Home
        Cold-Probe Users Manual - University of Colorado Denver
         Contents
1.     z   el   j     Compressor O Inlet  psi  213 0 Htr Res  Ohms  F 40  j Valve 2 Channel 9 1 Cool Pack Errors      F valve 3 Channel  2     Eject Air O lon Pump  Torr  Mda Mas Eoaea  S     Valve 4 Channel a8    Turbo  Torr    2 31E 8     z   Ion Pump O     Ion Valve Channel 44    High Limit w   10 00      Accelerate Warm Up ie acon  Test Valves  F4  Sars TA Low Limit  Y        X   CoolPack Configuration  Enable Eject Air on Error    Load Config  F5  Disable Eject Air on Error Potente a                         Ion Pump On          Error Path 9 C  CryobayErrors      Ion Pump Off Data Path  amp C  CryobayData    Exit         Enable Yacuum Error Checking  Save Config  F6  Disable Vacuum Error Checking Error  amp  Data File Paths    Figure 4 The layout of the CryoBay Monitor Window  During routine operation  the region  outlined in red should be monitored carefully        Temp  is the temperature of the cold head inside the probe  and is essentially the temperature of the  RF coil  This value should not change by more than     0 5 K when an experiment starts  Once steady  state has been reached this value should be stable at 25 K     Heater  is the amount of power  in Watts  required to heat the super cooled He gas to 25 K  This value  should not change by more than 1 Watt compared to its starting value  If too much power is used  during an experiment then the probe will not be able to maintain a steady temperature and damage  may occur to the probe and your sample     Timer is th
2.  Before starting your acquisition with the cold probe  you should do the following     e dsp  i  or dsp  to make sure that it is set to    i        qcomp    y     or qcomp  to check    oversamp 16  or oversamp  to check    rof2 2 4  not  gt  5    gradientshaping  y      CHECK your decoupling and spin lock powers  decoupling and spin lock times  and your   acquisition time to be sure that you do not exceed the limits specified for the probe  Seeprevious   page     e After starting your acquisition  CLOSELY monitor the CryoBay Monitor window  paying close  attention to the Heater power  and the Temperature  You should monitor this for at least 5  minutes after the acquisition begins    e Ifthe heater drops more than 1 2 Watts from the baseline  i e  if    Heater    drops from 3 9 to 2 7    ABORT the acquisition  aa   and check your decoupler powers  spin lock powers  decoupling  times  and spin lock times  see above   Remember that in most cases  the acquisition time  at   also represents a decoupling time    e For long  multi day  experiments  check the information in the CryoBay Monitor Window each  time you check on your acquisition     10    
3.  channels  tune only  no match adjustment   the cabling for probe tuning  is the same as for a conventional probe     Cold Probe Operation on the 600  4  Setting Up Experiments     Variable Temperature  Sample Temperature      e The cold probe has a VT range of 0 to  50 degrees Celsius  DO NOT EXCEED THESE LIMITS    e Do not set the FTS below  15 degrees C    e ALWAYS ensure that there is VT gas flow to the probe  20 L min   This is very important to prevent  the sample from freezing     Radiofrequency Power Handling Issues     As with any probe  using too much Rf power for too long a time can damage the probe  The cold probe  is more efficient  in terms of B1 field vs  Rf power   therefore  the user must be cognizant of the correct  calibrations for this probe  The power levels for the cold probe must be adjusted from those values  used in experiments performed on the regular probe  The values below represent safe operating limits  for   C and    N Decoupling  and  H Spin Locks  If multiple spin locks and multi nucleus decoupling is  used  you must adjust your pulse sequence parameters so that you do not too put much power into the  probe     e Standard Pulse Widths  1H 8 2 us at 55 dB  13C 14 us at 59 dB  15N 40 us at 60 dB  e Safe levels for C Decoupling   dowr  43  pw  80 4 us  3 1 KHz field  for a maximum of 12 Omsec  dowr  45  pw   63 9 us  3 9 KHz field  for a maximum of 60 msec    e Safe levels for   C Spinlock  owr  52dB  8 7 KHz  for a maximum of 25 msec    e Safe l
4.  particularly those obtained from Lewis  Kay in Toronto do not use this parameter  Damage to the probe may occur if you use one of these  experiments without any modification    Cold Probe Operation on the 600    Instrument Cabling   The preamplifer for the proton observe channel is now located inside the cryogenically cooled    region of the probe  Consequently a different receiver pre amplifier module  called a    Cryopreamp  Driver     has replaced the original preamp  The way this system is setup for operation and probe  tuning is different     Probe Tuning   In addition to cabling issues  the cold probe has a different user interface for probe tuning  A single    brass knob is used to tune all channels  A larger knob is used to position the tuning stick to engage  the tuning element for each channel  Tuning channels are available for  H Tune   H Match     C  Tune   N Tune  and   H Tune     Cold Probe Operation on the 600    2  Description of the Cold Probe System    The layout of the cold probe system and the individual components are shown in Figure 2  The basic  theory of how this woks and what each component does is as follows     Figure 2 Layout of the Cold Probe System       Close Cycle Chiller   He refrigerator      r             ae N     Vibration Cryogenic  Vibration isolation plate damping pier probe    e Cryobay  o Helium gas is compressed by a compressor inside the CryoBay and this is fed into the e  Closed Cycle Chiller  CCC    which is the box located in the
5.  pit of the 600  where it  undergoes thermal expansion   e Closed Cycle Chiller  o Inside the CCC  thermal expansion of the He gas is used to cool the cold head to below 25  Kelvin  A small amount of the Helium gas is routed through the cold head in the CCC where  it is cooled to below 25 Kelvin  and then flows through the flexible stainless steel line to the  side arm of the probe  and up through the    probe  Sample tube   e Probe  o Within the probe  the cold He gas is used to Cryo cooled r f  coil   cool the tuning elements  the coils  and the Cryogenic heat exchanger   receiver pre amplifier to 25 K  Figure 3  The  coldhead    probe contains a vacuum chamber at  10      10   Torr to thermally isolate the cryogenic Vacuum chamber   circuit from the outside world  As a result the probe body   sample and the magnet are kept at or near       room temperature  while the probe is at 25 K  Hniena oneee  saptar    w  e  A    WARNING  Even though the sample is relatively well   cold He  E    insulated from the cryogenic temperatures inside the gas supply P  probe  it is imperative that dry air or nitrogen is kept Warm He gas return       flowing through the VT port in the probe  Without VT Vacuum chamber valve       air flow  the sample will cool at a rate of  1  C  per  minute  until it freezes      Figure 3  Diagram of the Cold Probe       e The Intelligent Temperature Controller  o The ITC is located inside the cryobay  and can be seen at the back of the cryobay  It uses a  va
6.  the Tuning VT Assembly of  the Cryo Probe  Right  Photo of the tuning assembly  installed in the probe  with the tuning selector set to    C  and the tuning rod in the    down    position       Rotate the    Function Selector    until the desired channel is visible in the Indicator Window  Cable the Preamp Tune Interface as required  described above    Selected the appropriate channel number on the reflected display meter as usual   Push the brass Tune Match knob up until it seats fully into the slot   Turn the brass knob to tune and minimize the reflected power on the tuning meter    Proton Channel Tuning   e Rotate the Function Selector to 1H T  1H Tune    Push the brass Tune Match knob up until it is fully seated   Adjust the 1H Tune for minimum reflected power   LOWER the Brass Tune Match knob  pull down    Rotate the Function Selector so that 1H M is in the window   Push the brass knob up to engage  and adjust the 1H Match for minimum reflected power   Repeat until the probe is optimally tuned and matched     Note that the tuning dip is extremely sharp with these probes  therefore  the sensitivity of the tuning is  very high  In contrast  the match is very broad and may need to be turned some distance  Typically  proton will tune to a reflected power of  2  and C and N to values   4 10  It takes a patience anda  careful touch to properly tune these probes     e After the probe is tuned  be sure to re cable the system for normal operation     e For tuning the    C and  N
7. Using the Cold Probe on the  Varian Inova 600 NMR Spectrometer    Original material March 2004  by Rich Shoemaker  Dept of Chemistry University of Colorado at Boulder  Modified by David Jones  Univ of Colorado Health Sciences Center  September 2006          WARNING        Read through the whole manual before using the cold probe    Failure to follow the guidelines in this handout could result in severe  damage to the Cold Probe and to your sample     All NMR experiments must be modified to account for the substantial  differences in the technology of cold probes     If you are using the cold probe for the first time  make sure that you check  with the manager that your experiments are acceptable for use on the cold  probe  Failure to do so will require that your group pay for any repair costs  that are incurred due to damage caused by your experiment     If you have any doubts about how to run the cold probe  ASK  it is better to  spend 10 minutes figuring it out  than 3 months waiting for the probe to  come back from the repair shop     Cold Probe Operation on the 600  1  Overview    This manual is intended for users who are experienced at operating Varian NMR spectrometers at a  relatively high level of proficiency  The intent is to point out the issues pertaining to the cold probes   how the use of the cold probe differs from a conventional probe  and the issues that directly affect how  one does spectroscopy with these probes and how to avoid damage to the probe are highl
8. adient amplitudes  can lead to a loss in gradient reproducibility  and a reduction in signal to noise     WARNING  Only pulse sequences that use the    zgradpulse    statement to apply gradients will use  gradient shaping  A number of earlier pulse sequences  particularly those obtained from Dr  Lewis  Kay   s group in Toronto use the    rgradient    parameter to apply gradients  DO NOT USE THESE  experiments on the cold probe  They must be edited to have all the rgradient statements replaced with  the    zgradpulse    statement    Data Acquisition Issues     Your data acquisition with the cold probe is affected by the long ring down times of the Rf pulses  To  deal with this  a new global  flag parameter called qcomp is used in conjunction with oversampling   dsp  1         Make sure that dsp    i   dsp      The parameter qcomp should be set to    y    by typing qcomp  y       The parameter rof2 should be set to 2 4 usec   lf digital signal processing is turned off at any time  dsp  n      then gcomp will also be set to    n      lt must be reset to    y    manually  after dsp  l  is executed    The software will reset the oversampling any time the spectral window  sw  is adjusted    The parameter oversamp indicates the amount of oversampling being used  If oversamp is too  large  acquisition errors can occur    e BEFORE STARTING YOUR ACQUISITION type oversamp  to check the amount of  oversampling being used  There is no real reason for it to be greater than 16     Summary  
9. cold probe assembly to support or steady yourself while  working under the probe  The probe is under a high vacuum  and any disturbance to this  vacuum will cause the probe to shut down and warm up to room temperature     e Atthe probe  the process of tuning is significantly different  Refer to Figure 8 below for the  following description which shows the tuning assembly at the bottom of the probe     e The tuning assembly consists of two parts  the first part is a small brass    tuning knob    which is  used to tune the selected channel     IMPORTANT  the tuning knob slides up and down into the probe to engage the actual tuning rods  inside the probe  The tuning knob MUST BE PULLED DOWN before a different channel can be  selected     Cold Probe Operation on the 600    e Before starting tuning  make sure that the tuning knob is pulled all the way down  Figure 8      e Use the    function selector knob    to select which part of the probe circuit is being tuned  The  window in the side of the function selector shows which part is being tuned  The options are   1H    M Proton Match 1H T Proton Tune  2H Deuterium Tune 13C Carbon Tune 15N Nitrogen Tune    Temperature  controller gas  flange screw    Temperature iar  one of three   controller LN e      connector hg     y   Heater  Function    a Va assembly  selector      knob   Label carrier    Flow cell port  2  plug removed be  gt    Function  Tune   Match    _    i indication  selector knob window    Figure 8  Above  Diagram of
10. e amount of time left until the probe will be ready for us  if it is in cool down mode     B  Sample Insertion and Ejection  The cold probe can accommodate both round and rectangular NMR tubes  Rectangular tubes    are used to reduce noise when the sample contains high concentrations of salt  There are two different  spinners for these probes  The spinner for the rectangular tubes has a rectangular slot as you might  expect  The spinner for the round tubes has a slot cut into the bottom of the spinner  See Figure 1      Cold Probe Operation on the 600    There is a danger of your sample freezing and causing damage to the probe if there is a loss of  temperature regulation if the VT gas flow is turned off   Therefore the probe is equipped with an emergency  sample eject system  As part of this system  the top of  the bore tube on magnet has been modified to include  a    Sample Catcher     This consists of a small brass  toggle switch that will prevent the sample going into the  magnet  In an emergency  the sample is very  forcefully ejected from the probe so that it pops up  above the catcher  and as it drops back down is caught  on the lip of the catcher     To insert your sample  place the spinner into the bore  tube so that it sits on the lip of the sample catcher   Slide the brass nut on the front of the catcher to the     Off    position  Figure5   your sample should slip down  over the catcher and float freely on the stream of eject Figure 5  Sample Catcher for  air  Th
11. ecise temperature by using an Intelligent Temperature    Controller  ITC  to heat super cooled helium gas to a stable temperature of 25 0 Kelvin under  normal operation  When RF power is inserted into the probe  especially when using broadband  decoupling  i e     N and    C  this causes heating of the probe and this is compensated by changing  the power in the heater controlled by the ITC  The heater power  in Watts  is displayed in the  CryoBay Monitor window  As the net RF power in an experiment increases  the heater power will  drop to maintain a steady probe head temperature of 25 0 Kelvin  If too much power is used  such  that lowering the heater power cannot maintain the probe head temperature  then the tuning will be  lost  and ultimately damage to the probe  or the receiver pre amplifier in the probe  could occur     Pulsed Field Gradient Performance   The inductance of the gradient coil in these probes is much higher  Adjustments have been made  to the gradient amplifier to compensate for this  in addition  the software has built in  rudimentary  shaping of the rise fall times of the gradient pulses to further improve the gradient performance   esp  the amplitude stability   A new parameter called    gradientshaping     a global    flag     parameter  is used to activate the shaping of the gradient pulses     WARNING Gradient shaping only works for pulse sequences that use the    zgradpulse    statement to  control gradients  Many early experimental pulse sequences 
12. en slide the switch back to the    On    position  Emergency Eject System  You can then proceed as normal        WARNING     The sample catcher must be in the    On    position if a sample is in the magnet   Serious damage to your sample and the probe may result if the catcher is left in the    Off     position        C  Instrument Probe Cabling TE    P       Cabling for    C     N  and   H is the same as a conventional eres   probe  including the location of the bandpass filters   H From HB Amplifier           cabling is different  Refer to Figure 6  and the following J5303   instructions to ensure that the system is cabled properly  ee   e The cable from the High Band Transmitter plugs into the To Probe  H CY  top port    HB XMTR  J5303       e HBPROBE XMTR  J5331  goes to the    1H    port on the  probe    e HB PROBE PREAMP  J5335  connects to the    1H RCVR     port on the probe     J5331    e OUTPUT  J5302  is connected to the same cable as a Ae PREBE BREEN  normal pre amplifier  Output cable connects to the mixer From Probe Fa  in the magnet leg interface   H Premap ee  Signal path  To HB pa  The transmitter pulse is routed to the probe from port RCVR nd       J5331 on the cryopreamp driver to the probe  This cable   a    fat    cable  also carries the DC voltage for gating the  Transmit Receive Switch  Figure 6  Cryoprobe Preamp Driver    h   a J5304          After pulsing  the NMR signal is amplified in the pre amplifier located inside the probe  and this  signal come
13. evels for    N Decoupling   dowr 2   42  pw 306 us  0 8 KHz field  fora maximum of 120msec    e Safe Levels for  H Spin Locks   TOCSY  spin lock pwr   44   pw  29 4 us  8 5 KHz field  for a maximum of 80 msec  ROESY  spin lock pwr   34  2 6 KHz field  for a maximum of 450msec    e Safe Levels for   H Decoupling   ow and dpwr 3  48 dB  pw 400 us  field strength 0 6 KHz  for a maximum of 60 msec    NOTE     For  C decoupling  you should use WURST 40 decoupling schemes instead of GARP   wherever possible  This will reduce the amount of probe heating     On the CryoBay Monitor  there is a value for    HEATER     This represents the number of watts being  used in the heater to maintain the probe temperature of 25 Kelvin  As more RF power is delivered to  the probe  the HEATER wattage will drop  PAY ATTENTION to this heater power when an experiment  is started  If the heater power drops by more than 1 2 Watts  STOP YOUR ACQUISITION  aa   Adjust  your power levels  spin lock times  acquisition times and relaxation delays to reduce this     Cold Probe Operation on the 600  Using Pulsed Field Gradients     To improve the stability of gradient pulses with the cold probe  all gradients are implemented as shaped  gradients so that the gradient is turned on and off in a smoothed fashion  albeit somewhat crude  This  shaping is turned on for all experiments by setting the parameter gradientshaping  y        lt is recommended that all gradient amplitudes used are less than 18 000  Larger gr
14. ighted      Note  Some of the figures have been used with permission from Varian       Doing NMR spectroscopy with a cold probe is essentially the same as with a conventional probe   however  certain properties of the cold probe must be understood to obtain optimal results  Each of  these will be address in further detail later in this document  The primary  issues that should be understood are        View from  Sample Spinners bottom  The salt tolerant cold probe can accept square tubes as well as round  tubes  To allow this to happen there are new spinners for use on the cold QO    probe  The spinner for square tubes is obvious  the spinner for round  tubes has a slot cut into the bottom  See Figure 1   Use of other spinners  will mean that your sample will not get into the probe     Figure 1  Spinner for use  with cold probe       Rf Pulse Ringdown Times   Due to the high    Q     Quality factor  of these probes  the ring down time of the RF coil after a pulse    is significantly longer  This means that the delay between the end of the pulse  and start of signal  detection is longer  Since pre acquisition delays can adversely effect the appearance of the  spectrum  mostly in frequency dependent phase errors  and other baseline issues   there are new  parameters that deal with this problem  The parameter    qcomp     for Q compensation  is used   and is dependent upon the settings of the    dsp    parameter     Rf Power Handling   The coil in the probe is maintained at a pr
15. riable heater inside the probe to maintain the desired temperature at the coldhead     4    Cold Probe Operation on the 600  3  Preparing for NMR Experiments on the Cold Probe     Doing NMR is not much different than a normal probe  The main issues are associated with how much  power you can use in your experiments for decoupling and spin locking etc     A  Monitoring the Status of the Cold Probe        Before starting an experiment check the status of the probe in the    CryoBay monitor window     Figure  4   Note the value of    Temp    and    Heater    readings in the box outlined in red  You need to know the  values of these settings when the probe is not in use  so that you can judge how much power you are  using in your experiments     I gt  CrunRaw Front Danol WY  2 rri    4     me selector   oo  ak CryoBay Monitor Mass flow  SLPM     VARIAN      Settings    l 2   26 28    Enable ITC  Te m p   2 5   O 4   Cold loop impedance Disable ITC    j na  Timer   Te  C L   1 1 8 Cooling Configuration DAQ Configuration Errors        j Device  1     LabVIEW Errors  CI P   Cooling ee  N Exhaust Ch pre Error 5012 occurred at an    ose rane      n unidentified location   Total Time  Mi Snlet___    F3  H eate r 3   12 saeh Mass Flow Ch massflow Possible reasons     PSI Time  Sec   CCC P1  Psi  Ton Pump Ch Jonpump_  Fault   Vacuum    CCC P2  Psi  IMG Yacuum Ch oe   z Scan Rate 4500   Set Point  k  z   Purge In O   Off Temp  k    NScans  250   pees    Exhaust  psi    181 9 ee Pn Weare 7
16. s back to J5335 on the cryopreamp driver  The signal is further amplified  and sent to  the mixer RCVR via the OUTPUT  J5302  port    Cold Probe Operation on the 600  D  Probe Tuning     For  C     N  and  H channels  the cabling to tune the probe is the same as with a normal probe  The     Probe    connector on the tune interface connects to the appropriate channel on the probe  and the  cable connected to the    Output    port of the BROADBAND Pre Amplifier is connected to the    Output     port of the tune interface     For the  H channel  THE CABLING SETUP FOR TUNING IS DIFFERENT   Figure 7    e Disconnect the thick cable connected to the top port of the cryopreamp driver  J5303    Figure 7 A  and B    e Using a short BNC cable  connect the now empty port J5303  Figure 7B  to the PROBE port of the  tune interface module  Figure 7C   Move the cable from J5202  OUTPUT  to the OUTPUT port of the tune interface   DO NOT REMOVE the cables going to the  H channel of the probe  J5331    The system is now ready to tune the  H channel of the probe        Figure 7 Cabling of 1H Preamp for normal operation and tuning   A  Normal operation  B  After removing  HB transmitter cable and  C  After bridging HB transmitter and    Probe    connector on Tune interface       CAUTION  Extreme care must be exercised when working under the magnet  Be very careful  not to touch the side arm or any of the vacuum connections on the probe  Do not lean on  pull  or otherwise press on any part of the 
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
CABLE TESTER CT100  GTS® et GTS®-DI  Philips Zenia SS071P User's Manual  ISTRUZIONI D`USO  VXF532 Instruction Manual  Tecumseh RKA5512YXA Drawing Data  取扱説明書  Venezuela – Quelle est la situation actuelle ? La partie  TIA-340/ 350S - Russo Power Equipment  American Standard Laundry Faucet 2086 User's Manual    Copyright © All rights reserved. 
   Failed to retrieve file