Home
        easy users guide to the hamamatsu c5680 streak camera
         Contents
1.  Rob Heathcote  amp  Margaret Notley    2008       OPERATING THE PROGRAM    The software for running the streak camera is called HPD TA 32   Before starting the program all components need to be powered  up  The current version of the software is 7 00    On starting the program the following window will appear     HPD TA 32  High Performance Digital Temporal Analyzer                          Camera type  C4742 95  ORCA  7 PC DIG     Streak devices     oo  Device GP IB Addr  Cancel  C5680 7 M5676 B      DG535 15                                                The camera type should be 4742 95 ORCA  PCDIG                   The streak devices should be C5680 M5676 on GPIB address 8   There may also be a DG535     If you are running the system with a DG535 Stanford Delay  Generator then the GPIB address of the device needs to match it  in the program  If a DG535 needs to be added or the address  changed click the Setup button     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide   9       Science  amp  Technology Facilities Council       Central Laser Facility    Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008    Device control setup     IF setup   cabling       External devices setup    170 addr  Fu DEAK  M GPIB installed at   0 asl Ait as Rate ni    GPIB board type       Keithley    HI    Device C5680 g    P
2. a    Science  amp  Technology Facilities Council       Central Laser Facility  Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008    EASY USERS GUIDE TO THE  HAMAMATSU C5680 STREAK CAMERA    VERSION 1 00       The aim of this manual is to train the user in the basic safe  operation of the Hamamatsu Streak Camera C5680  Not to be  confused with the Hamamatsu Femtosecond Streak Camera   FESCA    For more details on particular functions in the software  please refer to the HPD TA User Manual     Any comments on this manual will be greatly appreciated   Rob Heathcote     September 2006  TRANSPORTATION    The streak camera should only be moved by a member of the  target area staff    Notes  The camera should only be lifted from its base  The handle  at the top is NOT designed to bare the weight of the camera  Take  special care not to put any load on the CCD camera or its mount or  the entrance slit housing  Only move the streak camera once all  cables have been removed     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide   9       a    Science  amp  Technology Facilities Council    Central Laser Facility  Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008    HARDWARE SETUP    The Streak camera components should be wired in the following  way  Once every thing is connected power can be br
3. h sweep rate on the DG535     Only applicable if a DG535 is used     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide   9    
4. he slit can be adjusted with the large screw  on the top of the entrance housing next to the main body  This will  normally not be required     F8 is a panic button which will instantaneously close the shutter  and reduce the gain to zero     In focus mode you will notice a bright horizontal band across the  screen even when the shutter is closed  This is the thermal  background dark current of the cathode  The dimensions of the  cathode are 0 15 mm x 5 33 mm  It is therefore important that  extreme care is taken when operating this camera  as the slit cannot  be moved to another region of the screen if the cathode is    damaged     If you need to capture an image in focus mode then a separate  trigger synchronous with the laser will need to be applied to the  Ext trig on the rear panel of the CCD controller  This is because  the Monitor Out only gives out a signal when the streak camera is  sweeping     Triggering Internal Delays    For the signal to appear at the centre of the screen  the external  trigger has to arrive at the camera at the exact times given in Table    1  before the light hits the slit     Sweep Internal     Accumulative  Range ns 2e ay ns r ns    re  i ae a    11    566      665         Table 1 Internal delay values    This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide     9    Science  amp  Technology Facilitie
5. lugin   M5676         M5679    Connected to  Addr   Streak camera GPIB   8   Use spectrograph    Spectrograph   None   13    Delay generator GPIB  15   Delay  generator Hone    i    i Counter board    Counter board typey     f DT2819    TROS    Connected to       Streak camega status port      A 6536 shutter box    lf il addr   installed at  560    if Use delay generator      Use delay  generator    Setup   Cancel    Use this box to set the correct GPIB  address to the DG535 you are using     Check this box if you are using a  DG535     All other setting should remain as in the diagram above     Once this is been done and the changes have taken place click OK    on the start window     To start the acquisition controller click the camcorder symbol at    the top of the screen     Device  Chromex 25015       Device  DG535       Device  DG535          This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide     9    Science  amp  Technology Facilities Council    Central Laser Facility  Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008             GENERAL NOTES    The entrance housing consists of transmissive input optics that  image the slit on to the cathode  To give the sharpest image for the  wavelength the camera is being operated in the camera focusing  ring may need to be adjusted     The vertical position of t
6. ment in conjunction with this practical use guide     9       a    Science  amp  Technology Facilities Council    Central Laser Facility    Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008    WINDOW EXPLANATIONS        gt  C5680 M5676 control Eg    Parameter name Parameter value       C5680 M5676 control    This window controls the streak camera     Changes the sweep range  Changes the operation mode    Controls the value on the MCP  Maximum of  63     Controls shutter    Changes between allowing single and  continuous triggers    The time after which the shutter automatically  closes when in focus mode    The manufacturer has recommended that the finder mode not be  used as it can easily result in damage to MCP if gain is set  gt 1     LUT control    This tool changes the contrast of the displayed images on the  screen  The asterisk star button in the right hand lower corner auto  scales to the min and max intensity values of the displayed image     C4742 95 12NRB Acquisition Control    This dialog box is the controls for the CCD camera  For most  applications LIVE and ACQUIRE are the only options that will be  needed  For the uses of ANALOG INTEGRATION and  PHOTON COUNTING please refer to the HPD TA manual     LIVE runs the camera either in free run or capture on every trigger     ACQUIRE lets the CCD capture one frame then stops     DG535 Control Window   This only applies for applications where a DG535 can be used     This dialog box allows you t
7. o control a DG535 delay box from the  computer  This has the added benefit that it can be setup to  remember the delay settings for each Sweep Speed     The delays are best altered using the up and down arrow keys  To  move in steps of 10 hold down either Ctrl  Shift or Alt while using  the arrow keys  For steps of 100 hold any combination of 2 of Alt   Shift and Ctrl and for steps of a 1000 hold all 3     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide     9       Science  amp  Technology Facilities Council       Central Laser Facility    Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008    MENU EXPANATIONS        If delays are typed in the return key will need to be pressed before  the value changes     To change the exponent of the range press    E       To change the setup of the DG535  select    manual input    to ON   This should give you all the options of the DG535 as running it    manually     Setup gt C4742 95 12NRB     C4 742 95 127NRB setup    Scan mode         ECD area  C Normal f  1024 x 1024    C External timing    Superpixel C 1280 x 1024    Edge trigger    Level trigger SB imming    fj 2x   iqger source  dxd  BNC C 8xB  D Sub  Digital 1 F    ie  i  i    Scan Area     Re oe   512 x 512      Camera information    C4742 95 127NHE  AD converter  12 bit  Normal type   Available binning  2x2  4
8. ought on  It is  recommended that the power to the controller units and Computer  be brought on individually  Before starting the software make a  note of the GPIB address of SRS Delay box  DG535   Address can  be found by pressing the GPIB menu button on the front panel       SRS Delay Generator DG535               BNC CABLE IN TO EXT  TRIG OF DG535  MAX 15V     OUTPUT BNC CABLE  FROM AB IN TO TRIG IN     TTL   SV SQ CAMERA CABLE    ESEN  BNC CABLE    MONITOR OUT TO  EXT TRIG    SERIAL CABLE          GPIB CABLE  DIGITAL INTERFACE CABLE    GPIB CABLE    When using a DG535 it is very important that as a 50Q load is  specified for the AB output  that a 50Q load  50Q terminator  be  present on  AB output as well     It is not always possible to run the streak camera with a DG535  because of timing issues  The minimum internal delay of a DG535  is 85ns and the maximum internal delay of the streak camera is  83 4ns   the 50ns range   The time between trigger and the light  needs to be greater than the combined cable and internal delays of  the units  If the delays are greater then the DG535 will have to be  exchanged for manual delay boxes such as Hamamatsu C1097     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide   9    a    Science  amp  Technology Facilities Council     lt  gt  Central Laser Facility  Manual edited and published by
9. s Council    Central Laser Facility             Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008    SCREEN LAYOUT    LUT controls  Look Up Table     Changes the way the image is displayed in the image  window  Effectively changes the contrast      HPD TA 32  High Perform ince Digital Temporal Analyzer     File Setup Acquisition Correctidhs Analysis Display Processing Window Info  AE ee   Ee ED fa  LUT Control    1000 2000 3000 4000    Live  0   Zoom x 1     Mi    i    aiii  i       Image window    In    live    the image will be refreshed in the live  window     In acquire each capture will appear in a new  window      ek                      C4742 95 12NRB acquisition control  X      eee T Analog Y Photon  Integration counting             m Exposure time   Exposure   af     gt   fiT ms     Gain  amp  Offset    Gain 4 ro Offset      oof     Streak trigger    Active Set count fi Elapsed         Real time background subtraction  RTBS     Do RT Backsub Get BG data                                    DG535 control C5680 M5676 control C4742 95 12 NRB acquisition  control   This window conttols the This is control for the main streak   Stanford Research Systems camera unit  This window controls the image     SRS  delay generator     This is only present if one can be  used   See hardware setup    acquisition     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equip
10. x4  8x68    OF  Cancel    The C4742 95 12NRB setup dialog box should be set as in the    diagram above        The reason why the CCD ts set to bin 2x2 is because the limiting factor in  resolution tis the streak unit itself  Running the CCD camera at its full  1024x1024 pixels does not gain you any more resolution from the system     This document is for guidance purposes only for use of CLF equipment at CLF labs  Please consult local rules for  health and safety of use of equipment in conjunction with this practical use guide     9    a    Science  amp  Technology Facilities Council       Central Laser Facility  Manual edited and published by Rob Heathcote  amp  Margaret Notley    2008       Setup gt Trigger setup   Trigger method for operate mode    In    Not syncronised    the CCD camera    free runs    and is similar to  the way the camera operates in focus mode     In    Sequential trigger    the streak camera triggers the CCD to  capture  Selecting this prompts you to connect a BNC from the     Monitor Out    on the back of the streak camera to the    Trigger in     on the back of the CCD controller     If you wish to trigger the CCD in focus mode a separate trigger has to be  applied to the back of the CCD controller from another source  In this    circumstance this menu does not apply     Setup gt Device Control Options  The only box that should be selected is the Azzo delay  delay box      Selecting this option allows the computer to remember the time    delay of eac
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Catálogo de compras  Akai AR250P User's Manual  jeudi 28 mars  Tyco H9301 User's Manual  Les données de consommation - Aquitaine Développement Innovation    DBI SALA 14 Ft. Fall Arrest Post Manual  User`s Manual - CollabTask - Collaborative Task Management  Tucano SARM55 mobile phone case      Copyright © All rights reserved. 