Home
        Room Analyzer PRO ENG
         Contents
1.           Room Analyzes A FaR     Freq  286 34H2 Mag     Room Mode Measurement  Instructions  Place the microphone to the left  in the room as shown in the  picture   Left Speaker Right Speaker Then press Measure     Subwoofer            7    verage 3 positions  Mic    DO    ESC   main        Cancel         With a    double click    in the Room  Analyzer window you change  information between Room Analyzer   1  and spectrogram  3      Room Analyzer    Shows the frequency response  Blue curve   room  modes  Green curve  and calculated result curve     White curve  Select what curve to display by    clicking the boxes       Modes EQ  if Measured    11 of 37                            lyzer RTA SPL          Measured frequency response  in real time  with a  resolution of 1 3 octave     Spectrogram     reverberation time       WS w12 bas vbid v  gg   XTZ  elp                                     Ele view Setup He    CEE  G d8  Type  2 H Mode o  1 675 Mode      Measured  Result smoot th  3ppo  Mades smooth   2ppo  Meas  smooth   Stimulus EQ 3ppo  Use flH  Q G  dB    ms   50 0  Store Curve  Load    Clear StimuusEQ    CopyRoomModes     Save      Export  lag  23 6508          Shows the result of the room analysis     The Time spectrum analyzer displays a colour  spectrum in frequency  level and time    The spectrum shows the reverberation time in the  room     12 of 37    File View Setup Help    RE  11    Full Range             FoomAnabzer  RTA   SPL            Shows the frequency res
2.         If the final result is not to your satisfaction  it could be  caused by measuring disturbance  Please redo the  measuring procedure      More information about Frequency response    RTA  see page 18  Full Range page 30     27 of 37       Manual adjustment  of curves  Drag  amp   Drop adjustment     Adjustment of the  time axis    These options are only displayed in    Advanced mode       For the advanced user there is a function available to  manually adjust the frequency response curve     By grabbing a peak or dip  and right clicking the mouse   you may correct the curve manually  By moving the  mouse sideways the frequency is altered        Releasing the mouse button will produce new  suggestions for the PEQ in order to achieve the desired  curve  The new PEQ suggestions are editable for amore New PEQ suggestions may be  precise adjustment  This is done by double clicking the imported to XTZ W12 DSP    value to change  or right click then select    properties     without overwriting data  imported earlier  These have  to be added manually     r    This requires more PEQ    s available     Important    Use these options with great care  An increase at  any frequency may cause substantial amplifier  load     Important     Room Analyzer measures the frequency response at time  zero  This point is indicated with a yellow line in the  spectrogram  The yellow time line can be moved by  clicking it  then holding the left mouse button down  while moving the mouse  The freq
3.      99    Only visible in    advanced mode    Export multi    Export ROOM  RTA or both information to XTZ ni test   XIZ  DSP Products    ROM RTA BOTH         15 of 37    This is about File menu and the options              File wiew Setup Help  BISA  ROM RTA BOTH           File  New  Open Loads earlier stored curves  Save Saves the measurement  Save as     Export curve Exports the measurement to a textfile  Export Screen Dump Exports the measurement to a picture   bmp   Print Prints the measurement Black White  View  Toolbar Shortcuts to    Save        Load    resp     Print    functionen  Statusbar Shows the marker in Frequency and magnitude  Color Theme Dark Bright   Changes color theme in the program Bright or Dark  Setup  Advanced mode More options settings for some functions  e Stimulus EQ  e Smoothing possibility  e Load Save  curves   e Change of scale  RTA window    e Export Room Analyzer  amp  RTA  Help  About XTZ    Software version    16 of 37    Preparation    Set up    Level adjustment  Output level    Make sure the software is installed and the equipment is See mounting  connected     Before the measuring procedure starts  make sure the  microphone is placed in ear height when adjusting the  output level         Start    Room Analyzer    PN Room Analyzer   XTZ    File View Ed Help  E  amp  v Advanced Mode       For advanced users  there is a choice called    Advanced  Mode  lt     More settings and features can be found in Advanced  mode     In order to achieve
4.  a reliable measuring procedure it is  important that the correct output level is set     Put the control on the output level in the middle   Now there should be a white noise in the speakers     The level meter to the right should be up to the green       G     If the level is too high or low  adjust it with the ipur Level  volume control on the system subwoofer        Click on    STOP    to turn off the white noise  Action  Now there is a possibility to make small adjustments on  the output level with the control bar  H    REMEMBER    HIGH SPL level can  damage your ears     17 of 37    Basic principles    Measuring methods    The subwoofer  system should have as neutral settings  as possible before the measurement begins  This way   the smallest possible adjustments can be used     The basic principle is that the measurement is made in    two steps     Step 1     Room Analyzer   Room modes are best treated with PEQ  Parametric  EQ   for more accurate precision    Use Room Analyzer at high resolution to find the    room modes  If the Room Analyzer suggests the use of    PEQ  Parametric Equalizer  you will need this in your  equipment     Step 2     RTA Full Range   After adjustments of the room modes  adjust the  frequency response in the sweet spot with the RTA    Full Range  frequency response  with graphic EQ  and or parametric EQ     There are different methods for measuring the    frequency response     1     Frequency response   RTA  A simple measurement which gives t
5.  di Poor odes  se 2            G  dB    Type  M1 5 43 6  24 00 Mode  W2 108 36 6 16 75 Mode    MODE  Room Modes  Is the first priority for Room Analyzer   EQ  A peak dip that needs treatment     The values that are presented in Found Room Modes are  values that directly can be used in a subwoofer with  PEQ     The values can be set manual  by right click the value   This can be exported to XTZ DSP products     The presented values in    Found Room Modes    can be  exported     Click on    Export    to export the presented value  It is possible to export the values by clicking on    ROOM  in the short cut menu    The exported settings can now be imported to 99W12  DSP by PC REMOTE  Software to do settings on 99W12  DSP     The Room Analyzer presents values  based on the   measurements  for use in a PEQ  These values can be dO DA  copied to the Stimulus EQ window  If a new         measurement is made with values in Stimulus EQ  there  will be a result as if there had been an adjustment with a  PEQ     Click on    Copy Room Modes    and the values are copied  to the Stimulus EQ window     Remove these by clicking on    Clear Stimulus EQ       With a graphic EQ  the frequency response can be  adjusted  with Frequency response   RTA Full  Range    The room compensation is now done     If the measures are made with a subwoofer  there will be  a difference when the complete system is connected    So it could be wise to do a Frequency response    RTA Full Range on the complete system
6.  microphone to the left in the room    as shown on the picture     Room Mode Measurement    Instructions    Place the microphone to the left  in the room as shown in the  picture    Right Speaker Then press Measure     Left Speaker    Subwoofer    V Average 3 positions    Mic      O       Click on    Measure    and the measuring process starts    A sinus sweep will be played     Move the microphone to the sweet spot   as shown in the picture     Room Mode Measurement    Instructions    Move the microphone to the  listening position as shown in the  picture     Left Speaker Sar Right Speaker Then press Measure        Titi ie  IF vi    Cancel       Click on    Measure    and part 2 starts     Move the microphone to the right in the room   as shown on the picture     Room Mode Measurement    Instructions      Move the microphone to the right  in the room as shown in the  p picture   Left Speaker aros Right Speaker Then press Measure   Mic    DO      i    CEER      Cancel       Click on    Measure    and measuring 3 starts     The measuring procedure is now finished     23 of 37          Read the difference  between 1 and 3 position  measurement   Technical  explanation    Room Analyzer    After a short while of calculations  the result is  presented     In the main window  the frequency response is now  shown    To the right in the smaller window  the reverberation  time  Ifyou double click in the main window the  information in the windows changes              There are three d
7. Analyzer    but a good indication how the  average sound is     Full Range is the most advanced tab in Room  Analyzer  Here is a high resolution full range  measurement  There are options about how  much room information that should be part of  the measurement  The different information  can be shown with 3 curves at the same time     RT60 is a reverberation measurement   We are doing it according to ISO 3382  Displayed with standing bars     We also have a water fall diagram  in both 2D   amp  3D  A lot of information can be found in this  diagram     Equalizer is a tool for raising or lowering  levels on different frequencies  A graphic  equalizer has pre defined bands  Frequency  area   A parametric equalizer can be adjusted  on a specific frequency  Q value etc     35 of 37    Smoothing    Connect Room  Analyzer to your  HIFI system    Smoothing determines the resolution of the  curves    This is a balance depending on equipment as  well as your individual choice     With the highest resolution it is impossible to  achieve a completely flat frequency response     With smoothing the resolution can be lowered  in steps  so it is more equal with the resolution  of the ear  in the low area     With a good listening room  a competent  system and good experience of sound  the  highest resolution can be used    For people with less experience  the  recommendation is to use a lower resolution        If Room Analyzer is to be connected to the  system  Front speakers   Sub  a Y cab
8. In the window you can find the modes the software  has found and the recommendations for  countermeasures                 nAverages    4  UdB      Referred to XTZ 99W12 DSP or other subwoofers Result amott  with PEQ  3ppo     Modes smooth   stimulu   3 pp        Use IHJ  Q G  dB     Meas  smooth      Ynnn T    Smoothing   Smoothing determines the resolution of the curves    This is a balance depending on equipment as well Gb   as your individual choice     Result smooth    With the highest resolution it is impossible to     T   achieve a completely flat frequency response   a    Meas  smooth     2 ppo  3 ppo and 6 ppo is the option   ppo   part per octave        Stimulus EQ   Only visible in    advanced  mode       Stimulus EQ Values are suited for XTZ 99W12  The Room Analyzer proposes some values for use DSP or other subwoofers with PEQ  with a PEQ  Parametric EQ   these values can be       copied to Stimulus EQ window  If making a new e ay eT   measurement with the Stimulus EQ  the result i ee o    would be like using a PEQ     Clear Stimulus EQ           Stimulus EQ   Only visible in    advanced  mode       14 of 37       Store curve     Full Range Fesul    Modes Et      Measured    An option to store 4 curves on the screen at the Free smoath  Roo time  Essen  Excellent tool to have when comparing different Meas  smooth    settings and or placements of the speaker   Gives you the result direct on the screen        Overlay    A click on    Show    turns on off the curves   
9. XTZ ROOM ANALYZER PRO    Software for room analyzing and measuring  frequency response      a mis nessa a E  U SCI i iia    i a      N A     e  eS       2009 XTZ AB   www xtz se             Introduction  About XTZ  Technical presentation  Preparations  Basics of sound   Installation and placement tips  Mounting and connecting  Functional reference  Area of use  Measurement  Frequency response   RTA  Room Analyzer  e 1 Position measurement  e 3 Position measurement    Spectrogram    Advanced features   Full Range   RT60   Technical explanations  Technical specifications    Service  amp  Support    2 of 37    Page     O N OF A O    10  11   17   18   20  21 25  21   23   26    28  29  32  33  35  35    Introduction    Congratulations to your purchase of XTZ Room  Analyzer    XTZ Room Analyzer provides many possibilities  in  order to utilize this product in the best way  please  read the manual carefully before using Room  Analyzer  It takes patience to optimize a hifi system  If  you lack experience from this kind of measurements  or have any questions  please contact our    free of  charge    support   Support info on the last page of the  user manual     30137    Philosophy    XTZ Goals    Contact    Our reference and ambition is to reproduce a natural  sound  keeping in mind that acoustics is always a  matter of taste    With Room Analyzer  this goal is easier to achieve   Our main goal is to provide the best value for  money    Our concept       Cut the numbers of middleme
10. contrary  produce hard  reflections that may lead to a blurry dialogue  Apart  from colouring the sound  the perspective of the  sound will also deteriorate  Reflections from the  room can roughly be compared to the reflections that  cause ghost pictures on a TV screen     When a loudspeaker is placed near a wall or floor  lower frequencies will be amplified  sometimes in a  not desirable way  since it may lead to indistinct  sound reproduction   This amplification becomes  even more obvious if the loudspeaker is placed near a  corner     However  sometimes this can be desirable  and it is  important to have good knowledge about floors   walls and ceilings     There are constructions specially developed for  placement near walls  according to the instructions     Be aware that furniture may vibrate  creating  disturbing sounds and rattles at loud levels     Quadratic rooms or rooms where the length is exactly  twice as long as the width should be avoided  since  they may create unwanted resonances     The way the loudspeakers are placed towards the  listener is of great importance     In order to get the best results the front speakers  should be placed symmetrically in front of the  listener  The distance between the front speakers  should be about 80  of the distance to the listener   In other words  the recommended angle between the  front speakers should be 45       You can place the loudspeakers angled in towards the  listener or straight ahead     8 of 37       Correct 
11. e to correct  Change of the sweet  spot or speaker position is recommended in  this case        peak    An extreme dip is shown as a valley in the  frequency response measurement        dip  1 Position When optimizing the sound without  measurement compromising     1 position measurement     should be used  Ex A high end system in a  3 Positions room with good acoustics   measurement   Average 3 positions  The risk with    1 position measurement    is that    the sweet spot could be in a dip  Sometimes it  is impossible to compensate for without  having other problems  One solution is to  change the sweet spot and or move the  speakers  subwoofer     Control measurement or measurement of the  speaker or specific positions in the room   Use     1 position measurement        In normal case  the recommendation is    3  positions measurement     The measurement is  made on 3 positions in the room  to give a  more accurate result  finding the room modes  easier     Correcting the main modes in the room  gives  a better average listening result    Parametric EQ  99W12 DSP  helps a lot  working on the main modes  Use both  methods and compare the results to find  what s best for your room     Spectrogram Shows how long time it takes for the intensity  of the sound has reduced to a certain level  All  room has a certain reverberation time  A long  Reverberation time affects the sound     A hard hand clap and the following echo is an  indication of Reverberation time     34 of 37    R
12. g number  1  amp  3 active   The volume is set by adjusting the knob to the right  A Mieden  higher spl level often produces better measurement  accuracy  a  L J  6  L  Output Level    31 of 37    Start The measurement starts by pressing Measure   Full Range 4 sweeps follows and the measurement is done   measurement    This gives you results in form of frequency response  curves where you can view the curves with different time  windows  RT60 information  the reverberation time of  the room acc  to IS03382  And the waterfall diagram in  3D or 2D               60d v    Waterfall        Overlay   mem   2  Gow    DEE                       Room    Analyzer   RTA   Full Range      When measurements are done at further distance at  lower frequencies  dips are likely to occur  These are the  phase cancellations that depend on the room itself   microphone  and speaker placement  Measurements on  loudspeakers in lower frequencies  20 500Hz  should  be done at close range  0 50cm  to get an accurate  result  At distance over 50cm the room will effect the  measurement a lot  The different between close range  measurement and from a distance is the influence from  the room     To get an accurate evaluation of all the presented    information  knowledge and experience is a must  but  feel free to ask us anything if you have any questions     32 of 37    Reverberation time   T  The most important room parameter is the  reverberation time   T     It is defined as a time interval required fo
13. he  complete frequency range     Room Analyzer  High resolution   1 or 3 position high resolution measurement     1 position measurement    For optimizing the sound without  compromising    1 position measurement     should be used    Ex A high end system in a room with good    acoustics     The    risk    with    1 position measurement    is that  the sweet spot could be in a dip  Sometimes it is  impossible to compensate without having other  problems  One solution is to change sweet spot    and or move the speakers subwoofer     1 position measurement can be used for     Room Analysis   Frequency response   Frequency response on a speaker  Reverberation analysis    18 of 37    For users with PEQ options    Room Analyzer finds the  largest and most significant  room modes  The reason for  not finding all existing modes  is that the number of  suggested PEQ filters will  then be infinite     The additional info  corresponds only to    1  position measurement     If  you are not satisfied with the  resulting frequency response  the    High resolution  measurement    can be  repeated  More PEQ filters  can be added to achieve a  better end result     You can also use the function     Manual adjustments of  curves     described under     Advanced functions     This  requires more PEQ filters  being available        3 position measurement    In normal case  the recommendation is    3  positions measurement     The measurement is  made on 3 positions in the room  to give a more  
14. height    Cables    Background noise    Finally    By angling the loudspeakers inwards  you decrease  unwanted reflections  enhancing the stereo  perspective     If the loudspeakers are small  they should be placed  on stands or tilted so that they are on the same level  as the ears     Some speakers have other recommendations from  the manufacturer     in that case  follow those     Try to keep cables as short as possible  By its  electrical properties  a long conductor will have more  influence on the sound than a short one  It may also  work as an antenna  receiving various signals that  may become a constant noise in an active subwoofer   Make sure that all connections are clean and not  oxidized  All connections should be mechanically  stable  both power  signal and loudspeaker cables   Signal cables should be separated from other cables     Important     External noise from trucks  train  traffic etc can  easily disturb the measurement process    Ifthe measurement result looks strange in any way  you should do the same procedure again once or  maybe twice  If you get the same result again you can  be sure that it is correct     Please remember that    good    sound is a matter of  taste  so you have to experiment to obtain your  personal goal  We wish you best of luck     9 of 37       Placing the  microphone    Mounting    Connections       and connecting    Download the software from our website and install it  according to the instructions    Connect the microphone t
15. icking on the right side  see picture      e Measured   Blue curve  e Modes EQ   Green curve  e Result   White curve    The standard curve  the blue  is the measured frequency    response   80       21 of 37          Acton      Meas                       Modes EQ  Iw Measured    The green curve shows the frequency response for room  modes and EQ adjustments  Recommended in Found  Room Modes   The values in    Found room modes    are  those who are shown in the green curve       Result          Wo Measured  The White curve is the theoretical frequency response  after using PEQ settings  On the chosen spot    Result   The picture shows both measured frequency response  and the theoretical curve   m Hesu     Modes E0  W Measured          Result    7 ModestEQ  iw Measured  h Ave rages      4  0dB          Result smooth     E ppo          Modes smooth          E ppo       Meas  smooth       E ppo              Smoothing determines the  resolution of the curves    This is a balance depending  If another resolution is wanted  use smoothing  2  3  amp  6 on equipment as well as your    ppo  points per octave  can be chosen  individual choice     Only shown in    Advanced mode       22 of 37    Room Analyzer  High resolution   3 Position measurement    Start the  measurements    Click on measure to start   3 Positions measurements  The square    Average 3 positions    should be marked     To obtain a good result  the measuring spots should not  be very close to each other     Place the
16. ifferent curves   Choose by clicking on the right side  see picture      e Measured   Blue curve  e Modes EQ   Green curve  e Result   White curve             The standard curve  the blue  is the measured frequency  response       Modes EQ  W Measured       The green curve shows the frequency response for room  modes and EQ adjustments  Recommended in Found  Room Modes   The values in    Found room modes    are  those who are shown in the green curve     80      Result    iw iM odes E g   W Measured          24 of 37    The White curve is theoretical frequency response     On the chosen spot    Result  How it would look if the    values from    Found room modes    are used with PEQ    The picture shows both measured frequency response  and the theoretical curve     80    Only shown in    Advanced mode       If another resolution is wanted  use smoothing  2 3  amp  6  ppo  points per octave  can be chosen    25 of 37            Modes EQ  W Measured              Result  T Modes EQ  i    Measured  hn Averages      d       PAP      Result smooth      epp x     Modes smooth      Sppo x     Meas  smooth     Smoothing determines the  resolution of the curves    This is a balance depending  on equipment as well as your  individual choice     Spectrogram In the little window up to the right  the reverberation          time is shown in the spectrogram     Double click    Double click to change place of the information   The Spectrogram shows a colour spectrum of frequency   magnitude a
17. le is  need     Connect the y cable to a free line input   AUX TV  on your system           36 of 37    Microphone    Bandwidth    Filter    Measurement  method    Service  amp  Guarantee    Support    USB microphone   adjusted to flat frequency  response  16Hz 20kHz     20 20000 Hz   RTA  16 250 Hz   Room Analyzer  20 20000 Hz   Full Range    Parametric EQ with variable Q    Chirp  log sine sweep  and FFT    If you need service contact your local dealer    ALWAYS pack the product part very  You are always welcome to contact usifyou carefully  Unfortunately damages  have problems with product by e mail  during transportation are very    common  If the packaging is  SUD POPEO ALRE insufficient  the transporting    company does not compensate for  damages    Always enclose a copy of the receipt  and a description of the defect     Webpage  www xtz se    Please contact our    free of charge    support if  you need installation advice  or if any  problems occur during the installation     Measurements and curves can be hard to  interpret  Send us a mail and we will help you  to analyse the result     Contact us by e mail support xtz se and  include your phone number if you require    help  and we will ring you back     There is a Macromedia Flash instruction on  the website  http   www xtz se     370137    
18. n      Put more money into product quality  and less on advertising       Cost effective manufacturing in large volumes      Provide perfect technical solutions    Website  www  xtz se  E mail  info xtz se    4 of 37    Technical presentation    Why a room  analyzing software     Why a room  measurement  software     The fact that the room adds colouration to the sound  is known  but it is often brought out of the light  because it is hard to handle with conventional  products     Every room has unknown and unique acoustic  characteristics that affect the sound quality  The  biggest problems are in the low frequency spectrum   A normal  not acoustically treated  room has several  resonances and dips in that spectrum caused by so  called standing waves  Depending on the listening  position these produce incorrectly balanced sound        Symbolic picture     The frequency response of a good loudspeaker should  be in the range of   3dB  but a normal room can  introduce variations of   15dB  This leads us to the  conclusion that even a very good loudspeaker can  sound poor in a  bad  room  Most regular rooms  suffer from severe acoustic problems     Regardless of the cost of the system  you always have  use for this instrument  helping you to optimize your  system  Room Analyzer provides an advanced   analysis  as the name indicates  of the problems that  exist in the room system  Symbolic picture        If you spend a great deal of money on your system  it  is our experience tha
19. nd time  It could be called a 2D water fall                    The spectrum shows the reverberation time  An uneven  spectrum shows a non linear frequency response   Laggings are proof of long reverberation time in the  room           The different information can be seen  by choosing  Result  Modes   EQ  Measured                    DP Modes E0  MW Measured          Two examples   A    bad    room  with long reverberation time  substantial  room modes     File View Setup Help    SR     70    65 j   60    55 L    50   ps   46         40  im fee     36    e    30    dB 16 315 63 125 He 250    Vi  Output Level      Mi 53 49 6  24 00 Mode  m2 108 36 6 16 75 Mode    Stimulus EQ   Use   f Hz  Q   G  dB        t  ms   0 0 PR    Load  Clear Stimulus EQ Copy Room Modes Save          A good room with short reverberation time and decent  frequency response                                                              Use f Hz  Q G  dB  Type 7  Result  FE E E A  oes  E  M2 43 8 0  3 36 EQ Measured  M3 66 8 0 330 EQ  M4 24 80 494 EQ Result smooth    3ppo  v  Modes smooth    2ppo     aae  e       Meas  s   ooth   Stimulus EQ    3 ppo    Use fiH Q G  dB     t ms  50 0  gt       Store Ci    Load       Clear Stimulus EQ f Copy Room Modes Save               ee      Room Analyzer RTA SPL          Time  159 6ms Freq  95 89Hz Mag  66 1 1dB    26 of 37    Found Room Modes    Export to  XTZ 99W12 DSP    Stimulus EQ    Final adjustments  with RTA Full Range    The Found Room Modes are presented here 
20. o the USB port on your  computer    The microphone can be connected directly to the  computer or to the base if you need a longer cable   Connect the RCA cable to the RCA connector on the  microphone  or by the connectors in the base  to an  input  AUX TV IN  on the amplifier or low level input  on the subwoofer     Sound cabel  Microphone    from    Sound cabel to amplifier  and or subwoofer       If Room Analyzer is connected to the system  Front  speaker   sub  a Y cable is needed  Connect this to  AUX TV or any other input     Front speaker   sub should be connected when the  final adjustments are being made  by Frequency  response   RTA  This is needed to see how much the  front speakers add to the measurement    Restart the computer     Room Analyzer  High resolution   The microphone should be placed in the height of 70   110cm from the floor     Frequency response   RTA   When measuring the full frequency span  the  microphone should be pointed towards the speaker    The microphone is sensitive of directions over 5kHz     10 of 37    http    www xtz se    The amplifier subwoofer  should be turned off while  connecting the microphone        CD R TAPE IMD   DVD LD          Main windows    FR Untitled   XTZ Room Analyzer Pro  Fie View Setup Help    ee as                v 5  Found Room Modes Died Level  Use fH  Q GIB  Twe    Rent    Result smooth   6 ppo v    Stimulus EQ Spo v  Us fH  Q G  8     O veday        Bi    Ee    gt     Clea Stmk EQ    CopyRoomModes   U           
21. of the reference level  for each frequency     Change the scale with the dropdown menus to the right   Perfect to use as a constant frequency response meter    when trying to find a good placement for a subwoofer or  speakers  adjustment of GEQ  Graphic EQ     Export   There are two ways of exporting the RTA values  to XTZ Export    99W12 DSP   either by the    EXPORT    button  which is Si  accessible after 7 8 seconds  or by the quick links in the  menu        The export function can be used in three ways  File View Setup Help  ROOM   The PEQ settings are exported  WHainaa  RTA   The GEQ settings are exported             BOTH   Both RTA and ROOM are exported  a   OBS     When doing room analyzing  the RTA measurements  should be done after the room analyzing     20 of 37    Start the Click on measure to start           measurement  1 Position measuring    Unclick    Average 3 positions       Click on    Measure    and the procedure starts   Room Mode Measurement A  Instructions         Move the microphone to the  listening position as shawn in the  icture   Left Speaker cabanar Right Speaker Then press Measure        Mic  es  Cancel    Room Analyzer After a short while of calculations  the result is    presented     In the main window  the frequency response is now  shown    To the right in the smaller window  the reverberation  time is shown     A double click in the main window and the information  in the windows changes     There are three different curves   Choose by cl
22. om effects     phase errors and reflexes are 50d Me   clearly visible   Raw     Unfiltered data   300 ms  Full FFT  L         Bro      Allinformation is displayed without any filtering  Every  little detail  reflex and phase cancellation is visible  This  measurement has the highest resolution     Waterfall diagram  CSD    300ms  slice  total l  ngd p   200ms  sliding CSD     There are three different 3Dviews available  Right   Front  amp  Left   and a 2D view    In the 3D chard  the X axel is showing frequency  The Y   axel is showing the magnitude and the Z axel is showing  the time     In the 2D Chard  X axel is showing frequency and Y axel  is Showing the time  The second X axel  to the right  is  showing the magnitude                                      30 of 37    Overlays    Volume settings          Overlay   1   Show 1    2   Show 2    3   Show 3    4   Show 4   Full Range measurement offers the possibility of having  4 curves stored on the screen  Storing a curve is easily  done by clicking on 1 2 3  or 4 in the overlay area   It is only the last measured curve that can be stored   Ifyou want 4 measurements stored and then make a  5 th measurement  Press Ctrl N  to prepair for a new  measurement  without losing the 4 th curve   Ifyou want to store a curve on a taken overlay  just press  the number and replace it   To show or hide a measurement  click    Show 1  Show 2   Show 3 or Show 4    to show or hide it   Oyerlay   1   Show   E   3   Show 3   E   Picture showin
23. oom Analyzer    Modes EQ    RTA    Full Range    EQ PEQ    A short reverberation time is always better  In  the midrange and tweeter this can be obtained  by adding curtains and carpets in the room     Room Analyzer shows the frequency response   With the highest resolution it is   Blue curve  with high resolution impossible to get an absolutely flat  frequency response  Therefore you  can adjust the resolution in different  smoothing steps  This represents  more than the ears resolution  The  hearing resolution is personal     It shows the exact response down to each  frequency  That is why all defects are revealed  with this resolution  The resolution can be set  to different levels  Smoothing     Room Analyzer shows the Room modes  green  curve   Room Modes are caused by the room  and the ideal is trying to remove all modes     The calculated value is shown in the white  curve  If 99 W12DSP is used  the response will  be as the white curve     MODE  Room Modes  Is the first priority for Room  Analyzer   EQ  A peak dip that needs treatment     MODE  Is a Room Mode that Room Analyzer  has found  This is the first priority for the  program  If there are less than 5 found modes   Room Analyzer will analyze the possibility to  use EQ to fix the curve     EQ  A peak dip that needs treatment    In second priority for Room Analyzer     The RTA shows the frequency response in  Real time     The RTA shows the energy with 1 3 octave  resolutions  not as high a resolution as    Room  
24. ponse from the  measurement    Time window options are Anechoic  Ambient  amp   Raw    Water fall can be shown in 3D and 2D   Output level   Adjusted with level control    The signal level shows to the right of the control     The level should be on green   Red  too high  Blue   Too low    Measure Start the measurements    By clicking on the Measure button  you start the  measuring procedure     Measuring window     Room Analyzer        x     Room Mode Measurement    Instructions  Move the microphone to the  listening position as shown in the    picture   Left Speaker Right Speaker Then press Measure       Average 3 positions    Es    Subwoofer    Cancel          1 position measuring     With    Average 3 positions    unmarked  it is a one  position measurement  in the sweet spot or specific    13 of 37          Room Analyzer    ATA            Full Range          L  Output Level          10       place     Room Mode Measurement    Instructions  Place the microphone to the left  in the room as shown in the    Left Speaker    i picture   Right Speaker Then press Measure     Mic    Subwoofer    Iv Average 3 positions    Measure    Cancel     Symbolic picture of the 3 measuring positions        a positions measuring     Mark    Average 3 positions     it gives an average  value calculated and measured on three positions     Outpul Level        Found Hoon M odas   Use   f Hz  Q G  dB  Type    V1 54 10 0  20 00 EQ  M Mades EQ  Iv 2 104 10 0  10 00 EQ Iv Measured       Room Modes   
25. r the sound  energy to decay 60 dB after the excitation has stopped   To get the reverberation time  RT   we need to measure  the energy decay curve after the sound source is  switched off    The energy decay curve is a irregular and noisy curve  that usually approximate with linear decay     To get a RT60 measurement  the excitation needs to be  at least 60db above the noise level  Normal noise level in  a room is  40dB  To get a reliable T6o value  there  needs to be a sound level of at least 100dB     r  dB       40       G0        RT 60       Show   Scale range   EEE  Cia        The RT60 is shown according to the ISO3382 standard   Reverberation time is normally measured between  125Hz  amp  4000Hz  Reverberation below 125Hz  is  normally called room modes  At lower frequencies the  result is unsure and for frequencies above 4000Hz  the  air absorption is essential to the result of the  measurement     In the bass area Room Analyzer tab is better to use  This  measurement has a higher resolution    The trends of the frequency above 4000Hz are similar to  the result of 4000Hz     33 of 37    Room Modes Room modes appear due to sound reflexes in  the room   Direct sound and reflexes makes a pattern in  the room and when same frequency is in  phase  you get room modes   Standing waves     Dips If the frequency is in counter phase it  produces dips in the frequency response    nodes     If extreme dips are found by Room Analyzer   1 position measurement   it may be  impossibl
26. s manufactured to match our  requirements and the software has built in  corrections for the deviations in frequency response     The software is specially designed for XTZ by a well  established audio software company    The software can be downloaded from our website   Size   1 Mb    6 of 37    Minimum system    P3   1 GHz    1024x768 resolution  256  colours    Windows XP  512MB ram     http    www xtz se    If no file download is  possible  we will provide you  with a CD     Unpacking    Accessories    Installation    Check that all items are included   1pcs Microphone med USB contact   1pcs Base microphone stand with USB contact  1pcs RCA cable  5m    1pcs Male male adapter  RCA    1pcs RCA Y cable   User manual    Download the software from XTZ website and follow http   www xtz se  the instruction of the installation    7 of 37    In which room do  you achieve the best  sound     Reflections    Amplification of bass  frequencies    Furniture    Room dimension    Placements of  speakers    Front speakers    Angling     This chapter contains general information about  loudspeaker placement and installation     These are general rules  so there may be exceptions     Regardless how good the equipment is  in the wrong  listening environment it will inevitably sound poor   There are some basic rules concerning proper  loudspeaker installation     Carpets  curtains and soft furniture absorb mid range  and high frequency sounds  and this is preferable   Big empty areas  on the 
27. secure measure  to find the room modes easier    Weight 30   40   30   on the measuring     3 positions measuring can be used for   e Room Analysis  e Reverberation analysis    Frequency response   Full Range   A measurement 20 20000Hz which gives the  complete frequency range  This is done with  sweep instead of noise  as in RTA    measurement     Time windows options are Anechoic  Ambient  and Raw     RT6o information of the reverberation time  is  given acc  to ISO3382    19 of 37    Room Mode Measurement    Left Speaker SOBRE    _   O Mic    Right Speaker  Mic       RTA          Ha  Room Analyzer RTA    A    P   A emma       RTA measurement is a full  range measurement  At  frequencies over 5kHz  the  microphone is more sensitive  for directions and should be  aimed towards the source           Click on the RTA tab   Click on    Measure    to start the measuring process and  stop it by clicking on    Paused        The RTA displays the frequency response in real time  and shows measured levels with 1 3 octave resolution   RTA has not got as high a resolution as    Room Analyzer      but it gives a good opinion about the hearing sound in  average  perfect to use as a constant frequency response  measuring unit  when adjusting the position of speakers  or adjusting a graphic equalizer     There is a listener mode  Click on    Listen    to activate it   This mode is not suitable for adjustments     Each bar shows the level on a fixed frequency  Below the  bars is information 
28. t it is a    must    to optimize the  room or    remove    the problem with DSP technology   Ex 99W12 DSP   For this kind of work  Room  Analyzer is perfect     A comparison with cooking  Using a timer and a  thermometer helps obtaining the same perfect result  every time  Room Analyzer gives a picture of which  frequencies in the bass region cause the problems and  hints of how to correct them  Room Analyzer can  measure the frequency response of any speaker     Room Analyzer provides more analysis   1  Frequency response with Room Analyzer    16 250 Hz    a  Find room modes   b  Reverberation Time diagram   Time Frequency Level    c  Simulates the use of PEQ   Parametric Equalizer     2  Frequency response with RTA  Real Time  Analyzer 20 20000 Hz     3  Full range measurement  a  Full range measurement  FFT   20 20000 Hz  b  Waterfall 3D 2D  c  RT 60 measurement    5 of 37    Computer System  demands    Microphone    Software    The analysis render the following areas of  use    e Give exact values for compensation for the room  modes  to a subwoofer with PEQ   ex XTZ 99 W12  DSP    e While adjusting placements of  speakers subwoofers   e While optimizing the room with absorbents   diffusers  sound traps and more   e Measuring speakers subwoofers    The system is based on PC technology     The system demands are Win XP   Room Analyzer works on Vista  but is recommended  to run on XP     Room Analyzer can be run on Mac  Under Bootcamp  or Fusion with XP     The microphone i
29. uency response at the  specific time  0 250 ms  is now shown in red     This explains why the frequency response may seem  unsatisfactory even after being adjusted with PEQ and  GEO at the time zero  The other way around  a perfectly  straight frequency response may not be ideal  since noise  and defects appearing later than time zero affect the  reproduction        The experienced frequency response equals the sum of  the energy over time at each frequency  as displayed by  the RTA     28 of 37    Full Range    Working areas    The Full Range tab contains a high resolution full  frequency measurement option  You can choose how the  data will be presented        l   E   Room Analyzer RTA Full Range       The tab also displays a 2D 3D waterfall  where the  measurement can be seen as function of time   O 200ms      Lasar    File View Setup Help    Wie ea mm             _  Room Analyzer RTA SPL       The Full Range tab has a split user interface    By clicking on the grey area between the two  measurements and drag  you can alter the size of the  working areas                          File View Setup Help    Woe sm                Room Analyzer RTA SPL       20 0f 37    Views    There are three different graphs that can be shown  The  different between the graphs are the time window  settings  Default view is Anecoic    Anechoic    Non reverb   300 4 ms  Multi Win   Displays the results with limited room effects    Ambient     Room information   300 8 ms  Ambi Win     Displays ro
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Samsung SC4170 دليل المستخدم  取扱説明書 - DXアンテナ  LWB's Drupal User Manual - Wek'eezhii Land and Water Board  User's Manual - BRILUX LED lighting  KM3050-4050-5050 service manual  Tarjeta PCI Express Dual Combo Serie-Paralelo  I<enmore  Conseil économique et social Distr. général  KOHLER K-1188-RE-0 Installation Guide      Copyright © All rights reserved. 
   Failed to retrieve file