Melanie 3 – Frequently Asked Questions 1/17 • Does
Contents
1. such files contain data only in gray levels even if they were obtained from a scanner with OD calibration capacity The scanner I m using contains a built in density calibration film to convert gray levels in OD Is it really impossible to use files where gray levels would be transformed in OD Normally there shouldn t be any problem for opening and using such tif files in Melanie Melanie takes into account the conversion tables obtained from a scanner with OD calibration capacity stored in the tif file and should therefore give values that correspond to the ODs However if you have difficulties with the reading or the interpretation of these files you can send them to us melanie support genebio com and we ll look at the problem SPOT DETECTION QUANTIFICATION When I use the auto detect option get SO many spots It is difficult to tell whether they are really spots or not think the computer is picking up any dark areas and considering them to be spots Any suggestions on how to manipulate the system Do you obtain 100 500 2000 5000 spots These numbers may all be normal depending on the type of organism tissue and the size of gel you re studying Of course you have to optimize the spot detection parameters see page 40 in the Melanie 3 User Manual in such a way that you only detect real spots Please note that the Saturation and Peakedness Increase parameters shouldn t normally be changed so there are only 4
2. However since it depends on the spot boundary Area spot detection differences e g use of different detection parameters and errors can considerably influence the results In certain cases the OD may therefore perform better To get an idea you might display the scatter plots using the Vol and the OD and see how well the values correlate between the different gels GEL ALIGNMENT What can I do when align gels using landmarks and the error message reads that the landmarks are poorly distributed This message is usually generated when the landmarks are not well distributed over the entire gel and when the alignment function didn t find a unique solution for interpolation of the landmarks One of the following operations might help to solve the problem v Add one or several landmarks in regions of the gel where no landmarks were defined previously y Select View gt Align Gels gt Other Alignment in the menu Choose the reference gel in the upcoming dialog box In the displayed Align Gels window change the alignment polynomial degree for example to First degree v Move one of the landmarks just slightly When aligning or matching a larger set of gels how does Melanie cope with the situation that a landmark is not present in all of the gels The landmarks do not have to be present in all gels Melanie will only align or match the landmarks that have the same name between the reference gel and the mat
3. Melanie offers so many possibilities If you just want to compare a few gels and you are looking for proteins that might be differently expressed you should use the tools described in Chapter 7 of the Melanie 3 User Manual pages 88 100 v First you can do a visual inspection by displaying the histograms on groups We recommend that you play around with some of the functionalities such as Sorted values Sort histograms by similarity and so on This already allows you to see some tendencies in your analysis v You can then display a report on groups while selecting the Separability and Correspondence Analysis as additional values to be computed see page 96 The Separability option will give the highest gap between two consecutively sorted spot values note that the values in a column can be sorted by clicking in the column header This is useful for splitting the group values in two classes different populations of gels Please note that groups selected as histograms or report entries can be re selected on the gels Melanie 3 Frequently Asked Questions 15 17 v Choose the correspondence analysis icon in the Group Reports toolbar to display a 2 dimensional graph in which similar gels might be clustered together Spots close to those gels are generally characteristic of the gels The Heuristic clustering utility allows you to automatically classify gels and the heuristic clustering report will give the gr
4. also do a calibration of the image capture device using calibration step wedges or calibration strips Such step wedges or strips have known intensity values assigned to it by the manufacturer of the scanning device If you scan such a strip at the same time as your gels you can assign the known intensity values to the different gray scales of the strip as if it were a gel use these values to calibrate the intensity Edit gt Gels gt Calibrate intensity and then use the slope and offset values to pass on the calibration to all gels captured at the same time Edit gt Gels gt Edit calibration e After calibration do the values in the reports and histograms correspond to the raw data or the calibrated data Once a gel has been calibrated using Melanie the program will consider and display the calibrated values in the analyses To get back to the raw data you have to choose the option Edit gt Gels gt Reset Calibration in the menu Melanie 3 Frequently Asked Questions 11 17 Why can I only use the intensity OD for calibration The main goal of the calibration function is to compensate for intensity variations caused by the scanner or by staining differences This is the reason why at least in Melanie 3 only the OD is used for calibration and not the other quantification values such as the volume However intensity based calibration also affects the volume as can be seen from the following equations used for
5. in other gels look if a group exists Since the number of spots and matches are generally different from gel to gel this will give a different number of groups The quality of your reference gel is therefore very important it should be absolutely representative of the gels to be analyzed and contain all spots or at least most of the spots common to the other selected gels In order to prevent getting very different numbers of groups when choosing different reference gels you should carefully check your results at each step of the analysis process and especially at the level of detection and matching Before doing an automatic matching you should verify if spot detection was done correctly and add delete spots where necessary After the automatic matching you should again check your results and look if pairs have to be removed or added PI AND Mw CALIBRATION How can I obtain approximated pl and MW values for the spots in my gel I have a reference gel with pl MW standards The principle of the Pi_Mw calibration in Melanie 3 is as follows First you have to define Pi_Mw annotations for a certain number of spots the more spots the better of course in your reference gel page 54 in the User Manual To do so Melanie 3 Frequently Asked Questions 8 17 you have to open your reference gel for which you know the pl and MW values as any other gel detect your spots and then add Pi_Mw annotations on different spots well dis
6. the calibration The new intensity values are calculated with the formula New_Intensity Gray_Slope Intensity Gray_Offset Therefore the new volume will be New_Volume New_Intensity Area Gray_Slope Intensity Gray_Offset Area Gray_Slope Volume Area Gray_Offset Area Gray_Slope Volume Gray_Offset Area SYNTHETIC GELS Can create a synthetic gel from 3 or more gels run on the same sample repeat the process for several samples and then compare the different synthetic gels It is indeed possible to create synthetic gels This option allows you to merge all spots from 2 or more gels into one synthetic gel image When based on several gels run on the same sample the synthetic gel is in fact an average gel which should contain all the spots that are representative of that sample To create such a synthetic gel you should first match all your gels Ideally you should do a pairwise matching That is if you have 3 gels called 1 2 and 3 the following matches should be performed 1 2 2 3 and 1 3 However in practice this can become very time consuming and the set of gels is therefore often matched to a single reference gel Note that the choice of the reference gel is very important because it will be used as a basis for the creation of the synthetic gel You should therefore choose the gel with the best quality image and the highest spot resolution as the reference gel Once you have matched your gel
7. 2 gels in a general way The superimposition of landmarks will never be perfect just as a regression line doesn t pass perfectly through the points Moreover the alignment doesn t change the relative positions translation of the gels if you moved or zoomed some of your gels for a particular reason you may prefer not to loose that information work In order to reposition your gels so that corresponding regions are superimposed you therefore have to use the double click on the region in question This is entirely normal By using the Shift key at the same time you can carry out this operation on all visible gels Melanie 3 Frequently Asked Questions 6 17 After aligning a group of gels is there a way to conduct a save function so that when this group of gels are reopened together there is no need to re align them Gel alignments imply relationships between different gels A gel is aligned relative to one specific other gel This is the reason why the gel alignments are not saved in the current version of Melanie Indeed the next time you open a gel you may want to align it with another gel So each time you re open your gels you should redo the alignment This shouldn t be much of a problem if you defined a sufficient number of annotations in each of your gels regularly distributed over the entire gel or have saved pairs between the gels You can use the pairs or annotations for aligning your gels and the proce
8. 7 OPENING FILES IMAGE FORMATS In the import function of Melanie 3 users can select TIFF GIF and Biorad SCAN When attempting to import a file from one of the Bio Rad software the Biorad SCAN option is grayed out in the import function Why The reason the Biorad SCAN option is grayed out in the import function is that Melanie 3 doesn t find the file read_scnfile dll in Melanie plugins One possibility is that the file is really absent the other is that you installed an older version of Melanie AFTER Melanie 3 In the latter case the program will look for the directory structure of the latest installed oldest version and will not find the appropriate folders the Help option shouldn t be working either in that case The solution is therefore to reinstall Melanie 3 after uninstalling it first What is the file extension of images that the Biorad SCAN option imports The file extension of the Biorad SCAN images should be gsc Is it possible for Melanie3 to import files from any other analysis programs such as PDQuest retaining spot identification information Melanie can only read the raw PDQuest image from files with the extension gsc with the Import gt Bio rad scan function But no spot identification information is retained No files from other programs can be imported The reason for this is that each program uses its own proprietary file format I understood that when opening tif files in Melanie
9. Melanie 3 Frequently Asked Questions 1 17 INSTALLATION SYSTEM REQUIREMENTS Does Melanie 3 run under Windows 2000 Yes Melanie 3 runs on the Microsoft Windows 95 98 NT 2000 and XP platforms To our knowledge no compatibility problems of Melanie 3 with any of these systems exist However if you encounter any problems please inform us and we will fix them as soon as possible Please note that if you are installing Melanie 3 on a Windows NT 2000 or XP environment you MUST log on as local administrator of the PC before installing the software How can check the exact version release of Melanie I m working with Click on the Help menu in the main Melanie window The exact version number is mentioned after the About Melanie item The Melanie Technical Support team will ask for this number when you report particular problems or bugs with your version SCANNING When I scan my gels at 600 dpi their file size is around 25 MB I import the gels in Melanie with a reduction factor of 4 which reduces the size to 2 5 MB Does this affect the quantitative analysis if use the same factor for all gels What do you recommend Can I use a higher reduction factor Your reduction factor of 4 seems a good choice It allows you to obtain reasonably sized files and should normally do without a significant loss of analysis quality Using a slightly higher reduction factor might not be recommended 2 5 MB is a very reasonable file si
10. by choosing File gt Import gt Calibration select the saved fitting report Each selected gel is now calibrated To check this you can display the new scatter plots All slopes should be close to 1 and all offsets be close to 0 You should be very careful when using this method because if there are a lot of proteins that are differently expressed you might introduce errors and get identical values for corresponding spots that are in fact up or down regulated Such a calibration should essentially be restricted to gels that belong to the same population or that represent the same sample v Another possibility is to use calibration annotations This works in the following way You pick some spots or pixels that are present in all of your gels and that should be unaffected by the different conditions you use in terms of protein expression Note that one should try to cover the entire raw value range You define calibration annotations for the reference gel you have to give values that correspond to optical densities for example or counts or other real world values When you select the other gels and choose Edit gt Gels gt Calibrate intensity Melanie will first ask you the name of the label category that contains the calibration values and then if you want to use the calibration values from the reference gel If you say yes the calibration values from the reference gel are propagated to your gel via the paired spots v You can
11. ched or aligned gel There is no problem if you have one or several landmarks missing in a gel Melanie will just not take these into account when aligning or matching In cases where the spots for the landmarks have been lost because of technical problems sometimes happens to border spots it is better not to add the landmark than putting it in a hypothetical spot position Melanie 3 Frequently Asked Questions 5 17 I would like to match two gels but as the spots do not superimpose exactly I used some landmarks to align the gels Still in some regions of the gels spots that should align because they are the same do not align perfectly How can match my gels It is very important to understand that the alignment is nothing else than a visual deformation which doesn t modify the underlying data of a gel that aims to minimize the distances between corresponding landmarks in 2 gels in a general way The superimposition of landmarks will never be perfect just as a regression line doesn t pass perfectly through the points Aligning your gels is just a visual method that can help you find the corresponding spots in two gels In the current version of Melanie Melanie 3 it can also help the matching process in difficult cases But it is not necessary to align the gels perfectly before doing the matching You could just put a few correctly distributed landmarks on your gels and immediately do the matching without aligning Or
12. d MW values on landmarked protein spots The pl and MW values for all protein spots in the pattern are then calculated The pl and MW values that entered for the landmarks do not stay the same in the spot report How can possible disagreement between input and output pl and MW values be corrected Normally the pl and MW values in the spot report should be the same as in the landmarks However this is only the case if you have attached your landmarks to actual spots and not just to pixel positions in the gel If the landmark is attached to a pixel the pl and MW values for the spot that lies close to it will be slightly different from that of the pixel to which the landmark was attached To link a landmark annotation to a spot select the annotation and then choose Edit gt Annotation gt Link with Spot in the menu You may also move the annotation onto the spot Select the annotation and drag it to the desired location You will then be asked if you want to link the annotation with the spot NORMALIZATION INTENSITY CALIBRATION BACKGROUND SUBTRACTION Upon group analysis observed that the intensity ratios of spots between gels didn t match my visual estimates This seems to be due to failure to correct for background intensity I m figuring out how to perform this correction Should use calibration annotations First of all you should be careful when making visual estimations Image analysis programs have been designed to prev
13. database available on the Internet using Melanie gels More information can be found on the following site http Awww expasy org ch2d make2ddb html ERROR MESSAGES Beware something strange has occurred This message appears when there is some kind of internal system error an error that has not been foreseen Normally you shouldn t get such messages If you can reproduce them you should definitely send us melanie support genebio com a detailed description of the problem as well as a list of the different operations you carry out before getting the error message so that we can try to understand the problem and find a solution Bad image for the gel This indicates that there is something with the TIFF file This may just be a tag that is not recognized or be something more serious which prevents you from opening the file When you get this message you should send us melanie support genebio com one of the problematic TIFF files so that we can look into it and try to find a solution
14. e been matched you can select groups spots that are representative of your sample Spots that are only present in one of the five gels for example are not representative of the sample and could be excluded from the analysis You can select groups by the number of spots they contain For example you can select all groups that have at least 4 or 5 spots that is select spots that are present in at least 4 or 5 gels To do so select all groups Select gt Groups gt All choose Select gt Groups gt Restrict Selection in the menu and select the option Keep the groups whose number of selected spots is gt 4 or 5 This type of filtering is described on page 87 of the User Manual Next you can choose Select gt Spots gt Inverse Selection to obtain all spots that don t satisfy this criterion Please note that you should check your results carefully since detection or matching problems may cause the exclusion of spots that are in fact present in all gels or include spots that are only present in one or two gels but wrongly matched with other spots Also when selecting groups the program only takes into account the spots that are present in the reference gel Any spot absent in the reference gel won t be considered even if it is present in all other gels e How do get Melanie to display potential differences among 2 DE gels It is difficult to give an easy and complete answer to this question because
15. e groups that contain a minimum number of spots for example by selecting groups that have the corresponding spot in at least 4 out of the 5 gels see page 87 in the User Manual Please note that you should check your results carefully since detection or matching problems may cause the exclusion of spots that are in fact present in all gels or include spots that are only present in one or two gels but wrongly matched with other spots In addition when selecting groups the program only takes into account the spots that are present in the reference gel Any spot absent in the reference gel won t be considered even if present in all other gels After you made sure that all gels and spots you would like to include are highlighted green select Tools gt Create Synthetic Gel in the menu bar select the reference gel from the upcoming dialog box and specify a file name for the synthetic gel in the Create Synthetic Gel window The resulting synthetic gel that contains all selected spots will be created and displayed If spots belong to a group only one spot per group will be present in the synthetic gel If you select 3 gels or more you will be asked whether you would like to keep only triple pairs In this case only spots that form a triangle of pairs will be included in the synthetic gel This means a spot S on gel 1 has to match a spot S on gel 2 and spot S on gel 3 The spot S on gel 2 also has to match spot S on gel 3 This t
16. ent Filter the background and detect spots in all gels Define landmark annotations pick a reference gel align and match all the images to this reference Merge all the images into one synthetic gel Set the synthetic image as a new reference align and match all gels to it Assign each gel to a disease class disease A B or C Analyze the reports Classes Statistical tests and the histograms on Classes Groups looking at the Vol and detect interesting spots or groups AAA Na Can you tell me if this scheme is appropriate The general principle of the above scheme is indeed appropriate We would just like to add a few remarks and suggestions We will describe a rather ideal way of doing things but you can of course make more approximations and simplify some things in order to speed up your analysis First of all making a single synthetic gel to which you can match all the others is a very good idea However you might proceed a little differently y You could first make a reference gel for each disease class That is a synthetic gel for disease class A a synthetic gel for disease class B anda synthetic gel for disease class C In each of these gels you should filter out the noise You can do this by using criteria for the inclusion of spots in your synthetic gel You could for example only include spots that are present in at least 10 out of the 20 gels that you have in a class see page 87 in the Melanie 3 User Manua
17. ent forms It is not just a cosmetic function e Could you please explain the meaning of the various parameters of the spot detection window If you want to know what the different parameters represent you should take a look at the explanations on page 40 in the Melanie User Manual If you are interested in the mathematical background you can find more information in the following article Melanie II a third generation software package for analysis of two dimensional electrophoresis images Il Algorithms Appel R D 1997 Electrophoresis 18 2735 2748 Please note that spot detection has not changed from Melanie II to Melanie 3 But it will be improved in the forthcoming version Melanie 4 and will require less parameters to be set e Ihave some performance problems with Melanie 3 In particular spot detection and saving of the spots is very slow more than a minute for detection and about 2 minutes for saving spots of a 2 3 MB gel ona Pentium Ill Is this normal No it is not On a Pentium III 700 MHz with 256 MB RAM about 2 to 4 seconds are needed for detection depending on whether image data was loaded or not and half a second for saving spots for a gel that s approximately 2 3 MB in size Therefore if spot detection and saving are slow you should check the following v Are the gels on a local hard disk on the same PC than the one Melanie is running on or are they on a networked disk Disk access could play a
18. ent this But it might indeed be necessary to correct your gels for background intensity However no calibration annotations are needed You can use the Filter Background option in the Tool gt Gels gt Filter Image menu and select one of the background subtraction methods in the upcoming window Since the program will create a copy of the gel after filtering you are asked to give the filtered gel a new name Note that the original image remains unaffected Is there a way to normalize gels Melanie not only calculates the OD Area and Vol for spots that have been detected It also calculates the normalized values OD OD divided by the total OD over the whole image and Vol Vol divided by the total Vol over the whole image The latter takes into account differences due to staining for example at least for gels with very similar spot patterns All spot quantification values are automatically included in the Spots Report select spots of interest and then choose Reports gt Spots Report gt Current The OD and Vol can also be considered for the statistics on groups I want to calibrate a set of gels in order to correct the pixel intensities for scanning protein loading and staining differences How to proceed There are different ways to calibrate gels Please note that calibrating is not always necessary If you look at OD or Vol values you already normalize your spot values with respect to the total OD or Vol of you
19. even try to match the gels without putting any landmarks Obviously the spots have to be detected first You also have to be sure to only put landmarks on spots for which you are 100 sure that they should be matched otherwise you may introduce matching errors Once you have matched the gels you can re do an alignment based on the pairs that have been found In this case the visual overlap of the gels may be better although this is not always the case In stacked mode pairs can be visualized by displaying the pair vectors View gt Stack gt Show pairs How can I after successful alignment of the gels retrieve a list of all spots with their original X and Y coordinates before alignment in addition to the new X and Y positions after alignment You should be aware that the gel alignment function is a purely visual tool This means that the gel is visually deformed in order to better superimpose the landmarks or pairs with those in the reference gel But the numerical values are not influenced by the alignment I think there is a problem with the alignment function Sometimes when aligning gels the landmarks are not correctly superimposed In those cases only a double click allows correct alignment It is very important to understand that the alignment is nothing else than a visual deformation which doesn t modify the underlying data of a gel that aims to minimize the distances between corresponding landmarks in
20. ind the representative spots in a group Choose the gel that has most of these representative spots y Once this is done you can match all gels in class A with the latter gel and display the scatter plots with OD values of all gels in the population against this gel In the Scatter plot window select the fitting report from the report icon and save this report filename txt y Select all gels if they got deselected in the meantime and choose File gt Import gt Calibration If you now display again the new scatter plots the fitting lines should have the form 1 0 x 0 or something very close to it So the overall intensities of the gels of the Disease A class should now be comparable y Be careful you should only use this type of calibration to normalize your spot intensities for gels that belong to the same population EXPORT Can Melanie 3 export coordinates to spot excision robots Yes Melanie 3 allows the export of spot coordinates to the Amersham Biosciences Ettan spot picker and the Genetix GelPix spot picker Moreover in principle the spot coordinates generated by Melanie can be used with any robot that accepts input files with coordinates am looking for a solution to put the data generated by 2 D gel analysis software on the Internet Is there a software associated with built into Melanie that can make the data stored with Melanie available on the Internet Yes with Make2ddb free you can make your own
21. l v Once you have done this you can make one final synthetic gel master gel out of these three synthetic gels This time you should include ALL the spots that are present in any of the three gels v Finally you can match all gels to this master gel which will act as a kind of spot position index reference and do a statistical analysis with the tools available in Melanie Chapter 8 pages 101 110 in the User Manual Be careful you need this master gel for matching your other gels to it but you shouldn t normally include it in one of your classes during the analysis Second within a class you could calibrate the spot intensities in your gels with respect to one particular gel in the class To do this we would suggest the following taking the gels of disease A as an example y Select the best quality gel in disease class A Match all the other gels to it and display the scatter plots with OD values If the data show high correlations and if all the fitting lines are close to identity 1 0 x 0 then it might not be necessary to calibrate your gels v If this is not the case then by looking at the values in the groups report try to find out which of the gels is most representative of disease class A in terms of spot values which of the gel has most spots with intensities close to the mean for the class You can use the procedures described on page 98 in the Melanie 3 Frequently Asked Questions 17 17 Manual to f
22. lf What is important and what takes time is finding the anchor points So when you get the message not enough input pairs this does not necessarily mean that you didn t define enough input pairs It might just mean that the program didn t find enough anchor points to continue the matching process This can be the case for very dissimilar or distorted gels In these cases it is very important to align your gels first careful the reference gel for alignment and matching should be identical Indeed although the alignment is optional it can speed up or improve the matching step for distorted or very dissimilar gels REFERENCE GEL GROUPS Why while working with the same set of gels do obtain a different number of groups each time I choose a different reference gel The reference gel is the central node of spot groups All spots in selected gels that are paired with the same spot in the reference gel form so called groups an essential concept in the Melanie software This means that when looking for all existing groups the program only takes into account the spots that are present in the reference gel Any spot absent in the reference gel will not be considered even if present in all other gels Therefore the use of different reference gels may give different analysis results For example when you choose a new reference gel the program will only consider the spots in this new reference gel and look if they are matched with spots
23. mel have associated pl and MW values because these gels were matched with the reference gel e havea question about the pl and MW calibration functions in Melanie 3 It is not clear to me exactly how the calculations are being performed to determine the pl and MW of all proteins in a 2D gel pattern and I would like to know more about the algorithms used for the calibration The algorithm does the following In case of the pl it looks up the two closest landmarks at the left and at the right of the spot for which the pl has to be determined and it then interpolates between these 2 points Since the program doesn t have any information about the experimental non linear pl scale the calculated values can only be approximate In case of the MW of the spots the procedure is very similar except that the program searches for the closest spots above and below the spot for which the MW has to be determined and that it makes a logarithmic interpolation the MW scale is logarithmic Extrapolating pl and MW values is a bit more complicated For example if the pl of a spot on the very right side of your gel has to be determined the program looks for the two closest spots on the left of the spot in question Provided that these two spots are sufficiently distant from each other in order to decrease the error the value for the spot in question can then be extrapolated Melanie 3 Frequently Asked Questions 9 17 I have input several pl an
24. n important role during detection especially when image data has to be loaded When saving the spot data the only operation is the disk access Since this really takes a lot of time at your location there might be a disk access or network problem Melanie 3 Frequently Asked Questions 4 17 v Make sure you do not saturate your memory by using loaded images or too high image depth i e true color This could be a problem In order to make the proper calculations and to compare our results with previously interpreted values we need to know which one of Melanie s quantification measures i e Vol OD Area etc or combination of measurements can best be used to describe the amount of protein in a spot This question is delicate to answer since everything depends on what you really want to do and on the type of problem you re treating If you want to compare your values with previously interpreted values you should employ the quantification measure used in the past If you use Melanie to re evaluate any previously obtained gels you should consider the following The OD and Vol are probably best because they normalize the spot values with respect to the overall intensities or volumes respectively in the gels This is particularly important if you want to compare spot values from different gels As to the choice between the two the Vol should normally be better because it represents the total amount of protein in a spot
25. oup IDs of the spots that are characteristic different between each class population of gels for each class All these tools are complementary and should be used together Once you know that there are 2 or more populations either by using the tools mentioned above or based on the experimental setup you can define classes and use the tools described in Chapter 8 of the User Manual pages 101 110 v v v Again the histograms on classes and histograms on classes groups options are very useful for this type of analysis The classes reports allow classifying the groups according to their overlapping value it measures the difference between the values in two classes Finally we recommend using the statistical tests described on page 108 They are quite powerful and indicate which differences are most significant e I m comparing two gels and would like to find the protein spots that are over or under expressed by more than 2 fold in one gel with respect to the other gel How to proceed 1 2 ae I 99 First match your gels see Chapter 6 page 79 in the Melanie 3 User Manual Select all groups by choosing Select gt Groups gt All from the menu you will be asked to select one gel as the reference gel All spots that are present in both gels will be selected green Display a groups report Analyze gt Groups Report Choose the value type to be displayed Vol for example Choose the
26. parameters left to be modified Optimizing these for your particular set of gels may need a little time but it is really worth it in order to get good and reproducible results If you have some artifacts left you can edit Melanie 3 Frequently Asked Questions 3 17 or delete these manually You can also do a spot filtering Choose Select gt Spots gt Restrict by Value in the menu page 47 in the User Manual in order to delete spots that are too small or too large in terms of Area OD Volume etc e would like to know when and why I should use the Fit with Gaussians option for spot detection Some people consider that spot contours and thus spot quantitation should be approximated by Gaussian shapes Some programs only provide this possibility We on the other hand think that detecting the shapes as they are using the Laplacian default in Melanie is more adequate However we want to leave the choice to the user We know that most of our customers use the default option and that it gives very good results It is up to you to decide which of the 2 detection options you find most adequate e If spots are fitted by Gaussians does this influence the quantification of the spots or is it rather cosmetics just to have a nicer looking interpretation of what a spot is supposed to be Spots fitted by Gaussians have different quantification values described on page 44 in the Melanie 3 User Manual since they have differ
27. r Manual and only include these groups in the synthetic gel Therefore you should check if you correctly selected all interesting spots and groups in all gels before creating the synthetic gel Melanie 3 Frequently Asked Questions 13 17 REPORTS When several gels have been matched with a reference gel how can I prepare a report list that includes all spots of the reference gel and all matching spots on the other gels Any Groups Report will only contain the matched spots and therefore unmatched peaks won t show up This isn t the case if you display a Spot Report which will contain all selected spots matched or not The trick is therefore to make the adequate spot selection first you should select all groups in all gels and then select all spots only in the reference gel Then re select all your gels and display the Spot Report The latter will contain both the matched spots in all the gels and the non matched spots in the reference gel You can customize your Spot Report by including the desired fields Spot ID spot value types pixel coordinates etc both for the gel and reference gel If you finally sort your data according to the R SpotID i e the GroupID you will get a nice report where all matched spots are grouped together l obtain strange results while generating spot reports the volume calculations don t seem correct since the same spot doesn t give the same results in two reports one on all spots the o
28. r gel This may be largely sufficient to get good results when comparing gels with a very similar spot pattern Anyway the major possibilities for calibration are v You display scatter plots first between your gels and the reference gel taking into account all the groups and choosing the OD as value type as explained Melanie 3 Frequently Asked Questions 10 17 on page 88 in the Melanie User Manual Such a scatter plot can give an idea on the relationship between the spot values from two gels by searching the linear dependence between the spot values from one gel and the values taken from another gel In most cases only relatively few protein expression changes are observed between gels only a limited number of proteins is differently expressed due to different experimental conditions or due to disease so most of the values in the two gels should be similar and therefore the fitting line should be close to identity If your slope is very close to 1 and your offset close to 0 when you look at OD values a value of 50 might be very low if maximum ODs in your gels correspond to 2000 for example this is only 2 5 then calibrating makes no sense If this is not the case you can do a Calibration of your gel In the scatter plot window obtained with OD as value type choose the Fitting report option in the Report icon and save this fitting report text file Then select the gels to be calibrated and import a calibration
29. riangle is illustrated on page 120 of the User Manual Please note that to use the triple pairs option the gels have to be matched pairwise before creating the synthetic gel More information about synthetic gels can be found in Chapter 10 of the User Manual page 119 Synthetic gels can be matched against each other or against any other gel in the usual way see description in Chapter 6 of the User Manual page 79 e After a synthetic gel has been created can more spots be added to it Yes To do so use the menu item Edit gt Spots gt Copy from Gel However these spots will be isolated ones neither matched nor grouped e have problems when creating a synthetic master gel from 3 synthetic gels each one of the latter being representative of about 30 gels A certain number of spots although correctly visualized detected and matched are not present any more in the final synthetic gel that should be a standard for the quality control There is only one reason why certain spots have not been included in your final master gel these spots were not selected when creating this master gel Indeed the user has to select the spots that he wants to add to the synthetic gel This allows the use of specific criteria for inclusion For example when 10 gels were matched for the creation of a synthetic gel one could select all groups that have the corresponding spot in at least 8 out of the 10 gels see page 87 in the Melanie 3 Use
30. s you should select all the spots and gels that you would like to include in the synthetic gel Gels can be selected by clicking on their selection banner on top of the gel To select more than one gel you can hold the Ctrl key and click on the banner of the additional gels If you want to select all gels in the window you can also choose Select gt Gels gt All in the menu To select all spots in the highlighted gels choose Select gt Spots gt All Other selection criteria for spots are also available see Chapter 3 of the User Manual page 46 Selected gels and spots are highlighted in green You can also first select spots in each individual gel This allows you to use different selection criteria for each gel This option is useful if you want to remove some spots or artifacts from a particular gel To do so highlight only one gel and select the spots of interest see also Chapter 3 of the User Manual page 46 After you have selected the spots for each gel you can select all gels to be used for the synthetic gel as described before Melanie 3 Frequently Asked Questions 12 17 When creating an average gel that should only contain spots representative of the sample spots present in all gels run on the same sample or almost all the gels it is useful to select only groups choose Select gt Groups gt All Thus you will eliminate isolated spots that are just artifacts You can also restrict your selection to th
31. ss is almost immediate GEL MATCHING I tried to auto match gels have downloaded from the ExPASy web server It is unsuccessful because the auto match entry does not seem enabled To use the auto match function Edit gt Pairs gt Auto Match Gel make sure that you have selected the gels you would like to match A first gel can be selected by clicking on its selection banner on top of the gel The gel will be highlighted in green indicating that it has been selected To select the next gel hold down the Shift or Ctrl key and click on the banner of the additional gel Now both gels should be highlighted in green and the auto match function should be enabled The only prerequisite for matching the gels is that the spots have already been detected Edit gt Spots gt Auto detect However for the master gels downloaded from the ExPASy server this should not be necessary Please note that the gels you would like to match should be in the same directory on your PC For getting familiar with the Melanie software it might be easier to use the sample gels that come with the program Program Files Melanie Trial Gels_Ecoli The Master Gels found on the ExPASy server are quite dissimilar to each other and it can therefore be difficult to match them automatically I would like to compare my gels with those found on the ExPASy server However the gel sizes are quite different Is it possible to change the size of the imported gels so that
32. staining and scanning artifacts and non representative spots Once you have detected spots you can do a first type of filtering that consists of selecting and possibly eliminating spots that do not satisfy certain criteria of size volume and so on To apply such a spot filtering select all gels Select gt Gels gt All and all spots Select gt Spots gt All then choose Select gt Spots gt Restrict Melanie 3 Frequently Asked Questions 14 17 by Value You will be asked for the value type Vol Area etc you would like to use for filtering and the thresholds to be employed Only spots verifying the conditions e g with an area smaller than 0 3 mm will be kept You can decide to delete these spots or just exclude them from your analysis A second filter could be applied at the group level In this case you should first match your gels That is you should find the pairs between corresponding spots in the different gels Detailed descriptions about the matching procedure can be found on pages 79 88 in the Melanie User Manual If you have very similar gels you can try to do a completely automatic matching by selecting your 5 gels picking a reference gel the best quality gel and choosing Edit gt Pairs gt Auto Match Gels in the menu If you don t find pairs it might be necessary to define a few landmark annotations and align your gels prior to matching see the Manual Once your gels hav
33. statistics mean 100 and M S D 100 as well as the additional values especially the Range ratio Melanie will display the Groups Report different statistical values for each of the groups see pages 90 99 in which you can sort your data see page 74 in the Manual In your case the most interesting column is the one named Ratio This Range Ratio is the maximum value divided by the smallest value in the sample specified by the percentage slider Since you only have 2 gels this is the ratio between the protein quantities in the two gels Click twice in the label of the Ratio column to sort the values in descending order Now all the spots that have a Range Ratio gt 2 are differently expressed by more than 2 times in one of the gels You can select these spots lines of the table by clicking on them and visualize the corresponding spots on the gels by using the navigation tools described in the User Manual pages 72 73 You can also display the histograms for the spots with Range Ratio gt 2 To do this select the corresponding lines in your Groups Report and click on the histogram icon in the Groups Report window Melanie 3 Frequently Asked Questions 16 17 I would like to use Melanie 3 for a rather large study aimed at identifying disease related proteins To do so have to compare a set of 60 gels prepared from patient samples with three different diseases 20 gels per disease have planned the following experim
34. their dimensions will be the same as my gels First of all it is important to know that you can compare gels of different sizes You should just define a sufficient number of landmarks put identical landmark annotations on corresponding spots in the different gels see pages 53 55 in the Melanie 3 User Manual and align the gels so that they visually overlap pages 31 32 You can then match pages 79 84 and compare the gels You can also reduce or enlarge a gel by choosing Tools gt Gels gt Scale by from the menu When matching the gels you might get the warning The pixel widths differ but you can ignore this message What can I do when getting the error message not enough input pairs while matching already matched 70 landmark annotations in my gels When Melanie 3 does a matching it will look for spots that can be used as anchors That is it looks for spots with sufficient intensity and that have an environment that is very similar to the one for a corresponding spot in the other Melanie 3 Frequently Asked Questions 7 17 gel The program needs to find a certain number of these anchor points before it can go on matching the rest of the pairs When you define landmarks and use them for the matching Melanie will first try to use these as anchor points but if the program considers that these points are ambiguous have ambiguous environments it will just ignore them and try to find anchor points by itse
35. ther on a few spots only Be careful you should be aware of the fact that each report is an instantaneous view of the gel and that at each modification adding splitting suppressing modifying spots you should display a new report in order to get the updated values Also if you suppress or add spots in Melanie 3 all the spots are renumbered and so a new report has to be created because the old one is not automatically updated This means that you should always do any spot editing filtering or matching before starting your final data analysis Please note the problem with the changing spot IDs will be solved in the forthcoming Melanie 4 In the next version of the program all spots will have a unique identifier which won t change when the spot is edited the number of a suppressed spot won t be used again and each new spot will get a new number ANALYSIS How can I get statistics such as median etc for a group You have to select your groups see the options in the Select gt Groups menu and page 87 in the Melanie 3 User Manual and display a groups report Analyze gt Groups Report When you do this you re asked for the type of statistics mean midrange etc A complete description can be found in the User Manual from page 90 onwards I have run 5 gels on the same sample and already did spot detection using the same parameters in all gels How can I simultaneously clean up my 5 gels that is filter out all
36. tributed over the whole gel by y Choosing the annotation tool v Clicking on a spot for which you know the pl and MW values y Selecting the Pi_Mw category in the Create Annotation window v Entering the known pl and MW values separated by a space a value of 1 means that no value is set y Clicking OK Do this for a sufficient number of spots Subsequently you have to match your gel to this reference gel That is you have to find the pairs between the corresponding spots in your 2 gels Melanie can do this Semi automatically page 79 and further in the Melanie 3 User Manual Nevertheless you may speed up and improve the matching process by defining a few landmark annotations pages 53 55 and then automatically pairing the identical annotations pages 79 81 This can be followed by a gel alignment visual deformation of one of the images in order to best superimpose the landmarks pages 31 32 and finally the automatic matching page 80 Once this is done the calculated pl and MW values for all spots in the gel matched to the reference gel are automatically available from the spot report pages 49 50 in the User Manual and pl and Mw grid lines can be displayed over the gel page 31 in the User Manual You can take a look at the example gels in the Program Files GeneBio Melanie 3 Analyzed_Ecoli folder The reference gel ECOLI mel contains annotations from the Pi_Mw category and the spots from gels 94 0002 mel and 94 0005
37. ze but everything depends of course on the physical size and resolution of your gels The final resolution at which you work 600 4 150 dpi in your case should clearly be higher than the spot resolution Also keep in mind that we recommend an image depth of at least 12 16 bit should be standard for most scanners currently in use How can I check the image resolution of my gels What is the ideal resolution for scanning my images The image resolution of your gels is given by the PixWidth and PixHeight These values can be found in the Gels Report Report gt Gels Report and are given in micrometers So a PixWidth of 200 in the Gels Report specifies that each pixel has a width of 0 200 mm This also means that you have 5 pixels per mm or 50 pixels per cm or 127 pixels per inch For the scan resolution a lot depends on the quality of the gels and the detection level required desired The larger the scan resolution the better the image However the larger the scan resolution the larger the image which means that you can quickly slow down any kind of manipulation analysis of the gel A good compromise is about 200 dpi as this gives good quality images without resulting in huge image files Most scanners that are now available scan way above this level and you can adjust the resolution when you actually scan the image Also keep in mind that the image depth has to be sufficient 12 16 bit Melanie 3 Frequently Asked Questions 2 1