Home
        JA-12 Fixed Angle Rotor Manual
         Contents
1.   Also ensure  that O rings are lightly but evenly coated with silicone vacuum  grease  335148      2  Load the filled containers symmetrically into the rotor   Refer  to TUBES on page 15 for information about containers   If fewer  than twelve tubes are being run  they must be arranged symmetri   cally in the rotor  see Figure 3   Opposing tubes must be filled to  the same level with liquid of the same density           Figure 3  Typical Examples of Arranging Tubes in the Rotor    OPERATION    INSTALLING THE ROTOR    JA 12 Rotor    e Precool the rotor in the centrifuge or in a refrigerator before use     especially before short runs   to ensure that the rotor reaches the  set temperature  A suggested precooling cycle is a minimum of  30 minutes at 2000 rpm at the required temperature        Ifyou are using a microprocessor controlled centrifuge  Avanti J  series and some J2 models   be sure to enter rotor code 12  Refer to  your centrifuge instruction manual for additional information     During acceleration in an Avanti J series centri   fuge  the JA 12 rotor may be misidentified as a  JA 14 rotor  If such misidentification occurs  it is  possible to set the run speed to 14 000 rpm   and operate the rotor at that speed  To prevent  misidentification  always select the JA 12 rotor  and a maximum speed of 12 000 rpm during run  setup  and do not change these settings  Do not  run the JA 12 rotor above 12 000 rpm at any  time  or labware failure may result     Rotor Drive Pin 
2.  observe the maximum speed limits and fill  volumes shown   Maximum fill volume is the maximum amount that  can be centrifuged in the container listed      Refer to Appendix A in Rotors and Tubes for information on the  chemical compatibilities of tube and accessory materials     The tubes listed in Table 3 should be used one  time only  Tube damage and sample leakage  will result from repeated use     Do not use glass tubes in this rotor     15    JA 12 Rotor    16          Temperature Limits    e Plastic containers have been centrifuge tested for use at  temperatures between 2 and 25  C  For centrifugation at other  temperatures  pretest containers under anticipated run conditions     e Ifplastic containers are frozen before use  make sure that they are  thawed to at least 2  C prior to centrifugation     Table 3  Tubes for the JA 12 Fixed Angle Rotor                   Manufacturers     Nominal Nominal Max  Fill Maximum   Dimensions Volume Volume Recommended   Manufacturer Description  mm   mL   mL  g force for tube  Corning Polypropylene  capped 30 x 115 50 45 9400 x g  Falcon Polypropylene  capped 30 x 115 50 45 9400 x g  Sarstedt Polypropylene  capped 29 x 115 50 40 8800 x g                         The JA 12 rotor generates g forces higher than the tube manufact   urers    recommended maximums  see Table 3   Running these   tubes above the recommended maximum g forces will cause white  striations known as crazing  see Figure 8   Moderate crazing will not  affect run resu
3.  rotational speeds higher than the maximums you  have determined to be appropriate for your protocols  in no case  above 12 000 rpm   When solutions more dense than 1 2 g mL are  centrifuged in this rotor  use equation  7  to calculate the reduced  maximum allowable rotor speed        1 2 g mL  7     reduced maximum speed   pa  p  P  density of tube contents    where p   maximum speed    CARE AND MAINTENANCE    MAINTENANCE    Dual locking  Rotor Lid     364846  a  _  Small O ring    Large O ring       961697      961696   Check    for  Corrosion          Y Rotor Body     I gt  NOTE    Do not use sharp tools on the rotor that could  cause scratches in the rotor surface  Corrosion  begins in scratches and may open fissures in the  rotor with continued use     e Periodically  at least monthly  inspect the rotor  especially    inside cavities  for rough spots or pitting  white powder deposits   frequently aluminum oxide   or heavy discoloration  If any   of these signs are evident  do not run the rotor  Contact your  Beckman Coulter representative for information about the Field  Rotor Inspection Program and the rotor repair center     e Before using the rotor  inspect the rotor drive pins to ensure that    they are not damaged  Damaged drive pins can prevent the rotor  from seating properly on the centrifuge drive hub  Avanti J series  centrifuges only   To inspect the drive pins  turn the rotor upside  down and look into the drive hole in the center of the rotor  If  the drive 
4. COULTER              2009 Beckman Coulter  Inc        J TB 051AG  July 2009    JA 12 Fixed Angle Rotor       Used In Beckman Coulter J2 and  Avanti   J Series Centrifuges    JA 12 Rotor    Pees lee ieee    SAFETY NOTICE    This safety notice summarizes information basic to the safe use of the rotor described in this  manual  The international symbol displayed above is a reminder to the user that all safety  instructions should be read and understood before operation or maintenance of this equip   ment is attempted  When you see the symbol on other pages throughout this publication  pay  special attention to the specific safety information presented  Observance of safety precau   tions will also help to avoid actions that could damage or adversely affect the performance of  the rotor  This rotor was developed  manufactured  and tested for safety and reliability as part  of a Beckman Coulter centrifuge rotor system  Its safety or reliability cannot be assured if  used in a centrifuge not of Beckman Coulter   s manufacture or in a Beckman Coulter centri   fuge that has been modified without Beckman Coulter   s approval     Handle body fluids with care because they can transmit disease  No known test offers  complete assurance that such fluids are free of micro organisms  Some of the most  virulent   Hepatitis  B and C  viruses  HIV  I V   atypical mycobacteria  and certain  systemic fungi   further emphasize the need for aerosol protection  Handle other infectious  samples accor
5. Orientation    Two metal pins inside the rotor drive hole engage with teeth on the  centrifuge drive hub to prevent the rotor from slipping during acceler   ation and deceleration  see Figure 4   When the rotor is properly  installed  the pins sit either in the drive hub grooves or on top of the  hub surface  Figure 5      In rotors manufactured since early 1997  the rotor pins are positioned  parallel to the BECKMAN name engraved at the center of the rotor  body  see Figure 6   Knowing the pin orientation before you install  the rotor will help to ensure that you position the rotor properly on the  hub  minimizing the chance of hub damage  Follow the steps below     JA 12 Rotor    SECKMAKy                      Rotor  Drive    Hole Drive Pins                Avanti J Centrifuge  Drive Spindle Hub          Figure 4  Rotor Drive Pin Location and Orientation    Drive Rotor  Hub Pins  2   Teeth  2     Grooves  4     Drive Hub Drive Hub    es Drive i Drive    Spindle Spindle  Assembly Assembly  Older Model Centrifuges     Newer Model Centrifuges      Be sure the pins in the rotor drive hole Be sure the pins in the rotor drive hole  are seated in the grooves are not sitting on top of the teeth  of the drive spindle hub as shown  on the drive spindle hub     Figure 5  Centrifuge Drive Hub Configurations    JA 12 Rotor    Top View Side View                                                        Figure 6  Installing the Rotor    Rotor Lid Configurations    The rotor has two available l
6. ack of this manual for warranty information                iim  C2 D gt   Annulus  ae    Figure 2  Fluid Containment Annulus    PREPARATION AND USE    Specific information about the JA 12 rotor is given here  Use the  Rotors and Tubes for Beckman Coulter J Series Centrifuges manual   publication JR IM  and the centrifuge instruction manual along with  this manual for complete rotor and accessory operation      I gt  NOTE    Although rotor components and accessories  made by other manufacturers may fit in the  JA 12 rotor  their safety in this rotor cannot be  ascertained by Beckman Coulter  Use of other  manufacturers    components or accessories in the  JA 12 rotor may void the rotor warranty and  should be prohibited by your laboratory safety  officer  Only the components and accessories  listed in this publication should be used in   this rotor     JA 12 Rotor    PRERUN SAFETY CHECKS    ROTOR PREPARATION    Read the Safety Notice page at the front of this manual before using  the rotor     1  Make sure that the rotor and lid are clean and show no signs of  corrosion or cracking     2  Refer to Appendix A in Rotors and Tubes for the chemical  compatibilities of all materials used     3  Verify that the tubes being used are listed in Table 3     For runs at other than room temperature  refrigerate or warm the  rotor beforehand for fast equilibration     1  Be sure that metal threads in the rotor are clean and lightly but  evenly lubricated with Spinkote    lubricant  306812 
7. ase     If the rotor  and or accessories  becomes contaminated with radioac   tive material  it should be decontaminated using a solution that will  not damage the anodized surfaces  Beckman Coulter has tested a  number of solutions and found two that do not harm anodized  aluminum  RadCon Surface Spray or IsoClean Solution  for  soaking     and Radiacwash      IlIi NOTE    IsoClean can cause fading of colored anodized  surfaces  Use it only when necessary and  remove it promptly from surfaces     While Beckman Coulter has tested these methods and found that they  do not damage components  no guarantee of decontamination is  expressed or implied  Consult your laboratory safety officer regarding  the proper decontamination methods to use     If the rotor or other components are contaminated with toxic or patho   genic materials  follow all appropriate safety and decontamination  procedures as outlined by your laboratory safety officer     3 In U S   contact Nuclear Associates  New York   in Eastern Europe and Commonwealth States  contact Victoreen GmbH  Munich    in South Pacific  contact Gammasonics Pty  Ltd   Australia   in Japan  contact Toyo Medic Co  Ltd   Tokyo    4 In U S   contact Biodex Medical Systems  Shirley  New York   internationally  contact the U S  office to find the dealer closest to you     22    JA 12 Rotor    STERILIZATION AND DISINFECTION       The rotor and all rotor components can be autoclaved at 121  C for  401  C up to an hour  Remove the lid from th
8. cement price   Current rotor list price x    CONDITIONS    1  Except as otherwise specifically provided herein  this war   ranty covers the rotor only and Beckman Coulter  Inc   shall not be liable for damage to accessories or ancillary  supplies including but not limited to  i  tubes   ii  tube  caps   iii  tube adapters  or  iv  tube contents     2  This warranty is void if the rotor has been subjected to  customer misuse such as operation or maintenance con   trary to the instructions in the Beckman Coulter rotor or  centrifuge manual     3  This warranty is void if the rotor is operated with a rotor  drive unit or in a centrifuge unmatched to the rotor charac   teristics  or is operated in a Beckman Coulter centrifuge  that has been improperly disassembled  repaired  or  modified     DISCLAIMER    IT IS EXPRESSLY AGREED THAT THE ABOVE WAR   RANTY SHALL BE IN LIEU OF ALL WARRANTIES OF  FITNESS AND OF THE WARRANTY OF MERCHANT   ABILITY AND THAT BECKMAN COULTER  INC   SHALL HAVE NO LIABILITY FOR SPECIAL OR CON   SEQUENTIAL DAMAGES OF ANY KIND WHATSO   EVER ARISING OUT OF THE MANUFACTURE  USE   SALE  HANDLING  REPAIR  MAINTENANCE  OR  REPLACEMENT OF THE PRODUCT        B    BECKMAN  COULTER     Beckman Coulter  Inc      250 S  Kraemer Blvd        Brea  California 92821  Sales and Service  1 800 742 2345       Internet  www beckmancoulter com      2009 Beckman Coulter  Inc   All rights reserved    
9. d to enter the temperature        uty  Lu       F  Set 2  Find the compensation value in Table   that corresponds with the  set temperature and run speed        3  Press  COMP ADJ   The word    COMP    flashes below the  TEMPERATURE display and the display flashes     12    JA 12 Rotor    4  Use the keypad to enter the required compensation value  Press  the      key to enter a minus sign  pressing it again will remove  the minus sign     5  Check the temperature display   If the entry is incorrect  press  ce   and reenter the digits      6  When the entry is correct  press  ENTER RECALL        gt  NOTE    To clear a COMP ADJ entry  press  COMP ADJ     0   and  ENTER RECALL      REMOVAL AND SAMPLE RECOVERY    A CAUTION    If disassembly reveals evidence of leakage  you  should assume that some fluid escaped the rotor   Apply all appropriate safety and decontamina   tion procedures to the centrifuge and  accessories     Dual locking Lid    1  Place one hand on the rotor to keep it from turning  Grasp the  round tie down knob with your other hand and turn the knob to the  left  counterclockwise  until the stem disengages from the drive                                                                                     x hub threads   2  Hold the knob or place your hands on the rotor and lift the rotor  straight up off the drive hub      3  Place the rotor on a counter or bench top under a safety hood     4  Grasp the daisy knob and turn it to the left  counterclockwise   until 
10. ding to good laboratory procedures and methods to prevent spread of disease   Because spills may generate aerosols  observe proper safety precautions for aerosol contain   ment  Do not run toxic  pathogenic  or radioactive materials in this rotor without taking  appropriate safety precautions  Biosafe containment should be used when Risk Group II  materials  as identified in the World Health Organization Laboratory Biosafety Manual  are  handled  materials of a higher group require more than one level of protection     The rotor and accessories are not designed for use with materials capable of developing  flammable or explosive vapors  Do not centrifuge such materials in nor handle or store them  near the centrifuge     Although rotor components and accessories made by other manufacturers may fit in the  JA 12 rotor  their safety in this rotor cannot be ascertained by Beckman Coulter  Use of other  manufacturers    components or accessories in the JA 12 rotor may void the rotor warranty  and should be prohibited by your laboratory safety officer  Only the components and acces   sories listed in this publication should be used in this rotor     Do not run an empty rotor  Place filled tubes in at least two opposing cavities  Make sure that  filled containers are loaded symmetrically into the rotor and that opposing tubes are filled to  the same level with liquid of the same density     If disassembly reveals evidence of leakage  you should assume that some fluid escaped the  r
11. e rotor and place the rotor  and lid in the autoclave upside down      tT       Ethanol  70     or hydrogen peroxide  6   may be used on all  rotor components  including those made of plastic  Bleach  sodium  hypochlorite  may be used  but may cause discoloration of anod   ized surfaces  Use the minimum immersion time for each solution   per laboratory standards            Qmm    While Beckman Coulter has tested these methods and found that they  do not damage the rotor or components  no guarantee of sterility or  disinfection is expressed or implied  When sterilization or disinfec   tion is a concern  consult your laboratory safety officer regarding  proper methods to use     Refer to publication IN 192  included in each box of tubes or bottles   for tube and bottle sterilization and disinfection procedures     STORAGE    When the rotor is not in use  store it in a dry environment  not in the  instrument  with the lid removed to allow air circulation so moisture  will not collect in the tube cavities     5 Flammability hazard  Do not use in or near operating centrifuges     23    JA 12 Rotor    24    RETURNING A ROTOR                                                                                                                   Before returning a rotor or accessory for any reason  prior permission   a Returned Goods Authorization form  must be obtained from  Beckman Coulter  Inc  This RGA form  which may be obtained from  your local Beckman Coulter sales office  should con
12. fect at the time of sale  Beckman Coulter  Inc  agrees  to correct either by repair  or  at its election  by replacement   any defects of material or workmanship which develop within  seven  7  years after delivery of a J series rotor to the original  buyer by Beckman Coulter  Inc  or by an authorized represen   tative  provided that investigation and factory inspection by  Beckman Coulter  Inc  discloses that such defect developed  under normal and proper use  Should a Beckman Coulter cen   trifuge be damaged due to a failure of a rotor covered by this  warranty  Beckman Coulter will supply free of charge all cen   trifuge parts required for repair     REPLACEMENT    Any product claimed to be defective must  if requested by  Beckman Coulter  Inc   be returned to the factory  transporta   tion charges prepaid  and will be returned to Buyer with the  transportation charges collect unless the product is found to be  defective  in which case Beckman Coulter  Inc  will pay all  transportation charges     A defective rotor will be replaced by Beckman Coulter  Inc  at  its then current list price less a credit based upon the age of the  rotor  years since date of purchase   The Buyer shall not  receive credit until the claimed defective rotor is returned to  Beckman Coulter   s Indianapolis  Indiana facility or delivered  to a Beckman Field Service representative     The replacement price  cost to Buyer  for the respective rotor  shall be calculated as follows     years       Repla
13. id configurations        The dual locking lid  364846  allows the lid to be attached to  and removed from the rotor outside the centrifuge  so that tubes  can be loaded and unloaded under a safety hood  The dual locking  lid consists of a daisy knob that attaches the lid to the rotor body   and a tie down knob that attaches the rotor to the centrifuge    Tie down Knob                                    Daisy Knob                drive hub     Lubricate  Threads  _7 Tie down Knob    The single locking lid  360991  has a tie down knob that secures    the rotor to the centrifuge drive spindle  The rotor must be placed  in the centrifuge before the lid is attached                                                     Lubricate    ee Threads    JA 12 Rotor    10    Rotor Installation      To prevent the rotor from sticking to the centrifuge drive hub     apply a thin coat of Spinkote lubricant to the rotor drive pins and  to the lid knob threads       Lightly coat the large and small O rings with silicone vacuum    grease       Dual locking lid only  Place filled tubes into the rotor  then place    the lid on the rotor and turn the daisy knob to the right  clockwise   until secure  Do not overtighten       Turn the rotor until the BECKMAN name is either horizontal or    perpendicular to the drive hub pins or teeth  see Figure 4   Or   rotate the drive spindle to align the hub pins or teeth with the rotor  drive pins       Carefully lower the rotor straight down onto the centrifuge dr
14. ides contact with the Field Rotor Inspection Program   and the rotor repair center     Wash the rotor and rotor components immediately if salts or other  corrosive materials are used or if spillage has occurred  Do not allow  corrosive materials to dry on the rotor     Under normal use  wash the rotor frequently  at least weekly  to  prevent buildup of residues     1  Remove the O rings before washing     2  Wash the rotor and lid in a mild detergent  such as Beckman  Solution 555     that won   t damage the rotor  The Rotor Cleaning  Kit  339558  contains two special plastic coated brushes and two  quarts of Solution 555 for use with rotors and accessories  Dilute  the detergent 10 to 1 with water       gt  NOTE    Do not wash rotor components in a dishwasher   Do not soak in detergent solution for long  periods  such as overnight     3  Thoroughly rinse the cleaned rotor and components with distilled  water     4  Air dry the rotor and lid upside down  Do not use acetone to dry  the rotor     5  Apply a thin  even coat of silicone vacuum grease to both O rings  before replacing them     21    JA 12 Rotor    DECONTAMINATION    Clean metal threads every 6 months  or as necessary  Use a brush and  concentrated Solution 555  Rinse and dry thoroughly  then lubricate  lightly but evenly with Spinkote to coat all threads     Periodically remove the O rings and wipe clean as necessary  Clean  the O ring grooves with a cotton tipped swab  Reapply a light film of  silicone vacuum gre
15. if intermediate values are required                                      Rotor Required Sample Temperature    C    Speed    rpm   20    10 2 5 10 20 40   12 000 N    6    6  7  7    6  9   10 000  3    4  5  5  5  5    9  5 000    1  3  3 4 4  5  9          An    N    indicates that the rotor cannot achieve the required temperature at this speed     JA 12 Rotor    11    JA 12 Rotor    Table 2  Temperature Compensation Settings  for the Model J2 HC Analog Centrifuge   Interpolate if intermediate values are required              Rotor Required Sample Temperature    C    Speed    rpm     20    10 2 5 10 20 40   12 000 N  3  2  2  2  2  1   10 000  2  2  1  1  1    1    1  5 000    1    1 0 0 0 0 0                                  An    N    indicates that the rotor cannot achieve the required temperature at this speed     Analog J2 Series Centrifuges  Models J2 HS and J2 HC     Ser 1  Turn the SET knob on the centrifuge control panel to the required  O temperature   2  Find the compensation value in Table 1 for Models J2 HS or  SOMPE Table 2 for Model J2 HC that corresponds to the required temper     ature and run speed  Set the COMP dial to that setting   Interpolate  if intermediate values are required      IIIi NOTE    Temperature settings for earlier model centri   fuges must be empirically determined     Microprocessor Controlled J2 Series Centrifuges   Models J2 MI and J2 MC     TEMPERATURE   C 1  Press the  TEMP  key on the centrifuge control panel and then use  the keypa
16. ime from maximum speed    rotor fully loaded  max  brake                 0 0000 eee ee 4 min  Rotor weight  maximum sample mass                    11 kg  24 Ib   Rotor mat  tiall oi 22s isdn da dca iieteta se beaded aluminum  Rotor entry code for microprocessor controlled instruments           12      The critical speed range is the range of speeds over which the rotor shifts so as to rotate about  its center of mass  Passing through the critical speed range is characterized by some vibration       Relative Centrifugal Field  RCF  is the ratio of the centrifugal acceleration at a specified radius  and speed  r     to the standard acceleration of gravity  g  according to the following formula     RCF   2   amp     where r is the radius in millimeters    is the angular velocity in radians per second   2 z RPM  60   and g is the standard acceleration of gravity  9807 mm s    After substitution       BEM    RCF   1 12r  ES    JA 12 Rotor    DESCRIPTION  This rotor has been manufactured in a registered ISO 9001 or 13485  facility for use with the specified Beckman Coulter centrifuges     ete The JA 12 fixed angle rotor  rated for 12 000 rpm  is designed to hold   364846  up to twelve 50 mL tubes at a 35 degree angle  see Figure 1   Used  l in Beckman Coulter J2 and Avanti   J series centrifuges  the JA 12         pepe i rotor develops centrifugal forces that can efficiently pellet cells   aiae bacteria  and food products  separate proteins  viruses  and sub   O ring cellular fract
17. ions  and do phase separation and binding studies      961697   The JA 12 rotor with dual locking lid was tested  to demonstrate  a eer containment of microbiological aerosols under normal operating  conditions of the associated Beckman Coulter centrifuge  when used  and maintained as instructed  The rotor is also available with a single   locking lid              The JA 12 rotor has a patented fluid containment annulus  located  below the O ring sealing surface  see Figure 2   The annulus helps to  retain liquid that may escape from leaking or overfilled tubes  thereby  minimizing the escape of liquid into the instrument chamber during  centrifugation         Single locking Lid Dual locking Lid  ee g g g Y    N    WI     A  Se Y           l    O  iio    ZB        Figure 1  The JA 12 Fixed Angle Rotor      Validation of microbiological containment was done at an independent third party testing facility  CAMR  Porton Down  UK  or USAMRIID   Ft  Detrick  MD  U S A    Improper use or maintenance may affect seal integrity and thus containment     JA 12 Rotor    The rotor and lid are made of anodized aluminum  Two O rings made  of ethylene propylene rubber  one in the rotor lid and one in the rotor  body  help to maintain atmospheric pressure inside the rotor during  centrifugation when they are properly lubricated  Pins in the rotor  drive hole mate with the centrifuge drive hub to prevent the rotor  from slipping during acceleration and deceleration     See the Warranty at the b
18. it disengages     5  Remove the lid and set it aside     13    JA 12 Rotor    14    Single locking Lid    1  Unscrew the tie down knob   2  Remove the lid and set it aside     3  Lift the rotor straight up off the drive hub     If the Rotor Sticks to the Centrifuge Drive Hub    If the rotor sticks to the centrifuge drive hub  use the removal tool   360998  to release it as follows  see Figure 7         Rotor Removal    Tool  360998  er                             ELF    Figure 7  Using the Rotor Removal Tool     I gt  NOTE    This procedure requires you to remove the rotor  lid while the rotor is in the centrifuge  If you  are using the dual locking lid  the biosafe seal  provided by this mechanism will be broken  when the lid is removed  Consult your labora   tory safety officer for the proper safety  procedures to follow in this case     TUBES    JA 12 Rotor    1  Remove the lid     2  Screw the removal tool into the threaded opening  As the tool  tightens down it will push against the drive hub  forcing the rotor  up and off the hub     3  Lift the rotor up and out of the centrifuge by the removal tool  handle or by holding the rotor with your hands     4  Unscrew the removal tool and remove it from the rotor      gt  NOTE    Never run the rotor with the removal tool in  place     Lubricate the centrifuge drive hub with a light coat of Spinkote to  prevent further sticking     The JA 12 rotor uses the tubes listed in Table 3  Be sure to use only  those items listed  and to
19. ive    spindle hub  Do not drop the rotor onto the spindle hub       Slowly turn the rotor around the drive spindle to make sure that it    is properly seated  In rare cases  the rotor pins can rest on top of  the spindle hub pins or teeth  which prevents the rotor from being  secured to the drive spindle  Turning the rotor after installing it  will move the pins into proper position       Secure the rotor to the centrifuge drive spindle as follows     Dual locking lid  press the lid knob down and turn it to the right   clockwise  until secure  no more than two full turns  Do not  overtighten     Single locking lid  place filled tubes into the rotor  then place the  lid on the rotor  Press the lid knob down and turn it to the right   clockwise  until secure  Do not overtighten     TEMPERATURE    TEMP   C          To ensure that the JA 12 rotor reaches the required temperature    during the run  follow the appropriate instructions below for the  centrifuge being used     Avanti J Series Centrifuges    Enter the temperature according to the instructions in your centrifuge  instruction manual  No additional input is required     J2 Series Centrifuges    Enter the required temperature and the appropriate temperature    compensation units  see Tables 1 and 2  on the centrifuge control    panel as follows     Table 1  Temperature Compensation Settings    for Analog  and Microprocessor Controlled Centrifuges   Models J2 HS  JS 21  J2 MI  J2 21M  J2 MC  and J2 21M E      Interpolate 
20. lts or cause tube leakage  However  repeated tube  usage and or long run times will eventually lead to tube failure  To  minimize the possibility of tube failure  observe the following  conditions     e Do not use tubes more than once   e Discard crazed tubes     Do not allow the run length at maximum speed to exceed 30 min     e Refer to Table 4 to find the g forces produced at various speeds  in the rotor  and limit run speed to the speed that produces the  maximum recommended g force for the tube in use     Beckman Coulter tests indicate that the Sarstedt tubes exhibit more  crazing than the Corning or Falcon  and that due to slight dimensional  differences  the Corning tubes exhibit less crazing than the Falcon     RUN TIMES    TIME HR MIN       rou  wu g       JA 12 Rotor       Figure 8  Examples of Tubes with Minimum and Moderate Crazing    The k factor of the rotor is a measure of the rotor   s pelleting effi   ciency  Beckman Coulter has calculated the k factors for all of its  rotors  at maximum speed with full tubes  using the following  formula     ants    max      min   102    Serr  1   o2 3600   where wis the angular velocity of the rotor in radians per second     0 105 x rpm   Fmax is the maximum radius  and rmin is the  minimum radius   After substitution    gx  2 533 x 10  h In  nax    min    2    rpm     17    JA 12 Rotor    18    2 Based on sedimentation in water at 20  C     In the JA 12 rotor  the k factor is 1244 for full 50 mL tubes at  maximum speed   F
21. or containers that nominally hold less than  50 mL  the k factors will be different  Calculate the k factors for  these containers using equation  2      Use the k factor in the following equation to estimate the run time t   in hours  required to pellet particles of known sedimentation coeffi   cient s  in Svedberg units  S      pak 3   S  For example  at 12 000 rpm E  coli  s   3600 S  will pellet in  1244    t  3600   30 min  4     Run times can be estimated for centrifugation at less than maximum  speed by adjusting the k factor as follows     2  k 12 000      Pg  t  adj rated   actual run speed     5     Run times can also be estimated from data established in prior experi   ments using a different rotor if the k factor of the previous rotor is  known  For any two rotors  a and b       2  6     where the k factors have been adjusted for the actual run speed used     For more information on k factors  see Use of k Factor for Estimating  Run Times from Previously Established Run Conditions  publication  DS 719      RUN SPEEDS    SPEED RPM RCF       a1 eT   Ll LL ot       JA 12 Rotor    The centrifugal force at a given radius in a rotor is a function of run  speed  Comparisons of forces between different rotors are made by  comparing the rotors    relative centrifugal fields  RCF   When rota   tional speed is selected so that identical samples are subjected to the  same RCF in two different rotors  the samples are subjected to the  same force  see Table 4      Do not select
22. otor  Apply all appropriate safety and decontamination procedures to the centrifuge and  accessories     Never exceed the maximum rated speed of the rotor and labware in use  Refer to the section  on RUN SPEEDS     Do not use sharp tools on the rotor that could cause scratches in the rotor surface  Corrosion  begins in scratches and may open fissures in the rotor with continued use     JA 12 Rotor                                           JA 12 FIXED ANGLE ROTOR  Axis of Rotation  SPECIFICATIONS   Maximum speed ecset erse chee cee erti enea one Sees 12 000 rpm  Critical speed range     2    unuunu 400 to 1000 rpm  Density rating at maximum speed               0 000 eee eee 1 2 g mL  Maximum allowable imbalance of opposing loads              2 grams  Relative Centrifugal Field  at maximum speed   at rmax  144 mm      0 0    ee ees 23 200 x g   at ray  108 mm        cee eee 17 400 x g   at rmin  71 mm     1    eee eens 11 500 x g  k factor at maximum speed            0 0 c cece eee eee eee 1244  Conditions requiring speed reduction                  see RUN SPEEDS  Number of tube cavities    0    unuunu eee tee eee 12  Available bottles and tubes             00    cee eee eee eee see Table 3  Nominal tube dimensions                0000s eee eee 30 x 115 mm  Nominal tube capacity    0    0  eee eee 50 mL  Nominal rotor capacity    0 0    eee eee eee 600 mL  Approximate acceleration time to maximum speed    rotor fully loaded     1 0 00    0 0 3 1 2 min  Approximate deceleration t
23. pins appear damaged  contact Beckman Coulter Field  Service     19    JA 12 Rotor    Table 4  Relative Centrifugal Fields for the JA 12 Rotor   Entries in this table are calculated from the formula  RCF   1 12 r  RPM 1000   and then rounded to three significant digits                    Relative Centrifugal Field  x g   Rotor  Speed At  max At fay At min   rpm   144 mm   108 mm   71 mm   12 000 23 200 17 400 11 500  11 000 19 500 14 600 9 620  10 000 16 100 12 100 7 950  9 000 13 100 9 800 6 440  8 000 10 300 7 740 5 090  7 000 7 900 5 930 3 900  6 000 5 810 4 360 2 860  5 000 4 030 3 020 1 990  4 000 2 580 1 940 1 270  3 000 1 450 1 090 716  2 000 645 484 318                         Relative Centrifugal Fields for the JA 12 Rotor    24 000       20 000       16 000       20       RCF  x g     12 000    8 000    4 000                                                 Speed  rpm              CLEANING    Rotor Cleaning  Kit  339558              i _    Small O ring    Se Large    O ring             JA 12 Rotor    e Regularly lubricate the metal threads in the rotor and on the  tie down bolt with a thin  even coat of Spinkote lubricant  Failure  to keep these threads lubricated can result in damaged threads     e Regularly apply silicone vacuum grease to the O rings  Replace  O rings about twice a year or whenever worn or damaged     Refer to Appendix A in Rotors and Tubes for chemical compatibili   ties of rotor and accessory materials  Your Beckman Coulter repre    sentative prov
24. r Sales  1 800 742 2345 in the United  States   or your local Beckman Coulter office  or see the High Perfor   mance  High Speed  High Capacity Rotors  Tubes  amp  Accessories  catalog  BR 8102  available at www beckmancoulter com  for  detailed information on ordering parts and supplies  For your  convenience  a partial list is given below     REPLACEMENT ROTOR PARTS    JA 12 rotor assembly  biosafe  with dual locking lid              360992  JA 12 rotor assembly  with single locking lid                   360993  Dual locking lidie  accesses Aes Res eee S 364846  Single locking lid  na agian Aige eG GAS BASS AR a 360991  Tie down kit  for use with dual locking lid                     364922  Tie down kit  for use with single locking lid                    364919  Lid O ring  ethylene propylene rubber                      4  961697  Rotor body O ring  ethylene propylene rubber                  961696  Rotor removal tool 1 0 2    cee eee eee eee 360998  SUPPLIES  TUDES ai 58 Sb Ab Ra see Rte beh etd ee eee oes see Table 3  Removal tooli  c es  los ha tis oe ea E ans 360998  Rotor Cleaning Kites cc  0c0  sbds cierre bees eae dee os 339558  Solution    sss qt  ese eat Rone hea ee Rode pede eens sod 339555  Silicone vacuum grease  1 OZ       0    ec eee ee eee 335148  Spinkote lubricant  2 0Z    0 0    eens 306812    25    J SERIES ROTOR WARRANTY    Subject to the conditions specified below and the warranty  clause of the Beckman Coulter  Inc   terms and conditions of  sale in ef
25. tain the following  information     e rotor serial number   e history of use  approximate frequency of use    e reason for the return     e original purchase order number  billing number  and shipping  number  if possible     e name and phone number of the person to be notified upon receipt  of the rotor or accessory at the factory  and    e name and phone number of the person to be notified about repair  costs  etc     To protect our personnel  it is the customer   s responsibility to ensure  that the parts are free from pathogens and or radioactivity  Steriliza   tion and decontamination must be done before returning the parts   Smaller items  such as tubes  bottles  etc   should be enclosed in a  sealed plastic bag     All parts must be accompanied by a note  plainly visible on the out   side of the box or bag  stating that they are safe to handle and that  they are not contaminated with pathogens or radioactivity  Failure to  attach this notification will result in return or disposal of the items  without review of the reported problem     Use the address label printed on the RGA form when mailing the  rotor and or accessories     Customers located outside the United States should contact their local  Beckman Coulter office     JA 12 Rotor    SUPPLY LIST     I gt  NOTE    Publications referenced in this manual can be  obtained by calling Beckman Coulter at 1 800   742 2345 in the United States  or by contacting  your local Beckman Coulter office     Contact Beckman Coulte
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
DeWalt DW820 Grinder User Manual  ITSスポットを活用した 物流効率化の官民実証実験を開始しました  Viper SC IP Router For Licensed Spectrum  Belzona® 4311  Marketing de la ville et implication de ses parties  Gabarit de fraisage Lamello  Part 4 How to use the WIFI module  FS-4/FS-4Pro Portable DTE Festplattenrekorder Benutzerhandbuch  Télécharger la fiche technique  Samsung DE46C Εγχειρίδιο χρήσης    Copyright © All rights reserved. 
   Failed to retrieve file