1.  Washer Disinfector       N Instruments with this symbol may be cleaned and disinfected with a thermal washer disinfector   To use a thermal washer disinfector  first read the operation instructions for the unit   Strictly follow the instructions attached for cleaning mode and solution                O  Improper cleaning modes and solutions will damage the instrument   Make sure the o rings on the motor s connection cylinder are OK  m i l l l E l l   Follow the disinfector manufacturer s recommendations for cleaning solutions and methods for dental instruments  Do not use strong acidic or alkaline solutions  Thoroughly  Push the TorqTech attachment straight onto the motor s connection cylinder until you hear it snap rinse off all cleaning solution     into place     O  Make sure the motor s connection cylinder is free of dust  dirt  and debris  7  Lubrication   Lubricate with Ls Spray  sold separately  before autoclaving and after moring and afternoon treatment sessions     O  Do not connect this attachment to a motor that rotates faster than 40 000 rpm           Remove the burr or point and take the TorqTech attachment off the motor   O  Do not drop the instrument or bang it against anything  This could cause it to malfunction       A WARNING    e Give the attachment a light tug before using it to make sure it is securely connected   e Wait for the motor to come to a complete stop before connecting or disconnecting the attachment   e Use a motor that conforms to ISO 13294  I
2. CE mM    0197 MORITA    TORQTECH Operation Instructions  ST DH    Straight Handpiece Attachment    2011 08 22 PUB  H5074 E 1    Thank you for purchasing a TorqTech straight motor handpiece attachment  For optimum safety and  performance  read this manual thoroughly before using the unit and pay close attention to the warnings  and notes  Keep this manual in a handy place for ready reference     The following symbols and expressions indicate the degree of danger and harm that could result  from ignoring the instructions they accompany     A WARNING       This warns the user of the possibility of extremely serious injury or complete  destruction of the instrument as well as other property damage including the  possibility of fire     This warns the user of the possibility of mild injury or damage to the  instrument     ANOTE     1   Usage Note     This alerts the user of important points concerning operation or the risk of  equipment damage     Disclaimer    m The J  Morita Mfg  Corp  will not be responsible for accidents  equipment damage  or bodily injury resulting from     1  repairs made by personnel not authorized by the J  Morita Mfg  Corp    2  any changes  modifications  or alterations  of its products   3  the use of products or equipment made by other manufacturers  except for those procured by the  J  Morita Mfg  Corp    4  maintenance or repairs using parts or components other than those specified by the J  Morita  Mfg  Corp  and other than in their original conditi
3. HNOLOGY PROMEDT CONSULTING GMBH  AltenhofstraBe 80  66386 St  Ingbert  Germany    The authority granted to the authorized representative  MEDICAL TECHNOLOGY PROMEDT Consulting GmbH  by J  Morita Mfg  Corp  is solely limited  to the work of the authorized representative with the requirements of the European Directive 93 42EEC for product registration and incident report           11 08  1 500    Printed in Japan    The TORQTECH attachments are not autoclaved before shipping  Do not fail to autoclave the attachments before using them  Wipe the attachment with a piece of gauze that has been dampened with ethanol and wrung out thoroughly        Make sure the chuck opening is rust free and clean  Make sure the motor joint is clean and not damaged   Connect the attachment to the motor and operate it  Make sure it rotates smoothly  emits a fine  misty spray  and there is no air or water leak   age from the motor joint  Also make sure that there is no abnormal noise  excessive vibration  or overheating     Q Do not use anything except ethanol  Do not use too much ethanol  it could seep inside and damage the instrument     Q Rotation and spray delivery will be impaired if debris gets inside the water or air lines        Never wipe the instrument with corrosive solutions such as electrolytic acids  strong alkaline solutions  or other special cleaning solutions    A WARN   N G commonly available     e Do not use the attachment if it is bent  warped or otherwise deformed      s   Thermal
4. SO 11498  otherwise there is a risk of electrical leakage     a    Open the chuck     The spray will not be effectively delivered if the chuck is closed      3  Put the nozzle on the LS Spray can     P    Insert the nozzle into the connection end of the TorqTech attachment  hold it  firmly and then spray for about 2 seconds          The instrument will fail if it is not lubricated and cleaned with LS Spray regularly     O  Do not use LS spray for any instruments except those with ball bearings     Remove O  Do not use any lubricant except LS Spray  Other lubricants will cause the connection o rings to swell and make it difficult to connect and disconnect  Absolutely never use AC  Cleaner  which is for air bearings  If you do  the bearings will burn out as soon as the instrument is rotated     The head could get quite cold if the spray is applied for more than 5 seconds  In this case  wait awhile before operating the handpiece  otherwise the bearings could be dam    2 aged     Q  Shake the LS spray can 2 or 3 times before use and always use it in an upright position  Use only the special spray nozzle provided     A WARNING    e Always cover the attachment with gauze before spraying  If spray accidentally gets into your eyes  wash them out with plenty of water and see a doctor  Use soap and water to wash    Open the chuck all the way off any spray on your skin   e Do not use the LS Spray in an enclosed arrea which is not properly ventilated  Do not point the spray nozzle at any
5. damaged     ANOTE    e If a burr has a welded or soldered joint  do not use it if the head wobbles        Operation   Temp   10   C  50   F  to 40   C  104   F   Relative Humidity  30   75    without condensation    Atmospheric Pressure  700   1 060 hPa  Storage  Temp    10  14   F  to 70   C  158   F   Relative Humidity  10   85    without condensation      Atmospheric pressure  700   1 060 hPa      Avoid frequent or prolonged exposure to direct sunlight     If instrument will not be used for a long time  autoclave it and store it in a clean  dry place       The working life of TORQTECH attachments is 7 years from the date of shipment provided they are  regularly and properly inspected and maintained     Disposal of Medical Devices      Any medical devices which could possibly be contaminated must be first decontaminated by the responsible  doctor or medical institution and then be disposed of by an agent licensed and qualified to handle medical  and industrial waste     Before inspecting or adjusting the instrument  make sure the main power switch is turned on and that the  main air and water valves are open  If the instrument does not work properly even after inspection and  adjustment  contact your local dealer or the J  Morita corp  and have it repaired     Burr or Point will not go in chuck   Make sure the burr or point is OK  Clean and lubricate the instrument with LS Spray  sold separately    If this does not solve the problem  have the instrument repaired     Burr do
6. es not rotate    Make sure motor is properly connected to its tube    Take the TorqTech attachment off the motor and step on the foot pedal to see if the motor runs    If the motor runs  put the attachment back on and see if the burr rotates  If it doesn t  have the TorqTech  attachment repaired    If the motor doesn t run  refer to the user manual for the motor     Spray is not emitted    Take the TorqTech attachment off the motor and step on the foot pedal to see if water comes out of the  motor joint  If water comes out  check the o rings on the motor joint and replace them if they are worn out  or damaged  Coat new o rings with a little oil  If the o rings are OK  clean the air and water holes on the  TorqTech attachment  If there is still no spray  have the TorqTech attachment repaired    If water doesn t come out  refer to the user manual for the motor     Overheating  Make sure flows for cooling air and spray are up to specifications  Also make sure the motor is operating  properly     m The user  hospital  medical institute or clinic  is responsible for inspection and maintenance of medical  devices     m The user  i e   clinic  hospital etc   must perform the inspection procedures specified in  Maintenance  and Inspection  every three months     m For parts that wear out and need periodic replacement  refer to the Inspection Items and replace parts  whenever necessary     m It is assumed that the user will perform regular maintenance and inspection procedures  Howeve
7. il to clean and lubricate the instrument before autoclaving it  After autoclaving  take the instrument out of the autoclave as soon as  it cools off     O  High autoclave temperatures can damage the o rings etc  Keep the handpiece from directly contacting heat sources or the inside of the autoclave by putting it into a sterilization  pouch     A WARNING    e To prevent the spread of serious  life threatening infections such as HIV and hepatitis B  the TorqTech attachments must be autoclaved after finishing the treatment of each patient     ANOTE    e Instruments are extremely hot after autoclaving  do not touch them until they cool off        Clean debris off the attachment with a brush and running water        
8. inimum  for 0 2 MPa instrument pressure   Conforms to ISO 3964 and DIN 13940   Conforms to ISO 13294 and ISO 11498   30 mm minimum   62 334 to 02 350 mm   44 5 mm      Use burrs that conform to ISO 1791 1   Specifications may be changed without prior notice due to improvements      Always use burrs and points with clean shanks  dirty shanks cause the chuck to wear out more quickly and lose its grip   ping power    Always carefully remove all cleaning fluids before inserting a burr into the chuck  some cleaning fluids are corrosive  and  if a burr or point with cleaning fluid on it is inserted  the chuck could be ruined     Q  Diamond points are easily bent  Even new burrs and points are sometimes poorly balanced  if a burr makes a loud  me   tallic noise during rotation  stop using it       If a dirty burr or one with cleaning fluid has been inserted by mistake  contact your local dealer or the J  Morita Office to  have the instrument inspected     A WARNING    e Never use the following types of burrs and points  they could come out and injure the patient or be swallowed   Those which do not conform to ISO standards  Those with scratched or bent heads or shanks  Those which are rusty     chipped or broken  Those which have been modified or altered such as those having shortened shanks     e Use burrs and points with shanks that are at least 30 mm long  If the cutting tip touches the end of the attachment  it could  get too hot and burn the patient  or the attachment could be 
9. lasses to prevent eye injury    e Never switch rotation direction from forward to reverse or vice versa while the motor is actually running    e Remove burrs and points before returning the instrument to its holder in the treatment unit  otherwise  there is a risk of  injury or damage    Crown removal and other similar operations produce considerable vibration and can cause the burr to get stuck in the  handpiece  Avoid this by frequently taking the bur out and putting it back in again  If a burr gets stuck  do not try to  take it out with a pair of pliers or other forceful means  this could damage the chuck and create a risk of a burr coming  out during operation  Contact J  Morita and have the instrument repaired     If the instrument makes abnormal noise  vibrates too much  or overheats  take out the burr and replace it with a new  one  Ifthe same problem persists  have the instrument repaired     Follow all recommendations of the burr manufacturer concerning burr rotation speed and direction        Straight Attachment ST DH    Rotation Speed Ratio ie    Maximum Motor Speed 42 000 rpm   Length 83 mm   Weight 130   Usable Burrs   Shank Type 2  HP     Chuck Type   Chuck Gripping Power  Cooling Air Flow  Spray Flow   Tip Air Flow  Connection Type  Connectable Motors  Usable Burr Length  Usable Burr Diameter  Maximum Burr Length    Mechanical Lock   Conforms to ISO 7785 2   Minimum 6 NL  min   from motor    50 ml   min   minimum  for 0 2 MPa instrument pressure   1 5 L min   m
10. on   5  operating the equipment in ways other than the operating  procedures described in this manual or resulting from the safety precautions and warnings in this manual not being  observed   6  workplace conditions and environment or installation conditions which do not conform to those stated  in this manual such as improper electrical power supply   7  fires  earthquakes  floods  lightning  natural disasters  or  acts of God    m The J  Morita Mfg  Corp  will supply replacement parts and be able to repair the product for a period of 10 years after  the manufacture of the product has been discontinued     m Warranty is honored only when a claim is accompanied by the warranty agreement with purchase date  product name  and model  serial number  inscribed on the body of the product  and name of the distributor     m Do not fail to thoroughly read the user manual for your motor and follow all the instructions and cautionary remarks in it   m The owner  clinic  hospital or medical institution  is responsible for use and maintenance of medical devices     m This instrument must not be used by anyone except legally qualified dentists and oral surgeons   practitioners     This instrument equipment must be only used for its specified dental treatment purpose        Caution  Federal law restricts this device to sale by or on the order of a dentist  for U S A          Spray Nozzle   Disinfect with ethanol     Spray Hole Wire   Disinfect by autoclave     a      For parts that wear ou
11. one and do not use the spray near an open flame   and pull the burr out     2    OK NO    Close chuck     Turn sheath until triangle  marks  AW  line up          Closed Open       1  1  Put an air nozzle  motor accessory  on the end of a threeway syringe    2  Put the nozzle in the connection end of the TorqTech attachment and blow air      Never rotate the attachment with the chuck open  through it for 20 or 30 seconds  Wipe off the oil that gets on the surface of the  AWARNING instrument with a piece of gauze    e Always insert burrs and points all the way in and give them a light tug to make sure they will not come out  Otherwise  a burr or point could come out and injure the patient or be swallowed  A N OTE  UCI SLRS CA eo RA lac Stop roiatng before Reg 1S  EL SC eect e Do not fail to eliminate excess oil after lubricating the attachment  otherwise  oil might seep into the motor and damage it  Also excess oil could cause the attachment to overheat     td N  p         Recommended Autoclaving Conditions  In a sterilization pouch           atleast 6 minutes at 134   C  273 2   F  or at least 60 minutes at 121   C  249 8   F       It may be hard to remove a burr or point if it is left in the attachment    Autoclaving and drying temperatures must not exceed 135   C  275   F     Minimum drying time after sterilization  10 minutes     Take out the burr or point and disconnect the TorqTech attachment from the motor           Do not sterilize in any way except autoclave  Do not fa
12. r   if necessary or desired  these may be performed by a trained professional qualified to repair  medical devices  Contact your local dealer or the J  Morita Office  Inspect the following items   1  Motor connection  2  Burr and point insertion  3  TorqTech rotation and spray  4  Overheating     m For repairs or other types of service contact your local dealer or the J  Morita Office        ul J  MORITA MFG  CORP     680 Higashihama Minami cho  Fushimi ku  Kyoto  612 8533 Japan     http   www jmorita mfg com    Distributors   J  MORITA CORPORATION   Tokyo Office   2 11 15 Ueno  Taito ku  Tokyo  110 8513 Japan  Osaka Office   3 33 18 Tarumi cho  Suita  Osaka  564 8650 Japan  http   asia morita com    http   oceania morita com    J  MORITA USA  Inc   9 Mason  Irvine  CA 92618 U S A     SIAMDENT CO   LTD    444 Olympia Thai Tower  3rd Floor  Ratchadapisek Road  Samsennok   Huay Kwang  Bangkok 10310  Thailand   Tel   66 2 512 6049  Fax   66 2 512 6099  http   www siamdent com  J  MORITA CORPORATION Australia and New Zealand   Suite 2 05  Aero 247 Coward Street Mascot NSW 2020  Australia   Tel   61 2 9667 3555  Fax   61 2 9667 3577  http   oceania morita com    Tel   1 949 581 9600  Fax    1 949 465 1095  http   www jmoritausa com    J  MORITA EUROPE GMBH  Justus Von Liebig Strasse 27A  D 63128 Dietzenbach Germany  Tel   49 6074 836 0  Fax   49 6074 836 299  http   www jmoritaeurope de    EU Authorized Representative under the European Directive 93 42 EEC    Tec  rer  MEDICAL TEC
13. t and need periodic replacement   refer to the Inspection Items and replace parts whenever  necessary     May be cleaned and disinfected with a thermal washer disinfector   Autoclavable up to 135   C        Identification Ring   Blue  1 1 rotation ratio     Lock Sheath        Model         Gear Ratio ECH    ST D  Manufacturer    rie          Type    Conforms to European medical device Directive 93   42 EEC  Serial Number         A WARNING    e After inserting a burr  always operate the handpiece outside of the oral cavity to make sure it runs smoothly   e If a burr comes out during use  stop using the attachment immediately and have the instrument repaired     e Never use a burr or point to pull up on a metal crown or other metal piece  This is very dangerous  the burr could come  out and injure the patient or be swallowed     e Never use the instrument with no or too little cooling air  this could cause it to heat up and possibly burn the doctor s  hand or the patient s mouth     e Always use a soft  feather   intermittent touch to avoid injury to the pulp   e Wait for the burr to stop rotating before taking it out of the oral cavity     e To avoid cutting your finger etc   wait until the instrument has completely stopped rotating before inserting or removing a  burr or point     e Never turn the lock sheath while the instrument is actually running  This could cause the whole straight attachment to  start rotating and create a dangerous situation     ANOTE    e Wear safety g
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Panasonic SDYD250 Use and Care Manual : Free Download, Borrow, and Streaming : Internet Archive  Philips F connectors SWV2554W  平成 24 年度知的財産権ワーキング・グループ等侵害対策  QIS Pilot - QIES Technical Support Office  FV2200-LNX 商品説明書 - FAST CORPORATION[株式会社ファースト]  Au cœur de l`image  465M Ozone Analyzer  Sistema DCU per FlexAct® COM  Titre : L`ail votre allié santé incontournable Il fait baisser la tension  Deutsch English Español Italiano    Copyright © All rights reserved. 