USER`S MANUAL

image

Contents

1. using water at 70 F 21 C This K factor may be used for single point calibration and provide acceptable accuracy However readings may not be accurate when using this calibration method in some situations One example is when using the meter under extreme temperature conditions or with fluids other than water For improved accuracy under such conditions we recommend that a K factor specific to the application be determined and used for calibration A single point calibration may yield acceptable accuracy in the middle of the flo range but five or more calibration points may yield a high level of accuracy especially at the lower end of the flow range 5 MAINTENANCE Proper handling and care will extend the life and service of the meter Turbine Rotor The meter is virtually maintenance free However it is important the rotor moves freely Keep the meter clean and free of contaminants If the rotor does not turn freely apply a penetrating lubricant on the rotor shaft and bearings Remove any debris or deposits from the rotor using a soft brush or small probe Be careful not to damage the turbine rotor or supports Caution Blowing compressed air through the turbine assembly could damage the rotor Battery Replacement The computer display is powered by two 3 volt lithium batteries which may be replaced while the meter is installed ASSURED AUTOMATION When batteries are removed or lose power the batch a
2. ASSURED 1 800 899 0553 AUTOMATION assuredautomation com Plastic Digital Flow Meter Rate amp 2 Totals 1 Resetable USER S MANUAL Installation amp Maintenance Assured Automation 19 Walnut Avenue Clark New Jersey 07066 Tel 732 381 2255 Fax 732 381 2383 ASSURED AUTOMATION 1 IMPORTANT NOTICE Use TM Series meters with water and other chemicals compatible with wetted components see Specifications Section Do not use to meter fuel or incompatible chemicals TM Series meters are available with either a computer for local electronic display or a conditioned signal output module to provide a digital signal to customer interfacing equipment TM Series meters with computer display measure in gallons or liters Refer to the Calibration Section for details These meters are not legal for trade applications TM Series meters are very sensitive to electric noise if operated within one to two inches of some electric motors or other sources of electronic noise 2 INSTALLATION Connections Install your meter in line either horizontally or vertically or at the end of the hose adjacent to the nozzle Installation to metal connections is not recommended Install as follows 1 Plan to install turbine with a minimum straight pipe length as follows 1 1 Upstream from the turbine allow a minimum straight pipe length of 10 times the internal diameter of the turbine 1 2 Downstream from the turbine allow a minimu
3. M Activate the Meter Turn the computer display ON by starting water flow or briefly pressing the DISPLAY button The Batch or Cumulative Total from last use will be displayed Press DISPLAY briefly to display the Batch Total Hold the DISPLAY button down for 3 seconds to reset the Batch Total to zero The computer display is programmed to turn off automatically if not used for 4 minutes Factory and Field Calibration Curves All calibration information is visible to the user as words in the upper part of the display above the numeric digits 1 800 899 0553 assuredautomation com All units are configured with a factory calibration curve Both gallons and liters are available GAL or LTR will be displayed Use the CALIBRATE and DISPLAY buttons to switch between gallons and liters This curve is NOT user adjustable the word PRESET is displayed to show this The factory calibration is stored permanently in the computer s memory The field calibration curve may be set by the user and can be changed or modified at any time using the calibration procedure described below in the Calibration Section Totals or flowrate derived from the field calibration are visible when the field calibration setting is selected CAL B will be visible on the top line Selecting a Different Calibration Setting You can switch between GAL and LTR modes at will without corrupting totalizer contents For ex
4. OR SIGNAL OUTPUT Square Wave Output To achieve square wave output reference Wiring Diagram 2 and use an Electronic Digital Meter Battery Kit sold separately for battery power The terminal block and battery location are located on the back side of the module Access as follows 1 Remove the four Phillips head screws from the front of the module and lift the module from the turbine 2 To change terminal block connections loosen the appropriate screws Reconnect the wires in the proper positions and tighten the screws 3 Install the batteries Make sure the positive post is in the correct position 4 Position the module on the turbine housing To avoid moisture damage make sure the O ring is fully seated Tighten the four screws on the front of the module WIRING DIAGRAM 2 SQUARE WAVE OUTPUT 6 VOLT SQUARE TERMINAL WAVE SIGNAL BLOCK Q CUSTOMER 2 INTERFACING 3 EQUIPMENT 4 BLACK COMMON GROUND WHITE 1 6 VOLT SQUARE WAVE 2 9 TO 35 VOLT DC INPUT NOT USED 3 COMMON GROUND 4 OPEN COLLECTOR SIGNAL OUTPUT NOT USED Contact us for the most recent product and technical information Visit our website for more details assuredautomation com 2 ASSURED AUTOMATION Ten feet 3m of cable is provided with the module Trim the cable to desired length or extend it as necessary Verify Meter Accuracy Before using check the meter s accuracy and verify calib
5. ample the computer can totalize 10 00 gallons If the user switches to LTR mode the display will immediately change to 37 50 the same amount in units of liters GAL LTR switching also works in FLOWRATE mode To select a different calibration setting first press and hold the CALIBRATE button Continue to hold it while also pressing and releasing the DISPLAY button You may then also release the CALIBRATE button The flag indicators in the top line of the display will change to show the newly selected calibration setting Calibration settings change in this order GAL LTR CAL B GAL etc While fluid is flowing only the GAL and LTR selections may be made However when NO fluid flow is occurring any setting may be selected 4 CALIBRATION Before Beginning Field Calibration For the most accurate results dispense at a flowrate which best simulates your actual operating conditions Avoid dribbling more fluid or repeatedly starting and stopping the flow This can result in less accurate calibrations Make sure you meet the meter s minimum flowrate requirements Flowrate 1 GPM 3 8 LPM 2 GPM 7 5 LPM TM Series Meters 1 2 inch meter 3 4 inch meter 1 inch meter 5 GPM 18 8 LPM 1 1 2 inch meter 10 GPM 37 5 LPM 2 inch meter 20 GPM 75 LPM The use of a uniformly dependable accurate calibration container is highly recommended for the most accurate results Due to high flowrate it is strongly recommended
6. g to hold CALIBRATE also press and hold the DISPLAY button Hold both buttons for about 3 seconds until you see a blinking dd CAL message Once the dd CAL message appears release both buttons You are now in field calibration mode 3 Once the buttons have been released from Step 2 the display will show the blinking message run 01 If you want to exit the calibration now before dispensing any fluid go to Step 11 4 If you want to continue with the calibration but have not dispensed any fluid yet make your final preparations to your pumping system but don t start pumping yet 5 Start your pumping system so that fluid flows through the meter The display will stop blinking and show the run 01 message Dispense into a container that allows you to judge the amount of fluid pumped When you have pumped the desired amount for example 10 gallons stop the fluid flow quickly 6 Once the flow has stopped briefly press and release both buttons At this point the computer display will change to 0000 00 with the left hand digit blinking 7 Enter the volume amount of fluid that you dispensed for example if your 10 gallon container is full enter 10 0 for gallons or 37 5 for liters To enter numbers use the CALIBRATE button to change the value of the digit that is blinking and use the DISPLAY button to shift the blink to the next digit 8 Once the correct number is entered brief
7. ly press and release both buttons The display will now change to a blinking run 02 message You have installed the new calcurve point You are ready to end calibration Step 10 or enter another new calibration point Step 9 Contact us for the most recent product and technical information Visit our website for more details assuredautomation com 1 800 899 0553 assuredautomation com 9 To enter another calibration point go back and repeat Steps 3 through 8 It is possible to set up to 15 cal curve points and the run message will increment each time you repeat the calibration process run 01 run 02 run 03 etc up to run 15 10 To end calibration press and hold both buttons for about 3 seconds until you see the CAL End message After you release the buttons the computer will resume normal operations with the new cal point s active 11 If you HAVE NOT dispensed any fluid you can exit calibration without changing the cal curve If the message run 01 is showing and you have not dispensed any fluid hold both buttons for about 3 seconds until you see a CAL End message After you release the buttons the computer will resume normal operation and the old curve if you entered one in the past is still intact Calibration with Conditioned Signal Output Module The K factor of your meter appears on the calibration report as the number of pulses per gallon The factor is determined during production
8. m straight pipe length of 5 times the internal diameter of the turbine 2 For Soc Fittings use only primer and solvents approved for PVC gluing For NPT Fittings wrap all connections with 3 to 4 wraps of thread tape Make sure the tape does not intrude into the flow path 3 Attach meter with arrow pointed in the direction of flow 4 For NPT Fittings hand tighten the meter at the housing ends Do no use a wrench or similar tool to tighten This can damage the housing Conditioned Signal Output Module Wiring This conditioned signal output module can be wired to provide an open collector signal output or 6 volt square wave output Open Collector Signal Output To achieve an open collector signal output reference Wiring Diagram 1 The terminal block is located on the back side of the module The module is factory assembled for open collector signal output Please provide the 820 ohm minimum resistor Ten feet 3m of cable is provided with the module Trim it to desired length or extend it as necessary Distances up to 5 000 feet 1 524m can be achieved for open collector signal output 1 800 899 0553 assuredautomation com WIRING DIAGRAM 1 OPEN COLLECTOR SIGNAL OUTPUT TERMINAL RED 9 TO 35 VOLT DC BLOCK 2 BLACK GROUND 4 WHITE OPEN COLLECTOR SIGNAL CUSTOMER INTERFACING EQUIPMENT 1 6 VOLT SQUARE WAVE NOT USED 2 9 TO 35 VOLT DC INPUT 3 COMMON GROUND 4 OPEN COLLECT
9. nd cumulative totals reset to zero but the field and factory calibrations are retained If the display becomes dim or blank replace the batteries as follows 1 Remove the four Phillips head screws from the face of the meter and lift the faceplate from the turbine 2 Remove the old batteries and clean any corrosion from the terminals 3 Install new batteries Make sure the positive post is in the correct position 4 When the batteries are replaced the faceplate will power ON Check the display to ensure normal functions have resumed before assembling again 5 Reseat batteries if necessary and position the faceplate on the turbine housing To avoid moisture damage make sure the O ring is fully seated Tighten the four screws on the faceplate Contact us for the most recent product and technical information Visit our website for more details assuredautomation com 1 800 899 0553 assuredautomation com
10. ration 1 Make sure there is no air in the system by starting the flow until it runs steadily Then stop the flow using a valve or nozzle 2 Meter an exact known volume into an accurate container For best results meter with one continuous full stream 3 Check the volume against the display or recording equipment If the amount metered is accurate further calibration is not necessary If not refer to the Calibration Section for further instructions 3 OPERATION Computer Display Batch and Cumulative Totals The computer maintains two totals The Cumulative Total provides continuous measurement and cannot be manually reset The Batch Total can be reset to measure flow during a single use The Cumulative Total is labeled with TOTAL 1 LOCKED indicating that this total is locked and cannot be manually zeroed Batch Total is labeled with TOTAL 2 When the Cumulative Total reaches a maximum reading of 999 999 it will automatically reset to zero Press the DISPLAY button briefly to switch between the batch cumulative total and flowrate NOTE Totalization counts total units without differentiating between gallons liters or field calibrated units Flowrate Feature To use this feature press and release DISPLAY until FLOWRATE appears to the left of the bottom line When FLOWRATE is displayed the numbers on the middle line reflect the rate of flow for example the current gallons per minute GPM or liters per minute LP
11. that calibration be completed with a combination of volume and weight using fine resolution scales For best results the meter should be installed and purged of air before field calibration Contact us for the most recent product and technical information Visit our website for more details assuredautomation com 3 ASSURED AUTOMATION Field Calibration with Computer Display Field Calibration and Factory Calibration are defined in the previous section Factory calibration settings are custom programmed into each computer during production using water at 70 F 21 C Readings using the standard factory calibration curves may not be accurate in some situations for example under extreme temperature conditions or with fluids other than water For improved accuracy under such conditions the flow computer allows for field calibration that is user entry of custom calibration parameters A single point calibration may yield acceptable accuracy in the middle of the flow range but five or more calibration points may yield a higher level of accuracy especially at the lower end of the flow range Up to 15 custom calibration points can be entered Dispense Display Field Calibration Procedures 1 Hold down CALIBRATE while pressing and releasing DISPLAY until the field calibration curve appears CAL B message will be displayed Release both buttons 2 To calibrate press and hold the CALIBRATE button While continuin