Monitor The monitor can be moved from its handle CAUTION Do not allow the patient to grab the monitor handle or monitor arm when getting sea
89. ress OR to confirm that you want to save the current position as an automatic position If you do not wish to save the position as an automatic position press Cancel If you wish to define another preset chair position resume programming from step 2 User s Manual Planmeca Sovereign Classic 75 PROGRAMMING 19 3 Instrument settings 19 3 1 Instrument speed power reduction JE A Q 76 Planmeca Sovereign Classic NOTE 1 Activate the instrument 2 Press Program 3 Press Instrument speed power reduction A pop up window opens 4 For W amp H and Minetto brushless micromotors only Press rpm to select whether you want to adjust the instrument speed power in percentage points or in rpm 5 Change the instrument speed power with the plus and minus buttons The minimum speed power level is 5 or 10 depending on the instrument and the maximum is 95 The adjustment step is 5 6 Press OK The pop up window closes If the preset value is for example 50 the foot control pedal movement controls the speed power between 5 50 instead of the normal range of 5 100 The instrument speed power reduction does not affect the air driven instruments for which quickstart has been selected User s Manual PROGRAMMING 19 3 2 Instrument spray 1 Activate the instrument Press Program E 3 Press Instrument spray pop up window opens 4 Adjust the flow rates for water and air w
90. rinsing only not for drinking When the chair stops in the rinsing position the indicator light remains on Press Rinsing position a second time to return to the working position The chair remembers this previous position even though it has been moved into the rinsing position Bowl rinsing starts automatically The indicator light flashes throughout the movement When the chair stops the operating light turns back on if it was on initially Alternatively you can press the Flexy button a second time to return to the working position If you hold your finger on the Rinsing position button when moving the chair to the rinsing position you will enter the position normally but when the chair stops in the rinsing position the indicator light goes off and the chair does not remember its previous position If you hold your finger on the Rinsing position button when the chair is in the rinsing position the indicator light goes off and the chair will forget the previous position the return function is inactivated The unit can be set up so that the Rinsing position indicator light goes off and the previous position is forgotten when you are driving from the rinsing position to a preprogrammed position Next time you press Rinsing position the chair will move to the rinsing position Contact your Planmeca dealer You can drive the chair to the rinsing position with the foot control by pushing down the right side knob To return
91. roX see Planmeca ProX User s Manual User s Manual PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 10 Zeiss OPMI pico microscope The Zeiss OPMI pico dental microscope can be mounted to the Planmeca Sovereign Classic dental unit pylon NOTE The Planmeca Intra X ray unit and the Zeiss OPMI pico dental microscope can not simultaneously be attached to the Planmeca Sovereign Classic dental unit CAUTION Do not drive the chair when a patient is under the microscope For more information about Zeiss OPMI pico see the manufacturer s own documentation User s Manual Planmeca Sovereign Classic 25 OPERATING LIGHT 8 OPERATING LIGHT SingLED13 eps NOTE NOTE CAUTION 26 Planmeca Sovereign Classic The Planmeca SingLED operating light features a no touch function through an infrared light cell detecting hand movements Wave your hand shortly in front of the sensors 1 to switch on off the operating light It will switch on off the light after you have moved your hand away from the sensor activation area Also you will hear a signal tone Once the operation light is lit keep your hand longer in front of the sensors to adjust the brightness of the light The adjustment range is from 5 to 100 in steps of 5 percentage points You will hear a signal tone when the maximum minimum brightness is reached The control panel or foot control can be used alongside the sensor to switch the operating light on o
92. s dedica nidad 74 19 2 Automatic enai POSINONS sisi ea 75 193 Instrument sellings nun ame ee 76 19 3 1 Instrument speed power reduction 22 2u022400024000nnnennnnn nenne nenne nenne nenne nnnnn nennen nennen 76 19 32 OS TUNES secta tetes ee 77 19 33 CID DOW a ae eier 77 19 84 Eibre ptie light4 LED NGA ana u A 78 19 3 5 Planmeca Lumion polymerisation light occoonncconcnncccnconcnconcnconancnnanonnncnnoncncnncnnnns 78 WER MM CSC ICS ios 79 195 Bowinnse and CUB anche 79 19 5 1 Adjusting cup fill and bowl rinse flow rates ooccccoccconcconncocnconncconoconnonanonanonanonons 79 ii Planmeca Sovereign Classic User s Manual TABLE OF CONTENTS 1952 DUraton orDOWI ASIN uta idas 80 153 DUGUON OIL cu OMA abc 80 196 Intensitviot operates dira latina 81 19 7 DOor pen ass stant call esse dos 82 19 8 COCKS arras 83 MI A ee a os na eK Se ee era sen 84 20 MAINTENANCE AND CLEANING 222 2220220002000a2000n00nan0nnnnnnanunnnunn nun nun 85 BOs GUSPIAO DES stos E A esmenat sois 85 20 2 Filling disinfectant containers music ais 86 20221 Filling PlanGlearContalnen rana 86 20 22 _ Filling Orotoll Plus contain acer aaa en 86 203 GONIINUOUSC EAN Neal nei 87 20 4 Flushing and cleaning programs uu244s044sneenne nenn nnennnnnnennnnennne nenne nenn nenn nenne nenn nennen 88 20 4 1 When to use cleaning programs uusssssensenenennnnennnennnnnnnnnnnnnnnennnnen
93. s on how to flush the waterlines see section 20 4 6 Suction cleaning on page 96 1 Remove the cup from the cup holder Turn the cup fill tube so that it is above the bowl and not above the cup holder 2 Press Maintenance to go into maintenance mode 3 Remove the instrument handpieces from the instruments and clean them according to the manufacturer s instructions 4 Open the cuspidor door User s Manual User s Manual MAINTENANCE AND CLEANING Place the instruments into the openings in the flushing holder DCI syringe Remove the instrument cover and use an insert before placing the syringe in the flushing holder Lock the syringe into place with a rotating movement when the syringe is in the holder Luzzani syringe The metallic syringe cover must be removed before placing the syringe into the flushing holder Note the orientation of the syringe lt does not require an insert and must be placed in the holder as shown below ps SC UM 52 e Balanced instrument arms Start the vvaterline cleaning by bending the instrument arms of all vvater consuming instruments to an angle of at least 90 and simultaneously selecting Waterline cleaning on the control panel Hanging tube instruments Start the waterline cleaning by selecting Waterline cleaning on the control panel Release the instrument arms if your dental unit is equipped with balanced instrument arms Follow the messages on the control panel
94. ss for 1 second to unlock A progress bar is displayed while you press VS A button for dental units with a VS A suction system For instructions see section 20 9 4 Weekly cleaning procedures on page 112 20 4 3 After cleaning When the monthly or weekly if continuous cleaning is disabled waterline cleaning program is finished measure the concentration of hydrogen peroxide in the water that comes from the instruments and the cup fill line with a test strip The hydrogen peroxide concentration must be lt 250 ppm Testing water with a test strip Before testing the water read the information provided by the test strip manufacturer 1 Add water in a clean cup from an instrument or the cup fill tube 2 Dip the test strip into the water Wait for a while before reading the result on the test strip The affect time is specified by the test strip manufacturer User s Manual Planmeca Sovereign Classic 91 MAINTENANCE AND CLEANING 20 4 4 Short instrument flushing NOTE NOTE 30 NOTE LES KISS O 92 Planmeca Sovereign Classic Result For information on how to interpret the results on the test strip see the information provided by the test strip manufacturer In normal operation the hydrogen peroxide concentration is lt 250 ppm Concentration levels higher than 250 ppm means that the water can not be used when treating patients If the concentration level is too high perform lon
95. ssic If the unit has been stored at temperatures below 10 C for more than a few hours time must be allowed for the unit to reach room temperature in the original packing before connecting the unit to the mains voltage Before using the dental unit ensure that the instruments have been properly flushed and that the suction tubes as well as the dental units waterlines have been cleaned as instructed in this manual Guide the patient to sit on the chair Make sure no one sits on the legrest the backrest or any other part of the dental unit Do not touch the patient when opening the cuspidor door A faulty or broken dental unit must not be used Drops of water on the touch screen might disturb the functioning of the screen Lock the touch screen from the Maintenance window before using a table top instrument Make sure that the cup fill tube is always properly in its position by the cup holder The light source of the SingLED operating light may cause retinal injury if viewed upon directly Protect the patients and dental treatment staff s eyes with protective glasses that block high energy visible light HEV light or limit the direct exposure to 10 minutes Electrosurgical knives generate strong radio frequency signals that can interfere with the functions of other electronic equipment Although the Planmeca Sovereign Classic dental unit is tested to be very interference tolerant and is not known to be disturbed b
96. ssistant call function is enabled At the same time the Door open button turns grey which means that the door open function is disabled If so configured you activate the disabled function from the foot control Please contact a qualified Planmeca service technician User s Manual 198 Clock User s Manual 9 NOTE PROGRAMMING Optionally adjust the door open assistant call time with the plus and minus buttons Only the value of the enabled function can be adjusted The minimum value is O seconds the maximum 250 seconds The value is adjusted in steps of 1 in value range O 30 and in steps of 5 in value range 30 250 The default value is 5 Press OK The pop up window closes Press Program Press Clock A pop up window opens Select the clock programming view by pressing Clock at the bottom of the programming window Press 12 24 to change the mode 12 hour 24 hour clock Change the time with the plus and minus buttons Press OK The pop up window closes and the new time is displayed on the control panel Planmeca Sovereign Classic 83 PROGRAMMING 19 9 Date LV pu 84 Planmeca Sovereign Classic Press Program Press Clock A pop up window opens Select the date programming view by pressing Date at the bottom of the programming window When the date is not visible on the control panel the items in the programming window are grey This means tha
97. system only 4 Suction tube cleaning holder User s Manual Planmeca Sovereign Classic 85 MAINTENANCE AND CLEANING 20 2 Filling disinfectant containers 20 2 1 Filling PlanClear container CAUTION CAUTION Planmeca A PlanClear D 20 2 2 Filling Orotol Plus container CAUTION CAUTION 86 Planmeca Sovereign Classic Two different disinfectants are used for the dental units flushing and cleaning programs e Planmeca PlanClear for instrument flushing and dental unit waterline cleaning e Orotol Plus for suction tube cleaning The disinfectant containers are filled manually according to the instructions below For more information on the flushing and cleaning programs see section 20 4 Flushing and cleaning programs on page 88 The PlanClear level must be between the minimum and maximum markings on the container When the PlanClear level is low help message HE4011 is displayed and you must fill the container The fill interval is approximately four weeks depending on how often the instruments and waterlines are flushed and cleaned The PlanClear container must be filled with Planmeca PlanClear disinfectant only Wear protective gloves and glasses when you fill the container The PlanClear container is marked with a label see adjacent picture Fill the container as follows 1 Open the cap of the PlanClear container 2 Fill the container with Planmeca PlanClear up to the maximum marki
98. t intensity can be programmed see section 19 3 4 Fibre optic light LED light on page 78 Planmeca Sovereign Classic 71 OPERATING INSTRUMENTS 18 5 Polymerisation light CAUTION CAUTION Satelec Mini LED Planmeca Lumion 72 Planmeca Sovereign Classic The polymerisation light generates optical radiation and proper safety measures should be taken when using the instrument For detailed information please refer to OEM documentation Do not use the polymerisation light on patients with cardiac pacemakers The polymerisation light can cause disturbance on the pacemaker s function When the polymerisation light is picked up from the instrument console or activated by bending the balanced instrument arm the text Polymerisation light active is displayed on the control panel To operate the polymerisation light first press the navigation button on the handpiece to choose the curing menu Then start the polymerisation cycle by pressing the Start Stop button on the handpiece The polymerisation cycle stops automatically at the end of the cycle but can be interrupted by pressing the Start Stop button Refer to the Satelec Mini LED User Manual To start the polymerisation cycle first select the desired curing mode by pressing the navigation button on the handpiece Then start the polymerisation cycle by pressing the start button on the handpiece The polymerisation cycle can be interrupted by pressing one
99. t they are disabled To show the date on the control panel and to enable programming press the grey Date button at the top of the programming window Date 28 02 2014 The button will turn blue and all items in the window will turn black which means that programming is enabled Select the date format Press the arrow to scroll through the different format options e dd mm yyyy day month year e mm dd yyyy month day year e yyyy mm dd year month day Change the day month and year with the plus and minus buttons Press OK The pop up window closes and the new date is displayed on the control panel User s Manual 20 20 1 MAINTENANCE AND CLEANING MAINTENANCE AND CLEANING Cuspidor parts The different parts of the cuspidor that require maintenance and cleaning are presented in the picture belovv 1 2 gt 3 4 3 5 hm A 6 o ES 7 y 1 Water container with automatic filling 5 PlanClear container 2 Instrument flushing holder 6 Orotol Plus container 3 Coarse filters 7 Deposit cup D rr CAS1 suction
100. tant should not be left in the waterlines for longer than one night At a regular interval a full cleaning of the waterlines is performed A full cleaning lasts about two minutes longer than the regular cleaning cycle and cleans the waterlines User s Manual Resetting counter User s Manual MAINTENANCE AND CLEANING more thoroughly than the regular cycle The dental unit has a counter that keeps track of when to perform the next full cleaning so the user always performs the same cleaning procedure irrespective of whether the next cleaning is full or regular The counter counts the days that are left until the next full cleaning When the counter reaches O days it means that the next time you start waterline cleaning a full cleaning of the waterlines is initiated The counter can be reset to O for example for maintenance purposes For instructions see section Resetting counter on page 99 After installation or if the dental unit has not been used for a long time you must perform waterline cleaning before starting dental treatment 1 Press Program 2 Press Maintenance The following window opens Program Maintenance Continuous cleaning Time to full cleaning 30 days 3 Inthe pop up window press Now to reset the counter to O days The next time you start Waterline cleaning a full cleaning cycle is run Pressing the Now button does not start the waterline cleaning For instructions on how to st
101. ted or getting up from the patient chair Clean the monitor according to the instructions in section 20 6 Cleaning dental unit surfaces on page 104 NOTE Never spray water on the monitor or the PC For more information see the monitors user s manual 20 Planmeca Sovereign Classic User s Manual PLANMECA SOVEREIGN CLASSIC DENTAL UNIT 7 6 USB connectivity The Planmeca Sovereign Classic dental unit has several USB ports for different purposes 1 The USB port on the instrument hose slot on the backside of the instrument console is dedicated for the intraoral camera A Planmeca service technician can relocate this USB port to any slot in the console 2 Optionally a USB port can be installed on the backside of the instrument console to offer a USB connection for downloading uploading your personal settings This feature will be available in the future 3 The USB port inside the cuspidor enables the use of an intraoral camera on the assistant side At installation the position for the intraoral camera is selected either USB port 1 or 3 and the port that is selected is linked to the external PC NOTE Connect only intraoral cameras supplied by Planmeca to the dental unit t 4 ps SC_UM_49 2 e User s Manual Planmeca Sovereign Classic 21 PL
102. ted at the same time The duration of the long flushing cycle is configurable The factory default is 9 minutes Please contact a qualified Planmeca service technician In units equipped with a water heater the long flushing fills the water lines and water heater with cool water and switches the water heater off The flushing cycle can be interrupted by pressing Cancel After the interruption the dental unit can be used normally 1 Turn the cup fill tube so that it is above the bowl and not above the cup holder 2 Press Maintenance to go into maintenance mode 3 Remove the instrument handpieces from the instruments and clean them according to the manufacturers instructions 4 Open the cuspidor door User s Manual Y la N 7 Y Mt H y U B B A Es N de User s Manual MAINTENANCE AND CLEANING Place the instruments into the openings in the flushing holder DCI syringe Remove the instrument cover and use an insert before placing the syringe in the flushing holder Lock the syringe into place with a rotating movement when the syringe is in the holder Luzzani syringe The metallic syringe cover must be removed before placing the syringe into the flushing holder Note the orientation of the syringe lt does not require an insert and must be placed in the holder as shown below ps SC UM 52 e Balanced instrument arms Start the flushing cycle by bending the instrument arms o
103. ter cleaning on page 91 The concentration of hydrogen peroxide in the water must be lt 250 ppm If it is higher gt 250 ppm perform long instrument flushing Long instrument flushing takes 9 minutes default For instructions on long instrument flushing see section 20 4 5 Long instrument flushing on page 94 The dental unit is now ready for normal operation 102 Planmeca Sovereign Classic User s Manual 20 4 8 Extensive flushing 7 eps User s Manual NOTE NOTE MAINTENANCE AND CLEANING Extensive flushing takes lt 30 minutes to complete Adjust the water flow of the cup fill tube so that the water does not splash into the bowl For instructions see section 19 5 1 Adjusting cup fill and bowl rinse flow rates on page 79 The flushing cycle can be interrupted by pressing Cancel After the interruption make sure that the water is clean before using the dental unit again WARNING Even though every effort has been taken to ensure patient safety even in case of malfunction or misuse always make sure that the unit is properly flushed before taking it into use 1 Turn the cup fill tube so that it is above the bowl and not above the cup holder 2 Press Maintenance to go into maintenance mode 3 Remove the instrument handpieces from the instruments and clean them according to the manufacturer s instructions 4 Open the cuspidor door 5 Place the instruments into the
104. the chair to the required position use the foot control s centre knob or the chair positioning buttons on the control panel in a non automatic mode The operating light is programmed to switch on or off in preprogrammed positions The light goes off when the chair begins to move to a position where the light has been programmed to be off but goes on only after the chair has reached the programmed position where the light has been programmed to be on For more information on how to program the automatic positions see section 19 2 Automatic chair positions on page 75 Planmeca Sovereign Classic 55 OPERATING PATIENT CHAIR 16 3 Trendelenburg position Na 16 4 Rinsing position gre Y 2 56 Planmeca Sovereign Classic NOTE NOTE NOTE NOTE NOTE NOTE To drive the patient chair to the Trendelenburg position press the Backrest down button until the backrest reaches its lower movement limit and then press the Backrest down button again The chair is driven to a position where the patient s feet are higher than the head Press Rinsing position to move the chair to the preprogrammed rinsing position The indicator light flashes throughout the movement Filling the cup starts automatically the bowl is rinsed and the operating light goes off Alternatively you can press the Flexy button to move the chair to the preprogrammed rinsing position The water in the dental unit is intended for
105. to adjust the duration of the bowl rinse and cup fill For instructions see sections 19 5 2 Duration of bowl rinsing on page 80 and 19 5 3 Duration of cup filling on page 80 Planmeca Sovereign Classic 79 PROGRAMMING 19 5 2 Duration of bowl rinsing a 19 5 3 Duration of cup filling NOTE Press Program Press Bowl rinse A pop up window opens Change the duration of the bowl rinsing with the plus and minus buttons The minimum value is 5 seconds and the maximum 240 seconds and it is adjusted in steps of 5 Press OK The pop up window closes Press Program Press Cup fill A pop up window opens Change the duration of the cup filling with the plus and minus buttons The minimum value is 2 seconds and the maximum is 10 seconds The duration is adjusted in steps of 0 5 seconds Press OK The pop up window closes The cup filling is not activated unless the cup is positioned in its place under the cup fill tube The duration of the cup filling can also be programmed as follows Place an empty cup to the cup holder gt 2 Press Program 80 Planmeca Sovereign Classic User s Manual 19 6 Intensity of operating light User s Manual C e i NOTE PROGRAMMING Press Cup fill Push the foot control pedal down The cup filling time will first be set to 2 seconds and will start to increase after pushing the foot control pedal for 2 seconds The
106. to prevent the disinfection solution from spilling Also put a cap on the empty suction tube connector s The duration of suction tube cleaning is lt 5 minutes You can interrupt the cleaning program by pressing Cancel After the interruption you must perform suction tube cleaning before you can use the dental unit again Help messages on the control panel guide you through this procedure 1 Remove the suction handpieces from the suction tubes and clean them according to the manufacturer s instructions 2 Press Maintenance to go into maintenance mode User s Manual ps o co m O D Ti 5 User s Manual 4 MAINTENANCE AND CLEANING 3 Open the cuspidor door and insert the suction tubes into the suction tube cleaning holder Select Suction cleaning on the control panel Follow the messages on the control panel that inform you of the cleaning cycle progress L Flushing suction tubes Time remaining 1 40 If you have not inserted all suction tubes in the cleaning holder help message HE5005 is displayed Required suction tubes are not selected when starting suction cleaning Insert the required suction tubes into the suction tube cleaning holder The suction tubes are cleaned with Orotol Plus disinfection solution When Orotol is pumped to the suction cleaning system the yellow Orotol icon is displayed on the control panel Planmeca Sovereign Classic 97 MAINTEN
107. to run extensive flushing by displaying help message HE4018 on the con trol panel Duration lt 30 minutes When the dental unit is equipped with a water heater the duration of all cleaning programs is slightly longer When continuous cleaning is disabled make sure that the water container is clean before starting the cleaning program lf necessary clean the container according to the instructions in section 20 8 2 Water container on page 109 Always feed cold water to the dental unit The water in the dental unit is intended for rinsing only not for drinking The main water tap must be turned off when the dental unit is not in use If cleaning solution is splashed on the surfaces of the dental unit remove the splashes instantly with water and mild soap to avoid stains The water must flow properly through the cup fill line otherwise the unit will not be flushed Make sure that the cup fill line is not totally closed The flow through the cup fill line is recommended to be approximately 1 dl 5 sec Planmeca Sovereign Classic 89 MAINTENANCE AND CLEANING NOTE NOTE NOTE 90 Planmeca Sovereign Classic Turn off the main water feed at the end of the day Before switching on the dental unit in the morning make sure that the main water feed is turned on In the Maintenance mode you can perform instrument flushing as well as cleaning of the suction tubes and the dental unit s waterlines During
108. to the working position push down the right side knob again For information on how to program the rinsing position see section 19 2 Automatic chair positions on page 75 User s Manual OPERATING DENTAL UNIT 17 OPERATING DENTAL UNIT 17 1 Language A selection of languages is available for the touch panel from software version 1 0 2 onward Follow the steps below to change the language 1 Press Program 2 Press Language to open a list of available languages The available languages are e English e Finnish e German e Spanish e French e Italian e Swedish e Hungarian e Czech e Danish e Norwegian e Russian e Japanese e Polish e Traditional Chinese e Simplified Chinese e Romanian e Arabic e Dutch e Portuguese 3 Select a language by pressing it Move up and down on the list by pressing the up down buttons The list of languages closes and the language of your choice is displayed on the touch panel NOTE The language can not be changed from the foot User s Manual control Planmeca Sovereign Classic 57 OPERATING DENTAL UNIT 17 2 Bowl rinse s 17 3 Cup fill NOTE NOTE 58 Planmeca Sovereign Classic Control panel Press Bowl rinse to rinse the bowl Bowl rinsing can be stopped before it stops automatically by pressing Bowl rinse again Flexy holder Press Bowl rinse cup fill to rinse the bowl Bowl rinsing can be stopped before it stops automat
109. trument speed power reduction to reduce the turbine speed or power to a preset level If the preset value is for example 50 the foot control pedal movement controls the speed power between 5 50 instead of the normal range of 5 100 The preset maximum value is displayed on the control panel when the turbine is picked up from the instrument console Press Instrument speed power reduction again to switch the power reduction off When the turbine s speed or power is reduced the indicator light is green Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function The instrument speed power reduction level can be programmed See section 19 3 1 Instrument speed power reduction on page 76 Instrument speed power reduction does not affect the air driven instruments for which quickstart has been selected Note that the speed power level depends on the instrument Planmeca Sovereign Classic 67 OPERATING INSTRUMENTS Instrument spray 68 Planmeca Sovereign Classic NOTE NOTE NOTE The instrument cooling spray setting can be selected to be water amp air air or no spray The amounts of water and air for the currently used spray type are displayed on the control panel Press nstrument spray once to switch on the water amp air spray A green indicator light shows that the water amp air spray is switched on Press Instrument spray a second ti
110. turbance on the pacemaker s function Standard pedal To drive the scaler push the foot control pedal to the right or to the left Wide pedal To drive the scaler push the foot control pedal down Pushing the pedal further to the right left or down will increase the speed or power of the scaler As you push the pedal the power output is displayed on the control panel To stop the scaler allow the pedal to return to the rest position The speed power level depends on the instrument Ensure that the pedal is in rest position when you activate the scaler There are two adjustable spray settings for the scaler water 1 and water 2 The currently used spray setting water 1 water 2 spray off can be changed by pressing Instrument spray Alternatively you can change the scaler spray setting with the foot control pedal Standard pedal Push the pedal down briefly to select the spray setting Wide pedal Push the pedal briefly to the left to select the spray setting Check the indicator light on the control panel to verify that pushing the left side knob down activates the correct function When water 1 is selected the Instrument spray indicator light is green When water 2 is selected the Instrument spray indicator light is yellow When the spray is off the Instrument spray indicator light is off User s Manual 100 LM scaler A NOTE EMS No Pain scaler WRESTO psc T VENDO Y I W
111. turning of suction arm yy AS PZ PA eg dU yo 7 ke _ e 550 21 7 640 25 2 centerline for left right outermost unit conversion cuspidor position 134 Planmeca Sovereign Classic User s Manual TECHNICAL SPECIFICATIONS Patient chair 670 26 4 CS 570 22 4 Horizontal 470 18 5 530 20 9 Entry lt AA 570 22 4 Horizontal 530 20 9 Entry i 670 26 4 575 22 6 User s Manual Planmeca Sovereign Classic 135 TECHNICAL SPECIFICATIONS 24 2 FCC Class B Notice for wireless foot control This device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference 2 This device must accept any interference received including interference that may Cause undesired operation Note This equipment has been tested and found to comply with the limits for a Class B digital device pursuant to Part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation
112. ument spray type Push pedal down briefly Push pedal briefly to left right Activate manual chip blow Push and hold pedal down Push and hold pedal to left right Activate momentary spray Push pedal down briefly while N A driving instrument NOTE NOTE NOTE User s Manual For more information see sections 18 2 Micromotor on page 63 18 3 Turbine on page 66 and 18 4 Scaler on page 70 If you want to change from a standard pedal to the wide pedal or vice versa please contact a qualified Planmeca service technician The differences in functionality only applies to the operation of some instruments not the operation of the dental unit chair or headrest When the functionality of the foot control pedal is different for the standard pedal and the wide pedal this is clearly indicated in this document When the document refers to the foot control pedal in general without making this differentiation the same functionality applies to both types of pedal although the illustration only presents the standard pedal Planmeca Sovereign Classic 47 FOOT CONTROL When an instrument is activated the functions of the standard pedal depend on the used instrument These instrument specific functions are presented below Micromotor turbine scaler polisher Change spray 1 2 OFF Micromotor turbine scaler Manual chip blow Micromotor activated for as long as the turbine pedal is pushed and held down
113. us cleaning is enabled factory default it is shown as a small green icon at the top of the control panel When continuous cleaning is disabled the icon is grey Even if continuous cleaning is enabled you must still run the cleaning programs as described in section 20 4 Flushing and cleaning programs on page 88 When you disable continuous cleaning a message is displayed that advises you to perform long flushing to remove excess disinfectant solution from the waterlines However negligence to perform long flushing does not prevent dental unit operation To disable enable the continuous cleaning follow the steps below 1 Press Program 2 Press Maintenance The following window opens Program Maintenance Continuous cleaning Time to full cleaning 30 days Planmeca Sovereign Classic 87 MAINTENANCE AND CLEANING 3 In the pop up window toggle the Continuous cleaning button to enable disable continuous disinfection A grey button means that disinfection is disabled and a blue button that disinfection is enabled NOTE The Time to full cleaning section in the window has to do with the Waterline cleaning program see section Resetting counter on page 99 4 Press OR to confirm the selection If you want to close the window without changing the setting press Cancel 20 4 Flushing and cleaning programs 20 4 1 When to use cleaning programs We recommend that you run the flushing and
114. vereign Classic dental unit is delivered with the following manuals e Users manual For dental care professionals Describes the dental unit and its different parts as well as instructs how to operate and clean the dental unit e Installation manual For service personnel Describes how to install the dental unit e Technical manual For service personnel Gives instructions for service situations Planmeca Sovereign Classic 1 ASSOCIATED DOCUMENTATION 2 Planmeca Sovereign Classic The intraoral X ray unit Planmeca ProX can be installed to Planmeca Sovereign Classic Planmeca ProX is an option that is available in selected market areas Please contact your local sales representative for details Planmeca ProX is delivered with the following manuals e Users manual For dental health care professionals Describes the intraoral X ray unit and its different parts as well as instructs how to operate and clean the X ray unit e Installation manual For service personnel Describes how to install the intraoral X ray unit e Technical manual For service personnel Gives instructions for service situations The Planmeca ProSensor sensor is delivered with the following manuals e Users manual For dental care professionals Describes the sensor that is intended to be used for capturing digital intraoral X ray images and instructs how to use it e Installation manual For service personnel Describes how to install the sensor
115. when the button is pressed again Mix 20 ml Orotol Plus liquid with 1 water and shake well Pour the solution into the bowl Let the solution affect overnight Rinse the suction system the next morning with 2 of water Rinse the bowl with water the next morning by pressing Bowl rinse To clean the suction handpiece unscrew it from the suction tube connector Then disassemble the handpiece as instructed below 1 Identification bushing 2 3 4 D rr saliva suction D rr high volume suction Cattani saliva 4 high volume suction User s Manual MAINTENANCE AND CLEANING 20 9 6 D rr CAS1 deposit cup NY NJ 1 77 Y ch If your dental unit is equipped with a D rr CAS1 suction system replace it when it is gt 90 full An indicator light next to the deposit cup indicates the degree of fullness of the deposit cup B u A a Yellow indicator light 2 90 full Red indicator light 100 full ps o 00 SC UM 42 e 00 LE INTE 20 9 7 VS A container If your dental unit is equipped with a VS A suction system empty the VS A fine filter when it is full N L A When you have emptied the fine filter make sure you f place it properly back in its holder to prevent the water from leaking onto the floor When the fine filter is clogged the dental unit will instruct you to empty the fine filter In some error situ
116. y signals generated by the XO electrosurgical knife we recommend that the dental unit be shut down when the electrosurgical knife is used The use of the XO electrosurgical knife may affect the function of an implanted pacemaker or defibrillator Please refer to the manufacturer s own documentation Do not use the scaler or the polymerisation light on patients with cardiac pacemakers The instrument can cause disturbance on the pacemaker s function Do not use the equipment in close conjunction with anaesthetic gas or in highly oxygenated environments oxygen content gt 25 User s Manual User s Manual CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION CAUTION A A A FOR YOUR SAFETY Do not perform other maintenance procedures than those instructed in this manual When servicing the unit always switch the unit off Note that in extreme conditions the temperature of the chair lift and backrest motors might rise significantly DO NOT TOUCH THE MOTORS Note that the mains voltage is always present at the mains terminal under the main control PCB cover even if the unit is switched off from its own mains switch The power supply must be disconnected externally from for example a fuse or a mains switch The fuse or main switch must be lockable into off position If the drain is blocked the dental unit might overflow with contaminated water and excessive water could flow onto the floo
117. y pressing the handle 21 3 Error messages The following table lists the error messages which are mainly intended to assist the technician General ERROR CODE EXPLANATION Failed to load settings Control panel version mismatch Software update package corrupted Incompatible wireless foot control software Incompatible foot control software Incompatible wireless foot control receiver software Incompatible SingLED software Incompatible instrument console software Incompatible Mamco A software Incompatible Mamco B software Incompatible Water Management System software Incompatible control panel software Incompatible headrest control software 118 Planmeca Sovereign Classic User s Manual HELP amp ERROR MESSAGES Instruments EXPLANATION Instrument power on failed Scaler current leakage Drive air pressure too low Spray air pressure too low Spray water pressure too low Drive air pressure too high Spray air pressure too high Spray water pressure too high Chair ERROR CODE EXPLANATION Seat position sensor error Seat motor error Seat movement error E3006 Backrest motor error Water system ERROR CODE EXPLANATION Water pump pressure error Water pump current error PlanClear pump error Orotol pump error Water container fill timeout PlanClear cont