        MICROFOCUS X-RAY SOURCE PROJECT*
         Contents
1.   Works supported by the National Science Foundation    Specifications    The source has a maximum voltage of 90 kV  a  maximum power of 80 watts  and a maximum current of 2  mA  The tungsten target releases x rays in a 60 degree  cone  meaning a large portion of the shielding will receive  hits from primary photons  For more information on the  source specifications see table 1     X Ray Source Specifications    Minimum Distance Focus Objet   4mm  Target Angle Viewing Angle 15 30    Exit Window Diameter    Tungsten  W   Window Material Beryllium    Cathode Type Dispenser Cathode  Window Position End Window    Cooling Method H 0 0 15 L min   15 psi   10  to  55  C    Window Thickness 245 um    X Ray Tube Integrated Sealed       Table 1  Microfocus Source Specifications    SHIELDING CALCULATIONS  Unshielded Dose Rate    Between calculations and experiments run using  existing hutches  it has been determined that the dose rate  at the source is 1x10   mrem hr  Keep in mind that the  dose decreases with distance as an inverse square  see  equation one      _ kSEBu  1       p4nr        Shield Thickness    In order to protect all personnel from this high dose  rate  shielding is required  The material used is lead  because of its attenuation quality as well as its low cost of  production  Due to the complexity of the shielding  calculation  the simplest way to determine lead thickness  is to use the program XOP  By inputting an energy range  and shield thickness  the output will b
2. MICROFOCUS X RAY SOURCE PROJECT     Dan Mancuso  CHESS  Cornell University  NY  USA    ABSTRACT    At the Cornell High Energy Synchrotron Source   CHESS   scientists in all fields and from all over the  world utilize the high powered x rays generated by the  synchrotron  The issue is that there is such a high demand  for beam time that not every group gets to do all the  experiments they want  especially if there are any failures  in existing machinery  Therefore an outside source  not  tied into the synchrotron  would be a benefit  For this  summer program the task has been to design and build all  required safety features for this x ray source  These  features include lead shielding  safety interlocks  radiation  signs  personnel badges  and a shutter box  The source  itself will be used in a variety of tasks  ranging from  capillary testing to phase contrast imaging     SOURCE INFORMATION    The source is the Nova600 Microfocus X Ray Source  from Oxford Instruments  The package includes the  source  a controller unit  and a water chiller  The chiller is  the K O Concepts model LCR Max  The controller is  used to control the current and voltage of the source as  well as insure all interlocks are being satisfied  It can also  be hooked up to a computer by RS 232C for ease of use   The chiller keeps the source at an acceptable temperature  by using water or some mixture of water and inhibited  propylene glycol  IPG  or alcohol  1         Figure 1  Nova600 Microfocus Source   
3. at which the two lead sheets connect  Figure 4  illustrates this design     Figure 4  Overhang  Frame Design    The material chosen for framing is 80 20  commonly  referred to as the industrial erector set  The reason for this  choice is simple  the diversity of the available parts as  wells as the many attachment points make realizing  designs extremely easy  Figure 4 displays the general  frame design  Because the beams are made through  extrusion  they are very structurally sound  Even though  the roofing beams are laid on their weak side  2  linstead  of 1 2  the support provided by the columns and the wall   as well as the materials high ultimate tensile strength  will  easily support the lead panels        Figure 4  Framing    Windows    The lead paneling comes in 48 inch by 48 inch  sheets  It is about a quarter inch thick with less than an  eighth of an inch of aluminum on both sides to protect  lead   Additionally  two windows will be installed so individuals    personnel from the potentially hazardous  on a tour or exhibition of the facilities can see the  experiments being run inside the hutch  The windows will  be half inch thick leaded glass  equivalent to one eighth  inch lead  They have less attenuation value than the  quarter inch thick lead panels  This is still acceptable  because the closest window is about two meters from the    source  so the dose rate is severely less   Door    The door is always the weakest link in a sealed room  because it is the only 
4. e a graph of photon  energy versus percent transmission  The graph below is  for 6mm thick lead  Figure 2   This is twice as thick as  required by federal standards  but because the public will  have access to the facility it is important to keep the    radiation level below that of background radiation        Figure 2  Photon Fluence vs Percent Transmission  XOP   Shielded Dose Rate    The percent transmission for 6mm thick lead is 2x10       This leads to a dose rate of 0 02 mrem hr if the shield  is directly against the source  Because the closest location  an x ray will strike the shield is about a meter away  the  dose will be significantly less  well within that of  background radiation     Room Design    Shield Design    No matter how thick the lead is  it will not matter if  there are any leaks in the wall  it must be a perfect dark  room  The theory is simple  no direct lines of sight  the  difficulty lies in the manufacturing  Any individual can  build a room in which there are no gaps or cracks  but it  takes work to make one with a constrained amount of  material and available space  The simplest solutions to  cracks are to overlap pieces or attach a piece to cover the  crack  We decided a covering piece would be most  appropriate  see figure 1     E    Figure 3  Gap Coverage       Where the roof meets the wall will be another gap  this  time an additional slice of lead is not required though  By  bending the top piece 90 degrees it can be used to cover  the line 
5. f  the door is somehow opened the shutter will close to keep  any individuals from being exposed to radiation  The  mechanism is a small button that depresses when the door  is shut  Additionally  if either interlock is tripped  the  controller must be manually reset to open the shutter   Lighted display panels will be installed to show what  parts of the interlock system are being met  The image  below is a sketch of those panel designs  they also  describe the interlock logic     O    O  Control              gt  Shutter        gt  Safe  Power Closed O  O O  StationOK        gt  WaterFlow            Power Supply  O Interlock  O O O O    Door Closed     gt  Keyin        gt  Secure Shutter Enable    Storage    O    Figure 7  Shutter Logic  Miscellaneous Safety Features    On the access point to the hutch a sign will be placed  which displays a radiation symbol and the words  CAUTION   X RAY RADIAITON  This sign will light  up when the source is in use and the shutter is open  The  light will be tied into the interlock system so if the bulb  burns out the source cannot be turned on     Once the hutch is fully built  the source will be put  inside and turned on to allow a careful examination of the  outside using a Geiger counter  Any leaks that were  missed in the design will be found and patched  After this  initial study  an additional search will be made once a  year to guarantee the shielding is meeting all guidelines     Below are the procedures for working with the  Micro
6. focus source     e The X ray generator will be disabled at all times  that it is not in use by removing the interlock and  enable keys     e Set up will come under the supervision of a  CHESS Safety Committee member who will  survey the area with a Geiger counter after the  unit has been installed     e Entries will be made into the CHESS Safety  Log book after the tests     e The Microfocus unit will only be maintained  under the supervision of the Permit Holder     e Any alteration to the Microfocus setup will have  to be approved by the CHESS safety committee  and supervised by the Permit Holder     e First time users of this system must first receive  instruction and then permission from the Permit  Holder     e A list of authorized users will be maintained in a  log book     e Upon startup of the X ray generator  the user  shall enter the date  operator name  operating   of max kV and mA   Microfocus Source   s log book     conditions into the    e The power supply key will be locked up in lock   box in the CHESS operations area when the  equipment is not in use     Issues  Asbestos    The image below is of the room in which the hutch is  being built  prior to any work  Note the black tabletop  it  is affixed to the cabinet with asbestos glue  Due to this it  could not be removed by any CHESS workers  a special  team had to be called in from the Cornell Environmental  Health and Safety  EH amp S         Figure 8  Rooms Initial State    Gas Water Lines    Once the asbesto
7. m on the frame  each panel weighs over 200 pounds  each     Acknowledgements    This work is supported by the National Science  Foundation under Grant No  0841213     References     1  K O Concepts  Inc      Operations Manual     July 2009   http    www k oconcepts com     2  Oxford Instruments     Nova600 Microfocus X Ray  Source User Manual     http   www oxfordxtg com       
8. point that cannot be permanently  sealed  The initial design called for a sliding door  This  had potential because all the existing CHESS hutches use  sliding doors  there is no moment arm to worry about   interlock buttons are easy to install  and it takes up no  additional room  It was decided not to use one because it  took the optimal space for the windows  it is significantly  more difficult to put together  and the geometry of the  room lends a hand towards a swinging door     Figure 5 shows the doors placement  within the 45  degree segment of the wall  This location is perfect  because it is directly behind the source  therefore only  secondary x rays will ever reach it  This is also the  location at which all controls will be placed  so a user    need only stand up and he or she will be at the entrance        Figure 5  Windows and Door    Safety  Keys    The door to the hutch will be locked with a kirk key   which utilizes the method of trapped key interlocking   This is when there are two locks but only one key  In this  case the door will have a lock and the power supply will  have a lock  so the door cannot be unlocked while the  power is on     Shutter vs  Switch    A recurring issue for a while has been the sources  warm up time  two hours  2   This means there is a wait  of two hours from when the source is turned on to when it  can be used  This is not a major problem in itself  but  every time it is turned off there is cooling that takes place   which means 
9. s was removed and the cabinets  were pulled out  the plumber could be called in to cut off  the gas and water lines  Unfortunately it took the plumber  a number of weeks to arrive  Once he came and removed  the pipes  there was still a nub left poking from the wall   so the shielding has to sit a few inches from the wall to fit  the cut pipes     Shipping Times    The major source of trouble with this project has  been shipping times  There is nothing to be done once  everything has been designed and all there is to do but  wait for materials  The reasons provided for the    continuing extensions have ranged from a truck    breakdown to just plain forgetting to ship the order   Company Difficulties    The companies that have manufactured the source   chiller  and framing material have not helped to make this  project easy  As previously stated  the framing material  took so long to arrive because the order was neglected   and a _ shutterbox is required because the source  manufacturer has relayed minimal circuitry information   There have also been a number of small issues that have  plagued the process  such as the chillers pressure gauge   The company that manufactured the chiller neglected to  test the pressure gauge prior to shipment  Fortunately the    replacement was very simple   Current State    At this point in time the lead paneling has yet to  arrive  Once it does it will be a simple matter of cutting  and bending the pieces  The difficulty will lie in placing  the
10. there is a wait once it is turned back on  while it resumes the appropriate temperature  Because  users will be entering the hutch throughout the day to  manage their experiments  the x rays will have to be  controlled     This can be accomplished in one of two ways  a  shutter box or a high voltage switch  By placing the  source in a lead shielded box with a tungsten shutter on it  the source can stay on while only the shutter has to open    and close  The high voltage switch completely turns off  the x rays while allowing the filament to stay on   preventing cooling from occurring  The switch is more  attractive becomes it appears more streamlined  there are  no mechanical pieces to worry about  but it was decided  against  This is largely the fault of the company from  which the source was purchased  In order to set up a high  voltage control switch there is a lot of information about  the circuits that is required  unfortunately the company  has provided little to no assistance in that category  As    such the solution is a shutter box        Figure 6  Shutterbox  Interlocks    Built into the controller are two interlocks  a thermal  interlock and a door interlock  The thermal one is  necessary because heat is the primary method from which  the source can become damaged  So  if the temperature  rises above 95 degrees Celsius or the flow rate drops  below 0 1 liters per minute  the unit will automatically  turn off     The door interlock is designed for personnel safety  i
    