Home
        MFI 10 Melt Flow Indexer User Manual
         Contents
1.      Because the Die Swell and Draw Down depend on an accurate measurement of the extrudate  it is  important that the cross section should be as circular as possible  The most common cause of  irregular cross section circularity is draught  causing asymmetric cooling  the equipment should be  placed as far away as possible from any draught source                    Desired Molecular  Process Characteristic Design  Blow Moulding High Melt Strength Low MFR  Film Blowing Broad molecular weight distribution  Thermoforming  Extrusion Low Die Swell Narrow molecular weight distribution  Injection Moulding Easy Flow Medium to High MFR  Moderate Temperatures Narrow molecular weight distribution  Rapid Cycles  2 0 THE MFI 10    The Melt Flow Rate apparatus has been designed in accordance with ISO 1133 and ASTM D1238   Method B requires the Displacement Transducer  AFRT   ISO 1133 now supersedes all European  Standards such as DIN 53735  Germany   NF  51 106  France   UNI 5640 74  Italy  UNE 53098   Spain   The apparatus is a dead weight extrusion plastometer with a microprocessor controlled melt  chamber  the barrel  in which  under standard set conditions  the test sample of polymer is heated  and extruded through a die with a known load  All component parts to perform the tests are calibrated  within stringent limits required by the above standards     There are two tests  the manual cut off  Procedure A  is suitable for materials with an MFR between  0 1 and 25g 10mins or for compa
2.      TRADEMARKS   AMETEK is a registered trademarks of AMETEK  Inc   LLOYD INSTRUMENTS is a trademark of AMETEK  Inc   DAVENPORT is a trademark of AMETEK  Inc    Other trademarks are the property of their respective owners     SUPPORT  AMETEK Lloyd Instruments Ltd  Steyning Way    Bognor Regis   West Sussex   UK   PO22 9ST   Tel   44  0  1243 833 370   Fax   44  0  1243 833 401  Email   techsupport ametek co uk    Internet  www lloyd instruments co uk    AMETEK Measurement and Calibration Technologies  8600 Somerset Drive  Largo  FL 33773    USA   Tel   1 727 536 7831   Tel   1 800 527 9999  within USA only   Email   chatillon fl lar ametek com    2 MFI10 User Manual    ICONS    WARNING    The raised hand icon warns of a situation or condition that may lead to personal injury    or death  Do not proceed until the warning is read and thoroughly understood  Warn   ing messages are shown in bold type     DANGEROUS VOLTAGE     gt     The lightning icon warns of the presence of an uninsulated dangerous voltage within    the product enclosure that might be of sufficient magnitude to cause serious shocks  or death  Never open the enclosures unless you are an authorized and qualified  LLOYD INSTRUMENTS    service personnel  Never open any enclosure when power  is connected to the system or its components     CAUTION     gt      The exclamation point icon indicates a situation or condition that may lead to equip     ment malfunction or damage  Do not proceed until the caution message 
3.    Part No  01 2713     I EK June 2012  TEST  amp  CALIBRATION INSTRUMENTS    MFI 10  Melt Flow Indexer    User Manual       Davenport nt    polymer test equipment    WARRANTY   This instrument is warranted against defects in workmanship  material and design for one  1  year  from date of delivery to the extent that AMETEK will  at its sole option  repair or replace the instrument  or any part thereof which is defective  provided  however  that this warranty shall not apply to instru   ments subjected to tampering or  abuse  or exposed to highly corrosive conditions     THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES WHETHER EXPRESS OR IMPLIED  AND AMETEK HEREBY DISCLAIMS ALL OTHER WARRANTIES  INCLUDING  WITHOUT LIMITA   TION  ANY WARRANTY OF FITNESS FORA PARTICULAR PURPOSE OR MERCHANTABILITY   AMETEK SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR CONSEQUENTIAL DAMAGES   INCLUDING  BUT NOT LIMITED TO  ANY ANTICIPATED OR LOST PROFITS     This warranty is voidable if the purchaser fails to follow any and all instructions  warnings or cautions  in the instrument s Instruction Manual     If a manufacturing defect is found  AMETEK will replace or repair the instrument or replace any defec   tive part thereof without charge  however  AMETEK   s obligation hereunder does not include the cost  of transportation  which must be borne by the customer  AMETEK assumes no responsibility for dam   age in transit  and any claims for such damage should be presented to the carrier by the purchaser
4.   Indicating thermometers used for testing to standardised conditions are calibrated over a range of  4  C in steps of 0 1   C  For example  an indicating thermometer for use at the standard test  temperature of 190   C is graduated from 188   C to 192   C     A standard range of these thermometers is available covering all the standard test temperatures laid  down in ISO 1133 and ASTM D1238  Long range thermometers can also be supplied covering the  working  range of 100   C to 300  C in three equal  70   x 1 0    overlapping ranges  100   170  C    165   235  C and 230   300  C     2 5 THE DIE   Made of tungsten carbide  these conform precisely to standard dimensions  8mm long  9 47mm  overall diameter  bore diameter 2 095mm  The bore is very highly finished and is tested to tolerances  well inside the closest called for in any published standard  particularly with regard to bell mouthing   The alternative standard die called for in BS 2782 method 105C  as jet B  with a bore diameter of  1 181mm  is also available on request as an extra  Please note that this is no longer called for in the  Method 720A which replaces 105C  or in ISO 1133 which replaces the BS Standard  or the ASTM  Standard  However it is still in use for polymers with exceptionally high flow rates     2 6 THE DIE PLATE   The die plate supports the die and is held in place underneath the barrel by three socket head  screws  The die plate is a piece of circular steel with three keyhole slots and a larger rou
5.  2 4  2 5  2 6  2 7  2 8  2 9  2 10  2 11  2 12  2 13  2 14  2 15  2 16  2 17  2 18  2 19    3 0  3 1  3 2  3 3    4 0    5 0  5 1  5 2  5 3  5 4  5 5    6 0    6 1  6 2    7 0    TABLE OF CONTENTS    INTRODUCTION o i 5  Melt Flow Rate    5  Other Applications of the Apparatus    5  Die SWelllscceescenstscecersesiovsrctesrapoeeccetead 6  Draw Down 6             Temperature Controller    7  Indicating Thermometers a  Dies          Die Plate   PISTON ER at aan armee  Control Panel   Automatic Flow Rate Timer     Using the Control Panel     Test PrOCEdUTES rsrsrsrsrs  TeSt SUD  urine  Remote Controle  Supervisor Mode   ais  Calibration Mode     Startup OptONS serres means  Ancillary Equipment     Additional Features           s  ss1011s11e11  Additional Requirements                 INSTALLATION o   28  Unpacking                  28  Temperature Control     ae 30  Cut Off Timer Operation    30  CALCULATION AND PRESENTATION  OF RESULTS nimes 32  OPERATING TECHNIQUES              33  Cutting Off the Extrudite              33  The Importance of Piston Grooves     33  Form of Test Sample    33  Cleaning                                       33  Differences in Test Condition                35    AUTOMATIC FLOW RATE TIMER    OPERATION  amet 36  Notes for Auto Flow Operation           36  Melt Density Calculations               36  GENERAL MAINTENANCE              36    4 MFI 10 User Manual    8 0 THERMOMETER                  0 00   37  8 1 Repairing Breaks in Mercury Column  Thermomete
6.  OFF THE EXTRUDATE   For accurate and repeatable results  the cutting off action should be done quickly and neatly  The tip  of the Cut Off knife should pass upwards along front side of the conical hole in the die plate  lightly  against the bottom of the die  cutting the extrudate in its passage and continuing down the other side  of the conical hole  The neatness of the cut may be gauged from watching the way the next portion  is extruded  If the cut has been clean and sharp  the emergent strand will point straight down  If not   there will be a tendency for it to curl  or even stick to one side of the plate hole and emerge in the  form of a loop     Until some practice in cutting off has been gained  it is helpful to smear a little silicone oil around the  inside of the conical hole  This will help to prevent miss cuts from sticking to the inside of the hole  It  is also important to avoid allowing the tip of the Cut Off knife to become hot by holding it in contact  with the side of the die plate orifice between cuts  as this will increase the tendency of the extrudate  to stick to the Cut Off knife  Keep the tip of the Cut Off knife sharp and unmarred     o NOTE  DO NOT USE IT AS A SCREWDRIVER     5 2 THE IMPORTANCE OF PISTON GROOVES   The wall pressure in the barrel INCREASES as the distance between the piston head and the die  DECREASES  Unless this distance is standardised for the purpose of measuring the extrudate flow  rate  very variable results can be obtained from
7.  a warning message is printed with the test results  It is calculated as the  difference between the maximum and minimum sample weight divided by the average weight of the  samples  Any value in the range 0  to 100  can be entered although ISO 1133 Procedure A  specifies that a maximum spread of 15  is allowed  This option is not offered if the Cut Off method  is not used     18 MFI 10 User Manual       Cut off time  Secs  60  Accept Change BkSp             The Standard specifies the Cut Off time for each sample according to the flow rate  An audible  beeper is sounded each time the set time period expires  The timer will then be reset automatically   Any value in the range 4 seconds to 999 seconds can be entered and will depend on the material  being tested and the weight applied  A Cut Off time should be chosen which gives a Cut Off length  of between 10mm and 20mm for the given sample  This option is not offered if the Cut Off method is  not used     MFI g 10mins  Extrudate cut off time intervals  S     0 1 to 0 5 240    gt 0 5 to 1 0 120   gt 1 0 to 3 5 60   gt 3 5 to 10 30   gt 10 5 to 15     NOTE      ISO 1133 Recommends the measurement of MFIs between 0 1g  10mins and  100g 10mins  although ASTM D1238 allows measurements up to 50g 10mins     NOTE   2  ISO 1133 recommends that all polymers with an MFR greater than 25g 10mins  use Procedure B or an automatic cut off device accurate to less than 0 1 second  This is to  ensure adequate repeatability     AA       Set Temper
8.  and allow to cool down   Then shake the thermometer firmly and decisively  with downward strokes to force the mercury back  down into the bulb  Repeat if necessary  If this does not work place in the freezer compartment of a  normal domestic refrigerator for about two minutes to retract the mercury into the bulb     RS Components also supply a cryogenic spray which is much more convenient  DO NOT USE  LIQUID NITROGEN  IT IS TOO COLD  DO NOT leave the thermometer in a cold place as this may  solidify the mercury  Then the thermometer should be returned to room temperature slowly as rapid  warming may burst the bulb owing to the pressure generated in forcing the Mercury through the  capillary  Whether to heat or cool depends on the position of the break  Those near the bulb are best  rejoined by cooling and visa versa     8 2 THERMOMETERS WITH CONTRACTION AND EXPANSION CHAMBERS   If the detached part of the column is above the contraction chamber  rejoin the expansion chamber  by heating as already explained  If there are breaks in the mercury in the contraction chamber  firmly  but smoothly swing the thermometer bulb downwards  as when shaking down a clinical thermometer   If this fails  tap the bulb vertically on the palm of the hand  Occasionally a small globule of mercury is  left in the neck of the contraction chamber  It can often most easily be reunited by heating the  thermometer to drive it some way up the stem and then chilling the bulb  The detached piece will  shoo
9.  bottom of the    weight  If the AFRT is not required  this message will not be displayed  This message will  automatically clear after 10 seconds     12 MFI10 User Manual       Pre heating Sample 2 30  6 00   Release Abort             The system will now continue preheating the sample for the time set in the test set up  This would  normally be a total of 6 minutes from when the sample is first entered into the barrel  The preheat  time elapsed is displayed with the total preheat time shown in parenthesis  If using the auto weight  loader  then at the end of the preheat time the auto weight loader arm will be moved to the bottom  position  This will release the weights and allow the test to start automatically  By pressing the key  marked Release  the weight loader arm can be lowered before the end of the preheat period  The  Release option is not displayed if the auto weight loader is not in use        Remove Piston Support  OK Abort       If the auto weight loader is not being used  then at the end of the preheat period the piston support  should be removed allowing the piston to fall freely  This message will automatically clear after 10  seconds        Pre Heat finished  start test  6 30  OK Abort             If the manual cut off method is being used the operator will now be waiting for the lower piston scribe  mark to enter the top of the barrel  As soon as this happens the Start key should be pressed and the  waste extrudate cut off at the same time  The time displaye
10.  the same sample  This is particularly true of those  materials such as polypropylene which are very pressure sensitive with respect to shear rate  The  purpose of the piston grooving  therefore  is to ensure that all samples are taken over substantially  the same range of pressures in each case  thereby achieving a significant improvement in the  repeatability of the results     A secondary benefit derived  sometimes erroneously described as the main purpose of the grooving   is that taking samples over a prescribed and limited part of the piston travel also ensures that they  are all taken over the same temperature gradients  which will no doubt be advantageous in the case  of those melt indexers where there is a pronounced temperature gradient along the barrel length     5 3 FORM OF TEST SAMPLE   The sample under test may be in any form which can be loaded into the barrel  However  it should   be noted that certain forms  such as films  powder  fibre or any very small particles  are to be avoided  if possible  This is because it is most difficult to exclude air during the charging of such samples   Included air will produce extrudate containing bubbles and may also promote oxidative degradation of  the sample     5 4 CLEANING   Immediately after each test  the apparatus must be cleaned  Not only is it easier to clean a machine  whilst still hot  but there is less likelihood of the cleaning being forgotten  There is also a real risk of  damage in trying to clean hardened po
11.  three keys while the Davenport sign  on message is displayed  Counting from the left the first key  followed by the third key and then the  second key must be pressed in sequence  A special password is required to enter this mode  When  this mode has been selected  operation is as normal but no limit is placed on the values which can  be entered  This facility is used for testing the barrel over temperature trip which needs to be set at  450  C     2 17 ANCILLARY EQUIPMENT   All equipment should be kept clean and stowed away after use  The long tools  including the piston   each slot into the holes in the top cap  The small tools should be kept in the drawer in the base of the  MFI  The draw contains cut outs sections for each tool so that the tools do not roll around and get  damaged  Missing items are easily identified     2 17 1 CHARGING TOOL  A non metallic rod for tamping the test charge in the cylinder     2 17 2 DIE EJECTOR  A steel rod with a special soft alloy tip to avoid any possibility of damaging the die  Used to push the  die up and out of the cylinder     2 17 3  CLEANING TOOL  For cleaning the bore of the cylinder     MFI 10 User Manual 25    2 17 4 CUT OFF KNIFE  Designed specially for cutting off the extrudate     2 17 5 CIRCULAR SPIRIT LEVEL  Fitted with a thermal insulating foot so that it can be placed on top of a hot barrel without suffering  damage  For levelling the apparatus     2 17 6 DIE BROACH  For cleaning the bore of the die     2 17 7 HEXAGON K
12. EXYGEN will start the connection process  If this is not completed the MFI 10 will remain in this  state for 10 minutes  after which it will automatically return to the Select Test option  During this time  the Abort button will also display the Select Test option  The Calibrate button will start the calibrate  mode  for more details of the calibrate mode see the Calibration Mode section of this manual  page  35   When the NEXYGEN has finished the connection process the next screen will be displayed       Under control of remote computer             automatically   When this message is displayed the MFI 10 will be controlled by the NEXYGEN MFI software and will  not respond to any local button press     2 14 SUPERVISOR MODE   This mode can be entered when the MFI is switched ON by pressing the first option key followed by  the third option key while the Davenport sign on message is displayed  This provides the following list  of system configuration options  Many of these options will never need to be changed        Enter Password  Accept Change BkSp             If a password has been set up  then it is necessary to enter the password before the following  options can be modified  This provides a level of security which enables only supervisors to modify  the configuration  If no password has been set up  this display is not shown        AFRT start pos above die  mm  50 0  Accept Change BkSp       This screen allows the user to enter the position at which the AFRT test will b
13. EY  For loosening the die plate retaining screws     2 17 8 TWO PISTON SUPPORT SLEEVES  For restricting the movement of the piston during the pre heat period     One of 71mm length for low flow rate polymers   One of 81mm length for high flow rate polymers     2 17 9 THERMOMETER SLEEVE  Brass cylinder with two vertical cut outs    Designed to hold the calibration thermometer absolutely straight and make sure the mercury bulb is  directly above the die orifice     2 17 10 DEAD WEIGHT  Weighing exactly 2 06kg  which together with the piston makes up the most common standard test  load of 2 16kg     2 17 11 CERAMIC INSERTS     gt  NOTE  These should be changed if chipped or polymer has built up     2 17 12 FUNNEL  To help loading with the loading of polymer     There is also a powder funnel available to special order with an extension into the barrel to avoid  polymer sticking to the side of the top of the barrel     2 17 13 DIE PLATES  PTFE coated to prevent polymer sticking to the 3 key hole slots to ensure quick and simple removal  and replacement     2 17 14 CALIBRATION THERMOMETER  90   angle  indicating  mercury in glass  4  C x 0 1  C    2 17 15 BOX OF CLEANING PATCHES  Used for cleaning the barrel     26 MFI10 User Manual    2 18 ADDITIONAL FEATURES   With the exception of the test loads  all items required to perform Melt Flow Index tests are housed  within the cabinet of the apparatus  thus being a totally stand alone work station using the minimum  amount of bench spa
14. Sp             This specifies the maximum time which may elapse between first charging the sample and taking the  first Cut Off or AFRT reading  If this time is exceeded  the user has the option of aborting the test  If  continued  a warning message is printed with the test results        Max Total Test Time  Mins  Secs  25 00  Accept Change BkSp             This specifies the maximum total duration for the test  If this time is exceeded  a warning message is  printed with the test results        Use Weight Loading System   Yes No       Asks whether the auto weight loader should be used for the test  The option is only offered if an auto  weight loader has at some time been connected to the machine        Pause Weight Loading System  81mm  Accept 71mm 81mm No Pause             If the auto weight loader is to be used  it can be programmed to pause at a position equivalent to  using either a 71mm or a 81mm spacer to support the weights  If the pause option is not required   this can also be selected     20  MFI10 User Manual       Start heating barrel when switched ON   Yes No             If this option is set to Yes  default setting   the barrel will automatically begin heating as soon as the  MFI is switched ON  If it is set to No  at switch ON a screen will display the prompt for the test set up  procedure  The barrel will only be heated when a test set up has been selected        Start heating barrel when switched ON   Accept f 4   gt              This screen allows the use
15. T FORCE so fast that the tool crashes into the die plate below  Slow down towards the  bottom of the barrel to prevent damaging the die plate  ceramic insert screws and screw holes  Use  as many pieces as necessary to ensure the barrel is thoroughly clean and shiny  A dirty barrel will  cause friction between the piston and the barrel wall and give false low results     E After each series of tests  or batch exchange  give the machine more thorough cleaning  removing the die plate and cleaning both sides of it and the bottom of the cylinder  in addition to the  routine cleaning process already detailed     F About every 100 hours use  or once a week for frequent users  remove the die retaining plate  and retaining screws and clean thoroughly  Check the ceramic insert for chips and damage  If the  insert is damaged or covered in blackened polymer  replace it     G a  The cylinder is hardened  so the bore may be occasionally cleaned with a patch loaded with  a mild metal polish  ON NO ACCOUNT however  use any coarser abrasive material than this     b  The corrosion resistant barrel is lined with Tungsten Carbide to avoid excessive wear by caustic   e g  PVC  or abrasive  glass filled  polymers     H The dies are made of tungsten carbide  and with reasonably careful use  will maintain their  accuracy over years of service  Dies with PE will last for years  but should be checked regularly  If  PVC or glass filled polymers are used  then the die dimensions should be checked far mo
16. acticable  the heat sink effect of the weights  The  loose weights are tared to give the correct total load with the correct piston  Weights purchased with  model 6  7  8 or later  or since Lloyd Instruments acquired DAVENPORT are tared for use with the  heavier pistons     MFI 10 User Manual 7    The shank of the piston is marked with two circumferential grooves so that the upper will be exactly  coincident with the top of the cylinder when the face of the piston head is precisely 20mm from the top  of the die  whilst the distance between the two grooves themselves is 30mm  The purpose of these  grooves is explained later in this manual     Please note  that in the early stages of the development of the Melt Index method of characterising  polymers  HOLLOW pistons weighing only 58g were extensively used     They were later abandoned as the need for heavier test loads became apparent  since they lacked  the necessary mechanical strength  Weights which have been tared for use with the hollow pistons  WILL NOT BE CORRECT for use with the later solid pistons  This fact should be considered if you  decide to use weights other than those supplied with the apparatus     2 8 CONTROL PANEL       Select Test      ASTM1238    Select Edit                MFI 10 Control Panel Layout    This is used to display information and to enter test set ups  The set temperature and actual tem   peratures are both displayed at all times  The lower group of green numbers shows the set  temperature  wh
17. age will only be displayed if the  maximum time has been exceeded  The user has the option to Continue or Abort the test  If  Continue is selected  a warning message will be printed if the results are sent to a printer        Cut Off Time 25 sec  60 Sec   Last Abort             If the cut off method is being used  this message will be displayed showing the time remaining until  the next cut off should be made  The figure in parenthesis is the total cut off time for this test    During the last 2 seconds prior to the cut off 2 short audible beeps will sound followed by a longer  beep telling the user to perform the cut off  This sequence of beeps is designed to improve the  accuracy of the manual cut off method by reducing user induced error  The beeps may optionally be  turned off by using supervisor mode  When the time reaches zero and the cut off beep is sounded   the display shows the following message  telling the user to cut the sample  When the last cut off  has been made the operator should press the Last key  All cut offs should be made before the upper  scribe mark on the piston enters the barrel  If the AFRT is also being used  the test will end automati   cally if the upper scribe mark on the piston enters the barrel     NOTE  The Automatic Cut Off device is optional equipment and is not supplied as standard  with the MF110     74 MFI10 User Manual       Cut Off Sample  Continue Abort             This message is displayed for 5 seconds  It is then cleared automatic
18. ally and the previous message  is displayed  The Standards both require at least 3 good cut offs to be made  Cut offs containing air  bubbles should be discarded and the good cut offs weighed and recorded in the order in which they  were cut        Performing AFRT Test  Abort             If the cut off method is not used  or if the cut off test is finished before the AFRT test  this message will  be displayed  This message will automatically be cleared when the AFRT test is finished or when the  upper piston scribe mark enters the barrel     At the end of the test  if the auto weight loader was used  it will be automatically raised to the top  position ready to begin a new test        Sample Weight  g    1  0 234  Finished Enter BkSp Abort             This screen  displayed if the cut off method was used  allows the user to enter the weight of each cut  off  The user can enter the weight to 5 decimal places followed by Enter  The Standards both require  the cut offs to be weighed to 4 decimal places  If the user is re entering the sample weights after   the results have been displayed  the Abort message is replaced by Delete  This allows the user to  remove any weights from the calculations        Cut off MFR  g 10 min  1 0525  Re Weight More Print New Test          This display shows the melt mass flow rate  MFR  of the sample tested using the Cut Off method  If  the Cut Off method was not used  this display is not shown  If the AFRT was also used  the More  option is provid
19. anguage used can be selected and saved as part of the test set up  This enables each test set   up to use a different language        AFRT Used   Yes No             Selects whether the AFRT is to be used for the test  The AFRT is required to calculate results in  accordance with ISO 1133 and ASTM D 1238 Procedure B        AFRT Distance  mm  2 00  Accept Change BkSp       If an AFRT is to be used  it is necessary to set up the distance over which the AFRT is to be timed    If a single measurement is to be taken  this can be anything up to the full test range of the AFRT   usually 30mm  If multiple readings are to be taken  the total must be less than the AFRT range  The  value entered is the distance for each reading  This option is bypassed if the AFRT is not used        Number of AFRT readings 5  Accept Change BkSp             This specifies the number of readings that will be taken at the distance entered in the previous  section  ISO 1130 Procedure B specifies that at least three measurements should be taken  This op   tion is bypassed if the AFRT is not used     MFI 10 User Manual 77       Material Melt Density  g cc  0 760  Accept Change BkSp Calculate             This allows the user to enter the melt density for the material to be tested if it is known  This value  is used to calculate the melt mass flow rate  MFR  from the melt volume flow rate  MVR  determined  from the AFRT test  The density shown is for the standard PE resin used to test all systems prior   to ship
20. ature    C  190 0  Accept Change BkSp       Determines the temperature at which the test is to be performed  Any value in the range 40  C to  400  C can be entered        Weight used   Kg  2 160  Accept Change BkSp             Allows the user to enter the value of weight which is to be used for the test  The value entered here  is used as a prompt when the operator is instructed to fit the weight  thus reducing the likelihood of  operator error if multiple weights are available  The weight is also quoted in the test report printed at  the end of the test     MFI 10 User Manual 19       Preheat Time  Mins  Secs  6 00  Accept Change BkSp             Allows entry of any preheat time in the range of 1 minute to 99 minutes in 1 second steps  This is the  total time from when the sample is first charged into the barrel to when the first cut offs can be taken   including the time spent charging the sample  The ISO 1133 Standard recommends 6   8 mins        Min Test Start Time  Mins  Secs  6 00  Accept Change BkSp             This is the minimum time that must elapse from when the sample is first charged into the barrel to  when the first cut offs can be made or AFRT readings taken  If this minimum time is not reached   a warning message will be displayed on the LCD  The user then has the option of aborting the test  or continuing  If to continued within the minimum time a warning message is printed with the test  results        Max Test Start Time  Mins  Secs  8 00  Accept Change Bk
21. calibration enabled        Enter new password  Accept Change BkSp None             Ifa password already exists  it can be confirmed by pressing Accept  A new password can be entered  by pressing Change  If a password is no longer required  selecting None will remove the password    The password may be up to eight characters long and made up using any of the numeric keys on the  keypad        Test new password  Accept Change BkSp             If the password has been changed the user will be prompted to re enter the password  The password  should be typed in exactly as before followed by Enter  If the password has been re entered correctly   it will be accepted  If the two entries are different  the user will be prompted to enter the password  again     MFI 10 User Manual 23    2 15 CALIBRATION MODE   Calibration mode can be entered by pressing the calibration key   the blank key at the bottom left  corner of the numeric keypad  from any of the test states  It cannot be entered from the test set up  state  Calibration mode allows calibration of the temperature controller and AFRT  Note that the  AFRT must be re calibrated whenever the internal working memory is cleared  The barrel tempera   ture should only be calibrated when the temperature control loop is active and the barrel tempera   ture is stable  The temperature calibration data is not erased when the internal working memory is  erased        Adjust until display reads OK  OK  0 1  C  0 1  C Follow             This disp
22. ce  To assist in both the cleaning of the barrel and manual operation  i e  Cut Off  principle  the automatic flow rate timer sensor unit is able to be swung away from the barrel housing  to a park position  thus giving easy access to the barrel entry and die plate  When required  the  automatic Flow Rate timer sensor unit is easily returned to its position near the barrel  The apparatus  has a built in over temperature protection relay  If the temperature of the barrel exceeds the pre set  Cut Off point  the heaters will be disconnected     2 19 ADDITIONAL REQUIREMENTS   Certain additional equipment is required for use of the apparatus but is not supplied as standard  equipment  commonly available in most laboratories   The principle item is an analytical  LABORATORY BALANCE capable weighing to 4 decimal places  If you are only using the automatic  Procedure B AND the melt density is known  this equipment will not be necessary     All users will require a top pan balance capable of weighing to 2 decimal places for weighing out the  sample charging mass     A Scotchbrite pad  flat rectangular pan scourer available from most supermarkets  is ideal for  polishing the die once the bulk of the polymer has been scraped off with the cut of knife     MFI 10 User Manual 27    3 0 INSTALLATION    3 1 UNPACKING   All packing material should be carefully removed from the MFI machine and a careful visual  inspection should be made to ensure that there is no obvious transit damage  The fo
23. ce over which a test is performed is programmed as  part of the test set up described in the relevant section of this manual  Standard distances are 2mm  and 20mm ISO 1133 and 6 25mm  1 4 inch  and 25 4mm  1 inch  for ASTM D1238     2 10 USING THE CONTROL PANEL   The top line of the LCD provides the user with information or instructions  The second line displays  messages which are used to define the function of four soft keys located directly under the LCD  display  Where messages are displayed on the second row  simply press corresponding key to select  the required option     2 11 TEST PROCEDURE       Davenport Melt Flow Indexer  V 1 0 30 1 09          This sign on message is displayed for 3 seconds following power ON  After this time the display will  automatically clear and the next message will be displayed  The second line shows the version num   ber of the embedded software and its release date  If the internal EPROM is replaced with a different  version number  the new version will be displayed  A change of version will cause the internal work   ing memory to be re initialised  This will result in the loss of all test set ups which have previously  been saved  Temperature calibration information will not be lost        Select Test Standard  Select Edit E  gt              This screen is displayed when the user needs to select one of the pre programmed test set ups or   to define a completely new one  It is also displayed when a test is aborted at any point  allowing t
24. d  in wooden crate  53kg  117 Ibs    5  C to 45  C  41  F to 129  F    0   80  RH non condensing  Well ventilated  draught free     20  C to 55  C  Substantially Dry   4  F to 131  F    CN1   I E C  Plug  Mains In    CN2   I E C  Socket  Mains Out  250VA max    CN3   9 Pin    D    Plug  RS232 interface    CN4   15 Way    D    Socket  Auto Cut Off  amp  Analogue O P   CN5   15 Way    D    Socket  Weight Loading System  amp    Analogue  O P    CN6   9 Way    D    Socket  Automatic Flow Rate Timer    1S01133 BS2782  Part 7  Method 720A   ASTM D1238  Procedures A  amp  B     Procedure B requires optional Automatic Flow Rate Timer   NFT   UNE   UNI   DIN    MFI 10 User Manual 39    10 0 CONNECTION LEADS    10 1 RS232 LEADS  MFI10 to PC Lloyd Part No  09 0695   supplied      9 Way    D    9 Way    D     SKT  Female  SKT  Female     RxD 2 2 RxD  TxD 3 X  3 TxD  DCD 1 1 DCD  DTR 4  X 4 DTR  RTS 7 7 RTS  CTS 8   a 8 CTS  GND5 o         OoO 5 GND    Baud rate 4800  8 bit  1 Stop bit  No parity    MFI10 to Citizen Printer Lloyd Part No  09 0644   optional      9 Way    D    25 Way    D     SKT  Female  SKT  Male   MFI10 End Printer End    TxD3 o            o 3RxD    CTS 8 o      o 20DTR  GND5 Oo         O 7GND    Baud Rate 4800  8 bit  1 Stop bit  No parity     NOTE  To maintain EMC compliance ONLY leads supplied by Lloyd Instruments should be used   y Check leads for signs of damage regularly  Replace leads that are damaged in any way     40 MFI10 User Manual    yy international Symbol
25. d is the total test time commencing from  when the sample was first entered into the barrel  The next screen for the cut off method is shown if  the test start time is not reached  If the AFRT is being used  the following message will be displayed  instead        Pre Heat finished  awaiting start 6 30  Abort             When the lower piston scribe mark enters the barrel  the MFI will automatically start to measure the  time taken for the AFRT to move through the specified distance     MFI 10 User Manual 13       Test start time not reached  6 30  Continue Abort             The test set up allows the user to enter a minimum time from when the sample is entered into to the  barrel to when the first measurements should be taken  This message will only be displayed if the  minimum time has not been reached  The user has the option to Continue or Abort the test  If  Continue is selected before the minimum time is reached  a warning message will be printed if the  results are sent to a printer  For most tests the minimum time will be 6 minutes  the same as the  preheat time  so this situation should not occur        Cut Off Waste  Abort             This message is only displayed if the cut off method is being used  It is automatically cleared after 5  seconds        Test Start time exceeded  Continue Abort             The test set up allows the user to enter a maximum time from when the sample is entered into to the  barrel to when the first measurements should be taken  This mess
26. ded if the MFI has been switched OFF  If the AFRT  has been disconnected or if the internal working memory has been initialized  it will be necessary to  re calibrate the AFRT  This is a simple procedure which will automatically be invoked when required   Follow the on screen instructions or see the section on AFRT calibration for more details  page 36         Moving Weight Loader  please Wait  Stop Stop Stop Stop             This screen is displayed whenever the weight loader arm is moving  If the test uses the auto weight  loader then is must be positioned in its upper position before the test can start  If it is not in the upper  position  it will be moved there automatically with the above message displayed     Pressing any option key will stop the weight loader before it reaches the top     i NOTE  The Automatic Weight Loader is optional equipment and is not supplied as standard  with the MF110     10 MFI10 User Manual       Heating Barrel  please Wait  Loader Abort       This screen is displayed while the temperature controller is setting the barrel to the required tempera   ture  When the barrel temperature is stable within  0 1  C for a minimum of 10 seconds  the buzzer  is sounded and the next message is displayed  If an auto weight loader is connected to the MFI 10 it  can be manoeuvred manually by selecting the Loader key        Move Weight Loader   81mm 71mm Bottom Finish                Move Weight Loader   Top Bottom 7imm Finish             When the auto weight l
27. e other  see notes on operating techniques      E Ifthe sample was thermally unstable during the test  the successive extrudate samples will have  a continually changing weight  Some materials cross link  forming high molecular weight groupings  and others degrade due to breaks in the chain  With cross linking the MFR tends to become lower  and the extrudate samples lighter  whilst with chain breaks the MFR will tend to increase so the  samples will become heavier  When a systematic change of mass of this nature is observed  it is  most useful to calculate the extent of the weight change  MELT FLOW RATE SPREAD is not only an  indication of the thermal instability of the sample  but is also used to indicate the significance of the  quoted Melt Flow Rate value     The MFR Spread is simply the range of extrudate sample weights expressed as a percentage of the  average and calculated from this equation     Total MFR Spread        Max Weight   Min Weight  X 100  Average Weight       Lower MFR Spread         Ave Weight   Min Weight  X 100  Average Weight       Upper MFR Spread            Max Weight   Ave Weight  X 100  Average Weight    Using the control panel numeric keypad individual sample weights can be entered into the MFI  The    MFI then calculates and displays the MFR using the equations given above  The MFR and total MFR  spread can be sent to the RS232 interface for printing on the optional ticket printer     32 MFI10 User Manual    5 0 OPERATING TECHNIQUES    5 1 CUTTING
28. ect what information is to be printed  Selecting All will print Test Set        Printing      Abort             up  Samples and Results in one go     This display is shown while printing is in progress  Pressing Abort will cancel printing     16  MFI10 User Manual    2 12 TEST SETUP   The following series of screens are used to set up the test method  Where Yes No options are  offered  the existing setting is displayed above the left hand key  Where values are to be entered   the previous value is offered as a default and can be selected by pressing Accept which is displayed  above the left hand key  As all existing defaults can be selected by pressing the left key  it is a simple  matter to edit one part of a test set up without having to remember all of the other settings  After  setting all of the options  they can be saved into non volatile memory  Up to ten test set ups can be  saved and recalled later  Two Standard Test Set ups are provided as examples  One uses manual  cut off method  the other uses the AFRT  These can be used as they are or modified to suit individual  requirements        Enter Password  Accept Change BkSp             If a user password has been set up in supervisor mode  then it is necessary to enter the password  before the test set up can be modified  This provides a level of security which enables only  supervisors to modify the test set up  If no password has been set up  this display is not shown        English Messages  OK  lt   gt        The l
29. ed for displaying additional results  otherwise this option is not shown     MFI 10 User Manual 15       AFRT MVR  cc 10 min  1 4700  Repeat More Print New Test             This display shows the melt volume flow rate  MVR  of the sample tested using the AFRT method   The More option is provided for displaying additional results        Melt Density  g cc  0 6410  Repeat More Print New Test             This display shows the Melt Density of the sample tested using a combination of the MFR obtained  from the Cut Off method and the MVR obtained from the AFRT method  If the test set up does not  specify that the melt density is calculated  this display is not shown        AFRT MFR  g 10 min  1 0525  Repeat More Print New Test             This display shows the melt mass flow rate of the sample tested using the AFRT method  It is  calculated using the result for the MVR and the value entered for sample melt density as part of the  test set up  If the AFRT was not used  or if the test set up specified that the melt density should be  calculated  this display is not shown      gt   NOTE  Selecting Repeat from any of the above screens will repeat the test using the same  test set up parameters  More will display the next results screen as applicable for the type   of test performed  Print will display the following screen and New Test will enable a different   test set u sesites    Select printout options  Test Setup Samples Results All             This display allows the user to sel
30. egin  The figure  specifies the distance between the lower edge of the piston head and the top of the die  ISO 1133  specifies that this distance should be 50mm and that is the default setting in our factory  The ASTM       AFRT end pos above die  mm  20 0  Accept Change BkSp             Standard requires the piston to be 48mm above the die    This screen allows the user to enter the position at which the AFRT test will end  The figure specifies  the distance between the lower edge of the piston head and the top of the die  ISO 1133 specifies  that this distance should be 20mm     22 MFI10 User Manual       Enable buzzer  Yes No             The audible beeper can be switched off by selecting No  This may be useful where several MFI  machines are operating in close proximity or the test has been fully automated by addition of the  AFRT and the auto weight loading system  Default setting is that the buzzer is enabled        Cut off warning Bips   Yes No             This option allows the user to select whether the two short bips which are sounded before a Cut Off is  to be made are to be enabled  Select No if vou do not require these bips  Default is bips enabled        Enable Calibration mode   Yes No             Calibration mode can be entered by pressing the calibration key    the blank key at the bottom left  hand corner of the numeric key pad   during any part of a test   not during test set up   This option  allows calibration mode to be disabled  Default setting is with 
31. he  user to select a different test type  The default test set up is the one which was current when the  machine was last switched OFF  The   and  gt  options scroll through the available test set ups  The  Edit button will allow the user to customise the test set up  This display will not normally be shown  after power ON     To force this screen to be shown at power ON the appropriate option in the test set up for the default  test must be set  See the section on setting up the test for more information     MFI 10 User Manual 9    Pressing select will advance to the next screen and tell the temperature controller to start heating the  barrel to the temperature specified in the selected test set up        No AFRT fitted please connect  Ignore Abort             This screen is only displayed when the test requires an AFRT and none is connected  Connect an  AFRT or select Ignore to perform test using cut off method        No weight loading fitted please connect  Ignore Abort             This screen is only displayed when the test requires the weight loader and none is connected   Connect a weight loader or select Ignore to perform test using manual weight loading method        Push AFRT to its lowest position  Abort             If the test uses an AFRT  the MFI needs to know the current AFRT position relative to the barrel  By  pushing the AFRT to the bottom of its stroke  a micro switch is activated which provides a reference  position for the MFI  This information is only nee
32. he option AFRT is being used the  MFI will start taking readings automatically     30 MFI10 User Manual    G Cut off and retain  in order  specimens of the extrudate at a succession of identical time    intervals     The time interval used should be chosen on the basis of the expected MFR of the sample  see table    but whenever possible cut at least 5 samples making quite certain that they are ALL taken when the  lower piston ring is INSIDE the barrel and the upper ring is still outside it     MELT FLOW RATE  GRAMS 600 SECS    0 1   0 5  0 5   1 0  1 0 3 5  3 5 10  10 25    SAMPLE MASS  GRAMS    4 5    CUT OFF INTERVAL    SECONDS  240  120  60  30   10 15    H On completion of the test and after all subsequent tests  clean the piston  die and barrel     See section OPERATING TECHNIQUES  Cleaning   4 0 CALCULATIONS AND PRESENTATION OF RESULTS    MFI 10 User Manual    31    A Weigh each Cut Off in the order of the extrusion  Discard any that have obvious air bubbles in  them     B Calculate the average weight of the extrudate samples to three significant figures   C Calculate the MFR from the equation    MFR  10W T   where    MFR   Melt Flow Rate in grams per 10 minutes     W the average extrudate sample weight   T the extrusion time per sample in minutes     D Report the value thus obtained to two decimal places  When reporting this result always quote  the testing conditions used  such as  the temperature  the load and whether the die used was of  standard geometry or of som
33. her the die or the bore of the barrel  or both  are not perfectly clean and they should be re examined  and cleaned if necessary     E The apparatus is supplied with a plug socket and mains lead  Fuses for the apparatus are  incorporated into the inlet socket  Power consumption of the apparatus will not exceed 1KW  An  auxiliary power outlet is fitted with a IEC socket  Amaximum of 1 amp is available at this socket     We are able to supply suitable connecting leads for this socket to enable other peripherals to be  powered directly from the apparatus     F The temperature controller will start to control the barrel temperature as soon as an electrical  supply is connected  unless the test set up has been customized to disable this feature    This controller is calibrated at the factory to give a controlled temperature between 190 and 230   C to  within  0 2   C  If itis necessary to perform tests at higher temperatures  for example 320   400  C it  is advisable to set the controller to the desired set point and check the barrel temperature  If an  appreciable difference is apparent  one of two procedures can be followed     1  An offset can be used by setting the set point of the controller to allow for the difference     2  The controller can be re calibrated against a known accurate thermometer to give the correct  reading at the required set point     The controller will control to   0 1   C of any set point     3 2 TEMPERATURE CONTROL  MFI 10 User Manual 29    With the a
34. ile the upper group of red numbers shows the actual barrel temperature  LED   s pro   vide quick status indication  The band of five LED   s to the right of the temperature displays indicate  the temperature relative to the set temperature     The centre green LED will be lit when the actual temperature is within  0 1  C of the set  temperature  The amber LED   s indicate that the temperature is between 0 1  C and 0 5  C  from the set temperature and the outer red LED   s will be lit when the temperature is more  than 0 5  C from the set temperature     The alphanumeric LCD provides user instructions and allows easy entry of the test param     eters  An audible beeper helps to prompt the operator and provides end of preheat and  sample cut off warning signals     8 MFI 10 User Manual    The numeric keypad is used for entering parameters  The blank key grouped with the numeric keys  is used to select calibration mode  The four keys directly under the LCD display a soft keys used to  select from the list of options displayed on the second line of the LCD     More detailed information on setting up tests and displayed messages is given in The MFI 10  Test  Set up section of this manual     2 9 AUTOMATIC FLOW RATE TIMER  Optional Equipment    This attaches to the side of the MFI and the lead plugs into the 9 way socket on the right hand side of  the column  The Automatic flow rate timer derives its power from the MFI  All readings are  automatically handled by the MFI  The distan
35. ing on the die     24 MFI10 User Manual       Fit die  piston and weight  Attach AFRT  OK 10 5432             By assembling the components in a clean condition as instructed  the MFI control software can  measure the distance from the reference micro switch to this position     Press OK to indicate that the system is correctly set up  Once this has been done  the system control  software can determine test start and end positions by moving the AFRT arm down to the micro  switch     2 16 STARTUP OPTIONS  It is possible to force certain actions when the machine is first switched ON by pressing certain keys  or sequences of keys while the Davenport sign on screen is displayed     Internal Working Memory Initialization can be forced by pressing the right hand option key during the  Davenport sign on message  This forces all internal working variables to be cleared and will result in  the loss of all test set ups and AFRT calibration  Temperature calibration will not be lost     Hardware Test Mode can be forced by pressing the left hand option key twice while the sign on  message is displayed  This mode is used for testing the internal electronics of the MFI and should  only be used by qualified test personnel     Supervisor Mode can be started by pressing the left hand option key followed by the third option key  from the left  See the section THE APPARATUS  Supervisor mode for details of options available in  the mode     System Test Mode can be entered by pressing a sequence of
36. is read and  thoroughly understood  Caution messages are shown in bold type     NOTE     2    CAUTION    HIGH TEMPERATURES AND MOLTEN PLASTIC ARE  INVOLVED IN THE MFI PROCESSES     ALL SPECIFIED SAFETY PROCEDURES SHOULD BE  STRICTLY ADHERED TO     READ THIS MANUAL BEFORE USING THE MFI MODEL  10     THE THERMOMETER S  SUPPLIED ARE GLASS AND  CONTAIN MERCURY  PLEASE STUDY SECTION 8 OF  THIS MANUAL     BEFORE USING THE MATERIALS TESTING MACHINE     The note icon indicates additional or supplementary information about the action   activity or concept  Notes are shown in bold type     General Safety   General safety precautions must be followed  when using this LLOYD INSTRUMENTS prod   uct  Failure to observe precautions and warn   ings may result in damage to the equipment  or  injury to personnel     It is understood that safety rules within compa   nies vary  Ifa conflict exists between the mate   rial contained in all LLOYD INSTRUMENTS     User   s Guides and the rules of a company using  a LLOYD INSTRUMENTS product  the more  stringent rules should take precedence     Safety Considerations   The MFI 10 is completely enclosed and provides  no potentially hazardous outputs  Safety con   siderations are related to the power connections  and physical mountings     Electronic and mechanical components housed  within the MFI 10 covers are to be serviced by  authorized LLOYD INSTRUMENTS representa   tives only     MFI 10 User Manual 3    1 0  1 1  1 2  1 3  14    2 0  2 1  2 2  2 3 
37. lay allows calibration of the temperature controller  The temperature can be calibrated at  any set temperature by placing a calibrated thermometer in the barrel of the MFI  For best results the  thermometer should be surrounded by a polymer having similar characteristics to the polymer to be  tested  If the system is calibrated again at the same set temperature  the previous calibration value  is overwritten  If the system is calibrated at a second set temperature  the calibration value stored at  the original set temperature is not affected  To calibrate the temperature at ambient  the set tempera   ture should be adjusted to a value around ambient  Calibration may be performed at a maximum of  fifty set temperatures  Pressing  0 1  C or  0 1  C will increment or decrement the displayed actual  temperature by 0 1  C  This is only adjusting the displayed value  not the value currently used by the  temperature control loop  When the displayed temperature is correct  Follow can be pressed  This  will tell the temperature control loop to use the displayed temperature reading for controlling the bar   rel temperature  This Follow message will change to display Hold  Further pressing of the  0 1  C or   0 1  C keys will change the displayed temperature and force the control loop to adjust the barrel  temperature accordingly  Pressing Hold will tell the temperature control loop to maintain the  temperature at the last set temperature regardless of further changes to the displayed 
38. le and the smaller die  Similar caution must be  exercised when considering test temperatures and test loads     MFI 10 User Manual 35    6 0 AUTOMATIC FLOW RATE TIMER OPERATION    6 1 NOTES OF AUTO FLOW OPERATION  The Auto Flow Rate Timer is calibrated to time the swept volume of a weighted piston travelling  through a pre selected distance  usually 2mm or 25mm     The Swept Volume of 2mm   0 142cc  The Swept Volume of 25mm   1 776cc    The Auto Flow Rate Time recorded on the display is the time taken to extrude either of the above  volumes and depends on the Standard being applied  In order to obtain the Melt Flow Rate from this  volumetric flow it is necessary to obtain the melt density at the test temperature and load from the  material supplier or by integration of Cut Off time with the use of the Auto Flow Rate Timer     6 2 MELT DENSITY CALCULATION    p  m  0 711 xi     where   p   the density  g cm     of the melt at the TEST TEMPERATURE   m   the average mass  g  of the cut off     the predetermined distance move by the piston   NOTE  The melt density is NOT the same as the  solid  density at 23  C usually quoted on  D gt  the certificate of conformity supplied with the resin    NOTE  The GD6250    Hostalen    resin with an MFI of 1 0 to 1 1g 10mins is used to test all  LY machines before leaving our factory has a density of 0 947 at 23  C but a melt density   of 0 760 at 190  C   7 0 GENERAL MAINTENANCE  Apart from regular  careful cleaning of the working parts as a
39. llowing items  should be in this package     Supplied items Quantity Part No  This User Manual 1 01 2713  Melt Flow Indexer model 10 1  depending on order   Die plate casting 1 73718  Ceramic die plate insert 2 73717  Standard die 2 096mm bore 1 338 10  Die broach standard diameter 1 338 101  Standard piston assembly 1 338 45  Weight 2 16kg 1 338 14  Cut off Knife 1 338 33  Charging tool 1 338 52  Die ejection tool 1 338 54  Cleaning tool 1 338 55  Cleaning patches 1 box 338 132  Filling funnel 1 730 79  Piston support sleeve 71mm 1 338 293  Piston support sleeve 81mm 1 338 294  Die tweezer   s 1 92 0404  RS232 lead MF110 to PC 1 09 0695  Standard thermometer 188 192  C 1 338 104 3  Thermometer sleeve 1 730 62  Circular spirit level 1 338 100  Spanner 16 17mm 1 91 0095  Hex driver 3 32 1 338 50  Adjustable foot assembly 5 02 0443  Cordset 1  depending on order     If there is any damage  or parts missing  please report this to your authorized DAVENPORT  representative     A Carefully clean all the parts  removing any preservative that may have been applied to protect    28  MFI10 User Manual    them during transit and storage  It is strongly recommended that at this stage the small loose parts  should be stored in the wooden tool tray in the base of the MFI in their appropriate recesses The  long tools and piston placed into the pockets located on the top of the unit     B Choose a suitable site  preferably out of reach of strong draughts  where the apparatus may be  set up m
40. lready described  the Melt Flow Rate  Apparatus requires little maintenance   In the rare event of component failure  this should be corrected by replacing the defective unit  e g   thermometer  If the apparatus is to be taken out of use for any length of time  it should be given a  very thorough clean  after which all exposed metal parts should be protected with a light greasing    before storing the apparatus ina DRY PLACE and under dust covers     Under these conditions the Melt Flow Indexer should give many years of trouble free service     36 MFI10 User Manual    8 0 THERMOMETERS   We take care to pack our thermometers so that they will not be damaged in transit  Even though it  may still be that  due to vibration  breaks have appeared in the mercury column  These can easily be  repaired and MUST be looked for before using the thermometer as  if one exists and is not noticed  a  false temperature indication will be given as well as the possibility of the mercury bulb breaking     8 1 REPAIRING BREAKS IN MERCURY COLUMN TYPES OF THERMOMETERS    The Thermometer       Expansion Chamber    183 190 132       A break in the mercury column can usually be detected by eye  If it is not it will be seen as soon as it  is put into a heated barrel  The mercury will shoot up beyond the maximum temperature very quickly   where as under normal operation the movement of the mercury column will be quite slow  If the  mercury does shoot up very quickly  immediately remove it from the barrel
41. lymer from the operating surfaces  It has been found  as a  result of experience in the field  that the bulk of cases where there has been apparent malfunction of  the apparatus  leading to anomalous results and un reliability  are due to faulty cleaning  usually over  some time  rather than to component failure  In particular  half hearted cleaning of the barrel and    or the die can easily lead to deposits of pyrolysed polymer  often remarkably difficult to see and rec   ognise  effectively reducing the cross sectional area and markedly affecting the results  The routine  below should be rigorously followed     MFI 10 User Manual 33    A Remove the piston and clean off any polymer adhering to the head  face and particularly the  shank  with a cloth  Place the cleaned piston in the housing provided on the apparatus     B Push the die UP the cylinder bore with the die ejector and out of the top  The die ejector will  pass through the hole in the die plate and it is not necessary to remove this  The passage of the die  through the barrel will remove the bulk of any polymer remaining after the test     C Clean the die thoroughly with a cloth and clean the bore with a diebroach  polishing with a  wooden peg  DO NOT HEAT THE DIE INA FLAME TO BURN OFF POLYMER     D Clean the bore of the barrel with the end of the cleaning tool wrapped in a cleaning patch which  has first been folded in half  Push the cleaning tool firmly  with a slight rotating motion  down the bar   rel but DO NO
42. ment  NB  The melt density is NOT the same as the  Solid  density and may be calculated    r   m divided by 0 711 x i  Where r   density in g cc of the melt at the test temperature and m   the  average mass in grams of the cut offs and i   predetermined distance moved by piston  The user  may also select Calculate  in which case both AFRT and Cut Off tests can be performed simultane   ously  The melt density is calculated from the results obtained from each type of test  If Calculate is  selected  the following option to use Cut Off method is automatically set to Yes  This option is not  offered if the AFRT is not used        Cut off method  Yes No             Selects whether the Cut Off method is to be used  If the AFRT is not used  this is automatically set to  Yes  It is also set to Yes if The AFRT is used and the option to calculate melt density is chosen  The  Cut Off method is required to calculate results in accordance with ISO 1133 Procedure A and ASTM  D1238        Use Auto Cut off   Yes No             If the Cut Off method is to be used and an auto cut off device has at some time been connected to  the MFI the test can be configured to use the auto cut off device  If Yes is selected  then during a  test  when a cut off is due to made  the auto cut off device will be automatically triggered        Allow MFR Spead     15 0  Accept Change BkSp       This sets the maximum MFR spread which is allowed from the Cut Off test  If the total spread is  above the entered value 
43. nd 100 1  depending on the MWD  of the sample and the simple arithmetic ratio   MFR 21 6kg MFR 2 16kg gives a measure of the  flow parameters     Depending on the MFRs at 2 16kg load  the  value at 21 6kg can be equivalent to shear rates  more than 100 secs 1 coming  possibly  into  the range of extrusion processes  The facility  also exists to measure flow parameters at dif   ferent temperatures giving an insight as to the  temperature sensitivity of the sample     The extrudate generated during the test can also  be used to determine other useful parameters of  the sample  The emerging sample temperature  and the die diameter are known and its rate of  cooling and ambient temperature gives    con   trolled    conditions and is therefore repeatable   Two important properties can be measured more  conveniently and with greater precision from   the Melt Flow Indexer than from commercial  processing equipment  These are Die Swell and  Draw Down     MFI 10 User Manual 5    1 3 DIE SWELL   Die Swell or swelling ratio  is an arithmetic ratio between the diameter of the cooled extrudate at  ambient temperature and the diameter of the die  This is an important factor for processes such as  pipe extrusion and blow moulding     1 4 DRAW DOWN   Draw Down is the extending of the polymer extrudate by its own weight as the extrudate lengthens  It  is  for example  a requirement of a polymer to have good draw characteristics in the blown moulding  process so that a good parison is produced
44. nd MFR  are indirectly proportional  so that  although a  sample of high MFR will process more readily  its  melt strength is likely to be low     1 2 OTHER APPLICATIONS OF THE  APPARATUS  In addition to determining the MFR  the ap   paratus can be used to measure other properties  of the polymer providing useful knowledge of its  behavior  By performing the MFR test under two  or more different loads  useful information about  the rheological properties of the material can be  found  It could be wrongly assumed that two  polymer samples with the same MFR will behave  in the same way with respect to output  shear  rate  when tested under the same conditions   When subjected to the higher pressures  shear  stress  required for a moulding or extrusion  process  they could perform quite differently  If  the samples have different Molecular Weight  Distributions  MWD  the increase in the shear  rate resulting from an identical increase in shear  stress will be different as well  In general  a       MFI 10 Melt Flow Indexing Machine    large MWD corresponds with a greater shear  sensitivity  i e  a higher rate of change in shear  rate per unit increase in shear stress  A useful  measure of the shear stress shear rate relation   ship  flow parameter  can be made by measur   ing the samples MFR at standard load and at a  higher load  Increasing the test load  for poly   ethylene   s  from the standard 2 16kg to 21 6kg  loads  increases the level of shear rate by factor  between 50 a
45. nd hole in  the middle into which the ceramic insert is placed  The non stick PTFE coating assist with cleaning   and a ceramic insert withstands high temperatures and provide a rigid flat surface for the die to sit on   The ceramic insert sits slightly proud of the die plate in contact with the die and bottom of the barrel   This concave underside enables the cut off knife to cut the extrudate right up to the base of the die     Insert screws to about half their length so that the die plate may be pushed up to the bottom of the  barrel and turned 10   to the engaged position  Do not over tighten the screws  it should be possible  to engage and remove the die plate without adjusting the screws and still have a tight fit between the  ceramic insert and the bottom of the die     2 7 THE PISTON   There are two pistons  the standard version for use with weights up to 10 kg and one with guide bush  fitted for weights of 10kg and above  The guide bush is a metal tapered collar which slides up and  down the shank of the piston and provides extra rigidity with higher loads to prevent vectored loads   The standard piston weighs 100g     The solid steel piston has a head with a diameter of 9 47mm and a length of 6 35mm  The shank has  a diameter of 8 89mm and terminates with a portion of further reduced diameter which is covered by  an insulating  heat resistant sleeve  This forms a spigot upon which the loading weights are placed   The purpose of this sleeve is to reduce  as far as pr
46. oader is to be driven manually  one of the above screens will be displayed   The exact options will depend on the actual position of the weight loader arm  Press Finished to end  manual operation of the weight loader  If the weight loader is in the wrong position it will automatically  reposition itself before continuing the test        Weight Loader Door Open  please Close  Abort             The weight loader arm can only be moved if both doors are closed  If either door is open the above  message is displayed        Load Sample  Compress Loader Abort       This screen instructs the operator to start charging the barrel with the sample to be tested  The same  sample mass of the test polymer should be loaded every time to improve repeatability of the test   especially in QC situations  Please refer to fig 72  As soon as the charging process has been started  the Compress button should be pressed  This starts the preheat timer and displays the next mes   sage  Note that the time spent compressing the sample is included in the preheat time     NOTE  While pouring sample into funnel  agitate the funnel slightly from side to side  This  Le will speed loading     MFI 10 User Manual 11       Compress Sample  OK Abort             The sample should now be compressed  Press OK when you are satisfied that the sample has been  compressed and there are no more air bubbles  Air bubbles create a    popping    sound  This process  should be continued for a maximum of one minute  After this 
47. ore or less permanently     Place the circular spirit level on top of the barrel  raise the two rear corner feet clear of the work  surface and level the apparatus as exactly as possible by means of the centre rear foot and two front  feet  Once levelled  lower the two rear corner feet to provide stability and lock all feet into place  Do  not move the apparatus  If it has to be moved  ensure that the levelling procedure is repeated in the  new site     NOTE  IT IS IMPORTANT FOR CORRECT FUNCTION THAT THE APPARATUS SHOULD        BE LEVEL     NOTE  It is advisable to check the levelling of the apparatus from time to time  whether it  has been moved or not  It is quite in order for this to be done whilst the machine is hot but in  i this case  take care not to leave the circular spirit level on the hot barrel for any  considerable length of time  Despite the insulated base of the level  enough heat will be  generated to burst it over a long period of time     C Loosen the die plate retaining screws and remove the die plate  Clean the cylinder bore  thoroughly using the cleaning tool and as many cleaning patches as may be necessary  finish off with  a polish using a dry patch     D Similarly clean the die  paying special attention to the bore which should be polished with a  small wooden stick   a pared down match stick   Drop the cleaned die into the bore  It must fall  freely and strike the die plate with a clearly audible click  If this does not happen it indicates that ei   t
48. orted by prompts on the control panel     A Allow the barrel temperature to reach a set point and stabilise for at least half an hour  Insert  the die into the barrel  Pre heat the piston  The LCD will prompt loading when the temperature has  re stabilised  Remove the piston    B Charge the cylinder with the test sample  introducing it in increments and tamping down each  increment with the charging tool to exclude as much air as possible  Complete the charging  operation in less than one minute  Preheat time will begin with the charging process     C On completion of the charging operation  insert the piston into the barrel and place the piston  support sleeve so that the piston is supported by the sleeve acting on the rim  with the base of the  sleeve square on the barrel  Two sleeves are provided  the longer for the MFRs greater than one and  the shorter for MFRs less than one     D Place the appropriate weight on the piston and manually force the test material through the die  until the top of the sleeve prevents further travel of the piston into the barrel  DO NOT purge at the  end of the pre heat period     E When the pre heat time has elapsed  remove the sleeve to start the test     CAUTION  The sleeve will be hot     FWatch the progress of the piston down the barrel  As soon as the lower circumferential ring  on the piston enters the barrel  Cut Off and press the OK button  discard the extrudate and AT THE  SAME TIME START THE TIMER for the beginning of the test  If t
49. pparatus connected to the mains  the temperature controller will start to heat the barrel  to set point  factory set to 190  C   The LED below the temperature displays will indicate that a  controlled output is being applied to the heaters  The set point on the controller can be changed  by means of the keys on the front of the unit  See the section THE MFI 10  Test Procedure of this  manual     The temperature shown on the digital display may be checked  if required  with a mercury in glass  thermometer following the procedure below     A Insert the thermometer sleeve into the barrel and ensure that it is fully down against the die  plate     B Insert a thermometer of suitable range into the thermometer sleeve  Any difference can be  re adjusted by re calibration  See the section THE MFI 10  Calibration Mode of this manual     C If ISO R 1133 standard is being followed  the temperature should be checked with the  thermometer placed directly into the melt of polymer in the barrel     D Calibration adjustments should be performed 4 minutes after the polymer and thermometer are  loaded into the barrel and all readings are stable     3 3 CUT OFF TIMER OPERATION   Although various standard methods are in general agreement  particularly regarding apparatus  specifications  they do differ in certain methodological aspects  The test method described below is   in our opinion  the best way to use the apparatus for the most consistent  reproducible and  meaningful results  and is supp
50. r to specify a name for the test set up  Any name up to ten characters long  may be entered  The UP and DOWN arrows are used to change the letter indicated by the underline  cursor     A list of alphanumeric characters and some special symbols is provided  Pressing the RIGHT arrow  will advance the cursor to the next letter  After the tenth letter  the cursor is returned to the start of the  name  The name entered is the one displayed when selecting the test set up at the start of a test   Pressing Accept advances to the next screen        Setup Complete  Save Options  Save Create New Abort       Having programmed all of the test set up options  they can now be saved into non volatile memory  to be called up later  Selecting Save will overwrite the previous version of this test set up  Selecting  Create New will not overwrite the previous version but will create a copy of it with the changes just  made     2 13 REMOTE CONTROL  The MFI 10 can be controlled from a computer using the NEXYGEN MFI software  p n 40 0734         Select Test  Remote  Select Edit  lt  lt   gt        This requires Version 1 6 or later of the embedded software  To do this the Remote test setup must  be selected  The Remote test setup can be edited by pressing the Edit button  however  only the  language option can be changed  The Remote test setup is selected by pressing the Select button   This will display the next screen     MFI 10 User Manual 21       Cannot connect to remote computer  Abort       N
51. re often  It  should be cleaned  as already described  by means of the die broach and a wooden peg  Suitable  plug GO NO GO gauges are available for checking that the bore diameter is within the correct  tolerance  If such a plug gauge is used to check dies in this way  great care must be taken to see that  the die is both perfectly clean and COLD  If there is any dirt in the bore there is a possibility that the  plug gauge will jam and  if this happens  the die will  in all probability  become damaged or chipped  whilst trying to remove it  Similarly  a plug gauge inserted into a hot die will very probably lock due  to differential thermal expansion  In this case make no attempt to force it but allow the whole to cool  down  then the gauge will become free     IIT IS ABSOLUTELY ESSENTIAL that no die should be used which has any visible imper    gt  fection in the bore  particularly any chip  however small  on the edge of the bore opening   The die ejector is tipped with a soft alloy to prevent damage to the dies     NOTE  Never use any die ejector which has lost its tip     34 MFI10 User Manual    5 5 DIFFERENCES IN TEST CONDITIONS   The standard die is not always considered suitable for use with materials with MFR greater than 25  grams 10 mins  In such cases  a die with a smaller diameter bore  1 181mm  may be used   the Jet  B of BS2782   It must never be forgotten  however  that it is impossible to make a strict comparison  between a result obtained with a different samp
52. rison of filled and unfilled polymers  The automatic method   Procedure B  requires that the melt density either be known or calculated using the    calculate     option in the embedded software  The AFRT automatically allows the flow rate to be measured and  is suitable for flow rates between 0 5 and 900g 10mins   according to ASTM but only 100g 10mins  according to ISO 1133      2 1 THE BARREL  OR CYLINDER    The barrel is a steel cylinder  50 8mm diameter and 162mm long  through which a smooth  straight  hole 9 55mm diameter has been bored and honed  The barrel is encased in a double layer of insulat   ing material and the whole is housed in a sheet metal case     2 2 THE HEATER   The barrel is heated by three cartridge heaters equally spaced around the central bore  Great care is  taken to ensure that all our barrel assemblies are produced to close tolerances to give the best heat  distribution and temperature stability     6 MFI10 User Manual    2 3 TEMPERATURE CONTROLLER   The temperature of the barrel is controlled by means of a platinum resistance thermometer  PT100   in conjunction with a micro processor based  three term temperature controller  This enables the  operator to select any temperature within a standard range of 40 400  C which will be accurately held  to  0 1  C  The selection and adjustment of the barrel temperature forms part of the test set up which  is entered from the console and saved when the machine is switched off     2 4 INDICATING THERMOMETER 
53. rs                               37  8 2 Thermometers with Contraction and  Expansion Chambers    37  8 3 Red Spirit Thermometers                      37  8 4 General Care of Thermometers          38  9 0 TECHNICAL SPECIFICATIONS        39  10 0 CONNECTION LEADG                      40    1 0 INTRODUCTION TO MELT  FLOW INDEXING    1 1 MELT FLOW RATE   The melt flow rate  MFR  of a thermoplastic  material is the measured gravimetric flow of   the sample extruded through a die following  specified conditions of die length  diameter  tem   perature and pressure  Different standardised  conditions of extrusion temperatures and pres   sures are used for different types of polymer but   for the purpose of comparison  samples of the  same type of polymer should be tested under  identical conditions     The MFI Apparatus may be regarded as a  rheometer operating at conditions of low shear   Although the shear stresses applied and the  resultant shear rates are much lower than those  used in the processes  the results obtained do  provide a useful indication as to the relative  ease with which different materials will flow  when processed  Since higher MFR values  indicate easier melt flow  a grade of polymer  with a high MFR is generally chosen when the  process envisaged involves relatively high rates  of shear  i e  injection moulding  MFR is also a  measure of the molecular weight of a sample  and is indicative of the mechanical strength of   a material  Average molecular weight a
54. s   WEEE Directive   This equipment contains electrical and electronic circuits  and should not be directly disposed of in a landfill site     RoHS   This product is RoHS and China RoHS compliant  This  symbol indicates the equipment contains some restricted  hazardous substances above the recommended level  and  may offer potential harm to the environment after 15 years  from date of manufacture        AMETEK    TEST  amp  CALIBRATION INSTRUMENTS www lloyd instruments com  UK Denmark Singapore   Lloyd Instruments Ltd AMETEK Denmark AMETEK Singapore Pte Ltd   Tel  44  0 1243 833 370 Tel  45 4816 8000 Tel  65 6484 2388  uk far general ametek co uk ametek ametek dk aspl ametek com sg   France USA China   AMETEK S A S  AMETEK Measurement  amp  Calibration Technologies AMETEK Inc    Shanghai   Tel  33  0 1 30 68 89 40 Tel  1  727  536 7831 Tel  86 21 5868 5111  general lloyd instruments ametek fr chatillon fl lar ametek com    AMETEK Inc    Beijing  Tel  86 10 8526 2111    Germany India   AMETEK GmbH AMETEK Instruments India Pvt Ltd  AMETEK Inc    Guangzhou  Tel  49  0 2159 9136 510 Tel  91 22 2836 4750 Tel  86 20 8363 4768  info mct de ametek de ametek ametek dk lloyd ametek com cn    Information in this document is subject to change without notice    2012 by AMETEK  Inc   www ametek com  All rights reserved     
55. t back down the capillary and fly into the contraction chamber     8 3 RED SPIRIT THERMOMETERS  It is not unusual for these to exhibit breaks  If they occur they can be rejoined by freezing the bulb     NOTE  DO NOT ATTEMPT TO REJOIN BY HEATING    MFI 10 User Manual 37    8 4 GENERAL CARE OF THERMOMETERS IN USE   Thermometers must be treated with care  Being made of glass  they are not only mechanically  fragile  but also cannot withstand excessive thermal shocks  either by heating or by cooling  It is  therefore always better to allow a thermometer to warm up and cool down gradually  If it is necessary  to change thermometers during use DO NOT put hot thermometers onto cold conductive surfaces   Try to warm up cold thermometers gradually  rather than plunging them directly into high temperature  environments     38 MFI10 User Manual    9 0 SPECIFICATIONS    MFI Specification  Temperature Range  Temperature Controller  Mounting   Electrical Supply    Overall Dimensions  Weight  with standard accessories  Operating Conditions    Storage Conditions  i p  amp  olp Connections    Standards    40  C to 400  C with 20 minutes warm up time from ambient to 190  C  All systems   0 1  C accuracy   All systems are bench mounted   MFI 10  amp  MFI 10 CB  230V ac  10  50   60Hz  Fuses 5A T    MFI 10 110  amp  MFI 10 110 CB  115V ac  10  50   60Hz Fuses 10A  T   Maximum power requirement 1 kW  via I E C  plug   CN 1    Height 600mm  Width 463mm  Depth 404mm   Unpacked 32kg  71 Ibs    Packe
56. temperature     Press OK when the displayed temperature matches the temperature of the calibration thermometer   To exit without changing the calibration data stored  press the calibration key again     NOTE  Load 2g of PE or the polymer you normally test into the barrel  Compress to expel  Sp air bubbles  Insert the thermometer holder  a brass cylinder about 8cm long with two vertical   sections cut out on either side about 2cm long  Then insert the thermometer to its full length   so that the 90   angle piece rests on top of the barrel cover  The brass holder ensures that   the thermometer remains vertical and straight  The 90   angle on the thermometer ensures that   the thermometer measures at the precise distance above the die orifice  Correct the temperature  controller if necessary  after 4 minutes has elapsed and both the thermometer and barrel  temperature are stable        Push AFRT to its lowest position  10 5432             The AFRT uses an optical encoder to measure distance moved  This cannot provide an absolute  position so the AFRT uses a micro switch as a reference mark  The micro switch is activated by  pushing the AFRT arm to its lowest position  Activation of the micro switch will zero the counter  associated with the optical encoder and will automatically be detected by the MFI software  The  number displayed in the lower right hand corner of the LCD indicates the AFRT arm position relative  to its position when attached to the weight with the piston rest
57. time the audible beeper will sound and  the next message will be displayed     will introduce bubbles and cover the side of the barrel with plastic  causing possible fric   tion  between the piston and barrel      gt  NOTE  Compress sample firmly  Do not move the compression tool up and down  This       Fit Piston on its Support  OK Abort       Having compressed the sample  the piston should be fitted with its support  This message will  automatically clear after 10 seconds  Two piston supports are provided  one 81mm long for high  flow rate polymers  up to 5g 10 mins  and one shorter  71mm  for low flow rate polymers under 5g 10       Fit Piston  OK Abort             Fit Weight  2 16Kg   ok Abort          mins  When using the auto weight loader  the message reads    The weight should now be fitted to the top of the piston  The figure in parenthesis is the weight en   tered as part of the test set up and is included here to assist the user  This message will  automatically clear after 10 seconds  When using the auto weight loader  the weights will be lowered  automatically without showing this message  The auto weight loader will automatically stop when it  reaches the pre set stop position corresponding to a spacer height of either 71mm or 81mm  If neither       Attach AFRT Sensor  OK Abort             option was set during the test set up the loader arm will continue to its bottom position   If the selected test requires the use of the AFRT  it should now be attached to the
    
Download Pdf Manuals
 
 
    
Related Search
    
Related Contents
Réflexions sur la Contraception Masculine  1 PONTIFÍCIA UNIVERSIDADE CATÓLICA DO PARANÁ DOMINUS  quelques idées pour la réussite d`un projet d`éducation et de  Allied Telesis AT-SB4108 network switch      N439217 man compact hammers D25133 LA.indd  Samsung MW3100H Mikrobølgeovne, 23 L Brugervejledning  取扱説明書を見る  LE-379 - Commell    Copyright © All rights reserved. 
   Failed to retrieve file