1. GrA Not documented P3 OTH LOCK on Keyboard lock enabled see chap 7 11 off Keyboard lock disabled ANM on Animal weighing enabled see chap 7 10 off Animal weighing disabled P4 tAr Restricted taring range Press and current setting will be displayed Using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing Confirm input by 143 SFB BA e 1013 39 English 9 Service maintenance disposal 9 1 Cleaning e Before cleaning disconnect the appliance from the operating voltage e Cleaning is possible by water jet and short time immersion e Do not apply aggressive detergents solvents etc 9 2 Service maintenance The appliance may only be opened by trained service technicians who are authorized by KERN Before opening disconnect from power supply 9 3 Disposal Disposal of packaging and appliance must be carried out by operator according to valid national or regional law of the location where the appliance is used 9 4 Error messages Error message Description Possible causes o e Remove from Maximum load exceeded emove load from scale or reduce preload Orr y Zeroing range exceeded e Object on the weighing due to switching on balance plate or pressing a normally e Overload when zeroing 4 max e Improper adjustment e Damaged weighing cell e Damaged electronics Sep B Value outside the A D e Weighing plate not chan
2. ON See chap 8 or by using the faster option of a key combination gt gt Place the load onto the weighing system and wait until it is fairly stable gt Press and at the same time a signal sounds meaning that the animal weighing function is enabled During the calculation of a mean average you can add or remove loads as the mean average will be continuously updated English gt To disable the animal weighing function press Ca and FE at the same time 7 11 Lock keyboard Go to menu item P3 OTH or P4 OTH gt LOCK see chap 8 and enable disable the keyboard interlock The enabled function will be locked after 10 minutes of inactivity K LCK will be displayed as soon as a key is pressed To cancel locking keep pressed ES and at the same time 2s until U LCK appears 7 12 Display background illumination gt Keep pressed 3s until setbl appears gt Press again and the current setting will be displayed SFB BA e 1013 31 gt Use to select desired setting bl on Background lighting is on continuously bl off Background illumination off bl Auto Automatic background illumination on when weighing plate is loaded gt Save entry by ag or cancel using E 7 13 Automatic switch off function AUTO OFF The instrument will switch off automatically after a set time when the display unit or weighing bridge has been idle gt Keep pressed 3s until setbl
3. EN 24 Call up menu item P5 Unt see chap 8 Press and the first weighing unit with the current setting will be displayed To enable on disable off the displayed weighing unit press 143 Acknowledge with EI The next unit with the current setting will be displayed To enable off disable on the displayed weighing unit press 143 Acknowledge with El Repeat sequence for each weighing unit Note tj and Hj cannot be activated at the same time only either or Return to weighing mode using itch over weighing unit Keep pressed the display changes over to the weighing units activated before e g kg S Ib SFB BA e 1013 7 6 Weighing with tare gt Deposit weighing vessel After successful stop check press the button The zero display and the indicator NET appear The weight of the container is now internally saved gt Weigh the material the net weight will be indicated gt The weight of the weighing container will be displayed as a minus number after removing the weighing container gt The tare procedure can be repeated as many times as necessary for example with initial weighing of several components for a mix add on weighing The limit is reached when the taring range see type plate capacity is full English gt Switch between gross weight and net weight by pressing the Ca key gt To delete the tare value remove load from weighing plate and press E
4. appears gt Call up AUTO OFF function nd gt Press ES current setting appears gt Use a to select desired setting of on AUTO OFF function disabled of 5 Weighing system will switch off after 5 min of 15 Weighing system will switch off after 15 min gt Save entry by 39 or cancel using ES 32 SFB BA e 1013 8 Menu Navigation in the menu Call up menu Switch on instrument and press E during self test Press in turn I Ea rad and the first menu block PO CHK will appear Select menu block With help of El the individual menu items can be selected one after the other Select setting Confirm selected menu item with aa The current setting will be displayed Change settings Acknowledge setting exit the menu Use the navigation keys see chap 2 1 to switch between the available settings gt Save entry by 3 or cancel using E Return to weighing mode To exit menu press ES repeatedly SFB BA e 1013 33 English Overview of not verifiable models Menu BOEK menu Men Available settings explanation Main menu Submenu PO CHK SET H Upper limit value input see chap 7 7 EN SET LO Lower limit value input see chap 7 7 ee with BEEP no Acoustic signal for weighing with tolerance range A switched off chap 7 7 ok Audio sound when load is within tolerance limits n
5. Notes on verified weighing systems Access to conductor plate e Removing verification seals e Howto open the display unit e To adjust access the menu set the jumper J over both pins J 22 SFB BA e 1013 7 Operation 7 1 Start up ON gt Press and the instrument will carry out a self test The instrument is ready for weighing when a weight display appears 7 2 Switching Off gt Press until the display disappears 7 3 Zeroing Resetting to zero corrects the influence of light soiling on the weighing plate Resetting range 2 max The instrument comprises an automatic zero setting function however the instrument can be reset to zero whenever needed as described below gt Remove load from weighing system gt Press El and the zero display as well as the ZERO indicator will appear 7 4 Simple weighing gt Place goods to be weighed on balance gt Wait for stability display STABLE gt Read weighing result Overload warning 1 Overloading exceeding the stated maximum load max of the device minus a possibly existing tare load must be strictly avoided The instrument may be damaged by overloading Exceeding of maximum load is indicated by as well as a signal sound Remove load from weighing system or reduce preload SFB BA e 1013 23 English 7 5 Switch over weighing unit only not verifiable models How to enable weighing units gt gt Sw
6. Printer Modea ar a beta entente aider 42 Continuous data Output serra 42 Instant iii a a ii 43 SFB BA e 1013 3 English 1 Technical Data KERN SFB 10K1HIP SFB 15K5HIPM SFB 20K2HIP Readability d 1g 59 29 Weighing range max 10 kg 15 kg 20 kg Minimum load Min 100 g Verification value e 5g Verification class III Reproducibility 1g 5g 2g Linearity 19 5g 29 adjustment weight not 10g 15kg 20 ig added class Warm up time 30 minutes 10 minutes 30 minutes Stabilization time typical 2 3 sec Weighing unit kg Auto Off Available options 5 15 min Ambient temperature 0 C 40 C 10 C 40 C 0 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 220 V 240 V 50 Hz Power pack secondary voltage 9V 800mA Rechargeable battery Standard Service life background light on for 70 h Service life background light off 90 h Charge time 12 h Dimensions display unit B x Dx H mm 266 x 165 x 96 Weighing surface mm 300 x 240 IP protection IP 67 as per DIN 60529 Interface RS 232 optional SFB BA e 1013 KERN SFB 30K10HIPM SFB 50K5HIP SFB 50K5LHIP Readability d 10g 5g 5g Weighing range max 30 kg 50 kg 50 kg Minimum load Min 200 g Verification value e 10g Verification
7. by using e gt Finish entry with i 2 2 Overview of displays Display Significance n Ee Rechargeable battery very low STABLE Stability display ZERO Zero display GROSS Gross weight NET Net weight AUTO Automatic add up enabled Kg Weighing unit M Adding LED Indicators for weighing with tolerance limits v 10 SFB BA e 1013 3 Basic Information General 3 1 Proper use The balance you purchased is intended to determine the weighing value of material to be weighed It is intended to be used as a non automatic balance i e the material to be weighed is manually and carefully placed in the centre of the weighing plate As soon as a stable weighing value is reached the weighing value can be read 3 2 Improper Use Do not use balance for dynamic weighing In the event that small quantities are removed or added to the material to be weighed incorrect weighing results can be displayed due to the stability compensation in the balance Example Slowly draining fluids from a container on the balance Do not leave permanent load on the weighing plate This may damage the measuring system Impacts and overloading exceeding the stated maximum load max of the weighing plate minus a possibly existing tare load must be strictly avoided Balance may be damage by this Never operate balance in explosive environment The serial version is not explosion
8. enten nend neden 13 Biz TAPO OE 15 6 3 Malns CONNSCUON sr A Ano 18 6 4 Rechargeable battery operation nnnrenresenee eenen snennenenneenvansenenseernedeenn denn nnne 18 6 5 Adjustme acres locos tenerte 18 6 6 Linearisation not verifiable models ananereeareeenenenenennn serre venvenenendennee 21 6 7 METIO EET 22 2 SFB BA e 1013 7 7 1 7 2 7 3 7 4 7 5 7 6 7 7 7 8 7 9 7 10 7 11 7 12 7 13 9 9 1 9 2 9 3 9 4 10 10 1 10 2 10 3 11 e ERE E er eatin ovis 23 Switching Oir ds 23 Zero arends O ends dend E ites 23 Simple WEIGMING sica nd 23 Switch over weighing unit only not verifiable models nonnen 24 Weighing With Tudor titulada 25 Weighing with tolerance range s ccccccneceeeeeeenesseceeeeeeeeeeeseeeseneanseceereeeeeeesennecenees 25 ManualdotalIZI esin ande cna te tt codec ones 28 Automatic adding pinsin etn ereen nd eneen 30 Animal weighing sneren ese salon eter reb ci 31 LOGE KEV DOAN Azeri d 31 Display background illumination coito tte 31 Automatic switch off function AUTO OFF cooooococcccccccconanononnnnnncnonannnnnnnnnnnnnnnnnnnnnnnnnnnnnnos 32 Mentiras 33 Service maintenance disposal aanne nennen enn nnnnnnenennenn 40 Cleaning rare retteketet eneen baren tanend 40 Service maintenance iii dad enden 40 DISPOSAL TEEN 40 Error messages ult le Canta tt ld ed o 40 Data output RS 232C optional nennen eene nenenennenennenenenen 41 Technical Dae it ditadia 41
9. gt After the adjustment the balance will carry out a self test Remove adjusting weight during selftest balance will return into weighing mode automatically An adjusting error or incorrect adjusting weight will be indicated by the error message repeat adjustment procedure O Prior to a new verification the access to the menu block P2 CAL must be 1 locked again by removing the jumper SFB BA e 1013 19 English Not verifiable models gt Call up menu item P3 CALSCalSnoLin see chap 8 gt Acknowledge with Ensure that there are no objects on the weighing plate i Lin i l Wait for stability display then press El gt The currently set adjustment weight will be displayed l Sl if i I wo gt To change by using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing gt Acknowledge with El gt Carefully place adjusting weight in the centre of the weighing plate Wait for stability display then press 143 gt After the adjustment the balance will carry out a self test Remove adjusting weight during selftest balance will return into weighing mode automatically An adjusting error or incorrect adjusting weight will be indicated by the error message repeat adjustment procedure 20 SFB BA e 1013 6 6 Linearisation not verifiable models Linearity shows the greatest deviation of a weight display on the scale to the value of the respective test weig
10. manufacturer SFB BA e 1013 43
11. pack secondary voltage 9V 800mA Rechargeable battery Standard Service life background light on for 70 h Service life background light off 90 h Charge time 12 h Dimensions display unit 266 x 165 x 96 B x Dx H mm Weighing surface mm 300 x 240 400 x 300 IP protection IP 67 as per DIN 60529 Interface optional RS232 SFB BA e 1013 KERN SFB 100K10HIP SFB 120K50HIPM Readability d 10g 50 g Weighing range max 100 kg 120 kg Minimum load Min 1 kg Verification value e 50 y Verification class HI Reproducibility 10g 50g Linearity 10g 50g adjustment weight not 100 ke 120 ke added class Warm up time 30 minutes 10 minutes Stabilization time typical 2 3 sec Weighing unit kg Auto Off Available options 5 15 min Ambient temperature 0 C 40 C 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 220 V 240 V 50 Hz Power pack secondary voltage 9V 800mA Rechargeable battery Standard Service life background light on for 70 h Service life background light off 90 h Charge time 12 h Dimensions display unit B x Dx H mm 266 x 165 x 96 Weighing surface mm 400 x 300 IP protection IP 67 as per DIN 60529 Interface optional RS232 SFB BA e 1013 English x O c Li 2 Appliance ove
12. protected The structure of the balance may not be modified This may lead to incorrect weighing results safety related faults and destruction of the balance The balance may only be used according to the described conditions Other areas of use must be released by KERN in writing 3 3 Warranty Warranty claims shall be voided in case e Our conditions in the operation manual are ignored e The appliance is used outside the described uses e The appliance is modified or opened e Mechanical damage or damage by media liquids natural wear and tear e The appliance is improperly set up or incorrectly electrically connected e The measuring system is overloaded SFB BA e 1013 11 English 3 4 Monitoring of Test Resources In the framework of quality assurance the measuring related properties of the balance and if applicable the testing weight must be checked regularly The responsible user must define a suitable interval as well as type and scope of this test Information is available on KERN s home page www kern sohn com with regard to the monitoring of balance test substances and the test weights required for this In KERN s accredited DKD calibration laboratory test weights and balances may be calibrated return to the national standard fast and at moderate cost 4 Basic Safety Precautions 4 1 Pay attention to the instructions in the Operation Manual Carefully read this operation manual before setup and commissioning even if
13. you are already familiar with KERN balances 4 2 Personnel training The appliance may only be operated and maintained by trained personnel 5 Transportation amp Storage 5 1 Testing upon acceptance When receiving the appliance please check packaging immediately and the appliance itself when unpacking for possible visible damage 5 2 Packaging return transport gt Keep all parts of the original packaging for a possibly required return gt Only use original packaging for returning gt Prior to dispatch disconnect all cables and remove loose mobile parts Reattach possibly supplied transport securing devices Secure all parts such as glass wind screen weighing platform power unit etc against shifting and damage 12 SFB BA e 1013 6 Unpacking and implantation 6 1 Installation Site Location of Use The balances are designed in a way that reliable weighing results are achieved in common conditions of use You will work accurately and fast if you select the right location for your balance On the installation site observe the following e Place the balance on a firm level surface e Avoid extreme heat as well as temperature fluctuation caused by installing next to a radiator or in the direct sunlight e Protect the balance against direct draughts due to open windows and doors e Avoid jarring during weighing e Protect the balance against high humidity vapors and dust English e Do not expose the ba
14. B BA e 1013 P3 CAL Configuration COUNT Display internal resolution DECI Position of the decimal dot DUAL Setting balance type capacity Max and readability d When DUAL is displayed press ES the current setting will appear Select desired setting by L off Single range balance on Dual range balance Confirm by 3 the display r1inc for entering the readability appears in a dual range balance for the first weighing range Press 2 and current setting will be displayed Use ES to select the desired setting and confirm by E Hinc will be displayed Press E Y the display r1CAP for entering capacity will appear at dual range balance for the first range Press E and current setting will be displayed Using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing Acknowledge with ES In a single range balance the entry of capacity readability is finished either in single range balance Press a the unit will return to the menu or In a dual range balance enter readability verification value and capacity of the second weighing range Press ca the display CAP2 for entering the capacity of the second weighing range will appear Press E and current setting will be displayed Using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing Confirm input by p Press the di
15. EE NO KERN amp Sohn GmbH j 8 Ziegelei 1 Phone 49 0 7433 9933 0 a er D 72336 Balingen Fax 49 0 7433 9933 149 email info kern sohn com Internet www kern sohn com Operating instructions Stainless steel scale KERN SFB Version 1 3 02 2011 GB SFB BA e 1113 KERN SFB Version 1 3 02 2011 Operating instructions stainless steel scale Table of Contents T Technical Data PE 4 2 Appliance overview ete 8 2 1 Keyboard OVEINEN rt dato 9 2 1 1 Numeric input via navigation keys nnen nnen enneeren anneer nn nenneerenenneeeen eneen 10 2 2 Overview of displays mas A deel dadde enke e 10 3 Basic Information General sssersesersernsnenssvnnsensnvonnensednnnensncnkkantaen nea 11 3 1 EAK STA EEEN 11 3 2 Improper SEL 11 3 3 Warranty etienne ea tele sleten nd dere eier Be dte el derne aa 11 3 4 Monitoring of Test Resources lt a nn ph ccngecenasapinhetineuteneeleect 12 4 Basic Safety Precauhons an rennende ene 12 4 1 Pay attention to the instructions in the Operation Manual neen 12 4 2 Personnel ral cui liiia 12 5 Transportation amp Storage oommommccccnnnnnnnnnnnnnnnnnnnnnnnnnnnnnnn nan 12 5 1 Testing upon acceptance oooocococoncconcnonncnononononnnnnn non nnnnnnnn nn 12 5 2 Packaging return transport bt ta iaa 12 6 Unpacking and implantation nnnnnnnnnnnnnnnnnnnnnnnnnnnnnennnnennnnnnnnnnnnnnnnnnnnnn 13 6 1 Installation Site Location of Use cion aia 13 6 2 Unpaeking implantaAlOf rme
16. G Audio sound when load is beyond tolerance limits P1 REF A2n0 Automatic zero point correction Autozero by changing the display digits selectable 0 5d 1d 2d 4d Zero point settings OAUto Zero setting range i Ne Load range where the display after switching on the balance is set to zero Selectable 0 2 5 10 20 50 100 OrAGE Zero setting range Load range where the display is set to zero by pressing Co Selectable 0 2 4 10 20 50 100 OtArE Automatic taring on off taring range adjustable in menu item DAuto SPEEd Not documented Zero Zero point setting P2 COM MODE CONT Continuous data output Interface ST1 One output for stable weighing value parameter STC Continuous data output of stable weighing values PR1 Output after pressing PR2 Manual totalizing see chap 7 8 Press and the weighing value will be added to the summation memory and issued AUTO For automatic add up see chap 7 9 This function is used to issue and add individual weighing values automatically to the summation memory on unloading of weighing scale ASK Remote control instructions R Read T Tare Z Zero reset wire Not documented BAUD Available Baudrate 600 1200 2400 4800 9600 Pr TE1 7 bits even parity 701 7 bits odd parity 8n1 8 bits no parity PTYPE tPUP Standard printer setting LP50 Not documented Lab Data output format Prt LANG eng Standard settings English chn 34 SF
17. class III Reproducibility 10g 5g 5g Linearity 10g 5g 5g ademen eht 30 ig 50 ig 50g added class Warm up time 10 minutes 30 minutes 30 minutes Stabilization time typical 2 3 sec Weighing unit kg Auto Off Available options 5 15 min Ambient temperature 10 C 40 C 0 C 40 C 0 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 220 V 240 V 50 Hz Power pack secondary voltage 9V 800mA Rechargeable battery Standard Service life background light on for 70 h Service life background light off 90 h Charge time 12 h Dimensions display unit B x Dx H mm 266 x 165 x 96 Weighing surface mm 300 x 240 400 x 300 IP protection IP 67 as per DIN 60529 Interface optional RS232 SFB BA e 1013 English KERN SFB 60K20HIPM SFB 60K20LHIPM Readability d 209 209 Weighing range max 60 kg 60 kg Minimum load Min 400g 400g Verification value e 20g 20g Verification class HI III Reproducibility 20g 20g Linearity 20g 20g adjustment weight no 60 kg 60 kg added class Warm up time 10 minutes Stabilization time typical 2 3 sec Weighing unit kg Auto Off Available options 5 15 min Ambient temperature 10 C 40 C Moist environment 0 95 non condensing Electric Supply Input voltage 220 V 240 V 50 Hz Power
18. e centre of the weighing plate Wait for stability display then press 143 When Ld 3 is displayed put the third adjustment weight max carefully in the centre of the weighing platform Wait for stability display then press 143 After the adjustment the balance will carry out a self test Remove adjusting weight during selftest balance will return into weighing mode automatically SFB BA e 1013 21 English 6 7 Verification General introduction According to EU directive 90 384 EEC balances must be verified if they are used as follows legally controlled area a For commercial transactions if the price of goods is determined by weighing b For the production of medicines in pharmacies as well as for analyses in the medical and pharmaceutical laboratory c For official purposes d For manufacturing final packages In cases of doubt please contact your local trade in standard Verification notes An EU Qualification Approval is in existence for verified weighing systems If a balance is used where obligation to verify exists as described above it must verified and re verified in regular intervals Reverification is carried out according to relevant national statutory regulations The validity for verification of balances in Germany is e g 2 years The legal regulation of the country where the balance is used must be observed 1 e Verification of the weighing system is invalid without the seal
19. er limit value input see chap 7 7 Weighing with SET LO Lower limit value input see chap 7 7 tolerance range A aj hi BEEP no coustic signal for weighing with tolerance see chap 7 7 range switched off ok Audio sound when load is within tolerance limits ng Audio sound when load is beyond tolerance limits P1 COM MODE CONT Continuous data output ST1 One output for stable weighing value Interface STC Permanent data output more stable parameter Weighing values z g Output after pressing PR2 Manual totalizing see chap 7 8 Press and the weighing value will be added to the summation memory and issued AUTO For automatic add up see chap 7 9 This function is used to issue and add individual weighing values automatically to the summation memory on unloading of weighing scale ASK Remote control instructions R Read T Tare Z Zero reset BAUD Available Baudrate 600 1200 2400 4800 9600 Pr TE1 7 bits even parity 701 7 bits odd parity 8n1 8 bits no parity PTYPE tPUP Standard printer setting LP50 Not documented SFB BA e 1013 37 English P2 CAL Configuration COUNT Display internal resolution DECI Position of the decimal dot DUAL Setting balance type capacity and readability verification value When DUAL is displayed press B the current setting will appear Select desired setting by A off Single range balance on Dual ran
20. ge balance Confirm by Es the display r1inc for entering the readability appears in a dual range balance for the first weighing range Press E current setting will be displayed Use ga to select the desired setting and confirm by Hinc will be displayed Press ES the display r1CAP for entering capacity will appear at dual range balance for the first range Press Co Current setting will be displayed Using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing Acknowledge with E In a single range balance the entry of capacity readability is finished either in single range balance Press CE the unit will return to the menu or In a dual range balance enter readability verification value and capacity of the second weighing range Press ca the display CAP2 for entering the capacity of the second weighing range will appear Press and current setting will be displayed Using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing Confirm input by El Press ES the display inC2 for entering the readability of the second weighing range will appear Press fe and current setting will be displayed rs Y and acknowledge by a Press 5 the unit will return to the menu Please note Max 2 gt Max 1 d2 gt d1 Select desired setting with 38 SFB BA e 1013 CAL Adjustment see chap 6 5
21. ger range installed e Damaged weighing cell e Damaged electronics Should other error messages occur switch balance off and then on again If the error message remains inform manufacturer 40 SFB BA e 1013 10 Data output RS 232C optional Weighing data can be issued according to menu settings either via the RS 232C interface or by pressing via the interface This data exchange is asynchronous using ASCII Code The following conditions must be met to provide successful communication between the weighing balance and the printer e Use a suitable cable to connect the weighing balance to the interface of the printer Faultless operation requires an adequate KERN interface cable e Communication parameters baud rate bits and parity of balance and printer must match For detailed description of interface parameters see chap 8 menu block P1 COM or P2 COM amp 2 10 1 Technical Data D u Connector 25 pin d subminiature bushing Pin 2 input Pin 3 output Pin 5 signal earth Baud rate 600 1200 2400 4800 9600 Parity 8 bits no parity 7 bits even parity 7 bits odd parity SFB BA e 1013 41 10 2 Printer mode e Standard printout weighing data ST Stable value US Instable value GS Gross weight NT Net weight lt lf gt Space line lt lf gt Space line Printout example KERN YKB 01N ST GS 1 000kg Printout summation memory kkkkkkkkkkkkkk lt lf gt Space l
22. gt Use the navigation keys See chap 2 1 1 for entering the lower limit e g 1000 kg the currently active digit will be flashing gt Confirm input by g gt Select SET H using 26 SFB BA e 1013 gt Press E the current setting for the upper limit will be shown gt Use the navigation keys See chap 2 1 1 to set the upper mi In E limit e g 1100 kg the currently active digit will be flashing gt Confirm input by El gt Select BEEP ME gt Press E the current setting for the audio sound will be shown gt Use no ok ng to select the desired setting gt Confirm input by l gt Press ES and the weighing system will go into tolerance weighing mode From here evaluation takes place whether the goods to be weighed are within the two tolerance limits Weighing with tolerance range gt Tare if using a weighing container gt Put on goods to be weighed tolerance control is started 1 e The tolerance control is not active when the weight is under 20d e To delete limits enter a value of 00 000 kg SFB BA e 1013 27 English 7 8 Manual totalizing With this function the individual weighing values are added into the summation memory by pressing and edited when an optional printer is connected eo Menu settings 1 P1 COM or P2 COM gt MODE gt PR2 see chap 8 e The totalisation function is not active when t
23. he scale s maximum weight Info about test weights can be found on the Internet at http www kern sohn com e Observe stable environmental conditions Stabilisation requires a certain warm up time 18 SFB BA e 1013 Verified models In verified balances the access to menu block P2 CAL is locked For menu 1 access both contacts of the printed circuit board must be short circuited by a jumper see chap 6 7 Attention After destruction of the seal the weighing system must be re verified by an authorised agency and a new verification wire seal mark fitted before it can be reused for applications subject to verification gt Switch on balance and during the selftest press gt Press subsequenty Caa B C ne trst menuso POCHF PO CHK will be displayed gt Press repeatedly until P2 CAL will be displayed gt Press ig the first menu item COUNT will be displayed gt Press repeatedly until CAL will be displayed gt Acknowledge with a Ensure that there are no objects on the weighing plate gt Wait for stability display then press El English gt The currently set adjustment weight will be displayed I CI ii it gt To change by using the navigation buttons see chap 2 1 1 select the desired setting the active digit is flashing gt Acknowledge with El gt Carefully place adjusting weight in the centre of the weighing plate Wait for stability display then press El
24. he weight is under 20d Add up gt Place goods to be weighed A Wait until the stability display STABLE appears then press el The weight value will be saved and a printout received if an optional printer is connected Remove the weighed good More weighed goods can only be added when the display zero gt Place goods to be weighed B Wait until the stability display appears then press ej The weight value will be added to the summation memory and possibly printed The number of weighing processes followed by the total weight will be shown for 2 sec gt Add more weighed goods as described before Please note that the weighing system must be unloaded between the individual weighing procedures gt This process can be repeated until the capacity of the weighing system is exhausted Display of the saved weighing data gt Press and the number of weighings followed by the total weight will be shown for 2 sec To receive a printout press 4 during this display 28 SFB BA e 1013 Delete weighing data gt Press p and p at the same time The data in the summation memory are deleted Printout example kkkkkkkkkkkkkkk NO 1 1 000kg NO 2 2 0 500kg kkkkkkkkkkkkkkk NO 3 3 0 700kg kkkkkkkkkkkkkkk 1 First weighing 2 Second weighing 3 Third weighing 65 65 65 6 4 Number of weighings total SFB BA e 1013 29 English 7 9 Automatic adding up With this function
25. ht according to plus and minus over the entire weighing range If linearity deviation is discovered during a testing instrument control you can improve this by means of linearization 1 e In balances with a resolution of gt 15 000 dividing steps carrying out a linearisation is recommended e Carrying out linearization is restricted to specialist staff possessing well acquainted with the workings of weighing scales e The test weights to be used must be adapted to the weighing scale s specifications see chapter testing instruments control e Observe stable environmental conditions Stabilisation requires a certain warm up time e After successful linearisation you will have to carry out calibration see chapter testing instruments control Call up menu item P3 CAL gt Cal gt Liner see chap 8 Confirm by pressing E the password query Pn will be displayed Press subsequently E El fa A Ensure that there are no objects on the weighing plate Wait for stability display then press 143 The adjustment points are freely selectable 10Max Max e g in a balance max 300 kg for the first adjustment point an adjustment weight of 30kg 40 kg 290 kg can be selected When Ld 1 is displayed place the first adjustment weight carefully in the centre of the weighing plate Wait for stability display then press 143 When Ld 2 is displayed place the second adjustment weight carefully in th
26. ine TOTAL NO 3 Number weighing processes TOTAL wgt 0 447KG Total of all individual weighings kkkkkkkkkkkkkk 10 3 Continuous data output con1 Weighing mode HEADER 1 HEADER 2 m WEIGHT DATA r WEIGHT UNIT T ma HEADER1 ST STABLE US UNSTABLE HEADER2 NT NET GS GROSS 42 SFB BA e 1013 11 Instant help In case of an error in the program process briefly turn off the balance and disconnect from power supply The weighing process must then be restarted from the beginning Help Fault The displayed weight does not glow The displayed weight is permanently changing The weighing result is obviously incorrect Possible cause The balance is not switched on Mains power failure mains cable defective Power supply interrupted Rechargeable batteries are inserted incorrectly or empty No rechargeable batteries inserted Draught air movement English Table floor vibrations Weighing plate has contact with other objects Electromagnetic fields static charging choose different location switch off interfering device if possible The display of the balance is not at zero Adjustment is no longer correct Great fluctuations in temperature Warm up time was ignored Electromagnetic fields static charging choose different location switch off interfering device if possible Should other error messages occur switch balance off and then on again If the error message remains inform
27. ing on the weight display H indicates that the battery is getting low Approximately 10 h of instrument usage are left afterwards it will shut off automatically Use the supplied battery charger for charging the battery Charge status of rechargeable battery is indicated by the LED display red Voltage has dropped below prescribed minimum green Rechargeable battery is completely charged yellow Charging storage battery To save battery life you can enable the automatic switch off function AUTO OFF see chap 7 13 6 5 Adjustment As the acceleration value due to gravity is not the same at every location on earth the balance must be coordinated in compliance with the underlying physical weighing principle to the existing acceleration due to gravity at its place of location only if the balance has not already been adjusted to the location in the factory This adjustment process must be carried out for the first commissioning after each change of location as well as in case of fluctuating environment temperature To receive accurate measuring values it is also recommended to adjust the balance periodically in weighing operation 1 e In balances with a resolution of lt 15 000 dividing steps an adjustment is recommended In balances with a resolution of gt 15 000 dividing steps a linearisation is recommended see chap 6 6 e The weight to be used depends on the capacity of the scale Carry out adjustment as near as possible to t
28. itial installation and after each change of work area it is necessary to level the weighing bridge A gt As the air bubble is located under the weighing plate remove it gt Level balance with foot screws until the air bubble of the water balance is in the prescribed circle 14 SFB BA e 1013 6 2 1 Tripod Assemble example models platform size 300 x 240 mm i 2 71 Attach the tripod to the platform acc to fig using the 4 screws 1 securing disks and washers Ensure that the cable is not damaged nor squeezed Screw in support screw 2 till it is safely fixed SFB BA e 1013 15 Remove display unit from holder for that remove the turning knobs 3 on the side Attach the tripod with the four raised counter sunk head screws 4 and the nuts on the holder of the display unit Re attach and position display unit using the turning knobs 3 16 SFB BA e 1013 Scope of delivery models platform size 400 x 300 mm Ez D u O Tripod tube Adapter display unit _ Tripod foot SFB BA e 1013 17 6 3 Mains connection Power is supplied via the external mains adapter The stated voltage value must be the same as the local voltage Only use original KERN mains adapters Using other makes requires consent by KERN 6 4 Rechargeable battery operation Before the first use the battery should be charged by connecting it to the mains power supply for at least 12 hours The symbol appear
29. l 7 7 Weighing with tolerance range You may determine an upper and lower limit for weighing with tolerance limits in order to ensure that the weighed load remains exactly within the fixed tolerance limits During tolerance checks such as dispensing portioning and sorting the instrument will indicate any lower deviation or exceeding of limits with the help of a visual signal or audio sound Acoustic signal The audio sound depends on the setting of the menu block BEEP Options e no Acoustic signal turned off e ok Acoustic signal sounds when load is within tolerance limits e ng Acoustic signal sounds when load is beyond tolerance limits SFB BA e 1013 25 English Optical signal Three colour pilot lamps indicate whether load is within the two tolerance limits The signal lamps provide the following information Goods to be weighed above tolerance limit Red signal lamp glowing v Goods to be weighed within tolerance range Green signal lamp glowing Goods to be weighed below tolerance limit Red signal lamp glowing Settings for tolerance weighing may be set either by calling up menu block PO CHK See chap 8 or by applying the faster option of pressing the key combination Settings gt Press ES and ES at the same time in weighing mode gt Press El the display used for entering the lower limit SET L appears gt Press eel current setting appears
30. lance to strong humidity for extended periods Non permitted condensation condensation of air humidity on the appliance may occur if a cold appliance is taken to a considerably warmer environment In this case acclimatize the disconnected appliance for ca 2 hours at room temperature e Avoid static charge of goods to be weighed or weighing container e Being of protection type IP 67 as per DIN EN 60529 the weighing scale is suitable for short term use in wet conditions Major display deviations incorrect weighing results may be experienced should electromagnetic fields e g due to mobile phones or radio equipment static electricity accumulations or instable power supply occur Change location or remove source of interference 6 2 Unpacking implantation Scope of delivery serial accessories e Balance see chap 2 e Transit Securing e Mains adapter e Rechargeable battery e Instruction Manual Carefully remove the balance from the packaging remove plastic cover assemble the tripod and the display unit see chap 6 2 1 and setup balance at the intended workstation SFB BA e 1013 13 Remove the transportation lock 1 Models platform size 300 x 240 mm Remove the marked screws 2 Models platform size 400 x 300 mm Remove the screw marked by the label EE gt Attention The sealed screws must not be unscrewed Accurate weighing results require a weighing bridge with perfect horizontal alignment During in
31. rview 1 2 3 4 5 6 T Battery status display Keyboard Weight display Tolerance tag see chap 7 7 Weighing unit Levelling screw Spirit level underneath weighing platform SFB BA e 1013 2 1 Keyboard overview Button Function gt Turn on off a3 tr Navigation key Zeroing Confirm entry 63 Navigation key A Taring At numeric input increase flashing digit Scroll forward in menu 65 x Navigation key gt Display sum total Digit selection to the right Qs Navigation key Add weighing value in summation memory Digit selection to the left o Al Calculate weighing data via interface Delete ESC Switch over gross weight net weight Back to menu weighing mode Ns Activate animal weighing function Activate weighing with tolerance limits ap Delete total added memory SFB BA e 1013 English 2 1 1 Numeric input via navigation keys gt Press current setting appears The first digit is flashing and can be changed gt If the first digit is not to be MO press and the second digit will start flashing Each time you press the display unit jumps to the subsequent digit returning to the first digit after the last digit has been pressed gt To change the selected flashing digit press repeatedly until the desired value appears Then select by using additional digits and change these
32. splay inC2 for entering the readability of the second weighing range will appear Press and current setting will be displayed Select desired setting with g and acknowledge by BE Press 1483 the unit will return to the menu Please note Max 2 gt Max 1 d2 gt d1 SFB BA e 1013 35 English CAL noLin Adjustment see chap 6 5 Liner For linearisation see chapter 6 6 GrA Not documented on Keyboard lock enabled see chap 7 11 KA KSA off Keyboard lock disabled ANM on Animal weighing enabled see chap 7 10 off Animal weighing disabled kg on P5 Unt off Switch over Ib on weighing unit off see chap 7 5 OZ on off tJ on off HJ on off P6 xcl Not documented P7 rst Use 143 to reset balance settings to factory default P8 uwb Not documented 36 SFB BA e 1013 Overview of verified models In verified balances the access to P2 CAL and P4 tAr is locked For menu access both contacts of the printed circuit board must be short circuited by a jumper see chap 6 7 Attention After destruction of the seal the weighing system must be re verified by an authorised agency and a new verification wire seal mark fitted before it can be reused for applications subject to verification Menu block Menu nem Available settings explanation Main menu Submenu PO CHK SET H Upp
33. the individual weighing values are automatically added into the summation memory when the balance is unloaded without pressing p and edited when an optional printer is connected e Menu settings P1 COM or P2 COM gt MODE gt AUTO see chap 8 o Indicator AUTO is displayed 1 Add up gt Place goods to be weighed A After the standstill control sounds a signal tone The weighing value is added to the summation memory followed by printing gt Remove the weighed good More weighed goods can only be added when the display zero gt Place goods to be weighed B After the standstill control sounds a signal tone The weighing value is added to the summation memory followed by printing The number of weighings followed by the total weight will be shown for 2 sec gt Add more weighed goods as described before Please note that the weighing system must be unloaded between the individual weighing procedures gt This process can be repeated until the capacity of the weighing system is exhausted O Display and delete the weighing data as well as printout examples 1 see chap 7 8 30 SFB BA e 1013 7 10 Animal weighing The animal weighing function is ideal for unstable loads The weighing system calculates and displays a stable mean average from several weighing values The animal weighing program may either be enabled by calling up menu block P3 OTH or P4 OTH ANM gt