Home
        2-wire Type pH/ORP(Redox)
         Contents
1.                                                Note  Display shows Return to maintenance    the value as at the time mene  of taking the sample     Set the value   using the  gt      ENT key    Select the flashing digit with the  gt  key   Increase its value by pressing         key   When the correct value is displayed   press ENT to enter the change                    YES                         IM 12B07D02 01E    Maintenance 7 1    7                             7 1  Periodic maintenance for the EXA transmitter   The transmitter requires very little periodic maintenance  The housing is sealed to IP65  NEMA 4X  standards  and remains closed in normal operation  Users are required only to make sure the front win   dow is kept clean in order to permit a clear view of the display and allow proper operation of the push   buttons  If the window becomes soiled  clean it using a soft damp cloth or soft tissue  To deal with more  stubborn stains  a neutral detergent may be used     NOTE   Never used harsh chemicals or solvents  In the event that the window becomes heavily stained or  scratched  refer to the Customer Maintenance Parts Lists for replacement part numbers     When you must open the front cover and or glands  make sure that the seals are clean and correctly fit   ted when the unit is reassembled in order to maintain the housing   s weatherproof integrity against water  and water vapor  The pH measurement uses high impedance sensors and may otherwise be prone to  problems 
2.                                    YES n YES                            Return to     measuring mode     measuring mode           IM 12B07D02 01E    6 6 Calibration    6 5 4  Sample calibration                          X lt APHSG 2  YOKOGAWA                Press the MODE key  The legend AUT CAL appears  and the YES NO  key prompt flags flash  Press NO     The display MAN CAL appears   Press NO SAMPLE appears  Press YES to start calibration                        HIITI   11               FILI EOI       TE    n         L    MINIT CII                 TIL          s                    NO                                                 Press YES at the same time as taking  sample for analysis                       PH 202 now continues to measure control   as before  SAMPLE flashes to indicate that    data is stored waiting for input of analyzed  value           IM 12B07D02 01E    Calibration 6 7                               When the laboratory analysis is completed  the data is entered by first pressing MODE   then following the sequence below                                                                                           Return to maintenance  menu                               For first calibration                         of a new sensor                                                    To calibrate                      ORP or rH                               Note  Display shows  the value as at the time  of taking the sample                          NO or  YES      
3.                        7 2   8  Troubleshooting as             eee ee ee 8 1       itt ee                          vide 8 2  8 1 1  Off line calibration checks e crt eterne nete          cua v ee Le Ra es 8 2  8 1 2  On line impedance                                                8 2   9  Spare Parts                                                                               9 1         12  07002 01      10  Appendix                             10 1    10 1  User setting ach ae      eat ase dete aaa               10 1  10 2  Configuration checklist for     202                                                               10 3  10 3  Setup for sensor                                                                   10 4  10 3 1                 EE 10 4  10 3 2  Selection of measurement and reference electrode                 10 4  10 3 3  Selecting a temperature                               10 4  10 4  Set up for other                                                       10 5  10 5  Set up for Pfaudler Type 18 sensor                                             10 6  10 5 1  General Set Up  rrr rni pi i ea etn deor        10 6  10 5 2  Calibration set                     4                      10 6  10 6  Device Description  DD  menu                                                                         10 7  Glossary  Dum iuc  bd qc                                                                                 11 1          1212125 M                 11 1  Tl 2  Wining di
4.                        Tov xperaleote o  nyiss oxeruk        Ex         TOTLK   YAWOOA TAPAKAAOVLE ETLKOWOV  OTE LE TO  TANTLETTEPO ypageto        Yokogawa      AVTLTPOTWTO TNS     IM 12B07D02 01E    CO    V amp etky n  vody      obsluhu pre pr  stroje s ATEX Ex  su    dispoz  cii v jazyku anglickom  nemeckom a  franc  zskom  V pr  pade potreby n  vodu pre Ex   pr  stroje vo Va  om n  rodnom jazyku  skontaktujte  pros  m miestnu kancel  riu firmy Yokogawa              V  echny u  ivatelsk   p    ru  ky pro v  robky  na   n     se vztahuje nev  bu  n   schv  len   ATEX Ex    jsou dostupn   v angli  tin    n  m  in   a francouz  tin     Po  adujete li pokyny t  kaj  c   se v  robk   s  nev  bu  n  m schv  len  m ve va  em lok  ln  m jazyku   kontaktujte pros  m va  i nejbli       reprezenta  n    kancel     Yokogawa     QD    Visos gamini   ATEX Fx kategorijos Eksploatavimo  instrukcijos teikiami angl    vokie  i   ir pranc  z    kalbomis  Nor  dami gauti prietais   Ex  dokumentacij   kitomis kalbomis susisiekite su  artimiausiu bendrov  s    Yokogawa    biuru arba  atstovu     Cw     Visas ATEX Ex kategorijas izstr  d  jumu  Lieto  anas instrukcijas tiek pieg  d  tas angiu  vacu  un fran  u valod  s  Ja     1                    Ex ieri  u  dokument  ciju cit   valoda  Jums ir j  sazin  s ar  firmas Jokogava  Yokogawa  tuvako ofisu vai  parstavi                K  ik ATEX Ex toodete kasutamisjuhendid on  esitatud inglise  saksa ja prantsuse keeles  Ex  seadmete muukeelse dokumen
5.                       Back to  the Top                                                                               IM 12B07D02 01E    4 6 Operation    4 5 3  Display functions pH  rH           Service Code 01 Set for pH        on parameter 2           Temp Display                                                                         6                rH Display                                                                         Service Code 02 Set to rH          p See Man Cal  Chapter 6  Current  Output                                         6                     Pot  pH Sensor                                           Slope  pH Sensor                                              See Man  Imp   check Chapter 5                                                                                                                As Pot ORP       See Temp Menu  Chapter 5  Impedance  Input 1  mu  29  Hold Menu  Chapter 5  Impedance  Input 2  Software  Release  Version                                  IM 12B07D02 01E    Parameter setting 5 1    5  PARAMETER SETTING       5 1  Maintenance mode  Standard operation of the EXA instrument involves use of the maintenance  or operating  mode to set up  some of the parameters     Access to the maintenance mode is available via the six keys that can be pressed through the flexible  window in the instrument cover  Press the MODE key once to enter this dialog mode   Note   At this stage the user will be prompted for pass code where this has 
6.               2 12  2 10  Control Drawing of PH202S FF PB Specification                                      2 13  2 11  Control Drawing of     2025 FF PB Specification                                                  vB ts    a p i                   D 2 14  2 12  Control Drawing of PH202S FF PB Specification                                                  FM Intrinsically safe 2                                           2 16  2 13  Control Drawing of PH202S FF PB Specification                                                       Non incendive Entity                                                                                2 18  2 14  Control Drawing of PH202S FF PB Specification                                                  FM Non ncendive FNICO  2 2                        2 19  2 15  Control Drawing of PH202S FF PB Specification  CSA                               2 20  3  Installation And                                 4  444  2  4   10001111  3 1  3 1  Installation and dimensions                                           3 1  3 1 1   Installation  Site    2                        3 1  3 1 2  Mounting         0                                                      cod rb a                 3 1  3 2  PIBDAIIOE                3 3  3 2 1  Cables  terminals and                                     3 3  3 3           Of SENSO                      3 4  3 3 1  General                                       3 4  3 3 2  Additional precautions for installations in hazardous ar
7.               Screen       Core Screen    4                 a  e  zt p er 21   9 017      leloj    210  8          2 22 222 22    arp p pr pa pape  15                Brown Co axial Cable                                                                      Co axial cable    EXA pH TRANSMITTER   CONVERTER    WF10 Cable          12  blue     Co axial cable   brown     11  red        Fig  3 11  Connection of WF10 extension cable and BA10 BP10 junction box    NOTE  See page 3 12 for termination for WF10 cable    3 6 4  Connection VP type sensor    in combination with EXA pH         gt  Connections normal pH    PD fF  lt a gt     pH  ref             LE       temp       A 15     13    0 14     11     12  5 3      63              gt Connections differential                                     LE       temp    A 15  B 14  C 13  D 17  E 11  F 12  S 3 or 63          IM 12B07D02 01E       3 12 Installation and wiring    Extension cable may be purchased in bulk quantities  cut to length  Then it is necessary to terminate the  cable as shown below    Termination procedure for WF10 cable    1  Slide 3 cm of heat shrink tube  9 x 1 5  over the cable end to be terminated    2  Strip 9 cm of the outer  black  insulating material  taking care not to cut or damage internal cores                    4  3        9      gt        heat shrink remove insulation          Fig  3 12a     3  Remove loose copper screening  and cut off the cotton packing threads as short as possible   4  Strip insulation fr
8.           IM12B07D02 01E    Parameter setting 5 9    HOLD value set     return to commissioning  menu                                                                             YES Set HOLD  fixed value              for mA output                IM12B07D02 01E              5 10 Parameter setting    5 2 3  Service                         PHOS             YOKOGAWA             Example  Service Code 01  Select main parameter    ri  LI for pH    i for ORP  With the  gt     ENT keys                         Wait screen is displayed    commissioning menu     briefly before returning to                   XPH  RP                                                                                                                                              XPHORP          IM12B07D02 01E       Parameter setting 5 11    5 3  Notes for guidance in the use of service coded settings   Don t set or input service code numbers other than the code numbers defined in this manual  Setting an  undefined service code may make the transmitter malfunction    When an undefined service code is input by some accident  push the MODE key and escape from the  service level     5 3 1  Parameter specific functions    Code 01    Code 02     pH ORP     PRM 2    Code 03  amp  04  Z1 CHK    Code 05     amp   Z2 CHK     CAL CK    Choose the main measuring parameter  The option of the ORP input is used  with an inert metal electrode as measuring sensor which gives a reading  directly in millivolts  This signal can then
9.          11 14  11 6 3  Supplement of temperature sensor                       11 15  11 6 4  Auto Return           etri                                                      11 15  Uu c CALEND          EE 11 15  11 6 6  How to cancel sample                                        sss 11 15  12  Appendix 3 QUALITY                                                                      11  12 1  12 1  PH202G  PH202SJ 2 Wire pH ORP                                                              12 1  12 2  P2023 2 M re pH ORP Transmitter                                     eg bebe dedos 12 5  12 3  PH202G      2025 2 Wire pH ORP Transmitter          2 ciento tins                              etc D   12 9  12 4  PH202G      2025 2 Wire pH ORP Transmitter                                                     Probus Om ANON            12 13  Customer Maintenance Parts List  PH202G Style   S3           CMPL 12B07D02 03E  Customer Maintenance Parts List  PH202S Style   S3           CMPL 12B07D02 23E                              i    IM 12B07D02 01E         12  07002 01                                              Electric discharge    The EXA analyzer contains devices that can be  damaged by electrostatic discharge  When servic   ing this equipment  please observe proper pro   cedures to prevent such damage  Replacement  components should be shipped in conductive pack   aging  Repair work should be done at grounded  workstations using   grounded soldering irons and wrist straps to avoid  elec
10.          IM12B07D02 01E    5 4 Parameter setting    5 1 3  Manual activation of HOLD                                                              lt  NO                                                                                                 YES YES                         Note  The HOLD feature must first be activated in the commissioning mode section 5 2 2                    IM12B07D02 01E    Parameter setting 5 5    5 1 4  Manual impedance check    Note  The manual impedance start   is available when the sensor impedance  measurement is enabled in Service  Code 3 and 4    This enables the impedance data to be  updated immediately after a maintenance  event  e g replacing an electrode                                         YOKOGAWA             MODE             c3                                                                            Return to measuring  mode after updating  impedance check             IM12B07D02 01E    5 6 Parameter setting    5 2  Commissioning mode  In order to obtain peak performance from the EXA  you must set it up for each custom application            Output range mA output is set as default to 0   14 pH     OUTP   For enhanced resolution in more stable measuring processes  it may be desir   able to select 5   10 pH range  for example   Service codes 31 and 35 can be used to choose output function on mA output     Hold The EXA transmitter has the ability to    hold    the output during maintenance peri     HOLD  ods  This parameter should 
11.          Nominal 50 A  O D   60 5 mm    2 inch pipe        4 eps          RRAR             m TR                e         Figure 3 4  Internal view of EXA wiring compartment    IM 12B07D02 01E    Installation and wiring 3 3    3 2  Preparation  Refer to figure 3 4  The power output connections and the sensor connections should be made in  accordance with the diagram on page 3 6  The terminals are of a plug in style for ease of mounting     To open the EXA 202 for wiring     Loosen the four frontplate screws and remove the cover   The terminal strip is now visible   Connect the power supply  Use the gland on the left for this cable   Connect the sensor input  using the gland on the right  see fig  3 5   Switch on the power   Commission the instrument as required or use the default settings   Replace the cover and secure frontplate with the four screws   Connect the grounding terminals tp protective earth   The optional hose connection is used to guide the cables coming from an immersion fitting through a  protective plastic tubing to the transmitter     Noo BON    3 2 1  Cables  terminals and glands   The PH202 is equipped with terminals suitable for the connection of finished cables in the size range   0 13 to 2 5 mm  26 to 14 AWG   The glands will form a tight seal on cables with an outside diameter in  the range of 6 to 12 mm  0 24 to 0 47 inches           Sensor  cable gland    Power Output  cable gland       terminal    Figure 3 5  Glands to be used for cabling    IM 12B07
12.          for maintenance interval time exceeded    System not maintained in Perform maintenance  preset time period Reset interval  E17   Output span too small  lt  1pH Incorrect configuration by user Reprogram  E18   Table values make no sense  E19  Programmed values outside acceptable limits  Incorrect configuration by user Reprogram  E20   All programmed data lost Fault in electronics Contact Yokogawa  Very severe interference  E21   Checksum error Software problem Contact Yokogawa  E23  Zero point outside limits Sensors are aged or polluted Check buffer solution  Mistake in calibration Recalibrate at pH7  Replace sensor             IM 12B07D02 01E    Spare parts 9 1    9  SPARE PARTS       See Customer Maintenance Parts List    IM 12B07D02 01E                                       10 1                                                                                                                10  APPENDIX  10 1  User setting table  FUNCTION SETTING DEFAULTS USER SETTINGS  Parameter specific functions  01   PH ORP   0 pH  02   PRM2 0 Off  03   Z1 CHK   1 1 1 High range  TC on  check on   04   72        0 0 1 Low range  TC off  check off no TC  05   CAL CK   1 1 AP on  Slope on  Temperature functions  10   T SENS   0 Pt1000  11    T UNIT 0   C  12    T ADJ None  13    T COMP   0 Off                  0 00 pH 10  C  Calibration functions  20    At SEC 5 Sec          0 02 pH  21    AS LOW    120 mV   AS HI 120 mV  22   SLLOW   70     SL HI 110    23    ITP 7 00 pH   SLOPE 10
13.         for 8 55kO NTC sensor max  120   C      Output signal     4 20 mA loop powered  isolated from input   maximum load 425    at 24 V DC  With the  possibility of 21 mA  FAIL  signal  burn up   and 3 6 mA  burn down when HART or  distributor comm  is non used   3 9 mA  burn  down when HART or distributor comm  is  used      E  Temperature compensation     Range     30   C to 140   C    for 8 55     sensor    10   C to 120   C    Sensor types  Pt100  Pt1000         PTC  5 1      PTC  8 55     NTC  3500 PTC   6 8     PTC  10kQ PTC   Automatic or manual compensation to Nernst   equation  Process compensation by configura    ble coefficient  Adjustable ITP  Iso thermal point  of intersection      F  Calibration    Semi automatic  using tables in transmitter   for pH 4  7  amp  9 buffer solutions  or using user   defined tables  with automatic check of meas   urement stability   Manual  using standard sample  by correcting  reading to value of standard   Calibration by slope and asymmetry potential  setting   IEC746 2     G  Logbook      Software record of important events and diag   nostic data  Available through HART   link     Specification 2 1    with key diagnostic information available in the  display     H  Serial communication    Bi directional HART   digital communication  superimposed on the 4 20 mA signal        Display    Custom liquid crystal display  with a main dis   play of 3 1 2 digits 12 5 mm high   Message display of 6 alphanumeric charac   ters  7 mm high  
14.         li ImayY 400  Applicable standard  IEC 60079 15 2001   Ui Vmax  31 5 V                100 mA  IEC 60079 0 2004               1 2 W       22 nF  Li 35 pH   FU                                                                          1         Certificate  IECEx KEM 06 0052    IECEx IECEx Intrinsically safe  see Note  Scheme Ex nA  nL        Temp  Class  T4  Amb  Temp    10 to 55      IECEx Intrinsically safe P TE  Amb          Mri d  Applicable standard  IEC60079 0        60079 11  Uiz31 5 V  Ci 22 nF Liz35 uH   Note  IEC60079 26     Eps  Certificate  IECEx KEM 06 0052X           Zone 0 Ex ia      Item Description Code  Temp  Class  2 pis T  Eu            Intrinsically safe Approval  TUS A  ELS Ci od F LEUR Applicable standard  FM3600  FM3610  FM3810  FES V                                     BITISH MEEPS Intrinsically Safe for Class    Division 1  Groups ABCD                Description Code Class    Zone 0  AEx ia IIC     Temp  Class  T4  Amb              10 to 55     or  CENELEC ATEX  KEMA  Intrinsically safe Approval Intrinsically Safe Apparatus Parameters  Applicable standard  EN60079 0  EN50020           V  Imax 250 mA   CENELEC Entit      ATEX    EN60079 26    1 Pmax 1 2 W  Ci 220 pF  Li 0 uH         Certificate           07        0048 X Facto FISCO Vmax 17 5 V  Imax 380 mA   Entity Ex ia       Group  Il  Category  1G P ry Pmax 5 32 W  Ci 220pF  Li 0 uH  Temp  Class  T4  Amb  Temp    10 to 55  C Mutual  FM               Ui 24 V     250 mA       1 2 W  Ci 220 pF  
15.        1    means  Analog output      LRV    means  Lower rangeval      URV    means  Upper rangeval             1 Process variables    gt         1  PV  2  Temp  3  PV   rnge    Level 3 menu             2 Diag Service      gt                                                         3 Basic setup                       4  More pr var      gt  1 Slope  2  Aspot  Note        PV               means        of output range  1  Status 1  Poweru  2  Hold      gt  1  Hold status p  2  Hold fnc 2  Powerdwn  3  Hold type 3  Defaults  4  Hold fix 4  Lg  Erased  5  Low range  3 Temp man       gt   1  T meas 6  High range  2  Man  temp 7  Hold on  4 Logbook      gt  1 Logbook conf           9  Hold off  2  Logbook 1 B Enron  3  Logbook 2                      5  Calibrate        gt  1  Sample Man Temp  2  Manual Cal  PV Aspot  6  Loop test Aspot 2  Slope  1 Tag Temp  coef  2 Device information 1  Date 21          2  Descriptor 22          3  Message Stab time   4  Write protect ETP   5  Manufacturer Zero pnt  6  Dev id New sensor  One p cal  Two p cal       Level 4 menu                    Note   HART protocol DD files can be downloaded by following URL   http   www yokogawa com an download an dl fieldbus 001en htm    IM 12B07D02 01E    Continued next page        Continued     Appendix 10 9       Level 1 menu    Level 2 menu    Level 3 menu Level 4 menu       4 Detailed setup           gt      5 Review                             3  Cal  specific       1  Param  specific    gt                
16.        Descriptor  Message  Write protect  Detailed Setup Param  Specific       second parameter       Impedance input1 Temp  comp   Imp  check    ON LINE MENU Imp  limits                                           lineata            Device setup Temp  comp  Primary value Imp  check  Analog output    Calibration check Imp  limits  Lower rangeval  Temp  Spec m Temp sensor  Upper rangeval  Temp unit  Temp comp   Temp coeff   Calibration                  Stability     Stable time  Stable pH  Aspot  Zeropoint    Aspot low limit  Aspot high limit                                                                                                             Slope  __  Slope value  Slope low limit  Slope high limit  ITP  Buffer Buffer 4            4           Buffer40  C    80       Buffer 7 Buffer 7 name    Buffer70         80       Output function           4 MA function Buffer 9 Buffer name  Burn function Buffer90         80   C  Table table 096       table 10096  Exa user interf  Error programming     Error 4 1     Error 16  Maintenance timer Timer on off  Reload value  Display       Display resolution  Auto return     Man  impedance  Review Manufacturer Passcode I    Maintenance  Distributor Commissioning  Model Service  Meas  type etc  etc                                      IM 12B07D02 01E    10 8 Appendix    Menu structure for HHT 375 shown below        ON LINE MENU Level 1 menu    Level 2 menu       Device setup  PV   AO1   LRV    URV    Note       PV    means  Primary value   
17.        Selection of the temperature compensation sensor  The default selection is  the Pt1000 Ohm sensor  which gives excellent precision with the two wire  connections used  The other options give the flexibility to use a very wide  range of other pH sensors     Celsius or Fahrenheit temperature scales can be selected to suit user  preference     With the process temperature sensor at a stable known temperature  the  temperature reading is adjusted in the main display to correspond  The  calibration is a zero adjustment to allow for the cable resistance  which will  obviously vary with length    The normal method is to immerse the sensor in a vessel with water in it   measure the temperature with an accurate thermometer  and adjust the reading  for agreement     Process compensation automatically allows for changes in the pH or ORP  of the process with temperature  The characteristic of each process will be  different  and the user should determine if this feature is to be activated  and  what compensation figure to choose     The compensation is given in pH per 10   C or mV per 10   C   The standard temperature is 25  C  fixed value  in setting the temperature  compensation     Example  For pure water with an alkali dose   e g  boiler feed water   a coefficient of approx  0 35pH can be expected  However   applications vary and a simple test will determine what if any  coefficient is suitable for the process     IM12B07D02 01E    5 14 Parameter setting            lt         
18.        gt                 1  Disp  res  2  Auto  Ret          3  Passcodes              1  Maintenance  2  Commissioning  3  Service             Level 5menu Level 6 menu         Change    Exit            Change    Exit            Change    Exit               Change    Exit                                                  4           0        80                         7           0 C   80 C             Buffer 9 name  0 C   80 C                IM 12B07D02 01E    10 10 Appendix         12  07002 01      Glossary    pH          H        ORP    rH    Asymmetry potential    Slope    ITP    Zero point    Glossary    This is a logarithmic function of the Hydrogen ion activity  concentration   This  provides a quick indication of the acidic or alkaline behavior of a dilute solution   Normally measured on a scale of 0 14 pH where low numerical values are acidic   0 is approximately 1 Normal acid  and high numbers are alkaline  14 is approxi   mately 1 Normal NaOH   The neutral point is pH 7    Defined by Nernst in the following equation  E   Eo   RT nF x Ln  H     E   measured potential    R   gas constant   T   absolute temperature   n   valence   F   Faraday number   Ln   Napierian logarithm    H     activity of the Hydrogen ion  Eo   Reference potential    Oxidation reduction potential is a measure of oxidizing power of a solution  The  greater the milli Volt value with a negative polarity  the greater the oxidizing  power  Reducing power is indicated by positive values of mV     T
19.      Intrinsically safe output parameters         maximum external parameters   terminal 11 and 17          14 4 V  lo 32 3 mA       0 12 W  Co 600  nF  Lo 34 mH   Note 2 Installation     Electrostatic charges on the display window shall  be avoided              external earth connection facility shall be  connected reliably              instrument modification or parts replacement  by other than authorized representative of  Yokogawa Electric Corporation and will void  NEPSI Intrinsically safe certification              user shall not change the configuration in  order to maintain ensure the explosion protection  performance of the equipment  Any change may  impair safety      For installation  use and maintenance of  the product  the end user shall observe the  instruction manual and the following standards    GB50257 1996  Code for construction and  acceptance of electric device for explosion  atmospheres and fire hazard electrical  equipment installation engineering    GB3836 13 1997  Electrical apparatus for  explosive gas atmospheres Part 13  Repair and  overhaul for apparatus used in explosive gas  atmospheres     GB3836 15 2000  Electrical apparatus for  explosive gas atmospheres  Part 15  Electrical  installations in hazardous area  other than  mines       GB3836 16 2006  Electrical apparatus for  explosive gas atmospheres  Part 16  Inspection  and maintenance of electrical installation  other  than mines       IM 12B07D02 01E    mA HART   communication  A  Input    
20.     Software  X National Instruments   NI FBUS configurator   L  Hardware  F BUS interfaces from National  Instruments  AT FBUS  PCMIA   FBUS    M  Other control systems   YOKOGAWA PRM  DTM    moou      IM 12B07D02 01E    2 6 Specification    2 3  Model and suffix codes    1  2 Wire pH ORP Transmitter  Non explosionproof type        Model    Suffix Code     Style    3        Option Code Description             PH202G              2 Wire             Transmitter   1                              mA with HART  Europe type       mA with HART  Canada type        mA with HART  North America type        Language             Option       Mounting Hardware  Hood    Tag Plate  Conduit Adapter                         SS Ce M PE Profibus   ID FF   TOTIS Japanese      2 English   IU Pipe  wall mounting bracket  Stainless steel   IPM Panel mounting bracket  Stainless steel   IH Hood for sun protection  Carbon steel       2 Hood for sun protection  Stainless steel   ISCT Stainless steel tag plate   IAFTG G 1 2    ANSI 1 2 NPT       Screw terminal   2   Epoxy baked finish   3                  1  The PH202G can be also used as ORP transmitter   Setting can be made      the field        2  It can be specified when the suffix code  A is selected      3  The housing is coated with epoxy resin     2  2 Wire pH ORP Transmitter  Explosionproof type     Model    Suffix Code     Style   S3        Option Code Description           202     2 Wire pH ORP Transmitter   1        Intrinsic safe mA with HART
21.     i amp   amp  18 H INSPECTION ITEM      R RESULT                         amp 7              GR 100M Q         500V DC  BETWEEN POWER SUPPLY TERMINALS      AND EARTH TERMINAL G   100M Q OR MORE 500V DC          11917     f amp    7     A8  G  f  100M Q         500V DC  BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G   100M Q OR MORE 500V DC                     f amp c7    A  sT Gm 600V      2    BETWEEN POWER SUPPLY TERMINALS      AND EARTH TERMINAL G   600V AC 2 SEC  AJ  011917            A sth  G  f  600V      2    BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G   600V AC 2 SEC      11917     f amp  BR                   600V      2    BETWEEN INPUT TERMINALS 11 TO 17  AND POWER SUPPLY TERMINALS       600V AC 2 SEC    ACCURACY    0 3     PT1000     amp  m INDICATION         riu          REFERENCE   Xi ACTUAL ERZ ERROR     92168    gt           10973   5 1      14983   4                   0 01pH ACCURACY    0 01        RABE    INPUT VOLTAGE EETA EAE        mV  REFERENCE ACTUAL ERROR        4141     L 00               03        X m INDICATION  mV   INPUT VOLTAGE    TE  SERI R     mV  REFERENCE ACTUAL ERROR    L9  0 1 14      1500 po       FARE       AMBIENT TEMP   amp  HUM        amp     KBE  APPROVED BY  QIC 12B07D02 21  Ed1  Mar 2007       Appendix 12 9    12 3      202        2025 2 Wire             Transmitter  Fieldbus Communication        Quality PH202G      2025  Inspection 2 Wire pH ORP Transmitter  Standards  Fieldbus Communication     1  Scope    This
22.    1 2 W  Ci 220 pF     0 uH CSA Non incendive safe Approval or       lECExintinsicalysale            PN B  Applicable standard  IEC 60079 0  IEC60079 11     C322 Nodo ABUS i             1    60073 26 1  680073 27    C22 2  No 213 M1987    22 2       61010 1   Scheme Certificate  IECEx KEM 07 0026X Class   Division 2 Groups ABCD pres or  155 Temp    cla _ y Amb     10 to 55  C ExnA fllii    emp  Class  T4  Amb  Temp    10 to   Si     Ui 17 5 V  11 380 mA  Pi 5 32 W       220 pF  Liz0 uH Temp  Class     pis Ln e 2   D  IECEx Type of protection      Entity  Ui Vmax  32 V  Ci 220 pF  Li 0 uH  Applicable standard  IEC 60079 15 2001   B FNICO  Ui Vmax  32 V  Ci 220 pF  Li 0 uH  IEC 60079 0 2004                Certificate  IECEx        07 0026X  Scheme Ex nA  nL       or  Temp  Class  T4  Amb  Temp    10 to 55  C  T6  Amb  Temp    10 to 40     E  Ui 32 V  Ci 220 pF  Li 0 uH D IM 12B07D02 01E                   IEC EPS    2 4 Specification    NEPSI Certification  PH202S K    NEPSI Intrinsically Safe Type   Cert No  GYJ081156X     Applicable Standard    GB3836 1 2000  GB3836 4 2000     Type of Protection and Marking Code   Ex ia        4   6     Ambient Temperature     T6   10 to 40  C    4     10 to 55  C   Note 1 Entity Parameters     Intrinsically safe input parameters   terminal   and       Maximum Input Voltage  Ui    31 5 V  Maximum Input Current  li    100 mA  Maximum Input Power  Pi    1 2 W  Maximum Internal Capacitance  Ci    22 nF  Maximum Internal Inductance  Li    35     
23.    Uo  lt  17 5 V   lo  lt  380 mA   Po  lt  5 32 W         gt  220 pF   Ccable  Lo20        Lcable    Electrical data of the PH202S B  amp  PH202S D  Type of protection    n        Supply and output circuit    Maximum input voltage Ui   32 V   Effective internal capacitance Ci  220pF  Effective internal inductance Li  0 pH   Sensor input circuit    Maximum output voltage Uo 14 4 V  Maximum output current 10 32 3 mA  Maximum allowed external capacitance Co   3 5 uF   Maximum allowed external inductance Lo   76 mH IM 12B07D02 01E    2 14 Specification    2 11  Control Drawing of PH202S FF PB Specification  FM Intrinsically safe Entity      FM Class I  DIV  1  Group ABCD          T4 for ambient temp   lt  55   C ee  i    PH202S F Max  cablelength  60 mtr   or PH202S P Cable dia    3   12 mm     l oz Sensor  FM Approved   Connections  barrier                Voc  Vt     24 V    15     2 2 80 eer eee eee     15        H    H  Pc       i certified   i certified                  Terminator                  1               Terminator   Ca  gt  220pF  Ccable    i    i  Laz0H  lcabe   i     B   O        Exe ul d           1 2600 fF                                       Division 1    Unclassified Location Classified Location    Sensor s  are of a passive type to be regarded as    simple apparatus     devices which neither  store nor generate voltages over 1 5 V  currents over 0 1 A  power over 25 mW or energy over  20 uJ  or are FM Approvals entity approved and meet connection requireme
24.    User   s EKA    Manual Model PH202G  Style   3   PH202S  Style   3     2 wire Type pH ORP Redox  C      Transmitter    IM 12B07D02 01E       vigilantplant        YOKOGAWA     IM 12B07D02 01E    11th Edition  Yokogawa Electric Corporation                                  5                    1  Introduction And General Description                                              1 1  1 1                                                                                         1 1                                                     1 3  2  PH202 Specifications qe                                      2 1  CNNC 1   eem 2 1  2 2  Operating specifications            a et                   2 2  2 3  Model and suffix COGGBS     saci vanecsaccatvecnbacdessinassaaddaceesddacedeanesowitasarasndecaxacadeccdes    2 6  2 4  Control Drawing of PH202S mA HART Specification  IECEXx                        2 7  2 5  Control Drawing of     2025 mA HART   Specification                                  2 8  2 6  Control Drawing of PH202S mA HART   Specification                                            FM Intrinsically safe design                                        2 9  2 7  Control Drawing of     2025 mA HART   Specification                                            FM Non incendive 4    9                           enne nnn 2 10  2 8  Control Drawing of PH202S mA HART   Specification                                 2 11  2 9  Control Drawing of PH202S FF PB Specification  IECEX                
25.    YEAR   MONTH   DAY    Code 62    ERASE    5 3 7  General    Code 70  LOAD    Code 79  CUST D    The settings should be adjusted to suit the communicating device connected to  the output  The communication can be set to HART   or to PH201 B distributor   for Japanese market only     Select adress 00 for point to point communication with 4 20 mA transmission   Address 01 to 15 are used in multi drop configuration  fixed 4 mA output    For the Yokogawa PC202 software package  the default settings match the  software as shipped     The clock calendar for the logbook is set for current date and time as  reference     Erase logbook function to clear the recorded data for a fresh start  This may  be desirable when re commissioning an instrument that has been out of service  for a while     The load defaults code allows the instrument to be returned to the default set  up with a single operation  This can be useful when wanting to change from  one application to another     Load customer defaults  This code allows the instrument to be returned to the  factory default set  except that buffer tables  code 24 25 26  are unchanged     5 3 8  Test and setup mode    Code 80  TEST    Not used    Note  attempting to change data in service code  80 and above without the proper instructions and  equipment  can result in corruption of the instrument setup  and will impair the performance of    the unit     IM12B07D02 01E    5 22 Parameter setting       Code   Display    Function    Function 
26.    set time interval 1   250 days  hold last or hold fix   set temperature coefficient   2 x 50 events       codes 24  25  amp  26  code 12   code 27   code 55    hold    code 13   code 61  62       IM 12B07D02 01E    10 4 Appendix    10 3  Setup for sensor compatibility    10 3 1  General   The inputs of the EXA transmitter are freely programmable for ease of installation  Standard glass pH  electrodes  Ag AgCl reference electrodes and Pt100 and Pt1000 temperature sensors need no special  programming  The EXA indicates a fault with a signal in the display field if there is a mismatch of sen   sors in the connection     10 3 2  Selection of measurement and reference electrode   The EXA PH202 is preprogrammed to accept industry standard glass electrodes and reference elec   trodes  The unit initiates checks for asymmetry and slope during calibration  The on line impedance  checking function has been upgraded in this most recent EXA release     The EXA is universally compatible with all types of electrodes  such as enamel and antimony  In such  systems  however  the specific isothermal point of intersection  ITP   slope  pH mV  and asymmetry  potential can be set for the type of electrode     10 3 3  Selecting a temperature sensor   The EXA PH202 reaches its highest accuracy when used with the Pt1000 temperature sensor  This ele   ment offers a 10 fold increase in resistance dependence over the Pt100 sensor  Choice of temperature  sensor is made in the Service Codes found in Ch
27.   3        11 12 14 17 13 LOW IMP oH M 15 16  SUPPLY L TEMP LE INPUT 2 INPUT 1                                    T      4     G 11 12 14 17 13 LOW IMP ficu ime  15 16  suPPLY   TEMP LE INPUT 2 INPUT 3                                  Fig  3 8  Jumper positions    IM 12B07D02 01E       3 8 Installation and wiring             or          G 11 12 14117 13                            15 16    SUPPLY  L TEMP LE INPUT 2 INPUT 1       Figure 3 9  Terminal identification labels    3 6  Sensor wiring   Refer to figure 3 10  which includes drawings that outline sensor wiring    For sensor wiring of PHSEFP  PH8ERP  PH8EHP  OR8EFG  OR8ERG      405      406      485   DPA405  DPA406  DPA485  HF405  DPAS405  DPAS485 and FU20  see APPENDIX 2  Chapter 11      The EXA analyzers can be used with a wide range of commercially available sensor types  both from  Yokogawa and other manufacturers  The sensor systems from Yokogawa fall into two categories  the  ones that use a fixed cable and the ones with separate cables     To connect sensors with fixed cables  simply match the terminal numbers in the instrument with the iden   tification numbers in the instrument on the cable ends     The separate sensors and cables are not numbered  but instead use a color coding system  The elec   trodes have a colored band incorporated in the label on the connection cap       Red for measuring electrodes  both pH and ORP      Yellow for reference electrodes     Blue for combined sensors with both measuring and 
28.   55  C  AMB TEMP    10  55  C AMB TEMP   410  55  C STYLE  STYLE STYLE No   No  No  A        Ta 90  GN SEE USER S MANUAL  BEFORE USE  YOKOGAWA   YOKOGAWA   YOKOGAWA    Made in Japan Tokyo 180 8750 JAPAN Made in Japan Tokyo 180 8750 JAPAN Made in Japan Tokyo 180 8750 JAPAN       O             O  n200 O  O O      FISCO field device     PH ORP TRANSMITTER PH ORP TRANSMITTER No  IECEx        07 0026X  MODEL   PH202S F MODEL   PH202S P  SUFFIX SUFFIX SEE CONTROL DRAWING  No  KEMA 07        0048 X   uo  ExiallC T4 IP65  SUPPLY FISCO SUPPLY FISCO SEE CONTROL DRAWING  17 5VDC     380mAI5 32W 17 5VDC     380mA 5 32W         or 24VDC     250     1 2     or 24VDC      250mA 1 2W AND AEx ia       OUTPUT   FF TYPE111     511     OUTPUT   PROFIBUS PA teil CONTROL DR  NING  Li 0 pH  Ci 220pF Li 0 uH  Ci 220pF APPROVED IKE025 A10 PA to P 7  AMB TEMP    10  55  C AMB TEMP   410  55  C CL I  DIV 1       ABCD  STYLE STYLE  5      Ex ia IIC T4  No  No  SEE CONTROL DRAWING  1881741        65          35                                                        2 sbstuton de composants  Made in Japan Tokyo 180 8750 JAPAN Made in Japan Tokyo 180 8750 JAPAN      ur TUB                                               nz00      2025        O  O O  PH ORP TRANSMITTER PH ORP TRANSMITTER FNICO field device        No  IECEx KEM 07 0026    MODEL   PH202S B MODEL   PH202S D EX PALM Tt for Taso  SUFFIX SUFFIX    CONTROL DRAWING  No  KEMA 07ATEX0049  EEx nA nL      T4         36 EEx nl  IC T6 for Ta 4o c  SEE CON
29.   ATEX    Intrinsic safe mA with HART  CSA    Intrinsic safe mA with HART  FM    Intrinsic safe mA with HART  NEPSI    Intrinsic safe Profibus  ATEX  CSA  FM    Intrinsic safe FF  ATEX  CSA  FM    Non incendive FF  ATEX  CSA  FM    3   Non incendive mA with HART  ATEX  CSA  FM    3   Non incendive Profibus  ATEX  CSA  FM    3     SS          Language       Option    Mounting Hardware  Hood    Tag Plate  Conduit Adapter          Japanese  English          Pipe  wall mounting bracket  Stainless steel   Panel mounting bracket  Stainless steel   Hood for sun protection  Carbon steel   Hood for sun protection  Stainless steel   Stainless steel tag plate   G 1 2   1 2 NPT   Epoxy baked finish   2             1  The     2025        be also used as ORP transmitter   Setting can be made in the field         2  The housing is coated with epoxy resin       3  When the instrument with Suffix Code   B  N  D  is used  take measures so that  the display window is not exposed to direct sunlight     IM 12B07D02 01E    Specification 2 7    2 4  Control Drawing of PH202S mA HART   Specification  IECEx                                                                                                                     Intrinsically safe design  IEC Ex standard EX ia       T4 for ambient temp   lt  55  C EX ia or ib  T6 for ambient temp   lt  40  C Certified safety barrier or power  Certificate nr  IECEx KEM 06 0052X        a      2025  pH ORP transmitter        compatible   24 volts DC Nominal  Su
30.   Note     0   9  where   Maintenance passcode On E  0   no passcode Commissioning passcode Off 0 Off  12111  22333  3 777   Commissioning passcode On    4 888  5 123  6 957   Service passcode      0 Off  7 331  8 546  9 847   Service passcode On  53  Err 4 1 Error setting Impedance low  input 1    Soft fail 0  Impedance low  input 1  Hard fail 1 1 Hard   Err 5 1 Impedance high  input 1  Soft fail 0  Impedance high  input 1  Hard fail 1 1 Hard   Err 4 2 Impedance low  input 2  Soft fail 0  Impedance low  input 2  Hard fail 1 1 Hard   Err 5 2 Impedance high  input 2  Soft fail 0  Impedance high  input 2  Hard fail 1 1 Hard   Err 07 Temperature too high Soft fail 0  Temperature too high Hard fail 1 1 Hard   Err 08 Temperature too low Soft fail 0  Temperature too low Hard fail 1 1 Hard   Err 09 pH out of range Soft fail 0  pH out of range Hard fail 1 1 Hard   Err 11 Wash recovery check Soft fail 0 0 Soft  Wash recovery check Hard fail 1   Err 16 Call for maintenance Soft fail 0 0 Soft  Call for maintenance Hard fail 1  54 Not used  55  CALL M Call for maintenance   Set time limit for calibration Off 0 0 Off  Set time limit for calibration On  Set valid maintenance period if On 250   days  56  DISP Display resolution Set pH decimal display 0 1 pH 0  Set pH decimal display 0 01pH 1 1 0 01 pH  57 59 Not used                                     IM12B07D02 01E    Parameter setting 5 21    5 3 6  Communication setup    Code 60                    ADDR     Code 61   HOUR   MINUT   SECND
31.   Press the     key  The message display will show    SERV     Press the  YES  key  The message display will show     CODE    and the data display will  show  00  with the first digit of O flashing    Press the   gt   key once  The second digit of 0 flashes    Press the     key once  The data display will show  01  with 1 flashing    Press the  ENT  key  The message display will show  PH ORP       Press the     key once  The data display will change from    0    to    1       Press the  ENT  key    Turn on power again     In this state  change the simulation input of the standard voltage generator as shown in Table 3   The corresponding data must be within the range                    Table 3  Simulation Input  mV  Data Display  mV    1500  1500  1  0 0  1   1500  1500  1                             Return to      mode by following the steps below        then the tests        completed   a   b     Press the     key  The message display will show     SERV       Press the  YES  key  The message display will show                 and the data display will  show  00  with the first digit of O flashing    Press the   gt   key once  The second digit of 0 flashes    Press the     key once  The data display will show    01    with 1 flashing    Press the  ENT  key  The message display will show  PH ORP       Press the     key once  The data display will change from    1    to    0       Press the  ENT  key     QIS 12B07D02 71E       IM 12B07D02 01E       Appendix 12 15       3 3  3 6 
32.   Two wire system 4 20 mA    B  Power supply    PH202G   up to 40 volts  PH202S   up to 31 5 volts  Note  The transmitter contains a switched  power supply  drawing its energy  from the 0 4 mA section of the signal   Consequently the 17 volt limit is  applied at 4 mA  The characteristic of  the unit is such that above about 7 mA  on the output  the terminal voltage can  drop to 14 5 volts without problem    see figure 2 2   C  Transmission  Isolated output of 4 to 20 mA DC   D  Signal   Maximum load 4250   see figure 2 1     Burn to signal failure acc   NAMUR Recommendation NE43   18 01 1994     E  Operating range   3 9 to 21mA    F  Communication                1200 Baud  FSK    modulated on 4 to 20 mA signal        Configuration   Local with 6 keys    H  Software   Firmware based on Yokogawa stack   1  Hardware   Yokogawa            Modem F9197UB    J  Other Control systems    Yokogawa PRM  Rosemount AMS   Siemens PDM  K  Hand Terminal   Rosemount HHT 275 375  L  Other control systems  Yokogawa PRM   Rosemount AMS  Siemens PDM    M  Output span    pH   min 1 pH  max 20 pH    max 9096 zero suppression             instrument is user  programmable for linear or non     linear pH ranges     N  Cable specification    0 5 mm diameter or 24 AWG over  maximum length of 1500 m  O  DD specification    The PH202 Device Description is  available enabling communications  with the Handheld Communicator  and compatible devices     PROFIBUS PA communications    mooo        Input signal  Di
33.   and    12 0     Check the current outputs  Corresponding to the data display  the  current output must be within the range shown in Table 1     When the data display shows    12 0     pressing the  ENT  key causes the message display to  show    RIPPLE    and the data display to remain unchanged  Press the  ENT  key again  The  message display will show    mA    and the data display    16 0                    Table 1  Data Display Current Output  mA   4 0 4  0 02  12 0 12  0 02  20 0 20  0 02                Temperature Indication Check    Following Section 3 3  press the  ENT  key until the message display shows    PT1000     In this  state  change the resistance value of the decade resistance box as shown in Table 2 and check  the data display  The corresponding value on the data display must be within the range shown  in Table 2     QIS 12B07D02 21E       IM 12B07D02 01E                         12 7          3 5    3 6    3 3                Table 2  Reference Temperature   Resistance Box Resistance Data Display                       20 921 6  20 0  0 3   25 1097 3  25 0  0 3   130 1498 3  130 0  0 3                        Indication Check    Following Section 3 4  press the  ENT  key until the message display shows    PH INP     In this  state  change the simulation input as shown in Table 3 by means of the standard  voltage current source and check the data display  The corresponding value on the data display  must be within the range shown in Table 3                 Tab
34.   calibrations and programmed data changes  By reference to this  log  users can easily evaluate diagnostic information to determine predictive maintenance schedules   For example  by monitoring the deterioration in the slope of the pH sensor  it can be changed before  a failure  or process shutdown  occurs     IM 12B07D02 01E    10 6 Appendix    10 5  Set up for Pfaudler Type 18 sensor    The PH202 is intended to measure with all sorts of pH sensors  including the Pfaudler Type 18 sensor   The Pfaudler design of dual membrane system uses two enamels of differing sensitivity  The first a pH  sensitive membrane  and the second one that responds to Na  and K  and acts as a reference     The analyzer has dual high impedance inputs which measure perfectly even with very high impedance  sensors  However  the impedance measuring system  diagnostics  needs to be set up for best perform   ance     10 5 1  General set up    1  Set impedance measuring hardware  This is done by the use of links on the terminals adjacent to  the input terminals  For the Pfaudler system  this means that the terminals should have the links  disconnected in order to set for HIGH HIGH impedance measuring     2  Set the impedance check in software  Use codes 03  amp  04 to enable the measurement and set for  high impedance and configure appropriate limits     Code 03 set to 1 0 1 low limit 1 Megaohm  high limit 1 Gigaohm  Code 04 set to 1 0 1 low limit 1 Megaohm  high limit 1 Gigaohm    3  Set the temperature
35.  0  added   and NEPSI Certification  added to page 2 4  P 2 6  NEPSI suffix code of   K  added to the PH202S MS   code  CMPL 12B07D02 23E of PH202S S3  revised to 2nd edition  some parts    no  deleted      
36.  10 to   55       10 to 131   F     C  Storage temperature     30 to  70        20 to 160   F     D  Humidity    10 to 90  RH  Non condensing     F  Housing   Case  Cast aluminum case with chemically  resistant coating   Cover  Polycarbonate window    Case color  Off white  Equivalent to Munsell  2 5Y8 4 1 2    Cover color  Deep sea Moss green   Equivalent to Munsell 0 6GY3 1 2 0         Power Output  Grounding cable gland    terminal    IM 12B07D02 01E    Cable and terminals    The PH202 is equipped with terminals  suitable for the connection of finished  cables in the size  0 13 to 2 5 mm  26 to  14 AWG   Cable entry  2 cable glands 1 2NPT  The cable  glands will form a tight seal on cables with  an outside diameter in the range of 6 to 12  mm  0 24 to 0 47 inches    Construction   Weather resistant to     65         NEMA      standards  Mounting   Pipe  wall or panel mounting  using  optional hardware  Weight   Approx  1 6 kg    G  Shipping details    Package size w x h x d  290 x 300 x 290 mm   11 5 x 11 8 x 11 5 in   Packed weight approx  2 5 kg  5lb     H  Data protection    EEPROM for configuration and logbook  and  lithium cell for clock        Watchdog timer    Checks microprocessor    J  Automatic safeguard    Return to measuring mode when no keystroke  is made for 10 min     K  Operation protection    3 digit programmable password     L  Sensor impedance checking     Independent impedance check on measuring  and reference sensor elements  with  temperature compensa
37.  After calibration   the unit checks the calculated asymmetry potential and the slope to determine if they are still within limits  specified by the software     The slow shift of asymmetry potential could signal a poisoning of the reference electrode system by the  process  The decrease of slope equals a decrease of sensitivity of the glass electrode or can show a  coating buildup at the electrode     The EXA makes a distinction among diagnostic findings  All errors are signaled by the FAIL flag in the  display  Only faults in the measuring circuit can be set as HARD FAIL  with    Burn up or Burn down    sig   nals on the mA output     What follows is a brief outline of some of the EXA troubleshooting procedures  followed by a detailed  table of error codes with possible causes and remedies     NOTE   The diagnostic function of the EXA gives a variable time interval between impedance checks  up  to 5 minutes  When trouble shooting  a manual impedance check can be initiated by following the  procedure in section 5 1 4     IM 12B07D02 01E    8 2 Troubleshooting    8 1  Diagnostics    8 1 1  Off line calibration checks  The EXA transmitter incorporates a diagnostic check of the asymmetry potential after a calibration has  been completed  This is a valid check for both manual and automatic calibration routines     The actual value can be called up from the DISPLAY routine in the maintenance menu  A large value  often indicates poisoning or pollution of the reference system used 
38.  If the asymmetry potential exceeds  programmable limits  the EXA generates an error    2      The EXA also performs diagnostics to check for the slope of the pH electrode after automatic calibration  is completed  The actual value of the slope can be called up on the DISPLAY routine in the maintenance  menu  SL   This value is an indication of the age of the electrode  If the value stays within the limits of 70  to 110 percent of the theoretical value  59 16 mV pH at 25  C   it is accepted  Otherwise  the unit gener   ates an error            Activation or deactivation of the asymmetry diagnostic check and slope check is made from the Service  Codes  See Chapter 5 or Chapter 10  Appendix      8 1 2  On line impedance checks   The EXA has a sophisticated impedance checking system  The sensors can be checked for their imped   ance over a very wide range  which makes the tool equally useful for glass  enamel  reference and metal   ORP  sensors  The measurement is temperature compensated for the characteristic of the pH glass  sensor     In order to measure accurately over such a wide range  it is necessary to split the range into two  This is  done by a pair of jumper settings  high range and low range can be set on either input  making the sys   tem extremely flexible     The following error message table gives a list of problems that are indicated when the high or low imped     ance limits are exceeded for a sensor  Such things as fouling  breakage and cable faults are readil
39.  Supply  Po   1 2 Watt  SENSOR S   terminals 11 17 Functional  earth  Hazardous area Safe area   lt i      gt    lt 2  Zone 0 or 1    Sensor s  are of a passive type to be regarded as    simple apparatus        Electrical data of the     2025     Supply and output circuit  terminals   and      Maximum input voltage Ui  31 5 V  Maximum input current I    100 mA   Maximum input power Pi   1 2 W   Effective internal capacitance C    22 nF   Effective 1nternal inductance Li   35 uH       Sensor input circuit  terminals 11 through 17    Maximum output voltage Uo   14 4 V  Maximum output current       32 3 mA   Maximum allowed external capacitance Co   600 nF   for PH202S E  C  U    Co   3 5 uF  for PH202S N    Maximum allowed external inductance Lo   34 mH  for PH202S E  C  U    Lo   76 mH  for     2025          Barriers and power supply specification must not exceed the maximum values as shown  in the diagram above  These safety descriptions cover most of the commonly used  industry standard barriers  1solators and power supplies              safety barrier shall be certified by notify body      Installation should be in accordance with local installation requirements      If use ordinary wirings  the general purpose equipment must have Nonincendive  Field Wiring terminal approved     The Hand Held Communicator must be of a ATEX certified intrinsically safe type 1n case  1t is used on the intrinsically safe circuit in the hazardous area or of a ATEX certified  non incendive type 
40.  Warning flags and units  pH and mV      J  Power supply    Nominal 24 volt DC loop powered system         202      17 to 40 volts  see Fig  2 1         202     17 to 31 5 volts  see Fig  2 2     K  Maximum load resistance   For the PH202G  see Fig  2 1   For the     2025  see Fig  2 2     1 1150      1000             gt            Possible    Load Resistance  Q   gt     0 1718 20 25 30 35 40  Voltage          gt   Fig  2 1  Supply voltage  load diagram for the PH202G    F2 1E eps    800                        200 Possible    Load Resistance             A      e       12 16 20 24 28 32  Voltage  V              Fig  2 2  Minimum terminal voltage for the PH202S     Note  In this manual a sign appears if it       concerns the PH202G  S  E   C   U   N   K     IM 12B07D02 01E    2 2 Specification    2 2  Operating specifications    A  Performance   pH      Linearity    0 01 pH    Repeatability   0 01 pH    Accuracy    0 01 pH  Performance   ORP     Linearity    1 mV     Repeatability    1 mV     Accuracy    1 mV    Performance   Temperature with Pt1000 Q          Balco  5k1Q  3500  6k8Q  PTC10kO  amp   8k55Q      Linearity    0 3         Repeatability   0 1         Accuracy    0 3       Performance   Temperature with Pt100 Q    Linearity     0 4   C    Repeatability   0 1         Accuracy    0 4   C    Note on performance specifications   The following tolerance is added to above  performance   mA output tolerance     0 02 mA of   4   20 mA     B  Ambient operating temperature    
41.  accordance  with the right table     T03 ai  S OFF  light Push  YES  key       IM 12B07D02 01E    Appendix 12 1  12  APPENDIX 3 QUALITY INSPECTION       12 1  PH202G  PH202SJ 2 Wire pH ORP Transmitter       Quality PH202G  PH202SJ    Inspection      2 Wire                 Standards ire pH O ransmitter    1  Scope  This inspection standard applies to the PH202G  PH202SJ 2 Wire pH ORP Transmitter     2  Inspection Items    2 1 Insulation resistance test   2 2 Current output test   2 3 Temperature indication check  2 4 pH indication check   2 5 ORP indication check    3  Inspection Methods  Standards and Conditions    e Connect the testing circuit as shown in Figure 1  Allow the instrument to warm up for at least  5 minutes before conducting the tests  For the connections for the insulation resistance test   follow the instructions in Section 3 1     e Performance tests should be done in the inspection mode where the tests from Section 3 3  through Section 3 5 take place in sequence and cannot be retraced  If the reconfirmation of  a test is needed  turn off the power to the transmitter  turn on the power again  and enter the  inspection mode to restart the tests     3 1 Insulation Resistance Test    As for the PH202G  follow the instructions below      1  Apply 500 V DC between the power supply terminals shorted together    and      and the  earth terminal       The insulation resistance must be 100      or greater      2  Apply 500 V DC between the input terminals shorted tog
42.  are available in English  German and French   Should you require Ex related instructions in your  local language  you are to contact your nearest  Yokogawa office or representative     Alle brugervejledninger for produkter relateret   til ATEX Ex er tilg  ngelige pa engelsk  tysk og  fransk  Skulle De   nske yderligere oplysninger  om h  ndtering af Ex produkter p   eget sprog  kan  De rette henvendelse herom til den n  rmeste  Yokogawa afdeling eller forhandler     Tutti i manuali operativi di prodotti ATEX con   trassegnati con Ex sono disponibili in inglese   tedesco e francese  Se si desidera ricevere 1 man   uali operativi di prodotti Ex in lingua locale  met   tersi in contatto con l ufficio Yokogawa pi   vicino o  con un rappresentante              Todos los manuales de instrucciones para los          ductos antiexplosivos de ATEX estan disponibles  en ingl  s  aleman y franc  s  Si desea solicitar las  instrucciones de estos articulos antiexplosivos en  su idioma local  debera ponerse en contacto con      oficina    el representante de Yokogawa mas  cercano     Alle handleidingen voor producten die te maken  hebben met ATEX explosiebeveiliging  Ex    zijn verkrijgbaar in het Engels  Duits en Frans   Neem  indien u aanwijzingen op het gebied van  explosiebeveiliging nodig hebt in uw eigen taal   contact op met de dichtstbijzijnde vestiging van  Yokogawa of met een vertegenwoordiger     Kaikkien          Ex  tyyppisten tuotteiden kayt               ovat saatavilla engl
43.  be interpreted to give information  about the oxidation state of the process solution  and derived information like  the absence of a compound  like Cyanide for example which is destroyed in  oxidizing solutions      Enable the use of a second measuring parameter simultaneously with pH  the  main parameter     With the correct sensor  e g FU20   ORP measurement is possible as  parameter 2 With the same sensor  rH measurement is possible as parameter  2  this is calculated from pH and ORP and is a value which gives the oxidizing  power of the solution while compensating for the effect of pH    This function is particularly useful for applications where both the pH and  oxidation reduction potential of the process need to be known  The availability  of both measurements in a single system is convenient    Note that in both cases a suitable sensor combination is needed to make this  possible  The Yokogawa   020  4 in 1  sensor can be used for this purpose  or  a combination of individual sensors  Contact your local Yokogawa sales office  for advice regarding applications and sensor selection     The EXA PH202 has an impedance check capable of monitoring the impedance  of all sorts of sensor systems  In order to  fine tune  this diagnostic tool it is  necessary to set it up to match the sensors used  The default settings give a  good setup for a conventional system comprising pH glass sensor and a  reference electrode  either as individual electrodes or as a combination style  s
44.  compensation sensor as 100 Ohm Platinum RTD with service code 10   Code 10 set to 1 100 Ohms Pt     The system will now respond properly to the Pfaudler type 18 sensor  and the other functions of the  EXA analyzer will need to be set in the normal way to suit the use to which the loop is being put   Output ranges  control functions and alarms should all be set as described elsewhere in this manual     10 5 2  Calibration set up    4  The alternative Zero point  calibration and display  according to IEC 60746 2 may be enabled in  service code 27  and set in the MAN CAL routine  A value of 10 5 pH is a good starting point for the  Pfaudler 18 sensor     5  Where lab test data are available for the sensor  service code 23 can be used to set values for ITP  amp   Slope  and As pot for parameter 2 when enabled       This method can be useful for the type 18 sensor  as it is not usual to perform regular calibrations on  this system as with normal sensors  This is because the system may well respond differently  to ordinary  buffers  than with the process solutions  The procedure is to determine the temperature response  ITP   and the sensitivity  Slope  of the sensor  and enter these values in code 23      Because this is a rather complex procedure  it is recommended instead to use the default settings of ITP    7 00  and Slope   100    and make a single point  MAN CAL  calibration in the process at the work   ing temperature  and at the normal operating  control setpoint  pH  Thi
45.  impedance limit    x None  K  M or G    High impedance limit    x None  K  M or G          Low   High   Temp comp off  Temp comp on  Imp check off    imp check on    Press NO to step through choice of units     press YES to select units  then use the     gt   AENT keys to set the value    Press NO to step through choice of units     press YES to select units  then use the        gt            keys to set the value                   T5 3 1E eps    Yokogawa sensors with the liquid earth electrode include the PH8EFP  PH8ERP  PH8EHP     OR8ERG  OR8EFG and FU20     The liquid earth electrode is inside the sensor  and at the potential of the measured solution  The  liquid earth electrode is also used for an impedance check  to determine if a functional sensor  is connected   Be sure to connect the wire from the liquid earth electrode to terminal 14 on the    transmitter                                impedance check detects when the glass electrode is broken  and also provides a check of  junction contamination and junction clogging  Such checks are qualitative rather than absolute   and    relatively loose high low limits for  acceptable  values are usually set     In the measurement in low conductivity water such as pure water  the impedance check does not  work  Set the impedance check function for the reference electrode to off in Service Code 04     If the sensor doesn t have a liquid earth electrode  it s not possible to perform an automatic   impedance check to determine if
46.  inspection mode      1  Entering Service Code 87  a  Press the     key  The message display will show   OUTP       Press the  NO  key until the message display shows     SERV         Press the  YES  key  The data display will show  00  with the first digit of 0 flashing      Press the     key eight times  The data display will show  80  with the first digit of 8 flashing      Press the   gt   key once  The data display will show  80  with the second digit of O flashing   Press the     key seven times  The data display will show    87    with the second digit of 7  flashing    g  Press the  ENT  key  The message display will show     PASS        2  Setting Password 070  a  Press the   gt   key once  The data display will show    000    with the second digit of 0 flashing   b  Press the     key seven times  The data display will show    070    with the second digit of 7  flashing   c  Press the  ENT  key  The message display will show  HIF    d  Press the  YES  key                               3  Checking the date and time  a  Press the  ENT  key  The data display will show the date in day month year order   b  Press the  ENT  key  The data display will show the time in hour minute second order   c  Press the  ENT  key  The message display will show    mA        This is the inspection mode  When the message display shows    mA     check the current outputs     Each time the  ENT  key is pressed  the value on the data display increases in steps of 4 0 like     4 0        8 0  
47.  inspection standard applies to the PH202G and     202  2 Wire pH ORP Transmitters   Fieldbus specification      2  Inspection Items    2 1 Insulation resistance test   2 2 Dielectric strength test  2 3 Temperature indication check  2 4 pH indication check   2 5 ORP indication check   2 6 Fieldbus communication functional check    Note  Items marked with an asterisk     may only be confirmed by a test certificate     3  Inspection Methods  Standards and Conditions    e Connectthe testing circuit as shown in Figure 1  Allow the instrument to warm up for at least  5 minutes before conducting the tests  For the connections for the insulation resistance test   follow the instructions in Section 3 1 and for the connections for the dielectric strength test   follow the instructions in Section 3 2    e Use test equipment shown in Figure 1  or equivalent  for the tests     3 1 Insulation Resistance Test     1  Apply 500 V DC between the power supply terminals shorted together    and      and the  earth terminal  G   The insulation resistance must be 100 MO or greater      2  Apply 500 V DC between the input terminals shorted together  11 to 17  and the earth  terminal  G   The insulation resistance must be 100 MO or greater     3 2 Dielectric Strength Test  Required Only for PH202S     Apply 600 V AC  an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz  or 60 Hz  between the terminals shown below  for at least 2 seconds  The insulation must  withstand this vo
48.  interface   certified   i certified      O Terminator        iTerminator        O          Safe area Zone 0 or 1  Hazardous area       Sensor s  are a thermocouple  RTD s  passive resistive switch devices  or is CSA entity  approved and meet connection requirements   e Electrical data of the PH202S F  amp  PH202S P     Supply and output circuit   Maximum input voltage Ui  24 V  Maximum input current li  250 mA  Maximum input power Pi  1 2 W  Effective internal capacitance Ci   220 pF   Effective internal inductance Li   0 pH   or  Maximum input voltage Ui  17 5 V  Maximum input current li  380 mA  Maximum input power Pi  5 32 W  Effective internal capacitance Ci   220 pF   Effective internal inductance Li   0          Sensor input circuit   Maximum output voltage Uo   14 4 V  Maximum output current lo   32 3 mA  Maximum allowed external capacitance Co   600 nF  Maximum allowed external inductance Lo   34 mH       Any CSA approved 1 5  interface may be used that meets the following requirements   Uo  lt  24    lo  lt  250 mA        lt  1 2 W        gt  220 pF   Ccable  Lo  gt  0 uH   Lcable  or  Uo  lt  17 5 V  lox  lt  380 mA  Po  lt  5 32 W  Co  gt  220 pF   Ccable  Lo  gt  0 pH   Lcable    Installation should be in accordance with Canadian Electrical Code  Part   or CEC  Part 1   Maximum safe area voltage should not exceed 250 Vrms        Electrical data of the PH202S B  amp  PH202S D  non incendive    For Class     Div 2  Group ABCD the CSA approved 1 5  interface is not 
49.  may NOT be connected to ground at the distributor         f the transmitter is mounted on a non conducting surface  e g  a brick wall  it is recommended to  ground the shield of the 2 wire cable at the distributor end     IM 12B07D02 01E    3 6 Installation and wiring    3 4 2  Connection of the power supply    The terminal strip is accessed as was described in   3 2 1  Use the left hand gland to insert the supply   output cable to the transmitter  Connect the supply to the terminals marked      and G as is indicated in    figures 3 8 and 3 9     3 4 3  Switching the instrument on    After all connections are made and checked  the power can be switched on from the distributor  Observe  the correct activation of the instrument at the display  If for any reason the display does not indicate a    value  consult the trouble shooting section                                                                                                               11 12 14 17 13        mO          we 15 16    12 M 17 1   LOW IMP   45 16  TEMP LE INPUT 2 INPUT 1 TEMP LE INPUT 2 INPUT 1   iink J Uink J  rd  bl bl  rd rd  bl rd  bl bl  Green bk Yellow Red       Blue  D  Colour code E      pil c        tem   pH ORP ombi  rd   red p  temp  pH Ref  bl   blue  bk   black ORP Ref  wt   white  SINGLE  ELECTRODES    Fig  3 7  Connection diagrams    IM 12B07D02 01E    Installation and wiring 3 7    3 5  Wiring the sensor system  3 5 1  Impedance measurement jumper settings    NOTE   It is important to dec
50.  terminal        2  Between the input terminals shorted together  11 to 17  and the earth terminal  G      3  Between the input terminals shorted together  11 to 17  and the power supply terminals  shorted together    and         3 3 Temperature Indication Check    Connect the instruments as shown in Figure 1  and set them as follows   Connect a shorting bar between terminals LOW and IMP     Decade resistance box  921 6 O  Standard voltage source  0      DC source  24 V DC    Change the resistance value of the decade resistance box as shown in Table 1  The  corresponding temperature indication must be within the range        YOKOGAWA   QIS 12B07D02 71E    Yokogawa Electric Corporation 1st Edition  Sep  2007  YK              IM 12B07D02 01E    12 14                         2 3                   Table 1  Reference Temperature   Resistance Box Resistance Data Display                       20 921 6  20 0  0 3   25 1097 3  25 0  0 3   130 1498 3  130 0  0 3                3 4 pH Indication Check    Set the instrument as follows   Decade resistance box  10000    Change the simulation input value of the standard voltage generator as shown      Table 2          corresponding pH data must be within the range                    Table 2  Check Point  pH  Simulation Input  mV  Data Display  pH   0  414 1 0 00  0 01  7 0 0 7 00  0 01  14  414 1 14 00  0 01                                3 5        Indication Check    Put the instrument      ORP mode by following the steps below   a   b   
51.  the display board   This menu is used to set such values as the output ranges and hold features  It also gives access to the  service menu   See table 4 1      LEVEL 3  Service   For more advanced configuration selections  press the button marked     then press    NO    repeatedly  until you reach  SERV  Now push the    YES    button  Selecting and entering  Service Code    numbers       the commissioning menu provide access to the more advanced functions  An explanation of the Service  Codes is listed in chapter 5 and an overview table is shown in chapter 10     Table 4 1  Operations overview                                       J Routine   Function         Chapter     Maintenance AUT CAL Calibration with programmed buffer solutions 6  MAN CAL Calibration with other buffer solutions 6  SAMPLE Grab sample calibration 6  DISP Read auxiliary data or set message display 4  MAN IMP Manual start of impedance check 5  TEMP Select automatic or manual compensation 5  HOLD Switch hold on off  when activated  5   Commissioning  OUTP Adjust the output range 5   HOLD Activate the hold function 5   Service  SERV Fine tune the specialized functions of the transmitter 5    Access to coded entries   from the commissioning   level    NOTE     All three levels may be separately protected by a password  See Service Code 52 in chapter 5  Service Code table for details on setting passwords     IM 12B07D02 01E    4 2 Operation       Output hold flag Manual temperature Fail flag  compensation fla
52.  the sensor is connected  Turn the functions of Service Codes 3 and  4 off when using such a sensor   Yokogawa sensors without liquid earth terminals include the HA405  HA406  DPA405  DPAS405   DPA406  HF405  HA485  DPA485 and DPAS485     IM 12B0702 01    E                   11 15    When the OR8EFG      OR8ERG sensor is used  set both INPUT1 and INPUT2 to the low impedance  measurement  Confirm the position of the jumper cable of the wiring diagram     Setting for ORP     dm Bb il                       G 1112141713 tow me   Sich     15 16  INPUT 2 INPUT 1    SUPPLY 2 TEMP LE                               The temperature compensation of the impedance measurement is for conventional pH glass sensors   When other sensors are used  switch this feature off     11 6 3  Supplement of temperature sensor  When the temperature sensor is not connected with the transmitter  when no connections are made  to terminals 11 and 12   set the manual temperature compensation  refer to Section 5 1 1    Yokogawa   s pH sensors  Models     405  DPA405  DPAS405 and     405  do not          a built in  temperature sensor     11 6 4  Auto Return  Manual HOLD is automatically turned OFF   Calibration mode or Commissioning mode is changed to Measurement mode automatically   If no key is operated for 10 minutes in any mode other than measurement mode or after 10 minutes  in Hold status  Auto Return  factory setting  On  1  in service code 50  will be activated to return the  transmitter to measurement m
53.  the signal returns to normal  This allows a latching alarm unit to record  the error  In the case of the EXA the diagnostics are extensive and cover the  whole range of possible sensor faults     The table function allows the configuration of an output curve by 21 steps   intervals of 5      The following example shows how the table may be configured to linearise the  output with a mA curve                    IM12B07D02 01E    5 18 Parameter setting                                                 Code   Display Function Function detail X Default values  mA Outputs  30 Not used  31  OUTP F mA output functions   pH 0 0  Code 01 set for pH pH  table  1  Parameter 2  with suitable sensor s   2  and when enabled in code 02   32  BURN Burn function No burnout 0 0   No Burn   Burnout downscale 1  Burnout upscale 2  Pulse burnout 3  33  34 Not used  35  TABLE Output table for mA   0  Linearisation table for     1      5  steps    5  The measured value is set    the main   10  display using the  gt      ENT keys  for  each of the 5  interval steps   vee Where a value is not known  that value may   90  be skipped  and a linear interpolation will   100  take place   36 39 Not used          IM12B07D02 01E       5 3 5  User interface    Code 50    Code 51    Code 52    Code 53    Code 54    Code 55    Code 56                                      55            4 1     CALL M     DISP    Parameter setting 5 19    When Auto return is enabled  the transmitter reverts to the measuring mode  f
54.  this key increases  the value by one unit  The value can not be decreased  so in order to obtain a  lower value  increase past nine to zero  then increase to the required number    ENT  When the required value has been set using the     amp    keys  press ENT to confirm   the data entry  Please note that the EXA does not register any change of data until  the ENT key is pressed                                          key This is the commissioning mode key  It is used to obtain access to the commissioning  menu  This can only be done with the cover removed or opened  Once this button has  been used to initiate the commissioning menu  follow the prompts and use the other  keys as described above     IM 12B07D02 01E    Operation 4 3    4 3  Setting passcodes    4 3 1  Passcode protection   In Service Code 52  EXA users can set passcode protection for each one of the three operating levels   or for any one or two of the three levels  This procedure should be completed after the initial commis   sioning  setup  of the instrument  The passcodes should then be recorded safely for future reference     When passcodes have been set  the following additional steps are introduced to the configuration and  programming operations     Maintenance  Press MODE key  The display shows 000 and  PASS   Enter a 3 digit passcode as set in Service Code 52 to obtain access to the Maintenance Mode    Commissioning  Press   key  The display shows 000 and  PASS   Enter a 3 digit passcode as set in Ser
55.  too high  Reference sensor fouled Clean or replace sensor   Limits set in service code 04   Liquid earth disconnected Check sensor immersion  Insufficient electrolyte Check electrolyte reservoir  E7   Temperature sensor open Process too hot or too cold Check process   gt  140  C  or  lt  10  C for 8k55  Wrong temperature sensor setting Check sensor  amp  setting  Temperature sensor damaged Check connections  E8   Temperature sensor shortened Process too cold or too hot Check process   lt   30   C  or  gt  120   C for 8k55  Wrong temperature sensor used Check sensor  amp  setting  Temperature sensor damaged Check connections  E9                         out of range   2 to 16 pH  Sensors disconnected Check cabling  Sensor wrongly connected Check cabling  Sensor s  defective Replace sensor s   E10   EEPROM write failure Fault in electronics Try again  if unsuccessful  contact Yokogawa  E11   Wash recovery check error Measuring sensor aged Replace measuring sensor   if communication is set to pH201 B in code   Sensor still coated after washing Check cleaning system  60  Defective wash system If needed adjust timings  E12   ORP   rH outside of preset limits Sensors disconnected or wrongly Check cabling  connected  E14  No valid calibration data  Data lost after switching from Recalibrate  pH to ORP  E15  Cable resistance to temperature Cable resistance too high Use Pt10000  sensor exceeds limit value  Corroded contacts Clean and reterminate  Wrong sensor programmed Reprogram  E16
56. 0 0     ASP 1D   0 0 mV   ASP mV mV  24   BUF ID 4 NIST 4  25   BUF ID 7 NIST 7  26    BUF ID 9 NIST 9  27   ZERO P   0 disabled  mA outputs  31    OUTP F   0 pH  ORP   32    BURN 0 off  35    TABLE 21 pt table   see code 31                         IM 12B07D02 01E    10 2 Appendix                                                                                     FUNCTION SETTING DEFAULTS USER SETTINGS   User interface   50  RET  1 on   51  MODE 0 off   52  PASS 0 0 0 all off   53  Err 4 1 1 hard fail   Err 5 1 1 hard fail   Err 4 2 1 hard fail   Err 5 2 1 hard fail   Err 07 1 hard fail   Err 08 1 hard fail   Err 09 1 hard fail   Err 11 0 soft fail   Err 16 0 soft fail   55  CALL M 0 250 days   56  DISP 1 0 01 pH   Communication   60  COMM  1 0 on write ena    ADDR  00 00   61  HOUR   62  ERASE   General   Test and setup mode   80  TEST                         IM 12B07D02 01E                              10 2  Configuration checklist for PH202G    Appendix 10 3                               display calibration log     enabled with logbook    diagnostics setup    Standard Configuration Options Reference for  change  Measured Variable s   primary inputs pH  ORP and Temp  pH range 0 14 pH any span within  2 16 pH    output          range linearized disabled 21 point table codes 31 amp  35  ORP range  500 to 500 mV spans up to 3000 mV between    output      1500 to 1500mV  Temperature range  30 140  C  Temperature unit Celsius Fahrenheit code 11  mA Outputs  analog output 4  20 m
57. 0 01                   ORP Indication Check    Put the instrument in ORP mode by following the steps below     a  Press the     key  The message display will show     SERV        b  Press the  YES  key  The message display will show     CODE    and the data display will  show    00    with the first digit of 0 flashing     Press the  ENT  key   Turn on power again     2079200    press the   gt   key once  The second digit of O flashes   press the     key once  The data display will show  01  with 1 flashing   Press the  ENT  key  The message display will show  PH ORP      Press the     key once  The data display will change from  0  to  1      In this state  change the simulation input of the standard voltage generator as shown in Table 3   The corresponding data must be within the range                 Table 3  Simulation Input  mV  Data Display  mV    1500  1500  1  0 0  1   1500  1500  1                Return to PH mode by following the steps below        then the tests        completed     a  Press the     key  The message display will show     SERV        b  Press the  YES  key  The message display will show                 and the data display will  show    00    with the first digit of 0 flashing     o 7920    Press the  ENT  key     press the   gt   key once  The second digit of O flashes   press the     key once  The data display will show    01    with 1 flashing   Press the  ENT  key  The message display will show  PH ORP      Press the     key once  The data displ
58. 1    Unclassified Location Classified Location    Sensor s  are of a passive type to be regarded as  simple apparatus     devices which  neither store nor generate voltages over 1 5 V  currents over 0 1 A  power over 25 mW or  energy over 20 uJ  or are FM Approvals entity approved and meet connection  requirements     Electrical data of the PH202S F  amp  PH202S P     Supply circuit  Ui   17 5 V  li   380 mA  Pi   5 32 W  Ci   220 pF  Li   0 pH     Sensor input circuit  Vt   14 4 V       32 3 mA  Ca  600 nF  La   34 mH    Any FM Approved FISCO barrier may be used that meets the following requirements   Voc or Vt  lt  17 5 V  loc or It 380 mA         or Pt  lt  5 32W   When installing this equipment  follow the manufacturer s installation drawing    Installation should be in accordance with ANSI ISA RP 12 06 01    Installation of  Intrinsically Safe Systems for Hazardous  Classified  Locations  and the National  Electrical Code  ANSI NFPA 70     Associated apparatus connected to the FISCO barrier must not use or generate more  than 250 Vrms or Vdc     e Resistance between FISCO Intrinsically Safe Ground and earth ground must be less than  1 0 Ohm   e      case of using cable glands in Outdoor location  they shall be UV rated or made of metal     e The FISCO concept allows the interconnection of several 1 5  apparatus not specifically  examined in such combination  The criterion for such interconnection is that the voltage   Vmax   the current  Imax  and the power  Pi  which I 
59. 2B07E01 01E   HA405 Solid electrolyte  Xerolyt        HA406 pH sensor      12  07  02 01     HA406 Solid electrolyte  Xerolyt      with temperature element    DPAS405 pH sensor      12B7G1 01E DPAS405 pH sensor for small culture tanks  DPA405 pH sensor      12  07  01 01     DPA405 pH sensor for chemical process    with temperature element    T2 eps               12B07D02 01E    11 2                     11 2  Wiring diagrams  11 2 1  Example of Non Explosionproof System    WTB10 PH202G PH201G SDBT  Terminal Box pH ORP Transmitter Distributor Distributor       Output Signal   1 to 5V DC     Output Signal  PH8EFP  1 to 5V DC   PH8ERP    PH8EHP                         f           2 L   1000 or less    Hold FAIL WASH    F2 3E eps    11 2 2  Example of Intrinsically Safe Explosionproof System  PH202S  pH ORP Transmitter    pH ORP  Sensor    Safety Barrier Distributor          PH8EFP Output  PH8ERP                       F2 4E eps    Ground to earth     1  Use a 2 conductor shielded cable with an outside diameter of 6 to 12 mm  Shield must be connected to internal terminal G of transmitter  and left unconnected at the other side      2  Transmitter must be grounded using external terminal  for general purpose version ground resistance of PH202G should not exceed 100V   Japanese Class D grounding        N WARNING    Use an appropriate DC power supply  such as from the     201   distributor  for the PH202   transmitter  Under no circumstances should you connect AC power such as 100V A
60. 4  Output function              5  User interface     gt              Model     Manufacturer    Distributor  Tag     Descriptor  Message  Date   Dev id     Write protect  Universal rev  Fid dev rev  Software rev  Hardware rev  Poll addr  Num req preams                                                  SV Meas        SV Param   Z1 setup          1  Z1 Comp  2  Z1 Meas   3  Z1 Err lim                           Z2 setup       L            1  21 low       3   2  Z1 high                                        2  Temp  specific     1           1  22 low       3     2  Z2             gt                Z1  means impedance input1    Z2  means impedance input2                                  Stab     means stability                 1  22 Comp  2  22            3  Z2 Err lim      gt      Cal  Check         1  Aspot chk  2  Slope chk    PV unit Note   Temp  sens  Temp  unit  Temp  comp    Stab  chk      1  Stab  time Note   2  Stab  val    Aspot            1  Aspot  2  Zeropnt   3  Aspot L  4  Aspot H    Slope               1  Slope  2  Slope L  3  Slope H  ITP  Buffers       gt  1  Buffer 1                1         id    2  Buffer1 values         N      Buffer 2                  1  Buf2 id    2  Buffer2 values                          Buffer 3     __            ONS    MA funct   Burn funct   mA table       1  Buf3 id    2  Buffer3 values   gt                    m         Error prog     L      3     eum    Err4 1   Err16          Maintenance 4   5    1  Maint  timer          Display     
61. 41 SEE CONTROL DRAWING  e 0 Ex ia           for      40     Hd      WARNING AVERTISSEMENT  SEE CONTROL DRAWING Substitution of La substitution de composants  CL I  DIV 1  GP ABCD components may peut rendre ce materiel     Ex ia      T4 impair suitability inacceptable pour les  IS CLI  DIV 1  GP ABCD Ex ia IIC T6 for      40     for class 1  Division 2  emplacements de  n  AEx              SEE CONTROL DRAWING Classe I  Division 2   APPROVED Type 4X WARNING                 95 AVERTISSEMENT Q  Install per CONTROL DRAWING ossi NR        2025           024   10    4 1 to P 4 2 intrinsic safety intrins  que   Made in Japan Made in Japan                                           Figure 1 1  Nameplate    IM 12B07D02 01E          1 2 Introduction                                                                                                                                                                                                                                                                                                                                                                                                                                                                      O O  DISSOLVED OXYGEN TRANSMITTER DISSOLVED OXYGEN TRANSMITTER PH ORP TRANSMITTER  MODEL PH202G F MODEL     202      MODEL PH202S K  SUFFIX SUFFIX SUFFIX  SUPPLY 2406      SUPPLY   9T032VDC         SUPPLY   9T032VDC      OUTPUT 4 20mA 06       OUTPUT   FF TYPE113 OUTPUT   PROFIBUS PA AMB TEMP   10 
62. 5 V       22 nF  Li 35 uH           1 2 W  Ci 22 nF  Li 35 uH               mme iid                            CENELEC ATEX  KEMA  Type of protection  n   IECEx IECEx Intrinsically safe  see Note  Applicable standard  EN60079 0 2006   CENELEC ATEX  KEMA  Intrinsically safe Approval CENELEC   Certificate                    Applicable standard                 EEX nA  nl        Group  II  Category  3G  Certificate  KEMA OGATEX0218 X Temp  Class  T4  Amb  Temp    10 to 55  C  CENELEC   T6  Amb  Temp    10 to 40  C  ATEX Ex ia       Group  Il  Category  1G  E Ui 31 5 V      22 nF  1 25 pH  Temp  Class  T4  Amb  Temp    10 to 55  C 2 V            6  Amb              10 to 40  C CSA                           safe Approval       Ui 31 5 V  1   100 mA       1 2 W  Ci 22 nF  Li 35      type of protection  n           E A          Applicable standard    22 2       0   1991   1         IECEx Intrinsically safe  see Note  Canadian 22 2  No 04 M2004  C22 2  No 157 M1992   CSA Intrinsically safe Approval Standards C22 2  No 213 M1987    22 2       61010 1  Applicable standard  C22 2  No 0 M1991  Association   Class 1  Division 2  Groups ABCD  C22 2  No 04 M2004  C22 2  No 157 M1992   CSA  Ex nA  nL        Canadian C22 2  No 61010 1 Temp  Class    4  Amb  Temp    10 to 55      Standards   Ex ia Class    Division 1  Groups ABCD  T6  Amb  Temp    10 to 40 C  Association  EX ia               C Ui Vmax  31 5 V  Ci 22 nF  Li 35 uH   CSA  Temp  Class     pre Tee 5 E ji IECEx Type of protection  n 
63. 5 uH       Sensor input circuit  terminals 11 through 17    Maximum output voltage Vt   14 4 V  Maximum output current It   32 3 mA   Maximum allowed external capacitance Ca   600 nF   Maximum allowed external inductance La   34 mH    If Hand Held Terminal  HHT  is not connected to the power supply lines of the PH202S   see figure 1    Any FM Approved barrier or power supply may be used that meets the following requirements               Vt    31 5V  IscorIt  lt  100 mA        gt 22nF Ccable  La  gt    5       Leable  If HHT is connected to the power supply lines of the PH202S  see figure 2    The Hand Held Terminal must be FM Approved  Refer to the manufacturers control drawing of  the HHT and the barrier power supply to determine the cable parameters    Voc or Vt    VHHT  lt  31 5      Isc or It    IHHT  lt  100 mA        222nF   Ccable  CHHT  La  gt    5       Lcable  LHHT  When installing this equipment  follow the manufacturer s installation drawing   Installation should be in accordance with ANSI ISA RP 12 06 01  Installation of Intrinsically Safe  Systems for Hazardous  Classified  Locations  and the National Electrical Code  ANSI NFPA 70    Control equipment connected to the barrier power supply must not use or generate more than  250 Vrms or Vdc       Resistance between Intrinsically Safe Ground and earth ground must be less than 1 0 Ohm     In case of using cable glands in Outdoor location  they shall be UV rated or made of metal   WARNING    Substitution of compone
64. 60   enabling the PH202 to communicate with the PH201G  Style B     In addition  set the wash function on  enabling wash function   WASH ON  on  parameter setting    In this time     on the LCD shows because  Wash  isn t made      MODE   of the text plate     IM 12B0702 01E    11 4 4  Setting of errors  FAIL contact   Code 53  Err 4 1  Err  16  9 items      Code  Display    User interface    Appendix 11 9    These set how errors are notified to the user  Either of two modes can be set   Hard  Fail  mode lights the  FAIL  indicator on the LCD panel  When  enable status output  contact  is set for the PH201G distributor  Style B   the FAIL contact remains closed and  the  wash  contact stops operating  If Burnout functions are enabled by Code 32  then    the output becomes Burnout Upscale or Burnout Downscale      Soft Fail  mode blinks the  FAIL  indicator on the LCD panel  When  enable status  output contact  is set for the PH201G distributor  Style B   the FAIL contact pulses and  the  wash  contact and maintenance contact continue to operate normally  In  Soft Fail   mode  notification that  periodic maintenance is due  still functions  service code 55  so      for this reason     it is better to set  Soft Fail  mode than  Hard Fail  mode     Function    Function detail    Ddefault values       53        Err 4 1     Err 07     Err 08     Err 09       Error setting       Impedance low  input 1   Impedance low  input 1   Impedance higt  input 1   Impedance high  input 1   Imp
65. 9315CN 1 Epoxy polyester Coating  2     Internal Works Assembly with amplifier assembly    1 For mA   HART     1                 1 For Profibus     1        mA   HART  Non incendive   3   Housing Assembly      1 Polyurethane Coating  mE 1 Epoxy polyester Coating  4 L9811FV 2 Cable Gland Assembly  5    1 Flat Cable  6   1 Pin Terminal Unit 3 terminals type  8     2 Stud  9    Analog Board Assembly      1                             1        FF Profibus     1        mA   HART  Non incendive   10    1 Digital Display Board     1                            1 For FF Profibus  11    1 Screw Assembly to fix amplifier  12 K9660YP 1 Stainless steel screw asssembly to fix cover  13    Adapter Assembly  K9414DH 1 For G1 2 screw when  AFTG specified  2 units    K9414DJ 1        1 2       screw when  ANSI specified  2 units    14     Comm  Board Assembly      1 For FF      1 For Profibus    Copyright 2008  1st Edition  Feb  2008        CM PL 1 2B07D02 23E  YOKOGAWA  lt  gt  Subject to change without notice  2nd Edition   Nov 2008  YK     Yokogawa Electric Corporation    Pipe Wall Mounting Hardware  Option Code    U     1 ae                               Panel Mounting Hardware   Option Code    SCT     Hood to sun protection                                Option Code             Item Parts No  Qty Description  1 K9171SS 1 Universal Mount Set   U   2 K9311BT 1 Tag Plate   SCT   3 K9311KA 1 Fitting Assembly   PM   4 K9311KG 1 Hood Assembly   H   K9660JA 1 Hood Assembly   H2        CMPL 12B07
66. A                     Use   gt    A  keys to    adjust and enter manual temperature setting                                                                                                       YES              XH Dil              II TINI  Hi Li X YES  NO        LIIN                                        V             YES             XWASH                F5 2 3E  eps                         The contact output for wash can be controlled by sending wash siganals  This function is available  with the PH201G  style B   Control the contact output in combination of the PH202 and the PH201G   style B   Set communication setting of the PH201G at the Service Code 60 of PH202    You can individually set the wash period  interval between wash cycles   tl  the wash time  length of  the wash phase   tw  and the recovery time  tR     IM 12B0702 01E    pH    Appendix 11 7    You can individually set the wash period  interval between wash cycles   tl  the wash time  length of  the wash phase   tw  and the recovery time  tR  During the recovery phase of the wash cycle  the detector checks if the sensor response is sluggish      the detector checks if the time to recover half the wash cycle pH change  1 2ApH  is within one third  of the recovery time  1 3tR  setting    to determine if the sensor requires manual cleaning or is near  the end of its useful life  Determine the recovery time setting accordingly  To disable this check  you  need to set service code 60 accordingly   The wash cycle p
67. A for pH pH ORP  parameter 2  code 01  02  31  output linearization disabled pH ORP codes 35  Communication  digital interface disabled HART 60  communication software external HHC or PC202 contact factory  variables on display pH ORP and temp pH ORP  parameter 2  mA output    display     SL        21  22 etc    burn out disabled burn low  3 6 or 3 9   high  21  on mA output   code 32  password protection disabled for maint  comm   serv level code 52  autoreturn return to measure      10 min    enable or disable code 50  add  functions in MAINT disabled Impedance check start code 51  Diagnostics  impedance checking active enable or disable code 03  amp  04  check on calibration data active enable or disable code 05  check on stability 0 02 pH per 5s choose stability level code 20    codes 03  04  amp  05          other sensors  2nd parameter  manual temp  comp     enamel sensors  Pfaudler   disabled  disabled    ITP  amp  impedance check setup  pH  amp  ORP  pH  amp rH  disable or enable    Compatibility  pH or ORP glass sensor metal electrode   pH or ORP code 01  temperature sensor Pt 10000 Pt1000  Pt100  etc code 10    codes 23  03  amp  04  code 02   temp        Special Features       buffer table configuration  temperature calibration   zero point calibration   call for maintenance   HOLD during maintenance  process temp  compensation  logbook          NIST standard  none  disabled    disabled  disabled       fully configurable   adjustment     20        disable or enable
68. C or similar AC  power supply line  To measure pH or ORP in hazardous locations  use the     2025 or PH202SJ with  intrinsic safety barriers     Grounding   Be sure to ground the transmitter by using the ground terminal on its case   Connect the G terminal inside the transmitter  to the shield wire of two core shield cable which is  conneced between the distributor and transmitter   For the PH202G transmitter  this does not apply to the     2025  if you cannot ground the     terminal on the transmitter case then connect this G terminal to the shield of the two wire cable  connecting the transmitter and distributor  and ground it at the distributor end     11 3  Sensor wiring  Refer to Figure 11 1 and 11 2  which includes drawings that outline sensor wiring   The     202 can be used with a wide range of commercially available sensor types  both from  Yokogawa and other manufacturers  The sensor systems from Yokogawa fall into two categories  the  ones that use a fixed cable and the ones with separate cables   To connect sensors with fixed cables  simply match the terminal numbers in the instrument with the  identification numbers in the instrument on the cable ends   The recommended procedure is to color code each end of the cables to match the sensors with the  color strips provided with each cable  This provides a quick way to identify the ends of the cables  belonging to a particular sensor when they are installed    The procedure for fixing the identification labels is des
69. CATION           INPUT VOLTAGE FTT Sii RE   mV  REFERENCE ACTUAL ERROR                       ORP INDICATION  INPUT VOLTAGE    SERI RE   mV                        er          o o      NOTES  Att            DATE AMBIENT TEMP   amp  HUM        amp     BAA KBE  INSPECTOR APPROVED BY  YOKOGAWA  9 QIC 12B07D02 01  Ed6  Mar 2007    IM 12B07D02 01E                     12 5    12 2      202  2 Wire pH ORP Transmitter                    202     Inspection    2 Wire pH ORP     Standards ire pH O ransmitter    1  Scope  This inspection standard applies to the PH202S 2 Wire pH ORP Transmitter     2  Inspection Items    2 1 Insulation resistance test   2 2 Dielectric strength test   2 3 Current output test   2 4 Temperature indication check   2 5 pH indication check   2 6 ORP indication check    Note  Items marked with an asterisk     may only be confirmed by a test certificate     3  Inspection Methods  Standards and Conditions  e Connect the testing circuit as shown in Figure 1  Allow the instrument to warm up for at least  5 minutes before conducting the tests  For the connections for the insulation resistance test   follow the instructions in Section 3 1 and for the connections for the dielectric strength test   follow the instructions in Section 3 2     e Performance tests should be done in the inspection mode where the tests from Section 3 3  through Section 3 6 take place in sequence and cannot be retraced  If the reconfirmation of  a test is needed  turn off the power to the transm
70. Code   Display Function Function detail Default values       Temperature measuring and compensation functions       10  T SENS Temperature sensor   Pt1000 0 Pt1000  Pt100  3kBalco  5k1  8k55  350   6k8    PTC10k       11  T UNIT Display in   C or   F   C 0         F     gt  OJIN OO A OO N o       12  T ADJ Calibrate temperature   Adjust to allow for cable resistance None       13  T COMP Set temp comp Compensation for process changes off 0 0 Off  Compensation for process changes on 1   T COEF Adjust process TC Set for TC in pH per 10   C  0 00   pH per  10  C       14 19 Not used                                     IM12B07D02 01E    Parameter setting 5 15    5 3 3  Calibration functions    Code 20    Code 21    Code 22     ALSEC  amp    APH    These functions are used to determine the stability level demanded by the  EXA as acceptance criteria for the automatic calibration  for general purpose  electrode systems with a fast response  Where heavy duty electrodes are  used  or when low temperatures are concerned  these values should be adjusted   When adjusting these settings  the longer the time interval and the smaller the  pH change  the more stable will be the reading  However  it is important to  bear in mind that the time taken to reach stability is an exponential function   and too ambitious a setting will cause the instrument to wait for a very long  time before accepting a calibration      AS LOW  amp  Limit values for the drift of an electrode system before an error i
71. D02 01E    3 4 Installation and wiring           Hand Held  Communicator                            Sensors       Input         2 5or10m       CURRENT  OUTPUT    Recorder Safety Barrier  PH2028 only ref pH ORP emp                    Figure 3 6  System configuration    3 3  Wiring of sensors    3 3 1  General precautions  Generally  transmission of signals from pH sensors is at a very low voltage and high impedance level   Thus a lot of care must be taken to avoid interference  Before connecting sensor cables to the transmit   ter make sure that next conditions are met        the sensor cables are not mounted in tracks together with high voltage and or power switching   cables       only standard coaxial electrode cables or extension cable are used       the transmitter is mounted within the distance of the sensor cables  max  10 m        the setup is kept flexible for easy insertion and retraction of the sensors in the fitting     3 3 2  Additional precautions for installations in hazardous areas  Make sure that the total of capacitance and inductances connected to the input terminals of the EXA      202  do not exceed the limits given in the certificate   This sets a limit to the cable and extensions used       The intrinsic safe version of the PH202S instrument        be mounted in Zone 0 or 1  PH202S B   N    D        The sensors can be installed in Zone 0 or Zone 1 if a safety barrier according to the limits given in  the system certificate is used       Ensure that the t
72. D02 23E 2nd Edition   Nov 2008  YK     Revision Record    Manual Title      Manual Number      Edition  1st  2nd    3rd    4th    5th    eth    7th    8th    9th    10th    Model PH202G  Style   3       202   Style   3     2 wire Type pH ORP Redox  Transmitter    Date  Apr  2000  May  2002    Feb  2004  May  2004  Dec  2004  Dec  2005  Nov  2006    Mar  2007    Oct  2007    Apr  2008    IM 12B07D02 01E    Remark  s    Newly published   CENELEC Certification changed to ATEX  KEMA   CSA Certification added   Apart from a general rewite of English    Style changed to S2    HART Communication function is additionally described    TIIS intrinsic safe version PH202S  Style    1  added  etc    Write in some error correction    Model name of TIIS intrinsic safe version PH202S  style    1  changed to  PH202SJ  style    1     All over revised     style of PH202G changed to S3  and PH202S changed to S2     PREFACE 1  Some of warning description modified  P  1 1  Some of  nameplate in Figure 1 1 changed  P  1 2  Some of nameplate in Figure 1 2  changed  P  2 2  EN 61000 3 3 deleted from  N  Regulatory compliance   P   2 3  Certificate no  of CENELEC ATEX and IECEx Scheme explosionproof  added  CSA explosionproof description added  P  2 4  CSA explosionproof  description added  P  2 5  P  layout changed  P  2 6  Miss spering of Model  and suffix code corrected  P  2 9  Control Drawing for mA HART Specification   FM Intrinsically safe design  corrected  P  2 10  Control Drawing for mA  HAR
73. E   The special gland is intended to be used to seal the multiple cables from the Yokogawa flow  fittings such as FF20 and FP20  The designated cables are WU20 sensor cables  which are  approximately 5 mm  0 2   in diameter  and 82895002 liquid earth cables  which are approximately  2 5 mm  0 1   in diameter     For sensor systems using a single cable  like the FU20  FU25  and the PR20  PD20  PF20 and PS20     the standard gland will accommodate the cable adequately  Single cables between approximately 6 mm  and 12 mm  0 28  and 0 47   can be sealed properly with these glands     IM 12B07D02 01E    Installation and wiring 3 11    3 6 3  Sensor cable connections using junction box  BA10  and extension cable  WF10     For the WTB10 terminal box  see APPENDIX     Where a convenient installation is not possible using the standard cables between sensors and transmit   ter  a junction box and extension cable may be used  The Yokogawa          junction box and the WF10  extension cable should be used  These items are manufactured to a very high standard and are neces   sary to ensure that the specifications of the system are not compromised  The total cable length should    not exceed 50 metres  e g  5 m fixed cable and 45 m    extension cable   In the case of systems using    dual high impedance sensors  e g  Pfaudler 18   then the cable length is restricted to 20 metres  fixed    cable only  no extension with WF 10              4            Screen    White Co axial cable          
74. ES        NEW SNS  Display flashes YES NO  answer YES if new sensor fitted or NO if not   Care  YES resets logbook calibration data              Insert the sensors in pH 7 buffer solution   Press YES to start calibration                                                                    Note  To start calibration with another solution  press NO to cycle through the other options    CAL7        CAL4        CAL9  NO  CAL      gt                              The instruments waits for  the reading to stabilise    The pH display flashes   When reading Is stable   the CAL END message  appears           Press YES for single point   As Pot  adjustment  The sinal  Press NO to proceed to calibrate Slope     singie     point  YES  NO          calibration  is now  complete   put sensors  back in the  process and  press NO to  switch off  HOLD and  return to  measuring  mode                                                                           YES    NO    Transfer to pH 4  buffer and press YES                       The instrument then waits for the reading  to stabilise   The pH display flashes      When the reading is stable  the CAL END  message appears  Press YES   WAIT flashes briefly then HOLD       The calibration is now complete  Put the sensors back in the process  and press NO to turn off HOLD and return to the measuring mode              IM 12B07D02 01E    6 4 Calibration    6 5 3  Manual calibration  2nd parameter calibration        12B6C3 32    Press the MODE key  The leg
75. H change is ApH  the difference between the pre wash pH and the pH during the  wash   If the difference between the normal pH value in process and the pH during the wash is small  then  half the difference  1 2ApH  will be too small to measure reliably  so this check should be disabled in  this case  The monitoring of waste water pH is an example of such an application   If you select water jet cleaning in such a case  the normal pH as well as the washingtime pH will  be around pH7  the difference will be approximately zero  too small for the recovery time check to  measure  so the recovery time check will not work normally and should be disabled    1  Wash Period    Good Electrode           1 2ApH      VIR           W  Wash Time   tR   Recovery Time                        Time           tl   Wash Period  0 1 to 36 0 hours  default 6 0 hours   tW   Wash Time  0 1     9 9 minutes  default   0 5 minutes         Recovery Time  0 1     9 9 minutes  default   0 5 minutes  F5 0 eps    IM 12B07D02 01E    11 8 Appendix    11 4 3  How to output manual wash signal                 gt  A ENT  CoD CD C                                                                                    MODE                                                                           1  J YES                                                   T5 1 5E eps    Note   Wash functuon is only avairable when using the PH201G style B  distributor   In using the function  it is necessary to set 2 0 or 2 1 at service code 
76. ITTER PH ORP TRANSMITTER  MODEL PH202 MODEL PH202S E  SUFFIX SUFFIX  SUPPLY               OUTPUT 4 20mADC                            10  55  C  STYLE  SUPPLY   24VDC     No   OUTPUT 4 20mADC           No  IECEx KEM 06 0052X  AMB TEMP    10   55     Zone 0 Ex la IIG T6 for Ta   0 c  STYLE SEE CONTROL DRAWING  No           06        0218 X  No  DLE D  SEE CONTROL DRAWING IP65  YOKOGAWA   YOKOGAWA      Made in Japan Tokyo 180 8750 JAPAN                   Made in Japan Tokyo 180 8750 JAPAN                                                                                                                                                                                                              nz200  PH ORP TRANSMITTER PH ORP TRANSMITTER O    No  IECEx KEM 06 0052X  MODEL PH202S U peto     2025    Ex nA nL       T4  Ex nA nL  IIC T6 for      40      SUFFIX    65  SEE CONTROL DRAWING  No           06        0219        EEx nA nL  IIC   4      36  EExnA nL IIC T6 for      40     IP65  SEE CONTROL DRAWING  NI CL I  DIV 2  GP ABCD AND  SUPPLY            SUPPLY 2406     CLI  ZN 2  GP IIC  T4  OUTPUT   4 20mADC     OUTPUT   4 20mADC       Type 4X  E 0 APPROVED Install per CONTROL DRAWING  AMB TEMP     10  55  C AMB TEMP  10  55  C       024   10 P 4 3 to P 4 4  STYLE STYLE   Ex nA nL      No NI CL I  DIV 2  GP ABCD  N    z T4    No  IECEx KEM 06 0052X 5 Be T6 for Ta 40  c  No  IECEx KEM 06 0052X Zone 0 Ex ia      T4 IP65 Type 3S  Zone 0 Ex ia      T4 Zone 0 Ex ia      T6 for Ta 40 C LR817
77. Li 0 uH       oy FM Non incendive safe Approval  CENELEC ATEX            Intrinsically safe Approval 2  10     e 0     Applicable standard      60079 0      50020 G ABCD  2 2                     EN60079 26      60079 27     POURS p Aone              Certificate           07        0048                    FISCO Ex ia       Group      Category  1G on incendive Safe Apparatus Parameters  Temp  Class  T4  Amb  Temp    10 to 55  C Entity   Vmax 32 V  Pmax 1 2 W   Ui 17 5 V  li 380 mA  Pi 5 32 W  Ci 220 pF  Li 0 uH    uH         CENELEC                     Type of protection                Ciz220 pF           O  Applicable standard      60079 0 2006   B        60079 15 2003 FM EPS  CENELEC   Certificate           07ATEX0049 m  ATEX EEx nA  nL        Group  Il  Category  3G Item Description Code  Temp  Class  n Amp     gs    pea CSA Intrinsically safe Approval  Ui 32 V       220 pF  Li 0 pH D Applicable standard  C22 2  No  0   1991          22 2        04   2004    22 2        157   1992   P  ATEX EPS   22 2  No  61010 1  Item Description Code E    m l  Division 1  Groups ABCD   r  LE 2           Temp  Class    4  Amb  Temp    10 to 55  C  pplicable standard   0   11            IECEx       60079 26 Canadian   Entity TEES ME EUN d  Scheme   Certificate  IECEx        07 0026X Standards Ui Vmax  17 5 V  l  max  380 mA   Entity Zone 0 Ex ia       P Association   FISCO Pi Pmax  5 32 W       220 pF  Li 0 uH  Temp  Class  T4  Amb  Temp    10 to 55      CSA           Ui 24 V  li 250 mA    
78. MINAL G   STRENGTH 600V AC 2 SEC         2025022          OI       EP    A Sint  G  600V      2 fl   ONLY FOR BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G     RERE INSPECTION ITEM   5 R RESULT        2025  600V      2 SEC          11 17   4       600V AC 2T   R  BETWEEN INPUT TERMINALS 11 TO 17  AND POWER SUPPLY TERMINALS       600V AC 2 SEC   E IR BER  0 3     ACCURACY             TEMPERATURE INDICATION    C     eee M   T                       14983   1  90                 A   001     ACCURACY   NIIT  pH INDICATION ANBE INDICATION  pH     INPUT VOLTAGE EX T                                               ACTUAL I                4141 LI   LLLI    ORP INDICATION          6 7  INDICATION  mV     INPUT VOLTAGE BAG MME Re  REFERENCE ACTUAL ERROR     1500                    D2 47          PROFIBUS PROFIBUS FUNCTION CHECK  FUNCTION    NOTES    Aft      SE   DATE AMBIENT TEMP   amp  HUM          LI   INSPECTOR APPROVED BY       QIC 12B07D02 71  YOKOGAWA 4 Ed1Sep  2007    HVI IADU JUZU IL       Customer Model PH202G  Style    3     L   pH ORP Transmitter  Parts List                         ltem Part No  Qty Description       1     Cover Assembly  K9315CA 1 Polyurethane Coating  K9315CN 1 Epoxy polyester Coating  2     Internal Works Assembly with amplifier assembly  K9661AD 1 For mA   HART  K9661AE 1 For FF  K9661AF 1 For Profibus  3     Housing Assembly  K9661HA 1 Polyurethane Coating  K9661HB 1 Epoxy polyester Coating  4 L9811FV 2 Cable Gland Assembly  5 K9660AQ 1 Flat Cabl
79. NALS      AND EARTH TERMINAL G  EN  STRENGTH 600V AC 2 SEC       2025004         11 17                F  G  fil 600V AC 2   Fi   ONLY FOR BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G       2025  600        2 5      AJ  11717       amp      nF        600   AC 2   fil  BETWEEN INPUT TERMINALS 11      17  AND POWER SUPPLY TERMINALS       600V AC 2 SEC    im E AIR        03  C ACCURACY    0 3  C  TEMPERATURE       amp  x INDICATION           INDICATION RESISTANCE        REFERENCE ZEB ACTUAL      ERROR     Q   0        9216      1097   1498 3   no           pH ACCURACY    0 01pH  pH INDICATION ABE  amp     INDICATION  pH     INPUT VOLTAGE Re AE         L            90  14      4 7111 1     42                               1mV ACCURACY    1mV  ORP INDICATION      X    INDICATION  mV   INPUT VOLTAGE EETA ESMA               400                     e             1500            74                     24                        FIELDBUS FIELDBUS FUNCTION CHECK  FUNCTION    NOTES    Att FARE RE   DATE AMBIENT TEMP   amp  HUM   REF RBA   INSPECTOR APPROVED BY       QIC 12B07D02 61  YOKOGAWA        1        2007    IM 12B07D02 01E                     12 13    12 4      202    PH202S 2 Wire pH ORP Transmitter  Profibus Communication        Quality PH202G  PH202S  Inspection 2 Wire pH ORP Transmitter  Standards  Profibus Communication     1  Scope    This inspection standard applies to the PH202G and     202  2 Wire pH ORP Transmitters   Profibus specification      2  Inspection It
80. Profibus Communication Functional Check    Check for normal function using Profibus equipment specified by Yokogawa     In the tests of Item 3 3 to 3 5  check the communication function using Profibus communication                                                                                                        PH202  SUPPLY             INPUT2 INPUT1    1      G 11 12 14 17 13 LOW IMP HIGH IMP 15 16     O O O                   O 9 9       Shorting Bar          o   Ground     O d                     Decade Standard  Resistance        Voltage Source         ANN   e        Dc  DA PA Coupler Source     R     24VDC  AAA e       R 502  1 Q  C 2uF 40 4 uF     _                                                                Profibus Equipment Specified by YOKOGAWA     Figure1 Testing Circuit and Test Equipment       QIS 12B07D02 71E          IM 12B07D02 01E    12 16                   K         TEST CERTIFICATE    RAR               5 BINo   PRODUCT NAME pH ORP TRANSMITTER TAG NO      2   MODEL PH202    F         HEES  ORDER NO  SERIAL NO     APPEARANCE    ie d SURE          507   5  GO 100M Q         500V DC  INSULATION BETWEEN POWER SUPPLY TERMINALS      AND EARTH TERMINAL G   RESISTANCE 100M        MORE   500V DC  Arg T 01017  38607    A Sint      100M9 OR MORE 500V DC  BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G   10049 OR MORE 500V DC                          amp  amp   7                   600V      2 0 RA  DIELECTRIC BETWEEN POWER SUPPLY TERMINALS    AND EARTH TER
81. S  apparatus can receive and remain  intrinsically safe  considering faults  must be equal to or greater that the voltage  Voc  Vt    the current  loc  It  and the power  Poc  Pt  which can be providede by the FM approved  FISCO barrier  In addition  the maximum unprotected residual capacitance  Ci  and  inductance  Li  of each apparatus  other than the terminator  connected to the Fieldbus  must be less than or equal to 5 nF and 10 uH respectively     IM 12B07D02 01E    Specification 2 17       In each 1 5  Fieldbus segment only one active source  normally the FM Approved FISCO  barrier  is allowed to provide the necessary power for the Fieldbus system       other  equipment connected to the bus cable has to be passive  not providing energy to the  system   except to a leakage current of 50     for each connected device  Seperately  powered equipment needs a galvanic isolation to insure that the 1 5  Fieldbus circuit  remains passive     e The cable used to interconnect the devices needs to comply with the following  parameters   Loop resistance R   15     150 O km  Inductance per unit length L   0 4     1 mH km  Capacitance per unit length C   80     200 nF km   C         line line   0 5       line screen if both line are floating           C    line line   C    line screen if the screen is connected to one line   Length of spur cable  max  30m  Length of trunk cable  max  1 km  Length of splice   max  1       WARNING    Substitution of components may impair Intrinsic Safet
82. Service Code 53 to    0    for    soft fail     to disable   Fail contact output  When you set Service Code 53 to    1    for    hard            set Service  Code 60 to    2 0    to enable Fail contact output of     201    Style B   or set Service  Code 60 to    0 1    to disable Fail contact output of PH201G  Style          ADDR  Select address 00 for point communication with 4 20 mA transmission  Addresses  01 to 15 are used in multi drop configuration  fixed 4 mA output      Display   Function Function detail Ddefault values    Communication       60  COMM   Communication Set communication Off  Set communication On  write enable  write protect    Set communication PH201G B    Without half time check by setting 2 0  With half time check by setting 2 1  Network address Set address 00 to 15                                   Valid for instrument using software version 3 1 or later  T5 3 6 eps  For earlier versions the default is 0 1     IM 12B07D02 01E    11 6 Appendix    11 4 2  The setting of sending WASH signal   WASH      WASH Commissioning mode has the setting ability of sending signal for WASH output  contact to the PH201G distributor  style B   Converning to On Off WASH selection  when  selecting  On   wash period  wash time and retention time  waiting time after wash   should be set           To apply wash function  use the PH201G  styleB    and    2 0    or  2 1  at the service code 60 must be set                                      C5 C2 C5  EXA              YOKOGAW
83. T Specification  FM Non incendive design  corrected  P  2 11  Control  Drawing for mA HART Specification  CSA  corrected  P  2 12  Control Drawing  for FF PB Specification  IECEx  corrected  P  2 13  Control Drawing for   FF PB Specification  ATEX  corrected  P  2 14  Control Drawing for FF PB  Specification  FM  corrected  P  2 16  Control Drawing for FF PB Specification   FM Intrinsically safe FISCO  corrected  P  2 18  Control Drawing for FF PB  Specification  FM Non incendive Entity  corrected  P  2 19  Control Drawing  for FF PB Specification  FM Non incendive FNICO  corrected  P  2 20  Control  Drawing for FF PB Specification  CSA  corrected  P  11 1  11 5  11 6  11 11   11 13 Some error corrected  Sec  12 APPENDIX 3 QUALITY INSPECTION  added  CMPL 12B07D02 03E   22   revised to 2nd edition because some part  no  changed    Style of PH202S changed to 53 and related description changed as follows   P 2 6  Style of PH202S changed to 53 for FM approval  P 3 1  Some of  dimensions in Figure 3 1 corrected  P 10 7  Appendix   Note of HART protrocol  DD files URL added  CMPL 12B07D02 23E 1st edition added for     2025 style  S3     IM12B07D02 01E    Edition Date  11th Oct  2009    IM12B07D02 01E    Remark  s    PREFACE   Zone 0  added to Warning label explanation  P 1 2  Name plate of  PH202S K  NEPSI  added to Figure 1 2  P 2 2  Some revision of N  Regulatory  compliance  description for EMC revised   P 2 3 to 2 4  Some revision of  IECEx Intrinsically safe description   Zone
84. TROL DRAWING  NI CL I  DIV 2  GP ABCD AND  CL I  ZN 2  GP       T4  SUPPLY 97032 0      SUPPLY   9T032VDC                     OUTPUT   FF TYPE 113 OUTPUT   PROFIBUS PA UMP                      AMB TEMP    10   55  C                    10  55  C Ex nA nL        STYLE STYLE NI CL I  DIV 2  GP ABCD  No No       T6 for      40          1  65          35       SEE CONTROL DRAWING  AVERTISSEMENT  YOKOGAWA                            of La subttuion de composants  Made in Japan Tokyo 180 8750 JAPAN Made in Japan Tokyo 180 8750 JAPAN noair                          ponr les       O  nz00                 200 ee Classe   Division 2           Figure 1 2  Nameplate    IM 12B07D02 01E       PH202S B  D                   Introduction 1 3    NOTE  Check that all the parts are present  including mounting hardware  as specified in the option  codes at the end of the model number   For a description of the model codes  refer to Chapter 2 of this manual under General  Specifications     Basic Parts List  Transmitter PH202  User   s Manual English  Optional mounting hardware when specified  See model code     NOTE  mounting screws and special grommet are packed in the terminal compartment  together with a  second link for impedance selection     1 2  Application   The EXA converter is intended to be used for continuous on line measurement in industrial installations   The unit combines simple operation and microprocessor based performance with advanced self diag   nostics and enhanced communicatio
85. a   ble measuring loop     3 3 5  Access to terminal and cable entry   1  To access terminals remove the front cover of the EXA PH202S by releasing the 4 captive screws    2  Thread the sensor cables into the connection space and connect the cables to the terminals as  indicated in the wiring diagram  Make sure all connections are firm and do not touch each other     3  Screw the gland securely and tighten it to keep out moisture  DO NOT use a wrench to tighten the  nut   4  The optional hose connection is used to guide the cables coming from an immersion fitting through a    protective plastic tubing to the transmitter     3 4  Wiring of power supply    3 4 1  General precautions    WARNING    Do not activate the power supply yet  First make sure that the DC power supply is according to the spec   ifications given     DO NOT USE ALTERNATING CURRENT OR MAINS POWER SUPPLY       The cable leading to the distributor  power supply  or safety barrier transports power to and output sig   nal from the transmitter  Use a two conductor shielded cable with a size of at least 1 25 mm and an  outside diameter of 6 to 12 mm  The cable gland supplied with the instrument accepts these diameters   The maximum length of the cable is 2000 metre  or 1500 metres when using the communications  This  ensures the minimum operating voltage for the instrument     Grounding        If the transmitter is mounted on a grounded surface  e g  a metal frame fixed in the soil  the shield of  the 2 wire cable
86. acitance        22 nF Effective internal inductance Li   35 uH     Sensor input circuit  terminals 11 through 17    Maximum output voltage       14 4 V  Maximum output current It   32 3 mA   Maximum allowed external capacitance Ca   2 294      Maximum allowed external inductance La   64 96mH               Hand Held Terminal must be FM Approved in case it is used in the classified location   When installing this equipment  follow the manufacturers installation drawing  Installation shall  be in accordance with Article 501 4 B  of the National Electrical Code   Non incendive field wiring may be installed in accordance with Article 501 of the National  Electrical Code       Grounding shall be in accordance with Article 250 of the National Electrical code     Incase of using cable glands in Outdoor location  they shall be UV rated or made of metal     WARNING    Substitution of components may impair suitability for Division 2    Do not remove or replace while circuit is live unless area is know to be non hazardous    Explosion Hazard     Do not disconnect equipment unless area is know to be non hazardous    Do not reset circuit breaker unless power has been removed from the equipment or the area is  know to be non hazardous    Application Doc  No   IKE024 A10 P 4 3 to P 4 4    IM 12B07D02 01E    2 8  Control Drawing of PH202S mA HART   Specification  CSA      Specification 2 11       Intrinsically safe design    CSA certified    CSA Ex ia Class I  Div 1  Group ABCD  T4 for ambien
87. agrams                           11 2  11 2 1  Example of Non Explosionproof System                  11 2  11 2 2  Example of Intrinsically Safe Explosionproof System                                                  11 2  11 3  Sensor WIFIFIE                    11 2  11 3 1  Connection CADIG ics      recte Er eo                   11 3  11 3 2  Sensor cable connection using terminal                      11 5  11 4  PH201G B Dedicated                                            11 5  11 4 1  Communication                           11 5  11 4 2  The setting of sending WASH signal                                   11 6  11 4 3  How to output manual wash                            11 8  11 4 4  Setting of errors  FAIL                                      11 9  11 5  Calibration in                                 11 10  11 5 1 Electrode  Check                                   11 10  11 5 2  Manual                                                                           11 11  11 5 3                                                                                    11 11  1125 4  Data                rti nitent                EE nae dA 11 11  11 5 5  Calibration  ProCedUre   eret rers                                            11 12  11 6  Supplement of troubleshooting                                                    11 13  11 6 1  Error Codes ioi E irre                          E a                           EE 11 13  11 6 2  On line impedance                         1            
88. ameter     pH  ORP    pH       02  PRM 2    Enable 2nd parameter    Off  ORP  rH    Off       03  Z1 CHK       7             7            Impedance check 1    Low impedance limit  x   None  K  M or G    High impedance limit  x None  K  M or G    Low   High   Temp comp off  Temp comp on  Imp check off    Imp check on    Press NO to step through choice of units     press YES to select units  then use the     gt     ENT keys to set the value    Press NO to step through choice of units     press YES to select units  then use the     gt     ENT keys to set the value     gt  OJN  gt       o    High       04  Z2 CHK    ZZ            7            Impedance check 2    Low impedance limit  x   None  K  M or G    High impedance limit  x   None  K  M or G    Low   High   Temp comp off  Temp comp on  Imp check off    Imp check on    Press NO to step through choice of units     press YES to select units  then use the     gt     ENT keys to set the value    Press NO to step through choice of units     press YES to select units  then use the     gt     ENT keys to set the value    0 0 1    Low    Off    1000    200            05  CAL CK    Calibration check    Asymmetry check off  Asymmetry check on  Slope check off    Slope check on    1 1       06 09                Not used                      IM12B07D02 01E       Parameter setting 5 13    5 3 2  Temperature compensation and measuring functions     Code 10    Code 11    Code 12    Code 13                5     T UNIT     T ADJ             
89. annin   saksan  ja  ranskankielisina  Mikali tarvitsette Ex  tyyppisten  tuotteiden ohjeita omalla paikallisella kielell  nnne   ottakaa yhteytta lahimpaan Yokogawa toimistoon  tai  edustajaan        gt     Todos os manuais de instru    es referentes aos  produtos Ex da ATEX est  o dispon  veis em Ingl  s   Alem  o e Franc  s  Se necessitar de instru    es na  sua l  ngua relacionadas com produtos Ex  dever    entrar em contacto com a delega    o mais pr  xima  ou com um representante da Yokogawa     Tous les manuels d instruction des produits  ATEX Ex sont disponibles en langue anglaise   allemande et frangaise  Si vous n  cessitez des  instructions relatives aux produits Ex dans votre  langue  veuillez bien contacter votre repr  sentant  Yokogawa le plus proche     Alle Betriebsanleitungen f  r ATEX Ex bezogene  Produkte stehen in den Sprachen Englisch   Deutsch und Franz  sisch zur Verf  gung  Sollten  Sie die Betriebsanleitungen f  r Ex Produkte in  Ihrer Landessprache ben  tigen  setzen Sie sich  bitte mit Ihrem   rtlichen Yokogawa Vertreter in  Verbindung     Alla instruktionsb  cker f  r ATEX Ex  explosions   s  kra  produkter   r tillg  ngliga      engelska  tyska  och franska  Om Ni beh  ver instruktioner f  r  dessa explosionss  kra produkter p   annat spr  k   skall Ni kontakta n  rmaste Yokogawakontor eller  representant            Ta eyyerpi  ra Aevrovpyytas                         pe ATEX Ex                         ota Ayyhik    leppavik   kar                   
90. apter 5 of this manual             Most Yokogawa sensor systems use an Iso thermal point  ITP  of pH7 and    zero point at pH7   This is the default condition for which the transmitter is set  It is only necessary to consider this  adjustment when installing a system with a different ITP  Antimony systems and Pfaudler probes are  good examples of systems with different ITP values  Service code 23 is used  This also permits the  setting of calibration data for precalibrated sensors       Temperature sensor  The Pt 10000 RTD sensor is now becoming the most commonly used for temperature compensation   The transmitter accepts inputs from several different temperature sensors to suit most sensor  Systems  Service code 10 19 are used to set the temperature parameters and the process  temperature coefficient       Temperature calibration  For best accuracy  the temperature sensor should be calibrated to compensate for connection cable  errors  See Service code 12     e pH Calibration  Traditionally  users select buffer solutions to suit the chosen output range  This is merely a  continuation of the days of analog instruments that used indicators driven by the mA output  With  digital technology  it is better to choose good buffer solutions and make an effective calibration than  to use commercial  adjusted  buffers which may have round number values  but are less effective  buffers with lower buffer capacity  It is for this reason that Yokogawa recommends that the NIST 4   7 and 9 s
91. ata display will show    87    with the second digit of 7  flashing   g  Press the  ENT  key  The message display will show     PASS          2  Setting Password 070  a  Press the   gt   key once  The data display will show    000    with the second digit of 0 flashing   b  Press the      key seven times  The data display will show    070    with the second digit of 7  flashing   c  Press the  ENT  key  The message display will show    HIF      d  Press the  YES  key      3  Checking the date and time  a  Press the  ENT  key  The data display will show the date in day month year order   b  Press the  ENT  key  The data display will show the time in hour minute second order   c  Press the  ENT  key  The message display will show    mA        This is the inspection mode  When the message display shows    mA     check the current outputs     Each time the  ENT  key is pressed  the value on the data display increases in steps of 4 0 like     4 0        8 0    and    12 0     Check the current outputs  Corresponding to the data display  the  current output must be within the range shown in Table 1     When the data display shows    12 0     pressing the  ENT  key causes the message display to  show    RIPPLE    and the data display to remain unchanged  Press the  ENT  key again  The  message display will show    mA    and the data display    16 0                    Table 1  Data Display Current Output  mA   4 0 4  0 02  12 0 12  0 02  20 0 20  0 02                3 3 Temperature I
92. atibility  The PH202G is not intrinsically safe  explosionproof   so never install it in a  hazardous area     Conductivity Use of Distoributor PH201G    transmitter   No use of contact output   Use of contact output        2026 Style     amp  Style    possible   Only Style    possible   Non Explosionproof type      1           For information about instruments related to the     202  refer to the following Instruction Manuals     Manual Name   IM No  O  Instruments mentioned    pH sensor  amp  holders IM 12B7K1 02E PH8ERP KCI Refillable pH sensor  for general purpose IM 12B7J1 01E PH8EFP       Filling type pH sensor  IM 12B7M2 01E PH8HG Guide pipe  IM 12B07M01 01E   PH8HS  PH8HSF Submersion type holder  IM 12B07N01 01E   PH8HF  PH8HFF Flow through type holder  IM 12B07WO03 01E   PH8AX Accessories    ORP sensor IM 12C07J01 01E   OR8EFG KCI filling type        sensor  for general purpose IM 12C04K01 01E   OR8ERG KCI Refillable ORP sensor  IM 12C04W02 01E   OR8AX Accessories  for pure water IM 12B07P01 01E   PH8HH holder for pH sensor for pure water  for fermentation use      Terminal box    IM 19D01B01 01E   WTB10 PH1  WTB10 PH2 Terminal Box  Ultrasonic oscillator IM 19C1B3 01E PUS400G Ultrasonic Oscillator   IM 12  502    PH8USF  Explosionproof type  PH201G distributor  Style B IM 19B01E04 02E   PH201G  style B  Distributor  SDBT  SDBS distributor IM 1B4T1 01E SDBT  SDBS Distributor  Attachment rack instrument IM 1B4F2 01E Instruments for rack attachment  HA405 pH sensor IM 1
93. ay will change from  1  to    0        QIS 12B07D02 61E       IM 12B07D02 01E                           12 11       3 6 Fieldbus Communication Functional Check    Check for normal function using Fieldbus equipment specified by Yokogawa     3 3                                                                                           PH202  SUPPLY   LE NPUT2 INPUT         G 11 14 17 13 LOW IMP HIGH IMP 15 16  9       C                                       pD Shorting Bar  U              o  Ground  C O O  2    Standard  Decade Voltage Source  Resistance Box        R C         509  1 Q  C 2uFXx0A4uF     R R    L 5mH  5   RL 50Q  1         gd  Ground  RL  r5       DC  Source  24VDC  Fieldbus Equipment Specified by YOKOGAWA   Figure 1 Testing Circuit and Test Equipment                            QIS 12B07D02 61E          IM 12B07D02 01E    12 12                            TEST CERTIFICATE    meh pH ORP1zi  28 47 No   PRODUCT NAME pH ORP TRANSMITTER TAG NO   T   MODEL PH202         Hees  ORDER_NO  SERIAL NO                  Cd    APPEARANCE  REH                   7                    100M Q         500V DC       R  amp  18 H INSPECTION ITEM      R RESULT    INSULATION BETWEEN POWER SUPPLY TERMINALS      AND EARTH TERMINAL G   RESISTANCE 100M Q OR MORE   500V DC          11917        G  Fl 100      OR MORE 500V DC  BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G   100M Q OR MORE 500V DC  MEE                                600V      2 fi   DIELECTRIC BETWEEN POWER SUPPLY TERMI
94. be set up to hold the last measured value  or a fixed  value to suit the process     Service This selection provides access to the service menu     SERV     What follows are pictorial descriptions of typical frontplate pushbutton sequences for each parameter set     ting function  By following the simple YES NO prompts and arrow keys  users can navigate through the  process of setting range  hold and service functions     IM12B07D02 01E    Parameter setting 5 7    ELN 5 2 1  Output range                 ACI 7  a        ENT                                        EXA PHOS YOKOGAWA                                                   Note  When                  is  enabled      codes 02 and 31   the output range is set        a similar way to pH                                                 X       YES    NO                                                                                                                                                             IM12B07D02 01E    5 8 Parameter setting    5 2 2  Hold                                              YOKOGAWA                                           YES                                    X                               X                         XH T FE                         HOLD deactivated  return  to commissioning menu                          X A       E                                      YES    9       XH                                   HOLD active  last measured  value                       YES               
95. been previously set up in  service code 52 in chapter 5     Automatic calibration See    calibration    section 6    AUT CAL   Manual calibration See    calibration    section 6    MAN CAL   Sample calibration See    calibration    section 6    SAMPLE   Display setting See    operation    section 4    DISP   Manual impedance check See    parameter setting      5 1 4 and   5 3 5 code 51    MAN IMP   Temperature Set automatic or manual compensation and adjust manual reading   TEMP   when pH is set in section 5 service code 01   See adjustment proce   dure in   5 1 1     Set automatic reading  when        is set in Section 5  service code 01    See adjustment procedure   5 1 2     Hold Manually switch on off HOLD  when enabled in commissioning menu   HOLD  section   See adjustment procedure in   5 1 3     IM12B07D02 01E    5 2 Parameter setting    5 1 1  Manual temperature selection and adjustment  pH selected in service code 01                                      Use                                              ENT          keys to    adjust and enter manual temperature setting          IM12B07D02 01E       Parameter setting 5 3    5 1 2  Process temperature measuring in ORP mode  ORP selected in service code 01                          EXAPHEO  YOKOGAMA                                                                                                                                Display return to  measuring mode with  temperature reading           MEASURE                      
96. caused by exposure of the circuitry to condensation     The EXA analyzer contains a logbook feature which needs a clock to provide the timings  The EXA  instrument contains a lithium cell  battery  to support the clock function when the power is switched  off  This cell needs to be replaced at 5 yearly intervals  or when discharged   Contact your nearest  Yokogawa service centre for spare parts and instructions     7 2  Periodic maintenance for the sensor system    NOTE   Maintenance advice listed here is intentionally general in nature  Sensor maintenance is highly  application specific     The sensor system must be kept clean to function well  This may require regular cleaning of the elec   trodes   The effect of dirty electrodes will be to slow the system response and perhaps corrupt the  measuring loop entirely   The frequency of cleaning and the method of cleaning will depend entirely on  the process     Where a refillable  flowing electrolyte  reference system is employed  make sure that the reservoir is  kept topped up  The rate of electrolyte consumption will again be process dependent  so experience will  show how often you must refill     The periodic recalibration of the sensor system is necessary to ensure best accuracy  This takes into  account the aging of the sensors  and the nonrecoverable changes that take place  These processes are  slow  however  If frequent recalibration is needed  it is usually because the cleaning process is not effec   tive  the calibra
97. cribed in detail in the instruction sheet  provided with the cable      IM 12B0702 01E    Appendix 11 3    11 3 1  Connection cable  To connect the other sensor systems  follow the general pattern of the terminal connections as  listed below   11 amp 12 Temperature compensation resistor input   Pt100  Pt1000                5 1            8 55kONTC  3500        6 8kOPTC  10                                                                                          13 Input       2  normally the reference element   14 Liquid earth  solution ground  connection  15 Input no  1  normally the measuring element   16 Screen  shield  for input no  1  17 Screen  shield  for input no  2  In connecting to a general purpose pH sensor In connecting to a gerneral purpose ORP sensor  WTB10 PH1  1 PH202G  S WTB10 PH1  1 PH202G  S  _ Terminal Box _ pH ORP transmitter _ Terminal        _ pH ORP transmitter  te 172 i 0150 i 2                   46 i                  H          PH8ERP 1 1 OR8ERG 1 i  PH8EHP i   E7013     0130 3    012     0120     i             Ono      LO     In connecting to a special purpose      sensor In connecting to a special purpose ORP sensor  WTB10 PH2  3     202      WTB10 PH1  1 PH202G  S  Terminal Box pH ORP transmitter Terminal Box pH ORP transmitter  t  2  HA405 01504 N T     c  HF405   i  DPA405            485 i O16 1  DPAS405 130    DPAS485 013                0120                      SA405  14 i Souo    i        1   Terminal        is used only when pH ORP transmitt
98. cturer s live  maintenance procedures     The cable used to interconnect the devices needs to comply with the following  parameters    Loop resistance R     15     150          Inductance per unit length L     0 4     1 mH km  Capacitance per unit length C   80     200 nF km    C   C line line   0 5       line screen if both line are floating     C      C    line line         line screen if the screen is connected to one line    Length of spur cable          30      Length of trunk cable  max  1 km   Length of splice   max  1       WARNING    Substitution of components may impair Intrinsic Safety    To prevent ignition of flammable or combustible atmospheres  disconnect power before  servicing or read  understand and adhere to the manufacturer s live maintenance  procedures     Application Doc  No         025   10 P 4 to P 5    IM 12B07D02 01E    2 16 Specification    2 12  Control Drawing of PH202S FF PB Specification       Intrinsically safe FISCO     FM Class     DIV  1  Group ABCD                for ambient temp   lt  55      Sensor    Connections  PH202S F Max  cablelength  60 mtr     or PH202S P Cable dia    3   12 mm   15                         FM Approved          Connection  FISCO barrier   a   i  d EM Approved i B FM Approved i  Voc  Vt    17 5 V    Terminator     Terminator        loc  It  380mA         90  1000   iR 90 1000    Poc  Pt    5 32 W ic 0 22yF                  iC 0 22yF i            Transmitter   Transmitter       d   O      d   O   O         Division 
99. detail    Default values       Communication          60  COMM  Communication Set HART   communication Off 1 0 On  Set HART   communication On  write enable write  write protect enable  Set communication PH201 B  Without half time check  With half time check   ADDR  Network address Set address 00 to 15 00  61  HOUR Clock setup Adjust to current date and time using   MINUT  gt     and ENT keys   SECND   YEAR   MONTH   DAY       62                Erase logbook    Press YES to clear logbook data       63 69                Not used                         Code   Display    Function    Function detail    Default values       General       70  LOAD    Load defaults    Reset configuration to default values       71 78    Not used       79  CUST D    Load Customer  Defaults    Reset configuration to default values  except buffer tables             Code  Display       Function    Function detail    Default values       Test and setup mode       80        5            Test and setup          Not used                      IM12B07D02 01E          Calibration 6 1    6  CALIBRATION                         202        be calibrated    three distinct ways     6 1  Automatic calibration   This method uses internally programmed buffer tables   from Service Codes 24  25 and 26   to calculate  the buffer value at the actual temperature during the calibration  In addition  the stability of the reading is  automatically calculated  and when the reading has stabilized fully automatic adjustments of 
100. e                        adc                                    duda ea Gels gaz 4 1  4 2  Explanation of operating                                              222  12 11 1  21       4 1            4 2  4 3  Setting passcodes mM x                        SQ 4 3  4 3 1  Passcode                                        4 3   4 4  Display                                                              4 3  4 5  Display functions RC  cT                s 4 4  4 5 1  Display functions pH                                         4 4  4 5 2  Display functions pH                                          4 5  4 5 3  Display functions pH                     4 6   SM Parameter sie                  5 1  5 1  caede   tec 5 1  5 1 1  Manual temperature selection and                                   5 2  5 1 2  Process temperature measuring in                                     5 3  5 1 3  Manual activation of                          4           emeret nnne nennen nennen ens 5 4  5 1 4  Manual impedance Check    rte et ctt x er der eet den ree o            aae 5 5   5 2  Commissioning MOdE mc           5 6  5 2 1 OUTPUT TANGO weiss                                5 7  5 2 2                                                      5 8  5 233  SGIVICQ                                                      5 10   5 3  Notes for guidance in the use of service coded settings                                 5 11  5 3 1  Parameter specific functions    e eer EE epe ra        5 11   5 3 2  Temperature compen
101. e  6 A1726JD 1 Pin Terminal Unit 3 terminals type  7 K9184AA 1 Screw Terminal Unit when  TB specified  8 K9661HR 2 Stud   9      Analog Board Assembly  K9661PA 1 For mA   HART  K9661PC 1 For FF Profibus   10    1 Digital Display Board  K9661AV 1 For mA   HART  K9661AW 1 For FF Profibus  11 K9660YQ 1 Screw Assembly to fix amplifier  12 K9660YP 1 Stainless steel screw assembly to fix cover  13    Adapter Assembly  K9414DH 1 For G1 2 screw when  AFTG specified  2 units    K9414DJ 1 For 1 2NPT screw when  ANSI specified  2 units     14   Comm  Board Assembly  K9661MA 1                9661     1 For Profibus      Do not exchange these parts  Call service personnel          Copyright 2007  1st Edition  Mar  2007  YK  CM PL 1 2B07D02 03E  YOKOGAWA  9 Subject to change without notice  2nd Edition   Aug 2007  YK     Yokogawa Electric Corporation    Pipe Wall Mounting Hardware  Option Code    U     1   X                               Panel Mounting Hardware   Option Code    SCT     Hood to sun protection   LE                    Option Code             ltem Parts No  Qty Description  1 K9171SS 1 Universal Mount Set   U   2 K9311BT 1 Tag Plate   SCT   3 K9311KA 1 Fitting Assembly   PM   4 K9311KG 1 Hood Assembly   H   K9660JA 1 Hood Assembly   H2        CMPL 12B07D02 03E 2nd Edition   Aug 2007              Customer Model PH202S  Style    3     L   pH ORP Transmitter  Parts List                            ltem Part No  Qty Description  1   Cover Assembly  K9315CA 1 Polyurethane Coating  K
102. e is the direct entry of  calibration data using the Service Code Menu  refer to Chapter 5   For example  the sensor may  be calibrated in a test laboratory and then transported to its location in the field and the sensor  parameters entered into the associated PH202   Service Code 23 allows you to access and enter ITP  Asymmetry Potential  or Zero Point  and Slope  parameters          12B07D02 01E    11 12 Appendix    11 5 5  Calibration Procedure   1  Manual Calibration  This involves calibration by direct entry of ORP sensor zero offset   1  Press the MODE key   2  MAN CAL is displayed   If you press NO then you will proceed to calibration with sample    3  Press YES   4  NEW SNS is displayed  asking if you wish to change the sensor parameters as for a new sensor   5  If you press YES then the existing calibration data stored in PH202 memory is deleted  If replacing  the sensor is not desired  press NO   6  In either case  ORP CAL is displayed   7  If you press YES then START is displayed  and you should immerse the sensor in the test  solution   If you press NO then you will proceed to Calibration with Sample   see  2  following    8  Next press YES again then 0000 mV is displayed  so enter a suitable zero offset value and press  ENT  CAL END is displayed   9  Press YES and WAIT is displayed for a moment  then the PH202 reverts to measurement mode      2  Calibration with Sample   This involves calibration by adjusting the PH202 to read the same as the ORP value derived b
103. earth disconnented  Insufficient electrolyte   Process too hot or too cold  Wrong temperature sensor setting  Temperaure sensor damaged    Check or replace sensor  Check sensor immersion  Check electrolyte reservoir  Check process   Check sensor  amp  setting  Check connections       Temperature sensor open   lt   30      or  gt 120      for 8 55        Process too hot      too cold  Wrong temperature sensor used  Temperaure sensor damaged    Check process  Check sensor  amp  setting  Check connections       Measurement out of range     2 to 16 pH for pH     1500 to 1500 mV for ORP     Sensors disconnected  Sensor wrongly connected  Sensor s  defective    Check cabling  Check cabling  Replace sensor s        EEPROM write failure    Fault in eledtronics    Try agein  if unsuccessful  contact Yokogawa       Wash recovery check error     if communication is set to PH201 Style B   in code 60     ORP   rH outside of preset limits    Measring sensor aged   Sensor still coated after washing  Defective wash system   Sensors disconnected or wrongly  connected    Replace measuring sensor  Check cleaning system   if needed adjust timings  Check cabling       No valid calibration data     Cable resistance to temperature  sensor exceeds limit value     Data lost after switching from  pH to ORP   Cable resistance too high  Corroded contacts   Wrong sensor programmed    Recalibrate    Use Pt1000  Clean and reterminate  Reprogram       Call for maintenance interval time exceeded    System no
104. eas                                             3 4  3 3 3  Installation in  Hazardous Area Non Incendive                 0022     0 0 0 00000 3 5  3 3 4  Liquid CANN                                               3 5  3 3 5  Access to terminal and cable                                   3 5  3 4  Wiring Of power supply saccanieweteidsistsdsceleeswencntdsaiadavanercteietedadecacarecerindierdsuadedeuse 3 5  3 4 1  General precautions                                       3 5  3 4 2  Connection of the power supply                                           3 6  3 4 3  Switching the instrument                            3 6  3 5  Wiring the Sensor                                               3 7  3 5 1  Impedance measurement jumper                                      3 7                3 8  3 6 1  Connection                                      3 9  3 6 2  Sensor cable connection with special                                    3 10  3 6 3  Sensor cable connections using junction box      10  and extension cable  WF 10     3 11  3 6 4  Connection VP type                                                                                   ETENEE E 3 11    In this manual a sign appears if it concerns the PH202G  S  E   C   U   N   K       IM 12B07D02 01E  11th Edition  Oct  2009 YK   All Rights Reserved   Copyright    2000  Yokogawa Electric Corporation    IM 12B07D02 01E    4  Operation  Display Functions And Setting                                       4 1    4 1  Operator interfac
105. ection and  safety of the human body  the product and the  system containing the product     IM 12B07D02 01E    The following safety symbols        used on the  product as well as in this manual     A DANGER    This symbol indicates that an operator must  follow the instructions laid out in this manual in  order to avoid the risks  for the human body  of  injury  electric shock  or fatalities  The manual  describes what special care the operator must  take to avoid such risks      N WARNING    This symbol indicates that the operator must  refer to the instructions in this manual in  order to prevent the instrument  hardware  or  software from being damaged  or a system  failure from occurring                         This symbol gives information essential for  understanding the operations and functions           This symbol indicates Protective Ground  Terminal         This symbol indicates Function Ground  Terminal  Do not use this terminal as the  protective ground terminal      rw    This symbol indicates Alternating current     This symbol indicates Direct current     IM 12B07D02 01E    Warranty and service   Yokogawa products and parts are guaranteed  free from defects in workmanship and material  under normal use and service for a period of   typically  12 months from the date of shipment  from the manufacturer  Individual sales organi   zations can deviate from the typical warranty  period  and the conditions of sale relating to the  original purchase order should be c
106. edance low  input 2   Impedance low  input 2   Impedance higt  input 2   Impedance high  input 2   Temperrature too high  Temperrature too high  Temperrature too low  Temperrature too low   pH out of range   pH out of range    Wash recovery check    Wash recovery check    Call for maintenance    Call for maintenance    Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail  Soft fail  Hard fail                      T5 3 5E eps    IM 12B07D02 01E    11 10 Appendix    11 5  Calibration in        Mode  For an ORP meter  unlike the case for a pH meter  there is no predefined  standard solution  that  you must use for calibration  For an ORP meter  normal maintenance is limited to a check of the  electrodes  Electrode checks and calibration are performed in the following circumstances     Electrode Check    When you change the ORP sensor  or when you stop continuous operation     When you wash the sensor tip or liquid path of the ORP sensor   Calibration    When the electrode potential is out of limits     When you want two or more ORP meters to read as near as possible to the same value   Note  There are two ways of calibrating the PH202 in ORP mode  but both are one point calibration   zero shift      1  Manual calibration  MAN CAL  and   2  Calibration with sample  SAMPLE     If you use the FU20 combination pH and ORP sensor  then you can calibrate ORP using t
107. elect YES NO then set value with  gt      ENT keys     Press      to change to            ASP Set As Pot For the main parameter  Press YES to confirm 1 mV resolution   then set value with  gt      ENT keys   24  BUF ID Buffer table 4 Buffer tables to NIST  formerly NBS   25  BUF ID Buffer table 7  see section 10 for table details   26  BUF ID Buffer table 9 User adjustable for special requirements  27  ZERO P Enable zero point in   Disable zero point  enable As Pot  0 Disabled  pH units Enable zero point  disable As Pot   28 29 Not used          IM12B07D02 01E       7  ITP    ero Point       pH       Parameter setting 5 17    5 3 4  mA output functions    Code 31    Code 32    Code 35    Table 5 2      OUTP F       BURN     TABLE    When      is set in code 01 as the main parameter  the output functions may be  set as follows     0  pH   1  pH  table    2  Parameter 2  ORP or rH as set in code 02    When        is set in code 01 as the main parameter  the output functions may  be set to    0  ORP   1  ORP  table     Diagnostic error messages can signal a problem by sending the output signals  upscale or downscale  21 mA or 3 6 mA when HART or distributor comm  is  non used  3 9 mA when HART or distributor comm  is used   This is called  upscale or downscale burnout  from the analogy with thermocouple failure  signalling of a burned out or open circuit sensor  The pulse burnout setting  gives a 21 mA signal for the first 30 seconds of an alarm condition  After the     pulse   
108. ems    2 1 Insulation resistance test   2 2 Dielectric strength test  2 3 Temperature indication check  2 4 pH indication check   2 5 ORP indication check   2 6 Profibus communication functional check    Note  Items marked with an asterisk     may only be confirmed by a test certificate     3  Inspection Methods  Standards and Conditions    e Connectthe testing circuit as shown in Figure 1  Allow the instrument to warm up for at least  5 minutes before conducting the tests  For the connections for the insulation resistance test   follow the instructions in Section 3 1 and for the connections for the dielectric strength test   follow the instructions in Section 3 2    e Use test equipment shown in Figure 1  or equivalent  for the tests     3 1 Insulation Resistance Test     1  Apply 500 V DC between the power supply terminals shorted together    and      and the  earth terminal  G   The insulation resistance must be 100 MO or greater      2  Apply 500 V DC between the input terminals shorted together  11 to 17  and the earth  terminal  G   The insulation resistance must be 100 MO or greater     3 2 Dielectric Strength Test  Required Only for PH202S     Apply 600 V AC  an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz  or 60 Hz  between the terminals shown below  for at least 2 seconds  The insulation must  withstand this voltage   The sensed current should be 10 mA       1  Between the power supply terminals shorted together    and      and the earth
109. end AUT CAL appears   and the YES NO key prompt flags flash  Press NO     The display MAN CAL appears   Press YES to start calibration               Press YES or NO at NEW SNS prompt                                                   EXA           YOKOGAWA                                                                       gt      Note  Press NO to start  calibration of zero point when  enabled in Service Mode               Put sensors in buffer  solution  Press YES           Set the value  using the  gt     ENT key     Select the flashing digit with the  gt  key           Increase its value by pressing the  key     When the correct value is displayed  press ENT to enter the change                              YES       WAIT is displayed  briefly then EXA returns  to measuring mode                    For 2 point  As Pot and Slope   Adjustment select second buffer  solution and adjust as for pH7 buffer              IM 12B07D02 01E    Calibration 6 5       Manual Calibration of Manual Calibration of mV   zero point according offset for ORP  2nd parameter     to IEC 60746 2  Where both pH and ORP  or rH   When enabled in service are measured  the offset  Asymmetry  code 27  potential  of the second parameter    is calibrated as shown below   When enabled in service  code 02                                                                                                                                      Adjust value Adjust value  with  gt  A   ENT Keys with  gt  A   ENT Keys  
110. ensor  The impedance limits will need to be adjusted to get the best from  systems using heavy duty  or fast response electrodes    The impedance measuring system has a very wide span requirement  As it can  measure in      and also in       109  there are hardware switches to set high  range  1MO to 2 GQ  or low range  1kO to 1MQ  measuring  As a default the  system is set to measure high impedances on input 1  the one normally   used for the pH glass sensor input  and low impedances on input 2  the one  normally used for the reference input   Examples of where these settings need  to be changed from the default  are Pfaudler enamel sensors which need two  high impedance settings  and Platinum sensors with a standard reference   which need two low impedance settings    The temperature compensation of the impedance measurement is for conven   tional pH glass sensors  When other sensors are used  switch this feature off     The calibration checking feature  when enabled  gives security against entering  wrong calibration data  For example when aged sensors are due for  replacement  the EXA flags an error message and prevents a calibration being  completed where the subsequent measurement can only exhibit errors and  drift    Limits are set for the maximum permissible Asymmetry potential  and Slope     IM12B07D02 01E    5 12 Parameter setting    Code   Display    Function    Function detail    Default values       Parameter specific functions       01  PH ORP    Select main par
111. er is installed far from the cable length of pH ORP sensor    2   The cable is specified by an option code of the terminal box    3   Use the box in combining PH202 with SA405     F3 6 1E eps    Figure 11 1 Wiring of a sensor    IM 12B07D02 01E    11 4 Appendix    In connection of the PH202G    and a special purpose pH  sensor having temperature element             Core cable number Terminal number of  of pH sensor pH ORP transmitter  HA406 PH202G  S  DPA406 pH ORP transmitter   14     16               15       13   Two core cable numbered by 14 are connected   13  O 13 to the No  14 terminal of the     202    S     Nothing is connected to the 16 terminal   12     7 of the PH202G  S    11  O 11        14  14       In connection of the PH202G  S and FU20 pH ORP sensor                          202       FU20 pH ORP transmitter  11 O     12    12  13 O 13  14 O 14  15 O 45  16    16 F3 6 2E eps  Figure 11 2 Optional Sensor wiring  NOTE  pH sensor ORP sensor    GE 15  Red  GE 15  Red   RE 13 Brown  RE 13  Brown   T1 11 Black  SE 14  Black   T2 12  White  G 16  Green   SE 14  Green    S  16 Yellow     When PH202    TB option specified for using  the sensor with folk terminal      for     200     400  OR200 OR400 type      Use the cable gland within the transmitter   For wiring  refer to right figure      Cables  terminals and glands  For the PH202G  PH202SJ use cable of outside diameter 6 to 12 mm  When using individual  electrodes    glass electrode  reference electrode    and se
112. et  menu  as  pot  or  Chapter 5  zero point    4  53               e                   Lt Lt Lt Efficiency            2   slope              No 4       Z SETHOLD i            hse ms PR re     line of display                NO           C        gt  A ENT      C                                       Im XOKI      wa     12C6C3 20 1 ai       IM 12B07D02 01E       4 5 2     Display functions       ORP     Operation 4 5                                                                                                  Measuring  Mode                                                                                                         DISPLAY As Pot  Display 2                See Man   Imp  Check  chapter 5    YES                                                      Service Code 01 Set for pH   Service Code 02 Set for ORP  on parameter 2        Logbook Scrolling   Logbook data is available  only in instruments with   PIN  for advanced function  The display can give information  about calibrations performed  with date and time    The example below shows   Asymmetry Potential     Scrolling of Data is also  available on Slope   As Pot ORP  Impedance Input 1  Impedance Input 2  When these functions  are enabled in Service  Codes                                                                                      See Temp  Menu  chapter 5                                                                      See Hold  Menu  chapter 5                                                         
113. ether  11 to 17  and the earth  terminal       The insulation resistance must be 100      or greater   As for the PH202SJ  follow the instructions below      1  Apply 125 V DC between the power supply terminals shorted together    and      and the  earth terminal       The insulation resistance must be 9 5      or greater      2  Apply 125 V DC between the input terminals shorted together  11 to 17  and the earth  terminal       The insulation resistance must be 100      or greater     3 2 Current Output Test    Connect the instruments as shown in Figure 1  and set them as follows   Connect a shorting bar between terminals LOW and IMP     Decade resistance box  921 6 O  Standard voltage current source  0 mV  DC source  24 V DC    Use the following procedure to enter the inspection mode      1  Entering Service Code 87   Press the     key  The message display will show     OUTP       Press the  NO  key until the message display shows     SERV       Press the  YES  key  The data display will show  00  with the first digit of 0 flashing    Press the     key eight times  The data display will show  80  with the first digit of 8 flashing   Press the   gt   key once  The data display will show    80    with the second digit of 0 flashing                    YOKOGAWA    QIS 12B07D02 01E    Yokogawa Electric Corporation ISt Edition  Reb  2001 YK   6th Edition  Mar  2007                         12B07D02 01E    12 2                      2 3    f  Press the      key seven times         d
114. ether the test  solution was fresh and whether it was mixed in the correct amounts  see mixing instructions  above         Tolerance                      Tolerance             0 10 20 30 40 50  Temperature   C F6 1E eps    Fig  11 3 ORP of Test Solution     7  Replace the sensor in its holder        WARNING    Be careful to make sulfuric acid solution     11 5 2  Manual Calibration  The procedure for calibration is described in 11 5 5  1    Calibration is performed by using a solution of known ORP value  and adjusting the PH202 to read  the same value  Alternatively you may calibrate the PH202 to read the same as another  standard   instrument in a given solution  The electrode check described in 11 5 1 above is performed if the  measured value appears to be well out of tolerance  Calibration of the ORP value is performed by  zero shift one point calibration     11 5 3  Calibration with Sample  The procedure is described in 11 5 5  2    Calibration with a sample enables measuring a representative sample  e g  by laboratory analysis   while the ORP sensor is on line and measuring the same sample solution  saving the measured data  in the memory of the PH202  and then adjusting the PH202 to read the same value  as obtained  e g  by laboratory analysis   So normal on line control or pH measurement is not affected  The ORP  measurement is calibrated by zero shift one point calibration     11 5 4  Data Entry  If the above methods of calibration present problems  another alternativ
115. fied i    Terminator p                    9       Terminator                i Transmitter   i Transmitter i    4          Safe area   Zone 0 or 1    Hazardous area    e Sensor s  are of a passive type to be regarded as  simple apparatus        e Electrical data of the PH202S F  amp  PH202S P     or    Supply and output circuit    Maximum input voltage Ui 24 V  Maximum input current li 250 mA  Maximum input power      1 2 W   Effective internal capacitance Ci  220 pF   Effective internal inductance Li  0 pH     FISCO field device   Maximum input voltage Uiz17 5 V  Maximum input current li 380 mA  Maximum input power Pi 5 32 W  Effective internal capacitance Ci  220 pF   Effective internal inductance Li  0 pH     Sensor input circuit    Maximum output voltage Uo 14 4 V  Maximum output current lo 32 3 mA  Maximum allowed external capacitance Co 600 nF   Maximum allowed external inductance Lo 34 mH    e Any 1 5  interface may be used that meets the following requirements     FISCO power supply   Uox24V            lt  17 5         lo  lt  250 mA lo  lt  380 mA         lt  1 2 W       lt  5 32 W         gt  220 pF                Lo20        Lcable    e Electrical data of the     2025     amp  PH202S D  Type of protection              Supply and output circuit    Maximum input voltage Ui   32 V   Effective internal capacitance Ci   220 pF  Effective internal inductance Li   0 pH   Sensor input circuit    Maximum output voltage Uo 14 4 V  Maximum output current lo 32 3 mA  Maximum al
116. g          Menu pointer flags    Units                GSEs               It COC  VLEET             YES                ENT    Main display        Commissioning  function menu  Message display                                                            yvvvvv  5             Key prompt flags            Selection keys  YES   Accept setting    NO   Change setting  PR mode key    P keys   ENT    Choose         CD     Broken line indicates area that can be    adjust      Adjust digit EXAM O2 YOKOGAWA    seen through front cover    ENT   Confirm change    Commissioning  mode access key    Measure Maintenance                                     Figure 4 1  PH202 operator interface    4 2  Explanation of operating keys  MODE key This key toggles between the measuring and maintenance modes  Press once to obtain  access to the maintenance function menu   AUT CAL  MAN CAL  DISP  SETPOINT  WASH  MAN IMP  TEMP  HOLD  Press again to return to the measuring mode  press twice when hold is activated     YES NO keys These are used to select choices from the menu   YES is used to accept a menu selection   NO is used to reject a selection  or to move ahead to the next option                          DATA ENTRY keys     P                gt       used as a    cursor    key  Each press on this key moves the cursor or flashing digit   one place to the right  This is used to select the digit to be changed when entering  numerical data    A   is used to change the value of a selected digit  Each press on
117. gital     Supply voltage  9 to 32 V DC     Operating current  26 0 mA     Operating values  According to IEC 1158 2    Bus connection    Fieldbus interface base on        1158 2 according to  FISCO Model  Power supply  Power supply is achieved de   pendant on the application by  means of segment coupler      Data transfer  According to PROFIBUS  PA    profile class B based on EN  50170 and DIN 19245 part 4      GSD file  The actual file can be down     loaded from www profibus   com Configuration  Local with  6 keys  Software  Firmware based on Siemens  DPC31 stack   Hardware         or PCMCIA interfaces from  Siemens      Other control  Siemens PDM systems    Electrical connection   Terminals acc  to IEC 1158 2      Fieldbus cable types     Twisted and shielded two   wire cable according to  recommendation based on IEC  1158 2 Cable diameter  6 to 12  mm  0 24 to 0 47 inch     Specification 2 5    FOUNDATION FIELDBUS H1 communications    Input signal  Digital    Supply voltage  9 to 32 V DC    Operating current  26 0 mA  base current     Operating values  According to IEC 1158 2    Bus connection    Fieldbus interface based on IEC  1158 2 according to FISCO Model    F  Power supply   Power supply is achieved  dependant on application by  means of segment coupler   G  Data transfer   FF specification Rev  1 4 Basic  device   H  Function blocks   3 x Al  Transducer  Resource      Files  Actual file can be downloaded from  our homepage   J  Configuration  locally with 6 keys   
118. gram above  These safety descriptions cover most of the commonly used  industry standard barriers  1solators and power supplies              Hand Held Communicator must be of a IECEx certified intrinsically safe type in  case it is used on the intrinsically safe circuit in the hazardous area or of a IECEx  certified nonincendive type in case it is used in the nonincendive circuit in the  hazardous area     IM 12B07D02 01E    2 8 Specification    2 5  Control Drawing of PH202S mA HART   Specification                                                                                                                           Intrinsically safe design  CENELEC standard EEX ia IIC  T4 for ambient temp   lt  55  C        ia or ib  T6 for ambient temp   lt  40  C Certified safety barrier or power  Certificate nr  KEMA 06ATEX0218 X        ae as  PH202S  pH ORP transmitter  P  24 volts DC Nominal  Supply Voltage                        o                 Uo   31 5 Volt DC     o G lo   100 mA            x     SENSOR S         terminals 11 17 Functional Functional  earth earth Load  Resistance  Hazardous area Safe area  4 Zone 0 or 1      lt   Intrinsically safe design  CENELEC standard EEx ia       T4 for ambient temp   lt  55  C EEx ia or ib Certified Repeater  T6 for ambient temp  lt  40  C Power Supply  Certificate       KEMA 06ATEX0218 X  HART Compatible   PH202S  pH ORP transmitter      o        Output                                              Uo   31 5 Volt DC  O Io  100 mA O      
119. he    2nd  parameter    setting as per the section 6 5 3  Manual Calibration     11 5 1 Electrode Check  To determine whether the        electrode is normal      not  a solution of known        is measured to  check if the measured value is within tolerance  see Fig  11 3   This is done in measurement mode     Electrode test solution  A solution of known ORP  Oxygen Reduction Potential  is used  Usually quinhydrone or ferric chloride  solution is used as the test solution   Yokogawa sells the following chemicals for making test solution  The procedure for mixing solution is  described below   Quinhydrone salts  P N K9024EC  and Ferrous  amp  Ferric salts  P N K9024ED    Procedure for Electrode Test  You determine whether or not the ORP sensor is normal by measuring the ORP of a test solution of  known ORP and determining if the measured value is within tolerance   Before starting normal operation  when you need to check if the ORP sensor is normal or not  use the  procedure below   A solution of known ORP is measured to check if the measured value is within tolerance   Here we describe how to mix test solutions using the quinhydrone salts supplied with the OR8AX  accessory kit  or the ferrous salts that are sold separately by Yokogawa      1  Mix test solution   lt To mix quinhydrone test solution gt   Pour one packet of quinhydrone into a wide mouth  250 ml  beaker and dissolve in pure water to  make 250 ml of solution  If the pure water is cold  the powder will not complete
120. his is a composite value that indicates the oxidizing power of a solution com   pensating for the influence of the acid or alkaline components  The scale is 0   55 rH  where oxidizing solutions provide the highest readings     This is the difference between the isothermal point of intersection and the zero  point     This is the sensitivity of the pH electrode  mV pH  usually expressed as a   of  the theoretical value  Nernst      This is the isothermal point of intersection  This is the value in pH at which the  temperature response of the system is at a null point  In other words  the point  of intersection of the temperature lines on a graph of milli volts vs pH  This point  is critical to the correct operation of the temperature compensation circuitry     This is the value of pH at which the electrode combination yields 0 mV as an  output                         2004  0                  12B07D02 01E    Appendix 11 1    11  APPENDIX 2       11 1                  Feasible combinations of the PH202G pH ORP transmitters with different styles of the PH201G  distributor are listed in the table below  The distributor has the usual distributor functions  supply  power to transmitter  receive current output from transmitter  and provide analog output  as well  as contact output functions  maintenance  wash and fail status signals   Since the two transmitters  provide different digital signals to control the distributor contact outputs  two distributor styles are  provided for comp
121. ide first which application and which settings are appropriate for the  installation  This decision is best made before the jumpers are installed  because the cables will  rest beside the jumpers in their installed positions     Table 3 1  Impedance measuring jumpers          Figure no  Jumper Settings Jumper Settings Application  amp  Sensor Connections  Input  1 Input  2  1 High Impedance Low Impedance Normal pH sensors    Glass sensor on Input  1  Reference sensor on Input  2       2 High Impedance High Impedance Special electrodes using  2 glass sensors   e g  Pfaudler 18        3 Low Impedance High Impedance ORP  pH compensated  and or rH  metal sensor on Input  1  pH glass  as reference  on Input  2       4 Low Impedance Low Impedance ORP  Redox measurement   metal sensor on Input  1  Normal reference on Input  2                      For convenience insulated jumper links are provided  Ordinary wire links can also be used  and are just  as effective     The following four jumper figure illustrations  figure 3 8  show the jumper positions related to the figure  numbers in the above table                                                                                              T                 T                1     G 11 12 14 17 13 tow me              15 16 Setting on shipment  SUPPLY   L TEMP LE INPUT 2 INPUT 1    T                             T      2     G 11 12 14 17 13 Sow me              15 16  SUPPLY TEMP LE INPUT 2 INPUT 1    T T                 T         T T
122. in accordance with Article  501 4 B  3     e Grounding shall be in accordance with Article 250 of the National Electrical code   e      case of using cable glands in Outdoor location  they shall be UV rated or made of metal     WARNING    Substitution of components may impair suitability for Division 2          not remove or replace while circuit is live unless area is know to be non hazardous    Explosion Hazard   Do not disconnect equipment unless area is know to be  non hazardous         not reset circuit breaker unless power has been removed from the equipment or the  area is know to be non hazardous    Application Doc  No   IKE025 A10 P 8    IM 12B07D02 01E    Specification 2 19    2 14  Control Drawing of PH202S FF PB Specification       Non incendive FNICO     FM Class    DIV  2  Group ABCD         FM Approved         Power Supply   E  Voc x 32 VDC iFM Approved                                T4 for ambient temp   lt  55   C Sensor  Connections    PH202S B Max  cablelength  60 mtr              2025    Cable dia   3   12 mm                        Connection          i Terminator j       iR   90 1000               iC 0 22yuF   iC 0 22yF                            564           i Transmitter     Transmitter 1    Division 2  Unclassified Location Classified Location    Sensor s  are of a passive type to be regarded as  simple apparatus   devices which  neither store nor generate voltages over 1 5 V  currents over 0 1 A  power over 25 mW or  energy over 20 uJ       are FM Ap
123. in case it is used in the non incendive circuit in the hazardous area     IM 12B07D02 01E    Specification 2 9    2 6  Control Drawing of PH202S mA HART   Specification  FM Intrinsically safe design        Intrinsically safe design    7 is  FM Class I  Div l Group ABCD    4 for ambient temp   lt  55     FM Approved safety barrier or  power supply  PH202S transmitter with Rint   300      HART compatible   24 volts DC Nominal          Supply Voltage                    B  Sensor s  For electrical data         terminals 11 17 see text below  NT               cablelength  60 mtr  rin  Cable dia    3   12 mm  earth                     Load  Resistance     q Classified Location Unclassified Location       Figure 1          Intrinsically safe design                                                          FM Approved  FM Class I  Div 1  Group ABCD  T4 for ambient temp   lt  55  C  Power Supply  PH202S transmitter  HART compatible                 Output      o  0                 u       to    Go           ro             Supply               Sensor s  For electrical data   terminals 11 17 seeitextibelows Functional  Max  cablelength  60 mtr  ea  Cable dia   3   12 mm   4 Classified Location    g   nclassified Location  Figure 2    Electrical data of the PH2028    Supply circuit  terminals   and      Maximum input voltage Vmax   31 5 V  Maximum input current Imax   100 mA   Maximum input power Pmax   1 2 W   Effective internal capacitance Ci   22 nF  Effective internal inductance Li   3
124. ion ground y    o 13 Reference  14 Solution ground 15 Glass  measure    16 Shield         15 Glass  measure   Blue       6Shield   o 47  Blue 17Shield              x Cable markers                                   FU20 FU25 4 in one sensor             Note  Connect cables to similarly marked  terminals  11 to 11  12 to 12  etc              11 Temperature i  12 Temperature   13 Reference    14 Solution ground    15 Glass                    16 Shield     17 Shield    Green   Red    11 Temperature         12                          13 Reference     14 Solution ground   15 Glass  measure   16 Shield i   17 Shield                             7  CESS Retractable sensor assembly PR20     Also PD20      20  amp    520   k                Cable markers    NOTE     Connect cables to similarly marked terminals   11 to 11  12 to 12  etc     Combined pH ref sensors    Figure 3 10a  Sensor wiring    IM 12B07D02 01E    3 10 Installation and wiring             o 11 Temperature   11 Temperature                Yellow          913 Temporaire   o 12                      2 Blue  Black   j    lue      13Reference                  1  Solution around 14 Solution ground       15 Metal  measure     Blue    lt 16 Shield    Blue    15 Metal  measure   17Shield       j    16 Shield    17 Shield                                   Cable markers         Cable markers  Separate electrodes    ORP REF electrode  Combined ORP pH electrode       Notes  11 Temperature    12                        1  A temperature 
125. is also signalled     Table 11 1  Error Codes    Error description  Buffer solution temoerature  outside the programmed range    Possible cause  Buffer solution too hot or too cold    Suggested remedy  Adjust buffer temperature  Check cabling       Measurement failed to stabilize   during the calibration    Sensors fouled  Sensors too slow  aged sensor      Clean sensors  Replace sensors       Asymmetry potential too high     Limits set in service code 21      Slope   sensitivity   is outside limits     Limits set in service code 22      Sensors are aged or polluted  Mistake in calibration    Measuring sensor aged  Poor insulation at the connector    Check buffer solution  Recalibrate at pH7  Replace sensor   Replace measuring sensor  Replace or dry cables       Impedance of input 1 too low     Limits set in service code 03      Measuring sensor broken  Dameged or damp connections    Replace measuring sensor  Replace or dry cable       Impedance of input 2 too low     Limits set in service code 04      Measuring sensor broken  Dameged connections    Replace measuring sensor  Replace cable       Impedance of input 1 too high    Limits set in service code 03      Measuring sensor disconnected  Sensors not immersed in process  Liquid earth disconnented    Check connections  Check porocess  Check connentions       Impedance of input 2 too high    Limits set in service code 04      Temperature sensor open   gt 140              10   C for 8 55kO    Reference sensor fouled   Liquid 
126. is manual should be passed on to the end  user    The contents of this manual are subject to  change without prior notice    The contents of this manual shall not be  reproduced or copied  in part or in whole   without permission    This manual explains the functions contained in  this product  but does not warrant that they are  suitable the particular purpose of the user   Every effort has been made to ensure accuracy  in the preparation of this manual    However  when you realize mistaken  expressions or omissions  please contact the  nearest Yokogawa Electric representative or  sales office    This manual does not cover the special  specifications  This manual may be left  unchanged on any change of specification   construction or parts when the change does  not affect the functions or performance of the  product     If the product is not used in a manner specified  in this manual  the safety of this product may be  impaired     Yokogawa is not responsible for damage to the   instrument  poor performance of the instrument   or losses resulting from such  if the problems are   caused by      Improper operation by the user      Use of the instrument in improper applications     Use of the instrument in an improper  environment or improper utility program     Repair or modification of the related instrument  by an engineer not authorized by Yokogawa     Safety and Modification Precautions     Follow the safety precautions in this manual  when using the product to ensure prot
127. itter  turn on the power again  and enter the  inspection mode to restart the tests     3 1 Insulation Resistance Test     1  Apply 500 V DC between the power supply terminals shorted together    and      and the  earth terminal  G   The insulation resistance must be 100 MO or greater      2  Apply 500 V DC between the input terminals shorted together  11 to 17  and the earth  terminal  G   The insulation resistance must be 100 MO or greater     3 2 Dielectric strength test    Apply 600 V AC  an AC voltage of substantially sinusoidal waveform with a frequency of 50 Hz  or 60 Hz  between the terminals shown below  for at least 2 seconds  The insulation must  withstand this voltage   The sensed current should be 10 mA       1  Between the power supply terminals shorted together   and      and the earth terminal  G    2  Between the input terminals shorted together  11 to 17  and the earth terminal  G      3  Between the input terminals shorted together  11 to 17  and the power supply terminals  shorted together    and            YOKOGAWA    QIS 12B07D02 21E    Yokogawa Electric Corporation 184 Edition  Mar  2007                        12B07D02 01E    12 6                      3 3    3 4       2 3    Current Output Test    Connect the instruments as shown in Figure 1  and set them as follows   Connect a shorting bar between terminals LOW and IMP     Decade resistance box  92160  Standard voltage current source  0 mV  DC source  24 V DC    Use the following procedure to enter the
128. lay  mV    1500  1500  1  0 0 1   1500  1500  1                Press the  ENT  key until the message display shows    READY      Press the  ENT  key again to end the tests     PH202G  PH202SJ    SUPPLY TEMP LE INPUT 2 INPUT 1  Tr 1    13     LOWIMP HIGH IMP         3000  Shorting bar    Resistance  Box                        Standard  Voltage Source       Figure 1 Testing Circuit and Test Equipment    QIS 12B07D02 01E       IM 12B07D02 01E       12 4                   R    OX    TEST CERTIFICATE         Mh             53535 AF       PRODUCT NAME pH ORP TRANSMITTER TAG_NO   T   MODEL PH202G  PH202SJ   FBCNo            ORDER       SERIAL            AIA A INSPECTION ITEM  4    RESULT                                                                7     Gm  INSULATION  1100   Q         500V DC PH202G  L19 5M Q         125V DC PH202SJ   RESISTANCE BETWEEN POWER SUPPLY TERMINALS      AND EARTH TERMINAL G    1100     OR MORE 500V DC PH202G    19 5  9 OR MORE 125V DC      2025     AJ 3   11917             F  G  Fl  L1100M             500V DC  PH202G  L1100M Q LE   125V DC  PH202SJ   BETWEEN INPUT TERMINALS 11 TO 17  AND EARTH TERMINAL G   L 100M 9 OR MORE 500V DC PH202G  L1100MQ OR MORE 125V DC  PH202SJ   HA Bit HARE   0 02mA ACCURACY    0 02      CURRENT OUTPUT AU      OUTPUT  mA     INDICATION  248   REFERENCE SEM ACTUAL    BERT      PT1000  TEMPERATURE INDICATION                       14983                                  HAZ    0 01pH ACCURACY   nsn    pH INDICATION          INDI
129. le 3  Check Point  pH  Simulation Input  mV  Data Display  pH   0  414 1 0 00  0 01  7 0 0 7 00  0 01  14  414 1 14 00  0 01                          Indication Check    Following Section 3 5  press the  ENT  key  The message display shows    ORP INP     In this state   change the simulation input as shown in Table 4 by means of the standard voltage current source  and check the data display  The corresponding value on the data display must be within the    range shown in Table 4                 Table 4  Simulation Input  mV  Data Display  mV    1500  1500  1  0 0  1   1500  1500  1                Press the  ENT  key until the message display shows    READY        Press the  ENT  key again to end the tests     SUPPLY    3000                                  24VDC                   1                   T 1    Decade  Resistance    PH202S    INPUT 2    13       LOWIMP HIGH IMP          Shorting bar                 Standard  Voltage Source            Testing Circuit and Test Equipment    INPUT 1    QIS 12B07D02 21E          IM 12B07D02 01E       12 8                            TEST CERTIFICATE         pH ORP zi  28 27 No   PRODUCT NAME pH ORP TRANSMITTER TAG NO          MODEL PH202S   E B  No  HRES  ORDER       SERIAL NO     APPEARANCE    8817   INSULATION  RESISTANCE    MEE  DIELECTRIC  STRENGTH    EDER    CURRENT OUTPUT    DERT  TEMPERATURE  INDICATION    pHm  pH INDICATION                     INDICATION    NOTES    Att   DATE          INSPECTOR    YOKOGAWA 4    IM 12B07D02 01E
130. lowed external capacitance Co   3 5uF   Maximum allowed external inductance Lo   76 mH    IM 12B07D02 01E    Specification 2 13    2 10  Control Drawing of PH202S FF PB Specification  ATEX     Ex ia        T4 for ambient temp   lt  55   C  Ui  24 V or Ui 17 5V  li   250      li   380 mA  i Pi   5 32 W               2025              2025       t   Sensor         Connections                    Safe area  O LS     1 5       1 5     Apparatus Pa interface i certified   d certified i  Terminator     LO i i  Safe area Zone 0 or 1       Hazardous area    Sensor s  are of a passive type to be regarded as    simple apparatus        Electrical data of the PH202S F  amp  PH202S P     Supply and output circuit     or    Maximum input voltage Ui 24 V  Maximum input current li 250 mA  Maximum input power Pi 1 2 W   Effective internal capacitance Ci  220 pF   Effective internal inductance   0 pH     FISCO field device   Maximum input voltage Ui 17 5 V  Maximum input current 11 380 mA  Maximum input power      5 32 W  Effective internal capacitance Ci  220 pF   Effective internal inductance Li  0 pH       Sensor input circuit     or    Maximum output voltage Uo 14 4V  Maximum output current 10 32 3 mA  Maximum allowed external capacitance Co 600 nF  Maximum allowed external inductance Lo 34 mH    Any 1 5  interface may be used that meets the following requirements           lt  24 V   lo  lt  250 mA         lt  1 2 W         gt  220 pF   Ccable  Lo  gt  0        Lcable    FISCO power supply
131. ltage   The sensed current should be 10 mA       1  Between the power supply terminals shorted together    and      and the earth terminal        2  Between the input terminals shorted together  11 to 17  and the earth terminal  G      3  Between the input terminals shorted together  11 to 17  and the power supply terminals  shorted together    and         3 3 Temperature Indication Check    Connect the instruments as shown in Figure 1  and set them as follows   Connect a shorting bar between terminals LOW and IMP     Decade resistance box  92160  Standard voltage source          DC source  24 V DC    change the resistance value of the decade resistance box as shown in Table 1  The  corresponding temperature indication must be within the range        YOKOGAWA   QIS 12B07D02 61E    Yokogawa Electric Corporation ee Ration  Apr                         IM 12B07D02 01E    12 10                      3 4    3 5       2 3                Table 1  Reference Temperature   Resistance Box Resistance Data Display                       20 921 6  20 0  0 3   25 1097 3  25 0  0 3   130 1498 3  130 0  0 3                        Indication Check    Set the instrument as follows   Decade resistance box     10000    Change the simulation input value of the standard voltage generator as shown      Table 2          corresponding pH data must be within the range                 Table 2  Check Point  pH  Simulation Input  mV  Data Display  pH   0  414 1 0 00  0 01  7 0 0 7 00  0 01  14  414 1 14 00  
132. ly dissolve and some  may float on the surface  but this will not cause any problems in practice    lt To mix ferrous  amp  ferric chloride test solution gt   Pour one packet of the light purple or light green powder to a wide mouth  250 ml  beaker and add  a 2 mol l solution of sulfuric acid to make 250 ml of solution  If you want to start with concentrated  sulfuric acid  mix one packet of the powder with approximately 150 ml of pure water in a wide mouth  beaker then  stirring vigorously  add 14 ml of concentrated sulfuric acid  Top up with pure water to  make 250 ml    2  Transfer about 50 100 ml of test solution to a clean 200 ml beaker    3  Remove the ORP sensor from its holder  If an immersion holder is used  first remove the holder then  remove the  sensor  protector from the holder    4  Wash off any of the measured solution adhering to the sensor using tap water  then wipe off any  moisture droplets adhering to the sensor    5  Immerse the sensor tip in the test liquid  see  2   above   wait for the ORP reading to stabilize  this  typically takes 5 to 10 minutes      IM 12B0702 01E                   11 11     6  After the display reading has stabilized  read and note it  Measure the temperature of the test  solution  and check the ORP of the test solution at that temperature using Fig  11 3   If the ORP reading is within the tolerance shown in Fig  11 3 then the sensor is normal   Note  If the measurement value is only slightly out of tolerance  you should check wh
133. n an ORP sensor is used  connect both terminals LOW and IMP on the upper row with a shorting  bar as well both terminals HIGH and IMP on the lower row with another one     11 3 2  Sensor cable connection using terminal box     you can t install a sensor close to the PH202 transmitter  you may need to connect to       intermediate terminal box  WTB10 PH1   PH2  and use an extension cable  You should order these  from Yokogawa  as these Yokogawa parts are specially designed so that system specifications will be  met  The total cable length  for 5 m electrode cable plus 15 m extension cable  may be up to 20 m   Refer to Figure 11 1 for the wiring    11 4  PH201G B Dedicated Distributor    The     202 supports proprietary bidirectional serial communications with the     201    Style B  for  remote maintenance  washing and remote diagnostic purposes  This allows PH202 status  under  maintenance abnormal  to be monitored remotely  and commands to be sent to an instrument for  wash  Prepare for the instrument separately  Service codes 60 relates to communication settings     11 4 1  Communication setup  Code 60  COMM  The settings should be adjusted to suit the communicating device connected to the  output  The communication can be set to HART or to PH201G  Style B  distributor   When used with our     201    Style B  you can enable or disable contact outputs   namely  Fail contact and Hold contact  The PH201G  Style B  can output Hold  contact and Fail contact signals  You can set 
134. ndication Check    Following Section 3 3  press the  ENT  key until the message display shows        1000     In this  state  change the resistance value of the decade resistance box as shown in Table 2 and check  the data display  The corresponding value on the data display must be within the range shown                in Table 2   Table 2  Reference Temperature   Resistance Box Resistance Data Display                       20 921 6  20 0  0 3   25 1097 3  25 0  0 3   130 1498 3  130 0  0 3                   3 4      Indication Check    Following Section 3 4  press the  ENT  key until the message display shows    PH INP     In this  state  change the simulation input as shown in Table 3 by means of the standard  voltage current source and check the data display  The corresponding value on the data display  must be within the range shown in Table 3     QIS 12B07D02 01E       IM 12B07D02 01E    Appendix 12 3          3 5    3 3                Table 3  Check Point  pH  Simulation Input  mV  Data Display  pH   0  414 1 0 00  0 01  7 0 0 7 00  0 01  14  414 1 14 00  0 01                          Indication Check    Following Section 3 5  press the  ENT  key  The message display shows    ORP INP     In this state   change the simulation input as shown in Table 4 by means of the standard voltage current source  and check the data display  The corresponding value on the data display must be within the    range shown in Table 4                 Table 4  Simulation Input  mV  Data Disp
135. nel mounting using two  2  self tapping screws  refer to Fig  3 2b        Surface mounting on a plate  using bolts from the back      Wall mounting on a bracket  for example  on a solid wall      Pipe mounting using a bracket on a horizontal or vertical pipe  nominal pipe diameter JIS 50A        202 7 95  Panel thickness Unit  mm  inch     155 6 1  ENT 1to 10 Panel mounting         N  0 04 to 0 39  bracket              TL 80    16                                EY Hood  Option   Option code     HO     3                                                                 a 115  4 5   Mise s  1 65  PANEL CUTOUT  screw  fo my T  68  2 7   9   173     138   0 35   Sensor cable inlet  6 81   5 43        gland  Pg13 5   cA M6 screw  32  1 26  depth   4211   Transmission signal cable inlet  4 76  5         Cable gland   Pg13 5 1        Fig  3 2    Panel mounting diagram                                                    M   49 Approx    1 93   55  Adapter  2 2   i        G1 2 screw   AFTG   1 2 NPT screw   ANSI  2           Fig  3 1  Housing dimensions and layout of glands    IM 12B07D02 01E    3 2 Installation and wiring                                  min  203 Unit  mm  inch    min  8 0  1549    30  1 18   ge  ts       B x  NIS       D  cle                   2   04    1 3 5  2s  0 16   0 14   SPACING PANEL CUTOUT PANEL CUTOUT  Fig  3 2b  Panel mounting using two  2  self tapping screws Unit  mm  inch   Pipe mounting Pipe mounting   Vertical   Horizontal   ae  200   7 87              
136. nfiguration allows easy start up  the  configuration should be adjusted to suit each particular application  An example of an adjustable item   is the type of temperature sensor used  The EXA can be adjusted for any one of eight different types of  temperature sensors     To record such configuration adjustments  write changes in the space provided in Chapter 10 of this  manual  Because the EXA is suitable for use as a monitor  a controller or an alarm instrument  program  configuration possibilities are numerous     Details provided in this user   s manual are sufficient to operate the EXA with all Yokogawa sensor    systems and a wide range of third party commercially available probes  For best results  read this man   ual in conjunction with the corresponding sensor user   s manual     IM 12B07D02 01E    1 4 Introduction    IM 12B07D02 01E                2      202 SPECIFICATIONS    2 1  General    A  Input specifications    Dual high impedance inputs  2 x 10120  with  provision for liquid earth connection  Suitable  for inputs from glass or enamel pH  amp  refer   ence sensors and ORP metal electrodes     B  Input ranges      pH    2 to 16 pH    ORP    1500 to 1500 mV    rH   0 to 55 rH      Temperature    30 C to 140       20 to 300 F     8  55   NTC sensor   10  C to 120  C  10 to  250   F     10kO PTC    20  C to 140  C  0 to 300  F   C  Output ranges      pH   min 1 max 20 pH    ORP   min 100 max 3000 mV    rH   min 2 max 55 rH      Temperature   min  25      max  200
137. ns capability to meet the most advanced requirements  The meas   urement can be used as part of an automated process control system  It can also be used to indicate  dangerous limits of a process  to monitor product quality  or to function as a simple controller for a dos   ing neutralization system     Yokogawa designed the EXA analyzer to withstand harsh environments  The converter may be installed  either indoors or outside because the IP65  NEMA 4X  housing and cabling glands ensure the unit is  adequately protected  The flexible polycarbonate window on the front door of the EXA allows pushbut   ton access to the keypad  thus preserving the water and dust protection of the unit even during routine  maintenance operations    A variety of EXA hardware is optionally available to allow wall  pipe  or panel mounting  Selecting a prop   er installation site will permit ease of operation  Sensors should normally be mounted close to the con   verter in order to ensure easy calibration and peak performance  If the unit must be mounted remotely  from the sensors  WF10 extension cable can be used up to a maximum of 50 metres  150 feet  with a           junction box  Except installations with dual high impedance sensors  where the maximum cable  length is 20 metres using integral cable only  no junction box      The EXA is delivered with a general purpose default setting for programmable items   Default settings  are listed in Chapter 5 and again in Chapter 10   While this initial co
138. nts   Electrical data of the PH202S F  amp  PH202S P     Supply circuit    Maximum input voltage Vmax   24 V   Maximum input current Imax   250 mA   Maximum input power Pi 1 2 W   Effective internal capacitance Ci   220 pF    Effective internal inductance Li   0 pH     Sensor input circuit    Maximum output voltage Vt   14 4 V    Maximum output current It   32 3 mA   Maximum allowed external capacitance Ca   600 nF   Maximum allowed external inductance La   34 mH    Any FM Approved barrier may be used that meets the following requirements   Voc or Vt  lt  24 V   loc or It  lt  250 mA                     lt  1 2 W         gt  220 pF   Ccable  La  gt  0 UH   Lcable    When installing this equipment  follow the manufacturer   s installation drawing    Installation should be in accordance with ANSI ISA RP 12 06 01    Installation of Intrinsically  Safe Systems for Hazardous  Classified  Locations    and the National Electrical Code   ANSI NFPA 70     Associated apparatus connected to the barrier must not use or generate more than   250 Vrms or Vdc    Resistance between Intrinsically Safe Ground and earth ground must be less than 1 0 Ohm   In case of using cable glands in Outdoor location  they shall be UV rated or made of metal     IM 12B07D02 01E    Specification 2 15    WARNING     Substitution of components may impair Intrinsic Safety     To prevent ignition of flammable or combustible atmospheres  disconnect power  before servicing or read  understand and adhere to the manufa
139. nts may impair Intrinsic Safety    To prevent ignition of flammable or combustible atmospheres  disconnect power before servicing  or read  understand and adhere to the manufacturer slive maintenance procedures     Application Doc  No   IKE024 A10 P 4 1 to P 4 2 IM 12B07D02 01E    2 10 Specification    2 7  Control Drawing of PH202S mA HART   Specification  FM Non incendive design         Nonincendive design             FM Class     Div 2  Group ABCD  T4 for ambient temp   lt  55  C FM Approved  power supply  PH202S transmitter Voc   31 5 VDC                ES      Lo                   0                                             Sensor s  F al data    or electric  terminals 11 17 see text below  Max  cablelength  60 mtr    Cable dia    3    12 mm                 Functional  eart         Load                                           Resistance  Classified Location U nclassified Location  Nonincendive design  FM Class I  Div 2  Group ABCD  T4 forambient temp   lt  55          2028 transmitter EM  Approved  power supply  Voc  amp  31 5 VDC     r o     to        o      o      o                       Sensor s   terminals 11 17  Max  cablelength  60 mtr   Cable dia   3    12 mm       or electric  see text below a  Functional  earth                1 Classified Location           Unclassified Location                  Electrical data of the PH202S      Supply circuit  terminals   and      Maximum input voltage Vmax   31 5 V  Maximum input power Pmax   1 2 W  Effective internal cap
140. ode  To disable Auto Return  set the service code 50 to Off  0                              you disable Auto Return function then the transmitter does not automatically return to Measurement  Mode  You need to press  MODE  key to return to Measurement mode     11 6 5  CALEND    After CAL END appears once during calibration  it blinks and calibration starts again    For the PH202  even after CAL END is displayed  a measurement stability check continues for the   time interval set by service code 20  If the reading changes significantly  during this interval   then   recalibration will start automatically  In auto calibration mode  AUT CAL  immediately after CAL END   is displayed  you can press YES or NO within time interval set by service code 20 to confirm or reject    repeat  the reading    This applies to both calibration points    1  If the reading does not stabilize  and CAL END appear  within 10 minutes of start of calibration   then error E1  measurement unstable during calibration  is generated    2  If you press YES or NO then you can continue immediately to the next step     3  If you don t press YES or NO and the pH is stable  within  pH  for time interval  t    then calibration  continues to the next step     11 6 6  How to cancel sample calibration    In case of stopping calibration at data aquisition situation Indication on LCD   Key Operation     waiting for measured value input with blinking   SAMPLE    after sample calibration start  cancel the calibration in
141. om the last 3 cm of the brown  and the white coaxial cores                    cotton threads             Fig  3 12b     5  Extract the coaxial cores from the braid  and trim off the black  low noise  screening material as short  as possible                       6  Insulate the overall screen and the 2 coaxial screens with suitable plastic tubing   7  Strip and terminate all ends with suitable  crimp  terminals and identify with numbers as shown   Fig  3 12c     8  Finally shrink the overall heat shrink tube into position     IM 12B07D02 01E    Operation 4 1    4  OPERATION  DISPLAY FUNCTIONS AND SETTING    4 1  Operator interface   This section provides an overview of the operation of the EXA operator interface  The basic procedures  for obtaining access to the three levels of operation are described briefly  For a step by step guide to  data entry  refer to the relevant section of this user s manual  Figure 4 1 shows the EXA operator inter   face     LEVEL 1  Maintenance   These functions are accessible by pushbutton through a flexible front cover window  The functions make  up the normal day to day operations that an operator may be required to complete  Adjustment of the  display and routine calibration are among the features accessible in this way   See table 4 1      LEVEL 2  Commissioning   A second menu is exposed when the EXA front cover is removed and the display board is revealed   Users gain access to this menu by pressing the button marked   in the lower right of
142. onsulted   Damage caused by wear and tear  inadequate  maintenance  corrosion  or by the effects of  chemical processes are excluded from this war   ranty coverage     In the event of warranty claim  the defective  goods should be sent  freight paid  to the service  department of the relevant sales organization for  repair or replacement  at Yokogawa discretion    The following information must be included in the  letter accompanying the returned goods      Part number  model code and serial number  Original purchase order and date   Length of time in service and a description of  the process   Description of the fault  and the circumstances  of failure   Process environmental conditions that may be  related to the installation failure of the device  A statement whether warranty or non warranty  service is requested   Complete shipping and billing instructions for  return of material  plus the name and phone  number of a contact person who can be  reached for further information     Returned goods that have been in contact with  process fluids must be decontaminated disin   fected before shipment  Goods should carry a  certificate to this effect  for the health and safety  of our employees  Material safety data sheets  should also be included for all components of  the processes to which the equipment has been  exposed              Documentation    This procedure is only applicable to the countries  in European Union     All instruction manuals for ATEX Ex related prod   ucts
143. otal of capacitance and inductances connected to the terminals of the EXA      2025 do not exceed the limits given      the certificate of the safety barrier or distributor       The cable used should preferably have a BLUE colour or marking on the outside       Installation for  sensors in Zone 0 or 1    Generally  the distributor with input output isolation has no external earth connection  If there is an  earth connection on the distributor and the external connection of the transmitter is connected to   protective  earth  the shield of the 2 wire cable may NOT be connected to  protective  earth at the  distributor too     IM 12B07D02 01E    Installation and wiring 3 5    3 3 3  Installation in  Hazardous Area Non Incendive                    2025           be installed      a Category 3  Zone 2  Div 2 area without the use of safety barri   ers  Maximum permissible supply voltage 31 5V    3 3 4  Liquid earth   In all circumstances  the sensor side of the measuring loop must be grounded to the measuring liquid          EXA PH202S uses advanced differential high impedance input circuits  This technique calls for a  grounding to the liquid  In addition to that the sensor checking circuits also use the liquid earth for meas   urement of impedance of the sensors  All Yokogawa fittings have provisions for this connection  It is usu   ally called liquid earth in all our manuals    A separate connection should be made to the terminal numbered 14 in all cases to get a proper and st
144. parate temperature sensor  rather than  an all in one sensor  use the circular packing with four holes  Plug the unused hole in the packing  with the black plug  When using this packing  you have to be careful to ensure that it is well enough  sealed to be waterproof     Requirement of connecting with external instruments    Terminal for pin cable terminal   Screw terminal  option  TB   Crimp contact   Pin shaped crimp contact Ring shaped or fork shaped   for cable with sleeve insulator crimp contact   Usable contact   max  2 5 mm Crimp contact shown as the figures          under this table  which meets M3 screw  Torque for fixing   0 5 N m or less 1 35 N m  recommended    Example of Weidmuller Co   Ltd  made  JST  Mfg  Co   Ltd  made    crimp contact      0 34 10    0 5 12  H1 12    1 5 125   VD1 25 3  Ring shape     VD1 25 S3A  Fork shape                      Note  Other crimp contact may be required   depending on core cable diameter   T3 2E eps                 NN    Pin shaped terminal Ring shaped terminal Fork shaped terminal    2 5mm or less      8 3mm or less   8 3mm or less          F37EPS    IM 12B0702 01E                   11 5    When the option of the screw          TB  is specified  the 2  block layout is as follows   12 17 LOW IMP          NC HIGH IMP    Note  Nothing is connected to NC F3 8E eps    When a pH sensor is used  connect both terminals LOW and IMP on the upper row with a shorting  bar and nothing to both terminals HIGH and IMP on the lower row    Whe
145. pears in the display  and the  YES NO key prompt flags flash  Press YES     NEW SNS  Display flashes YES NO  answer YES if new sensor fitted or NO if not   Care  YES resets logbook calibration data                 Insert the sensors in pH 7 buffer solution   Press YES to start calibration                                         YOKOGAWA 4                                                           Note       start calibration with another solution  press      to cycle through the other options  YES   CAL7  NO  CAL4        CAL9  NO  CAL 7                       The instruments waits for  the reading to stabilize    The pH display flashes   When reading Is stable   the CAL END message    appears  rol       L  HL    Press YES for single point   As Pot  adjustment   Press NO to proceed to calibrate Slope                 l          YES After briefly displaying WAIT  the display  returns to the normal readout           NO                      Transfer to second buffer      4  and press YES                          NO    The display now flashes the  pH value  The instrument  waits for the reading to stabilize             When the reading is stable  the CAL END  message appears  Press YES     After briefly displaying WAIT  the display returns to  the normal readout                 IM 12B07D02 01E    Calibration 6 3           6 5 2  Automatic calibration with HOLD active       12B6C3 31    Press the MODE key  AUT CAL appears in the display  and the  YES NO key prompt flags flash  Press Y
146. pply Voltage          o            0    Q          O 0   0  Uo   31 5 Volt DC     o G     Io   100 mA      0                SENSOR S       terminals 11 17 Fun  tional Functional  earth earth Load  Resistance  4 Ianue area p Safe area  Intrinsically safe design             standard EX ia IIC  T4 for ambient temp   lt  55  C EX ia or ib Certified Repeater  T6 for ambient temp  lt  40  C Power Supply  Certificate nr  IECEx KEM 06 0052X  HART Compatible       2025  pH ORP transmitter       0     Output     ro O   O      to On                 Uo   31 5 Volt DC            100 mA O       Supply      o Po   1 2 Watt o      SENSOR S   terminals 11 17 unctional  earth     lt q Hazardous area 5 4 Safe area  Zone 0 or 1    Sensor s  are of a passive type to be regarded as    simple apparatus        Electrical data of the PH2028     Supply and output circuit  terminals   and      Maximum input voltage Ui  31 5 V  Maximum input current I    100 mA   Maximum input power Pi   1 2 W   Effective internal capacitance Ci   22 nF   Effective 1nternal inductance Li   35 uH       Sensor input circuit  terminals 11 through 17    Maximum output voltage Uo   14 4 V  Maximum output current Io   32 3 mA   Maximum allowed external capacitance Co   600 nF   for PH202S E  C  U           3 5 uF  for PH202S N    Maximum allowed external inductance Lo   34 mH  for PH202S E  C  U    Lo   76 mH  for     2025          Barriers and power supply specification must not exceed the maximum values as shown  in the dia
147. provals entity approved and meet connection  requirements    Electrical data of the PH202S B  amp  PH202S D      Supply circuit  Vmax 32 V  Pi 5 32 W        220 pF  Li  0          Sensor input circuit  Vt 14 4 V  It 32 3 mA         2 29 uF  La   64 96 mH   When installing this equipment  follow the manufacturers installation drawing   Installation shall be in accordance with Article 501 4 B  of the National Electrical   Code  ANSI NFPA 79     Non incendive field wiring may be installed in accordance with Article 501 4 B  3   Grounding shall be in accordance with Article 250 of the National Electrical code    In case of using cable glands in Outdoor location  they shall be UV rated or made of metal     WARNING      Substitution of components may impair suitability for Division 2      Do not remove or replace while circuit is live unless area is know to be non hazardous     Explosion Hazard   Do not disconnect equipment unless area is know to be  non hazardous     Do notreset circuit breaker unless power has been removed from the equipment or the  area is know to be non hazardous    Application Doc  No   IKE025 A10 P 9    IM 12B07D02 01E    2 20 Specification    2 15  Control Drawing of PH202S FF PB Specification  CSA      CSA Ex ia Class     DIV  1  Group ABCD  T4 for ambient temp   lt  55   C      Ui 24V or Ui 17 5V  li   250 mA li   380 mA    Pi 1 2W Pi  5 32 W    PH202S F  or PH202S P                  Sensor  Connections  Safe area  O 1     1 5           1 5  i  Apparatus  
148. r activates the    SAMPLE    calibration routine  at the same time as taking a representative  process sample  After determining the pH of this sample by independent methods   in the lab for exam   ple  the reading can be adjusted  While the sample is being analyzed  EXA holds the sample data in  memory  while continuing to control and read pH normally     6 4  Data entry   In special circumstances  users can directly enter the calibration data in the service code menu  see  chapter 5   This is appropriate where the manufacturer provides calibration data for each probe  as with  the Pfaudler sensors  or where electrodes are laboratory calibrated for subsequent installation on the  plant    Service Code 23 allows the values of ITP  asymmetry potential  or zero point  and slope to be entered     NOTE            sure to use fresh  pure standard solution to ensure that the calibration is accurate      If the temperature of the standard solution changes then its pH will also change   Ensure that the temperature of the standard solution is stable at calibration time            using a pH sensor which isn t incorporated with a temperature element for automatic temperature  compensation  the temperature of measured solution should be set on the PH202 transmitter      When an antimony sensor is applied  measured temperature value also should be set     IM 12B07D02 01E    6 2 Calibration    6 5  Calibration procedures    6 5 1  Automatic calibration       Press the MODE key  AUT CAL ap
149. reference elements in the same body    Green for temperature sensors    The recommended procedure is to color code each end of the cables to match the sensors with the color  strips provided with each cable  This provides a quick way to identify the ends of the cables belonging to  a particular sensor when they are installed   The procedure for fixing the identification labels is described  in detail in the instruction sheet provided with the cable      IM 12  07002 01      Installation and wiring 3 9  3 6 1  Connection cable  There are two types of connection cable  one for single sensors and one for combined sensors  The  former is a coaxial cable and has only two connections     Red to measuring element    Blue to screen  shield     The latter is a triaxial cable with three connections   it has an extra white wire termination  these wires  are connected      Red to measuring element     Blue to reference     White to screen  shield     To connect the other sensor systems  follow the general pattern of the terminal connections as listed  below     11  amp  12 Temperature compensation resistor input   13 Input no  2  normally the reference element   17 Screen  shield  for input no  2   14 Liquid earth  solution ground  connection   15 Input no  1  normally the measuring element   16 Screen  shield  for input no  1          pH transmitter  Green     Red 11 Temperature       11 Temperature   12 Temperature      Yellow    Red    12 Temperature   __ 13 Reference 7    f  14 Solut
150. required  and the  sensor input circuit is non incendive having the parameters   Maximum output voltage Uo   14 4 V  Maximum output current lo   32 3 mA  Maximum allowed external capacitance Co   3 5 uF  Maximum allowed external inductance Lo   76 mH    IM 12B07D02 01E    Installation and wiring 3 1    3  INSTALLATION AND WIRING  3 1  Installation and dimensions    3 1 1  Installation site   The EXA transmitter is weatherproof and can be installed inside or outside  It should  however  be  installed as close as possible to the sensor to avoid long cable runs between sensor and transmitter  In  any case  the cable length should not exceed 50 meters  162 feet   Select an installation site where   Mechanical vibrations and shocks are negligible   No relay power switches are in the direct environment   Access is possible to the cable glands  see figure 3 1    The transmitter is not mounted in direct sunlight or severe weather conditions   When the instrument with Suffix Code   B  N  D  is used  take measures so that the display window is  not exposed to direct sunlight   Maintenance procedures are possible  avoiding corrosive environments     The ambient temperature and humidity of the installation environment must be within the limits of the  instrument specifications   See chapter 2      3 1 2  Mounting methods   Refer to figures 3 2 and 3 3  Note that the EXA transmitter has universal mounting capabilities        Panel mounting using optinal bracket  refer to Fig  3 2a      Pa
151. rom anywhere in the configuration menus  when no button is pressed during  the set time interval of 10 minutes     The manual impedance check  on demand  can be setup for operation in the  maintenance mode   Through the closed front cover      Passcodes can be set on any or all of the access levels  to restrict access to  the instrument configuration     Error message configuration  Two different types of failure mode can be set     Hard fail gives a steady FAIL flag in the display  A Fail signal is transmitted on  the outputs when enabled in code 32     Soft fail gives a flashing FAIL flag in the display  The call for maintenance is   a good example of where a SOFT fail is useful  A warning that the regular  maintenance is due  should not be used to shut down the whole measurement   Not used   Call for maintenance is a trigger to signal that the system has been in service  for longer than the set time without calibration  The user can set up to 250 days    as a routine service interval     The display resolution can be set to either 0 01pH or 0 1pH  Not applicable to  the ORP  mV  display     IM12B07D02 01E    5 20 Parameter setting                               Code   Display Function Function detail X Y 2 Default values  User interface  50  RET  Auto return Auto return to measuring mode Off 0  Auto return to measuring mode On 1 1 On  51  MODE Mode setup Manual impedance check Off 0 0 Off  Manual impedance check On 1  52  PASS Passcode Maintenance passcode Off 0 0 0 0   Off
152. s ensures that the desired control  point will be measured accurately  even if there may be small deviations when there is a big deviation  from the setpoint  This of course has no effect on the accuracy of a control loop  The special construc   tion of the Pfaudler sensor ensures that there is practically no drift in the calibration  All that is necessary  is to keep the sensor membranes clean  This is best done by cleaning with low pressure steam  which  restores the original condition of the sensor  including the original calibration values     IM 12B07D02 01E                   10 7    10 6  Device Description  DD  menu structure  The Device Description  DD  is available from Yokogawa or the HART foundation  An example is shown  below of the ON LINE menu structure  This manual makes no attempt to explain the operation of the  Hand Held Terminal  HHT   For detailed operating instructions  refer to the HHT user   s manual and the  on line help structure  For menu structure of HHT 375  see next page     Level 1 menu Level 2 menu Level 3 menu Level 4 menu Level 5 menu                                                                            Process variab              Process value  Second process value  Temperature  0                   Slope                     Aspot2  Imp1  Imp2  Diag Service Status  Hold         Hold on off  Hold type  Hold value  Temp Man   Manual temp value  Manual temp on off   Logbook1  Logbook2  Basic setup          Tag  Unit  Device info     Date    
153. s signalled     AS HI     ZP LOW  amp    ZP HI     SL LOW  amp    SL HI    Code 23             SLOPE    Code 24   25   amp  26    Code 27     amp              BUF ID     ZERO P    when a calibration is done  These default values should be adjusted to suit the  application  this will be especially important with enamel or Antimony probes    In case in service code 27 the Asymmetry Potential is disabled and the Zero  Point is used  service code 21 is used for entering the limits of the Zero Point     Limit values for zero point  if enabled in service code 27     Limit values for acceptable slope  sensitivity  calibrations     Values can be entered directly in this section  These data can be provided by  the manufacturer of the probe  or by the users laboratory etc  They are  determined independently of the measuring loop    Note  it is not necessary to enter this data in most cases as the EXA  automatically does this while performing a calibration  The feature is  used in the case of special electrode systems and where calibration in  the process environment is not possible     The following buffer calibration tables are programmed into the EXA  They are  the primary buffer standards according to NIST  formerly NBS  and various other  national standards  We strongly recommend the use of these buffer solutions as  they give the best buffer capacity  reliability and accuracy when calibrating     Table 5 1                                These tables may be adjusted in the case tha
154. sation and measuring functions                                                    5 13  5 3 3  Calibration TUMCt ONS      5 15  5 3 4  MA output                                                                 5 17  5 3 5  User Interface    oci d arto             5 19  5 3 6  Communication                                                                       5 21  5 3 7  General e 5 21  5 3 8  Test and set  p          arae      5 21   6  Calibration      6 1  6 1  Automatic calibration                             6 1  6 2  Man  al calbration                                               E 6 1  6 3                                                                 6 1                    mom 6 1  6 5  Calibration                 c 6 2  0 5 1  Automatic  calibration    ctr tnt t o        6 2  6 5 2  Automatic calibration with HOLD active                   ssssssssssesseneens 6 3  6 5 3  Manual calibration  2nd parameter                                        6 4  6 5 4  Sample calibration                          6 6  7                                   m 7 1  7 1  Periodic maintenance for the EXA transmitter                                           7 1  7 2  Periodic maintenance for the sensor                                                  4   1      7 1  7 3  Calibration procedures are described in step by step detail in chapter 6              However  follow these       6                                                                                                           
155. sensor maybe  E 13 Reference   connected to 11  amp  12  for     7 14 Solution ground QRP  temperature indication     15 pH I 2  See section 5 3 1 for    16 Shield   impedance checking of glass     17 4 reference sensors                 ORP rH wiring                                            Note                or rH setting in service  code 02                    Figure 3 10b  Sensor wiring    3 6 2  Sensor cable connection with special grommet   In order to seal multiple sensor cables into EXA  a special grommet is provided that is designed to  accommodate one  two or three sensor cables  5 mm dia   plus a liquid earth cable  2 5 mm dia    In the  pack with the grommet are blanking pieces to close any unused holes  When correctly assembled  the  grommet maintains the IP65  NEMA 4X rating of the EXA PH202 housing     Refer to figure 3 5 to assemble the grommet connections    First remove the nut and standard rubber seal from the selected gland   Discard the seal  This will be replaced later by the special grommet   Thread the cables through the nut and the gland   Connect the cables to their designated terminals   Arrange the cables to avoid tangles and insert the grommet between the gland and the nut   The grommet is split to permit the cables to be mounted after connection   This also ensures even  length adjustment     Ensure that any unused holes are filled with the blanking pieces    Tighten the nut to form a firm seal   Hand tight is sufficient      ON OURWN gt     NOT
156. slope and  asymmetry are made  This eliminates the question of how long the operator should allow prior to adjust   ment  A menu driven prompt system conducts the operator through the simple  foolproof routine     Default settings for the buffer solutions are the standard NIST  formerly NBS  recognised solutions    4         7    and    9     These are known as primary buffers  They have a much better buffer capacity than the     commercial    or adjusted buffers  Yokogawa strongly recommends the use of these buffers to provide  the best pH calibration     6 2  Manual calibration  In this method  the operator decides on the actual value to enter  Manual calibration is most often used  for single point adjustment of the asymmetry potential  by comparison method     Manual calibration can also be used to perform a full 2 point calibration with solutions other than the  NIST buffers that are listed in the calibration tables  In this case  the solutions are applied sequentially as  in the AUT CAL method  but the user determines the adjustment of reading and stability     NOTE   During manual calibration the temperature coefficient is still active  This means that the readings  are referred to 25   C  This makes grab sample calibration easy and accurate  However  if the  manual calibration technique is used for buffer calibration  the temperature coefficient must be set  to zero in maintenance mode in the    TEMP    routine  see chapter 5      6 3  Sample calibration   The operato
157. t  terminals 11 through 17    Maximum output voltage Voc   14 4     Maximum output current Isc   32 3 mA   Maximum allowed external capacitance Ca   600 nF  Maximum allowed external inductance La   34 mH     Barriers and power supply should be CSA certified  The specifications must not exceed  the maximum values as shown in the diagram above   Installation should be in accordance with Canadian Electrical Code  Part I   Maximum safe area voltage should not exceed 250 VRMS   For Class I  Div  2  Group ABCD the CSA certified barrier is not required  and the Sensor  input circuit  terminals 11 through 17  is non incendive having the parameters    Maximum output voltage Voc   14 4 V  Maximum output current Isc   32 3 mA   Maximum allowed external capacitance Ca   3 5 uF   Maximum allowed external inductance La   76 mH   The Hand Held Communicator must be of a CSA certified intrinsically safe type  in case it is used on the intrinsically safe circuit in the hazardous area  or of a  CSA certified  non incendive type in case it is used on the non incendive circuit in the  hazardous area     100 mA     IM 12B07D02 01E       2 12 Specification    2 9  Control Drawing of PH202S FF PB Specification  IECEx                           Ex ia       T4 for ambient temp   lt  55   C    Ui  24V or Ui 17 5V     250 mA liz 380 mA    Pi 1 2W       5 32          2025                2025                       Connections  Safe area  O 1 5  p  deo    POS     Apparatus   interface    certified   i certi
158. t maintained in  preset time period    Perform maintenance  Reset interval       Output span too small  lt 1pH    Incorrect configuration by user    Reprogram          Table values make no sense  Programmed values outside acceptable limits    Incorrect configuration by user    Reprogram       Programmed data lost    Fault in electronics  Very severe interference    Reprogram       Checksum error    Software problem    Contact Yokogawa       Zeropoint outside limits       Sensors are aged or poluted  Mistake in calibration       Note 1  see section 11 6 2  Note 2  see section 11 6 3        Check buffer solution  Recalibration at pH7  Replace sensor    T11 1E eps    IM 12B07D02 01E    11 14 Appendix    11 6 2  On line impedance checks  This check can be applied to a sensor with liquid earth electrode  In using a sensor having no liquid  earth electrode  set impedance check off on service code 3 and 4     Display Function Function detail Ddefault values    Parameter specific    functions       03  71           Impedance check 1    Low impedance limit    x None  K  MorG    High impedance limit    x None  K  M or G    Low   High   Temp comp off  Temp comp on  Imp check off    Imp check on    Press NO to step through choice of units     press YES to select units  then use the     gt          keys to set the value    Press NO to step through choice of units     press YES to select units  then use the     gt   NENT keys to set the value        72              Impedance check 2    Low
159. t temp   lt  55  C  T6 for ambient temp   lt  40  C    safety barrier or power supply   HART compatible                                                                                              202  pH ORP transmitt        24 volts DC Nominal  Supply Voltage      ro        ro T         o    Sensor s      o G Suitable values are   terminals 11 17 Vmax   31 5 VoltDC  o Imax   100 mA       LI  For electrical data   see text below          Functional Functional  earth earth  Resistance     Hazardous area p   4 Safe area  Intrinsically safe design  CSA Ex ia Class I  Div 1  Group ABCD CSA certified  T4 for ambient temp   lt  55  C Power Supply  T6 for ambient temp   lt  40  C  HART compatible         2028   pH ORP transmitter                 Output     _o            o           Sensor s  Suitabl 1       Lo litable values are         terminals 11 17    im niis  max   31 5 VoltDC     Lo Imax   100 mA O       Supply  Pmax   1 2 Watt      O           For electrical data   see text below       Functional  earth  4 Hazardous area    P Safe area                  Sensor s  are thermocouples  RTD   s  passive resistive switch devices  or are CSA entity approved  and meet connection requirements     Electrical data of the     2025     Supply and output circuit  terminals   and      Maximum input voltage Vmax   31 5 V  Maximum input current Imax    Maximum input power Pmax   1 2 W   Effective internal capacitance Ci   22 nF  Effective internal inductance Li   35 uH     Sensor input circui
160. t the user wishes to use other  calibration solutions  The                of the buffer can be changed at the  BUF ID  prompt  The other values can then be adjusted in sequence     As an alternative to Asymmetry Potential  the Zero point can be used to define  and calibrate the EXA pH unit    Note that this method conforms to the DIN standard for instruments   No  IEC 60746 2     IM12B07D02 01E    5 16 Parameter setting       Code Display Function    Function detail    Default values       Calibration functions                                                                         20  ALSEC Stability check time 5 sec    APH Stability check pH 0 02 pH  21  AS LOW As Pot low limit  120 mV   As Pot   AS HI As Pot high limit 120 mV  21  ZP LOW Zero Point low limit 5 00 pH   Zero    ZP HI Zero Point high limit 9 00 pH  22  SL LOW Slope low limit 70 96   SL HI Slope high limit 110 96  23  ITP Set ITP Preset calibration data from manufacturer 7 00 pH   pH   SLOPE Set slope or from laboratory determinations  100     ASP 1D Set As Pot For the main parameter 0 0 mV  A Press YES to confirm 0 1 mV resolution   Select YES NO then set value with  gt      ENT keys   Y Press NO to change to  ASP    ASP Set As Pot For the main parameter  Press YES to confirm 1 mV resolution   then set value with  gt      ENT keys    ASPmV Set As Pot ORP For parameter 2  when activated in  service code 02   23  ASP 1D Set As Pot  ORP  For the main parameter   ORP  fi Press YES to confirm 0 1 mV resolution   S
161. tandard buffers be used to calibrate solutions  The temperature responses of these are  pre programmed into Service codes 24  25  and 26 in the EXA PH202  Where other buffers are used  with the semi automatic calibration function  their temperature response should be programmed into  the relevant code     IM 12B07D02 01E                   10 5    10 4  Set up for other functions    Current outputs  Transmission signals for the measured parameters and FAIL signals can be set up in service codes  31  32 and 35       Diagnostic Checks  Impedance checks  response time and stability checks are all included in the PH202  In order to get the  best performance from each of these features  the transmitter should be fine tuned according to experi   ence in the installation  and for the particular sensors selected  Service codes 3  4  5  amp  20 all contribute to  the diagnostics  Please note that the default settings provide an excellent starting point and provide most  valuable information about the performance of the electrode system       Communications  The proprietary HART  FSK  communication link allows remote configuration and data retrieval  through the PC202 communication package  This is an excellent tool for the maintenance engineer   quality engineer or plant manager  Service codes 60 62 are used to set up the communications     e Logbook  In combination with the communications link  a    logbook    is available to keep an electronic record of  events such as error messages
162. tati cel mai apropiat  birou sau reprezentant Yokogawa                 manwali kollha      1 istruzzjonijiet ghal prodotti  niarbuta ma    ATEX      huma disponibbli bl Ingli     bil   ermani   u bil Fran  i    Jekk tkun te  tie    struzzjonijiet marbuta ma    Ex fil lingwa lokali tieg  ek   g  andek tikkuntattja lill eqreb rappre  entan jew  uffi    ju ta  Yokogawa    1  INTRODUCTION AND GENERAL DES    CRIPTION    Introduction 1 1    The Yokogawa EXA 202 is a 2 wire transmitter designed for industrial process monitoring  measurement  and control applications  This user   s manual contains the information needed to install  set up  operate  and maintain the unit correctly  This manual also includes a basic troubleshooting guide to answer typi     cal user questions     Yokogawa can not be responsible for the performance of the EXA analyzer if these instructions are not    followed           1 1  Instrument check          Upon delivery  unpack the instrument          carefully and inspect it to ensure that  it was not damaged during shipment   If damage is found  retain the original                      packing materials  including the outer  box  and then immediately notify the          carrier and the relevant Yokogawa          sales office        Make sure the model number on the  textplate affixed to the side of the  instrument agrees with your order   Examples of nameplates are shown                                                                                    TRANSM
163. tatsiooni   saamiseks p    rduge l  hima Iokagava  Yokogawa   kontori v6i esindaja poole     IM 12B07D02 01E    Wszystkie instrukcje obstugi dla urzadzen   w wykonaniu przeciwwybuchowym Ex   zgodnych z wymaganiami ATEX  dostepne       w Jezyku angielskim  niemieckim i  francuskim  Jezeli wymagana jest instrukcja   obstugi w Panstwa lokalnym je zyku  prosimy     kontakt 7 najblizszym biurem Yokogawy     Vsi predpisi in navodila za ATEX Ex sorodni  pridelki so pri roki v angli amp   ini  nems  ini ter  franco amp   ini  Ee so Ex sorodna navodila potrebna v  va  em tukejnjem jeziku  kontaktirajte va   najblisi  Yokogawa office ili predstaunika     o    Az ATEX Ex m  szerek g  pk  nyveit angol  n  met    s francia nyelven adjuk ki  Amennyiben helyi  nyelven k  rik az Ex eszk  z  k le  r  sait  k  rj  k  keress  k fel a legk  zelebbi Yokogawa irod  t  vagy   k  pviseletet                                                                               ATEX                                                                                                                                                                                                                                                                                                                                                            Yokogawa     Toate manualele de instructiuni pentru produsele  ATEX Ex sunt in limba engleza  germana si  franceza  In cazul in care doriti instructiunile in  limba locala  trebuie sa contac
164. than for NIST solutions     NOTE   NIST  formerly NBS  buffers are available as consumable items from any Yokogawa sales office  under the following part numbers   6C232 4 01 pH at 25  C   6C237 6 87 pH at 25  C  A box contains 5 packets of powder  Each makes a 200 ml solution   6C236 9 18 pH at 25  C     IM 12B07D02 01E    Troubleshooting 8 1    8  TROUBLESHOOTING                     is    microprocessor based analyzer that performs continuous self diagnostics to verify that it  is working correctly  Error messages resulting from faults in the microprocessor systems itself are few   Incorrect programming by the user can be corrected according to the limits set in the following text     In addition  the EXA also checks the electrodes to establish whether they are still functioning within  specified limits  The transmitter checks the glass electrode impedance for a low value to determine if it is  broken or cracked  and for a high impedance to check for internal breakage or disconnection     The reference system is prone to more faults than the glass electrode in general  The unit measures the  impedance value and compares it to the programmed value in memory to determine acceptance during  testing  A high impedance signals pollution or poisoning of the reference electrode diaphragm     Also  the EXA checks the electrodes during calibration to determine if the reaction time is suitable for pH  measurement  A specially timed check can be activated following each cleaning cycle 
165. tion  Display of sensor  impedance on message line of display  FAIL  flag in event of    out of limits    impedance  and  the possibility of 21 mA or  3 6 mA or 3 9 mA   error signal     M  Signal processing  pH ORP      The PH202 can measure pH              Using  the FU20 allows simultaneous measurement  and display of pH and         It also allows  display and out put of pH         EMC Conformity standards             00  EN 61326 1 Class A  Table 2   For use in industrial locations   EN 61326 2 3  EN 61326 2 5  pending     A CAUTION    This instrument is a Class A product  and it is  designed for use in the industrial environment   Please use this instrument in the industrial  environment only     Specification 2 3    O  Intrinsically safe  Refer to conrol drawings                                                                                                                                                           Item Description Code Item Description Code  FM Intrinsically safe Approval FM Non incendive safe Approval  AUD 422114020 Applicable standard  FM3600  FM3611  FM3810  ntrinsically        Tor              Division 1  Groups Non incendive for Class    Division 2  Groups ABCD  Factory  Class I  Zone 0  AEx ia      Factory  2005   DT    Mutual        Temp  Class  T4  Amb  Temp    10 to 55  C   Mutual  FM    Temp  Class  T4  Amb  Temp    10 to 55      Intrinsically Safe Apparatus Parameters Non incendive Safe Apparatus Parameters  Vmax 31 5 V  Imax 100 mA  Vmax 31 
166. tion is not well executed or the pH readings are temperature dependent  Monthly calibra   tions should be sufficient for most applications     If a film remains on the pH sensor after cleaning  or if the reference junction is partly plugged  then  measuring errors can be interpreted as a need for recalibration  Because these changes are reversible  with correct cleaning and or proper selection or adjustment of the electrolyte flow through the junction   make sure that these items are correct before recalibrating the system     IM 12B07D02 01E    7 2                           7 3  Calibration procedures        described in step by step detail in chapter 6  However  follow these  guidelines     1  Before starting a calibration  make sure the electrode system is properly cleaned so that electrodes  are fully functional  They must then be rinsed with clean water to avoid contamination of the  calibration solution    2  Always use fresh buffer solutions to avoid the possibility of introducing errors from contaminated or  aged solutions  Buffers supplied as liquids have a limited shelf life  especially alkaline buffers which  absorb CO2 from the air    3  Yokogawa strongly recommends NIST  primary  buffer standards in order to ensure the best  accuracy and best buffer capacity is available  Commercially adjusted buffers  e g  7 00  9 00  or 10 00 pH  are a compromise as a standard  and are often supplied without the temperature  dependency curve  Their stability will be much worse 
167. trostatic discharge     Installation and wiring    The EXA analyzer should only be used with equip   ment that meets the relevant international and  regional standards  Yokogawa accepts no respon   sibility for the misuse of this unit     A CAUTION    The instrument is packed carefully with shock  absorbing materials  nevertheless  the instrument  may be damaged or broken if subjected to strong  shock  such as if the instrument is dropped  Handle  with care     Although the instrument has a weatherproof  construction  the transmitter can be harmed if it  becomes submerged in water or becomes exces   sively wet     Do not use an abrasive material or solvent when  cleaning the instrument     Do not modify the PH202 transmitter        WARNING    Electrostatic charge may cause an explosion haz   ard  Avoid any actions that cause the generation of  electrostatic charge  e g   rubbing with a dry cloth     Warning label      A WARNING AVERTISSEMENT  8  HRObtnibsoc                                              POTENTIAL ELECTROSTATIC CHARGING HAZARD    SEE INSTRUCTIONS   DANGER POTENTIEL DE CHARGES  ELECTROSTATIQUES   VOIR INSTRUCTIONS       Because the enclosure of the pH ORP transmitter               2025      P   F are made of aluminium  if it  is mounted in an area where the use of category   1 G Zone 0 apparatus is required  it must be  installed such  that  even in the event of rare inci   dents  ignition sources due to impact and friction  sparks are excluded     Notice      Th
168. vice Code 52 to obtain access to the Commissioning Mode     Service  From the commissioning menu  select  SERV by pressing YES key  The display shows 000 and  PASS   Enter a 3 digit passcode as set in Service Code 52 to obtain access to the Service Mode     NOTE   See Service Code 52 for the setting of passcodes     4 4  Display examples  The following pages show the sequence of button presses and screens displayed when working in some  standard configurations              or less options will be made available by the configuration of some service codes  For instance the    impedance measurement screens do not appear when impedance checking is switched off in service  codes 03 and 04     IM 12B07D02 01E    4 4 Operation  4 5  Display functions    4 5 1  Display functions pH  default                          Display Access        See Auto Impedance of input 1    calibration  Chapter 6                                             See Manual  calibration  Chapter 6  Impedance of input 2             MONI                                                           Sample  calibration  Chapter 6  Software release number                                              YES              Process temperature                                                           See Manual                                                                                                                                                              Impedance   MAN check Chapter 5  Current output   See Temp Offs
169. y         prevent ignition of flammable or combustible atmospheres  disconnect power  before servicing or read  understand and adhere to the manufacturer s live  maintenance procedures     Application Doc  No   IKE025 A10 P 6 to P 7    IM 12B07D02 01E    2 18 Specification    2 13  Control Drawing of PH202S FF PB Specification  FM Non incendive Entity      FM Class I  DIV  2  Group ABCD  T4 for ambient temp   lt  55   C            Sensor  Connections    PH202S B Max  cablelength  60 mtr     or PH202S D Cable dia   3   12 mm   Sensor   FM Approved        Connection  Power Supply     Voc  lt  32 VDC i FM Approved E     FM Approved i    i Terminator 01    R   90  1000    ic 0 22uF         Terminator            90  1000 i  1   0  2 2      i         Division 2    Unclassified Location Classified Location       Sensor s  are of a passive type to be regarded as  simple apparatus     devices which  neither store nor generate voltages over 1 5 V  currents over 0 1 A  power over 25 mW or  energy over 20 uJ  or are FM Approvals entity approved and meet connection  requirements     e Electrical data of the PH202S B  amp  PH202S D     Supply circuit  Vmax 32 V  Pi 1 2 W  Ci  220 pF  Li  0 uH    Sensor input circuit  Vt 14 4 V  It 32 3 mA  Ca 2 29uF  La 64 96 mH  When installing this equipment  follow the manufacturers installation drawing   Installation shall be in accordance with Article 501 4 B  of the National Electrical Code   ANSI NFPA 79   Nonincendive field wiring may be installed 
170. y   laboratory analysis  for example    1  Press the MODE key    2  MAN CAL is displayed    3  Press NO    4  SAMPLE is displayed   To abort calibration  press NO and S OFF is displayed  then press YES to  revert to MEAS mode     5  While SAMPLE is displayed  and when the measured value stabilizes  acquire some sample liquid  for laboratory analysis and at the same time press YES  S ON is displayed  so press YES again to  confirm  SAMPLE is displayed  flashing  on thedisplay  and the PH202 waits for a value    derived  by laboratory analysis or the like    to be entered    6  To enter the value derived by laboratory analysis or the like  while SAMPLE is lashing press the  MODE key  LAB CAL is displayed   To abort calibration  press NO and you will be returned to step   4  above     7  Press YES  NEW SNS is displayed  If you want to change the sensor  press YES  otherwise press  NO  ORP CAL is displayed    8  Press YES  START is displayed  so press YES again    9  0000 mV is displayed  so enter a suitable zero offset value and press ENT    10  CAL END is displayed    11  Press YES to exit Calibration with Sample     IM 12B0702 01E    11 6  Supplement of troubleshooting  11 6 1  Error Codes    Appendix 11 13    The following error message table gives a list of problems that are indicated when the high or low  impedance limits are exceeded for a sensor  Such things as fouling  breakage and cable faults are  readily detected  The non immersion of the sensors in the process fluid 
171. y  detected  The non immersion of the sensors in the process fluid is also signalled     IM 12B07D02 01E    Table 8 1  Error Codes    Troubleshooting 8 3                                                                                     Code   Error description Possible cause Suggested remedy  EO   Buffer solution temperature Buffer solution too hot or too cold Adjust buffer temperature  outside the programmed range Check cabling  E1 Measurement failed to stabilize  Sensors fouled Clean sensors  during the calibration Sensors too slow  aged sensor  Replace sensors  E2   Asymmetry potential too high  Sensors are aged or polluted Check buffer solution   Limits set in service code 21   Mistake in calibration Recalibrate at pH7  Replace sensor         Slope  sensitivity  is outside limits  Measuring sensor aged Replace measuring sensor   Limits set in service code 22   Poor insulation at the connector Replace or dry cables  E4 1   Impedance of input 1 too low  Measuring sensor broken Replace measuring sensor   Limits set in service code 03   Damaged or damp connections Replace or dry cable  E4 2   Impedance of input 2 too low  Reference sensor broken Replace reference sensor   Limits set in service code 04   Damaged connections Replace cables  E5 1   Impedance of input 1 too high  Measuring sensor disconnected Check connections   Limits set in service code 03   Sensors not immersed in process Check process  Liquid earth disconnected Check connections  E5 2   Impedance of input 2
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Supply List  DuraVision FDH3601 Manuel d`utilisation  Viewsonic LED LCD VA1911A-LED  Fujitsu ESPRIMO Q9000  LTC 0435 & LTC 0455 - Bosch Security Systems  Bosch Appliances HBL 74 Oven User Manual  Skill Assessment - McGraw  DC-Mike Actuators  2004 Nissan Frontier Owner Guide  困ったときのQ&A - 121ware.com    Copyright © All rights reserved. 
   Failed to retrieve file