Home
        HumaStar 300
         Contents
1.        Diagnostics Worldwide    5 3 4 Testing Procedure   The Module has to be calibrated daily together with the other components    1  Cal  B    approx  400 ul in a sample cup should be placed on the Calibration Rack   2  Place daily also 400 ul of Clean S solution in its position     3  Select to Calibrate the ISE  it requires about 2 min     a  Clean S solution will be aspirated to clean and rinse the electrodes   b  Followed by purging several times the electrodes with Cal  A  reagent     c  It calibrates the electrodes with Cal A     the first calibration point   d  It aspirates the Cal B twice to calibrate the second point   e  The Analyzer checks that the 2 calibration points are correct according to the preset values in the system       an OK will be displayed    f  Should the calibration fail for one of the electrolytes  there will be displayed a message  The calibration  has to be repeated    g  For further information see the Maintenance Program    h  If Quality Control is desired  include a Serum control during the Calibration     Note  If the calibration of one of the parameters failed  the system will print a     next to this parameter  The  determination of all the tests will proceed  the parameter that failed the calibration and if not repeated will use  the data saved from the previous calibration     5 3 5 Urine Tests    If there are Urine samples to be analyzed  these have to be first pre diluted 1 10 with its proper diluent  Use the  Prepper  In any case 
2.      ISEo      ISE Out Range      RPL      Error in the triplicate Standard Calibration        4 8 2 Sample Pre dilution   In routine testing a 3 ul sample is normally used  Even in the case of Automatic Repetition of Result a 1 5 ul is used  and gives perfectly acceptable results    However  automatic Pre dilution of sample can be used for any method that may require it  or in the case of small  volumes or pediatric samples  that may need it  The sample Pre dilution should be programmed in the Method and  executed before the start of routine testing    The sample Pre dilution should be done on both the Samples and their Standards and Calibrators  When doing Pre   dilution  make sure that for every sample there is an empty pre dilution cup    Make sure that there is a full bottle with pre dilution liquid under the Sampling Arm  Make sure to select the  Dilution Ratio before Start  There is only one pre dilution Ratio possible with each run  If there are more then ONE  pre dilution ratio necessary in a Worklist  the other tests will have to be done on a separate run    Urine for ISE     remember that all Urine samples have to pre diluted 1 10 prior to testing    In case the analyzer is stopped and new samples have been added that require pre dilution  these will be  automatically pre diluted when START is given     Human  I A        Diagnostics Worldwide  57    58 Human HumaStar 300 User Manual    5 Bar Code Reader  Ise Module  Data Exchange    Both the Bar Code Reader as well
3.     8  Clear Page     to delete all results without saving   9  Save and Clear all Samples   10  Repeat all Samples     that have been programmed    11  Clear all Samples     delete all samples without saving    2 7 _ Navigation  The following commands are available     Human  L   zeg    Diagnostics Worldwide  23    Ele Edit Graph   Navigation 2                      y Previous page     Newt Page Samples  Previous Section ao ae  3 in ie 2  2  Close Register 23  2  A 24  2   de 25   2   26   21    2 7 1    24 Human HumaStar 300 User Manual    2 8 Help    The following commands are available     File Edt Graph Navigation   2  Help   Tell me How     Help   What s this     About       Sar    Pos    21  22  23  24       Di    26        2 8 1    2 9 Analyzer Status  In part 3 ofthe Main Menu there are 3 icons that identify the Analyzer Status       1  START     when this icon is green   the analyzer is READY to execute a command  To START click    onthe green triangle  When the triangle is gray     the analyzer is in operation and cannot accept  a new command     2  STOP Intelligent     Enables to STOP the analyzer during operation  without disturbing the  E operational cycle     There are two colours possible   RED     analyzer is operating and cannot accept other commands  GREY     analyzer cannot accept any command     3  Alarm   When illuminated indicates the presence of an ERROR or FLAG  Click on the ICON to  A open a window to display the error message     4  Arrow close to th
4.    To run a check  click on its CHECK BOX  e     then click START   The analyzer will perform the requested operation  an    OK    will be displayed and the time of operation   Wait for the green light START to be ON  before requesting the next operation        ISE MODULE CHECK  In case of difficulty during Calibration  before repeating  try the suggested procedure below     1  Run twice the ISE Clean cycle    fill the sample cup with the Clean solution and place it into its position    2  Run 3 times the ISE Purge cycle     Then repeat the Calibration  If the problem persists     call technical assistance  For additional details see description  in Chapter 5     72 Human HumasStar 300 User Manual    6 3 Measurement System  From the Main Menu click on    F8        MAINTENANCE   All Auto checks are indicated OK as shown below     Ele Edt Graph Navigation 2    Measurement System    Set Acton    Time    Instrument On Line 11 20    Int Hardware   Fil Hydraulics  Waste Tank Level  Wath Tank Level  Ine  Liquid Level  Test Level Bath  Drain Bath   Fil Incub  Bath  Wash All Cuvettes  Clean All Cuvettes  Read Temper     C   Read Al Fitters  Read 340 nm      Read 380 nm      Read 405 nm  2   Read 510 nm      Read 546 nm  Si  Read 578 nm           11 20  11 20  11 20  11 20  11 20  11 20    Ready A E    11 49 42 AM  AutoCheck OK        6 3 1  Measurement System    How to use the above program    To run a check  click on its CHECK BOX  e     then click START     The analyzer will perform
5.   Click F6 and the Book will open to all available programmed Standards and Calibrators    The standards are arranged in alphabetical order on the right  Each programmed standard or Calibrator has its own  page which displays the name of the standard   its concentration   the absorbance measured or its calculated  Factor and the date of its last calibration  There are no limits of the number of Standards that can be programmed   To Program a new Standard   click on Edit and again on  Insert Page  to open a window  Introduce the name of the  new Standard and click OK  A new page will be added for the Standard book  Click on the first line with the asterisk  to open the opposite page displaying all the available programmed methods  Click the check box of the method  corresponding to that Standard and click again on  Reference  line below to introduce the concentration value for  that standard     The same standard solution may be used for more than one method  A multi Calibrator with all the substrates can  be programmed the same way  Introduce the concentrations for each of the methods one at a time  To change the  concentration value of a standard  highlight it and change its concentration value on the Reference line    To Delete a standard   simply highlight it and press Delete on the keyboard  A confirmation window will open   choose Yes or No accordingly     TO PROGRAM A MULTI   STANDARD OR CALIBRATOR   make sure to open a separate page for each concentration  level  as de
6.   Onthe monitor all 40 positions are coloured White     free positions  8  Inside the Figure are displayed the Free positions in White and the others as follows     Yellow Position Accepted  Green Sample Identified   Red Sample NOT     Identified  White Free Position       Es Aeoistere Dion 7   162817    6009900     D                   A    O  O en 10  10   Simpie NOT ienned    oO  Re        a C  Actepted Poskions       41 2016   4  E     90656009            5 1 2 3  To Assign Sample Positions    60 Human HumaStar 300 User Manual    Sample not identified  maybe dueto    Missing Sample    Bar Code not readable    No Bar Code present    All identified samples   their positions become GREEN     9  Click on    Accepted Posit     To transfer its work list and the position becomes Yellow     position accepted     10  The non identified Samples become    Not assigned     These can be assigned manually by    Drag and Drop    or  click on    Assign  Manual        11  In both cases the positions become first green and the yellow     To exit back to the Main Menu  click    Cancel     See Fig  5 1 2 4         09 20 2001 152928    Fle Beyim  Dot    Sample Rack Load    Free Postions  Sample Kertifund  Sample NOT Meretiod  Accepted Posimons       5 1 2 4  To Accepted Sample Positions    5 2 Reagent Identification  To identify the Reagents in their correct position as programmed on the Reagent ON Line Page    1  From the main Menu click F9     OPTIONS and activate Reagent Bar Code Identifica
7.   first turn plate                     gt      Inc  Time 1 Read Time  4   1 Add reagent into cuvette  2 Within 9 to 40 seconds Read Reagent cuvette Blank   M1  3 ADD SAMPLE to preheated reagent and mix  4 Incubation Time 1   variable   Example 62 sec   or any other time from the table   5 Measure Sample M2   First Measurement  6 Read Time   78 sec   or any other time      Second Measurement    M3    This is a two point measurement  CALCULATION OF RESULT     M3   M2  X FACTOR       FIXED TIME TEST   2 reagents    Read Reagent cuvette   M1  Add Reagent 2    Add Reagent  blank RB  Add Sample    9 40 seconds 62  88  114  140  166  192  218  244  270  296  322    eagent preheating to 37  C   About 90  Max Incubation   322 seconds  first turn plate    gt   gt  Ss   Inc  Time 1 Inc  Time 2 Read Time  4 6 8       1 Add reagent into cuvette   2 Within 9 to 40 seconds Read Reagent cuvette Blank   M1   3 ADD SAMPLE to preheated reagent and mix   4 Incubation Time 1   variable   Example 62 sec   or any other time from the table   5 Add reagent 2 and mix   6 Incubation Time 2   variable   Example 62 sec   or any other time from the table   7 Measure Sample M2   First Measurement   8 Read Time   78 sec   or any other time      Second Measurement    M3    This is a two point measurement   CALCULATION OF RESULT    M3   M2  X FACTOR   In case of RB BIAS the calculation becomes   CALCULATION OF RESULT     M3   M2   BIAS  x FACTOR       100 Human HumaStar 300 User Manual    KINETIC TEST   Singl
8.  1  Use of Profiles during Patient Entry     2  On Line Print Results    3  Automatic Repetition of Results   4  STD calibration in triplicate   5  Select primary tube   6  Show factor in standard section    94 Human HumaStar 300 User Manual    Options not included in the Standard Analyzer  are     1  Bar Code reader for Reagents   2  Bar Code for positive Sample Identification     These options requires special software and modules to be implemented and available only upon request   For Detailed Information see Chapter 3 7     Human  I o        Diagnostics Worldwide  95    F8   MAINTENANCE   da    Click F8 to open BOOK to MAINTENANCE   The Maintenance is divided into 3 parts     the Reagent System     the Sample System and     the Measurement  System  There is a complete auto check every time the analyzer is turned ON     This maintenance program has been developed to enable to check and control every operation in the system  This  makes possible to quickly identify any problem or failure on the system ON LINE and eventually make repairs  immediately or via a phone or modem line     A description and explanation of the Maintenance program is given in the Service Manual  Herewith  there is only a  list of checks and controls that the analyzer system goes through every time the analyzer is turned ON as well as all  the controls that are performed during operation     It is possible to check every moving part in the system as well as do maintenance that the analyzer requires fr
9.  10 0 47  1 47 0 43  Human  PR o    Diagnostics Worldwide    65    5 4 Data Exchange with Host Computer   The SOFTWARE of the HumaStar 300 produced by HUMAN enables the exchange of data with other computers via  a serial port RS232 connection    There are two options available     1  Transfer of Worklist  download from Host   2  Transfer of Results  Upload to Host     5 4 1 Host Computer Setup  The  Host Computer to be connected to the HumaStar 300 should have the following set up     19200 n 8 1 XON XOFF     5 4 2 Connection from Host Computer To HumaStar 300    The Host Computer should be connected to the first serial  COM 1  by means of a serial cable whose connections  are described below     HOST COMPUTER  DTE  HumaStar  DTE   DB9  DB25 DB9  2 E Ar ei 3  3 De E O 2  5 aee 5    5 4 3 Protocol  The exchange of data is organised in packs  These packs can be generally classified as     A prologue pack  sent at the beginning of the file to signal the number of data lines that are going to be  transmitted  Following is a description of a prologue pack     HeadChar             NumberOfLines            CheckSum   CR  Data pack  sent after a prologue pack  following is its description   HeadChar            DataRecord             CheckSum   CR         there is a space with a character ASCII 32  hex 20   HeadChar is a single character line header     R     Result  for uploading     P     Patient  in case of downloading     NumberOfLines is the number of lines  Data packs  that are 
10.  12D9    35 U L  0  35 200501201512 1672    Cardio 0 GLUC 2  108 mg dl   65  105 200501201512 1814    2  1 968 mg dl    8  1 4 200501201512 18BD    Cardio 0 BilBic 2  1 59 mg dl   0  1 2 200501201512 1902    0 BilBic 2  1 58 mg dl   0  1 2 200501201512 1C1A  OJAST 0  34 U L  0  35 200501201512 198A  O GLUC O  103 mg dl   65  105 200501201512 1B26  O CREAT 2  1 937 mg dl    7  1 2 200501201512 1BCF       R 12345678 Verdi Mario 18 0   O CREAT 1   453 mg dl    7  1 2 200501201512 159F  R 12345678 Verdi Mario 18 0   O BilBic O   34 mg dl   0  1 2 200501201512 15EA   R 12345678 Verdi Mario 18 0   O AST O  7JU L  0  35 200501201512 1363   R 12345678 Verdi Mario 18 0   0 GLUC 1  29 mg dl   65  105 200501201512 1507    NOTE     To activate LIS communication  select  F9  print ON     LINE    Human  PA o    Diagnostics Worldwide    69    70 Human HumasStar 300 User Manual    6 Maintenance Program   In this section are described all the controls and checks that the system does when turned ON and its maintenance   The maintenance is divided in three parts    1  Reagent System   2  Sample System   3  Measurement System    This program enables the operator to run a complete diagnostic check of the system without the need of a service  engineer     When turned ON  the analyzer goes through a complete functional Auto check and when everything is in order  the  green triangle START is lighted indicating that the analyzer is ready for operation        During the Auto check  should there be an 
11.  Any damage so incurred shall be reported  to the freight carrier for settlement or claim     Human  I A    Diagnostics Worldwide  3    1 3 Installation    1 3 1 Unpacking  Shipping and packing materials have been selected to provide maximum protection during transportation under  normal handling conditions     NOTICE  once the carrier has taken possession of the system for transportation from the factory  the carrier  assumes all liability until delivery  All claims for damage due to transportation must be filed with the carrier as  soon as possible                                            1 3 1 1    Examine the shipping crate and cartons for indications of damage  e g  crushed or indented walls  holes or gouges   water damage  etc  Have the carrier note any such damage on the delivery receipt  this will simplify formulating a  claim ifany ofthe instruments or parts are damaged     NOTICE  the system serial number is identified as the serial number of the Analyzer     Open the carton from the top and remove the instrument with care as shown on the graph above  It is  recommended that two people help with the removal of the instrument from its crate and its plastic bag  Save the  crate and packing material until you are sure you have received everything and all is in good working condition     1 3 2 Installation   The Analyzer system must be installed in the laboratory by specialised personnel  At the time of installation the  system will be checked to ensure proper operati
12.  Calibration in triplicate   to run a standard in triplicate from the same sample cup during Calibration  Will  report a Mean value as the result     4  Automatic Repetition of Results   enables to repeat automatically a test if its result is out of Linear Range  LIN   or Substrate Depletion  SDP      5  Samples and Reagents with Bar Code   enables the identification of samples and reagents   if the Bar Code  Reader is installed     6  Use of Primary Tubes   has to be activated when working with primary tubes     7  Display Factor in the Standard Program   activate to display the calculated Factor during Calibration   instead of Optical Density ofthe Standard     44 Human HumasStar 300 User Manual    3 7 2 Special Options  upon request     1  Bar Code Reader for Samples and Reagents   installation has to be made in factory or by qualified  personnel     2  Identification of Samples with Optical Pen   to identify the Bar Code of Samples  An Optical Pen can be used in  parallel with the keyboard     File Edt Graph Navigation 2 3 10 06 11 43 22 AM       Y  Automatic repetition of results      Select primary tube  iw  Show factor in standard section           Ready   A    HE P  M4242aMauoCheck 0K 7       3 7 2 1    Diagnostics Worldwide  45    46 Human HumaStar 300 User Manual    4 Analyzer Operation    4 1 Analyzer Operation  Check before Start    Before turning ON the analyzer  make sure     a  Tofillthe Wash Solution Container with fresh solution and place the liquid sensor fo
13.  Check        48 Human HumaStar 300 User Manual    If Bar Code Reader is built in   it will check the reagent positions  if they are as programmed on the  Reagent On Line page  If the Reagent is not recognized there will be a Warning     A    After the Reagents have been recognized  the Sampling Probe will enter into each of the programmed reagent  bottles and check its volume  The quantity of reagent will be displayed in Test  as number of tests available in  the reagent bottle     When the analyzer is not ready or the operational check failed  the small red warning triangle will be ON  Click  on the arrow next to it     to identify the reason  Check Maintenance and trouble shooting guide to solve the  problem before proceeding     A    In case of Warning or Error  make sure that a Test using 2 reagents has been assigned 2 positions onthe Rack  and both Reagents are present     Make sure to have enough Reagent for the daily use  To add Reagent     click on its Check Box to deactivate it   add reagent and then click again to activate  Click START to repeat the Reagent volume check only on that  reagent     To proceed operation     click START      gt       Start Operations es 2  x        Searching operations     No maintenance requesl  Rack A  1 10   Pal  no warnings present      No reagent request    no incongruences present      No calibration request    standards present      Predil  Ratio    Start Operation for above Rack  No  Change Predilution cups betore to start   Pre
14.  Empty   Alphanumeric  Sample Type 1 digit 0   Serum Plasma   1   Urine   2  FCS  List of Tests  3  Variable Length Alphanumeric  4    1  Can be empty only ifthe    Family Name    is not empty   2  Can be empty only if    PID    is not empty   3  The DataRecord must contain a list of tests for the patient    4  Each test has to be incorporated between the strings       ASCII 176  and       ASCII 167  and must be max 6 characters   Example   P 60126    P 123 Jimmy Denver 77 1  Private Patient 0   AST    BilBic     CREAT    GLUC   1AF2   P 1234 Frank Rossi 18 1  Epat  O   AST    BilBic    CREAT    GLUC   16A1   P 12345 Elisa Ribas 34 0   0   AST     BilBic    CREAT    GLUCS   14F8   P 123456 Walter Giotto 88 1  Cardio 0   AST    BilBic    CREAT    GLUC   1890   P 1234567 Coppi Sandra 33 0  Private Patient 0   AST    BilBic    CREAT    GLUC   1BA9  P 12345678 Verdi Mario 18 0   O   AST    BilBic     CREAT    GLUC   15B2   Where    P 123 Jimmy Denver 77 1  Private Patient O   AST    BilBic     CREAT    GLUC   1AF2    P   space  are prologue to DataRecord  space  cod  ascii32 hex20   123  identification of sample  max 13 characters   Jimmy  Fam Name  max 16 characters    Denver  First name  max 12 characters    77  Age   1  Sex  0 Female 1 Male    Empty  Accept Time   Private Patient   Comments  max 22 characters   0  Type of Sample  see table         start test name       cod  ascii176    AST  test name  max 6 characters       end test name      cod  ascii167    next test always 
15.  Incubation   322 seconds  first turn plate     IncTime 1 Inc  Time 2    4 6  1 Add reagent into cuvette  2 Within 9 to 40 seconds Read Reagent cuvette Blank   M1  3 ADD SAMPLE to preheated reagent and mix  4 Incubation Time 1   variable   Example 62 sec   or anyother time from the table   5 Measure Sample M2   before adding Reagent REAGENT 2   after 62 sec   6 Incubation Time 2     Example   260 sec   7 Measure final reaction   M3    CALCULATION OF RESULT    M3   M2  X FACTOR       Note  Sample Blank is measured just before adding Reagent 2  Incubation 1   and Incubation 2 are variable    102 Human HumaStar 300 User Manual    BICHROMATIC DETERMINATION  2 3    Read Reagent cuvette   M1  Read End Point   M2    Add Reagent  blank RB  Add Sample    9 40 seconds    eagent preheating to 37  C   About 9C Max Incubation   322 seconds  first turn plate     Add reagent into cuvette   Within 9 to 40 seconds Read Reagent cuvette Blank   M1   at 2 different wavelengths    ADD SAMPLE to preheated reagent and mix   Incubation Time 1   variable   See Incubation Table      Read final reaction of sample   M2   at 2 different wavelengths   Incubation of reactiion can be anywhere from 36 to 322 seconds at intervals of 26 seconds  Set incubation as necessary for the method from the incubation Table 1    CALCULATION    filter M2   Afilter M1  x FACTOR  Where  A filter    is the difference in mAbs between the 2 wavelengths    NOTE  Bichromatic determination can be single or two reagents     FIXED 
16.  Intelligent STOP    Display Bar that identifies the error or warning   For Detailed Information see Chapter 4 3     96 Human HumaStar 300 User Manual    FLAGS   WARNINGS and ERRORS   Q    The following Flags and Error Warnings will be displayed during analyzer operation       1    1     RSE UM qa ty N    10     12   13        Rept        test to be repeated      Low        result below normal range     High     result above normal range      OD       Optical Density out of range      lt RB gt       Reagent blank out of range     SbDp        Substrate Depletion       lt NL gt    Non linear Kinetic measurement     Std         No standard available      Log         error in Log Logit determination      lt 0      negative result      Lin        Result above Linearity Limit     Rgt       Missing Reagent      Smp      Missing Sample   Error in ISE module      ISEC    ISE Cal  A Air      ISEa    ISE Air Error      ISEn    ISE Noise Error      ISEd    ISE Drift Error      ISEo    ISE Out Range    For detail see Chapter 7     Human  I o    Diagnostics Worldwide  97    98 Human HumaStar 300 User Manual    8 Appendix     Reaction details                                                          END POINT   Single Reagent  1 2 3 4 5  z   5  5 EE 2  o OU    q v O E     Yor d  s  385 S  30m  2 e St E S  9 40 seconds 10  36  62  88  114  140  166  192  218  244  270  296  322  Ara  Reagent preheating to 37  C   About 90  Max Incubation   322 seconds  first turn plate   1 Add reagent into cu
17.  Patient Report    Normally the print done during operation is per patient  OPTIONS     is activated On Line  Patient   s results are  printed one after the other  there might be several patients on the same page  This print is ideal for routine testing  in case one wants to check     x  Cancel   O List by Test a     O List by Sample OK  O Print by Report    From Position    1        To Position  10         4 7 1  The results before printing the Final Report   However  the final Patient Report can be printed any time  as follows     1  From the Main Menu click on    File    and the    Print     There are 3 possibilities of print     a  Print by Test  b  Print by Patient  c  Print Final Patient Report    Choose accordingly    56 Human HumaStar 300 User Manual    4 8 Special Notes  4 8 1 Flags and or Errors displayed On Line        Rept        sample to be repeated      Low        result below normal range     High     result above normal range      OD       Optical Density out of range      lt RB gt       Reagent blank out of range     SbDp        Substrate Depletion       lt NL gt    Non linear Kinetic measurement     Std         No standard available      Log         error in Log Logit determination      lt  0      negative result      Lin        Result above Linearity Limit     Rgt       Missing Reagent      Smp       Missing Sample   Error in ISE module      ISEc      ISE Cal  A Air      ISEa      ISE Air Error      ISEn      ISE Noise Error      ISEd      ISE Drift Error 
18.  Reagents     f  Stabil   hrs    stability time of reagent once placed on the analyzer  There will be a flag when the time is  up  To deactivate stability time enter   999      g  Lot Numbers   enter the lot number indicated on the reagent bottle   h  Bottle Type    there are two sizes N  1   50 ml   N  2   5ml     i  LD  Reagents   insert identification number only if the Bar Code Reader is installed  It checks the bottle  position inside the Reagent Tray     9  FILTERS  nm     a  First  amp  Second    Click on the line to open the window that displays the 8 built in interference filters   Choose according to the method     b  Second Filter   Used only in Bichromatic Methods     d Bichromatic Factor  not active parameter    Human  PA o    Diagnostics Worldwide  33    Page 3 includes           2 21 05 5 12 53 PM    AST GOT    Read Time  sec   168    CHECK ABSORBANCE  mAbs   P Blank  mir   1000     max  2200  Substr  Depl    min   300    CALCULATION  Calibrate  Factor  Factor  1822 000  Reag  Bias  No    LINEAR CORRELATION  Intercept  0 00  Slope  1 00        gt   15 11 21 PM  Auto Check OK          3 3 4  10  REACTION   a  Type  click to open the options  and choose one   a  EndPoint  b  Fixed Time  c  Kinetic    b  Read Time  sec    Measurement Time  Click to open window and choose from the pre programmed and  optimized times  Only for Kinetic and Fixed Time Methods   a  For End Point measurements the time is fixed   b  For Fixed time and Kinetic measurement times can vary f
19.  Registers  Contains the following sub programs     File   Registers Tools 2  Operation           F Data from Archive  Methodology    Profiles     Standards  Controls QC     Maintenance    Options    2 4 1    The following functions can be programmed from Registers or simply a click on its ICON or on its functional key    from F2 to F9   NOTE  The program METHODS  FA  is protected by a password     2 5 Tools  Contains the following sub programs        File Register    Tools 2  v Esser Mode       Instrument   nit  Preprer  FR Rees culate Results    Repot Definition    iid   E a nte  a  Y       2 5 1    1  Expert Mode   Enables to introduce the password to access METHODS  F4     2  Init instrument     To repeat an initial Auto Check of the Analyzer    20 Human HumaStar 300 User Manual    3  Prepper    This function enables to automatically pre dilute the sample OFF LINE  Follow instructions indicated on the  monitor     2x     op Insert initial and final positions to use prepper   sp select prediution ratio and then press OK  _ Cancel      From position  1 40 ft  To position  1 40 ft  Gun pe E  Samples Ring  Internat   zl       4  Load Sample Rack  Identification of samples with the built in BAR CODE READER  optional      Ele Edt Graph Navigation 2 3 8 06 3 55 30 PM    J Available Reagents I Reagents On Line  Cite   Set Reagent Test  Stbihrs Bott  Cool      Test lm Y TAG 999 1   Dua SE           Human  D    Diagnostics Worldwide  21    5  Recalculate Results    Enables to recal
20.  again after 52 sec   to check for substrate depletion   M2  Incubation 2   example 78 sec   or any other time    Read time   for Kinetic measurement     example 130 sec   or any other time    M3    CALCULATION is a Linear Regression based on the 130 sec measurement  multiplied by its FACTOR   END POINT   2 reagents   POST R2  3 5    Add Reagent 1   Add Sample   Dispense Reagent 2   Read Sample cuvette Blank      Read End Point    9 40 seconds    Reagent preheating to 37  C   Abq Max Incubation   322 seconds  first turn plate      gt   ey  Inc  1   62  Incubation 2   260   4  1 Add reagent into cuvette  2 ADD SAMPLE to preheated reagent and mix  3 Add REAGENT 2   example after 62 sec   4 Incubation Time 1   variable   See Incubation Table   Example 62 sec   5 Within 9 to 40 seconds Read Reagent cuvette Blank   M1  6 Measure final reaction   M2    CALCULATION OF RESULT    M2   M1  X FACTOR       104 Human HumaStar 300 User Manual    HUMAN   Gesellschaft fur Biochemica und Diagnostica mbH     Max Planck Ring 21   65205 Wiesbaden   Germany    Tel    49 61 22 99 88 0   Fax   49 61 22 99 88 100    e Mail  human human de   www human de    Human     EEE    
21.  as the ISE module are optional and are available directly from the factory  They    can also be installed at a later date at the customer s side by an authorised service engineer     5 1 Bar Code Reader for Samples and Reagent  5 1 1 The Bar Code can read the following codes   UPC EAN   Code 128    1        Code 39          w N    Interleaved 2 of 5  Codabar     gt     5    Per additional information contact your local agent     5 1 2 To identify the samples with the Bar Code     File Registers   Tools 2  v Expert Mode          Instrument Init  Prepper    S FS Recalculate Results    Report Definition  Z  F4 Date and Time  A  F5   FS   AM    5 1 2 1  Select Bar Code Reader    From the Main Menu go to OPTIONS  F9    Select    Samples with Bar Code     See Fig 5 1 2 1    Insert Sample Tray  The Bar Code will identify automatically the samples   See Fig  5 1 2 2   A window will open displaying the Sample Tray and commands to identify them     e Ad S    Human  I A    Diagnostics Worldwide  59       File Registers Optore 2 09 30 2001 15228    Sample Rack Load    00000     O   Q  O     S    O FeeePostions   DD Sorel weruhee  Sara NOT Vermeil  O  AmepedPostions    A go    O       5 1 2 2  Sample Tray Bar Code Identification  5  Ifthe samples have been transferred via Serial port RS 232  these should contain as follows     a  ID number of the sample   b  Test to be performed  6  Make sure that the identification label with its Code on the Primary tube is visible on the Sample Tray  7
22.  diluter   Check reading cuvettes  Check washing probe   Clean probe   s washing chamber  General cleaning and tubing  check    Replace sensor level   s flat cable    General revision    HELP ON LINE     TELL ME HOW  Welcome to the HELP ON LINE    Click for the following information     1  Reagents On Line    2  Calibration    3  Patient Entry    4  STAT Operation       86 Human HumaStar 300 User Manual    HumaStar 300     Maintenance Program    15days    30days    4months    6months    EN              Y    O    E    O    lyear    2year    5  Archive       6  Print Results    7  Methodologies    8  To Program a New Method    9  Standards and Calibrators    10  Profile    11  Controls    12  Ouality Control Program    13  Options    14  Maintenance    15  Special Functions    16  Warnings and Errors    Ov ZF QQraRre       REAGENTS ON LINE   2    SE       Click the mouse on Operations   F2 Reagents On Line     All the chemistries that have been performed during the last run   are displayed on the right  There are a total of 30  ON LINE reagents  that are identified from 1 to 30 on that page  To see reagents 20 through 30 scroll the bar on the  right    To choose the methods to run today  simply click the mouse on its check box    If a new reagent or method is needed and is not on that list  click the mouse on free position  which is identified by  an asterisk to highlight it and to open the left side page  where all the remaining programmed chemistry methods  are displayed    C
23.  for each individual method    8  Various possibilities of performing particular tests    9  Many checks and controls to assure reliability and accuracy of operation     Human     mme    Diagnostics Worldwide  27    28 Human HumaStar 300 User Manual    3 Software  3 1 Software    HumaStar 300   has an extremely easy and user friendly software with graphic presentation all incorporated into a   Book  which further simplifies its use  A  HELP  On   Line guides the operator during programming and operation     Most of the special functions have been described in Chapter 2  herewith are described in detail the various parts of  the program     Methods  F4  are protected by a password that can be defined by the user at the time of analyzer installation        File Registers Tools   3 3 06 9 44 10 AM    E   FA    HumaStar 300    F3    F4    F5    FE    F7    F      Fg    e Qtr SS E    Ready          Human            Diagnostics Worldwide  29    3 2 F3 Archive    y    Click on its ICON to open the options offered     ww Archive Selection            CG  Internal Archive Cancel          Archive on Diskette OK  O Test Statistic   O Test Counter   O Last Warnings            3 2 1    3 3 F4  Methods    S      Click on its ICON or F4 to open the window with the 2 choices          Regular Methods     O Calculated Methods  DK         3 3 1    1  Regular Methods   access to all the Chemistry methods in alphabetical order     2  Calculated Methods   user programmable formula to calculate RATI
24.  introduced  This operation is done  automatically through the Maintenance program  Simply press first Drain Bath and then Fill Bath     NOTE  USE ONLY BI DISTILLED WATER FOR THE INCUBATION BATH    3     4                 16   MONITORA                  11 ON OFF  SWITCH       15   VOLTAGE  SELECTOR    AIR PUMP    1 4 1 1    ISE     CALIBRATOR    A      Place the  Calibrator A  bottle close to the ISE module and close it with its level  sensor cap  When using the ISE module  the reagent will automatically fill the system  For details see ISE  Calibration    WASTE     It is suggested to empty the Waste Container daily and rinse it with a detergent to avoid bacteria  formation     6 Human HumaStar 300 User Manual    15 Description of the Analyzer    1 5 1 Layout   The analyzer consists in three main parts which are assembled on a single support    REAGENTS   SAMPLES   REACTION MEASUREMENT including cuvette washing system  See Figure 1 4 3 1 each of  these parts will be described in detail below               REACTION PLATE  with 39 cuvettes and Built Photometer       Reagent Dispensing Arm       Sample Dispensing Arm    4 STAT      16 Place for  Standard and Controls          NY J  Washing well    LS E           Primary tube       30 ON LINE REAGENT SAMPLE TRAY    1 5 1 1  Layout    Human  PN    Diagnostics Worldwide  7    1 5 2 Analytical Plate and its incubation thermostatic bath    The reaction takes place in an analytical place containing 39 quartz cuvettes  The plate is inc
25.  the requested operation  an OK will be displayed with the time of operation  Wait for the    Green Triangle START to be On  before requesting the next operation   6 4 Daily Checks    During SHUT DOWN of the analyzer  there will be an automatic wash of the cuvettes  In case of a non correct Shut  Down of the Analyzer  when analyzer is turned ON again  after having done the necessary Auto checks  it will wash    the cuvettes   Daily maintenance     a  Empty Waste container     b  Fill the Wash Container with fresh solution   Make sure to use good quality distilled water  bad quality of    water may give ERRORS with the level sensor    c  Fillthe container with bi distilled water for the incubation bath   d  If ISE Module is installed     check the presence of CAL A reagent   e  Remove all reagents    Note     WHEN ANALYZER IS TURNED OFF     THE COOLING OF REAGENT TRAY IS ALSO OFF     Human  A        Diagnostics Worldwide  73    6 5 Maintenance every TWO weeks  Change the bi distilled water in the Thermostat and Incubation Bath     1  Empty and then refill with fresh bi distilled water     use the Maintenance program    2  Depending on the tests performed  however  it is suggested to wash the cuvettes with a diluted Solution of  Extran or a similar product used for laboratory glassware  Use the Clean procedure found in the Maintenance  program    6 6 Operation to be done every 2 months   Clean Thermostat and Incubation Bath   1  This should be done using a liquid detergent for 
26.  to loose the O ring  3      Unscrew the syringe body  2  including its piston      Make sure to fit the O ring  3  properly when mounting the new syringe      gt     5    Note  Do this operation only if necessary  if there is a leaking or broken syringe        6 7 6 1  Diluter Syringe    1  Piston 6  Syringe holder  2  Syringe body 7  Tip tubing locking screw  3  O ring 8  Tubingto Wash solution  4  Piston locking screw 9  Tubingto Probe   5  Locking screw syringe    Human  I o    Diagnostics Worldwide  79    80 Human HumaStar 300 User Manual    7 Trouble Shooting  Spare Parts    7 1 Trouble Shooting Guide    SYMPTOM  Analyzer does not Turn On    Photometer lamp is OFF  Program does not load    Printer does not turn ON    Printer does not print On Line    Printer does not print Patient Reports  or other print outs    Tubing leading to the two Probes are  empty    Air bubbles in the tubing leading to the  Probe    Photometer does not read    Diluter Module stops    Diluter does not aspirate or dispense    Syringe has difficulty in filling or  emptying    Water Blank has too high an  Absorbance    Results are not reproducible    CAUSES AND REMEDIES  check power supply cord if inserted  check fuse  check voltage input  Make sure that the Monitor is ON  check power supply before changing lamp  if computer and instrument is ON     change lamp  Turn OFF the analyzer   Wait 10 sec and Turn ON again  Make sure that the power cord is plugged in  Make sure it is connected to the analy
27.  two pages  click the triangle on the bottom of the page     7  SAMPLE       File Edit  Graph Navigation 2                  AST GL       SAMPLE REAC          Volume  ul   30 Type   Predil  ratio  Read  REAGENTS First Second CHECI  Features   none  R Bla  Volumes  ul 300 3     Incub   sec   72 0 Subst  Cooling  Yes No  Stabil   hrs   999 999 CALCL  Lot Numbers  Calibr  Bottle Type  Facto  Reag  ID First   ID Second   FILTERS  nm  LINEA  First  amp  Second  340   Interc  Bichromatic Factor  Slope    Ready A  BP  nza    a  Volume from 3 to 70 ul     b  Predil     Ratio    Option enables the automatic pre dilution of Samples  Standards and Controls before    starting a run  To select the dilution ratio  click on the line to open the 15 pre programmed dilution ratios  2   4  5  10  20  30  40  50  60  70  80  90  100  150 and 200     8  REAGENTS  Features     Click on the line to display the available options      none    Single Reagent Standard Method     Mono Diff   Single Reagent Fixed Differential Method   the first measurement M1 is taken 6 sec  after the addition of the sample  Sample Blank   The final measurement M2 is taken after the  programmed incubation time     Bi Reag   Two Reagents Standard Method     Diff  Fix   Two Reagents Fixed Differential Method   the first measurement M1   is taken 6 sec after  the addition of sample to the first reagent R1  sample Blank   The second reagent R2 is added after  the programmed Incubation 1  final measurement M2 is taken after th
28. Control 43  3 6 2     To Delete or Chan Bera  ControlV alle unse 43  316 3    Oualityi Control  Programas cases POB PEO O sc a an 44   ET el EE 44  3 7 1 Standard Options   can be activated by the operator when needed               mer 44  34 2     Special Options UN II  45   4 Analyzer  Operationen allen essen are REISENDEN teen 47   dd SEET EE 47   EA LE A A EES 47  4 2 1 Check Reagent Volumes for the Daily Run   48  4 2 2 Prepper   the Auto  Pre dilution system EEN 50   KERSCHT CL 50  4 3 1 The Calibration page    arar 51  4 3 2 To program a Serum Control for Calibration ENEE  51  43 3   Multi Standard Eallbration COPA OR In PA or ada 51  Ee UE TEE EN  E ET EN  E Een 51    4 3 7 Check Calibration    51  4 3 8 To Repeat a Standard or Calibrator   4 3 9 To Repeat a Serum Control    BER          4 3 10 To Display the graph of a Calibration or a Kinetic Control       44 Patient Emtnm  EEN BEE  4 4 1 To Program Patients using Profiles  BEE  Ee EE EEN 54  A A NO 55  4 6 1 To Add Samples During Operation EEN 55  4 6 2 To Repeat a Test  a Sample or All Tests ENEE 55  4 6 3 To Recalculate a Test against a New Factor ue 56  4 7 MAP TIME Of a RR ao 56  48  Special NOTES S Saa Saaai RRRS Ea SRE E ERES ES EES a SESS 57  481   Hags apndiorkrrorts gdleplavegd On Dip  smsen 57  4 8 2      Sample Pre dilutionz  WI N een nes 57   5 Bar Code  Reader se Module  Data EXchange n u    ensneeesee nennen 59  5 1 Bar Code Reader for Samples and Reagent EEN 59  5 1 1 The Bar Code can read the follow
29. GGT  GLUC  IRON  LDH  MAGN  TB  T P     UA    3 6 1 To Program a New Control    2 21 05 5 27 07 PM    3322  05 27 10    StdDev Q C  page H  06 19  03 21  12 23  5 25  07 27  3498 23  10 31  04 35  03 37  3 39  10 41  10 43  30 45  06 47  06 51  53  57  59            Click  Edit  and the    Insert Page     Introduce the name of the new Control and click OK  Introduce the Expiration    date and its Lot Number     a  Click on the line with an asterisk     on the right  On the left are displayed all the available methods  Click on the    check box of the desired method to highlight     b  Click the mouse on  Mean  and enter the Mean Value for that method  including the value of the plus and  minus corresponding to 2 Standard Deviations  or the values indicated on the sheet supplied with the Control     Program one by one all the methods that are part of that Control     Example  GLUCOSE  the theoretical value is 100 mg dl with     6     3 6 2 To Delete or Change a Control Value    Simply highlight the Control and press  Canc  on the keyboard  A confirmation window will open  choose  accordingly  To Delete the Page or the whole Control   click  Edit  and choose  Delete Page   A confirmation    window will open  choose accordingly     To Change the Value of a single method   simply click on its value and change accordingly  To exit   press the book  cover to close the book or Navigation to close the book  The programmed controls are automatically saved     Human  A        Diagnosti
30. HumaStar 300      User Manual          ee ET UTA  AG ELIT atol Meares    CEA FE  AR Sa IRA de    ERARA       agt       CE    Human  EEE    No     Rev   Date  01 2008 07  02 2010 08    REVISION LIST OF THE MANUAL    REVISION DESCRIPTION  First edition for modified HumaStar 300 from S N A520331  Correction of typing errors    Human  PA A        Diagnostics Worldwide    1 INTRODUCTION    This manual is considered as a part of the instrument  it has to be at the operator s hand as well as at the  maintenance operator s availability  For accurate installation  use and maintenance  please read the following  instructions carefully  In order to avoid instrument or personal damages  carefully read the  GENERAL SAFETY  WARNINGS     describing the suitable operating procedures  In case of breakdowns or any troubles with the  instrument  apply to the local Technical Service     2 USER WARRANTY    HUMAN warrants that instruments sold by one of its authorised representatives shall be free of any defect in  material or workmanship  provided that this warranty shall apply only to defects which become apparent within  one year from the date of delivery of the new instrument to the purchaser     The HUMAN representative shall replace or repair any defective item at no charge  except for transportation  expenses to the point of repair     This warranty excludes the HUMAN representative from liability to replace any item considered as expendable in  the course of normal usage  e g   lamps  valv
31. No maintenance request  Rack A  1 10   Pal  no warnings present        No reagent request     no incongruences present      No calibration request     standards present     Predil  Ratio     Start Operation for above Rack  N    Change Predilution cups before to start     Press OK to confirm and start operation     Cancel            4 6 1    The analyzer can be started to run the routine Patient samples or STATs any time after the analyzer has been  calibrated for the day  Place at least one Sample Tray with the programmed samples and Press START  the GREEN  TRIANGLE below      A confirmation window will be opened  press OK     4 6 1 To Add Samples During Operation   The analyzer is protected with a cover during operation not to disturb the moving parts  To add new samples or  STATS onto the Sample Tray     the analyzer has to be STOPPED  Click on the RED square next to the START  The  Sampling will STOP while the rest of the analyzer will continue operation  Wait until the message displays  Sampling stopped      only then open the cover and introduce the desired samples  Sample Tray or even Reagents if  necessary  Press START to continue operation  Make sure that the Plexiglas cove is closed again     Note  DO NOT INTRODUCE SAMPLES OR REAGENTS WHILE THE ANALYZER IS IN OPERATION  IT MAY DAMAGE  THE SAMPLING ARM  KEEP THE PLEXIGLAS COVER CLOSED DURING OPERATION     4 6 2 To Repeat a Test  a Sample or All Tests  To repeat a Sample or a Test  proceed as follow     1  Select the d
32. OS and calculated methods on tests  performed   See Par  4 2 2      30 Human HumaStar 300 User Manual    To Program a Method  The parameters of a method are described on three pages  Page one contains the following information to be  programmed        2721705 5 12 28 PM                   TestName  AST GOT  Meas  Unit UAL  Decimals  0              REFERENCE    Bay  0 35  Girl  0 35  er Age Limit  15  Male  0 35  Female  0 35  Age over  60  Male  0 35    Female  0       LINEARITY LIMIT        QUALITY CONTROL  Repeat control every  hrs  1                EIA PM  Auto Check DK     3 3 2  1  Name of Test   Introduce a name for the Test  2  Units   units of measurement  mg dl  U I  g l  etc     3  Decimals   decimals in final result   max  3   4  Reference Values   enter Min  and Max of Normal of reference values according to age and sex   5  Linearity Limit     maximum linearity value for the method  according to the reagent manufacturer  If the value    is not defined all results will be flagged as  LIN   non linear      NOTE  When the Automatic Repetition of Results is active  all results that are above the programmed   Linearity Limit   will be automatically repeated using half the sample volume indicated in the  method and the result corrected     6  Quality Control     Repeat Control every hour     enter the hours to automatically repeat the Serum  Control for the Quality Control Program  To exclude   enter  99      Human     ze    Diagnostics Worldwide  31    To open the next
33. TIME TEST   2 reagents POST R2          E      D  be      D      Ka  E      o  Ed  el    Dispense Sample  Dispense Reagent 2    9 40 seconds    Pre heating Incubation Time 1 Read Time    1 Dispense reagent 1 into cuvette  anywhere from 50 to 300 ul  No reagent blank reading   2 Dispense sample  Incubation time 1 variable from 36 seconds on   3 Dispense reagent 2 into cuvette  Total volume R2 sample R1 must be at least 300 ul   4 First reading will be taken 13  after  5 Second reading after the Read time   Introduce READ TIME for all Kinetic and Fixed Time   Introduce INCUBATION TIME 2 for End Point    This is a two point measurement   CALCULATION OF RESULT    M2   M1  X FACTOR   In case of RB BIAS the calculation becomes   CALCULATION OF RESULT     M2   M1   BIAS  x FACTOR       Human  A        Diagnostics Worldwide  103    KINETIC TEST   2 reagents   POST R2    4 5    M2    Read Reagent cuvette blank Mi  Measure solution within 10   6 N    Dispense Reagent 1  DispenseSample  Check substrate depletion      Dispense Reagent 2 w  3    9 40 seconds    36  62  ss  114  140  166  192  218 de A 296  322    Reagent preheating to 37  C   About 90  Max Incubation   322 seconds  first turn plate    pe    S  Inc  1   62  Inc  2   78  Read Time   130   Dispense reagent 1 into cuvette    ADD SAMPLE to preheated reagent and mix   Incubation Time 1 of sample    example 62 sec   or any time    Add Reagent 2   Starter    Within 9 to 40 seconds Read Sample cuvette Blank   M1   Measure solution
34. aStar 300 User Manual    Cat     Number  17903 5  17903 6  17903 7  17903 8  17903 9  17903 4  17903 2  17903 1  17903 3  17903 10  17903 11    Description  Calibrant A  500 ml   Calibrant B  125 ml    Cleaner  125 ml   Diluent for Urine  125 ml   Pump Cassette  Reference Electrode  Electrode for K  Electrode for Na  Electrode for Cl    Compression seals  box of 5     O ring  box of 4     Table 5 3 6 1  Results obtained with the ISE Module    Mean  SD  CV     Mean  SD  CV     First Instrument    Na  134  132  132  132  132  132  131  131  131  132    131 90  0 88  0 66    Third Instrument    Na  150  151  151  150  151  151  153  152  150  151    151 00  0 9428  0 6244    K  4 10  4 00  4 00  4 00  4 10  4 10  4 10  4 10  4 10  4 00    4 06  0 05  1 27    K  6 90  6 90  7 00  7 00  7 00  7 00  7 00  7 10  7 10  7 00    7 00  0 07  0 95    Cl  98  97  97  97  97  97  97  97  97  96    97 00  0 47  0 49    Cl  111  110  110  110  110  110  110  111  110  110    110 20  0 42  0 38    Mean  SD  CV     Mean  SD  CV     Second Instrument    Na  135  131  131  131  131  131  132  132    131 75  1 39  1 05    Fourth Instrument    Na  150  150  149  149  153  151  151  150  150  151    150 40  1 17  0 78    Qty   4   2   2   2   2   1   1   1   1   1   1   K Cl  3 80 98  3 80 97  3 80 98  3 80 98  3 80 97  3 80 97  3 80 97  3 80 98  3 80 97 50  0 00 0 53  0 00 0 55  K Cl  6 90 110  6 90 110  6 90 110  6 90 111  7 10 109  7 10 110  7 10 110  7 10 110  7 10 110  7 10 110  7 02 110 00  0
35. ace the Syringe    79  7 Trouble Shooting  Spare Parts    81  7 1 Trouble Shooting Guide     81  7 2 Spare parts and accessories     84  7 3 Suggestions and Warnings      85  7 4     HUMaStar 300     Maintenance  Prograrny         srvasercsvssescsrasssoossasesoanssssoassssesostansoonnssscounsassoonnssssounsassossniusesousasescesusseseusasescoguasesgend 86  8 Apperidix  Redetion UE 99   Human          Diagnostics Worldwide    iv iv Human HumaStar 300 User Manual    1 General Description    HumaStar 300 is an automatic random access Chemistry Analyzer to run your daily workload  a single test  a  profile  an organ panel or STAT with the same speed and efficiency     HumasStar 300 incorporates the latest in the art of smart technology and a sophisticated software that simplifies  the work of the operator and offers maximum versatility and reliability of operation  The analyzer offers many new  and desirable features  including a throughput of 300 chemistry test single reagent plus over 180 electrolytes per  hour  5  at the same time  making it ideal in any size laboratory        Fig 1 1      Human  wm Bi Wer  Diagnostics Worldwide   al    1 1 Main Characteristics    Humastar 300   is fully controlled by a built in Pentium computer and performs the tests in a true random access  mode  where all tests for sample ONE are processed before starting on sample number TWO  Each sample is  treated like a STAT and a Patient Report is automatically printed     THROUGHPUT   300 Chemistry te
36. agent Identification Number if Bar Code Reader is built in   11  Select Primary Filter 1   Select Filter 2  only if Bichromatic   12  Type   End Point    13  Measurement Time   it is programmed by default        14  Enter the values in mAbs  for the Reagent Blank  both Min and Max  A check will be made during  operation   Example  Glucose   Min 0 Max 150 mAbs     15  Substrate Depl min     Parameter active only for Kinetic    16  Calculation   select between Standard  Factor o Multistandard    17  If against Factor   enter the factor value    18  If against Standard   introduce its value in the Standard program F6   See graphic Presentation of the Method in Addendum     1b  End Point   Two Reagent Test  Program as above points 1and 2  3  Features  Bi Reagent  4  Program the volume of the second Reagent R2  from 3 ul to 500 pl    5a  Program incubation time 1    time before the addition of R2   5b  Program incubation time 2    time before final measurement   All other parameters remain the same   See graphic Presentation of the Method in Addendum     Human  PA        Diagnostics Worldwide  35    2   Bichromatic Determination  1  Bichromatic Determinations can be End Point  Fixed time or Kinetic   2  In point 10  above   enter both   Filter 1 and Filter 2   All other parameters remain the same   See graphic Presentation of the Method in Addendum     36 Human HumaStar 300 User Manual    3  Fixed Time Determinations     This is a two point measurement   using a single or two Reag
37. anomaly  an errors or warning  these will be notified by the red Triangle        The Warning massages will be displayed by clicking on its arrow  Some of the messages are common such as   missing Wash Solution  Incubation Temperature not yet at 37 C  etc    Other Warnings or flags   such as INIT hardware  Filter identification  etc      require to consult the Maintenance  program and check accordingly     6 1 Reagent System  From the Main Menu click on    F8           MAINTENANCE     as shown on the figure below  the main Auto checks of the Reagent and Sample System are all  OK     EN       Fie Edt Graph Navigation 2       Reagent System   Set Actor Tee  Instrument On Line 11 18  Ind Hardware 11 19  FA Hydiaados 11 19  Wash Probe 11 19  Ind Code Reader  Test Digo  Test Liquid Sensor  Pump P4 ON    Pump P4 OFF  Text Purp P2  Home Z Avis  Home Plate    Home Asm  Homa Diner  Arm On Reagerk  Asm On Cuvette    Arm on Wath  M mee On  5 sec      6 1 1  Reagent System    Human  SP        Diagnostics Worldwide  Zi    6 2 Sample System    3710 06 11 49 46 AM    Sample System  Set Acton Tme  Instrument On Line 11 19  Ind Hardware 11 19  Fil Hydraubes 1 20  Wash Probe 1120  Int Coda Reader  Test D  ser  Test Sample Laval  Test Pump Pi  Test Predl Level  Test Clean Level  Test Cal B Level  ISE modi On Lime  ISE Maintenance  ISE Purge  ISE Clean  Home Z Aust  Home Piste  Home Asm  Home Dater       Ready A E  gt   imasaamasChexor z       6 2 1  Sample System    How to use the above program 
38. ar 300 User Manual    2 Start Up  2 1 Start Up    Turn ON the Analyzer  The analyzer goes through an Auto check program and any anomaly will be automatically  flagged     When ready  the green triangle START will be lit   2 2 Main Menu Display  The Main Display is divided into 4 parts     Part 1     contains all the commands for operation    Part 2     contains all the icons of the 8 sub programs or Registers   Part 3     indicates the Status of the analyzer    Part 4   is the book     OPERATION     de Registers Tools   3 3 06 9 44 10 AM  ne          HumasStar 300    E UA RS    Ready       Human  SP        Diagnostics Worldwide  17    2 3 File  Click   to obtain a number of options as shown below     File Registers Tools 2          Communications  Move Archives  Print  gt   Printer Setup b    2 3 1    1  Communications  Enables to send and receive data  Click to open the window that offers the below displayed options        Communications  COM   19200 n 8 1  E gs  CC gx          BM Ready to receve tests  Fress OK tostat O by Serial Port    Dee  h O by Diskette   p   ot    Store ir  Bae       Receive Worklist  O Transmit Worklist  O Transmit Results       2 3 2    The data can be transferred two ways  via the Serial Port or via a floppy diskette  Choose the way and proceed to     a  Receive a Work list  b  Transfer a Work list  c  Transfer Results    The description above on the left  1  will indicate how it will be done when pressed OK  The lower part in the  window  2  will 
39. be  automatically saved and placed alphabetically among the other existing methods in the system     To delete a method  open that method     click on EDIT and again on  Delete Page   A confirmation window will  open     click accordingly     For Detailed Information see Chapter 3 3     92 Human HumaStar 300 User Manual    Y toa  F5   To Program a PROFILE   e    Click on F5   to open the Profile page     Click on any existing Profile or any empty line with an asterisk  to open the Left side page that displays all the  programmed Methodologies available in the system  An existing Profile will have its methods marked with  v  in  their corresponding check boxes    To program a new Profile   click on an empty line on the right to open the opposite page with all the available  programmed methods  Click on the line  Name  on the left and introduce the name of the new Profile  Go back to  the right and click on the selected line to introduce that name  Click on the check box of the methods that are to be  part of that new Profile  There are no limits of the number of Profiles that can be programmed     To exit   click the mouse on the BOOK cover  The Profiles are automatically saved when the book is closed   To delete a Profile   simply highlight the Profile and press Key Delete on the keyboard  A window will open for  confirmation   press Yes or No accordingly     Click on F5   to open the Book to the Profile page   For Detailed Information see Chapter 3 4     F6     STANDARDS    
40. can be connected to the outlets AUX located in the back of the analyzer     PARALLEL PORT   PRINTER          o SE ES 0    va vas    au           KEYBOARD coma  wos   gt     1 4 1    Human  P    Diagnostics Worldwide    5    1 4 1 Hydraulic Connections    Follow example in Figure 1 4 1 1    1   a     b   c     2   a     b     WASH SOLUTION   Fill container with 5 liters of dist  water and add 1 500ul of Wash Additive cat  no   18971  Mix slowly to avoid foaming  Close the container with its screw cap containing the level sensor     It will be flagged when the solution level in the container gets low  however  there will be enough solution  to finish the programmed workload     It is suggested to prepare the Wash Solution fresh every day and add to the container as necessary    Once a week it is suggested to rinse thoroughly the Wash Container with distilled water before adding  the freshly prepared Wash Solution    INCUBATION BATH   Fill the container with bi distilled water and close it with its screw cap and its level  sensor    When the Analyzer is turned ON  the incubation bath will be automatically filled including its thermostat   Both the temperature and the level of liquid in the incubation bath are continuously monitored    Missing bi distilled water in the bath is automatically signaled and flagged    To avoid formation of bacteria or other undesirable matter inside the water bath  it is important that  every two weeks this bi distilled water be drained and fresh one
41. concentrations have to be of the same Lot Number and have the same  ID     The Calibration Curve can be programmed two ways   a  Automatic Pre dilution from a concentrated Primary Standard    Enter the name and ID  the number of calibration points  A page for each concentration will be automatically  pre set  Enter the concentration of the primary Standard  There will be an automatic geometric pre dilution of  the standard for the number of points programmed     b  Calibration Curve with individual Concentration Points     Enter the name and ID  the number of calibration points  A page for each concentration level will be  automatically pre set  Enter the concentration of each individual calibration point on a separate page     3 5 2 To Delete or modify a Standard Value  Simply highlight the Standard and press key  Canc  on the keyboard  A confirmation window will open  choose  Yes    or No  accordingly  To delete the Page    click  Edit  and  Delete Page   A confirmation window will open  choose  accordingly     To change the Value of a Standard   click on its Value and change accordingly     To save the modification   close the book     42 Human HumaStar 300 User Manual    3 6 F7  Controls    From the Main Menu click on Controls   F7 to open its page  There is no limit of the number of Controls that can be    programmed     Fie Edt Graph Navigation 2    Control Definition    Lot number  Expiabion date     Method  Bie  BILTD  Bil ot  hem  BUN  CAL  CHE  CHOL  CREAT  D BILI  
42. cs Worldwide  43    3 6 3 Quality Control Program  After each calibration the results of all Controls are automatically memorized for the Quality Control Program     To check  view or print   open the Book  From the Main Menu click on F7     Controls  Open the page to a desired  control and double click on a method  The OC pages will open     The right displays the values obtained during Calibration and the left displays the curve and the values of  Mean     SD  Standard Deviation     CV  Coefficient of Variation  and  Th CV   which is the difference between the  analyzed value   compared to the Mean value programmed for that method     The OC graph and its values are displayed for 60 days and are automatically updated daily  To cancel a wrong value  or an error  highlight it and press  Canc    Ele Edt Graph Navi  gt  4 3 2006 9 16 00 AM  Gluc   114 220 mg dl    Data Concentrazione Riferimento   12  4139   117  34 134   118  413   118  34 134   ne 194134   116  34 134   114 194 134   116  34 134   110  94 134   121  34 134   102  34 134   101  94 134   107  34 134   102  34 134   111  34 134   104  34 134   107  34 134   107  94 134   104  34 134     074 A    20 4 2004 5 5 2004 10 7 2004          3 6 3 1    3 7 Options  Available Standard Options     3 7 1 Standard Options   can be activated by the operator when needed     1  Use of Profiles     enables to use the Profiles during Patient Entry   2  On Line Print Results   enables to print while analyzer is in operation     3 
43. culate results that have not yet been saved  Select the test  introduce the new FACTOR into the  appropriate window and press OK     EE       2 Select test  insert a factor and then press OK Cancel  to recalculate   Warming  the selected test will be declared vs  factor   ALT       2 5 4    Note     The new Factor will be also memorised in the method  If the method was originally against a    Standard     it  will be automatically modified and become against FACTOR     6  Report Definition  Enables to personalise the print of the final Patient Report     Fle Registers Tods 2 2 21 05 501 07 PM  ax  Pinter Font Name  DEES    x     Ge    Header       HUMAN Gemary          OC Lato abiy  Body  Use mouse to size columns     Test Name Result Flag Unk  Value Flag Measurament Unit  160 High mgd  Bold Bold Regular  Right Center Left  RH  H    Quality Control        2 5 5    22 Human HumaStar 300 User Manual    2 6 EDIT  The following commands are available     Ele   Edit Graph Navigation 2    x    RAN arm    11 Save  amp  Clear Page   2 Repeat Page  Clear Page    Save  amp  Clear All Samples  Repeat All Samples  Clear All Samples       2 6 1    1  Save Batch for      enables to program a batch   a number of samples for the same test     2  Edit Results       3  Repeat Row   To repeat a Single test or Single Sample   4  Insert Page   5  Delete Page    to delete a page   6  Save and Clear Page     to save and delete results from a page  7  Repeat Page     to repeat all tests of that page
44. d Control to open the left side page  The methods that can be calibrated with that control will be  displayed  including their    Reference Values    and the results obtained during the last calibration  As above  simply  click against the methods to be calibrated  A  V  inside its check box indicates that the method has been  programmed     NOTE  There can be more then one standard for the same method  To make sure that its value is correct     click  on its line and its value will be displayed on the left side page     There can also be one standard solution or Calibrator that is used to calibrate more then one method     simply click  on its line to display its values on the left and choose accordingly     4 3 3 Multi Standard Calibration   Make sure to program each standard concentration individually  Each concentration requires its sample cup and a  place on the Sample Tray 1 to 14  Place the standards with an increasing concentration with the lowest  concentration first     In case of Automatic Geometric Standard pre dilution     place only the Primary Standard followed by the number of  empty cups necessary for the curve dilution     4 3 4 ISE Module Calibration   Only if ISE Module is installed  The calibration of the ISE module has to be done daily  The Calibration is done  together with all the other Standards and Controls  but has to beprogrammed  Make sure there is a  V  in the  corresponding check box  Should the Calibration fail  go to Maintenance  F8   Purge 3 
45. e Edit Graph Navigation 2 2 21 05 539 53PM    Standard Control Details E Calibration    Standard Set Standard Ctl Type preDil Date  Bias Reag bias May23    Alb    o  y    Last value Reference  6    16  ISECAB  std       4 3 1    Calibration   identified with an enlarging lens  it contains         Serum Controls     for a proper operation of the analyzer and for the Monthly Quality Control Program it is  recommended to run daily at least one or two Serum Controls       Standards and Calibrators    4 3 1 The Calibration page  There are 14 Check Boxes corresponding to 14 places on the Sample Calibration reserved to hold the Standards and  Controls  plus four Stats    1  The Calibration page   displays all the Standards and Calibrators programmed in the system  To program a new  Standard   simply click on an empty line on the right to open the left side page  Click on Type     to select  between a Standard or Control     2  Click on Name   using arrows up and down  choose the desired Standard or Control  Click on every Check box  that needs calibration     3  Click on its programmed position on the right to confirm  The name and value of the programmed test will be  displayed  A  V  mark inside the check box indicates that it has been programmed     Note  A method can be calibrated with more then one Standard or Calibrator  but the last one will be used for  calculation     50 Human HumaStar 300 User Manual    4 3 2 To program a Serum Control for Calibration   Click on the desire
46. e Level Sensor sometimes     Check wash solution     if volume is low  air bubbles enter the  does not work properly     flags    no system and may produce errors in the Level sensor     fill hydraulics  sample        but there is sample in the with fresh wash solution   cup     Dirty Sample Probe and  or with a sticking drop        Foam or air bubbles on top of the sample      Wash solution has too much foam that disturbs      Check the Level Sensor connection    82 Human HumaStar 300 User Manual    Reagent Level Sensor sometimes does  not work properly     flags     no Reagent         but there is reagent in the container    The same as above for the Sample  Make sure not to shake  but slowly swirl the reagents when    preparing  Shaking reagents creates foam and air bubbles that  disturb proper operation   Make sure that the Reagent Container is not too full  there should    be NO Reagent in the container neck  where the Level Sensor does  work     Human  AP TT       Diagnostics Worldwide  83    7 2 Spare parts and accessories    Part No  OTY Description  17939 1 Adapter for Sample Cups for cat  no  17960 x  17944 1 Adapter for Tubes 12x75  17960 1 1 Sample Tray Segment A for 14 x 100 tubes  17960 2 1 Sample Tray Segment B for 14 x 100 tubes  17960 3 1 Sample Tray Segment C for 14 x 100 tubes  17960 4 1 Sample Tray Segment D for 14 x 100 tubes  17961 1 1 Sample Tray Segment A for 12 5 x 75 tubes  17961 2 1 Sample Tray Segment B for 12 5 x 75 tubes  17961 3 1 Sample Tray Se
47. e Triangle     click to open a scrolling window that displays all the messages     2 10 Sub Programs     F2 a F9  From the Main Menu click on any of the ICONS to access the sub programs     F2     OPERATION                         gt  1  REAGENTS a  User programmable positions of the Reagents  El ON LINE    b  Reagent volume check     number of tests displayed  ii F2 c  Reagent stability displayed and checked  d  Bar Code identification of Reagents  optional     2  CALIBRATION   to calibrate and or standardise the analyzer using standards   calibrators and serum controls   ae  x 3  STATS   program to run Stats  4  SAMPLES   possibility to program up to 240 Patients at a time  a  User programming using keyboard and mouse  i   b  Transfer worklist from host or from diskette   10 c  Identification by built in Bar Code Reader  Optional     F3   ARCHIVE  F3                   Human  PA o    Diagnostics Worldwide  25    is  Archive Selection         Internal Archive Cancel        O Archive on Diskette OK    O Test Statistic     Test Counter     Last Warnings       2 10 1  1  Internal Archive   saved Patient data on hard disk  Data can be checked  printed  transferred or saved   Choose accordingly   2  Archive on diskette   enables the user to find data saved on diskette   3  Statistics   enables calculation of statistical data on a group of results  4  Test Counter   shows the number of test performed for each method  5  Flags  amp  Errors   memorises all flags and errors for 90 day
48. e programmed Incubation 2     Diff  Var    Two Reagents Variable Differential Method   the first measurement M1 is taken 12 sec  before the addition of reagent R2  The final measurement M2 is taken after the programmed  Incubation 2     Post R2   Two Reagents Method   the first measurement M1 is made after   6 sec the addition of  Reagent 2  The final measurement M2 is made after the programmed Incubation 2     Premix   Two Reagent Method   where Reagent 1 can be less then 300 pl  However  the final reagent  volume has to be 300 ul or more  The first measurement M1 is taken 6 sec after the addition of  reagent 2  Reagent Blank   Sample is added and final measurement M2 is taken after the  programmed Incubation 1  Incubation 2 for this method must be 0     32 Human HumaStar 300 User Manual    a  Volume First Reagent  R1   ul    The volume of the first reagent is 300 pul  the only exception is with  method Post R2 and Premix     b  Volume Second Reagent  R2   ul    by default none    This volume is programmable from 3 to 500 ul and  is dispensed after Incubation 1   except Premix   The volume of  R1   R2  should not be more then  600  pl     c  Incubation 1  sec    incubation time after the addition of sample into R1  Click to open and choose one of  the pre programmed optimized times     d  Incubation 2  sec    incubation time after the addition of the second reagent R2  Click to open and  choose one of the pre programmed optimized times     e  Cooling   select    Yes o No    
49. e reagent    3 4  q       2       Read Reagent cuvette  Substrate depletion check M3 ul    Add Reagent  Add Sample    9 40 seconds Weg 10  36  62  88  114  140  166  192  218  244  270  296  322    Reagent preheating to 37  C   About 90 Max Incubation    gt  seconds  first turn plate     Inc  1   114  Read Time   130     1 Add reagent into cuvette   2 Within 9 to 40 seconds Read Reagent cuvette Blank   M1   3 ADD SAMPLE to preheated reagent and mix   4 Measure solution within 10 sec after addition of sample   M2   5 Measure solution again 52 sec later to check for Substrate depletion   M3   6 Incubation of sample   example 114 sec   or any other time    7 Read time   for Kinetic measurement     example 130 sec   or any other time    M4   8 CALCULATION is a Linear Regression based on the 130 sec measurement  multiplied  by its FACTOR     KINETIC TEST   2 reagents    3l 1a  N       M3 o     Read Reagent cuvette blank Mi N  Measure solution within 10    6 N    Check substrate depletion    Add Reagent 2    Add Reagent  Add Sample    9 40 ee A 10  36  62  ss  114  140  166  192  218 a 296  322    Reagent preheating to 37  C   About 90  Max Incubation   322 seconds  first turn plate    e iaa    gt   Inc  1 62    Inc  2 78  Read Time   130   1 Add reagent into cuvette    2 Within 9 to 40 seconds Read Reagent cuvette Blank   M1  3 ADD SAMPLE to preheated reagent and mix  4 Incubation Time 1 of sample    example 62 sec   or any time   5 Add Reagent 2   Starter   6 Measure solution wit
50. ed Time or Kinetic     2  The Optimized Incubation Times are displayed in the Table below   a  Tests with a Single Reagent   the Incubation time indicates the time from the addition of sample to the    final measurement     b  Two Reagent Tests   Incubation 1   is the time from the addition of the Sample to the dispensing of the  Second Reagent  Incubation 2   is the time after the addition of the Second Reagent and the final    measurement     FIRST TURN OF REACTION PLATE    Incubation time 1 in  second    36  62  88  114  140  166  192  218  244  270  296  322    Longer incubation and Read Time    Read Time 8  Incubation    time 2 in second    26   52  78  104  130  156  182  208  234  260  286  312    LONGER INCUBATION AND READ TIME    Incubation time 1 in    Max incubation Timel    491 sec     second    348  374  400  426  452  478  491  517  543  569  595  621    Read Time  amp  Incubation  time 2 in second    Max incubation Time2    585 sec     338  364  390  416  442  481  507  533  559  585  611    It is possible to use a longer incubation and reading times  However  this will slow down the throughput of the    analyzer     Human  PA o    Diagnostics Worldwide    39    3 3 3 Calculated Methods   F4   From the Main Menu click on F4 and select Calculate Methods    These are methods whose results are derived from a mathematical formula calculation based on results from other  tests  An example is A G Ratio     File Edit Graph Navigation 2    Alb Glob    FORMULA  Expressi
51. elate the results with other methods    or instruments  By default  the Slope is 1 and Intercept is 0     3 3 1 To Introduce or Delete a Method    6     From the Main Menu click  Tools   then  Expert Mode   Introduce your Password   Click on F4    Methods to open the method Book     Click  Edit  above  A window will open that enables to      Insert page    to introduce a new method      Delete Page    to cancel an existing method        Insert Page        a new page will be opened  Enter a short name or Code  max 6 characters   Click on the Code  and program the new Method as described above  all 3 pages  Click on another method or close the book to  automatically save the new method     To delete an existing method  select the method to be cancelled and click on    Delete Page     Awindow will  open to request confirmation to cancel  Answer Yes o No  accordingly     To Exit program  click on    Tools    and choose accordingly     3 3 2 Examples of how to Program the Methods     1a  End Point   Single Reagent  1  Volume of Sample     from 3 ul to 70 ul with a 1 ul increment     2  Pre dilution Ratio   only if needed  See the 15 pre programmed Ratios    3  Features    None    4  Volume Reagent 1     from 300 ul to 600 ul    5  Incubation Time 1     Insert incubation time from table from 36 to 621 sec   6  Select if Reagent needs refrigeration    7  Enter Stability of Reagent or 999    8  Enter the Lot number   only if necessary    9  Enter Bottle Type   1 or 2     10  Enter Re
52. ents     xi       Abs ContrNormal   BUN    431 96 8 1485 1743 196 2251 2509 2767 3025 3288    Cancel   Print   Data details  Ai 43 1 1 3475 0 0000    3 3 5    Program as above for Single or Two reagents  with the exception of     12  Type   enter Fixed Time   13  Read Time  sec    enter the time from the optimized table  from 26 to 416 sec   15  Enter the value for substr  Depl min    the default value is 300    16  Calculation   enter Factor  Standard  Multi standard  etc   All other parameters remain the same   See graphic Presentation of the Method in Addendum     Human  z   ze    Diagnostics Worldwide  ER    4  Kinetic     Single or Two Reagents Determination   the result is based on many readings and calculated by linear regression    and linearity checks      xl       Abs ContrNormal   AST  al    1 5425        1 4850   1        1 3125     A  1 1975     1 1400  38C   41 8 93 7 147 1 198 4 2258 251 7 2775 303 3 3291 355 3813    Cancel   Print   Data details   E 41 8 1 4777 0 0000    3 3 6   1  Program as above for Single or Two Reagents with exception of    12  Type  enter Kinetic    13  Read Time  enter the time from the optimized table from 26 to 416 sec    14  Reagent Blank Check  mAbs      Enter the values since they are checked to determine if the reaction is  positive or negative  By default the Negative Reaction will have a minimum Reagent Blank higher then  900 mAbs  The Positive Reaction Kinetic should have a Min  Reagent Blank lower then 700 mAbs    15  Subst  D
53. epl  min   enter the value in mAbs for maximum linearity check   It will depend on the linearity of the method  measurement time and volumes of sample and reagent used   When the measured mAbs min is higher then programmed  the result will be flagged as  SDP    Substrate  Depletion    16  Calculation   although mostly determined by a Factor  it can also be determined by a Standard  If Factor is    used  enter its value on the line below   All other parameters remain the same   See graphic Presentation of the Method in Addendum     5  Multi Standard o Log Logit    1   2     They can be End Points  Fixed Time o Kinetic   Program all parameters as above  with special attention to     a  The method has to be calibrated with at least 2 and usually with 6 standards with different  concentration  A maximum of 11 different concentration standards can be used for a Calibration Curve     b  Make sure to program each Standard concentration individually    c  The Standards have to be introduced with an increasing concentration   All other parameters remain the same   See graphic Presentation of the Method in Addendum     38 Human HumaStar 300 User Manual    6  Incubation and Measurement Times    Both Incubation and the Measurement times have been optimized for maximum efficiency and they are based on    the operational cycle of the analyzer which is 12 sec     1  Maximum throughput of the Analyzer is obtained when running Single Reagent Chemistries  independent of  type   End Points  Fix
54. es  syringes  glassware  fuses  diskettes  tubing etc     The HUMAN representative shall be relieved of any liability under this warranty if the product is not used in  accordance with the manufacturers instructions  altered in any way not specified by HUMAN  not regularly  maintained  used with equipment not approved by HUMAN or used for purposes for which it was not designed     HUMAN shall be relieved of any obligation under this warranty  unless a completed installation   warranty  registration form is received by HUMAN within 15 days of installation of this product     This warranty does not apply to damages incurred in shipment of goods  Any damage so incurred shall be re ported  to the freight carrier for settlement or claim     3 INTENDED USE OF THE INSTRUMENT    The instrument has to be used for the expected purposes and in perfect technical conditions  by qualified  personnel  in working conditions and maintenance operations as described in this manual  according to the  GENERAL SAFETY WARNINGS  This manual contains instructions for professional qualified operators     4 GENERAL SAFETY WARNINGS    Use only chemical reagents and accessories specified and supplied by HUMAN and or mentioned in this manual   Place the product so that it has proper ventilation     The instrument should be installed on a stationary flat working surface  free from vibrations   Do not operate in area with excessive dust    Work at room temperature and humidity  according to the specificatio
55. esired Sample in the Worklist  1 to 240    2  Onthe left are displayed all the results of the given Patient     click on a Test result     a  Click on    Edit    and    Repeat Row      that single test will be repeated    b  Click on the desired Sample or Patient on the right  and    Repeat Row        all the tests for that given Patient will  be repeated    c  Click on the Patient on the right and then    Repeat Page        All 10 samples for that given test will be repeated      Click on the Patient on the right and then    Repeat all Samples        all tests on that page will be repeated     Human  I o        Diagnostics Worldwide  55    4 6 3 To Recalculate a Test against a New Factor    GC Select test  insert a factor and then press OK         to recalculate    Warming  the selected test will bz declared vs   factor     LUC v   fi OK         4 6 3 1    It is possible to recalculate the results of a whole test by introducing a new Factor   a  From the Main Menu click on Tools      Click on    Recalculate Result         A window will open to introduce the New Factor             o    an    Introduce the Code Name for the Test to be recalculated     e  Introduce the New Factor and press OK     Note  All results  obtained that way  will be flagged with the word Edit  indicating that these are edited  Results  The method     in case it was previously against a Standard   will be corrected in the  methodology against    Factor    and its new factor introduced     4 7 Print of a
56. gment C for 12 5 x 75 tubes  17961 4 1 Sample Tray Segment D for 12 5 x 75 tubes  17931 1 1 Sample Cup Adapter for cat  no  17961 x  18720 31 1 Plastic Sample Cups   Bag of 2000    18971 1 Wash Additive  WAD  4 x 25 ml   17942 1 Container for ISE Standard A   17903 10 1 Compression Seals  5 pcs     17903 11 1 Sensor O rings  4 pcs     17903 1 1 Na Electrode   17903 2 1 K Electrode   17903 3 1 Cl Electrode   17903 4 T Reference Electrode   17903 5 1 Standard A  500ml    17903 6 1 Standard B  100ml    17903 7 1 Cleaner  125ml    17903 8 1 Urine Diluent  125ml    17903 9 1 Pump Cassette   17915 1 1 Syringe with Piston   for Sample Diluter  17915 2 1 O ring for Syringe No 4  No 5  4 pcs   17916 1 1 Syringe with Piston   Reagent Diluter  17916 2 1 Teflon Tubing with Connector for Probe  17934 1 Kit Tubing for peristaltic pump  P 3  7  8   17933 1 1 Pump Tube Type 1 for P1  2  4  5  1 pcs    17953 1 Tube Connector for pumps P1  P2  17920 32 1 Photometer Lamp   17938 S 1 Reagent sample probe   17916 S 1 Teflon tubing w connector for probe  17950 5 1 Reagent container 50 ml with cap  17905 5 1 Reagent container 5 ml with cap  17906 5 1 Adapter for reagent container 5 ml  17950 6 1 Extractor for reagent container 50  amp  5 ml    84 Human HumaStar 300 User Manual    7 3    Suggestions and Warnings    SUGGESTIONS    To avoid to decrease the throughput of the analyzer we suggest to program methods with a total time lower  than 322 sec     2  Before the start of the daily work  refill co
57. going to be transmitted  its format is 5 digits with  leading space     DataRecord is described in the Uploading and Downloading chapters  Each field in the DataRecord must be distinct  from the others with a separator char          ASCII 124   hex 7C      CheckSum is computed by adding the ASCII value of each pack characters from the HeadChar up to  but not  including  the     before the CheckSum itself  It is transmitted in 4 digits in hex format with leading zeroes  e  g   OEA2   Following is an example of CheckSum for a download prologue pack     P 20122   Char ASCI Dec ASCI Hex  P 80 50  CH 32 20  a 32 20  Sr 32 20  ir 32 20  we 32 20  2    50 32  290 122    CR stands for Carriage Return  ASCII 13   hex OD     Each transmission must start with a Prologue pack that indicates the total number of line to follow  Example     Prologue HeadChar              2            CheckSum   CR  Data HeadChar            DataRecord           CheckSum   CR    66 Human HumaStar 300 User Manual    Data HeadChar            DataRecord           CheckSum   CR    5 4 4 To Transfer a Worklist  Downloading from HOST or from a diskette   The HeadChar is  P  and the DataRecord are formatted from the Table below     Parameters Dimension Max Values acceptable  PID  1  13 chars Empty   Alphanumeric  Family Name  2  16 chars Empty   Alphanumeric  First Name 12 chars Empty   Alphanumeric  Age 2digit Empty   Number  Sex 1 digit O   Female   1  Male  Accept Time 4 chars Format    HHMM     Comments 22 chars
58. h     1  Patient ID   2  Last Name   3  First Name   4  Comments  open line to any desired comments   5  Sex  amp  Age  use arrows to choose sex    6  Introduce Age   7  Type  amp  Time   8  Choose between     Serum Plasma  CSF or Urine     2  Below are displayed all the programmed methods in the system  Choose the methods by clicking on its check  box  Make sure that the programmed method has the    v    inside  Once all the desired methods have been  programmed for a Patient  click on a new line on the right for the next patient  The data introduced is  automatically saved     4 4 1 To Program Patients using Profiles  From the main window click    Options     A window will open     click on    Use Profiles     All the programmed Profiles  will to be displayed during Patient Entry on the right side  below its 10 Patients     Human  P    Diagnostics Worldwide  53    4 5 To Programa Stat    De Edt Graph Navigation 2        11 47 00 AM  AutoCheck OK   4 5 1    The procedure is exactly the same as programming a  Patient sample  except it is done on the special STAT  page  The page is marked with   a Red Cross with a  Syringe on it  There are four places reserved for STATS  marked A  B  C and D  Program and place the 4  samples into their corresponding positions on the  Sample Tray  These samples have a preference and  will be processed immediately     ki    54 Human HumaStar 300 User Manual    4 6 ToRun Patient Samples    w  Start Operations           Searching operations  H    
59. he data and or Clear the pages for the next run   a single page or all the samples    To transfer or receive data   click on FILE to open the menu  Communication  that enables to move  to receive and  to transfer data    For details see Chapter 2    Click on F3 to open the BOOK on ARCHIVE  which offers a number of important functions     For details see Chapter 3 2     90 Human HumaStar 300 User Manual         PRINT RESULTS     3    Results are automatically printed during operation per Patient   However  there is a PRINT program that enables to obtain different print outs that can be of great help to the  operator     Click on FILE from the Main Menu and then PRINT  The following print possibilities are offered   a  Page the right side page on Patient Entry    b  Calibration     data of the last calibration performed    c  Test Pending   a list of all tests programmed but not yet executed   worklist    d  Printout Results     a window will open to display the options to print by     Test     Sample     or Final Patient Report     For details see Chapter 4 7     Human  A TT       Diagnostics Worldwide  91    F4   METHODOLOGIES      Click on F4   METHODOLOGY    There are two types of Methods   1  Regular Methods     with an unlimited number of user programmable methods     2  Calculated Methods     such as Test Ratios  also user programmable  The user may program or modify any  method  both the Regular as well as the Calculated  The Parameters of each Regular method is program
60. heck  Prepare a fresh  solution or use pure dist  Water   Make sure that solution was dispensed into the cuvette   Reaction cuvette dirty    try Blank with another cuvette   Repeat water blank     Go to Maintenance and activate to repeat the  water blank    Check photometer lamp alignment  Dirty solution in the incubation bath     change    Dirty reaction cuvettes  Go to Measuring System maintenance and  activate check box    Clean cuvettes       Reagent volume in cuvette too small  Make sure there is at least a    Human  PA         Diagnostics Worldwide  81    minimum of 300   l  When checking Reagent Blank      Check sample and reagent volumes      Old photometer lamp     change      Badly connected Probes or torn tubing     check      Dirty Sample Probe     clean      Dirty or old wash solution     prepare fresh      Dirty solution in the incubation bath     change      The cuvette Washing station is not working properly  Make sure  there is wash solution and the probes go down all the way to the  bottom of the cuvettes       Make sure that the Air pump is working properly and drying the  cuvettes as it should      Make sure that the Sample Liquid Level Sensor works properly  Serum controls not within expected     Check controls for expiration date and correct lot number  range     Check method  it may have been inadvertently reprogrammed      Make sure that method conforms with reagent suppliers  specifications      Make sure the controls are not deteriorated      Re assa
61. hin 10 sec after addition of Reagent 2   M2  7 Measure solution again after 52 sec   to check for substrate depletion   M3  8 Incubation 2   example 78 sec   or any other time   9 Read time   for Kinetic measurement     example 130 sec   or any other time    M4  10 CALCULATION is a Linear Regression based on the 130 sec measurement  multiplied       Human            Diagnostics Worldwide  101       DIFFERENTIAL TEST   FIXED    1 2 3    fs 7  M2   M3   2 E   g EE  S E ak E     cr   5 St al  amp  be  F S   E  5 E     ro ba UI  X x 5     E 5  3 SE 3  E S   lt  xo  lt  pe x  9 40 seconds Ea 10  36  62  88  114  140  166  192  218  244  270  296  322    Reagent preheating to 37  C   About 90  Max Incubation   322 seconds  first turn plate   dic r Inc  Time 2    1 Add reagent into cuvette  2 Within 9 to 40 seconds Read Reagent cuvette Blank   M1    3 ADD SAMPLE to preheated reagent and mix  4 SAMPLE BLANK is measured automatically   within 10 sec after mixing of sample with the reagent   M2  5 Incubation Time 1   variable   See Incubation Table   6 Add Reagent 2  starter  if necessary  7 Measure final reaction of sample   M3  Set incubation as necessary for the method from the incubation Table 1 and or 2    CALCULATION of RESULT    M3   M2  x FACTOR    DIFFERENTIAL TEST   VARIABLE  USE ONLY WITH TWO REAGENTS     Read Reagent cuvette   M1  Add Reagent 2  starter     Add Reagent   1  blank RB   Add Sample  Read End Point    9 40 seconds    Reagent preheating to 37  C   About 90   Max
62. i standard calibrator should be programmed the same way as a control  where each concentration point  should be individual checked to be programmed     To Introduce a new Standard or Control into the Calibration Page  Click on an empty line from 1 to 14  identified by  an asterisk  to highlight it and to open the left side page     Click the line  Type  on the left and scroll to choose between Standard or Control     All Standards and Controls  programmed in the system are memorized  Click the line  Name    then scroll to find  the desired Standard or Control  The Method or the Control will be identified and the previously programmed  values displayed  Click again the empty position on the right to assign   ts place on the Sample Plate     Once all the Standards and the Serum Controls have been programmed   click the green triangle START  A  confirmation window will open  click OK to Start Calibration    After the calibration is completed  check the results  When in doubt  repeat a standard or control accordingly  The  Controls will be used for the monthly Quality Control Program   The green triangle START will be ON and the analyzer is ready for the next operation     For details see Chapter 4 3     88 Human HumaStar 300 User Manual    PATIENT ENTRY   A    Patient Entry can be done at any time  While the analyzer is working or calibrating  Patient Entry can also be done  via RS 232 interface  by a diskette previously prepared  or simply type in the Patient data from the keyb
63. in to the position highlighted on the  right  The reagent will be programmed     c  The tests are performed in the same sequence as their Reagent position on the Tray     The Reagent On Line Page contains the following information     1  Alittle Check Box with a number     identifies the position on the Reagent Rack  Reagent     the name or code of the reagent   Test     indicating the number of tests available in the reagent bottle   Stb hrs      stability of the reagent in hours   Bottle type   1 50mlor 2 5 ml   Cool    the    Yes    indicates that the reagent should be placed in the Tray     WN    ul    9   gt      e e    Note  The Tray is cooled to maintain better the reagent stability  It is not a refrigerator and is turned OFF  together with the analyzer     7  To Exit     click on    Navigation      than click again on    Close register    or simply click on the Book cover    4 2 1 Check Reagent Volumes for the Daily Run                lig Fo DE Upeai wm   Searching opershors  Ce F3   Na martenance request   3  no warrungs present     amp  F4 Reagent Volume Chack   Press OK to conti and stat operation  SE  q    Se  e     CW   Ready A   E  gt   EPM  Option  changed          4 2 1 1    Open the BOOK to Reagents On Line and make sure that all the reagents that are to be used for the day are  programmed  there is a Y check in its box      1  When the GREEN triangle below is ON     click START     gt     2  A window will open  Check Reagent Volume  Click OK to START Reagent
64. indicate what is being transferred or received     ATTENTION     The system will transmit all the data that it holds in memory  It is suggested to cancel all the old  results before transferring new data  To cancel results after having transferred them  go to F2    Work List     Choose  Modify and then click on    Save and delete all Samples or clear all samples        18 Human HumaStar 300 User Manual    2  Tosave a monthly Archive    Enables to transfer data by month from the Hard Disk onto a floppy diskette  It is suggested to use a separate  diskette for each month     21x       Select month to move  insert a diskette Fr  into floppy drive and then press OK to  continue     My  5        3  Print  There are the following options     a  Page  page shown in the book   b  All Pages   like all pages of the methods   c  Calibration     Results of the last calibration performed    d  Test Pending  All programmed tests not yet performed or Work List   e  Print Result     results for a given work list  A window will open     Follow instructions as indicated       Results Printout Selection x   L          O List by Test  O List by Sample    O Print by Report               From Position  11 X    To Position           2 3 4    f  Sample Reports     print a selected number of reports from a Worklist or from Archive     4  Printer Setting  enables to install or choose a new printer    5  Exit  to end operation and to Turn OFF the Analyzer    Human  P    Diagnostics Worldwide  19    2 4
65. ing check boxes     c  To Exit   click on the book cover or on  Navigation  to close the book  The Profile is automatically saved     3 4 2 To Delete a Profile    Simply highlight the Profile to be cancelled and press key  Canc  on the keyboard  A window will open for  confirmation   press Yes or No accordingly     Human  A ze    Diagnostics Worldwide  41    3 5 F6  Standards    From the Main Menu click on Standards     F6 to open its page  There is no limit to the number of Standards and  Calibrators that can be programmed     Fie Edt Graph Navigation 2  Standard Definition    Relaranca Lot MunStd ID  Expiration date 02 21 10    Avalable tests  ALB Method Vabe Factor Last Date  ALK  GLUC 100  ALT  AMY  AST  BiDe  BILTD  Bill ot  BUN  CAL  CHE  CHOL  CKNAC  CREAT  D BILI  GGT  IRON  LOH       3 5 1 To Program a New Standard   Click on  Edit  and then    Insert Page     In the window enter the name of the Standard  its lot number  if available   and the number of calibration points  On the right select the method  insert the expiration date and click on the  line with the asterisk      On the left side are displayed all the available methods  Click on the method and then on  the right side on the asterisk      On the top left enter the concentration for the Standard  Then click on the right to  save the value  If it is a Multi calibrator  it is possible to enter several methods that are standardized against a  single Calibrator     Note  In case of a Calibration curve all the 
66. ing codes  EEN 59  5 1 2 To identify the samples with the Bar Code  EEN 59  52   Reagent ldentification u    ee men 61  A UE 62  Sisal   OOMEN so EAEAN EE EAE AAA 62    ii iii Human HumaStar 300 User Manual    5 3 2 AAA 63            5 3 3 Important NOTES   eussssasseissascassaseassnssassnsssassasssassasscassnsscassnsscnssnssnssnsscnssnschsehdesnssnnssnssnsshsehssshssnsschsehssshsed  esdsehisnnsshissnsshisteee 63  5 34  Testing Procedure  unse 64  e A A Se SS EAA E ENEE 64  53167  Trouble Sno  tins the SE MOU e AANEREN 64  5 4 Data Exchange with Host Computer EEN 66  5 41  OA US A to 66  5 4 2 Connection from Host Computer To HumaStar 200 ENEE 66  E RO 66  544 To Transfer a Worklist  Downloading from HOST or from a diskette  inicie  67  5 4 5 Transfer Results  Transmission to HOST or Diskette        ccecccsssssccsssseecssssssccsssssecssssuseccssssessessusessssssecsssseecsessseecsssseees 68  6 Maintenance RR sn 71  6 1      Reagent SyStem ii a A 71  6 2 Sample System EEN 72  EE RE EE E A ee ee 73  6 4 A TO 73  6 5 Maintenance every TWO weeks EEN 74  6 6 Operation to be done every 2 months  74  EE INTEN DEE 75  6 7 1 Change Photometer Lamp  See 6 7 1 1  cocidos 75  6 7 2 Drain and Fill the Incubation bath  see 6 7 2 1   cccccsssssssssssecsssssessssssecssssssecssssscssssssecssssuscssssssccesssssccessueecessueeesssseeesses 76  6 7 3 Check Incubation Temperature 76  6 74    Change the Sampling Probe  See 6 7 4 1  sssini 77  6 7 5 Change Tubing on Peristaltic Pump    78  6 7 6 Repl
67. interventions     HUMAN   Gesellschaft f  r Biochemica und Diagnostica mbH     Max Planck Ring 21   65205 Wiesbaden   Germany    Tel    49 61 22 99 88 0   Fax   49 61 22 99 88 100    e Mail  tech support human de   www human de    Human  PP o    Diagnostics Worldwide    Contents    1    General Description    RR 1  Lo A NR 2  EC EN El Lu E 3  A GBR 4   E dn 4       13 2 Installation    A  13 3 Environmental  Reg ultetretttes ee A  1 34      Operating Temperature DIE eet A  1 3 5 Power Requirements ENEE 4  O A O 5  LAI  veraltet Ee HOENEN 6  15    A RN tits 7  EE e EE 7  1 5 2 Analytical Plate and its incubation thermostatic bath  EEN 8  1 9 3    Cuvette EIN E 9  154  ENEE Washing Kreeser 9  1 55  Samples 10  1 5 6    A RAT 11  o Sample YI eier 12  A EE UE AAA RR 12  15 9  D NE poi Gr SE a RR RR RR RR RR O O 13  1 5 10 N    E 14  1 5 11 Dlterro Eeer 15       2 1   2 2   2 3   2 4   2 5   2 6   227    ET 24   Dido E 25   2 9  Analyzer EH 25   2 10  UR E eer 25   2 11 User Personalized Programs    arar 26   2 12 Important fUNCtIONS  cinc ani 27  TOEAU A A NN 29   E  EE 29   3 2 e ATA O 30   A A A 30  3 341  To Mtroquce orDelete  Method iia OS 35  3 3 2 Examples of how to Program the Methods  ENEE 35  333  E EE 40   ES E Wl 41    34 To prostamia ne WR FONO eege 41    342     To  Delete Oe UE 41    3 5  E E 42  3 51 ToPrograma New Standard AAA TN 42  3 5 2  Mo Deleteormodihya Standard Va  lUe  issnin EEE 42   A AR O QUE  O SPA NPR ROS DO NRO ANNE ANI EN 43  3 6 1 To Program a New 
68. laboratory glassware such as Extran from Merck or any other  similar product  Use it instead of the bi distilled water    2  Use the procedure in Maintenance program to DRAIN and FILL Liquid Bath with the detergent    3  Leave the detergent liquid in the bath for at least an hour    4  Rinse well several times with bi distilled water and finally fill it with good quality bi distilled water     74 Human HumasStar 300 User Manual    6 7    6 7     Analyzer Maintenance  1 Change Photometer Lamp  See 6 7 1 1      Operation to be done with Analyzer OFF     Oy A E         Remove the rear panel    Disconnect the lamp cord from the power supply board   Remove screw holding lamp    Remove knob ADJ Horizz    Remove lamp and place the new one      Note  do not touch the glass part of the lamp with your fingers     use the plastic envelope to hold the glass  part     Proceed in reverse to lock well the screw     8  Turn ON the Analyzer     10   11   12     13     14     Pipette 500 ul of distilled Water into cuvette 38    Go to Maintenance  F8      Reading System  select    Photometer Test       Click INIT to initialize the system    Click START  Cuvette N   38 will position itself in front of the photometer and measurement will start at 340  nm for approximately 80 sec       The   Transmission will be displayed  Adjust the horizontal knob  loosen and rotate the vertical cam 1 or 2 to  obtain an optimum value of 60 e 90    the measurement time will stop when the   Transmission will be  s
69. lick on the desired reagent on the left   then click again into a free position on the right  The new desired reagent  will be automatically programmed  A  v  mark is its check box means the reagent is programmed for the next run     The reagent volume in each container is constantly monitored during operation  Before starting the actual run  an  automatic Reagent Volume Check is performed  The number of Tests Available in every reagent container is  displayed together with other pertinent information     To check the reagent volume  click START  the green triangle  and window will open  If no maintenance or other  operational requests are displayed   click OK to confirm      REAGENT VOLUME CHECK      Human     TT       Diagnostics Worldwide  87    Should the START triangle not be visible   close the Book or change page to display it   The  REAGENTS ON LINE  page will display for all the checked reagents     1  The position of the reagent on the Reagent Plate    2  The number of the tests available in each reagent container  3  The size or type of the container to be used   4  The stability of the reagent in hours   5  If reagent requires cooling or not    6  Double click on a Reagent     to display its last Calibration Curve    Reagent volume checks are made only once a day when programming the daily workload or at any time when  reagents are added or changed    Once the reagent check has been completed     the instrument will be READY for the next operation and the Green  t
70. med on  3 pages     Page 1   includes the name  Units  Reference Values  Linearity Limits and Quality Control checks     Page 2   contains the parameters for sample and reagent volumes  special features  incubation times and the  wavelength filters  etc     Page 3   Contains parameters for Reagent Blank checks  substrate depletion for enzymes  reaction types   measurement  including various calculation modes and linear correlation    The Calculated Methods are programmed separately They offer a standard formula enabling the operator to  program a combination of various ratios that are automatically calculated and results presented in the final Patient  Report     For more details see the Chapter 3 3   There is also a set of methods as an example supplied with the analyzer   Click on F4     METHODOLOGY    Note  To access the Methodology program requires a Password  From the Main Menu click    TOOLS    and then     Expert Mode        a window will open that enables to introduce the password     F4   To Program a New Method   e    Click F4     Methodology to open the Book  after having introduced the password  Click EDIT and again  Insert Page   for new method    A window will open enabling to introduce a short name or code for the new method with maximum of 6  characters    Introduce the new name and click OK  Proceed to program all the parameters for that method  After having  introduced all the data  simply click on another method or close the diary book  The new method will 
71. mmed  These samples can be inserted onto the  analyzer sample tray  at a later time or whenever convenient  To introduce new samples while analyzer is in  operation     press the Intelligent STOP  the red square      For details see Chapter 4 4     STAT OPERATION    If during routine operation it is necessary to run a STAT     click F2   to open the Book  Click on page 3     STAT  which  is marked with a syringe and a red cross     To introduce the STAT sample into the sample plate  it is necessary to STOP the analyzer  Click on the RED square    STOP and wait until the analyzer stops and the green triangle START is lit     This is an intelligent stop   the Reagent probe stops dispensing and the Sample probe finishes its programmed  cycle  The already prepared samples continue to be measured and processed in the reaction plate  No samples is  lost     There are 4 positions on the inside of the sample plate reserved for STATS  They are marked D  C  B and A  to  processed in this sequence     While waiting for the analyzer to STOP  which will take about a minute and half  go to the STAT page and program  the desired STAT samples  Up to 4 STAT samples can be programmed at a time  The procedure is exactly the same  as programming a Patient sample     When the green triangle START is lit  open the plexiglas cover and place the Stat sample into their programmed  positions on the Sample Tray  Click START to continue operation     The START OPERATION window will open  Make sure that 
72. mpletely the container for the pre dilution    3  Use mono reagent methods   4  Do not fill the reagent containers of 50 ml with less than 2 5 ml of reagent and the reagent containers of 5 ml  with less than 0 5 ml of reagent   The reagent could place itself in the corners of the container modifying the  real number of the test     5  To activate LIS communication  select  F9  print ON     LINE   6  There are 4 positions on the inside of the sample plate reserved for STATS  They are marked D  C  B and A  to  processed in this sequence    WARNING    DURING THE OPERATION     A   B   c   D   E   F   G     Do not open the cover in Plexiglas   Do not remove the pre dilution container   Do not remove reagent and sample tray   Do not move the caps of the Wash and Incubation containers also if flagged  wait the END OF RUN    Do not delete METHODS  STANDARDS or an CONTROLS   Do not deselect  ii  STANDARDS and CONTROLS during the calibration    Do not modify the setup of clock and calendar for avoiding to modify the data of calibrations and controls     Human  PA  Seeg    Diagnostics Worldwide  85    7 4  OPERATION    Change distillated water for  incubation analytical chamber  Clean analytical chamber  Cleanm reading cuvettes  Replace syringe type 4 and 5   cylinder and piston    Replace tubing pumps P 3  P7  P8  Replace tubing pumps    P1  P2  P4  P5  Replace pumps cartridge P 3  P7     P8   Replace pump P6   Replace photometer lamp  Clean optical assay   Check probe alignment  Check
73. n cuvettes to assure a non stop operation     DIRECT READING   a built in multi filter photometer measures each sample every 12 seconds at 8 different  wavelengths  There is a reagent cuvette blank for every sample to eliminate the possibility of errors     1 S E  MODULE   is built in for the determination of Na  K and Cl together with the routine chemistries   Optional     AUTOMATIC REPETITION OF RESULTS   results  that are outside of linearity range or suspect of substrate depletion   are automatically repeated with half its sample size     REAGENT VOLUME   only 300 ul of reagent is sufficient to run any test   KEYBOARD     a full keyboard for easy programming and patient data input     USER FRIENDLY SOFTWARE     with graphic presentation  makes the analyzer easy and simple to operate  The  operator is guided step by step through all operations  including an On Line HELP program     ON LINE HELP     a program to help the operator during operation and programming     OPEN SYSTEM     an unlimited number of Chemistries and Immunoassay methods  Profiles  Standards  Calibrators  and Controls are all user programmable  including automatic calculations of Ratio methods     ON LINE QUALITY CONTROL     checks the precision during standardisation and presents graphically a 60 day  Quality Control program with calculated MEAN  Standard Deviation   CV and theor CV      STATS   are processed within 6 minutes without interfering with the routine testing   PATIENT REPORT     are user pers
74. ns listed in this manual   Do not operate this instrument with covers and panels removed     Only use the power cord specified for this product  with the grounding conductor of the power cord connected to  earth ground     Use only the fuse type and rating specified by the manufacturer for this instrument  use of fuses with improper  ratings may pose electrical and fire hazards     To avoid fire or shock hazard  observe all ratings and markings on the instrument    Do not power the instrument in potentially explosive environment or at risk of fire    Prior to cleaning and or maintaining the instrument  switch off the instrument and remove the power cord   For cleaning use only materials specified in this manual  otherwise parts may become damaged    It is recommended always to wear protective apparel and eye protection while using this instrument   Respective warning symbols  if appearing in this manual  should be carefully considered     Human  P        Diagnostics Worldwide    5 DISPOSAL MANAGEMENT CONCEPT    The currently valid local regulations governing disposal must be observed  It is in the responsibility of the user to  arrange proper disposal of the individual components     All parts which may comprise potentially infectious materials have to be disinfected by suitable validated  procedures  autoclaving  chemical treatment  prior to disposal  Applicable local regulations for disposal have to be  carefully observed     The Instruments and electronic accessories  with
75. o check the temperature inside the cuvette  proceed as follows    1  The analyzer should be ON at least 10 min  before starting the test    2  Dispense 500 ul of dist  Water into a reaction cuvette     3  Place a thermometer probe into the cuvette and wait about 90 sec  before starting to measure   Use a  thermometer with a small 1mm thermocouple not to distort the measurement      76 Human HumasStar 300 User Manual    6 7 4 Change the Sampling Probe  See 6 7 4 1     Operation to be done with Analyzer OFF    1  Removethe Sample Probe and remove the screw to lock the probe   2  Disconnect tubing by rotating the TIP anti clockwise    3  Place the new Probe and make sure to lock well the TIP        6 7 4 1  Sampling Arm        Human  2  ze    Fe  Diagnostics Worldwide    77    6 7 5 Change Tubing on Peristaltic Pump   Operation to be done with Analyzer OFF    1  Remove the screws  4    2  Remove the support  5    3  Change tubing  3  cod  No  K10034 01   4  To reassemble  repeat the operation in reverse     Note  Use only the original tubing from Human  Other types of tubing will alter the volumes dispensed and  cause inconveniences in operation of the analyzer     IN       6 7 5 1  Peristaltic Pump    78 Human HumasStar 300 User Manual    6 7 6 Replace the Syringe    Operation to be done with Analyzer OFF    Follow the figure below    1  Remove the plastic locking screw of the Piston  4    Remove the locking screw  5   to easier remove the syringe     UU N         Make sure not
76. oard  Up to  240 Patients can be programmed at a time with an unlimited number of chemistries  However  there are only 30  On Line Reagents at a time     Click F2 to open the Book  Click on page 4  indicated as  1 10   Samples   Each page contains 10 samples  corresponding to one segment on the Sample plate  The number on the page corresponds to the number on the  Sample Segment    To program a Sample   click on the first line  or any line  to open the left side page  that displays the patient   Introduce the patient demographics or sample data     All the available programmed tests or methods in the system are displayed  To choose a methods  click its check  box  A  v  mark inside its check box indicates that the method has been programmed     It is also possible to program by Profile  Instead of programming one test at a time  click on a Profile and then the  corresponding sample number  The data introduced will be automatically saved  To program by Profile activate in  Options F9      Once a certain number of samples has been programmed  place them on the Sample Tray and START the analyzer    Click the green triangle      A START window will open  If no maintenance or other operational requests are displayed  it will identify the Racks  and the sample numbers that are programmed   click OK to confirm  If there are sample that require pre dilution  do not forget to introduce its dilution ratio before START     While the analyzer is working more patient samples can be progra
77. ol   highlight it and simply change its values   To exit     press the book cover  The programmed Controls are automatically saved     For Detailed Information see Chapter 3 6        F7   Quality Control Program      Click F7 to open the BOOK with all the available programmed Controls  All Serum Controls performed during  calibration are automatically memorized for the monthly Quality Control Program     Each method has its own OC page and its own graph  To check  view or print the OC page  click twice on the desired  method to open its OC program pages    The right side displays the values obtained during calibration and the left displays the curve with the values of  Mean   Standard Deviation and  Coefficient of Variation    Click twice on the graph to display a larger version of the graph with its details displayed in the window below   Scroll to display its OC values  the graph and its values    The Q C  values are displayed for only 60 days and are automatically updated daily    The QC sheets and the graphs can be printed    For Details Information see Chapter 3 6     F9   OPTIONS      Click on F9 to open OPTION page     There are a number of options available with the system  Some of those options are part of the standard software   There are others that require the addition of modules or extra parts and are optional not available with the  standard Analyzer system    The Standard Option     can be activated or deactivated directly by the user as necessary  These are   
78. om  time to time     For Detailed Information see Chapter 6     SPECIAL FUNCTIONS   N    There are some special function that apply to allthe programs in the system  there are    Use ofthe mouse     The mouse is an integral part in the operation of the analyzer  All operations with the exception  of Patient and other data Entry are done with the help of the mouse  In the instructions the use of the mouse will  be referred as CLICK     point with the mouse to the indicated item and CLICK to execute the order     Menu Items     On the top right ofthe Main Window  there are 5 special functional menu keys that are an important  part in the operation of the analyzer  Some of these functions are operative only in combination with certain  programs     These are as follows     FILE     It offers a number of sub commands  such as  Communication   Move archive   Print    Printer Set up and  Exit to shut down the system     REGISTERS     contains the 8 programs from F2 to F9     TOOLS     contains the following sub programs  Expert Mode   Instrument Init   Prepper for sample pre  dilution   To  Load Sample Rack   Recalculate Results   Report Definition   Set Date and Time     EDIT     offers 13 functions necessary during programming and operation of the analyzer   NAVIGATION     offers some commands useful during programming     HELP     On Line   Status of the Analyzer     There are 3 important icons on the bottom of the Main Menu that identify the status of  the analyzer     START   
79. on  A  A B    Use A  B  Cor D to indicate below Tests      A Test  ALB  B Test  T P   C Test   D Test        3 3 3 1  To program a test click on    Edit    and then    Insert Page     Enter the name of the new method followed by the  parameters requested on page 1  On page 2  enter the formula to calculate the new method   A total of 4 tests are available to calculate the new method   A  B  C  and D   Example is  A G Ratio  A   Albumin and B   Tot  Protein   On the line  Expression    enter the formula A  A     B      To cancel a method  click on its name  Go to  Modify  and select  Cancel Page        40 Human HumaStar 300 User Manual    3 4 F5   Profiles    From the Main Menu click on Profiles     F5 to open   ts page  On the left are displayed all the available programmed  methods in the system  There is no limit to the number of Profiles that can be programmed     Ele Edt Graph Navigation 2 2 21 05 5 19 02 PM    Name Diabetes    Avalable tests  D BILI   GGT  vi GLUC  vi IRON  v LDH    MAGN    PHOS  CJTBIU  v  TP  Test  Teitfla  v  Ted  TRG  vi UA        15 16 42 PM  Auto Cheek OK  D    3 4 1    3 4 1 To program a new Profile  Click on the asterisk     on the right side page   a  Click on the left on the line Name  enter the name of the new Profile  Go back to right page and click on  the previously selected line to introduce the new name     b  On the left are displayed all the available methods  Choose the desired methods for that Profile  by  clicking on their correspond
80. onalised  printed and automatically saved on a hard disk   BI DIRECTIONAL INTERFACE     via built in a RS 232 C connection to EDP systems     2 Human HumasStar 300 User Manual    1 2 Warranty    HUMAN warrants that instruments  sold by it or by one of its authorised dealers  shall be free of defects in material  or workmanship  provided that this warranty shall apply only to defects which become apparent within one year  from the date of delivery of the new instrument to the purchaser     HUMAN shall replace or repair any defective item at no charge  except for transportation expenses to the point of  repair     This warranty excludes HUMAN from liability to replace any item considered as expendable in the course of normal  usage  e g   lamps  valves  syringes  glassware  fuses  diskettes  tubing etc     HUMAN shall be relieved of any liability under this warranty if the product is not used in accordance with the  manufacturer s instructions  not regularly maintained  used with equipment not approved by HUMAN or used for  purposes for which it was not designed     HUMAN shall be relieved of any obligation under this warranty  unless   1  Acompleted installation warranty registration form is received by HUMAN within 15 days of installation of this  product     2  The Buyer  within the applicable period of time  returns the defective product or part thereof  freight pre paid at  Buyer s expense  to HUMAN     This warranty does not apply to damages incurred in shipment of goods 
81. ons  During installation  at least one person in the laboratory will  be trained in operation and maintenance of the Analyzer     The HumaStar 300 is a highly sophisticated  high precision  sensitive instrument  Proper installation will ensure  optimum performance     1 3 3 Environmental Requirements  The system should be mounted on a table or workbench in an area free from vibration  draughts  dust  strong  magnetic fields or direct sunlight     1 3 4 Operating Temperature Limits  Ambient room temperature from 15   Cto 32  C max  humidity 65   non condensing     1 3 5 Power Requirements   A standard 115 230 V   50 60 Hz   500 VA power is required  as indicated on the back of the instrument  A 2 wire  outlet is used to assure proper electrical grounding    If the laboratory power supply varies by more than 10   It is recommended to install an external stabiliser an On   Line UPS with a minimum rating is 600 VA     4 Human HumaStar 300 User Manual    14 Assembly  The Analyzer is supplied assembled and ready to use  However  it is important that the installation be done by  authorised personnel only  The instrument should be internally examined to check that no damage has occurred to    any of the electronic boards or mechanical parts during shipment     Then proceed to install the external parts as follows  Monitor  Printer  Mouse  Keyboard  Waste Container  Wash  Solution Container  Container of liquid for the Incubation Bath  See Figure 1 4 1     The Monitor and the printer 
82. or  carry over  A standard    B    sol n and a    Clean    sol n cups are placed on the calibration Rack necessary for calibration        5 3 1  ISE MODULE    5 3 1 Some Characteristics   Characteristics       Whole Serum or Plasma is used       Small Sample size     only 60 ul       Electrodes are mounted close to the electronics to minimise electronic noise and improve precision       Sample deposit near the ISE sensors for a rapid electrolyte result        Rapid operation     only 30 sec cycle time        Simple Maintenance       Easily accessible pumps and electrodes for maintenance and service      Two bubble detectors to assure reliability       Two point calibration to assure precision       One point calibration with every sample       Maintenance free Electrodes for 6 months or 10 000 sample warranty      Disposable reference electrodes  no additional filling required     62 Human HumaStar 300 User Manual    5 3 2 Technical Specifications    Sample   Volume Sample   Reproducibility   Serum   Na    K  Cl    Analysis Time     Throughput serum      Serum  plasma or dilute Urine   60 ul  160 ul for diluted Urine  1 10   Maximum imprecision  within run   Typical Carry over     serum     CV lt 1 5   100 160 mmol L   lt 0 5   CV lt 2   3 00 6 00 mmol L   lt 1 5   CV lt 2   80 0 120 0 mmol L   lt 1 0     serum   30 seconds  including one point calibration  urine   60 seconds  including one point calibration    90  x3  samples per hour  urine   45 samples per hour    5 3 3 Im
83. orldwide    9    1 5 5 Sampling Arm   The Sampling arm holds the probe that is connected to the diluter  which aspirates and transfers both the sample  and the reagent into the reaction cuvette and mixes at the same time  During its operating cycle the probe can  move to its various positions   Samples  Reaction cuvette  Washing Position  Sample Pre dilution Container and  into the ISE Module        1 5 5 2  Reagent Arm    10 Human HumaStar 300 User Manual    1 5 6 Reagent compartment  REAGENT CAPACITY   30 On Line reagents   each reagent can be cooled or left at room temperature   The large reagent bottles hold 50 ml and the smaller reagent bottles hold 5 ml     REAGENT COVER          1 5 6 1  Reagent Tray    Human  L   zeg    Diagnostics Worldwide  11    1 5 7 Sample Tray   SAMPLE CAPACITY   40 On Line primary tubes or small sample cups  on 4 independent sector  See Fig  1 4 9 1   plus  20 positions for Standards  Calibrators  Controls and STATs  Each sample has a corresponding place for a cup for  automatic pre dilution if necessary     For details see Chapter 6     PREDILUTION CUP         PRIMARY TUBE         SAMPLE SEGMENT     SAMPLE  SEGMENT       14 PLACES FOR  STANDARD AND CONTROLS    1 5 7 1  Sample Tray    1 5 8 Probe Washing Station   It is of utmost importance that the Probe is properly cleaned after each operation to avoid both carry over and  contamination  This washing station has been especially conceived to wash the probe both internally  P1 and P2   and exte
84. out batteries  power packs etc   must be disposed of according to  the regulations for the disposal of electronic components     Batteries  power packs and similar power source have to be dismounted from electric electronic parts and disposed  off in accordance with applicable local regulations     6 INSTRUMENT DISINFECTION    Analytical instruments for in vitro diagnostic involve the handling of human samples and controls which should be  considered at least potentially infectious  Therefore every part and accessory of the respective instrument which  may have come into contact with such samples must equally be considered as potentially infectious     Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated  parts  Before the instrument is removed from the laboratory for disposal or servicing  it must be  decontaminated disinfected  Decontamination disinfection should be performed by a authorised well trained  personnel  observing all necessary safety precautions  Instruments to be returned have to be accompanied by a  disinfection certificate completed by the responsible laboratory manager  If a disinfection certificate is not  supplied  the returning laboratory will be responsible for charges resulting from non acceptance of the instrument  by the servicing centre  or from authority   s interventions     7 NOTICE    Every effort has been made to avoid errors in text and diagrams  however  HUMAN GmbH assumes no  res
85. ponsibility for any errors which may appear in this publication  It is the policy of HUMAN GmbH to improve  products as new techniques and components become available  HUMAN GmbH therefore has to reserve the right  to change specifications if necessary in the course of such improvements     NOTICE    Analytical instruments for in vitro diagnostic application involve the handling of human samples and controls  which should be considered at least potentially infectious  Therefore every part and accessory of the respective    instrument which may have come into contact with such samples must equally be considered as potentially  infectious     emm    UY    BIOHAZARD       The    BIOHAZARD    warning label must be affixed to instrument prior to first use with biological material      Servicing Note    Before doing any servicing on the instrument it is very important to thoroughly disinfect all possibly contaminated  parts  Before the instrument is removed from the laboratory for disposal or servicing  it must be decontaminated   Decontamination should be performed by authorised well trained personnel only  observing all necessary safety  precautions  Instruments to be returned have to be accompanied by a decontamination certificate completed by  the responsible laboratory manager  If a decontamination certificate is not supplied  the returning laboratory will  be responsible for charges resulting from non acceptance of the instrument by the servicing centre  or from  authority s 
86. portant Notes  1  Power Supply of the ISE Module    The ISE module remains ON at all times  even after the analyzer is turned OFF  The ISE electrodes have to be kept  moist at all time even during the standby periods  Every 30 min  a small volume of Calibrant A is automatically  pumped through the electrodes to keep them moist and ready to operate  The ISE module is directly connected to    the power supply   2  Reagents    1     Calibrant A     Cal  A   is used as reagent wash and one point calibration    is being consumed at a rate of  approx  24 ml a day  That includes 100 samples  of Na  K  and Cl  in an 8 hour day  plus 16 hours of standby     2     Calibrant B       Standard Cal  B     for calibration only      is used as the second point calibration  It is used as a    Standard during calibration     consumption approx  400 ul a day     3  Cleaner     used only once or twice a day to clean the electrodes and to eliminate protein build up  It is used as a    sample during calibration     consumption approx  400 ul a day     4  Urine Diluent    Needed only for the dilution of the Urine samples  Urine samples must be pre diluted 1 10 with    this diluent before testing        ISE MODULE    SERIAL CABLE    5 3 3 1  Diagram of the ISE Module    Method for the ISE Module    The determination of the three methods Sodium  Potassium and Chloride  does not need any special method     Introduce only the Reference Values  allthe rest is programmed inside the module     Human  A TT
87. r control    To fill the smaller container with the liquid for the incubation bath  To empty the Waste Container     then place 20 ml of Sod  Hypochlorite     CAL  A      only if ISE Module is built in  place a full container with Calibrant A next to the module and close  with its liquid sensor     anos       Turn ON the Analyzer   it will perform all the necessary Auto Checks  when the green triangle is lit   the analyzer is  ready for operation     4 2 Reagents  Click on F2     OPERATION  The BOOK will open on the Reagents On Line Page     On the right are displayed all the reagents that have an assigned place on the Reagent Tray  Click the mouse on any  line to open the left side page  On the left are displayed all the remaining programmed methods that have not  been assigned a place on the Reagent Tray     File Edt Graph Navigation 2   Available Reagents Reagents On Line    Set Reagent Test Sib by  Bott Cool E  EA  v2  v3  va   v  5   v  6  vi   EA  EE   w 10  ivi   9  12         py  xe  5 a      8888888888888    Nada y ot pa ya       Human  z   ze    Diagnostics Worldwide  47    To program Reagents for the daily run    1  Simply click on the Check Box for the reagent to be used    2  If the reagent has not yet been assigned a position on the Reagent Rack or it is necessary to change its  position  proceed as follows   a  Click on an empty line with an asterisk Cl on the right to highlight the line     b  Click on the left on the method to be programmed     then click aga
88. riangle START will be ON     For details see Chapter 4 2     SH    F2   CALIBRATION      24  Once the Reagents have been checked     the next step is CALIBRATION   Click the mouse on Operations   F 2  The Book will open Click on the enlarging lens to open the Calibration page     There are 14 positions displayed  corresponding to the 14 places in the center of the Sample Plate reserved for  Standards  Calibrators and Controls  numbered 1 though 14  Places 15 and 16 are reserved for the ISE use only    All the previously used Standards and Controls are displayed  To choose a Standard for calibration  click on its  corresponding check box  A  v  mark inside a check box indicates that the corresponding Standard Calibrator or  Control has been programmed     To check the details of a standard or calibrator  simply click on its name to open the left side page  The standard is  identified including its concentration and its optical density or Factor obtained during its previous calibration     To check the details of a Serum Control  click on its name and the left side page will display all the chemistries  including their reference values as well as the value obtained during the last calibration     To program a Serum Control for Calibration   click the mouse on the desired control to open the left side page  All  the methods are displayed  Click the check box of the methods to be calibrated  A  v  mark inside its box will    indicate that the method has been programmed     A mult
89. rnally by pump P4 and  a final air jet cleans and dries the tip of the probe from any residue of wash  solution     A    Te  ve PSN SS   AD      3         WASH SOLUTION    1 5 8 1  Washing Station    12 Human HumaStar 300 User Manual    1 5 9 Diluter   A high precision Diluter connected to the Probe transfers both the sample and reagent  User programmable  volumes are possible from 3 ul to 70 ul for sample and 3 ul to 600 ul for reagent  with 1ul increments    The peristaltic pumps P1 and P2 mounted in series with the Diluter  washes both the Probe and its connecting  tubing after each operation  The volume of wash has been optimised to assure a perfectly clean operation        1 5 9 1    Human  EE i    Diagnostics Worldwide    13    1 5 10 Peristaltic pumps        4 PERISTALTIC PUMPS  TO RINSE THE PROBE    1 5 10 1    The four peristaltic pumps below serve as follows   P1   Washes the sample tubing and its Probe inside  P2   Washes the reagent tubing and its Probe inside  P4   Washes both probes externally    P5   Solution to wash the cuvettes    14 Human HumaStar 300 User Manual    1 5 11 Photometer  A multi filter photometer using 8 narrow band interference filters has been conceived to offer maximum    sensitivity  accuracy and stability of measurement   The following 8 standard interference filters have been built in  340  380  405  510  546  578  620 and 700 nm     674 020 00       1 5 11 1  Photometer    Human  Ke L   zeg  Diagnostics Worldwide    15    16 Human HumaSt
90. rom 26 to 416 seconds     11  CHECK ABSORBANCE  mAbs     enter the  Min   and  Max    forthe Reagent Blank  These values are important in both calibration and routine testing   especially important in the determination of Enzyme Kinetics  Any anomaly will be flagged     a  Substr  Depl  min   Substrate Depletion used only for Enzyme Kinetic Determinations  Enter the max  value  in mAbs  of the method linearity  Every Kinetic measurement is checked for Substrate Depletion and  Linearity  A measurement is taken 6 sec after the addition of sample into the reagent and again 60 sec  later  If enzyme activity is larger then programmed  it will be flagged as  SBD   When the  Automatic  Repetition of Results  is active  the test will be repeated automatically with half its sample size and result  corrected accordingly     12  CALCULATION  a  Calibrate    click to open the options     a  Factor  b  Standard  c  Multi Standard    34 Human HumaStar 300 User Manual    c  Log Logit  d  Log Logit Cubic    b  Factor  If method is to be determined against a known Factor  enter its value  The tests that are  determined against a Standard  or Calibration Curve   select accordingly  The actual values for  those parameters should be programmed in F6   Standards     c  Reag  Bias Subtr   Select  Yes or No   This is a Reagent Blank correction performed during calibration and  will be subtracted during result calculation     13  LINEAR CORRELATION  enables to introduce a Slope and Intercept to corr
91. s     2 11 User Personalized Programs    F4     METHODS   protected by Password   HumaStar 300 is an open system and  F4 enables the user to program an unlimited number of methods    i F5     PROFILES     unlimited number of user programmable profiles and organ panels      FS for easy patient entry    F6   STANDARDS   unlimited number of user programmable Standards  Calibrators  F6 and Calibration Curves    4 F7     CONTROL     unlimited number of user programmable Controls for an On Line   F7 Quality Control program with graphic presentation of data   F8     MAINTENANCE   enables to check every movable and operating part in the    s FS system  The program is divided into three parts     S Reagent System   S Sampling System   S Reaction Measurement System   Click on its corresponding ICON    In the next Chapters there will be a more detailed description     26 Human HumaStar 300 User Manual    Eh Es    F9     OPTIONS   there are Options that enable the User to activate or deactivate if needed     Upon request the following Options can be built into the system     1  Bar Code Reader for samples and Reagents    2 12 Important functions    1  User programmable calculated methods  Ratio tests    2  Possibility to display calibration curves and Quality Control Curves    3  User programmable graphic print of Patient Report   4  Communication with Host computer in real time   5  Re calculation of data of an entire test    6  Statistics on performed tests    7  Automatic pre dilution
92. scribed above  Also  make sure to identify a given multi standard or calibrator with a Lot number or  some identification that is common to all the different concentrations for a particular calibration curve    To exit   click on the book cover  The programmed Standards are automatically saved     For Detailed Information see Chapter 3 5     Human  PA T                   Diagnostics Worldwide  93       F7     CONTROLS      Click with the mouse on Controls or press key F7 and automatically the Diary Book will open to programmed  Controls    To Program a new Control   click Edit and again  Insert Page  to open a window and to introduce the name of the  new Control    A new page will be opened for that Control  Click on the first line with the asterisk     to open the opposite page  which displays all the available programmed methods  Click on a check box of the desired method  once  highlighted  click again on the line below indicated as  MEAN   Introduce on that line  the Mean value for that  method  including its plus and minus value found on the data sheet for that control  Program one by one all the  methods that are part of that Control    To Delete or Change a Control Value   simply highlight the Control and press key  Delete   A confirmation window  will open  press Yes or No accordingly    To Delete the Page or the Control completely   click on Edit and choose  Delete Page   A window will open for  confirmation  click Yes or No accordingly  To change a value for a Contr
93. ss OK tc confirm and start operation     Coml         4 2 1 2    A window will open to START operation     Start the daily run  on the Sample Rack indicated with the Check Box     Ifa Sample Pre dilution is necessary it will be indicated in the box Pred Ratio     Note  If there are methods in the daily Worklist that require Sample Pre dilution  introduce the Pre dilution       Ratio before START of the analyzer  Only one Sample pre dilution ratio is possible per run  The analyzer  will perform the tests on both the diluted and the non diluted samples     Br  Insert initial and final positions to use prepper   KH select predilution ratio and then press OK  Cancel    From position  1 40   tT  Topostion 1 4   fF  Predil Ratio   12 sl  Samples Ring   internal y        4 2 1 3    Human  z   ze    Diagnostics Worldwide  49    4 2 2 Prepper   the Auto  Pre dilution system   The Sample pre dilution has to be made before starting the daily run  From the Main Menu go to    Tools    and click  on Prepper  A window will open that enables to program the pre dilution ratio as well as the positions of the  samples    a  Samples in the Internal     empty pre dilution cup in the external row   b  Samples in the External     empty pre dilutions cups in internal row    c  Inserta full bottle with pre dilution liquid into its place and click OK    Follow instructions  they are self explanatory     4 3 Calibration  Once the reagents and their volumes have been checked     the next step is     Fil
94. st per hour independent of method used  plus 180 electrolytes at the same time  by a built in ISE module for Na  K and Cl   Optional      SAMPLE CAPACITY   The analyzer is conceived to process up to 240 Patients at a time  with automatic loading and  programming while the analyzer is in operation  40 On Line primary tubes or small sample cups  on 4 independent  sector  plus 20 positions for Standards  Calibrators  Controls and Stats  Each sample has a corresponding place for a  cup for automatic pre dilution if necessary     REAGENT CAPACITY   30 On Line reagents   each reagent can be cooled or left at room temperature     METHODS   an unlimited number of user programmable methods can be programmed with a single or two  reagents  Up to 30 different single reagent methods or 15 two reagent methods can be processed at a time  or a  combination of the two     REAGENT VOLUME   is automatically monitored with an On Line inventory  The number of tests available in each  reagent bottle is displayed and its stability checked  A missing reagent is automatically flagged     AUTOMATIC SAMPLE PRE DILUTION   on request prior to start operation     SAMPLING PROBE   with its built in mixer and liquid level sensor transfers both reagent and sample and  automatically mixes to start reaction     BAR CODE READER   to identify for samples and reagents  Optional      WALK AWAY OPERATION   Once the patients are programmed  all operations are fully automatic including  automatic wash of the reactio
95. sult RPL  Should the new Calibration not be  acceptable  and it is not repeated  for calculation of that method will be used the previously memorized result     4 3 8 To Repeat a Standard or Calibrator  To repeat a STD or Calibrator  it has to be reprogrammed  Go back to the Calibration Page and check the box  V  of  the one to be repeated  All the other standards etc should be deactivated  Press START to repeat calibration     4 3 9 To Repeat a Serum Control  To repeat a Control for one or more methods  simply deactivate their check boxes by removing the  V  on the right  and then reactivate  Deactivate all the ones on the left that do not need repetition  Click START to repeat     4 3 10 To Display the graph of a Calibration or a Kinetic Control    Go to    Reagents on Line    Page and click twice on any one of the Calibrated methods  its corresponding graph will  be displayed     Human  PA o    Diagnostics Worldwide  51    Eds Edt Gaph Nangaton 2 Y 28 65    22202 FM  xl        4 3 10 1    a  The graphs remain in memory until a new Calibration for that method is done or that calibration has been    cancelled  The values of the individual measurement points can be displayed by clicking on their arrows on the  right     b  Note  If the O D  values of a Calibration curve are available  the curve can be displayed in all 3 algorithms as  Multi Standard  Log Logit  or Cubic     c  Itis also possible to view and display the graph of a Patient result  Click twice on a Patient Resul
96. t    then it must  be a valid code      2  Can be empty only if the    P ID    field id not empty     68 Human HumaStar 300 User Manual    Example     R 240138    R 123 Jimmy Denver 77 1  Private Patient O AST 2  77 U L   0  35 200501201512 18DC   R 123 Jimmy Denver 77 1  Private Patient O GLUC 2  145 mg dl   65  105 200501201512 1A77  R 123 Jimmy Denver 77 1  Private Patient O CREAT 2  3 001 mg dl    8  1 4 200501201512 1BOB  R 123 Jimmy Denver 77 1  Private Patient 0 BilBic 2  2 1 mg dl   O  1 2 200501201512 1B28   R 1234 Frank Rossi 18 1  Epat  O BilBic 2  1 6 mg dl   O  1 2 200501201512 16DB   R 1234 Frank Rossi 18 1  Epat  O AST 2  36 U L   0  35 200501201512 1486   R 1234 Frank Rossi 18 1  Epat  0 GLUC 0  102 mg dl   65  105 200501201512 161D   R 1234 Frank Rossi 18 1  Epat  O CREAT 2  1 917 mg dl    8  1 4 200501201512 16C8   R 12345 Elisa Ribas 34 0   O CREAT 2  1 879 mg dl    7  1 2 200501201512 1523   R 12345 Elisa Ribas 34 0   O BilBic 2  1 59 mg dl   0  1 2 200501201512 156A   R 12345 Elisa Ribas 34 0   O AST O  34 U L   0  R 12345 Elisa Ribas 34 0   O GLUC 2  105 mg dl   65  105 200501201512 1479       R 123456 Walter Giotto 88 1   R 123456 Walter Giotto 88 1   R 123456 Walter Giotto 88 1   R 123456 Walter Giotto 88 1   R 1234567 Coppi Sandra 33 0  R 1234567 Coppi Sandra 33 0  R 1234567 Coppi Sandra 33 0  R 1234567 Coppi Sandra 33 0    Cardio O AST O   Cardio O CREAT   Private Patient     Private Patient   Private Patient        Private Patient    35 200501201512
97. t and its graph  will be displayed  Individual Patient graphs are not memorized  when their results are saved in the archive     52 Human HumaStar 300 User Manual    44 Patient Entry    Patient Entry can be done at any time  while the analyzer is working or calibrating  Patient Entry can also be done  via RS 232 serial port interface or by a previously prepared diskette  It can be done using a Bar Code Pen to identify  the sample or simply type in all the Patient data on the keyboard  Up to 240 Patients can be programmed at a time  with a maximum of 30 different methods     De Edt Boh Navigation 2  Sample Data    Sanpe ID 4455 Type pe Note  LastName  Carl 4455 s p  j FutName Smith  Comments Pivate  Sext  ge Male  Tye b Tine  Scrum Pls     ICH 4  K J  CKNAC   Na  v      ALB GL Y  iv ALK v  vi     15 40 45 PM  Auo Check OK        4 4 1       Click on F2 to open the BOOK  Click on page 4  1 10     Samples    or Patient Entry  Each page contains 10 Patients  corresponding to one Rack in the Sample Compartment  The number on the page corresponds to the number on  the Sample tray  Patients can be programmed one by one  by Profiles or Organ Panels or in a Batch mode  Program  the desired Test or Tests and click Edit and then Save for  indicating the number of Patients in that Batch  To  program a Patient   see figure     Click on the first line  or any line  on the right to open the left side page  The display is self explanatory   1  Introduce the Patient demographics  starting wit
98. the STAT check box is activated together with all the  other sample Racks that are still to be processed  Click OK to START the analyzer  The Stat samples are the first  ones to be processed  followed automatically by all the remaining programmed samples     For details see Chapter 4 5     F3     ARCHIVE    Click on F3 to open the BOOK on ARCHIVE  which offers a number of important functions     1  Internal Archive     stores all Patient data on the built7in hard disk  The data can be viewed  printed  transferred  or saved for future reference     2  Archive on a Diskette     enables to find Patient data from a monthly diskette     Human  PA        Diagnostics Worldwide  89    3  Test Statistics     on the work performed  Highlight a batch of results to obtain automatically the calculation of  Mean  Standard Deviation and   CV     4  Test Counter     displays the number and the type of tests performed   5  Last Warning     displays all the warnings and or errors encountered in the past 90 days of analyzer operation     The Patient Archive is stored in an alphabetical order  by date or    D  number  A Patient data can be found by its         Patiente ID       Patient s Last name        Patient   s First name         Date of testing    Make sure to SAVE all the daily work load before turning OFF the analyzer  or prior to start a new workload    To save Patient data   click on OPERATION     F2 and enter the Patient data page Click on EDIT to open the menu that  enables to Save t
99. the Urine samples are always run separately before the routine     5 3 6 Trouble Shooting the ISE Module  In case of difficulty in calibration go to F8     MAINTENANCE  From there it is possible to clean and wash properly the    electrodes  Proceed as follows    1  Open the Sampling System page    2  Activate   ISE Clean and click START  Repeat this operation at least twice   3  Activate     ISE Purge and click START  Repeat this operation at least twice   4  Activate     Calibrate and click START     This operation might be necessary if the Module has not been used for a while or when the electrodes are close to  expiration               Sample Tube        Sample Detector Optically  Detects Air Bubble in Flow    Sample      Na  Electrode    Signal to Microprocessor    pong noir a measures      Na  Potential  mv     Microprocessor measures  K  Potential  mV     K  Electrode    Microprocessor measures    Ci  Electrode  gt       Cr Potential  mV     Fiow ihru dans     Reference Electrode Body  Assam    by   AgCI Coated Silver Pin    High Concentration KCI    Sample Detector Optically   Signal to Microprocessor    Detects Air Bubble in Flow  Pump Tube  Low Resistance Ceramic Plug    lonic Contact thru High Concentration KGI  Internal Filling Solution    d    To P  2 Membrane    5 3 6 1  Electrode Measurement Diagram    For one year operation with HumaStar 300    The following is a spare part list recommended for one year operation  With the built in ISE Module       64 Human Hum
100. times and try again  It is  suggested that after every 50 samples the module should be CLEAN ed and PURGE See details in Analyzer  Maintenance     4 3 5 Reag  BIAS Subt     If requested  it should be programmed in Position O and the corresponding methods on the left should have their  check boxes with  V   The Bias value will be used in calculation and will compensate all samples  Standards and  Controls accordingly  Make sure that it has been programmed as YES in the Method     4 3 6 Start Calibration  Once all the Standards and the Serum Controls have been programmed     click the GREEN triangle START  A  confirmation window will open  press OK to Start Calibration     4 3 7 Check Calibration   After the calibration is done and printed  Check carefully every value and make sure that all the results are within  the limits of the programmed controls  There is an automatic check done by the analyzer against the previous  Calibration  If the new result varies more then 10  from the previous Calibration the result will be automatically  flagged as HIGH or LOW  When in doubt  repeat a standard or control accordingly  The Controls will be used for the  monthly Quality Control Program     The Standardization of a method can be made on a single Standard or that standard in triplicate from the same  cup  A Mean will be calculated and used for calculation  A minimum of 2 results have to be within 3    otherwise it  will be repeated  When printed there will be a Warning against that re
101. tion   2  Make sure that in the Method has been entered the Identification Code for each reagent    3  Should the Code not be programmed  it will not be identified  even though the bottle has a correct label   Proceed to assign manually the position     4  Open the Reagent Tray and insert the bottle  The Reagents positions are automatically identified    5  All information regarding the Reagents is displayed on    Reagents On Line    page    6  Ifthe Code is not legible  it will not be accepted  To use that Reagent  it must be reprogrammed in the Method   7  If the Code on the Reagent Bottle does not correspond to the one programmed in the Method  the following    Warning will be displayed        Error in Reagent Bar Code     impossible to identify the following Reagents   Example   GLU  CRE    etc    The Reagent position will be deactivated and it will be necessary to reprogram its position and recheck again     Human  AI o        Diagnostics Worldwide  61    5 3 Ise Module    HumaStar 300     can be supplied with a built in ISE Module for the determination of Sodium   Na  Potassium   K  and Chloride     Cl in serum  plasma and Urine    The ISE Module is calibrated daily together with all the other methods  It requires only a single reagent that is     Calibrator A     The samples to be analyzed are automatically transferred directly into the Module by the Sampling  Arm  After each determination there is an automatic wash with    Cal  A    reagent to eliminate contamination 
102. ubated in a water bath  at 37  C    0 2  C  The temperature in the bath is maintained constant by a thermostat that continuously circulates  the water by means of the pump P6  The measurement of the reaction is made directly inside the cuvettes through  two quartz windows mounted on the circumference of the Incubation chamber  that enables the photometric light  to go though to a detector     a  OL  E TETAS       1 5 2 1  Analytical Plate        e    OUT    1 5 2 2  Incubation Chamber    8 Human HumaStar 300 User Manual    1 5 3  Cuvette Washing System   It is of utmost importance that each cuvette is properly washed and dried to assure reliable results  since both the  reaction and the measurement take place inside that cuvette  To assure perfect cleaning  the cuvette has a circular  design with a spherical closure at the bottom to avoid right angles which are the difficult ones to clean     1 5 4 Cuvette Washing Probe  It   s a four way system  the first one empties the cuvette  the second purges with wash solution several times and  the third to blows air to eliminate any residue from the inside wall of the cuvette     The Probe as it goes down into the cuvette activates the aspirating pump P3 and P8  while wash solution is forced  inside the cuvette several times and re aspirated  Finally forced air is blown into the cuvette to eliminate the    residues of the wash solution        CUVETTE WASHING STATION    1 5 4 1  Cuvette Washing System    Human  z   ze    Diagnostics W
103. ubstituted with an OK  Lock well the screws 3 or 4   If necessary the operation can be repeated without the  dispensing of fresh dist  Water     Check the value of all 8 filters     they all should be between 40 and 90    Remount the back panel     ADJ  Horizz  Vertic       6 7 1 1    Human  P    Diagnostics Worldwide  75    6 7 2 Drain and Fill the Incubation bath  see 6 7 2 1     To be done with Analyzer ON    Drain    1  Make sure that the Waste Tubing is inside the WASTE container    2  Go to Maintenance  F8      Analytical System  select    Drain bath       3  Click    START     This operation takes about 3 minutes and will be over when the START triangle will be green  again    To Fill    1  Make sure the container is full with bi distilled water    2  Select    Fill bath       3  Click START  The operation takes about 3 minutes  and will be over when the START triangle will be green  again    4  Incase there is not enough bi dist  Water  there will be a Warning on the monitor  Add bi dist  Water into the  container and repeat points 2 and 3 above        Se   IN    ouT    OVERFLOW    6 7 2 1            Note  During bath filling the pump will be noisy  The noise will stop the minute the pump will be full with liquid    again     6 7 3 Check Incubation Temperature    Operation to be done with Analyzer ON    The incubation temperature is set to 37  C with the reaction cuvettes filled with 500 ul of dist  Water  The  temperature is continuously checked during operation    T
104. vette  2 Within 9 to 40 seconds Read Reagent cuvette Blank   Mi  3   ADD SAMPLE to preheated reagent and mix  4  Incubation Time 1   variable   See Incubation Table     5 Measure final reaction of sample   M2       Incubation of reactiion can be anywhere from    36 to 322 seconds at intervals of 26 seconds       Set incubation as necessary for the method from the incubation Table 1          CALCULATION OF RESULT     M2   M1  x    FACTOR       END POINT   2 reagents                                                                         1 2 3 5 7  Ex  5 as E   o D  bo v O E     Kor eg  s  EE 3  30  2 X 3e E S  9 40 seconds 10  36  62  88  114  140  166  192  218  244  270  296  322  Reagent preheating to 37  C   About 90  Max Incubation   322 seconds  first turn plate       gt     SD  Inc  1   62  Incubation 2   260   4 6  1 Add reagent into cuvette  2  Within 9 to 40 seconds Read Reagent cuvette Blank   M1  3   ADD SAMPLE to preheated reagent and mix  4  Incubation Time 1   variable   See Incubation Table   Example 62 sec   5 Add REAGENT 2   example after 62 sec   6  Incubation 2   variable   Example   260 sec   7 Measure final reaction   M2          CALCULATION OF RESULT    M2   M1 X FACTOR                                                          Human     T       Diagnostics Worldwide  99    FIXED TIME TEST   Single reagent    Read Reagent    Add Reagent  cuvette blank  RB M1   Add Sample    9 40 seconds    eagent preheating to 37  C   About 90  Max Incubation   322 seconds
105. with a start and end character     Each field should be separated by a          cod  ascii124 cod  hex7C   1AF2  Checksum hex CR  Carriage Return  cod  ascii13 cod  hex0D     Human  I A    Diagnostics Worldwide  67    5 4 5 Transfer Results  Transmission to HOST or Diskette   The HeadChar is  R  and the DataRecord are formatted from the Table below     Transfer to Host DataRecord    Parameters Accepted Values   Length   PID  1  Empty   Alphanumeric    13 chars  Fam  Name  2  Empty   Alphanumeric    16 chars  Name Empty   Alphanumeric     12 chars   Age Empty   Alphanumeric    2 digits  Sex O   Female     1 digit   1   Male  Accept  Time    HHMM    format   4  Comments Empty   Alphanumeric     22 charat   Test Type O   Serum Plasma 1 digit   1   Urine   2  FCS  Test Name Alphanumeric   6 chars   Type of Result  1   To repeat     2 digits   0   OK    1   Result below normal range  2   Result above normal range  3   Optical Density out of range  4   Reagent Blank out of range  5   Substrate depletion  6   Kinetic measurement not linear  7   Standard not available  8   Error in test Log  Logit  9  Negative Result  10   Result above Linearity Limit  11   Missing reagent  12   Missing Sample  13   Error in the ISE module  Result Use dot  as decimal separator  Units of Measurement Alphanumeric  Min Normal Range  Max Normal Range  Date of testing Format Date    YYYYMMDDHHMM        1  Can be empty only if    Fam  Name    field is not empty  but if data must be stored in    Pending Lis
106. y by an alternative system  method or different reagent      If substrates     recommend to re standardize analyzer using fresh  primary standards or calibrators      Dirty solution in the incubation bath     change      Make sure that the incubation bath is at 37  C      Make sure the factors used are correct      Make sure that the values of expected range are for the method  used on this analyzer   Syringe leaks     Check piston      Check the    O        ring      Check the OUT  tubing for blockage      Make sure the Probe is not blocked      Change Syringe and piston  Air bubbles inside the syringe     Make sure the piston is tight and there are no leaks      Make sure that the wash solution is correctly prepared and has  Tween 20 in it   one drop per liter of dist  Water       Make sure that the INLET connection into the syringe is tight or its  tubing torn  Samples are being contaminated     Reagent and or Sample Probe is badly connected or leaking      Missing wash solution or old  dirty or contaminated wash solution      prepare fresh       Dirty cuvettes     wash them first with a detergent and then with  the wash solution      Cuvettes are not being washed properly     check the washing  station      Make sure that the air pump operates correctly      Make sure that the washing probes do not leak      Make sure the washing probes go down properly all the way to the  bottom of the cuvettes during wash      Make sure that the waste is properly eliminated    The sampl
107. zer  check ON OFF switch  Make sure it is set ON LINE  Make sure check box in Options is enabled for Print ON LINE  Check printer cable  Make sure that the check box in Options is Disabled to print On   Line  make sure it is connected to the analyzer  make sure the printer is On Line  Check wash container  it may be empty  Leak or torn tubing     check   change tubing if necessary  Leaking syringe     check and change piston if necessary  Go to Maintenance and activate check box     Fill  Hydraulics    and or    Wash Probes    and click START    Make sure that all tubing is full of wash solution and there are no  air bubbles in the system     check proper operation of the peristaltic pumps  Torn tubing     replace   Check connections to the Probe   Leaking Syringe     replace   Torn tubing in peristaltic pump     change   Check    O    ring in syringe   Check if photometer lamp is ON   Try to read on another wavelength   Check if filter drum moves   Check if water was dispensed to read WATER BLANK    Try to lubricate the screw with very thin oil  Clean syringe with a good laboratory detergent  Change the whole syringe   Check with    Test Diluter    in F8     Maintenance    Squeezed or torn syringe tubing   Loose or badly connected Probe   Blocked probe     clean   Wash solution container empty     check   Make sure there is Wash Solution in the container    Check for leaking piston or blocked tubing   Remove syringe and clean   Dirty or old wash solution used for Blank     c
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
ECUsim 2000 User Guide  A-61750_de    Fostex HP-P1 User's Manual    Copyright © All rights reserved. 
   Failed to retrieve file