Skytron Elite 3100 Surgical Table
Contents
1. TABLE DOWN Figure 2 7 Elevation Function f Slide Function To move the table top toward the foot end of the table push the SLIDE FOOT button See figure 2 8 SLIDE A FOOT END Figure 2 8 Slide Foot Function The table top will move longitudinally up to 23 1 2 inches To return the table top to normal position push the SLIDE HEAD button See figure 2 9 HEAD END Figure 2 9 Slide Head Function g Kidney Bridge Model 3100 To raise the kidney bridge push the KIDNEY UP button See figure 2 10 KIDNEY Figure 2 10 Kidney Bridge Up Function 6 To lower the kidney bridge push the KIDNEY DOWN button h Return Function Pushing the RETURN button will return the table top to a level position from lateral tilt and or trendelenburg positions See figure 2 11 RETURN LU Figure 2 11 Return Function 2 3 Leg Section For perineal access the leg section can be easily removed from the table To remove the section rotate both locking handles outward and pull the leg section straight out Reverse the procedure to install the section See figure 2 12 LOCK HANDLE 2 Figure 2 12 Leg Section Removal 2 4 X ray Cassette Tray X ray cassettes can be inserted into a positionable tray at the head end of the table See figure 2 13 The cassette c
2. 3 Pressure check with a gauge the pres sure relief valve 4 Check all mechanical adjustments and adjust as necessary 5 Check hydraulic fluid level 6 Lubricate the slider assembly 3 2 Service Table maintenance can be performed by trained maintenance personnel using SKYTRON autho rized replacement parts and service techniques Service instructions and parts are available from SKYTRON To obtain service instructions replacement parts factory service or preventive maintenance con tracts contact your nearest SKYTRON repre sentative or write SKYTRON 5000 36th Street S E Grand Rapids MI 49512 1 800 SKYTRON 10 FEN A A A ANI N 5000 36th Street S E Grand Rapids MI 49512 1 800 SKYTRON or 1 616 957 0500 FAX 1 616 957 5053 az LEJA J mip Yan T Wii Y F ier BINI
3. ELITE SERIES SURGICAL TABLES OPERATORS MANUAL MODEL ELITE 3100 amp 3100A MODEL 3100 SKYTRON STANDARD LIMITED WARRANTY SKYTRON a Division of the KMW Group Inc SKYTRON warrants all new products sold by it directly or through a dealer or other authorized representative with exception to replacement parts spares bulbs surgical lights pads and accessory items surgical tables to be free from defects in material or workmanship under normal use and service for a period of two 2 years This warranty shall include the cost of repair or replacement of defective parts including the cost of service labor and travel time to the site of equipment use Delays caused by the user in accessing the equipment for repair will be chargeable at the normal hourly rate for service by SKYTRON s authorized service representative The warranty period shall begin with the initial operation or one 1 year after receipt of the product whichever shall occur first Replacement parts spares bulbs surgical lights pads and accessory items surgical tables are warranted to be free from defects in material or workmanship under normal use and service for a period of ninety 90 days from receipt by the ultimate user with exception to replacement parts supplied by SKYTRON for products under warranty which shall be covered for any remaining period of the original produ
4. TABLE DOWN SLIDE FOOT END II SLIDE HEAD END KIDNEY BRIDGE H tt KIDNEY BRIDGE DOWN RETURN r BRAKE RELEASE POWER ON OH REEE S SKYTRON PRESS TO LOCK BRAKES Figure 2 2 Pendant Control Unit a Floor Lock Brake System To activate the brakes without affecting table positioning press the TABLE UP button See figure 2 3 The eleva tion cylinder will not function until the brakes are completely extended An electronic timer in the brake circuit will activate the brake system until the brakes are completely set approximately 8 10 seconds Tewo AE TABLE UP RELEASE Figure 2 3 Brake System Activation Press the BRAKE UNLOCK button on the pendant control to release the four self leveling brake feet in orderto move the table See figure 2 3 The brake delay circuit automatically retracts the brake sys tem with just one press of the BRAKE UNLOCK button It takes approximately 7 8 seconds to totally release the system WARNING DO NOT unlock brakes when a patient is on the table An uneven patient weight load may cause instability b Emergency Brake Release In case of a power failure or an electrical problem within the table an emergency brake release system can be used to move the table The control lever for this function is located on the side ofthe table base and is identified by an EMERGENCY BRAKE RE LEASE label Turn the lever counterclockwise to release the brakes
5. AND THOSE OF THE OPERATING ROOM PERSONNEL ARE CLEAR OF ALL MOV ING PARTS BEFORE MOVING THE TABLE Proper restraints should always be used for patient safety Certain accessories can be damaged when chang ing the position of the table top sections Always look first to see if a desired movement is going to interfere with any accessories in use The operator has the ultimate responsibility of preventing damage or possible injury In general common sense will dictate when there is a poten tial hazard The following precautions should be reviewed by all personnel prior to operating the table WARNING DO NOT use the table in the presence of FLAMABLE GASES NOTE Activating any function button will acti vate the brakesystem Usingthe TABLE UP function to set the brakes provides a visual assurance that the brakes are locked As the brake cylinders are extending the entire table will move slightly When the table top begins to elevate the brakes are fully locked WARNING DO NOT unlock brakes when a patient is on the table An uneven patient weight load may cause instability NOTE In case of a power failure or an electri cal problem within the table the emer gency brake release system can be used to move the table The control lever for this function is located on the side of the table base and is identified byan EMERGENCY BRAKE RELEASE label Turn the lever counterclockwise to release the brakes NOTE The Emergency
6. Brake Release Valve must be closed before any hydraulic function will operate IMPORTANT The Emergency Brake Release Valve must be tightened securely when notin use lf the Emergency Brake Release Valve has been operated the BRAKE UN LOCK button on the pendant control will have to be pressed before brakes will lock again NOTE The table will lift and articulate a patient weight of 500 pounds assuming normal positioning of the patient SECTION I INTRODUCTION LEG SECTION X RAY CASSETTE TRAVEL INDICATOR CASSETTE TRAY POSITIONING GUIDE KIDNEY BRIDGE X RAY CASSETTE POSITIONING CRANK PENDANT CONTROL ELECTRICAL ENCLOSURE Figure 1 1 Model 3100 Surgical Table 1 1 General SKYTRON s Model 3100 and 3100A surgical tables are electro hydraulically operated surgical tables with special x ray equipment compatibility features See figure 1 1 A longitudinal top slide capability of 23 1 2 inches allows clearance for C arm coverage of 58 3 4 inches without the need to reposition the patient X ray cassettes can be loaded into a movable tray from the head end of the table A manual hand crank system allows cassette positioning through out an unobstructed range of 65 inches A travel indicator next to the hand crank indicates where the center of the cassette tray is located in relation to the positioning guide on the side of the table for precise x ray positioning The
7. See figure 2 4 Figure 2 4 Emergency Brake Release NOTE The Emergency Brake Release Valve must be closed before any hydraulic function will operate IMPORTANT The Emergency Brake Release Valve must be tightened securely when notin use lf the Emergency Brake Release Valve has been operated the BRAKE UN LOCK button on the pendant control will have to be pressed before brakes will lock again c Trendelenburg To place the surgical table inatrendelenburg head down position press the TREND button See figure 2 5 Tilt of up to 20 may be obtained To place the table in a reverse trendelenburg head up position press the REV TREND button Tilt of up to 20 may be obtained REVERSE TRENDELENBURG TRENDELENBURG ae Figure 2 5 Trendelenburg Positioning d Lateral Tilt To achieve lateral tilt right as viewed from the head end of the table press the TILT RIGHT button figure 2 6 Tilt of up to 20 may be obtained To achieve lateral tilt left press the TILT LEFT button Tilt of up to 20 may be obtained LATERAL T E LATERAL TILT LEFT Figure 2 6 Lateral Tilt Positioning e Elevation To raise table top press the TABLE UP button figure 2 7 The table will lift a patient weight of 500 pounds up to a maximum height of 40 To lower the table top press the TABLE DOWN button The table top will go down to a minimum height of 29
8. full radiographic table top allows clear sharp x rays throughout the cassette travel range The hydraulically operated built in kidney bridge on the Model 3100 is made up of movable table top sections This allows total unobstructed radio graphic capability with the kidney bridge in use The kidney bridge is not available on Model 3100A The push button pendant control unit operates the kidney bridge function Model 3100 as well as the trendelenburg lateral tilt top slide elevation and the floor lock brake system functions 2 Manual controls are provided for cassette tray positioning emergency brake release and leg section removal 1 2 Power Requirements The Elite 3100 series tables require a 120VAC 60 Hz electrical power supply The table is equipped with a 15 foot long power cord with a standard three prong hospital grade plug The electrical protection fuses are located on the front edge of the base See figure 1 2 The main power ON OFF switch is located on the electrical enclosure ELECTRICAL ENCLOSURE POWER CORD MAIN POWER SWITCH Figure 1 2 Power Switch and Fuse Location 1 3 Pendant Control Unit The hand held pendant control unit figure 1 3 has a non slip rubber cover which assures a positive grip during use A spring clip hanger located on the back of the control allows it to be stored on the table side rails CLIP FUNCTION BUTTONS POWER INDICATOR Figure
9. patient weight load may cause instability NOTE In case of a power failure or an electri cal problem within the table the emer gency brake release system can be used to move the table The control lever for this function is located on the side of the table base and is identified byan EMERGENCY BRAKE RELEASE label Turn the lever counterclockwise to release the brakes NOTE The Emergency Brake Release Valve must be closed before any hydraulic function will operate IMPORTANT The Emergency Brake Release Valve must be tightened securely when notin use lf the Emergency Brake Release Valve has been operated the BRAKE UN LOCK button on the pendant control will have to be pressed before brakes will lock again SECTION II OPERATION 2 1 Electrical Power a Check to be sure power cord is plugged into 120VAC outlet IMPORTANT Prior to operating the table review the SPECIAL USER ATTENTION section on page 1 b Depress Main Power ON OFF switch on the electrical enclosure See figure 2 1 The green Power On indicator light on the pendant control unit should now be illuminated POWER CORD Dl gt N MAIN POWER SWITCH ELECTRICAL ENCLOSURE Figure 2 1 Main Power Switch 2 2 Pendant Control Unit The hand held pendant control unit figure 2 2 activates the following table functions TRENDELENBURG TRENDELENBURG LATERAL TILT LEFT LATERAL TILT RIGHT TABLE UP I 1i
10. 1 3 Pendant Control Unit The function push buttons are identified with ab breviated descriptions for all functions See figure 1 4 The trendelenburg and table up buttons are red the remaining buttons are all black REVERS TRENDELENBURG meo TRENDELENBURG LATERAL TILT LEFT LATERAL TILT RIGHT TABLE UP I I TABLE DOWN SLIDE FOOT END Il SLIDE HEAD END KONEV BRIDGE UP hi H KIDNEY an DOWN RETURN 7 BRAKE RELEASE POWER ON O IT inpicaror RON PRESS TO LOCK BRAKES Figure 1 4 Function Push Buttons 1 4 Floor Lock Brake System The floor lock brake system consists of four self leveling hydraulic brake cylinders which raise and support the table base off from the casters Press the TABLE UP button on the pendant control to set the table s brakes An electronic timer in the brake circuit will activate the brake system until the brakes are completely set approximately 8 10 seconds NOTE Activating any function button will acti vate the brakesystem Usingthe TABLE UP function to set the brakes provides a visual assurance that the brakes are locked As the brake cylinders are extending the entire table will move slightly When the table top begins to elevate the brakes are fully locked To unlock the brakes press the BRAKE UNLOCK button and release The brakes will retract auto matically in approximately 7 8 seconds WARNING DO NOT unlock brakes when a patient is on the table An uneven
11. ECIAL USER ATTENTION ciiise srania an ae aara a a a a aai a a a a 1 SECTION INTRODUCTION 00 sinnini ananiita aeni ad aaa dadai Aa A aaien 2 ti Genar A an ee E E een 2 1 2 POWer Requirements sau see einen 2 1 3 Pendant Control Unit s ssssrisssrsurssnnanuandnnnndunaaN ANKAN ANANA AANEEN AANKAN ANANA KANAN NAN KANAANAN nenn nenn nennen nenn 3 1 4 Floor Lock Brake System ea ee 3 SECTION II OPERATION uuu ccccccccececccceseeeecessueeeeeeseeaeuusessueeuseseueeueueeuseuaueeseuaeeseseueeaseeeasaaseeenea 4 21 Electical POWER untada ara Een Taerar ee lern 4 2 2 Pendant Control Un 4 2 3 Leg SECON ci sch daedoendeeadsdeveeeauseederetedoxeseueecdecauiers 6 2 4 Xray Cassette AAA o Pe 7 SECTION III MAINTENANCE 00 ccccccecccccsesesecseeeeseueesceeeseccseeneeeeseeeeeeeeeceaesseaseaecaeeeeaueneeecsoesereres 8 3 1 Preventive Maintenance mmm nennen nenn ernennen nenn rnnnn nenn nennen ren 8 3 2 2 AV 01 rd aten 8 SPECIAL USER INFORMATION The extreme positioning capabilities of the 3100 series tables requires special attention for possible interference points when using multiple function positioning As with the operation of any surgical table acertain amount of care should be exercised to position the patient safely Although the thick pads and sheets substantially protect the patient pinch points located at the joints of the top section should always be considered BE SURE THAT THE ARMS HANDS AND FINGERS OF THE PATIENT
12. an be positioned for x ray coverage of up to 65 inches from the head end of the table The removable tray attaches to the drive mecha nism with magnets To position the cassette tray within the table rotate the manual crank handle located on the side of the table A positioning guide onthe sides of the table corresponds with the travel indicator for precise cassette positioning Use the positioning guide if the guide is not visible due to table draping a measuring tape can be used to determine the desired location for the cassette Insert the cassette into the tray and rotate the manual crank until the desired position is achieved on the travel indicator CASSETTE TRAY TRAVEL INDICATOR NS MANUAL POSITIONING GUIDE CRANK HEAD END OFTABLE Figure 2 13 X ray Cassette Tray SECTION Ill MAINTENANCE 3 1 Preventive Maintenance The following preventive maintenance checks and services are recommended to ensure the service ability and proper operation of your SKYTRON surgical table a During normal cleaning a general visual ex amination should be made checking for leaks loose bolts or parts and cracked chipped or missing paint Any necessary repairs should be made b Semi annually the following checks and ser vices should be performed 1 Check all hydraulic fittings mini valves and slave cylinders for proper operation and any signs of leaks 2 Check the hydraulic speed controls and adjust if necessary
13. ct warranty or for 90 days whichever is of greater benefit to the ultimate user SKYTRON s responsibility and liability shall be limited to the repair or replacement of any part which we SKYTRON determine to be defective within the applicable warranty period Minor adjustments required as a result of normal wear during the use of the product within the warranty period are not covered under warranty The labor portion of this warranty is covered by SKYTRON s Authorized Service Agent Repairs made by others are not authorized nor covered by SKYTRON with respect to labor costs SKYTRON shall not be liable for any other expense loss or damage whether direct incidental consequential or exemplary arising in connection with the sale or use of or the inability to use SKYTRON products NO EXPRESS WARRANTY IS GIVEN BY SKYTRON WITH RESPECT TO ITS PRODUCTS EXCEPT AS SPECIFICALLY SET FORTH HEREIN ANY WARRANTY IMPLIED BY LAW INCLUDING ANY WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE IS EXPRESSLY LIMITED TO THE TWO YEAR AND 90 DAY TERMS SET FORTH ABOVE THE FOREGOING STATEMENTS OF WARRANTY ARE EXCLUSIVE AND IN LIEU OF ALL OTHER REMEDIES No dealer agent employee or other representative of SKYTRON is authorized to extend or enlarge this warranty 4 00 Although current at the time of publication SKYTRON s policy of continuous development makes this manual subject to change without notice TABLE OF CONTENTS Title Page SP