1. 100 120VAC Not Applicable 200 240VAC Complies other than domestic and those directly connected to the public low voltage power supply network that supplies buildings used for domestic purposes 2 Electromagnetic Immunity The CXDI 50G is intended for use in the electromagnetic environment specified below The user of the CXDI 50G should assure that it is used in such an environment Safety Information Immunity Test EC 60601 ESmplanes Electromagnetic Environment Guidance Test Level Level Electrostatic dis 1 2 4 6 kV 2 4 6 kV Floors should be wood concrete or charge ESD contact contact ceramic tile If floors are covered with syn IEC 61000 4 2 thetic material the relative humidity 1 2 4 8 kV 2 4 8 kV should be at least 30 air air Electrical fast tran 12 kV for 12 KV for Mains power quality should be that of a sient burst power supply power supply typical commercial or hospital environ IEC 61000 4 4 lines lines ment 11 kV for 11 kV for input output input output lines lines Surge 11 kV differ 1 kV differ Mains power quality should be that of a IEC 61000 4 5 ential mode ential mode typical commercial or hospital environ ment 12 kV com 12 kV com mon mode mon mode Voltage dips short 596 Ur lt 5 Ur Mains power quality should be that of a interruptions and volt 29596 dip in 295 dip in typical commercial or hospital environ
2. Prone to vibration Insecure place Dusty environment Saline or sulfurous environment High temperature or humidity Freezing or condensation Installation Operation N WARNING Do not connect the instrument with anything other than specified Otherwise it may result in fire or electric shock Safety Information Power Supply Be sure to turn OFF the power of each instrument before connecting or dis connecting the cables ZN WARNING Also do not handle them with wet hands Otherwise you may get an electric shock that may result in death or serious injury Be sure to hold the plug or connector to disconnect the cable N WARNING If you pull the cable the core wire may be damaged resulting in fire or electric shock Do not cut or process the cables Also do not place anything heavy including the instrument on it step on it pull it bend it or bundle it ZN WARNING Otherwise the cable may be damaged which may result in fire or electric shock Do not turn ON the system power when condensation is formed on the instru ZN WARNING ment Otherwise it may result in fire or electric shock Because the instrument s cable is long take care so cables do not get tangled N CAUTION during use Also be careful not to get your feet caught in the cable Handling Always be sure to keep checking the condition of the system and the patient A to ensure th
3. ing exposures using a phan tom or a resolution chart Good Bad Good Bad Good Bad Cone MOM EE or if there is any problem General 5 1 4 Calibration Perform calibration when exposure conditions have changed significantly See the CXDI Series Operation Manual for the procedure for calibration 5 2 Cleaning 5 2 Cleaning 5 2 1 CFRP Carbon Fiber Reinforced Plastic Plate of the Sensor Unit Wipe the CFRP plate of the sensor unit with ethanol or glutaralde hyde solution to disinfect it each time a different patient uses the instrument in order to prevent infection If the optional grid unit is being used disinfect its surface N WARNING _ Please consult a specialist for the procedure for disinfection If you are using disinfectant other than those specified above or you are mixing another disinfectant with ethanol please also con sult a specialist because they may harm the CFRP plate and grid unit Disinfect the CFRP plate of the sensor unit each time a different patient uses the instru ment If the optional grid unit is being used disinfect its surface 5 2 2 Cover When the instrument is going to be cleaned be sure to turn OFF the power of each instrument and unplug the power cable from N WARNING the AC outlet Never use alcohol benzine thinner or any other flammable clean ing agents Otherwise fire or electric shock may result Clean the cover
4. Ensure that the grid unit is locked before lifting the sensor unit Otherwise the grid unit will fall and be damaged To remove the grid unit While sliding the locks on top of the grid unit outward lift and remove the grid unit 18 Canon CANON INC Medical Equipment Group 30 2 Shimomaruko 3 chome Ohta ku Tokyo Japan Telephone 81 3 3758 2111 CANON U S A INC CANON MEDICAL SYSTEMS 15955 Alton Parkway Irvine CA 92618 3616 U S A Telephone 1 949 753 4160 CANON EUROPA N V Medical Products Division Bovenkerkerweg 59 61 1185 XB Amstelveen The Netherlands Telephone 31 20 545 8926 PUB L IE 4134A 0406P0 001 Printed in Japan
5. N WARNING _ If itis not repaired properly it may cause fire electric shock or accident A For safety reasons be sure to inspect the instrument before using CAUTION it In addition carry out a regular inspection at least once a year 5 1 Inspection In order to ensure that the instrument is used safely and normally please be sure to inspect the instrument before use If any problem is found during the inspection please take measures indicated in this chap ter If problem still cannot be corrected please contact Canon representative or distributor It is recommended that a record of the inspection be kept by making copies of the check lists in this section or making a separate check list 5 1 Inspection 5 1 1 Daily Inspection 5 1 1 1 Before Turning ON the Power For safety reasons be sure to turn OFF the power of each instru N WARNING ment when the following inspections are going to be performed Otherwise it may result in electric shock Result Inspection Date Date Date Remedy Check that cables are not Contact C distrib damaged or cover of cables Good Bad Good Bad Good Bad A E tor if there is any problem 1s not torn ho O Check that the plugs and locks of connectors are not Good Bad Good Bad Good Bad Fully serine cablesiand lock them loose 3 Check that the cover or parts Lave O are not damaged a
6. and EN60601 1 This unit can be installed in the patient environment These marks indicate that the instrument Y must be handled with care Do not jolt or apply excessive load to the instrument ay a o si O o 1 Canon CXDI 50G ce x Wake FP vA Ed SERRATIS FAF3 30 2 ST RT ER HS HIER ENS TU751 224227 Ed FIIWFIAIFTt CXDI 50 ERRBRRBS 21500BZZ00363000 ERRBO752 BEERS PERT SMES WEES S N NO CANON INC 30 2 SHIMOMARUKO 3 CHOME OHTA KU TOKYO JAPAN MANUFACTURED MADE IN JAPAN EMCili amp IEC 60601 1 2 2001 This product conforms with DHHS radiation o performance standard 21CFR chapter 1 subchapter J 2 os N SRPMATABESAUTC EAU Do not jolt or apply excessive load Name Label 9 Contents Safety InTOrmalanzse RES DE Ee Ee uu ot dice ou nfi Ld tau Se 1 Regulations sae EE ta RUP pm tete pea bat fess ad fec det esas 1 Guidance and Manufacturer s Declaration for EMC Directive iese esse ee ee ee ee ke ee ee ee Re ee ee 2 General safety Information MEEL DE Re GRIP GE RI RAE OX ORUM 6 Labels and Markings on the Instrument sssssssse een eene 9 EST TNR 1 2 Notes for Using the Instrument eee 2 3 Deseriplion A MO EAE 5 A Operation dec hoe Mec teats oscar Vow de Setar ES v
7. die 1 Installation parts essen 1 set Options Digital Radiography CXDI 50 Power Box with Remote switch Grid unit for CXDI 50 Grid 40 line cm 110 cm 4 1 40 line cm 110 cm 8 1 40 line cm 150 cm 6 1 40 line cm 180 cm 10 1 9 Dimensions Sensor Unit Q EM d I mM ae i i jn zu ME ER NS LI lid RS Hr Se en aA N N i J Y M B 491 zl 23 Unit mm 16 Appendix Attaching the Optional Grid Unit Attach the optional grid unit to the sensor unit by the procedure indicated below NOTE Structure of grid unit is delicate Do not drop knock over bend or apply force or jolt to it Otherwise the unit may be damaged When the grid unit is not in use remove it from the sensor unit and store it in a safe place where it will not fall Even if it is not extremely damaged its characteristics may be changed which may cause a problem in image quality 1 Blow off the dust on the front and back of the grid unit 2 Place the sensor unit horizontally on a table 3 Hold the grid unit with both hands and engage the hooks on the bottom of the grid unit to the holes on the bottom of the sensor unit 17 Appendix Aftaching the Optional Grid Unit 4 Gently set the grid unit over the sensor unit A click sound is heard when the grid unit is locked properly NOTE
8. Canon DIGITAL RADIOGRAPHY CXDI 50G User s Manual Before using the instrument be sure to read this manual thoroughly Also read the manuals of other instruments in the system Keep the manual where it is easily accessible PLEASE NOTE 1 The user is responsible for the use and maintenance of the product We suggest that a member of the user s staff be designated as being in charge of mainte nance so as to ensure that the product is kept in a safe and good condition Also medical products must be used only by a qualified person 2 Roentgenography image processing reading of image and storage of data must be per formed in accordance with the law of the country where the product is being used Also the user is responsible for maintaining the privacy of image data 3 Inno event will Canon be liable for direct or indirect consequential damage arising out of the use of this product Canon will not be liable for loss of image data due to any reason 4 This product may malfunction due to electromagnetic waves caused by portable personal telephones transceivers radio controlled toys etc Be sure to avoid having objects such as these which affect this product brought near the product 5 Canon reserves the right to change the specifications configuration and appearance of the product without prior notice European Union and EEA only This symbol indicates that this product is not to be disposed of with your household
9. RE MECHANICAL MEDICAL EQUIPMENT WITH RESPECT AND OTHER SPECIFIED HAZARDS ONLY IN SNSSIF TO ELETRICAL SHOCK FIRE AND ACCORDANCE WITH CAN CSA C22 2 NO 601 1 S o MECHANICAL HAZARDS ONLY IN MEDICAL EQUIPMENT CERTIFIED FOR CANADA U ACCORDANCE WITH UL60601 1 C a L US lt CONTROL NUMBER 41C4 gt Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle marked Hospital Only or Hospital Grade For EU Countries The following mark shows compliance of the instrument with Directive 93 42 EEC CE This instrument has been classified into EN55011 Group 1 Class A This instrument is a CLASS EQUIPMENT according to EN 60601 1 1 Safety Information Guidance and Manufacturer s Declaration for EMC Directive Electromagnetic Emissions The CXDI 50G is intended for use in the electromagnetic environment specified below The user of the CXDI 50G should assure that it is used in such an environment Emission Test Compliance Electromagnetic Environment Guidance RF emissions GROUP 1 The CXDI 50G uses RF energy only for its internal func CISPR 11 tion Therefore its RF emissions are very low and are not likely to cause any interference in nearby electromagnetic equipment RF emissions Class A The CXDI 50G is suitable for use in all establishments CISPR 11 Harmonic emissions IEC 61000 3 2 Not Applicable Voltage fluctuations flicker emissions IEC 61000 3 3
10. age variations on Ur for 0 5 Uy for 0 5 ment If the user of the CXDI 50G requires power supply input cycle cycle continued operation during power mains lines interruptions it is recommended that the IEC 61000 4 11 40 Ur 40 Ur CXDI 50G be powered from an uninter 60 dip in 60 dip in ruptible power supply or a battery U1 for 5 Ur for 5 cycles cycles 70 Ur 70 Ur 30 dip in 30 dip in Ur for 25 Uy for 25 cycles cycles 596 Ur 596 Ur 295 dip in U1 for 5 sec 295 dip in Ur for 5 sec Power frequency 50 60Hz magnetic field IEC 61000 4 8 3 A m 3 A m Power frequency magnetic fields should be at levels characteristic of a typical loca tion in a typical commercial or hospital environment NOTE Ur is the a c mains voltage prior to application of the test level Safety Information Immunity Test IEC BOROT Test Compliance Electromagnetic Environment Guidance Level Level Conducted RF 3 Vrms 3Vrms Portable and mobile RF communications IEC 61000 4 6 150 kHz to 80 equipment should be used no closer to MHz any part of the CXDI 50G including cables than the recommended separation distance calculated from the equation Radiated RF 3 V m 3 V m applicable to the frequency of the trans IEC 61000 4 3 80 MHz to 2 5 mitter GHz Recommended separations distance d 1 2 P d 1 2 P 80 MHz to 800 MHz d 2 3 P 800MHz to 2 5 GHz where P is t
11. ain ze aute afa o NS 6 4 1 Turning ON OFF the Power of the System esse ees eed ee ee de ee ee ee ee Re ee emen 6 4 2 Calibratlon ito at RE N a E Dd ER RE EN 6 5 Inspection and Maintenance ce coco dusk Ei etis dead 7 5 1 Te eie OE ee aaa 7 FASE ERA eite ne E EE NE MO EE EE N 10 6 Service Information onere ES Ge EE EER EE GEE aa EE ER I HU MM MEL 11 27 cler Teilercli o st rare Dies 12 7 1 Main Specifications RE AE OE ue OER ER N 12 Ne Te ER EE nel EE EE 13 8 COMPONEN Sa E A ee UP otc 15 DIMENSIONS ER notiert radi Eg ed EDE Iba e er ee eb fu ode aduer 16 Appendix Attaching the Optional Grid Unit ee EE EE nn 17 1 Overview The Canon Digital Radiography CXDI 50G is a portable X ray digital camera that can take images of any part of the body It directly converts the X ray images captured by the LANMIT Large Area New MIS Sensor and TFT sensor into a high resolution digital images 2 Notes for Using the Instrument 1 2 Handling Handle the instrument carefully as it may be damaged if something is hit against it dropped or receives a strong jolt Also handle the optional grid unit with care too Do not pull the cable Also do not pull the sensor unit when the cable is tangled with something Other wise the cable may be damaged which may result in fire or electric shock Before Exposure Sudden heating of the room in cold areas will cause condensation to form
12. by the following procedure if it is dirty 1 5 Turn OFF the power of the power box Shut down the control computer and press side 0 of the power switch Turn OFF the power of each instrument Turn OFF the power of each instrument if connected Unplug the power cables Unplug the power cables of each instrument from the AC outlet Wipe the cover using neutral detergent Wipe the cover with a piece of cloth soaked in neutral detergent diluted in water and wrung dry Wipe out neutral detergent Wipe the cover with a piece of cloth soaked in water and wrung dry whenever neur tral detergent has been used ejes 6 Service Information 1 2 Repair If problem cannot be solved even after taking the measures indicated in chapter 5 contact Canon representative or distributor for repair Please refer to the name label and let us have the following information Name of the unit It is indicated on the name label Serial number 6 digit number indicated on the name label Phenomenon In detail Limit for Supplying Performance Parts for Repair Performance parts of this product will be stocked for eight years after discontinu ance of production to allow for repair Parts required to maintain the functioning of the product Mi 7 Specifications 7 1 Main Specifications Sensor Unit Purpose Pixel size Image matrix size Number of pixels Dynamic range Resolution Gray scale Environme
13. er of each instrument unplug the power cable from the AC outlet and contact Canon representative or distributor When there is smoke odd smell or abnormal sound When liquid has been spilled into the instrument or a metal object has entered through an opening When the instrument has been dropped and it is damaged Maintenance and Inspection N WARNING For safety reasons be sure to turn OFF the power of each instrument when the inspections indicated in this manual are going to be performed Otherwise it may result in electric shock N WARNING When the instrument is going to be cleaned be sure to turn OFF the power of each instrument and unplug the power cable from the AC outlet Never use alcohol benzine thinner or any other flammable cleaning agents Otherwise fire or electric shock may result N WARNING The instrument must be repaired by a qualified engineer only If itis not repaired properly it may cause fire electric shock or accident N CAUTION For safety reasons be sure to inspect the instrument before using it In addition carry out a regular inspection at least once a year Safety Information Labels and Markings on the Instrument The CXDI 50G has a few labels and markings on it Contents of those and positions where they are attached are indicated below Sensor Unit Front his mark indicates that this is a Type B Applied Part according to UL60601 1
14. ey are normal during the use of the instrument WARNING If any problem is found take appropriate measures such as stopping the operation of the instrument as required Never disassemble or modify the product as it may result in fire or electric A shock Also since the instrument incorporates parts that may cause electric WARNING shocks and other hazardous parts touching them may cause death or serious injury Do not hit or drop the instrument The instrument may be damaged if it N WARNING receives a strong jolt which may result in fire or electric shock if the instru ment is used without it being repaired Do not spill liquid or chemicals onto the instrument or in cases where the A patient is injured allow it to become wet with blood or other body fluids as CAUTION doing so may result in fire or electric shock In such situation protect the instrument with disposable covering as necessary Wipe the CFRP plate of the sensor unit with ethanol or glutaraldehyde solu tion to disinfect it each time a different patient uses the instrument in order to N CAUTION prevent infection If the optional grid unit is being used disinfect its surface Please consult a specialist for the procedure for disinfection Turn off the power of each instrument for safety when they are not going to be NCAUTION used i SN Safety Information When Problem Occurs N WARNING Should any of the following occur immediately turn OFF the pow
15. he maximum output power rat ing of the transmitter in watts W accord ing to the transmitter manufacturer and d is the recommended separation distance in metres m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey should be less than the compli ance level in each frequency range Interference may occur in the vicinity of equipment marked with the following sym bol Cp NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflections from structures object and people Field strengths from fixed transmitters such as base stations for radio cellular cordless tele phones and land mobile radios amateur radio AM and FM radio broadcast and TV broadcast can not be predicted theoretically with accuracy To assess the electromagnetic environment due to fixed RF transmitters an electromagnetic site survey should be considered If the measured field strength in the location in which the CXDI 50G is used exceeds the applicable RF compliance level above the CXDI 50G should be observed to verify normal operation If abnormal performance is observed additional measures may be necessary such as reorienting or relocating the CXDI 50G P Over the frequency range 150 kHz to 80 MHz field strengths should be less than 3 V m 4 Safety Informa
16. labels Ignoring such cautions or warnings while handling the product may result in injury or accident Be sure to read and fully understand the manual before use Keep this manual for future reference Meaning of Caution Signs This indicates a potentially hazardous situation which if not heeded could N WARNING result in death or serious injury to you or others This indicates hazardous situation which if not heeded may result in minor or ZN CAUTION moderate injury to you or others or may result in machine damage NOTE This is used to emphasize essential information Be sure to read this information to avoid incorrect operation Installation and Environment of Use N WARNING Do not use or store the instrument near any flammable chemicals such as alcohol thinner benzine etc Also this instrument is not a category AP or APG equipment If chemicals are spilled or evaporate it may result in fire or electric shock through contact with electric parts inside the instruments Also some disinfectants are flammable Be sure to take care when using them N CAUTION Do not install the instrument in a location with the conditions listed below Otherwise it may result in failure or malfunction fall or cause fire or injury Close to facilities where water is used Where it will be exposed to direct sunlight Close to air conditioner or ventilation equipment Close to heat source such as a heater
17. maged 3 Description Sensor Unit This unit converts the X rays into digital signals a d b C e f Bo M ui EK OT TA 7 rn anne Sun REA L L Name Description Grip Hold this grip when carrying the sensor unit CFRP Carbon Fiber Reinforced Plastic plate The part of the patient s body to which an image is to be taken should be placed against this plate This plate should be disinfected each time a different patient uses the instrument in order to prevent infection Sensor cable Connect this cable to the power box BUSY lamp Blinks when the sensor unit is busy communicating SENSOR lamp Blinks when the sensor unit is ready for exposure or when error has occurred POWER lamp Turns ON when the power of sensor unit is ON 4 Operation 4 1 Turning ON OFF the Power of the System When tuming ON OFF the power of the sensor unit refer to the Power Box option Oper ation Manual 4 2 Calibration Calibration is important to ensure that a good image is achieved with the CXDI 50G by obtaining the calibration data of the sensor unit Perform calibration when exposure condi tions have changed significantly NOTE For procedure for calibrating refer to Calibrating the Instrument in the CXDI Series Operation Manual 5 Inspection and Maintenance The instrument must be repaired by a qualified engineer only
18. nd not Good Bad Good Bad Good Bad a Canon rat bir D 1 tor if there is any problem S loose Yn 5 1 1 2 After Turning ON the Power Result Inspection Date Date Date Remedy Check that power lamp is lit Good Bad Good Bad Good Bad Cooner oe poseneaple dno sensor cable properly P Take measures if error mes i sage is displayed by refer 5 O Perform test exposure Good Bad Good Bad Good Bad ring to the CAD Series Operation Manual Contact Canon or distributor if the problem cannot be solved 5 1 Inspection 5 1 2 Monthly Inspection Perform the following inspection periodically more than once a month Contact Canon representative or distributor 1f there is any problem or if you cannot do it Result Inspection Date Date Date Remedy Check the performance of the instrument by perform Operation Manual for the ing exposures using a phan Good Bad Good Bad Good Bad procedure for self test Con tom or a resolution chart or tact Canon or distributor if perform self test there is any problem Refer to the CXDI Series General 5 1 3 Yearly Inspection Perform the following inspection periodically more than once a year Contact Canon repre sentative or distributor if there is any problem or if you cannot do it Result Inspection Date Date Date Remedy Check the performance of the instrument by perform
19. ntal requirements Power Dimensions and mass General radiography 160 um x 160 um 2208 x 2688 pixels Approx 5 9 million pixels Approx 80 dB 3 1 Ip mm 12 bit 4 096 gray scale Operation Temperature 5 to 35 C Humidity 30 to 75 RH no condensation Storage and transportation Temperature 30 to 50 C Humidity 10 to 60 RH no condensation Atmospheric pressure 700 to 1060 hPa Supplied from the power box 491 W x 477 H x 23 D mm 4 8 kg without sensor cable md 7 2 Characteristics 7 2 Characteristics 1 Reguired patient doses Equivalent to exposure dose of 200 or 400 speed film screen system 2 Sensitometric characteristics and dynamic range Exposure output characteristics 100000 10000 2 1000 a 3 io 6 10 0 001 001 01 1 10 100 Exposure mR Dynamic range is determined by the levels of system noise and saturation The level of uncertainty is estimated to be less than 410 ao 7 2 Characteristics 3 Presampling MTF 0 8 0 6 0 4 Presampling MTF 0 2 0 1 2 3 4 5 Spatial frequency Ip mm The level of uncertainty is estimated to be less than 42 or 0 015 whichever is greater 4 DOE DQE f X 0 01 0 0 5 1 1 5 2 2 5 3 Spatial frequency Ip mm The level of uncertainty is estimated to be less than 10 or 0 02 whichever is greater cdi 8 Components Sensor UNI ar ESE SE GE GEE GR RO e e
20. on the instrument In this case wait until condensation disappears before performing expo sure Ifthe instrument is used with condensation formed on it problems may occur in the quality of the instrument When an air conditioner is going to be used be sure to raise lower the temperature gradually so that a difference in temperature in the room and in the instrument does not occur to prevent forming of condensation 2 Notes for Using the Instrument 3 During Exposure Do not apply excessive weight to the sensor unit Otherwise the sensor may be damaged Limit of Load Uniform load 150 kg over the whole area of sensor unit surface Local load 100 kg on an area 40 mm in diameter Be sure to use the sensor unit on a flat place so it will not bend Other wise the sensor may be damaged Perform exposure after checking that the exposure conditions are optimally suited to this product The sensitivity of this product s sensor differs from that of the series product CDXI 50C 2 Notes for Using the Instrument 4 During Cleaning Do not use anything other than neutral detergent for cleaning the cover of the instru ment Otherwise the coating will be corroded Storage Be sure to store the sensor unit and the optional grid unit in a safe place where it will not fall or drop Others Be sure to reconnect the cables to the proper connectors Otherwise the instrument may malfunction or may be da
21. tion Recommended Separation Distances The CXDI 50G is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled The user of the CXDI 50G can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communica tions equipment transmitters and the CXDI 50G as recommended below according to the maximum output power of the communications equipment Separtion distance according to frequency of transmitter m Rated maximum output power of ey 150 kHz 80 MHz 80 MHz 800 MHz 800 MHz 2 5 GHz w AJP AJP ABP 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 12 23 10 38 38 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the equation applicable to the frequency of the transmitter where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer NOTE 1 At 80 Mhz and 800 MHz the separation distance for the higher frequency reange applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people Safety Information General Safety Information Follow the safety instructions in this manual and all warnings and cautions printed on the warning
22. waste according to the WEEE Directive 2002 96 EC and your national law This product should be handed over to a designated collection E point e g on an authorized one for one basis when you buy a new similar product or to an authorized collection site for recycling waste electrical and electronic equipment EEE Improper handling of this type of waste could have a possible negative impact on the environment and human health due to potentially hazardous substances that are generally associated with EEE At the same time your cooperation in the correct disposal of this product will contribute to the effective usage of natural resources For more information about where you can drop off your waste equipment for recycling please contact your local city office waste authority approved WEEE scheme or your household waste disposal service For more information regarding return and recycling of WEEE products please visit www canon europe com EEA Norway Iceland and Liechtenstein System and product names in this manual are the trademarks of each manufacturer that devel oped them CANON INC 2006 All rights reserved Under copyright laws this manual may not be copied in whole or in part without the written consent of Canon Safety Information Regulations This instrument conforms to IEC 60601 1 2 2001 For U S A and Canada This instrument is a CLASS EQUIPMENT according to UL60601 1 WITH RESPECT TO ELETRIC SHOCK FI