        G1328-90013 - Agilent Technologies
         Contents
1.          Notices       Agilent Technologies  Inc  2006   2011   2012    No part of this manual may be reproduced  in any form or by any means  including elec   tronic storage and retrieval or translation  into a foreign language  without prior agree   ment and written consent from Agilent  Technologies  Inc  as governed by United  States and international copyright laws     Manual Part Number  61328 90013    Edition  10 2012  Printed in Germany    Agilent Technologies  Hewlett Packard Strasse 8  76337 Waldbronn    This product may be used as a com   ponent of an in vitro diagnostic sys   tem if the system is registered with  the appropriate authorities and com   plies with the relevant regulations   Otherwise  it is intended only for gen   eral laboratory use     Warranty    The material contained in this docu   ment is provided    as is     and is sub   ject to being changed  without notice   in future editions  Further  to the max   imum extent permitted by applicable  law  Agilent disclaims all warranties   either express or implied  with regard  to this manual and any information  contained herein  including but not  limited to the implied warranties of  merchantability and fitness for a par   ticular purpose  Agilent shall not be  liable for errors or for incidental or  consequential damages in connection  with the furnishing  use  or perfor   mance of this document or of any  information contained herein  Should  Agilent and the user have a separate  written agreement w
2.  Exchanging the When cable damaged orwhenno 10 minutes See    Position Sensing Switch    on  position sensing start signal is sent when page 32  switch switching to the inject position  26 1260 Infinity Manual Injector    Maintenance 4    Flushing the Manual Injector    CAUTION Damage through crystal formation    The use of aqueous buffers or salt solutions can lead to crystal formation which may  cause scratches on the injection seal      gt  Always rinse the valve with water after using aqueous buffers or salt solutions        1 Switch the valve to the INJECT position   2 Use the pump to flush the sample loop and seal grooves     3 Use the needle  port cleaner  supplied with the valve  and syringe to  flush the needle port and vent capillary                 Needle port        Syringe    Needle port cleaner    Figure 11 Needle port Cleaner    1260 Infinity Manual Injector 27    4 Maintenance    Cleaning the Manual Injector    The manual injector base should be kept clean  Cleaning should be done  with a soft cloth slightly dampened with water or a solution of water and  a mild detergent     28 1260 Infinity Manual Injector    Maintenance    Injection Valve Seal    When e Poor injection volume reproducibility    Leaking injection valve    Tools required Description  Hexagonal key  9 64 inch    Parts required   p n Description  1 5068 0052 Rotor seal  PEEK    4          1 Loosen the three stator screws  2 Remove the stator head                 1260 Infinity Manual Injector    
3.  Sample loops stainless steel    Description  Sample loop 5 pL  Sample loop 10 pL  Sample loop 20 pL  Sample loop 50 pL  Sample loop 100 pL  Sample loop 200 pL  Sample loop 500 pL  Sample loop 1 mL  Sample loop 2 mL    Sample loop 5 mL    Sample loops PEEK    Description  Sample loop 5 pL  Sample loop 10 pL  Sample loop 20 pL  Sample loop 50 pL  Sample loop 100 pL  Sample loop 200 pL  Sample loop 500 pL  Sample loop 1 mL  Sample loop 2 mL    Sample loop 5 mL    1260 Infinity Manual Injector    1260 Infinity Manual Injector    Appendix   J    Agilent Technologies on Internet 42    Additional information    ea Agilent Technologies    41    6 Appendix    Agilent Technologies on Internet    For the latest information on products and services visit our worldwide  web site on the internet at     http   www chem agilent com    42 1260 Infinity Manual Injector    Index    A    Agilent on internet 42  APG remote 13    B  buffers 27    C    capillary connections 14  cleaning the manual injector 28    D   damaged packaging 8  delivery checklist 8  description 6    E  exchanging  injection seal 29  position sensing switch 32  F  flow cell    solvent information 19  flow connections 14  flushing the manual injector 27    information  oninternet 42    injecting sample 6  INJECT 6  22    1260 Infinity Manual Injector    injection seal 29  tefzel 20  vespel 20  installation 10  installing the manual injector 10  internet 42  42    L    leak channel 16  leak drainage 16    leaks 14  LOAD 6  2
4.  of the two vent capillaries  ports 5 and 6  and the needle port must be at  the same level to prevent siphoning  see Figure 7 on page 15         1 Connect the pump outlet capillary to port 2   2 Connect the column compartment inlet capillary to port 3     3 Connect the sample loop between ports 1 and 4     14 1260 Infinity Manual Injector    Installing the Manual Injector 2    low Connections    4 Connect one vent capillary  supplied with valve  to port 5 and one to  port 6     fA    From pump outlet                Sample loop                Waste                               Vent tube          To column          ae     Figure 6 Flow Connections         a Vent capillaries  ports 5 and 6           lt   anaa  Vent capillaries and needle port at the same level    Figure 7 Vent Capillaries    1260 Infinity Manual Injector 15    2 Installing the Manual Injector    Leak Drainage      WARNING   Leaking injector fittings    In the event of a leak  solvent will drop into the leak channel in the baseplate  from  where it is channelled to the front and back of the baseplate      gt  Check the manual injector fittings periodically for signs of leakage                                                                                                                                                                                            Leak channel       Figure 8 Leak Drainage    16 1260 Infinity Manual Injector    1260 Infinity Manual Injector    3  Using the Manual Injector    Warning
5. 2  M    make before break 6    needle type 21  needle port cleaner 27  needles 21    P    position sensing switch 32  position sensor cable 13  precision 22    repair procedures  injection seal 29  position sensing switch 32    S    salt solutions 27    Index    sample loops 6  sample volume 22    solvents  acids 19  buffers 27  ethers 19  rinsing 27    salt solutions 27  steel corrosive 19  start cable 13    U    unpacking  damaged packaging 8  delivery checklist 8    Ww    worldwide web 42    43    www agilent com    In This Book    This manual contains user information about  the Agilent 1260 Infinity Manual Injector   The manual describes the following     introduction to the manual injector   e installing the manual injector   e using the manual injector   maintenance of the manual injector   e parts and materials  and      additional information        Agilent Technologies 2006   2011  2012    Printed in Germany  10 2012    G1328  90013       fe Agilent Technologies    
6. 260 Infinity Manual Injector 37    5    Parts and Materials for Maintenance    Injection Valve Assembly    Item    wo    38    Manual Front Loading Injector Valve  600 bar  5067 4191  with starts  cable  complete assembly   including operating instructions  needle port  cleaner  vent tubes  x2  and fittings  5 64 and 9 64 inch hex keys  Includes  items 1 8     p n   1535 4045  1535 4046  5068 0052  5068 0118  0100 1859  0100 1860  5068 0020  8710 0060  0490 1849  1400 3166    Description   Bearing Ring   Qty 1  replacement   Isolation seal   Rotor seal  PEEK   Stator ring   Stator face   Stator head   Stator Screws  10 pack   Hex key wrench  9 64 inch  Position sensing switch    Ring stand  mounting bracket    1260 Infinity Manual Injector    Parts and Materials for Maintenance 5  Injection Valve Assembly             Figure 13    _Injection Valve Assembly    NOTE Accuracy of sample loops    The actual volume of a sample loop can differ by     10  for a 20ul loop  Smaller loops can  have a greater deviation and bigger loops a smaller one  Use partial loop filling if you must  know the actual injected volume        1260 Infinity Manual Injector 39    5    p n  0101    Parts and Materials for Maintenance     1248    0100 1923    0100 1922    0100 1924    0100 1921    0101   0101   0101   0101   0101     p n    0101   0101   0101   0101   0101   0101   0101   0101   0101   0101     40    1247  1246  1245  1244  1243    1241  1240  1239  1238  1242  1227  1236  1235  1234  1230   
7. 29    4 Maintenance       3 Remove the stator ring        5 Install the new seal  Ensure the seal is positioned as  shown     Rotor pin    Needle hole    Stator ring             4 Remove the seal        6 Install the stator ring  Ensure the pin in the stator ring is  aligned with the hole in the valve body and the position  sensoring switch is back in place              30    1260 Infinity Manual Injector    Maintenance 4    Injection Valve Seal          7 Insert the stator face onto the stator head  8 Install the stator head onto the valve  Ensure the pin in  the stator ring is aligned with the hole in the stator head        9 Secure the stator head in place with the stator screws   Tighten each screw alternately    4 turn until the stator  head is secure                 1260 Infinity Manual Injector 31    4 Maintenance    Position Sensing Switch    When      Tools required Description    Hexagonal key  9 64 inch    Parts required   p n  1 0490 1849    No start signal when switching to the inject position    Description    Position sensing switch       1 Loosen the three stator screws           2 Remove the stator head                    32    1260 Infinity Manual Injector    Maintenance 4    iak       Position Sensing S          3 Remove the stator ring        5 Insert the new sensing switch into the stator ring           4 Disconnect the sensor cable from the start cable  Pull the  sensing switch out of the stator ring        6 Install the stator ring  Ensure the pin in th
8. Manual Injector 27  Cleaning the Manual Injector 28  Injection Valve Seal 29  Position Sensing Switch 32    Parts and Materials for Maintenance 35    Manual Injector 36  Injection Valve Assembly 38    Appendix 41  Agilent Technologies on Internet 42    1260 Infinity Manual Injector    1260 Infinity Manual Injector    Introduction       Introduction to the Manual Injector 6    Operation and mechanical hardware    ee Agilent Technologies    1 introduction    Introduction to the Manual Injector    Sample is loaded into the external 20 ul sample loop through the injection  port at the front of the valve  The valve has a ceramic stator and PEEK  injection seal  PEEK is compatible with pH 0 14  incompatible with some  concentrated mineral acids  A make before  break passage in the stator  ensures flow is not interrupted when the valve is switched between the  INJECT and LOAD positions  and back again  see also    Needles    on   page 21 and    Flow Connections    on page 14      The valve is mounted on a steel mounting pole  and can be installed at  the left  or right hand side of the LC system         Mounting pole         Injection port    Figure 1 The Manual Injector installed to the mounting pole    6 1260 Infinity Manual Injector    R 1260 Infinity Manual Injector   e    e   r    a   o e     0   e        90 r Installing the Manual Injector      ry     e    Unpacking the Manual Injector  Damaged Packaging 8  Delivery Checklist 8   Installing the Manual Injector  Flow Connection
9. ct  or loss of important data  Do not  proceed beyond a CAUTION  notice until the indicated condi   tions are fully understood and  met     A WARNING notice denotes a  hazard  It calls attention to an  operating procedure  practice   or the like that  if not correctly  performed or adhered to  could  result in personal injury or  death  Do not proceed beyond a  WARNING notice until the indi   cated conditions are fully under   stood and met           1260 Infinity Manual Injector    In This Guide    In This Guide    This manual covers the Agilent 1260 Infinity Manual Injector  G1328C      1 Introduction    Operation and mechanical hardware    2 Installing the Manual Injector    Installation of the manual injector    3 Using the Manual Injector    How to use the manual injector    4 Maintenance    Instructions on simple  routine repair procedures    5 Parts and Materials for Maintenance    Detailed illustrations and lists for identification of parts and materials    6 Appendix    Additional information    1260 Infinity Manual Injector    Contents    Contents    Introduction 5  Introduction to the Manual Injector 6    Installing the Manual Injector 7  Unpacking the Manual Injector 8  Installing the Manual Injector 10  Flow Connections 14   Leak Drainage 16    Using the Manual Injector 17    Warnings and Cautions 18   Solvent Information 19   Information on Injection Seal Material 20  Needles 21   Injecting Sample 22    Maintenance 25    Overview of Maintenance 26  Flushing the 
10. ctor baseplate   see Figure 4 on page 12                                                                                                                                                                                                                           Figure 4 Installing the System    12 1260 Infinity Manual Injector    Installing the Manual Injector                    APG remote connector                   Start cable                Position sensor cable                                                    D00DDD                      Valve  see 1                              Figure 5 Installing the Start Cable    2       1  See Figure 6 on page 15       6 Connect the capillaries to the manual injector  see    Flow  Connections    on page 14      1260 Infinity Manual Injector 13    2 Installing the Manual Injector    Flow Connections      WARNING   Toxic  flammable and hazardous solvents  samples and reagents    The handling of solvents  samples and reagents can hold health and safety risks      gt  When working with these substances observe appropriate safety procedures  for  example by wearing goggles  safety gloves and protective clothing  as described in  the material handling and safety data sheet supplied by the vendor  and follow good  laboratory practice      gt  The volume of substances should be reduced to the minimum required for the  analysis      gt  Do not operate the instrument in an explosive atmosphere        CAUTION Prevent siphoning     gt  The outlets
11. e stator ring is  aligned with the hole in the valve body and the position  sensoring switch is back in place           1260 Infinity Manual Injector    33    4 Maintenance  Position Sensing Switch       7 Insert the stator face onto the stator head     9 Secure the stator head in place with the stator screws   Tighten each screw alternately    4 turn until the stator  head is secure              34       8 Install the stator head onto the valve  Ensure the pin in  the stator ring is aligned with the hole in the stator head                              1260 Infinity Manual Injector    1260 Infinity Manual Injector    r    e  e e         ee  5  7 e    Parts and Materials for Maintenance   amp    o         Manual Injector 36    Injection Valve Assembly 38    Detailed illustrations and lists for identification of parts and materials    te Agilent Technologies    35    5 Parts and Materials for Maintenance    Manual Injector    Item p n Description   1 5067 4191 Manual Front Loading Injector Valve  600 bar   2 5001 3738 Mounting pole  stainless steel   3 G1328 44121 Base plate   4 5042 8553 Organizer plate  5042 8576 Catch tube cap   5 5043 0208 Name plate  5190 1501 Valve syringe  fixed needle  50 pL  G1328 87600 Connection capillary  0 17 mm id  500 mm  0100 1677 Start cable    Manual injector valve see    Injection  Valve Assembly    on page 38     36    1260 Infinity Manual Injector    Parts and Materials for Maintenance 5  Manual Injector       Figure 12 Manual Injector    1
12. ith warranty  terms covering the material in this  document that conflict with these  terms  the warranty terms in the sep   arate agreement shall control     Technology Licenses    The hardware and or software described in  this document are furnished under a license  and may be used or copied only in accor   dance with the terms of such license     Restricted Rights Legend    If software is for use in the performance of a  U S  Government prime contract or subcon   tract  Software is delivered and licensed as     Commercial computer software    as  defined in DFAR 252 227 7014  June 1995    or as a    commercial item    as defined in FAR  2 101 a  or as    Restricted computer soft   ware    as defined in FAR 52 227 19  June  1987  or any equivalent agency regulation  or contract clause  Use  duplication or dis   closure of Software is subject to Agilent  Technologies    standard commercial license  terms  and non DOD Departments and  Agencies of the U S  Government will    receive no greater than Restricted Rights as  defined in FAR 52 227 19 c  1 2   June  1987   U S  Government users will receive  no greater than Limited Rights as defined in  FAR 52 227 14  June 1987  or DFAR  252 227 7015  b  2   November 1995   as  applicable in any technical data     Safety Notices    CAUTION    A CAUTION notice denotes a  hazard  It calls attention to an  operating procedure  practice  or  the like that  if not correctly per   formed or adhered to  could  result in damage to the produ
13. nformation on Injection Seal Material    The manual injector is supplied with a PEEK injection seal  PEEK is  compatible with pH 0 14  incompatible with some concentrated mineral  acids  see    Injection  Valve Assembly    on page 38      20 1260 Infinity Manual Injector    Using the Manual Injector    Needles    CAUTION Needle can damage valve     gt  Always use the correct needle size     3       Use needles with 0 028 inch outer diameter  22 gauge  x 2 inch long  needle  without electro taper  and with 90   point style  square tip      1260 Infinity Manual Injector    21    3 Using the Manual Injector    Injecting Sample    LOAD PositionLOAD Position    In the LOAD position  see Figure 9 on page 22   the pump is connected  directly to the column  ports 2 and 3 connected   and the needle port is  connected to the sample loop  At least 2 to 3 sample loop volumes  more  if better precision is required  of sample should be injected through the  needle port to provide good precision  The sample fills the loop  and  excess sample is expelled through the vent tube connected to port 6             From Pump    To column    Needle port connected from rear    loop         Waste  Figure 9 LOAD Position  INJECT Position  In the INJECT position  see Figure 10 on page 23   the pump is connected  to the sample loop  ports 1 and 2 connected   All of the sample is washed  22    1260 Infinity Manual Injector    Using the Manual Injector 3    out of the loop onto the column  The needle port i
14. o a  20 uL loop  With larger than half the loop volume  some of the sample is  lost out Vent Line 6  This is because sample flows down the center of the  loop at twice the average velocity due to the laminar flow effect     1 In INJECT  use the Needle Port Cleaner to flush out the needle port  with about 1 mL of mobile phase to flush out contamination from the  earlier injection  This liquid will exit out Vent Line 5     2 Insert the syringe into the needle port  You should feel slight resistance  as the needle passes through the needle seal before it stops against the  stator face     3 Load the sample slowly onto the loop   4 Leave the syringe in and turn the handle to INJECT     24 1260 Infinity Manual Injector    1260 Infinity Manual Injector    4  Maintenance    Overview of Maintenance 26  Flushing the Manual Injector 27  Cleaning the Manual Injector 28  Injection Valve Seal 29    Position Sensing Switch 32    Instructions on simple  routine repair procedures    ee Agilent Technologies    25    4 Maintenance    Overview of Maintenance    Table2 Overview of Repair Procedures       Procedure Typical Frequency    Time Required    Notes          Flushing the injector After using aqueous buffers or 5 minutes See    Flushing the Manual  salt solutions Injector    on page 27  Exchanging the After approximately 10000 to 10 minutes See    Injection Valve Seal    on page 29  injection valve seal 20000 injections  or when the  valve performance shows  indication of leakage or wear 
15. ow Cell    Long term operation at pH  gt  11 should be avoided  Never leave strongly  alkaline solutions in the flow cell without flow     Solvents    Always filter solvents through 0 4 um filters  small particles can  permanently block filters  frits and capillaries  Avoid the use of the  following steel  corrosive solvents     Solutions of alkali halides and their respective acids  for example   lithium iodide  potassium chloride  and so on      e High concentrations of inorganic acids like sulfuric acid  especially at  higher temperatures  replace  if your chromatography method allows  by  phosphoric acid or phosphate buffer which are less corrosive against  stainless steel      Halogenated solvents or mixtures which form radicals and or acids  for  example     2 CHCly   O  gt  2 COCl   2 HCl    This reaction  in which stainless steel probably acts as a catalyst   occurs quickly with dried chloroform if the drying process removes the  stabilizing alcohol     e Chromatographic grade ethers  which can contain peroxides  for  example  THF  dioxane  di isopropylether   Such ethers should be  filtered through dry aluminium oxide which adsorbs the peroxides     e Solutions of organic acids  acetic acid  formic acid  and so on  in  organic solvents  For example  a 1   solution of acetic acid in methanol  may attack steel       Mixtures of carbon tetrachloride with 2 propanol or THF  dissolve  stainless steel     1260 Infinity Manual Injector 19    3 Using the Manual Injector    I
16. r plate  5042 8553  1  Catch tube cap  5042 8576  1       1260 Infinity Manual Injector    Installing the Manual Injector 2    Table 1 Manual Injector Checklist       Description Quantity  Valve syringe  fixed needle  50 uL  5190 1501  1  User Manual  G1328 90013  1       1260 Infinity Manual Injector 9    2 Installing the Manual Injector    Installing the Manual Injector    CAUTION Defective on arrival  problems    If there are signs of damage  please do not attempt to install the module  Inspection by  Agilent is required to evaluate if the instrument is in good condition or damaged      gt  Notify your Agilent sales and service office about the damage      gt  An Agilent service representative will inspect the instrument at your site and  initiate appropriate actions        The manual injector can be installed at the left  or right hand side of the instrument stack     1 Place the baseplate on the bench     2 Connect the two organizer plates to the base plate        Figure 2 Connecting the Organizer Plates    3 Screw the mounting pole into one of the two holes in the organizer  plate     10 1260 Infinity Manual Injector    Installing the Manual Injector 2    4 Slide the manual injector onto the mounting pole  see Figure 3 on  page 11   Tighten the mounting screw     Mounting screw       Figure 3 Installing the Mounting Pole and Manual Injector    1260 Infinity Manual Injector 11    2 Installing the Manual Injector    5 Install other system modules on top of the manual inje
17. s    Leak Drainage    8    10  14    16    Installation of the manual injector    if Agilent Technologies    2    Installing the Manual Injector    Unpacking the Manual Injector    Damaged Packaging    Upon receipt of your manual injector  inspect the shipping containers for  any signs of damage  If the containers or cushioning material are  damaged  save them until the contents have been checked for completeness  and the manual injector has been mechanically checked  If the shipping  container or cushioning material is damaged  notify the carrier and save  the shipping material for the carriers inspection     Delivery Checklist    Ensure all parts and materials have been delivered with the manual  injector  The delivery checklist is shown in Table 1 on page 8  To aid in  parts identification  please see    Parts and Materials for Maintenance    on  page 35  Please report missing or damaged parts to your local Agilent  Technologies sales and service office     Table 1  Manual Injector Checklist       Description Quantity       Manual Front Loading Injector Valve  600 bar  5067 4191   1  A 20 uL sample loop is connected between Ports 1 and 4  The valve is supplied with a  start cable  Hex Keys  x2   Mounting Screws  x2  Vent Lines x2   Needle Port Cleaner   x1   Long Nuts  x3   Extra Long Nut  x1   Ferrules  x4  and a mounting bracket     Mounting pole  stainless steel  5001 3738  1  Connection capillary  0 17 mm id  500 mm  G1328 87600  1  Base plate  G1328 44121  1  Organize
18. s and Cautions 18   Solvent Information 19   Information on Injection Seal Material 20  Needles 21   Injecting Sample 22    How to use the manual injector    Apg Agilent Technologies    3 Using the Manual Injector    Warnings and Cautions    Ejection of mobile phase    When using sample loops larger than 100 pL  mobile phase may be ejected from the  needle port as the mobile phase in the sample loop decompresses      gt  Please observe appropriate safety procedures  for example  goggles  safety gloves  and protective clothing  as described in the material handling and safety data sheet  supplied by the solvent vendor  especially when toxic or hazardous solvents are  used        Splashing of solvent     gt  When using the Needle Port Cleaner  empty the syringe slowly to prevent solvent  from splashing back at you      gt  Please observe appropriate safety procedures  for example  goggles  safety gloves  and protective clothing  as described in the material handling and safety data sheet  supplied by the solvent vendor  especially when toxic or hazardous solvents are  used        CAUTION    Potential damage to the valve     gt  Rinse the valve with water after using buffer solutions to prevent crystals from  forming  which can cause scratches on the rotor seal        18    See    Flushing the Manual Injector    on page 27     1260 Infinity Manual Injector    Using the Manual Injector 3    Solvent Information    Observe the following recommendations on the use of solvents     Fl
19. s connected to the vent  tube  port 5         From pump       To column         Needle port connected from rear         Sample loop        Waste    Figure 10 INJECT Position    Complete Loop Filling    In complete  filling  the volume of sample injected is set by the volume of  the loop  this includes the valve passages   This method produces the  highest precision     At least 2 to 3 sample loop volumes  more if better precision is required   of sample should be injected through the needle port to provide good  precision     The sample fills the loop  and excess sample is expelled through the vent  tube connected to port 6     An excess of sample is needed because mobile phase near the wall of the  loop is displaced slowly due to the laminar flow effect     1 Turn the handle to the LOAD position    1260 Infinity Manual Injector 23    3 Using the Manual Injector    2 Insert the syringe into the needle port  You should feel slight resistance  as the needle passes through the needle seal before it stops against the  stator face     3 Load the sample slowly onto the loop  Repeat this step for higher  precision     4 Leave the syringe in and turn the handle to INJECT     Partial Loop Filling    If you only have small quantities of sample  this is the method of choice   In the partialfilling method the volume of sample injected is set by the  syringe  In this method  no more than half a loop volume of sample should  be loaded into the loop  For example  load no more than 10 uL int
    